EVOLUTION TYPICAL_WS2 RAIL_TBL2_08-14-2023.pdf
PDF 846 KB Posted
- Attached to
- Z2DB--Renovate Imaging and Catheterization Lab 570 21 220 Fresno VA Medical Center Fresno
- Solicitation number
- 36C26126R0045
About this file
This document is a set of technical drawings and specifications for a Carestream DRX-Evolution X-ray system room layout, dated August 14, 2023, prepared according to 2023 standards.
The drawings detail the installation and configuration requirements for a typical medical X-ray examination room, including comprehensive equipment plans, electrical requirements, structural support specifications, and cabling matrices. Key components specified include a Computer X-Ray Control unit, Generator and PDU, Overhead Tube Crane assembly with bridge travel, wall-mounted stand (premium model), ceiling-mounted rail system, table-fixed elevating equipment, and associated support infrastructure. The equipment plan indicates a room footprint of approximately 20 feet by 14.75 feet with specific mounting provisions for the tube crane bridge system, patient entry area, and control room integration. Environmental requirements specify operation at 64-86°F with 30-65% non-condensing relative humidity, with ceiling height flexibility of 8 feet to 10 feet 3 inches. Electrical specifications mandate 400/480V AC, 60 Hz, 3-phase power with maximum line voltage variation of ±10% and phase imbalance of ±2%. The generator requires 80 kW output capacity across three voltage options (380V, 400V, or 480V). Detailed cabling matrices specify conductor types, sizes, run distances, and connections for DTC cables, wall stand cables, console PC connections, and table connections. Structural requirements include leveled support structures within 0.039-inch tolerance, flush-mounted ceiling fixtures, and parallel rail alignment within 0.0313-inch tolerance, with floor capability to support the equipment's operating weight. All electrical installations must comply with NEC codes and state and local regulations.
View the file
Other files
Show 40 more
50 files listed here. See the rest on GovTribe.
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
AutoCAD SHX Text
CHECKED BY:
AutoCAD SHX Text
PROJECT MGR:
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED
AutoCAD SHX Text DOCUMENT SET INDEX: COVER SHEET...........C1 EQUIPMENT PLAN.......A1 ELECTRICAL PLAN.......E1 ELECTRICAL REQ'MENTS..E2 ELECTRICAL TABLES.....E3 STRUCTURAL UNISTRUT...S1 STRUCTURAL CEILING....S2 STRUCTURAL FLOOR......S3
AutoCAD SHX Text
PROJECT:
AutoCAD SHX Text B
AutoCAD SHX Text
TITLE:
AutoCAD SHX Text
OF
AutoCAD SHX Text
SHT.
AutoCAD SHX Text
REV.
AutoCAD SHX Text
CUSTOMER SIGNATURE:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
SIZE:
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
SHEET:
AutoCAD SHX Text
SEE DRAWING SCALE
AutoCAD SHX Text THESE DRAWINGS ARE CONCEPTUAL UNTIL CARESTREAM RECEIVES A PURCHASE ORDER.DRAWINGS ARE SUBJECT TO CHANGE UNTIL A POST PURCHASE ORDER ON-SITE KICK OFF MEETING IS HELD BETWEEN A CARESTREAM PROJECT MANAGER AND CUSTOMER CONSTRUCTION TEAM.CARESTREAM PROVIDES THESE DRAWINGS AS A CUSTOMER COURTESY AND ARE NOT TO BE USED AS OFFICIAL ARCHITECTURAL OR CONSTRUCTION DOCUMENTS.CARESTREAM ACCEPTS NO LIABILITY OR IMPLIES ANY WARRANTY FOR THE ACCURACY OF THE SITE WHERE THE EQUIPMENT WILL BE STORED OR USED.
AutoCAD SHX Text
NO/MS
AutoCAD SHX Text 08.14.2023
AutoCAD SHX Text
RA
AutoCAD SHX Text
NO/MS
AutoCAD SHX Text 08.14.2023
AutoCAD SHX Text
EVOLUTION SYSTEM
AutoCAD SHX Text
TYPICAL ROOM LAYOUT
AutoCAD SHX Text
WALL STAND-2 ON RAIL
AutoCAD SHX Text
TABLE-2
AutoCAD SHX Text C1
AutoCAD SHX Text T-1
AutoCAD SHX Text
DRX-EVOLUTION
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text T-1
AutoCAD SHX Text
08.14.2023 NO/MS
AutoCAD SHX Text
DESCRIPTION
AutoCAD SHX Text
TYPICAL ROOM LAYOUT PER 2023 STANDARDS
AutoCAD SHX Text
DATE/DFTR
REV
AutoCAD SHX Text
TUBE @ @ 90°WSV
AutoCAD SHX Text
36"DOOR
AutoCAD SHX Text
TUBE @ 0 @ 0°WSH & T
AutoCAD SHX Text
48"DOOR
AutoCAD SHX Text
CONTROL ROOM
AutoCAD SHX Text
LIMIT
AutoCAD SHX Text
LIMIT
AutoCAD SHX Text
40"SID
AutoCAD SHX Text
72"SID
AutoCAD SHX Text
AutoCAD SHX Text -180
AutoCAD SHX Text
AutoCAD SHX Text -90
AutoCAD SHX Text
CHECKED BY:
AutoCAD SHX Text
PROJECT MGR:
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED
AutoCAD SHX Text
PROJECT:
AutoCAD SHX Text B
AutoCAD SHX Text
TITLE:
AutoCAD SHX Text
OF
AutoCAD SHX Text
SHT.
AutoCAD SHX Text
REV.
AutoCAD SHX Text
CUSTOMER SIGNATURE:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
SIZE:
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
SHEET:
AutoCAD SHX Text
SEE DRAWING SCALE
AutoCAD SHX Text
NO/MS
AutoCAD SHX Text 08.14.2023
AutoCAD SHX Text
RA
AutoCAD SHX Text
NO/MS
AutoCAD SHX Text 08.14.2023
AutoCAD SHX Text
EVOLUTION SYSTEM
AutoCAD SHX Text
TYPICAL ROOM LAYOUT
AutoCAD SHX Text
WALL STAND-2
AutoCAD SHX Text
TABLE-2
AutoCAD SHX Text A1
AutoCAD SHX Text T-1
AutoCAD SHX Text
DRX-EVOLUTION
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
COMPUTER-X-RAY CONTROL
AutoCAD SHX Text
CTRL
AutoCAD SHX Text
AutoCAD SHX Text 16.1
AutoCAD SHX Text 13.2
AutoCAD SHX Text 4.0
AutoCAD SHX Text 15.5
AutoCAD SHX Text
ASSY-OVERHEAD TUBE CRANE
AutoCAD SHX Text
OTC
AutoCAD SHX Text
AutoCAD SHX Text 714.0
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
(INCLUDES 4M BRIDGE,TUBE, AND COLLIMATOR)
AutoCAD SHX Text
RAIL-CEILING,LONGITUDINAL
AutoCAD SHX Text
RAIL
AutoCAD SHX Text
AutoCAD SHX Text 146.0
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
(AT 6M MAX RAIL LENGTH)
AutoCAD SHX Text
WALL STAND-PREMIUM
AutoCAD SHX Text
WS
AutoCAD SHX Text
AutoCAD SHX Text 443.0
AutoCAD SHX Text 23.2
AutoCAD SHX Text 89.4
AutoCAD SHX Text
AutoCAD SHX Text
RAIL-FLOOR,WALL STAND
AutoCAD SHX Text
RAIL
AutoCAD SHX Text
AutoCAD SHX Text 43.0
AutoCAD SHX Text 6.7
AutoCAD SHX Text .71
AutoCAD SHX Text
AutoCAD SHX Text
RAIL-CEILING,WALL STAND
AutoCAD SHX Text
RAIL
AutoCAD SHX Text
AutoCAD SHX Text 76.0
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
CHAIN-CABLE MGMT,IGUS
AutoCAD SHX Text
IGUS
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
WINDOW - GLASS, LEAD LINED
AutoCAD SHX Text
LED
AutoCAD SHX Text
AutoCAD SHX Text
(AT 6M MAX RAIL LENGTH)
AutoCAD SHX Text
(AT 6M MAX RAIL LENGTH)
AutoCAD SHX Text
HGHT IN
AutoCAD SHX Text ENERGY CONSUMPTION - STANDBY BTU PER HR: WALL STAND ONLY, 1670.13 DUAL BUCKY SYSTEM, 1931.98
AutoCAD SHX Text OPERATING ENVIRONMENT: TEMPERATURE: 64-86°F (18-30°C) RELATIVE HUMIDITY: 30-65% (NON-CONDENSING)
AutoCAD SHX Text
ENVIRONMENTAL REQUIREMENTS FOR DRX-E
AutoCAD SHX Text
SUPPLIED & INSTALLED BY CUSTOMER OR CONTRACTOR
AutoCAD SHX Text
NO
AutoCAD SHX Text
ITEM
AutoCAD SHX Text
SUPPLIED/INSTALLED BY CARESTREAM HEALTH
AutoCAD SHX Text
WEIGHT LB
AutoCAD SHX Text
WIDTH IN
AutoCAD SHX Text
DPTH IN
AutoCAD SHX Text
EQUIPMENT LEGEND
AutoCAD SHX Text
NO
AutoCAD SHX Text
ITEM
AutoCAD SHX Text CEILING HEIGHT RANGE: 8'-10.3" TO 9'-3.5" (2.70m-2.83m)
AutoCAD SHX Text THIS CEILING HEIGHT RANGE PROVIDES FULL FUNCTIONALITY OF THE CARESTREAM SYSTEM. ANY DEVIATION MUST BE DISCUSSED WITH THE CARESTREAM PROJECT MANAGER.
AutoCAD SHX Text EXISTING CEILING HEIGHT: 0'-0" (X.Xm)
AutoCAD SHX Text COUNTERTOP PROVIDED BY CUSTOMER AT RECOMMENDED 38" A.F.F.WITH LOWER SHELF AT 4" A.F.F.
AutoCAD SHX Text
TABLE FLOAT
AutoCAD SHX Text
EQUIPMENT PLAN
AutoCAD SHX Text
PATIENT ENTRY
AutoCAD SHX Text 20'-0"
AutoCAD SHX Text 20'-0"
AutoCAD SHX Text 9'-8"
AutoCAD SHX Text 9'-8"
AutoCAD SHX Text 0'-4"
AutoCAD SHX Text 0'-4"
AutoCAD SHX Text 16'-9"
AutoCAD SHX Text 16'-9"
AutoCAD SHX Text 0'-3"
AutoCAD SHX Text 0'-3"
AutoCAD SHX Text 14'-0"
AutoCAD SHX Text 14'-0"
AutoCAD SHX Text 0'-2"
AutoCAD SHX Text 0'-2"
AutoCAD SHX Text
BRIDGE
AutoCAD SHX Text
TABLE-FIXED,ELEVATING(CAST)
AutoCAD SHX Text
TBL
AutoCAD SHX Text
AutoCAD SHX Text 696.7
AutoCAD SHX Text 33.0
AutoCAD SHX Text
AutoCAD SHX Text 98.7
AutoCAD SHX Text
NOTES: : 72" SID TO FLOOR IS OBTAINABLE
AutoCAD SHX Text
GENERATOR AND PDU
AutoCAD SHX Text G/P
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text 25.4
AutoCAD SHX Text 47.6
22.7
AutoCAD SHX Text
CONTROL ROOM
AutoCAD SHX Text
ELECTRICAL PLAN ELECTRICAL INSTALLED BY CUSTOMER
AutoCAD SHX Text
CHECKED BY:
AutoCAD SHX Text
PROJECT MGR:
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED
AutoCAD SHX Text
PROJECT:
AutoCAD SHX Text B
AutoCAD SHX Text
TITLE:
AutoCAD SHX Text
OF
AutoCAD SHX Text
SHT.
AutoCAD SHX Text
REV.
AutoCAD SHX Text
CUSTOMER SIGNATURE:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
SIZE:
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
SHEET:
AutoCAD SHX Text
SEE DRAWING SCALE
AutoCAD SHX Text
NO/MS
AutoCAD SHX Text 08.14.2023
AutoCAD SHX Text
RA
AutoCAD SHX Text
NO/MS
AutoCAD SHX Text 08.14.2023
AutoCAD SHX Text
EVOLUTION SYSTEM
AutoCAD SHX Text
TYPICAL ROOM LAYOUT
AutoCAD SHX Text
WALL STAND-2
AutoCAD SHX Text
TABLE-2
AutoCAD SHX Text E1
AutoCAD SHX Text T-1
AutoCAD SHX Text
DRX-EVOLUTION
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
SYM.SUPPLIED/INSTALLED BY CUSTOMER/CONTRACTOR UNLESS NOTED
AutoCAD SHX Text
ELECTRICAL LEGEND
AutoCAD SHX Text
OTC
AutoCAD SHX Text
OTC
AutoCAD SHX Text
CARESTREAM OTC CABLE ENTRY TO CD
AutoCAD SHX Text NOTE: ALL ELECTRICAL DUCTS AND CONNECTIONS MUST CONFORM TO THE N.E.C. CODES, IN ADDITION TO STATE AND LOCAL REGULATIONS FOR THE COUNTRY OR LOCALITY IN WHICH THE EQUIPMENT IS INSTALLED. SITE IS RESPONSIBLE TO DETERMINE ELECTRICAL DESIGN INCLUDING NUMBER OF DIVIDERS IN CABLE TROUGH PER CODE
AutoCAD SHX Text
CO3
AutoCAD SHX Text
NETWORK DROP, 14" A.F.F.
AutoCAD SHX Text JUNCTION BOX- 8"x8" FLUSH MOUNTED IN CONTROL ROOM WALL, BOTTOM AT 14" A.F.F.
AutoCAD SHX Text
JB2
AutoCAD SHX Text R1
AutoCAD SHX Text RISER- 10"x3 " FLUSH MOUNTED WITH REMOVABLE COVERS 12" FLUSH MOUNTED WITH REMOVABLE COVERS FROM CD TO FINISHED FLOOR, PROVIDE DIVIDERS PER LOCAL CODE
AutoCAD SHX Text
WL
AutoCAD SHX Text
EPO
AutoCAD SHX Text
CD
AutoCAD SHX Text CEILING DUCT- 10"x3 " ABOVE FINISHED CEILING, WITH 12" ABOVE FINISHED CEILING, WITH REMOVABLE COVERS FACING UP, PROVIDE DIVIDERS PER LOCAL CODE
AutoCAD SHX Text
FL
AutoCAD SHX Text
CO1
AutoCAD SHX Text
GEN
AutoCAD SHX Text
GENERATOR- SEE POWER SPECIFICATIONS ON SHEET E2 SUPPLIED/INSTALLED BY CARESTREAM
AutoCAD SHX Text
CO2
AutoCAD SHX Text
TBL
AutoCAD SHX Text
CARESTREAM TABLE BASE SHOWN FOR REFERENCE
AutoCAD SHX Text EMERGENCY POWER OFF- OPTIONAL SWITCH TO CUSTOMER SHUNT BREAKER IN MAIN DISCONNECT, EXACT LOCATION PER LOCAL CODE OR BY CUSTOMER
AutoCAD SHX Text
WS
AutoCAD SHX Text R2
AutoCAD SHX Text
RISER- 2X 2 " ELEC.CONDUIT FROM CD TO JB2 12" ELEC.CONDUIT FROM CD TO JB2
AutoCAD SHX Text CUSTOMER PROVIDED 1 " SEAL-TIGHT, INCLUDE 12" SEAL-TIGHT, INCLUDEFITTINGS FOR MAIN POWER TO CARESTREAM GENERATOR, WIP LENGTH PER LOCAL CODE,INSTALLED BY CUSTOMER ELECTRICIAN. SEE SHEET E2 FOR POWER REQUIREMENTS
AutoCAD SHX Text
CARESTREAM WALL STAND CABLE ENTRY
AutoCAD SHX Text WARNING LIGHT- OPTIONAL "X-RAY IN USE" LIGHT PER CODE, CUSTOMER TO PROVIDE FIXTURE AND 24V POWER / CONTACTS TO CSH GENERATOR. SEE SHEET E2 SCHEMATIC DIAGRAM
AutoCAD SHX Text
CUT OUT- 3" WITH GROMMET ON FACE PLATE OF JB2
AutoCAD SHX Text
CUT OUT- 3" KNOCK OUT OR 3" GROMMETED SPACE IN COVERPLATES OF R1, 36" A.F.F.
AutoCAD SHX Text
JB2
AutoCAD SHX Text C02
AutoCAD SHX Text R1
AutoCAD SHX Text
120 VAC QUAD OUTLET, 14" A.F.F.
AutoCAD SHX Text
WL
AutoCAD SHX Text C01
AutoCAD SHX Text
CO3
AutoCAD SHX Text
CUT OUT- 3" ON BOTTOM FACE OF CD
AutoCAD SHX Text NOTES: : ALL OUTLETS, WHETHER DUPLEX OR QUAD, SHOWN ON CARESTREAM SITE SPECIFIC DRAWINGS, ARE EXCLUSIVELY FOR CARESTREAM EQUIPMENT. ANY CUSTOMER REQUIRED OUTLETS MUST BE APPROVED SEPARATELY.
AutoCAD SHX Text 1'-0"
AutoCAD SHX Text 1'-0"
AutoCAD SHX Text 14'-0 "14"
AutoCAD SHX Text 14'-0 "14"
AutoCAD SHX Text 5'-11 "12"
AutoCAD SHX Text 5'-11 "12"
AutoCAD SHX Text 8'-7 "14"
AutoCAD SHX Text 8'-7 "14"
AutoCAD SHX Text 13'-5 "14"
AutoCAD SHX Text 13'-5 "14"
AutoCAD SHX Text 14'-0"
AutoCAD SHX Text 14'-0"
AutoCAD SHX Text 8'-6"
AutoCAD SHX Text 8'-6"
AutoCAD SHX Text
FL
AutoCAD SHX Text
TBL
AutoCAD SHX Text
CD
AutoCAD SHX Text
CD
AutoCAD SHX Text
EPO
AutoCAD SHX Text
GEN
AutoCAD SHX Text
WS
AutoCAD SHX Text
CO4
AutoCAD SHX Text R2
AutoCAD SHX Text
CO4
CUT OUT- 2" WITH GROMMET ON COVER PLATES OF R1,8" BELOW FINISHED CEILING
2"
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text H
AutoCAD SHX Text
AutoCAD SHX Text H
AutoCAD SHX Text
RUN
AutoCAD SHX Text
AutoCAD SHX Text 4"
AutoCAD SHX Text
AutoCAD SHX Text 5"
AutoCAD SHX Text
PER CODE
AutoCAD SHX Text
PER CODE
AutoCAD SHX Text
PER CODE
AutoCAD SHX Text
PER CODE
AutoCAD SHX Text
PER CODE
AutoCAD SHX Text 50'-0"
AutoCAD SHX Text 2 1/8"
AutoCAD SHX Text 3"
AutoCAD SHX Text 33'-0"
AutoCAD SHX Text 2"
AutoCAD SHX Text 3"
AutoCAD SHX Text 30'-0"
AutoCAD SHX Text 1 1/8"
AutoCAD SHX Text 3"
AutoCAD SHX Text 45'-0"
AutoCAD SHX Text
RJ45
AutoCAD SHX Text 3"
AutoCAD SHX Text 57'-0"
AutoCAD SHX Text
RJ45
AutoCAD SHX Text N/A
AutoCAD SHX Text 28'-0"
AutoCAD SHX Text
RJ45
AutoCAD SHX Text N/A
AutoCAD SHX Text 50'-0"
AutoCAD SHX Text
RJ45
AutoCAD SHX Text 2.5"
AutoCAD SHX Text 43'-0"
AutoCAD SHX Text
RJ45
AutoCAD SHX Text
H=HOSPITAL E=EMERGENCY STOP COMMUNICATION CABLE OCTOBER.07.2009
AutoCAD SHX Text 2.5"
AutoCAD SHX Text
PDU TO CONTROL
AutoCAD SHX Text
PDU TO OTC
AutoCAD SHX Text
E-OTC TO JB (CSH)
AutoCAD SHX Text
CONTROL TO OTC
AutoCAD SHX Text
PDU TO WALL STAND
AutoCAD SHX Text
E-WS TO JB(CSH)
AutoCAD SHX Text
PDU TO TABLE
AutoCAD SHX Text
E-TABLE TO JB(CSH)
AutoCAD SHX Text
HOSPITAL'S NETWORK TO CONTROL DESK
AutoCAD SHX Text
HOSPITAL'S MAIN TO PDU
AutoCAD SHX Text
MIN. CONDUIT SIZE (INCH)
AutoCAD SHX Text
MAX. CONNCT'R SIZE (INCH)
AutoCAD SHX Text
MAX. CONDUIT LENGTH (FT.)
AutoCAD SHX Text
CABLE RUN (POINT TO POINT)
AutoCAD SHX Text NOTE: ALL CABLES TERMINATE AT MAIN CONNECTION POINT FOR EACH MAJOR ASSEMBLY UNLESS SHOWN OTHERWISE.
AutoCAD SHX Text
MAIN POWER INTERCONNECTIONS AND WIRE SIZE
AutoCAD SHX Text
PER CODE
AutoCAD SHX Text ALL ELECTRICAL CONNECTIONS SHOULD CONFORM TO N.E.C. AND TO STATE AND LOCAL CODES. MAIN POWER REQUIREMENT VOLTAGE: 400/480V AC, 60 HZ, 3 PHASE WYE. MAX. LINE VOLTAGE VARIATION: ±10% MAX. PHASE IMBALANCE: ±2% MAX. LINE IMPEDANCE: 0.1 Ohms @ 400V, 0.15 OHMS @480V RECOMMENDED DISTRIBUTION POWER: INCOMING APPARENT POWER:105kva LOCKABLE CIRCUIT DISCONNECT OR TIME DELAYED FUSES: MAX. WIRE SIZE 2/0 AWG PER CODE FROM MAIN DISCONNECT TO PDU.
AutoCAD SHX Text
ELECTRICAL POWER REQUIREMENTS FOR DRX-E
AutoCAD SHX Text
CHECKED BY:
AutoCAD SHX Text
PROJECT MGR:
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED
AutoCAD SHX Text
PROJECT:
AutoCAD SHX Text B
AutoCAD SHX Text
TITLE:
AutoCAD SHX Text
OF
AutoCAD SHX Text
SHT.
AutoCAD SHX Text
REV.
AutoCAD SHX Text
CUSTOMER SIGNATURE:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
SIZE:
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
SHEET:
AutoCAD SHX Text
SEE DRAWING SCALE
AutoCAD SHX Text
NO/MS
AutoCAD SHX Text 08.14.2023
AutoCAD SHX Text
RA
AutoCAD SHX Text
NO/MS
AutoCAD SHX Text 08.14.2023
AutoCAD SHX Text
EVOLUTION SYSTEM
AutoCAD SHX Text
TYPICAL ROOM LAYOUT
AutoCAD SHX Text
WALL STAND-2
AutoCAD SHX Text
TABLE-2
AutoCAD SHX Text E2
AutoCAD SHX Text T-1
AutoCAD SHX Text
DRX-EVOLUTION
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
WARNING LIGHT "X-RAY IN USE" WIRING DIAGRAM
AutoCAD SHX Text
SUPPLIED AND INSTALLED BY CUSTOMER - CONNECTED BY CSH TO GENERATOR:
AutoCAD SHX Text
GENERATOR
AutoCAD SHX Text
24 VOLT TRANSFORMER
AutoCAD SHX Text
RELAY COIL
AutoCAD SHX Text Maximum Output Power
AutoCAD SHX Text Nominal Line Voltage (V)ac
AutoCAD SHX Text Main Disconnect (A)
AutoCAD SHX Text Recommended Distribution on Transformer Rating (kVA)
AutoCAD SHX Text Max. Input Line Resistance (@ full-load RMS)
AutoCAD SHX Text
GENERATOR - CARESTREAM X-RAY ELECTRICAL REQUIREMENTS
AutoCAD SHX Text Instantaneous Current (A)
AutoCAD SHX Text BTU/Hr
AutoCAD SHX Text 3-Phase, 380 - 480 V (ac)
AutoCAD SHX Text
80 KW
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text 0.13
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text 0.17
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text 0.11
Electri- Cord
AutoCAD SHX Text Claymount
AutoCAD SHX Text Alpha
AutoCAD SHX Text Alpha
AutoCAD SHX Text Belden
AutoCAD SHX Text L-Com
AutoCAD SHX Text L-Com
AutoCAD SHX Text Various
AutoCAD SHX Text Manuf. Part No.
AutoCAD SHX Text Manuf.
AutoCAD SHX Text
OTC CABLES
AutoCAD SHX Text Various
AutoCAD SHX Text
Q57712 JCR2
AutoCAD SHX Text 13650
AutoCAD SHX Text 65C07C4
AutoCAD SHX Text 5006C
AutoCAD SHX Text
AutoCAD SHX Text 7C2592
AutoCAD SHX Text
TC2599
AutoCAD SHX Text N/A
AutoCAD SHX Text N/A
AutoCAD SHX Text CSH Part No.
AutoCAD SHX Text 2G4200
AutoCAD SHX Text 2G4201
AutoCAD SHX Text 2G4202
AutoCAD SHX Text 2G4275
AutoCAD SHX Text 2G3790
AutoCAD SHX Text 2G3799
AutoCAD SHX Text 2G3721
AutoCAD SHX Text 2G4120
AutoCAD SHX Text 2G4121
AutoCAD SHX Text Cable Name
AutoCAD SHX Text Tube Stator
AutoCAD SHX Text OTC Can Out
AutoCAD SHX Text SCB to PDU Ethernet
AutoCAD SHX Text GND - Electronics Assy to PDU
AutoCAD SHX Text GND - Tube to PDU
AutoCAD SHX Text Power 240 VAC
AutoCAD SHX Text Tube High Voltage
AutoCAD SHX Text Console Estop/Auto Position to TMC
AutoCAD SHX Text OTC Echain Out
AutoCAD SHX Text Cable Rating
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text 75k VDC
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text Applied Voltage
AutoCAD SHX Text
24 VDC
AutoCAD SHX Text
12 VDC
AutoCAD SHX Text N/A
AutoCAD SHX Text N/A
AutoCAD SHX Text 240 V
AutoCAD SHX Text 75k VDC
AutoCAD SHX Text
AutoCAD SHX Text
24 VDC
AutoCAD SHX Text
12 VDC
AutoCAD SHX Text From
AutoCAD SHX Text Console J1
AutoCAD SHX Text
SCB P11
AutoCAD SHX Text Electronics Assy G4
AutoCAD SHX Text Tube
AutoCAD SHX Text
PDU
AutoCAD SHX Text
PDU
AutoCAD SHX Text
PDU
AutoCAD SHX Text Junction Box
AutoCAD SHX Text Junction Box
AutoCAD SHX Text To
AutoCAD SHX Text TMC Board P23/J26000
AutoCAD SHX Text
PDU
AutoCAD SHX Text PDU Ground
AutoCAD SHX Text PDU Tank Ground
AutoCAD SHX Text Tube
AutoCAD SHX Text Tube
AutoCAD SHX Text WS/Table Echain in
AutoCAD SHX Text Electronics AY PS
AutoCAD SHX Text WS/Table Can in
AutoCAD SHX Text Usable Length
AutoCAD SHX Text
15.805 m (50.58 ft)
AutoCAD SHX Text
12.055 m (38.58 ft)
AutoCAD SHX Text
15.855 m (50.74 ft)
AutoCAD SHX Text
12.605 m (40.34 ft)
AutoCAD SHX Text
15.833 m (50.74 ft)
AutoCAD SHX Text Function
AutoCAD SHX Text Emergency Stop
AutoCAD SHX Text Communication
AutoCAD SHX Text Ground
AutoCAD SHX Text Ground
AutoCAD SHX Text Anode Cathode X-ray Tube
AutoCAD SHX Text Stator X-ray Tube
AutoCAD SHX Text OTC Power
AutoCAD SHX Text
12.730 m (40.74 ft)
AutoCAD SHX Text
12.505 m (40.02 ft)
AutoCAD SHX Text
9.705 m (31.06 ft)
AutoCAD SHX Text Emergency Stop
AutoCAD SHX Text
10.055 m (32.18 ft)
AutoCAD SHX Text Communication
AutoCAD SHX Text Various
AutoCAD SHX Text Various
AutoCAD SHX Text Various
AutoCAD SHX Text Manuf. Part No.
AutoCAD SHX Text
CONSOLE PC
AutoCAD SHX Text
Q57712 JCR
AutoCAD SHX Text N/A
AutoCAD SHX Text N/A
AutoCAD SHX Text N/A
AutoCAD SHX Text N/A
AutoCAD SHX Text CSH Part No.
AutoCAD SHX Text 8H1110
AutoCAD SHX Text 8H1111
AutoCAD SHX Text 8H1108
AutoCAD SHX Text 8H1109
AutoCAD SHX Text 6H1305
AutoCAD SHX Text 8H1107
AutoCAD SHX Text Cable Name
AutoCAD SHX Text Communication
AutoCAD SHX Text Table GRND
AutoCAD SHX Text Console Ground
AutoCAD SHX Text Control
AutoCAD SHX Text PC Power
AutoCAD SHX Text Monitor Power
AutoCAD SHX Text Cable Rating
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text Applied Voltage
AutoCAD SHX Text N/A
AutoCAD SHX Text N/A
AutoCAD SHX Text
12 VDC
AutoCAD SHX Text
240 VAC
AutoCAD SHX Text
240 VAC
AutoCAD SHX Text
12 VDC
AutoCAD SHX Text From
AutoCAD SHX Text
PDU
AutoCAD SHX Text
PDU
AutoCAD SHX Text
PDU
AutoCAD SHX Text To
AutoCAD SHX Text Console
AutoCAD SHX Text Console
AutoCAD SHX Text Console
AutoCAD SHX Text Monitor
AutoCAD SHX Text
PC
AutoCAD SHX Text Usable Length
AutoCAD SHX Text
AutoCAD SHX Text 20 m (65.6 ft)
AutoCAD SHX Text Function
AutoCAD SHX Text Ground
AutoCAD SHX Text Ground
AutoCAD SHX Text Monitor Power
AutoCAD SHX Text PC Power
AutoCAD SHX Text L-Com
AutoCAD SHX Text
TC259
AutoCAD SHX Text 8H2554
AutoCAD SHX Text
AutoCAD SHX Text
12 VDC
AutoCAD SHX Text
PDU
AutoCAD SHX Text Manuf.
AutoCAD SHX Text Various
AutoCAD SHX Text
Q57712 JCR
AutoCAD SHX Text Communication
AutoCAD SHX Text
PDU
AutoCAD SHX Text
PDU
AutoCAD SHX Text
PDU
AutoCAD SHX Text
PC
AutoCAD SHX Text
PC
AutoCAD SHX Text 20 m (65.6 ft)
AutoCAD SHX Text 20 m (65.6 ft)
AutoCAD SHX Text 20 m (65.6 ft)
AutoCAD SHX Text 20 m (65.6 ft)
AutoCAD SHX Text 20 m (65.6 ft)
AutoCAD SHX Text Communication
AutoCAD SHX Text Communication
AutoCAD SHX Text Communication
AutoCAD SHX Text Electri- Cord
AutoCAD SHX Text Electri- Cord
AutoCAD SHX Text Electri- Cord
AutoCAD SHX Text Alpha
AutoCAD SHX Text Belden
AutoCAD SHX Text L-Com
AutoCAD SHX Text Various
AutoCAD SHX Text Manuf. Part No.
AutoCAD SHX Text Manuf.
AutoCAD SHX Text
WALL STAND
AutoCAD SHX Text
CABLING MATRIX
AutoCAD SHX Text
TC2592
AutoCAD SHX Text
TC2592
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text N/A
AutoCAD SHX Text
AutoCAD SHX Text CSH Part No.
AutoCAD SHX Text 2G4190
AutoCAD SHX Text 2G4156
AutoCAD SHX Text 2G4157
AutoCAD SHX Text 2G4158
AutoCAD SHX Text 2G4159
AutoCAD SHX Text 2G4177
AutoCAD SHX Text 2G4178
AutoCAD SHX Text 2G4179
AutoCAD SHX Text Cable Name
AutoCAD SHX Text WS CAN out
AutoCAD SHX Text WS GND - PDU to PS/TMC
AutoCAD SHX Text WS AEC - PDU to WS Bucky
AutoCAD SHX Text WS Ethernet - PDU to Detector
AutoCAD SHX Text WS CAN in
AutoCAD SHX Text WS Echain in
AutoCAD SHX Text WS Echain Out
AutoCAD SHX Text Cable Rating
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text Applied Voltage
AutoCAD SHX Text
24 VDC
AutoCAD SHX Text
12 VDC
AutoCAD SHX Text N/A
AutoCAD SHX Text
240 VAC
AutoCAD SHX Text
12 VDC
AutoCAD SHX Text
24 VDC
AutoCAD SHX Text From
AutoCAD SHX Text
PDU
AutoCAD SHX Text
PDU
AutoCAD SHX Text Junction Box
AutoCAD SHX Text To
AutoCAD SHX Text TMC Board G3
AutoCAD SHX Text
AEC
AutoCAD SHX Text WS - Power Supply
AutoCAD SHX Text
WS TMC P18
AutoCAD SHX Text Detector
AutoCAD SHX Text Usable Length
AutoCAD SHX Text
19.65 m (62.99 ft)
AutoCAD SHX Text
22.015 m (70.45 ft)
AutoCAD SHX Text
19.5 m (62.4 ft)
AutoCAD SHX Text
22.015 m (70.45 ft)
AutoCAD SHX Text Function
AutoCAD SHX Text Emergency Stop
AutoCAD SHX Text
AEC
AutoCAD SHX Text Wall Stand Power
AutoCAD SHX Text
9.3 m (29.76 ft)
AutoCAD SHX Text Emergency Stop
AutoCAD SHX Text Various
AutoCAD SHX Text Alpha
AutoCAD SHX Text Manuf. Part No.
AutoCAD SHX Text
TABLE
AutoCAD SHX Text
Q57712 JCR
AutoCAD SHX Text
TC2592
AutoCAD SHX Text N/A
AutoCAD SHX Text
AutoCAD SHX Text CSH Part No.
AutoCAD SHX Text 2G4447
AutoCAD SHX Text Cable Name
AutoCAD SHX Text Table Can in
AutoCAD SHX Text Table GND - PDU to Table PS/DNC
AutoCAD SHX Text Table Echain in
AutoCAD SHX Text Cable Rating
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text Applied Voltage
AutoCAD SHX Text N/A
AutoCAD SHX Text
12 VDC
AutoCAD SHX Text
240 VAC
AutoCAD SHX Text
24 VDC
AutoCAD SHX Text
12 VDC
AutoCAD SHX Text From
AutoCAD SHX Text
PDU
AutoCAD SHX Text To
AutoCAD SHX Text DMC Board
AutoCAD SHX Text
AEC
AutoCAD SHX Text Power Supply
AutoCAD SHX Text DMC Board P17
AutoCAD SHX Text Detector
AutoCAD SHX Text Usable Length
AutoCAD SHX Text 24 m (56.8 ft)
AutoCAD SHX Text Function
AutoCAD SHX Text Ground
AutoCAD SHX Text
AEC
AutoCAD SHX Text Manuf.
AutoCAD SHX Text
PDU
AutoCAD SHX Text
PDU
AutoCAD SHX Text
PDU
AutoCAD SHX Text DMC Board J14
AutoCAD SHX Text 18 m (57.6 ft)
AutoCAD SHX Text 14 m (44.8 ft)
AutoCAD SHX Text Table Power
AutoCAD SHX Text Communication
AutoCAD SHX Text Electri- Cord
AutoCAD SHX Text Belden
AutoCAD SHX Text L-Com
AutoCAD SHX Text L-Com
AutoCAD SHX Text Belden
AutoCAD SHX Text
Q57712 JCR
AutoCAD SHX Text
TC2599
AutoCAD SHX Text WS AEC - PDU to WS Power Supply
AutoCAD SHX Text
12 VDC
AutoCAD SHX Text
12 VDC
AutoCAD SHX Text
PDU
AutoCAD SHX Text
PDU
AutoCAD SHX Text Junction Box
AutoCAD SHX Text Junction Box
AutoCAD SHX Text Junction Box
AutoCAD SHX Text
WS TMC P17
AutoCAD SHX Text
WS TMC J25
AutoCAD SHX Text
WS TMC J24
AutoCAD SHX Text
9.3 m (29.76 ft)
AutoCAD SHX Text
9.35 m (29.97 ft)
AutoCAD SHX Text
9.35 m (29.97 ft)
AutoCAD SHX Text Communication
AutoCAD SHX Text Ground
AutoCAD SHX Text Communication
AutoCAD SHX Text Communication
AutoCAD SHX Text L-Com
AutoCAD SHX Text L-Com
AutoCAD SHX Text
TC2599
AutoCAD SHX Text
AutoCAD SHX Text 2G4449
AutoCAD SHX Text 2G4450
AutoCAD SHX Text 2G3791
AutoCAD SHX Text 2G3800
AutoCAD SHX Text 2G4178
AutoCAD SHX Text Table AEC - PDU to Table Bucky
AutoCAD SHX Text Table Power - PDU to Table Power Supply
AutoCAD SHX Text Table Ethernet - PDU to Detector
AutoCAD SHX Text
12 VDC
AutoCAD SHX Text Junction Box
AutoCAD SHX Text Junction Box
AutoCAD SHX Text 18 m (57.6 ft)
AutoCAD SHX Text 18 m (57.6 ft)
AutoCAD SHX Text 14 m (44.8 ft)
AutoCAD SHX Text Communication
AutoCAD SHX Text Emergency Stop
AutoCAD SHX Text
CHECKED BY:
AutoCAD SHX Text
PROJECT MGR:
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED
AutoCAD SHX Text
PROJECT:
AutoCAD SHX Text B
AutoCAD SHX Text
TITLE:
AutoCAD SHX Text
OF
AutoCAD SHX Text
SHT.
AutoCAD SHX Text
REV.
AutoCAD SHX Text
CUSTOMER SIGNATURE:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
SIZE:
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
SHEET:
AutoCAD SHX Text
SEE DRAWING SCALE
AutoCAD SHX Text
NO/MS
AutoCAD SHX Text 08.14.2023
AutoCAD SHX Text
RA
AutoCAD SHX Text
NO/MS
AutoCAD SHX Text 08.14.2023
AutoCAD SHX Text
EVOLUTION SYSTEM
AutoCAD SHX Text
TYPICAL ROOM LAYOUT
AutoCAD SHX Text
WALL STAND-2
AutoCAD SHX Text
TABLE-2
AutoCAD SHX Text E3
AutoCAD SHX Text T-1
AutoCAD SHX Text
DRX-EVOLUTION
AutoCAD SHX Text
BRIDGE TRAVEL
AutoCAD SHX Text
CHECKED BY:
AutoCAD SHX Text
PROJECT MGR:
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED
AutoCAD SHX Text
PROJECT:
AutoCAD SHX Text B
AutoCAD SHX Text
TITLE:
AutoCAD SHX Text
OF
AutoCAD SHX Text
SHT.
AutoCAD SHX Text
REV.
AutoCAD SHX Text
CUSTOMER SIGNATURE:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
SIZE:
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
SHEET:
AutoCAD SHX Text
SEE DRAWING SCALE
AutoCAD SHX Text
NO/MS
AutoCAD SHX Text 08.14.2023
AutoCAD SHX Text
RA
AutoCAD SHX Text
NO/MS
AutoCAD SHX Text 08.14.2023
AutoCAD SHX Text
EVOLUTION SYSTEM
AutoCAD SHX Text
TYPICAL ROOM LAYOUT
AutoCAD SHX Text
WALL STAND-2
AutoCAD SHX Text
TABLE-2
AutoCAD SHX Text S1
AutoCAD SHX Text T-1
AutoCAD SHX Text
DRX-EVOLUTION
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
UNISTRUT PLAN UNISTRUT INSTALLED BY CUSTOMER
AutoCAD SHX Text
ITEM
AutoCAD SHX Text
COMPONENT DESCRIPTION: SUPPLIED AND INSTALLED BY CUSTOMER/CONTRACTOR
AutoCAD SHX Text NEW UNISTRUT: FLUSH MOUNTED WITH FINISHED CEILING. UNISTRUT P1000 SERIES OR EQUIVALENT REQUIRED.
AutoCAD SHX Text
STRUCTURAL LEGEND
AutoCAD SHX Text
STRUCTURAL NOTES
AutoCAD SHX Text
1. CEILING TILES MUST NOT BE LOWER THAN UNISTRUT AT CEILING TILES MUST NOT BE LOWER THAN UNISTRUT AT FINISHED CEILING. 2. WIRELESS ANTENNAS SHOULD NOT BE MOUNTED TO THE CEILING INSIDE/WITHIN THE CSH "BRIDGE TRAVEL" AREA.
AutoCAD SHX Text
1. THE FACILITY STRUCTURE SUPPORTING ALL EQUIPMENT LOADS THE FACILITY STRUCTURE SUPPORTING ALL EQUIPMENT LOADS MUST BE VERIFIED BY CUSTOMER'S STRUCTURAL ENGINEER OF RECORD. THIS INCLUDES LOADS FOR WALL, CEILING AND FLOOR. 2. SUPPORT STRUCTURE MUST BE LEVELED AT FINISHED CEILING SUPPORT STRUCTURE MUST BE LEVELED AT FINISHED CEILING HEIGHT WITHIN 0.039 INCH (1.000 MM). 3. THE EQUIPMENT WILL ATTACH TO SUPPORT STRUCTURE RAILS THE EQUIPMENT WILL ATTACH TO SUPPORT STRUCTURE RAILS THAT ARE PARALLEL WITH EACH OTHER TO WITHIN 0.0313 INCH (0.795 MM). 4. SEE SPECIFICATION 8G8804 FOR STRUCTURAL INFORMATION. SEE SPECIFICATION 8G8804 FOR STRUCTURAL INFORMATION.
AutoCAD SHX Text
GENERAL NOTES
AutoCAD SHX Text 2'-4"
AutoCAD SHX Text 2'-4"
AutoCAD SHX Text 14'-0"
AutoCAD SHX Text 14'-0"
AutoCAD SHX Text 11'-9 "34"
AutoCAD SHX Text 11'-9 "34"
AutoCAD SHX Text 1'-5 "12"
AutoCAD SHX Text 1'-5 "12"
AutoCAD SHX Text 17'-7 "12"
AutoCAD SHX Text 17'-7 "12"
AutoCAD SHX Text 1'-6"
AutoCAD SHX Text 1'-6"
AutoCAD SHX Text 15'-2"
15'-2"
CHECKED BY:
AutoCAD SHX Text
PROJECT MGR:
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED
AutoCAD SHX Text
PROJECT:
AutoCAD SHX Text B
AutoCAD SHX Text
TITLE:
AutoCAD SHX Text
OF
AutoCAD SHX Text
SHT.
AutoCAD SHX Text
REV.
AutoCAD SHX Text
CUSTOMER SIGNATURE:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
SIZE:
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
SHEET:
AutoCAD SHX Text
SEE DRAWING SCALE
AutoCAD SHX Text
NO/MS
AutoCAD SHX Text 08.14.2023
AutoCAD SHX Text
RA
AutoCAD SHX Text
NO/MS
AutoCAD SHX Text 08.14.2023
AutoCAD SHX Text
EVOLUTION SYSTEM
AutoCAD SHX Text
TYPICAL ROOM LAYOUT
AutoCAD SHX Text
WALL STAND-2
AutoCAD SHX Text
TABLE-2
AutoCAD SHX Text S2
AutoCAD SHX Text T-1
AutoCAD SHX Text
DRX-EVOLUTION
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
CEILING PLAN EQUIPMENT INSTALLED BY CSH
AutoCAD SHX Text
ITEM
AutoCAD SHX Text
COMPONENT DESCRIPTION: SUPPLIED AND INSTALLED BY CARESTREAM HEALTH INC.
AutoCAD SHX Text
OTC CEILING RAIL
AutoCAD SHX Text
STRUCTURAL LEGEND
AutoCAD SHX Text
IGUS CHAIN
AutoCAD SHX Text
WALL STAND CEILING RAIL
AutoCAD SHX Text RESTRICTED CEILING AREA FOR CARESTREAM EQUIPMENT. NO CEILING MOUNTED FIXTURES, INCLUDING HVAC, LIGHTING AND SPRINKLERS.
AutoCAD SHX Text
1. REFLECTED CEILING PLAN REFERENCES LOCATION OF THE CARESTREAM CEILING MOUNTED EQUIPMENT NOT TO INTERFERE WITH LIGHTS, DIFFUSERS AND SPRINKLER HEADS.
AutoCAD SHX Text
GENERAL NOTES
AutoCAD SHX Text 1'-2"
AutoCAD SHX Text 1'-2"
AutoCAD SHX Text 2'-4"
AutoCAD SHX Text 2'-4"
AutoCAD SHX Text 1'-0"
AutoCAD SHX Text 1'-0"
AutoCAD SHX Text 1'-0"
AutoCAD SHX Text 1'-0"
AutoCAD SHX Text 2'-0 "12"
AutoCAD SHX Text 2'-0 "12"
AutoCAD SHX Text 10'-7 "12"
AutoCAD SHX Text 10'-7 "12"
AutoCAD SHX Text 12'-8"
AutoCAD SHX Text 12'-8"
AutoCAD SHX Text 17'-8"
AutoCAD SHX Text 17'-8"
AutoCAD SHX Text 1'-4"
AutoCAD SHX Text 1'-4"
AutoCAD SHX Text 0'-8 "34"
AutoCAD SHX Text 0'-8 "34"
AutoCAD SHX Text 17'-5"
AutoCAD SHX Text 17'-5"
AutoCAD SHX Text 1'-2 "12"
AutoCAD SHX Text 1'-2 "12"
AutoCAD SHX Text 7'-10 "12"
AutoCAD SHX Text 7'-10 "12"
AutoCAD SHX Text 3'-5 "18"
AutoCAD SHX Text 3'-5 "18"
AutoCAD SHX Text 8'-5 "12"
AutoCAD SHX Text 8'-5 "12"
AutoCAD SHX Text 0'-8"
AutoCAD SHX Text 0'-8"
AutoCAD SHX Text 1'-2 "12"
AutoCAD SHX Text 1'-2 "12"
AutoCAD SHX Text 1'-4 "12"
AutoCAD SHX Text 1'-4 "12"
AutoCAD SHX Text 11'-9 "34"
AutoCAD SHX Text 11'-9 "34"
AutoCAD SHX Text 1'-5 "12"
AutoCAD SHX Text 1'-5 "12"
AutoCAD SHX Text 13'-4"
13'-4"
FLOOR PLAN EQUIPMENT INSTALLED BY CSH
AutoCAD SHX Text
CHECKED BY:
AutoCAD SHX Text
PROJECT MGR:
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C1 AND SHOULD NOT BE SEPARATED
AutoCAD SHX Text
PROJECT:
AutoCAD SHX Text B
AutoCAD SHX Text
TITLE:
AutoCAD SHX Text
OF
AutoCAD SHX Text
SHT.
AutoCAD SHX Text
REV.
AutoCAD SHX Text
CUSTOMER SIGNATURE:
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
SIZE:
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
SHEET:
AutoCAD SHX Text
SEE DRAWING SCALE
AutoCAD SHX Text
NO/MS
AutoCAD SHX Text 08.14.2023
AutoCAD SHX Text
RA
AutoCAD SHX Text
NO/MS
AutoCAD SHX Text 08.14.2023
AutoCAD SHX Text
EVOLUTION SYSTEM
AutoCAD SHX Text
TYPICAL ROOM LAYOUT
AutoCAD SHX Text
WALL STAND-2
AutoCAD SHX Text
TABLE-2
AutoCAD SHX Text S3
AutoCAD SHX Text T-1
AutoCAD SHX Text
DRX-EVOLUTION
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
ITEM
AutoCAD SHX Text
COMPONENT DESCRIPTION: SUPPLIED AND INSTALLED BY CARESTREAM HEALTH INC.
AutoCAD SHX Text
STRUCTURAL LEGEND
AutoCAD SHX Text FLOOR MUST BE FLAT AND LEVEL WITHIN 0.125 INCH (3.175MM) IN BOTH DIRECTIONS OVER THE ENTIRE EQUIPMENT AREA. FOR SYSTEMS WITH THE OPTIONAL WALL STAND RAIL, THE FLOOR MUST BE LEVEL AND PARALLEL TO THE CEILING SUPPORT STRUCTURE TO WITHIN 0.2 INCH (5.0MM) OVER THE LENGTH OF THE RAIL. FLOORING MUST BE CAPABLE OF SUPPORTING THE OPERATING WEIGHT OF THE EQUIPMENT.
AutoCAD SHX Text
STRUCTURAL NOTES
AutoCAD SHX Text
TABLE BASE PLATE
AutoCAD SHX Text
GENERATOR
AutoCAD SHX Text
WALL STAND FLOOR RAIL
AutoCAD SHX Text 16'-0"
AutoCAD SHX Text 16'-0"
AutoCAD SHX Text 3'-0"
AutoCAD SHX Text 3'-0"
AutoCAD SHX Text 1'-1 "78"
AutoCAD SHX Text 1'-1 "78"
AutoCAD SHX Text 2'-1 "34"
AutoCAD SHX Text 2'-1 "34"
AutoCAD SHX Text 4'-10 "12"
AutoCAD SHX Text 4'-10 "12"
AutoCAD SHX Text 3'-6 "14"
AutoCAD SHX Text 3'-6 "14"
AutoCAD SHX Text 11'-6"
AutoCAD SHX Text 11'-6"
AutoCAD SHX Text CRITICAL TO FUNCTION FOR ITEMS 1 & 2 FLOOR UNDER WALL STAND FLOOR RAIL AND TABLE BASE PLATE TO BE FLAT AND LEVEL WITH A TOTAL TOLERANCE OF "18"
AutoCAD SHX Text 0'-6 "34"
0'-6 "34"
| Sheets and Views |
| EVOLUTION TYPICAL_WS2 RAIL_TBL2_08-14-2023-C1 |
| EVOLUTION TYPICAL_WS2 RAIL_TBL2_08-14-2023-A1 |
| EVOLUTION TYPICAL_WS2 RAIL_TBL2_08-14-2023-E1 |
| EVOLUTION TYPICAL_WS2 RAIL_TBL2_08-14-2023-E2 |
| EVOLUTION TYPICAL_WS2 RAIL_TBL2_08-14-2023-E3 |
| EVOLUTION TYPICAL_WS2 RAIL_TBL2_08-14-2023-S1 |
| EVOLUTION TYPICAL_WS2 RAIL_TBL2_08-14-2023-S2 |
| EVOLUTION TYPICAL_WS2 RAIL_TBL2_08-14-2023-S3 |
File details come from the government source that posted it. Updated .