SolmeteX Warranty Solmetex Hg5®-HV Air Water Separators are warranted against defects in material and workmanship for a period of two (2) years from the date of purchase, established by proof of purchase or formal warranty registration. During the warranty period SolmeteX will at its option repair or replace products that prove to be defective.
Solmetex Hg5®-HV Collection Containers are warranted against defects in material and workmanship for a period of one (1) year from the date of purchase, established by proof of purchase or formal warranty registration. During the warranty period SolmeteX will at its option repair or replace products that prove to be defective.
Labor, transportation and service charges are not included.
Limitations of Warranty The warranty shall not apply to defects resulting from improper installation, maintenance, abuse, unauthorized modification, or operation outside of the environmental specifications for the product or damages that occur due to improper repackaging of equipment for return to SolmeteX.
USE OF LINE CLEANERS HAVING A pH LESS THAN 6 OR GREATER THAN 10 WILL VOID THESE WARRANTIES.
For a complete list of recommended cleaners, visit our web site www.solmetex.com.
No other warranty is expressed or implied. SolmeteX specifically disclaims the implied warranties of merchantability and fitness for particular purpose.
Exclusive Remedies The remedies provided herein are the buyer’s sole and exclusive remedies. SolmeteX shall not be liable for any direct, indirect, special, incidental or consequential damages, whether based on contract, tort or any other legal theory.
Hg5®-HV Manual – Rev – April 27, 2012 Page 1 of 7
SolmeteX Hg5®-HV Amalgam Separator Type 2 – Maximum Flow 1000 ml/min
Installation and Maintenance Instructions
Table of Contents Warnings and Warranty ......................Page 1 Installation ....................................Page 2 Installation Diagram ..........................Page 3 Collection Container Replacement ............Page 4 Maintenance (Vacuum Line Cleaners) ........Page 6 Parts ...........................................Page 6 Trouble Shooting ..............................Page 7 Thoroughly read and understand instructions prior to installing, operating and servicing the Hg5®-HV Amalgam Separator.
The waste stream treated by the SolmeteX Hg5®-HV is generated by a dental vacuum system and as such may contain concentrations of solid and soluble mercury and silver. Because of this, any spills should be considered hazardous and should be handled in accordance with standard hazardous materials (HAZMAT) handling procedures.
Full Collection Containers are a HAZMAT and should be handled, stored and disposed of according to regulations applying to hazardous waste containing mercury. Always wear protective gear when handling full SolmeteX Hg5®-HV collection containers (latex gloves, safety glasses or face shield) and dispose of per local regulations and codes.
50 Bearfoot Road Northborough, MA 01532 www.solmetex.com
Instructions When choosing the location to mount the Hg5®-HV, remember that you must leave room to:
Complete the piping.
Access the front and sides of the unit for collection container replacement.
Have the unit mounted higher than the pump and drain.
The wall mounted support bracket must be mounted to the wall prior to installation of the Hg5®-HV unit.
The support bracket should be stud mounted in order to support the filled weight of approximately 100 pounds.
The inlet and outlet connections are 3” Schedule 40 PVC Pipe. It is recommended that flexible pipe couplings be used to connect the unit to the existing vacuum lines. Since pipe sizes can vary from site to site, bushings may be required to adapt the unit to the vacuum line piping at your facility.
Remove all components from package(s) make sure you have everything, prior to proceeding.
The unit is piped into the vacuum line between the treatment rooms and the vacuum source. The line from the treatment rooms is piped to the inlet of the Hg5®-HV and the outlet is piped to the vacuum source. DO NOT connect the unit to the outlet of the separating tank (dry vacuum system) or water ring pump (wet vacuum system).
Determine the location of unit with respect to existing plumbing; pump location, drain and ease of collection container change out.
Mount the back plate to an existing wall using lag bolts. The structure and bolts should be capable of supporting 100 pounds. Mount directly into the studs using the pre-drilled holes in the backplate. If no studs are available, molly bolts may be used providing they and the wall are capable of supporting the load.
After mounting the bracket, lower the Hg5®-HV onto the bracket. It will sit between the two side supports. Use the four ¼-20 bolts supplied with the unit to secure the unit to the bracket.
After the unit is installed, the two collection containers can be installed. (See included installation instructions).
Write the installation date on each collection container.
Complete the registration form and return to SolmeteX.
Environmental Specifications Overall Dimensions: 17.625” W x 28” H x 22.5” D Minimize water lift height.
Maximum Temperature = 120o F (52o C).
Maximum Vacuum = 15” Hg (51 kPa) For Dental Use ONLY
Hg5®-HV Manual – Rev – April 27, 2012 Page 2 of 7
Hg5®-HV Installation
Outlet Outlet
Inlet
Front View
Air Water Separator
Wall Mount Bracket
Collection Containers
28”
17 5/8”
Inlet
28”
22.5”
14”
Side View
Hg5®-HV Installation Diagram
Hg5®-HV Manual – Rev – April 27, 2012 Page 3 of 7
DO NOT connect Hg5-HV to outlet of tank.
Treatment
Rooms
Air Water Separating Tank
Tank Inlet Any Dry
Vacuum Unit
Any Water Ring Pump
DO NOT connect Hg5-HV to outlet of the water ring pump.
Pump Inlet
OR
Hg5-HV Inlet
Hg5-HV Outlet
Warning The waste stream treated by the Hg5®-HV is discharge from a dental vacuum system and as such may contain concentrations of solid and soluble mercury and silver. Because of this any spills should be considered hazardous and should be handled in accordance with standard hazardous materials (HAZMAT) handling procedures. Full Collection Containers are a HAZMAT and should be handled, stored and disposed of according to regulations applying to hazardous waste containing mercury. Always wear protective gear when handling full Hg5® collection containers (latex gloves, safety glasses or face shield) and dispose of per local regulations and codes.
NOTE: SolmeteX offers a recycle kit for the safe disposal/recycle of full collection containers. Please contact your dental distributor to purchase an Hg5® Recycle Kit for the safe disposal of your full collection container.
2. SAVE THE SHIPPING CARTON FOR STORAGE OF YOUR FULL
COLLECTION CONTAINER!
Contents:
a. One (1) Hg5® Instruction Sheet.
b. One (1) Hg5® Collection Container.
c. One (1) “zip lock” style Disposal Bag.
Hg5®-HV Manual – Rev – April 27, 2012 Page 4 of 7
Retaining Pins (2)
Retaining Pins (2)
Before you begin, observe the upper chamber of your Hg5®-HV Amalgam Separator. If liquid is present in the upper chamber, wait until the liquid has drained down into the collection container before proceeding to replace the collection container.
THIS PROCEDURE SHOULD BE PERFORMED ONLY
WHEN NO PATIENTS ARE BEING WORKED ON AS
LOSS OF CHAIRSIDE VACUUM MAY OCCUR.
LEAVE THE VACUUM SYSTEM RUNNING DURING
THE ENTIRE COLLECTION CONTAINER CHANGE
OUT PROCEDURE!
1. Sanitize the vacuum system using the sanitization procedure established at your facility. It is good practice to leave the vacuum pump running for 5 to 10 minutes every time the system is sanitized in order to allow excess liquid time to evacuate the lines.
Upper Chamber
Collection Containers
Hg5®-HV Collection Container Replacement
Hg5®-HV Manual – Rev – April 27, 2012 Page 5 of 7
4. With the vacuum system still running, Install your new collection container:
a. The collection container can be installed only one way. Note the alignment of the fittings on the collection container with the receiver holes in the upper unit.
b. Gently insert the fittings into the receiver holes. This can be accomplished by rocking the collection container slightly. The O- Rings on the fittings should seat easily into the upper unit receivers.
c. Continue to push the collection container into the upper unit until it is fully seated.
d. Insert the retaining pins into their holes until seated. The collection container is now locked into place.
5. Fill out the information label:
a. Enter the “Date Installed” on the new collection container.
b. Enter the “Next Change Due” on the new collection container.
NOTE: YOUR COLLECTION CONTAINER MUST BE CHANGED
ONCE EVERY TWELVE (12) MONTHS OR WHEN THE SEDIMENT
LEVEL REACHES THE FULL LINE, WHICHEVER OCCURS FIRST.
CHECK THE SEDIMENT LEVEL OF THE
COLLECTION CONTAINER WEEKLY.
6. Note the vacuum gauge reading. The reading should be the same as before the collection container change.
Hg5®-HV Collection Container Replacement
3. With the vacuum system still running, remove the full collection container from the Hg5®-HV Unit as follows:
a. Open the slider lock bag (supplied) and slip it over the collection container to be removed.
b. Support the collection container with one hand, removing the retaining pins with the other hand. Tilt the collection container to release the element side receiver first (NOTE: The slotted pipe side of the collection containers faces the back side of the upper chamber) and then pull down to release the entire collection container.
WARNING
Hold bag securely in the unlikely event that the upper chamber may contain some residual liquid that may drain during removal of the collection container.
c. Remove the plastic caps from the inlet and outlet fittings of the new collection container.
d. Place these caps into the inlet and outlet fittings of the full collection container.
e. Squeeze out all excess air and zip the disposal bag closed.
MAINTENANCE
Line Rinsing & Vacuum Line Cleaners Cleaners should have these qualities: Non-Foaming, De-Odorizing, Sanitizing & pH between 6 and 10.
USE OF LINE CLEANERS OUTSIDE OF THIS pH RANGE WILL VOID THE WARRANTY.
For a complete list of recommended cleaners, visit our web site www.solmetex.com.
WITH OTHER DENTAL VACUUM SYSTEMS: Follow manufacturers instructions.
Note: Plan to rinse no more than 5 rooms every 10 minutes. Limit the total maximum flow to the Hg5®-HV during rinsing to 1 liter per minute. Rinsing too much or too rapidly can overfill the Hg5®-HV and could affect the unit’s efficiency.
Collection Container Replacement - Follow instruction packed with replacement collection container.
CHECK THE SEDIMENT LEVEL OF THE COLLECTION CONTAINER WEEKLY.
PARTS
Description SolmeteX Part Number
Hg5-HV Mercury Removal Unit HG5-HV-001
Hg5-HV Collection Container & Recycle Kit HG5-002CR (2 required)
Hg5-HV Recycle Kit HG5-002R
Hg5-HV Wall Mount Bracket HG5-003
Hg5-HV Retaining Pins HG5-009 (4 required)
Hg5-HV Air/Water Separator w/Manifold HG5-HV-AWSEP
Hg5®-HV Maintenance & Parts
Hg5®-HV Manual – Rev – April 27, 2012 Page 6 of 7
Hg5®-HV Manual – April 27, 2012 Page 7 of 7
Hg5®-HV Trouble Shooting
Trouble Shooting
The Hg5®-HV is designed to provide years of trouble-free service, with minimal attention. In the unlikely event of system related problems, please consult the troubleshooting and maintenance chart below.
Problem: Little or No vacuum at the hand piece
1. Check sediment level of collection container, if full change the collection container.
2. If the vacuum gauge reads normally but there is little or no vacuum to the hand piece, there is probably a clog or a leak between the hand piece and the Hg5®-HV.
3. If the vacuum gauge reads lower than normal these are the possible causes:
Did you just replace the collection container?
Yes:
Check installation of the new collection container. If the vacuum is low there is a possibility that the o-rings on the collection container did not seal into the receivers. Remove the collection container, check the o-rings and re-install per the collection container installation procedure.
No If you have a solid collection cup (installed by others) - Check & Clean or replace element/screen if necessary.
Check operation of the vacuum pump.
b. Listen for vacuum leaks.
c. Check all connections for breaks or cracks.
Check all flexible hose for kinks, breaks, or loose clamps.
Problem: Water in the Upper Chamber (Air Water Separator)
1. Check sediment level of collection container, if full change collection container.
2. Check the pH of vacuum line cleaner, if below 6 or above 10 change collection container and line cleaner. Visit www.solmetex.com for a recommended list of cleaners.
3. The transition from the Upper Chamber to the Collection Container could be clogged and require cleaning.
Procedure:
a. Turn off vacuum to the system.
b. Remove the rubber coupling at the top of the air/water separator.
c. Prepare a solution of line cleaner per manufacturer’s specification and pour into the air/water separator.
d. Replace the rubber coupling at the top of the air/water separator.
e. Turn the vacuum on, and check the system for proper operation.
f. Remove the collection container (with the vacuum on).
g. Re-Insert the collection container and check to see that water exits the upper chamber.
3. Call your dental dealer.
a. If the problem cannot be solved easily, call or e-mail your dealer for support.
Technical Bulletin: DEN-001Y Revised April 27, 2012
A Division of Layne Christensen
Vacuum Line Cleaning Solution pH and Hg5® Performance
According to the American Dental Association (ADA) and regulators requiring Amalgam Separators, oxidizing line cleaners should not be used. Oxidizing cleaners can breakdown amalgam particles. Cleaners containing bleach (chlorine bleach, sodium hypochlorite) should never be used as chlorine is known to cause mercury to be released from amalgam.
Regulators have also become aware of pH related to line cleaners. Sewage treatment plants require discharges from their users to be between 5.5 and 10 on a pH scale. Some new amalgam separator regulations now require the use of line cleaners with neutral pH. Acidic (low pH) line cleaners have been known to hinder the operation of Amalgam Separators. Basic (high pH) cleaners can cause premature wear on materials used in amalgam separators.
Recommendations:
SolmeteX recommends the use of vacuum line cleaners that have a pH between 6 and 10. There are several cleaners with a pH near neutral (pH 7). Most of these rely on enzymatic processes. Below is a list of readily available cleaners that have pH ranges most compatible with the Hg5®. If the line cleaner you are currently using is not on the list feel free to contact SolmeteX for further information.
NAME MANUFACTURER pH (a, b)
PureVac SC Sultan 9.4 Recommended
BioVac Micrylium 8.5 Approved
Bio-Pure eVacuation System Cleaner Bio-Pure Products 6.6 – 7.4 Approved
BLAST Preferred Dist., Inc. 6 – 7 Approved
Citrizyme Pascal Company 7.7 Approved
EmPower Metrex 6.5 – 8.6 Approved
Enzymax Hu-Friedy 8.5 - 9 Approved
FRESH-VAC Ecolab, Inc/Huntington 6.5 – 8 Approved
Hayes Evacuation Cleaner Hayes Hendpiece 7.5 Approved
Medical Enzyme Detergent Enzyme Solutions 6.5 – 7 Approved
Microvac Microbex 9 – 10 Approved
Neutravac Biotrol 7 – 8 Approved
Precision Clense Plus Midmark 9 – 10 Approved
ProEZ Cetrol Internation 7.4 – 8.3 Approved
Quala Enzyme Industries 7.3 – 9.3 Approved
Sani-Soak Ultra Enzyme Industries 8.5 Approved
SlugBuster RAMVAC 8 – 8.5 Approved
SRG Evacuation Line Cleaner Mydent International 6.4 Approved
Super Vac 40 Plus EPR Industries 7 – 7.5 Approved
VacuCleanse Evacuation Infection Control Tech 7.1 Approved
a) The pH data is taken from manufacturer MSDS sheets.
b) The Massachusetts Department of Environmental Protection requires the use of line cleaners with pH between 6.5 and 9.
Technical Bulletin – DEN-001Y April 27, 2012