531-325 - Specs-Appx-C.pdf
PDF 2 MB Posted
- Attached to
- Y1DZ--TIERED EVALS - Construct Primary Care/Mental Health Building Minor Construction - Boise 531-325
- Solicitation number
- 36C77624R0096
About this file
This document appears to be a set of technical drawings and specifications for a patient lift system to be installed as part of the Construct Primary Care / Mental Health Building project at the Boise VA Medical Center. The key details include:
The system will use Guldmann GH-3 ceiling-mounted patient lifts with a 550 lb capacity. The drawings provide specifications for the lift system layout, attachment to the building's steel web truss structure, power supply, and safety requirements. The system will be installed in multiple patient rooms and other areas across the facility. The drawings indicate the project is part of Project 531-325 "Construct Primary Care/Mental Health Building Minor Construction" at the Boise VA Medical Center. The associated federal contract opportunity is Solicitation Number 36C77624R0096, which seeks construction services for this project.
View the file
Other files
Show all 40
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
CONSTRUCT PRIMARY CARE / MENTAL HEALTH BUILDING PROJECT #: 531-325
BOISE VA MEDICAL CENTER - BOISE, ID CONSTRUCTION DOCUMENTS
APPENDIX C PATIENT LIFT DETAILS
APPENDIX C:
PATIENT LIFT DETAILS
Guldmann Inc.
14401 McCormick Drive Tampa FL, 33626 Unit A
Toll Free: 1(800)-664-8834 Fax: 1(813)-880-9558
Email: Info@Guldmann.net
NOTES
APPROVED BY:
PROJECT NAME:
SHEET DESCRIPTION:
SHEET NO:
SIGNOFF / APPROVAL NOTES
1. The facility structure capacity must be able to support the load specified on the installation drawing.
2. Electrical power source provided by the facility, position to be verified in the field.
3. Guldmann Inc. reserves the right to slightly modify the track layout if necessary without Facility approval.
This includes lowering the system to avoid ceiling objects.
4. Guldmann Inc. will not be held responsible for any damages, inconvenience, resulting from hitting any non visible conduits during our installation process of which we have not been advised.
5. Modifications of door headers by the facility.
6. Signed drawings are required prior to ordering and installation of ceiling lifts.
This signature of each drawing confirms agreement of all the conditions listed above.
7. Systems are tested at 1.5 times capacity.
550 lbs. tested at 825 lbs.; 770 lbs. tested at 1155 lbs.; 825 lbs. tested at 1238 lbs.; 1100 lbs. tested at 1650 lbs.
MY SIGNATURE BELOW VERIFIES THAT I HAVE:
Reviewed all _____ pages, with a total quantity of _____ room(s), _____ system(s) for the approbation of layout for each room and the approbation of structure for each room.
CUSTOMER
APPROVAL: _____________________________________ DATE: _______________
DATE: DRAWN BY:
ACCURATE TO 4" BASED
ON PDF CONVERSION
Boise VA Boise PCMH
500 W Fort St.
Boise, ID 83702
PROJECT ID:102460
Boise PCMH
FACILITY SIGNOFF
GUL101
TRC04/19/2021
7 2 2
HOIST = 550 LBS1.
FLUSH RECESSED / TYPE G2.
CEILING HEIGHT = 9' - 0"3.
AFF TO B.O.S = 14' - 6"
ATTACHMENT TYPE = WEB TRUSS4.
SPAN = 5' - 4 1/4"
Revision Schedule
Date Issued by Letter Description
05/28/2021 TRC A RM. 118 RE-CONFIGURED / RMS 163 & 145 REMOVED AND SET TO ALT
06/14/2021 CML B NEW BACKGROUNDS / RM 136 REPLACED WITH RM 163
06/18/2021 CML C ROOMS 136 & 145 READDED / PROJECT SPLIT IN TWO
EACH FIXED RAIL TO
HAVE x1 PARALLEL & x2 PERPENDICULAR
STABILIZERS
ALL JOINTS MUST HAVE x1 PARALLEL & x1
PERPENDICULAR STABILIZER
SEE SPAN CHART
S E
E S
P A
N C
H A
R T
SEE SPAN CHART
± 22°
STABILIZERS CAN BE ROTATED 180°
Fixed rail systems Fixed rail systems must be connected by Guldmann Service Team or by a qualified installer.
A Guldmann class I transformer must be used.
Installation to follow existing rules for authorized electricians and according to IEC/UL 60601-1.
Terminals on transformers Class I Double-insulated UL approved cable must be used (Cable specification: UL category: AVLV2, min.300V, min.
80° C, min. 18AWG (1 mm2 )) Transformer DK-13991 In:
Terminal 1 = N Terminal 2 = 100-115V AC Terminal 3 = 230V AC Terminal 4 = Protective ground Out:
Terminal 5 = 33V AC Terminal 6 = 33V AC
POWER TO BE WITHIN 15 FT OF END OF FIXED RAIL
Guldmann permanent mounted ceiling aluminium rail
Hoist with trolleys movable in rail
Protective Plast Enclosure
UL approved 400V cable, green, min. 14 AWG
Protective Ground
M1
UL approved 400V double isolated cable, min. 18 AWG
ON/OFF
UL approved 400V double isolated cable, min. 18 AWG
T ra n s fo rm e r
D K
-1
Main power Permanent installation Specification of breaker:
> 1 Amp capacity Line must always be broken
L1
AC/DC charger
Main Motor Control
Emergency Stop
Safety Circuit Emergency down
Motor Function Selector Switch0 1
V A
C
U S
V
A C
P ro te c ti ve
G ro u n d
V
A C
V
A C1 A
T
.5
A T
Green Light ON (or ON / OFF - sign)
MAX
PROJECTION
MAX DISTANCE BETWEEN VERTICAL SUPPORTS
A
B
C
A
B
C
A
B
C
A
B
C
A
B
C
A
B
C
X
X
X
RAIL
RAIL
RAIL
RAIL
RAIL
RAIL
INCH
METER
,23
,7
1,2
39 49 59 68 78 118 137 147 157 177 187 206 215 236 246 264
1,0 1,25 1,5 1,75 2,0 3,0 3,5 3,75 4,0 4,5 4,75 5,25 5,5 6,0 6,25 6,75
MAX. 175 KG (385 LBS.)
MAX. 205 KG (450 LBS.)
MAX. 255 KG (560 LBS.)
MAX. 300 KG (660 LBS.)
MAX. 350 KG (770 LBS.)
MAX. 375 KG (825 LBS.)
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
A-RAIL
WEIGHT: 2.6 KG/M
(1.75 LBS/FT)
B-RAIL
WEIGHT: 5.0 KG/M
(3.36 LBS/FT)
C-RAIL
WEIGHT: 6.0 KG/M
(4.03 LBS/FT)
(2 1/2")
62 mm.
(2
/4
,5 m m
(2 1/2")
62 mm.
(5 m m
(2 1/2")
62 mm.
(6
/3 m m
A
B
C
RAIL MAX. 400 KG (880 LBS.)
X
X
X
A
B
C
RAIL MAX. 500 KG (1100 LBS.)
X
X
X
7,25
8,0
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X m m
/2 m m
/4
,5 m m
M
IN
m m
/4
DEPTH = 205 mm (8 1/4")
345 mm (13 3/4")
A
B
C
D
E
F
G
H
I
J
K
Ceiling bracket standard
A / B / C
Basic High M
Basic High M
Dimensions
205 mm
580 mm
345 mm
24 mm
58.5 / 125 / 170 mm
156 mm
184 mm
284 mm
2500 mm
970 mm
340 mm
8"
22.8"
13.5"
1.5"
2.2 / 4.9 / 6.6"
6.29"
7.2"
11.38"
98"
38"
13.3"
H
A
E D
F
C
G I
B
K J
SPRING PLATE
AVAILABLE FOR A & B RAIL
FLUSH CEILING PROFILE LIP IS 7/16" X 3/8"
CUT 12 - 15 MM OF THE 2 LONGITUDE
SCREW FOR
PROFILE END
Guldmann Inc.
14401 McCormick Drive Tampa FL, 33626 Unit A
Toll Free: 1(800)-664-8834 Fax: 1(813)-880-9558
Email: Info@Guldmann.net
NOTES
APPROVED BY:
PROJECT NAME:
SHEET DESCRIPTION:
SHEET NO:
DATE: DRAWN BY:
Boise PCMH
GEN. INFO
GUL102
TRC04/19/2021
PROJECT ID:102460
NTS
BRACING + SUPPORT
NTS
GH3 TRANSFORMER
HOIST = 550 LBS1.
FLUSH RECESSED / TYPE G2.
CEILING HEIGHT = 9' - 0"3.
AFF TO B.O.S = 14' - 6"
ATTACHMENT TYPE = WEB TRUSS4.
NTS
RAIL SPAN CHART
Revision Schedule
Date Issued by Number
SPAN = 5' - 4 1/4"
NTS
GH3 HOIST DIMS
NTS
GH3 PATIENT SLING DIMS
NTS
FLUSH CEILING PROFILE
DRAWINGS APPROVED FOR THE GULDMANN
GH-3 LIFTER SYSTEM STANDARD INSTALLATION
OF SUPPORT AND CONNECTION TO STEEL WEB
TRUSS.
SYSTEMS INSTALLED FOR 1100 POUNDS WILL BE
TESTED AT 1650 POUNDS.
SYSTEMS INSTALLED FOR 770 POUNDS WILL BE
TESTED AT 1155 POUNDS.
SYSTEMS INSTALLED FOR 550 POUNDS WILL BE
TESTED AT 825 POUNDS.
STEEL WEB TRUSS
TAG ITEM NUMBER DESCRIPTION
10 113024 3/8" NYLOCK NUT
6,7,12
A
GULDMANN RAIL
EXISTING STEEL TRUSS
M A
X
V A
R
IE
S
40° TO 90°
9 P2787 8-1/4" BEAM CLAMP
8 11101273 3/8" X 1-1/4" X 1/8" WASHER (THICK)
7 411764 3/8" X 1-1/4" BOLT
6 377882 MQM-F WINGNUT 3/8"
5 550435 CEILING BRACKET KIT - GULDMANN
4 550481-US 1" Ø-6013-T6 AL TUBE - GULDMANN
3 550439 STABILIZER BRACKET KIT - GULDMANN
2 HTHR037 12B7 3/8" B7 THREADED ROD
1 3584301 2-7/16" X 10' B2B STRUT SLOTTED
0º
90º
HVAC / MECH
BASEPLATE x4
STABILIZERS x3
STABILIZERS
± 22°
BRIDGE SECTION
BRIDGE / RAIL TOP
8,10
11 411752 3/8" HEX NUT
SPAN
12 370632 MQZ-L BASE PLATE 3/8"
OPTIONAL A
DETAIL A
OPTIONAL
ATTACHMENT
12 10
TRUSS
13 377885 MQA-F SADDLE NUT 3/8" (OPTIONAL A)
Guldmann Inc.
14401 McCormick Drive Tampa FL, 33626 Unit A
Toll Free: 1(800)-664-8834 Fax: 1(813)-880-9558
Email: Info@Guldmann.net
NOTES
APPROVED BY:
PROJECT NAME:
SHEET DESCRIPTION:
SHEET NO:
DATE: DRAWN BY:
SEE 3/GUL102 FOR RAIL SPANS & ADDITIONAL DIMENSIONS
Boise PCMH
ATTACHMENT
GUL103
TRC04/19/2021
PROJECT ID:102460
HOIST = 550 LBS1.
FLUSH RECESSED / TYPE G2.
CEILING HEIGHT = 9' - 0"3.
AFF TO B.O.S = 14' - 6"
ATTACHMENT TYPE = WEB TRUSS4.
Revision Schedule
Date Issued by Number
SPAN = 5' - 4 1/4"
GUL104
/8
H
O
IS
T
/2
T R
A V
E R
S E
C E
IL
IN
G
8" MIN.1
/8
/2
/8
27" MAX.
Guldmann Inc.
14401 McCormick Drive Tampa FL, 33626 Unit A
Toll Free: 1(800)-664-8834 Fax: 1(813)-880-9558
Email: Info@Guldmann.net
NOTES
APPROVED BY:
PROJECT NAME:
SHEET DESCRIPTION:
SHEET NO:
DATE: DRAWN BY:
SEE 3/GUL102 FOR RAIL SPANS & ADDITIONAL DIMENSIONS
Boise PCMH
PROJECT DETAILS
GUL104
TRC04/19/2021
PROJECT ID:102460
HOIST = 550 LBS1.
FLUSH RECESSED / TYPE G2.
CEILING HEIGHT = 9' - 0"3.
AFF TO B.O.S = 14' - 6"
ATTACHMENT TYPE = WEB TRUSS4.
Revision Schedule
Date Issued by Number
06/14/2021 CML B
06/18/2021 CML C
SPAN = 5' - 4 1/4"
1" = 1'-0"
SYSTEM ELEVATION
1 1/2" = 1'-0"
SYSTEM / CEILING
/8
6' - 2 1/2" 3' - 1 5/8"
9' - 10 1/8"
Guldmann Inc.
14401 McCormick Drive Tampa FL, 33626 Unit A
Toll Free: 1(800)-664-8834 Fax: 1(813)-880-9558
Email: Info@Guldmann.net
NOTES
APPROVED BY:
PROJECT NAME:
SHEET DESCRIPTION:
SHEET NO:
DATE: DRAWN BY:
SEE 3/GUL102 FOR RAIL SPANS & ADDITIONAL DIMENSIONS
ACCURATE TO 4" BASED
ON PDF CONVERSION
Boise PCMH
PROCEDURE RM 01 - 136
GUL105
TRC04/19/2021
PROJECT ID:102460
B rail
Ceiling bracket, Extended
Hoist
Stabilizers
KEY
Hoist Transformer
HOIST = 550 LBS1.
FLUSH RECESSED / TYPE G2.
CEILING HEIGHT = 9' - 0"3.
AFF TO B.O.S = 14' - 6"
ATTACHMENT TYPE = WEB TRUSS4.
SPAN = 5' - 4 1/4"
3/4" = 1'-0"
PROCEDURE RM 01 - 136
Revision Schedule
Date Issued by Number
06/14/2021 CML B
06/18/2021 CML C
10' - 4"
/4
/2
8' - 2 1/8" 2' - 1 1/2"
GUL104
Guldmann Inc.
14401 McCormick Drive Tampa FL, 33626 Unit A
Toll Free: 1(800)-664-8834 Fax: 1(813)-880-9558
Email: Info@Guldmann.net
NOTES
APPROVED BY:
PROJECT NAME:
SHEET DESCRIPTION:
SHEET NO:
DATE: DRAWN BY:
SEE 3/GUL102 FOR RAIL SPANS & ADDITIONAL DIMENSIONS
ACCURATE TO 4" BASED
ON PDF CONVERSION
Boise PCMH
WOMENS HEALTH 01 - 118
GUL106
TRC04/19/2021
PROJECT ID:102460
B rail
Ceiling bracket, Extended
Hoist
Stabilizers
KEY
Hoist Transformer
HOIST = 550 LBS1.
FLUSH RECESSED / TYPE G2.
CEILING HEIGHT = 9' - 0"3.
AFF TO B.O.S = 14' - 6"
ATTACHMENT TYPE = WEB TRUSS4.
SPAN = 5' - 4 1/4"
3/4" = 1'-0"
WOMENS HEALTH 01 - 118
Revision Schedule
Date Issued by Number
A B C D E F G H I J K L M
Guldmann Inc.
14401 McCormick Drive Tampa FL, 33626 Unit A
Toll Free: 1(800)-664-8834 Fax: 1(813)-880-9558
Email: Info@Guldmann.net
NOTES
APPROVED BY:
PROJECT NAME:
SHEET DESCRIPTION:
SHEET NO:
DATE: DRAWN BY:
SEE 3/GUL102 FOR RAIL SPANS & ADDITIONAL DIMENSIONS
ACCURATE TO 4" BASED
ON PDF CONVERSION
Boise PCMH
OVERALL
GUL107
TRC04/19/2021
PROJECT ID:102460
3/32" = 1'-0"
OVERALL
HOIST = 550 LBS1.
FLUSH RECESSED / TYPE G2.
CEILING HEIGHT = 9' - 0"3.
AFF TO B.O.S = 14' - 6"
ATTACHMENT TYPE = WEB TRUSS4.
SPAN = 5' - 4 1/4"
Revision Schedule
Date Issued by Number
05/28/2021 TRC A
Guldmann Inc.
14401 McCormick Drive Tampa FL, 33626 Unit A
Toll Free: 1(800)-664-8834 Fax: 1(813)-880-9558
Email: Info@Guldmann.net
NOTES
APPROVED BY:
PROJECT NAME:
SHEET DESCRIPTION:
SHEET NO:
SIGNOFF / APPROVAL NOTES
1. The facility structure capacity must be able to support the load specified on the installation drawing.
2. Electrical power source provided by the facility, position to be verified in the field.
3. Guldmann Inc. reserves the right to slightly modify the track layout if necessary without Facility approval.
This includes lowering the system to avoid ceiling objects.
4. Guldmann Inc. will not be held responsible for any damages, inconvenience, resulting from hitting any non visible conduits during our installation process of which we have not been advised.
5. Modifications of door headers by the facility.
6. Signed drawings are required prior to ordering and installation of ceiling lifts.
This signature of each drawing confirms agreement of all the conditions listed above.
7. Systems are tested at 1.5 times capacity.
550 lbs. tested at 825 lbs.; 770 lbs. tested at 1155 lbs.; 825 lbs. tested at 1238 lbs.; 1100 lbs. tested at 1650 lbs.
MY SIGNATURE BELOW VERIFIES THAT I HAVE:
Reviewed all _____ pages, with a total quantity of _____ room(s), _____ system(s) for the approbation of layout for each room and the approbation of structure for each room.
CUSTOMER
APPROVAL: _____________________________________ DATE: _______________
DATE: DRAWN BY:
ACCURATE TO 4" BASED
ON PDF CONVERSION
Boise VA Boise PCMH
500 W Fort St.
Boise, ID 83702
PROJECT ID:102460.1
Boise PCMH
FACILITY SIGNOFF
GUL101
TRC04/19/2021
7 2 2
HOIST = 550 LBS1.
FLUSH RECESSED / TYPE G2.
CEILING HEIGHT = 9' - 0"3.
AFF TO B.O.S = 14' - 6"
ATTACHMENT TYPE = WEB TRUSS4.
SPAN = 5' - 4 1/4"
Revision Schedule
Date Issued by Letter Description
05/28/2021 TRC A RM. 118 RE-CONFIGURED / RMS 163 & 145 REMOVED AND SET TO ALT
06/14/2021 CML B NEW BACKGROUNDS / RM 136 REPLACED WITH RM 163
06/18/2021 CML C ROOMS 136 & 145 READDED / PROJECT SPLIT IN TWO
EACH FIXED RAIL TO
HAVE x1 PARALLEL & x2 PERPENDICULAR
STABILIZERS
ALL JOINTS MUST HAVE x1 PARALLEL & x1
PERPENDICULAR STABILIZER
SEE SPAN CHART
S E
E S
P A
N C
H A
R T
SEE SPAN CHART
± 22°
STABILIZERS CAN BE ROTATED 180°
Fixed rail systems Fixed rail systems must be connected by Guldmann Service Team or by a qualified installer.
A Guldmann class I transformer must be used.
Installation to follow existing rules for authorized electricians and according to IEC/UL 60601-1.
Terminals on transformers Class I Double-insulated UL approved cable must be used (Cable specification: UL category: AVLV2, min.300V, min.
80° C, min. 18AWG (1 mm2 )) Transformer DK-13991 In:
Terminal 1 = N Terminal 2 = 100-115V AC Terminal 3 = 230V AC Terminal 4 = Protective ground Out:
Terminal 5 = 33V AC Terminal 6 = 33V AC
POWER TO BE WITHIN 15 FT OF END OF FIXED RAIL
Guldmann permanent mounted ceiling aluminium rail
Hoist with trolleys movable in rail
Protective Plast Enclosure
UL approved 400V cable, green, min. 14 AWG
Protective Ground
M1
UL approved 400V double isolated cable, min. 18 AWG
ON/OFF
UL approved 400V double isolated cable, min. 18 AWG
T ra n s fo rm e r
D K
-1
Main power Permanent installation Specification of breaker:
> 1 Amp capacity Line must always be broken
L1
AC/DC charger
Main Motor Control
Emergency Stop
Safety Circuit Emergency down
Motor Function Selector Switch0 1
V A
C
U S
V
A C
P ro te c ti ve
G ro u n d
V
A C
V
A C1 A
T
.5
A T
Green Light ON (or ON / OFF - sign)
MAX
PROJECTION
MAX DISTANCE BETWEEN VERTICAL SUPPORTS
A
B
C
A
B
C
A
B
C
A
B
C
A
B
C
A
B
C
X
X
X
RAIL
RAIL
RAIL
RAIL
RAIL
RAIL
INCH
METER
,23
,7
1,2
39 49 59 68 78 118 137 147 157 177 187 206 215 236 246 264
1,0 1,25 1,5 1,75 2,0 3,0 3,5 3,75 4,0 4,5 4,75 5,25 5,5 6,0 6,25 6,75
MAX. 175 KG (385 LBS.)
MAX. 205 KG (450 LBS.)
MAX. 255 KG (560 LBS.)
MAX. 300 KG (660 LBS.)
MAX. 350 KG (770 LBS.)
MAX. 375 KG (825 LBS.)
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
A-RAIL
WEIGHT: 2.6 KG/M
(1.75 LBS/FT)
B-RAIL
WEIGHT: 5.0 KG/M
(3.36 LBS/FT)
C-RAIL
WEIGHT: 6.0 KG/M
(4.03 LBS/FT)
(2 1/2")
62 mm.
(2
/4
,5 m m
(2 1/2")
62 mm.
(5 m m
(2 1/2")
62 mm.
(6
/3 m m
A
B
C
RAIL MAX. 400 KG (880 LBS.)
X
X
X
A
B
C
RAIL MAX. 500 KG (1100 LBS.)
X
X
X
7,25
8,0
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X m m
/2 m m
/4
,5 m m
M
IN
m m
/4
DEPTH = 205 mm (8 1/4")
345 mm (13 3/4")
A
B
C
D
E
F
G
H
I
J
K
Ceiling bracket standard
A / B / C
Basic High M
Basic High M
Dimensions
205 mm
580 mm
345 mm
24 mm
58.5 / 125 / 170 mm
156 mm
184 mm
284 mm
2500 mm
970 mm
340 mm
8"
22.8"
13.5"
1.5"
2.2 / 4.9 / 6.6"
6.29"
7.2"
11.38"
98"
38"
13.3"
H
A
E D
F
C
G I
B
K J
SPRING PLATE
AVAILABLE FOR A & B RAIL
FLUSH CEILING PROFILE LIP IS 7/16" X 3/8"
CUT 12 - 15 MM OF THE 2 LONGITUDE
SCREW FOR
PROFILE END
Guldmann Inc.
14401 McCormick Drive Tampa FL, 33626 Unit A
Toll Free: 1(800)-664-8834 Fax: 1(813)-880-9558
Email: Info@Guldmann.net
NOTES
APPROVED BY:
PROJECT NAME:
SHEET DESCRIPTION:
SHEET NO:
DATE: DRAWN BY:
Boise PCMH
GEN. INFO
GUL102
TRC04/19/2021
PROJECT ID:102460.1
NTS
BRACING + SUPPORT
NTS
GH3 TRANSFORMER
HOIST = 550 LBS1.
FLUSH RECESSED / TYPE G2.
CEILING HEIGHT = 9' - 0"3.
AFF TO B.O.S = 14' - 6"
ATTACHMENT TYPE = WEB TRUSS4.
NTS
RAIL SPAN CHART
Revision Schedule
Date Issued by Number
06/18/2021 CML C
SPAN = 5' - 4 1/4"
NTS
GH3 HOIST DIMS
NTS
GH3 PATIENT SLING DIMS
NTS
FLUSH CEILING PROFILE
DRAWINGS APPROVED FOR THE GULDMANN
GH-3 LIFTER SYSTEM STANDARD INSTALLATION
OF SUPPORT AND CONNECTION TO STEEL WEB
TRUSS.
SYSTEMS INSTALLED FOR 1100 POUNDS WILL BE
TESTED AT 1650 POUNDS.
SYSTEMS INSTALLED FOR 770 POUNDS WILL BE
TESTED AT 1155 POUNDS.
SYSTEMS INSTALLED FOR 550 POUNDS WILL BE
TESTED AT 825 POUNDS.
STEEL WEB TRUSS
TAG ITEM NUMBER DESCRIPTION
10 113024 3/8" NYLOCK NUT
6,7,12
A
GULDMANN RAIL
EXISTING STEEL TRUSS
M A
X
V A
R
IE
S
40° TO 90°
9 P2787 8-1/4" BEAM CLAMP
8 11101273 3/8" X 1-1/4" X 1/8" WASHER (THICK)
7 411764 3/8" X 1-1/4" BOLT
6 377882 MQM-F WINGNUT 3/8"
5 550435 CEILING BRACKET KIT - GULDMANN
4 550481-US 1" Ø-6013-T6 AL TUBE - GULDMANN
3 550439 STABILIZER BRACKET KIT - GULDMANN
2 HTHR037 12B7 3/8" B7 THREADED ROD
1 3584301 2-7/16" X 10' B2B STRUT SLOTTED
0º
90º
HVAC / MECH
BASEPLATE x4
STABILIZERS x3
STABILIZERS
± 22°
BRIDGE SECTION
BRIDGE / RAIL TOP
8,10
11 411752 3/8" HEX NUT
SPAN
12 370632 MQZ-L BASE PLATE 3/8"
OPTIONAL A
DETAIL A
OPTIONAL
ATTACHMENT
12 10
TRUSS
13 377885 MQA-F SADDLE NUT 3/8" (OPTIONAL A)
Guldmann Inc.
14401 McCormick Drive Tampa FL, 33626 Unit A
Toll Free: 1(800)-664-8834 Fax: 1(813)-880-9558
Email: Info@Guldmann.net
NOTES
APPROVED BY:
PROJECT NAME:
SHEET DESCRIPTION:
SHEET NO:
DATE: DRAWN BY:
SEE 3/GUL102 FOR RAIL SPANS & ADDITIONAL DIMENSIONS
Boise PCMH
ATTACHMENT
GUL103
TRC04/19/2021
PROJECT ID:102460.1
HOIST = 550 LBS1.
FLUSH RECESSED / TYPE G2.
CEILING HEIGHT = 9' - 0"3.
AFF TO B.O.S = 14' - 6"
ATTACHMENT TYPE = WEB TRUSS4.
Revision Schedule
Date Issued by Number
SPAN = 5' - 4 1/4"
C E
IL
IN
G
/2
T R
A V
E R
S E
/8
H
O
IS
T 1
/8
/2
/8
8" MIN.
27" MAX.
Guldmann Inc.
14401 McCormick Drive Tampa FL, 33626 Unit A
Toll Free: 1(800)-664-8834 Fax: 1(813)-880-9558
Email: Info@Guldmann.net
NOTES
APPROVED BY:
PROJECT NAME:
SHEET DESCRIPTION:
SHEET NO:
DATE: DRAWN BY:
SEE 3/GUL102 FOR RAIL SPANS & ADDITIONAL DIMENSIONS
Boise PCMH
PROJECT DETAILS
GUL104
TRC04/19/2021
PROJECT ID:102460.1
HOIST = 550 LBS1.
FLUSH RECESSED / TYPE G2.
CEILING HEIGHT = 9' - 0"3.
AFF TO B.O.S = 14' - 6"
ATTACHMENT TYPE = WEB TRUSS4.
Revision Schedule
Date Issued by Number
06/14/2021 CML B
06/18/2021 CML C
SPAN = 5' - 4 1/4"
1" = 1'-0"
SYSTEM ELEVATION
1 1/2" = 1'-0"
SYSTEM / CEILING
/8
9' - 10 1/8"
3' - 1 7/8" 6' - 2 1/2"
Guldmann Inc.
14401 McCormick Drive Tampa FL, 33626 Unit A
Toll Free: 1(800)-664-8834 Fax: 1(813)-880-9558
Email: Info@Guldmann.net
NOTES
APPROVED BY:
PROJECT NAME:
SHEET DESCRIPTION:
SHEET NO:
DATE: DRAWN BY:
SEE 3/GUL102 FOR RAIL SPANS & ADDITIONAL DIMENSIONS
ACCURATE TO 4" BASED
ON PDF CONVERSION
Boise PCMH
PROCEDURE RM 01 - 145
GUL105
TRC04/19/2021
PROJECT ID:102460.1
B rail
Ceiling bracket, Extended
Hoist
Stabilizers
KEY
Hoist Transformer
HOIST = 550 LBS1.
FLUSH RECESSED / TYPE G2.
CEILING HEIGHT = 9' - 0"3.
AFF TO B.O.S = 14' - 6"
ATTACHMENT TYPE = WEB TRUSS4.
SPAN = 5' - 4 1/4"
3/4" = 1'-0"
PROCEDURE RM 01 - 145
Revision Schedule
Date Issued by Number
05/28/2021 TRC A
1' - 10 3/8" 8' - 2 1/2"
10' - 4 1/8"
Guldmann Inc.
14401 McCormick Drive Tampa FL, 33626 Unit A
Toll Free: 1(800)-664-8834 Fax: 1(813)-880-9558
Email: Info@Guldmann.net
NOTES
APPROVED BY:
PROJECT NAME:
SHEET DESCRIPTION:
SHEET NO:
DATE: DRAWN BY:
SEE 3/GUL102 FOR RAIL SPANS & ADDITIONAL DIMENSIONS
ACCURATE TO 4" BASED
ON PDF CONVERSION
Boise PCMH
WOMENS HEALTH 02 - 163
GUL106
TRC04/19/2021
PROJECT ID:102460.1
3/4" = 1'-0"
WOMENS HEALTH 02 - 163
B rail
Ceiling bracket, Extended
Hoist
Stabilizers
KEY
Hoist Transformer
HOIST = 550 LBS1.
FLUSH RECESSED / TYPE G2.
CEILING HEIGHT = 9' - 0"3.
AFF TO B.O.S = 14' - 6"
ATTACHMENT TYPE = WEB TRUSS4.
SPAN = 5' - 4 1/4"
Revision Schedule
Date Issued by Number
A B C D E F G H I J K L M
Guldmann Inc.
14401 McCormick Drive Tampa FL, 33626 Unit A
Toll Free: 1(800)-664-8834 Fax: 1(813)-880-9558
Email: Info@Guldmann.net
NOTES
APPROVED BY:
PROJECT NAME:
SHEET DESCRIPTION:
SHEET NO:
DATE: DRAWN BY:
SEE 3/GUL102 FOR RAIL SPANS & ADDITIONAL DIMENSIONS
ACCURATE TO 4" BASED
ON PDF CONVERSION
Boise PCMH
OVERALL
GUL107
TRC04/19/2021
PROJECT ID:102460.1
3/32" = 1'-0"
OVERALL
HOIST = 550 LBS1.
FLUSH RECESSED / TYPE G2.
CEILING HEIGHT = 9' - 0"3.
AFF TO B.O.S = 14' - 6"
ATTACHMENT TYPE = WEB TRUSS4.
SPAN = 5' - 4 1/4"
Revision Schedule
Date Issued by Number
05/28/2021 TRC A
File details come from the government source that posted it. Updated .