ELECTRICAL_EQUIPMENT_DRAWING_-_HARMONYAIR_SURGICAL_LIGHTING_SYSTEM__A-SERIES.pdf

PDF 824 KB Posted

Attached to
Z2DB--Renovate Imaging and Catheterization Lab 570 21 220 Fresno VA Medical Center Fresno
Solicitation number
36C26126R0045
Issued by

About this file

This is an electrical equipment drawing for the HarmonyAIR Surgical Lighting System A-Series manufactured by STERIS Corporation. The document provides comprehensive installation specifications and technical requirements for integrating the surgical lighting system into healthcare facilities.

The system consists of an Integrated Wall Control (IWC) with integrated power supplies, ceiling-mounted components, lightheads, and optional ambient lighting and camera equipment. Maximum configuration allows up to four lightheads, two optional lighthandle HD cameras, two optional ambient lights, and two ceiling plates (primary and secondary mounts). The IWC rough-in box must be mounted 60 inches minimum above finished floor to comply with NFPA requirements for flammable anesthetics, in a 10-inch square opening with 3.5-inch to 3.875-inch depth. AC power input is 120-240VAC, 7-3A, 50/60Hz with 40-watts nominal lighthead consumption and 10-watts for LED ambient light. The system requires redundant grounding methods with earth ground connections to building steel, separate AC and DC junction boxes within 18 inches of the ceiling plate, and conduit runs per specified standards. Standard DC power cables are supplied in 50-foot or 75-foot lengths; longer runs require customer-supplied cable sized per provided tables (minimum 12 AWG for 6-75 feet, scaling to 4 AWG for 250-328 feet). Two configurations are detailed: Type 1 with single ceiling plate supporting up to three lightheads, and Type 2 with dual ceiling plates supporting up to four lightheads. All installation must comply with national and local electrical codes with allowable earth leakage current of 5 milliamps maximum.

View the file

Other files

File Type Posted
36C26126R0045 Amendment 0004.pdf PDF
Wage Determination CA20260018 dtd 7 14 2026.pdf PDF
RFI Responses Part II 36C26126R0045.pdf PDF
2A02B Shielding Design X Ray Room.pdf PDF
2A12B Shielding Design Cath Lab.pdf PDF
2B04 Shielding Design X Ray Room.pdf PDF
2A25B Shielding Design RF Room.pdf PDF
36C26126R0045 Amendment 0003.pdf PDF
2B02 Shielding Design X Ray Room.pdf PDF
RFI Responses 36C26126R0045 2.pdf PDF
X1_ACM_1.PDF PDF
QC-072_1.PDF PDF
Q2022-304 VA Fresno Cath Lab rev03.pdf PDF
0654C6R1.pdf PDF
SEC 21793 VA FRESNO RF 3 12.22.2021 ELECT DWGS 100 CD Binder1.pdf PDF
C3 Montr Carrier.PDF PDF
Shielding Design X-Ray Room 2A02B.pdf PDF
C2-EMS.PDF PDF
11028.pdf PDF
Fresno VA FA R3 N-WES250405_Rev00 08.25.25.pdf PDF
VA FRESNO CATH LAB_R-RAD3_26_E_PF1_03-18-2026.pdf PDF
EVOLUTION TYPICAL_WS2 RAIL_TBL2_08-14-2023.pdf PDF
Privacy-Curtains-And-Track-Curtain-Accessories-Product-Sheet-PS-CCACH.pdf PDF
C3-Monitor Carrier.PDF PDF
USA_Data_Sheets Flow Splitter 651 06.pdf PDF
GE-Logiq-E10-Brochure-1.pdf PDF
VA FRESNO CATH LAB_R-RAD1_26_E_PF1_03-17-2026.pdf PDF
P300Datasheet.pdf PDF
FS with stops.pdf PDF
VA FRESNO CATH LAB_R-RAD2_26_E_PF1_03-18-2026.pdf PDF
Partial Brand Name JA Cath Lab_RD.pdf PDF
Addendum 0002 - Specifications Added.pdf PDF
Addendum 0002 - Drawings Added.pdf PDF
Shielding Design X-Ray Room 2B04.pdf PDF
36C26126R0045 Amendment 0002.pdf PDF
Shielding Design X-Ray Room 2B02.pdf PDF
Shielding Design RF Room 2A25B.pdf PDF
Privacy-Curtains-And-Track-Standard-Track-Product-Sheet-PS-CC6062.pdf PDF
C1-Anesthesia.PDF PDF
VHA ICRA-1.5 (July 2 2026).pdf PDF
570-21-220 Sign in Sheet - Site Visit 6_8_26.pdf PDF
36C26126R0045 0001.docx DOCX document
7. Factor 2 Past Performance Questionnaire.docx DOCX document
6. Pre Proposal Site Visit Flyer 36C26126R0045.pdf PDF
3. 570-21-220 Drawings.pdf PDF
4. Requests for Information RFI_ Form.docx DOCX document
5. Price Breakdown Worksheet.docx DOCX document
8. WD CA20260018 dtd 5 18 2026.pdf PDF
36C26126R0045_2.docx DOCX document
1. 570-21-220 - Specifications.pdf PDF
Show 40 more

Latest 50 of 51 listed here. See the rest on GovTribe.

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

1 TABLE OF CONTENTS

3 SYSTEM GROUNDING NOTES AND CONDUIT REQUIREMENTS NOTES

4 CONFIGURATIONS

5 INTERGRATED WALL CONTROL (IWC) INFORMATION

6 ROUGH-IN BOX INFORMATION - IWC

TABLE OF CONTENTS

SHEET NO CONTENTS

7 IWC BACK VIEW

8 HARMONYAIR SURGICAL LIGHTING SYSTEM A-SERIES CABLES

9 CONDUIT REQUIREMENTS

12 COMPONENTS MOUNTED ON THE PRIMARY CEILING PLATE

11 TYPE 2 CONFIGURATION

2 GENERAL NOTES

STERIS Corporation

Mentor, OH

APPLICABLE TO EQUIPMENT DRAWINGS

ALL DIMENSIONS ARE IN INCHES

ALSO REFER TO GENERAL NOTES

R E

V

D W

G

N O

C

A D

D W

G

ACCOUNT NAME

ACCOUNT LOCATION

FILE NAME

PRINTED DATE

SHEET

A

ELECTRICAL EQUIPMENT DRAWING

HarmonyAIR SURGICAL LIGHTING SYSTEM, A-SERIES

10 TYPE 1 CONFIGURATION

Released

GENERAL NOTES

1. IWC IS THE "INTEGRATED WALL CONTROL WITH INTEGRATED POWER SUPPLIES", ALSO KNOWN AS THE CONTROL CENTER.

2. RECOMMENDED HEIGHT OF ROUGH-IN BOX 60" MIN. ABOVE FINISHED FLOOR TO COMPLY WITH NFPA REQUIREMENT FOR FLAMMABLE

ANESTHETICS. CONTROL BOX MAY BE MOUNTED BELOW 60" IF NO FLAMMABLE ANESTHETICS ARE UTILIZED.

3. OPENING IN WALL TO RECEIVE ROUGH-IN BOX MUST BE 10" SQUARE AND 3 1/2" MIN. TO 3 7/8" MAX. 3 5/8 NOMINAL DEPTH. DEPTH IS MEASURED

FROM FRONT SURFACE OF WALL. (254mm SQ. & 89mm TO 99mm, 92mm NOMINAL DEPTH) ROUGH-IN BOX MUST BE INSTALLED LEVEL, WITH

OUTSIDE SURFACE FLUSH WITH FINISHED WALL.

4. ROUGH-IN BOX MUST BE SECURELY MOUNTED WITHIN WALL. ENTIRE ROUGH-IN BOX AND IWC WEIGHS APPROXIMATELY 10 LBS (4.5 KGS).

1/4-20 CORNER HOLES ARE NOT FOR MOUNTING ROUGH-IN BOX TO THE WALL. THEY ARE FOR MOUNTING THE IWC TO THE ROUGH-IN BOX.

5. LABEL THE ENDS OF ALL WIRES FOR FUTURE IDENTIFICATION.

6. GROUNDING IS CRITICAL FOR SAFETY & EFFICACY OF THE SYSTEM. SEE SYSTEM GROUNDING ON PAGE 3.

7. EQUIPMENT MUST BE INSTALLED PER NATIONAL AND LOCAL CODES.

8. FOR RECOMMENDED CONDUIT AND WIRE LENGTHS, REFER TO IWC CABLE AND CONDUIT REQUIREMENTS CHARTS IN THIS DOCUMENT.

9. CEILING ACCESS PANEL HIGHLY RECOMMENDED, AND LOCATED WITHIN ARMS REACH OF THE HUB LOCATION FOR SERVICE ACCESSIBILITY.

10. JUNCTION BOX (NOT SUPPLIED BY STERIS) MUST BE MOUNTED WITHIN 18" OF CEILING PLATE CENTER FOR AC & DC CONNECTIONS. SEPARATE

JUNCTION BOXES ARE REQUIRED FOR AC & DC CONNECTIONS.

11. THIS DOCUMENT IS FOR THE HARMONYAIR SURGICAL LIGHTING SYSTEM A-SERIES MOUNTING ONLY. THIS DOCUMENT MUST BE COMBINED

WITH INSTALLATION INSTRUCTIONS AND MECHANICAL EQUIPMENT DRAWINGS IN ORDER TO UNDERSTAND ALL INSTALLATION REQUIREMENTS.

12. ALLOWABLE EARTH LEAKAGE CURRENT IS 5 MILLIAMPS.

13. STERIS ASSUMES NO RESPONSIBILITY FOR CHANGES MADE THROUGH FAILURE TO OBSERVE THESE INSTRUCTIONS.

14. IT IS RECOMMENDED TO HAVE A DEDICATED BREAKER AS WELL AS A DEDICATED GROUND FOR THE IWC & CEILING PLATE.

15. THE NOMINAL POWER CONSUMPTION IS LISTED FOR THE HARMONYAIR SURGICAL LIGHTING SYSTEM "A-SERIES" FAMILY OF LIGHTING.

WATTS BTU/HR

LIGHTHEAD 40* 136*

LED AMBIENT LIGHT 10 34

THE INCOMING POWER RATING FOR A FOUR LIGHTHEAD SYSTEM INCLUDING AN AMBIENT LIGHT IS 120-240 VAC, 7-3A, 50/60 Hz.

16. ALL MEASUREMENTS ARE IN INCHES UNLESS OTHERWISE SPECIFIED.

17. COMPLETE ROUGH-IN BOX 10101691 FOR IWC [REF. LAS0034] FURNISHED BY STERIS.

18. A 3/4" CONDUIT CAN FIT UP TO 2 STERIS SUPPLIED DC POWER CABLES (12 AWG, 2 CONDUCTOR). IF LARGER AWG CABLES ARE NEEDED DUE TO

RUNS LONGER THAN 75ft. THEN A LARGER CONDUIT MAY BE REQUIRED (CUSTOMER DETERMINED AND SUPPLIED), SEE TABLE 1 ON PAGE 3.

*LIGHTHEAD RANGES FROM 20-140 WATTS (68-478 BTU/HR) BECAUSE OF AUTO-INTENSITY ADJUSTMENTS

STERIS Corporation

Mentor, OH

APPLICABLE TO EQUIPMENT DRAWINGS

V

D W

G

N O

C

A D

D W

G

ACCOUNT NAME

ACCOUNT LOCATION

FILE NAME

PRINTED DATE

SHEET

A

ELECTRICAL EQUIPMENT DRAWING

HarmonyAIR SURGICAL LIGHTING SYSTEM A-SERIES

SYSTEM GROUNDING NOTES

1. NEC GROUNDING BASIC PRINCIPLES -- NATIONAL ELECTRICAL CODE SPECIFICALLY STATES THAT SYSTEMS WILL BE GROUNDED FOR THREE

REASONS:

1.) TO PROTECT FROM LIGHTNING AND OVER-VOLTAGE BY ACCIDENTAL CONTACT

2.) TO REFERENCE THE VOLTAGE AND STABILIZE IT WITH RESPECT TO GROUND (EMI/EMC)

3.) TO FACILITATE FUSING AND GROUND FAULT DEVICES

2. GROUNDING METHOD - A REDUNDANT GROUND METHOD IS USED FOR IWC IN ORDER TO MAKE SURE THAT TOUCH LEAKAGE CURRENT IS

MAINTAINED BELOW 500 MICRO-AMPS. AN EARTH GROUND SHALL BE RUN WITH AC MAINS CONNECTION TO THE ROUGH-IN BOX. A GROUND

WIRE OR CONDUIT CONNECTION SHALL BE RUN FROM ROUGH-IN BOX TO THE CEILING PLATE. AN EARTH GROUND SHALL BE BROUGHT FROM

A SEPARATE GROUND SOURCE TO THE CEILING PLATE.

3. THE GROUNDING POINT FOR ALL DEVICES SHOULD BE THE BUILDING STEEL. ALL HOSPITALS ARE REQUIRED TO HAVE A GRID GROUNDING

SYSTEM WHICH SHOULD INCLUDE THE BUILDING STEEL. THIS DOES NOT NECESSARILY MEAN THAT "ALL" STEEL PARTS ARE INCLUDED IN THE

GRID AND HAVE BONDED GROUND STRAPS. THE OTHER CONSIDERATION IS THAT ALL GROUND STRAPS ARE INTACT. A RESISTANCE CHECK

FROM SYSTEM ELECTRICAL GROUND TO A RELIABLE BUILDING GROUND POINT SHOULD SHOW CONTINUITY. IF CONTINUITY DOES NOT EXIST

FROM THE HUB TO THE BUILDING GROUND NOTIFY THE CUSTOMER SITE MAINTENANCE DEPARTMENT TO CORRECT THE PROBLEM.

5. STERIS "HARMONYAIR SURGICAL LIGHTING SYSTEM A-SERIES" GROUNDING SCHEME IS SHOWN ON PAGES 10 TO 11. ADDITIONAL GROUNDING

REQUIREMENTS ARE LISTED IN THE INSTALLATION INSTRUCTIONS 11007640.

CONDUIT REQUIREMENTS NOTES

1. (AC1) SITE AC POWER FOR IWC.

2. FOR DC POWER, RUNS OTHER THAN THE SUPPLIED (50ft. OR 75ft. OPTIONAL) WILL REQUIRE CUSTOMER SUPPLIED CABLE, SEE TABLE 1.

3. HD VIDEO FIBER OUTPUT CAN GO TO SITE SPECIFIC A/V EQUIPMENT, OR DIRECTLY TO MONITOR (CUSTOMER DETERMINED).

4. ALL CONDUIT, AC SUPPLY LINES, AND GROUND WIRES ARE TO BE SUPPLIED BY CUSTOMER ON SITE.

STERIS Corporation

Mentor, OH

APPLICABLE TO EQUIPMENT DRAWINGS

V

D W

G

N O

C

A D

D W

G

ACCOUNT NAME

ACCOUNT LOCATION

FILE NAME

PRINTED DATE

SHEET

A

ELECTRICAL EQUIPMENT DRAWING

HarmonyAIR SURGICAL LIGHTING SYSTEM A-SERIES

RUN LENGTH FROM

IWC TO HUB

LENGTH IN

75-100

250-328 150-250 100-150

6-75

FEET

LENGTH IN

22-30

75-100

45-75 30-45

2-22

METERS

ACCEPTABLE

4* 6* 8*

AWG

MIN.

(LONG RUNS) CABLE CHART

DC POWER

TABLE 1:

* FOR 4-8 AWG WIRE BOTH ENDS MUST BE SPLICED TO A SHORT LENGTH OF 10 AWG

WIRE IN ORDER TO ACCOMMODATE THE TERMINAL BLOCK SIZE RESTRICTION IN THE

IWC AND HUB BREAKOUT BOARD.

STERIS COMPONENT

MAXIMUM

NUMBER

PER SYSTEM COMMENTS

LIGHTHEADS

AMBIENT LIGHT (OPTIONAL)

MAXIMUM 3 LIGHTHEADS PER HUB BOARD

- REFER TO MECHANICAL EQUIPMENT DRAWINGS FOR MOUNTING REQUIREMENTS OF MECHANICAL COMPONENTS.

- REFER TO SHEETS 10-12 FOR SPECIFIC CONFIGURATIONS. NOT ALL POSSIBLE CONFIGURATIONS ARE SHOWN.

ONE MOUNTED PER CENTRAL SPINDLE.

MAXIMUM CONFIGURATION PER IWC

LIGHTHANDLE HD CAMERA (OPTIONAL)

2 WILL ONLY FUNCTION IF INSERTED INTO HD

COMPATIBLE LIGHTHEAD.

STERIS Corporation

Mentor, OH

APPLICABLE TO EQUIPMENT DRAWINGS

V

D

W G

N

O

C A

D D

W G

ACCOUNT NAME

ACCOUNT LOCATION

FILE NAME

PRINTED DATE

SHEET

A

ELECTRICAL EQUIPMENT DRAWING

HarmonyAIR SURGICAL LIGHTING SYSTEM A-SERIES

CONFIGURATION NOTES

**(SEE FIGURES 1, AND 2 ON SHEETS 10-11)**

GENERAL NOTES:

1. WIRING ENTERS THROUGH CEILING PLATE AREA. TERMINATE ELECTRICAL METALLIC TUBING (EMT) INTO LOW VOLTAGE JUNCTION BOX

WITHIN 18" OF CEILING PLATE AND GROUND TO CEILING PLATE.

2. 1" EMT MAY BE USED IF SYSTEM CONFIGURATION PERMITS.

DRAWING NOTES:

3. SIGNAL CONDUIT SIZE FROM A/V EQUIPMENT IS DETERMINED AND SUPPLIED BY A/V VENDOR.

4. ALL CONDUIT RUNS AND AC SUPPLY LINES ARE TO BE SUPPLIED BY CUSTOMER ON SITE.

5. FIBER CAN BE RUN THRU CONDUIT, LC FIBER CONNECTOR AVAILABLE AT CEILING PLATE.

6. THE DAISY CHAIN HARNESS INCLUDES AN INTEGRATED FLEXIBLE METAL CONDUIT THAT MUST BE CONNECTED TO EITHER A CUSTOMER

SUPPLIED GROUNDED JUNCTION BOX AT BOTH CEILING PLATES OR OPTIONAL HUB BREAKOUT BOARD ENCLOSURE STERIS P/N 10103591 (1

PER HUB).

CEILING PLATE (SPINDLE) 2 DENOTED AS (PRIMARY AND SECONDARY MOUNT)

STERIS Corporation

Mentor, OH

APPLICABLE TO EQUIPMENT DRAWINGS

V

D

W G

N

O

C A

D D

W G

ACCOUNT NAME

ACCOUNT LOCATION

FILE NAME

PRINTED DATE

SHEET

A

ELECTRICAL EQUIPMENT DRAWING

HarmonyAIR SURGICAL LIGHTING SYSTEM A-SERIES

WALL CAVITY ABOVE AND BELOW IWC MUST BE KEPT CLEAR OF ANY OBSTRUCTIONS TO PREVENT INTERFERENCE WITH CONDUITS.

IWC WITHOUT BEZEL

IWC INFORMATION

16.24

13.11

16.5

13.36

2.9

BEZEL MOUNTING SCREWS

BEZEL MOUNTING SCREWS

IWC WITH

BEZEL

USER INTERFACE

(POE)

SERVICE (ETHERNET)

Mentor, OH

APPLICABLE TO EQUIPMENT DRAWINGS

V

D

W G

N

O

C A

D D

W G

ACCOUNT NAME

ACCOUNT LOCATION

FILE NAME

PRINTED DATE

SHEET

A

ELECTRICAL EQUIPMENT DRAWING

HarmonyAIR SURGICAL LIGHTING SYSTEM A-SERIES

60" (1524mm) MIN.

TO FLOOR

SEE GENERAL NOTE 2,

SHEET 2

9.75

9.50

MOUNTING HOLES

FOR ROUGH-IN BOX

TYP BOTH SIDES

3 WIRE INPUT AC SUPPLY -

2 POWER, 1 GROUND

(10 AWG GROUND WIRE)

MUST BEPROVIDED AT

THIS 1/2" [13mm] KNOCKOUT

120V-240VAC, 7-3A,

50/60Hz

3/4" (19mm) X 1" (25mm) KNOCKOUT.

FRONT VIEW

USE THE (2) AVAILABLE

3/4" CONDUITS ONE FOR

DC POWER AND ONE FOR

COMMUNICATION.

11.15

11.25

3.88 MAX

INTEGRATED WALL CONTROL (IWC) ROUGH-IN BOX INFORMATION

(WALL CONTROL NOT SHOWN)

FINISHED WALL

3X 1/2" (13mm) X 3/4" (19mm) KNOCKOUT.

2X 1/2" (13mm) KNOCKOUT.

3X 1/2" (13mm) KNOCKOUT.

2X 3/4" (19mm) X 1" (25mm) KNOCKOUT.

2X 1/2" (13mm) X 3/4" (19mm) KNOCKOUT.

NOTES:

- WALL CAVITY ABOVE AND BELOW IWC MUST BE KEPT CLEAR OF ANY OBSTRUCTIONS TO PREVENT INTERFERENCE WITH CONDUITS.

- LIGHTHEAD DC POWER SUPPLIES HAVE AC INPUT AUTORANGING FEATURE FOR 120V-240VAC

Mentor, OH

APPLICABLE TO EQUIPMENT DRAWINGS

V

D

W G

N

O

C A

D D

W G

ACCOUNT NAME

ACCOUNT LOCATION

FILE NAME

PRINTED DATE

SHEET

A

ELECTRICAL EQUIPMENT DRAWING

HarmonyAIR SURGICAL LIGHTING SYSTEM A-SERIES

BACK VIEW

IWC INFORMATION

HUB 1 (COMM)

HUB 2 (COMM)

ORI (ETHERNET)

HOSPITAL NETWORK

(ETHERNET)

ACT (FUTURE USE)

TO HUB 1

24 VDC -B

TO HUB 1

24 VDC -A

o f 1

S T

E R

IS

C o rp o ra tio n

M e n to r, O H

A P

P L

IC

A

B L

E T

O E

Q U

IP

M

E N

T D

R A

W

IN

G S

A L

L D

IM

E

N S

IO

N

S A

R E

IN

IN

C H

E S

A L

S O

R E

F E

R T

O G

E N

E R

A L

N O

T E

S

REV. DWG. NO.CAD DWG

A C

C O

U N

T N

A M

E

A C

C O

U N

T L

O C

A T

IO

N

F

IL

E N

A M

E

P R

IN

T

E D

D A

T E

S H

E E

T

11014539A

E L

E C

T R

IC

A

L E

Q U

IP

M

E N

T D

R A

W

IN

G

H a rm o n y A

IR

S

U R

G

IC

A L

L

IG

H T

IN

G

S Y

S T

E M

A -S

E R

IE

S

* NON STANDARD AND MUST BE PURCHASED SEPERATELY.

END-2END-1

HUB 1

50 FT.COMMUNICATION CABLE

2 PER

SYSTEM

75 FT.

QUANTITY

50 FT.

LENGTHDESCRIPTION

POWER CABLE, 12 AWG,

JACKETED, LONG RUN (PLENUM RATED)

POWER CABLE, 12 AWG, JACKETED,

2 CONDUCTOR

CARRY LIGHTHEAD

POWER FROM THE

POWER SUPPLY

TO THE HUB

PURPOSE

CONNECTIONS

LAS0029

KITEQUIPMENT #ACCESSORY *

LONG RUN

DBL POWER CABLE

PLENUM RATED

IWC

POWER CABLES

INCLUDED WITH:

PART #HOW ARE THE CABLES SUPPLIED?

10100130

11004734

10100131

CABLE

HARMONYAIR SURGICAL LIGHTING SYSTEM A SERIES CABLES

HUB 1 OR

HUB 2

(SPLIT MOUNT)

IWC

CARRY SIGNALS

BETWEEN IWC

AND THE HUB

1 PER SINGLE MOUNT

2 PER SPLIT MOUNT

COMMUNICATION CABLES

IWC

IWC

LAS0029

LONG RUN COMM KIT

PLENUM RATED

11004635 75 FT.

COMMUNICATION CABLE, LONG RUN

(PLENUM RATED)

11006248

11006248

11006243LAS0025

LAS0025 11006243

LAS0028DAISY CHAIN

POWER

11006245 10100268

14 AWG FMC CABLE W/

INTEGRATED HUB CONNECTORS

15 FT.

N/A

1 PER

SPLIT MOUNT

CARRY DC POWER

FROM HUB 1

TO HUB 2

HUB 2HUB 1

N/A

HARNESS, MASTER CONTROL TO

HUB BREAKOUT BOARD, RJ45

50 FT.LAS0028DAISY CHAIN

POWER

11006245 10100130

R e le a s e d

IWC

SEE SHEET 8

50 FT. (75 FT.)C2 3/4"

C1 50 FT. (75 FT.)3/4"

COMMUNICATION / CAT 5

COMMUNICATION / CAT 5

COMMUNICATION SIGNALS

1,4IWC

NOTESTO

IWC

SITE AC POWER

QTY FROM

NOT BY STERIS

PART NUMBER

JUNCTION BOX, AT PRIMARY HUB

STERIS

CONNECTIONS PAGE

3/4" DC POWER FOR LIGHTHEADS

SITE DETERMINED

SIZE STD (LONG RUN) *

1/2" AC POWER FOR IWC

CABLE DESCRIPTION

AC POWER

LABEL

AC1

DC POWER

DC1

CONDUIT CABLE

CONDUIT REQUIREMENTS (FIGURES 1,2)

2,4, (COMMUNICATION

CABLES)

PRIMARY HUB

SECONDARY HUB1 COMM / CAT 5 CABLE

50 FT.(75 FT.)

1 COMM / CAT 5 CABLE IWC

SEE SHEET 8

(COMMUNICATION

CABLES)

o f 1

S T

E R

IS

C o rp o ra tio n

M e n to r, O H

A P

P L

IC

A

B L

E T

O E

Q U

IP

M

E N

T D

R A

W

IN

G S

A L

L D

IM

E

N S

IO

N

S A

R E

IN

IN

C H

E S

A L

S O

R E

F E

R T

O G

E N

E R

A L

N O

T E

S

REV. DWG. NO.CAD DWG

A C

C O

U N

T N

A M

E

A C

C O

U N

T L

O C

A T

IO

N

F

IL

E N

A M

E

P R

IN

T

E D

D A

T E

S H

E E

T

11014539A

E L

E C

T R

IC

A

L E

Q U

IP

M

E N

T D

R A

W

IN

G

H a rm o n y A

IR

S

U R

G

IC

A L

L

IG

H T

IN

G

S Y

S T

E M

A -S

E R

IE

S

1 PER SYSTEM

A & B POWER CABLE KIT

SEE SHEET 8

(POWER CABLES)

(CONDUIT OPTIONAL)

(CONDUIT OPTIONAL)

* MAXIMUM LENGTH FOR DC CABLES OR CAT5 COMMUNICATION CABLES IS 100M, 328ft. (CUSTOMER SUPPLIED).

R e le a s e d

Mentor, OH

APPLICABLE TO EQUIPMENT DRAWINGS

V

D

W G

N

O

C A

D D

W G

ACCOUNT NAME

ACCOUNT LOCATION

FILE NAME

PRINTED DATE

SHEET

A

ELECTRICAL EQUIPMENT DRAWING

HarmonyAIR SURGICAL LIGHTING SYSTEM A-SERIES

IWC

AC1 1

2" AC1

4"

PRIMARY

CEILING PLATE

C1 DC1

HUB

BREAKOUT

BOARD

4"

NOTES:

1. GROUNDING OF CEILING PLATE - CEILING PLATE MUST BE GROUNDED TO THE SAME

POTENTIAL AS THE IWC. TO ACHIEVE THIS, CONNECT AT LEAST ONE JUNCTION BOX WITH

A GROUND WIRE, 12" (305 MM) OR LESS TO THE CEILING PLATE

CONDUIT AND ELECTRICAL INFORMATION

FIGURE 1

TYPE 1 CONFIGURATION

IWC WITH A SINGLE CEILING PLATE

THIS CONFIGURATION ALLOWS UP TO 3 LIGHTHEADS

PER CEILING PLATE

X/X"

POSSIBLE CABLE ALTERNATIVES.

OPTIONAL EQUIPMENT, CABLES AND CONDUIT.

OPTIONAL EQUIPMENT.

COMMUNICATION CABLES AND CONDUIT.

DC SUPPLY CABLES AND CONDUIT.

RECOMMENDED CONDUIT SIZE.

AC SUPPLY CABLES AND CONDUIT.

LEGEND

REFER TO CONFIGURATION NOTES PAGE 4 FOR

SPECIFIC INFORMATION.#

XXX REFER PAGE 9 FOR CONDUIT REQUIREMENTS

ACX

DCX

CX

EMT GROUNDED AT THIS END

VIDEO RACK

OR

MONITOR

HD VIDEO THROUGH FIBER 3 5

Mentor, OH

APPLICABLE TO EQUIPMENT DRAWINGS

V

D

W G

N

O

C A

D D

W G

ACCOUNT NAME

ACCOUNT LOCATION

FILE NAME

PRINTED DATE

SHEET

A

ELECTRICAL EQUIPMENT DRAWING

HarmonyAIR SURGICAL LIGHTING SYSTEM A-SERIES

IWC AC1 1

2" AC1

4"

PRIMARY

CEILING PLATE

C1

DC1

HUB

BREAKOUT

BOARD

VIDEO RACK

OR

MONITOR

4"

SECONDARY

CEILING PLATE

HUB

BREAKOUT

BOARD

VIDEO RACK

OR

MONITOR

C2

CONDUIT AND ELECTRICAL INFORMATION

HD VIDEO FIBER

PRIMARY MOUNT SECONDARY MOUNT

FIGURE 2

TYPE 2 CONFIGURATION

IWC WITH 2 CEILING PLATES

THIS CONFIGURATION ALLOWS FOR

UP TO 4 LIGHTHEADS

X/X"

POSSIBLE CABLE ALTERNATIVES.

OPTIONAL EQUIPMENT, CABLES AND CONDUIT.

OPTIONAL EQUIPMENT.

COMMUNICATION CABLES AND CONDUIT.

DC SUPPLY CABLES AND CONDUIT.

RECOMMENDED CONDUIT SIZE.

AC SUPPLY CABLES AND CONDUIT.

LEGEND

REFER TO CONFIGURATION NOTES PAGE 4 FOR

SPECIFIC INFORMATION.#

XXX REFER PAGE 9 FOR CONDUIT REQUIREMENTS

ACX

DCX

CX

EMT GROUNDED AT THIS END

HUB DAISY CHAIN HARNESS P/N 10100268

NOTES:

1. GROUNDING OF CEILING PLATE - CEILING PLATE MUST BE GROUNDED TO THE SAME

POTENTIAL AS THE IWC. TO ACHIEVE THIS, CONNECT AT LEAST ONE JUNCTION BOX WITH

A GROUND WIRE, 12" (305 MM) OR LESS TO THE CEILING PLATE

4"

CEILING PLATE

Ø17.95" (456mm)

HUB BREAKOUT BOARD

CLAMPING BRACKET

ASSEMBLY

(STERIS P/N 1013426)

MAIN SPINDLE LOCATION

AUXILIARY SPINDLE

(4 POSSIBLE LOCATIONS)

MONITOR

CONDUIT

MOUNTING

LOCATION

EARTH GROUNDING

LOCATION (GROUND SCREW

(PROVIDED WITH HUB BREAKOUT

ASSEMBLY)

STERIS Corporation

Mentor, OH

APPLICABLE TO EQUIPMENT DRAWINGS

V

D

W G

N

O

C A

D D

W G

ACCOUNT NAME

ACCOUNT LOCATION

FILE NAME

PRINTED DATE

SHEET

A

ELECTRICAL EQUIPMENT DRAWING

HarmonyAIR SURGICAL LIGHTING SYSTEM A-SERIES

COMPONENTS MOUNTED ON CEILING PLATE

(ACTUAL COMPONENTS INSTALLED DEPENDS ON CONFIGURATION)

OPTIONAL

ENCLOSURE

(STERIS P/N 10103591)

OPTIONAL ENCLOSURE FOR HUB BREAKOUT BOARD

(BEAM CLAMP MOUNT ALLOWS MULTIPLE MOUNTING ORIENTATIONS

ON EITHER CEILING PLATE PLATE OR STRUCTURAL PLATE)

TOP VIEW

HUB BREAKOUT BOARD CLAMPING

BRACKET (CAN BE MOUNTED

ANYWHERE AROUND THE

CEILING PLATE PERIMETER)

File details come from the government source that posted it. Updated .