MP-WVA120_R22_PP-201.pdf
PDF 1 MB Posted
- Solicitation number
- W912DS24R0020
About this file
This document appears to be a technical drawing for the lab process piping in the second floor of Building 120 at the Watervliet Arsenal in New York. The drawing provides details on the layout of gas cylinder storage cabinets, gas manifolds, air compressor and air dryer equipment, and piping runs for various gases including nitrogen, air, argon, carbon dioxide, hydrogen, and oxygen. The drawing includes a legend, general notes, and equipment schedules. It is part of the technical specifications for the solicitation to repair Building 120, which is a $25M-$100M project by the U.S. Army Corps of Engineers to renovate and modernize the 102,000 square foot multi-use building. The project will consolidate maintenance and wood shop spaces on the ground level and construct a new lab facility on the second floor.
View the file
Other files
Show 40 more
Latest 50 of 59 listed here. See the rest on GovTribe.
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
D
E
F
G
87654321
GAS CYLINDERS
HYDROGEN CHLORIDE
GAS CYLINDERS
FORMING GAS
GAS CABINET
PROPANE
GAS CYLINDERS
P-10
FREIGHT
ELEVATOR
611 SF
MECHANICAL ROOM
PROCESS LAB
ELEVATOR
VESTIBULE
TOILET TOILET
TOILETTOILET
CHEMICAL STORAGE
CORRIDOR
A
A 1 A
B
B
172-1
213201 202
212C
115A
193 188 133
123 124 122194
120U119U 121U
P
SS
201P
157P
213C
202P
212P
84P
199P
200P
211P
163P 162P
161P
187,127,128,131,132,138,114, 129O,130O,143O,124O,220O
174U,19,116,117,140, 118O,135O,136O,139O141O,142O
P
133P
PROCESS PREP. AREA
SS
SS
SS
FH-BE-1
84-1
163 162 161
HD-MS
CFH-08
CFH-07
CFH-10
CFH-09 CFH-11
CFH-12 CFH-14 CFH-16
CFH-15CFH-13
CFH-01
CFH-02
CFH-04
CFH-03 CFH-05
CFH-06
CFH-18 CFH-20 CFH-22
CFH-21CFH-19CFH-17
F H
-M L
FH-PL2-(2)
FH-P162-(2)FH-P163-(2)
FH-PL4-(1)
FH-PL3-(1)
FH-P161-(2)
F H
-P C
FH-PL1-(1)
SDP
SDP
SDP
S D
P
1/2"-CU2-HS-N2
1/2"-CU2-HS-AIR
1/2"-CU2-HS-N2
BOP - 9' - 1"
1/2"-CU2-HS-N2
1/2"-CU2-HS-AIR
BOP - 9' - 6"
1/2"-CU2-HS-N2
1/2"-CU2-HS-AIR
BOP - 9' - 1"
AIR
COMPRESSOR
AIR DRYER AIR
RECEIVER
SPARE GAS
CYLINDER
STORAGE
1/2"-CU2-HS-ARG
1/2"-CU2-HS-N2
4' - 7" 2' - 2"
BOP - 12' - 3"
3' - 11 1/2"
/2
1/2"-CU2-HS-N2
1/2"-CU2-HS-ARG
1/2"-CU2-HS-AIR
BOP - 9' - 9"
BOP - 12' - 8"
2' - 10"
1' - 8 1/2"
3' - 5 1/2" 2' - 0" 4' - 2" 2' - 0"
1' - 0"
3' - 9"
BOP - 10' - 0"
1/2"-CU2-HS-AIR
1/2"-CU2-HS-N2
1/2"-CU2-HS-ARG
BOP - 9' - 0"
1/2"-CU2-HS-ARG
BOP - 8' - 8"
1' - 0" 3' - 1"
1' - 3"
1"-CU2-HS-ARG
1"-CU2-HS-N2
BOP - 11' - 10"
1' - 5"
2' - 2"
1/2"-CU2-HS-AIR
1/2"-CU2-HS-N2
1/2"-CU2-HS-ARG
1/2"-CU2-HS-AIR
BOP - 9' - 1"
7' - 8" 8' - 2 1/2" 3' - 7" 1' - 3"1' - 6"
3/4"-ST2-CO2
3/4"-ST2-H2
3/4"-ST2-O2
0' - 8" 3' - 11 1/2" 4' - 7"
0' - 8"
1' - 3"
1' - 4 1/2"
0' - 8"
5' - 1 1/2"
1' - 0"
2' - 0"
0' - 8"
2' - 2"
0' - 8"
GAS CYLINDERS
CARBON DIOXIDE
GAS CYLINDERS
ARGON
GAS CYLINDERS
NITROGEN
GAS CYLINDERS
ZERO AIR
/2
/2
1' - 6" 8' - 0"
2' - 10" 7' - 8" 2' - 2"
3/4"-ST2-P10
1' - 0"
1' - 6" 7' - 8" 1' - 6" 8' - 2" 2' - 1"
BOP - 8' - 8"
1/2"-CU2-HS-N2
1/2"-CU2-HS-AIR
BOP - 9' - 1"
1/2"-CU2-HS-AIR
1/2"-CU2-HS-ARG
1/2"-CU2-HS-N2
BOP - 9' - 1"
1/2"-CU2-HS-AIR
1/2"-CU2-HS-ARG
1/2"-CU2-HS-N2
BOP - 9' - 1"
BOP - 9' - 10"
1/2"-ST2-O2
1/2"-ST2-CO2
1/2"-ST2-H2
BOP - 9' - 1"
BOP - 8' - 8"
1/2"-CU2-HS-N2
1/2"-CU2-HS-ARG
1/2"-CU2-HS-AIR
1/2"-ST2-HCL
BOP - 8' - 8"
1"-CU2-HS-N2
1"-CU2-HS-ARG
1"-CU2-HS-AIR
BOP - 8' - 8"
1/2"-CU2-HS-N2
1/2"-CU2-HS-AIR
BOP - 9' - 1"
1/2"-CU2-HS-N2
1/2"-CU2-HS-ARG
1/2"-CU2-HS-AIR
BOP - 11' - 4"
BOP - 9' - 10"
BOP - 9' - 1"
1/2"-CU2-HS-AIR
1/2"-CU2-HS-ARG
1/2"-CU2-HS-N2
BOP - 9' - 1"
1/2"-CU2-HS-AIR
1/2"-CU2-HS-N2
1/2"-CU2-HS-AIR
1/2"-CU2-HS-N2
1/2"-CU2-HS-ARG
BOP - 9' - 1"
BOP - 9' - 1"
BOP - 9' - 1"
1/2"-CU2-HS-N2
1/2"-CU2-HS-ARG
1/2"-CU2-HS-AIR
BOP - 9' - 1"
3/4"-ST2-AR
3/4"-ST2-N2
3/4"-ST2-AIRZ
3/4"-ST2-P10
BOP - 11' - 5"
BOP - 10' - 9"
3/4"-ST2-P10
3/4"-ST2-AIRZ
3/4"-ST2-N2
3/4"-ST2-AR
BOP - 11' - 8"
1"-CU2-HS-ARG
1"-CU2-HS-N2
1"-CU2-HS-AIR
3/4"-ST2-CO2
3/4"-ST2-LPG
3/4"-ST2-FG
3/4"-ST2-CH4
3/4"-ST2-H2
3/4"-ST2-O2
BOP - 10' - 6"
BOP - 11' - 5"
BOP - 10' - 10"
1/2"-ST2-CO2
1/2"-ST2-CH4
1/2"-ST2-O2
1/2"-ST2-H2
1/2"-CU2-HS-N2
1/2"-CU2-HS-ARG
1/2"-CU2-HS-AIR
BOP - 10' - 1"
BOP - 10' - 1"
BOP - 13' - 3"
BOP - 10' - 10"
1/2"-ST2-LPG
1/2"-CU2-HS-N2
BOP - 10' - 1"
" 1/2"-ST2-LPG
1/2"-CU2-HS-N2
BOP - 10' - 1"
1/2"-ST2-CH4
1/2"-ST2-O2
1/2"-ST2-LPG
1/2"-CU2-HS-N2
1/2"-CU2-HS-ARG
1/2"-CU2-HS-AIR
BOP - 10' - 1" 1/2"-ST2-CO2
1/2"-ST2-CH4
1/2"-ST2-O2
1/2"-ST2-H2
1/2"-CU2-HS-N2
1/2"-CU2-HS-ARG
1/2"-CU2-HS-AIR
BOP - 10' - 1"
BOP - 10' - 1"
BOP - 10' - 3"
3/4"-ST2-P10
3/4"-ST2-AIRZ
3/4"-ST2-N2
3/4"-ST2-AR
BOP - 8' - 8"
3/4"-ST2-O2
3/4"-ST2-H2
3/4"-ST2-CH4
3/4"-ST2-FG
3/4"-ST2-LPG
3/4"-ST2-CO2
BOP - 8' - 8"
BOP - 10' - 6"
1"-CU2-HS-AIR
BOP - 11' - 6"
3/4"-ST7-HCL
1/2"-ST7-HCL
3/4"-ST7-HCL
3/4"-ST7-HCL
1"-CU2-HS-N2
1"-CU2-HS-ARG
1"-CU2-HS-AIR
BOP - 11' - 5"
BOP - 8' - 8"
BOP - 9' - 9"
GAS CABINET
OXYGEN
GAS CABINET
HYDROGEN
GAS CABINET
METHANE
/2
1' - 3"
3' - 0"
0' - 5"
3' - 0 1/2"
1' - 11 1/2"
1' - 6" 1' - 4"
3/4"-ST2-O2
3/4"-ST2-H2
3/4"-ST2-HCL3' - 1" 0' - 8" 5' - 3"
0' - 4"
0' - 4"
9' - 0"
" 1
/2
AIR COMPRESSOR SCHEDULE
TAG MANUFACTURER MODEL SERVICE TYPE
AIR
FLOW
(CFM)
PRESSURE
(PSIG)
PUMP
(HP)
ELECTRICAL
WEIGHT
(LBS)
REMARKS
V PH HZ FLA MOCP
AC-1 INGERSOLL RAND RS37n MECHANICAL ROOM 1ME2 AIR-COOLED, ROTARY SCREW, VARIABLE SPEED 200 100 50 480 3 60 71 125 2524
AIR DRYER SCHEDULE
TAG MANUFACTURER MODEL SERVICE TYPE
RATED
FLOW
(SCFM)
PRESSURE
DROP
INLET/
OUTLET
SIZE (IN.)
OPERATING TEMP (°F)
REFRIGERANT
TYPE
ELECTRICAL
WEIGHT
(LBS)
REMARKS
MIN. MAX. V PH HZ MCA
AD-1 INGERSOLL RAND NVC-200 MECHANICAL ROOM 1ME2 CYCLING REFRIGERATED 200 1.6 1.5" 113 120 R404A 460 3 60 3.9 A 620
AIR FILTER SCHEDULE
TAG
MANUFACTURER
MODEL AREA SERVED TYPE
PARTICULATE
REMOVAL
AF-1 INGERSOLL RAND FA400IG MECHANICAL ROOM 1ME2 INLINE 0.03 MICRON
AF-2 INGERSOLL RAND FA400IH MECHANICAL ROOM 1ME2 INLINE 0.01 MICRON
AIR RECEIVER SCHEDULE
TAG SERVICE TYPE MAWP MDMT
CAPACITY
(GAL)
DIAMETER
(IN.)
HEIGHT
(IN.)
WEIGHT
(LBS)
REMARKS
AR-1 MECHANICAL ROOM 1ME2 VERTICAL 165 PSI @ 400°F -20°F @ 165 PSI 500 36"Ø 117" 915
US Army Corps of Engineers ®
SHEET ID
U S
A R
M Y
C O
R P
S O
F E
N G
IN
E
E R
S D
R A
W N
B Y
D E
S
IG
N E
D B
Y
S U
B M
IT
T
E D
B Y
S
IZ
E
C H
E C
K E
D B
Y x
IS
S
U E
D A
T E
C O
N T
R A
C T
N O
S O
L
IC
IT
A
T
IO
N N
O
A
M A
R K
D E
S C
R
IP
T
IO
N
F ile
P a th
P lo t D a te
2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
B
C
D
E
F
G
H
J
K
L
M
N
P
WVA B120
IS
S
U E
F O
R C
O N
S T
R U
C T
IO
N
A u to d e s k D o c s
/A U
S -0
-U
S A
C E
N
Y W
V A
B ld g R e ca p /M
P -W
V A
R
.r v t
/2
/2
:5
:0
P
M
PP-201
L A
B P
R O
C E
S S
P
IP
IN
G
S
E C
O N
D F
L O
O R
P L A
N
W A
T E
R V
L
IE
T A
R S
E N
A L N
Y
N Y
D
IS
T R
IC
T
A R
C A
D
IS
U .S
I N
C
R
O U
T E
S
U
IT
E
C L
IF
T O
N P
A R
K N
Y
G H u ll
W D a v is
J H e il
J H e il
/2
/2
W D
S F
W D
S D
I D
C -M
R E
C A
P
IT
A L
IZ
A T
IO
N
O F
B U
IL
D
IN
G
B
P R
O J E
C T
N O
W
D
S F
SCALE: 1/4" = 1'-0"
LAB PROCESS PIPING - SECOND FLOOR PLAN COLUMNS 1-8, D-G
0 4' 8' 16'
PLAN NORTH
A. THE CONTRACTOR SHALL FIELD VERIFY ALL DIMENSIONS PRIOR TO THE
FABRICATION AND INSTALLATION OF PROCESS PIPING.
B. PIPE SUPPORTS ARE NOT SHOWN FOR CLARITY. PROVIDE PIPE SUPPORT
PER PROJECT SPECIFICATIONS.
C. PROVIDE A MINIMUM 1-INCH CLEARANCE BETWEEN BARE PIPING OR PIPE
INSULATION.
D. THE CONTRACTOR SHALL BE RESPONSIBLE FOR PIPING INSTALLATION IN
SUCH A MANNER AS TO CONFORM TO AN EXISTING STRUCTURE. AVOID
OBSTRUCTIONS, PRESERVE HEADROOM AND KEEP OPENINGS AND
PASSAGEWAYS CLEAR AND ACCESSIBLE WITHOUT ADDITIONAL COST TO THE
OWNER.
E. ALL PIPING SHALL BE SUPPORTED FROM BUILDING STRUCTURE ONLY. NO
PIPING SHALL BE SUPPORTED FROM OTHER PIPING OR DUCTWORK.
F. COORDINATE ALL PIPING WITH ALL OTHER WORK TO AVOID CONFLICTS.
INSTALL ALL PIPING TO BEST SUIT FIELD CONDITIONS AND COORDINATE WITH
THE INSTALLATION WORK OF OTHER TRADES. THE DRAWINGS ARE
DIAGRAMMATIC AND SHALL NOT BE SCALED TO DETERMINE THE EXACT
LOCATION OF MECHANICAL WORK.
G. THE CONTRACTOR SHALL INCLUDE ANY PIPE TURNS OR ANY ADDITIONAL
APPURTENANCE NECESSARY FOR THE PROPER FUNCTION OF THE SYSTEM
EVEN IF THEY ARE NOT SHOWN ON THE DRAWINGS.
H. FIELD ROUTE ALL FINAL CONNECTIONS TO THE LAB EQUIPMENT.
GENERAL PLAN NOTESKEYNOTE LEGEND
KEY VALUE KEYNOTE TEXT
EACH PAIR OF GAS CYLINDERS USED FOR SERVICE SHALL BE CONNECTED TO A GAS CHANGE OVER MANIFOLD SUITABLE FOR THE GAS SERVICE INDICATED WITH UP TO TWO CONNECTED BOTTLES ON THE SOURCE SIDE. PROVIDE TWO (2) CYLINDER STORAGE CABINETS FOR GASES IN SERVICE AS SHOWN ON PLAN (OXYGEN, HYDROGEN, METHANE AND PROPANE). GAS CHANGE OVER MANIFOLDS SHOULD INCLUDE INSTRUMENTATION FOR A MANUAL CHANGE OVER VALVE, PRESSURE REDUCING REGULATOR (0-100 PSIG), ADJUSTABLE PRESSURE RELIEF DEVICE (50-150 PSIG), AND PRESSURE INDICATOR FOR BOTH THE HIGH AND LOW PRESSURE SIDE OF THE REGULATOR. PROVIDE SHUTOFF VALVES ON MAIN DISTRIBUTION LINES FOR CYLINDER SUPPLIED GASES. ALL GAS FITTINGS SHOULD BE CGA FITTINGS AND COORDINATED WITH GAS BOTTLE SUPPLIER VENDOR TO INCLUDE TWO HOSES FOR CONNECTING GAS BOTTLES.
3 EACH TWO (2) BOTTLE CYLINDER STORAGE RACK TO BE SECURELY ANCHORED PER VENDOR RECOMMENDATIONS. EACH GAS BOTTLE RACK CONTAINS A GAS CHNAGE OVER MANIFOLD SUITABLE FOR THE GAS SERVICE INDICATED WITH UP TO TWO CONNECTED BOTTLES ON THE SOURCE SIDE. GAS CHANGE OVER MANIFOLDS SHOULD INCLUDE INSTRUMENTATION FOR A MANUAL CHANGE OVER VALVE, PRESSURE REDUCING REGULATOR (0-100 PSIG), ADJUSTABLE PRESSURE RELIEF DEVICE (50-150 PSIG), AND PRESSURE INDICATOR FOR BOTH THE HIGH AND LOW PRESSURE SIDE OF THE REGULATOR. ALL GAS FITTINGS SHOULD BE CGA FITTINGS AND COORDINATED WITH
GAS BOTTLE SUPPLIER VENDOR TO INCLUDE TWO HOSES FOR CONNECTING GAS BOTTLES.
4 EACH TWO (2) BOTTLE CYLINDER STORAGE RACK TO BE SECURELY ANCHORED PER VENDOR RECOMMENDATIONS. TWO (2) BOTTLE CYLINDER STORAGE RACKS ARE FOR STORAGE ONLY AND CONTAIN NO MANIFOLDS OR PIPING. QUANTITY FOURTEEN (14) GAS STORAGE
RACKS ARE REQUIRED FOR THE PROJECT.
GAS MONITORING SENSORS TO BE LOCATED IN THIS AREA. MONITORING FOR O2, LEL, CO2 AND HCL IS REQUIRED.
KEYPLAN
6 6 6
EACH LAB BENCH TO INCUDE SPIGOTS FOR HOUSE NITROGEN AND HOUSE AIR CONNECTION.
EACH LAB BENCH TO INCUDE SPIGOTS FOR HOUSE NITROGEN AND PROPANE CONNECTION.
5 5
A. THE CONTRACTOR SHALL FIELD VERIFY ALL DIMENSIONS PRIOR TO THE
FABRICATION AND INSTALLATION OF PROCESS PIPING.
B. PIPE SUPPORTS ARE NOT SHOWN FOR CLARITY. PROVIDE PIPE SUPPORT
PER PROJECT SPECIFICATIONS.
C. PROVIDE A MINIMUM 1-INCH CLEARANCE BETWEEN BARE PIPING OR PIPE
INSULATION.
D. THE CONTRACTOR SHALL BE RESPONSIBLE FOR PIPING INSTALLATION IN
SUCH A MANNER AS TO CONFORM TO AN EXISTING STRUCTURE. AVOID
OBSTRUCTIONS, PRESERVE HEADROOM AND KEEP OPENINGS AND
PASSAGEWAYS CLEAR AND ACCESSIBLE WITHOUT ADDITIONAL COST TO THE
OWNER.
E. ALL PIPING SHALL BE SUPPORTED FROM BUILDING STRUCTURE ONLY. NO
PIPING SHALL BE SUPPORTED FROM OTHER PIPING OR DUCTWORK.
F. COORDINATE ALL PIPING WITH ALL OTHER WORK TO AVOID CONFLICTS.
INSTALL ALL PIPING TO BEST SUIT FIELD CONDITIONS AND COORDINATE WITH
THE INSTALLATION WORK OF OTHER TRADES. THE DRAWINGS ARE
DIAGRAMMATIC AND SHALL NOT BE SCALED TO DETERMINE THE EXACT
LOCATION OF MECHANICAL WORK.
G. THE CONTRACTOR SHALL INCLUDE ANY PIPE TURNS OR ANY ADDITIONAL
APPURTENANCE NECESSARY FOR THE PROPER FUNCTION OF THE SYSTEM
EVEN IF THEY ARE NOT SHOWN ON THE DRAWINGS.
H. FIELD ROUTE ALL FINAL CONNECTIONS TO THE LAB EQUIPMENT.
I. FIELD ROUTED DROPS TO BE PROVIDED WITH POINT-OF-USE PRESSURE
REGULATORS AS NEEDED FOR EQUIPMENT AND INSTRUMENTS (10-30 PSIG).
REFER TO SPECIFICATION SECTION 40 05 13 FOR REGULATOR
SPECIFICATION.
12" = 1'-0"
AIR COMPRESSOR SCHEDULE
12" = 1'-0"
AIR DRYER SCHEDULE
12" = 1'-0"
AIR FILTER SCHEDULE12" = 1'-0"
AIR RECEIVER SCHEDULE
R
T A
S U
B M
IS
S
IO
N
/0
/2
A
M E
N D
M E
N T
/2
/2
PROVIDE A THREE (3) CYLINDER STORAGE CABINET FOR STORAGE OF OXYGEN GAS CYLINDERS NOT IN SERVICE. THREE (3) CYLINDER
STORAGE CABINETS ARE FOR STORAGE ONLY AND CONTAIN NO MANIFOLDS OR PIPING.
AC-2 W30IE-A-116 2ME1 SCROLL, OIL-FREE 124 116 40 460 TBD TBD 1967 AD-2 EH200 2ME1 DESICCANT
DESICCANT
ACTIVATED ALUMINA 913
AR-2 2ME1 200 30" 72" 505 AF-3
AF-4
FA230IH
FA230IH
2ME1
2ME1
0.2 MICRON
0.2 MICRON
JHeil Stamp
JHeil Date 07/25/2024
| Sheets |
| PP-201 - LAB PROCESS PIPING - SECOND FLOOR PLAN |
| 2024-07-25T14:52:49-0400 | |
| Jeffrey Heil | |
| I am approving this document. |
File details come from the government source that posted it. Updated .