This document is a detailed technical specification for a STERIS 3 Basin 120" Adjustable Height Reprocessing Sink, with comprehensive engineering drawings and utility requirements. The sink is a stainless steel (304 series) medical equipment design intended for endoscope and medical instrument processing, featuring an adjustable height range of 33"-41" with a total unit width of 120", depth of 28", and three basins.
Key technical specifications include multiple water connections (hot, cold, pure water), compressed air and water gun interfaces, electrical requirements of 120V 15A GFCI protection, and a LINAK DL2 lift system with 562 lbs weight capacity per adjustment. The sink requires specific compressed air quality meeting ISO 8573-1:2010 standards, with precise pressure (29-145 psi), flow rate, and air purity parameters. Installation demands include maintaining specific wall clearances, floor drain positioning, and seismic anchoring considerations. The shipping weight is approximately 690 lbs, with operating dimensions allowing flexibility in medical facility environments.
View the file
Other files
Show 40 more
Latest 50 of 61 listed here. See the rest on GovTribe.
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
7/8 [22] MIN
SEE NOTE #2
28-1/16 [712]
UNIT DEPTH17 [432]
BASIN
3X 30 [762]
BASIN
23-1/16 [585]
DRAIN BOARD
13 [330]
DRAIN BOARD
120 [3048] UNIT WIDTH
33 [838]
MINIMUM
WORKING
HEIGHT
3/4 MAX
[19]
11-9/16 [293]
BACKSPLASH
4 [102]
BACKSPLASH
42 [1041]
MAXIMUM
WORKING
HEIGHT
52-17/32 [1335]
MAXIMUM
HEIGHT
T I T L E
3 BASIN 120" ADJUSTABLE HEIGHT REPROCESSING SINK
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:
A 1.000 Kosir, Joe SINK50120 F.7 11002996 F This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
REVISION: ECN #:
REVISION HISTORY DEPICTS CHANGES FROM
PRIOR REVISIONF 106432
INDEX DESCRIPTION OF CHANGE
1 SHEET 1- NOTE 2 ADDED MINIMUM. NOTE 4 CHANGED 300 TO 304
2 SHEET 3- CHANGED UTILITY REQUIREMENTS FROM 1/2" NPSM TO 3/8" COMPRESSION (2 PLACES) 3 SHEET 3- CHANGED UTILITY REQUIREMENTS FROM QUICK CONNECTOR FLEX COIL TO 3/8" COMPRESSION 4 SHEET 3- CHANGED UTILITY REQUIREMENTS FROM 1/2" NPSM WITH ...ADAPTER (INCLUDED)C TO 1/2" NPT
5 SHEET 3- CHANGED UTILITY REQUIREMENTS FROM 1/4" NPFT HOSE INCLUDED TO 1/4" NPT
6 SHEET 3- CHANGED NOTE 5 FROM "PREVENTED" TO "PREVENTION"
7 SHEET 4- CHANGED WALL AND FLOOR VIEW DRAIN DETAILS AND NOTES
8 SHEET 4- MOVED SEISMIC ANCHOR VIEW TO SHEET 6
9 SHEET 6- ADDED NOTE 2
DETAIL A
BASIN DEPTH 10" [254] DETAIL A
F
F
NOTES:
1. UNLESS OTHERWISE NOTED, ALL DIMENSIONS ARE IN INCHES [MM].
2. INSTALLER MUST MAINTAIN A MINIMUM GAP OF .875" TO ALL ADJACENT WALLS AS PINCH POINT PROTECTION.
3. UNIT TO BE INSTALLED AGAINST A WALL, FOR BACK TO BACK INSTALLATION A PONY WALL IS REQUIRED.
4. 304 SERIES STAINLESS STEEL CONSTRUCTION.
5. PONY WALL HEIGHT MUST BE 53"[1346].
6. WALL MUST BE FREE OF OBSTRUCTIONS TO 54"[1372].
ADJUSTABLE HEIGHT FEET (4)
FRONT VIEW
TOP VIEW
RIGHT SIDE VIEW
SHOWN AT
MAX HEIGHT
56-1/4 [1429]A
63-3/4 [1619]WG
[686]
13-1/2 [343]
75 [1905]
SUPPLY CONNECTIONS
TO BE IN THIS ZONE
22-3/8 [568]
29 [737]PW
40-1/2 [1029]HW
48-1/2 [1232]CW
71-1/2 [1816]HW
79-1/2 [2019]CW
91 [2311]PW
MAXIMUM
80-3/4 [2051]
MINIMUM
72-3/4 [1848]
6-3/8 [162]
T I T L
A 1.000 Kosir, Joe SINK50120 F.7 11002996 F This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
FRONT VIEW
RIGHT SIDE VIEW
SHOWN AT
MAX HEIGHT
NOTE: WORKFLOW CAN BE REVERSED BY MOVING PURE WATER FAUCET
FROM RIGHT SIDE TO HOLE ON LEFT THAT IS CAPPED.
ADDITIONALLY THE COMPRESSED AIR AND WATER GUN FIXTURES SHOULD BE SWAPPED.
DB
HW HOT WATER
CW COLD WATER
PW PURE WATER
WG WATER GUN
A COMPRESSED AIR
DB DRAIN BASKET (BASIN)
D DRAIN
T L
A 1.000 Kosir, Joe SINK50120 F.7 11002996 F This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
NOTES:
1. SUPPLY LINES FOR EACH CONNECTION ARE SHIPPED WITH EACH SINK.
2. STERIS RECOMMENDS THE FOLLOWING WATER QUALITY PARAMETERS:
1. 100-200 PPM (500PPM MAX) TOTAL DISOLVED SOLIDS.
2. 70-120 PPM (180PPM MAX) TOTAL ALKALINITY.
3. 6.8-7.0PH (7.5 MAX).
4. 0.1-0.5PPM (1.0PPM MAX) TOTAL SILICA.
3. STERIS DOES NOT SUPPLY A PRESSURE REGULATOR FOR USE WITH AIR GUN.
4. STERIS DOES NOT SUPPLY A PRESSURE REDUCING VALVE FOR USE WITH WATER GUN.
5. BACK FLOW PREVENTION BY CUSTOMER AGAINST BACK SIPHONAGE MAY BE REQUIRED (CHECK LOCAL PLUMBING CODE).
6. FLOOR SINK OUTLET NEEDS TO BE ABLE TO ACCOMODATE WASTE FROM ALL BAYS (CHECK LOCAL PLUMBING CODE).
7. REFERENCES FOR PARAMETERS ARE TO MANUFACTURER'S RECOMMENDATION AND SHOULD BE FOLLOWED ACCORDINGLY.
F
F
F
F
F
F
ADJ. HEIGHT PROCESSING TRIPLE SINK 120" X 28" X 33"-41"
UTILITY REQUIREMENTS CHART (PLUMBING)
UTILITY PARAMETER REQUIREMENT NOTES
HOT WATER
(HW)
CONNECTION TYPE/SIZE 3/8" COMPRESSION
PRESSURE RANGE 15 TO 125 PSI
OPERATING FLOW RATE 1.6 GPM @ MAX PSI
TEMPERATURE RANGE 40 F TO 180 F
WATER QUALITY SEE NOTE #2
COLD WATER
(CW)
CONNECTION TYPE/SIZE 3/8" COMPRESSION
PRESSURE RANGE 15 TO 125 PSI
OPERATING FLOW RATE 1.6 GPM @ MAX PSI
TEMPERATURE RANGE 70 F MAX
WATER QUALITY SEE NOTE #2
PURE WATER
(PW)
CONNECTION TYPE/SIZE 1/2" NPT
PRESSURE RANGE 20 TO 125 PSI
OPERATING FLOW RATE .25 TO 2.00 GPM 15%
TEMPERATURE RANGE 40 F TO 140 F
WATER QUALITY
WATER GUN
(WG)
CONNECTION TYPE/SIZE 3/8" COMPRESSION
PRESSURE RANGE 0 TO 44 PSI
OPERATING FLOW RATE
TEMPERATURE RANGE 40 F TO 125 F
WATER QUALITY SEE NOTE #2
AIR GUN
(A)
CONNECTION TYPE/SIZE 1/4" NPT CONNECTION MUST BE MADE
PER LOCAL CODE
PRESSURE 30 PSI MAX
TIP 1
CONSUMPTION AT 30 PSI 2 SCFM
TIP 2
TIP 3
TIP 4
TIP 5
TIPS 4 & 5 MUST BE OPERATED AT 30 PSI OR LESS TO COMPLY WITH OSHA DIRECTIVE
BASIN DRAIN
(BD) CONNECTION TYPE
1-1/2" ABS P-TRAP (INCLUDED)
2" COPPER P-TRAP (ACCESSORY)
DRAIN RATE AT 20 GPM
EACH BASIN (ESTIMATED)
17-15/16 [455]
17-15/16 [455]
[787]
CENTER TO CENTER
DRAIN LOCATION
[787]
CENTER TO CENTER
DRAIN LOCATION
9-3/8 [238]
WALL CONNECTION
HEIGHT FROM FLOOR
[152]
[152]
65-3/8 [1662]
T I T L
A 1.000 Kosir, Joe SINK50120 F.7 11002996 F This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
FLOOR DRAIN CONNECTIONS
TO BE IN THIS ZONE
FRONT VIEW
FLOOR NOTES:
1. FLOOR DRAIN IS PREFERED.
2. DRAIN CANNOT BE LOCATED WITHIN 6" OF LEGS.
DRAIN
WALL NOTES:
1. EACH DRAIN INDIVIDUALLY TRAPPED.
2. WHEN TRAPS ARE NOT REQUIRED, ELBOWS CAN BE PURCHASED LOCALLY.
3. ROUGH-IN TOLERANCE 3/8"
DRAIN LOCATION, (WALL)
RIGHT SIDE VIEW
SHOWN AT
MAX HEIGHT
DRAIN
F
F
F
F
DRAIN LOCATION, (FLOOR)
FRONT VIEW RIGHT SIDE VIEW
SHOWN AT
MAX HEIGHT
[483]
[152]
1 [25]
[76]
T I T L
A 1.000 Kosir, Joe SINK50120 F.7 11002996 F This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
15A 120VAC GFCI
PROTECTED OUTLET
UTILITY (ELECTRICITY)
CONNECTION POSITION
RECOMMENDED
PARTIAL FRONT VIEW
INCLUDES 10'[3048] CORD
ADJ. HEIGHT REPROCESSING TRIPLE SINK 120" X 28" X 33" TO 41"
UTILITY SPECIFICATIONS CHART (ELECTRICAL)
ELECTRICITY
UTILITY PARAMETER REQUIREMENT NOTES
ELECTRICITY 120V 15A GFCI PROTECTED OUTLET
LINAK DL2 LIFT SYSTEM
LINAK DL2
LIFT SYSTEM
WEIGHT CAPACITY 562LBS [2500 N] EACH
ADJUSTMENT RANGE 8.0" [203,2]
LIFT SPEED UP TO 3/8"[10] PER SECOND
VOLTAGE 120VAC, 3A WITH OVERLOAD
PROTECTION
14 [356]
SEE NOTE BELOW
1-1/2 [38]
4X 9/16 [14]
FOR 1/2-13
SEISMIC ANCHORS
102-7/16 [2602]
SEE NOTE BELOW
95-1/16 [2415]
[762]
6-13/16 [173]
[406]
T I T L
A 1.000 Kosir, Joe SINK50120 F.7 11002996 F This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
BACK WALL
F
NOTE:
1. CONCRETE FLOOR MUST REMAIN FREE OF DEFECTS IN THESE AREAS FOR
INSTALLATION OF SEISMIC ANCHORS (NOT PROVIDED BY STERIS).
2. REFER TO EQUIPMENT SPECIFIC SEISMIC DRAWING FOR MORE INFORMATION.
SEISMIC ANCHOR HOLE POSITIONS
PHYSICAL SPECIFICATIONS
SHIPPING DIMENSIONS L X W X H 132L X 36W X 48H
SHIPPING WEIGHT 690 LBS
OPERATING DIMENSIONS L X W X H 120" X 28" X 44-9/16" TO 52-9/16"
[3048 X 711 X 1132 TO 1412]
OPERATING WEIGHT EMPTY 490 LBS
WORKING HEIGHT 33"[838] TO 41"[1041]
3X 30 [762]
BASIN
17 [432]
BASIN
97-1/2 [2477]
UNIT WIDTH
33 [838]
MINIMUM
WORKING
HEIGHT
41 [1041]
MAXIMUM
WORKING
HEIGHT
28-1/16 [712]
UNIT DEPTH
7/8 [22]MIN
SEE NOTE #2
23-1/16 [585]
DRAIN BOARD 3/4 MAX
[19]
28-15/16 [734]
MINIMUM
52-9/16 [1335]
MAXIMUM
HEIGHT
4 [102]
BACKSPLASH
11-9/16 [293]
BACKSPLASH
T I T L E
3 BASIN 97.5" ADJUSTABLE HEIGHT REPROCESS SINK
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:
A 0.022 Kosir, Joe SINK5097 F.3 11002999 F
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
REVISION: ECN #:
REVISION HISTORY DEPICTS CHANGES FROM
PRIOR REVISIONF 106432
INDEX DESCRIPTION OF CHANGE
1 ADD 2" COPPER P-TRAP AS ACCESSORY UNDER UTILITY REQUIREMENTS ALONG WITH ESTIMATED DRAIN RATE 2 SHEET 3- CHANGED UTILITY REQUIREMENTS FROM 1/2" NPSM TO 3/8" COMPRESSION (2 PLACES) 3 SHEET 3- CHANGED UTILITY REQUIREMENTS FROM QUICK CONNECTOR FLEX COIL TO 3/8" COMPRESSION 4 SHEET 3- CHANGED UTILITY REQUIREMENTS FROM 1/2" NPSM WITH ...ADAPTER (INCLUDED)C TO 1/2" NPT
5 SHEET 3- CHANGED UTILITY REQUIREMENTS FROM 1/4" NPFT HOSE INCLUDED TO 1/4" NPT
6 SHEET 3- CHANGED NOTE 5 FROM "PREVENTED" TO "PREVENTION"
7 SHEET 4- CHANGED WALL AND FLOOR VIEW DRAIN DETAILS AND NOTES
8 SHEET 4- MOVED SEISMIC ANCHOR VIEW TO SHEET 6
9 SHEET 6- ADDED NOTE 2
DETAIL A
NOTES:
1. UNLESS OTHERWISE NOTED, ALL DIMENSIONS ARE IN INCHES [MM].
2. INSTALLER MUST MAINTAIN A MINIMUM GAP OF .875" TO ALL ADJACENT WALLS AS PINCH POINT PROTECTION.
3. UNIT TO BE INSTALLED AGAINST A WALL, FOR BACK TO BACK INSTALLATION A PONY WALL IS REQUIRED.
4. 304 SERIES STAINLESS STEEL CONSTRUCTION.
5. PONY WALL HEIGHT MUST BE 53"[1346].
6. WALL MUST BE FREE OF OBSTRUCTIONS TO 54"[1372].
TOP VIEW
BASIN DEPTH 10" [254]
ADJUSTABLE HEIGHT FEET (4)
FRONT VIEW
DETAIL A RIGHT SIDE VIEW
SHOWN AT
MAX HEIGHTF
F
17-3/4 [451]PW
29-1/4 [743]HW
37-1/4 [946]CW
45 [1143]A
52-1/2 [1334]WG
60-1/4 [1530]HW
68-1/4 [1734]CW
79-3/4 [2026]PW
27 [686]
6-3/8 [162]
13-1/2 [343]
75 [1905]
SUPPLY CONNECTIONS
TO BE IN THIS ZONE
11-1/8 [282]
MAXIMUM
80-3/4 [2051]
MINIMUM
72-3/4 [1848]
T I T L E
3 BASIN 97.5" ADJUSTABLE HEIGHT REPROCESS SINK
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:
A 0.016 Kosir, Joe SINK5097 F.3 11002999 F
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
HW HOT WATER
CW COLD WATER
PW PURE WATER
WG WATER GUN
A COMPRESSED AIR
DB DRAIN BASKET (BASIN)
D DRAIN
NOTE: WORKFLOW CAN BE REVERSED BY MOVING PURE WATER FAUCET
FROM RIGHT SIDE TO HOLE ON LEFT THAT IS CAPPED.
ADDITIONALLY THE COMPRESSED AIR AND WATER GUN FIXTURES SHOULD BE SWAPPED.
RIGHT SIDE VIEW
SHOWN AT
MAX HEIGHT
FRONT VIEW
DB
T L E
3 BASIN 97.5" ADJUSTABLE HEIGHT REPROCESS SINK
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:
A 0.016 Kosir, Joe SINK5097 F.3 11002999 F
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
ADJ. HEIGHT PROCESSING TRIPLE SINK 97.5" X 28" X 33"-41"
UTILITY REQUIREMENTS CHART (PLUMBING)
UTILITY PARAMETER REQUIREMENT NOTES
HOT WATER
(HW)
CONNECTION TYPE/SIZE 3/8" COMPRESSION
PRESSURE RANGE 15 TO 125 PSI
OPERATING FLOW RATE 1.6 GPM @ MAX PSI
TEMPERATURE RANGE 40 TO 180 F
WATER QUALITY SEE NOTE #2
COLD WATER
(CW)
CONNECTION TYPE/SIZE 3/8" COMPRESSION
PRESSURE RANGE 15 TO 125 PSI
OPERATING FLOW RATE 1.6 GPM @ MAX PSI
TEMPERATURE RANGE 70 F MAX
WATER QUALITY SEE NOTE #2
PURE WATER
(PW)
CONNECTION TYPE/SIZE 1/2" NPT
PRESSURE RANGE 20 TO 125 PSI
OPERATING FLOW RATE .25 TO 2.00 GPM ±15%
TEMPERATURE RANGE 40° F TO 140°F
WATER QUALITY
WATER GUN
(WG)
CONNECTION TYPE/SIZE 3/8" COMPRESSION
PRESSURE RANGE 0 TO 44 PSI
OPERATING FLOW RATE
TEMPERATURE RANGE 40° F TO 125°F
WATER QUALITY SEE NOTE #2
AIR GUN
(A)
CONNECTION TYPE/SIZE 1/4" NPT CONNECTION MUST BE MADE
PER LOCAL CODE
PRESSURE 30 PSI MAX
TIP 1
CONSUMPTION AT 30 PSI 2 SCFM
TIP 2
TIP 3
TIP 4
TIP 5
TIPS 4 & 5 MUST BE OPERATED AT 30 PSI OR LESS TO COMPLY WITH OSHA DIRECTIVE
BASIN DRAIN
(BD) CONNECTION TYPE
1-1/2" ABS P-TRAP (INCLUDED)
2" COPPER P-TRAP (ACCESSORY)
DRAIN RATE AT 20 GPM
EACH BASIN (ESTIMATED)
NOTES:
1. SUPPLY LINES FOR EACH CONNECTION ARE SHIPPED WITH EACH SINK.
2. STERIS RECOMMENDS THE FOLLOWING WATER QUALITY PARAMETERS:
1. 100-200 PPM (500PPM MAX) TOTAL DISOLVED SOLIDS.
2. 70-120 PPM (180PPM MAX) TOTAL ALKALINITY.
3. 6.8-7.0PH (7.5 MAX).
4. 0.1-0.5PPM (1.0PPM MAX) TOTAL SILICA.
3. STERIS DOES NOT SUPPLY A PRESSURE REGULATOR FOR USE WITH AIR GUN.
4. STERIS DOES NOT SUPPLY A PRESSURE REDUCING VALVE FOR USE WITH WATER GUN.
5. BACK FLOW PREVENTION BY CUSTOMER AGAINST BACK SIPHONAGE MAY BE REQUIRED (CHECK LOCAL PLUMBING CODE).
6. FLOOR SINK OUTLET NEEDS TO BE ABLE TO ACCOMODATE WASTE FROM ALL BAYS (CHECK LOCAL PLUMBING CODE).
7. REFERENCES FOR PARAMETERS ARE TO MANUFACTURER'S RECOMMENDATION AND SHOULD BE FOLLOWED ACCORDINGLY.
F
F
F
F
17-15/16 [455]
17-15/16 [455]
[787]
CENTER TO CENTER
DRAIN LOCATION
[787]
CENTER TO CENTER
DRAIN LOCATION
9-3/8 [238]
WALL CONNECTION
HEIGHT FROM FLOOR
[152]
65-3/8 [1662]
[153]
T I T L E
3 BASIN 97.5" ADJUSTABLE HEIGHT REPROCESS SINK
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:
A 0.016 Kosir, Joe SINK5097 F.3 11002999 F
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
FLOOR NOTES:
1. FLOOR DRAIN IS PREFERED.
2. DRAIN CANNOT BE LOCATED WITHIN 6" OF LEGS.
FRONT VIEW
FRONT VIEW
DRAIN LOCATION, (FLOOR)
FLOOR DRAIN CONNECTIONS
TO BE IN THIS ZONE
DRAIN ℄
RIGHT SIDE VIEW
SHOWN AT
MAX HEIGHT
WALL NOTES:
1. EACH DRAIN INDIVIDUALLY TRAPPED.
2. WHEN TRAPS ARE NOT REQUIRED, ELBOWS CAN BE PURCHASED LOCALLY.
3. ROUGH-IN TOLERANCE ±3/8"
DRAIN LOCATION, (WALL)
RIGHT SIDE VIEW
SHOWN AT
MAX HEIGHT
DRAIN ℄
F
F
[483]
[152]
[76]
T I T L E
3 BASIN 97.5" ADJUSTABLE HEIGHT REPROCESS SINK
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:
A 0.022 Kosir, Joe SINK5097 F.3 11002999 F
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
ADJ. HEIGHT REPROCESSING TRIPLE SINK 97.5" X 28" X 33" TO 41"
UTILITY SPECIFICATIONS CHART (ELECTRICAL)
ELECTRICITY
UTILITY PARAMETER REQUIREMENT NOTES
ELECTRICITY 120V 15A GFCI PROTECTED OUTLET
LINAK DL2 LIFT SYSTEM
LINAK DL2
LIFT SYSTEM
WEIGHT CAPACITY 562LBS [2500 N] EACH
ADJUSTMENT RANGE 8.0" [203,2]
LIFT SPEED UP TO 3/8"[10] PER SECOND
VOLTAGE 120VAC, 3A WITH OVERLOAD
PROTECTION
PARTIAL FRONT VIEW
UTILITY (ELECTRICITY)
CONNECTION POSITION
RECOMMENDED
INCLUDES 10'[3048] CORD
15A 120VAC GFCI
PROTECTED OUTLET
1-1/2 [38]
14 [356]
SEE NOTE BELOW
⌀9/16 [14]
FOR 1/2-13
SEISMIC ANCHORS
102-7/16 [2602]
SEE NOTE BELOW
95-1/16 [2415]
6-13/16 [173]
[406]
[762]
TOLERANCES: +/- INCH (MM) IF NOT INDICATED OTHERWISE MATERIAL / NOTES
DIMENSION 0-48 48+ DIMENSION 0-4 4-24 24+
NA
FINISH: NA
MACHINING FRAC.
DEC.
1/64 0.005
1/32 0.015
PLASTIC FRAC.
DEC.
1/64 0.005
1/32 0.020
1/16 N/A
SHEET METAL / SAWING 1/32 1/16 DIMENSION 0-12 12-60 60+ COMPONENT/ASSEMBLY MUST BE RoHS COMPLIANT
DIMENSION 0-1 1+ WELDED ASSEMBLY 1/32 1/16 3/32 DIMENSION 0-3/8 3/8-3/4 3/4 +
BORING 0.001 0.002 SPOT WELD 1/16 3/32 HOLE DIAMETER 0.005 0.015 1/32
SURFACE FINISH 63 machined surface (Visual) SQUARENESS 1/16 total 1/8 total ANGLE 1° WEIGHT: [.2] LBS.
T I T L E
3 BASIN 97.5" ADJUSTABLE HEIGHT REPROCESS SINK
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:
A 0.022 Kosir, Joe SINK5097 F.3 11002999 F
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
PHYSICAL SPECIFICATIONS
SHIPPING DIMENSIONS L X W X H 132L X 36W X 48H
SHIPPING WEIGHT 580 LBS
OPERATING DIMENSIONS L X W X H 97.5" X 28" X 44-9/16" TO 52-9/16"
[2477 X 711 X 1132 TO 1412]
OPERATING WEIGHT EMPTY 380 LBS
WORKING HEIGHT 33"[838] TO 41"[1041]
NOTE:
1. CONCRETE FLOOR MUST REMAIN FREE OF DEFECTS IN THESE AREAS FOR
INSTALLATION OF SEISMIC ANCHORS (NOT PROVIDED BY STERIS).
2. REFER TO EQUIPMENT SPECIFIC SEISMIC DRAWING FOR MORE INFORMATION.
SEISMIC ANCHOR HOLE POSITIONS
SCALE 0.050
BACK WALL
ENDODRY™
Drying and Storage Cabinet | Compressed Air Requirements
The ENDODRY™ Drying and Storage Cabinet requires compressed air which must meet specific incoming air quality requirements. This ensures optimal cabinet performance with continuous air flow to safely eliminate all residual water from the endoscopes being dried, which may otherwise promote the growth of organisms.
ENDODRY Cabinet Air Requirements
1. Quality: ISO 8573-1:2010 minimum class 1.4.1, recommended class 1.2.1
• Particle Size: Incoming air filtered down to 1 micron in size
• Moisture Content: Minimum pressure dew point (PDP) of +3°C or better required and no liquid water, -40°C or better recommended
• Oil Content: < 0.01mg of oil per cubic meter of compressed air (liquid, aerosol or vapor oil)
2. Pressure: The below pressures are required at the point of cabinet connection.
Consult with the Facilities Maintenance Department or qualified air vendor to identify potential losses in air supply.
• ENDODRY Cabinet for ADVANTAGE PLUS™ Hookups: 29 – 145 psi (2 – 10 bar)
• ENDODRY Cabinet for DSD Hookups: 73-145 psi (5 – 10 bar)
3. Volume Flow: The below volume flows are required at the point of cabinet connection.
• ENDODRY Cabinet for ADVANTAGE PLUS Hookups: Max demand of 4.2 scfm (120 l/min ANR) per cabinet
• ENDODRY Cabinet for DSD Hookups: Max demand of 7 scfm (200 l/min ANR) per cabinet
Given the air volume flow demand difference, the ENDODRY Cabinet for ADVANTAGE PLUS Hookups is recommended when using or moving to the ADVANTAGE PLUS™ Automated Endoscope Reprocessor platform. This recommendation also applies to non-Medivators AER’s, due to lower air consumption and potential ADVANTAGE PLUS Endoscope Reprocessor upgrade later.
ENDODRY Cabinet Air Grade
Instrument Air (NFPA 99) is the recommended grade of air to use with the ENDODRY Cabinet. Additional particulate filtration may be required to meet the ISO class listed above. If house/facility grade air is used (or other inferior air source), additional air treatment may also be needed to achieve the moisture and oil content listed above. Medical facilities will need to consult their Facilities Maintenance Department or their qualified air vendor to identify additional filtration needs to meet the air requirements listed above.
© 2018 Cantel 50098-1660-EN / A
Air Responsibility
• The medical facility is responsible for selecting and supplying their compressed air solution.
• The facility should consult their internal Facilities Maintenance Department or a qualified air vendor.
• The facility should have an air backup plan, as air compressors require service and maintenance.
• To assist facilities, Medivators has compiled a list of compressors below that can be purchased directly from either the manufacturer or an authorized dealer. However, Medivators has not validated and does not endorse any of the listed compressors, and is not responsible for air compressor warranty, service and maintenance. The list has been compiled based on no additional air demand and supply plumbing of 100ft or less. Work with the air supply vendor or Facilities Maintenance Department for the specific application.
ADVANTAGE PLUS™ Reprocessor Hookups DSD Hookups
# of Cabinets
BeaconMedaes (w/ Additional
Dryer)
Powerex (w/ Dryer)
# of Cabinets
BeaconMedaes (w/ Additional Dryer)
Powerex (w/ Dryer)
1 LES03-145T-RD SES0308TM** 1 LES05-145T-RD SES0508TM**
2 LES05-145T-RD SES0508TM** 2 LES10-145RD LSE1504
3 LES07-145T-RD LSE1504 3 LES15-145RD LSE1504
4 LES10-145T-RD LSE1504 4 LES20-145RD LSE2005
5 LES15-145RD LSE1504 5 LES20-145RD LSE2005
Additional capacities are available. Contact Vendor for additional details.
Compressor sizes assume the inclusion of a desiccant drying solution to obtain a pressure dew point (PDP) of -40°C or better.
Systems are available with 208V/230V/460V **Single-phase systems may need additional receiver Footprint Ranges from 26”x40” to 66”x30” depending on compressor size (with heights from 46” to 62”) Powerex Contact Information: 1-888-769-7979 or http://www.powerexinc.com/ BeaconMedaes Contact Information: 1-888-463-3427 (Option 2 - Tech Support, Option 3 - customer service) or www.beaconmedaes.com (includes regional directory)
Medivators Inc.
14605 28th Avenue North Minneapolis, MN
55447-4822 USA
Tel: +1.800.444.4729 Fax: +1.763.553.3387
Medivators BV Sourethweg 11 6422 PC Heerlen The Netherlands Tel: +31.45.5.471.471 Fax: +31.45.5.429.595
ESCAD Medical GmbH Zur Ohmdwiesen 5 88633 Heiligenberg Germany Tel: +49.7554.9999.500 Fax: +49.7554.9999.558 www.cantelmedical.com
Cantel Medical Asia/Paci�c Pte. Ltd.
1A International Business Park #05-01 Singapore 609933 Tel: +65.6227.9698 Fax: +65.6225.6848
ADVANTAGE PLUS™ and ENDODRY™ are trademarks of Medivators Inc.
Medivators Headquarters 14605 28th Avenue North Minneapolis, Minnesota 55447 Tel: 1.800.444.4729 www.cantelmedical.com
E
D
C
B
A
87654321
D
C
B
A en
STATUS:DOCUMENT:
MATERIAL:
UNLESS OTHERWISE SPECIFIED
DIMENSIONS ARE IN MILLIMETERS
TOLERANCES.
DIN ISO 2768-1 class: m DIN ISO 2768-2 class: K
FINISH:
CHECKEDNAMEDATEDESCRIPTIONREV
DATE
QUALITY CONTROL
CHECKED
DRAWN
NAME
A3
DO NOT SCALE DRAWING
TITLE:
DWG NO.
SCALE:
11.08.17Stebner
30 00 20 001
Installation requirements -horizontal storage cabinet
1/11
Klein
1:10
WEIGHT: g
09.08.17
Copyright according to ISO 16016 and DIN 34 Passing on and copying of this document, utilization and communication of its contents is not permitted, unless expressly allowed.
Offenders will be liable for compensation.
All rights reserved for IP application.
ISO 128-30
Methode 1 a typing mistakes and packaging adjusted b Adjustment of the temperatures in technical data
Storage cabinet with drying function for flexible endoscopes/ Lagerungsschrank mit Trocknungsfunktion für flexible Endoskope
Cabinet with article no.: / Schränke mit Artikel Nr.:
#015120-#015125 (ENDODRY)
#015126-#015131 (ENDODRY one-click)
#015760-#015765 (endoSTORE BHT)
#015906-#015911 (endoSTORE one-click)
17.05.18 Klein Stebner
Klein Stebner11.09.18 w h d
E
D
C
B
A
87654321
D
C
B
A en
STATUS:DOCUMENT:
MATERIAL:
UNLESS OTHERWISE SPECIFIED
DIMENSIONS ARE IN MILLIMETERS
TOLERANCES.
DIN ISO 2768-1 class: m DIN ISO 2768-2 class: K
FINISH:
CHECKEDNAMEDATEDESCRIPTIONREV
DATE
QUALITY CONTROL
CHECKED
DRAWN
NAME
A3
DO NOT SCALE DRAWING
TITLE:
DWG NO.
SCALE:
11.08.17Stebner
30 00 20 001
Installation requirements -horizontal storage cabinet
2/11
Klein
WEIGHT: g
09.08.17
Copyright according to ISO 16016 and DIN 34 Passing on and copying of this document, utilization and communication of its contents is not permitted, unless expressly allowed.
Offenders will be liable for compensation.
All rights reserved for IP application.
ISO 128-30
Methode 1
Definition of the locally connections / Definition der bauseitigen Anschlüsse min. permitted floor load / minimale Nutzlast Boden: 4,4 kN/m² level and stable surface necessary / ebener und stabiler Untergrund erforderlich
Technical specifications / Technische Daten:
Dimension cabinet/ Außenmaße:
w: 600 mm h: 1900mm d: 540mm Weight/Gewicht: 145 kg
Dimension standard packaging / Maße für Standardverpackung:
w: 800 mm h: 2200 mm d: 1200 mm Weight/Gewicht: 220 kg
DL
EDV
STDPE
Notice / Hinweis Comment / Bemerkung
Electrical disconnection / Trennung vom Stromnetz
The power cable of the cabinet serves as a separation device from the power supply.
If this can not be achieved properly, a suitable device must be installed on the premise (e.g. central emergency stop): / Das Netzkabel des Schrankes dient als Trenneinrichtung vom Stromnetz. Ist dieses nicht einwandfrei zu erreichen, muss hausseitig eine geeignete Einrichtung installiert werden (bspw.
zetraler Not-Halt).
Electrical connections / Elektroanschlüsse
Electrical connections must be installed in accordance with applicable standards. / Elektroanschlüsse sind nach geltenden Vorschriften (z.B. VDE) zu installieren.
Reference / Bezug All dimensions refer to the finished floor and the finished wall. / Alle Maßangaben beziehen sich auf Fertigfußboden und Fertigwand.
Technical Data / Technische Daten Working Condition /
Betriebsbedingung Temperature/Temperatur: 50 °F-95 °F / 10 °C-35 °C Humidity / Luftfeuchtigkeit: 10 %rel.-75 %rel.
Noise level/ Geräuschpegel
55dB(A) +-5dB(A) frontal 1m distance from a single cabinet in quite environment / frontal 1m Abstand, einzelner Schrank, in ruhiger Umgebung
Waste heat/ Abwärme max. 256Btu/h /max.75W
Mark / Bezeichnung
Connection / Anschluss
Comment / Bemerkung
STD power socket / Netzsteckdose 100-240VAC, 50/60Hz,16A, max. 75W, <1kWh/d
EDV
network connection /
Netzwerkanschluss
(RJ45)
TCP-Ports: 80; 20; 21; 445; 1433; 1034; 37223 UDP-Ports: 80; 1900 LM- authentication for network logons
DL
compressed air supply / Druckluftanschluss
(DIN 13260-2,Profil Nr.2 or open end/hose barb
DN6mm (1/4"))
#015120-#015125, #015760-#015765: #015126-#015131, #015906-#015911:
2-10 bar (29-145 psi)
3.5 CFM (ANR) resp. 0.6 CFM @ 72.5 psi /
100l/min (ANR), bzw. 17 l/min @ 5bar Ø2.2CFM (ANR) / Ø63 l/min (ANR)
5-10 bar (72,5-145 psi) 7 CFM (ANR) resp. 1.2 CFM @ 72.5 psi / 200 l/min (ANR) bzw. 33 l/min @ 5 bar Ø3.3CFM (ANR) / Ø94 l/min (ANR)
ISO 8573-1:2010 min. class 1.4.1 on site supply/bauseitige Leitung min. DN10 connection tube/Anschlussschlauch DN6 max. L=2000
PE
potential earth / Potentialausgleich
(DIN 42801 6mm Plug) 4mm²
E
D
C
B
A
87654321
D
C
B
A en
STATUS:DOCUMENT:
MATERIAL:
UNLESS OTHERWISE SPECIFIED
DIMENSIONS ARE IN MILLIMETERS
TOLERANCES.
DIN ISO 2768-1 class: m DIN ISO 2768-2 class: K
FINISH:
CHECKEDNAMEDATEDESCRIPTIONREV
DATE
QUALITY CONTROL
CHECKED
DRAWN
NAME
A3
DO NOT SCALE DRAWING
TITLE:
DWG NO.
SCALE:
11.08.17Stebner
30 00 20 001
Installation requirements -horizontal storage cabinet
3/11
Klein
WEIGHT: g
09.08.17
Copyright according to ISO 16016 and DIN 34 Passing on and copying of this document, utilization and communication of its contents is not permitted, unless expressly allowed.
Offenders will be liable for compensation.
All rights reserved for IP application.
ISO 128-30
Methode 1
Overview variants / Varianten Übersicht
E
D
C
B
A
87654321
D
C
B
A en
STATUS:DOCUMENT:
MATERIAL:
UNLESS OTHERWISE SPECIFIED
DIMENSIONS ARE IN MILLIMETERS
TOLERANCES.
DIN ISO 2768-1 class: m DIN ISO 2768-2 class: K
FINISH:
CHECKEDNAMEDATEDESCRIPTIONREV
DATE
QUALITY CONTROL
CHECKED
DRAWN
NAME
A3
DO NOT SCALE DRAWING
TITLE:
DWG NO.
SCALE:
11.08.17Stebner
30 00 20 001
Installation requirements -horizontal storage cabinet
4/11
Klein
1:20
WEIGHT: g
09.08.17
Copyright according to ISO 16016 and DIN 34 Passing on and copying of this document, utilization and communication of its contents is not permitted, unless expressly allowed.
Offenders will be liable for compensation.
All rights reserved for IP application.
ISO 128-30
Methode 1
A
A shape fillet radius max. 50mm Hohlkehlen Radius max. 50mm 12
0 110
R520 R52
A-A
DL/STD/EDV/PE
E
D
C
B
A
87654321
D
C
B
A en
STATUS:DOCUMENT:
MATERIAL:
UNLESS OTHERWISE SPECIFIED
DIMENSIONS ARE IN MILLIMETERS
TOLERANCES.
DIN ISO 2768-1 class: m DIN ISO 2768-2 class: K
FINISH:
CHECKEDNAMEDATEDESCRIPTIONREV
DATE
QUALITY CONTROL
CHECKED
DRAWN
NAME
A3
DO NOT SCALE DRAWING
TITLE:
DWG NO.
SCALE:
11.08.17Stebner
30 00 20 001
Installation requirements -horizontal storage cabinet
5/11
Klein
1:20
WEIGHT: g
09.08.17
Copyright according to ISO 16016 and DIN 34 Passing on and copying of this document, utilization and communication of its contents is not permitted, unless expressly allowed.
Offenders will be liable for compensation.
All rights reserved for IP application.
ISO 128-30
Methode 1
420 600
B
B shape fillet radius max. 50mm Hohlkehlen Radius max. 50mm 4
0 110
R5
R
R520
R520
B-B
DL/STD/EDV/PE DL/STD/EDV/PE
Double wing door not possible.
Scanner module collides.
Flügeltüren nicht möglich.
Scanner Module kollidiert.
E
D
C
B
A
87654321
D
C
B
A en
STATUS:DOCUMENT:
MATERIAL:
UNLESS OTHERWISE SPECIFIED
DIMENSIONS ARE IN MILLIMETERS
TOLERANCES.
DIN ISO 2768-1 class: m DIN ISO 2768-2 class: K
FINISH:
CHECKEDNAMEDATEDESCRIPTIONREV
DATE
QUALITY CONTROL
CHECKED
DRAWN
NAME
A3
DO NOT SCALE DRAWING
TITLE:
DWG NO.
SCALE:
11.08.17Stebner
30 00 20 001
Installation requirements -horizontal storage cabinet
6/11
Klein
1:20
WEIGHT: g
09.08.17
Copyright according to ISO 16016 and DIN 34 Passing on and copying of this document, utilization and communication of its contents is not permitted, unless expressly allowed.
Offenders will be liable for compensation.
All rights reserved for IP application.
ISO 128-30
Methode 1
R520
#015120 Endodry os OL #015126 Endodry one-click os OL #015760 endoSTORE BHT os OL #015906 endoSTORE one-click os OL
E
D
C
B
A
87654321
D
C
B
A en
STATUS:DOCUMENT:
MATERIAL:
UNLESS OTHERWISE SPECIFIED
DIMENSIONS ARE IN MILLIMETERS
TOLERANCES.
DIN ISO 2768-1 class: m DIN ISO 2768-2 class: K
FINISH:
CHECKEDNAMEDATEDESCRIPTIONREV
DATE
QUALITY CONTROL
CHECKED
DRAWN
NAME
A3
DO NOT SCALE DRAWING
TITLE:
DWG NO.
SCALE:
11.08.17Stebner
30 00 20 001
Installation requirements -horizontal storage cabinet
7/11
Klein
1:20
WEIGHT: g
09.08.17
Copyright according to ISO 16016 and DIN 34 Passing on and copying of this document, utilization and communication of its contents is not permitted, unless expressly allowed.
Offenders will be liable for compensation.
All rights reserved for IP application.
ISO 128-30
Methode 1
501 49
#015121 Endodry os OR #015127 Endodry one-click os OR #015761 endoSTORE BHT os OR #015907 endoSTORE one-click os OR
E
D
C
B
A
87654321
D
C
B
A en
STATUS:DOCUMENT:
MATERIAL:
UNLESS OTHERWISE SPECIFIED
DIMENSIONS ARE IN MILLIMETERS
TOLERANCES.
DIN ISO 2768-1 class: m DIN ISO 2768-2 class: K
FINISH:
CHECKEDNAMEDATEDESCRIPTIONREV
DATE
QUALITY CONTROL
CHECKED
DRAWN
NAME
A3
DO NOT SCALE DRAWING
TITLE:
DWG NO.
SCALE:
11.08.17Stebner
30 00 20 001
Installation requirements -horizontal storage cabinet
8/11
Klein
1:20
WEIGHT: g
09.08.17
Copyright according to ISO 16016 and DIN 34 Passing on and copying of this document, utilization and communication of its contents is not permitted, unless expressly allowed.
Offenders will be liable for compensation.
All rights reserved for IP application.
ISO 128-30
Methode 1
R520
R520
501 49
#015122 Endodry pt OL/UL #015128 Endodry one-click pt OL/UL #015762 endoSTORE BHT pt OL/UL #015908 endoSTORE one-click pt OL/UL
E
D
C
B
A
87654321
D
C
B
A en
STATUS:DOCUMENT:
MATERIAL:
UNLESS OTHERWISE SPECIFIED
DIMENSIONS ARE IN MILLIMETERS
TOLERANCES.
DIN ISO 2768-1 class: m DIN ISO 2768-2 class: K
FINISH:
CHECKEDNAMEDATEDESCRIPTIONREV
DATE
QUALITY CONTROL
CHECKED
DRAWN
NAME
A3
DO NOT SCALE DRAWING
TITLE:
DWG NO.
SCALE:
11.08.17Stebner
30 00 20 001
Installation requirements -horizontal storage cabinet
9/11
Klein
1:20
WEIGHT: g
09.08.17
Copyright according to ISO 16016 and DIN 34 Passing on and copying of this document, utilization and communication of its contents is not permitted, unless expressly allowed.
Offenders will be liable for compensation.
All rights reserved for IP application.
ISO 128-30
Methode 1
R520
#015123 Endodry pt OR/UL #015129 Endodry one-click pt OR/UL #015763 endoSTORE BHT pt OR/UL #015909 endoSTORE one-click pt OR/UL
E
D
C
B
A
87654321
D
C
B
A en
STATUS:DOCUMENT:
MATERIAL:
UNLESS OTHERWISE SPECIFIED
DIMENSIONS ARE IN MILLIMETERS
TOLERANCES.
DIN ISO 2768-1 class: m DIN ISO 2768-2 class: K
FINISH:
CHECKEDNAMEDATEDESCRIPTIONREV
DATE
QUALITY CONTROL
CHECKED
DRAWN
NAME
A3
DO NOT SCALE DRAWING
TITLE:
DWG NO.
SCALE:
11.08.17Stebner
30 00 20 001
Installation requirements -horizontal storage cabinet
10/11
Klein
1:20
WEIGHT: g
09.08.17
Copyright according to ISO 16016 and DIN 34 Passing on and copying of this document, utilization and communication of its contents is not permitted, unless expressly allowed.
Offenders will be liable for compensation.
All rights reserved for IP application.
ISO 128-30
Methode 1
R520
501 49
#015124 Endodry pt OL/UR #015130 Endodry one-click pt OL/UR #015764 endoSTORE BHT pt OL/UR #015910 endoSTORE one-click pt OL/UR
E
D
C
B
A
87654321
D
C
B
A en
STATUS:DOCUMENT:
MATERIAL:
UNLESS OTHERWISE SPECIFIED
DIMENSIONS ARE IN MILLIMETERS
TOLERANCES.
DIN ISO 2768-1 class: m DIN ISO 2768-2 class: K
FINISH:
CHECKEDNAMEDATEDESCRIPTIONREV
DATE
QUALITY CONTROL
CHECKED
DRAWN
NAME
A3
DO NOT SCALE DRAWING
TITLE:
DWG NO.
SCALE:
11.08.17Stebner
30 00 20 001
Installation requirements -horizontal storage cabinet
11/11
Klein
1:20
WEIGHT: g
09.08.17
Copyright according to ISO 16016 and DIN 34 Passing on and copying of this document, utilization and communication of its contents is not permitted, unless expressly allowed.
Offenders will be liable for compensation.
All rights reserved for IP application.
ISO 128-30
Methode 1
R
#015125 Endodry pt OR/UR #015131 Endodry one-click pt OR/UR #015765 endoSTORE BHT pt OR/UR #015911 endoSTORE one-click pt OR/UR
| 1
USER MANUAL
WASHER-DISINFECTOR
CUSTOMER
REQUIREMENTS
NORTH AMERICA
ENDODRY™
Storage & Drying Sytem
2 |
50098-678-EN Rev E
© 2019 All rights reserved. This publication is protected by copyright. Copying, disclosure to others, or the use of this publication is prohibited without the express written consent of Medivators Inc.
ADVANTAGE PLUS™
is a trademark of Medivators Inc.
DSD EDGE™
is a trademark of Medivators Inc.
ENDODRY™
is a trademark of Medivators Inc.
All other company or product names referenced in this document are trademarks of their respective owner.
Medivators reserves the right to make changes in the specifications shown herein without notice or obligation. Contact your Medivators representative or Medivators customer service for more information.
| 3SITE REQUIREMENTS
This purpose of this document is to provide the facility with requirements to ensure correct installation and subsequent proper performance of the ENDODRY™ Storage and Drying System.
READ THE ENTIRE DOCUMENT.
APPLICABLE DOCUMENTS
• ENDODRY Storage and Drying System User Manual.
Any deviation from these specifications may cause operational issues that would not be covered by the Warranty.
4 |
BY SIGNING BELOW CUSTOMER ACKNOWLEDGES THAT:
1. Customer reviewed these Site Requirements for the ENDODRY™ Storage and Drying System as indicated in this document.
2. The customer assumes complete responsibility for meeting the specifications and requirements set forth in the Site Requirements documentation to ensure proper installation and performance of the ENDODRY Storage and Drying System. Failure of the customer to comply with these Site Requirements and any deviation from the specifications set forth herein may compromise the functionality of the system, and/or may cause operational issues which will not be covered by the Cantel Medical warranty, and are hereby disclaimed by Cantel Medical.
3. The customer assumes complete responsibility for meeting the specifications and requirements set forth in the Site Requirements documentation by the agreed upon installation date between the Customer and Cantel Medical. If the Cantel Medical installation team encounters any delays due to site requirements not meeting specifications and requirements, the Customer will be charged Medivators current rates for stand-by fees, until the site requirements are met and installation activity resumes; or, if it is determined that the delay results in rescheduling the installation to a later date, the Customer will be charged for all expenses at current Medivators rates (flights, lodging, and car rental, if applicable), and travel and labor time related to the return installation dispatch.
4. If installation is canceled or moved 48 hours prior to the agreed upon installation date, no fee will be charged.
5. For customers who request or accept Medivators space planning suggestions for the site prior to installation (“Space Planning Services”), Customer agrees that the Disclaimer, indicated in this document on Page 9, will apply to such Space Planning Services.
Customer Representative
Title
Signature Date Email
| 5
AIR SUPPLY
1. Quality: ISO 8573-1:2010 class 1.2.1
• Particle Size: Incoming air filtered down to <1 micron in size.
• Moisture Content: Pressure dew point (PDP) of -40°C or better required and no liquid water.
• Oil Content: < 0.01mg of oil per cubic meter of compressed air (liquid, aerosol or vapor oil).
2. Pressure: The below pressures are required at the point of cabinet connection. Facility should consult with their Maintenance Department or their qualified air vendor to identify potential issues with the air supply.
• ENDODRY™ Cabinet for ADVANTAGE PLUS™ Endoscope Reprocessor: 58-145 psi (4-10 bar).
• ENDODRY Cabinet for DSD EDGE™ Automated Endoscope Reprocessor: 72-145 psi (5-10 bar).
3. Volume: The below volumes are required at the point of cabinet connection.
• ADVANTAGE PLUS Endoscope Reprocessor/ENDODRY Cabinet: Minimum requirement of 4.2 scfm (120 l/min ANR) per cabinet.
• DSD EDGE Automated Endoscope Reprocessor/ENDODRY Cabinet: Minimum requirement of 7 scfm (200 l/min ANR) per cabinet.
4. Flexible Air Hose: A 5 foot (1.5 Meter) line is included with each cabinet.
• One end will be pre-configured to connect to ENDODRY™ Drying Cabinet.
• Open end will be ¼” or 6mm inner diameter and it is the responsibility of the facility to make that connection to their air supply line.
The compressors listed below are suggestions that the facility can purchase directly from either the compressor manufacturer or authorized dealer. Medivators is not responsible for air compressor warranty, service and maintenance.
ADVANTAGE HOOKUPS DSD HOOKUPS
# of Cabinets
BeaconMedaes (Requires Additional Dryer)
Powerex (w/ Dryer)
# of Cabinets
BeaconMedaes (Requires Additional Dryer)
Powerex (w/ Dryer)
1 LES03-145T-RD-xxx SES0308TM** 1 LES05-145T-RD-xxx SES0508TM**
2 LES05-145T-RD-xxx SES05058TM** 2 LES10-145RD LSE1504
3 LES07-145T-RD-xxx LSE1504 3 LES15-145RD LSE1504
4 LES07-145T-RD-xxx LSE1504 4 LES20-145RD LSE2005
5 LES15-145RD LSE1504 5 LES20-145RD LSE2005
6 |
ELECTRICAL SUPPLY
• 100-240 VAC ± 10%, single phase, 50/60 Hz, 15 amp circuit (120v), 10 amp circuit (230v).
• Power input: 75W
• A NEMA 5-15 power cord for the North America market and a CEE 7 power cord for the European market is included with the ENDODRY™ Storage and Drying System. The cord is approximately 6 feet (1.8 meters) in length.
• A GFI/RCD protected outlet located on the wall above the ENDODRY Storage and Drying System, see Floor & Wall Section.
• Medivators equipment must be on dedicated circuits.
NETWORKING
• The ENDODRY Storage and Drying System is a networkable piece of equipment and has the capabilities to communicate to other Medivators technologies via a LAN interface (RJ45, 10/100 Mbit).
• A 6.5 foot (2 meter) cable is included with the cabinet.
• A RJ45 wall mounted network jack located on the wall above the ENDODRY Storage and Drying System, see Floor & Wall section.
DIMENSIONS
Shipping Dimensions
• Crate Dimensions: H84”x W47”x D31”
• Crate Weight: 583 lbs.
Reprocessor Dimensions
• Dimensions: H74.8” X W23.6” X D21.3”
• Weight: 375 lbs (170 kg)
FLOOR & WALL
• The floor must be level.
• For optimal user-safety and operation, it is recommended that the ENDODRY Storage and Drying System always be placed directly on the floor, and not on an elevated platform.
• The back of the one-sided ENDODRY Storage and Drying System will be placed flush against the wall. All electrical, network and air connections must be placed a minimum of 78.7 inches (200 cm) off of the floor above the cabinet.
• The two-sided ENDODRY Storage and Drying System is designed to fit into an opening in the wall. The rough opening for a single unit is H75.2” X W24” (191cm x 61cm). All electrical, network and air connections must be placed a minimum of 78.7 inches (200 cm) off of the floor above the cabinet. The electrical, network and air connections must be located on the loading side of the ENDODRY Storage and Drying System.
| 7
.6
21.3"
41.7"
EN
D
O D
R Y
St or ag e an d
D ry in g
Sy st em
TM
El ec tri ca l C on ne ct io n
Ai r C on ne ct io n
N et w or k C on ne ct io n
2 3
O ne
-S id ed V er si on
C 3D
V ie w
M in im um fa ci lit y w al l c on ne ct io ns he ig ht is
8.
7” fr om fl oo r.
O pt io ns fo r D oo rs
O pe n Le ft -O
L O pe n
R ig ht -O
R
B To p Vi ew
A Fr on t 3
D V ie w
.6
.3
3.
9” w w w .e nd oi nf ec tio np re ve nt io n.
co m to d ow nl oa d th e R ev it fil e.
Fo r c la rif ic at io n an d/ or q ue st io ns pl ea se c on ta ct T ec hn ic al
S up po rt at
.8
.4
.4
9.
4.5”12” 4.8”
1 3
.8
8 |
62.2"
21.3"
.6
EN
D
O D
R Y
St or ag e an d
D ry in g
Sy st em
TM
El ec tri ca l C on ne ct io n
Ai r C on ne ct io n
N et w or k C on ne ct io n
2 3
Tw o-
Si de d Ve rs io n (P as s-
Th ru
O pt io ns fo r D oo rs
O pe n Le ft/ O pe n
Le ft
-O
LU
L O pe n
R ig ht /O pe n
Le ft
-O R
U L
O pe n Le ft/ O pe n
R ig ht
O
LU
R O pe n
R ig ht /O pe n
R ig ht
O R
U R
B To p Vi ew
A Fr on t 3
D V ie w
W et
S id e
D ry
S id e
W et
S id e
C 3D
V ie w
W et
S id e D ry S id e
1 3
3.
9"
M in im um fa ci lit y w al l c on ne ct io ns he ig ht is
8.
7” fr om fl oo r.
*M ea su re m en t u si ng a
” w al l.
4.5” 12”4.8”
.6
*1
5.
3"
.8 w w w .e nd oi nf ec tio np re ve nt io n.
co m to d ow nl oa d th e R ev it fil e.
Fo r c la rif ic at io n an d/ or q ue st io ns pl ea se c on ta ct T ec hn ic al
S up po rt at
.8
.4
.4
9.
| 9
SPACE-PLANNING SERVICES
For customers who request or accept Medivators space planning suggestions for the site prior to installation (“Space Planning Services”), Customer agrees that the following Disclaimer will apply to such Space Planning Services:
CUSTOMER ACKNOWLEGES AND ACCEPTS THAT SPACE PLANNING SERVICES ARE OFFERED AND
PROVIDED “AS IS” WITH NO REPRESENTATION OR WARRANTY, EXPRESS OR IMPLIED, OF ANY KIND WHATSOEVER, INCLUDING, WITHOUT LIMITATION, ANY WARRANTIES OR REPRESENTATIONS AS TO THE ADEQUACY OF ANY DESIGN, MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE, OR THAT ANY OF THE SERVICES WILL MEET CUSTOMER’S NEEDS OR REQUIREMENTS OR BE CAPABLE OF IMPLEMENTATION.
IN NO EVENT SHALL MEDIVATORS BE LIABLE TO CUSTOMER OR ANY THIRD PARTY FOR LOSS OF PROFITS, LOSS OF USE, OR ANY DIRECT, INDIRECT, INCIDENTAL, CONSEQUENTIAL OR SPECIAL DAMAGES OF ANY KIND, WHETHER OR NOT A PARTY WAS, IS, OR SHOULD BE AWARE OF SUCH LOSS OR DAMAGES.
CUSTOMER AGREES THAT IT WILL RELY ON, USE AND OR OTHERWISE EMPLOY ARCHITECTS, ENGINEERS, CONSTRUCTION PROFESSIONALS AND OTHER VENDORS OR SERVICE PROVIDERS (COLLECTIVELY, “PROJECT PROFESSIONALS”) REGARDING THE FEASIBILITY OR IMPLEMENTATION OF ANYSPACE PLANNING SERVICES PROVIDED BY MEDIVATORS. By accepting the Space Planning Services, (i) Customer agrees to hold Medivators and its affiliates entirely free from any liability, including financial responsibility, for all claims, losses, damages, suits, actions, costs, liabilities, or expenses incurred by Customer or its agents or affiliates as a result of any design defect, any requirement of corrective or other action or its use, reliance on, modification or implementation of any of the Space Planning Services; (ii) Customer acknowledges that Medivators is not a licensed architect or engineer or otherwise in the business of construction or space design and Customer accepts the risks involved in using such Space Planning Services from Medivators; and (iii) Customer agrees to communicate to the Project Professionals the content of all the Space Planning Services including, without limitation, all designs and plans, specifications and attachments, and to have such Project Professionals check, review, verify and approve such content prior to incorporating, using or otherwise implementing any Space Planning Service.
10 |
ENDODRY™ STORAGE AND DRYING SYSTEM
PRE-SITE REQUIREMENTS CHECKLIST
INITIAL AND SIGN AT THE TIME OF PRE-SITE INSPECTION
Customer Site Requirements Criteria Initials
ENDODRY™ Cabinet for ADVANTAGE PLUS™ Endoscope Reprocessor provides air supply of 58-145 psi (4-10 bar).
ENDODRY Cabinet for DSD EDGE™ Automated Endoscope Reprocessor provides air supply of 72-145 psi (5-10 bar).
Air supply provides a flow rate of 4.2 scfm (120 l/min ANR) for the Advantage and 7 scfm (200 l/min ANR) for the DSD.
Air supply maximum particle size is <1 micron, maximum dew point is -40°C, maximum oil content is 0.01 mg/m3.
Air Supply plumbed to the wall a minimum of 78.7 inches (200 cm) off of the floor above the cabinet.
GFI/RCD protected outlet located on the wall a minimum of 78.7 inches (200 cm) off of the floor above the cabinet.
RJ45 wall mounted network jack located on the wall a minimum of 78.7 inches (200 cm) off of the floor above the cabinet (ENDODRY™ Cabinet for ADVANTAGE PLUS™ Endoscope Reprocessor only).
Installation site has a level floor.
Installation site accommodates the dimensions of H74.8” x W23.6” x D21.3” (190cm x 60cm x 54cm)
Customer Site
Customer Representative
Title
Signature Date Email
Serial #
Serial #
Serial #
Serial #
Serial #
Serial #
Medivators Customer and Technical Support: Toll Free: 1-800-444-4729 · FAX: 1-866-421-7696
Medivators BV Sourethweg 11 6422PC Heerlen The Netherlands Tel: +31.45.5.471.471
Cantel Medical Asia/Pacic Pte. Ltd.
1A International Business Park #05-01 Singapore 609933 Tel: +65.6227.9698 www.medivators.com
Medivators Inc.
14605 28th Avenue North Minneapolis, MN 55447 USA Toll Free: +1.800.444.4729
ESCAD Medical GmbH Zur Ohmdwiesen 5 88633 Heiligenberg Germany Tel: +49.7554.9999.500 Fax: +49.7554.9999.558
METHOD #2
METHOD #1
41 [1041]
WALL OPENING
'A'
39-3/8 [1000]
UNIT WIDTH
89-1/16 [2263]
UNIT
HEIGHT
31-1/2 [800]
LOAD HT.
4 [102] TO 8 [203]
WALL THICKNESS
90-1/2 [2299]
WALL
OPENING
5-13/16 [148]
TRIM ZONE
TYP
5-7/16 [137]
TRIM ZONE
SERVICE AREA
SEE NOTE 4
'B'
CHAMBER LENGTH UNIT LENGTH - DIM. 'A'
MINIMUM SERVICE AND LOAD AREA - DIM. 'B'
WITHOUT ALUS WITH ALUS
39" [991] 52-1/8" [1324] 78" [1981] 147-1/16" [3735]
51" [1295] 64-1/16" [1627] 102" [2591] 182-7/8" [4645]
63" [1600] 75-9/16" [1919] 126" [3200] 218-11/16" [5555]
T I T L E
EQUIPMENT DRAWING, AMSCO 600 , 26.5" X 26.5" X 39", 51", & 63" LG.
SLIDING, SD, RECESSED WALL, FACILITY STEAM HEAT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:
A - Basch, Tyler 10307698 11001835 H
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
NOTES
1. THE FLOOR AREA (IN SERVICE CLEARANCE ARES) MUST BE SMOOTH AND HAVE A LEVEL TOLERANCE WITHIN ±1/2 deg OR ±1/16" PER FOOT, WITHOUT ANY OBSTRUCTIONS SUCH AS PIPING: WIRING, FLOOR SINKS, ETC.
2. SERVICE AND LOADING AREAS CLEARANCE OF UNIT SHOULD BE AT UNIT WIDTH.
3. SERVICE AREA SHOWN ARE SOLELY REPRESENTATION OF THE SPACE NEEDED TO SAFELY PERFORM PREVENTATIVE MAINTENANCE AND REPAIRS SERVICES. CUSTOMER IS RESPONSIBLE TO ENSURE CLEARANCE MEET OR EXCEED LOCAL AND NATIONAL CODE REQUIREMENTS. ADHERENCE TO LOCAL AND NATIONAL CODES AND PROCUREMENT OF PERMITS
ARE THE RESPOSIBILITY OF THE CUSTOMER UNLESS AGREED TO IN WRITING WITH STERIS.
4. PROVIDE SERVICE ACCESS FROM BACK OF UNIT IF EQUIPPED WITH FOLLOWING ACCESSORIES:
a. 30" [760] IF EQUIPPED WITH STERI-GREEN PLUS.
b. 18" [460] IF EQUIPPED WITH STERI-GREEN.
5. ZERO CLEARANCE NEEDED OUTSIDE OF SERVICE AND LOADING AREAS.
6. ALUS EQUIPMENT DRAWING P/N: 11028378.
7. LOADING EQUIPMENT CART AND TRANSFER CARRIAGE EQUIPMENT DRAWING P/N: 11012876.
8. ALL DIMENSIONS ARE INCHES AND MILLIMETERS [MM].
TOP VIEW
NO CLEARANCE
NEEDED ON SIDES
OF STERILIZER
NO CLEARANCE
NEEDED ON REAR
EXCEPT AS NOTED
ON NOTE 4
FRONT VIEW
SERVICE ACCESS
DOOR
SIDE VIEW
FINISHED
FLOOR
FINISHED
WALL
12-7/8 [327]
21-1/4 [540]
28-15/16 [735]
34-1/16 [865]
36-13/16 [935]
3-3/4 [95]
76-3/4 [1949]
2-15/16 [75]
CW COLD WATER
S STEAM
D DRAIN
A COMPRESSED AIR
T I T L E
EQUIPMENT DRAWING, AMSCO 600 , 26.5" X 26.5" X 39", 51", & 63" LG.
SLIDING, SD, RECESSED WALL, FACILITY STEAM HEAT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:
A - Basch, Tyler 10307698 11001835 H
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
TOP VIEW
CW
S
A
D
CW-S-A
D
FRONT VIEW
CW
D
S
A
SIDE VIEW
4-1/4 [108]
5-9/16 [142]
19-3/16 [487]
87-5/8 [2226]
17-3/4 [451]
T I T L E
EQUIPMENT DRAWING, AMSCO 600 , 26.5" X 26.5" X 39", 51", & 63" LG.
SLIDING, SD, RECESSED WALL, FACILITY STEAM HEAT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:
A - Basch, Tyler 10307698 11001835 H
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
E STERILIZER ELECTRIC
N NETWORK
TOP VIEW
SIDE VIEWFRONT VIEW
E N
E N
N E
25-5/16 [644]
20-5/8 [524]
26-5/8 [676,4]
T I T L E
EQUIPMENT DRAWING, AMSCO 600 , 26.5" X 26.5" X 39", 51", & 63" LG.
SLIDING, SD, RECESSED WALL, FACILITY STEAM HEAT
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER: REVISION:
A 1/40 Basch, Tyler 10307698 11001835 H
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SHEET
JKR JACKET RELIEF VALVE
DISCHARGE PIPE - 1-1/8 ODT
CHR CHAMBER RELIEF VALVE
DISCHARGE PIPE - 1-1/8 ODT
IMPORTANT: FOR OPTIMAL INSTALLATION, COMPRESSED AIR,
WATER, STEAM, NETWORK, & POWER UTILITIES SHOULD BE
SUPPLIED BY CUSTOMER FROM ABOVE THE UNIT IN THE
LOCATION INDICATED ON THE DRAWING. CONSULT WITH
STERIS IF A DEVIATION IS REQUIRED.
TOP VIEW
(FRONT OF UNIT)
RIGHT SIDE VIEWLEFT SIDE VIEW
INCOMING MECHANICAL
UTILITIES LOCATIONINCOMING ELECTRICAL
UTILITIES LOCATION
JKR
CHR
JKR
CHR
MECHANICAL
UTILITY REQUIREMENTS CHART
UTILITY PARAMETER SPECIFICATIONS NOTES
STERILIZER
COLD WATER
(CW)
SEE NOTE 11
CONNECTION SIZE/TYPE 3/4" NPT
INSTALLATION SPECIFICATIONS:
1. PIPE SIZES LISTED UNDER PLUMBING REQUIREMENTS
INDICATE THE EQUIPMENT TERMINATION SIZES ONLY.
SIZE PIPING TO EQUIPMENT DEPENDING ON LENGTH OF
PIPE RUN FROM PRESSURE REGULATING STATION FOR
THE STEAM LINE AND MAIN WATER HEADERS. SUPPLY
THE SPECIFIED SERVICE PRESSURE AND FLOW RATE AT
EQUIPMENT TERMINALS, INCLUDE EFFECT OF COINCIDENT
DRAW OF MULTIPLE UNIT INSTALLATIONS.
2. DOOR OPENING DURING PRE-INSTALLATION SHALL BE AT
MINIMUM 80"x40" [2032mm x 1016mm] (HxW).
3. PROVIDE PIPING, SHUTOFF VALVE, PIPE PLUGGED TEE,
AND UNION IN STEAM AND WATER SUPPLY CONNECTIONS
BETWEEN EQUIPMENT AND STUB OUTS. PLUGGED TEE
CAN LATER BE USED FOR TEST PRESSURE GAUGE
CONNECTION. ARRANGE CONNECTION PIPING TO ALLOW
ACCESS TO MACHINE…