This document is a facility requirements drawing set for the OR1 NEO-IP Series System installation at VA Medical Center West Palm Beach, FL, prepared by Karl Storz Endoscopy-America. The drawings are coordination and planning documents (not for construction) that specify technical facility requirements, electrical infrastructure, equipment specifications, and site readiness conditions necessary for integrating the OR1 NEO-IP integrated operating room system.
The facility requirements encompass comprehensive audiovisual and electrical infrastructure including an A/V equipment rack (WRK-3727 medium rack with 900 lbs seismic load capacity), conduit riser systems, junction boxes, power receptacles (20A circuits at 120VAC for rack and display power), data/telecommunications connectivity, ceiling-mounted speaker assemblies (six TCM-XA speakers with grille assemblies), and specialized wall-mounted equipment including touch panels, displays, microphones, and interface boxes. Equipment specifications include detailed thermal and power load data (e.g., NEO-IP 48 assembly with 1185 BTU heat load, AIDA printer with 1160 BTU, medical UPS with 304 BTU/hr), dimensional drawings, and interface requirements. Critical facility constraints mandate equipment operation between 32°F to 104°F ambient temperature with surrounding air not exceeding 90°F, requiring proper ventilation. The site readiness checklist establishes prerequisites including completed conduit installation with pull strings, energized permanent power, ceiling speaker cutout preparation, wall mount installation, network cable termination, boom equipment installation and powering, and owner-furnished equipment availability before Karl Storz deployment crew arrival. All construction work including conduit, electrical outlets, millwork, and structural support is specified as the client's contractor responsibility unless otherwise noted.
View the file
Other files
Show all 36
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
PLOT STAMP:SATRE, SHAUN |DATE: 5/21/2025 10:31 AM
C O
V E
R S
H E
E T
VA Medical Center West Palm Beach, FL
INDEX OF DRAWINGS KARL STORZ ENDOSCOPY-AMERICA
STAFFORD, TEXAS
REVS NOTES:
0. INITIAL RELEASE 2024/11/06
1. UPDATE PER NEW BACKGROUNDS
2. REMOVE FUTURE NOTE & CAMERA
DRAWING ACCEPTANCE SIGN-OFF
NAME:_______________________SIGNATURE:__________________________
TITLE:________________________DATE:_____________
Customer or customer's representative has reviewed and approved the drawings which will be used as the basis for the conduit and electrical requirements for installation of the KARL STORZ Integration Equipment.
AV.IP.6.0.0 COVER SHEET
AV.IP.6.1.1 LEGENDS, SCHEDULES & RISERS
AV.IP.6.2.1 REQUIREMENTS DETAILS #1
AV.IP.6.3.1 OVERALL FLOOR PLAN - L - Partial
AV.IP.6.4.01 FLOOR PLAN & RCP - IR LAB #1
AV.IP.6.5.01 KARL STORZ PROJECT MANAGEMENT REQUIREMENTS
IR LAB #1
FACILITY REQUIREMENTS
OR1 NEO - IP SERIES SYSTEM®
DATE:
PLT SC
DATE:REVISIONS:
us ed fo r t he p ro je ct a nd s ite s pe ci fic al ly
Id en tif ie d he re in a nd is n ot to b e us ed o n an y ot he r p ro je ct
It is to b e re tu rn ed u po n re qu es t.
Th is d ra w in g is th e pr op er ty o f K A
R L
S
TO
R Z
E N
D O
S C
O P
Y, a nd is n ot to b e re pr od uc ed o r c op ie d in w ho le o r i n pa rt.
I t i s on ly to b e
C O
N
FI
D E
N
TI
A L
- N O
T
TO
B E
D
IS
TR
IB
U
TE
D O
U
TS
ID
E
O F
K A
R L
S
TO
R Z
AVIP_20180502
DRAWING NUMBER
CHK. BY:
OF
DATE:
JOB NO.
STAFFORD, TX 77477-5753
13803 N. PROMENADE
FX. 281.566.4109
PH. 281.566.4100
ENG. BY:
DRAWN BY:
FILE NAME:
VA
M ed ic al C en te r - W es t P al m
B ea ch , F
L
2024/11/06
2410-08-AVIP
KS-2410-08
O R
N
E O
-I P
S E
R
IE
S S
Y S
T E
M F A
C
IL
IT
Y
R E
Q U
IR
E
M E
N T S
N.T.S
SSatre
6.0.0F O
R C
O O
R D
IN
A
T
IO
N O
N L Y
N
O T F
O R
C O
N S
T R
U C
T
IO
N
AV.IP.
The designs shown on these drawings are planning concept designs only, are not drawn to scale, are not drawn as architectural plans or detailed engineering drawings, do not represent fully detailed working designs, and do not represent or incorporate detailed structural or engineering analysis.
These drawings are provided "as-is" and are intended to be used only as planning concept guides, and not for any other purpose. All warranties and representations of any kind, whether expressed, implied or statutory, including warranties of merchantability and fitness for a particular purpose, with regard to the designs or drawings by a recipient in any manner other than as herein intended. Any recipient of these drawings hereby agrees to indemnify KARL STORZ against any and all claims, liability, damages and costs (including attorneys' fees) arising out of, or resulting from, any use of these designs or drawings other than as herein intended.
0 REV. 0 2024/11/06
1 REV. 1 2024/12/03
2 REV. 2 2025/05/21
2 REVS REC. FROM D.S. ON 2025/05/21
A/V EQUIPMENT RACK (AVR): RACK SUPPLIED BY STOR=. STRUCTURAL SUPPORT BY C.C.(NOT STOR=)
RACK POWER RECEPTACLE: LOCATED AT A/V RACK FOR STOR= RACK POWER AND ASSOCIATED A/V EQUIPMENT. RACK RECEPTACLE TO BE
PROMINENTLY LABELED
RACK POWER
DISPLAY POWER RECEPTACLE AT RACK: LOCATED AT RACK FOR BOOM MOUNTED DISPLAYS AND/OR TOUCH PANELS WITH REMOTE ��VDC
POWER SUPPLIES. DISPLAY RECEPTACLE SHALL BE PROMINENTLY LABELED
DISPLAY POWER
BOOM JUNCTION BOXES: MOUNT ADJACENT TO BOOM SUPPORT STRUCTURE. PROVIDE CEILING ACCESS BELOW BOXES IF CEILING SPACE AT BOX LOCATION IS INACCESSIBLE AFTER FINISHED CEILING IS INSTALLED. SEE BOOM JUNCTION BOX DETAIL.
POWER CIRCUITS & RECEPTACLES IN EQUIPMENT HANDLER BOOMS: STOR= EQUIPMENT TYPICALLY REQUIRES (�)��A, 1��VAC CIRCUITS WITH (�) RECEPTACLES EACH. SEE BOOM SUPPLIER DRAWINGS FOR ELECTRICAL REQUIREMENTS AT BOOMS.
DATA/TELEPHONE BOX: DATA AND TELECOM CONNECTIVITY REQUIRED FOR STOR= INTEGRATION. PROVIDE SUITABLE IDENTIFICATION LABELS.
SEE TELECOM/DATA SCHEDULE FOR REQUIREMENTS.
BOOM MOUNTED SURGICAL LIGHT CAMERA: CONDUIT FROM THE SURGICAL LIGHT CAMERA CONTROL UNIT (VIDEO/CONTROL) OUTPUTS TO THE NURSE STATION IS REQUIRED FOR STOR= INTEGRATION. VERIFY OUTPUT LOCATION WITH SURGICAL LIGHT CAMERA CONTROL UNIT SUPPLIER.
SHEET NOTES �
NEO�IP BASE(AVR) EQUIPMENT
IN EQUIPMENT ROOM
J 1.�
JUNCTION BOX SCHEDULE
J�BOX � SI=E(MINIMUM) TYPE (NOTE 1 BELOW) MOUNTING/COMMENT (NOTE � BELOW)
J�1.� ��� X ��� X �� D
(MINIMUM)
NEMA 1, SCREW COVER,
EQUAL TO HOFFMAN A�SE SERIES
WALL, FLUSH, SI=E ACCORDING TO CONDUITS
REQUIREMENTS
J�BM.� �� X �� X �� D
NEMA 1, SCREW COVER,
EQUAL TO HOFFMAN A�SE SERIES
ABOVE CEILING, ADJACENT TO BOOM, ACCESS REQUIRED
JUNCTION BOX SCHEDULE NOTES:
1. ALL BOXES AND COVER PLATES ARE PROVIDED BY THE CLIENT
S CONTRACTOR UNLESS NOTED OTHERWISE.
�. MOUNTING REQUIREMENTS SHOWN ARE THE DEFAULT REQUIREMENTS UNLESS OTHERWISE NOTED/DETAILED.
POWER & HEAT LOAD SCHEDULE PER SYSTEM (1 SYSTEMS PER A/V RACK)
CIRCUIT
REF. �
VOLTS AMP DESTINATION LOAD
ELECT LOAD
(WATTS)
HEAT LOAD
(BTU)
1 1�� VAC �� RECEPTACLE # A/V RACK A/V RACK(IP) 11�1.�� ����.1�
� 1�� VAC �� RECEPTACLES # A/V RACK (�)DISPLAYS ��� 11��.�1
� 1�� VAC �� RECEPTACLES # NSTP & AIDA NSTP/AIDA/PRINTER 11�� ����.��
� 1�� VAC �� RECEPTACLES # WALL DISPLAY(S) (�)LCD ��� 1���.��
POWER SCHEDULE NOTES:
1. ALL A/V POWER CIRCUITS MUST COME FROM THE SAME ELECTRICAL PANEL THAT PROVIDES POWER TO THE PATIENT
VICINITY OUTLETS, BOOMS AND SERVICE COLUMNS
�. ALL A/V POWER CIRCUITS ARE TO HAVE FULL SI=E DEDICATED NEUTRALS.
�. CIRCUITS SHOWN ARE FOR DEDICATED USE WITH STOR= EQUIPMENT ONLY
�. A/V CIRCUITS SHALL BE ON THE SAME PHASE AND FEED FROM THE SAME ISOLATED POWER PANEL THAT FEEDS ALL THE
PATIENT�CRITICAL EQUIPMENT.
�. CIRCUIT BREAKER TRIP CURVE CHARACTERISTICS SHALL BE EQUAL TO OR GREATER THAN SQUARE�D QOB�HM SERIES.
A/V SYSTEM CONDUIT RISER
SCALE: N.T.S.
DATA SCHEDULE (CLIENT CABLING BY C.C.)
RUN � LOCATION C.C. TO PROVIDE
D1 DT OUTLET AT RACK (�) HOSPITAL NETWORK (FOR STOR= NETWK EQUIP/VIDEO CONF.) RJ���
D� DT OUTLET AT AIDA (�) HOSPITAL NETWORK (FOR STOR= NETWK EQUIP/VIDEO CONF.) RJ���
DATA TELECOM SCHEDULE NOTES:
1. CONNECTIONS REQUIRED FOR VIDEO STREAMING, REMOTE SUPPORT AND SOFTWARE UPDATE DELIVERY.
SYMBOL LEGEND
SYMBOL DESCRIPTION BOX (NOTE 1 BELOW) MOUNTING/COMMENT
CEILING ACCESS PANEL (NOTE � BELOW) FOR REQUIRED ACCESS TO BOXES ABOVE CLG.
EQUIPMENT LABEL DENOTES EQUIPMENT AT LOCATION SHOWN
CONDUIT OR CABLE RUN � CONDUIT/CABLE RUN REFERENCE �
CONDUIT STUB OUT CONDUIT STUB OUT WITH BUSHING
A/V IN/OUTPUT WALL BOX �S, TR�1G, 1G�B.C.P. WALL, FLUSH, RECEPTACLE HT., SEE �/AV.IP.�.�.1
A/V IN/OUTPUT WALL BOX �S, TR�1G, 1G�B.C.P. WALL, FLUSH, ��� A.F.F. HT., SEE �/AV.IP.�.�.1
A/V WALL DISPLAY WALL BOX �S, TR�1G, 1G�B.C.P. WALL, FLUSH, SEE ��/AV.IP.�.�.1
A/V PLASTIC WALL BOX �S, TR�1G, 1G�B.C.P. (PLASTIC) WALL, FLUSH, ��� A.F.F. HT., SEE �/AV.IP.�.�.1
JUNCTION BOX SEE J�BOX SCHEDULE SEE J�BOX SCHEDULE
CEILING SPEAKER CUTOUT (PER TEMPLATE) HOLE CUT IN SHEET ROCK BY C.C.(NOT STOR=). CUTOUT TEMPLATE
PROVIDED BY STOR=. SEE �/AV.IP.�.�.1
SURGICAL IN LIGHT CCU OUTPUT SEE VENDOR SPECIFICATIONS COORDINATE W/SURGICAL LIGHT VENDOR
DATA DROP A/R, TYPICAL 1G WALL, FLUSH, RECEPTACLE HT. (NOTE �)
DUPLEX POWER RECEPTACLE A/R, TYPICAL 1G WALL, FLUSH, RECEPTACLE HT.
QUAD POWER RECEPTACLE A/R, TYPICAL �G WALL, FLUSH, RECEPTACLE HT.
AP
1G
1G���
1G�D
1G�P
J
S1
ILC
SYMBOL LEGEND NOTES
1. ALL BOXES AND COVER PLATES ARE PROVIDED BY THE CLIENT
S CONTRACTOR UNLESS NOTED OTHERWISE.
�. MOUNTING REQUIREMENTS SHOWN ARE THE DEFAULT REQUIREMENTS UNLESS OTHERWISE NOTED/DETAILED.
�. BOXES/OUTLETS/RECEPTACLES LOCATED ABOVE THE CEILING MUST BE WITHIN 1� INCHES OF THE CEILING AND BE ACCESSIBLE FROM THE
SPACE BELOW. CLIENT / C.C. TO PROVIDE SUITABLE CEILING ACCESS PANELS AT DEVICE LOCATIONS. (ACCESS PANELS EQUAL TO WILLIAMS
BROS. MODEL WBFRC).
�. �RECEPTACLE HEIGHT� DENOTES MOUNTING HEIGHT OF REGULAR RECEPTACLES SPECIFIED FOR THIS PROJECT.
SWITCH HEIGHT� DENOTES MOUNTING HEIGHT OF REGULAR SWITCHES SPECIFIED FOR THIS PROJECT.
�. SEE DATA/TELECOM SCHEDULE FOR CLIENT CABLING REQUIREMENTS.
�. SEE STOR= STREAMCONNECT SYSTEM REQUIREMENTS FOR CLIENT REQUIREMENTS.
�. C.C. TO SUSPEND SPEAKER ASSEMBLY FROM STRUCTURE ABOVE CEILING.
�. BOX/OUTLET NUMBER: �1.�� FIRST NUMBER DENOTES BOX/OUTLET NUMBER. ��� DENOTES ROOM SEQUENCE.
BOX TYPES:
1G SINGLE GANG METALLIC BACK BOX (MINIMUM DEPTH ��1/��).
�G � GANG METALLIC BACK BOX (MINIMUM DEPTH ��1/��).
�G � GANG METALLIC BACK BOX (MINIMUM DEPTH ��1/��).
�S �� X �� SQUARE BACK BOX WITH TRIM RING AS NOTED (MINIMUM DEPTH ��1/��).
TR�1G TRIM RING FOR MOUNTING 1 GANG PLATE.
TR��G TRIM RING FOR MOUNTING � GANG PLATE.
�G�B.C.P. BLANK COVER PLATE, METALLIC, GANG SI=E AS DENOTED BY �.
RA
CK
J
UN
CT
IO
N
BO
X
C
�A
1G�P
BT/AUDIO
INPUT
.�1
1� 1/
C
TO
P W
R
CK
T
1�
1G�D
��� WALL
DISPLAY
.�1
FLEX
W
IR
E
OU
T
OF
C
ON
DU
IT
(T
YP
IC
AL
J�
BM
D
P1
J
DP .�
EQUIPMENT
HANDLER
BOOM
J�
BM
B
IMAGING
DISPLAY/
SHIELD/LIGHT
BOOM
J�
BM
C
IMAGING
DISPLAY/
SHIELD/LIGHT
BOOM
J�
BM
A
P�
J
DP .�
ANESTHESIA
BOOM
TO
P
W R
CK
T
1G
TOUCH
PANEL
OUTLET
.�1.�
1� 1/
C
CONTROL ROOM
1� 1/
C
1� 1/
C
1� 1/
C
1� 1/
C
1� 1/
C
TO
P W
R
CK
T
TO
D
AT
A
(D
(S
EE
D
AT
A
SC
HE
DU
LE
1G
AIDA/PRT
INPUT
1� 1/
C
1�
A
1G
PACS
INPUT
TO
P
W R
CK
T
(D
IS
PL
AY
P W
R)
(R
AC
K
PW
R)
TO
D
AT
A
(D 1)
(S
EE
D
AT
A
SC
HE
DU
LE
) TO
P W
R
CK
T
S1
S1
S1
S1
�A
�B �C 1/
C
1/
C
1/
C
1/
C
CEILING
SPEAKERS
FINISHED FLOOR
FINISHED CEILING
NSTPNSTP
NOTES:
1. MAXIMUM �� FT. CONDUIT LENGTH (1�� FT. CABLE LENGTH) ON RUNS FROM RACK TO OR EQUIPMENT.
�. SEE FIBER TRUNK & PATCH PANEL SPECIFICATIONS IF CONDUIT/CABLE LENGTH EXCEEDS MAXIMUMS.
�. DEDICATED � PPò � � PPò, MAX.� �,� PP GROUND WIRE FROM GROUND PANEL TO A/V RACK PROVIDED BY C.C. (NOT STOR=).
�. C.C. TO PROVIDE SCREW/PLATE ISOLATION GASKETS WHERE REQUIRED.
�. KARL STOR= OR INTEGRATION EQUIPMENT MUST OPERATE BETWEEN ���F TO 1���F. THE AIR TEMPERATURE AROUND THE EQUIPMENT
(CABINET/ENCLOSURE) SHOULD NOT EXCEED ���F TO ENSURE THAT THE EQUIPMENT DOES NOT EXCEED THE 1���F MAXIMUM OPERATING TEMPERATURE.
GENERAL NOTES & ABBREVIATIONS
GENERAL CONDUIT & BOX NOTES
1. C.C. TO PROVIDE BLANK COVER PLATES FOR ALL JUNCTION BOXES.
�. C.C. TO PROVIDE SCREW/PLATE ISOLATION GASKETS WHERE REQUIRED.
�. C.C. TO PROVIDE CABLE PROTECTION BUSHINGS ON ALL BOXES AND CONDUIT STUB�OUTS.
�. ALL CONDUIT TO BE MINIMUM �/�� METALLIC U.N.O.
�. MAXIMUM LENGTH FOR FLEX CONDUIT IF USED FOR STOR= A/V CABLING IS �� FEET.
�. EXTRA PULL BOX IS REQUIRED AFTER BENDS TOTAL ����
�. CONDUITS TO BE DE�BURRED, CLEANED, CAPPED, AND FURNISHED WITH NYLON PULL ROPE.
�. CONDUIT STUB�OUTS, JUNCTION BOXES, AND OUTLETS SHALL BE SUITABLY LABELED FOR IDENTIFICATION OF FUNCTION, J�BOX �, AND/OR
CONDUIT NUMBER.
GENERAL NOTES:
1. THESE DRAWINGS ARE FOR REQUIREMENTS REFERENCE ONLY. NOT FOR CONSTRUCTION.
�. SUPPLY AND INSTALLATION OF ALL CONDUIT, POWER CIRCUITS, BACK BOXES, ELECTRICAL DEVICES, ELECTRICAL WIRING IS TO BE PROVIDED BY THE
CLIENT
S ELECTRICAL CONTRACTOR UNLESS NOTED OTHERWISE.
�. KARL STOR= OR INTEGRATION EQUIPMENT MUST OPERATE BETWEEN ���F TO 1���F. THE AIR TEMPERATURE AROUND THE EQUIPMENT
(CABINET/ENCLOSURE) SHOULD NOT EXCEED ���F TO ENSURE THAT THE EQUIPMENT DOES NOT EXCEED THE 1���F MAXIMUM OPERATING
TEMPERATURE.
�. BE AWARE THE ´ABOVE THE CEILINGµ LOCATION HAS MANY RESTRICTIONS RELATED TO PLENUM RATED CABLES, MOUNTING DEVICES AND STERILE
FIELD =ONES.
ABBREVIATIONS:
A.F.F. ABOVE FINISHED FLOOR
A/R AS REQUIRED
B.C.P. BLANK COVER PLATE
B.F.C. BELOW FINISHED CEILING
C.C. CLIENT
S CONTRACTOR (NOT KARL STOR=)
C.P. COVER PLATE
CRE CONTROL ROOM EQUIPMENT
HT. HEIGHT
I.P. ISOLATED PWR. (UNGROUNDED NEUTRAL PWR)
NSE NURSE STATION EQUIPMENT
NSTP NURSE STATION TOUCH PANEL
N/A NOT APPLICABLE
N/R NOT REQUIRED
N/U NOT USED
O.F.E. OWNER FURNISHED EQUIPMENT
O.R. OPERATING ROOM
RJB RACK JUNCTION BOX
S.F. STERILE FIELD
T.B.D. TO BE DETERMINED
U.N.O. UNLESS NOTED OTHERWISE
SLC SURGICAL LIGHT CAMERA
MIN. MINIMUM
CKT CIRCUIT
AVR
1� 1/
C
1G
PHILLIPS
IN/OUTPUTS
IM
AG
IN
G
VE
ND
ER
CA
BL
E
DU
CT
IN
G M
AY
BE
US
ED
IN
L
IE
U
OF
S
PE
C.
C
�A
1G���
WIRELESS
MIC
INPUT
.�1
C
1G���
PTT MIC
INPUT
.�1
1� 1/
C
1�
1G
HIS
INPUT
S1
S1
A 1/
C
1/
C
CEILING
SPEAKERS
1� 1/
C
TO
P W
R
CK
T
1�
1G�D
��� WALL
DISPLAY
1� 1/
C
1�
1G
HIS
INPUT
1G
DISPLAY
POWER
1G
1� C
1� C
PLOT STAMP:SATRE, SHAUN _DATE: �/�1/���� 1�:�� AM
L E
G E
N D
S S
C H
E D
U L E
S
R
IS
E R
S
DATE:
PLT SC
DATE:REVISIONS:
us ed fo r t he p ro je ct a nd s ite s pe ci fic al ly
Id en tif ie d he re in a nd is n ot to b e us ed o n an y ot he r p ro je ct
It is to b e re tu rn ed u po n re qu es t.
Th is d ra w in g is th e pr op er ty o f K A
R L
S
TO
R Z
E N
D O
S C
O P
Y, a nd is n ot to b e re pr od uc ed o r c op ie d in w ho le o r i n pa rt.
I t i s on ly to b e
C O
N
FI
D E
N
TI
A L
- N O
T
TO
B E
D
IS
TR
IB
U
TE
D O
U
TS
ID
E
O F
K A
R L
S
TO
R Z
AVIPB��1�����
TKH GHVLJQV VKRZQ RQ WKHVH GUDZLQJV DUH SODQQLQJ FRQFHSW
GHVLJQV RQO\, DUH QRW GUDZQ WR VFDOH, DUH QRW GUDZQ DV
DUFKLWHFWXUDO SODQV RU GHWDLOHG HQJLQHHULQJ GUDZLQJV, GR QRW
UHSUHVHQW IXOO GHWDLO ZRUNLQJ GHVLJQV, DQG GR QRW UHSUHVHQW RU
LQFRUSRUDWH GHWDLO VWUXFWXUDO RU HQJLQHHULQJ DQDO\VLV. TKHVH
GUDZLQJV DUH SURYLGHG �DV�LV� DQG DUH LQWHQGHG WR EH XVHG
RQO\ DV SODQQLQJ FRQFHSW JXLGHV, DQG QRW IRU DQ\ RWKHU
SXUSRVH. AOO ZDUUDQWLHV DQG UHSUHVHQWDWLRQV RI DQ\ NLQG,
ZKHWKHU H[SUHVVHG, LPSOLHG RU VWDWXWRU\, LQFOXGLQJ ZDUUDQWLHV
RI PHUFKDQWDELOLW\ DQG ILWQHVV IRU D SDUWLFXODU SXUSRVH, ZLWK
UHJDUG WR WKH GHVLJQV RU GUDZLQJV E\ D UHFLSLHQW LQ DQ\
PDQQHU RWKHU WKDQ DV KHUHLQ LQWHQGHG. AQ\ UHFLSLHQW RI WKHVH
GUDZLQJV KHUHE\ DJUHHV WR LQGHPQLI\ KARL STOR= DJDLQVW DQ\
DQG DOO FODLPV, OLDELOLW\, GDPDJHV DQG FRVWV (LQFOXGLQJ
DWWRUQH\V
IHHV) DULVLQJ RXW RI, RU UHVXOWLQJ IURP, DQ\ XVH RI
WKHVH GHVLJQV RU GUDZLQJV RWKHU WKDQ DV KHUHLQ LQWHQGHG.
DRAWING NUMBER
CHK. BY:
OF
DATE:
JOB NO.
STAFFORD, TX 77477-5753
13803 N. PROMENADE
FX. 281.566.4109
PH. 281.566.4100
ENG. BY:
DRAWN BY:
FILE NAME:
� REV. � ����/11/��
1 REV. 1 ����/1�/��
� REV. � ����/��/�1
VA
M
HG
LF
DO
C
HQ
WH
U � W
HV
W P
DO
P
B
HD
FK
, F
L
2024/11/06
2410-08-AVIP
KS-2410-08
O R
N E
O
I P
S E
R
IE
S S
Y S
T E
M F A
C
IL
IT
Y
R E
Q U
IR
E
M E
N T S
N.T.S
SSatre
6.1.1F O
R C
O O
R D
IN
A
T
IO
N O
N L Y
N
O T F
O R
C O
N S
T R
U C
T
IO
N
AV.IP.
SIDE VIEW ISO VIEW
TOP VIEW
(WxHxD): 4.6"[117] x 2"[51] x 6.3"[160]
NOTE:
1. MIC WIRE 10' APPR.
2. MIC NECK 14"
PUSH-TO-TALKPTT
DETAIL - DESK TOP MIC
SCALE: N.T.S
'H'
FRONT VIEW
'W'
SIDE VIEW
'D'
NODE
StreamConnect Node (WxHxD): 5.9"[149] x 1.1"[29] x 8"[203]
MAX. WT. (LOADED) : 2.5 LBS
HEAT LOAD: 102.4 BTU
SCALE: N.T.S
DETAIL - StreamConnect NODE
OUTLET
MOUNT
14" [356]
SIDE VIEW FRONT VIEW
TYPICAL MOUNTING
HEIGHT = 6'-0" [1828] A.F.F.
FLOOR
BRACKET
STRUCTURE BLOCKING
NOTES:
1. DIMENSIONS FOR REFERENCE ONLY
2. C.C. TO PROVIDE ANCHORAGE TO SUPPORT DISPLAY & BRACKET.
3. C.C. TO MOUNT BRACKET TO STRUCTURE. (BRACKET SUPPLIED BY STORZ)
1G-D 4IN X 4IN BACKBOX
DUPLEX RECEPTACLE
'EQ' 'EQ'
WALL MOUNTED LCD DISPLAY
MODEL DIMENSION WEIGHT POWER VESA
FW65BZ40H 57.09"[1470] W x 32.8"[833] H x 2.76"[70] D 48.94LBS 214W(0.5W) 300x300mm
TYPE DESCRIPTION WEIGHT DEPTH
ST660 TILT: +15, -5 DEGREES 8LBS 2.6"
SEE OSHPD "OPM-0211-13" FOR ANCHORAGE(CA)
BRAVIA
SONY
DETAIL - WALL MOUNTED 4K DISPLAY - 65"
SCALE: N.T.S
DISP
DVI FIBER TXTX/RX
HEAT LOAD: 17.06 BTU
NOTES:
1. 3.57" W x 1.58" H x .65" D
DETAIL - Module-DVI-Fiber TX/RX
SCALE: N.T.S
9" [229]
5.4" [138]
CANARYBOXCANB
FRONT VIEW SIDE VIEW
WT. : 2 lbs (951g)(w/DESK STAND) HEAT LOAD: 27 BTU/hr4.2"
[108]
PLATE FRONT
CANARYBOX AUDIO
FROM CANARY
TO CANARY
NOTE: CANARYBOX SHOWN WITH 1G WALL/BOOM PLATE
PLATE SIDE
DETAIL - CANARYBOX & 1G WALL PLATE
SCALE: N.T.S
4" [103]
FRONT VIEW
7.2" [183]
TOP VIEW
4.6" [117]
WMR
WIRELESS MIC REC
MAX. WT. : 10.1 oz
MAX. PWR: .55 W
HEAT LOAD: 1.9 BTU
DETAIL - WIRELESS MIC RECEIVER
SCALE: N.T.S
J RACK J-BOX(J-1.#)
1.#
DISPLAY PWR
RACK PWR
D1
24.3" [616] 27.6" [702]
FRONT VIEW
WRK-3727
TOP VIEW
.9
CA
ST
ER
S)
A/V RACKAVR
NOTE:
1. WEIGHT & HEAT LOADS VARIES BASE
ON SYSTEM REQUIREMENTS.
2. CASTER BASE RACK REQUIRES 8" MIN.
ON ALL SIDES FOR COOLING.
3. AFFIXED RACK REQUIRES 24" MIN. ON
BACK & ONE SIDE FOR SERVICE.
SEISMIC CERTIFIED LOAD
CAPACITY: 900 lbs.
.9
SIDE VIEW
DETAIL - WRK-37-27 RACK (MEDIUM)
SCALE: N.T.S
B-MIC
FRONT VIEW
BOOM MICROPHONE
SEE SPEC SHEET FOR MORE DETAILS
15" [380]
GN50E
DET - BOOM MICROPHONE
SCALE: N.T.S
FRONT
SDI
SIDE
(HxWxD): 3.9"[99] x 2.8"[71] x 1.9"[48]HD/3G/12G-SDI
SCALE: N.T.S
DET - HD/3G/12G-SDI WPLT
'H'
FRONT VIEW
'W'
SIDE VIEW
'D'
NODE
StreamConnect Node (WxHxD): 5.9"[149] x 1.1"[29] x 8"[203]
MAX. WT. (LOADED) : 2.5 LBS
HEAT LOAD: 102.4 BTU
SCALE: N.T.S
DET - StreamConnect NODE
NEO 4K SYSTEM
6" [152]
CABLE
CLEAR
AVR
0]
6" MINIMUM CABLE CLEARANCE
REQUIRED
FRONT VIEW SIDE VIEW
TOP VIEW
9.1" [230]
MAX. WT. (LOADED) : 34.9 LBS
HEAT LOAD: 950 BTU
Air Humidity: 10%-80% RH, non-condensing
17.3" [440] 18" [457]
NOTE:
1.DEDICATED 4 mm² - 6 mm² MAX.Ø 5,9 MM
GROUND WIRE FROM GROUND PANEL
PROVIDED BY C.C. (NOT STORZ).
DET - NEO IP
SCALE: N.T.S
MIN
CABLE
CLEAR
POWER SUPPLY
6"
PS16B
9]
6" MINIMUM CABLE CLEARANCE
REQUIRED
FRONT VIEW SIDE VIEW
TOP VIEW
2.
1"
[5 3]
MAX. WT. (LOADED) : 9.6 LBS
HEAT LOAD: 655 BTU
17.3" [440] 16" [407]
B
DET - NEO PS16B
SCALE: N.T.S
NEO-IP 48 ASSM
6" [152] CABLE CLEAR
ASSM
.7
6]
FRONT VIEW SIDE VIEW
TOP VIEW
WEIGHT(UNLOAD):6.85 lbs MAX. WT.(LOAD): 25.6 lbs
(SWITCHER INSIDE
HEAT LOAD: 1185 BTU)
17.6" [447]
27.8" [706]
NOTES:
1. NOISE DAMPENING HOUSING FOR
SWITCHER.
3.4" [86]
DET - NEO-IP 48 ASSEMBLY
SCALE: N.T.S
UPM WITH ISOLATION
TRANSFORMER
SIDE VIEW
TOP VIEW
3.4" [86]17.3" [438] 20.7" [526]
6" [152]
CABLE
CLEAR
WEIGHT: 65 LBS.
PWR: 89 WATTS
HEAT LOAD: 304 BTU/HR
3" CLEARANCE ON ALL SIDES FOR
PROPER VENTILATION
FRONT VIEW
POWERVAR 58100-28R/58144-28R
DET - MEDICAL UPM
SCALE: N.T.S
UPM
8.1" [207]
FRONT VIEW
SONY UP-DR80MD
12.5" [317]
SIDE VIEW
SONY UP-DR80MD
16.8" [427]
TOP VIEW
SONY UP-DR80MD
.8
5]
AIDA PRINTER
MAX. WT. (LOADED) : 25.3 LBS
HEAT LOAD: 1160 BTU
NOTES:
1. 3" [76] MIN. CABLE CLEARANCE
APTR
NOTES:
1. CONNECTED TO AIDA THRU USB CABLE
DET - AIDA PRINT (SONY)
SCALE: N.T.S
7.6" [192]
FRONT VIEW
12" [305]
SIDE VIEW
14.9" [378]
TOP VIEW
.9
8]
IMAGE CAPTURE
MAX. WT. (LOADED) : 37.6 LBS
HEAT LOAD: 2757 BTU
AIDA
NOTES:
1. SHOWN WITH XPAND
6" [152]
CABLE
CLEAR
12" SmartScreen
AIDA WD 300
DET - AIDA BELLA
SCALE: N.T.S
'H'
FRONT VIEW
'W'
SIDE VIEW
Sager Electronics P/N: GSM120A-12-KS for (NEO IP TC1 & TC+)
'D'
NEO IP TCx & TC+ POWER SUPPLY (WxHxD): 1.4"[36] x 2.6"[66] x 6.6"[168]
HEAT LOAD: 102W - 348 BTU/HR
PWR
NOTES:
1. Output plug (Power 6 Pin Dual Row with Locking clip) Molex 39-01-2060
2. Mating plug (on TCx/TC+) Molex 39-01-3063
3. Customer side not provided with power supply
2 3 1 4
POWER LED
DET - NEO IP TCx/TC+ PWR SUPPLY
SCALE: N.T.S
NURSE STATION TOUCH PANELNSTP
NOTES:
1. (WxHxD): 23"[584] x 13.8"[368] x 4.4"[112] WITHOUT STAND
2. MAX. WT.: 20.8 LBS WITHOUT STAND
3. MOUNTING: VESA 100
'H'
FRONT VIEW
24" TOUCH PANEL
'W'
(WxHxD): 23"[584] x 16.4"[417] x 7.4"[188]
MAX. WT.: 25 LBS
POWER CONSUMPTION: <40W (MAX.)
MOUNTING: VESA 100
SIDE VIEW
24" TOUCH PANEL
'D'
WM101
+-MENU
DET - 24" NS TOUCH PANEL
SCALE: N.T.S
NEO IP THERMAL SPECS:
1. NEO IP SYSTEM COMPONENTS HAVE BEEN TESTED AND CERTIFIED TO PERFORM
RELIABLY AT A MAXIMUM AMBIENT AIR TEMPERATURE OF 40°C (104° F). THAT
MEANS THAT THE AIR TEMPERATURE AROUND THE NEO IP SYSTEM
COMPONENTS (SURROUNDING AIR TEMPERATURE ONE INCH DISTANCE FROM
THE COMPONENT CASES, NOT EXHAUST FAN OPENINGS) SHOULD NOT EXCEED
104°F. (THAT SAID…THE DEVICES WILL OPERATE ABOVE 104°F. AMBIENT BUT
WE CANNOT GUARANTEE FOR HOW LONG.)
2. NEO IP COMPONENTS CANNOT BE PLACED INSIDE SMALL CLOSED AREAS WITH
NO AIRFLOW…THIS WILL CAUSE A SLOW RISE IN TEMPERATURE POTENTIALLY
EXCEEDING THE MAX OPERATING TEMPERATURE OF 104°F.…. VENTILATION
MUST BE PROVIDED….INSTALLATION OF DOOR OR DRAWER COVER GRATES
ALLOW OUTSIDE ROOM AIR TO PASS THROUGH THE DOOR OR DRAWER COVER
INTO THE ENCLOSED AREA WHERE NEO IP COMPONENTS ARE INSTALLED.
3. SEE NEO IP SERVICE MANUAL FOR PROVIDE ADDITIONAL INSTRUCTIONS
CONCERNING THIS SUBJECT.
DET - NEO IP THERMAL SPECS
SCALE: N.T.S
FRONT SIDE
NOTE:
WUIS3619 FAILSAFE/2N1 PLATE
2N1 FRONT
2 N 1
POWER
VIDEO
DISPLAY
PORT
HDMI
SCALE: N.T.S
DET - ENDO CAMERA BOOM PLATES
REMOVABLE COVER
MIN. 18" X 18"[457]
ACCESS PANEL
WITHIN 3'[915] OF
A/V JUNCTION BOX
BOOM JUNCTION BOX LOCATED WITHIN 36"[915] FROM BOOM, A
MINIMUM OF 12"[305] AND A MAXIMUM OF 18"[457] FROM
CEILING. C.C. TO PROVIDE CABLE PROTECTION BUSHINGS FOR A/V
CABLE EXIT, SAME SIZE AS ENTRY CONDUITS.
CONDUIT AS SCHEDULED
CEILING
A/V CABLES (WITH
SERVICE LOOP)
BOOM SHROUD
BOOM
BOOM
STRUCTURE
DET - BOOM TYPICAL REQUIREMENTS
SCALE: N.T.S
'4S
BACKBOX
'TR-1G'
TRIM RING-
1 GANG
'1G-BCP'
1 GANG- BLANK
COVER PLATE
ASSEMBLY
WALL
4S
1G-BCP
TR-1G
ASSEMBLY SIDE
WALL OPENING CUT
TO BOX OPENING
NOTE:
1. 4"[102] SQUARE BACK BOX MINIMUM DEPTH 3-1/2"[88].
3/4"[19] MIN. CONDUIT
DET - 4S BOX WITH 1G TRIM RING
SCALE: N.T.S
ASSEMBLY
NOTE:
1. 4"[102] SQUARE BACK BOX MINIMUM DEPTH 3-1/2"[88].
WALL
4S
'4S'
BACK BOX
2G-BCP
TR-2G
ASSEMBLY SIDE
'TR-2G'
TRIM RING-
2 GANG
4IN X 4IN BACKBOX
'2G-BCP'
2GANG-BLANK
COVER PLATE
WALL OPENING CUT
TO BOX OPENING
3/4"[19] MIN. CONDUIT
DET - 4S BOX WITH 2G TRIM RING
SCALE: N.T.S
FRONT VIEW
RACK J-BOXES
NOTE: C.C. TO PROVIDE GROUND TERMINALS INSIDE J-BOX FOR RACK &
OR1 NURSE STATION.
CABLE OUT OF CONDUIT
TO RACK
18"[457]
TYP.
NOTE:
C.C. TO PROVIDE FRONT COVER WITH 4"
CABLE 'MOUSE HOLE' CUT-OUT & BUSHING
FOR CABLES
FLOOR
J RACK J-BOX(J1.#)
1.#
DISPLAY PWR
RACK PWR
(A/V RACK)
LOCATE DATA DROP
INSIDE J1 IF POSSIBLE. D1
DET - RACK J-BOX ELEVATION (TYP.)
SCALE: N.T.S
5.2" [132]
CEILING
NON-FLEX CONDUIT
1/2"[13] FLEX CONDUIT:
PROVIDE MINIMUM 3'[915] LENGTH BELOW
CEILING (TO ALLOW FOR INSTALLATION/SERVICE)
MAXIMUM 25'[7620] TOTAL FLEX LENGTH
SPEAKER AND GRILLE (PROVIDED AND
INSTALLED BY STORZ)
SIDE VIEW
S1
9.32" [237]
NOTES:
1. EACH SPEAKER WT. : 10.64 LBS. (AMP INCLUDED)
2. SPEAKER CUTOUT DIMENSIONS 9.32"[237] DIA.
3. C.C. (NOT STORZ) TO PROVIDE SPEAKER CUTOUT AND 4”[102] MINIMUM
AIRSPACE IN CEILING.
4. C.C. (NOT STORZ) TO PROVIDE SEISMIC WIRES TO SECURE SPEAKERS IN
ACCORDANCE WITH LOCAL CODE.
TCM-XA
SCALE: N.T.S
7 DET - CLG. SPEAKER/AMP (C-IC6PL)
FRONT VIEW
PLATE OVER TRIM
RING
LOW VOLTAGE
TRIM RING
'1G-BCP'
1 GANG- BLANK
COVER PLATE
SIDE VIEW
'LV-TR-1G'
LOW VOLTAGE
TRIM RING
NOTE: STUB CONDUIT 2" MIN. & 4" MAX. ABOVE TRIM RING.
WALL OPENING CUT TO
OPENING SIZE OF TRIM RING
WALL
TRIM RING
CONDUIT STUB IN
PLATE
PROTECTIVE
BUSHING
DET - LV TRIM RING AND PLATE
SCALE: N.T.S
NOTES:
1. UL APPROVED 1G/4S PLASTIC BOX W/1G-TR/1G-BCP
'4S'
BACK BOX
(PLASTIC)
'TR-1G'
TRIM RING-
1 GANG
(PLASTIC)
'1G-BCP'
1 GANG- BLANK
COVER PLATE
(PLASTIC)
ASSEMBLY
WALL
4S
1G-BCP
TR-1G
ASSEMBLY SIDE
WALL OPENING CUT
TO BOX OPENING
3/4" MIN. CONDUIT
SCALE: N.T.S
DET - 4S BOX WITH 1G TRIM RING (PLASTIC)
EN/DECODER RACK KIT
MAX. WT.: 13 LBS
ERK
FRONT VIEW SIDE VIEW
TOP VIEW
18.9" [481] 4.9"
[124]
7" [177]
R
ST
C O
N F
PW
R
VID
TX
R X
LIN
K
NEO IP-D
R
ST
C O
N F
PW
R
VID
TX
R X
LIN
K
NEO IP-D
R
ST
C O
N F
PW
R
VID
TX
R X
LIN
K
NEO IP-D
R
ST
C O
N F
PW
R
VID
TX
R X
LIN
K
NEO IP-D
R
ST
C O
N F
PW
R
VID
TX
R X
LIN
K
NEO IP-D
R
ST
C O
N F
PW
R
VID
TX
R X
LIN
K
NEO IP-D
R
ST
C O
N F
PW
R
VID
TX
R X
LIN
K
NEO IP-D
R
ST
C O
N F
PW
R
VID
TX
R X
LIN
K
NEO IP-D
DET - ENDCODER RACK KIT
SCALE: N.T.S
RST CONF
PWR VID TX RX
LINK
NEO IP-D
ENC
FRONT SIDE
(WxHxD): 6.2"[158] x 1.6"[41] x 5.2"[132]
MAX. WT. (LOADED) : 1.4 LBS
DET - 10G Encoders & Decoders
SCALE: N.T.S
SIDE VIEW
'D'
'H'
32" (TM343) DISPLAY
(WxHxD): 30"[762] x 18.3"[462] x 3.6"[91] DISPLAY WT.: 29.1 LBS(34.6 lbs with bracket & decoder) HEAT LOAD: 569.8 BTU/hr
DISP
NOTES:
1. VESA STANDARD 100 x 100 mm
'W'
FRONT VIEW
EIZO EX3242 (TM343)
PRESET INPUT MENU
F1 F2 F3 F4
ENTER
DET - 32" (TM343) BOOM DISPLAY
SCALE: N.T.S
'H'
FRONT VIEW
'W'
SIDE VIEW
XP POWER P/N: AHM250PS48T for (EIZO 32" TM343 DISPLAY)
'D'
TM343 POWER SUPPLY
(WxHxD): 1.5"[38] x 3.5"[89] x 8.8"[224]
HEAT LOAD: 167W - 569.8 BTU/HR
PWR
NOTES:
1. CLASS 1 DEVICE.
2. ATTACHED CORD 35"[889] LONG.
3. PLACE AT A/V RACK LOCATION.
DET - TM343 DISPLAY PWR SUPPLY
SCALE: N.T.S
FRONT
4N1
SIDE
'H'
'W'
(WxHxD): 2.85"[72] x 3.82"[97] x 4.63"[117.49]
'D'
POWER
LINK
IR
DVI
COMPOSITE
VGA
S-VIDEO
4 N 1
DET - 4N1 INTERFACE (1G) WPLT
SCALE: N.T.S
FRONT
2N1
SIDE
'H'
'W' 'D'
2 N 1
POWER
VIDEO
DISPLAY
PORT
HDMI
(WxHxD): 3"[76] x 4.6"[117] x 3.6"[92]
SCALE: N.T.S
DET - 2N1 INTERFACE (1G) WPLT
FRONT
BT/AUD
SIDE
(WxHxD): 2.8"[71] x 4.5"[114] x 1.8"[46]
NOTE:
1. UN/CSA APPROVED 18CI PLASTIC
BOX/1G-TR/1G-BCP
AUDIO
RECEIVER
AUDIO IN
4”
KBSOUND
SCALE: N.T.S
DET - BLUETOOTH/AUDIO WPLT RECIEVER
FRONT SIDE
GROM
(HxWxD): 4.5"[114] x 2.8"[71] x .5"[13]
SCALE: N.T.S
DET - FLEX GROMMET 1G PLATE
FRONT SIDE
GROM
(HxWxD): 4.5"[114] x 4.5"[114] x .5"[13]
SCALE: N.T.S
DET - FLEX GROMMET 2G PLATE
DISPLAY POWER
J-BOX
1" FLEX CONDUIT (BY E.C.)
CLASS 1 POWER CABLES
4#10 THHN STRANDED (BY STORZ)
I" CONDUIT W/BUSHING (BY E.C.)
FACTORY PWR SUPPLY 95" CABLE & CONNECTOR
DISPLAY POWER
SUPPLY
AV RACK / NURSE STATION
1 GANG DISPLAY POWER
PLATE WITH MATING
CONNECTOR FOR DISPLAY
POWER SUPPLY FACTORY
CABLE (BY STORZ)
STORZ
DISPLAY
LEAD WIRE HOOK UP
PIN # COLOR FUCTION
RED +24VDC
WHITE +24VDC
BLACK RETURN
GREEN RETURN
1 DRAIN GROUND
4# 10 THHN STRANDED
COLOR FUCTION
RED +24VDC
BLACK GROUND
GREEN EARTH
WHITE EARTH-SHIELD
FRONT
2 1
SIDE
DISPLAY POWER
24VDC
AVR
FDP
DP.#J
1G 1.#
DET - FLORIDA DISPLAY POWER (AHCA)
SCALE: N.T.S
PLOT STAMP:SATRE, SHAUN |DATE: 5/21/2025 10:32 AM
R E
Q U
IR
E
M E
N T S
D E
T A
IL
S
WARNING:
KARL STORZ OR INTEGRATION EQUIPMENT MUST OPERATE BETWEEN 32°F TO 104°F. THE AIR TEMPERATURE AROUND THE EQUIPMENT (CABINET/ENCLOSURE) SHOULD NOT EXCEED 90°F TO ENSURE THAT THE EQUIPMENT DOES NOT EXCEED THE 104°F MAXIMUM OPERATING TEMPERATURE.
DATE:
PLT SC
DATE:REVISIONS:
us ed fo r t he p ro je ct a nd s ite s pe ci fic al ly
Id en tif ie d he re in a nd is n ot to b e us ed o n an y ot he r p ro je ct
It is to b e re tu rn ed u po n re qu es t.
Th is d ra w in g is th e pr op er ty o f K A
R L
S
TO
R Z
E N
D O
S C
O P
Y, a nd is n ot to b e re pr od uc ed o r c op ie d in w ho le o r i n pa rt.
I t i s on ly to b e
C O
N
FI
D E
N
TI
A L
- N O
T
TO
B E
D
IS
TR
IB
U
TE
D O
U
TS
ID
E
O F
K A
R L
S
TO
R Z
AVIP_20180502
The designs shown on these drawings are planning concept designs only, are not drawn to scale, are not drawn as architectural plans or detailed engineering drawings, do not represent full detail working designs, and do not represent or incorporate detail structural or engineering analysis. These drawings are provided "as-is" and are intended to be used only as planning concept guides, and not for any other purpose. All warranties and representations of any kind, whether expressed, implied or statutory, including warranties of merchantability and fitness for a particular purpose, with regard to the designs or drawings by a recipient in any manner other than as herein intended. Any recipient of these drawings hereby agrees to indemnify KARL STORZ against any and all claims, liability, damages and costs (including attorneys' fees) arising out of, or resulting from, any use of these designs or drawings other than as herein intended.
DRAWING NUMBER
CHK. BY:
OF
DATE:
JOB NO.
STAFFORD, TX 77477-5753
13803 N. PROMENADE
FX. 281.566.4109
PH. 281.566.4100
ENG. BY:
DRAWN BY:
FILE NAME:
0 REV. 0 2024/11/06
1 REV. 1 2024/12/03
2 REV. 2 2025/05/21
VA
M ed ic al C en te r - W es t P al m
B ea ch , F
L
2024/11/06
2410-08-AVIP
KS-2410-08
O
O
I P
S E
R
IE
S S
Y S
T E
M F A
C
IL
IT
Y
R E
Q U
IR
E
M E
N T S
N.T.S
SSatre
6.2.1F O
R C
O O
R D
IN
A
T
IO
N O
N L Y
N
O T F
O R
C O
N S
T R
U C
T
IO
N
AV.IP.
2 REMOVE ROOM CAMERA
11.8
11.8
12 13 14
DD
EE
FF
GG
C.8C.8
E.8E.8
F
C 1350
MD 26+
C 15
C
L 21
C 1800
L 21
C 1950
A
B
C
D
C us to m er 's
P hy si o
(P H
Y)
O pt io na l ( U
PS
C oo rd in at e fin al lo ca tio n w ith cu st om er if or de re d
STERILE
SUPPLIES
2B-234
CONTROL
ROOM
2B-238
Corridor 2B-C7
Corridor 2B-C4
HYBRID OR
2B-237
TRANSITION
AREA
2B-236
TRANSITION
AREA
2B-223
SEMI-RESTRICTED
AREA
2B-235
ICR
2B-239
EQPT.
ALCOVE
2B-238A
PLOT STAMP:SATRE, SHAUN |DATE: 5/21/2025 10:31 AM
O V
E R
A L L F
L O
O R
P L A
N
L
P a rt ia l
1 OVERALL FLOOR PLAN - L - PARTIAL
SCALE: 1/8" = 1'-0"
BACKGROUND REF.: WPB VA RENOVATE IR CATH LAB_FLOOR PLAN 2024-11-12.DWG RECEIVED 2024/11/12
SEE AV.EN.6. .4 01
DATE:
PLT SC
DATE:REVISIONS:
us ed fo r t he p ro je ct a nd s ite s pe ci fic al ly
Id en tif ie d he re in a nd is n ot to b e us ed o n an y ot he r p ro je ct
It is to b e re tu rn ed u po n re qu es t.
Th is d ra w in g is th e pr op er ty o f K A
R L
S
TO
R Z
E N
D O
S C
O P
Y, a nd is n ot to b e re pr od uc ed o r c op ie d in w ho le o r i n pa rt.
I t i s on ly to b e
C O
N
FI
D E
N
TI
A L
- N O
T
TO
B E
D
IS
TR
IB
U
TE
D O
U
TS
ID
E
O F
K A
R L
S
TO
R Z
AVIP_20180502
The designs shown on these drawings are planning concept designs only, are not drawn to scale, are not drawn as architectural plans or detailed engineering drawings, do not represent full detail working designs, and do not represent or incorporate detail structural or engineering analysis. These drawings are provided "as-is" and are intended to be used only as planning concept guides, and not for any other purpose. All warranties and representations of any kind, whether expressed, implied or statutory, including warranties of merchantability and fitness for a particular purpose, with regard to the designs or drawings by a recipient in any manner other than as herein intended. Any recipient of these drawings hereby agrees to indemnify KARL STORZ against any and all claims, liability, damages and costs (including attorneys' fees) arising out of, or resulting from, any use of these designs or drawings other than as herein intended.
DRAWING NUMBER
CHK. BY:
OF
DATE:
JOB NO.
STAFFORD, TX 77477-5753
13803 N. PROMENADE
FX. 281.566.4109
PH. 281.566.4100
ENG. BY:
DRAWN BY:
FILE NAME:
0 REV. 0 2024/11/06
1 REV. 1 2024/12/03
2 REV. 2 2025/05/21
VA
M ed ic al C en te r - W es t P al m
B ea ch , F
L
2024/11/06
2410-08-AVIP
KS-2410-08
O
O
I P
S E
R
IE
S S
Y S
T E
M F A
C
IL
IT
Y
R E
Q U
IR
E
M E
N T S
N.T.S
SSatre
6.3.1F O
R C
O O
R D
IN
A
T
IO
N O
N L Y
N
O T F
O R
C O
N S
T R
U C
T
IO
A/V EQUIPMENT SCHEDULE NOTES:
1. EQUIPMENT LISTED IS FOR COORDINATION PURPOSES ONLY. (BASED ON STORZ'S UNDERSTANDING OF EQUIPMENT & LOCATIONS). REFER TO CONTRACT SPECIFICATIONS
FOR ACTUAL REQUIREMENTS.
WALL MOUNTED WIRELESS MICROPHONE RECEIVER INPUT
WALL MOUNTED PUSH TO TALK MICROPHONE INPUT
WALL MOUNTED REMOTE (DP)COMPUTER INPUT PACS COMPUTER & DATA DROP
NURSE'S TOUCH PANEL A/V TOUCH PANEL/KEYBOARD/MOUSE
WALL MOUNTED REMOTE (DP)COMPUTER INPUT HIS COMPUTER, POWER & DATA DROP
WALL MOUNTED LARGE FORMAT FLAT PANEL DISPLAY BRACKET ANCHORAGE BY C.C. (NOT STORZ)
WALL MOUNTED (FOR DATA TELECOM CONNECTION) SEE DATA / TELECOM SCHEDULE
CEILING MOUNTED (6) CEILING SPEAKER ASSEMBLY,
CEILING SPEAKER HOLE TEMPLATE
HOLE CUT IN SHEETROCK BY C.C. (NOT STORZ).
CUTOUT TEM1G PLATE PROVIDED BY STORZ.
A/V EQUIPMENT RACK A/V EQUIPMENT RACK (NEO - IP) ,
UPS / ISOLATION TRANSFORMER,
10G NETWORK (24/48 PORT) SWITCHER,
NEO-PS16+ POWER SUPPLY,
(8) NEO ENCODER/DECODER
WALL MOUNTED BLUETOOTH AUDIO RECEIVER INPUT 1G PLATE PLASTIC BOX & TRIM RING
WALL MOUNTED AIDA BELLA (AIDA/PRINTER) DATA DROP
EQUIPMENT BOOM AUX INPUT (ACC/FAILSAFE) 1G PLATE (ENDO),
(2) AUX INPUT 2N1 (DisplayPort/HDMI) 1G PLATE (ENDO/AUX), AUX INPUT 4N1 (CV/SV/VGA/DVI) 1G PLATE, AUX INPUT (HD/3G/12G-SDI) 1G PLATE
DISPLAY MOUNT & SURGICAL LIGHT
IMAGING
DISPLAY/LIGHT/SHIELD BOOM
AUX INPUT 2N1 (DisplayPort/HDMI) 1G PLATE, AUX INPUT 4N1 (CV/SV/VGA/DVI) 1G PLATE, BOOM MICROPHONE
DISPLAY MOUNT & SURGICAL LIGHT
IMAGING
DISPLAY/LIGHT/SHIELD BOOM
AUX INPUT 2N1 (DisplayPort/HDMI) 1G PLATE, AUX INPUT 4N1 (CV/SV/VGA/DVI) 1G PLATE, BOOM MICROPHONE
DISPLAY MOUNT & SURGICAL LIGHT
ANESTHESIA BOOM AUX INPUT 2N1 (DisplayPort/HDMI) 1G PLATE, AUX INPUT 4N1 (CV/SV/VGA/DVI) 1G PLATE, AUX INPUT (HD/3G/12G-SDI) 1G PLATE
VITALS COMPUTER
SYMBOL LOCATION/DEVICE EQUIPMENT BY STORZ EQUIPMENT BY OTHERS
A/V EQUIPMENT SCHEDULE
WMR
PTT
PACS
NSTP
HIS
DISP
DAT
CSPK
AVR
AUD-R
AIDA
D
C
B
A
11.8
11.8
C 1350
MD 26+
C 15
C
L 21
C 1800
L 21
C 1950
A
B
C
D
J .
NEO IP
DISP PWR
RACK PWR
1 01
D1
AVR
DAT
1G
-D
DI
SP
HI
S
1G
1G
-D
DI
SP 55
1G
NS
TP
1G -P
AU
D-
R
PACS
0141G
HIS
0141G
PTT
0111G-36
WMR
0121G-36
IMG
0161G
IMAGING VENDER
CABLE DUCTING MAYBE
USED IN LIEU OF SPEC.
DA
T
AI
DA
D2 1G
. 0181G 0171G
FDP
C us to m er 's
P hy si o
(P H
Y)
O pt io na l ( U
PS
C oo rd in at e fin al lo ca tio n w ith cu st om er if or de re d
STERILE
SUPPLIES
2B-234
CONTROL
ROOM
2B-238
HYBRID OR
2B-237
TRANSITION
AREA
2B-236
SEMI-RESTRICTED
AREA
2B-235
ICR
2B-239
EQPT.
ALCOVE
2B-238A
11.8
11.8
C 1350
MD 26+
C 15
C
L 21
C 1800
L 21
C 1950
JJ
.DP 01BM 01
A
J
BM 02
B
J
BM 03
C
JJ
.DP 02BM 04
D
(TYP. 4X)
CSPK S1
S1 S1
S1
(TYP. 4X)
CSPK
S1
S1
C us to m er 's
P hy si o
(P H
Y)
O pt io na l ( U
PS
C oo rd in at e fin al lo ca tio n w ith cu st om er if or de re d
STERILE
SUPPLIES
2B-234
CONTROL
ROOM
2B-238
HYBRID OR
2B-237
TRANSITION
AREA
2B-236
SEMI-RESTRICTED
AREA
2B-235
ICR
2B-239
EQPT.
ALCOVE
2B-238A
PLOT STAMP:SATRE, SHAUN |DATE: 5/21/2025 10:32 AM
F L O
O R
P L A
N
R C
P
I R
L A
B
1 FLOOR PLAN - IR LAB #1
SCALE: N.T.S
BACKGROUND REF.: N-SOU240607.DWG.DWG RECEIVED 2024/12/03
2 REFLECTED CEILING PLAN - IR LAB #1
SCALE: N.T.S
BACKGROUND REF.: 940537-1-05.DWG.DWG RECEIVED 2024/12/03
DATE:
PLT SC
DATE:REVISIONS:
us ed fo r t he p ro je ct a nd s ite s pe ci fic al ly
Id en tif ie d he re in a nd is n ot to b e us ed o n an y ot he r p ro je ct
It is to b e re tu rn ed u po n re qu es t.
Th is d ra w in g is th e pr op er ty o f K A
R L
S
TO
R Z
E N
D O
S C
O P
Y, a nd is n ot to b e re pr od uc ed o r c op ie d in w ho le o r i n pa rt.
I t i s on ly to b e
C O
N
FI
D E
N
TI
A L
- N O
T
TO
B E
D
IS
TR
IB
U
TE
D O
U
TS
ID
E
O F
K A
R L
S
TO
R Z
AVIP_20180502
The designs shown on these drawings are planning concept designs only, are not drawn to scale, are not drawn as architectural plans or detailed engineering drawings, do not represent full detail working designs, and do not represent or incorporate detail structural or engineering analysis. These drawings are provided "as-is" and are intended to be used only as planning concept guides, and not for any other purpose. All warranties and representations of any kind, whether expressed, implied or statutory, including warranties of merchantability and fitness for a particular purpose, with regard to the designs or drawings by a recipient in any manner other than as herein intended. Any recipient of these drawings hereby agrees to indemnify KARL STORZ against any and all claims, liability, damages and costs (including attorneys' fees) arising out of, or resulting from, any use of these designs or drawings other than as herein intended.
DRAWING NUMBER
CHK. BY:
OF
DATE:
JOB NO.
STAFFORD, TX 77477-5753
13803 N. PROMENADE
FX. 281.566.4109
PH. 281.566.4100
ENG. BY:
DRAWN BY:
FILE NAME:
0 REV. 0 2024/11/06
1 REV. 1 2024/12/03
2 REV. 2 2025/05/21
VA
M ed ic al C en te r - W es t P al m
B ea ch , F
L
2024/11/06
2410-08-AVIP
KS-2410-08
O
O
I P
S E
R
IE
S S
Y S
T E
M F A
C
IL
IT
Y
R E
Q U
IR
E
M E
N T S
N.T.S
SSatre
6.4.01F O
R C
O O
R D
IN
A
T
IO
N O
N L Y
N
O T F
O R
C O
N S
T R
U C
T
IO
N
AV.IP.
2 REMOVE ROOM CAMERA
KARL STORZ SITE READINESS CHECKLIST
OVERALL SITE STATE:
Upon KARL STORZ’S arrival at the facility, the site must be clean with all site prep work completed, the boom equipment installed, and the nurse station millwork complete. All owner furnished equipment must be available for installation and testing.
Although other trades may work in the same space, priority must be given to the floor space requirements of staging equipment and materials and the ability of the crew to maintain their workflow.
GENERAL SITE PREP WORK:
□ All required conduits must be marked and installed with pull strings.
□ All facility provided network cables must be installed and tested.
□ All electrical outlets on equipment handlers, nurse millwork and a/v rack location must have power.
□ Permanent power must be energized.
□ Ceiling speaker cutouts must be prepared.
□ Wall mounts for flat panel displays and touch panels must be installed.
□ All network data drops need to be installed.
□ Equipment locations must be well ventilated as per the BTU specification provided.
MILLWORK:
□ Millwork must be installed with proper venting for calculated heat loads per the KARL STORZ drawings.
□ Cover plate of large junction box behind a/v rack has mouse hole cut in proper location.
□ Facility provided carpenter must be available for any grommet placement.
BOOM EQUIPMENT:
□ Boom equipment must be installed and available for cabling by the KARL STORZ team.
□ Boom equipment must be powered.
□ In-Light Camera boxes must be available.
OWNER FURNISHED EQUIPMENT:
□ OR equipment such as C-Arms, Ultrasound, CT Machines and Microscopes must be available for testing and integration.
□ Computers and workstations must be available for cable integration and testing.
□ Videoconference telephone or network connections must be live.
□ If applicable, all lighting control equipment must be installed and tested by the facility.
ENVIRONMENTAL CONDITIONS:
□ Site is safe to work.
□ If site has specific attire requirements and/or PPE requirements, the customer must provide access to appropriate attire and PPE for the crew.
□ The OR must be clean and dust free.
□ Delivery access must be available.
□ Area to store all KARL STORZ materials for duration of installation identified.
□ Trash disposal must be available.
□ All safety training must be scheduled prior to arrival.
□ Any sub-contractor conflicts need to be addressed and conflicting schedules need to be addressed to minimize KARL STORZ downtime.
□ Adequate working hours must be available to the product deployment crew.
Use this document to ensure that all teams and spaces are aligned for the deployment of your KARL STORZ integrated operating room.
Failure to meet readiness can impact your installation schedule.
Thank you for your partnership, KARL STORZ
97208100 OR1 1 2.0 us 11/2022/SE-E-US
PLOT STAMP:SATRE, SHAUN |DATE: 5/21/2025 10:32 AM
K A
R L S
T O
R Z
P R
O
JE
C T M
A N
A G
E M
E N
T R
E Q
U
IR
E M
E N
T S
DATE:
PLT SC
DATE:REVISIONS:
us ed fo r t he p ro je ct a nd s ite s pe ci fic al ly
Id en tif ie d he re in a nd is n ot to b e us ed o n an y ot he r p ro je ct
It is to b e re tu rn ed u po n re qu es t.
Th is d ra w in g is th e pr op er ty o f K A
R L
S
TO
R Z
E N
D O
S C
O P
Y, a nd is n ot to b e re pr od uc ed o r c op ie d in w ho le o r i n pa rt.
I t i s on ly to b e
C O
N
FI
D E
N
TI
A L
- N O
T
TO
B E
D
IS
TR
IB
U
TE
D O
U
TS
ID
E
O F
K A
R L
S
TO
R Z
AVIP_20180502
The designs shown on these drawings are planning concept designs only, are not drawn to scale, are not drawn as architectural plans or detailed engineering drawings, do not represent full detail working designs, and do not represent or incorporate detail structural or engineering analysis. These drawings are provided "as-is" and are intended to be used only as planning concept guides, and not for any other purpose. All warranties and representations of any kind, whether expressed, implied or statutory, including warranties of merchantability and fitness for a particular purpose, with regard to the designs or drawings by a recipient in any manner other than as herein intended. Any recipient of these drawings hereby agrees to indemnify KARL STORZ against any and all claims, liability, damages and costs (including attorneys' fees) arising out of, or resulting from, any use of these designs or drawings other than as herein intended.
DRAWING NUMBER
CHK. BY:
OF
DATE:
JOB NO.
STAFFORD, TX 77477-5753
13803 N. PROMENADE
FX. 281.566.4109
PH. 281.566.4100
ENG. BY:
DRAWN BY:
FILE NAME:
0 REV. 0 2024/11/06
1 REV. 1 2024/12/03
2 REV. 2 2025/05/21
VA
M ed ic al C en te r - W es t P al m
B ea ch , F
L
2024/11/06
2410-08-AVIP
KS-2410-08
O
O
I P
S E
R
IE
S S
Y S
T E
M F A
C
IL
IT
Y
R E
Q U
IR
E
M E
N T S
N.T.S
SSatre
6.5.01F O
R C
O O
R D
IN
A
T
IO
N O
N L Y
N
O T F
O R
C O
N S
T R
U C
T
IO