0654C6R1.pdf
PDF 831 KB Posted
- Attached to
- Z2DB--Renovate Imaging and Catheterization Lab 570 21 220 Fresno VA Medical Center Fresno
- Solicitation number
- 36C26126R0045
About this file
This document is an architectural floor plan and equipment layout specification for a Catheterization (Cath) Lab at the VA Central California Health Care System in Fresno, California. The plan details the proposed installation of medical ceiling-mounted equipment manufactured by STERIS Corporation, including specifications for the HarmonyAIR A-Series surgical lighting and monitoring system.
The equipment layout specifies four primary components: a Tandem HarmonyAIR Adjustable Arm Spring with anesthesia accessory arm (C1), a Tandem HarmonyAIR Fixed surgical light with accessory arm (C2), a HarmonyAIR Monitor Carrier (C3), and a HarmonyAIR A-Series Integrated Wall Control (K1). The electrical contractor is responsible for installing two 0.75-inch conduit runs from K1 to C2, with a 0.5-inch conduit for AC power supply (120V-240VAC, 7A-3A, 50/60Hz) to the wall control unit. Critical specifications include conduit runs not exceeding 25 feet, 1-inch conduit for DC supply light head cables on long runs, and floor-level tolerance of ±1/16 inch per 39 inches, with ceiling height of 106 5/16 inches. A revision note dated 09 September 2022 indicates K1 relocation, and the drawing is based on Philips-Azurion 7 C20 FlexArm 4300mm imaging equipment. The document includes a disclaimer that STERIS ceiling equipment locations are preliminary and must be reviewed against current imaging vendor drawings, and that indicated service clearances do not meet minimum safety clearances recommended in STERIS equipment drawings.
View the file
Other files
Show 40 more
50 files listed here. See the rest on GovTribe.
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Le ve l o f f lo or
1 16
2m m ) p er
9" (1 m ) w ith in th is a re a.
Ta bl e Sa fe ty
Zo ne
M ot or iz ed M ov em en t L in e
C1
C2
C3
41316"
1 8
90" 78"
C1 C3
2"-33"
78" 90"
90" C2
16"
5 1
C E I L I N G
H E I G H T
5 1
C E I L I N G
H E I G H T
K1
EQUIPMENT LIST
TAG QTY DESCRIPTION
C1 1 Tandem HarmonyAIR Adj Arm Spring - A Series-Anesthesia, Accessory Arm(s)
C2 1 Tandem HarmonyAIR Fixed - A Series-Equipment, Surgical Light(s) & Accessory Arm(s)
C3 1 HarmonyAIR Monitor Carrier-Monitor Carrier
K1 1 HarmonyAIR A-Series Integrated Wall Control
A SERIES LIGHT CONDUIT SCHEDULE
FROM - TO QUANTITY SIZE RESPONSIBILITY
K1 - C2 2 .75" ELECTRICAL CONTRACTOR
LIGHT CONDUIT SCHEDULE NOTES
1. .5" CONDUIT FOR AC POWER TO K1; 3 COND. 10AWG WIRE
INPUT AC SUPPLY - 120V-240VAC, 7A-3A, 50/60Hz
2. LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET
3. CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT
ALSO REFERENCED WITHIN THE EQUIPMENT DRAWINGS FOR
THE SPECIFIED SURGICAL LIGHT WALL CONTROL SYSTEM
4. DC SUPPLY LIGHTHEAD CABLES REQUIRE 1" CONDUIT WHEN
DOUBLE RUNS ARE REQUIRED (LONG RUNS)
5. REFER TO APPROPRIATE EQUIPMENT SPECIFICATIONS TO
ENSURE THAT THE MOST CURRENT & DETAILED INFORMATION
OF THIS LIGHT SYSTEM IS UTILIZED
MOUNT C2 & C3
LIMITED UPPER ARM
ROTATIONS AT WALL
SCALE: 1/4" = 1'-0"
PROPOSED EQUIPMENT LAYOUT1
SCALE: 1/4" = 1'-0"
FOOT END EQUIPMENT ELEVATION2
SCALE: 1/4" = 1'-0"
FOOT END EQUIPMENT ELEVATION3
Clea / Table
AD
Ta bl e Fl oo r P la te
N1N13 N2
N1
10'-11"
'-9
Le ve l o f f lo or
1 16
2m m pe r 3
9" (1 m ) w ith in th is a re a.
x2 N
1x2
REVISION HISTORY
REVISION # DESCRIPTION PLANNER/DATE
1 UPDATED VENDOR CAD BACKGROUND; RELOCATED K1 JTP/09SEP2022
* THE SERVICE CLEARANCES INDICATED DO NOT MEET THE MINIMUM SAFETY CLEARANCES RECOMMENDED IN THE EQUIPMENT DRAWINGS PROVIDED BY STERIS CORPORATION
Planning & Design Group
STERIS
5960 Heisley Road
Mentor, OH 44060 USA 800-548-4873
Dr
Ck
Date
Scale
H- C
STERIS drawings do not supersede the drawings from the architect of record.
ALL STERIS CEILING EQUIPMENT LOCATIONS ARE
PRELIMINARY AND MUST BE REVIEWED AGAINST MOST
CURRENT IMAGING VENDOR DRAWINGS. THIS
DRAWING BASED ON:
DATE REVISION # DESCRIPTION
F 08/16/2022 PHILIPS - AZURION 7 C20 FLEXARM 4300MM
-N-WES190300FF
1/4" = 1'-0"
12AUG2022
JTP
6R1
PARTIAL FLOOR PLAN
CATH LAB
PROPOSED EQUIPMENT LAYOUT
VA CENTRAL CALIFORNIA
HEALTH CARE SYSTEM
FRESNO, CALIFORNIA
| Sheets and Views |
| LAYOUT |
File details come from the government source that posted it. Updated .