C3 Montr Carrier.PDF
PDF 2 MB Posted
- Attached to
- Z2DB--Renovate Imaging and Catheterization Lab 570 21 220 Fresno VA Medical Center Fresno
- Solicitation number
- 36C26126R0045
About this file
This is a technical specification and installation document for a HarmonyAIR Monitor Carrier system being installed at VA Fresno in Cath Lab 1. The project is scheduled for installation completion by May 12, 2021, with pre-installation items required by January 12, 2021. STERIS is the vendor, with Lee Jacobsen serving as the Account Manager and Jay Chernault as the Project Manager.
The installation includes a single HarmonyAIR Monitor Carrier with an Extended Short 1600mm upper arm (Philips Azurion compatible), part number HCES160005, featuring a DA 102 heavy spring (62x40, 2600mm) and electro-mechanical brakes. The system is designed to support a maximum load of 815 lbs with a moment of 4,988 ft-lbs. The electrical requirements include two circuits: a 6-amp circuit for spring arm up/down controls and electro-mechanical brakes, and a 20-amp circuit for monitor power with four IEC320-C13 outlets. The structural support requires a 28-inch square mounting plate (1-inch A36 steel provided by STERIS) integrated into the building's structural framework at a mounting height of 17.47 inches from the finished ceiling. Two ceiling openings are required—one 23-inch square opening for the EMS unit and one 24x24-inch access panel for maintenance. The document includes detailed engineering specifications, bolt hole patterns, gas supply requirements (50-95 PSIG recommended), and comprehensive electrical and structural installation drawings spanning 16 pages of technical specifications and schematics.
View the file
Other files
Show 40 more
50 files listed here. See the rest on GovTribe.
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Account Name: VA Fresno Room Application Cardiac Cath Room Numbers: Cath Lab 1 Quantity: 1 Date Required: 5/12/2021 Date Pre-installation items required: 1/12/2021 STERIS PO Number: Order Item No:
Customer PO: Customer Item Number:
STERIS Account Manager: Lee Jacobsen Voice Mail: 916 730-2029 STERIS Project Manager: Jay Chernault Voice Mail: 714 743-0837
VA Fresno
Fresno, CA
10/13/2022 1 of 16Item no: C3H-number:
HarmonyAIR Monitor Carrier-Monitor Carrier
Table Of Contents
SHT. NO. CONTENTS
1 Account Information
2 Table of Contents
3 Elevation and Plan Overviews
4 Electrical Panel Details
5 No Charge Preinstall Items
6 Support Structure Mounting Height Detail
7 Support Structure Detail Single
8 Structure Plate Bolt Hole Pattern
9 Support Structure Engineering Requirements
10 Ceiling Access Requirements
11 Ceiling Access Requirements - Single
12 Structural Plate Instructions
13 Gas Details Tandem EMS United States Page 1
14 Gas Details Tandem EMS United States Page 2
15 Electrical Details EMS United States Page 1
16 Electrical Details EMS United States Page 2
Special Instructions:
Description Part Number Qty/Unit
HarmonyAIR Monitor Carrier Canopy Mount Ceiling Kit (Single Mount)
HCKIT00003 1
Drop One HarmonyAir Monitor Carrier Extended Short 1600mm Upper Arm Philips Compatibility Kit - Azurion Version
HCES160005 1
DA 102 in Heavy Spring (62x40) (2600mm) (Extended Short)
Electro-Mechanical Brakes 1
VA Fresno
Fresno, CA
10/13/2022 2 of 16Item no: C3H-number:
Accessory Placement to be determined at the time of installation by the STERIS PDM & AM.
Each grid line above represents 3" distance.
Mount Specific Load / Moment Values Arm Description Load (lbs) Moment (ft-lbs)
Drop #1 HarmonyAIR DA 102 in Heavy Spring 62x40 in (1600x1000mm) (300mm Bearing)
815 4988
TOTAL 815 4988
For Max load/moment values see"Support Structure Engineering Requirements" detail page in equipment drawing section.
For siesmic zones consult project specific calculations. Contact your local STERIS Project Design Manager. (sterisarchportal.com)
VA Fresno
Fresno, CA
10/13/2022 3 of 16Item no: C3H-number:
HarmonyAIR Monitor Carrier Electrical Circuit(s)
EMS CIRCUIT TYPE AMP QTY
EMS (MONITOR CARRIER)
SPRING ARM UP/DOWN CONTROLS AND E/M BRAKES
6 1
EMS (MONITOR CARRIER)
MONITOR(S) POWER - (EM) - 4x IEC320-C13
20 1
VA Fresno
Fresno, CA
10/13/2022 4 of 16Item no: C3H-number:
Description Part# Qty/Unit Total Qty.
HarmonyAIR Above Ceiling Pre-Install Electric , Gas Kit
QR0000000000022 1 1
VA Fresno
Fresno, CA
10/13/2022 5 of 16Item no: C3H-number:
VA Fresno
Fresno, CA
10/13/2022 6 of 16Item no: C3H-number:
10/13/2022 7 of 16Item no: C3H-number:
10/13/2022 8 of 16Item no: C3H-number:
10/13/2022 9 of 16Item no: C3H-number:
10/13/2022 10 of 16Item no: C3H-number:
10/13/2022 11 of 16Item no: C3H-number:
10/13/2022 12 of 16Item no: C3H-number:
10/13/2022 13 of 16Item no: C3H-number:
10/13/2022 14 of 16Item no: C3H-number:
10/13/2022 15 of 16Item no: C3H-number:
10/13/2022 16 of 16Item no: C3H-number:
File details come from the government source that posted it. Updated .