This is a Data Item Description (DID) that establishes format and content requirements for calibration certificates and reports accompanying calibrations performed on Department of Defense test, measurement, and diagnostic equipment. DI-QCIC-80798C, approved January 23, 2017, supersedes the previous version DI-QCIC-80798B and applies to contracts requiring calibration of test equipment prior to delivery. The DID aligns with ISO/IEC 17025 standards for general requirements for the competence of testing and calibration laboratories.

The calibration certificate or report must include fourteen mandatory content elements: title; laboratory name, address, and calibration location if different; unique identification and page identification; client or customer name and address; identification of the method used; description and unambiguous identification of calibrated items; dates of receipt and performance; reference to sampling plans and procedures; test or calibration results with units of measurement; names, functions, and signatures of authorizing personnel; a statement limiting results to items tested or calibrated where relevant; environmental or operational conditions affecting results; measurement uncertainty or statement of compliance with metrological specifications; and evidence of measurement traceability. Contractor format is acceptable, and the applicable document versions shall be specified in the contract.

View the file

Other files

File Type Posted
1 - AFMC_260_First_Article_Requirements.pdf PDF
8 - DID_Certification_Data_for_Non_Nuclear_Level_I.pdf PDF
9 - DID_Computer_Software_Product_End_Items.pdf PDF
10 - DID_Conference_Agenda.pdf PDF
16 - DID_Government_Property_Inventory_Report.pdf PDF
21 - DID_Request_for_Variance.pdf PDF
31 - Personnel_and_Skill_Requirement.pdf PDF
Exhibit A - CDRLs.pdf PDF
2 - AFMC_158_Packaging_Requirements.pdf PDF
7 - DID_Certificate_of_Compliance.pdf PDF
22 - DID_Status_Report.pdf PDF
28 - Item_Description_Antenna.pdf PDF
3 - DD_Form_1653_Transportation_Data_for_Solicitations.pdf PDF
4 - DD_Form_2169_Special_Packaging_Instructions.pdf PDF
12 - DID_Contract_Depot_Maintenance_Production_Report.pdf PDF
13 - DID_Contractors_Corrective_Action_Plan.pdf PDF
25 - IUID_Checklist.pdf PDF
26 - PWS_Updated_26_Jan_23.pdf PDF
29 - DD254_Security_Classification.pdf PDF
32 - Pricing_Matrix.xlsx XLSX spreadsheet
FA8119-23-R-0011 P00008.pdf PDF
6 - DID_CAV_Reporting.pdf PDF
11 - DID_Conference_Minutes.pdf PDF
14 - DID_Counterfeit_Prevention_Plan.pdf PDF
15 - DID_Engineering_Change_Proposal.pdf PDF
17 - DID_Presentation_Material.pdf PDF
18 - DID_Product_Engineering_Design_Data_and_Associated_Lists.pdf PDF
19 - DID_Production_Surge_Plan.pdf PDF
20 - DID_Report_of_Shipping_Item_and_Packaging_Discrepancy.pdf PDF
23 - DID_Test_Inspection_Report.pdf PDF
24 - GFP.pdf PDF
26_PWS_Updated_20_March_23.pdf PDF
27 - Item_Description_Radome.pdf PDF
30 - Justification_for_Qualification_Requirements.pdf PDF
33 - Salient_Characteristics_Radome.pdf PDF
Show all 35

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

DATA ITEM DESCRIPTION

Title: CALIBRATION CERTIFICATE/REPORT

Number: DI-QCIC-80798C Approved Date: 20170123 AMSC Number: F9766 Limitation: N/A DTIC Applicable: No GIDEP Applicable: No Preparing Activity: 36 Project Number: QCIC-2017-003 Applicable Forms: N/A Use/Relationship: The Calibration Certificate/Report documents calibration certificates or reports accompanying the calibrations performed on Department of Defense (DoD) test, measurement, and diagnostic equipment.

a. This Data Item Description (DID) contains the format and content preparation instructions for the data product generated by the specific and discrete task requirement as delineated in the contract.

b. This DID is applicable to contracts requiring calibration of test equipment prior to delivery.

This DID is consistent with ISO/IEC 17025, General requirements for the competence of testing and calibration laboratories, requirements for calibration reports or certificates.

(Copies of this document are available online at http://www.iso.org.)

c. This DID supersedes DI-QCIC-80798B.

Requirements:

1. Reference documents. The applicable issue of the documents cited herein, including their approval dates and dates of any applicable amendments, notices, and revisions, shall be as specified in the contract.

2. Format. Contractor format is acceptable.

3. Content. The calibration certificate or report shall include:

a. Title (e.g., calibration certificate or calibration report);

b. The name and address of the laboratory and the location where the calibrations were carried out, if different from the address of the laboratory;

c. Unique identification of the calibration certificate or test report (such as the serial number), and on each page an identification in order to ensure that the page is recognized as a part of the test report or calibration certificate, and a clear identification of the end of the calibration certificate or test report;

d. The name and address of the client or customer;

e. Identification of the method used;

f. A description of, the condition of, and the unambiguous identification of the item(s) tested or calibrated;

g. The date of receipt of the test or calibration item(s) where this is critical to the validity and application of the results, and the date(s) of performance of the test or calibration;

DISTRIBUTION STATEMENT A. Approved for public release. Distribution is unlimited.

Source: https://assist.dla.mil -- Downloaded: 2017-03-13T13:39Z Check the source to verify that this is the current version before use.

DI-QCIC-80798C

h. Reference to the sampling plan and procedures used by the laboratory or other bodies where these are relevant to the validity or application of the results;

i. The test or calibration results with, where appropriate, the units of measurement;

j. The name(s), function(s) and signature(s) or equivalent identification of person(s) authorizing the calibration certificate or test report;

k. Where relevant, a statement to the effect that the results relate only to the items tested or calibrated;

l. The conditions (e.g., environmental) under which the calibrations were made that have an influence on the measurement results;

m. The uncertainty of measurement or a statement of compliance with an identified metrological specification or clauses thereof;

n. Evidence that the measurements are traceable.

End of DI-QCIC-80798C.