Attachment P - 2W IR CATH EP_Draeger-Boom Install_Site Specific.pdf

PDF 9 MB Posted

Attached to
Z1DA--636A8-21-001 CON Remodel 2W for IR/CATH/EP 36C263-22-AP-3804
Solicitation number
36C26323R0002
Issued by

About this file

This document package includes technical drawings and specifications for a medical facility renovation project to remodel areas for interventional radiology, cardiac catheterization, and endoscopy at a Department of Veterans Affairs medical center. The project involves installing new medical equipment mounting structures, electrical and medical gas supply lines, and lighting in multiple procedure rooms. Key equipment to be installed includes C-arms, surgical lights, booms, and supply columns from Draeger and Polaris. The technical drawings provide detailed mounting and connection requirements for this equipment in areas 2W151, 2W152, 2W153, as well as electrical panel schedules and medical gas outlet specifications by room.

View the file

Other files

File Type Posted
Attachment N - Questions Request Form-all questions combined_Response_12132022.doc DOC document
Attachment S- 2W IR CATH EP_Luminos Agile Max_Site Specific.pdf PDF
Attachment O - 2W IR CATH EP_Cath Lab_Site Specific.pdf PDF
36C26323R0002 0001.docx DOCX document
Attachment U - Shielding Documents.pdf PDF
636A8-21-001 Pre-Proposal Conference Sign-in List 11222022.pdf PDF
Evaluation Factors -Amendment 0001.docx DOCX document
Attachment R- 2W IR CATH EP_IR Lab_Site Specific.pdf PDF
Attachment J.1 636A8-21-001 Sole Source List.pdf PDF
Attachment T - 2W IR CATH EP_Multi-Use Rad Room_Site Specific.pdf PDF
Attachment J - Brand Name Only JAs_Redacted.pdf PDF
Attachment H - 2W for IR-CATH-EP CDs Drawings 4 of 5.pdf PDF
Attachment F - 2W for IR-CATH-EP CDs Drawings 2 of 5.pdf PDF
Attachment C - 2W IR CATH EP_CDs_ SPECS Vol1.pdf PDF
Attachment K - Wage Determination IA20210068 dated 09-02-2022.pdf PDF
Attachment I - 2W for IR-CATH-EP CDs Drawings 5 of 5.pdf PDF
36C26323R0002.docx DOCX document
Attachment A - SOW 636A8-21-001 final.pdf PDF
Attachment M - Past Performance Questionnaire - 3 Pages.docx DOCX document
Attachment B - Price Schedule Breakdown Sheet - 1 page.xlsx XLSX spreadsheet
Attachment N - Questions Request Form.doc DOC document
Attachment G - 2W for IR-CATH-EP CDs Drawings 3 of 5.pdf PDF
Attachment L - NARA Contract Language 2021.pdf PDF
Attachment D - 2W IR CATH EP_CDs_ SPECS Vol2.pdf PDF
Attachment E - 2W for IR-CATH-EP CDs Drawings 1 of 5.pdf PDF
Show all 25

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

TITLE:

Date:

Name:

Rev04

AUTHOR:

12-16-21

EH

02-10-22

SHEET:

3D-REF

DESCRIPTIONS3D REF

EP Lab 2W151

CATH Lab 2W152

IR Lab 2W153

°m m

R1000

R1000 m m

[1 m m

[1

4m m [8

.8 m m

[5 0.

2" m m

[9 7.

0" m m

[7 6.

8"

7m m [3

1"

533mm [21.0"]

Max Weight Tensile Force Torque at Ceiling

360°

0m m [2

9.

5"

Detail Mount

"A" Column

600mmx600mm [23.5" x 23.5"]

Flat Cover Panel

View 1:16 "A"

4m m [6

.8 m m [2

.5

130.1kg

330°

9m m [6

.3

2m m [1

3.

1"

1250mm [49.2"]

1000mm [39.4"]

(T ub e Le ng th

Movita plus (P os t L en gt h)

360°

3728N 6532Nm

This drawing is for illustrative purposes only.

Line Denotes Regulator/Supplied by TBD/or by Draeger.

Line Denotes Electrical/Supplied by Others.

Line Denotes Medical Gases/Supplied by TBD/or by Draeger.

Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others

Line Denotes Structure Supplied by Draeger.

Line Denotes Electrical/Supplied by TBD/or by Draeger.

AIR AIRO

O

2 VAC VAC VACO

Red denotes Structure by Others

301mm [11.9"]

Finished Floor

Deck

Finished Ceiling

Rough-In Views 1:16Rough-In Notes:

1)

2)

3)

Deck (by Others)

Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)

Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)

Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)

4m m [6

.8

533mm [21"]

Top View

Elevation View

(See "Rough-In Notes" #3)

A A

533mm [21"] mm

[49 .2"

Top View 1:20 "A"

1000mm [39.4"]

R1250mm[R49.2"]

R2250mm [R88.6"]

View L

View F

View R

Top View

Detail Mount "A" Placement Column

Column Detail View M1:10

Bottom View

View B

1 / Ambia Column

Mount Location "A" Ambia C/M 500 (article-number: OP75471)

EP Lab - Mount Location "A"

EHART

16DEC2021

1) Frame Color: Medical White W1100/S-0500-N

2) Service Plate Color: Anodized

Column

R

Q

P

2 3 4

N

M

L

K

J

5 6 7 8 9 10 11 12

H

G

F

E

D

C

B

A

2 3 4 5 6 7 8 9 10 11 12

13 14 15 16 1817 19 size A020 21 22 23

R

Q

P

N

M

L

K

J

13 14 15 16 1817 19

H

G

F

E

D

20 21 22 23

C

B

A

U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.

A ll rig ht s re se rv ed

R ef er to p ro te ct io n no tic e

IS

O

Type:

3D File:

System:

Date:

Name:

Author: Approved by: Scale:

Date:

Name:

View: 1:16 Others: See Details

3DFL - VA Iowa City 08FEB2022 1041

Quantity of Units: Project Name:

3D-Tool #:

Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):

VA Iowa City EP CATH IR

US United States120 60

Date: Author: Revision Text: Rev #:Notes:

Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented

Sample units for the definition of the US Country Standard.

Units partly against current Project Planning Specifications:

Couplings for Flammable Gases above Electrical Components (Sockets).

Gas Couplings Hoses.

1)

2)

PRESUBMITTALS ONLY

NOT FOR CONSTRUCTION

Rm Name, Rm # + Mount #:

Mount File #:

1st SubmissionEHart22DEC2021

287LBS/130KGMax. Load Ambia Column Data:

176LBS/80KG

22LBS/10KG

Shelf Max. Load (ea.)

Rails Max. Load (set)

Shelving Data:

01 - CATH LABs

Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 2

Medical Air Beacon Medaes Chemetron 0 Vacuum Beacon Medaes Chemetron 2

N2O Beacon Medaes Chemetron 0 AGSS/WAGD Beacon Medaes Chemetron 0

CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0

High Voltage Emergency Quad Power/UPS Hubbell Circuits / 2 1 Emergency Quad Power Hubbell Circuits / 3 1

Emergency Duplex Power Hubbell Circuit / 0 0 Normal Quad Power Hubbell Circuit / 0 0

Normal Duplex Power Hubbell Circuits / 4 & 5 2 Electric Brakes Hubbell Dual Brakes, Circuit / 1 1 Surgical Light Polaris Circuit / 0 0

Low Voltage Single Blank Provision Hubbell 1 Triple Blank Provision Hubbell 0

RJ45 Hubbell Cat.6 / 4 1 Draeger EQPM Connection Hubbell Circuit / 1 1

Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0

Misc. Other 0

DI EQPM UpdateEHart10JAN2022

DI EQPM UpdateEHart08FEB2022

UPS

Measurements to be Verified on Site. (V.I.F.)

Plan View N.T.S.

DI EQPM UpdateEHart10FEB2022 04

AutoCAD SHX Text o533mm

AutoCAD SHX Text o533mm m m

[1 m m

[1

°m m

R1000

R1000

4m m [8

.8 m m

[6 6.

6" m m

[9 7.

0" m m

[7 6.

8"

360°

4m m [1

3.

1"

Detail Mount

"B" Head

4m m [6

.8 m m [2

.5

330°

9m m [6

.3

2m m [1

3.

1"

1250mm [49.2"]

1000mm [39.4"]

(T ub e Le ng th

Movita plus (P os t L en gt h)

877mm [34.5"]

360°

3728N 6532Nm130.1kg

7m m [3

1"

533mm [21.0"]

Max Weight Tensile Force Torque at Ceiling

600mmx600mm [23.5" x 23.5"]

Flat Cover Panel

This drawing is for illustrative purposes only.

Line Denotes Regulator/Supplied by TBD/or by Draeger.

Line Denotes Electrical/Supplied by Others.

Line Denotes Medical Gases/Supplied by TBD/or by Draeger.

Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others

Line Denotes Structure Supplied by Draeger.

Line Denotes Electrical/Supplied by TBD/or by Draeger.

AIR AIRO

O

2 VAC VAC VACO

Red denotes Structure by OthersView 1:16 "B"

Finished Floor

Deck

Finished Ceiling

Rough-In Views 1:16Rough-In Notes:

1)

2)

3)

Deck (by Others)

Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)

Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)

Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)

4m m [6

.8

533mm [21"]

Top View

Elevation View

(See "Rough-In Notes" #3)

A A

533mm [21"] mm

[49 .2"

Top View 1:20 "B"

1000mm [39.4"]

R1250mm[R49.2"]

R2250mm [R88.6"]

View L

View F

View R

Top View

Detail Mount "B" Placement Column

Column Detail View M1:10

Bottom View

View B

View U

Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 2

Medical Air Beacon Medaes Chemetron 1 Vacuum Beacon Medaes Chemetron 2

N2O Beacon Medaes Chemetron 1 AGSS/WAGD Beacon Medaes Chemetron 1

CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0

High Voltage Emergency Quad Power Hubbell Circuit / 2 1 Emergency Duplex Power/UPS Hubbell Circuits / 3 & 4 2

Normal Quad Power Hubbell Circuit / 0 0 Normal Duplex Power Hubbell Circuits / 5 & 6 2

Electric Brakes Hubbell Dual Brakes, Circuit / 1 1 Surgical Light Polaris Circuit / 0 0

Low Voltage Single Blank Provision Hubbell 2 Triple Blank Provision Hubbell 0

RJ45 Hubbell Cat.6 / 4 1 Draeger EQPM Connection Hubbell Circuit / 1 1

Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0

Misc. Other 0

1 / Ambia Head

Mount Location "B" Ambia H/M 750 (article-number: OP75484)

EP Lab - Mount Location "B"

EHART

16DEC2021

1) Equipment Color: Medical White W1100/S-0500-N

2) Frame Color: Medical White W1100/S-0500-N

3) Service Plate Color: Anodized

R

Q

P

2 3 4

N

M

L

K

J

5 6 7 8 9 10 11 12

H

G

F

E

D

C

B

A

2 3 4 5 6 7 8 9 10 11 12

13 14 15 16 1817 19 size A020 21 22 23

R

Q

P

N

M

L

K

J

13 14 15 16 1817 19

H

G

F

E

D

20 21 22 23

C

B

A

U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.

A ll rig ht s re se rv ed

R ef er to p ro te ct io n no tic e

IS

O

Type:

3D File:

System:

Date:

Name:

Author: Approved by: Scale:

Date:

Name:

View: 1:16 Others: See Details

3DFL - VA Iowa City 08FEB2022 1041

Quantity of Units: Project Name:

3D-Tool #:

Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):

VA Iowa City EP CATH IR

US United States120 60

Date: Author: Revision Text: Rev #:Notes:

Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented

Sample units for the definition of the US Country Standard.

Units partly against current Project Planning Specifications:

Couplings for Flammable Gases above Electrical Components (Sockets).

Gas Couplings Hoses.

1)

2)

PRESUBMITTALS ONLY

NOT FOR CONSTRUCTION

Rm Name, Rm # + Mount #:

Mount File #:

1st SubmissionEHart22DEC2021

287LBS/130KGMax. Load Ambia Head Data:

176LBS/80KG

22LBS/10KG

Shelf Max. Load (ea.)

Rails Max. Load (set)

Shelving Data:

01 - CATH LABs

Head

View 1:16 "B"

DI EQPM UpdateEHart10JAN2022 02

UPS

UPS

Measurements to be Verified on Site. (V.I.F.)

Plan View N.T.S.

DI EQPM UpdateEHart08FEB2022 03

DI EQPM UpdateEHart10FEB2022 04

AutoCAD SHX Text o533mm m m

[4 6.

6" m m

[8 5.

1" m m

[8 9.

6" m m

[9 6.

8" m m

[9 2.

4" m m

[7 0.

7" m m

[4 9.

1" ]17 m m [6

8.

4"

] 21 m m

[8 3.

2" m m [2

.0 m m [2

.0

4m m [8

.0

6m m [1

2.

4" m m

[1 m m

[1

4m m [6

.8

3m m [1

1.

1"

View 1:16 "C" m m [2

.5

1450mm [57.1"]

(T ub e Le ng th

360°

330°

360°

340°

910mm

[35.8"]

0°

Polaris 600 incl. Control Panel + prepared Sterilizable Handle set E

360°

360°

1600mm [63.0"]

Lifting reduced 868mm/34.2" as shown (max. lifting of arm 1179mm/46.4")

7m m [3

1"

533mm [21.0"]

Max Weight Tensile Force Torque at Ceiling

600mmx600mm [23.5" x 23.5"]

Flat Cover Panel

Red denotes Structure by Others

This drawing is for illustrative purposes only.

Line Denotes Electrical/Supplied by Others.

Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others

Line Denotes Structure Supplied by Draeger.

Line Denotes Electrical/Supplied by TBD/or by Draeger.

1885N 1661Nm

Lifting reduced 979mm/38.6" as shown (max. lifting of arm 1179mm/46.4")

910mm [35.8"]

360°

Finished Floor

Deck

Finished Ceiling

Rough-In Views 1:16Rough-In Notes:

1)

2)

3)

Deck (by Others)

Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)

Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)

Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)

4m m [6

.8

533mm [21"]

Top View

Elevation View

(See "Rough-In Notes" #3)

A A

533mm [21"]

1600mm [63.0"]

910mm [35.8"]

1450mm [57.1"]

Top View 1:20 "C"

R1600mm

[R63.0"]

R2510mm

[R98.8"]

910mm

[3 5.8"]

R1450mm [R57.1"]

R2360mm[R92.9"]

1 / P600 and 1/X-Ray Shield

Mount Location "C"

Polaris 600 (article-number: OP75084)

X-Ray Shield (article-number: OV2013-ML681-04 OT50001)

EP Lab - Mount Location "C"

EHART

16DEC2021

R

Q

P

2 3 4

N

M

L

K

J

5 6 7 8 9 10 11 12

H

G

F

E

D

C

B

A

2 3 4 5 6 7 8 9 10 11 12

13 14 15 16 1817 19 size A020 21 22 23

R

Q

P

N

M

L

K

J

13 14 15 16 1817 19

H

G

F

E

D

20 21 22 23

C

B

A

U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.

A ll rig ht s re se rv ed

R ef er to p ro te ct io n no tic e

IS

O

Type:

Type:

3D File:

System:

Date:

Name:

Author: Approved by: Scale:

Date:

Name:

View: 1:16 Others: See Details

3DFL - VA Iowa City 08FEB2022 1041

Quantity of Units: Project Name:

3D-Tool #:

Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):

VA Iowa City EP CATH IR

US United States120 60

Date: Author: Revision Text: Rev #:Notes:

Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented

Sample units for the definition of the US Country Standard.

Units partly against current Project Planning Specifications:

Couplings for Flammable Gases above Electrical Components (Sockets).

Gas Couplings Hoses.

1)

2)

PRESUBMITTALS ONLY

NOT FOR CONSTRUCTION

Rm Name, Rm # + Mount #:

Mount File #:

1st SubmissionEHart22DEC2021

01 - CATH LABs

Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 0

Medical Air Beacon Medaes Chemetron 0 Vacuum Beacon Medaes Chemetron 0

N2O Beacon Medaes Chemetron 0 AGSS/WAGD Beacon Medaes Chemetron 0

CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0

High Voltage Emergency Quad Power Hubbell Circuit / 0 0 Emergency Duplex Power Hubbell Circuit / 0 0

Normal Quad Power Hubbell Circuit / 0 0 Normal Duplex Power Hubbell Circuit / 0 0

Electric Brakes Hubbell Dual Brakes, Circuit / 0 0 Surgical Light Polaris 600 Circuit / 1 1

Low Voltage Single Blank Provision Hubbell 0 Quad Blank Provision Hubbell 0

RJ45 Hubbell Cat.6 / 4 0 Draeger EQPM Connection Hubbell Circuit / 0 0

Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0

Misc. Other 0

DI EQPM UpdateEHart10JAN2022 02

Measurements to be Verified on Site. (V.I.F.)

Plan View N.T.S.

DI EQPM UpdateEHart08FEB2022 03

DI EQPM UpdateEHart10FEB2022 04

AutoCAD SHX Text o533mm m m

[9 5.

8" m m

[4 4.

1" ] 16 m m [6

5.

8"

7m m [3

2.

2" m m

[9 7.

0" m m

[7 6.

8" m m

[1 m m

[1

7m m

[3 1"

] 17 4m m [6

.8 m m [2

.5

1m m [6

.3 m m

[4 7.

6"

9m m [6

.3

6m m [8

.5

533mm [21.0"]

600mmx600mm [23.5" x 23.5"]

Flat Cover Panel

View 1:16 "D"

(by Draeger)

(T ub e Le ng th

1400mm [55.1"]

1000mm[39.4"]Downward Movement 34"

°m m

R1000

R1000

360°

Detail Mount

"D" Column

330°

2m m [1

3.

1"

1250mm [49.2"]

1000mm [39.4"]

Movita plus (P os t L en gt h)

360°360°

0m m [1

1.

0"

(T ub e

Le ng th

7930Nm180.0kg

Max Weight Tensile Force Torque at Ceiling

Max Weight

4218N

(GD60)

(Ambia)

406mm [16.0"]

This drawing is for illustrative purposes only.

Line Denotes Regulator/Supplied by TBD/or by Draeger.

Line Denotes Electrical/Supplied by Others.

Line Denotes Medical Gases/Supplied by TBD/or by Draeger.

Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others

Line Denotes Structure Supplied by Draeger.

Line Denotes Electrical/Supplied by TBD/or by Draeger.

AIR AIRO

O

2 VAC VAC VACO

Red denotes Structure by Others

Finished Floor

Deck

Finished Ceiling

Rough-In Views 1:16Rough-In Notes:

1)

2)

3)

Deck (by Others)

Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)

Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)

Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)

4m m [6

.8

533mm [21"]

Top View

Elevation View

(See "Rough-In Notes" #3)

A A

533mm [21"]

Top View N.T.S.

"D"

R1400mm[R55.1"]

1400mm [55.1"] m m

[3 9.

4"

R2400mm [R94.5"] mm

[49 .2"

1000mm [39.4"]

R1250mm[R49.2"]

R2250mm [R88.6"]

View L

View F

View R

Top View

Detail Mount "D" Placement Column

Column Detail View M1:10

Bottom View

View B

Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 0

Medical Air Beacon Medaes Chemetron 2 Vacuum Beacon Medaes Chemetron 2

N2O Beacon Medaes Chemetron 0 AGSS/WAGD Beacon Medaes Chemetron 0

CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0

High Voltage Emergency Quad Power Hubbell Circuit / 2 1 Emergency Duplex Power/UPS Hubbell Circuits / 3 & 4 2

Emergency Duplex Power Hubbell Circuit / 5 1 Normal Quad Power Hubbell Circuit / 0 0

Normal Duplex Power Hubbell Circuits / 6 2 Electric Brakes Hubbell Dual Brakes, Circuit / 1 1

GD60 GD60 Circuit / 7 1

Low Voltage Single Blank Provision Hubbell 4 Quad Blank Provision Hubbell 1

RJ45 Hubbell Cat.6 / 4 1 Draeger EQPM Connection Hubbell Circuit / 1 1

Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0

Misc. Other 0

1 / Ambia Column and 1 / GD60 Pro-Flex

Mount Location "D"

Ambia (article-number: OP75478)

GD60 Pro-Flex (article-number: DCR-2018-001701)

EP Lab - Mount Location "D"

EHART

16DEC2021

1) Frame Color: Medical White W1100/S-0500-N

2) Service Plate Color: Anodized

Column

R

Q

P

2 3 4

N

M

L

K

J

5 6 7 8 9 10 11 12

H

G

F

E

D

C

B

A

2 3 4 5 6 7 8 9 10 11 12

13 14 15 16 1817 19 size A020 21 22 23

R

Q

P

N

M

L

K

J

13 14 15 16 1817 19

H

G

F

E

D

20 21 22 23

C

B

A

U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.

A ll rig ht s re se rv ed

R ef er to p ro te ct io n no tic e

IS

O

Type:

Type:

3D File:

System:

Date:

Name:

Author: Approved by: Scale:

Date:

Name:

View: 1:16 Others: See Details

3DFL - VA Iowa City 08FEB2022 1041

Quantity of Units: Project Name:

3D-Tool #:

Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):

VA Iowa City EP CATH IR

US United States120 60

Date: Author: Revision Text: Rev #:Notes:

Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented

Sample units for the definition of the US Country Standard.

Units partly against current Project Planning Specifications:

Couplings for Flammable Gases above Electrical Components (Sockets).

Gas Couplings Hoses.

1)

2)

PRESUBMITTALS ONLY

NOT FOR CONSTRUCTION

Rm Name, Rm # + Mount #:

Mount File #:

1st SubmissionEHart22DEC2021

331LBS/150KGMax. Load Ambia Column Data:

176LBS/80KG

22LBS/10KG

Shelf Max. Load (ea.)

Rails Max. Load (set)

Shelving Data:

01 - CATH LABs

DI EQPM UpdateEHart10JAN2022 02

UPS

UPS

Measurements to be Verified on Site. (V.I.F.)

Plan View N.T.S.

DI EQPM UpdateEHart08FEB2022 03

DI EQPM UpdateEHart10FEB2022 04

AutoCAD SHX Text o533mm m m

[1 m m

[1 m m [2

.0 m m

[8 9.

6" m m

[9 6.

8"

4m m [8

.0 m m

[4 9.

1" m m

[8 3.

2"

4m m [6

.8 m m

[6 8.

4"

3m m [1

1.

1"

7m m [3

1"

View 1:16 "E" m m [2

.5

1450mm [57.1"]

(T ub e Le ng th

360°

330°

360°

340°

910mm

[35.8"]

0°

Polaris 600 incl. Control Panel + prepared Sterilizable Handle set E

360°

360°

1600mm [63.0"]

Lifting reduced 868mm/34.2" as shown (max. lifting of arm 1179mm/46.4")

1885N 1661Nm

533mm [21.0"]

Max Weight Tensile Force Torque at Ceiling

600mmx600mm [23.5" x 23.5"]

Flat Cover Panel

Red denotes Structure by Others

This drawing is for illustrative purposes only.

Line Denotes Electrical/Supplied by Others.

Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others

Line Denotes Structure Supplied by Draeger.

Line Denotes Electrical/Supplied by TBD/or by Draeger.

m m

[4 6.

6" m m

[8 5.

1"

6m m [1

2.

4" m m [9

2.

4" m m

[7 0.

7"

Lifting reduced 979mm/38.6" as shown (max. lifting of arm 1179mm/46.4")

910mm [35.8"]

360° m m [2

.0

Finished Floor

Deck

Finished Ceiling

Rough-In Views 1:16Rough-In Notes:

1)

2)

3)

Deck (by Others)

Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)

Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)

Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)

4m m [6

.8

533mm [21"]

Top View

Elevation View

(See "Rough-In Notes" #3)

A A

533mm [21"]

1600mm [63.0"]

910mm [35.8"]

1450mm [57.1"]

Top View 1:20 "E"

R1600mm

[R63.0"]

R2510mm

[R98.8"]

910mm

[3 5.8"]

R1450mm [R57.1"]

R2360mm[R92.9"]

1 / P600 and 1/X-Ray Shield

Mount Location "E"

Polaris 600 (article-number: OP75084)

X-Ray Shield (article-number: OV2013-ML681-04 OT50001)

EP Lab - Mount Location "E"

EHART

16DEC2021

R

Q

P

2 3 4

N

M

L

K

J

5 6 7 8 9 10 11 12

H

G

F

E

D

C

B

A

2 3 4 5 6 7 8 9 10 11 12

13 14 15 16 1817 19 size A020 21 22 23

R

Q

P

N

M

L

K

J

13 14 15 16 1817 19

H

G

F

E

D

20 21 22 23

C

B

A

U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.

A ll rig ht s re se rv ed

R ef er to p ro te ct io n no tic e

IS

O

Type:

Type:

3D File:

System:

Date:

Name:

Author: Approved by: Scale:

Date:

Name:

View: 1:16 Others: See Details

3DFL - VA IOWA CITY 10JAN2022 0945

Quantity of Units: Project Name:

3D-Tool #:

Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):

VA Iowa City EP CATH IR

US United States120 60

Date: Author: Revision Text: Rev #:Notes:

Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented

Sample units for the definition of the US Country Standard.

Units partly against current Project Planning Specifications:

Couplings for Flammable Gases above Electrical Components (Sockets).

Gas Couplings Hoses.

1)

2)

PRESUBMITTALS ONLY

NOT FOR CONSTRUCTION

Rm Name, Rm # + Mount #:

Mount File #:

1st SubmissionEHart22DEC2021

01 - CATH LABs

Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 0

Medical Air Beacon Medaes Chemetron 0 Vacuum Beacon Medaes Chemetron 0

N2O Beacon Medaes Chemetron 0 AGSS/WAGD Beacon Medaes Chemetron 0

CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0

High Voltage Emergency Quad Power Hubbell Circuit / 0 0 Emergency Duplex Power Hubbell Circuit / 0 0

Normal Quad Power Hubbell Circuit / 0 0 Normal Duplex Power Hubbell Circuit / 0 0

Electric Brakes Hubbell Dual Brakes, Circuit / 0 0 Surgical Light Polaris 600 Circuit / 1 1

Low Voltage Single Blank Provision Hubbell 0 Quad Blank Provision Hubbell 0

RJ45 Hubbell Cat.6 / 4 0 Draeger EQPM Connection Hubbell Circuit / 0 0

Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0

Misc. Other 0

DI EQPM UpdateEHart10JAN2022 02

Measurements to be Verified on Site. (V.I.F.)

Plan View N.T.S.

DI EQPM UpdateEHart08FEB2022 03

DI EQPM UpdateEHart10FEB2022 04

AutoCAD SHX Text o533mm

9m m [3

8.

2" m m

[1

.0 m m

[8 2.

8" m m

[1 m m

[1

°m m

R1000

R1000

4m m [8

.8

360° m m

[4 7.

6"

Detail Mount

"F" Column

View 1:16 "F"

(by Draeger)

4m m [6

.8 m m [2

.5

330°

9m m [6

.3

2m m [1

3.

1"

1250mm [49.2"]

1000mm [39.4"]

(T ub e Le ng th

Movita plus (P os t L en gt h)

360°

533mm [21.0"]

Max Weight Tensile Force Torque at Ceiling

600mmx600mm [23.5" x 23.5"]

Flat Cover Panel

This drawing is for illustrative purposes only.

Line Denotes Regulator/Supplied by TBD/or by Draeger.

Line Denotes Electrical/Supplied by Others.

Line Denotes Medical Gases/Supplied by TBD/or by Draeger.

Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others

Line Denotes Structure Supplied by Draeger.

Line Denotes Electrical/Supplied by TBD/or by Draeger.

AIR AIRO

O

2 VAC VAC VACO

Red denotes Structure by Others

406mm [16.0"]

3728N 6532Nm130.1kg

5m m [2

5"

Finished Floor

Deck

Finished Ceiling

Rough-In Views 1:16Rough-In Notes:

1)

2)

3)

Deck (by Others)

Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)

Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)

Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)

4m m [6

.8

533mm [21"]

Top View

Elevation View

(See "Rough-In Notes" #3)

A A

533mm [21"] mm

[49 .2"

Top View 1:20 "F"

1000mm [39.4"]

R1250mm[R49.2"]

R2250mm [R88.6"]

View L

View F

View R

Top View

Detail Mount "F" Placement Column

Column Detail View M1:10

Bottom View

View B

Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 3

Medical Air Beacon Medaes Chemetron 1 Vacuum Beacon Medaes Chemetron 3

N2O Beacon Medaes Chemetron 1 AGSS/WAGD Beacon Medaes Chemetron 1

CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0

High Voltage Emergency Quad Power/UPS Hubbell Circuit / 2 1 Emergency Quad Power Hubbell Circuits / 3 & 4 2

Emergency Duplex Power Hubbell Circuit / 0 0 Normal Quad Power Hubbell Circuit / 0 0

Normal Duplex Power Hubbell Circuits / 5 & 6 2 Electric Brakes Hubbell Dual Brakes, Circuit / 1 1 Surgical Light Polaris Circuit / 0 0

Low Voltage Single Blank Provision Hubbell 2 Quad Blank Provision Hubbell 0

RJ45 Hubbell Cat.6 / 4 1 Draeger EQPM Connection Hubbell Circuit / 1 1

Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0

Misc. Other 0

1 / Ambia Column

Mount Location "F" Ambia C/L1000 (article-number: OP75478)

CATH Lab - Mount Location "F"

EHART

16DEC2021

1) Frame Color: Medical White W1100/S-0500-N

2) Service Plate Color: Anodized

Column

R

Q

P

2 3 4

N

M

L

K

J

5 6 7 8 9 10 11 12

H

G

F

E

D

C

B

A

2 3 4 5 6 7 8 9 10 11 12

13 14 15 16 1817 19 size A020 21 22 23

R

Q

P

N

M

L

K

J

13 14 15 16 1817 19

H

G

F

E

D

20 21 22 23

C

B

A

U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.

A ll rig ht s re se rv ed

R ef er to p ro te ct io n no tic e

IS

O

Type:

3D File:

System:

Date:

Name:

Author: Approved by: Scale:

Date:

Name:

View: 1:16 Others: See Details

3DFL - VA Iowa City 08FEB2022 1041

Quantity of Units: Project Name:

3D-Tool #:

Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):

VA Iowa City EP CATH IR

US United States120 60

Date: Author: Revision Text: Rev #:Notes:

Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented

Sample units for the definition of the US Country Standard.

Units partly against current Project Planning Specifications:

Couplings for Flammable Gases above Electrical Components (Sockets).

Gas Couplings Hoses.

1)

2)

PRESUBMITTALS ONLY

NOT FOR CONSTRUCTION

Rm Name, Rm # + Mount #:

Mount File #:

1st SubmissionEHART22DEC2021

331LBS/150KGMax. Load Ambia Column Data:

176LBS/80KG

22LBS/10KG

Shelf Max. Load (ea.)

Rails Max. Load (set)

Shelving Data:

01 - CATH LABs

DI EQPM UpdateEHart10JAN2022 02

UPS

Measurements to be Verified on Site. (V.I.F.)

Plan View N.T.S.

DI EQPM UpdateEHart08FEB2022 03

DI EQPM UpdateEHart10FEB2022 04

AutoCAD SHX Text o533mm m m

[5 2.

6" m m

[9 1.

1" m m

[9 8.

4" m m

[7 6.

7" m m

[1 m m

[9 5.

6" m m

[1

.8 m m

[5 5.

1" ]18 m m [7

4.

4"

] 22 m m

[8 9.

2" m m

[1 m m [2

.0

6m m [1

2.

4" m m [2

.0

4m m [8

.0 "]17

4m m

[6 .8

3m m [1

1.

1"

533mm [21.0"]

Max Weight Tensile Force Torque at Ceiling

600mmx600mm [23.5" x 23.5"]

Flat Cover Panel

View 1:16 "G" m m [2

.5

1450mm [57.1"]

(T ub e Le ng th

360°

330°

360°

340°

910mm

[35.8"]

0°

Polaris 600 incl. Control Panel + prepared Sterilizable Handle set E

360°

360°

1600mm [63.0"]

Lifting reduced 868mm/34.2" as shown (max. lifting of arm 1179mm/46.4")

1885N 1661Nm

Red denotes Structure by Others

This drawing is for illustrative purposes only.

Line Denotes Electrical/Supplied by Others.

Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others

Line Denotes Structure Supplied by Draeger.

Line Denotes Electrical/Supplied by TBD/or by Draeger.

5m m [2

5"

Lifting reduced 979mm/38.6" as shown (max. lifting of arm 1179mm/46.4")

910mm [35.8"]

360°

Finished Floor

Deck

Finished Ceiling

Rough-In Views 1:16Rough-In Notes:

1)

2)

3)

Deck (by Others)

Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)

Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)

Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)

4m m [6

.8

533mm [21"]

Top View

Elevation View

(See "Rough-In Notes" #3)

A A

533mm [21"]

1600mm [63.0"]

910mm [35.8"]

1450mm [57.1"]

Top View 1:20 "G"

R1600mm

[R63.0"]

R2510mm

[R98.8"]

910mm

[3 5.8"]

R1450mm [R57.1"]

R2360mm[R92.9"]

1 / P600 and 1/X-Ray Shield

Mount Location "G"

Polaris 600 (article-number: OP75084)

X-Ray Shield (article-number: OV2013-ML681-04 OT50001)

CATH Lab - Mount Location "G"

EHART

16DEC2021

R

Q

P

2 3 4

N

M

L

K

J

5 6 7 8 9 10 11 12

H

G

F

E

D

C

B

A

2 3 4 5 6 7 8 9 10 11 12

13 14 15 16 1817 19 size A020 21 22 23

R

Q

P

N

M

L

K

J

13 14 15 16 1817 19

H

G

F

E

D

20 21 22 23

C

B

A

U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.

A ll rig ht s re se rv ed

R ef er to p ro te ct io n no tic e

IS

O

Type:

Type:

3D File:

System:

Date:

Name:

Author: Approved by: Scale:

Date:

Name:

View: 1:16 Others: See Details

3DFL - VA Iowa City 08FEB2022 1041

Quantity of Units: Project Name:

3D-Tool #:

Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):

VA Iowa City EP CATH IR

US United States120 60

Date: Author: Revision Text: Rev #:Notes:

Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented

Sample units for the definition of the US Country Standard.

Units partly against current Project Planning Specifications:

Couplings for Flammable Gases above Electrical Components (Sockets).

Gas Couplings Hoses.

1)

2)

PRESUBMITTALS ONLY

NOT FOR CONSTRUCTION

Rm Name, Rm # + Mount #:

Mount File #:

1st SubmissionEHART22DEC2021

01 - CATH LABs

Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 0

Medical Air Beacon Medaes Chemetron 0 Vacuum Beacon Medaes Chemetron 0

N2O Beacon Medaes Chemetron 0 AGSS/WAGD Beacon Medaes Chemetron 0

CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0

High Voltage Emergency Quad Power Hubbell Circuit / 0 0 Emergency Duplex Power Hubbell Circuit / 0 0

Normal Quad Power Hubbell Circuit / 0 0 Normal Duplex Power Hubbell Circuit / 0 0

Electric Brakes Hubbell Dual Brakes, Circuit / 0 0 Surgical Light Polaris 600 Circuit / 1 1

Low Voltage Single Blank Provision Hubbell 0 Quad Blank Provision Hubbell 0

RJ45 Hubbell Cat.6 / 4 0 Draeger EQPM Connection Hubbell Circuit / 0 0

Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0

Misc. Other 0

DI EQPM UpdateEHart10JAN2022

DI EQPM UpdateEHart08FEB2022

Measurements to be Verified on Site. (V.I.F.)

Plan View N.T.S.

DI EQPM UpdateEHart10FEB2022 04

AutoCAD SHX Text o533mm m m

[1

.8 m m

[5 0.

1" ] 18 m m [7

1.

8"

4m m [6

.8

1m m [6

.3 m m

[1 m m

[1

0m m [1

1.

4"

533mm [21.0"]

600mmx600mm [23.5" x 23.5"]

Flat Cover Panel

View 1:16 "H" m m [2

.5

(T ub e Le ng th

360°

1400mm [55.1"]

1000mm[39.4"]Downward Movement 34"

Max Weight Tensile Force Torque at Ceiling

3728N 6532Nm

5m m [2

5"

Red denotes Structure by Others

This drawing is for illustrative purposes only.

Line Denotes Electrical/Supplied by Others.

Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others

Line Denotes Structure Supplied by Draeger.

Line Denotes Electrical/Supplied by TBD/or by Draeger.

Finished Floor

Deck

Finished Ceiling

Rough-In Views 1:16Rough-In Notes:

1)

2)

3)

Deck (by Others)

Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)

Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)

Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)

4m m [6

.8

533mm [21"]

Top View

Elevation View

(See "Rough-In Notes" #3)

A A

533mm [21"]

Top View 1:20 "H"

R1400mm[R55.1"]

1400mm [55.1"] m m

[3 9.

4"

R2400mm[R94.5"]

1 / GD60 Pro-Flex

Mount Location "H" GD60 Pro-Flex (article-number: DCR-2018-001701)

CATH Lab - Mount Location "H"

EHART

16DEC2021

R

Q

P

2 3 4

N

M

L

K

J

5 6 7 8 9 10 11 12

H

G

F

E

D

C

B

A

2 3 4 5 6 7 8 9 10 11 12

13 14 15 16 1817 19 size A020 21 22 23

R

Q

P

N

M

L

K

J

13 14 15 16 1817 19

H

G

F

E

D

20 21 22 23

C

B

A

U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.

A ll rig ht s re se rv ed

R ef er to p ro te ct io n no tic e

IS

O

Type:

3D File:

System:

Date:

Name:

Author: Approved by: Scale:

Date:

Name:

View: 1:16 Others: See Details

3DFL - VA Iowa City 08FEB2022 1041

Quantity of Units: Project Name:

3D-Tool #:

Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):

VA Iowa City EP CATH IR

US United States120 60

Date: Author: Revision Text: Rev #:Notes:

Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented

Sample units for the definition of the US Country Standard.

Units partly against current Project Planning Specifications:

Couplings for Flammable Gases above Electrical Components (Sockets).

Gas Couplings Hoses.

1)

2)

PRESUBMITTALS ONLY

NOT FOR CONSTRUCTION

Rm Name, Rm # + Mount #:

Mount File #:

1st SubmissionEHart22DEC2021

01 - CATH LABs

Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 0

Medical Air Beacon Medaes Chemetron 0 Vacuum Beacon Medaes Chemetron 0

N2O Beacon Medaes Chemetron 0 AGSS/WAGD Beacon Medaes Chemetron 0

CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0

High Voltage Emergency Quad Power Hubbell Circuit / 0 0 Emergency Duplex Power Hubbell Circuit / 0 0

Normal Quad Power Hubbell Circuit / 0 0 Normal Duplex Power Hubbell Circuit / 0 0

Electric Brakes Hubbell Dual Brakes, Circuit / 0 0 GD60 GD60 Circuit / 1 1

Low Voltage Single Blank Provision Hubbell 0 Quad Blank Provision Hubbell 0

RJ45 Hubbell Cat.6 / 4 0 Draeger EQPM Connection Hubbell Circuit / 0 0

Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0

Misc. Other 0

DI EQPM UpdateEHart10JAN2022

DI EQPM UpdateEHart08FEB2022

Measurements to be Verified on Site. (V.I.F.)

Plan View N.T.S.

DI EQPM UpdateEHart10FEB2022 04

AutoCAD SHX Text o533mm m m [2

.0 m m

[1

.5 m m

[1

.8

4m m [8

.0 m m

[5 5.

1" ]18 m m [7

4.

4"

] 22 m m

[8 9.

2" m m

[1

4m m [6

.8 m m [2

.5 m m

[1

3m m [1

1.

1"

Max Weight Tensile Force Torque at Ceiling

View 1:16 "I"

(T ub e Le ng th

360°

330°

360°

340°

910mm

[35.8"]

0°

Polaris 600 incl. Control Panel + prepared Sterilizable Handle set E

360°

1600mm [63.0"]

Lifting reduced 868mm/34.2" as shown (max. lifting of arm 1179mm/46.4")

1885N 1661Nm

533mm [21.0"]

600mmx600mm [23.5" x 23.5"]

Flat Cover Panel

Red denotes Structure by Others

This drawing is for illustrative purposes only.

Line Denotes Electrical/Supplied by Others.

Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others

Line Denotes Structure Supplied by Draeger.

Line Denotes Electrical/Supplied by TBD/or by Draeger.

5m m [2

5"

Finished Floor

Deck

Finished Ceiling

Rough-In Views 1:16Rough-In Notes:

1)

2)

3)

Deck (by Others)

Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)

Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)

Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)

4m m [6

.8

533mm [21"]

Top View

Elevation View

(See "Rough-In Notes" #3)

A A

533mm [21"]

1600mm [63.0"]

910mm [35.8"]

Top View 1:20 "I"

R1600mm

[R63.0"]

R2510mm

[R98.8"]

1 / P600

Mount Location "I" Polaris 600 (article-number: OP75084)

CATH Lab - Mount Location "I"

EHART

16DEC2021

R

Q

P

2 3 4

N

M

L

K

J

5 6 7 8 9 10 11 12

H

G

F

E

D

C

B

A

2 3 4 5 6 7 8 9 10 11 12

13 14 15 16 1817 19 size A020 21 22 23

R

Q

P

N

M

L

K

J

13 14 15 16 1817 19

H

G

F

E

D

20 21 22 23

C

B

A

U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.

A ll rig ht s re se rv ed

R ef er to p ro te ct io n no tic e

IS

O

Type:

3D File:

System:

Date:

Name:

Author: Approved by: Scale:

Date:

Name:

View: 1:16 Others: See Details

3DFL - VA Iowa City 08FEB2022 1041

Quantity of Units: Project Name:

3D-Tool #:

Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):

VA Iowa City EP CATH IR

US United States120 60

Date: Author: Revision Text: Rev #:Notes:

Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented

Sample units for the definition of the US Country Standard.

Units partly against current Project Planning Specifications:

Couplings for Flammable Gases above Electrical Components (Sockets).

Gas Couplings Hoses.

1)

2)

PRESUBMITTALS ONLY

NOT FOR CONSTRUCTION

Rm Name, Rm # + Mount #:

Mount File #:

1st SubmissionEHart22DEC2021

01 - CATH LABs

Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 0

Medical Air Beacon Medaes Chemetron 0 Vacuum Beacon Medaes Chemetron 0

N2O Beacon Medaes Chemetron 0 AGSS/WAGD Beacon Medaes Chemetron 0

CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0

High Voltage Emergency Quad Power Hubbell Circuit / 0 0 Emergency Duplex Power Hubbell Circuit / 0 0

Normal Quad Power Hubbell Circuit / 0 0 Normal Duplex Power Hubbell Circuit / 0 0

Electric Brakes Hubbell Dual Brakes, Circuit / 0 0 Surgical Light Polaris 600 Circuit / 1 1

Low Voltage Single Blank Provision Hubbell 0 Quad Blank Provision Hubbell 0

RJ45 Hubbell Cat.6 / 4 0 Draeger EQPM Connection Hubbell Circuit / 0 0

Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0

Misc. Other 0

DI EQPM UpdateEHart10JAN2022

DI EQPM UpdateEHart08FEB2022

Measurements to be Verified on Site. (V.I.F.)

Plan View N.T.S.

DI EQPM UpdateEHart10FEB2022 04

AutoCAD SHX Text o533mm

1m m [6

.3 m m

[9 5.

8"

0m m [1

1.

0"

4m m [6

.8 m m [2

.5 m m

[4 4.

1" ] 16 m m [6

5.

8"

7m m [3

2.

2" m m

[9 7.

0" m m

[7 6.

8" m m

[4 7.

6"

9m m [6

.3

4m m [8

.4

Finished Floor

Deck

Finished Ceiling

View 1:16 "J"

(T ub e Le ng th

1400mm [55.1"]

1000mm[39.4"]Downward Movement 34"

°m m

R1000

R1000

360°

330°

2m m [1

3.

1"

1250mm [49.2"]

1000mm [39.4"]

Movita plus (P os t L en gt h)

360°360°

(T ub e Le ng th

7930Nm180.0kg

Max Weight Tensile Force Torque at Ceiling

Max Weight

4218N

(GD60)

(Ambia) m m

[1 m m

[1

7m m

[3 1"

This drawing is for illustrative purposes only.

Line Denotes Regulator/Supplied by TBD/or by Draeger.

Line Denotes Electrical/Supplied by Others.

Line Denotes Medical Gases/Supplied by TBD/or by Draeger.

Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others

Line Denotes Structure Supplied by Draeger.

Line Denotes Electrical/Supplied by TBD/or by Draeger.

AIR AIRO

O

2 VAC VAC VACO

Red denotes Structure by Others

533mm [21.0"]

600mmx600mm [23.5" x 23.5"]

Flat Cover Panel

Detail Mount

"J" Column

406mm [16.0"]

Rough-In Views 1:16Rough-In Notes:

1)

2)

3)

Deck (by Others)

Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)

Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)

Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)

4m m [6

.8

533mm [21"]

Top View

Elevation View

(See "Rough-In Notes" #3)

A A

533mm [21"]

Top View N.T.S.

"J"

R1400mm[R55.1"]

1400mm [55.1"] m m

[3 9.

4"

R2400mm [R94.5"] mm

[49 .2"

1000mm [39.4"]

R1250mm[R49.2"]

R2250mm [R88.6"]

View L

View F

View R

Top View

Detail Mount "J" Placement Column

Column Detail View M1:10

Bottom View

View B

View L

View F

View R

View B

1 / Ambia Column and 1 / GD60 Pro-Flex

Mount Location "J"

Ambia (article-number: OP75478)

GD60 / Pro-Flex (article-number: DCR-2018-001701)

IR Lab - Mount Location "J"

EHART

16DEC2021

1) Frame Color: Medical White W1100/S-0500-N

2) Service Plate Color: Anodized

Column

R

Q

P

2 3 4

N

M

L

K

J

5 6 7 8 9 10 11 12

H

G

F

E

D

C

B

A

2 3 4 5 6 7 8 9 10 11 12

13 14 15 16 1817 19 size A020 21 22 23

R

Q

P

N

M

L

K

J

13 14 15 16 1817 19

H

G

F

E

D

20 21 22 23

C

B

A

U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.

A ll rig ht s re se rv ed

R ef er to p ro te ct io n no tic e

IS

O

Type:

Type:

3D File:

System:

Date:

Name:

Author: Approved by: Scale:

Date:

Name:

View: 1:16 Others: See Details

3DFL - VA Iowa City 08FEB2022 1041

Quantity of Units: Project Name:

3D-Tool #:

Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):

VA Iowa City EP CATH IR

US United States120 60

Date: Author: Revision Text: Rev #:Notes:

Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented

Sample units for the definition of the US Country Standard.

Units partly against current Project Planning Specifications:

Couplings for Flammable Gases above Electrical Components (Sockets).

Gas Couplings Hoses.

1)

2)

PRESUBMITTALS ONLY

NOT FOR CONSTRUCTION

Rm Name, Rm # + Mount #:

Mount File #:

1st SubmissionEHart22DEC2021

331LBS/150KGMax. Load Ambia Column Data:

176LBS/80KG

22LBS/10KG

Shelf Max. Load (ea.)

Rails Max. Load (set)

Shelving Data:

01 - CATH LABs

DI EQPM UpdateEHART10JAN2022 02

DI EQPM UpdateEHART08FEB2022 03

UPS

UPS

Measurements to be Verified on Site. (V.I.F.)

Plan View N.T.S.

DI EQPM UpdateEHart10FEB2022 04

Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 0

Medical Air Beacon Medaes Chemetron 2 Vacuum Beacon Medaes Chemetron 2

N2O Beacon Medaes Chemetron 0 AGSS/WAGD Beacon Medaes Chemetron 0

CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0

High Voltage Emergency Quad Power Hubbell Circuit / 2 1 Emergency Duplex Power/UPS Hubbell Circuits / 3 & 4 2

Emergency Duplex Power Hubbell Circuit / 5 1 Normal Quad Power Hubbell Circuit / 0 0

Normal Duplex Power Hubbell Circuits / 6 2 Electric Brakes Hubbell Dual Brakes, Circuit / 1 1

GD60 GD60 Circuit / 7 1

Low Voltage Single Blank Provision Hubbell 4 Quad Blank Provision Hubbell 1

RJ45 Hubbell Cat.6 / 4 1 Draeger EQPM Connection Hubbell Circuit / 1 1

Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0

Misc. Other 0

AutoCAD SHX Text o533mm m [2

.0 m m

[9 5.

5" m m

[9 6.

8"

4m m [8

.0 m m

[4 9.

1" ]17 m m [6

8.

4"

] 21 m m

[8 3.

2"

4m m [6

.8

Polaris 600 incl. Control Panel + prepared Sterilizable Handle set E

3m m [1

1.

1"

View 1:16 "K" m m [2

.5

(T ub e Le ng th

360°

330°

360°

340°

910mm

[35.8"]

0°

360°

1600mm [63.0"]

Lifting reduced 868mm/34.2" as shown (max. lifting of arm 1179mm/46.4")

1885N 1661Nm

533mm [21.0"]

Max Weight Tensile Force Torque at Ceiling

600mmx600mm [23.5" x 23.5"]

Flat Cover Panel

Red denotes Structure by Others

This drawing is for illustrative purposes only.

Line Denotes Electrical/Supplied by Others.

Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others

Line Denotes Structure Supplied by Draeger.

Line Denotes Electrical/Supplied by TBD/or by Draeger.

Finished Floor

Deck

Finished Ceiling m m

[1 m m

[1

7m m

[3 1"

Rough-In Views 1:16Rough-In Notes:

1)

2)

3)

Deck (by Others)

Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)

Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)

Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)

4m m [6

.8

533mm [21"]

Top View

Elevation View

(See "Rough-In Notes" #3)

A A

533mm [21"]

1600mm [63.0"]

910mm [35.8"]

Top View 1:20 "K"

R1600mm

[R63.0"]

R2510mm

[R98.8"]

1 / P600

Mount Location "K" Polaris 600 (article-number: OP75084)

IR Lab - Mount Location "K"

EHART

16DEC2021

R

Q

P

2 3 4

N

M

L

K

J

5 6 7 8 9 10 11 12

H

G

F

E

D

C

B

A

2 3 4 5 6 7 8 9 10 11 12

13 14 15 16 1817 19 size A020 21 22 23

R

Q

P

N

M

L

K

J

13 14 15 16 1817 19

H

G

F

E

D

20 21 22 23

C

B

A

U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.

A ll rig ht s re se rv ed

R ef er to p ro te ct io n no tic e

IS

O

Type:

3D File:

System:

Date:

Name:

Author: Approved by: Scale:

Date:

Name:

View: 1:16 Others: See Details

3DFL - VA Iowa City 08FEB2022 1041

Quantity of Units: Project Name:

3D-Tool #:

Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):

VA Iowa City EP CATH IR

US United States120 60

Date: Author: Revision Text: Rev #:Notes:

Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented

Sample units for the definition of the US Country Standard.

Units partly against current Project Planning Specifications:

Couplings for Flammable Gases above Electrical Components (Sockets).

Gas Couplings Hoses.

1)

2)

PRESUBMITTALS ONLY

NOT FOR CONSTRUCTION

Rm Name, Rm # + Mount #:

Mount File #:

1st SubmissionEHart22DEC2021

01 - CATH LABs

Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 0

Medical Air Beacon Medaes Chemetron 0 Vacuum Beacon Medaes Chemetron 0

N2O Beacon Medaes Chemetron 0 AGSS/WAGD Beacon Medaes Chemetron 0

CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0

High Voltage Emergency Quad Power Hubbell Circuit / 0 0 Emergency Duplex Power Hubbell Circuit / 0 0

Normal Quad Power Hubbell Circuit / 0 0 Normal Duplex Power Hubbell Circuit / 0 0

Electric Brakes Hubbell Dual Brakes, Circuit / 0 0 Surgical Light Polaris 600 Circuit / 1 1

Low Voltage Single Blank Provision Hubbell 0 Quad Blank Provision Hubbell 0

RJ45 Hubbell Cat.6 / 4 0 Draeger EQPM Connection Hubbell Circuit / 0 0

Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0

Misc. Other 0

DI EQPM UpdateEHART10JAN2022 02

DI EQPM UpdateEHART08FEB2022 03

Measurements to be Verified on Site. (V.I.F.)

Plan View N.T.S.

DI EQPM UpdateEHart10FEB2022 04

AutoCAD SHX Text o533mm

°m m

R1000

R1000

4m m [8

.8

7m m [3

2.

2" m m

[9 7.

0" m m

[7 6.

8"

360° m m

[4 7.

6"

Detail Mount

"L" Column

View 1:16 "L"

4m m [6

.8 m m [2

.5

130kg

330°

9m m [6

.3

2m m [1

3.

1"

1250mm [49.2"]

1000mm [39.4"]

(T ub e Le ng th

(P os t L en gt h)

360°

3728N 6532Nm

533mm [21.0"]

Max Weight Tensile Force Torque at Ceiling

600mmx600mm [23.5" x 23.5"]

Flat Cover Panel

This drawing is for illustrative purposes only.

Line Denotes Regulator/Supplied by TBD/or by Draeger.

Line Denotes Electrical/Supplied by Others.

Line Denotes Medical Gases/Supplied by TBD/or by Draeger.

Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others

Line Denotes Structure Supplied by Draeger.

Line Denotes Electrical/Supplied by TBD/or by Draeger.

AIR AIRO

O

2 VAC VAC VACO

Red denotes Structure by Others

Movita plus

406mm [16.0"]

Finished Floor

Deck

Finished Ceiling m m [1 m m

[1

7m m

[3 1"

Rough-In Views 1:16Rough-In Notes:

1)

2)

3)

Deck (by Others)

Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)

Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)

Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)

4m m [6

.8

533mm [21"]

Top View

Elevation View

(See "Rough-In Notes" #3)

A A

533mm [21"] mm

[49 .2"

Top View 1:20 "L"

1000mm [39.4"]

R1250mm[R49.2"]

R2250mm [R88.6"]

View L

View F

View R

Top View

Detail Mount "L" Placement Column

Column Detail View M1:10

Bottom View

View B

View L

View F

View R

View B

Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 3

Medical Air Beacon Medaes Chemetron 1 Vacuum Beacon Medaes Chemetron 3

N2O Beacon Medaes Chemetron 1 AGSS/WAGD Beacon Medaes Chemetron 1

CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0

High Voltage Emergency Quad Power/UPS Hubbell Circuit / 2 1 Emergency Quad Power Hubbell Circuits / 3 & 4 2

Emergency Duplex Power Hubbell Circuit / 0 0 Normal Quad Power Hubbell Circuit / 0 0

Normal Duplex Power Hubbell Circuits / 5 & 6 2 Electric Brakes Hubbell Dual Brakes, Circuit / 1 1 Surgical Light Polaris Circuit / 0 0

Low Voltage Single Blank Provision Hubbell 2 Quad Blank Provision Hubbell 0

RJ45 Hubbell Cat.6 / 4 1 Draeger EQPM Connection Hubbell Circuit / 1 1

Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0

Misc. Other 0

1 / Ambia Column

Mount Location "L" Ambia C/L1000 (article-number: OP75478)

CATH Lab - Mount Location "L"

EHART

16DEC2021

1) Frame Color: Medical White W1100/S-0500-N

2) Service Plate Color: Anodized

Column

R

Q

P

2 3 4

N

M

L

K

J

5 6 7 8 9 10 11 12

H

G

F

E

D

C

B

A

2 3 4 5 6 7 8 9 10 11 12

13 14 15 16 1817 19 size A020 21 22 23

R

Q

P

N

M

L

K

J

13 14 15 16 1817 19

H

G

F

E

D

20 21 22 23

C

B

A

U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.

A ll rig ht s re se rv ed

R ef er to p ro te ct io n no tic e

IS

O

Type:

3D File:

System:

Date:

Name:

Author: Approved by: Scale:

Date:

Name:

View: 1:16 Others: See Details

3DFL - VA Iowa City 08FEB2022 1041

Quantity of Units: Project Name:

3D-Tool #:

Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):

VA Iowa City EP CATH IR

US United States120 60

Date: Author: Revision Text: Rev #:Notes:

Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented

Sample units for the definition of the US Country Standard.

Units partly against current Project Planning Specifications:

Couplings for Flammable Gases above Electrical Components (Sockets).

Gas Couplings Hoses.

1)

2)

PRESUBMITTALS ONLY

NOT FOR CONSTRUCTION

Rm Name, Rm # + Mount #:

Mount File #:

1st SubmissionEHart22DEC2021

331LBS/150KGMax. Load Ambia Column Data:

176LBS/80KG

22LBS/10KG

Shelf Max. Load (ea.)

Rails Max. Load (set)

Shelving Data:

01 - CATH LABs

DI EQPM UpdateEHART10JAN2022 02

DI EQPM UpdateEHART08FEB2022 03

UPS

Measurements to be Verified on Site. (V.I.F.)

Plan View N.T.S.

DI EQPM UpdateEHart10FEB2022 04

AutoCAD SHX Text o533mm m [2

.0 m m

[4 6.

6" m m

[8 5.

1"

6m m [1

2.

4" m m [9

2.

4" m m

[7 0.

7" m m [2

.0 m m

[8 9.

6" m m

[9 6.

8"

4m m [8

.0 m m

[4 9.

1" ]17 m m [6

8.

4"

] 21 m m

[8 3.

2"

4m m [6

.8 m m

[1 m m

[1

3m m [1

1.

1"

Max Weight Tensile Force Torque at Ceiling

View 1:16 "M"

1450mm [57.1"]

(T ub e Le ng th

360°

330°

360°

340°

910mm

[35.8"]

0°

Polaris 600 incl. Control Panel + prepared Sterilizable Handle set E

360°

360°

1600mm [63.0"]

Lifting reduced 868mm/34.2" as shown (max. lifting of arm 1179mm/46.4")

1885N 1661Nm m m [2

.5

7m m [3

1"

533mm [21.0"]

600mmx600mm [23.5" x 23.5"]

Flat Cover Panel

Red denotes Structure by Others

This drawing is for illustrative purposes only.

Line Denotes Electrical/Supplied by Others.

Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others

Line Denotes Structure Supplied by Draeger.

Line Denotes Electrical/Supplied by TBD/or by Draeger.

Lifting reduced 979mm/38.6" as shown (max. lifting of arm 1179mm/46.4")

910mm [35.8"]

360°

Finished Floor

Deck

Finished Ceiling

Rough-In Views 1:16Rough-In Notes:

1)

2)

3)

Deck (by Others)

Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)

Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)

Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)

4m m [6

.8

533mm [21"]

Top View

Elevation View

(See "Rough-In Notes" #3)

A A

533mm [21"]

1600mm [63.0"]

910mm [35.8"]

1450mm [57.1"]

Top View 1:20 "M"

R1600mm

[R63.0"]

R2510mm

[R98.8"]

910mm

[3 5.8"]

R1450mm [R57.1"]

R2360mm[R92.9"]

1 / P600 and 1/X-Ray Shield

Mount Location "M"

Polaris 600 (article-number: OP75084)

X-Ray Shield (article-number: OV2013-ML681-04 OT50001)

IR Lab - Mount Location "M"

EHART

16DEC2021

R

Q

P

2 3 4

N

M

L

K

J

5 6 7 8 9 10 11 12

H

G

F

E

D

C

B

A

2 3 4 5 6 7 8 9 10 11 12

13 14 15 16 1817 19 size A020 21 22 23

R

Q

P

N

M

L

K

J

13 14 15 16 1817 19

H

G

F

E

D

20 21 22 23

C

B

A

U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.

A ll rig ht s re se rv ed

R ef er to p ro te ct io n no tic e

IS

O

Type:

Type:

3D File:

System:

Date:

Name:

Author: Approved by: Scale:

Date:

Name:

View: 1:16 Others: See Details

3DFL - VA Iowa City 08FEB2022 1041

Quantity of Units: Project Name:

3D-Tool #:

Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):

VA Iowa City EP CATH IR

US United States120 60

Date: Author: Revision Text: Rev #:Notes:

Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented

Sample units for the definition of the US Country Standard.

Units partly against current Project Planning Specifications:

Couplings for Flammable Gases above Electrical…

This is the start of the file's text. The full file is on GovTribe.

File details come from the government source that posted it. Updated .