Seed 2_557-20-101 -Cardiac_D-Spect Specs.pdf

PDF 317 KB Posted

Attached to
VISN 7 CONSTRUCTION MATOC 36C24726R0026 - SOLICITATION
Solicitation number
36C24726R0026
Issued by

About this file

This document is a technical specifications sheet for the D-SPECT® Cardio Imaging System, a cardiac imaging device manufactured by Spectrum Dynamics Medical, Inc.

The D-SPECT® Cardio is a pixelated crystal detector system with nine detectors utilizing integrated tungsten collimators and cadmium zinc telluride (CZT) semi-conductor crystals. Key detector specifications include a spatial resolution of 2.46 mm pixel pitch, 1024 pixels per detector (9216 total), energy range of 40-170 keV with extrinsic energy resolution of ≤6.2% at 140 keV, detection efficiency of 80% at 140 keV, and maximum count rate exceeding 1,350,000 counts per second with less than 5% count loss. The system features integrated ECG gating with 8 or 16 frames per R-R interval and automatic beat centering. The patient chair/bed accommodates a maximum static weight of 454 kg (1000 lbs) and dynamic weight of 240 kg (529 lbs), with adjustable height between 60-90 cm from the floor. System dimensions are 190 cm height, 225 cm depth in supine mode, and 155 cm width, with a total weight of 670 kg. Operating conditions require ambient temperature between 15-24°C (59-75°F) and 30-65% relative humidity. The system requires 880 VA nominal power with dual UPS options: 120 VAC (95-151 VAC, 60 Hz) or 230 VAC (181-290 VAC, 50-60 Hz). Minimum room requirements are 9'5" x 8'8" for upright imaging and 9'5" x 11' for supine imaging.

View the file

Other files

File Type Posted
36C24726R0026 0013.pdf PDF
RFI LOG AMENDMENT 0013_ SEED 1 AND 2.pdf PDF
Seed 2_Project 557-20-101_27 13 23_Comm Optical Fiber Backbone Cabling.pdf PDF
RFI LOG AMENDMENT 0012_ SEED 3.pdf PDF
Seed 2_Project 557-20-101_27 10 00_Update Structured Cabling.pdf PDF
RFI LOG AMENDMENT 0011_ SEED 2 REVISED.pdf PDF
36C24726R0026 0012.pdf PDF
RFI LOG AMENDMENT 0011_ SEED 2.pdf PDF
36C24726R0026 0011.pdf PDF
36C24726R0026 0010.pdf PDF
SEED 1_PROJECT 534-23-910 INFECTION CONTROL RISK ASSESSMENT.pdf PDF
36C24726R0026 0009.pdf PDF
RFI LOG AMENDMENT 0009_SEED 1_ SEED 2_ MATOC RFP.pdf PDF
SEED 1_PROJECT 534-23-910 INFECTION CONTROL DETAILS.pdf PDF
ATTACHMENT 18 - SEED 3 WAGE DETERMINATION TUSKEGEE AL20260078 5 18 2026.pdf PDF
36C24726R0026 0008.pdf PDF
ATTACHMENT 14 - SEED 2_WAGE DETERMINATION GA20260095 05 18 2026.pdf PDF
RFI LOG AMENDMENT 0007_Seed 3 - 619A4-25-102.pdf PDF
36C24726R0026 0007.pdf PDF
Seed 3_ 619A4-25-102 - Correct Egress Door Deficiencies Final Drawings.pdf PDF
RFI LOG AMENDMENT 0006_SEED 2 - Revised.pdf PDF
ATTACHMENT 11 - SEED 1_WAGE DETERMINATION SC20260023 05 18 2026.pdf PDF
Seed 2_557-20-101 - Project Bldg 5 As-Builts.pdf PDF
Seed 2_557-20-101 - Photos_RFI 262 263.docx DOCX document
Seed 3_619A4-25-102 Wage Rates AL20260078 01 02 2026.pdf PDF
Seed 1_534-23-910 SOW Rev1 - Final.pdf PDF
Seed 2_557-20-101 -Typical - Symbia Intevo Bold.pdf PDF
Seed 2_557-20-101 - 2024-06-19_SD Rpt_ GC85A_Physicist Rp.pdf PDF
Seed 2_557-20-101 -HME_DR_GM85_Datasheet_v42_230209_FDA.pdf PDF
Seed 2_557-20-101 -GEHC-SP_SIGNA-VOYAGER_8-278F.pdf PDF
Seed 2_557-20-101 -5418654-1EN Rev 22.pdf PDF
Seed 2_557-20-101- CT_GE Revolution Apex spec sheet_brosh.pdf PDF
Seed 2_557-20-101 -HME_DR_GC85A_Datasheet_v43_230316_FDA.pdf PDF
RFI LOG AMENDMENT 0006_SEED 2-557-20-101.pdf PDF
36C24726R0026 0006.pdf PDF
Seed 2_557-20-101 -gehealthcare-brochure_idxa_pdf.pdf PDF
Seed 2_557-20-101 - Zone Valve.jpg JPG image
Seed 2_557-20-101 - Monitors.jpg JPG image
Seed 2_557-20-101 Ultrasound_RS85PrestigeTechSpecv2051RS85145rev01.pdf PDF
Seed 2_557-20-101 -GEHC-SP_REVAPEX_6-120F.pdf PDF
Seed 2_557-20-101 -Zone Valve A133-5.jpg JPG image
Seed 2_557-20-101 - Persona RF PREMIUM Brochure Fuji.pdf PDF
Seed 2_557.20-101 - REVIT and CAD Screenshot.docx DOCX document
Seed 2_557-20-101 -GC85A Site planning - Samsung Healthcare.pdf PDF
Seed 2_557-20-101 -Signa Voyager 1_5 MRI.pdf PDF
Seed 2_557-20-101 - VA Radiology - CD100 Final Bid Documents - Combined Set - SIGNED.pdf PDF
Seed 2_557-20-101 -Probe Cabinet specs.pdf PDF
Seed 2_557-20-101 - Med Gas Master Alarm Panel.jpg JPG image
RFI LOG AMENDMENT 0006_MATOC RFP.pdf PDF
RFI LOG AMENDMENT 0005_Seed 1.pdf PDF
Show 40 more

Latest 50 of 92 listed here. See the rest on GovTribe.

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

D-SPECT®Cardio Imaging System

Detector Specifications Detector Properties

Detector spatial resolution (pixel pitch) 2.46 mm Detector Type Pixelated Crystals

Intrinsic Spatial Resolution Collimator Type Tungsten

FWHM in UFOV NA Number of Detectors 9

FWTM in UFOV NA Camera Field-of-View(FOV) 15.74 cm x patient adjustable

*D-SPECT uses Integrated Collimators* Detector Field-of-View(UFOV) 15.74 x 3.94 cm (6.2 x 1.6 in)

Intrinsic Spatial Linearity Number of Pixels per Detector 1024

Differential in UFOV NA Total Number of Pixels 9216

Integral in UFOV NA Detector Range of Motion >110°

*Linearity defined by pixelated structure with NO drifts Pixel Size 2.26mm x 2.26 mm

Extrinsic Energy Resolution Energy Range 40 - 170 keV

FWHM at 140keV ≤6.2% Lead Shielding 175 keV

Intrinsic Energy Resolution Crystal Properties

FWHM at 140keV NA Crystal Material CZT (Cadmium Zinc Telluride)

Extrinsic Uniformity Crystal Type Semi-Conductor

Differential in UFOV <3% Decay Time 300ns

Integral in UFOV <4% Detection Efficiency 80% @140keV

Intrinsic Uniformity Crystal Thickness 5.0 mm (0.2 in)

Differential in UFOV NA Patient Chair/Bed Integral in UFOV NA Max Patient Weight (Static) 454 kg (1000 lb)

*D-SPECT uses Integrated Collimators* Max Patient Weight (Dynamic) 240 kg (529 lb)

Max Count Rate with Scatter Lowest chair position from floor 60 cm (23.62 in)

(Per camera with <5% count Loss) >1,350,000 cps Highest chair position from floor 90 cm (35.43 in)

Reconstructed Spatial Resolution Dimensions and Weight FWHM with Scatter at 15cm 5.0 mm Height 190 cm (74.8 in)

ECG Gating Depth (Supine) 225 cm (88.58in)

Number of Frames per R-R 8 or 16 Depth (Upright) 155 cm (61.02 in)

Automatic Beat Centering Yes Width 155 cm (61.02 in)

Integrated ECG Yes Weight 670 kg (1477 lb)

Flood Loading (No Patient) 610 kg/m2 (125 lb/ft2)

TECHNICAL SPECIFICATIONS

System Component Condition

Ambient

Temperature

Relative Humidity

(Non-Condensing)

Operation 15 -24° C 30 -65 %

D-SPECT Cardio 59 -75° F

Max Temp Change Rate 3° C per hour

5.4° F per hour

Storage and Transportation 0 - 30° C 30 -65 %

32 -86° F

Heat Dissipation Handling 6800 BTU/Hr

D-SPECT Cardio Power Nominal Input Voltage Allowed Input Voltage Frequency

(VA) (VAC) (VAC)

With 120V UPS 880 VA 120 VAC 95-151 VAC 60Hz

With 230V UPS 880 VA 230 VAC 181-290 VAC 50-60 Hz

The D-SPECT Cardio has a small footprint that easily fits into compact spaces

Minimum room requirements for upright imaging – 9’5” x 8’8”

Minimum room requirements for supine imaging – 9’5” x 11’

Spectrum Dynamics Medical, Inc.

Spectrum Dynamics Medical HK LTD

Spectrum Dynamics Medical SA

301 N. Cattlemen Rd. Suite 109 5 Canton Road Suite 1729A Rue de Lausanne, 31

Sarasota, FL 34232 USA Ocean Center 1110 Morges, Switzerland

Tel 941 256 3660 Tsim Sha Tsui, Kowloon, Hong Kong Tel +41 21 544 4710

Fax 941 256 3662 Tel +852 580 14151 Fax +41 21 544 4711 info@spectrum-dynamics.com Fax +852 580 14146 www.spectrum-dynamics.com

*Spectrum Dynamics Medical reserves the right to make changes in specifications without notice or warning*

D-SPECT® is a registered trademark of Spectrum Dynamics Medical.

@2017 All rights reserved. Aug 2018 MAN0052_C

PHYSICAL, ENVIROMENTAL, AND POWER SPECIFICATIONS