Q2022-304 VA Fresno Cath Lab rev03.pdf
PDF 5 MB Posted
- Attached to
- Z2DB--Renovate Imaging and Catheterization Lab 570 21 220 Fresno VA Medical Center Fresno
- Solicitation number
- 36C26126R0045
About this file
This document is a technical drawing package for a CleanSuite modular operating room installation at VA Fresno Cath Lab in Fresno, California, prepared by BASX Solutions.
The drawing package encompasses six coordinated design drawings (M1.0 through M6.0) detailing the complete installation specifications. M1.0 provides general notes and room layout overview, establishing that all work must comply with the 2018 International Building Code as amended by California. Structural steel materials include ASTM A500 Grade B hollow structural sections (46 KSI yield strength), ASTM A653 sheet metal (36 KSI), and ASTM A53 pipe, with all connections utilizing Grade 5 bolts and FCAW-G welding per AWS standards. The contractor must verify all dimensions prior to work commencement and notify the engineer of discrepancies. M2.0 details module layout with dimensions showing supply air velocity of 907 FPM and access panel configurations. M3.0 specifies equipment layout with three STERIS EMS mounting plates positioned at 14 inches from the finished ceiling. M4.0 provides structural layout and suspension point details, indicating gravity and seismic support holes (.688" and .563" diameter respectively) with a system tolerance of ±0.125". M5.0 covers grid assembly and electrical requirements, specifying 28 Nortek air diffusers (distributed as 10 Type A, 10 Type B, 2 Type E, 2 Type F, 2 Type G, and 2 Type H units) with a system delivering 2,700 CFM at 22.30 air changes per hour across 819 square feet. M6.0 documents lighting calculations showing table illuminance of 515 fc maximum and 479 fc minimum, with an estimated perimeter circuit draw of 1.545A at 277VAC and interior circuit draw of 2.25A at 277VAC, using clear-diffused LED lenses on two independent circuits with 0-10V dimming control.
View the file
Other files
Show 40 more
Latest 50 of 51 listed here. See the rest on GovTribe.
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Clea / Table
10'-11"
SP
L e v e l o f fl o o r
1 _
(2 m m p e r
(1 m w it h in t h is a re a
ATY
N1N1N2
N1
L e v e l o f fl o o r
1 _
(2 m m p e r
(1 m w it h in th is a re a x
N x
ISO X
ISO Y
N
CLEANSUITE
DRAWING LIST
THIS DRAWING PACKAGE, SPECIFICATIONS, AND CONCEPTS CONTAINED HEREIN ARE THE SOLE PROPERTY OF BASX SOLUTIONS, AND MAY NOT BE REPRODUCED OR USED IN ANY FASHION WITHOUT THE PRIOR WRITTEN PERMISSION OF BASX SOLUTIONS.
DRAWING NO. 1 GENERAL NOTESM1.0
VA FRESNO CATH LAB
DATE 04/02/2026
EQUIPMENT LAYOUT
MODULE LAYOUTDRAWING NO. 2
DRAWING NO. 3
M2.0
M3.0
DRAWING NO. 4
DRAWING NO. 5
DRAWING NO. 6
STRUCTURAL LAYOUTM4.0
GRID ASSEMBLY & ELECTRICAL
LIGHTING CALCULATIONS
M5.0
M6.0
GENERAL NOTES
1. THE CONTRACTOR SHALL VERIFY ALL DIMENSIONS PRIOR TO
STARTING WORK. THE ENGINEER SHALL BE NOTIFIED OF ANY
DISCREPANCIES.
2. THE CONTRACTOR SHALL PROVIDE AND MAINTAIN ADEQUATE
ERECTION SHORING AND BRACING AS REQUIRED FOR
STABILITY OF ALL STRUCTURES DURING ALL PHASES OF
CONSTRUCTION.
3. IN NO CASE SHALL WORKING DRAWINGS BE SCALED FROM
PLANS, SECTIONS OR DETAILS ON THE DRAWINGS.
4. THE CONTRACT STRUCTURAL DRAWINGS REPRESENT THE
FINISHED STRUCTURES. THEY DO NOT INDICATE METHODS OF
CONSTRUCTION. THE CONTRACTOR SHALL PROVIDE ALL
MEASURES NECESSARY TO PROTECT EXISTING AND NEW
STRUCTURES DURING CONSTRUCTION. SUCH MEASURES
SHALL INCLUDE, BUT ARE NOT LIMITED TO, BRACING,
SHORING, ETC, FOR ALL CONSTRUCTION PHASE LOADS.
5. ALL WORK SHALL BE PREFORMED IN COMPLIANCE WITH 2018
INTERNATIONAL BUILDING CODE (IBC) AS AMENDED BY THE
STATE OF CALIFORNIA.
6. SEE STRUCTURAL DRAWINGS FOR CLEANSUITE SUPPORT
DETAILS
STRUCTURAL STEEL
1. CODES AND STANDARDS COMPLY WITH:
1.A. AISC "SPECIFICATION FOR STRUCTURAL STEEL
BUILDINGS & COMMENTARY".
1.B. AISC "CODE OF STANDARD PRACTICE" EXCLUDING SECTIONS
7.5.4 AND 7.11.5.
1.C. AWS "STRUCTURAL WELDING CODE", EXCLUDING ITEMS
CONFLICTING WITH AISC.
2. MATERIAL SHALL CONFORM AS FOLLOWS:
2.A. TUBE STEEL HOLLOW STRUCTURAL SECTION:
ASTM A500 FY= 46KSI, GRADE B.
2.B. SHEET METAL: ASTM A653 FY= 36KSI
2.C. PIPE:ASTM A53, FY=36 KSI
2.D. BOLTS: (GRADE 5 EQUIVALENT)
2.E. STRUCTURAL STEEL SHALL BE FABRICATED AND ERECTED IN
ACCORDANCE WITH THE LATEST EDITION OF AISC
"BUILDINGS"(ANSI/AISC 360) AND THE LATEST EDITION OF
AISC (ANSI/AISC 341), AS "PROVISIONS FOR STRUCTURAL
STEEL BUILDINGS
3. CONSTRUCTION:
3.A. FABRICATED IN AN APPROVED FABRICATORS SHOP.
3.B. FABRICATED BEAMS WITH INCIDENTAL CAMBER UP, UNO.
3.C. ALL STEEL NOT ENCASED IN CONCRETE, MASONRY, OR
FIREPROOFING SHALL BE SHOP PRIMED AND PAINTED PER
SPECIFICATION.
3.D. ANY ABRASIONS OR UNPAINTED AREAS SHALL BE TOUCHED UP
AFTER ERECTION.
3.E. SUBMIT SHOP DRAWINGS TO ARCHITECT (STRUCTURAL ENGINEER)
FOR REVIEW AND, UPON REQUEST, TO GOVERNING CODE
AUTHORITY. INDICATE AN ERECTION SEQUENCE OF WELDING TO
MINIMIZE LOCKED UP STRESSED OR DISTORTIONS FOR MOMENT
RESISTING STEEL FRAMES.
4. WELDED CONECTIONS:
4.A. WELD MATERIAL:
4.A.a TENSILE STRENGTH: 70 KSI 4.A.b YIELD STRENGTH: 58 KSI AS WELDED WITH 75/.25% 4.A.c WELD TYPE: FCAW-G FILL METAL E71T-1M/T-9M
5. BOLTED CONNECTIONS:
TOP, BOTTOM AND CORNERS OF CLEANSUITE CONNECTIONS:
5.A.a GRADE 5 EQUIVALENT DIA 5/8" BOLT, 24.00" OC.
5.A.b USE A HARDENED WASHER BENEATH THE TURNED ELEMENT OF THE
BOLT OR NUT. HARDENED WASHER TO BE AT LEAST 5/16" THICK
CONFORMING TO ASTM F436.
5.A.c HOLES HAVE A STANDARD OVERSIZE CALL OUT UNLESS OTHERWISE
NOTED.
5.A.d TIGHTEN BOLTS SNUG UNTIL ALL PLIES OF THE JOINT ARE IN FIRM
CONTACT.
5.A.e ENLARGE BOLT HOLES BY REAMING IF NECESSARY. DO NOT TORCH
CUT.
STRUCTURAL STEEL CONT'D
FULL SIZE DRAWING IS 30"x42"
DO NOT SCALE DRAWING
WORK ID NO.
DRAWING NO.
DRAWN
CHECKED
CROSSCHECK
SAFETY
PLOT SCALE DRAWING SCALE
JOB NUMBER
OLD FILE NAMEEQT CODE
PLANNING
ENGINEERING
APPROVED
APPROVALS DATE
AS BUILT DATEDATEAPPROVEDREV. NO. REVISION DESCRIPTION
1 : 1
THIS DRAWING IS THE PROPERTY OF
BASX AND CAN NOT BE
REPRODUCED IN WHOLE OR IN PART,
NOR DELIVERED TO OTHERS WITHOUT
THE EXPRESS WRITTEN PERMISSION
OF BASX iOR.
STAMP:
3500 SW 21ST PLACE
REDMOND, OREGON 97756
Phone (541) 647-6650 www.basxsolutions.com
4/9/2026 10:32:19 AM
As indicated
M1.0
Copyright
VA FRESNO CATH LAB
VA CENTRAL CALIFORNIA
FRESNO, CA
Q2022-304
GENERAL NOTES
3/8" = 1'-0"M1.0
1 ROOM LAYOUT
M1.0
2 OVERVIEW
M2.0
ISO X
ISO Y
M-1
INDICATES ISO CENTER
INDICATES AIR FLOW
SUPPLY AIR VELOCITY = 907 FPM
M-2
M-3
West East
125"
76.35"
SUPPLY
AIR
OPENING
C
G
South
North
M-4
M-5 M-6
48.65"
.3
.3
.3
.3
.6
STANDARD ACCESS COVER
STANDARD ACCESS OPENING
14"
NOTES:
1. ACCESS PANEL MAY BE MODIFIED
2. ACCESS PANELS TO BOOM ALCOVES CAN BE FIELD-MODIFIED BY OTHERS
TO ALLOW FOR EQUIPMENT, ELECTRICAL AND/OR MED GAS
CLEARANCES
3. SIZE OF ACCESS PANELS TO BOOM ALCOVES ARE MAXIMIZED
.2
10.25"
FINISH CEILING
106.31250
ISO Y WestEast
48.65" 76.35"
NOTCH 12"
118.31250
NOTCH 8"
114.31250
.7
.2
36"8.5"
FINISH CEILING
106.312505
.7
CGSouth North
79.675"86"64.3"
NOTCH 12"
118.31250
NOTCH 8"
114.31250
36" 4.75"96.5" 48"
FINISH CEILING
106.31250
ISO YWest East
48.65"76.35"
NOTCH 12"
118.31250
NOTCH 8"
114.312501
.6
FINISH CEILING
106.31250
C G SouthNorth
64.3"86"79.675"
NOTCH 12"
118.31250
NOTCH 8"
114.31250
1 12ga PLENUM SIDEWALL
2 PERIMETER GRID
3 HSS4x2x1/4
NOTE:
USE GMAW-S WITH ER70S-6 WIRE
AND 90/10 SHIELDING GAS FOR
CONNECTING PLENUM WALL
(A653) TO HSS (A500)
.125
.125
TYP
1/2" = 1'-0"M2.0
5 MODULE LAYOUT
1" = 1'-0"M2.0 M2.0
6 TYPICAL ACCESS PANEL
FULL SIZE DRAWING IS 30"x42"
DO NOT SCALE DRAWING
WORK ID NO.
DRAWING NO.
DRAWN
CHECKED
CROSSCHECK
SAFETY
PLOT SCALE DRAWING SCALE
JOB NUMBER
OLD FILE NAMEEQT CODE
PLANNING
ENGINEERING
APPROVED
APPROVALS DATE
AS BUILT DATEDATEAPPROVEDREV. NO. REVISION DESCRIPTION
1 : 1
THIS DRAWING IS THE PROPERTY OF
BASX AND CAN NOT BE
REPRODUCED IN WHOLE OR IN PART,
NOR DELIVERED TO OTHERS WITHOUT
THE EXPRESS WRITTEN PERMISSION
OF BASX iOR.
STAMP:
3500 SW 21ST PLACE
REDMOND, OREGON 97756
Phone (541) 647-6650 www.basxsolutions.com
4/9/2026 10:32:28 AM
As indicated
M2.0
VA FRESNO CATH LAB
VA CENTRAL CALIFORNIA
FRESNO, CA
Q2022-304
MODULE LAYOUT
1/2" = 1'-0"M2.0
1 NORTH ELEVATION
1/2" = 1'-0"M2.0
2 EAST ELEVATION
1/2" = 1'-0"M2.0
3 SOUTH ELEVATION
1/2" = 1'-0"M2.0
4 WEST ELEVATION
6" = 1'-0"M2.0
TUBE TO SIDEWALL
ATTACHMENT
M2.0
7 MODULE ASSEMBLY
ISO X
ISO Y
INDICATES ISO CENTER
73"
23.5625" 23.5625"
8"8"
24"
92.375" 78"
92.375"
.1
.8
.9
1"28"
1/2" = 1'-0"M3.0
2 EQUIPMENT LAYOUT
FULL SIZE DRAWING IS 30"x42"
DO NOT SCALE DRAWING
WORK ID NO.
DRAWING NO.
DRAWN
CHECKED
CROSSCHECK
SAFETY
PLOT SCALE DRAWING SCALE
JOB NUMBER
OLD FILE NAMEEQT CODE
PLANNING
ENGINEERING
APPROVED
APPROVALS DATE
AS BUILT DATEDATEAPPROVEDREV. NO. REVISION DESCRIPTION
1 : 1
THIS DRAWING IS THE PROPERTY OF
BASX AND CAN NOT BE
REPRODUCED IN WHOLE OR IN PART,
NOR DELIVERED TO OTHERS WITHOUT
THE EXPRESS WRITTEN PERMISSION
OF BASX iOR.
STAMP:
3500 SW 21ST PLACE
REDMOND, OREGON 97756
Phone (541) 647-6650 www.basxsolutions.com
4/9/2026 10:32:35 AM
As indicated
M3.0
VA FRESNO CATH LAB
VA CENTRAL CALIFORNIA
FRESNO, CA
Q2022-304
EQUIPMENT LAYOUT
M3.0
1 CLEANSUITE
EQUIPMENT/BOOM PLATE SCHEDULE
Count Mark Manufacturer Description
Plate Mounting Height (From Finished Ceiling to
Bottom of Plate) Size of Beauty Ring (When applicable)
3 STERIS EMS MOUNTING PLATE 14.00000
Grand total: 3
1" = 1'-0"M3.0
STERIS EMS MOUNTING
PLATE
ISO X
ISO Y
TYP
West East
C
G
South
North
P4100
KEY NOTE
INDICATES GRAVITY SUSPENSION PONT (.688" DIA HOLE)
INDICATES SEISMIC KICKER LOCATIONS (.563" DIA HOLE)
INDICATES ISO CENTER
NOTES:
1. GRAVITY SUPPORT AND SEISMIC SUPPORT PROVIDED BY OTHERS.
SEE PLAN AND KEY NOTES FOR CONNECTION HOLE DIAMETER AND LOCATIONS
2. SEISMIC KICKER HOLES PROVIDED UPON REQUEST TO EACH INDIVIDUAL MODULE CORNER
3. OVERALL SYSTEM DIMENSIONS HAVE A TOLERANCE OF +/- 0.125"
.8
.1
.9
.7
.2
.2
.1
.0
.9
74.14" 1.105" 1.105" 46.44"
107.375" 63" 30"
107.375"
71.14" 2.605" 2.605" 43.44"
FINISH CEILING
106.31250
NOTCH 12"
118.31250NOTCH 8"
114.31250
NOTE:
REFER TO STRUCTURAL CALCS FOR
ADDITIONAL INFORMATION
1. SUSPENSION HOLE
2. SEISMIC KICKER HOLE
.5
1"
6.5"
SEISMIC NON-SEISMIC
0.06" 3.25"
.2
.0
0.06" 3.25"
.2
.0
.5
1" 6.5" 1"
1 HSS TUBE STEEL
2 IMBEDED P5500 UNISTRUT ASSEMBLY FOR IMAGING EQUIPMENT RAIL
3 GRID LINE
4 HSS AT MODULE SPLIT
5 3/8" DIA BOLT W/ 1/2" PL WASHER
6 INDUSTRY SPECIFIC GRID EXTRUSION FLUSH WITH FINISHED CEILING
7 IMAGING EQUIPMENT RAIL BY OTHERS
STANDARD STRUT DETAIL
STRUCTURAL STRUT AT MODULE SPLIT DETAIL REQUIRED UNISTRUT HEIGHT IS MEASURED FROM
FINISHED FLOOR TO BOTTOM OF UNISTRUT.
SEE EQUIPMENT PACKAGE
.3
.6
0.08" 0.67"
0.125"
2.25"
0.125"
0.67" 0.08"
.9
.6
.0
2.25"
1 4
1/2" = 1'-0"M4.0
STRUCTURAL LAYOUT /
SUSPENSION POINTS
FULL SIZE DRAWING IS 30"x42"
DO NOT SCALE DRAWING
WORK ID NO.
DRAWING NO.
DRAWN
CHECKED
CROSSCHECK
SAFETY
PLOT SCALE DRAWING SCALE
JOB NUMBER
OLD FILE NAMEEQT CODE
PLANNING
ENGINEERING
APPROVED
APPROVALS DATE
AS BUILT DATEDATEAPPROVEDREV. NO. REVISION DESCRIPTION
1 : 1
THIS DRAWING IS THE PROPERTY OF
BASX AND CAN NOT BE
REPRODUCED IN WHOLE OR IN PART,
NOR DELIVERED TO OTHERS WITHOUT
THE EXPRESS WRITTEN PERMISSION
OF BASX iOR.
STAMP:
3500 SW 21ST PLACE
REDMOND, OREGON 97756
Phone (541) 647-6650 www.basxsolutions.com
4/9/2026 10:32:44 AM
As indicated
M4.0
VA FRESNO CATH LAB
VA CENTRAL CALIFORNIA
FRESNO, CA
Q2022-304
STRUCTURAL LAYOUT
1/2" = 1'-0"M4.0
MODULE TO MODULE
SECTION
M4.0
CLEANSUITE - STRUCTURAL
ASSEMBLY
3" = 1'-0"M4.0
TRUSS CORNER SUSPENSION
POINT DETAIL
3" = 1'-0"M4.0
GEN 1 PLENUM CORNER
SUSPENSION POINT DETAIL
1 1/2" = 1'-0"M4.0
4 HYBRID RAIL HANGER DETAIL
SYSTEM SIZING
A: 819 SF
H: 8'-10.3125"
ACH: 22.30
CFM: 2700
DIFA: 91.41 SF
VELOCITY: 29.54 FPM
A
94.7"
B B A
A B B A
E F F E
A B B A
A B B A
A B B A
G H H G
27.1"
2.5"
17.75"17.75"
2.5"
27.1"29"
.3
.3
.3
76.35"
.3
48.65"
.6
6.5
6 3.5
6.5
5 5
.5
1.5
1.5 1.5
1.5
1.51.5
1.51.5
1.5
1.5 1.5
1.5 1.5
1.5 1.5
6 3.5
6 3.5
1.5
1.5
1.5
1.5
1.5
1.5
1.5
1.5
1.5114
.5
ELECTRICAL
CONNECTION LOCATION
NOTE:
REFER TO ELECTRICAL DRAWINGS FOR ADDITIONAL INFORMATION
NOTE:
1. LED LENS IS CLEAR-DIFFUSED
2. LIGHTING ON (2) CIRCUITS (PERIMETER AND INTERIOR)
(1) 120/277VAC POWER
(1) 0-10V LED DIMMING CONTROL
3. LOW VOLTAGE AND ELECTRICAL WIRING FOR BOOMS ARE WIRED DIRECTLY TO
BOOM AND PROVIDED BY OTHERS
4. ESTIMATED (FLA) CIRCUIT VALUES:
ESTIMATED PERIMETER FLA = 1.545A @ 277VAC
ESTIMATED INTERIOR FLA = 2.25A @ 277VAC
FULL SIZE DRAWING IS 30"x42"
DO NOT SCALE DRAWING
WORK ID NO.
DRAWING NO.
DRAWN
CHECKED
CROSSCHECK
SAFETY
PLOT SCALE DRAWING SCALE
JOB NUMBER
OLD FILE NAMEEQT CODE
PLANNING
ENGINEERING
APPROVED
APPROVALS DATE
AS BUILT DATEDATEAPPROVEDREV. NO. REVISION DESCRIPTION
1 : 1
THIS DRAWING IS THE PROPERTY OF
BASX AND CAN NOT BE
REPRODUCED IN WHOLE OR IN PART,
NOR DELIVERED TO OTHERS WITHOUT
THE EXPRESS WRITTEN PERMISSION
OF BASX iOR.
STAMP:
3500 SW 21ST PLACE
REDMOND, OREGON 97756
Phone (541) 647-6650 www.basxsolutions.com
4/9/2026 10:32:46 AM
1/2" = 1'-0"
M5.0
VA FRESNO CATH LAB
VA CENTRAL CALIFORNIA
FRESNO, CA
Q2022-304
GRID ASSEMBLY & ELEC
1/2" = 1'-0"M5.0
1 GRID LAYOUT
1/2" = 1'-0"M5.0
2 LIGHTING LAYOUT
AIR DIFFUSER SCHEDULE
Count Type Mark Manufacturer Description Comments
10 A NORTEK GUILLOTINE EQUALIZER / CLEANSCREEN
10 B NORTEK GUILLOTINE EQUALIZER / CLEANSCREEN
2 E NORTEK GUILLOTINE EQUALIZER / CLEANSCREEN
2 F NORTEK CLEANSCREEN
2 G NORTEK GUILLOTINE EQUALIZER / CLEANSCREEN
2 H NORTEK GUILLOTINE EQUALIZER / CLEANSCREEN
Grand total: 28
M5.0
3 GRID ASSEMBLY
GENERAL NOTES
1. THESE RESULTS ARE ESTIMATED.
2. CALCULATION GRID HEIGHT ABOVE THE FLOOR: 3 FT (STANDARD TABLE HEIGHT)
3. EXTERNAL LIGHTING THAT IS OUTSIDE THE iOR FOOTPRINT IS NOT INCLUDED.
4. SHARED LIGHT SUCH AS BOOM LIGHTS OR TASK LIGHTING IS NOT INCLUDED.
5. WALL LOCATIONS ARE APROXIMATE.
6. LIGHT LOSS FACTOR: .78 FOR 1FT WHITE LED UNLESS NOTED OTHERWISE. SEE LIGHTING FIXTURE SCHEDULE
7. ONLY DIRECT LIGHT IS CALCULATED.
8. SEE SHEET M2.0 ELEVATIONS FOR CEILING HEIGHT.
CALCULATIONS HAVE BEEN PERFORMED ACCORDING TO IES STANDARDS AND GOOD PRACTICE,
SOME DIFFERENCES BETWEEN MEASURED VALUES AND CALCULATED RESULTS MAY OCCUR DUE TO
TOLERANCES IN CALCULATION METHODS, TESTING PROCEDURES, COMPONENT PERFORMANCE,
MEASUREMENT TECHNIQUES AND FIELD CONDITIONS SUCH AS VOLTAGE AND TEMPERATURE
VARIATIONS. INPUT DATA USED TO GENERATE THE ATTACHED CALCULATIONS SUCH AS ROOM
DIMENSIONS, REFLECTANCES, FURNITURE AND ARCHITECTURAL ELEMENTS SIGNIFICANTLY AFFECT THE LIGHTING CALCULATIONS. IF THE REAL ENVIRONMENT CONDITIONS DO NOT MATCH THE INPUT
DATA, DIFFERENCES WILL OCCUR BETWEEN MEASURED VALUES AND THE CALCULATED VALUES.
DISCLAIMER
LIGHTING FIXTURE SCHEDULE
Manufacturer Model Total Light Loss Factor OmniClass Number OmniClass Title BASX 0.78 23.80.70.11 Luminaries for Internal Lighting
FULL SIZE DRAWING IS 30"x42"
DO NOT SCALE DRAWING
WORK ID NO.
DRAWING NO.
DRAWN
CHECKED
CROSSCHECK
SAFETY
PLOT SCALE DRAWING SCALE
JOB NUMBER
OLD FILE NAMEEQT CODE
PLANNING
ENGINEERING
APPROVED
APPROVALS DATE
AS BUILT DATEDATEAPPROVEDREV. NO. REVISION DESCRIPTION
1 : 1
THIS DRAWING IS THE PROPERTY OF
BASX AND CAN NOT BE
REPRODUCED IN WHOLE OR IN PART,
NOR DELIVERED TO OTHERS WITHOUT
THE EXPRESS WRITTEN PERMISSION
OF BASX iOR.
STAMP:
3500 SW 21ST PLACE
REDMOND, OREGON 97756
Phone (541) 647-6650 www.basxsolutions.com
4/9/2026 10:32:47 AM
As indicated
M6.0
VA FRESNO CATH LAB
VA CENTRAL CALIFORNIA
FRESNO, CA
Q2022-304
LIGHTING CALCULATIONS1/2" = 1'-0"M6.0
1 LIGHTING CALCULATION
ElumTools General Use Illuminance Results
Project Name Project Number Calculation Points Name Maximum Minimum VA FRESNO CATH LAB Q2022-304 Table 515 fc 479 fc
VA FRESNO CATH LAB Q2022-304 Hybrid 1 416 fc 26 fc
| Sheets |
| M1.0 - GENERAL NOTES |
| M2.0 - MODULE LAYOUT |
| M3.0 - EQUIPMENT LAYOUT |
| M4.0 - STRUCTURAL LAYOUT |
| M5.0 - GRID ASSEMBLY & ELEC |
| M6.0 - LIGHTING CALCULATIONS |