COVER PAGE OF SELLER’S SUBMITTAL
City Name: Hanford, WA - MSA
Pall Project No.: 01.000287
Job Number: Project L-897, Subcontract 70816
Task Register 6 (Task 1 – VI Submittal Package)
To: (Contractor)
Mission Support Alliance, LLC (MSA)
PO Box 650
Richland, WA 99352
From: (Subcontractor)
Pall Water
A Division of Pall Corporation
839 State Route 13
Cortland, NY 13045
Patty Willis
Phone: 607-758-1699
Name of Project: Hanford, WA, Central Plateau WTF
Pall Document Number: NYCS000015175 Rev: 01
Description of Item: Flow Instrumentation
Specification Section: 40 73 02
Deviations: See document NYCS000015383
INDEX:
Page 1: Cover Page Information
Page 2: Revision History and Approvals
Page 3: Tab 1 - NYCS000015448,CSS,Flow Instrumentation,01.287 R01
Pages 4-15: Tab 2 - Brooks GT1600 Rotometer Cut Sheet R00
Pages 16-19: Tab 3 - Hach TU5X00 Flow Sensor Cut Sheet R00
Pages 20-23: Tab 4 - Turck Flow Switch Cut Sheet R00a
Pages 24-61: Tab 5 - 0.5 Inch Rosemount 8750W Magnetic Flow Meter Cut Sheet R00
Pages 62-101: Tab 6 - 6-Inch & 8-Inch Rosemount 8750W Magnetic Flow Meter Cuts
Pages 102-125:Tab 7 - FCI ST100L Air Flow Meter Cut Sheet R00
Date: 12/19/2019
PALL ADVANCED SEPARATIONS SYSTEMS
DIVISION OF PALL CORPORATION, CORTLAND, NY
Rev Name Date Created By
Approved By
Created By
Approved By
Created By
Approved By
Created By
Approved By
Revision Description
Title
Document # Revision griffinju Text Box These commodities, technology, or software may only be exported from the United States in accordance with the Export Administration Regulations. Diversion contrary to U.S. law is prohibited.
Doc Number: NYCS000015448R01 Doc Title: NYCS000015448,CSS,Flow Instrumentation,01.287 R01.docx
Components Specification Sections
The following specification sections are relevant to the approval of submittal NYCS000015175, Flow Instrumentation,01.287:
L‐897, Attachment No. 2 ‐ HNF‐SPEC‐63538, Section 40 73 02, Process Instrumentation Meters & Transmitters.
Any exceptions and clarifications to the specification section above can be found in submittal NYCS000015383, L‐897, Section 40 73 02, Es and Cs,01.287.
Pall has taken exceptions and clarifications relevant to the flow instrumentation proposed.
PALL CORPORATION
Membrane Filtration (MF) System
FLOW INSTRUMENTATION CUT SHEET
Rotometers (Flow Indicators):
3.5 to 35 Liters/Hour:
BROOKS P/N: 16 4 0 A G6 A 1 AA 1 3 D 1 0 A
Part Number Key:
16 Base Model, Horizonatl Inlet & Outlet 4 127mm Scale Length, 127 – Low Flow 0 Horizontal Inlet, Horizontal Outlet Connections A Always “A” (Default)
G6 R-6-127-AT Tube, ¼” Ball Float, Carboloy
3.5 to 35 LPH Range, Water
A 316 SST Fittings 1 ¼” Connection Size
AA Standard NPT(F) Connection 1 Viton O-Ring 3 Direct Reading Scale D +/- 2 Percent Full Scale Accuracy 1 Valve at Inlet 0 No Alarms A No Approvals
9 to 90 Liters/Hour:
BROOKS P/N: 16 4 0 A H6 A 1 AA 1 3 D 1 0 A
Part Number Key:
16 Base Model, Horizonatl Inlet & Outlet 4 127mm Scale Length, 127 – Low Flow 0 Horizontal Inlet, Horizontal Outlet Connections A Always “A” (Default)
H6 R-6-127-BT Tube, ¼” Ball Float, Carboloy 9 to 90 LPH Range, Water
A 316 SST Fittings 1 ¼” Connection Size
AA Standard NPT(F) Connection 1 Viton O-Ring 3 Direct Reading Scale D +/- 2 Percent Full Scale Accuracy 1 Valve at Inlet 0 No Alarms A No Approvals
Thermal Mass Flow
GT1600 Model
• Configurable to retro-fit GT10xx, GT130x and Full-View
• 360 degrees rotatable viewing angle
• High quality materials for safety, in- and outdoor durability
• 316 stainless steel frame
• Polycarbonate safety shield
• 316/316L dual certified stainless steel process fittings
• Alarms for high- and low flow (optional for all flow ranges, either at time of order, or as add-on in the field)
• Flanged or threaded connections, available horizontal and vertical
• Panel- and wall mount options
• Easy in-situ maintenance: Clean or replace tube and float without removal from the process piping
• Adjust the scale to compensate for process variation
• Optional integral needle valve
DATA SHEET- data
Variable Area
Model GT1600
1BrooksInstrument.com
Flow Accuracy Standard: ±10%, ± 5%, ±2% Full Scale, Class 2.5 acc VDI/VDE Optional: ±1% Full Scale, Class 1.6 acc VDI/VDE
Repeatability ≤ 0.5% Full Scale
Ambient Temperature Limits ‐4°F to 125°F | ‐20°C to 52°C
Scales Transparant scale. Adjustable Nominal Lengths: 75mm, 127mm and 250mm Choice of direct reading units, millimeter or percentage of maximum flow with factor tag.
Operating Fluid Temperature Limits 33 to 250°F | 1°C to 121°C
Mounting options In‐line Panel‐mount (see Dimensions) Wall‐mount (see Dimensions)
Features
View GT1600 Product Page
Industrial Glass Tube Variable Area Flow Meters
Product Specifications
Capacities and Pressure Drops See Capacities
HomR Text Box
3.5 to 35 LPH: BROOKS P/N 16 4 0 A G6 A 1 AA 1 3 D 1 0 A 9 to 90 LPH: BROOKS P/N: 16 4 0 A H6 A 1 AA 1 3 D 1 0 A psi bar psi bar 02 300 20.7 240 16.5 06 300 20.7 240 16.5 07 300 20.7 240 16.5 08 250 17.0 240 16.5 09 200 13.8 200 13.8 10 175 12.1 175 12.1
Metering Tube (wetted) Process fittings (wetted)
Float Stops (wetted) O rings (wetted) Housing Window & safety shield Hardware
Tube size 02&06 Tube size 07 10
Pressure Ratings
Float (wetted)
Materials of construction
Maximum Operating Pressure (PSIG | bar) at Fluid Temperature Tube size Threaded process connections Flanged process connections
Horizontal IN/OUT Horizontal IN/OUT
Vertical IN/ Horizontal OUT
Process connections NPT Threaded Flanged ANSI 150# RF
Vertical IN/ Horizontal OUT
Vertical IN/OUT
Polycarbonate with UV inhibitor 316 stainless steel
Valve Optional integral needle valve
Div1/Div2 (CSA) – gas/dust. Protection method I.S. / Non incendive Zone1/Zone2 (CSA/ATEX/IECex) – gas/dust. Protection method Ex m (no barrier)
Alarms Ring initiator high / low alarm
Hazardous area approvals
Material Certification to DIN 3.1 Declaration of Compliance 2.1 Oxygen Service
Reed switch high / low alarm
Vertical IN/OUT Horizontal IN/ Vertical OUT Horizontal IN/ Vertical OUT
316/316L (dual certified stainless steel)
Viton®, Buna N, Kalrez®, EPDM
Sizes 2&6: Carboloy® or 316/316L (dual certified stainless steel) Sizes 7 10: 316/316L (dual certified stainless steel) Teflon®
Borosilicate glass
Certifications International Calibration Certificate
316 stainless steel
Product Specifications
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Line
HomR Line
HomR Rectangle
Product Specifications - Capacities
±0.9% of S.P. (20-100% F.S.), ±0.18% of F.S. (2-20% F.S.)
>1100 slpm F.S.:
±1.0% of F.S.
AA1:L48
127-Low Flow Pressure Drop Pressure Drop Alarm inch W.C./kPa inch W.C./kPa Type
R-2-127-AAAAT 316SS 1.9 0.11 0.7/0.17 1.0 0.11 6.7 0.8/0.19
R-2-127-AAAAT Carboloy 3.7 0.22 1.1/0.27 1.0 0.2 12 1.2/0.3
R-2-127-AAT Carboloy 7.5 0.45 1.2/0.29 1.0 0.38 23 1.5/0.37
R-2-127-DT 316SS 15 0.93 0.9/0.22 1.0 0.68 40 1.0/0.24
R-2-127-DT Carboloy 25 1.5 1.5/0.38 1.0 1 61 1.7/0.42
R-2-127-AT 316SS 39 2.3 1.1/0.27 1.0 1.4 86 1.2/0.3
R-2-127-AT Carboloy 59 3.5 1.9/0.47 1.0 2 120 2.1/0.52
R-2-127-BT 316SS 110 6.8 1.8/0.45 1.0 3.9 230 2.0/0.21
R-2-127-BT Carboloy 170 10 3.0/0.75 1.0 5.6 340 3.3/0.83 R-2-127-CT Carboloy 270 16 6.1/1.51 1.0 9 540 4.8/1.2
R-6-127-AT 316SS 410 25 6.1/1.53 1.0 14 850 6.8/1.7
R-6-127-AT Carboloy 620 37 10.5/2.61 1.0 20 1200 11.6/2..9
R-6-127-BT 316SS 1000 65 30.1/7.5 1.0 35 2100 33.3/8.3
R-6-127-BT Carboloy 1500 95 57.8/14.4 1.0 49 2900 64.2/16
250-High Flow Pressure Drop Pressure Drop inch W.C./kPa inch W.C./kPa
7-XV-11A-A 0.48 100 8/2 1.0 1.8 3.1 10/2.5
7-XS-23-A 0.77 170 16/4.0 1.0 3.3 5.6 17/4.2
8-XV-8-A 1.00 240 5/1.5 3.7 4.4 7.5 5/1.3
8-XV-14-A 1.40 320 8/2 5.4 5.8 9.9 8/2
8-XV-31-A* 2.00 460 16/4 7.0 14 24 17/4.3
9-XS-33-A 3.20 730 4/1 2.3 13 22 8/2
9-XV-87-A* 3.90 890 14/3.5 17 28 48 16/4
9-XS-87-A* 5.10 1100 18/4.5 3.5 36 62 19/4.8
10-XV-64-A 6.20 1400 12/3 15 25 43 14/3.5
10-XS-64-A 7.80 1700 16/4 3.7 32 54 18/4.5
10-XS-138-A* 10.00 2400 30/7.5 5.5 80 130 36/9
10-XJ-238-A* 21.00 4800 104/26 1.0 150 270 16/4
127-High Flow 7-XV-11A-A 0.41 93 8/2 1.0 1.8 3.1 10/2.5
7-XS-23-A 0.66 150 16/4.0 1.0 2.7 4.6 17/4.2
8-XV-8-A 0.99 220 5.0/1.25 3.7 4 6.9 6.0/1.5
8-XV-14-A 1.3 310 8.0/2 5.4 5.3 9.1 10/2.5
8-XV-31-A* 1.7 400 24/6 7.0 11 20 28/7
9-XS-33-A 3.0 690 7.0/1.7 2.3 12 21 8.0/2.0
9-XV-87-A* 3.6 830 14/3.5 17 26 44 16/4.0
9-XS-87-A* 4.5 1000 17/4.2 3.5 32 55 20/5.0
10-XV-64-A 6 1300 9/2.25 15 24 41 10/2.5
10-XS-64-A 7.4 1600 12/3 3.7 30 52 13/3.25
10-XS-138-A* 9.8 2200 29/7.25 5.5 68 110 33/8.25
10-XJ-238-A* 20 4700 104/26 1.0 140 250 16/4
075-High Flow 8-XV-8-A 0.82 180 8/2 3.3 5.6 9/2.25
8-RJ-10 1.5 340 9/2.25 6.3 10 11/2.7
8-RJ-23 2.4 540 13/3.25 9.9 16 15/3.7
8-RJ-30 3.1 710 19/4.8 13 22 22/5.5
8-RJ-39 4.7 1000 34/6
10-RJ-80 12 2700 18/4.5
10-RJ-180 21 4900 71/17.6
10-RJ-37 24 40 18/4.5
10-RJ-83 36 62 35/8.7
10-RJ-90 55 93 71/17.6
Ring Initiator
Reed Switch
Size 02
Size 06
SLPM ln/h
Size 07 R-7M-25-1FT
Size 08 R-8M-25-4FT
Size 09 R-9M-25-3FT
Size 10 R-10M-25-3FT
Full Scale - Water Full Scale - Air
Tube Float GPM l/h V.I.C.
cSt
Tube Float cc/min l/h V.I.C.
cSt
SCFM m3n/h
Size 07 R-7M-127-1FT
Size 08
Size 10 R-10M-75-3
R-8M-75-1
Size 08 R-8M-127-4FT
Size 09 R-9M-127-3FT
Size 10 R-10M-127-3FT
HomR Rectangle
HomR Rectangle
HomR Text Box
Product Specifications - Dimensions
Product Description - Dimensions inch mm inch mm inch mm inch mm inch mm inch mm
H-IN / H-OUT 1020 8.59 218.3 0.59 15.0
V-IN / V-OUT 1024 9.84 250.0 - -
H-IN / V-OUT 1026 9.22 234.1 0.59 15.0
V-IN/ H-OUT 1027 9.22 234.1 0.59 15.0
H-IN / H-OUT 1110 8.59 218.3 0.59 15.0
V-IN / V-OUT 1114 9.63 244.5 - -
H-IN / V-OU 1T 140 9.11 231.4 0.59 15.0
V-IN/ H-OUT 1144 9.11 231.4 0.59 15.0
H-IN / H-OUT 1020 8.59 218.3 3.50 75.0
V-IN / V-OUT 1024 10.63 270.0 - -
H-IN / V-OU 1T 026 9.61 244.1 2.95 75.0
V-IN/ H-OUT 1027 9.61 244.1 2.95 75.0
H-IN / H-OUT 1140 8.59 218.3
V-IN / V-OUT 1144 377.8 - -
H-IN / V-OU 1T 146 11.73
V-IN/ H-OUT 1147 11.73
H-IN / H-OU 1T 020 17.50 444.5 0.91 23.0
V-IN / V-OUT 1024 19.69 500.0 - -
H-IN / V-OU 1T 026 18.59 472.3 0.91 23.0
V-IN/ H-OUT 1027 18.59 472.3 0.91 23.0
H-IN / H-OUT #7/8 16.50 419.1
H-IN / H-OUT #9/10 17.50 444.5
V-IN / V-OUT #7/ 18 7.00 431.8
V-IN / V-OUT #9/10 17.25 438.2
H-IN / V-OUT #7/ 18 6.75 425.5
H-IN / V-OUT #9/1 10 7.38 441.3
V-IN/ H-OUT #7/ 18 6.75 425.5
V-IN/ H-OUT #9/10 17.38 441.3
V-IN / V-OUT Retrofit only(4) GT1307 1307 16.94 430.2 - -
H-IN / H-OUT #7/8 3.94 100.0
H-IN / H-OUT #9/1 40 .72 120.0
V-IN / V-OUT 1024 19.69 500.0 - -
H-IN / V-OUT #7/8 3.94 100.0
H-IN / V-OUT #9/1 40 .72 120.0
V-IN/ H-OUT #7/ 38 .94 100.0
V-IN/ H-OUT #9/10 4.72 120.0
H-IN / H-OUT #7/
22.88
8 3 .50 88.9
H-IN / H-OUT #9/10 4.00 101.6
V-IN / V-OUT #7/
7.00 419.1
V-IN / V-OUT #9/10
17.25 444.5
H-IN / V-OUT #7/8 3 .50 88.9
H-IN / V-OUT #9/10 4 .00 101.6
V-IN/ H-OUT #7/8 3 .50 88.9
V-IN/ H-OUT #9/10 4 .00 101.6
Note (3) - Standard dimension
Note (4) - Retro-fit dimensions are for replacement of GT10xx, GT130x and Full-View meters in e in new in .
6.0
Full-ViewRetrofit only(4)
Flanged 150lbs RF
Notes
Dimension A B C L M N
Meter type Tube size
Process on
Or n H(orizontal)
V(er al R) etro-fit Model
50.0 4.00 101.6 1.00 25.4
127-Low Flow 02
NPT- GF T1000Standard dimensions(3)
1.97 0.24
1110 0.91 23.0
1114 - -
70.0 8.002.76
0.91 23.0
1117 0.91
18.59 472.3
GT1000Standard dimensions(3)
Full-ViewRetrofit only(4)
250-High Flow 07
Standard dimensions(3)
Full-ViewRetrofit only(4)
Full-View 1140Retrofit only(4)
GT1000Flanged 150lbs RF
NPT- GF T1000Standard dimensions(3)
7.0203.2 1.63 41.28 0.28
23.0
17.50 444.5
18.59 472.3
P Q
3.00 101.6 14.88
298.1 298.1
3.00 101.6
3.00 101.6
1847.24 37.61.48
393.515.49 69.72.74
581.0
22.50 571.5
19.69 500.0
20.00 508.0
19.69 500.0
20.00 508.0 inch mm inch mm inch mm inch mm inch mm inch mm inch mm inch mm
H IN / H OUT 12.31 312.7 0.91 23.0
V IN / V OUT 14.50 368.2
H IN / V OUT 13.41 340.5 0.91 23.0
V IN/ H OUT 13.95 354.4 0.91 23.0
H IN / H OUT #7/8 11.50 292.1
H IN / H OUT #9/10 12.50 317.5
V IN / V OUT #7/8 12.00 304.8
V IN / V OUT #9/10 12.25 311.2
H IN / V OUT #7/8 11.75 298.5
H IN / V OUT #9/10 12.38 314.3
V IN/ H OUT #7/8 11.75 298.5
V IN/ H OUT #9/10 12.38 314.3
V IN / V OUT Retrofit only(4) GT1306 1306 11.75 298.5
H IN / H OUT #7/8 3.94 100.0
H IN / H OUT #9/10 4.72 120.0
V IN / V OUT #7/8
V IN / V OUT #9/10
H IN / V OUT #7/8 3.94 100.0
H IN / V OUT #9/10 4.72 120.0
V IN/ H OUT #7/8 3.94 100.0
V IN/ H OUT #9/10 4.72 120.0
H IN / H OUT #7/8 11.50 292.1 3.50 88.9
H IN / H OUT #9/10 12.50 317.5 4.00 101.6
V IN / V OUT #7/8 17.88 454.0
V IN / V OUT #9/10 17.50 444.5
H IN / V OUT #7/8 14.69 373.0 3.50 88.9
H IN / V OUT #9/10 15.00 381.0 4.00 101.6
V IN/ H OUT #7/8 14.69 373.0 3.50 88.9
V IN/ H OUT #9/10 15.00 381.0 4.00 101.6
H IN / H OUT 9.02 229.1 0.83 21.0
V IN / V OUT 11.20 284.6
H IN / V OUT 10.11 256.9 0.83 21.0
V IN/ H OUT 10.11 256.9 0.83 21.0
V IN / V OUT Retrofit only(4) GT1305 1305 7.75 196.9
H IN / H OUT #7/8 3.94 100.0
H IN / H OUT #9/10 4.72 120.0
V IN / V OUT #7/8
V IN / V OUT #9/10
H IN / V OUT #7/8 3.94 100.0
H IN / V OUT #9/10 4.72 120.0
V IN/ H OUT #7/8 3.94 100.0
V IN/ H OUT #9/10 4.72 120.0
Note (3) Standard dimension are for new installations
Note (4) Retro fit dimensions are for replacement of GT10xx, GT130x and Full View meters in existing installation. Not for use in new installations.
Model
Dimension A B C L M N
1110 0.91 23.0
NPT F Standard dimensions(3)
1116 0.91 23.0
1117 0.91 23.0
Retrofit only(4)
70.0 N/A2.76
10.11 256.9
10.11 256.9
13.41 340.5
13.41 340.5
Standard dimensions(3)
12.31 312.7
14.50 368.2
9.02 229.1
11.20 284.6
Retrofit only(4)
NPT F Standard dimensions(3)
127 High Flow
075 High Flow 08
Flanged 150lbs RF
Full View
GT1000
Full View
Notes
Flanged 150lbs RF
Standard dimensions(3)
Meter type Tube size
Process connection
Orientation H(orizontal) V(ertical) Retro fit
Product Specifications - Dimensions
Product Approvals Overview
M e c h a n ic a l
R e e d
S w it c h
In d u c ti v e A la rm
Declaration Declaration
CRN
ATEX
IECEX
Explosion safety "Intrinsic Safety (ia)
UL File E73889 Vol3 Sec
Explosion safety "Intrinsic Safety (ia)"
ATEX
Pepperl + Fuchs
PTB 99 ATEX 2128 X
Pepperl + Fuchs Control Draw ing:116- 0165G
TCF: 203104000-1604
Explosion safety "Intrinsic Safety
(ia)"
Inductive Ring Sensor
Non-Hazardous Locations Pow er Supply Current consumption
Ambient Temperature
Hazardous Location
ATEX
FM Approvals
Pepperl + Fuchs Model: RC10-14-N3-Y53478 Pepperl + Fuchs Model: RC15-14-N3-Y53479
Nominal Voltage 8V Operating Voltage 5...25V Active area clear : 3mA (at 8V) Active area obscured: 0.5...0.95mA (at 8V)
0°C to 40°C
II 2 G Ex ia IIC T6…T1 Gb IP67 Refer to ATEX Certif icate for: Input parameters, Max Ambient Temperature, Special conditions for use
Class I, Division 1, Group A, B, C, Class II, Division
Standards used for evaluation:
Ambient Temperature range:
Input Pow er:
Special conditions for safe
EN 60079-0 : 2012+A11: 2013, EN 60079-18 : 2015
IEC 60079-0 : 2011, IEC 60079-18 : 2014
-20 °C to +65 °C (Standard Version) -20 °C to +55 °C (With Junction box Version) 30V, 250mA, 3W Refer to IOM
Ambient Temperature ratings:
Input parameters:
Special conditions for safe use:
-20° C ≤ Tamb ≤ 65° C Vmax = 30V, Imax = 100mA, Ci = 0μF, Li = 0μH Refer to IOM
IS Class I, II, III, Div 1, Groups A, thru G
Ex mb IIC T6 Gb Ex mb IIIC T85°C Db (Standard Version)
KIWA 18ATEX0013 X
IECEx KIWA 18.0008X
Declarations Mark
Meter Options
Standards/Directives/Marking Status/Certificate
Explosion safety "Constructional safety (c)"
ATEX II2G Ex h IIC T6…T4 Gb II2D Ex h IIIC T120°C Db Special conditions for safe use:
Refer to IOM
EU Declaration of Conformity
EMC Directive (2014/30/EU) RoHS Directive (2011/65/EU)
EMC Directive (2014/30/EU)
Reed Sw itch Alarms are classif ied as “Simple Apparatus” w hen used in Intrinsically Safe circuits. They comply w ith the requirements of EN60079-11 clause 5.7 – Simple apparatus.
Pressure Equipment Directive (2014/68/EU) EMC Directive (2014/30/EU) The equipment uses a reed sw itch sensor & is outside the scope of the directive since the inherent nature of the physical characteristics of w hich is such that:
(i) it is incapable of generating or contributing to electromagnetic emissions w hich exceed a level allow ing radio and telecommunication equipment and other equipment to operate as intended; and
(ii) it operates w ithout unacceptable degradation in the presence of the electromagnetic disturbance normally consequent upon its intended use.
Canadian Registration Number (CRN) Explosion safety "encapsulation (m)"
II 2 G Ex mb IIC T6 Gb II 2 D Ex mb IIIC T85°C Db (Standard Version)
Ex mb IIC T5 Gb Ex mb IIIC T100°C Db (With Junction box Version)
II 2 G Ex mb IIC T5 Gb II 2 D Ex mb IIIC T100°C Db (With Junction box Version) use:
1, Group E, F, G, Class III, Division 1 Class I, Zone 0, Group IIC T6
Product Approvals Overview
Model Code
Code Description Code Option Option Description1
Request a Quote
¹
I.-II. Base Model Number 16 Horizontal Inlet and Outlet
III. Body - Flow/Scale Length Body Flow Scale Length 1 250-High Flow 250mm 2 127-High Flow 127mm 3 075-High Flow 75mm 4 127-Low Flow 127mm
IV. Connection Orientation 0 Horizontal Inlet - Horizontal Outlet 4 Vertical Inlet - Vertical Outlet 6 Horizontal Inlet - Vertical Outlet 7 Vertical Inlet - Horizontal Outlet
V. Model Revision A Initial release of global model code
VI.-VII. Maximum Flowrate 250 - High Flow Water Air Code GPM l/h SCFM m3n/h Tube Float
JC 0.48 100 1.8 3.1 J R-7M-25-1FT C 7-XV-11A-A
JF 0.77 170 3.3 5.6 J R-7M-25-1FT F 7-XS-23-A
KC 1.00 240 4.4 7.5 K R-8M-25-4FT C 8-XV-8-A
KF 1.40 320 5.8 9.9 K R-8M-25-4FT F 8-XV-14-A
KJ 2.00 460 14* 24* K R-8M-25-4FT J 8-XV-31-A*
LC 3.20 730 13 22 L R-9M-25-3FT C 9-XS-33-A
LF 3.90 890 28* 48* L R-9M-25-3FT F 9-XV-87-A*
LJ 5.10 1100 36* 62* L R-9M-25-3FT J 9-XS-87-A*
MC 6.20 1400 25 43 M R-10M-25-3FT C 10-XV-64-A
MF 7.80 1700 32 54 M R-10M-25-3FT F 10-XS-64-A
MJ 10.00 2400 80* 130* M R-10M-25-3FT J 10-XS-138-A*
MM 21.00 4800 150* 270* M R-10M-25-3FT M 10-XJ-238-A*
* These codes require a back pressure of 30 psig /2 bar
127 - High Flow Water Air Code GPM l/h SCFM m3n/h Tube Float
SC 0.41 93 1.8 3.1 S R-7M-127-1FT C 7-XV-11A-A
SF 0.66 150 2.7 4.6 S R-7M-127-1FT F 7-XS-23-A
TC 0.99 220 4 6.9 T R-8M-127-4FT C 8-XV-8-A
TF 1.3 310 5.3 9.1 T R-8M-127-4FT F 8-XV-14-A
TJ 1.7 400 11* 20* T R-8M-127-4FT J 8-XV-31-A*
UC 3 690 12 21 U R-9M-127-4FT C 9-XS-33-A
UF 3.6 830 26* 44* U R-9M-127-4FT F 9-XV-87-A*
UJ 4.5 1000 32* 55* U R-9M-127-4FT J 9-XS-87-A*
VC 6 1300 24 41 V R-10M-127-3FT C 10-XV-64-A
VF 7.4 1600 30 52 V R-10M-127-3FT F 10-XS-64-A
VJ 9.8 2200 68* 110* V R-10M-127-3FT J 10-XS-138-A*
VM 20 4700 140* 250* V R-10M-127-3FT M 10-XJ-238-A*
* These codes require a back pressure of 30 psig/2 bar
075 - High Flow Water Air Code GPM l/h SCFM m3n/h Tube Float
PC 0.82 180 3.3 5.6 P R-8M-75-1 C 8-XV-8-A
PF 1.5 340 6.3 10 P R-8M-75-1 F 8-RJ-10
PJ 2.4 540 9.9 16 P R-8M-75-1 J 8-RJ-23
PM 3.1 710 13 22 P R-8M-75-1 M 8-RJ-30
PQ 4.7 1000 N/A N/A P R-8M-75-1 Q 8-RJ-39
RC 12 2700 N/A N/A R R-10M-75-3 C 10-RJ-80
RF 21 4900 N/A N/A R R-10M-75-3 F 10-RJ-180
RJ N/A N/A 24 40 R R-10M-75-3 J 10-RJ-37
RM N/A N/A 36 62 R R-10M-75-3 M 10-RJ-83
RQ N/A N/A 55 93 R R-10M-75-3 Q 10-RJ-90
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Text Box
Model Code
Code Description Code Option Option Description1¹
VI.-VII. Maximum Flowrate (continued) 127 - Low Flow Water Air Code GPH l/h SLPM In/h Tube Float
A3 1.9 0.11 0.11 6.7 A R-2-127-AAAAT 3 316SS
A6 3.7 0.22 0.2 12 A R-2-127-AAAAT 6 CARBOLOY
B6 7.5 0.45 0.38 23 B R-2-127-AAT 6 CARBOLOY
C3 15 0.93 0.68 40 C R-2-127-DT 3 316SS
C6 25 1.5 1 61 C R-2-127-DT 6 CARBOLOY
D3 39 2.3 1.4 86 D R-2-127-AT 3 316SS
D6 59 3.5 2 120 D R-2-127-AT 6 CARBOLOY
E3 110 6.8 3.9 230 E R-2-127-BT 3 316SS
E6 170 10 5.6 340 E R-2-127-BT 6 CARBOLOY
F6 270 16 9 540 F R-2-127-CT 6 CARBOLOY
G3 410 25 14 850 G R-6-127-AT 3 316SS
G6 620 37 20 1200 G R-6-127-AT 6 CARBOLOY
H3 1000 65 35 2100 H R-6-127-BT 3 316SS
H6 1500 95 49 2900 H R-6-127-BT 6 CARBOLOY
VIII. Fitting Material A 316 Stainless Steel B 316 Stainless Steel w/ CRN
IX. Connection Size 1 1/4” Low Flow 2 1/2” High Flow 3 3/4” High Flow 4 1” High Flow 5 1” Vertical / 3/4” Horizontal. Size 10 only.
X. - XI. Connection Type AA Standard / GT1000 Retrofit NPT - Female AB Standard / GT1000 Retrofit ANSI #150 RF Flange BA Retrofit GT130x NPT - Female CA Retrofit Full-View 11xx NPT - Female CB Retrofit Full-View 11xx ANSI 150# RF Flange DC Retrofit GT1000 Rc - Female\
XII. O-Ring Material 1 Viton® fluoroelastomer 2 Buna 3 Kalrez®
4 EPDM
XIII. Scale Inscription 1 MM Scale 2 Percent Scale 3 Direct Reading Scale 4 Dual Scale
XIV. Meter Accuracy A ±10% B ±5% Full Scale C ±5% w/calibration D ±2% Full Scale E ±2% Full Scale w/calibration F ±1% Full Scale w/calibration G 2.5 VDI/VDE w/calibration H 1.6 VDI/VDE w/calibration
XV. Valve Type/Location 0 None 1 Valve at Inlet 2 Valve at Outlet
XVI. Alarm 0 None 1 1 Switch/Sensor 2 2 Switches/Sensors
XVII. Approvals A None B ATEX - Zone 1 & Zone 2, non-Electrical C UL / FM - Hazardous location D ATEX - Zone 1 & Zone 2, Electrical
Sample Standard Model Code
I-II III IV V VI-VII VIII IX X-XI XII XIII XIV XV XVI XVII
16 1 4 A D0 A 2 AA 1 3 D 0 0 A
Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Text Box
Additional Services Available
· CONDULET JUNCTION BOX ASSEMBLY
Op�on Code Descrip�on
0 None 1 1 Switch/Sensor 203Z023AAA 2 2 Switches/Sensor 203Z023AAA from 16xxG Model Code, posi�on 16 Part Numbers for Relays Alarm
· RELAY
Op�on Code Descrip�on
110VAC
1 or 2 relay
220VAC
1 or 2 relay
24VAC
1 or 2 relay
0 None 1 1 Switch/Sensor 029G008ZZZ or 029G010ZZZ or 029G012ZZZ 2 2 Switches/Sensor 029G008ZZZ or 029G010ZZZ or 029G012ZZZ
Alarm from 16xxG Model Code, posi�on 16 Part Numbers for Relays
· FRONT PANEL MOUNTING KIT
#2-6 778Z018AAA 127mm
#7-10 778Z019AAA 250mm
· BACK PANEL MOUNTING KIT
#2-6 778Z020AAA 127mm
#7-10 778Z021AAA 250mm
· 3.1 Material Cert
· International Calibration Certificate (ICC)
· NACE MR0175 MR0103
· PMI
· Clean for Oxygen Service 2.1
· Declaration of Compliance 2.1
· Pressure Test Certificate 2.2
· Commercial Cleaning
Brooks is committed to assuring all of our customers receive the ideal flow solution for their application, along with outstanding service and support to back it up. We operate first class repair facilities located around the world to provide rapid response and support. Each location utilizes primary standard calibration equipment to ensure accuracy and reli-ability for repairs and recalibration and is certified by our local Weights and Measures Authorities and traceable to the relevant International Standards.
Visit www.BrooksInstrument.com to locate the service location nearest to you.
START-UP SERVICE AND IN-SITU CALIBRATION
Brooks Instrument can provide start-up service prior to operation when required. For some process applications, where ISO-9001 Quality Certification is important, it is mandatory to verify and/or (re)calibrate the products periodically. In many cases this service can be provided under in-situ conditions, and the results will be traceable to the relevant international quality standards.
CUSTOMER SEMINARS AND TRAINING
Brooks Instrument can provide customer seminars and dedicated training to engineers, end users, and maintenance per-sons. Please contact your nearest sales representative for more details. Due to Brooks Instrument’s commitment to continuous improvement of our products, all specifications are subject to change without notice.
TRADEMARKS
Brooks ......................................................................Brooks Instrument, LLC All other trademarks are the property of their respective owners.
DS-VA-GT1600-eng/2019-9
Service and Support
Hach TU5X00sc Series Turbidimeter Flow Sensor:
Hach P/N: LQV160.99.00002
USAGE:
In-Line paddle-wheel type flow sensor to provide local measurement of Hach TU5X00sc Turbidimeter (ie, TU5300sc & TU5400sc) sample water flow rate on Hach SC200 controller
Integral 25-Inch signal cable plugs into port on Hach TU5X00sc series Turbidimeters.
Flow Sensor for TU5300 sc and TU5400 sc Laser Turbidimeter Product #: LQV160.99.00002
USD Price: $152.00 Available
Adding a flow measurement along with your TU5300 sc or TU5400 sc turbidity measurement provides critical information about the status of your drinking water process. A TU5 series turbidimeter's flow sensor accessory can provide alerts to flow issues, giving you important information to resolve problems and get your plant running smoothly.
Product details pdf footer17
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Line
HomR Text Box
SAMPLE
WATER INLET
HomR Oval
HomR Text Box
FE-*
HomR Oval
HomR Text Box
AE-*
HomR Oval
HomR Text Box
AIT
HomR Polygonal Line
HomR Line
HomR Text Box
SAMPLE WATER
TO DRAIN
HomR Line
HomR Text Box
HACH FLOW
SENSOR
HomR Text Box
HACH TU5X00 TURBIDIMETER
SENSOR
HomR Text Box
HACH SC200 CONTROLLER
HomR Oval
HomR Line
HomR Text Box
25" FLOW SENSOR CABLE
HomR Oval
HomR Line
HomR Oval
HomR Line
HomR Text Box
1/4" OD SAMPLE WATER TUBING
HomR Polygonal Line
HomR Text Box
(MANUAL)
FLOW REGULATOR VALVE
HomR Line
HomR Text Box
5.25' TURBIDIMETER
SENSOR CABLE
Polygon
Figure 3 Installation overview with accessories
1 Service bracket 6 Flow sensor (accessory)
2 Automatic cleaning module (accessory) 7 Bubble trap overflow
3 SC controller 8 Sample outlet
4 Bubble trap (accessory) 9 Sample inlet
5 Flow regulator5 10 TU5300 sc or TU5400 sc
5 Not used with the bubble trap.
12 English
HomR Oval
HomR Text Box
HACH TU5X00
FLOW SENSOR
(FE-)
HomR Line
HomR Line
HomR Line
HomR Line
HomR Line
HomR Line
HomR Line
HomR Text Box
HACH SC200
(AIT-)
HomR Text Box
HACH TU5X00
(AE-)
HomR Rectangle
HomR Line
HomR Line
HomR Text Box
(NOT USED)
HomR Line
HomR Line
HomR Text Box
(NOT USED)
HomR Text Box
(NOT USED)
HomR Text Box
(NOT USED)
HomR
(NOT USED)
Text Box
HACH TU5X00 FLOW SENSOR
HomR Line
HomR Text Box
HACH TU5X00
TURBIDIMETER
SENSOR
HomR Line
HomR Text Box
HACH FLOW SENSOR INTEGRAL SIGNAL CABLE
(PLUGS INTO HACH TU5X00 TURB. SENSOR)
HomR Line
HomR Text Box
SAMPLE WATER
INLET LINE
Line
Flow Switch (Insertion Type):
FLOW SWITCH:
TURCK P/N: FCS-N1/2 A4P-AP8X-H1141/L100
THERMAL FLOW SWITCH: 4-INCH (100mm) PROBE w/ ½” NPT PROCESS
CONNECTION, 316 SS WETTED PARTS, PNP SIGNAL OUTPUT, 24VDC
POWER, 4-PIN M12 (MALE) SIGNAL PORT
CABLE PLUG:
MENCOM P/N: MDC-4FP-FW11
PLUG CONNECTOR: 4-PIN M12 (FEMALE) FIELD-WIRING CONNECTOR,
POLYAMIDE BODY w/ NICKEL-PLATED BRASS NUT, COPPER CONTACTS w/ GOLD PLATING, ACCEPTS #18-#22 AWG CONDUCTORS / 5-10mm (0.20”-
0.40”) OD CABLE
TURCK
DC Self-Contained Flow Monitors
19 TURCK Inc. 3000 Campus Drive Minneapolis, MN 55441 Application Support: 1-800-544-7769 Fax: (763) 553-0708 www.turck.com
FCS-N1/2 A4P-AP8X-H1141
• DC PNP/NPN transistor output, short-circuit protected
• Flow monitor and signal processor in one compact housing
• Dynamic pressure resistance up to 870 psi
• Simple adjustment by means of potentiometer
This DC self-contained flow device has a flow monitor and signal processor incorporated into one compact housing. It is ideal for monitoring liquid flow speeds in the 2 to 295 ft/min range.
The flow monitor is designed for installation directly into the pipeline by means of a pipe-tee or welded pipe socket. It is suitable for use in a 1/2 inch pipe diameter or larger.
The flow monitor probe is made of 316TI stainless steel and the housing is made of plastic. The flow monitor is rated for dynamic pressures up to 870 psi. The outputs of the monitors feature a short-circuit protected PNP transistor.
Setpoint adjustments, in reference to flow speed, are achieved by means of an adjuster located under the protective screw cap on the front of the device.
The output is de-energized when the flow speed is below the setpoint; the output is energized when flow speed is above the setpoint. Each of the six LEDs display the difference the actual flow speed in comparison to the flow speed that the setpoint is adjusted to. The LEDs operate as follows:
Red LED: flow is below set point Yellow LED: flow is at or above set point Green LEDs: each LED indicates a percentage of flow above the setpoint
Notes: Available with relay output. Consult factory.
homr homr
TURCK FLOW SWITCH (100mm / 4-INCH PROBE LENGTH):
P/N FCS-N1/2 A4P-AP8X-H1141/L100
TURCK Inc. 3000 Campus Drive Minneapolis, MN 55441 Application Support: 1-800-544-7769 Fax: (763) 553-0708 www.turck.com 20
Flow Monitors DC Self Contained
FCS-N1/2 A4P-AP8X-H1141
Type FCS-N1/2 A4P-AP8X-H1141 ID Number M6871032
Operating Voltage 19.2 - 28.8 VDC (including ripple) Power / Current consumption ≤65 mA
Output PNP transistor, short-circuit and reverse polarity protected
Switching current ≤400 mA Voltage drop ≤1.5 V at 400 mA
Temperature Range (flow) -25° to +80°C (-13° to +176°F) cont.
up to +100°C (+212° F) for approx.10 min.
Operating Range (flow rate) Water 1-150 cm/s (2-295 ft/min) Oil 3-300 cm/s (6-590 ft/min)
Time Delay Before Availability 2-15 s (8 s typical) Switch ON time (above setpoint) 1-13 s (2 s typical) Switch OFF time (below setpoint) 1-15 s (2 s typical) Temperature gradient max. 250°C/min. (450°F/min.)
Pressure Rating 1450 psi (100 bar)
LED Indications Below setpoint, output de-energized red At setpoint, output energized yellow Above setpoint, output energized green (4), in addition to yellow LED)
Probe Material (DIN 2 462/17 440) 316TI stainless steel (1.4571) Enclosure (DIN 40 050) IP 67, NEMA 1,3,4,4x,12,13 Temperature rating -25° to +80°C (-13° to +176°F) Torque 100 Nm (73.7 ft-lb)
Housing Material PBT Enclosure (DIN 40 050) IP 67, NEMA 1, 3, 4, 4x, 12, 13 Operating temperature -25° to +80°C (-13° to +176°F)
Connection 3-wire eurofast quick disconnects RK 4T-* or WK 4T-* (Page 59)
* = Length in meters
FCS-N1/2 A4P-AP8X-H1141
homr Rectangle homr
NOTE: PIN 2 NOT USED
H-20 Mencom Corporation TEL: 770-534-4585 www.mencom.com
MDC Series (Micro-DC) Cordsets
Micro-DC Series (M12) Field Wireable Plugs PG 11 Threads 18-24 AWG
Face View of Male
# of Poles
Wire AWG Rating PG Size Cable Range Female Straight Male Straight
18-24
250V 4A
PG 11
PG 11*
PG 11*
PG 11*
PG 11*
5.0 - 10 mm
4.5 - 5.5 mm x 2
4.0 - 5.0 mm x 2
3.5 - 4.5 mm x 2
2.5 - 3.5 mm x 2
MDC-4FP-FW11
MDC-4FP-FW11-2x5 MDC-4FP-FW11-2x4.5 MDC-4FP-FW11-2x4 MDC-4FP-FW11-2x3
MDC-4MP-FW11
MDC-4MP-FW11-2x5 MDC-4MP-FW11-2x4.5 MDC-4MP-FW11-2x4 MDC-4MP-FW11-2x3
18-24
50V 4A
PG 11
PG 11*
PG 11*
PG 11*
PG 11*
5.0 - 10 mm
4.5 - 5.5 mm x 2
4.0 - 5.0 mm x 2
3.5 - 4.5 mm x 2
2.5 - 3.5 mm x 2
MDC-5FP-FW11
MDC-5FP-FW11-2x5 MDC-5FP-FW11-2x4.5 MDC-5FP-FW11-2x4 MDC-5FP-FW11-2x3
MDC-5MP-FW11
MDC-5MP-FW11-2x5 MDC-5MP-FW11-2x4.5 MDC-5MP-FW11-2x4 MDC-5MP-FW11-2x3
20-24
50V 4A
PG 11
PG 11*
PG 11*
PG 11*
PG 11*
5.0 - 10 mm
4.5 - 5.5 mm x 2
4.0 - 5.0 mm x 2
3.5 - 4.5 mm x 2
2.5 - 3.5 mm x 2
MDC-8FP-FW11
MDC-8FP-FW11-2x5 MDC-8FP-FW11-2x4.5 MDC-8FP-FW11-2x4 MDC-8FP-FW11-2x3
MDC-8MP-FW11
MDC-8MP-FW11-2x5 MDC-8MP-FW11-2x4.5 MDC-8MP-FW11-2x4 MDC-8MP-FW11-2x3
PG 11 Male StraightPG 11 Female Straight
These connectors are easily wired by inserting the individual wires into the screw terminal contacts. The connectors are water tight to IP 67 and are fully mate-able with other “Micro DC” style connectors. Other multi-hole grommets are available for the PG11 version, please consult the factory for part number and availability.
* These part numbers have a two hole grommet that allows two different cables to be installed into one fi eld wireable connector.
Specifi cations:
Body: Black Polyamide Insulator: Black Polyamide Contacts: Copper Alloy with Gold Plating Coupling Nut: Nickel Plated Brass Seal: NBR Temperature: -25ºC to +85ºC (-13ºF to +185ºF) Protection: IP67 - Single opening grommet IP65 - Two hole grommet Approvals: U.S.: UL Pending Canada: C-UL Pending
2.6
.8
2.7
.8
HomR Polygonal Line
HomR Text Box
4-PIN MICRO-DC (FEMALE) PLUG: MEMCON P/N MDC-4FP-FW11
HomR Rectangle
HomR Line
HomR Line
HomR Oval
HomR Line
HomR Line
FLOW INSTRUMENTATION CUT SHEET
½-Inch Flanged Mag. Flow Meter w/ 316L SS Electrodes/GND Rings:
Rosemount P/N: 8750W D M R 2 A 4 F T S A 005 C A1 DA1 M4 G1 AX Q4
Part Number Key:
8750W Model: Rosemount 8750W Utility Magnetic Flow Meter D Sensor Design Revision: Revision "D"
M Transmitter Class: Revision 4 Electronics R Transmitter Mount: Remote Field Mount (Order Cable Separately)
2 Transmitter Power: DC Power Supply (12 to 42VDC) A Transmitter Outputs: 4-20mA; Digital HART; Scalable Pulse
4 Conduit Entries: 1/2-14 NPT, Additional Entry F Sensor Style: Flanged T Lining Material: PTFE
S Electrode Material: 316L Stainless-Steel A Electrode Type: 2 Measurement Electrodes – Standard
005 Line Size: ½ Inch (DN15) C Flange Type and Material: Slip-On, Raised-Face, Carbon Steel
A1 Flange Rating: ASME B16.5, Class 150 DA1 Adv. Diagnostics Suite 1: Process Diagnostics HPN, Grounding/Wiring, Electrode Coating
M4 Transmitter Display: Local Operator Interface G1 Ground Ring: 316L Stainless-Steel (Qty 2) AX Discrete I/O: Two Discrete Channels (DI/DO 1, DO 2) Q4 Quality Cert: Calibration Certificate per ISO 10474 3.1B
Remote Flow Transmitter Signal Cable:
Rosemount P/N: 08732-0065-2001 (order by FT)
FLOW INSTRUMENTATION CUT SHEET
½-Inch Flanged Mag. Flow Meter w/ Hastalloy Electrodes/GND Rings:
Rosemount P/N: 8750W D M R 2 A 4 F T H A 005 C A1 DA1 M4 G2 AX Q4
Part Number Key:
8750W Model: Rosemount 8750W Utility Magnetic Flow Meter D Sensor Design Revision: Revision "D"
M Transmitter Class: Revision 4 Electronics R Transmitter Mount: Remote Field Mount (Order Cable Separately)
2 Transmitter Power: DC Power Supply (12 to 42VDC) A Transmitter Outputs: 4-20mA; Digital HART; Scalable Pulse
4 Conduit Entries: 1/2-14 NPT, Additional Entry F Sensor Style: Flanged T Lining Material: PTFE
H Electrode Material: Hastalloy C-276 A Electrode Type: 2 Measurement Electrodes – Standard
005 Line Size: ½ Inch (DN15) C Flange Type and Material: Slip-On, Raised-Face, Carbon Steel
A1 Flange Rating: ASME B16.5, Class 150 DA1 Adv. Diagnostics Suite 1: Process Diagnostics HPN, Grounding/Wiring, Electrode Coating
M4 Transmitter Display: Local Operator Interface G2 Ground Ring: Hastalloy C-276 (Qty 2) AX Discrete I/O: Two Discrete Channels (DI/DO 1, DO 2) Q4 Quality Cert: Calibration Certificate per ISO 10474 3.1B
Remote Flow Transmitter Signal Cable:
Rosemount P/N: 08732-0065-2001 (order by FT)
Product Data Sheet 00813-0300-4750, Rev CB
May 2019
Rosemount® 8750W Magnetic Flowmeter System for Utility, Water, and Wastewater Applications
■ Best in class value with performance, reliability, and diagnostics suited for monitoring applications
■ Reliable all welded coil housing and lightweight sensor design rated to IP68
■ Available in 4–20mA with HART®, FOUNDATION™ Fieldbus, Modbus® RS-485, Process Diagnostics, and SMART™ Meter Verification to improve reliability and performance
■ Available with drinking water certifications
HomR Text Box
ROSEMOUNT P/N: 8750W D M R 2 A 4 F T S A 005 C A1 DA1 M4 G1 AX Q4
8750W D M R 2 A 4 F T H A 005 C A1 DA1 M4 G2 AX Q4
HomR Text Box
316 SS ELECTRODES &
GROUNDING RINGS
HomR Text Box 1/2-Inch FLANGED MAG. FLOW METER WITH REMOTE TRANSMITTER (24VDC POWER):
HomR Line
HomR Line
HomR Line
HomR Line
HomR Line
HomR Line
HomR Line
HomR Line
HomR Text Box
HASTALLOY ELECTRODES &
GROUNDING RINGS
Product Selection Guide The Rosemount 8750W Magnetic Flowmetersensor is available in a flanged style and transmitter is available in remote and integral transmitter configurations to ensure compatibility with all utility, water, and wastewater applications.
Transmitter selection
Transmitter General characteristics
Field mount ■ Integral and remote configurations available
■ HART/Analog and Pulse outputs available
■ FOUNDATION™ Fieldbus and pulse output available
■ Modbus RS-485 and Pulse output available
■ Advanced Diagnostics available
■ LCD display (optional)
— With optional optical switch local operator interface(1)
■ Two discrete channels (optional)
Wall mount ■ Wall mount configuration
■ HART/Analog and Pulse outputs available
■ Modbus RS-485 and Pulse output available
■ FOUNDATION™ Fieldbus and pulse output available
■ Advanced Diagnostics available
■ Local LCD display (optional)
— With optional 15 button tactile key pad(1)
■ Two discrete channels (optional)
(1) HART or Modbus protocol only.
Contents Product Selection Guide
Magmeter Diagnostics
Magnetic Flowmeter Sizing
Ordering Information
Product Specifications
Product Certifications
Dimensional drawings
May 2019
2 www.rosemount.com
Sensor selection Table 1: Sensor Selection
Sensor General characteristics
Flanged ■ Flanged process connections Welded coil housing
■ ½-in. (15 mm) to 48-in. (1200 mm)
■ Standard, reference, and bullet-nose electrodes available
Selecting materials For guidance on selecting materials, refer to the Rosemount™ Magnetic Flowmeter Material Selection Guide (00816-0100-3033) , available at www.emerson.com/rosemount.
Magmeter Diagnostics
Rosemount diagnostics reduce cost & improve output by enabling new practices
Rosemount Magnetic Flowmeters provide device diagnostics that detect and warn of abnormal situations throughout the life of the meter - from installation to maintenance and meter verification. With Rosemount Magnetic Flowmeter diagnostics enabled, plant availability and throughput can be improved, and costs through simplified installation, maintenance and troubleshooting can be reduced.
Table 2: Magnetic flowmeter diagnostics
Diagnostic name Diagnostic category Product capability
Basic diagnostics
Tunable Empty Pipe Process Standard
Electronics Temperature Meter Health Standard
Coil Fault Meter Health Standard
Transmitter Fault Meter Health Standard
Reverse Flow Process Standard
Coil current Maintenance Standard
Electrode saturation Process/Maintenance Standard
Advanced diagnostics
High Process Noise Process Suite 1 (DA1)
Grounding and Wiring Fault Installation Suite 1 (DA1)
Coated Electrode Detection Process Suite 1 (DA1)
Commanded Smart Meter Verification Meter Health Suite 2 (DA2)
Continuous Smart Meter Verification Meter Health Suite 2 (DA2)
May 2019 www.rosemount.com 3 http://www.emerson.com/rosemount HomR Rectangle
Table 2: Magnetic flowmeter diagnostics (continued)
Diagnostic name Diagnostic category Product capability
4-20 mA Loop Verification(1) Installation Suite 2 (DA2)
(1) Available with HART output only.
Options for accessing diagnostics
Rosemount Magmeter Diagnostics can be accessed through the Local Operator Interface (LOI), ProLink® III v3.1, a HART Field Communicator(1), and AMS® Suite: Intelligent Device Manager(1). Contact your local Rosemount representative to activate diagnostics or for diagnostic availability on existing transmitters.
Access diagnostics through the LOI for quicker installation, maintenance, and meter verification
Rosemount Magnetic Flowmeter Diagnostics are available through the LOI to simplify maintenance.
Access diagnostics through ProLink III v. 3.0 (HART)/ProLink III v. 3.1 (HART, Modbus)
Simplify maintenance and troubleshooting practices by utilizing ProLink III v3.0/v3.1 to access diagnostics and troubleshooting information, log variable data, run SMART Meter Verification, and print verification reports.
Access diagnostics through AMS Intelligent Device Manager(1) for the ultimate value
The value of the diagnostics increases significantly when AMS Intelligent Device Manager is used. AMS Intelligent Device Manager provides a simplified screen flow and procedures for how to respond to the diagnostic messages.
Magnetic Flowmeter Sizing Selecting the appropriate sensor size is an important step when considering a magnetic flowmeter. The physical properties of the process fluid, as well as the fluid velocity should be considered. It may be necessary to select a flow sensor that is larger or smaller than the adjacent piping to ensure the fluid velocity is in the recommended flow range for the application.
■ Operation outside these guidelines may also give acceptable performance.
Table 3: Sizing guidelines
Application Velocity range (ft/s) Velocity range (m/s)
Normal Service 0–39 0–12
Preferred Service 2–20 0.6–6.1
Note Operation outside these guidelines may also give acceptable performance.
To convert flow rate to velocity, use the appropriate factor listed in Table 4 and the following equation:
Velocity = Flow Rate
Factor
(1) Available with HART output only.
May 2019
4 www.rosemount.com
Example: English units Example: SI units
Magmeter Size: 4 in. (factor from Table 4 = 39.679) Normal Flow Rate: 300 GPM
Velocity = 300 (gpm)
39.679
Velocity = 7.56 ft/s
Magmeter Size: 100 mm (factor from Table 4 = 492.78)Normal Flow Rate: 800 L/min
Velocity = 800 (L/min)
492.78
Velocity = 1.62 m/s
Table 4: Line size vs. conversion factor
Nominal line size—Inches (mm) Gallons per minute factor Liters per minute factor
½ (15) 0.947 11.762
1 (25) 2.694 33.455
1½ (40) 6.345 78.806
2 (50) 10.459 129.89
2 ½ (65) 14.923 185.33
3 (80) 23.042 286.17
4 (100) 39.679 492.78
5 (125) 62.356 774.42
6 (150) 90.048 1,118.3
8 (200) 155.93 1,936.5
10 (250) 245.78 3,052.4
12 (300) 352.51 4,378.0
14 (350) 421.70 5,237.3
16 (400) 550.80 6,840.6
18 (450) 697.19 8,658.6
20 (500) 866.51 10,761
24 (600) 1,253.2 15,564
30 (750) 2006.0 24,913
36 (900) 2,935.0 36,451
40 (1000) 3,652.1 45,357
42 (1050) 4,115.1 51,107
48 (1200) 5,407.6 67,159
May 2019 www.rosemount.com 5
HomR Rectangle
HomR Text Box
ROSEMOUNT P/N: 8750W D M R 2 A 4 F T S A 005 C A1 DA1 M4 G1 AX Q4
8750W D M R 2 A 4 F T H A 005 C A1 DA1 M4 G2 AX Q4
HomR Line
Table 5: Line size vs. velocity/rate
Nominal line size in inches (mm)
Minimum/maximum flow rate
Gallons per minute Liters per minute at 0.04 ft/s
(low-flow cutoff) at 1 ft/s
(min range setting) at 3 ft/s at 39.37 ft/s
(max range setting) at 0.012 m/s
(low-flow cutoff) at 0.3 m/s
(min range setting) at 1 m/s at 12 m/s
(max range setting)
½ (15) 0.038 0.947 2.841 37.287 0.141 3.529 11.76 141.15
1 (25) 0.108 2.694 8.081 106.05 0.401 10.04 33.45 401.46
1½ (40) 0.254 6.345 19.04 249.82 0.946 23.64 78.81 945.67
2 (50) 0.418 10.459 31.38 411.77 1.559 38.97 129.89 1,558.7
2½ (65) 0.597 14.923 44.77 587.51 2.224 55.60 185.33 2,224.0
3 (80) 0.922 23.042 69.13 907.17 3.434 85.85 286.17 3,434.0
4 (100) 1.587 39.679 119.04 1,562.2 5.913 147.84 492.78 5,913.4
5 (125) 2.494 62.356 187.07 2,454.9 9.293 232.33 774.42 9,293.0
6 (150) 3.602 90.048 270.14 3,545.2 13.42 335.50 1,118.3 13,420
8 (200) 6.237 155.93 467.79 6,138.9 23.24 580.96 1,936.5 23,238
10 (250) 9.831 245.78 737.34 9,676.3 36.63 915.73 3,052.4 36,629
12 (300) 14.10 352.51 1,057.5 13,878 52.54 1,313.4 4,378.0 52,535
14 (350) 16.87 421.71 1,265.1 16,603 62.85 1,571.2 5,237.3 62,848
16 (400) 22.03 550.80 1,652.4 21,685 82.09 2,052.2 6,840.6 82,087
18 (450) 27.89 697.19 2,091.6 27,448 103.90 2,597.6 8,658.6 103,903
20 (500) 34.66 866.51 2,599.5 34,114 129.14 3,228.4 10,761 129,137
24 (600) 50.13 1,253.2 3,759.6 49,339 186.77 4,669.2 15,564 186,769
30 (750) 80.24 2,006.0 6,018.0 78,976 298.96 7,474.0 24,913 298,959
36 (900) 117.40 2,935.0 8,805.1 115,553 437.42 10,935 36,451 437,416
40 (1000) 146.09 3,652.1 10,956 143,785 544.29 13,607 45,357 544,286
42 (1050) 164.60 4,115.1 12,345 162,011 613.28 15,332 51,107 613,278
48 (1200) 216.30 5,407.6 16,223 212.898 805.91 20,148 67,159 805,908
Upstream and downstream piping To ensure specified accuracy over widely varying process conditions, install the sensor with a minimum of five straight pipe diameters upstream and two pipe diameters downstream from the electrode plane.
May 2019
6 www.rosemount.com
HomR Rectangle
HomR Text Box
ROSEMOUNT P/N: 8750W D M R 2 A 4 F T S A 005 C A1 DA1 M4 G1 AX Q4
8750W D M R 2 A 4 F T H A 005 C A1 DA1 M4 G2 AX Q4
HomR Line
Figure 1: Upstream and downstream straight pipe diameters
A. Five pipe diameters (upstream) B. Two pipe diameters (downstream) C. Flow direction
Installations with reduced upstream and downstream straight runs are possible. In reduced straight run installations, the meter may not meet accuracy specifications. Reported flow rates will still be highly repeatable.
Sensor grounding
A reliable ground path is required between the sensor and the process fluid. Optional grounding rings or process reference electrodes are available with the sensor to ensure proper grounding. See Table 5 and Table 6.
Ordering Information
Rosemount 8750W Magnetic Flowmeter Platform
The Rosemount 8750W Magnetic Flowmeter is available in a flanged sensor design. The sensors are fabricated from stainless and carbon steel and welded to provide a sealed coil housing that protects against moisture or other contaminants. Sizes range from 1/2-in. (15 mm) to 48-in. (1200 mm). The field mount transmitter has a die cast aluminum housing for excellent reliability. The wall mount transmitter features an easy to use operator interface. Both transmitter styles are available with advanced diagnostics to provide the best insight into the process and the meter's health.
Note The starred (★) offerings represent the most common options, and should be selected for best delivery.
Model code structure
Example model code with one selection out of each category:
8750W D M T 1 A 1 F P S A 010 C A1 Z5 DA2 AX M4 BD G5 B6 R15 V1 Q4 HR7 WG YF
Table 6: Requirements - select one from each available choice
Example code Category
8750W Base model—Magnetic Flowmeter System (utility, water, and wastewater)
D Sensor design revision—Revision "D"
M Transmitter class (Table 8)
T Transmitter mount (Table 9)
1 Transmitter power (Table 10)
May 2019 www.rosemount.com 7
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Text Box
HomR Line
HomR Text Box
ROSEMOUNT P/N: 8750W D M R 2 A 4 F T S A 005 C A1 DA1 M4 G1 AX Q4
8750W D M R 2 A 4 F T H A 005 C A1 DA1 M4 G2 AX Q4
HomR Line
HomR Line
HomR Text Box R
Table 6: Requirements - select one from each available choice (continued)
Example code Category
A Transmitter outputs (Table 11)
1 Conduit entries (Table 12)
F Sensor style (Table 13)
P Lining material (Table 14)
S Electrode material (Table 15)
A Electrode type (Table 16)
010 Line size (Table 17)
C Flange type and material (Table 18)
A1 Flange rating (Table 19)
Table 7: Options - select only as needed
Example code Category
Z5 Hazardous area certifications (Table 20)
DA2 Advanced diagnostics (Table 21)
AX Discrete input/output (Table 22)
M4 Display (Table 23)
BD Certifications (Table 24)
G5 Grounding rings (Table 25)
B6 Miscellaneous (Table 26)
R15 Submergence protection (Table 27)
V1 Special paint (Table 28)
Q4 Quality certificates (Table 29)
HR7 Revision configuration (Table 30)
WG Witness inspection (Table 31)
Yx Quick Start Guide language (Table 32)
Requirements
Table 8: Rosemount 8750W transmitter class
Code Description
M Revision 4 electronics ★
0 Spare sensor, no transmitter
Table 9: Rosemount 8750W transmitter mount
Code Description
T Integral field mount ★
R Remote field mount ★
May 2019
8 www.rosemount.com
HomR Text Box
HomR Text Box T
HomR Text Box
HomR Text Box
DA1
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Text Box
ROSEMOUNT P/N: 8750W D M R 2 A 4 F T S A 005 C A1 DA1 M4 G1 AX Q4
8750W D M R 2 A 4 F T H A 005 C A1 DA1 M4 G2 AX Q4
HomR Text Box S, H
HomR Rectangle
HomR Line
HomR Line
HomR Line
HomR Line
HomR Text Box
8750W D M R 2 A 4 F T S A 005 C A1 DA1 M4 G1 AX Q4
8750W D M R 2 A 4 F T H A 005 C A1 DA1 M4 G2 AX Q4
HomR Rectangle
HomR Rectangle
HomR Rectangle
Table 9: Rosemount 8750W transmitter mount (continued)
Code Description
W Remote wall mount ★
Table 10: Rosemount 8750W transmitter power
Code Description
1 AC Power Supply (90 -250VAC, 50/60Hz) ★
2 DC Power Supply (12 - 42VDC) ★
0 Spare sensor, no transmitter
Table 11: Rosemount 8750W transmitter outputs
Code Description
A 4-20mA output with digital HART protocol & scalable pulse output ★
F FOUNDATION™ Fieldbus & scalable pulse output ★
M Modbus RS-485 electronics, scalable pulse ★
0 Spare sensor, no transmitter
Table 12: Rosemount 8750W conduit entries
Code Description Integral mount quantity Remote mount quantity
1 ½–14 NPT 2 4 ★
2 M20–1.5 adapters 2 4 ★
4(1) ½–14 NPT, additional entry 3 5
5(1) M20–1.5, additional entry 3 5
0 Spare sensor, integral mount only, no transmitter
N/A N/A
(1) Not available with the wall mount transmitter.
Table 13: Rosemount 8750W sensor style
Code Description
F Flanged ★
0 Spare transmitter, no sensor
Table 14: Rosemount 8750W lining material
Code Description
T(1) PTFE ★
P(2) Polyurethane ★
N(3) Neoprene ★
0 Spare transmitter, no sensor
(1) Available in ½ - to 36-in. (15 mm to 900 mm) line sizes.
(2) Available in line sizes 1- to 36-in., 42-in., and 48-in. (25 mm to 900 mm, 1050 mm, and 1200 mm) line sizes.
(3) Available in line sizes 1- to 48-in. (25 mm to 1200 mm) line sizes.
May 2019 www.rosemount.com 9
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Text Box
8750W D M R 2 A 4 F T S A 005 C A1 DA1 M4 G1 AX Q4
8750W D M R 2 A 4 F T H A 005 C A1 DA1 M4 G2 AX Q4
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
Table 15: Rosemount 8750W electrode material
S 316L stainless steel ★
H Nickel alloy 276 (UNS N 10276) ★
0 Spare transmitter, no sensor
Table 16: Rosemount 8750W electrode type
Code Description
A 2 Measurement electrodes – standard ★
B(1) 2 Measurement electrodes – bulletnose ★
E 2 Measurement electrodes plus 1 reference electrode – standard ★
F(1) 2 Measurment electrodes plus 1 reference electrode – bulletnose ★
0 Spare transmitter, no sensor
(1) Not available in ½-in. (15mm).
Table 17: Rosemount 8750W line size
Code Line size Liner availability
In this table, the starred (★) offerings represent available liner based on line size. Consult factory for additional Flange Type/Rating sensor availability
PTFE code T Poly code P Neoprene code N
005 ½-in. (15 mm) ★
010 1-in. (25 mm) ★ ★ ★
015 1½-in. (40 mm) ★ ★ ★
020 2-in. (50 mm) ★ ★ ★
025 2½-in. (65 mm) ★ ★
030 3-in. (80 mm) ★ ★ ★
040 4-in. (100 mm) ★ ★ ★
050 5-in. (125 mm) ★ ★
060 6-in. (150 mm) ★ ★ ★
080 8-in. (200 mm) ★ ★ ★
100 10-in. (250 mm) ★ ★ ★
120 12-in. (300 mm) ★ ★ ★
140 14-in. (350 mm) ★ ★ ★
160 16-in. (400 mm) ★ ★ ★
180 18-in. (450 mm) ★ ★ ★
200 20-in. (500 mm) ★ ★ ★
240 24-in. (600 mm) ★ ★ ★
300 30-in. (750 mm) ★ ★ ★
May 2019
10 www.rosemount.com
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Text Box
8750W D M R 2 A 4 F T S A 005 C A1 DA1 M4 G1 AX Q4
8750W D M R 2 A 4 F T H A 005 C A1 DA1 M4 G2 AX Q4
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
HomR Rectangle
Table 17: Rosemount 8750W line size (continued)
Code Line size Liner availability
In this table, the starred (★) offerings represent available liner based on line size. Consult factory for additional Flange Type/Rating sensor availability
PTFE code T Poly code P Neoprene code N
360 36-in. (900 mm) ★ ★ ★
400 40-in. (1000 mm) ★
420 42-in. (1050 mm) ★ ★
480 48-in. (1200 mm) ★ ★
000 Spare transmitter, no sensor
Table 18: Rosemount 8750W flange type and material
Code…