C3-Monitor Carrier.PDF
PDF 2 MB Posted
- Attached to
- Z2DB--Renovate Imaging and Catheterization Lab 570 21 220 Fresno VA Medical Center Fresno
- Solicitation number
- 36C26126R0045
About this file
This document is a technical specification and installation guide for a HarmonyAIR Monitor Carrier system for cardiac catheterization lab use at VA Fresno in Fresno, California. The installation is scheduled for May 12, 2021, with pre-installation items required by January 12, 2021. The project involves a single Cath Lab 1 room application with one monitor carrier unit featuring an extended short 1600mm upper arm with Philips compatibility (Azurion version), DA 102 heavy spring configuration (62x40 inches, 2600mm), and electro-mechanical brakes.
The system requires specific infrastructure including a structural mounting plate (28" x 28" reference dimensions) with defined bolt hole patterns for both single and tandem mount configurations, a 23-inch square ceiling opening with a 24x24-inch access panel, and gas and electrical pre-installation components. Electrical requirements include two circuits: a 6-amp circuit for spring arm controls and electro-mechanical brakes, and a 20-amp circuit for monitor power with four IEC320-C13 outlets. The maximum load is 815 pounds with a moment of 4988 ft-lbs. Gas specifications include multiple medical gas types (CO2, N, O2, medical air, N2O, WADD/EVAC, vacuum, and instrument air) with pressure ratings between 50-189 PSIG. All structural design, support bracing, and installation responsibility rests with the customer's building engineer of record, with STERIS providing specific components including the canopy mount ceiling kit, bearing flange, J-box branch circuits, and gas riser bracket.
View the file
Other files
Show 40 more
Latest 50 of 51 listed here. See the rest on GovTribe.
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Account Name: VA Fresno Room Application Cardiac Cath Room Numbers: Cath Lab 1 Quantity: 1 Date Required: 5/12/2021 Date Pre-installation items required: 1/12/2021 STERIS PO Number: Order Item No:
Customer PO: Customer Item Number:
STERIS Account Manager: Lee Jacobsen Voice Mail: 916 730-2029 STERIS Project Manager: Jay Chernault Voice Mail: 714 743-0837
VA Fresno
Fresno, CA
10/13/2022 1 of 16Item no: C3H-number:
HarmonyAIR Monitor Carrier-Monitor Carrier
Table Of Contents
SHT. NO. CONTENTS
1 Account Information
2 Table of Contents
3 Elevation and Plan Overviews
4 Electrical Panel Details
5 No Charge Preinstall Items
6 Support Structure Mounting Height Detail
7 Support Structure Detail Single
8 Structure Plate Bolt Hole Pattern
9 Support Structure Engineering Requirements
10 Ceiling Access Requirements
11 Ceiling Access Requirements - Single
12 Structural Plate Instructions
13 Gas Details Tandem EMS United States Page 1
14 Gas Details Tandem EMS United States Page 2
15 Electrical Details EMS United States Page 1
16 Electrical Details EMS United States Page 2
Special Instructions:
Description Part Number Qty/Unit
HarmonyAIR Monitor Carrier Canopy Mount Ceiling Kit (Single Mount)
HCKIT00003 1
Drop One HarmonyAir Monitor Carrier Extended Short 1600mm Upper Arm Philips Compatibility Kit - Azurion Version
HCES160005 1
DA 102 in Heavy Spring (62x40) (2600mm) (Extended Short)
Electro-Mechanical Brakes 1
VA Fresno
Fresno, CA
10/13/2022 2 of 16Item no: C3H-number:
Accessory Placement to be determined at the time of installation by the STERIS PDM & AM.
Each grid line above represents 3" distance.
Mount Specific Load / Moment Values Arm Description Load (lbs) Moment (ft-lbs)
Drop #1 HarmonyAIR DA 102 in Heavy Spring 62x40 in (1600x1000mm) (300mm Bearing)
815 4988
TOTAL 815 4988
For Max load/moment values see"Support Structure Engineering Requirements" detail page in equipment drawing section.
For siesmic zones consult project specific calculations. Contact your local STERIS Project Design Manager. (sterisarchportal.com)
VA Fresno
Fresno, CA
10/13/2022 3 of 16Item no: C3H-number:
HarmonyAIR Monitor Carrier Electrical Circuit(s)
EMS CIRCUIT TYPE AMP QTY
EMS (MONITOR CARRIER)
SPRING ARM UP/DOWN CONTROLS AND E/M BRAKES
6 1
EMS (MONITOR CARRIER)
MONITOR(S) POWER - (EM) - 4x IEC320-C13
20 1
VA Fresno
Fresno, CA
10/13/2022 4 of 16Item no: C3H-number:
Description Part# Qty/Unit Total Qty.
HarmonyAIR Above Ceiling Pre-Install Electric , Gas Kit
QR0000000000022 1 1
VA Fresno
Fresno, CA
10/13/2022 5 of 16Item no: C3H-number:
VA Fresno
Fresno, CA
10/13/2022 6 of 16Item no: C3H-number:
10/13/2022 7 of 16Item no: C3H-number:
10/13/2022 8 of 16Item no: C3H-number:
10/13/2022 9 of 16Item no: C3H-number:
10/13/2022 10 of 16Item no: C3H-number:
10/13/2022 11 of 16Item no: C3H-number:
10/13/2022 12 of 16Item no: C3H-number:
10/13/2022 13 of 16Item no: C3H-number:
10/13/2022 14 of 16Item no: C3H-number:
10/13/2022 15 of 16Item no: C3H-number:
10/13/2022 16 of 16Item no: C3H-number: