This memorandum provides recommendations for shielding requirements for a medical imaging room in a Veterans Affairs medical center. It details that the room will be used for up to two fluoroscopy exams per week and calculates that approximately 30 exams could be performed before reevaluation of shielding needs. No additional shielding is currently required given the limited use, but the room must be posted as a radiation area when imaging equipment is in use.

The related federal contract opportunity is for renovation of approximately 11,000 square feet of space on the 7th floor of the main hospital building at the Clement J. Zablocki VA Medical Center in Milwaukee, Wisconsin. The project scope involves demolishing existing interior spaces and renovating the area to relocated the GI procedure clinic. It will also update mechanical, electrical, plumbing and other building systems. The contract value is estimated between $2-5 million. The solicitation is scheduled to post on October 15, 2020 to the SAM.gov platform. The NAICS code is 236220 and it is set aside for Service-Disabled Veteran-Owned Small Businesses. The bid opening is scheduled for November 16, 2020 and a pre-bid conference will be held.

View the file

Other files

File Type Posted
36C25221B0002 0003.docx DOCX document
S06 - ET101 - SEVENTH FLOOR PLAN - TECHNOLOGY - Amd03.pdf PDF
36C25221B0002 0002.docx DOCX document
S06 - Appendix E - Physicist Report.pdf PDF
S06 - AE501 - BUILDING DETAILS.pdf PDF
S06 - EP602 - SCHEDULES POWER.pdf PDF
S06 - ET101 - SEVENTH FLOOR PLAN - TECHNOLOGY.pdf PDF
S06 - DG107 - SEVENTH FLOOR MEDICAL GAS PLAN.pdf PDF
S06 - 08 71 13.11 - LOW ENERGY DOOR OPERATORS Amend No2.pdf PDF
S06 - AE002 - WALL TYPES.pdf PDF
S06 - AE101 - SEVENTH FLOOR ARCHITECTURAL PLAN.pdf PDF
S06 - ET501 - DETAILS - TECHNOLOGY.pdf PDF
S06 - 11 73 00 - CEILING MOUNT PATIENT LIFT Amend No2.pdf PDF
S06 - GC101 - INFECTION CONTROL_PHASING PLAN.pdf PDF
S06 - EP401 - POWER RISERS.pdf PDF
S04 - Site Visit Sign-In Sheet - 695-19-107.pdf PDF
36C25221B0002 0001.docx DOCX document
S02 - Wage Determination Effective 10.2.2020.pdf PDF
S02 - J_A 695-19-107_Door Hardware.pdf PDF
S02 - J_A 695-19-107_HVAC BAS.pdf PDF
S02 - 695-19-107_2020.09.24_Volume 2 Specifications 100percent.pdf PDF
S02 - 695-19-107_2020.09.24_Volume 1 Specifications 100percent.pdf PDF
S02 - J_A 695-19-107_Interior Signage.pdf PDF
S02 - J_A 695-19-107_Firestop Systems.pdf PDF
S02 - 695-19-107 Drawing Set Updated.pdf PDF
S02 - J_A 695-19-107_Medical Gas Systems.pdf PDF
36C25221B0002_1.docx DOCX document
S02 - J_A 695-19-107_Card Readers.pdf PDF
S02 - J_A 695-19-107_Nurse Call.pdf PDF
S02 - J_A 695-19-107_Fire Alarm System.pdf PDF
Show all 30

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Department of Memorandum Veterans Affairs

Date: June 1, 2020

From: Alex Paul Radiation Safety Officer

Subj: 7AN – Room 7207 Shielding

To: Biomedical Engineering

ACCOMPANYING 36C25221B0002 0002

The following assumptions were derived from Pulmonary Staff and NCRP Report No.

147:

Current weekly patient load is 2 exams per week.

Per calculations, 30 exams could be performed before a reevaluation is needed.

Distance was calculated from drawings provided by Guidon Design.

If above assumptions have changed or if any dimensions relating to the placement of equipment or walls are changed, the required shielding will need to be recalculated.

RECOMMENDATIONS

No shielding is currently required for the limited C-Arm use in room 7207. The room must be posted with Caution – Radiation Area when the C-Arm is in use.

Alex Paul Health Physicist/Radiation Safety Officer

CC: Jon Wardecke James Beier Robert Steldt Meaghen Krebsbach Linda Sue Schwartz David Korman

Digitally signed by Alex Paul 1656846 Date: 2020.06.01 10:11:30 -05'00'