This document is an architectural floor plan and equipment layout specification for a Catheterization (Cath) Lab at the VA Central California Health Care System in Fresno, California. The plan details the proposed installation of medical ceiling-mounted equipment manufactured by STERIS Corporation, including specifications for the HarmonyAIR A-Series surgical lighting and monitoring system.

The equipment layout specifies four primary components: a Tandem HarmonyAIR Adjustable Arm Spring with anesthesia accessory arm (C1), a Tandem HarmonyAIR Fixed surgical light with accessory arm (C2), a HarmonyAIR Monitor Carrier (C3), and a HarmonyAIR A-Series Integrated Wall Control (K1). The electrical contractor is responsible for installing two 0.75-inch conduit runs from K1 to C2, with a 0.5-inch conduit for AC power supply (120V-240VAC, 7A-3A, 50/60Hz) to the wall control unit. Critical specifications include conduit runs not exceeding 25 feet, 1-inch conduit for DC supply light head cables on long runs, and floor-level tolerance of ±1/16 inch per 39 inches, with ceiling height of 106 5/16 inches. A revision note dated 09 September 2022 indicates K1 relocation, and the drawing is based on Philips-Azurion 7 C20 FlexArm 4300mm imaging equipment. The document includes a disclaimer that STERIS ceiling equipment locations are preliminary and must be reviewed against current imaging vendor drawings, and that indicated service clearances do not meet minimum safety clearances recommended in STERIS equipment drawings.

View the file

Other files

File Type Posted
36C26126R0045 Amendment 0004.pdf PDF
Wage Determination CA20260018 dtd 7 14 2026.pdf PDF
2B04 Shielding Design X Ray Room.pdf PDF
2A25B Shielding Design RF Room.pdf PDF
2B02 Shielding Design X Ray Room.pdf PDF
2A12B Shielding Design Cath Lab.pdf PDF
2A02B Shielding Design X Ray Room.pdf PDF
RFI Responses Part II 36C26126R0045.pdf PDF
RFI Responses 36C26126R0045 2.pdf PDF
QC-072_1.PDF PDF
X1_ACM_1.PDF PDF
Q2022-304 VA Fresno Cath Lab rev03.pdf PDF
VA FRESNO CATH LAB_R-RAD3_26_E_PF1_03-18-2026.pdf PDF
VA FRESNO CATH LAB_R-RAD2_26_E_PF1_03-18-2026.pdf PDF
P300Datasheet.pdf PDF
FS with stops.pdf PDF
Partial Brand Name JA Cath Lab_RD.pdf PDF
Addendum 0002 - Specifications Added.pdf PDF
C3-Monitor Carrier.PDF PDF
Shielding Design X-Ray Room 2A02B.pdf PDF
SEC 21793 VA FRESNO RF 3 12.22.2021 ELECT DWGS 100 CD Binder1.pdf PDF
Privacy-Curtains-And-Track-Curtain-Accessories-Product-Sheet-PS-CCACH.pdf PDF
C2-EMS.PDF PDF
Addendum 0002 - Drawings Added.pdf PDF
C3 Montr Carrier.PDF PDF
Shielding Design RF Room 2A25B.pdf PDF
VHA ICRA-1.5 (July 2 2026).pdf PDF
Privacy-Curtains-And-Track-Standard-Track-Product-Sheet-PS-CC6062.pdf PDF
C1-Anesthesia.PDF PDF
Shielding Design X-Ray Room 2B02.pdf PDF
Shielding Design X-Ray Room 2B04.pdf PDF
VA FRESNO CATH LAB_R-RAD1_26_E_PF1_03-17-2026.pdf PDF
USA_Data_Sheets Flow Splitter 651 06.pdf PDF
GE-Logiq-E10-Brochure-1.pdf PDF
Fresno VA FA R3 N-WES250405_Rev00 08.25.25.pdf PDF
11028.pdf PDF
EVOLUTION TYPICAL_WS2 RAIL_TBL2_08-14-2023.pdf PDF
ELECTRICAL_EQUIPMENT_DRAWING_-_HARMONYAIR_SURGICAL_LIGHTING_SYSTEM__A-SERIES.pdf PDF
36C26126R0045 Amendment 0002.pdf PDF
570-21-220 Sign in Sheet - Site Visit 6_8_26.pdf PDF
36C26126R0045 0001.docx DOCX document
8. WD CA20260018 dtd 5 18 2026.pdf PDF
5. Price Breakdown Worksheet.docx DOCX document
3. 570-21-220 Drawings.pdf PDF
36C26126R0045_2.docx DOCX document
7. Factor 2 Past Performance Questionnaire.docx DOCX document
6. Pre Proposal Site Visit Flyer 36C26126R0045.pdf PDF
4. Requests for Information RFI_ Form.docx DOCX document
2. 570-21-220 - Basis of Design.pdf PDF
1. 570-21-220 - Specifications.pdf PDF
Show 40 more

50 files listed here. See the rest on GovTribe.

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Le ve l o f f lo or

1 16

2m m ) p er

9" (1 m ) w ith in th is a re a.

Ta bl e Sa fe ty

Zo ne

M ot or iz ed M ov em en t L in e

C1

C2

C3

41316"

1 8

90" 78"

C1 C3

2"-33"

78" 90"

90" C2

16"

5 1

C E I L I N G

H E I G H T

5 1

C E I L I N G

H E I G H T

K1

EQUIPMENT LIST

TAG QTY DESCRIPTION

C1 1 Tandem HarmonyAIR Adj Arm Spring - A Series-Anesthesia, Accessory Arm(s)

C2 1 Tandem HarmonyAIR Fixed - A Series-Equipment, Surgical Light(s) & Accessory Arm(s)

C3 1 HarmonyAIR Monitor Carrier-Monitor Carrier

K1 1 HarmonyAIR A-Series Integrated Wall Control

A SERIES LIGHT CONDUIT SCHEDULE

FROM - TO QUANTITY SIZE RESPONSIBILITY

K1 - C2 2 .75" ELECTRICAL CONTRACTOR

LIGHT CONDUIT SCHEDULE NOTES

1. .5" CONDUIT FOR AC POWER TO K1; 3 COND. 10AWG WIRE

INPUT AC SUPPLY - 120V-240VAC, 7A-3A, 50/60Hz

2. LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET

3. CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT

ALSO REFERENCED WITHIN THE EQUIPMENT DRAWINGS FOR

THE SPECIFIED SURGICAL LIGHT WALL CONTROL SYSTEM

4. DC SUPPLY LIGHTHEAD CABLES REQUIRE 1" CONDUIT WHEN

DOUBLE RUNS ARE REQUIRED (LONG RUNS)

5. REFER TO APPROPRIATE EQUIPMENT SPECIFICATIONS TO

ENSURE THAT THE MOST CURRENT & DETAILED INFORMATION

OF THIS LIGHT SYSTEM IS UTILIZED

MOUNT C2 & C3

LIMITED UPPER ARM

ROTATIONS AT WALL

SCALE: 1/4" = 1'-0"

PROPOSED EQUIPMENT LAYOUT1

SCALE: 1/4" = 1'-0"

FOOT END EQUIPMENT ELEVATION2

SCALE: 1/4" = 1'-0"

FOOT END EQUIPMENT ELEVATION3

Clea / Table

AD

Ta bl e Fl oo r P la te

N1N13 N2

N1

10'-11"

'-9

Le ve l o f f lo or

1 16

2m m pe r 3

9" (1 m ) w ith in th is a re a.

x2 N

1x2

REVISION HISTORY

REVISION # DESCRIPTION PLANNER/DATE

1 UPDATED VENDOR CAD BACKGROUND; RELOCATED K1 JTP/09SEP2022

* THE SERVICE CLEARANCES INDICATED DO NOT MEET THE MINIMUM SAFETY CLEARANCES RECOMMENDED IN THE EQUIPMENT DRAWINGS PROVIDED BY STERIS CORPORATION

Planning & Design Group

STERIS

5960 Heisley Road

Mentor, OH 44060 USA 800-548-4873

Dr

Ck

Date

Scale

H- C

STERIS drawings do not supersede the drawings from the architect of record.

ALL STERIS CEILING EQUIPMENT LOCATIONS ARE

PRELIMINARY AND MUST BE REVIEWED AGAINST MOST

CURRENT IMAGING VENDOR DRAWINGS. THIS

DRAWING BASED ON:

DATE REVISION # DESCRIPTION

F 08/16/2022 PHILIPS - AZURION 7 C20 FLEXARM 4300MM

-N-WES190300FF

1/4" = 1'-0"

12AUG2022

JTP

6R1

PARTIAL FLOOR PLAN

CATH LAB

PROPOSED EQUIPMENT LAYOUT

VA CENTRAL CALIFORNIA

HEALTH CARE SYSTEM

FRESNO, CALIFORNIA

Sheets and Views
LAYOUT

File details come from the government source that posted it. Updated .