This document contains a technical exhibit outlining measurement capabilities for fiber optics and infrared testing, along with details of a related federal contract opportunity for the Air Force Primary Standards Laboratory (AFPSL) Recompete. The technical exhibit specifies the parameter ranges and uncertainties for fiber optics measurements of power, spectrum analysis, and source testing. It also outlines radiometric temperature and spectral radiance measurement capabilities for infrared testing. The federal contract opportunity is a solicitation from the Department of the Air Force Materiel Command Lifecycle Management Center seeking proposals for operation and sustainment of the AFPSL. A bidder's library with relevant documents will be made available after January 2020, with a final RFP anticipated to be posted on January 31, 2020. Proposals will be evaluated to select an awardee to continue supporting AFPSL testing requirements.
View the file
Other files for this federal contract opportunity
Show all 50
AFPSL Recompete has more files on GovTribe.
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Technical Exhibit 5e 29 August 2018
AFPSL MEASUREMENT CAPABILITIES
Air Force Critical Information
FIBER OPTICS MEASUREMENT AREA
Parameter Range Uncertainty Reference Std.
| Power Sensitivity | Non-Connectorized 0 to -60 dBm: | ||
| 800 - 900 nm | ±0.25 dB | 1% | |
| 1250 - 1350 nm | ±0.25 dB | 1% | |
| 1500 - 1600 nm | ±0.25 dB | 1% | |
| Connectorized +3 to -60 dBm | |||
| 800 - 900 nm | ±2% | 1% | |
| 1250 - 1350 nm | ±2% | 1% | |
| 1500 - 1600 nm | ±2% | 1% | |
| Spectrum Analyzers | 365 nm to 1530 nm | ±0.5 nm | 0.05 nm |
(Wavelength) Sources (LEDs & LDs) Spectral Properties ±2 nm 0.5 nm
INFRARED MEASUREMENT AREA
| Radiometric Temperature | 200 - 1200 C | ±1.40 C | ±0.22 C |
| with Emissivity | 0.005 | 0.00002 |
(Blackbodies)
| Spectral Radiance | <300 nm | ±4% | ±1.7% |
| (Lamps) | 300-2400 nm | ±3% | ±1.1% |
| Spectral Irradiance | <350 nm | ±5% | ±2.2% |
| (Lamps) | 350 – 1600 nm | ±4% | ±1.9% |
| >1600 nm | ±8% | ±6.5% | |
| Noncontact Thermometry | 20 - 800 C | ±5% of F.S. | |
| 850 - 2300 C | 6 C | 2.5 C | |
| 2300 - 2500 C | 8 C | 2.5 C | |
| Spectral Transmission | Wavelength (185 - 3200 nm) | ±0.6 nm | |
| Transmittance | |||
| (10% - 100%) | ±2.5% of Value | ||
| (1.0% - 10%) | ±5% of Value | ||
| (0.1 % - 1.0%) | ±8% of Value | ||
| (less than 0.1 %) | ± 15% of Value | ||
| Spectral Transmission | Wavelength (2.5 - 25 m) | ±0.6 m | |
| Transmittance | |||
| (10% - 100%) | ±2.5% of Value | ||
| (1.0% - 10%) | ±5% of Value | ||
| (0.1% - 1.0%) | ±8% of Value | ||
| (less than 0.1%) | ±15% of Value |
LASER MEASUREMENT AREA
Parameter Range Uncertainty Reference Std.
| Peak Power (Irradiance) | 1.06 m, 1.0 - 100 mW/cm2 | ±2.5 dB | ±7.0% |
| 1 W/cm2 - 1.1 W/cm2 | ±15.0% | ±7.0% | |
| Fluence (Energy/Pulse) | 1.06 m, 10 - 7000 fJ/cm2 | ±10.0% | |
| Average Power (Q-Switched) | 1.06 m, 1.0 mW - 3 W | ±5.0% | ±1.0% |
| CW Power | 488 nm, 1.0 mW - 2.0 W | ±5.0% | ±1.0% |
| 514.5 nm, 1.0 mW - 2.0 W | ±5.0% | ±1.0% | |
| 632.8 nm, 20 nW - 60 mW | ±5.0% | ±1.0% | |
| 1.06 m, 1.0 mW - 2.0 W | ±5.0% | ±1.0% | |
| 10.6 m, 200 mW - 150 W | ±5.0% | ±2.0% | |
| 10.6 m, 50 W - 1500 W | ±5.0% | ±3.0% | |
| CW Power (On-Site) | 1 - 11 m, 50 W - 200 kW | ±5.0% | ±3.0% |
PHOTOMETRICS MEASUREMENT AREA
| Luminous Intensity | 25 - 2000 Candela | ±2.0% | 0.5% |
| Color Temperature | 2000 - 3200 K | ±20 K | From Spectral Irradiance |
| Luminous Directional Intensity | 0.43 - 0.7 fL/fC | ±1.5% | 0.7% |
| UV Irradiance | 365 nm, 0 -3 mW/ cm2 | ±7%% | 5% |
RADIAC MEASUREMENT AREA
| Alpha Standards | 0 - 13,000,000 CPM | ±5.0% | ±1.0% |
| Gamma Standards | 1 mR/hr - 10,000 R/hr | ±5.0% | ±1.0% |
| Alpha Meters | 0 - 13,000,000 CPM | ±20.0% | ±1.0% |
| Gamma Meters | 1 mR/hr - 500 R/hr | ±10.0% | ±1.0% |
| Neutron Meters | 1 - 15,000 mRem/hr | ±15.0% | ±5.0% |
DC/LOW FREQUENCY MEASUREMENT AREA
Parameter Range Uncertainty Reference Std.
| Range/Uncertainty | |||
| Capacitance | 0.01pf - 1000 pf @ 1000 Hz | ±0.001% to ±0.002% | 1000 pf @ 1000 Hz/± 0.0004% |
| 1 pf @ 1000 Hz (Coaxial) | ±0.1% | 1 pf @ 1000 Hz/± 0.052% | |
| Resistance | 0.001 to 1T | ±0.0002% to ±0.2% | 10 k/± 0.000015% |
| 10 k | ±0.0001%(w/historical data) | 10 k/± 0.000015% | |
| 1 | ±0.0006%(w/historical data) | 1 /± 0.00005% | |
| DC Current | 0.00001 A to 300 A | ±0.004 to ±0.02% | 0.00001 A to100 A/±0.0004% to |
| ±0.002% | |||
| AC Current | 10 mA to 30 mA (100 kHz) | ±0.015% | 10 mA to 30 mA/±0.01% |
| 10 mA to 20 A (1 kHz, 1 0 kHz) | ±0.01% to ±0.03% | 10 mA to 20 A/±0.005% to ±0.015% | |
| Inductance | 100 mH @ 1 kHz | ±0.03% | 100 mH @ 1 kHz/±0.02% |
| Thermal Voltage Converters | 0.45V to 100 V (1 MHz to 100 MHz) | ±0.06% to ±1.1% | Same/±0.02% to ±0.5% |
| Thermal Transfer | 0.0022V to 1000 V (10 Hz to 30 MHz) | ±0.0022% to ±0.62% | Same/±0.001% to ±0.5% |
Standard Multi-range
| AC Flatness | Input Resistance: 50 | ±1.0% | Measured w/DMM |
| (Peak-to-Peak Detector) | Sensitivity: 1% P-P change to | ±1 mV | Measured w/DMM |
| produce a 10 mV DC change | |||
| in output | |||
| Frequency Response: | ±0.2% to ±1.6% | 50 Hz to 500 MHz/±0.08% to ±1.2% | |
| (50 kHz reference) | |||
| 0.1 MHz to 500 MHz | |||
| Helmholtz Coils | 0 to 1000 Gauss | ±0.1 % to ±2.5% | Absolute Standard |
(magnetic field strength)
| Reference Magnets | 0 to 20 kGauss | ±0.6% to ±2.5% | Absolute Standard | |
| DC Voltage (low) | Standard Cells: 1.018 V | ±0.0001%(w/historical data) | 1.018 V/±0.00005% | |
| Zener Reference: 10 V | ±0.0001%(w/historical data) | 10 V/±0.00005% | ||
| DC Voltage (high) | 5 kV to 100 kV | ±0.025% | 5 kV to 75 kV/±0.01% | |
| AC Voltage (high) | 15 kV to 100 kV | ±1.0% | 5 kV to 75 kV/±0.03% | |
| DC Voltage Ratio | 0 to 1.0 | ±0.05 ppm (terminal linearity) | N/A | |
| 10: 1 ratio | ±0.2 ppm | N/A | ||
| 100: 1 ratio | ±0.5 ppm | N/A | ||
| AC Voltage Ratio | Transfer Ratio Corrections | |||
| 0 to 1.0 (50 Hz to 10 kHz) | ||||
| (50 Hz to 1 kHz) | ±0.0001% | (50 Hz to 1 kHz)/±0.00005% | ||
| (1 kHz to 10 kHz) | ±0.0001% x (Freq(kHz)2 ) | (1 kHz to 10 kHz)/±0.00005% x (Freq(kHz)2) | ||
| Risetime | 25 Picoseconds | ±7 Picoseconds | 25 Picoseconds/±5 Picoseconds | |
| Precise Time and | Time: Universal Time Coordinate | ±10 nanoseconds of UTC | USNO | |
| Frequency | steered by USNO | |||
| Frequency: 100 kHz to 10 MHz | ±1.1 X 10-13 fractional | USNO | ||
| frequency offset | ||||
| Phase Angle | 1 Hz to 100 kHz | ±(5+0.1 R) millidegrees to ±(48+R) | N/A | |
| millidegrees where R is the ratio | ||||
| of the larger to the smaller output | ||||
| voltage (i.e. reference voltage, | ||||
| variable phase voltage) |
RF/MICROWAVE MEASUREMENT AREA
| Parameter | Range | Uncertainty | Reference | |||
| Coaxial Power | 0.1 to 10 MHz | 1.0 to 1.25% | 0.5% | |||
| 0.01 to 18 GHz | 1.5 to 2.6% | 0.8 to 1.7% | ||||
| 18 to 26.5 GHz | 4.4% | 2.4% | ||||
| 26.5 to 40 GHz | 4.0% | 2.1% | ||||
| Waveguide Power | 8.2 to 40 GHz | 1.75 to 3.6% | 1.5 to 2.4% | |||
| 33 to 50 GHz | 2.6 to 4.5% | 1.65 to 3.54% | ||||
| High Power | 1 MHz to 1.3 GHz | 3.0 to 3.5% | 2.4 to 2.9% | |||
| 1 to 1000 Watts | ||||||
| Attenuation | 1.25 MHz (Var) | 0.01 dB of charted value | 0.002/6 dB*1 | |||
| 30 MHz | 0.01 dB of charted value | 0.005/10 dB*1 | ||||
| 2.5 MHz to 26.5 GHz (0 to 100 dB) | 0.015+0.005/10 dB step+1digit*1 | 0.003+0.003/10 dB*1 | ||||
| 2.5 MHz to 26.5 GHz (100 to 120 dB) | 0.015+0.05/10 dB step+1digit* | 0.003+0.003/10 dB*1 | ||||
| 2.5 MHz to 26.5 GHz (120 to 127 dB) | 0.05+0.02/10 dB step+1digit | 0.003+0.003/10 dB*1 | ||||
| +0.18 dB noise*1 | ||||||
| 26.5 to 40 GHz (0 to 50 dB) | 0.045*2 | 0.01 dB of charted value | ||||
| Coaxial Reflection | BNC: 100 kHz to 100 MHz | 0.01 to 0.03*2 | <0.00025 @ 30 MHz | |||
| Coefficient | 14 mm: 100 kHz to 8.5 GHz | 0.007 to 0.04*2 | 0.0016 to 0.0020*2 | |||
| N: 100 kHz to 18 GHz | 0.007 to 0.04*2 | 0.0011 to 0.0020*2 | ||||
| 7 mm: 100 kHz to 18 GHz | 0.005 to 0.025*2 | 0.0016 to 0.0018*2 | ||||
| 3.5 mm: 100 kHz to 26.5 GHz | 0.01 to 0.07*2 | 0.0015 to 0.0030*2 | ||||
| 2.92 mm: 100 kHz to 40 GHz (SWR<2) | 0.025 to 0.15*2 | 0.0009 to 0.0012*1 | ||||
| 2.4 mm: 100 kHz to 40 GHz (SWR<2) | 0.015 to 0.09*2 | 0.0009 to 0.0012*1 | ||||
| Waveguide Reflection | 2.6 to 40 GHz (SWR <2) | 0.015 to 0.06*2 | <1.01413*2 | |||
| Coefficient | 40 GHz to 50 GHz (SWR <2) | <1.1 | <1.01413*2 | |||
| VHF Omnidirectional | 0 to 359.99 degrees | 0.09 degrees | 0.01 @ 0 | |||
| Ranging | 0.02 @ all other bearing | |||||
| angles | ||||||
| AM (Instrument Landing | 90, 150 Hz | (0.1 to 0.2)% | 0.05% | |||
| System) | (Based on a modulation | |||||
| depth of 0 to 50%) | ||||||
| AM Modulation | 20 Hz to 100 kHz | 1 to 3%*2 | 0.3% of reading + 1 digit*2 | |||
| (1 50 kHz to 1300 MHz | ||||||
| carrier frequency) | ||||||
| FM Modulation | 20 Hz to 200 kHz | 1 to 5%*2 | 0.75% rdg*2 | |||
| (250 kHz to 1300 MHz | ||||||
| carrier frequency) | ||||||
| Phase Modulation | 200 Hz to 20 kHz | 3*3 | 0.1% (Total Harmonic | |||
| (1 50 kHz to 1300 MHz | Distortion) | |||||
| carrier frequency) | ||||||
| Power Density | 2450 MHz | 2 dB | 0.5 dB | |||
| Coaxial Thermal Noise | 0.01 to 8 GHz | 0.23 dB*4 | 0.23 dB*4 | |||
| 8 to 26 GHz | 0.32 dB*4 | 0.32 dB*4 | ||||
| 26.5 GHz | 0.35 dB | 0.35 dB | ||||
| Waveguide Thermal Noise | 8.2 GHz to 12.4 GHz | 0.2 dB | 0.05 dB*4 | |||
| 12.4 to 18 GHz | 0.25 dB | 0.09 dB*4 | ||||
| Phase Noise | 5 MHz to 1.6 GHz | 2 dB (0.01Hz to 1MHz offset) | 0.3 to 2.0 dB (Spec An linearity) | |||
| 1.6 GHz to 18 GHz | 4 dB (1MHz to 40MHz offset) | 0.3 to 2.0 dB (Spec An linearity) | ||||
| Phase | 0.045 to 26.5 GHz | 0.2 to 180 degrees | 0.2 to 180 degrees | |||
| (Reflection/Transmission) | (varies with frequency, | (varies with frequency, | ||||
| connector type, and | connector type, and | |||||
| reflection/transmission | reflection/transmission | |||||
| coefficient) | coefficient) |
*1 Plus mismatch error *2 Varies with frequency and reflection coefficient *3 Varies with carrier frequency and modulation rate *4 Varies with frequency
FORCE MEASUREMENT AREA
Parameter Range Uncertainty Reference Std.
Dead Weight Machines 0- 100000 lbf ±0.003% rdg
GAS FLOW MEASUREMENT AREA
| Nozzle Stand | 0 - 440 scfm | ±0.5% rdg | NIST Cal'd Nozzles | ±0.30% rdg |
| Laminar Stds | 0.02 - 1 000 lpm | ±1% rdg | NIST Traceable Element | ±0.25% rdg |
| Volumetric Displacement | 0.02 - 50 lpm | ±1% rdg | Bell Provers | ±0.30% rdg |
| Air Velocity | 30 - 9000 fpm | ±4% rdg | Wind Tunnel | ±2.0% rdg |
MASS MEASUREMENT AREA
| Mass (minimum) | 1 milligram | ±5 micrograms |
| Mass (maximum) | 50 Kilograms | ±9.85 milligrams |
PRESSURE MEASUREMENT AREA
| Pneu. Pressure Piston Gauge | 0.2-700 psi | ±0.006% rdg |
| Hydraulic Pressure Piston Gauge | 50-40000 psi | ±0.01% rdg |
| Hydraulic Pressure Piston Gauge | 6-15000 psi | ±0.01% rdg |
| Oxygen Pressure Piston Gauge | 500-15000 psi | ±0.01 % rdg |
SOUND MEASUREMENT AREA
| Microphones (re 1 V/Pa) | 50 - 10 kHz | ±0.15 dB | NIST Cal'd Microphone | +0.09 dB |
| Sound Dosimeters | 2 Hz - 10 kHz | ±0.30 dB | NIST Cal'd Microphone | +0.09 dB |
| Sound Calibrators | 1 kHz | ±0.50 dB | NIST Cal'd Microphone | +0.09 dB |
| Sound Level Meters | 2 Hz - 10 kHz | ±0.50 dB | NIST Cal'd Microphone | +0.09 dB |
| Piston Phones | 250 Hz | ±0.30 dB | NIST Cal'd Pistonphone | +0.10 dB |
TEMPERATURE MEASUREMENT AREA
| Thermocouples | 0 to 11000 C | ±10 C | NIST Cal'd Thermocouple | ±0.10 C |
| Platinum Resistance Therm. | -190 to 6610 C | ±0.010 C | NIST Cal'd SPRT | <±0.000 C |
| Optical Pyrometer | 900 to 22000 C | ±1% of Scale | NIST Cal'd Pyrometer | ±1.0% Scale |
VACUUM MEASUREMENT AREA
| Low Vacuum(Capacitance Gage) | 10-3 Torr to 1 Torr | ±0.001 to ±0.007 Torr | NIST Cal'd Gage | ±0.3% rdg |
| Low Vacuum | 1 Torr to 100 Torr | ±0.00625 to ±0.65 Torr | NIST Cal'd Gage | ±0.3% rdg |
| High Vacuum (Ion Gauge) | 10-8 Torr to 10-4Torr | ±20% to ±40% rdg | NIST Cal'd Gage | ±6 to 25% rdg |
| High Vacuum(Spin Rotor Gauge) | 10-6 Torr to 10-2 Torr | ±5% rdg | NIST Cal'd Gage | ±1.0% rdg |
HUMIDITY MEASUREMENT AREA
| Dewpoint | -70 to 200 C | ±1.50 C | NIST Cal'd Chilled Mirror | ±0.30 C |
| Relative Humidity | 5 to 95% | ±1.5% rdg | Dual Pressure/Temp System | ±1.0 % |
VIBRATION MEASUREMENT AREA
Accelerometers 5 Hz to 10 kHz ±3.0% of output NIST Cal'd Accelerometer ±1 to 2% of output
DIMENSIONAL MEASUREMENT AREA
| Parameter | Range | Uncertainty | Reference Std. | |
| Gage Blocks | 0 to 0.090 inch | ±3 in | ±2 in | |
| ±6 in | ±2 in | |||
| 0.100 to 4 in | ±2 in | ±2 in | ||
| ±6 in | ±2 in | |||
| 5 to 12 in | ±3 in | ±2 in | ||
| 5 to 20 in | ±1 in/in | ±1 in/in | ||
| Horizontal Length | 0 to 12 in | ±20 in | ±2 in | |
| 13 to 36 in | ±2 in/in | ±1 in/in | ||
| Circularity | 0.0625 to 4 in | ±4 in | ±2 in | |
| 4 to 12 in | ±4 in | ±1 in/in | ||
| Cylindricity | 1 to 4 in | ±30 in | ±2 in | |
| 4 to 11 in | ±30 in | ±1 in/in | ||
| Flatness | 4 to 12 in | ±0.5 in | ±0.5 in | |
| 1 to 12 in | ±1.5 in | ±0.5 in | ||
| above 12 to 16 in | ±4.0 in | ±0.5 in | ||
| Surface Finish | 2 to 100 in | ±2% of Nominal | ±0.51 in | |
| (Depth of SCRATCH) | ||||
| 8 to 250 in | ±2% of Nominal | ±0.59in | ||
| (Ra) | ||||
| Internal/External Diameters | 0.125 to 4.0 in | ±6 in | ±2 in | |
| 4 to 7 in | ±8 in | ±1 in/in | ||
| Internal/External Thread | 0 to 4 in | ±40 in | ±2 in |
Geometry
| Thread Wires | 80 to 4 pitch diameter | ±20 in | ±2 in |
| Steel/Glass Scales | 0 to 4 in | ±50 - 100 in | ±2 in |
| 4 to 18 in | ±50 - 100 in | ±1 in/in |
OPTICS MEASUREMENT AREA
| Angle - Polygon Face | 0 - 360 degrees | ±0.25 arc sec | Absolute Calibration |
| Angle - Angle Blocks | 0 - 45 degrees | ±0.4 arc sec | 0.25 arc sec |
| Angle - Indexing Table | 0 - 360 degrees | ±0.5 arc sec | Absolute Calibration |
| Pentaprism | 90 degrees | ±0.25 arc sec | Absolute Calibration |
Angle - Theodolite (First Order)
| Horizontal Circle | 0 - 360 degrees | ±0.4 arc sec | Absolute Calibration | |
| Collimation | ±2 arc sec | 1 arc sec | ||
| Vertical Leveling | ±1 arc sec | Absolute Calibration | ||
| Angle - Tilt Angle Generator | 0 - 100 arc sec | ±0.2% of nominal | 0.1 arc sec | |
| 100 - 1000 arc sec | ±0.3% of nominal | 0.2 arc sec | ||
| Angle - Autocollimator | 10 arc sec | ±0.01 arc sec* | Modeling technique | |
| 40 arc sec | ±0.03 arc sec* | Modeling technique | ||
| 150 arc sec | ±0.05 arc sec* | Modeling technique | ||
| 300 arc sec | ±0.5 arc sec | 0.2 arc sec | ||
| 1800 arc sec | ±2 arc sec | 0.5 arc sec | ||
| Mirror Flatness | 1 -10 inch diameter | ±1.0 in | 0.5 in | |
| Mirror Flatness (Rectangular) | up to 18 in | ±1.0 in | 0.5 in | |
| Laser Frequency Stability | 546.1 - 632.8 nm | ±5 x 10-10 | Absolute Calibration | |
| (Power 3mw or less) | ||||
| Length - Electronic Distance | 0 to 100 m | ±5 mm | Absolute Calibration | |
| Measuring | above 100 - 1000 m | ±7 mm | Est. based on Comm data |
*Based on AAMACS operation; without AAMACS, 0.25 arc sec