This document is a VA Responses to Requests for Information (RFI's) Part II addressing clarifications for Solicitation 36C26126R0045 concerning the Catheterization Lab and Imaging Renovation at the Fresno VA (Project Number: 570-21-220).

The RFI responses address two key issues: First, Amendment 0002 contained a mislabeled shielding design file for X-Ray Room 2B04 that actually contained information for Room 2B02. The VA directed bidders to Amendment 0003 for all corrected and properly named shielding reports, including the correct "Shielding Design Xray Room 2B04" document. Second, regarding Ultrasound Rooms 1, 2, and 3, the VA confirmed that these rooms do not require lead shielding and therefore no shielding reports are necessary for these spaces. This clarification document serves to correct documentation errors and provide definitive requirements for radiation shielding specifications as part of the imaging renovation project.

View the file

Other files

File Type Posted
36C26126R0045 Amendment 0004.pdf PDF
Wage Determination CA20260018 dtd 7 14 2026.pdf PDF
2A02B Shielding Design X Ray Room.pdf PDF
2A12B Shielding Design Cath Lab.pdf PDF
2B04 Shielding Design X Ray Room.pdf PDF
2A25B Shielding Design RF Room.pdf PDF
36C26126R0045 Amendment 0003.pdf PDF
2B02 Shielding Design X Ray Room.pdf PDF
RFI Responses 36C26126R0045 2.pdf PDF
X1_ACM_1.PDF PDF
QC-072_1.PDF PDF
Q2022-304 VA Fresno Cath Lab rev03.pdf PDF
0654C6R1.pdf PDF
SEC 21793 VA FRESNO RF 3 12.22.2021 ELECT DWGS 100 CD Binder1.pdf PDF
C3 Montr Carrier.PDF PDF
Shielding Design X-Ray Room 2A02B.pdf PDF
C2-EMS.PDF PDF
11028.pdf PDF
Fresno VA FA R3 N-WES250405_Rev00 08.25.25.pdf PDF
VA FRESNO CATH LAB_R-RAD3_26_E_PF1_03-18-2026.pdf PDF
EVOLUTION TYPICAL_WS2 RAIL_TBL2_08-14-2023.pdf PDF
ELECTRICAL_EQUIPMENT_DRAWING_-_HARMONYAIR_SURGICAL_LIGHTING_SYSTEM__A-SERIES.pdf PDF
Privacy-Curtains-And-Track-Curtain-Accessories-Product-Sheet-PS-CCACH.pdf PDF
C3-Monitor Carrier.PDF PDF
USA_Data_Sheets Flow Splitter 651 06.pdf PDF
GE-Logiq-E10-Brochure-1.pdf PDF
VA FRESNO CATH LAB_R-RAD1_26_E_PF1_03-17-2026.pdf PDF
P300Datasheet.pdf PDF
FS with stops.pdf PDF
VA FRESNO CATH LAB_R-RAD2_26_E_PF1_03-18-2026.pdf PDF
Partial Brand Name JA Cath Lab_RD.pdf PDF
Addendum 0002 - Specifications Added.pdf PDF
Addendum 0002 - Drawings Added.pdf PDF
Shielding Design X-Ray Room 2B04.pdf PDF
36C26126R0045 Amendment 0002.pdf PDF
Shielding Design X-Ray Room 2B02.pdf PDF
Shielding Design RF Room 2A25B.pdf PDF
Privacy-Curtains-And-Track-Standard-Track-Product-Sheet-PS-CC6062.pdf PDF
C1-Anesthesia.PDF PDF
VHA ICRA-1.5 (July 2 2026).pdf PDF
570-21-220 Sign in Sheet - Site Visit 6_8_26.pdf PDF
36C26126R0045 0001.docx DOCX document
7. Factor 2 Past Performance Questionnaire.docx DOCX document
6. Pre Proposal Site Visit Flyer 36C26126R0045.pdf PDF
3. 570-21-220 Drawings.pdf PDF
4. Requests for Information RFI_ Form.docx DOCX document
5. Price Breakdown Worksheet.docx DOCX document
8. WD CA20260018 dtd 5 18 2026.pdf PDF
36C26126R0045_2.docx DOCX document
1. 570-21-220 - Specifications.pdf PDF
Show 40 more

Latest 50 of 51 listed here. See the rest on GovTribe.

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

VA RESPONSES TO REQUESTS FOR INFORMATION (RFI’S) PART II

SOLICITATION# 36C26126R0045

CATHETERIZATION LAB AND IMAGING RENOVATION, AT THE FRESNO VA,

PROJECT NUMBER: 570-21-220

RFI #

1: Amendment 0002 file "ShieldingDesignXRayRoom2B04.pdf" is mislabeled — title block and content are for

X-Ray Room 2B02, duplicating the separately provided 2B02 report. No shielding report has been issued for

Room 2B04. Please issue the correct shielding design report for X-Ray Room 2B04.

Response: See Amendment 0003 for all attached Shielding reports re-posted and named properly, also

“Shielding Design Xray Room 2BO4”

2: Amendment 0002 did not include shielding reports for Ultrasound Rooms 1, 2, or 3. Please confirm whether these rooms require lead shielding. If so, please provide the corresponding shielding reports..

Response: Ultrasound rooms do not require lead shielding.

END OF RFI RESPONSES PART II