This document is a technical specification and installation guide for a HarmonyAIR Monitor Carrier system for cardiac catheterization lab use at VA Fresno in Fresno, California. The installation is scheduled for May 12, 2021, with pre-installation items required by January 12, 2021. The project involves a single Cath Lab 1 room application with one monitor carrier unit featuring an extended short 1600mm upper arm with Philips compatibility (Azurion version), DA 102 heavy spring configuration (62x40 inches, 2600mm), and electro-mechanical brakes.

The system requires specific infrastructure including a structural mounting plate (28" x 28" reference dimensions) with defined bolt hole patterns for both single and tandem mount configurations, a 23-inch square ceiling opening with a 24x24-inch access panel, and gas and electrical pre-installation components. Electrical requirements include two circuits: a 6-amp circuit for spring arm controls and electro-mechanical brakes, and a 20-amp circuit for monitor power with four IEC320-C13 outlets. The maximum load is 815 pounds with a moment of 4988 ft-lbs. Gas specifications include multiple medical gas types (CO2, N, O2, medical air, N2O, WADD/EVAC, vacuum, and instrument air) with pressure ratings between 50-189 PSIG. All structural design, support bracing, and installation responsibility rests with the customer's building engineer of record, with STERIS providing specific components including the canopy mount ceiling kit, bearing flange, J-box branch circuits, and gas riser bracket.

View the file

Other files

File Type Posted
36C26126R0045 Amendment 0004.pdf PDF
Wage Determination CA20260018 dtd 7 14 2026.pdf PDF
RFI Responses Part II 36C26126R0045.pdf PDF
2A02B Shielding Design X Ray Room.pdf PDF
2A12B Shielding Design Cath Lab.pdf PDF
2B04 Shielding Design X Ray Room.pdf PDF
2A25B Shielding Design RF Room.pdf PDF
36C26126R0045 Amendment 0003.pdf PDF
2B02 Shielding Design X Ray Room.pdf PDF
RFI Responses 36C26126R0045 2.pdf PDF
X1_ACM_1.PDF PDF
QC-072_1.PDF PDF
Q2022-304 VA Fresno Cath Lab rev03.pdf PDF
0654C6R1.pdf PDF
SEC 21793 VA FRESNO RF 3 12.22.2021 ELECT DWGS 100 CD Binder1.pdf PDF
C3 Montr Carrier.PDF PDF
Shielding Design X-Ray Room 2A02B.pdf PDF
C2-EMS.PDF PDF
11028.pdf PDF
Fresno VA FA R3 N-WES250405_Rev00 08.25.25.pdf PDF
VA FRESNO CATH LAB_R-RAD3_26_E_PF1_03-18-2026.pdf PDF
EVOLUTION TYPICAL_WS2 RAIL_TBL2_08-14-2023.pdf PDF
ELECTRICAL_EQUIPMENT_DRAWING_-_HARMONYAIR_SURGICAL_LIGHTING_SYSTEM__A-SERIES.pdf PDF
Privacy-Curtains-And-Track-Curtain-Accessories-Product-Sheet-PS-CCACH.pdf PDF
USA_Data_Sheets Flow Splitter 651 06.pdf PDF
GE-Logiq-E10-Brochure-1.pdf PDF
VA FRESNO CATH LAB_R-RAD1_26_E_PF1_03-17-2026.pdf PDF
P300Datasheet.pdf PDF
FS with stops.pdf PDF
VA FRESNO CATH LAB_R-RAD2_26_E_PF1_03-18-2026.pdf PDF
Partial Brand Name JA Cath Lab_RD.pdf PDF
Addendum 0002 - Specifications Added.pdf PDF
Addendum 0002 - Drawings Added.pdf PDF
Shielding Design X-Ray Room 2B04.pdf PDF
36C26126R0045 Amendment 0002.pdf PDF
Shielding Design X-Ray Room 2B02.pdf PDF
Shielding Design RF Room 2A25B.pdf PDF
Privacy-Curtains-And-Track-Standard-Track-Product-Sheet-PS-CC6062.pdf PDF
C1-Anesthesia.PDF PDF
VHA ICRA-1.5 (July 2 2026).pdf PDF
570-21-220 Sign in Sheet - Site Visit 6_8_26.pdf PDF
36C26126R0045 0001.docx DOCX document
7. Factor 2 Past Performance Questionnaire.docx DOCX document
6. Pre Proposal Site Visit Flyer 36C26126R0045.pdf PDF
3. 570-21-220 Drawings.pdf PDF
4. Requests for Information RFI_ Form.docx DOCX document
5. Price Breakdown Worksheet.docx DOCX document
8. WD CA20260018 dtd 5 18 2026.pdf PDF
36C26126R0045_2.docx DOCX document
1. 570-21-220 - Specifications.pdf PDF
Show 40 more

Latest 50 of 51 listed here. See the rest on GovTribe.

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Account Name: VA Fresno Room Application Cardiac Cath Room Numbers: Cath Lab 1 Quantity: 1 Date Required: 5/12/2021 Date Pre-installation items required: 1/12/2021 STERIS PO Number: Order Item No:

Customer PO: Customer Item Number:

STERIS Account Manager: Lee Jacobsen Voice Mail: 916 730-2029 STERIS Project Manager: Jay Chernault Voice Mail: 714 743-0837

VA Fresno

Fresno, CA

10/13/2022 1 of 16Item no: C3H-number:

HarmonyAIR Monitor Carrier-Monitor Carrier

Table Of Contents

SHT. NO. CONTENTS

1 Account Information

2 Table of Contents

3 Elevation and Plan Overviews

4 Electrical Panel Details

5 No Charge Preinstall Items

6 Support Structure Mounting Height Detail

7 Support Structure Detail Single

8 Structure Plate Bolt Hole Pattern

9 Support Structure Engineering Requirements

10 Ceiling Access Requirements

11 Ceiling Access Requirements - Single

12 Structural Plate Instructions

13 Gas Details Tandem EMS United States Page 1

14 Gas Details Tandem EMS United States Page 2

15 Electrical Details EMS United States Page 1

16 Electrical Details EMS United States Page 2

Special Instructions:

Description Part Number Qty/Unit

HarmonyAIR Monitor Carrier Canopy Mount Ceiling Kit (Single Mount)

HCKIT00003 1

Drop One HarmonyAir Monitor Carrier Extended Short 1600mm Upper Arm Philips Compatibility Kit - Azurion Version

HCES160005 1

DA 102 in Heavy Spring (62x40) (2600mm) (Extended Short)

Electro-Mechanical Brakes 1

VA Fresno

Fresno, CA

10/13/2022 2 of 16Item no: C3H-number:

Accessory Placement to be determined at the time of installation by the STERIS PDM & AM.

Each grid line above represents 3" distance.

Mount Specific Load / Moment Values Arm Description Load (lbs) Moment (ft-lbs)

Drop #1 HarmonyAIR DA 102 in Heavy Spring 62x40 in (1600x1000mm) (300mm Bearing)

815 4988

TOTAL 815 4988

For Max load/moment values see"Support Structure Engineering Requirements" detail page in equipment drawing section.

For siesmic zones consult project specific calculations. Contact your local STERIS Project Design Manager. (sterisarchportal.com)

VA Fresno

Fresno, CA

10/13/2022 3 of 16Item no: C3H-number:

HarmonyAIR Monitor Carrier Electrical Circuit(s)

EMS CIRCUIT TYPE AMP QTY

EMS (MONITOR CARRIER)

SPRING ARM UP/DOWN CONTROLS AND E/M BRAKES

6 1

EMS (MONITOR CARRIER)

MONITOR(S) POWER - (EM) - 4x IEC320-C13

20 1

VA Fresno

Fresno, CA

10/13/2022 4 of 16Item no: C3H-number:

Description Part# Qty/Unit Total Qty.

HarmonyAIR Above Ceiling Pre-Install Electric , Gas Kit

QR0000000000022 1 1

VA Fresno

Fresno, CA

10/13/2022 5 of 16Item no: C3H-number:

VA Fresno

Fresno, CA

10/13/2022 6 of 16Item no: C3H-number:

10/13/2022 7 of 16Item no: C3H-number:

10/13/2022 8 of 16Item no: C3H-number:

10/13/2022 9 of 16Item no: C3H-number:

10/13/2022 10 of 16Item no: C3H-number:

10/13/2022 11 of 16Item no: C3H-number:

10/13/2022 12 of 16Item no: C3H-number:

10/13/2022 13 of 16Item no: C3H-number:

10/13/2022 14 of 16Item no: C3H-number:

10/13/2022 15 of 16Item no: C3H-number:

10/13/2022 16 of 16Item no: C3H-number: