This document is a shielding evaluation memorandum for the installation of a new Siemens Biograph Vision PET/CT scanner in Room 4503 at a Veterans Affairs medical facility. The memorandum specifies the shielding criteria and requirements based on the scanner model and expected workload of 40 patients per week. It requires a minimum of 5.5 inches of lead for the walls, 3 inches for the door, and 0.5 inches for the windows to meet regulatory radiation protection standards. The lead barriers must be rated to block 511 keV photons. The evaluation was performed by a qualified medical physicist and certifies that the specified shielding is sufficient based on an annual workload of 40 patients per week being imaged with average activities of 10-15 mCi of fluorine-18 radiotracers over 30-60 minute scan times.
View the file
Other files
Show all 44
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Memorandum
Date: 13 October 2020
From: Medical Physicist, Radiology Service
Subj: Shielding Evaluation for Siemens Biograph Vision PET/CT 2 Scanner Room 4503
To: Ernest Belezzuoli, M.D., Nuclear Medicine Service (115) cc: Amilcare Gentili, MD
Sara Johnson, MS Michael Degnan
Floor plan drawings (AE-104 FLOOR PLAN and AE-204 RCP1) were received from Engineering Services on 10/1/2020 for the proposed installation of the new Siemens Biograph Vision PET/CT Scanner in Room 4503. Siemens x-ray scatter specifications (A-102) were also received.
Workload data and sample protocols were obtained from Nuclear Medicine Service and Biomedical Engineering.
The shielding requirements are summarized below and on the attached room sketch. This report should be maintained for the life of this facility to ensure adequate documentation of the shielding design criteria. Please ensure that a copy of this report is given to the owner/operator. The Radiation Shielding Plan is based on the requirements of the Nuclear Regulatory Commission, AAPM Task Group 108, California Radiation Control Regulations Title 17, Report 147 of the National Council on Radiation Protection & Measurements and applicable provisions of Chapter 31C, California Building Code (Title 24, California Code of Regulations) relating to radiation protection.
It is highly recommended that the lead materials be physically inspected at the time of installation and documentation be made. The Qualified Medical Physicists shall perform radiation scatter measurements during Acceptance Testing of the scanner after its installation.
The following criteria were applied in the shielding calculations and radiation safety specifications based on information supplied by Nuclear Medicine personnel, Siemens and Engineering Services. If there are any changes in the workload values, the scanner model, the room layout or area occupancy the shielding calculations may no longer be valid. Please contact me immediately if there are any changes or revisions contemplated for this room layout or its proposed use. These calculations reflect conservative shielding criteria consistent with good practice. Shielding is rated for a workload with 100% positron emitters using the half-life (110 min) of F-18 FDG 10 mCi and F-18 Furibapiz 2 doses 15 mCi total. Use of shorter-lived isotopes (N-13, O-15, C-11, Rb- 82, Ga-88 5 mCi) will not affect these requirements. I further declare that I am a Medical Radiation Physicist certified by the American Board of Radiology and registered with the State of California.
Lead thickness greater than the minimum specified may be used. All lead equivalent barriers (walls, doors and windows) must be rated at 511 keV.
Shielding Criteria:
1. Total Workload: 40 pats/wk, 8 hrs/day, 40 hrs/wk, mixture of 2D and 3D acquisitions, whole body 65%
2. Administered activity: 10-15 mCi (366-555 MBq)
3. Patients will void (~15% of administered activity) after the 60 min uptake before scanning
4. Average imaging time: 30-60 min
5. 1 hour integrated dose from patient at 1 m: 3.0 μGy/hr/37 MBq
6. Body absorption factor for PET: 0.34,
7. CT techniques: 120 kVp, 166 mAs body, 362 mAs head (per PET/CT 1), 96,140 mAs/wk
8. All Use Factors are 0 for primary beam, 1 for scattered radiation
9. Uncontrolled areas: <0.02 mSv (2 mR) per week
10. Controlled areas: Target value of < 0.02 mSv (2 mR) per week SDVA ALARA Level 1
11. Floor-to-Floor Distance: 17 ft
12. Concrete floors: 5.1 inches. No additional barrier required for ceiling or floor (occupancy 1.0 for each). The interstitial space above the scan room must be restricted during use.
13. Attenuation of x-rays and photons in the scanner gantry was included in barrier calculations.
14. A thicker lead barrier than that specified in this Report may be installed. If this is done, full documentation in the permanent record is required.
Thank you for the opportunity to review this project and to be of service. Please contact us at your convenience for any questions about this report.
Sincerely, Michael P. André, Ph.D., FAIUM, FAAPM David J. Hall, Ph.D.
Certified, American Board of Radiology Certified, American Board of Radiology
Corridor
Corridor
PET/CT
SPECT/CT
Corridor
VASD SIEMENS BIOGRAPH VISION PET/CT SCANNER RECOMMENDED BARRIERS
Room 4503
Barrier Location Or Wall
(Ref. Drawing)
Type of Area
Distance ft (m)
Occupancy Factor
XUT CT
Scatter
(mSv/wk)
PET
(mSv/wk) at barrier
CT Required Minimum
Barrier
PET Barrier Recommended Total Barrier
(A1) Shared Control Rm wall, door, window
Controlled
0.1 mSv/wk
(4.9)
1 T
0.11 23 0.4 mm Pb 1 lb
1.3 mm 3#
1.6 mm 4# 8 ft high
(B) Control Rm doorway, open
Controlled
0.1 mSv/wk
1/5 T
0.06 23 No additional No additional
(C1) East Corridor wall, former door
Uncontrolled
0.02 mSv/wk
(5.5)
0.06 18 No additional No additional --
(C2) East Corridor wall
Uncontrolled
0.02 mSv/wk
0.08 26 0.3 mm Pb 1 lb
0.3 mm Pb 0.8 mm 2# 8 ft high
(D) East Corridor wall
(1.8)
0.65 32 0.8 mm Pb 2 lb
3.7 mm Pb 10#
3.9 mm 10# 8 ft high
(E) East Corridor wall
Uncontrolled
0.02 mSv/wk
(1.8)
1/5 T 0.39 32 0.67 mm Pb 2 lb
3.7 mm Pb 10#
3.9 mm 10# 8 ft high
(F1) South Corridor wall
Uncontrolled
0.02 mSv/wk
(1.5)
1/5 T 0.85 47 0.9 mm Pb
2.5 lb
6 mm Pb 16#
6.3 mm 16# 8 ft high
(F2) Receptionist alcove 4076 southeast
Uncontrolled
0.02 mSv/wk
(3.7)
1 T 0.77 40 0.8 mm Pb 2 lb
5.5 mm Pb 16#
6.3 mm 16# 8 ft high
(F3) Reading Room south Uncontrolled
0.02 mSv/wk
(4)
1 T 0.63
34 0.8 mm Pb 2 lb
4.2 mm Pb 12#
No additional F1 satisfies
(G) Reading Room west
Uncontrolled
0.02 mSv/wk
(2.1)
1 T 0.43 118 0.7 mm Pb
1.5 lb
3 mm Pb 8#
3 mm 8# 8 ft high
(H1) Equipment Rm 4520 west
Uncontrolled
0.1 mSv/wk
(1.8)
1/20 T 0.12 8 0.5 mm Pb 2 lb
2.3 mm Pb 6#
3 mm 8# 8 ft high
(H2) Corridor 4521 beyond Equip Rm
Uncontrolled
0.1 mSv/wk
(3.7)
1/5 T 0.17 24 0.5 mm Pb 2 lb
2.3 mm Pb 6#
H1 satisfies
(I) West Door 4503.1 Uncontrolled
0.1 mSv/wk
(4.9)
1/5 T <0.02 <0.02 No additional No additional No additional
(J) Corridor wall Uncontrolled
0.1 mSv/wk
(4.9)
1/5 T 0.12 5 0.4 mm Pb 0.4 mm Pb 0.8 mm 2# 8 ft high
(K) Control Doorway 4504.1
Uncontrolled
0.1 mSv/wk
(5.5)
1/5 T <0.02 <0.02 No additional No additional No additional barrier required
(L) Control wall west Uncontrolled
0.1 mSv/wk
(5.8)
1/5 T <0.02 <0.02 No additional No additional No additional barrier required
Ceiling
14 1 T 0.38 29.7 2.6” concrete
2.2” concrete
5.1” concrete existing
No additional Floor
14 1T 0.38 29.7 2.6” concrete 2.2” concrete
5.1” concrete existing
No additional Workload: 40 pats/wk
1 HVL broad beam geometry at 511 keV = 5 mm Pb (12 lb/sq ft)
CT Radiation Scatter Plots
| Shielding Criteria: |
| 1. Total Workload: 40 pats/wk, 8 hrs/day, 40 hrs/wk, mixture of 2D and 3D acquisitions, whole body 65% |
| VASD SIEMENS Biograph Vision PET/CT SCANNER RECOMMENDED BARRIERS |
Or Wall