FA520521R00030002_Atch5_Revised Cut Sheets for JSN S0442 and S0960.pdf

PDF 4 MB Posted

Attached to
Project No. 1090686, Hospital Modernization Misawa Air Base, Japan
Solicitation number
FA520521R0003
Issued by

About this file

This document package includes cut sheets summarizing specifications for two medical equipment items, the Amsco 7052HP Single-Chamber Washer/Disinfector and related accessories. The cut sheets provide detailed technical specifications, dimensions, utility and electrical requirements, noise levels, heat output, and other performance characteristics for the washer/disinfector and accessories. Also included is information on a related federal contract opportunity for Project No. 1090686 to modernize the hospital at Misawa Air Base in Japan, which may involve procurement of the washer/disinfector and accessories described in the cut sheets.

View the file

Other files

File Type Posted
FA520521R00030004.pdf PDF
FA520521R00030004_Atch3_Revised Specifications.pdf PDF
FA520521R00030004_Atch1_Consolidated List of Govt Response.pdf PDF
FA520521R00030004_Atch2_Addendum No 2.pdf PDF
FA520521R00030004_Atch5_Revised Soliciation Page 1 and 21.pdf PDF
FA520521R00030004_Atch4_Revised Drawing Sheets.pdf PDF
FA520521R00030003_Atch2_Revised Soliciation Page 1 and 21.pdf PDF
FA520521R00030003.pdf PDF
FA520521R00030002_Atch2_Revised Specifications.pdf PDF
FA520521R00030002_Atch7_Revised Soliciation Page 1 and 21.pdf PDF
FA520521R00030002_Atch6_Consolidated List of Govt Response.pdf PDF
FA520521R00030002_Atch3_Revised Drawing Sheets.pdf PDF
FA520521R00030002.pdf PDF
FA520521R00030002_Atch1_Addendum No 1.pdf PDF
FA520521R00030002_Atch4_Revised Submittal Register.pdf PDF
FA520521R00030001.pdf PDF
Atch2_Drawings_PDF3of10_Structural.pdf PDF
Atch2_Drawings_PDF8of10_Plumbing.pdf PDF
Atch7-2_Structural Interior Design Volume 1 thru 3.pdf PDF
Atch2_Drawings_PDF6of10_Equipment.pdf PDF
RFP FA520521R0003_Hospital Modernization.pdf PDF
Atch2_Drawings_PDF4of10_Architectural.pdf PDF
Atch6_List of Contractor Furnished US Material.pdf PDF
Atch7-3_Mobile Sterile Processing Department Trailer Information.pdf PDF
Atch9_Past Performance Questionnaire Cover Letter.pdf PDF
Atch1_Specifications_Volume 2 of 2.pdf PDF
Atch2_Drawings_PDF10of10_Electrical.pdf PDF
Atch2_Drawings_PDF7of10_FireProtection.pdf PDF
Atch4_Construction Timeline.pdf PDF
Atch1_Specifications_Volume 1 of 2.pdf PDF
Atch2_Drawings_PDF2of10_Civil.pdf PDF
Atch2_Drawings_PDF1of10_General.pdf PDF
Atch8_Hazardous Material Survey Report.pdf PDF
Atch2_Drawings_PDF9of10_Mechanical.pdf PDF
Atch7-1_Equipment Data and Cut Sheets Logistics Category A.pdf PDF
Atch3_Other Contract Requirements.pdf PDF
Atch2_Drawings_PDF5of10_Interiors.pdf PDF
Atch5_Submittal Register.pdf PDF
Atch10_Past Performance Questionnaire.docx DOCX document
Show all 39

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Attachment 5 to

FA520521R00030002

Date: 15 Apr 21 jpthurman Text Box Manufacturer to provide all accessories to properly seal between the unit and wall for proper room pressure requirements to be achieved.

Amsco 7052HP single-chamber washer/disinfector

AMSCO® 7052HP SINGLE-CHAMBER

WASHER/DISINFECTOR

10283376 Rev E (10/27/20)

Item ________________________ Location(s)___________________

(Typical - details may vary.)

STANDARDS

The AMSCO 7052HP Single-Chamber Washer/Disinfector complies with the following standards:

As certified by Intertek:

• CAN/CSA-C22.2 No. 61010-1, 3rd Edition (R2017)

• CAN/CSA-C22.2 No. 61010-2-040 2nd Edition Dated 07/01/2016

• UL 61010-1, 3rd Edition (2012-05-11)

The Selections Checked Below Apply To This Equipment

POWER*

Steam-Heated Unit Electric-Heated Unit

VOLTAGE*

200 to 208 V, 60 Hz (Steam) 200 to 208 V, 50 Hz (Steam) 380 to 400 V, 60 Hz 380 to 415 V, 50 Hz 460 to 480 V, 60 Hz

LANGUAGE PACKAGE

French (FD355) Spanish (FD356) German (FD357) Italian (FD358) Japanese (FD383) Polish (FD384)

* Careful consideration must given to voltage selection prior to ordering. Later changes require substantial field modification.

† Refer to SD867 for Material Handling Accessories.

Refer to 10726323 for AMSCO® Top Enclosure Panels Accessories.

Refer to SD857 for SCS Conveyor System Accessories.

Refer to 10370003 for ATLAS WAV System Accessories.

ACCESSORIES†

Impact Printer (FD094) Thermal Printer (FD012) Seismic Tie-Down Kit (FD354) Condensate Return To Drain (FD353) Two-Level Manifold Rack (FD74-800) Two-Level Manifold Rack with Equal Space

(FD79-100)

Three-Level Manifold Rack (FD74-900) Four-Level Single-Chamber Manifold Rack

(FD75-000)

Four-Level Manifold Rack (FD75-100) Five-Level Manifold Rack (FD75-200) Multi-Function Rack for Large Items (FD75-500) Multi-Function Rack for Small Items (FD75-600) General Purpose Basket (FD77-100) RAS 12 Rack (FD256) Self Docking Transfer Cart (FD199) Transfer Cart Adapter (FD209) Set of Dry Contacts (FD362) Handheld Bar Code Reader (FD363) Conductivity Probe (FD350) Additional Wall Flange (FD351)

ACCESSORIES (CONT’D)†

Installation Kit – Steam (FD088) Installation Kit – Electric (FD037) Additional Chemical Pump (FD349) Offset Installation Air Management Kit (FD352) IQ-OQ-PQ Protocols (FD359) Machine Identifier (11000866) Drain Acu-CoolDown (FD361) ConnectAssure Technology for Data Export

(CAT3030)

ProConnect® Technical Support Services (Remote

Monitoring, Priority Technical Support, Customer Care Center Access, Equipment Performance Reports).

Available in U.S. and Canada only. (GP09-167)

OPTIONS

Acu-Rinse Reservoir (FD000054) Acu-Wash Reservoir (FD000055) 40-Amp Min Circuit Protection (FD000056) RAS Cycle (FD000057)

APPLICATION

The AMSCO 7052HP Single-Chamber Washer/ Disinfector with touch screen control is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps), utensils (such as bedpans and urinals, trays, bowl, basin, kidney dishes), rubber and plastic goods, theatre shoes and other similar and related articles found in healthcare facilities.

DESCRIPTION

The AMSCO 7052HP Single-Chamber Washer/ Disinfector is a cabinet-type mechanical washer that includes:

• Color touch screen control system and visual progress indicator

• Audible warning system with adjustable volume

• Space saving, power vertical sliding doors

• Eight factory-loaded (Instruments, Orthopedic Instruments, Utensils, Plastic Goods, Rigid M.I.S., Anesthesia / Respiratory Goods, Gentle and Decontamination) and up to 20 Custom cycles. Each cycle is effective, fast, energy efficient and may be customized (within allowed range of parameters) to meet specific operating requirements

• LED status indicator lighting

• Compact space saving footprint

• Conforms to seismic building code requirements

• Patented spray arm system providing entire chamber coverage

• 180 degree reverse installable

Provide seal around perimeter of all plate, wall, ceiling openings and at hoods to ceiling.

S0960

Governing Directive for the affixing of the CE mark:

• Medical Devices Directive 93/42/EEC as amended by 2007/

47/EC

• Machinery Directive 2006/42/EC

Standards Applied to Demonstrate Conformity to the Directive:

• IEC 61010-1 3rd Edition (2012-05-11), EN 61326-1, 2013, IEC 61326-1 Second Edition (July 2012) and IEC 61010-2- 040:2015

• EN ISO 15883-1:2009+A1:2014, ANSI/AAMI ST15583-1:2009

and CAN/CSA-Z15883-1:09 (R2019) Washer-Disinfectors – General Requirements, Definitions and Tests

• EN ISO 15883-2:2009, ANSI/AAMI ST15583-2:2013 and CAN/CSA-Z15883-2-09 (R2019) Requirements and Tests for Washer-Disinfectors Employing Thermal Disinfection for Surgical Instruments, Anesthesia Equipment, Holloware, Utensils, Glassware, etc.

SIZE (W X H X L)

Exterior dimensions:

• 42 x 80-3/4 x 32" (1067 x 2051 x 813 mm)

Interior chamber dimensions:

• 29-1/2 x 29-1/4 x 26-1/2" (749 x 743 x 673 mm)

Load height:

• 31" (787 mm) above finished floor

FEATURES

Vertical sliding power door:

• Door is constructed of double pane tempered glass to allow operator to view chamber interior with door closed.

• Color LED light illuminates top portion of each door to indicate machine status.

• Door is pneumatically activated using touch screens located on the control panel (one each side of unit) and is equipped with a built-in safety system.

• Each door is mounted on a compressed seal that reduces heat loss and increases heating capability.

• If a power failure occurs, door can be opened manually. A door interlock feature is provided to prevent cross-contamination.

• Unit can be configured in single- or double-door operation.

Stainless-steel pump is powered by a dual-speed motor. High pump speed provides equivalent capacity of a 14 hp (10.4 kW) motor, 165 U.S. gal/min at 196 ft (625 L/min at 60 m) head pressure.

Low pump speed provides equivalent capacity of a 3.5 HP (2.6 kW) motor, 90 U.S. gal/min at 46 ft (341 L/min at 14 m) head pressure. Pump impeller is mounted directly on motor shaft and does not require additional bearings. Pump motor is equipped with a drip-proof frame, magnetic starter, overload protection and sealed bearings (not requiring periodic lubrication).

Pump, spray system and all recirculating piping are of stainless-steel construction.

Wash chamber:

• Constructed of 16-gauge, 316L stainless steel (No. 4 finish), argon-welded and polished. Chamber is of sanitary-type design (horizontal fixed surfaces are sloped, without overlapping sheet metal) for complete drainability and reducing hard-to-clean locations.

• Single-walled, insulated construction of chamber exterior reduces heat loss and noise level to the work area.

• Rotary spray assemblies are positioned (one at top and one at bottom of chamber, each measuring 24-3/8" [620 mm] long) to reach all load surfaces. Spray assemblies are easily removable without tools to aid in cleaning and maintenance.

Sump:

• Constructed of 316L stainless steel with 7.2 gal (27 L) capacity with removable stainless-steel filter preventing debris from entering pump and piping system.

• Heating coil (steam or electric) at the bottom of the wash chamber (sump) raises and maintains water temperature up to 185°F (85°C) during the Wash phase and up to 194°F (90°C) during the Thermal Rinse phase.

Chemical peristaltic injection pumps:

• Pre-Programmed cycles and pumps are optimized to use Prolystica® Ultra Concentrate (PUC) HP chemicals. Each ultra concentrated product is 10 times the concentration of a traditional product; so,10 times less chemical is injected per cycle.

• Unit comes with one enzyme and one detergent injection peristaltic pump which automatically add a selected quantity of chemical from 1/40 to 2 oz/gal (0.2 to 16 mL/L).

• One pump is dedicated to automatically add a selected quantity of lubricant during Thermal Rinse phase.

• Pumps are positioned near chemical containers. A low-level sensor is included to indicate when detergent level is low, or when insufficient chemical is available for next cycle.

• 100’ (30 m) extension tubes are included to pump chemicals from canisters installed in a remote location to unit.

• Control monitors volume of chemicals injected and indicates if this parameter meets specified criteria during all specific phases.

Top Utility Connections (except drain connections) facilitate installation. All utilities (including vent, steam, electric, water and compressed air) are connected on washer top. Top water injection gives a secure air gap of 31-1/2" (800 mm). Backflow preventers are not required for hot and pure water.

Front Service Panels provide easy access to piping, valves, electrical components and wiring. Unit can be configured as either a Vented System that exhausts chamber vapors to building exhaust system through 3.0" (76 mm) OD vent connection located at top of washer or as a Non-Vented Drying System where chamber vapors are exhausted through cold water condenser to facility room.

Drying System uses uniquely designed four-sided inflow drying pattern to produce high-flow air curtain. Air curtain provides broad, efficient, drying within wash chamber. Recycled and non-recycled air is manifolded and circulated through piping and accessory providing energy efficient system while ensuring complete chamber air coverage. Fresh air is drawn through a HEPA filter. Drying system includes a 3 hp (2.2 kW) blower, to remove vapor from chamber prior to opening doors, and three electric heaters totaling 11.4 kW to heat and maintain chamber air temperature. When Acu-Drying is selected, drying time self-adjusts to meet drying time of loaded rack and items.

Pure Water Stainless-Steel Supply Valve is supplied on washer/disinfector so Pure Water Rinse Supply may be used with factory programmed cycles and optional Acu-Rinse Reservoir.

Drain Discharge Cool Down ensures water drained at end of each phase, from chamber sump to building drain system, does not exceed 140°F (60°C). If water temperature in sump is higher than 140°F (60°C), cold water is automatically added to reduce temperature of water discharged into building drain system.

Sampling Valve is supplied in washer/disinfector recirculation piping for North American Customers that are not able to install sampling valves on facility supply lines near washer/disinfector.

Barrier Wall Kit Flanges for one wall (clean side) installation come standard. An optional kit to enclose second wall (dirty side) is also available (FD351).

Cycle Graph enables monitoring of injection volume, water circulating pressure and water temperature during a cycle.

Graphs are visible during cycle and can also be displayed at cycle completion. Data can be exported using ConnectAssure Technology or USB port.

CONTROL SYSTEM

The AMSCO 7052HP Single-Chamber Washer/Disinfector is equipped with a touch screen Customer interface. This 8.4" (213 mm) color touch screen is mounted at eye level beside chamber door (both sides of unit) and tilted for better visibility, allowing easy monitoring of all wash cycles. Control system monitors and controls all phases of each programmed cycle.

The control system features:

• Cycle selection is simplified through use of Icons representing devices and cycle confirmation (on-screen) to ensure correct cycle is selected.

• Service mode for preventive maintenance testing and to facilitate troubleshooting.

• Built-in service diagnostic program to permit system calibration and verification of component operations.

• Security lock-out feature that enables cycles and temperatures to be locked and unchangeable without proper access code.

• Cycle data is stored as a protection against power disruption. Data may be downloaded from controller using supplied USB ports.

• Permits operator to monitor current washer/disinfector status (including current chamber temperature, circulating pressure and time remaining in phase).

• Indicates any abnormal conditions with audible and visual alarm system.

CYCLE DESCRIPTION

ADVISORY NOTE: STERIS does not intend, recommend or represent in any way that this AMSCO 7052HP Single-Chamber Washer/Disinfector be used for the terminal disinfection or sterilization of any regulated medical device. AMSCO 7052HP Washer/Disinfector is intended only to perform an initial step in the processing of soiled, reusable medical devices. If medical devices contact blood or compromised tissues, such devices must be terminally processed in accordance with good hospital practices before each use in human patients.

Once treatment cycle is selected, washer/disinfector automatically processes load through standard phases (additional phases are included in certain treatment cycles depending on unit configuration) as described on Cycle Charts and in supplied Operator Manual.

AMSCO 7052HP Single-Chamber Washer/Disinfector enables Customers to use Dual Wash phase combining PUC HP enzyme and PUC HP detergent in the same filling which accelerates cycle time and reduces water consumption.

Custom cycles are available but must be validated by the Customer (not STERIS responsibility).

SAFETY FEATURES

Vertical chamber doors are equipped with an obstruction sensor to detect any door obstruction. If obstruction is present, door automatically opens.

The washer/disinfector is equipped with a safety lockout feature so a cycle cannot start unless the door is fully closed.

If the door is opened during a cycle, all utility services to chamber are shut off, and the cycle stops.

Door interlock feature is provided to prevent cross-contamination. Door interlock feature allows only one door to be opened at a time whenever power is on. When the cycle is in process, door interlock prevents either door from being opened. Access to the load is then restricted.

Safety stop pushbutton(s), one on the load and one on the unload side, automatically stop all unit operation when pressed.

Building electrical supply disconnect switch must be used to shut off power to the unit before servicing.

ACCESSORIES AND OPTIONS

Printer (FD094 or FD012), if provided, produces an easy-to-read printed record of whether load was properly processed at preset temperature, as well as a complete list of alarm and abort in-cycle messages. A paper out LED is included.

Acu-Rinse Reservoir (if equipped) pre-heats Pure Water for Rinse and Thermal Rinse phases significantly reducing cycle times. Constructed of 316L stainless steel.

Acu-Wash Reservoir recovers a portion of water used in Thermal Rinse treatment and stores it for use in Wash treatment of next cycle reducing overall cycle water consumption.

RAS Cycle (Robotic Assisted Surgery Cycle) (FD000057 or FD258) is used with RAS 12 Rack (FD256) for effective cleaning, rinsing, intermediate level disinfection and drying of reusable da Vinci X/Xi and S/Si EndoWrist instruments.

da Vinci® and EndoWrist® are registered trademarks of Intuitive Surgical, Inc.

For Further Information, contact:

STERIS Corporation 5960 Heisley Road Mentor, OH 44060-1834 • USA 440-354-2600 • 800-548-4873 www.steris.com

10283376 ©2020, STERIS Corporation. All rights reserved. Rev E (10/27/20)

This document is intended for the exclusive use of STERIS Customers, including architects or designers. Reproduction in whole or in part by any party other than a Customer is prohibited.

The base language of this document is ENGLISH. Any translations must be made from the base language document.

CUSTOMER IS RESPONSIBLE FOR COMPLIANCE WITH

APPLICABLE LOCAL AND NATIONAL CODES AND

REGULATIONS.

Steam Condensate Return To Drain (FD353) allows for connection of steam condensate return outlet to drain when steam condensate return line is not available in building. Cold water is also injected in drain piping when condensate return water temperature is too high. Condensate return cool down keeps temperature in drain piping below 140°F (60°C).

Seismic Tie-Down Kit (FD354) includes a seismic report for proper installing and securing of washer to building floor. Unit is designed to comply with seismic building code.

Additional Chemical Pump (FD349; up to two pumps) is available, giving flexibility to wash with neutral process, alkaline process, or to vary chemical used, depending on load type.

Drain Acu-CoolDown (FD361) precisely injects correct amount of potable water into facility drain system, minimizing volume of cold water needed.

Offset Installation Air Management Kit (FD352) controls airflow at top of door assemblies.

ConnectAssure Data Export (CAT3030) exports XML cycle records from equipment to Customer tracking system.

ProConnect® Technical Support Services maximizes operational efficiencies with secure, Internet-based, real-time equipment monitoring. Data from Customer equipment is used by STERIS to provide pro-active Customer alert notifications, technical support and predictive maintenance. (ProConnect Technical Support Services is available in U.S. and Canada only.) Refer to Tech Data sheet SD983, PROCONNECT TECHNICAL SUPPORT SERVICES, for details.

INSTALLATION

The washer/disinfector is designed as a fully enclosed cabinet for freestanding or recessed installation. Clearance between top of the unit and the ceiling must be at least 14" (356 mm).

If the system is recessed through one or two barrier walls, stainless-steel barrier flanges are included to provide a finished wall appearance.

Once installed, system is designed to allow for easy access for maintenance purposes.

All configuration modifications are available upon request.

Contact STERIS to receive a quotation.

PREVENTIVE MAINTENANCE

Customers are encouraged to contact STERIS concerning our annual maintenance program. Under the terms of the program, preventive maintenance, adjustments and replacement of worn parts are provided on a scheduled basis to help ensure optimal equipment performance and help minimize untimely or costly schedule interruptions. STERIS maintains a worldwide staff of well-equipped, factory-trained technicians to provide these services, as well as on-site installation, training and expert repair services. Contact STERIS for details.

NOTES

1. Machine segments (unit is bolted onto a wooden skid) are shipped in one cardboard carton (W x H x L): 50 x 91 x 42" (1270 x 2311 x 1067 mm).

2. Customer must ensure the washer/disinfector stands on a noncombustible floor (floor should be level).

3. Customer must provide utility connections with shutoff disconnects within 2’ (0.6 m) of equipment perimeter and below ceiling deck or purchase Installation Kit from STERIS.

4. STERIS recommends vacuum breakers (not provided by STERIS) be installed on service lines, and disconnect switches (with lockout in OFF position; not provided by STERIS) be installed in electric supply lines near equipment.

5. For all ventilation ducting from the washer/disinfector, STERIS recommends installation of a dedicated, corrosion-proof, flexible watertight duct (3" [76 mm] OD) to the exterior of the building, sloped toward the washer/disinfector.

6. Minimum ceiling height for door removal is 94" (2388 mm).

7. STERIS recommends illumination of the service area along with providing a convenience outlet for maintenance.

8. Noise level of 60.6 dB; Operating weight of 1543 lb (700 kg).

9. Unit is shipped with an Operator Manual. Maintenance Manual is available for purchase.

10. ConnectAssure Data Export requires an Ethernet drop and Ethernet connection within 10 ft of the equipment. The Customer must supply the network cable.

UTILITY REQUIREMENTS

IMPORTANT: Refer to proper equipment drawing for installation details and specifications:

• Electric-Heated Unit: 10292222, 10292223, 10292224 or

10292225

• Steam-Heated Unit: 10292218, 10292219, 10292220 or 10292221

80-3/4 [2051]

UNIT

HEIGHT

31 [787]

LOADING

HEIGHT

26-1/4 [667]

LOADING

CLEARANCE

HEIGHT

8-1/2 [216]

[381]

91 [2311]

RAISED DOOR

HEIGHT

34 [864]

MAX LENGTH

32 [813]

UNIT LENGTH

90 [2286]

SERVICE AREA

HEIGHT

45 [1143]

WALL OPENING

*(NOTE F)

1-1/2 [38]

MINIMUM

94 [2388] *(NOTE G)

RECOMMENDED

98 [2489] *(NOTE H)

MAX 1 [25]

ADJUSTMENT WITH

WASHER LEGS

(UNIT HEIGHT 81-3/4 [2077] MAX)

MINIMUM

94 [2388] *(NOTE G)

RECOMMENDED

96 [2438] *(NOTE H)

42 [1067]

UNIT WIDTH

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

SERVICE AREA

REQUIRED FOR

DRYING MOTOR

REMOVAL

TYPICAL INSTALLATION

NOTES:

(A) UNLESS OTHERWISE SPECIFIED, ALL DIMENSIONS ARE IN INCHES [mm].

(B) SERVICE ACCESS SIDE OF UNIT IS RECOGNIZED BY THE TOUCH SCREEN AT THE RIGHT OF

WASH CHAMBER. ELECTRICAL AND CONTROL COMPONENTS, AS WELL AS MOST OF

SERVICABLE PARTS ARE ACCESSIBLE FROM THAT SIDE OF UNIT. NON-SERVICE SIDE IS

EQUIPPED WITH THE TOUCH SCREEN ON THE LEFT OF THE WASH CHAMBER.

(C) CUSTOMER MUST BE SURE THAT MACHINE STANDS ON A NON-COMBUSTIBLE FLOOR.

(D) WEIGHTED EQUIVALENT SURFACE SOUND PRESSURE LEVEL: 64.0 dB(A) CALCULATED AS DESCRIBED IN ISO 3746.

(E) VERTICAL BARRIER WALL FLANGES ARE PROVIDED WITH UNIT TO FILL GAP BETWEEN

UNIT AND WALL ON BOTH SIDES. DIMENSIONS ARE 3-1/2 in [89 mm] WIDTH X 80 in [2032 mm] LONG.

(F) HORIZONTAL BARRIER WALL FLANGES ARE PROVIDED WITH UNIT TO FILL GAP BETWEEN

UNIT AND WALL ON BOTH SIDES. DIMENSIONS ARE 49 in [1245 mm] WIDTH X 16 in [406 mm] HEIGHT.

(G) MINIMUM REQUIRED DIMENSION FOR DOOR REMOVAL WITH LIMITED ADJUSTMENT AVAILABLE: +0 in [0 mm] THE STANDARD BARRIER WALL FLANGES CAN'T BE USED AT THAT DIMENSION. -3/4 in [19.1 mm]

(H) RECOMMENDED DIMENSION FOR DOOR REMOVAL WITH FULL LEVELING ADJUSTMENT AVAILABLE: +1 in [25.4 mm] THE STANDARD BARRIER WALL FLANGES CAN BE USED AT THAT DIMENSION. *(NOTE E AND F) -3/4 in [19.1 mm]

FRONT VIEW

SERVICE ACCESS

SIDE OF UNIT

*(NOTE B)

TOP VIEW

PASS THROUGH

DOOR

SIDE VIEWLEVEL FLOOR

*(NOTE C)

W A

LL

O

P E

N

IN

G H

E

IG

H T

C E

IL

IN

G H

E

IG

H T

FO

R

D O

O R

R E

M O

V A

L

Page 1 of 10 10292220.C

Released S0960

2X 33-7/16 [849]

2X 27-11/16 [704]

5-11/16 [144]

7-1/8 [181]

4X 15 [381]

2-1/8 [54]

17-3/8 [441]

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

*NOTES:

(A) CONCRETE FLOOR MUST REMAIN FREE OF DEFECTS IN THESE AREA, AS PER ESR-1917 & ESR-3187.

FOR INSTALLATION OF SEISMIC ANCHORS (NOT PROVIDED BY STERIS).

(B) THE FLOOR SINK AND DRAIN SHALL NOT OCCUPY THESE AREAS. 5 in x 5 in [127 mm x 127 mm] FLOOR SINK WITH Ø4 in [102 mm] DRAIN IS RECOMMENDED WHEN USING SEISMIC MOUNTS.

SEISMIC MOUNTS

SEISMIC ANCHORAGE

FOR DETAIL SEE FD354

QTY 4 ANCHORS

FINISHED WALL

PREFERENCE

LOCATION OF

FLOOR SINK

*(NOTE A AND B)

Page 2 of 10 10292220.C

80-3/4 [2051]

7-1/8 [181]

[381]

31-1/8 [791]

[533]

23-1/32 [585]

20-17/32 [521]

19-1/2 [495]

3-7/8 [99]

[457]

15-1/2 [394]

32-11/32 [822]

HW HOT WATER

CW COLD WATER

PW PURE WATER

A COMPRESSED AIR

D FLOOR DRAIN LOCATION

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

UTILITY (PLUMBING)

CONNECTION POSITION

FRONT VIEW

A

D

PW

HW

CW

SIDE VIEW

HW

PW

CW

A

D

CR

STOP VIEW

HW

CW

PW

A

CR

S

D

S STEAM

CR CONDENSATE RETURN

S

CR

Page 3 of 10 10292220.C

79-1/4 [2013]

35-13/16 [910]

22-3/8 [568]

[636]

38-1/8 [967]

39-5/8 [1007]

E ELECTRICAL

N NETWORK (SEE NOTE)

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

UTILITY (ELECTRICITY)

CONNECTION POSITION

NOTE:

NETWORK CONNECTION (IF APPLICABLE)

- ACTIVE NETWORK DROP WITH CAT 5/6 CABLE WITHIN 10 ft [3 m] OF UNIT

- STATIC OR DHCP RESERVED IP ADDRESS (FOR EACH UNIT)

- NETWORK CABLE (SUPPLIED BY OTHER)

FRONT VIEW SIDE VIEW

E

N

E

N

TOP VIEW

E

N

Page 4 of 10 10292220.C

6-5/32 [156]

[178]

81-29/32 [2080]

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

EX VENTILATION EXHAUST

UTILITY (VENTILATION)

CONNECTION POSITION (OPTIONAL)

FRONT VIEW SIDE VIEW

EX

SUGGESTED DUCTWORK

(BY OTHERS)

EX

DRAIN 3/4 in [20 mm] O.D.

(ROUTED TO WASHER DRAIN)

DRY CONTACT (SEE PAGE 10 *(NOTE 3))

VAPOR CONDENSATE

COLLECTING POINT

TOP VIEW

EX

VENTILATION CONNECTION (OPTIONAL)

(SEE PAGE 9 *(NOTE 1))

EX

BY OTHERS

Page 5 of 10 10292220.C

60 [1524]

REF

8-1/2 [216]

132 [3350]

MAX.

17 [432]

INSTRUMENT

(With PUC HP)

UTENSIL

(With PUC HP)

CHEMICAL

(CH)

ENZYME CONSUMPTION PER CYCLE

(SEE PAGE 7; *(NOTE 4)) 0.74 oz [22 ml] 0.74 oz [22 ml]

DETERGENT CONSUMPTION PER CYCLE

(SEE PAGE 7; *(NOTE 4)) 0.74 oz [22 ml] 0.74 oz [22 ml]

LUBRICANT CONSUMPTION PER CYCLE

(SEE PAGE 7; *(NOTE 4)) 0.2 oz [6 ml] N/A

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

*NOTES:

(A) MAXIMUM CONTAINER HEIGHT FROM FLOOR IS 24 in [610 mm]. AMBIENT TEMPERATURE SHALL NOT EXCEED 40 °C [104 °F]

DOSING STATION

CEILING

ROUTE CHEMICAL LINES

AND SIGNAL CABLE

THROUGH PIPE TO REACH

BOTTOM OF UNIT

CHEMICAL CONTAINERS

*(NOTE A)

12 ft [3.6 m] OF INJECTION LINE

(FROM CONTAINERS TO

DOSING STATION)

DOSING STATION

WEIGHT : 15 lb [6.8 kg]

2X FD349 SHOWN

UP TO 5 TUBES

100 ft [30 m] OF INJECTION LINE

AND SIGNAL CABLE

DETERGENT LINES SHALL NOT BE IN CONTACT WITH HOT SURFACES.

STERIS RECOMMENDS CUSTOMER TO ROUTE CHEMICAL LINES AND

SIGNAL CABLE THROUGH A 3 in [76 mm] ID CONDUIT.

EXCESS LENGTH MUST BE CUT BEFORE CONNECTING TO

WASHER/STATION TO AVOID FORMING BUNDLES THAT COULD

AFFECT THE CHEMICAL INJECTION SYSTEM'S PERFORMANCE

(DO NOT CUT SIGNAL CABLE).

Page 6 of 10 10292220.C

UTILITY REQUIREMENTS CHART (PLUMBING)

UTILITY PARAMETER REQUIREMENT NOTES

HOT WATER

(HW)

CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]

1. STERIS REQUIRES TO INSTALL WATER HAMMER

ARRESTOR, UNION & PRESSURE GAUGE.

2. CUSTOMER MUST PROVIDE UTILITY

CONNECTIONS WITH SHUTOFFS DISCONNECTS

WITHIN 2 FEET OF THE PERIMETER OF THE

EQUIPMENT AND BELOW THE CEILING DECK.

*STERIS OFFERS FLEXIBLE HOSES

ACCESSORY (FD088 OR FD037)

WITH INSTRUCTIONS.

3. SAMPLING VALVES INSTALLATION AS CLOSE AS

POSSIBLE TO WASHER WATER CONNECTIONS IS

RECOMMENDED.

PRESSURE RANGE

15 - 50 psig DYNAMIC

[103-345 kPa] MAX. 90 psig STATIC [620 kPA]

OPERATING FLOW RATE 6.8 US GPM [25.7 L/min] TO

14.1 US GPM [53.4 L/min]

TOTAL CONSUMPTION PER

INSTRUMENT CYCLE

*(NOTE 4)

9.9 US GAL [37.5 L]

WITH ACU-WASH RESERVOIR:

6.6 US GAL [25.0 L]

TEMPERATURE RANGE 110 °F [43 °C] TO 150 °F [65.5 °C]

WATER QUALITY *(NOTE 5) HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)

COLD WATER

(CW)

*(NOTES 6)

CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]

4. CONSUMPTION BASED ON INSTRUMENT CYCLE WITH

PROLYSTICA ULTRA CONCENTRATE HIGH

PERFORMANCE CHEMICALS (PUC HP) WITH

THE FOLLOWING UTILITY CONDITIONS:

THREE-LEVEL MANIFOLD RACK

(CW) COLD WATER PRESSURE: 30 psig DYNAMIC [207 kPa]

(CW) COLD WATER TEMPERATURE: 70 °F [21 °C]

(HW) HOT WATER TEMPERATURE: 150 °F [65 °C]

(PW) PURE WATER TEMPERATURE 70 °F [21 °C]

(S) STEAM SUPPLY: 80 psig DYNAMIC [552 kPa]

UTENSIL CYCLE CONSUMPTION:

TWO-LEVEL MANIFOLD RACK

CW: 10.4 US GAL [39.4 L]

HW: 9.6 US GAL [36.3 L]

PW: 6.4 US GAL [24.2 L]

S: 14.1 lb [6.4 kg]

5. STERIS RECOMMENDS THE FOLLOWING WATER

QUALITY PARAMETERS:

TOTAL DISSOLVED SOLIDS: 100-200 ppm (MAX. 500 ppm) TOTAL ALKALINITY (CaCO3) : 70-120 ppm (MAX. 180 ppm) pH : 6.8-7 (MAX. 7.5) TOTAL SILICA : 0.1-0.5 ppm (MAX. 1 ppm)

PRESSURE RANGE

30 - 50 psig DYNAMIC

[206-345 kPa] MAX. 90 psig STATIC [620 kPa]

OPERATING FLOW RATE 10.7 US GPM [40.5 L/min] TO

14.1 US GPM [53.4 L/min]

TOTAL CONSUMPTION PER

INSTRUMENT CYCLE

*(NOTE 4)

FOR SUMP FILLING: 10.5 US GAL [39.8 L]

(ADD THE FOLLOWING CONSUMPTION

FOR OPTIONS/CONFIGURATIONS)

CONDENSATE RETURN COOL DOWN:

6.2 US GAL [23.5 L]

NON-VENTED SYSTEM:

12.9 US GAL [48.8 L]

DRAIN COOLDOWN:

14.5 US GAL [54.9 L]

DRAIN ACU-COOLDOWN:

10.9 US GAL [41.3 L]

TEMPERATURE RANGE 70 °F [21 °C] MAXIMUM

WATER QUALITY *(NOTE 5) HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)

PURE WATER

(PW)

CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]

6. BACK FLOW PREVENTER AGAINST BACK

SIPHONAGE MAY BE REQUIRED (LOCAL

PLUMBING CODE)

7. PW IS OPTIONAL. IF UNAVAILABLE, UNIT WILL

PROCESS THERMAL RINSE USING HOT WATER

SUPPLY.

PRESSURE RANGE 5 - 30 psig DYNAMIC [30 - 206 kPa] MAX. 90 psig STATIC [620 kPa]

OPERATING FLOW RATE 5.5 US GPM [21 L/min] TO

12.6 US GPM [48 L/min]

TOTAL CONSUMPTION PER

INSTRUMENT CYCLE

*(NOTE 4 AND NOTE 7)

6.6 US GAL [25.0 L]

TEMPERATURE RANGE 70 °F [21.1 °C] TO 150 °F [65.5 °C]

RESISTIVITY

(CONDUCTIVITY)

0.1-0.5 M -cm (2-10 µS/cm) pH 6.8 - 7.5

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

Page 7 of 10 10292220.C

UTILITY REQUIREMENTS CHART (PLUMBING)

COMPRESSED

AIR

(A)

CONNECTION SIZE / TYPE 1/8 in / FEMALE NPT [1/8 in / MALE BSPT]

1. AIR PURITY PER ISO-8573-1: 2010 [7:4:4]

MAX. PARTICLE DENSITY: 10 mg/m³

MAX. DEW POINT FOR WATER CONTENT:

37 °F [3 °C]

MAX. OIL CONCENTRATION FOR THE OIL

CONTENT: 5 mg/m³

2. STERIS RECOMMENDS A REFRIGERATED AIR

DRYER WHERE ENVIRONMENTAL DEW POINT

CONDITIONS ARE HIGHER THAN

RECOMMENDED.

3. CUSTOMER MUST PROVIDE UTILITY

CONNECTIONS WITH SHUTOFFS DISCONNECTS

WITHIN 2 FEET OF THE PERIMETER OF THE

EQUIPMENT AND BELOW THE CEILING DECK.

PRESSURE RANGE MIN 80 psig DYNAMIC [552 kPa] MAX. 125 psig STATIC [860 kPa]

OPERATING FLOW RATE 1.75 SCFM [50 L/min]

DRAIN

(D)

FLOOR DRAIN SIZE MIN 4 in [102 mm] O.D. WITH 8 in x 8 in [204 mm x 204 mm] FLOOR SINK

4. WASHER DRAIN OUTLET SIZE 1-1/2 in / MALE NPT

5. MAX OFFSET OF 3/4 in [19 mm] FROM

OUTLET (PAGE 3, D).

6. FOR DRAIN REQUIREMENTS WHEN USING

SEISMIC MOUNTS, SEE PAGE 2.

OPERATING FLOW RATE GRAVITY DRAIN : 50 US GPM

[189.3 L/min]

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

CHEMICAL

DOSING

SYSTEM

CEILING ROUTING

CONDUIT 3 in CONDUIT

11. STERIS RECOMMENDS TO ROUTE CHEMICAL

LINES AND SIGNAL CABLE THROUGH A

3 in [76 mm] ID CONDUIT (BY OTHERS).

100 ft [30 m] OF INJECTION LINE AND

SIGNAL CABLE ARE PROVIDED WITH THE UNIT.

STEAM INLET

(S)

CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]

7. STERIS REQUIRES AT LEAST 97% SATURATED

DRY PLANT STEAM, FREE OF DIRT, RESIDUES

AND CONTAMINANTS.

8. STERIS RECOMMENDS TO INSTALL A STEAM

TRAP AS CLOSE AS POSSIBLE TO UNIT

CONNECTION.

9. PROPER OPERATION OF THE WASHER

REQUIRES A MINIMUM PRESSURE DIFFERENTIAL

OF 21.5 psig [148 kPa] BETWEEN STEAM

INLET AND CONDENSATE RETURN. IN CASES OF

LOW STEAM INLET PRESSURE, CONDENSATE

CAN BE ROUTED TO DRAIN TO INCREASE

DIFFERENTIAL.

*STERIS OFFERS CONDENSATE RETURN

TO DRAIN ACCESSORY (FD353)

10. CUSTOMER MUST PROVIDE UTILITY

CONNECTIONS WITH SHUTOFFS DISCONNECTS

WITHIN 2 FEET OF THE PERIMETER OF THE

EQUIPMENT AND BELOW THE CEILING DECK.

PRESSURE RANGE

30 - 80 psig DYNAMIC

[206-552 kPa] MAX. 90 psig STATIC [620 kPa]

OPERATING

MASS FLOW RATE

169 lb/h [76.5 kg/h] AT 30 psig [207 kPa]

271 lb/h [123 kg/h AT 80 psig [550 kPa]

TOTAL CONSUMPTION

PER INSTRUMENT CYCLE

(SEE PAGE 7, *(NOTE 4))

14.1 lb [6.4 kg]

CONDENSATE

RETURN

(CR)

PRESSURE RANGE MAX. 8.5 psig [68.9 kPa] BACK PRESSURE (17 ft [5.2 m] ELEVATION)

OPERATING FLOW RATE PEAK : 0.54 US GPM [2.05 L/min]

CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]

Page 8 of 10 10292220.C

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

UTILITY REQUIREMENTS CHART (ELECTRIC)

UTILITY

PARAMETER AND REQUIREMENT NOTES

CONNECTION

SIZE/TYPE

NOMINAL

AMPERAGE

RECOMMENDED

CIRCUIT

ELECTRICITY

(EL)

200-208V 60 Hz, 3-Ph

1 in [27 mm] 1-1/4 in [35 mm]

CONDUIT

36 A 50 A

1. ALL CONNECTIONS SHOULD BE IN

ACCORDANCE WITH LOCAL CODES.

2. NEUTRAL WIRE NOT REQUIRED.

3. ALWAYS FOLLOW LOCAL ELECTRICAL CODES

AND SAFETY-RELATED WORK PRATICES TO

DETERMINE THE VALUE OF PROTECTION.

4. WIRING ON THE EQUIPMENT TERMINATES AT

A JUNCTION BOX. WIRING BETWEEN JUNCTION

BOX AND BUILDING SERVICE LINES ARE NOT

SUPPLIED BY STERIS.

5. DISCONNECT SWITCHES WITH OFF POSITION

LOCKOUT (SUPPLIED BY OTHERS) MUST BE

INSTALLED IN ELECTRIC SUPPLY LINES NEAR THE

EQUIPMENT.

6. STERIS RECOMMENDS A LIGHT IN SERVICE

AREA.

200-208V 60 Hz, 3-Ph

WITH 40A

MIN. CIRCUIT

PROTECTION

1 in [27 mm] 1-1/4 in [35 mm]

CONDUIT

28.5 A 40 A

200-208V 50 Hz, 3-Ph

1 in [27 mm] 1-1/4 in [35 mm]

CONDUIT

35 A 50 A

460-480V 60 Hz, 3-Ph

1 in [27 mm] 1-1/4 in [35 mm]

CONDUIT

16.5 A 30 A

380-400V 60 Hz, 3-Ph

1 in [27 mm] 1-1/4 in [35 mm]

CONDUIT

19 A 30 A

380-400-415V 50 Hz, 3-Ph

1 in [27 mm] 1-1/4 in [35 mm]

CONDUIT

18.5 A 30 A

Page 9 of 10 10292220.C

UTILITY REQUIREMENTS CHART

VENTILATION

EXHAUST

(EX)

OPTIONAL

CONNECTION SIZE 3 in [76 mm]

OUTSIDE DIAMETER

1. NOT REQUIRED WITH NON VENTED UNIT.

2. STERIS REQUIRES CORROSION-RESISTANT AND

WATER-TIGHT DUCT:

- MIN OPERATING TEMPERATURE 210 °F [99 °C]

- SLOPED TOWARDS THE WASHER AND NO DEAD LEGS

- INSTALL DRAIN PART AT LOWEST POINT NEAR UNIT

3. EXHAUST FAN NORMALLY OPEN-AUXILIARY CONTACTS

IS AVAILABLE TO SUPPLY A SIGNAL TO THE FACILITY HVAC

CONTROL SYSTEM WHENEVER THE WASHER VENTILATION

EXHAUST FAN IS IN OPERATION.

PRESSURE RANGE 1/2 in [50 mm]

VACUUM AT CONNECTION

OPERATING FLOW RATE 60 SCFM [1.7 m³/min]

TEMPERATURE RANGE MAX. 240 °F [116 °C]

PHYSICAL SPECIFICATIONS

SHIPPING DIMENSION W x H x L 50 in x 91 in x 42 in [1270 mm x 2311 mm x 1067 mm]

1. LEVELING FEET ARE PROVIDED FOR PROPER INSTALLATION.

2. PERMISSIBLE ENVIRONMENTAL CONDITIONS.

THIS EQUIPMENT IS DESIGNED TO GIVE OPTIMAL RESULTS

UNDER THE FOLLOWING CONDITIONS:

- INDOOR USE ONLY;

- ALTITUDE OF OPERATION UP TO 6562 ft [2000 m];

- TEMPERATURE OF 41 TO 104 °F [5 to 40 °C];

- MAXIMUM RELATIVE HUMIDITY IS 80% FOR TEMPERATURE UP

TO 88 °F [31 °C], DECREASING LINEARLY TO 50% RELATIVE

HUMIDITY AT 104 °F [40 °C];

- POLLUTION DEGREE 2: EQUIPMENT MUST BE INSTALLED IN AN

ENVIRONMENT WHERE NORMALLY ONLY NON-CONDUCTIVE

POLLUTION OCCURS BUT WHERE OCCASIONAL, TEMPORARY

CONDUCTIVITY CAUSED BY CONDENSATION CAN BE EXPECTED

(AS DETERMINED ACCORDING TO INTERNATIONAL STANDARD

EN/IEC 61010-1).

3. STERIS ASSUMES NO RESPONSIBILITY FOR CHANGES MADE TO,

OR FAILURE TO OBSERVE THE SPECIFICATIONS ON EQUIPMENT

DRAWING AND NOTE PAGES. SPECIFICATIONS AND

DESCRIPTIONS ARE SUBJECT TO CHANGE WITHOUT NOTICE.

SHIPPING WEIGHT 1268 lb [575 kg]

OPERATING DIMENSION W x H x L 42 in x 80.75 in x 32 in [1067 mm x 2051 mm x 813 mm]

OPERATING WEIGHT

(WITH WATER AND FULLY LOADED) 1254 lb [569 kg]

AVERAGE FLOOR LOAD

OVER WASHER FOOTPRINT 0.93 psi [6.43 kPa]

MAX. FLOOR LOAD

AT WASHER FOOT PADS

(3 in [76.2 mm] DIAMETER PADS) 67 psi [462 kPa]

NOISE LEVEL:

A-WEIGHTED EQUIVALENT

SURFACE SOUND PRESSURE LEVEL

60.6 dB(A)

CALCULATED AS DESCRIBED

IN ISO 3746.

OPENING AREA FROM

CLEAN SIDE TO SOILED SIDE

46.5 in² [300 cm²]

*(NOTE 4)

4. OPENING AREA CAN BE REDUCED TO 13 in² [84 cm²]

WITH INSTALLATION OF AIR MANAGEMENT PANELS

ACCESSORY (FD352)

CHAMBER DIMENSION W x H x L

VOLUME

26.5 in x 26.25 in x 25.5 in [673 mm x 667 mm x 648 mm]

10.3 ft³ [0.29 m³]

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

TYPICAL HEAT LOSS AT 75 °F [24 °C]

CONTINUOUS OPERATION

NON-VENTED UNIT

WITHOUT ACU-RINSE RES.

VENTED UNIT (OPTIONAL)

WITHOUT ACU-RINSE RES.

CLEAN SIDE 2049 Btu/h [600 W]

2049 Btu/h [600 W]

SOILED SIDE 6557 Btu/h [1920 W]

4747 Btu/h [1390 W]

TOTAL 8606 ± 854 Btu/h [2520 ± 250 W]

6796 ± 854 Btu/h [1990 ± 250 W]

VALUES INCLUDE HEAT LOSS DURING DOOR OPENING FOR LOADING AND UNLOADING OPERATIONS. "TOTAL" WILL VARY DEPENDING ON DISINFECTION TIME AND TEMPERATURE, DRYING TIME AND TEMPERATURE AND USE RATE OF UNIT.

AN ADDITIONAL HEAT LOSS FOR LOAD COOL DOWN IN CLEAN SIDE CAN BE ADDED TO THE TOTAL HEAT LOSS OF THE WASHER:

MAXIMUM 1713 Btu [1807 kJ]

Page 10 of 10 10292220.C

FA520521R00030002 Atch 5 Revised Cut Sheets for JSN S0442 and S0960
JSN S0442
JSN S0960

File details come from the government source that posted it. Updated .