This document is a technical data sheet for the Samsung GC85A Smart Digital Radiography system, a ceiling-mounted X-ray imaging platform designed for general radiography applications.
The GC85A system comprises four primary components: an X-ray tube assembly available in two configurations (Type A with 600 kHU anode heat storage and Type B with 300 kHU), a high voltage generator available in four types (Types A-D) with maximum outputs ranging from 50-82 kW at 150 kVp and 630-1000 mA, a ceiling-mounted fully automated tube head unit with 12.1-inch touch screen display featuring 180 cm vertical travel range and 240-340 degree rotation capability, and multiple AccE detector options including wireless portable detectors (F4335-AW, S4335-AW, S4343-AW, S3025-AW) and standard detectors (S4335-AWM, S4343-AWM, S3025-AWM). Detector specifications vary by model, with pixel matrices ranging from 2536 x 2024 to 3040 x 3036 pixels, 16-bit A/D conversion, DQE between 60-76%, and wireless/tethered operation capabilities. The system integrates a 6-way patient table with motorized elevation (40 mm/sec speed), optional wall stand accessories (Types A and B) with motorized positioning, automatic collimation with 4-axis blade control and copper prefiltering, and smart stitching functionality for spine and long bone imaging with acquisition times under 15 seconds for multi-image exams. The imaging workstation S-Station features Intel Core Xeon processing, 8-16GB RAM, 256GB SSD with 1TB HDD storage, Windows 10 operating system, and comprehensive DICOM connectivity including verification, modality worklist, MPPS, storage, commitment, grayscale print, and query/retrieve functions. Installation requirements specify three-phase 380/400/480 VAC power at 50/60 Hz with 120 kVA rating, operating temperature range of 10-40°C, and relative humidity of 30-75%. Optional accessories include barcode scanners, patient compression belts, detector battery chargers, portable grids, quality assurance packages, and advanced software modules such as SimGrid, bone suppression, lung nodule detection, and wireless barcode support.
View the file
Other files
Show 40 more
Latest 50 of 92 listed here. See the rest on GovTribe.
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
SAMSUNG SMART
DIGITAL RADIOGRAPHY
* GC85A configuration may vary by needs.
ACCELERATING INTELLIGENCE
GC85A
GC85A Technical Data1 GC85A Technical Data 2
X-ray Tube Assembly
High Voltage Generator
GC85A
Type A Type B
Tube construction Rhenium-Tungsten faced Molybdenum Rhenium-Tungsten faced Molybdenum
Anode heat storage capacity 600 kHU (420 kJ) 300 kHU (210 kJ)
Target angle 12° 12°
Focal spot size 0.6 mm/1.2 mm 0.6 mm/1.2 mm
Permanent filtration 1.1 mm Al @75 kV 0.9 mm Al @75kV Anode Speed Minimum 3200 min-1 (60 Hz)
Minimum 2700 min-1 (50 Hz)
Type A Type B Type C Type D
Max. Output* 82 kW, 150 kVp, 1000 mA 52 kW, 150 kVp, 640 mA 80 kW, 150kVp, 1000 mA 50 kW, 150 kVp, 630 mA
Exposure Voltage 40 ~ 150 kV (1 kV step) 40 ~ 150 kV (1 kV step) 40 ~ 150 kV (1 kV step) 40 ~ 150 kV (1 kV step) Line voltage 380/400/480 VAC, 3-Phase, 50/60 Hz
380/400/480 VAC,
3-Phase, 50/60 Hz
380/400/480 VAC,
3-Phase, 50/60 Hz
380/400/480 VAC,
3-Phase, 50/60 Hz mA range 10 ~ 1000 mA 10 ~ 640 mA 10 ~ 1000 mA 10 ~ 630 mA
Time range 0.001 ~ 10 sec 0.001 ~ 10 sec 0.001 ~ 10 sec 0.001 ~ 10 sec mAs range 0.1~500 mAs 0.1~500 mAs 0.1~1000 mAs 0.1~1000 mAs
Generator frequency Max. 30 kHz Max. 30 kHz Max. 450 kHz Max. 450 kHz Generator cabinet dimensions (LxWxH)
620x405x630 (HVG), 620x450x432 (H/W)
620x405x630 (HVG), 620x450x432 (H/W)
596x410x480 (HVG), 620x450x432 (H/W)
596x410x480 (HVG), 620x450x432 (H/W)
* Select 1 type of 4 kinds of High Voltage Generator The configurations may vary by country
Handles, buttons, jog and touch screen are all in one layer. The THU integrates a 12.1 inch wide touch screen to increase user convenience and optimize procedure.
Ceiling-mounted Tube Head Unit
Type Fully automated Patient coverage along the longitudinal /transverse axis Max. Longitudinal/Transverse: 902.6cm/363.6cm(9 x 4 rail)
Vertical travel range 180 cm
Tube rotation around the horizontal axis 240° (±120°)
Tube rotation around the vertical axis 340° (-157° ~ +183°)
Vertical movement speed 100 mm/sec
Soft Handling(Motorized Power Assist System) Yes
Multi-functional display with color touch screen 12.1 inch Wide touch screen
- Automatic SID measurement, Display of tube rotation angle
- Automatic/Manual modification of kV, mAs and ms, Focal spot selection
- System standby, System movement check, Projection setting, Provision of screen information
- Image collection, Image preview, positioning help
- X-ray projection related information entry, Intra-radiology lab image check possible after X-ray photography
- Auto-repositioning setting, Smart Stitching function related information entry
Camera for patient monitoring and stitching Yes
GC85A
AccE Detector
Wireless Portable Detector
Detector model F4335-AW
Detector type Amorphous Silicon TFT / Cesium Iodide scintillator (CsI)
Dimensions 460 x 384 x 15 mm
Active detector matrix (Effective Pixel matrix) 3,040 × 2,466 pixels
Effective area 425.60 mm × 345.24 mm
Pixel pitch 140 ㎛
A/D conversion (Pixel depth) 16 bits
Spatial resolution 3.57 lp/mm
Detective Quantum Efficiency (DQE) 76% (0.0 lp/mm, Typical)
Modulation Transfer Function (MTF) 86% (0.5 lp/mm, Typical)
Installation Wireless / Tethered
Image Acquisition Tethered / Wireless (802.11 a/b/g/n & ac) @2.4/5 GHz
AED Available, v3.0
Weight Approx. 2kg (w/o Battery Set*) Max.load capacity 400 kg for uniform load, 200 kg for local load (40 mm in diameter disk at the center) Dust/Water-resistance IP54
Self diagnosis Automatic detection & alarming of line artifact
Shock logging Real-time detection, logging and alarming of shock level
Battery charging time Typ. 6 hrs. (tethered), Typ. 3 hrs. (battery charger) Battery Operation Time 500 images for 4 hrs.
Waiting for more than 10 hrs under deep sleep mode
*The Battery Set includes the battery and the battery cover
Wireless Portable Detector
Detector model S4335-AW
Detector type Amorphous Silicon TFT / Cesium Iodide scintillator (CsI)
Dimensions 460 x 384 x 15 mm
Active detector matrix (Effective Pixel matrix) 3,040 × 2,466 pixels
Effective area 425.6 mm × 345.24 mm
Pixel pitch 140 ㎛
A/D conversion (Pixel depth) 16 bits
Spatial resolution 3.57 lp/mm
Detective Quantum Efficiency (DQE) 76% (0.0 lp/mm, Typical)
Modulation Transfer Function (MTF) 86% (0.5 lp/mm, Typical)
Installation Wireless / Tethered
Image Acquisition Tethered / Wireless (802.11 a/b/g/n & ac) @2.4/5 GHz
AED Available, v3.0
Weight Approx. 3 kg Max.load capacity 400 kg for uniform load, 200 kg for local load (40 mm in diameter disk at the center) Max.bend capacity > 200 kg @ 3-point bending condition
Dust/Water-resistance IP54
Self diagnosis Automatic detection & alarming of line artifact
Detector shock logging* Real-time detection, logging and alarming of shock level
Battery charging time Typ. 6 hrs. (tethered), Typ. 3 hrs. (battery charger) Battery Operation Time 500 images for 4 hrs.
Waiting for more than 10 hrs under deep sleep mode
* This feature is only for F4335-AW, S4335-AW and S4343-AW detector
GC85A Technical Data3 GC85A Technical Data 4
GC85A GC85A
AccE Detector Wireless Portable Detector
Detector model S4343-AW
Detector type Amorphous Silicon TFT / Cesium Iodide scintillator (CsI)
Dimensions 460 x 460 x 15 mm
Active detector matrix (Effective Pixel matrix) 3,040 × 3,036 pixels
Effective area 425.60 mm × 425.04 mm
Pixel pitch 140 ㎛
A/D conversion (Pixel depth) 16 bits
Spatial resolution 3.57 lp/mm
Detective Quantum Efficiency (DQE) 76% (0.0 lp/mm, Typical)
Modulation Transfer Function (MTF) 86% (0.5 lp/mm, Typical)
Installation Wireless / Tethered
Image Acquisition Tethered / Wireless (802.11 a/b/g/n & ac) @2.4/5 GHz
AED Available, v3.0
Weight Approx. 3.6 kg Max.load capacity 400 kg for uniform load, 200 kg for local load (40 mm in diameter disk at the center) Max.bend capacity > 200 kg @ 3-point bending condition
Dust/Water-resistance IP54
Self diagnosis Automatic detection & alarming of line artifact
Detector shock logging* Real-time detection, logging and alarming of shock level
Battery charging time Typ. 6 hrs. (tethered), Typ. 3 hrs. (battery charger) Battery Operation Time 500 images for 4 hrs.
Waiting for more than 10 hrs under deep sleep mode
* This feature is only for F4335-AW, S4335-AW and S4343-AW detector
Wireless Portable Detector
Detector model S3025-AW
Detector type Amorphous Silicon TFT / Cesium Iodide scintillator (CsI)
Dimensions 350 x 287 x 15 mm
Active detector matrix 2536 x 2024 pixels (W x H)
Effective area 314.5 mm × 251.0 mm (W x H)
Pixel pitch 124 ㎛
A/D conversion (Pixel depth) 16 bits
Spatial resolution 4.0 lp/mm
Detective Quantum Efficiency (DQE) 75% (0 lp/mm, Typical)
Modulation Transfer Function (MTF) 87% (0.5 lp/mm, Typical)
Installation Wireless only
Image Acquisition Wireless (802.11 n/ac) @2.4/5 GHz
AED N/A
Weight 2.1 kg (Typical)
Max.load capacity 400 kg for uniform load, 200 kg for local load (40 mm in diameter disk at the center)
Dust/Water-resistance IP67
Self diagnosis Automatic detection & alarming of line artifact
Battery charging time 2 hrs. (AC-DC adaptor)
Battery Operation Time 270 images for 7.5 hrs
AccE Standard Detector Wireless Portable Detector
Detector model S4335-AWM
Detector type Amorphous Silicon TFT / Cesium Iodide scintillator (CsI)
Dimensions 460 x 384 x 15 mm
Active detector matrix (Effective Pixel matrix) 3,040 × 2,466 pixels
Effective area 425.6 mm × 345.24 mm
Pixel pitch 140 ㎛
A/D conversion (Pixel depth) 16 bits
Spatial resolution 3.57 lp/mm
Detective Quantum Efficiency (DQE) 65% (0.0 lp/mm, Typical)
Modulation Transfer Function (MTF) 88% (0.5 lp/mm, Typical)
Installation Wireless / Tethered
Image Acquisition Tethered / Wireless (802.11 a/b/g/n & ac) @2.4/5 GHz
AED Available, v3.0
Weight Approx. 3.2 kg Max.load capacity 400 kg for uniform load, 200 kg for local load (40 mm in diameter disk at the center) Max.bend capacity > 200 kg @ 3-point bending condition
Dust/Water-resistance IP54
Self diagnosis Automatic detection & alarming of line artifact
Battery charging time Typ. 6 hrs. (tethered), Typ. 3 hrs. (battery charger) Battery Operation Time 500 images for 4 hrs.
Waiting for more than 10 hrs under deep sleep mode
Wireless Portable Detector
Detector model S4343-AWM
Detector type Amorphous Silicon TFT / Cesium Iodide scintillator (CsI)
Dimensions 460 x 460 x 15 mm
Active detector matrix (Effective Pixel matrix) 3,040 × 3,036 pixels
Effective area 425.60 mm × 425.04 mm
Pixel pitch 140 ㎛
A/D conversion (Pixel depth) 16 bits
Spatial resolution 3.57 lp/mm
Detective Quantum Efficiency (DQE) 65% (0.0 lp/mm, Typical)
Modulation Transfer Function (MTF) 88% (0.5 lp/mm, Typical)
Installation Wireless / Tethered
Image Acquisition Tethered / Wireless (802.11 a/b/g/n & ac) @2.4/5 GHz
AED Available, v3.0
Weight Approx. 3.6 kg Max.load capacity 400 kg for uniform load, 200 kg for local load (40 mm in diameter disk at the center) Max.bend capacity > 200 kg @ 3-point bending condition
Dust/Water-resistance IP54
Self diagnosis Automatic detection & alarming of line artifact
Battery charging time Typ. 6 hrs. (tethered), Typ. 3 hrs. (battery charger) Battery Operation Time 500 images for 4 hrs.
Waiting for more than 10 hrs under deep sleep mode
GC85A Technical Data5 GC85A Technical Data 6
Automatic collimation
Blade control 4-axis motorized control
Copper prefilter 0.1 mm ; 0.2 mm ; 0.3 mm
Collimation control Manual or Automatic
Lamp LED lamp > 160lux
Collimation Size Fixed Maintain the user-selected Collimation size
Collimation Size Undo Return the previous collimation size in accordance with each protocol
Flexible and accurate function reduces radiation dose and improves functions.
Automatic Collimator with Each Blade Control
Patient table* 460 mm×460 mm, 85 lp/cm, 10:1, SID 110 cm, Carbon cover
460 mm×460 mm, 92 lp/cm, 10:1, SID 110 cm, Carbon cover
Wall stand* 460 mm×460 mm, 85 lp/cm, 10:1, SID 140 cm, Carbon cover
460 mm×460 mm, 92 lp/cm, 10:1, SID 140 cm, Carbon cover
460 mm×460 mm, 85 lp/cm, 10:1, SID 180 cm, Carbon cover
460 mm×460 mm, 92 lp/cm, 10:1, SID 180 cm, Carbon cover
*Optional (Select 1 of 2 kinds of Grid)
Scatter Radiation Grid
6-Way movement
Table top dimensions (L×W) 2,410 mm×812 mm
Tabletop material HPL
Table Top X-ray density 1.44 mmAl
Elevating (Up / Down) range 545~900 mm
Longitudinal (Left & Right) movement range ±480 mm
Transverse (Front & Rear) movement range ±140 mm Longitudinal Patient coverage (without patient repositioning) 2,078 mm
Bucky movement range 688 mm Max. Patient weight 200 kg (Dynamic load off center) / 310 kg (Dynamic load center) /
350 kg (Static load center) Elevating speed from min. to max height (No load) 40 mm/sec
Foot switch Wireless foot switch (Elevating Up/ Down, Floating Table Top) Auto tracking Longitudinal tracking of detector with longitudinal travel and rotation of tube Auto Centering Automatically adjusts and centers the detector and the tube
Shared BuckyTM Both Detectors (F4335AW, S4343-AW and S4335-AW) are available on a bucky
Available Detector Size 17 x 17 inch / 14 x 17 inch
Detector Installation Fixed or Removable
Grid configuration 85lp/cm or 92lp/cm or none(SID 110cm / 140cm / 180cm)
The flexible 6-way table can endure heavy weight. Thin table legs make the best use of fixed space, and the out-standing traffic line will improve workflow.
Patient Table
GC85A
AccE Standard Detector
Wireless Portable Detector
Detector model S3025-AWM
Detector type Amorphous Silicon TFT / Cesium Iodide scintillator (CsI)
Dimensions 350 x 287 x 15 mm
Active detector matrix 2536 x 2024 pixels (W x H)
Effective area 314.5 mm × 251.0 mm (W x H)
Pixel pitch 124 ㎛
A/D conversion (Pixel depth) 16 bits
Spatial resolution 4.0 lp/mm
Detective Quantum Efficiency (DQE) 60% (0 lp/mm, Typical)
Modulation Transfer Function (MTF) 90% (0.5 lp/mm, Typical)
Installation Wireless only
Image Acquisition Wireless (802.11 n/ac) @2.4/5 GHz
AED N/A
Weight 1.95 kg (Typical)
Max.load capacity 400 kg for uniform load, 200 kg for local load (40 mm in diameter disk at the center)
Dust/Water-resistance IP67
Self diagnosis Automatic detection & alarming of line artifact
Battery charging time 2 hrs. (AC-DC adaptor)
Battery Operation Time 270 images for 7.5 hrs
GC85A Technical Data7 GC85A Technical Data 8
Response
- Without additional absorber 800 pC/μGy · m2
- With additional absorber(0.5 mm Al) 920 pC/μGy · m2
Response versus radiation quality -6 % / +0 % (50~150 kV, acc. IEC 60580)
Quality equivalent filtration 0.2 mm Al @ 70 kV
Transparency > 70 %
Active area (max.) 1 ~ 200 cm2
Chamber voltage 300 V
Distance of the electrodes 6 mm
Stabilization time 5 min
Transportation Temperature -20 ~ 60 °C
Transportation Humidity 10 ~ 80 % (max. 20g/m3 ; non-condensing)
*Optional
Dose Area Product (DAP)* DAP provides patient level dose estimates and recommended exposure. Also connected to PACS system, relevant information is stored on PACS, enabling cumulative dose tracking.
S/W Optional Items
GC85A
Type A Type B
Up / Down range 280~1,850 mm (Motorized)
Receptor tilting -20°~+90° (Motorized) N/A
Detector rotate range 0°~+90° N/A
Receptor Rotation Manual and 0°, 45°, 90° detent N/A
Auto tracking Vertical and rotational tracking of tube with vertical travel and tilting of detector
Auto Centering Automatically adjusts and centers the detector and the tube
Foot switch* Wireless foot switch (Elevating Up/ Down)
LED Indicator The change of the light on the column shows the step of the test directly
Operating Panel Multi-functional operating panel
Motorized fast Up/ Down
Motorized slow Up/ Down (Precise Positioning with tumbler switch)
Motorized Receptor Tilting (-20°~ 90°) Preprogrammed exam positions -Receptor Bottom~Top/ Tilting (-20°~90°), Parking/Table/ Stand/ Stand tilting Mode
Auto Centering, Auto Sync On/Off, Auto Run
- Auto Sync : Enable the synchronization of the detector and Tube
- Auto Run : Moves the system to a preset position defined at the THU (Tube Head Unit) Collimator size lock
Available Detector Size 17 x 17 inch
Detector Installation Fixed or Removable
Grid configuration 85lp/cm or 92lp/cm or none(SID 110cm / 140cm / 180cm)
Wall Stand
AEC function prevents excessive radiation exposure.
Automatic Exposure Control (AEC)
Minimum Response Time Less than 1 mS Output Sensitivity(Gain Range) Adjustable 0.046 ~ 0.91 V/ μGy @76 kV. Additional output sensitivities available upon request Ionization Chamber Potential +75 VDC ±10 V (internally generated)
Output Reproducibility Less than ±0.045 Coefficient of Variation
Output Linear ramp with no more than ±5 % deflection in full output scale.
Field Matching Outputs of multi-field chambers are individually adjustable to within 5 % of one another.
X-ray % Transmission No less than 85 % from 50 kV to 150 kV with 2.5 mm to 3.0 mm total aluminum equivalent beam filtration from the x-ray tube and collimator Power Supply Requirement ±11.4 VDC ~ ±15.75 VDC@ 0.1 A unless specified otherwise
Operating Temperature 10 ~ 40 °C
Operating Humidity 10 ~ 60 % relative humidity non-condensing
Operating Atmospheric Pressure 860 ~ 1060 hPA
Transport and Storage Temperature Range -40 ~ 70 °C
Transport and Storage Humidity Range 10 % ~ 95 % relative humidity non-condensing
Transport and Storage Atmospheric pressure 860 ~ 1060 hPA
Description
SimGridTM Enhancing the visibility of image without physical grid by compensating the decrease in contrast caused by scatter radiation
SimGrid™ In-bucky Extending SimGrid™ function to Stand/Table protocols
Bone Suppression Improving the clarity of soft tissues by suppressing the appearance of bones in chest images
Auto Lung Nodule Detection ALND helps users find lung nodules on x-ray Chest PA images
S-Enhance Enhancing visibility of foreign body in chest, abdomen and l-spine images
Remote View Transferring configured image through a web browser using the IP address
QAP Package Quality Assurance Program and Al Filter/Phantom
S-DAP Software DAP(Dose Area Product)
S-ShareTM1 Sharing detectors between GC85A and Samsung DR system Value-up Package Voice Guide / Image Auto Rotation / Additional Patient Information / Prior Exam Review /
S-Align™ Info. To DICOM / Detector Connection Alarm / Bariatric Exposure Management / Easy Protocol Management / Smart Crop Management
Mirror View Provide secured screen sharing of S-Station to the display device using Wi-Fi CERTIFIED Miracast™
WBS Activation Support verified wireless barcode scanner
- Refer to the verified scanner list
- Interface(Host/Wireless Scanner) : USB Serial/Bluetooth(2.4GHz-2.5GHz, ISM)
Vision Auto Enhancing patient positioning and system setting workflow with live camera Vision Stitching Enhancing stitching workflow with live camera
1) There are restrictions on sharing between detectors, please refer to separate guide
GC85A Technical Data9 GC85A Technical Data 10
Power On/Off switch For Ceiling suspension, Patient table, Wall stand and S-Station
Exposure button 2-Step(Ready, Shot) button
X-ray exposure Indicator Standby : None, Ready : Green, Exposure : Yellow
Emergency stop button Yes
Control Interface Box (CIB)
Network Specifications
Yes
Ceiling X, Y, Z Moving
Auto Centering, Auto Sync, Auto Detector
Table Up & Down, Table Top free
Stand Up & Down, Fast Up & Down, Motorized Receptor tilting
Preprogrammed exam positions, THU control
Parking, Table, Stand, Wall Stand tilting Mode
Collimator lamp control
Vertical & Horizontal enlargement / Reduction of collimator
Collimator Size Fixed, Undo(return to pre-position)
Remote controller
DICOM functions
- DICOM Verification (SCU)
- DICOM Modality worklist (SCU) Interface with HIS / RIS with auto refresh option
- DICOM MPPS (SCU) Send the status of exams to HIS / RIS
- DICOM Storage (SCU) Send Image (DX or CR IOD) and GSPS to PACS
- DICOM Storage commitment (SCU) Send commitment status
- DICOM Grayscale print (SCU) Support DICOM printers
- DICOM Query / Retrieve (SCU) Query and retrieve DX and CR images from PACS
- DICOM GSPS (SCU) Send overlay information along with the image
- DICOM Media exchange (DICOM DIR) Patient images export to DVD/CD
-DICOM DOSE Structured Report (SCU) Send Dose Information to DOSE Server
IHE Integration Profile
- Scheduled Workflow Acquisition Modality
- Patient Information Reconciliation Acquisition Modality
- Radiation Exposure Monitoring Acquisition Modality
- Consistent Time Time Client
- Portable Data for imaging Portable Media Creator
Remote Maintenance System (RMS) The remote management system analyses performance and emerging technical issues.
Real-time use information Rounds of filming, distribution of use, operating ration, new & re-filming rate and accumulated radiation dose.
Remote access Real-time monitoring system enables quick responses to problematic situations.
Remote assistant service allows auto-diagnosis and remote repair.
GC85A
Stitching Range at Patient Table max 1,095 mm for 3-images(17x14, 17x17 detector), 1,015 mm for 4-imag-es(14x17 detector)
Stitching Range at Wall Stand max 1,780 mm for 5-images Average acquisition time for a 3-images exam
< 15 sec (Table)
< 12 sec (stand) Image pasting and processing time for a 3-image exam
< 6 sec from last exposure (Stand / Table)
Grid base stitching both table and stand Grid configuration 85lp/cm or 92lp/cm or none(SID 110cm / 140cm / 180cm)
*Optional
The X-ray tube and the detector move automatically when a full body image is being captured. It captures 2 or more images consecutively and then stitches the images into one. The Smart Stitching function can be operated using in two different methods; Tube Angle and Collimator Control.
Smart Stitching* - Spine and long bone imaging
Hardware Type A
- CPU Intel Core Xeon
- RAM 8G or 16G
- Storage SSD 256 GB + HDD 1 TB
- Image storage capacity F4335-AW/AWM : ~ 14,000 images S4343-AW/AWM : ~ 12,000 images S4335-AW/AWM : ~ 14,000 images S3025-AW/AWM : ~ 30,000 images
- Hard drive encryption
- Operating system Microsoft Windows 10
- Accessories Keyboard, Mouse
Software
- Image acquisition times* (X-ray exposure to image)
F4335-AW(Tethered) : Preview < 2 sec, Final < 4 sec F4335-AW(Wireless) : Preview < 2 sec, Final < 5 sec S4343-AW/AWM(Tethered) : Preview < 2 sec, Final < 5 sec S4343-AW/AWM(Wireless) : Preview < 2 sec, Final < 6 sec S4335-AW/AWM(Tethered) : Preview < 2 sec, Final < 4 sec S4335-AW/AWM(Wireless) : Preview < 2 sec, Final < 5 sec S3025-AW/AWM(Wireless) : Preview < 3.5 sec, Final < 6 sec
* It depends on the image processing parameters and wireless condition
- Post processing
S-VueTM
- Protocol(Body part) adaptive multi scale image processing
- Advanced noise reduction processing
- Excess contrast reduction processing for minimizing metal artifact
- Display functions Window level control, Zoom / Magnifier. Flip / Rotate, Invert, Shutter (Fixed / Auto / Manual), Grid View Annotations (Marker / Free text / Arrow / Line / Length / Angle / Cobbs angle / Rectangle / Ellipse) Layout (1×1, 1×2, 2×1, 2×2, Full Screen View)
- Auto cropping Auto shutter (Rectangle / Polygon)
- Auto APR (Anatomical Programmed Radiography) matching
X-ray conditions, Mechanical position, Image processing parameters, Marker etc
- Multi-language English, French, German, Italian, Spanish, Russian, Simplified Chinese, Portuguese, Turkish, Finnish, Swedish, Czech, Polish
Display*
Type A Type B Type C
Size 24” 23” 27” Resolution 1920 x 1080 1920 x 1080 2560 x 1440 Touch No Yes No Dicom No Yes Yes
*Optional (Select 1 of 3 kinds of monitors)
Imaging Workstation - S-Station
GC85A Technical Data11 GC85A Technical Data 12
Ceiling Suspension SCALE : 1/50 (A3)
Premium Table SCALE : 1/50 (A3)
GC85A
Room Considerations
GC85A
Accessories P-bar for lateral patient for Stand
Patient hand grips for Table
Barcode Scanner*
Patient compression belt for Table*
Lateral detector holder*
Grid holder*
Auto Stitching Stand*
Detector Weight Distribution Cap* (43x43 cm size)
Trolley* (220 kg, 786 mm)
Cone Rail*
QAP* - SW V3.0, 21 mm AL Filter, Phantom (Pehamed co. DIGRAD A+K)
DVD-RW* (External USB type)
RFID Reader (Support 125kHz, 13.56MHz) (Recommended card type : HID Prox / iCLASS ID)
Detector Accessory* Detector Battery Charger
Detector Battery
Grid* Portable Grid
- 4335 : 404 x 480 mm, 85 lp/cm, 10:1, SID 110 cm, Carbon cover 404 x 480 mm, 92 lp/cm, 10:1, SID 110 cm, Carbon cover
- 4343 : 480 x 480 mm, 85 lp/cm, 10:1, SID 110 cm, Carbon cover 480 x 480 mm, 92 lp/cm, 10:1, SID 110 cm, Carbon cover
*Optional
Accessories
The entire system is powered via a single line voltage connection.
Power connection 3-phase, 380/400/480 VAC, 50/60 Hz Power rating 120 kVA
5 A (Nominal) 180/150 A (Momentary)
Wall stand 558 mm×877 mm×2,275 mm (Stroke : 280~1,850 mm) Approx. 180 kg
Ceiling suspension 4,008 mm×3,010 mm×1,253 mm (The rail size(W X L) can be changed by depending on room size) Approx. 390 kg
Patient table 2,410 mm×812 mm×545 mm (Vertical Stroke : 355 mm) Max. 350 kgf supported Approx. 210 kg
System cabinet Type A & B: 620mmX405mmX630mm, Approx. 96kg(HVG), 620mmX450mmX432mm, Approx. 62kg(H/W)
Type C & D: 596mmX410mmX480mm, Approx. 63kg(HVG), 620mmX450mmX432mm, Approx. 62kg(H/W)
Operating Temperature Temperature range : +10~+40 °C
Operating Humidity Relative humidity : 30~75 %
Operating Atmospheric Pressure Air pressure : 70~106 kPa
Storage / Transport
Transportation Temperature - Temperature range : -10~+55 °C
Transportation Humidity - Humidity range : 10~80 %
Transportation Atmospheric Pressure - Pressure range : 70~106 kPa
Rail Longitudinal 3 ~ 5 m (0.5 m steps), 6.5m (Special Order)
Transverse 2.5 ~ 4 m (0.5 m steps)
Installation Data
GC85A Technical Data13 GC85A Technical Data 14
Wall Stand – Type. A SCALE : 1/50 (A3)
Wall Stand – Type. B SCALE : None
GC85A
Room Considerations
GC85A
Room Considerations
Stitching Foot Step
Stitching Stand
314.5x500x100 [mm]
ⓒ 2022 Samsung Electronics All rights reserved.
Samsung Electronics reserves the right to modify the design, packaging, specifications, and features shown herein, without prior notice or obligation.
About Samsung Electronics Co., Ltd.
Samsung Electronics Co., Ltd. inspires the world and shapes the future with transformative ideas and technologies. The company is redefining the worlds of TVs, smartphones, wearable devices, tablets, cameras, digital appliances, printers, medical equipment, network systems, semiconductor and LED solutions. For the latest news, please visit the Samsung Newsroom at news.samsung.com.
For more information For more information about GC85A, visit www.samsunghealthcare.com
GC85A Datasheet v4.3-230316-EN
File details come from the government source that posted it. Updated .