This document contains multiple engineering reports summarizing shielding requirements for radiology areas in a Veterans Affairs Medical Center. The reports evaluate shielding needs for an Interventional Radiology Lab, Electrophysiology Lab, Cardiac Catheterization Lab, Radiology Fluoroscopy area, and Radiology Multi-Purpose area. The reports identify specific barrier locations, required shielding thicknesses of lead or concrete to limit public radiation exposure to below regulatory dose limits, and other comments. Shielding requirements are determined using methodology from NCRP Report 147 on structural shielding for medical x-ray facilities. All shielding should extend 7 feet above the finished floor. The related federal contract opportunity is to remodel areas 2W for Interventional Radiology, Cardiac Catheterization, and Electrophysiology procedures for the Department of Veterans Affairs Veterans Health Administration VISN 23.
View the file
Other files
Show all 25
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Environmental Health and Safety Office University of Iowa 122 Grand Avenue Court Iowa City, Iowa 52242-0000 319-335-8501 ehs.research.uiowa.edu
October 28, 2021
TO: Thaddeus Longo General Engineer Facility Engineering Group Iowa City VAMC
FROM: Laurie Scholl/Gordon Axt Radiation Safety
RE: Shielding Requirements for VAMC IR LAB 2W153, EP LAB 2W151 and CATH LAB 2W152
Shielding requirements for the IR LAB 2W153, EP LAB 2W151 and CATH LAB 2W152 are summarized in the tables below. Determinations were made using methodology in NCRP 147 "Structural Shielding Design for Medical X-Ray Imaging Facilities". All shielding should extend to a height of 7 ft. above the finished floor.
The dose limit of 0.02 mGy/wk. corresponds to the IDPH limit of 100 mrem/yr. for the public. Distance is from the point of the x-ray tube to 1 ft. beyond the barrier.
IR LAB 2W153
Barrier Description Shielding Required
Comments
A Exterior Wall N/A N/A
B Stair Area Wall 1/16” Lead Equivalent
C Control Rm Wall 2W153A.3
1/16” Lead Equivalent
DR1 Doorway 2W153.1
1/16” Lead Equivalent
DR2 Doorway 2W153A.1
1/16” Lead Equivalent
E Control Room Wall 2W152A
1/16” Lead Equivalent
Ceiling Ceiling 1 3/4” Concrete
Floor Floor 1 3/4” Concrete https://ehs.research.uiowa.edu/
EP LAB 2W151
Comments
A Exterior Wall
N/A N/A
B Control Rm Wall 2W151A.3
1/16” Lead Equivalent
C Corridor Wall 2C03B
1/16” Lead Equivalent
DR Doorway 2W151.1
1/16” Lead Equivalent
Equip E1 Equipment Wall 2W159.1/2W159.2
1/16” Lead Equivalent
Elec E2 Electrical Rm 2W179
1/16” Lead Equivalent
CATH LAB 2W152
Comments
A Exterior Wall N/A N/A
B Control 2W152.3
1/16” Lead Equivalent
C Corridor 2C03B 1/16” Lead Equivalent
DR Doorway 2W152.1 1/16” Lead Equivalent
E Control Rm 2W151A 1/16” Lead Equivalent
Summary of Shielding Requirements:
Please let me know if you have any questions.
University of Iowa
122 Grand Avenue Court
Iowa City, Iowa 52242-0000
January 3, 2022
General Engineer
RE: Shielding Requirements for VAMC Radiology Fluoroscopy – 2W130
Shielding requirements for the Radiology Fluoroscopy 2W130 area are summarized in the tables below.
Determinations were made using methodology in NCRP 147 "Structural Shielding Design for Medical X-Ray
Imaging Facilities". All shielding should extend to a height of 7 ft. above the finished floor.
Summary of shielding requirements:
Fluoroscopy 2W130
Barrier Description Shielding Required Comments
A Corridor 2C06 1/16” Lead Equivalent
B Multi-Purpose Room 2W131 1/16” Lead Equivalent
D EKG2 2S39 1/16” Lead Equivalent
E Control Room Wall 1/16” Lead Equivalent
W Control Rm Window FL3 0.5 mm Lead Equivalent
F/J Doorway Toilet 2W130A/
EKG Tech Area 2S43A
1/16” Lead Equivalent
G Nurse MGR Office 2W129C 1/16” Lead Equivalent
H IR MGR Office 2W129D 1/16” Lead Equivalent
I Door 2W130.1 1/16” Lead Equivalent
Please contact Radiation Safety if you have any questions
University of Iowa
122 Grand Avenue Court
Iowa City, Iowa 52242-0000
December 17, 2021
General Engineer
RE: Shielding Requirements for VAMC Radiology Multi-Purpose – 2W131
Shielding requirements for the Radiology Multi-Purpose 2W131area are summarized in the tables below.
Determinations were made using methodology in NCRP 147 "Structural Shielding Design for Medical X-Ray
Imaging Facilities". All shielding should extend to a height of 7 ft. above the finished floor.
Summary of Shielding Requirements:
Multi-Purpose Room – 2W-131
Comments
A Corridor 2C06 1/16” Lead Equivalent
B Door 2W131.1 1/16” Lead Equivalent
C Ultrasound 2W132 1/16” Lead Equivalent
D Echo2 2S38 1/16” Lead Equivalent
E Control Room Wall 1/16” Lead Equivalent
W Control Room Window 0.7 mm Lead equivalent
F Corridor 2CS11 1/16” Lead Equivalent
G Fluoroscopy Room 2W130 1/16” Lead Equivalent
Please contact Radiation Safety if you have any questions.
| SheildingVAMC IRCATH 102021 |
| SheildingVAMCFluoroscopy 2W130REV-GAfinal |
| SheildingVAMCMuti-Purpose 2W131REV-GA |