ELECTRICAL_EQUIPMENT_DRAWING_-_HARMONYAIR_SURGICAL_LIGHTING_SYSTEM__A-SERIES.pdf
PDF 824 KB Posted
- Attached to
- Z2DB--Renovate Imaging and Catheterization Lab 570 21 220 Fresno VA Medical Center Fresno
- Solicitation number
- 36C26126R0045
About this file
This is an electrical equipment drawing for the HarmonyAIR Surgical Lighting System A-Series manufactured by STERIS Corporation. The document provides comprehensive installation specifications and technical requirements for integrating the surgical lighting system into healthcare facilities.
The system consists of an Integrated Wall Control (IWC) with integrated power supplies, ceiling-mounted components, lightheads, and optional ambient lighting and camera equipment. Maximum configuration allows up to four lightheads, two optional lighthandle HD cameras, two optional ambient lights, and two ceiling plates (primary and secondary mounts). The IWC rough-in box must be mounted 60 inches minimum above finished floor to comply with NFPA requirements for flammable anesthetics, in a 10-inch square opening with 3.5-inch to 3.875-inch depth. AC power input is 120-240VAC, 7-3A, 50/60Hz with 40-watts nominal lighthead consumption and 10-watts for LED ambient light. The system requires redundant grounding methods with earth ground connections to building steel, separate AC and DC junction boxes within 18 inches of the ceiling plate, and conduit runs per specified standards. Standard DC power cables are supplied in 50-foot or 75-foot lengths; longer runs require customer-supplied cable sized per provided tables (minimum 12 AWG for 6-75 feet, scaling to 4 AWG for 250-328 feet). Two configurations are detailed: Type 1 with single ceiling plate supporting up to three lightheads, and Type 2 with dual ceiling plates supporting up to four lightheads. All installation must comply with national and local electrical codes with allowable earth leakage current of 5 milliamps maximum.
View the file
Other files
Show 40 more
Latest 50 of 51 listed here. See the rest on GovTribe.
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
1 TABLE OF CONTENTS
3 SYSTEM GROUNDING NOTES AND CONDUIT REQUIREMENTS NOTES
4 CONFIGURATIONS
5 INTERGRATED WALL CONTROL (IWC) INFORMATION
6 ROUGH-IN BOX INFORMATION - IWC
TABLE OF CONTENTS
SHEET NO CONTENTS
7 IWC BACK VIEW
8 HARMONYAIR SURGICAL LIGHTING SYSTEM A-SERIES CABLES
9 CONDUIT REQUIREMENTS
12 COMPONENTS MOUNTED ON THE PRIMARY CEILING PLATE
11 TYPE 2 CONFIGURATION
2 GENERAL NOTES
STERIS Corporation
Mentor, OH
APPLICABLE TO EQUIPMENT DRAWINGS
ALL DIMENSIONS ARE IN INCHES
ALSO REFER TO GENERAL NOTES
R E
V
D W
G
N O
C
A D
D W
G
ACCOUNT NAME
ACCOUNT LOCATION
FILE NAME
PRINTED DATE
SHEET
A
ELECTRICAL EQUIPMENT DRAWING
HarmonyAIR SURGICAL LIGHTING SYSTEM, A-SERIES
10 TYPE 1 CONFIGURATION
Released
GENERAL NOTES
1. IWC IS THE "INTEGRATED WALL CONTROL WITH INTEGRATED POWER SUPPLIES", ALSO KNOWN AS THE CONTROL CENTER.
2. RECOMMENDED HEIGHT OF ROUGH-IN BOX 60" MIN. ABOVE FINISHED FLOOR TO COMPLY WITH NFPA REQUIREMENT FOR FLAMMABLE
ANESTHETICS. CONTROL BOX MAY BE MOUNTED BELOW 60" IF NO FLAMMABLE ANESTHETICS ARE UTILIZED.
3. OPENING IN WALL TO RECEIVE ROUGH-IN BOX MUST BE 10" SQUARE AND 3 1/2" MIN. TO 3 7/8" MAX. 3 5/8 NOMINAL DEPTH. DEPTH IS MEASURED
FROM FRONT SURFACE OF WALL. (254mm SQ. & 89mm TO 99mm, 92mm NOMINAL DEPTH) ROUGH-IN BOX MUST BE INSTALLED LEVEL, WITH
OUTSIDE SURFACE FLUSH WITH FINISHED WALL.
4. ROUGH-IN BOX MUST BE SECURELY MOUNTED WITHIN WALL. ENTIRE ROUGH-IN BOX AND IWC WEIGHS APPROXIMATELY 10 LBS (4.5 KGS).
1/4-20 CORNER HOLES ARE NOT FOR MOUNTING ROUGH-IN BOX TO THE WALL. THEY ARE FOR MOUNTING THE IWC TO THE ROUGH-IN BOX.
5. LABEL THE ENDS OF ALL WIRES FOR FUTURE IDENTIFICATION.
6. GROUNDING IS CRITICAL FOR SAFETY & EFFICACY OF THE SYSTEM. SEE SYSTEM GROUNDING ON PAGE 3.
7. EQUIPMENT MUST BE INSTALLED PER NATIONAL AND LOCAL CODES.
8. FOR RECOMMENDED CONDUIT AND WIRE LENGTHS, REFER TO IWC CABLE AND CONDUIT REQUIREMENTS CHARTS IN THIS DOCUMENT.
9. CEILING ACCESS PANEL HIGHLY RECOMMENDED, AND LOCATED WITHIN ARMS REACH OF THE HUB LOCATION FOR SERVICE ACCESSIBILITY.
10. JUNCTION BOX (NOT SUPPLIED BY STERIS) MUST BE MOUNTED WITHIN 18" OF CEILING PLATE CENTER FOR AC & DC CONNECTIONS. SEPARATE
JUNCTION BOXES ARE REQUIRED FOR AC & DC CONNECTIONS.
11. THIS DOCUMENT IS FOR THE HARMONYAIR SURGICAL LIGHTING SYSTEM A-SERIES MOUNTING ONLY. THIS DOCUMENT MUST BE COMBINED
WITH INSTALLATION INSTRUCTIONS AND MECHANICAL EQUIPMENT DRAWINGS IN ORDER TO UNDERSTAND ALL INSTALLATION REQUIREMENTS.
12. ALLOWABLE EARTH LEAKAGE CURRENT IS 5 MILLIAMPS.
13. STERIS ASSUMES NO RESPONSIBILITY FOR CHANGES MADE THROUGH FAILURE TO OBSERVE THESE INSTRUCTIONS.
14. IT IS RECOMMENDED TO HAVE A DEDICATED BREAKER AS WELL AS A DEDICATED GROUND FOR THE IWC & CEILING PLATE.
15. THE NOMINAL POWER CONSUMPTION IS LISTED FOR THE HARMONYAIR SURGICAL LIGHTING SYSTEM "A-SERIES" FAMILY OF LIGHTING.
WATTS BTU/HR
LIGHTHEAD 40* 136*
LED AMBIENT LIGHT 10 34
THE INCOMING POWER RATING FOR A FOUR LIGHTHEAD SYSTEM INCLUDING AN AMBIENT LIGHT IS 120-240 VAC, 7-3A, 50/60 Hz.
16. ALL MEASUREMENTS ARE IN INCHES UNLESS OTHERWISE SPECIFIED.
17. COMPLETE ROUGH-IN BOX 10101691 FOR IWC [REF. LAS0034] FURNISHED BY STERIS.
18. A 3/4" CONDUIT CAN FIT UP TO 2 STERIS SUPPLIED DC POWER CABLES (12 AWG, 2 CONDUCTOR). IF LARGER AWG CABLES ARE NEEDED DUE TO
RUNS LONGER THAN 75ft. THEN A LARGER CONDUIT MAY BE REQUIRED (CUSTOMER DETERMINED AND SUPPLIED), SEE TABLE 1 ON PAGE 3.
*LIGHTHEAD RANGES FROM 20-140 WATTS (68-478 BTU/HR) BECAUSE OF AUTO-INTENSITY ADJUSTMENTS
STERIS Corporation
Mentor, OH
APPLICABLE TO EQUIPMENT DRAWINGS
V
D W
G
N O
C
A D
D W
G
ACCOUNT NAME
ACCOUNT LOCATION
FILE NAME
PRINTED DATE
SHEET
A
ELECTRICAL EQUIPMENT DRAWING
HarmonyAIR SURGICAL LIGHTING SYSTEM A-SERIES
SYSTEM GROUNDING NOTES
1. NEC GROUNDING BASIC PRINCIPLES -- NATIONAL ELECTRICAL CODE SPECIFICALLY STATES THAT SYSTEMS WILL BE GROUNDED FOR THREE
REASONS:
1.) TO PROTECT FROM LIGHTNING AND OVER-VOLTAGE BY ACCIDENTAL CONTACT
2.) TO REFERENCE THE VOLTAGE AND STABILIZE IT WITH RESPECT TO GROUND (EMI/EMC)
3.) TO FACILITATE FUSING AND GROUND FAULT DEVICES
2. GROUNDING METHOD - A REDUNDANT GROUND METHOD IS USED FOR IWC IN ORDER TO MAKE SURE THAT TOUCH LEAKAGE CURRENT IS
MAINTAINED BELOW 500 MICRO-AMPS. AN EARTH GROUND SHALL BE RUN WITH AC MAINS CONNECTION TO THE ROUGH-IN BOX. A GROUND
WIRE OR CONDUIT CONNECTION SHALL BE RUN FROM ROUGH-IN BOX TO THE CEILING PLATE. AN EARTH GROUND SHALL BE BROUGHT FROM
A SEPARATE GROUND SOURCE TO THE CEILING PLATE.
3. THE GROUNDING POINT FOR ALL DEVICES SHOULD BE THE BUILDING STEEL. ALL HOSPITALS ARE REQUIRED TO HAVE A GRID GROUNDING
SYSTEM WHICH SHOULD INCLUDE THE BUILDING STEEL. THIS DOES NOT NECESSARILY MEAN THAT "ALL" STEEL PARTS ARE INCLUDED IN THE
GRID AND HAVE BONDED GROUND STRAPS. THE OTHER CONSIDERATION IS THAT ALL GROUND STRAPS ARE INTACT. A RESISTANCE CHECK
FROM SYSTEM ELECTRICAL GROUND TO A RELIABLE BUILDING GROUND POINT SHOULD SHOW CONTINUITY. IF CONTINUITY DOES NOT EXIST
FROM THE HUB TO THE BUILDING GROUND NOTIFY THE CUSTOMER SITE MAINTENANCE DEPARTMENT TO CORRECT THE PROBLEM.
5. STERIS "HARMONYAIR SURGICAL LIGHTING SYSTEM A-SERIES" GROUNDING SCHEME IS SHOWN ON PAGES 10 TO 11. ADDITIONAL GROUNDING
REQUIREMENTS ARE LISTED IN THE INSTALLATION INSTRUCTIONS 11007640.
CONDUIT REQUIREMENTS NOTES
1. (AC1) SITE AC POWER FOR IWC.
2. FOR DC POWER, RUNS OTHER THAN THE SUPPLIED (50ft. OR 75ft. OPTIONAL) WILL REQUIRE CUSTOMER SUPPLIED CABLE, SEE TABLE 1.
3. HD VIDEO FIBER OUTPUT CAN GO TO SITE SPECIFIC A/V EQUIPMENT, OR DIRECTLY TO MONITOR (CUSTOMER DETERMINED).
4. ALL CONDUIT, AC SUPPLY LINES, AND GROUND WIRES ARE TO BE SUPPLIED BY CUSTOMER ON SITE.
STERIS Corporation
Mentor, OH
APPLICABLE TO EQUIPMENT DRAWINGS
V
D W
G
N O
C
A D
D W
G
ACCOUNT NAME
ACCOUNT LOCATION
FILE NAME
PRINTED DATE
SHEET
A
ELECTRICAL EQUIPMENT DRAWING
HarmonyAIR SURGICAL LIGHTING SYSTEM A-SERIES
RUN LENGTH FROM
IWC TO HUB
LENGTH IN
75-100
250-328 150-250 100-150
6-75
FEET
LENGTH IN
22-30
75-100
45-75 30-45
2-22
METERS
ACCEPTABLE
4* 6* 8*
AWG
MIN.
(LONG RUNS) CABLE CHART
DC POWER
TABLE 1:
* FOR 4-8 AWG WIRE BOTH ENDS MUST BE SPLICED TO A SHORT LENGTH OF 10 AWG
WIRE IN ORDER TO ACCOMMODATE THE TERMINAL BLOCK SIZE RESTRICTION IN THE
IWC AND HUB BREAKOUT BOARD.
STERIS COMPONENT
MAXIMUM
NUMBER
PER SYSTEM COMMENTS
LIGHTHEADS
AMBIENT LIGHT (OPTIONAL)
MAXIMUM 3 LIGHTHEADS PER HUB BOARD
- REFER TO MECHANICAL EQUIPMENT DRAWINGS FOR MOUNTING REQUIREMENTS OF MECHANICAL COMPONENTS.
- REFER TO SHEETS 10-12 FOR SPECIFIC CONFIGURATIONS. NOT ALL POSSIBLE CONFIGURATIONS ARE SHOWN.
ONE MOUNTED PER CENTRAL SPINDLE.
MAXIMUM CONFIGURATION PER IWC
LIGHTHANDLE HD CAMERA (OPTIONAL)
2 WILL ONLY FUNCTION IF INSERTED INTO HD
COMPATIBLE LIGHTHEAD.
STERIS Corporation
Mentor, OH
APPLICABLE TO EQUIPMENT DRAWINGS
V
D
W G
N
O
C A
D D
W G
ACCOUNT NAME
ACCOUNT LOCATION
FILE NAME
PRINTED DATE
SHEET
A
ELECTRICAL EQUIPMENT DRAWING
HarmonyAIR SURGICAL LIGHTING SYSTEM A-SERIES
CONFIGURATION NOTES
**(SEE FIGURES 1, AND 2 ON SHEETS 10-11)**
GENERAL NOTES:
1. WIRING ENTERS THROUGH CEILING PLATE AREA. TERMINATE ELECTRICAL METALLIC TUBING (EMT) INTO LOW VOLTAGE JUNCTION BOX
WITHIN 18" OF CEILING PLATE AND GROUND TO CEILING PLATE.
2. 1" EMT MAY BE USED IF SYSTEM CONFIGURATION PERMITS.
DRAWING NOTES:
3. SIGNAL CONDUIT SIZE FROM A/V EQUIPMENT IS DETERMINED AND SUPPLIED BY A/V VENDOR.
4. ALL CONDUIT RUNS AND AC SUPPLY LINES ARE TO BE SUPPLIED BY CUSTOMER ON SITE.
5. FIBER CAN BE RUN THRU CONDUIT, LC FIBER CONNECTOR AVAILABLE AT CEILING PLATE.
6. THE DAISY CHAIN HARNESS INCLUDES AN INTEGRATED FLEXIBLE METAL CONDUIT THAT MUST BE CONNECTED TO EITHER A CUSTOMER
SUPPLIED GROUNDED JUNCTION BOX AT BOTH CEILING PLATES OR OPTIONAL HUB BREAKOUT BOARD ENCLOSURE STERIS P/N 10103591 (1
PER HUB).
CEILING PLATE (SPINDLE) 2 DENOTED AS (PRIMARY AND SECONDARY MOUNT)
STERIS Corporation
Mentor, OH
APPLICABLE TO EQUIPMENT DRAWINGS
V
D
W G
N
O
C A
D D
W G
ACCOUNT NAME
ACCOUNT LOCATION
FILE NAME
PRINTED DATE
SHEET
A
ELECTRICAL EQUIPMENT DRAWING
HarmonyAIR SURGICAL LIGHTING SYSTEM A-SERIES
WALL CAVITY ABOVE AND BELOW IWC MUST BE KEPT CLEAR OF ANY OBSTRUCTIONS TO PREVENT INTERFERENCE WITH CONDUITS.
IWC WITHOUT BEZEL
IWC INFORMATION
16.24
13.11
16.5
13.36
2.9
BEZEL MOUNTING SCREWS
BEZEL MOUNTING SCREWS
IWC WITH
BEZEL
USER INTERFACE
(POE)
SERVICE (ETHERNET)
Mentor, OH
APPLICABLE TO EQUIPMENT DRAWINGS
V
D
W G
N
O
C A
D D
W G
ACCOUNT NAME
ACCOUNT LOCATION
FILE NAME
PRINTED DATE
SHEET
A
ELECTRICAL EQUIPMENT DRAWING
HarmonyAIR SURGICAL LIGHTING SYSTEM A-SERIES
60" (1524mm) MIN.
TO FLOOR
SEE GENERAL NOTE 2,
SHEET 2
9.75
9.50
MOUNTING HOLES
FOR ROUGH-IN BOX
TYP BOTH SIDES
3 WIRE INPUT AC SUPPLY -
2 POWER, 1 GROUND
(10 AWG GROUND WIRE)
MUST BEPROVIDED AT
THIS 1/2" [13mm] KNOCKOUT
120V-240VAC, 7-3A,
50/60Hz
3/4" (19mm) X 1" (25mm) KNOCKOUT.
FRONT VIEW
USE THE (2) AVAILABLE
3/4" CONDUITS ONE FOR
DC POWER AND ONE FOR
COMMUNICATION.
11.15
11.25
3.88 MAX
INTEGRATED WALL CONTROL (IWC) ROUGH-IN BOX INFORMATION
(WALL CONTROL NOT SHOWN)
FINISHED WALL
3X 1/2" (13mm) X 3/4" (19mm) KNOCKOUT.
2X 1/2" (13mm) KNOCKOUT.
3X 1/2" (13mm) KNOCKOUT.
2X 3/4" (19mm) X 1" (25mm) KNOCKOUT.
2X 1/2" (13mm) X 3/4" (19mm) KNOCKOUT.
NOTES:
- WALL CAVITY ABOVE AND BELOW IWC MUST BE KEPT CLEAR OF ANY OBSTRUCTIONS TO PREVENT INTERFERENCE WITH CONDUITS.
- LIGHTHEAD DC POWER SUPPLIES HAVE AC INPUT AUTORANGING FEATURE FOR 120V-240VAC
Mentor, OH
APPLICABLE TO EQUIPMENT DRAWINGS
V
D
W G
N
O
C A
D D
W G
ACCOUNT NAME
ACCOUNT LOCATION
FILE NAME
PRINTED DATE
SHEET
A
ELECTRICAL EQUIPMENT DRAWING
HarmonyAIR SURGICAL LIGHTING SYSTEM A-SERIES
BACK VIEW
IWC INFORMATION
HUB 1 (COMM)
HUB 2 (COMM)
ORI (ETHERNET)
HOSPITAL NETWORK
(ETHERNET)
ACT (FUTURE USE)
TO HUB 1
24 VDC -B
TO HUB 1
24 VDC -A
o f 1
S T
E R
IS
C o rp o ra tio n
M e n to r, O H
A P
P L
IC
A
B L
E T
O E
Q U
IP
M
E N
T D
R A
W
IN
G S
A L
L D
IM
E
N S
IO
N
S A
R E
IN
IN
C H
E S
A L
S O
R E
F E
R T
O G
E N
E R
A L
N O
T E
S
REV. DWG. NO.CAD DWG
A C
C O
U N
T N
A M
E
A C
C O
U N
T L
O C
A T
IO
N
F
IL
E N
A M
E
P R
IN
T
E D
D A
T E
S H
E E
T
11014539A
E L
E C
T R
IC
A
L E
Q U
IP
M
E N
T D
R A
W
IN
G
H a rm o n y A
IR
S
U R
G
IC
A L
L
IG
H T
IN
G
S Y
S T
E M
A -S
E R
IE
S
* NON STANDARD AND MUST BE PURCHASED SEPERATELY.
END-2END-1
HUB 1
50 FT.COMMUNICATION CABLE
2 PER
SYSTEM
75 FT.
QUANTITY
50 FT.
LENGTHDESCRIPTION
POWER CABLE, 12 AWG,
JACKETED, LONG RUN (PLENUM RATED)
POWER CABLE, 12 AWG, JACKETED,
2 CONDUCTOR
CARRY LIGHTHEAD
POWER FROM THE
POWER SUPPLY
TO THE HUB
PURPOSE
CONNECTIONS
LAS0029
KITEQUIPMENT #ACCESSORY *
LONG RUN
DBL POWER CABLE
PLENUM RATED
IWC
POWER CABLES
INCLUDED WITH:
PART #HOW ARE THE CABLES SUPPLIED?
10100130
11004734
10100131
CABLE
HARMONYAIR SURGICAL LIGHTING SYSTEM A SERIES CABLES
HUB 1 OR
HUB 2
(SPLIT MOUNT)
IWC
CARRY SIGNALS
BETWEEN IWC
AND THE HUB
1 PER SINGLE MOUNT
2 PER SPLIT MOUNT
COMMUNICATION CABLES
IWC
IWC
LAS0029
LONG RUN COMM KIT
PLENUM RATED
11004635 75 FT.
COMMUNICATION CABLE, LONG RUN
(PLENUM RATED)
11006248
11006248
11006243LAS0025
LAS0025 11006243
LAS0028DAISY CHAIN
POWER
11006245 10100268
14 AWG FMC CABLE W/
INTEGRATED HUB CONNECTORS
15 FT.
N/A
1 PER
SPLIT MOUNT
CARRY DC POWER
FROM HUB 1
TO HUB 2
HUB 2HUB 1
N/A
HARNESS, MASTER CONTROL TO
HUB BREAKOUT BOARD, RJ45
50 FT.LAS0028DAISY CHAIN
POWER
11006245 10100130
R e le a s e d
IWC
SEE SHEET 8
50 FT. (75 FT.)C2 3/4"
C1 50 FT. (75 FT.)3/4"
COMMUNICATION / CAT 5
COMMUNICATION / CAT 5
COMMUNICATION SIGNALS
1,4IWC
NOTESTO
IWC
SITE AC POWER
QTY FROM
NOT BY STERIS
PART NUMBER
JUNCTION BOX, AT PRIMARY HUB
STERIS
CONNECTIONS PAGE
3/4" DC POWER FOR LIGHTHEADS
SITE DETERMINED
SIZE STD (LONG RUN) *
1/2" AC POWER FOR IWC
CABLE DESCRIPTION
AC POWER
LABEL
AC1
DC POWER
DC1
CONDUIT CABLE
CONDUIT REQUIREMENTS (FIGURES 1,2)
2,4, (COMMUNICATION
CABLES)
PRIMARY HUB
SECONDARY HUB1 COMM / CAT 5 CABLE
50 FT.(75 FT.)
1 COMM / CAT 5 CABLE IWC
SEE SHEET 8
(COMMUNICATION
CABLES)
o f 1
S T
E R
IS
C o rp o ra tio n
M e n to r, O H
A P
P L
IC
A
B L
E T
O E
Q U
IP
M
E N
T D
R A
W
IN
G S
A L
L D
IM
E
N S
IO
N
S A
R E
IN
IN
C H
E S
A L
S O
R E
F E
R T
O G
E N
E R
A L
N O
T E
S
REV. DWG. NO.CAD DWG
A C
C O
U N
T N
A M
E
A C
C O
U N
T L
O C
A T
IO
N
F
IL
E N
A M
E
P R
IN
T
E D
D A
T E
S H
E E
T
11014539A
E L
E C
T R
IC
A
L E
Q U
IP
M
E N
T D
R A
W
IN
G
H a rm o n y A
IR
S
U R
G
IC
A L
L
IG
H T
IN
G
S Y
S T
E M
A -S
E R
IE
S
1 PER SYSTEM
A & B POWER CABLE KIT
SEE SHEET 8
(POWER CABLES)
(CONDUIT OPTIONAL)
(CONDUIT OPTIONAL)
* MAXIMUM LENGTH FOR DC CABLES OR CAT5 COMMUNICATION CABLES IS 100M, 328ft. (CUSTOMER SUPPLIED).
R e le a s e d
Mentor, OH
APPLICABLE TO EQUIPMENT DRAWINGS
V
D
W G
N
O
C A
D D
W G
ACCOUNT NAME
ACCOUNT LOCATION
FILE NAME
PRINTED DATE
SHEET
A
ELECTRICAL EQUIPMENT DRAWING
HarmonyAIR SURGICAL LIGHTING SYSTEM A-SERIES
IWC
AC1 1
2" AC1
4"
PRIMARY
CEILING PLATE
C1 DC1
HUB
BREAKOUT
BOARD
4"
NOTES:
1. GROUNDING OF CEILING PLATE - CEILING PLATE MUST BE GROUNDED TO THE SAME
POTENTIAL AS THE IWC. TO ACHIEVE THIS, CONNECT AT LEAST ONE JUNCTION BOX WITH
A GROUND WIRE, 12" (305 MM) OR LESS TO THE CEILING PLATE
CONDUIT AND ELECTRICAL INFORMATION
FIGURE 1
TYPE 1 CONFIGURATION
IWC WITH A SINGLE CEILING PLATE
THIS CONFIGURATION ALLOWS UP TO 3 LIGHTHEADS
PER CEILING PLATE
X/X"
POSSIBLE CABLE ALTERNATIVES.
OPTIONAL EQUIPMENT, CABLES AND CONDUIT.
OPTIONAL EQUIPMENT.
COMMUNICATION CABLES AND CONDUIT.
DC SUPPLY CABLES AND CONDUIT.
RECOMMENDED CONDUIT SIZE.
AC SUPPLY CABLES AND CONDUIT.
LEGEND
REFER TO CONFIGURATION NOTES PAGE 4 FOR
SPECIFIC INFORMATION.#
XXX REFER PAGE 9 FOR CONDUIT REQUIREMENTS
ACX
DCX
CX
EMT GROUNDED AT THIS END
VIDEO RACK
OR
MONITOR
HD VIDEO THROUGH FIBER 3 5
Mentor, OH
APPLICABLE TO EQUIPMENT DRAWINGS
V
D
W G
N
O
C A
D D
W G
ACCOUNT NAME
ACCOUNT LOCATION
FILE NAME
PRINTED DATE
SHEET
A
ELECTRICAL EQUIPMENT DRAWING
HarmonyAIR SURGICAL LIGHTING SYSTEM A-SERIES
IWC AC1 1
2" AC1
4"
PRIMARY
CEILING PLATE
C1
DC1
HUB
BREAKOUT
BOARD
VIDEO RACK
OR
MONITOR
4"
SECONDARY
CEILING PLATE
HUB
BREAKOUT
BOARD
VIDEO RACK
OR
MONITOR
C2
CONDUIT AND ELECTRICAL INFORMATION
HD VIDEO FIBER
PRIMARY MOUNT SECONDARY MOUNT
FIGURE 2
TYPE 2 CONFIGURATION
IWC WITH 2 CEILING PLATES
THIS CONFIGURATION ALLOWS FOR
UP TO 4 LIGHTHEADS
X/X"
POSSIBLE CABLE ALTERNATIVES.
OPTIONAL EQUIPMENT, CABLES AND CONDUIT.
OPTIONAL EQUIPMENT.
COMMUNICATION CABLES AND CONDUIT.
DC SUPPLY CABLES AND CONDUIT.
RECOMMENDED CONDUIT SIZE.
AC SUPPLY CABLES AND CONDUIT.
LEGEND
REFER TO CONFIGURATION NOTES PAGE 4 FOR
SPECIFIC INFORMATION.#
XXX REFER PAGE 9 FOR CONDUIT REQUIREMENTS
ACX
DCX
CX
EMT GROUNDED AT THIS END
HUB DAISY CHAIN HARNESS P/N 10100268
NOTES:
1. GROUNDING OF CEILING PLATE - CEILING PLATE MUST BE GROUNDED TO THE SAME
POTENTIAL AS THE IWC. TO ACHIEVE THIS, CONNECT AT LEAST ONE JUNCTION BOX WITH
A GROUND WIRE, 12" (305 MM) OR LESS TO THE CEILING PLATE
4"
CEILING PLATE
Ø17.95" (456mm)
HUB BREAKOUT BOARD
CLAMPING BRACKET
ASSEMBLY
(STERIS P/N 1013426)
MAIN SPINDLE LOCATION
AUXILIARY SPINDLE
(4 POSSIBLE LOCATIONS)
MONITOR
CONDUIT
MOUNTING
LOCATION
EARTH GROUNDING
LOCATION (GROUND SCREW
(PROVIDED WITH HUB BREAKOUT
ASSEMBLY)
STERIS Corporation
Mentor, OH
APPLICABLE TO EQUIPMENT DRAWINGS
V
D
W G
N
O
C A
D D
W G
ACCOUNT NAME
ACCOUNT LOCATION
FILE NAME
PRINTED DATE
SHEET
A
ELECTRICAL EQUIPMENT DRAWING
HarmonyAIR SURGICAL LIGHTING SYSTEM A-SERIES
COMPONENTS MOUNTED ON CEILING PLATE
(ACTUAL COMPONENTS INSTALLED DEPENDS ON CONFIGURATION)
OPTIONAL
ENCLOSURE
(STERIS P/N 10103591)
OPTIONAL ENCLOSURE FOR HUB BREAKOUT BOARD
(BEAM CLAMP MOUNT ALLOWS MULTIPLE MOUNTING ORIENTATIONS
ON EITHER CEILING PLATE PLATE OR STRUCTURAL PLATE)
TOP VIEW
HUB BREAKOUT BOARD CLAMPING
BRACKET (CAN BE MOUNTED
ANYWHERE AROUND THE
CEILING PLATE PERIMETER)
File details come from the government source that posted it. Updated .