Attachment P - 2W IR CATH EP_Draeger-Boom Install_Site Specific.pdf
PDF 9 MB Posted
- Solicitation number
- 36C26323R0002
About this file
This document package includes technical drawings and specifications for a medical facility renovation project to remodel areas for interventional radiology, cardiac catheterization, and endoscopy at a Department of Veterans Affairs medical center. The project involves installing new medical equipment mounting structures, electrical and medical gas supply lines, and lighting in multiple procedure rooms. Key equipment to be installed includes C-arms, surgical lights, booms, and supply columns from Draeger and Polaris. The technical drawings provide detailed mounting and connection requirements for this equipment in areas 2W151, 2W152, 2W153, as well as electrical panel schedules and medical gas outlet specifications by room.
View the file
Other files
Show all 25
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
TITLE:
Date:
Name:
Rev04
AUTHOR:
12-16-21
EH
02-10-22
SHEET:
3D-REF
DESCRIPTIONS3D REF
EP Lab 2W151
CATH Lab 2W152
IR Lab 2W153
°m m
R1000
R1000 m m
[1 m m
[1
4m m [8
.8 m m
[5 0.
2" m m
[9 7.
0" m m
[7 6.
8"
7m m [3
1"
533mm [21.0"]
Max Weight Tensile Force Torque at Ceiling
360°
0m m [2
9.
5"
Detail Mount
"A" Column
600mmx600mm [23.5" x 23.5"]
Flat Cover Panel
View 1:16 "A"
4m m [6
.8 m m [2
.5
130.1kg
330°
9m m [6
.3
2m m [1
3.
1"
1250mm [49.2"]
1000mm [39.4"]
(T ub e Le ng th
Movita plus (P os t L en gt h)
360°
3728N 6532Nm
This drawing is for illustrative purposes only.
Line Denotes Regulator/Supplied by TBD/or by Draeger.
Line Denotes Electrical/Supplied by Others.
Line Denotes Medical Gases/Supplied by TBD/or by Draeger.
Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others
Line Denotes Structure Supplied by Draeger.
Line Denotes Electrical/Supplied by TBD/or by Draeger.
AIR AIRO
O
2 VAC VAC VACO
Red denotes Structure by Others
301mm [11.9"]
Finished Floor
Deck
Finished Ceiling
Rough-In Views 1:16Rough-In Notes:
1)
2)
3)
Deck (by Others)
Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)
Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)
Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)
4m m [6
.8
533mm [21"]
Top View
Elevation View
(See "Rough-In Notes" #3)
A A
533mm [21"] mm
[49 .2"
Top View 1:20 "A"
1000mm [39.4"]
R1250mm[R49.2"]
R2250mm [R88.6"]
View L
View F
View R
Top View
Detail Mount "A" Placement Column
Column Detail View M1:10
Bottom View
View B
1 / Ambia Column
Mount Location "A" Ambia C/M 500 (article-number: OP75471)
EP Lab - Mount Location "A"
EHART
16DEC2021
1) Frame Color: Medical White W1100/S-0500-N
2) Service Plate Color: Anodized
Column
R
Q
P
2 3 4
N
M
L
K
J
5 6 7 8 9 10 11 12
H
G
F
E
D
C
B
A
2 3 4 5 6 7 8 9 10 11 12
13 14 15 16 1817 19 size A020 21 22 23
R
Q
P
N
M
L
K
J
13 14 15 16 1817 19
H
G
F
E
D
20 21 22 23
C
B
A
U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.
A ll rig ht s re se rv ed
R ef er to p ro te ct io n no tic e
IS
O
Type:
3D File:
System:
Date:
Name:
Author: Approved by: Scale:
Date:
Name:
View: 1:16 Others: See Details
3DFL - VA Iowa City 08FEB2022 1041
Quantity of Units: Project Name:
3D-Tool #:
Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):
VA Iowa City EP CATH IR
US United States120 60
Date: Author: Revision Text: Rev #:Notes:
Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented
Sample units for the definition of the US Country Standard.
Units partly against current Project Planning Specifications:
Couplings for Flammable Gases above Electrical Components (Sockets).
Gas Couplings Hoses.
1)
2)
PRESUBMITTALS ONLY
NOT FOR CONSTRUCTION
Rm Name, Rm # + Mount #:
Mount File #:
1st SubmissionEHart22DEC2021
287LBS/130KGMax. Load Ambia Column Data:
176LBS/80KG
22LBS/10KG
Shelf Max. Load (ea.)
Rails Max. Load (set)
Shelving Data:
01 - CATH LABs
Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 2
Medical Air Beacon Medaes Chemetron 0 Vacuum Beacon Medaes Chemetron 2
N2O Beacon Medaes Chemetron 0 AGSS/WAGD Beacon Medaes Chemetron 0
CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0
High Voltage Emergency Quad Power/UPS Hubbell Circuits / 2 1 Emergency Quad Power Hubbell Circuits / 3 1
Emergency Duplex Power Hubbell Circuit / 0 0 Normal Quad Power Hubbell Circuit / 0 0
Normal Duplex Power Hubbell Circuits / 4 & 5 2 Electric Brakes Hubbell Dual Brakes, Circuit / 1 1 Surgical Light Polaris Circuit / 0 0
Low Voltage Single Blank Provision Hubbell 1 Triple Blank Provision Hubbell 0
RJ45 Hubbell Cat.6 / 4 1 Draeger EQPM Connection Hubbell Circuit / 1 1
Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0
Misc. Other 0
DI EQPM UpdateEHart10JAN2022
DI EQPM UpdateEHart08FEB2022
UPS
Measurements to be Verified on Site. (V.I.F.)
Plan View N.T.S.
DI EQPM UpdateEHart10FEB2022 04
AutoCAD SHX Text o533mm
AutoCAD SHX Text o533mm m m
[1 m m
[1
°m m
R1000
R1000
4m m [8
.8 m m
[6 6.
6" m m
[9 7.
0" m m
[7 6.
8"
360°
4m m [1
3.
1"
Detail Mount
"B" Head
4m m [6
.8 m m [2
.5
330°
9m m [6
.3
2m m [1
3.
1"
1250mm [49.2"]
1000mm [39.4"]
(T ub e Le ng th
Movita plus (P os t L en gt h)
877mm [34.5"]
360°
3728N 6532Nm130.1kg
7m m [3
1"
533mm [21.0"]
Max Weight Tensile Force Torque at Ceiling
600mmx600mm [23.5" x 23.5"]
Flat Cover Panel
This drawing is for illustrative purposes only.
Line Denotes Regulator/Supplied by TBD/or by Draeger.
Line Denotes Electrical/Supplied by Others.
Line Denotes Medical Gases/Supplied by TBD/or by Draeger.
Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others
Line Denotes Structure Supplied by Draeger.
Line Denotes Electrical/Supplied by TBD/or by Draeger.
AIR AIRO
O
2 VAC VAC VACO
Red denotes Structure by OthersView 1:16 "B"
Finished Floor
Deck
Finished Ceiling
Rough-In Views 1:16Rough-In Notes:
1)
2)
3)
Deck (by Others)
Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)
Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)
Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)
4m m [6
.8
533mm [21"]
Top View
Elevation View
(See "Rough-In Notes" #3)
A A
533mm [21"] mm
[49 .2"
Top View 1:20 "B"
1000mm [39.4"]
R1250mm[R49.2"]
R2250mm [R88.6"]
View L
View F
View R
Top View
Detail Mount "B" Placement Column
Column Detail View M1:10
Bottom View
View B
View U
Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 2
Medical Air Beacon Medaes Chemetron 1 Vacuum Beacon Medaes Chemetron 2
N2O Beacon Medaes Chemetron 1 AGSS/WAGD Beacon Medaes Chemetron 1
CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0
High Voltage Emergency Quad Power Hubbell Circuit / 2 1 Emergency Duplex Power/UPS Hubbell Circuits / 3 & 4 2
Normal Quad Power Hubbell Circuit / 0 0 Normal Duplex Power Hubbell Circuits / 5 & 6 2
Electric Brakes Hubbell Dual Brakes, Circuit / 1 1 Surgical Light Polaris Circuit / 0 0
Low Voltage Single Blank Provision Hubbell 2 Triple Blank Provision Hubbell 0
RJ45 Hubbell Cat.6 / 4 1 Draeger EQPM Connection Hubbell Circuit / 1 1
Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0
Misc. Other 0
1 / Ambia Head
Mount Location "B" Ambia H/M 750 (article-number: OP75484)
EP Lab - Mount Location "B"
EHART
16DEC2021
1) Equipment Color: Medical White W1100/S-0500-N
2) Frame Color: Medical White W1100/S-0500-N
3) Service Plate Color: Anodized
R
Q
P
2 3 4
N
M
L
K
J
5 6 7 8 9 10 11 12
H
G
F
E
D
C
B
A
2 3 4 5 6 7 8 9 10 11 12
13 14 15 16 1817 19 size A020 21 22 23
R
Q
P
N
M
L
K
J
13 14 15 16 1817 19
H
G
F
E
D
20 21 22 23
C
B
A
U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.
A ll rig ht s re se rv ed
R ef er to p ro te ct io n no tic e
IS
O
Type:
3D File:
System:
Date:
Name:
Author: Approved by: Scale:
Date:
Name:
View: 1:16 Others: See Details
3DFL - VA Iowa City 08FEB2022 1041
Quantity of Units: Project Name:
3D-Tool #:
Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):
VA Iowa City EP CATH IR
US United States120 60
Date: Author: Revision Text: Rev #:Notes:
Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented
Sample units for the definition of the US Country Standard.
Units partly against current Project Planning Specifications:
Couplings for Flammable Gases above Electrical Components (Sockets).
Gas Couplings Hoses.
1)
2)
PRESUBMITTALS ONLY
NOT FOR CONSTRUCTION
Rm Name, Rm # + Mount #:
Mount File #:
1st SubmissionEHart22DEC2021
287LBS/130KGMax. Load Ambia Head Data:
176LBS/80KG
22LBS/10KG
Shelf Max. Load (ea.)
Rails Max. Load (set)
Shelving Data:
01 - CATH LABs
Head
View 1:16 "B"
DI EQPM UpdateEHart10JAN2022 02
UPS
UPS
Measurements to be Verified on Site. (V.I.F.)
Plan View N.T.S.
DI EQPM UpdateEHart08FEB2022 03
DI EQPM UpdateEHart10FEB2022 04
AutoCAD SHX Text o533mm m m
[4 6.
6" m m
[8 5.
1" m m
[8 9.
6" m m
[9 6.
8" m m
[9 2.
4" m m
[7 0.
7" m m
[4 9.
1" ]17 m m [6
8.
4"
] 21 m m
[8 3.
2" m m [2
.0 m m [2
.0
4m m [8
.0
6m m [1
2.
4" m m
[1 m m
[1
4m m [6
.8
3m m [1
1.
1"
View 1:16 "C" m m [2
.5
1450mm [57.1"]
(T ub e Le ng th
360°
330°
360°
340°
910mm
[35.8"]
0°
Polaris 600 incl. Control Panel + prepared Sterilizable Handle set E
360°
360°
1600mm [63.0"]
Lifting reduced 868mm/34.2" as shown (max. lifting of arm 1179mm/46.4")
7m m [3
1"
533mm [21.0"]
Max Weight Tensile Force Torque at Ceiling
600mmx600mm [23.5" x 23.5"]
Flat Cover Panel
Red denotes Structure by Others
This drawing is for illustrative purposes only.
Line Denotes Electrical/Supplied by Others.
Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others
Line Denotes Structure Supplied by Draeger.
Line Denotes Electrical/Supplied by TBD/or by Draeger.
1885N 1661Nm
Lifting reduced 979mm/38.6" as shown (max. lifting of arm 1179mm/46.4")
910mm [35.8"]
360°
Finished Floor
Deck
Finished Ceiling
Rough-In Views 1:16Rough-In Notes:
1)
2)
3)
Deck (by Others)
Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)
Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)
Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)
4m m [6
.8
533mm [21"]
Top View
Elevation View
(See "Rough-In Notes" #3)
A A
533mm [21"]
1600mm [63.0"]
910mm [35.8"]
1450mm [57.1"]
Top View 1:20 "C"
R1600mm
[R63.0"]
R2510mm
[R98.8"]
910mm
[3 5.8"]
R1450mm [R57.1"]
R2360mm[R92.9"]
1 / P600 and 1/X-Ray Shield
Mount Location "C"
Polaris 600 (article-number: OP75084)
X-Ray Shield (article-number: OV2013-ML681-04 OT50001)
EP Lab - Mount Location "C"
EHART
16DEC2021
R
Q
P
2 3 4
N
M
L
K
J
5 6 7 8 9 10 11 12
H
G
F
E
D
C
B
A
2 3 4 5 6 7 8 9 10 11 12
13 14 15 16 1817 19 size A020 21 22 23
R
Q
P
N
M
L
K
J
13 14 15 16 1817 19
H
G
F
E
D
20 21 22 23
C
B
A
U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.
A ll rig ht s re se rv ed
R ef er to p ro te ct io n no tic e
IS
O
Type:
Type:
3D File:
System:
Date:
Name:
Author: Approved by: Scale:
Date:
Name:
View: 1:16 Others: See Details
3DFL - VA Iowa City 08FEB2022 1041
Quantity of Units: Project Name:
3D-Tool #:
Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):
VA Iowa City EP CATH IR
US United States120 60
Date: Author: Revision Text: Rev #:Notes:
Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented
Sample units for the definition of the US Country Standard.
Units partly against current Project Planning Specifications:
Couplings for Flammable Gases above Electrical Components (Sockets).
Gas Couplings Hoses.
1)
2)
PRESUBMITTALS ONLY
NOT FOR CONSTRUCTION
Rm Name, Rm # + Mount #:
Mount File #:
1st SubmissionEHart22DEC2021
01 - CATH LABs
Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 0
Medical Air Beacon Medaes Chemetron 0 Vacuum Beacon Medaes Chemetron 0
N2O Beacon Medaes Chemetron 0 AGSS/WAGD Beacon Medaes Chemetron 0
CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0
High Voltage Emergency Quad Power Hubbell Circuit / 0 0 Emergency Duplex Power Hubbell Circuit / 0 0
Normal Quad Power Hubbell Circuit / 0 0 Normal Duplex Power Hubbell Circuit / 0 0
Electric Brakes Hubbell Dual Brakes, Circuit / 0 0 Surgical Light Polaris 600 Circuit / 1 1
Low Voltage Single Blank Provision Hubbell 0 Quad Blank Provision Hubbell 0
RJ45 Hubbell Cat.6 / 4 0 Draeger EQPM Connection Hubbell Circuit / 0 0
Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0
Misc. Other 0
DI EQPM UpdateEHart10JAN2022 02
Measurements to be Verified on Site. (V.I.F.)
Plan View N.T.S.
DI EQPM UpdateEHart08FEB2022 03
DI EQPM UpdateEHart10FEB2022 04
AutoCAD SHX Text o533mm m m
[9 5.
8" m m
[4 4.
1" ] 16 m m [6
5.
8"
7m m [3
2.
2" m m
[9 7.
0" m m
[7 6.
8" m m
[1 m m
[1
7m m
[3 1"
] 17 4m m [6
.8 m m [2
.5
1m m [6
.3 m m
[4 7.
6"
9m m [6
.3
6m m [8
.5
533mm [21.0"]
600mmx600mm [23.5" x 23.5"]
Flat Cover Panel
View 1:16 "D"
(by Draeger)
(T ub e Le ng th
1400mm [55.1"]
1000mm[39.4"]Downward Movement 34"
°m m
R1000
R1000
360°
Detail Mount
"D" Column
330°
2m m [1
3.
1"
1250mm [49.2"]
1000mm [39.4"]
Movita plus (P os t L en gt h)
360°360°
0m m [1
1.
0"
(T ub e
Le ng th
7930Nm180.0kg
Max Weight Tensile Force Torque at Ceiling
Max Weight
4218N
(GD60)
(Ambia)
406mm [16.0"]
This drawing is for illustrative purposes only.
Line Denotes Regulator/Supplied by TBD/or by Draeger.
Line Denotes Electrical/Supplied by Others.
Line Denotes Medical Gases/Supplied by TBD/or by Draeger.
Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others
Line Denotes Structure Supplied by Draeger.
Line Denotes Electrical/Supplied by TBD/or by Draeger.
AIR AIRO
O
2 VAC VAC VACO
Red denotes Structure by Others
Finished Floor
Deck
Finished Ceiling
Rough-In Views 1:16Rough-In Notes:
1)
2)
3)
Deck (by Others)
Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)
Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)
Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)
4m m [6
.8
533mm [21"]
Top View
Elevation View
(See "Rough-In Notes" #3)
A A
533mm [21"]
Top View N.T.S.
"D"
R1400mm[R55.1"]
1400mm [55.1"] m m
[3 9.
4"
R2400mm [R94.5"] mm
[49 .2"
1000mm [39.4"]
R1250mm[R49.2"]
R2250mm [R88.6"]
View L
View F
View R
Top View
Detail Mount "D" Placement Column
Column Detail View M1:10
Bottom View
View B
Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 0
Medical Air Beacon Medaes Chemetron 2 Vacuum Beacon Medaes Chemetron 2
N2O Beacon Medaes Chemetron 0 AGSS/WAGD Beacon Medaes Chemetron 0
CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0
High Voltage Emergency Quad Power Hubbell Circuit / 2 1 Emergency Duplex Power/UPS Hubbell Circuits / 3 & 4 2
Emergency Duplex Power Hubbell Circuit / 5 1 Normal Quad Power Hubbell Circuit / 0 0
Normal Duplex Power Hubbell Circuits / 6 2 Electric Brakes Hubbell Dual Brakes, Circuit / 1 1
GD60 GD60 Circuit / 7 1
Low Voltage Single Blank Provision Hubbell 4 Quad Blank Provision Hubbell 1
RJ45 Hubbell Cat.6 / 4 1 Draeger EQPM Connection Hubbell Circuit / 1 1
Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0
Misc. Other 0
1 / Ambia Column and 1 / GD60 Pro-Flex
Mount Location "D"
Ambia (article-number: OP75478)
GD60 Pro-Flex (article-number: DCR-2018-001701)
EP Lab - Mount Location "D"
EHART
16DEC2021
1) Frame Color: Medical White W1100/S-0500-N
2) Service Plate Color: Anodized
Column
R
Q
P
2 3 4
N
M
L
K
J
5 6 7 8 9 10 11 12
H
G
F
E
D
C
B
A
2 3 4 5 6 7 8 9 10 11 12
13 14 15 16 1817 19 size A020 21 22 23
R
Q
P
N
M
L
K
J
13 14 15 16 1817 19
H
G
F
E
D
20 21 22 23
C
B
A
U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.
A ll rig ht s re se rv ed
R ef er to p ro te ct io n no tic e
IS
O
Type:
Type:
3D File:
System:
Date:
Name:
Author: Approved by: Scale:
Date:
Name:
View: 1:16 Others: See Details
3DFL - VA Iowa City 08FEB2022 1041
Quantity of Units: Project Name:
3D-Tool #:
Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):
VA Iowa City EP CATH IR
US United States120 60
Date: Author: Revision Text: Rev #:Notes:
Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented
Sample units for the definition of the US Country Standard.
Units partly against current Project Planning Specifications:
Couplings for Flammable Gases above Electrical Components (Sockets).
Gas Couplings Hoses.
1)
2)
PRESUBMITTALS ONLY
NOT FOR CONSTRUCTION
Rm Name, Rm # + Mount #:
Mount File #:
1st SubmissionEHart22DEC2021
331LBS/150KGMax. Load Ambia Column Data:
176LBS/80KG
22LBS/10KG
Shelf Max. Load (ea.)
Rails Max. Load (set)
Shelving Data:
01 - CATH LABs
DI EQPM UpdateEHart10JAN2022 02
UPS
UPS
Measurements to be Verified on Site. (V.I.F.)
Plan View N.T.S.
DI EQPM UpdateEHart08FEB2022 03
DI EQPM UpdateEHart10FEB2022 04
AutoCAD SHX Text o533mm m m
[1 m m
[1 m m [2
.0 m m
[8 9.
6" m m
[9 6.
8"
4m m [8
.0 m m
[4 9.
1" m m
[8 3.
2"
4m m [6
.8 m m
[6 8.
4"
3m m [1
1.
1"
7m m [3
1"
View 1:16 "E" m m [2
.5
1450mm [57.1"]
(T ub e Le ng th
360°
330°
360°
340°
910mm
[35.8"]
0°
Polaris 600 incl. Control Panel + prepared Sterilizable Handle set E
360°
360°
1600mm [63.0"]
Lifting reduced 868mm/34.2" as shown (max. lifting of arm 1179mm/46.4")
1885N 1661Nm
533mm [21.0"]
Max Weight Tensile Force Torque at Ceiling
600mmx600mm [23.5" x 23.5"]
Flat Cover Panel
Red denotes Structure by Others
This drawing is for illustrative purposes only.
Line Denotes Electrical/Supplied by Others.
Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others
Line Denotes Structure Supplied by Draeger.
Line Denotes Electrical/Supplied by TBD/or by Draeger.
m m
[4 6.
6" m m
[8 5.
1"
6m m [1
2.
4" m m [9
2.
4" m m
[7 0.
7"
Lifting reduced 979mm/38.6" as shown (max. lifting of arm 1179mm/46.4")
910mm [35.8"]
360° m m [2
.0
Finished Floor
Deck
Finished Ceiling
Rough-In Views 1:16Rough-In Notes:
1)
2)
3)
Deck (by Others)
Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)
Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)
Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)
4m m [6
.8
533mm [21"]
Top View
Elevation View
(See "Rough-In Notes" #3)
A A
533mm [21"]
1600mm [63.0"]
910mm [35.8"]
1450mm [57.1"]
Top View 1:20 "E"
R1600mm
[R63.0"]
R2510mm
[R98.8"]
910mm
[3 5.8"]
R1450mm [R57.1"]
R2360mm[R92.9"]
1 / P600 and 1/X-Ray Shield
Mount Location "E"
Polaris 600 (article-number: OP75084)
X-Ray Shield (article-number: OV2013-ML681-04 OT50001)
EP Lab - Mount Location "E"
EHART
16DEC2021
R
Q
P
2 3 4
N
M
L
K
J
5 6 7 8 9 10 11 12
H
G
F
E
D
C
B
A
2 3 4 5 6 7 8 9 10 11 12
13 14 15 16 1817 19 size A020 21 22 23
R
Q
P
N
M
L
K
J
13 14 15 16 1817 19
H
G
F
E
D
20 21 22 23
C
B
A
U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.
A ll rig ht s re se rv ed
R ef er to p ro te ct io n no tic e
IS
O
Type:
Type:
3D File:
System:
Date:
Name:
Author: Approved by: Scale:
Date:
Name:
View: 1:16 Others: See Details
3DFL - VA IOWA CITY 10JAN2022 0945
Quantity of Units: Project Name:
3D-Tool #:
Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):
VA Iowa City EP CATH IR
US United States120 60
Date: Author: Revision Text: Rev #:Notes:
Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented
Sample units for the definition of the US Country Standard.
Units partly against current Project Planning Specifications:
Couplings for Flammable Gases above Electrical Components (Sockets).
Gas Couplings Hoses.
1)
2)
PRESUBMITTALS ONLY
NOT FOR CONSTRUCTION
Rm Name, Rm # + Mount #:
Mount File #:
1st SubmissionEHart22DEC2021
01 - CATH LABs
Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 0
Medical Air Beacon Medaes Chemetron 0 Vacuum Beacon Medaes Chemetron 0
N2O Beacon Medaes Chemetron 0 AGSS/WAGD Beacon Medaes Chemetron 0
CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0
High Voltage Emergency Quad Power Hubbell Circuit / 0 0 Emergency Duplex Power Hubbell Circuit / 0 0
Normal Quad Power Hubbell Circuit / 0 0 Normal Duplex Power Hubbell Circuit / 0 0
Electric Brakes Hubbell Dual Brakes, Circuit / 0 0 Surgical Light Polaris 600 Circuit / 1 1
Low Voltage Single Blank Provision Hubbell 0 Quad Blank Provision Hubbell 0
RJ45 Hubbell Cat.6 / 4 0 Draeger EQPM Connection Hubbell Circuit / 0 0
Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0
Misc. Other 0
DI EQPM UpdateEHart10JAN2022 02
Measurements to be Verified on Site. (V.I.F.)
Plan View N.T.S.
DI EQPM UpdateEHart08FEB2022 03
DI EQPM UpdateEHart10FEB2022 04
AutoCAD SHX Text o533mm
9m m [3
8.
2" m m
[1
.0 m m
[8 2.
8" m m
[1 m m
[1
°m m
R1000
R1000
4m m [8
.8
360° m m
[4 7.
6"
Detail Mount
"F" Column
View 1:16 "F"
(by Draeger)
4m m [6
.8 m m [2
.5
330°
9m m [6
.3
2m m [1
3.
1"
1250mm [49.2"]
1000mm [39.4"]
(T ub e Le ng th
Movita plus (P os t L en gt h)
360°
533mm [21.0"]
Max Weight Tensile Force Torque at Ceiling
600mmx600mm [23.5" x 23.5"]
Flat Cover Panel
This drawing is for illustrative purposes only.
Line Denotes Regulator/Supplied by TBD/or by Draeger.
Line Denotes Electrical/Supplied by Others.
Line Denotes Medical Gases/Supplied by TBD/or by Draeger.
Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others
Line Denotes Structure Supplied by Draeger.
Line Denotes Electrical/Supplied by TBD/or by Draeger.
AIR AIRO
O
2 VAC VAC VACO
Red denotes Structure by Others
406mm [16.0"]
3728N 6532Nm130.1kg
5m m [2
5"
Finished Floor
Deck
Finished Ceiling
Rough-In Views 1:16Rough-In Notes:
1)
2)
3)
Deck (by Others)
Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)
Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)
Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)
4m m [6
.8
533mm [21"]
Top View
Elevation View
(See "Rough-In Notes" #3)
A A
533mm [21"] mm
[49 .2"
Top View 1:20 "F"
1000mm [39.4"]
R1250mm[R49.2"]
R2250mm [R88.6"]
View L
View F
View R
Top View
Detail Mount "F" Placement Column
Column Detail View M1:10
Bottom View
View B
Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 3
Medical Air Beacon Medaes Chemetron 1 Vacuum Beacon Medaes Chemetron 3
N2O Beacon Medaes Chemetron 1 AGSS/WAGD Beacon Medaes Chemetron 1
CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0
High Voltage Emergency Quad Power/UPS Hubbell Circuit / 2 1 Emergency Quad Power Hubbell Circuits / 3 & 4 2
Emergency Duplex Power Hubbell Circuit / 0 0 Normal Quad Power Hubbell Circuit / 0 0
Normal Duplex Power Hubbell Circuits / 5 & 6 2 Electric Brakes Hubbell Dual Brakes, Circuit / 1 1 Surgical Light Polaris Circuit / 0 0
Low Voltage Single Blank Provision Hubbell 2 Quad Blank Provision Hubbell 0
RJ45 Hubbell Cat.6 / 4 1 Draeger EQPM Connection Hubbell Circuit / 1 1
Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0
Misc. Other 0
1 / Ambia Column
Mount Location "F" Ambia C/L1000 (article-number: OP75478)
CATH Lab - Mount Location "F"
EHART
16DEC2021
1) Frame Color: Medical White W1100/S-0500-N
2) Service Plate Color: Anodized
Column
R
Q
P
2 3 4
N
M
L
K
J
5 6 7 8 9 10 11 12
H
G
F
E
D
C
B
A
2 3 4 5 6 7 8 9 10 11 12
13 14 15 16 1817 19 size A020 21 22 23
R
Q
P
N
M
L
K
J
13 14 15 16 1817 19
H
G
F
E
D
20 21 22 23
C
B
A
U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.
A ll rig ht s re se rv ed
R ef er to p ro te ct io n no tic e
IS
O
Type:
3D File:
System:
Date:
Name:
Author: Approved by: Scale:
Date:
Name:
View: 1:16 Others: See Details
3DFL - VA Iowa City 08FEB2022 1041
Quantity of Units: Project Name:
3D-Tool #:
Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):
VA Iowa City EP CATH IR
US United States120 60
Date: Author: Revision Text: Rev #:Notes:
Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented
Sample units for the definition of the US Country Standard.
Units partly against current Project Planning Specifications:
Couplings for Flammable Gases above Electrical Components (Sockets).
Gas Couplings Hoses.
1)
2)
PRESUBMITTALS ONLY
NOT FOR CONSTRUCTION
Rm Name, Rm # + Mount #:
Mount File #:
1st SubmissionEHART22DEC2021
331LBS/150KGMax. Load Ambia Column Data:
176LBS/80KG
22LBS/10KG
Shelf Max. Load (ea.)
Rails Max. Load (set)
Shelving Data:
01 - CATH LABs
DI EQPM UpdateEHart10JAN2022 02
UPS
Measurements to be Verified on Site. (V.I.F.)
Plan View N.T.S.
DI EQPM UpdateEHart08FEB2022 03
DI EQPM UpdateEHart10FEB2022 04
AutoCAD SHX Text o533mm m m
[5 2.
6" m m
[9 1.
1" m m
[9 8.
4" m m
[7 6.
7" m m
[1 m m
[9 5.
6" m m
[1
.8 m m
[5 5.
1" ]18 m m [7
4.
4"
] 22 m m
[8 9.
2" m m
[1 m m [2
.0
6m m [1
2.
4" m m [2
.0
4m m [8
.0 "]17
4m m
[6 .8
3m m [1
1.
1"
533mm [21.0"]
Max Weight Tensile Force Torque at Ceiling
600mmx600mm [23.5" x 23.5"]
Flat Cover Panel
View 1:16 "G" m m [2
.5
1450mm [57.1"]
(T ub e Le ng th
360°
330°
360°
340°
910mm
[35.8"]
0°
Polaris 600 incl. Control Panel + prepared Sterilizable Handle set E
360°
360°
1600mm [63.0"]
Lifting reduced 868mm/34.2" as shown (max. lifting of arm 1179mm/46.4")
1885N 1661Nm
Red denotes Structure by Others
This drawing is for illustrative purposes only.
Line Denotes Electrical/Supplied by Others.
Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others
Line Denotes Structure Supplied by Draeger.
Line Denotes Electrical/Supplied by TBD/or by Draeger.
5m m [2
5"
Lifting reduced 979mm/38.6" as shown (max. lifting of arm 1179mm/46.4")
910mm [35.8"]
360°
Finished Floor
Deck
Finished Ceiling
Rough-In Views 1:16Rough-In Notes:
1)
2)
3)
Deck (by Others)
Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)
Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)
Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)
4m m [6
.8
533mm [21"]
Top View
Elevation View
(See "Rough-In Notes" #3)
A A
533mm [21"]
1600mm [63.0"]
910mm [35.8"]
1450mm [57.1"]
Top View 1:20 "G"
R1600mm
[R63.0"]
R2510mm
[R98.8"]
910mm
[3 5.8"]
R1450mm [R57.1"]
R2360mm[R92.9"]
1 / P600 and 1/X-Ray Shield
Mount Location "G"
Polaris 600 (article-number: OP75084)
X-Ray Shield (article-number: OV2013-ML681-04 OT50001)
CATH Lab - Mount Location "G"
EHART
16DEC2021
R
Q
P
2 3 4
N
M
L
K
J
5 6 7 8 9 10 11 12
H
G
F
E
D
C
B
A
2 3 4 5 6 7 8 9 10 11 12
13 14 15 16 1817 19 size A020 21 22 23
R
Q
P
N
M
L
K
J
13 14 15 16 1817 19
H
G
F
E
D
20 21 22 23
C
B
A
U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.
A ll rig ht s re se rv ed
R ef er to p ro te ct io n no tic e
IS
O
Type:
Type:
3D File:
System:
Date:
Name:
Author: Approved by: Scale:
Date:
Name:
View: 1:16 Others: See Details
3DFL - VA Iowa City 08FEB2022 1041
Quantity of Units: Project Name:
3D-Tool #:
Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):
VA Iowa City EP CATH IR
US United States120 60
Date: Author: Revision Text: Rev #:Notes:
Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented
Sample units for the definition of the US Country Standard.
Units partly against current Project Planning Specifications:
Couplings for Flammable Gases above Electrical Components (Sockets).
Gas Couplings Hoses.
1)
2)
PRESUBMITTALS ONLY
NOT FOR CONSTRUCTION
Rm Name, Rm # + Mount #:
Mount File #:
1st SubmissionEHART22DEC2021
01 - CATH LABs
Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 0
Medical Air Beacon Medaes Chemetron 0 Vacuum Beacon Medaes Chemetron 0
N2O Beacon Medaes Chemetron 0 AGSS/WAGD Beacon Medaes Chemetron 0
CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0
High Voltage Emergency Quad Power Hubbell Circuit / 0 0 Emergency Duplex Power Hubbell Circuit / 0 0
Normal Quad Power Hubbell Circuit / 0 0 Normal Duplex Power Hubbell Circuit / 0 0
Electric Brakes Hubbell Dual Brakes, Circuit / 0 0 Surgical Light Polaris 600 Circuit / 1 1
Low Voltage Single Blank Provision Hubbell 0 Quad Blank Provision Hubbell 0
RJ45 Hubbell Cat.6 / 4 0 Draeger EQPM Connection Hubbell Circuit / 0 0
Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0
Misc. Other 0
DI EQPM UpdateEHart10JAN2022
DI EQPM UpdateEHart08FEB2022
Measurements to be Verified on Site. (V.I.F.)
Plan View N.T.S.
DI EQPM UpdateEHart10FEB2022 04
AutoCAD SHX Text o533mm m m
[1
.8 m m
[5 0.
1" ] 18 m m [7
1.
8"
4m m [6
.8
1m m [6
.3 m m
[1 m m
[1
0m m [1
1.
4"
533mm [21.0"]
600mmx600mm [23.5" x 23.5"]
Flat Cover Panel
View 1:16 "H" m m [2
.5
(T ub e Le ng th
360°
1400mm [55.1"]
1000mm[39.4"]Downward Movement 34"
Max Weight Tensile Force Torque at Ceiling
3728N 6532Nm
5m m [2
5"
Red denotes Structure by Others
This drawing is for illustrative purposes only.
Line Denotes Electrical/Supplied by Others.
Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others
Line Denotes Structure Supplied by Draeger.
Line Denotes Electrical/Supplied by TBD/or by Draeger.
Finished Floor
Deck
Finished Ceiling
Rough-In Views 1:16Rough-In Notes:
1)
2)
3)
Deck (by Others)
Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)
Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)
Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)
4m m [6
.8
533mm [21"]
Top View
Elevation View
(See "Rough-In Notes" #3)
A A
533mm [21"]
Top View 1:20 "H"
R1400mm[R55.1"]
1400mm [55.1"] m m
[3 9.
4"
R2400mm[R94.5"]
1 / GD60 Pro-Flex
Mount Location "H" GD60 Pro-Flex (article-number: DCR-2018-001701)
CATH Lab - Mount Location "H"
EHART
16DEC2021
R
Q
P
2 3 4
N
M
L
K
J
5 6 7 8 9 10 11 12
H
G
F
E
D
C
B
A
2 3 4 5 6 7 8 9 10 11 12
13 14 15 16 1817 19 size A020 21 22 23
R
Q
P
N
M
L
K
J
13 14 15 16 1817 19
H
G
F
E
D
20 21 22 23
C
B
A
U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.
A ll rig ht s re se rv ed
R ef er to p ro te ct io n no tic e
IS
O
Type:
3D File:
System:
Date:
Name:
Author: Approved by: Scale:
Date:
Name:
View: 1:16 Others: See Details
3DFL - VA Iowa City 08FEB2022 1041
Quantity of Units: Project Name:
3D-Tool #:
Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):
VA Iowa City EP CATH IR
US United States120 60
Date: Author: Revision Text: Rev #:Notes:
Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented
Sample units for the definition of the US Country Standard.
Units partly against current Project Planning Specifications:
Couplings for Flammable Gases above Electrical Components (Sockets).
Gas Couplings Hoses.
1)
2)
PRESUBMITTALS ONLY
NOT FOR CONSTRUCTION
Rm Name, Rm # + Mount #:
Mount File #:
1st SubmissionEHart22DEC2021
01 - CATH LABs
Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 0
Medical Air Beacon Medaes Chemetron 0 Vacuum Beacon Medaes Chemetron 0
N2O Beacon Medaes Chemetron 0 AGSS/WAGD Beacon Medaes Chemetron 0
CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0
High Voltage Emergency Quad Power Hubbell Circuit / 0 0 Emergency Duplex Power Hubbell Circuit / 0 0
Normal Quad Power Hubbell Circuit / 0 0 Normal Duplex Power Hubbell Circuit / 0 0
Electric Brakes Hubbell Dual Brakes, Circuit / 0 0 GD60 GD60 Circuit / 1 1
Low Voltage Single Blank Provision Hubbell 0 Quad Blank Provision Hubbell 0
RJ45 Hubbell Cat.6 / 4 0 Draeger EQPM Connection Hubbell Circuit / 0 0
Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0
Misc. Other 0
DI EQPM UpdateEHart10JAN2022
DI EQPM UpdateEHart08FEB2022
Measurements to be Verified on Site. (V.I.F.)
Plan View N.T.S.
DI EQPM UpdateEHart10FEB2022 04
AutoCAD SHX Text o533mm m m [2
.0 m m
[1
.5 m m
[1
.8
4m m [8
.0 m m
[5 5.
1" ]18 m m [7
4.
4"
] 22 m m
[8 9.
2" m m
[1
4m m [6
.8 m m [2
.5 m m
[1
3m m [1
1.
1"
Max Weight Tensile Force Torque at Ceiling
View 1:16 "I"
(T ub e Le ng th
360°
330°
360°
340°
910mm
[35.8"]
0°
Polaris 600 incl. Control Panel + prepared Sterilizable Handle set E
360°
1600mm [63.0"]
Lifting reduced 868mm/34.2" as shown (max. lifting of arm 1179mm/46.4")
1885N 1661Nm
533mm [21.0"]
600mmx600mm [23.5" x 23.5"]
Flat Cover Panel
Red denotes Structure by Others
This drawing is for illustrative purposes only.
Line Denotes Electrical/Supplied by Others.
Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others
Line Denotes Structure Supplied by Draeger.
Line Denotes Electrical/Supplied by TBD/or by Draeger.
5m m [2
5"
Finished Floor
Deck
Finished Ceiling
Rough-In Views 1:16Rough-In Notes:
1)
2)
3)
Deck (by Others)
Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)
Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)
Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)
4m m [6
.8
533mm [21"]
Top View
Elevation View
(See "Rough-In Notes" #3)
A A
533mm [21"]
1600mm [63.0"]
910mm [35.8"]
Top View 1:20 "I"
R1600mm
[R63.0"]
R2510mm
[R98.8"]
1 / P600
Mount Location "I" Polaris 600 (article-number: OP75084)
CATH Lab - Mount Location "I"
EHART
16DEC2021
R
Q
P
2 3 4
N
M
L
K
J
5 6 7 8 9 10 11 12
H
G
F
E
D
C
B
A
2 3 4 5 6 7 8 9 10 11 12
13 14 15 16 1817 19 size A020 21 22 23
R
Q
P
N
M
L
K
J
13 14 15 16 1817 19
H
G
F
E
D
20 21 22 23
C
B
A
U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.
A ll rig ht s re se rv ed
R ef er to p ro te ct io n no tic e
IS
O
Type:
3D File:
System:
Date:
Name:
Author: Approved by: Scale:
Date:
Name:
View: 1:16 Others: See Details
3DFL - VA Iowa City 08FEB2022 1041
Quantity of Units: Project Name:
3D-Tool #:
Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):
VA Iowa City EP CATH IR
US United States120 60
Date: Author: Revision Text: Rev #:Notes:
Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented
Sample units for the definition of the US Country Standard.
Units partly against current Project Planning Specifications:
Couplings for Flammable Gases above Electrical Components (Sockets).
Gas Couplings Hoses.
1)
2)
PRESUBMITTALS ONLY
NOT FOR CONSTRUCTION
Rm Name, Rm # + Mount #:
Mount File #:
1st SubmissionEHart22DEC2021
01 - CATH LABs
Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 0
Medical Air Beacon Medaes Chemetron 0 Vacuum Beacon Medaes Chemetron 0
N2O Beacon Medaes Chemetron 0 AGSS/WAGD Beacon Medaes Chemetron 0
CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0
High Voltage Emergency Quad Power Hubbell Circuit / 0 0 Emergency Duplex Power Hubbell Circuit / 0 0
Normal Quad Power Hubbell Circuit / 0 0 Normal Duplex Power Hubbell Circuit / 0 0
Electric Brakes Hubbell Dual Brakes, Circuit / 0 0 Surgical Light Polaris 600 Circuit / 1 1
Low Voltage Single Blank Provision Hubbell 0 Quad Blank Provision Hubbell 0
RJ45 Hubbell Cat.6 / 4 0 Draeger EQPM Connection Hubbell Circuit / 0 0
Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0
Misc. Other 0
DI EQPM UpdateEHart10JAN2022
DI EQPM UpdateEHart08FEB2022
Measurements to be Verified on Site. (V.I.F.)
Plan View N.T.S.
DI EQPM UpdateEHart10FEB2022 04
AutoCAD SHX Text o533mm
1m m [6
.3 m m
[9 5.
8"
0m m [1
1.
0"
4m m [6
.8 m m [2
.5 m m
[4 4.
1" ] 16 m m [6
5.
8"
7m m [3
2.
2" m m
[9 7.
0" m m
[7 6.
8" m m
[4 7.
6"
9m m [6
.3
4m m [8
.4
Finished Floor
Deck
Finished Ceiling
View 1:16 "J"
(T ub e Le ng th
1400mm [55.1"]
1000mm[39.4"]Downward Movement 34"
°m m
R1000
R1000
360°
330°
2m m [1
3.
1"
1250mm [49.2"]
1000mm [39.4"]
Movita plus (P os t L en gt h)
360°360°
(T ub e Le ng th
7930Nm180.0kg
Max Weight Tensile Force Torque at Ceiling
Max Weight
4218N
(GD60)
(Ambia) m m
[1 m m
[1
7m m
[3 1"
This drawing is for illustrative purposes only.
Line Denotes Regulator/Supplied by TBD/or by Draeger.
Line Denotes Electrical/Supplied by Others.
Line Denotes Medical Gases/Supplied by TBD/or by Draeger.
Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others
Line Denotes Structure Supplied by Draeger.
Line Denotes Electrical/Supplied by TBD/or by Draeger.
AIR AIRO
O
2 VAC VAC VACO
Red denotes Structure by Others
533mm [21.0"]
600mmx600mm [23.5" x 23.5"]
Flat Cover Panel
Detail Mount
"J" Column
406mm [16.0"]
Rough-In Views 1:16Rough-In Notes:
1)
2)
3)
Deck (by Others)
Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)
Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)
Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)
4m m [6
.8
533mm [21"]
Top View
Elevation View
(See "Rough-In Notes" #3)
A A
533mm [21"]
Top View N.T.S.
"J"
R1400mm[R55.1"]
1400mm [55.1"] m m
[3 9.
4"
R2400mm [R94.5"] mm
[49 .2"
1000mm [39.4"]
R1250mm[R49.2"]
R2250mm [R88.6"]
View L
View F
View R
Top View
Detail Mount "J" Placement Column
Column Detail View M1:10
Bottom View
View B
View L
View F
View R
View B
1 / Ambia Column and 1 / GD60 Pro-Flex
Mount Location "J"
Ambia (article-number: OP75478)
GD60 / Pro-Flex (article-number: DCR-2018-001701)
IR Lab - Mount Location "J"
EHART
16DEC2021
1) Frame Color: Medical White W1100/S-0500-N
2) Service Plate Color: Anodized
Column
R
Q
P
2 3 4
N
M
L
K
J
5 6 7 8 9 10 11 12
H
G
F
E
D
C
B
A
2 3 4 5 6 7 8 9 10 11 12
13 14 15 16 1817 19 size A020 21 22 23
R
Q
P
N
M
L
K
J
13 14 15 16 1817 19
H
G
F
E
D
20 21 22 23
C
B
A
U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.
A ll rig ht s re se rv ed
R ef er to p ro te ct io n no tic e
IS
O
Type:
Type:
3D File:
System:
Date:
Name:
Author: Approved by: Scale:
Date:
Name:
View: 1:16 Others: See Details
3DFL - VA Iowa City 08FEB2022 1041
Quantity of Units: Project Name:
3D-Tool #:
Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):
VA Iowa City EP CATH IR
US United States120 60
Date: Author: Revision Text: Rev #:Notes:
Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented
Sample units for the definition of the US Country Standard.
Units partly against current Project Planning Specifications:
Couplings for Flammable Gases above Electrical Components (Sockets).
Gas Couplings Hoses.
1)
2)
PRESUBMITTALS ONLY
NOT FOR CONSTRUCTION
Rm Name, Rm # + Mount #:
Mount File #:
1st SubmissionEHart22DEC2021
331LBS/150KGMax. Load Ambia Column Data:
176LBS/80KG
22LBS/10KG
Shelf Max. Load (ea.)
Rails Max. Load (set)
Shelving Data:
01 - CATH LABs
DI EQPM UpdateEHART10JAN2022 02
DI EQPM UpdateEHART08FEB2022 03
UPS
UPS
Measurements to be Verified on Site. (V.I.F.)
Plan View N.T.S.
DI EQPM UpdateEHart10FEB2022 04
Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 0
Medical Air Beacon Medaes Chemetron 2 Vacuum Beacon Medaes Chemetron 2
N2O Beacon Medaes Chemetron 0 AGSS/WAGD Beacon Medaes Chemetron 0
CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0
High Voltage Emergency Quad Power Hubbell Circuit / 2 1 Emergency Duplex Power/UPS Hubbell Circuits / 3 & 4 2
Emergency Duplex Power Hubbell Circuit / 5 1 Normal Quad Power Hubbell Circuit / 0 0
Normal Duplex Power Hubbell Circuits / 6 2 Electric Brakes Hubbell Dual Brakes, Circuit / 1 1
GD60 GD60 Circuit / 7 1
Low Voltage Single Blank Provision Hubbell 4 Quad Blank Provision Hubbell 1
RJ45 Hubbell Cat.6 / 4 1 Draeger EQPM Connection Hubbell Circuit / 1 1
Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0
Misc. Other 0
AutoCAD SHX Text o533mm m [2
.0 m m
[9 5.
5" m m
[9 6.
8"
4m m [8
.0 m m
[4 9.
1" ]17 m m [6
8.
4"
] 21 m m
[8 3.
2"
4m m [6
.8
Polaris 600 incl. Control Panel + prepared Sterilizable Handle set E
3m m [1
1.
1"
View 1:16 "K" m m [2
.5
(T ub e Le ng th
360°
330°
360°
340°
910mm
[35.8"]
0°
360°
1600mm [63.0"]
Lifting reduced 868mm/34.2" as shown (max. lifting of arm 1179mm/46.4")
1885N 1661Nm
533mm [21.0"]
Max Weight Tensile Force Torque at Ceiling
600mmx600mm [23.5" x 23.5"]
Flat Cover Panel
Red denotes Structure by Others
This drawing is for illustrative purposes only.
Line Denotes Electrical/Supplied by Others.
Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others
Line Denotes Structure Supplied by Draeger.
Line Denotes Electrical/Supplied by TBD/or by Draeger.
Finished Floor
Deck
Finished Ceiling m m
[1 m m
[1
7m m
[3 1"
Rough-In Views 1:16Rough-In Notes:
1)
2)
3)
Deck (by Others)
Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)
Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)
Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)
4m m [6
.8
533mm [21"]
Top View
Elevation View
(See "Rough-In Notes" #3)
A A
533mm [21"]
1600mm [63.0"]
910mm [35.8"]
Top View 1:20 "K"
R1600mm
[R63.0"]
R2510mm
[R98.8"]
1 / P600
Mount Location "K" Polaris 600 (article-number: OP75084)
IR Lab - Mount Location "K"
EHART
16DEC2021
R
Q
P
2 3 4
N
M
L
K
J
5 6 7 8 9 10 11 12
H
G
F
E
D
C
B
A
2 3 4 5 6 7 8 9 10 11 12
13 14 15 16 1817 19 size A020 21 22 23
R
Q
P
N
M
L
K
J
13 14 15 16 1817 19
H
G
F
E
D
20 21 22 23
C
B
A
U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.
A ll rig ht s re se rv ed
R ef er to p ro te ct io n no tic e
IS
O
Type:
3D File:
System:
Date:
Name:
Author: Approved by: Scale:
Date:
Name:
View: 1:16 Others: See Details
3DFL - VA Iowa City 08FEB2022 1041
Quantity of Units: Project Name:
3D-Tool #:
Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):
VA Iowa City EP CATH IR
US United States120 60
Date: Author: Revision Text: Rev #:Notes:
Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented
Sample units for the definition of the US Country Standard.
Units partly against current Project Planning Specifications:
Couplings for Flammable Gases above Electrical Components (Sockets).
Gas Couplings Hoses.
1)
2)
PRESUBMITTALS ONLY
NOT FOR CONSTRUCTION
Rm Name, Rm # + Mount #:
Mount File #:
1st SubmissionEHart22DEC2021
01 - CATH LABs
Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 0
Medical Air Beacon Medaes Chemetron 0 Vacuum Beacon Medaes Chemetron 0
N2O Beacon Medaes Chemetron 0 AGSS/WAGD Beacon Medaes Chemetron 0
CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0
High Voltage Emergency Quad Power Hubbell Circuit / 0 0 Emergency Duplex Power Hubbell Circuit / 0 0
Normal Quad Power Hubbell Circuit / 0 0 Normal Duplex Power Hubbell Circuit / 0 0
Electric Brakes Hubbell Dual Brakes, Circuit / 0 0 Surgical Light Polaris 600 Circuit / 1 1
Low Voltage Single Blank Provision Hubbell 0 Quad Blank Provision Hubbell 0
RJ45 Hubbell Cat.6 / 4 0 Draeger EQPM Connection Hubbell Circuit / 0 0
Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0
Misc. Other 0
DI EQPM UpdateEHART10JAN2022 02
DI EQPM UpdateEHART08FEB2022 03
Measurements to be Verified on Site. (V.I.F.)
Plan View N.T.S.
DI EQPM UpdateEHart10FEB2022 04
AutoCAD SHX Text o533mm
°m m
R1000
R1000
4m m [8
.8
7m m [3
2.
2" m m
[9 7.
0" m m
[7 6.
8"
360° m m
[4 7.
6"
Detail Mount
"L" Column
View 1:16 "L"
4m m [6
.8 m m [2
.5
130kg
330°
9m m [6
.3
2m m [1
3.
1"
1250mm [49.2"]
1000mm [39.4"]
(T ub e Le ng th
(P os t L en gt h)
360°
3728N 6532Nm
533mm [21.0"]
Max Weight Tensile Force Torque at Ceiling
600mmx600mm [23.5" x 23.5"]
Flat Cover Panel
This drawing is for illustrative purposes only.
Line Denotes Regulator/Supplied by TBD/or by Draeger.
Line Denotes Electrical/Supplied by Others.
Line Denotes Medical Gases/Supplied by TBD/or by Draeger.
Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others
Line Denotes Structure Supplied by Draeger.
Line Denotes Electrical/Supplied by TBD/or by Draeger.
AIR AIRO
O
2 VAC VAC VACO
Red denotes Structure by Others
Movita plus
406mm [16.0"]
Finished Floor
Deck
Finished Ceiling m m [1 m m
[1
7m m
[3 1"
Rough-In Views 1:16Rough-In Notes:
1)
2)
3)
Deck (by Others)
Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)
Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)
Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)
4m m [6
.8
533mm [21"]
Top View
Elevation View
(See "Rough-In Notes" #3)
A A
533mm [21"] mm
[49 .2"
Top View 1:20 "L"
1000mm [39.4"]
R1250mm[R49.2"]
R2250mm [R88.6"]
View L
View F
View R
Top View
Detail Mount "L" Placement Column
Column Detail View M1:10
Bottom View
View B
View L
View F
View R
View B
Services Type Manufacturer Style Qty Gas Oxygen/O2 Beacon Medaes Chemetron 3
Medical Air Beacon Medaes Chemetron 1 Vacuum Beacon Medaes Chemetron 3
N2O Beacon Medaes Chemetron 1 AGSS/WAGD Beacon Medaes Chemetron 1
CO2 Beacon Medaes Chemetron 0 N2 Beacon Medaes Chemetron 0
High Voltage Emergency Quad Power/UPS Hubbell Circuit / 2 1 Emergency Quad Power Hubbell Circuits / 3 & 4 2
Emergency Duplex Power Hubbell Circuit / 0 0 Normal Quad Power Hubbell Circuit / 0 0
Normal Duplex Power Hubbell Circuits / 5 & 6 2 Electric Brakes Hubbell Dual Brakes, Circuit / 1 1 Surgical Light Polaris Circuit / 0 0
Low Voltage Single Blank Provision Hubbell 2 Quad Blank Provision Hubbell 0
RJ45 Hubbell Cat.6 / 4 1 Draeger EQPM Connection Hubbell Circuit / 1 1
Light Control Hubbell Circuit / 0 0 RGBW Light Control Hubbell Circuit / 0 0
Misc. Other 0
1 / Ambia Column
Mount Location "L" Ambia C/L1000 (article-number: OP75478)
CATH Lab - Mount Location "L"
EHART
16DEC2021
1) Frame Color: Medical White W1100/S-0500-N
2) Service Plate Color: Anodized
Column
R
Q
P
2 3 4
N
M
L
K
J
5 6 7 8 9 10 11 12
H
G
F
E
D
C
B
A
2 3 4 5 6 7 8 9 10 11 12
13 14 15 16 1817 19 size A020 21 22 23
R
Q
P
N
M
L
K
J
13 14 15 16 1817 19
H
G
F
E
D
20 21 22 23
C
B
A
U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.
A ll rig ht s re se rv ed
R ef er to p ro te ct io n no tic e
IS
O
Type:
3D File:
System:
Date:
Name:
Author: Approved by: Scale:
Date:
Name:
View: 1:16 Others: See Details
3DFL - VA Iowa City 08FEB2022 1041
Quantity of Units: Project Name:
3D-Tool #:
Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):
VA Iowa City EP CATH IR
US United States120 60
Date: Author: Revision Text: Rev #:Notes:
Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented
Sample units for the definition of the US Country Standard.
Units partly against current Project Planning Specifications:
Couplings for Flammable Gases above Electrical Components (Sockets).
Gas Couplings Hoses.
1)
2)
PRESUBMITTALS ONLY
NOT FOR CONSTRUCTION
Rm Name, Rm # + Mount #:
Mount File #:
1st SubmissionEHart22DEC2021
331LBS/150KGMax. Load Ambia Column Data:
176LBS/80KG
22LBS/10KG
Shelf Max. Load (ea.)
Rails Max. Load (set)
Shelving Data:
01 - CATH LABs
DI EQPM UpdateEHART10JAN2022 02
DI EQPM UpdateEHART08FEB2022 03
UPS
Measurements to be Verified on Site. (V.I.F.)
Plan View N.T.S.
DI EQPM UpdateEHart10FEB2022 04
AutoCAD SHX Text o533mm m [2
.0 m m
[4 6.
6" m m
[8 5.
1"
6m m [1
2.
4" m m [9
2.
4" m m
[7 0.
7" m m [2
.0 m m
[8 9.
6" m m
[9 6.
8"
4m m [8
.0 m m
[4 9.
1" ]17 m m [6
8.
4"
] 21 m m
[8 3.
2"
4m m [6
.8 m m
[1 m m
[1
3m m [1
1.
1"
Max Weight Tensile Force Torque at Ceiling
View 1:16 "M"
1450mm [57.1"]
(T ub e Le ng th
360°
330°
360°
340°
910mm
[35.8"]
0°
Polaris 600 incl. Control Panel + prepared Sterilizable Handle set E
360°
360°
1600mm [63.0"]
Lifting reduced 868mm/34.2" as shown (max. lifting of arm 1179mm/46.4")
1885N 1661Nm m m [2
.5
7m m [3
1"
533mm [21.0"]
600mmx600mm [23.5" x 23.5"]
Flat Cover Panel
Red denotes Structure by Others
This drawing is for illustrative purposes only.
Line Denotes Electrical/Supplied by Others.
Line Denotes Support Structure/Supplied by Others Line Denotes Main Building Feeds from Source/Supplied by Others
Line Denotes Structure Supplied by Draeger.
Line Denotes Electrical/Supplied by TBD/or by Draeger.
Lifting reduced 979mm/38.6" as shown (max. lifting of arm 1179mm/46.4")
910mm [35.8"]
360°
Finished Floor
Deck
Finished Ceiling
Rough-In Views 1:16Rough-In Notes:
1)
2)
3)
Deck (by Others)
Stuctural Framework and Bracing, see "Rough-In Notes" #1 (by Others)
Structural Mounting Plate, refer to Plan View and "Rough-In Notes" #3 (by Others) Finished Ceiling (by Others) Draeger Ceiling Flange (by Draeger)
Require a Clear Ceiling Opening of 21" x 21" / [533mm x 533mm] (by Others) For 23.5" x 23.5" / [600mmx600mm] Cover Panel (by Draeger)
4m m [6
.8
533mm [21"]
Top View
Elevation View
(See "Rough-In Notes" #3)
A A
533mm [21"]
1600mm [63.0"]
910mm [35.8"]
1450mm [57.1"]
Top View 1:20 "M"
R1600mm
[R63.0"]
R2510mm
[R98.8"]
910mm
[3 5.8"]
R1450mm [R57.1"]
R2360mm[R92.9"]
1 / P600 and 1/X-Ray Shield
Mount Location "M"
Polaris 600 (article-number: OP75084)
X-Ray Shield (article-number: OV2013-ML681-04 OT50001)
IR Lab - Mount Location "M"
EHART
16DEC2021
R
Q
P
2 3 4
N
M
L
K
J
5 6 7 8 9 10 11 12
H
G
F
E
D
C
B
A
2 3 4 5 6 7 8 9 10 11 12
13 14 15 16 1817 19 size A020 21 22 23
R
Q
P
N
M
L
K
J
13 14 15 16 1817 19
H
G
F
E
D
20 21 22 23
C
B
A
U se o f t hi s do cu m en t / c on te nt s is fo rb id de n w ith ou t e xp re ss a ut ho rit y.
A ll rig ht s re se rv ed
R ef er to p ro te ct io n no tic e
IS
O
Type:
Type:
3D File:
System:
Date:
Name:
Author: Approved by: Scale:
Date:
Name:
View: 1:16 Others: See Details
3DFL - VA Iowa City 08FEB2022 1041
Quantity of Units: Project Name:
3D-Tool #:
Dräger-Country Code: Target Country:Voltage (V): Frequency (Hz):
VA Iowa City EP CATH IR
US United States120 60
Date: Author: Revision Text: Rev #:Notes:
Maximum load and torque requirements for this system have been acknowledged and it is confirmed by signing this document, that proper provisions at the constructions site have been made or will be implemented
Sample units for the definition of the US Country Standard.
Units partly against current Project Planning Specifications:
Couplings for Flammable Gases above Electrical…
This is the start of the file's text. The full file is on GovTribe.
File details come from the government source that posted it. Updated .