2B02 Shielding Design X Ray Room.pdf

PDF 802 KB Posted

Attached to
Z2DB--Renovate Imaging and Catheterization Lab 570 21 220 Fresno VA Medical Center Fresno
Solicitation number
36C26126R0045
Issued by

About this file

This is a Radiation Shielding Design Calculations report for X-Ray Room 2B02 at VA Central California in Fresno, California, dated September 30, 2022, prepared by Jeremy L. Corwin, MHP, DABHP, a Medical Physicist.

The report specifies minimum radiation shielding requirements for a new radiographic system with digital radiography (DR) capability, based on a workload of 140 patients per week and regulatory exposure limits of 0.10 mGy/week for controlled areas and 0.02 mGy/week for uncontrolled areas. Key shielding specifications include: Chest Bucky/Exterior Wall (Wall A) requires no additional shielding due to existing 6 inches of concrete; Mechanical Room Wall (Wall B), Storage Room Wall (Wall C), Control Wall (Wall D), and Cross-Table/Adjacent X-Ray Wall (Wall G) each require 0.79 mm lead shielding; Hallway Door (Wall F) requires 0.79 mm lead shielding; Hallway Wall (Wall E) requires no additional shielding with existing two layers of 5/8-inch gypsum wallboard; Floor and Ceiling require no additional shielding with existing 4-inch concrete slabs. All shielding must extend from floor to 7 feet height unless otherwise noted. The Control Window must be at least 0.79 mm lead equivalence and 1 foot by 1 foot minimum size, with the exposure switch positioned at least 40 inches from the control area edge and maintaining at least 7.5 square feet of unobstructed floor space. The design references NCRP Report 147 and notes that shielding thickness may increase if occupancy, x-ray system type, or workload changes. The report includes detailed recommendations for shielding penetrations (pipes, conduits, electrical panels) and a commercial lead shielding schedule. California registration requirements for x-ray machines and optional county health department submissions (Los Angeles and San Diego Counties) are noted.

View the file

Other files

File Type Posted
36C26126R0045 Amendment 0004.pdf PDF
Wage Determination CA20260018 dtd 7 14 2026.pdf PDF
2B04 Shielding Design X Ray Room.pdf PDF
2A25B Shielding Design RF Room.pdf PDF
2A12B Shielding Design Cath Lab.pdf PDF
2A02B Shielding Design X Ray Room.pdf PDF
RFI Responses Part II 36C26126R0045.pdf PDF
Q2022-304 VA Fresno Cath Lab rev03.pdf PDF
RFI Responses 36C26126R0045 2.pdf PDF
QC-072_1.PDF PDF
X1_ACM_1.PDF PDF
C3 Montr Carrier.PDF PDF
Shielding Design RF Room 2A25B.pdf PDF
VHA ICRA-1.5 (July 2 2026).pdf PDF
Privacy-Curtains-And-Track-Standard-Track-Product-Sheet-PS-CC6062.pdf PDF
C1-Anesthesia.PDF PDF
VA FRESNO CATH LAB_R-RAD3_26_E_PF1_03-18-2026.pdf PDF
VA FRESNO CATH LAB_R-RAD2_26_E_PF1_03-18-2026.pdf PDF
P300Datasheet.pdf PDF
FS with stops.pdf PDF
Partial Brand Name JA Cath Lab_RD.pdf PDF
Addendum 0002 - Specifications Added.pdf PDF
C3-Monitor Carrier.PDF PDF
Shielding Design X-Ray Room 2A02B.pdf PDF
SEC 21793 VA FRESNO RF 3 12.22.2021 ELECT DWGS 100 CD Binder1.pdf PDF
Privacy-Curtains-And-Track-Curtain-Accessories-Product-Sheet-PS-CCACH.pdf PDF
C2-EMS.PDF PDF
Addendum 0002 - Drawings Added.pdf PDF
0654C6R1.pdf PDF
Shielding Design X-Ray Room 2B02.pdf PDF
Shielding Design X-Ray Room 2B04.pdf PDF
VA FRESNO CATH LAB_R-RAD1_26_E_PF1_03-17-2026.pdf PDF
USA_Data_Sheets Flow Splitter 651 06.pdf PDF
GE-Logiq-E10-Brochure-1.pdf PDF
Fresno VA FA R3 N-WES250405_Rev00 08.25.25.pdf PDF
11028.pdf PDF
EVOLUTION TYPICAL_WS2 RAIL_TBL2_08-14-2023.pdf PDF
ELECTRICAL_EQUIPMENT_DRAWING_-_HARMONYAIR_SURGICAL_LIGHTING_SYSTEM__A-SERIES.pdf PDF
36C26126R0045 Amendment 0002.pdf PDF
570-21-220 Sign in Sheet - Site Visit 6_8_26.pdf PDF
36C26126R0045 0001.docx DOCX document
8. WD CA20260018 dtd 5 18 2026.pdf PDF
5. Price Breakdown Worksheet.docx DOCX document
3. 570-21-220 Drawings.pdf PDF
36C26126R0045_2.docx DOCX document
7. Factor 2 Past Performance Questionnaire.docx DOCX document
6. Pre Proposal Site Visit Flyer 36C26126R0045.pdf PDF
4. Requests for Information RFI_ Form.docx DOCX document
2. 570-21-220 - Basis of Design.pdf PDF
1. 570-21-220 - Specifications.pdf PDF
Show 40 more

50 files listed here. See the rest on GovTribe.

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Radiation Shielding Design Calculations March 21st, 2018 Facility X-Ray Room XX

Radiation Shielding Design

X-RAY ROOM 2B02

9/30/2022

VA CENTRAL CALIFORNIA

2615 E. CLINTON AVE.

FRESNO, CA 93703

Jeremy L. Corwin, MHP, DABHP Medical Physicist jeremy@corwinhp.com mailto:jeremy@corwinhp.com

Radiation Shielding Design Calculations September 30th, 2022 VA Central California - Fresno, CA X-Ray Room 2B02

Shielding Design Summary

Shielding calculations were performed for the new x-ray system that will be installed in X-Ray Room 2B02 at VA Central California in Fresno, California.

A summary table of shielding calculations is at enclosure 1. An annotated floorplan with shielding requirements is at enclosure 2. The following is a detailed explanation of the required shielding on each wall.

Outline of Required Shielding

The following shielding specifications are the minimum amount of shielding required to reduce radiation exposures to below the current regulatory limits. Installation of shielding that is thicker than specified below is acceptable. Unless noted, all shielding must extend from the floor to a height of 7 feet.

Chest Bucky / Exterior Wall - Wall A. The Chest Bucky / Exterior Wall does not require additional shielding. The existing 6” of concrete will provide adequate shielding.

Mechanical Room Wall - Wall B. The Mechanical Room Wall requires 0.79 mm lead shielding.

Storage Room Wall - Wall C. The Storage Room Wall requires 0.79 mm lead shielding.

Control Wall / Window - Wall D. The Control Wall requires 0.79 mm lead shielding.

The Control Window must also be at least 0.79 mm lead equivalence and be at least 1 foot by 1 foot in size. Please note that the x-ray exposure switch must be mounted in such a manner that it is impossible for an operator to make exposures from an unprotected area. This can be accomplished by placing the switch at a permanent position at least 40 inches from the edge of the control area. Also, there should be at least 7.5 square feet of unobstructed floor space in the control area.

Hallway Wall - Wall E. The Hallway Wall does not require additional shielding. The existing two layers of 5/8” gypsum wallboard (≈ 28 mm total) will provide adequate shielding.

Hallway Door - Wall F. The Hallway Door shall have at least 0.79 mm lead shielding.

Cross-Table / Adjacent X-Ray Wall - Wall G. The Cross-Table / Adjacent X-Ray Wall requires 0.79 mm lead shielding.

Note: The shielding design for X-Ray Room 2B04 (dated September 30th, 2022) is also calling for 0.79 mm lead shielding on this wall. As long as 0.79 mm lead shielding is installed on one side of this wall, both x-ray rooms will be adequately shielded from each other.

Floor: The Floor does not require additional shielding. The existing 4” concrete slab will provide adequate shielding.

Note: The Floor requires at least 1.36 mm lead equivalence directly below the table.

According to NCRP Report No. 147, a shielding equivalence of 0.85 mm lead can be considered already in place by the attenuation of the grid, detector, and image receptor supporting structures (Xpre). For the primary beam shielding calculation, the contribution of Xpre was considered in addition to the existing 4” concrete slab.

Ceiling: The Ceiling does not require additional shielding. The existing 4” concrete slab will provide adequate shielding.

Methods and Assumptions

The shielding requirements above were based on the following information:

The proposed x-ray system is a radiographic system with a table and an upright image receptor. The facility will use Digital Radiography (DR). Calculations were performed for a DR system.

For the purpose of shielding calculations, the maximum allowable exposure limit for controlled areas is 0.10 mGy/week while the limit for uncontrolled areas is 0.02 mGy/week.

The maximum workload of 28 patients per 8 hours was provided by the facility. The provided workload was modified for a 40 hour work week to make 140 patients/week. This modified workload was used for the shielding calculations.

The occupancies of surrounding areas are noted on the enclosed summary table of shielding calculations and the annotated floorplan.

The primary reference for these shielding calculations is NCRP Report 147, Structural Shielding Design for Medical X-Ray Imaging Facilities.

Please note the following:

These shielding recommendations are based on information provided by the facility at the time of this shielding design. If changes occur regarding the occupancy of surrounding areas, the type of x-ray system being used, or an increase in workload, additional shielding may be required.

California Radiation Control Regulations require that all x-ray machines be registered with the State of California, Department of Public Health. The Radiation Machine Registration can be completed online by following the link below.

https://cdph-rhb.powerappsportals.us/

It is believed that Los Angeles County and San Diego County are the only counties that require submission of shielding designs to their respective health departments.

Please check with your local county health department to determine if shield design plans need to be submitted for approval before construction begins.

Please keep a copy of this shielding plan at your facility indefinitely.

Thank you for allowing us to perform your shielding calculations. If you have any questions, please call me at (360) 736-6066.

Jeremy L. Corwin, MHP, DABHP Medical Physicist

CORWIN HEALTH PHYSICS INC

Enclosures:

Summary of Shielding Calculations Annotated Floorplan Appendix A - Recommendations for Shielding Penetrations Appendix B - Schedule of Commercial Lead Shielding https://cdph-rhb.powerappsportals.us/

Average kVp Rad-Table Rad-Chest Total

Radiographic System X-Ray See NCRP 147 140 350 84 434 Digital Radiography (DR) will be used

Wall Wall Primary or Type of Occupancy Distance to Description Description Secondary Occupancy Factor Occupied Area (m) Thickness (mm) Type

Chest Bucky / Exterior Wall A Primary Uncontrolled 1/40 2.44 61.12 Concrete Chest Bucky / Exterior Wall A Secondary Uncontrolled 1/40 1.52 8.94 Concrete Mechanical Room Wall B Secondary Uncontrolled 1/20 1.68 44.28 GWB Storage Room Wall C Secondary Uncontrolled 1/20 1.37 0.21 Lead Control Wall / Window D Secondary Controlled 1 1.17 0.55 Lead Hallway Wall E Secondary Uncontrolled 1/5 4.57 24.35 GWB Hallway Door F Secondary Uncontrolled 1/8 4.88 0.04 Lead Cross-Table / Adjacent X-Ray Wall G Primary Uncontrolled 1/2 5.18 0.47 Lead Cross-Table / Adjacent X-Ray Wall G Secondary Uncontrolled 1/2 4.19 0.22 Lead Floor N/A Primary Uncontrolled 1 4.42 1.36 Lead Floor N/A Secondary Uncontrolled 1 3.35 31.55 Concrete Ceiling N/A Secondary Uncontrolled 1 2.59 40.30 Concrete

Notes:

1The Door shall have at least 0.79 mm lead.

All shielding extends from the floor to a height of seven feet unless otherwise noted.

3The Floor requires at least 1.36 mm lead equivalence directly below the table. According to NCRP Report No. 147, a shielding equivalence of 0.85 mm lead can be considered already in place by the attenuation of the grid, detector, and image receptor supporting structures (Xpre). For the primary beam shielding calculation, the contribution of Xpre was considered in addition to the existing 4” concrete slab. Please see the Outline of Required Shielding for additional details regarding Xpre.

Summary of Shielding Calculations VA Central California - Fresno, CA

X-Ray Room 2B02 September 30th, 2022

28 mm GWB 0.79 mm Lead

Type of X-Ray Machine Room ID Patients Per Week Workload Information

(mA-Minutes/week)

4" Concrete None

28 mm GWB 0.79 mm Lead2

None

28 mm GWB 0.79 mm Lead2

28 mm GWB None Unknown 0.79 mm lead door1

4" Concrete None3

2The shielding design for X-Ray Room 2B04 (dated September 30th, 2022) is also calling for 0.79 mm lead shielding on this wall. As long as 0.79 mm lead shielding is installed on one side of this wall, both x-ray rooms will be adequately shielded from each other.

Shielding Specifications Calculated

Existing Shielding Required ShieldingShielding

6" Concrete None 6" Concrete None

28 mm GWB 0.79 mm Lead

4" Concrete

28 mm GWB 0.79 mm Lead

Adjacent X-Ray Room

Exterior

Hallway

Control Area

Storage Room

Mechanical Room

Scatter Radiation

Chest Bucky / Exterior Wall - Wall A. The Chest Bucky / Exterior Wall does not require additional shielding. The existing 6” of concrete will provide adequate shielding.

Control Wall / Window - Wall D.

The Control Wall requires 0.79 mm lead shielding. The Control Window must also be at least 0.79 mm lead equivalence and be at least 1 foot by 1 foot in size. See the Outline of Required Shielding for additional information.

0.79 mm Lead Placement

Cross-Table / Adjacent X-Ray Wall - Wall G.

The Cross-Table / Adjacent X-Ray Wall requires

0.79 mm lead shielding.

Note: The shielding design for X-Ray Room 2B04 (dated September 30th, 2022) is also calling for

0.79 mm lead shielding on this wall. As long as

0.79 mm lead shielding is installed on one side of this wall, both x-ray rooms will be adequately shielded from each other.

Hallway Door - Wall F.

The Hallway Door shall have at least 0.79 mm lead shielding.

Hallway Wall - Wall E.

The Hallway Wall does not require additional shielding. The existing two layers of 5/8” gypsum wallboard ( 28 mm total) will provide adequate shielding.

Storage Room Wall - Wall C.

The Storage Room Wall requires

0.79 mm lead shielding.

Mechanical Room Wall - Wall B. The Mechanical Room Wall requires 0.79 mm lead shielding.

Floor: The Floor does not require additional shielding. The existing 4” concrete slab will provide adequate shielding. See the Outline of Required Shielding for additional information.

Ceiling: The Ceiling does not require additional shielding. The existing 4” concrete slab will provide adequate shielding.

Appendix A – Recommendations for Shielding Penetrations

Shielding Screw/Nails: Penetrations made by screws/nails do not require lead “tabs” to cover the top of the screws/nails. The steel nails/screws generally attenuate radiation equally, or more effectively than the lead displaced by the screws/nails.

Shielding Joints: The joints between lead sheets should be constructed so that their surfaces are in contact with an overlap of not less than 1 cm. This can be accomplished by using lead battens, or by using drywall that is manufactured with the lead sheet extending beyond the edge of the drywall for adequate overlap.

Shielding Penetrations: Air conditioning ducts, electrical conduit, plumbing, and other infrastructure will penetrate shielded walls, floors, and ceilings. The shielding of the x-ray room shall be so constructed that the protection is not impaired by these openings or by service boxes, etc., embedded in barriers. The duct/conduit shielding thickness shall be equivalent to the displaced shielding thickness. Below are general shielding details for duct/pipe penetrations into a shielded surface where H is the diameter of the penetration.

1. Straight pipe/conduit penetration: Situations where there is a straight pipe/conduit penetrating a lead shielded surface require lead to line (wrap) the pipe/conduit twice the distance as the diameter/thickness of the pipe or conduit. See Detail 1 below.

Detail 1:

2. Curved elbow penetration: Situations where there is a curved pipe/conduit penetrating lead shielding requires lead to line (wrap) to line the entire top/backside of the elbow and pipe for 3 times the distance of the diameter of the opening. See Detail 2 below.

Detail 2:

3. Flat Panel Penetrations (electrical panels etc.). Flat panel penetrations can be shielded by either wrapping the entire structure (panel/box) being installed or by shielding behind the panel/box. If the lead shielding is installed behind the panel, the lead must extend beyond each side of the panel (penetration) by twice the thickness of the wall. For example, if the wall is 4 inches thick, the lead shielding must extend 8 inches beyond each edge of the cut-out section of the wall.

4. 90 degree elbow running out of the wall: Situations where there is a 90 degree elbow penetrating a lead shielded surface requires lead to line the top of the elbow up to the 90 degree turn, and then the back side of the pipe for twice the distance of the diameter/thickness of the pipe PLUS the thickness of the wall. See Detail 3 below:

Detail 3:

The above recommendations and details are adequate for shielding most common shielding penetrations. There may be exceptions where these methods are not appropriate for providing adequate radiation protection. Please contact Corwin Health Physics if there are other questions on how to appropriately shield a space.

Reference: The primary reference for these recommendations is NCRP Report 147.

Appendix B – Schedule of Commercial Lead Shielding

Thickness (mm)

Thickness (inches)

Weight in Pounds per Square Foot (PSF)

0.79 1/32 2

1.00 5/128 2.5

1.19 3/64 3

1.58 1/16 4

1.98 5/64 5

2.38 3/32 6

3.00 -- 7.5

3.17 1/8 8

3.97 5/32 10

4.76 3/16 12

5.55 7/32 13

6.35 1/4 16

8.47 1/3 20

9.52 3/8 24

12.70 1/2 30

19.05 3/4 45

25.40 1 60