VASD Siemens PET CT 2 Rm 4503 Final2 Shield Design.pdf

PDF 737 KB Posted

Attached to
Y1DA--CONSTRUCTION 664-CSI-115 4W Nuclear Medicine PET CT II
Solicitation number
36C25821B0012
Issued by

About this file

This document is a shielding evaluation memorandum for the installation of a new Siemens Biograph Vision PET/CT scanner in Room 4503 at a Veterans Affairs medical facility. The memorandum specifies the shielding criteria and requirements based on the scanner model and expected workload of 40 patients per week. It requires a minimum of 5.5 inches of lead for the walls, 3 inches for the door, and 0.5 inches for the windows to meet regulatory radiation protection standards. The lead barriers must be rated to block 511 keV photons. The evaluation was performed by a qualified medical physicist and certifies that the specified shielding is sufficient based on an annual workload of 40 patients per week being imaged with average activities of 10-15 mCi of fluorine-18 radiotracers over 30-60 minute scan times.

View the file

Other files

File Type Posted
RFIs 1-68 Amendment 0007.docx DOCX document
36C25821B0012 0007.docx DOCX document
36C25821B0012 0006.docx DOCX document
RFIs 1-67 Amendment 0006.docx DOCX document
RFIs 1-63 Amendment 0005.docx DOCX document
36C25821B0012 0005.docx DOCX document
270526 Grounding and Bonding for Comm Systems.pdf PDF
270533 Raceways and Boxes for Communications Systems.pdf PDF
271000 Control Comm and Cable Wiring.pdf PDF
271500 Communications Structured Cabling.pdf PDF
270511 Requirements for Comm Installations.pdf PDF
Wage Determination CA20210001 June 25 2021.pdf PDF
36C25821B0012 0004.docx DOCX document
RFIs 1-60 Amendment 0004 14 Jul 2021.docx DOCX document
2020-1106 Siemens Set.pdf PDF
Hazardous Materials Inspection Report 01-2015Part-3.pdf PDF
RFIs 1-60 13 Jul 2021 Amendment 0003.docx DOCX document
Hazardous Materials Inspection Report 01-2015Part-4.pdf PDF
36C25821B0012 0003.docx DOCX document
Hazardous Materials Inspection Report 01-2015Part-6.pdf PDF
Hazardous Materials Inspection Report 01-2015Part-1.pdf PDF
Hazardous Materials Inspection Report 01-2015Part-5.pdf PDF
Hazardous Materials Inspection Report 01-2015Part-2.pdf PDF
Site Walk Sign-In Sheet.pdf PDF
Questions Asked by Potential Bidder.docx DOCX document
36C25821B0012 0002.docx DOCX document
VASD Siemens PET_CT_2 Rm 4503.Final2 Shield Design.pdf PDF
datasheet_gh3_us.pdf PDF
Questions and Responses from Bidders.docx DOCX document
RFIs.docx DOCX document
2020-1001 FP set.pdf PDF
2020-1001 Electrical Set - Copy.pdf PDF
36C25821B0012 0001.docx DOCX document
2020-1001 Plumbing set.pdf PDF
2020-1001 Mechanical Set.pdf PDF
FINAL20201204 Plumbing Set.pdf PDF
Wage Determination June 10 2021.docx DOCX document
FINAL20201204 Architectural Set.pdf PDF
FINAL20201204 Mechanical Set.pdf PDF
FINAL20201204 Structural Set.pdf PDF
36C25821B0012.docx DOCX document
Final IFC Specification Manual.pdf PDF
Final17X001 Elec 20201203.pdf PDF
Atch 1 Self Performance Calculations.pdf PDF
Show all 44

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Memorandum

Date: 13 October 2020

From: Medical Physicist, Radiology Service

Subj: Shielding Evaluation for Siemens Biograph Vision PET/CT 2 Scanner Room 4503

To: Ernest Belezzuoli, M.D., Nuclear Medicine Service (115) cc: Amilcare Gentili, MD

Sara Johnson, MS Michael Degnan

Floor plan drawings (AE-104 FLOOR PLAN and AE-204 RCP1) were received from Engineering Services on 10/1/2020 for the proposed installation of the new Siemens Biograph Vision PET/CT Scanner in Room 4503. Siemens x-ray scatter specifications (A-102) were also received.

Workload data and sample protocols were obtained from Nuclear Medicine Service and Biomedical Engineering.

The shielding requirements are summarized below and on the attached room sketch. This report should be maintained for the life of this facility to ensure adequate documentation of the shielding design criteria. Please ensure that a copy of this report is given to the owner/operator. The Radiation Shielding Plan is based on the requirements of the Nuclear Regulatory Commission, AAPM Task Group 108, California Radiation Control Regulations Title 17, Report 147 of the National Council on Radiation Protection & Measurements and applicable provisions of Chapter 31C, California Building Code (Title 24, California Code of Regulations) relating to radiation protection.

It is highly recommended that the lead materials be physically inspected at the time of installation and documentation be made. The Qualified Medical Physicists shall perform radiation scatter measurements during Acceptance Testing of the scanner after its installation.

The following criteria were applied in the shielding calculations and radiation safety specifications based on information supplied by Nuclear Medicine personnel, Siemens and Engineering Services. If there are any changes in the workload values, the scanner model, the room layout or area occupancy the shielding calculations may no longer be valid. Please contact me immediately if there are any changes or revisions contemplated for this room layout or its proposed use. These calculations reflect conservative shielding criteria consistent with good practice. Shielding is rated for a workload with 100% positron emitters using the half-life (110 min) of F-18 FDG 10 mCi and F-18 Furibapiz 2 doses 15 mCi total. Use of shorter-lived isotopes (N-13, O-15, C-11, Rb- 82, Ga-88 5 mCi) will not affect these requirements. I further declare that I am a Medical Radiation Physicist certified by the American Board of Radiology and registered with the State of California.

Lead thickness greater than the minimum specified may be used. All lead equivalent barriers (walls, doors and windows) must be rated at 511 keV.

Shielding Criteria:

1. Total Workload: 40 pats/wk, 8 hrs/day, 40 hrs/wk, mixture of 2D and 3D acquisitions, whole body 65%

2. Administered activity: 10-15 mCi (366-555 MBq)

3. Patients will void (~15% of administered activity) after the 60 min uptake before scanning

4. Average imaging time: 30-60 min

5. 1 hour integrated dose from patient at 1 m: 3.0 μGy/hr/37 MBq

6. Body absorption factor for PET: 0.34,

7. CT techniques: 120 kVp, 166 mAs body, 362 mAs head (per PET/CT 1), 96,140 mAs/wk

8. All Use Factors are 0 for primary beam, 1 for scattered radiation

9. Uncontrolled areas: <0.02 mSv (2 mR) per week

10. Controlled areas: Target value of < 0.02 mSv (2 mR) per week SDVA ALARA Level 1

11. Floor-to-Floor Distance: 17 ft

12. Concrete floors: 5.1 inches. No additional barrier required for ceiling or floor (occupancy 1.0 for each). The interstitial space above the scan room must be restricted during use.

13. Attenuation of x-rays and photons in the scanner gantry was included in barrier calculations.

14. A thicker lead barrier than that specified in this Report may be installed. If this is done, full documentation in the permanent record is required.

Thank you for the opportunity to review this project and to be of service. Please contact us at your convenience for any questions about this report.

Sincerely, Michael P. André, Ph.D., FAIUM, FAAPM David J. Hall, Ph.D.

Certified, American Board of Radiology Certified, American Board of Radiology

Corridor

Corridor

PET/CT

SPECT/CT

Corridor

VASD SIEMENS BIOGRAPH VISION PET/CT SCANNER RECOMMENDED BARRIERS

Room 4503

Barrier Location Or Wall

(Ref. Drawing)

Type of Area

Distance ft (m)

Occupancy Factor

XUT CT

Scatter

(mSv/wk)

PET

(mSv/wk) at barrier

CT Required Minimum

Barrier

PET Barrier Recommended Total Barrier

(A1) Shared Control Rm wall, door, window

Controlled

0.1 mSv/wk

(4.9)

1 T

0.11 23 0.4 mm Pb 1 lb

1.3 mm 3#

1.6 mm 4# 8 ft high

(B) Control Rm doorway, open

Controlled

0.1 mSv/wk

1/5 T

0.06 23 No additional No additional

(C1) East Corridor wall, former door

Uncontrolled

0.02 mSv/wk

(5.5)

0.06 18 No additional No additional --

(C2) East Corridor wall

Uncontrolled

0.02 mSv/wk

0.08 26 0.3 mm Pb 1 lb

0.3 mm Pb 0.8 mm 2# 8 ft high

(D) East Corridor wall

(1.8)

0.65 32 0.8 mm Pb 2 lb

3.7 mm Pb 10#

3.9 mm 10# 8 ft high

(E) East Corridor wall

Uncontrolled

0.02 mSv/wk

(1.8)

1/5 T 0.39 32 0.67 mm Pb 2 lb

3.7 mm Pb 10#

3.9 mm 10# 8 ft high

(F1) South Corridor wall

Uncontrolled

0.02 mSv/wk

(1.5)

1/5 T 0.85 47 0.9 mm Pb

2.5 lb

6 mm Pb 16#

6.3 mm 16# 8 ft high

(F2) Receptionist alcove 4076 southeast

Uncontrolled

0.02 mSv/wk

(3.7)

1 T 0.77 40 0.8 mm Pb 2 lb

5.5 mm Pb 16#

6.3 mm 16# 8 ft high

(F3) Reading Room south Uncontrolled

0.02 mSv/wk

(4)

1 T 0.63

34 0.8 mm Pb 2 lb

4.2 mm Pb 12#

No additional F1 satisfies

(G) Reading Room west

Uncontrolled

0.02 mSv/wk

(2.1)

1 T 0.43 118 0.7 mm Pb

1.5 lb

3 mm Pb 8#

3 mm 8# 8 ft high

(H1) Equipment Rm 4520 west

Uncontrolled

0.1 mSv/wk

(1.8)

1/20 T 0.12 8 0.5 mm Pb 2 lb

2.3 mm Pb 6#

3 mm 8# 8 ft high

(H2) Corridor 4521 beyond Equip Rm

Uncontrolled

0.1 mSv/wk

(3.7)

1/5 T 0.17 24 0.5 mm Pb 2 lb

2.3 mm Pb 6#

H1 satisfies

(I) West Door 4503.1 Uncontrolled

0.1 mSv/wk

(4.9)

1/5 T <0.02 <0.02 No additional No additional No additional

(J) Corridor wall Uncontrolled

0.1 mSv/wk

(4.9)

1/5 T 0.12 5 0.4 mm Pb 0.4 mm Pb 0.8 mm 2# 8 ft high

(K) Control Doorway 4504.1

Uncontrolled

0.1 mSv/wk

(5.5)

1/5 T <0.02 <0.02 No additional No additional No additional barrier required

(L) Control wall west Uncontrolled

0.1 mSv/wk

(5.8)

1/5 T <0.02 <0.02 No additional No additional No additional barrier required

Ceiling

14 1 T 0.38 29.7 2.6” concrete

2.2” concrete

5.1” concrete existing

No additional Floor

14 1T 0.38 29.7 2.6” concrete 2.2” concrete

5.1” concrete existing

No additional Workload: 40 pats/wk

1 HVL broad beam geometry at 511 keV = 5 mm Pb (12 lb/sq ft)

CT Radiation Scatter Plots

Shielding Criteria:
1. Total Workload: 40 pats/wk, 8 hrs/day, 40 hrs/wk, mixture of 2D and 3D acquisitions, whole body 65%
VASD SIEMENS Biograph Vision PET/CT SCANNER RECOMMENDED BARRIERS

Or Wall

File details come from the government source that posted it. Updated .