2020-1001 Electrical Set - Copy.pdf

PDF 4 MB Posted

Attached to
Y1DA--CONSTRUCTION 664-CSI-115 4W Nuclear Medicine PET CT II
Solicitation number
36C25821B0012
Issued by

About this file

This solicitation is for construction services to fully renovate approximately 1,200 square feet of an existing Nuclear Medicine department to prepare the space for installation of a new PET CT scanner. The scope of work includes renovations in accordance with provided specifications, drawings, and statement of work. The soliciting agency is the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 22. The solicitation number is 36C25821B0012 and is titled "Y1DA--CONSTRUCTION 664-CSI-115 4W Nuclear Medicine PET CT II."

View the file

Other files

File Type Posted
RFIs 1-68 Amendment 0007.docx DOCX document
36C25821B0012 0007.docx DOCX document
36C25821B0012 0006.docx DOCX document
RFIs 1-67 Amendment 0006.docx DOCX document
RFIs 1-63 Amendment 0005.docx DOCX document
36C25821B0012 0005.docx DOCX document
VASD Siemens PET CT 2 Rm 4503 Final2 Shield Design.pdf PDF
270526 Grounding and Bonding for Comm Systems.pdf PDF
270533 Raceways and Boxes for Communications Systems.pdf PDF
271000 Control Comm and Cable Wiring.pdf PDF
271500 Communications Structured Cabling.pdf PDF
270511 Requirements for Comm Installations.pdf PDF
Wage Determination CA20210001 June 25 2021.pdf PDF
36C25821B0012 0004.docx DOCX document
RFIs 1-60 Amendment 0004 14 Jul 2021.docx DOCX document
2020-1106 Siemens Set.pdf PDF
Hazardous Materials Inspection Report 01-2015Part-3.pdf PDF
RFIs 1-60 13 Jul 2021 Amendment 0003.docx DOCX document
Hazardous Materials Inspection Report 01-2015Part-4.pdf PDF
36C25821B0012 0003.docx DOCX document
Hazardous Materials Inspection Report 01-2015Part-6.pdf PDF
Hazardous Materials Inspection Report 01-2015Part-1.pdf PDF
Hazardous Materials Inspection Report 01-2015Part-5.pdf PDF
Hazardous Materials Inspection Report 01-2015Part-2.pdf PDF
Site Walk Sign-In Sheet.pdf PDF
Questions Asked by Potential Bidder.docx DOCX document
36C25821B0012 0002.docx DOCX document
VASD Siemens PET_CT_2 Rm 4503.Final2 Shield Design.pdf PDF
datasheet_gh3_us.pdf PDF
Questions and Responses from Bidders.docx DOCX document
RFIs.docx DOCX document
2020-1001 FP set.pdf PDF
36C25821B0012 0001.docx DOCX document
2020-1001 Plumbing set.pdf PDF
2020-1001 Mechanical Set.pdf PDF
FINAL20201204 Plumbing Set.pdf PDF
Wage Determination June 10 2021.docx DOCX document
FINAL20201204 Architectural Set.pdf PDF
FINAL20201204 Mechanical Set.pdf PDF
FINAL20201204 Structural Set.pdf PDF
36C25821B0012.docx DOCX document
Final IFC Specification Manual.pdf PDF
Final17X001 Elec 20201203.pdf PDF
Atch 1 Self Performance Calculations.pdf PDF
Show all 44

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

CONSTRUCTION DOCUMENTSOffice of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

664-CSI-1154W NUCLEAR MEDICINE PET- CT 2

SAN DIEGO, CA

IDS GROUP INC.

1 PETERS CANYON RD. STE. 130

IRVINE, CA 92606

949.387.8500

FULLY SPRINKLERED

C U s e rs \A le ja n d ro .A g u ila r\ D o c u m e n ts

\1 X

.0

-V

A S

D -N u c -M e d

E le c R

T V ra m i.

e lh a s s a n id s g i.

c o m .r v t

/2 /2

:5

:4

P

M

EP-001

ELECTRICAL GEN. NOTES, LEGEND &

ABBREVIATION

10/02/20 RO GA

SINGLE LINE

CONDUIT ONLY

CIRCUIT BREAKER

AMP SWITCH

COMMUNICATION

CABINET

AMPERE

MOUNTED

MINIMUM

MECHANICAL

MAIN CIRCUIT BREAKER

LABORATORY INFORMATION

MANAGEMENT SYSTEM

FEET OR FOOT

FIXTURE

DEMOLITION/DEMOLISH

ELECTRICAL

NATIONAL ELECTRICAL

HEATING, VENTILATING &

DISCONNECT

DRAWING

GROUND FAULT CIRCUIT INTERRUPTER

KILOVOLT AMPERE

NATIONAL ELECTRICAL

MAIN LUGS ONLY

INTERMEDIATE METAL CONDUIT

AMERICAN WIRE GAUGE

CATALOG/CATEGORY

AMPERE INTERRUPTING CAPACITY

ARC FAULT, AMP FUSE

ABOVE FINISHED FLOOR

SHORT CIRCUIT AMPERES

INCH(ES)

NEMA

EMT

EM

FF

HP

GND

FT

IMC

HVAC

EQUIP

FIXT

GFCI

MAX.

MIN.

ISC

IN

KVA

MLO

NEC

NECA

KW

LIMS

MCB

MECH.

MTD

EQUIPMENT

EMERGENCY

ELECTRICAL METALLIC TUBING

GROUND

FINISHED FLOOR

NATIONAL ELECTRICAL CODE

MAXIMUM

KILOWATT

DWG

AIC

AS

AFF

C/B

COMM

CAB

AF

CAT

CKT

CLG

CU

AWG

CO, EC

C

A

ELECT.

(D)

EA

DISC.

CIRCUIT

COPPER

CEILING

CONDUIT

EACH

BLDG BUILDING

(E), EXIST EXISTING

KCMIL,MCM THOUSAND CIRCULAR MILS

NORMALLY CLOSEDNC

FLEXIBLE METALLIC CONDUIT (STEEL)FLEX

JUNCTION BOXJB,J-BOX

LEDFLUOR

DOWNDN

BC BARE COPPER

INTERNATIONAL RESIDENTIAL CODE

INTERNATIONAL BUILDING CODE

IBC

NOT IN CONTRACT

PANEL

POWER

RECEPTACLE

REQUIRED

SECTION

QUANTITY

NEVADA POWER COMPANY

NORMALLY OPEN

NOT TO SCALE

SCHEDULE

OVERCURRENT PROTECTION

NATIONAL FIRE CODE

NEUTRAL

NIC

NEUT

NFC

NL

PH

NPCO

SPEC

SP

OCP

SECT

SCHED

REQ'D

RECEP

PWR

PNL

NTS

QTY

NO

NIGHT LIGHT

SINGLE POLE

SPECIFICATION

PHASE

POLEP

SW SWITCH

SWBD SWITCHBOARD

SWGR SWITCH GEAR

SYS SYSTEM

TEMP TEMPORARY

TELE TELEPHONE

XFMR TRANSFORMER

UBC UNIFORM BUILDING CODE

UL UNLESS NOTED OTHERWISE

V VOLT OR VOLTAGE

VP VAPOR PROOF

TYP. TYPICAL

W WATT, WIRE

WP UL LISTED WEATHERPROOF,

NON-FUSIBLENF

POLYVINYL CHLORIDEPVC

VD VOLTAGE DROP

VA VOLT AMPERE

MANUFACTURERS ASSOCIATION

CONTRACTOR'S ASSOCIATION

28. ALL PENETRATIONS OF FIRE RATED FLOOR OR WALL ASSEMBLIES SHALL BE PROTECTED

BY MATERIALS AND INSTALLATION DIAGRAMS THAT CONFORM TO UL LISTING FOR

"THROUGH-PENETRATION FIRE STOP SYSTEMS".FIRESTOPPING SHALL MATCH F, T, W, AND L

RATINGS OF FLOOR AND WALL ASSEMBLIES.

29. CONTRACTOR SHALL ENGAGE THE SERVICES FOR A STATE LICENSED FIRE ALARM

MANUFACTURER/INSTALLER TO PREPARE ALL DESIGN DRAWINGS AND CALCULATIONS

REQUIRED FOR SYSTEM APPROVAL BY THE AUTHORITY HAVING JURISDICTION.

SUBMIT ALL PLANS AND PROVIDE ALL PERMITS REQUIRED FOR A COMPLETE AND

OPERABLE APPROVED LIFE SAFETY SYSTEM.

30. ALL FIRE ALARM DEVICE WIRING SHALL BE CLASSIFIED PER NFPA 72. FIRE ALARM DEVICE WIRING

SHALL BE MINIMUM #14 AWG COPPER OR PER SYSTEM MANUFACTURER REQUIREMENTS.

PROVIDE MINIMUM 3/4" SEPARATE RACEWAY SYSTEM OR AS REQUIRED FOR LIFE SAFETY SYSTEM

WIRING CONFIGURATION.

31. UPON COMPLETION OF THE INSTALLATION OF LIFE SAFETY SYSTEM WIRING AND

DEVICES, A PERFORMANCE TEST OF THE ENTIRE LIFE SAFETY SHALL BE

PERFORMED TO THE SATISFACTION OF THE AUTHORITY HAVING JURISDICTION.

32. ALL EQUIPMENT ELECTRICAL TERMINATIONS TO UNDERGO A TORQUE TEST.

ELECTRICAL CONTRACTOR IS RESPONSIBLE FOR MANUFACTURER'S RECOMMENDED

TORQUE DOCUMENTATION AND TOOLS TO PERFORM TORQUE TEST. CONTRACTOR SHALL APPLY

UL TABLE 486A-B WHERE THE MANUFACTURER DOES NOT PROVIDE TORQUE

RECOMMENDATIONS.

33. ALL UNDERGROUND SERVICE CONDUITS SHALL BE SEALED PER NEC ARTICLE 230-8.

34. FLOOR MOUNTED ELECTRICAL EQUIPMENT SHALL BE MOUNTED ON A 4" HIGH

CONCRETE PAD.

35. INSTALL TRANSFORMER FOLLOWING MANUFACTURER'S RECOMMENDATIONS FOR

VENTILATION CLEARANCES.

36. PRIOR TO WORK COORDINATE ELECTRICAL REQUIREMENTS FOR ALL PLUMBING AND MECHANICAL

EQUIPMENT WITH FINAL CONTRACTOR SELECTION. THE CONTRACTOR SHALL SIZE

DISCONNECTS BASED UPON CIRCUIT BREAKER RATINGS AND PROVIDE FUSING AS

REQUIRED PER EQUIPMENT MANUFACTURER RECOMMENDATIONS AND U.L. LISTING

REQUIREMENTS.

37. PROVIDE #10 AWG CONDUCTORS FOR 20 AMPERE, 120V BRANCH CIRCUITS LONGER

THAN 75' AND #8 AWG CONDUCTORS FOR 20 AMPERE, 120V BRANCH CIRCUITS

LONGER THAN 120'. PROVIDE #10 AWG CONDUCTORS FOR 20 AMPERE, 277V

BRANCH CIRCUITS LONGER THAN 200'.

UNDERGROUND PULL SECTIONUGPS

AIR CONDITIONING

NEMA 3R

TELEVISION OUTLET

TELEPHONE/COMPUTER OUTLET-MOUNTED 450MM (18")AFF

UNLESS OTHERWISE NOTED

ELECTRICAL STRIP MOLD-NEMA 5-20R

RECEPTACLES AT 600MM(2'-0")INTERVALS

SERVICE CABLE TERMINATION

CIRCUIT BREAKER

FUSE

FUSED DISCONNECT SWITCH

SWITCH

SURGE SUPPRESSOR

CURRENT TRANSFORMER

POTENTIAL TRANSFORMER

GROUNDING ELECTRODE

POWER METER

MOTOR

GENERATOR

SHUNT TRIP

GROUND FAULT INTERRUPT

TRANSFER SWITCH

DUPLEX RECEPTACLE, NEMA 5-20R-20AMP

MOUNTED 450MM (18")AFF UNLESS OTHERWISE NOTED

CONTACT (NORMALLY OPEN)

CONTACT (NORMALLY CLOSED)

TIME SWITCH

CONTROL SWITCH

PUSH BUTTON

DUPLEX RECEPTACLE WITH GROUND FAULT INTERRUPTER,

NEMA 5-20R-20AMP-MOUNTED 450MM (18")AFF UNLESS

OTHERWISE NOTED

DUPLEX RECEPTACLE, NEMA 5-20R-20AMP

MOUNTED ABOVE COUNTER TOP

UNDERWRITERS LABORATORYU.N.O.

DUPLEX RECEPTACLE WITH GROUND FAULT INTERRUPTER,

NEMA 5-20R-20AMP-MOUNTED ABOVE COUNTER TOP

WEATHERPROOF DUPLEX RECEPTACLE WITH GFI, NEMA

5-20R-20AMP-MOUNTED 450MM (18")AFF UNLESS

OTHERWISE NOTED

QUADRAPLEX OUTLET, NEMA 5-20R-20AMP-MOUNTED

450MM (18")AFF OR QUADRAPLEX OUTLET, NEMA

5-20R-20AMP-PEDESTAL-MOUNTED

QUADRAPLEX OUTLET, NEMA 5-20R-20AMP-MOUNTED

ABOVE COUNTER TOP

QUADRAPLEX OUTLET WITH GROUND FAULT INTERRUPTER,

NEMA 5-20R-20AMP-MOUNTED 450MM (18")AFF UNLESS

OTHERWISE NOTED

QUADRAPLEX OUTLET WITH GROUND FAULT INTERRUPTER,

NEMA 5-20R-20AMP-MOUNTED ABOVE COUNTER TOP

DUPLEX RECEPTACLE, NEMA 5-20R-20AMP-EMERGENCY

POWER-MOUNTED 450MM (18")AFF UNLESS OTHERWISE

NOTED

QUADRAPLEX RECEPTACLE, NEMA

5-20R-20AMP-EMERGENCY POWER

SPECIAL RECEPTACLE

WALL MOUNTED TELEPHONE/COMPUTER OUTLET-

MOUNTED AT 1200MM (48")AFF UNLESS OTHERWISE NOTED

SPEAKER-CEILING MOUNTED

INTERCOM OUTLET

NURSE CALL DOME LIGHT-CEILING MOUNTED

NURSE CALL DOME LIGHT-WALL MOUNTED

NURSE CALL DUTY STATION

EMERGENCY NURSE CALL

NURSE CALL STAFF STATION

VOLUME CONTROL-WALL MOUNTED

JUNCTION BOX-PURPOSE AND LOCATION AS NOTED

SINGLE POLE SWITCH

SINGLE POLE SWITCH-SUFFIX OF a,b OR c INDICATES

SEPARATE CONTROL OR FIXTURES WITH SAME DESIGNATION

DIMMER SWITCH

THREE WAY SWITCH

DOOR SWITCH

FUSED OR UNFUSED DISCONNECT SWITCH

2'x2' LED FIXTURE

1'x4' LINEAR LED FIXTURE

2'x4' FLUORESCENT FIXTURE

WALL MOUNTED FLUORESCENT FIXTURE

2'x2' LED FIXTURE-EMERGENCY POWER

2'x4' FLUORESCENT FIXTURE-EMERGENCY POWER

WALL MOUNTED FLUORESCENT FIXTURE-EMERGENCY

POWER

WALL MOUNTED LIGHT FIXTURE-TYPE AS NOTED

LIGHT FIXTURE-TYPE AS NOTED

LIGHT FIXTURE-TYPE AS NOTED EMERGENCY POWER

CIRCUIT BREAKER

BATTERY POWERED CLOCK

EMERGENCY BATTERY UNIT WITH HEADS

DUAL TECHNOLOGY OCCUPANCY SENSOR - CEILING

DUAL TECHNOLOGY OCCUPANCY SENSOR - WALL

A

GFI

GFI-A

WP

A

GFI

GFI-A

w

D

E

S a

D os

NOTES NOTES (CONT'D.)

ST

TVSS

G

M

GFI

S

I

N

N

N

N

N

V

J

DS

CB

C

POWER

LIGHTING SIGNAL

STANDARD ABBREVIATIONS

GENERAL ELECTRICAL NOTES SHEET NUMBER SHEET NAME

EP-001 ELECTRICAL GEN. NOTES, LEGEND & ABBREVIATION

EP-002 SINGLE LINE DIAGRAM

EP-003 ELECTRICAL PANEL SCHEDULES

EP-004 VOLTAGE DROP CALCULATIONS & LOAD SUMMARY

EP-005 TITLE 24 COMPLIANCE DOCUMENTS

EP-100 ELECTRICAL 4TH FLOOR - OVERALL

EP-101 ELECTRICAL 4TH INTERSTITIAL - ENLARGED PLANS

EP-102 ELECTRICAL 4TH FLOOR - AREA OF WORK DEMOLITION

EP-103 4TH FLOOR-AREA OF WORK REMODEL- LIGHTING

EP-104 4TH FLOOR-AREA OF WORK REMODEL-POWER

EP-105 ELECTRICAL ENLARGED PLAN - INTERSTITIAL

EP-106 ELECTRICAL ENLARGED PLAN - EXTERIOR

EP-300 ELECTRICAL LIGHTING CONTROL

EP-301 ELECTRICAL LIGHTING CONTROL SINGLE LINE DIAGRAM

EP-501 PET/CT SCANNER

EP-502 PET/CT SCANNER

EP-503 PET/CT SCANNER

EP-504 PET/CT SCANNER

EP-505 PET/CT SCANNER

EP-600 LIGHTING DETAILS

EP-601 POWER DETAILS

EP-602 POWER DETAILS

EP-603 TELECOM DETAILS

EP-701 4TH FLR- AREA OF WORK REMODEL - PHOTOMETRICS

EP-702 4TH FLR- AREA OF WORK REMODEL - EGRESS PHOTOMETRICS

ISSUED FOR 100% C.D. 2020-10-02

TV

CB CODE BLUE INITIATION STATION

LIGHTINGLTG

PANELBOARD - SURFACE MOUNT

NOT FOR CONSTRUCTION

REVISION SYMBOL

SCHEDULED EQUIPMENT

KEYNOTE SYMBOL

EQUIPMENT TAG

DIAGRAM TAG

WIRELESS APPLICATION PROTOCOL

SWITCH BANK

XX

SHEET

UNIT

X

WAP

MISCELLANEOUS

SB

SINGLE POLE THERMAL RATED MOTOR STARTER SWITCH

M

DOUBLE POLE THERMAL RATED MOTOR STARTER SWITCH

38. FUSES SHALL BE PROVIDED WITH REJECTION TYPE FUSE HOLDERS.

NO PIPING, DUCTS OR EQUIPMENT FOREIGN TO ELECTRICAL EQUIPMENT SHALL BE PERMITTED

TO BE LOCATED WITHIN THE DEDICATED SPACE ABOVE THE ELECTRICAL EQUIPMENT.

CONTRACTOR SHALL PROVIDE DIMENSIONED SHOP DRAWINGS FOR ALL ELECTRICAL

EQUIPMENT.

CONTRACTOR SHALL COORDINATE WITH ALL DICIPLINES AND PROVIDE DIMENSIONED

SUBMITTAL DRAWINGS FOR ALL CONDUIT, PIPE AND WALL DUCTWORK, SUBMITTAL DRAWINGS

SHALL INCLUDE SIZES, DIMESSIONS AND ELEVATIONS FROM BUILDING WALLS, FLOORS AND

DECKS.

CONTRACTOR SHALL COORDIANTE WITH ALL DISCIPLINES AND ROUTE ELECTRICAL BRANCH

CIRCUIT CONDUITS WITH MECHANICAL PIPING WHERE POSSIBLE.

EMERGENCY POWER OFF (EPO) PUSH BUTTON

EXIT SIGN - CEILING MOUNTED

LINEAR LED STRIP FIXTURE

LINEAR LED STRIP FIXTURE - EMERGENCY POWER

TELEPHONE BACKBOARD

OS

TRANSFORMER

T

FUSE (> 600V)

DOOR SWITCHDS

EMERGENCY POWER OFF PUSH BUTTON

(HEIGHT NOTED ON PLANS)

EPO

39.

40.

41.

CONDUIT CAP-OFF

HOMERUN TO PANELBOARD 1/2"C W/3#12

CONDUCTORS UNO

RACEWAY TURNED DOWN

RACEWAY TURNED UP

RACEWAY W/#12 CONDUCTORS UNO

CONDUIT ROUTED UNDERFLOOR / UNDERGROUND

CONDUIT WITH TICKS FOR NUMBER & SIZE OF

CONDUCTOR#10

WIRING

1. ELECTRICAL WORK SHALL BE PERFORMED IN A WORKMANLIKE MANNER IN ACCORDANCE WITH

THE NECA INSTALLATION STANDARDS TO THE SATISFACTION OF THE ARCHITECT AND

ENGINEER.

2. ALL WORK, MATERIALS AND EQUIPMENT SHALL CONFORM TO THE CURRENTLY ACCEPTED

EDITION OF ALL APPLICABLE NATIONAL, STATE AND CITY CODES AND ORDINANCES.

3. ALL ELECTRICAL SYSTEM COMPONENTS SHALL BE LISTED OR LABELED BY UL OR NATIONALLY

RECOGNIZED TESTING FACILITY AS ALLOWED BY AUTHORITY HAVING JURISDICTION.

4. WHERE AN APPARENT DISCREPANCY EXISTS BETWEEN THE REQUIREMENTS OF THE GENERAL

NOTES AND/OR INFORMATION PORTRAYED IN THE ELECTRICAL DRAWINGS, THE CONTRACTOR

SHALL INCLUDE IN HIS BID THE COST OF THE GREATER QUALITY OR QUALITY.

5. CONTRACTOR SHALL VISIT JOB SITE PRIOR TO BID AND VERIFY EXISTING CONDITIONS.

CONTRACTOR SHALL INCLUDE IN BASE BID ALL COSTS REQUIRED FOR PERMITS AND

INSPECTIONS.

6. CONTRACTOR SHALL VERIFY, WITH OWNER'S REPRESENTATIVE PRIOR TO SUBMITTING BID,

ALLOWABLE WORKING HOURS, EMPLOYEE PARKING AREAS, MATERIAL DELIVERY, STORAGE

REQUIREMENTS, DEMOLITION AND REMOVAL OF CONSTRUCTION DEBRIS, AS WELL AS DAILY

CLEAN UP REQUIREMENTS. INCLUDE ALL COSTS IN BID FOR DUST BARRIERS, INFECTION

CONTROL, DUMPSTERS ETC. AS REQUIRED FOR THE DURATION OF THE PROJECT. PERFORM

WORK AS DIRECTED BY OWNER'S REPRESENTATIVE AND ARCHITECT.

7. ALL ELECTRICAL SYSTEMS SHALL BE TESTED FOR PROPER OPERATION. IF TESTS SHOW THAT

WORK IS DEFECTIVE, CONTRACTOR SHALL MAKE ALL NECESSARY CORRECTIONS AT NO

ADDITIONAL COST TO OWNER.

8. CONTRACTOR SHALL GUARANTEE WORK AGAINST DEFECTS IN MATERIALS AND WORKMANSHIP

WHICH MAY OCCUR UNDER NORMAL USE FOR A PERIOD OF ONE YEAR AFTER OWNER'S

ACCEPTANCE. DEFECTS SHALL BE PROMPTLY CORRECTED BY CONTRACTOR WITHOUT

ADDITIONAL CHARGE TO OWNER.

9. AS PART OF THE CLOSE-OUT FOR THE PROJECT, THE CONTRACTOR SHALL PROVIDE AS-BUILT

DRAWINGS TO ARCHITECT. DRAWINGS SHALL INCLUDE ACCURATE CONDUIT AND DEVICE

LOCATIONS DIMENSIONED FROM PERMANENT LANDMARKS SUCH AS BUILDING WALLS.

10. DO NOT SCALE ELECTRICAL DRAWINGS. VERIFY EXACT LOCATION OF ALL DEVICES, JUNCTION BOXES, LIGHTING FIXTURES WITH ARCHITECTURAL AND INTERIOR DESIGN DRAWINGS PRIOR TO

INSTALLATION. CONTRACTOR SHALL VERIFY THE EXACT LOCATION OF MECHANICAL

EQUIPMENT AND OTHER EQUIPMENT REQUIRING ELECTRICAL CONNECTION PRIOR TO ROUGH-

IN. EVERY OUTLET HEIGHT SHALL BE VERIFIED ON EACH WALL WITH THE INTERIOR PLANNING

AND DESIGN DRAWINGS. COORDINATE WITH CABINET SHOP DRAWINGS TO ENSURE PROPER

HEIGHT AND LOCATION WITH RESPECT TO MILLWORK, EQUIPMENT.

11. THESE DRAWINGS INDICATE THE FINISHED REQUIREMENTS FOR THE ELECTRICAL SYSTEMS, EQUIPMENT, LIGHTING FIXTURES, OUTLETS AND DEVICES. DUE TO STRUCTURAL CONDITIONS, MECHANICAL DUCT, PIPING CONFLICTS, OR OTHER LEGITIMATE REASONS, THE CONTRACTOR

MAY DESIRE TO INSTALL THE WORK INDICATED IN A MANNER DIFFERENT FROM THAT SHOWN.

SUCH CHANGES SHALL BE PRESENTED TO THE OWNER'S REPRESENTATIVE FOR REVIEW AND

APPROVAL PRIOR TO PROCEEDING. UPON APPROVAL, THE WORK SHALL BE PERFORMED AND

THE AS-BUILT DRAWINGS SHALL BE REVISED TO ACCURATELY REFLECT THE WORK AS

ACTUALLY INSTALLED.

12. RACEWAY SYSTEMS ARE SHOWN DIAGRAMMATICALLY. ACTUAL LOCATION AND ROUTING OF

ALL, SHALL BE DETERMINED BY CONTRACTOR TO SUIT FILED CONDITIONS

13. PROVIDE DEDICATED NEUTRAL FOR EACH NEW CIRCUIT PER NEC 210.4. SHARED NEUTRALS

SHALL NOT BE ALLOWED ETWEEN LOADS FROM SEPERATE OVERCURRENT DEVICES.

HANDLETIES SHALL NOT BE ALLOWED AS A METHOD OF COMPLIANCE WITH NEC 210.4 BETWEEN

SEPERATE DEVICES. HOME RUN CONDUCTORS MAY BE COMBINED INTO ONE CONDUIT WITH

SEPARATE NEUTRALS FOR EACH CIRCUIT. NO RACEWAY OR CABLE SHALL CONTAIN MORE THAN

NINE (9) CURRENT CARRYING CONDUCTORS. NEUTRAL CONDUCTORS SHALL BE CONSIDERED

CURRENT CARRYING CONDUCTORS. WHERE MULTIPLE CONDUCTORS IN EXCESS OF THREE (3)

ARE INDICATED ON THESE DRAWINGS, THEY SHALL BE DERATED AS REQUIRED BY NEC ARTICLE

310 REQUIREMENTS.

14. RACEWAYS SHALL BE INSTALLED CONCEALED (IN CMU OR OTHER WALL) WHENEVER POSSIBLE.

RACEWAYS INSTALLED EXPOSED SHALL BE ROUTED OUT OF PUBLIC VIEW AS MUCH AS

POSSIBLE. RACEWAYS SHALL BE RUN PARALLEL WITH, OR AT RIGHT ANGLE TO WALLS.

15. PROVIDE APPROVED EXPANSION FITTINGS WHERE RACEWAYS CROSS BUILDING EXPANSION

JOINTS. PROVIDE BONDING JUMPER(S) SIZED PER CODE WHERE REQUIRED. PROVIDE FITTINGS

REQUIRED FOR A COMPLETE INSTALLATION. REFER TO ARCHITECTURAL DRAWINGS FOR

EXPANSION JOINT LOCATION(S).

16. MINIMUM RACEWAY SIZE SHALL BE 3/4". MINIMUM CONDUCTOR SIZE SHALL BE #12 AWG U.N.O.

TYPICAL FOR ALL INTERIOR. MINIMUM CONDUCTOR SIZE FOR EXTERIOR SHALL BE #10 AWG

U.N.O. ALL POWER RELATED CONDUITS SHALL HAVE A GROUND WIRE INSTALLED IN EACH RUN

PER NEC 250.66.

17. CONTRACTOR SHALL PROVIDE PULL CORDS IN SPARE CONDUITS. WHERE MORE THAN ONE

CONDUIT TERMINATES IN A JUNCTION BOX, THE CONTRACTOR SHALL IDENTIFY EACH CONDUIT

AND JUNCTION BOX IN A MANNER ALLOWING IDENTIFICATION AFTER ALL WALL FINISHES HAVE

BEEN APPLIED.

18. CONTRACTOR SHALL PROVIDE ALL RACEWAY SYSTEMS INDICATED ON THE DRAWING PER NEC

REQUIREMENTS AND GENERAL NOTES. ANY DEVIATION FROM THE WIRING METHODS

INDICATED SHALL BE ALLOWED ONLY BY SPECIFIC WRITTEN APPROVAL FROM EITHER THE

ARCHITECT, ENGINEER OR OWNER. CONTRACTOR'S BID SHALL INCLUDE ALL COSTS FOR

RACEWAY SYSTEMS AS SPECIFIED UNLESS SPECIFIC WRITTEN APPROVAL FOR AN

ALTERNATIVE WIRING METHOD IS OBTAINED FROM EITHER THE ARCHITECT, ENGINEER OR

OWNER AND IS SUBMITTED AS PART OF CONTRACTOR'S FORMAL BID PROPOSAL.

19. CONTRACTOR SHALL BE RESPONSIBLE FOR CORRECT SIZE AND INSTALLATION OF OUTLET,

PULL AND JUNCTION BOXES IN ACCORDANCE WITH NEC 314-16. BOXES SHALL BE MINIMUM 4"

SQUARE BY 1-1/2" DEEP OR AS INDICATED ON THE DRAWINGS. BOXES SHALL BE RECESSED

WITH COVER PLATE TO SUIT THE INTENDED APPLICATION.

20. PRIOR TO INSTALLATION, CONTRACTOR SHALL REVIEW THE COMPLETE SET OF CONSTRUCTION

DOCUMENTS FOR CONFLICTS WITH OTHER TRADES. CONTRACTOR SHALL BE RESPONSIBLE TO

COORDINATE WORK WITH OTHER TRADES TO AVOID CONFLICT DURING INSTALLATION.

CONTRACTOR SHALL MAKE MINOR ADJUSTMENTS IN EQUIPMENT LOCATION AND ROUTING AS

NECESSARY AT NO ADDITIONAL COST TO THEOWNER.

21. CONTRACTOR SHALL BE RESPONSIBLE TO PROPERLY CUT AND PATCH EXISTING

CONSTRUCTION AS REQUIRED TO INSTALL NEW ELECTRICAL WORK. ALL PATCHING SHALL BE

OF THE SAME MATERIALS, WORKMANSHIP AND FINISH AS THE EXISTING WORK AND SHALL

ACCURATELY MATCH ALL SURROUNDING WORK TO THE SATISFACTION OF THE ARCHITECT.

22. ALL ELECTRICAL EQUIPMENT SHALL HAVE SUFFICIENT GUTTER SPACE AND LUGS TO

ACCOMMODATE QUANTITY AND SIZE OF CONDUCTORS REQUIRED. CONTRACTORS SHALL

PROVIDE EQUIPMENT WITH OVERSIZED ENCLOSURES WHERE REQUIRED.

23. THE CONTRACTOR SHALL BE RESPONSIBLE TO PROVIDE TYPE WRITTEN UPDATED PANEL

DIRECTORY WITH DATE AND INCLUDING ROOM NUMBER FOR ALL NEW EXISTING LOADS PER

NEC 408.4 TO BE MOUNTED ON INSIDE OF ALL PANEL DOOR COVERS. DIRECTORY SHALL

REFLECT ALL ADDITIONS OR MODIFICATIONS TO EXISTING PANELS AND SHALL REFLECT ACTUAL

"AS-BUILT" CONDITIONS. HANDWRITTEN DIRECTORIES OR MODIFICATIONS TO DIRECTORIES

ARE NOT ACCEPTABLE. LOAD DESCRIPTION SHALL NOT INCLUDE "EXISTING" OR "EXISTING

LOAD". ALL LOADS SHALL BE VERIFIED BY THE CONTRACTOR. PROVIDE DIRECTORIES AS A

CLOSE-OUT SUBMITTAL PER SPEICIFICATION SECTION 26 24 16 PANELBOARDS.

24. VERIFY DEVICE COLOR AND MOUNTING ORIENTATION (VERTICAL OR HORIZONTAL) WITH

ARCHITECTURAL AND INTERIOR DESIGN DRAWINGS PRIOR TO ORDERING ANY EQUIPMENT AND

PROVIDE DEVICES AS REQUIRED. UNLESS NOTED OTHERWISE, DEVICES AND DEVICE PLATES

SHALL BE WHITE IN COLOR.

25. WHERE MOTORS ARE INSTALLED IN SUSPENDED CEILINGS, CONTRACTOR SHALL PROVIDE

DISCONNECT SWITCH IN SUSPENDED CEILING WITHIN REACH FROM ACCESS POINT.

26. CONTRACTOR SHALL VERIFY ACTUAL MOTOR AND APPLIANCE RATING AND LOADS.

CONTRACTOR TO PROVIDE MOTOR OVERLOAD ELECTRICAL COMPONENTS BASED ON

NAMEPLATE RATING. REFLECT ANY CHANGES IN THE AS-BUILT DRAWINGS.

27. CONTRACTOR SHALL SUBMIT SHOP DRAWINGS TO ARCHITECT FOR REVIEW OF THE

FOLLOWING EQUIPMENT:

A. ELECTRICAL SWITCHGEAR: SWITCHBOARDS, PANELS, MOTOR CONTROL CENTERS AND

SAFETY SWITCH

B. TIME/CURRENT TRIP CURVES

C. LIGHTING FIXTURES: INDOOR/OUTDOOR AS PSECIFIED, PHOTOMETRIC PERFORMANCE

DATA AND LAMPS

D. DEVICES: SWITCHES, RECEPTACLES, MOTOR CONTROLLERS AND DEVICE PLATES

INCLUDING SHOP DRAWINGS FOR ALL LIGHTING CONTROLS.

E. LIFE SAFETY /FIRE ALARM SYSTEMS: CONTROL PANEL, ANNUNCIATOR PANEL,INITIATION

AND NOTIFICATION DEVICES / APPLIANCES, SYSTEM WIRING REQUIREMENTS AND

DIAGRAMS, SYSTEM LOCA CALCULATIONS, STAND-BY BATTERY CALCULATIONS, AND

AUXILIARY POWER SUPPLY

LIGHTING FIXTURE SCHEDULE

Type Mark Description Lamp Mounting Volts Manufacturer Model Wattage Notes

F1 2'X2' LED FIXTURE WITH 0-10V

DIMMING

LED RECESSED 120V DAYBRITE 2LTG27L840-2-21-UNV-D

IM-BAC

46 W

F1E 2'X2' LED FIXTURE WITH 0-10V

DIMMING

LED RECESSED 120V DAYBRITE 2LTG27L840-2-21-UNV-D

IM-BAC

46 W CONNECTED TO LIFE SAFETY

CIRCUIT

F2 6" LED DOWNLIGHT WITH 0-10V

DIMMING

LED RECESSED 120V LIGHTOLIER C6RN-BAC /

C6L25840WZ10U-BAC /

C6RDLCL-BAC

0 W

F2E 6" LED DOWNLIGHT WITH 0-10V

DIMMING

LED RECESSED 120V LIGHTOLIER C6RN-BAC /

C6L25840WZ10U-BAC /

C6RDLCL-BAC

0 W

F3 4' LINEAR LED STRIP FIXTURE WITH

0-10V DIMMING

LED SURFACE 120V DAYBRITE FSS440L840-UNV-DIM-B

AC

31 W

F3E 4' LINEAR LED STRIP FIXTURE WIRH

0-10V DIMMING

LED SURFACE 120V DAYBRITE FSS440L840-UNV-DIM-B

AC

31 W

X1 SINGLE FACE COMPACT EXIT SIGN GREEN SURFACE 120V CHLORIDE CLXNGW2C PROVIDE UNIVERSAL

MOUNTING KIT FOR SIDE,

END, TOP AND FACE

MOUNTING.

X-RAY X-RAY IN USE WARNING SIGN LED RECESSED 120V CHLORIDE AMS-120-GW-"X-RAY IN

USE"-CMB

COORDINATE REQUIRED

MOUNTING BRACKET FOR

SIDE OR CEILING MOUNT

Revisions: Date:

CEM1

H 11

5CE

L 8

PP4D

J5 5

4A

J3 4.2

4XC

E2 9.3

4XE

4CEXA

J5 5

225 A

150 AT

EMG SWITCH GEAR

800 A

800 AT

ATS 5

NE

800 A

70 AF

EDB1

800 A

800 AT

800 A

800 AT

1200A

1200AT

A B

100 A

70 AT

400 A

400 AT

100 A

100 AT

1200A TIE BREAKER

H 11

2CE

C D

1600 A

1200AT

225 A

150 AT

1200A TIE BREAKER

SWBD. B

225 A

150 AT

100 A

100 AT

100 A

70 AT

800 A

800 AT

SWBD. D

800 A

800 AT

225 A

200 AT

150 A

200 AF

J5 4.2

4XB

L 5.1

4XAA

L 5.1

4XA

D4A

N5 9.5

4D

60 A

60 AF

60 A

100 AF

L8

J5 4.2

4XCA

O1 9.5

4XQ

N 8

4XR

J3 5.1

4XD

150 A

200 AF

150 A

200 AF

150 A

200 AF

D4XA

L8

D4AA

N5 9.5

60 AF

60 A

4CCD

RM 4054

150 KVA

J5 14.1

150 A

200 AF

112.5 KVA

T16

T

D2 9.5D2 9.3

4XG

4XF

E2 10

4XH

200 A

200 AF

D4XB

H11

T17

4XK

J5 9.5

4XS

K 14.1

4XL

J5 13.3

4XM

J 13.3

4XJ

100 A

100 AF

175 A

200 AF

150 A

200 AF

T

800 A

800 AT800 AT

800 A 800 A

800 AT

100 AF

800 A

100 AF

800 A

CEM2-603

600 AF 250A400 AF

800 A

800 AF

800 A

600 AF 800 AF 100A225A

4CEXD

100 AF

100 A

H11

200 AF

150 A

200 AF

150 A

200 AF

175 A

100 AF

100 A

D4CEXA-234

T

4CEXB

E2 9.5

4CEXC

J5 13.3 O 9.5

T15-449

112.5 KVA

E N ATS 12

J5 5

4CEXA1

J5 5

4CEXAA

600 AF

800 A

BASEMENT

2ND INT'STIT.

1ST INT'STIT.

3RD INT'STIT.

4TH INT'STIT.

5TH INT'STIT.

4CEXE

9 N

ISC = 3,682A

ISC = 3,621A

ISC = 2,806A

ISC = 2,909 A

ISC = 5,408 A

ISC = 11,040A

ISC = 10,083AISC = 2,423A

NORMAL POWERCRITICAL POWER

SIEMENS

EQUIPMENT

PANEL

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

CONSTRUCTION DOCUMENTSOffice of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

664-CSI-1154W NUCLEAR MEDICINE PET- CT 2

SAN DIEGO, CA

IDS GROUP INC.

1 PETERS CANYON RD. STE. 130

IRVINE, CA 92606

949.387.8500

FULLY SPRINKLERED

C U s e rs \A le ja n d ro .A g u ila r\ D o c u m e n ts

\1 X

.0

-V

A S

D -N u c -M e d

E le c R

T V ra m i.

e lh a s s a n id s g i.

c o m .r v t

/2 /2

:5

:4

P

M

EP-002

SINGLE LINE DIAGRAM

10/02/20 RO GA

1. AS REQUIRED ALL OVERSIZED FEEDERS THAT WERE ADJUSTED IN SIZE TO COMPENSATE FOR VOLTAGE DROP SHALL BE PROVIDED WITH ADAPTER LUGS OR SPLICE BOX. ADAPTER LUGS SHALL BE PROVIDED IF SIZE IS AVAILABLE, OTHERWISE PROVIDE CABLE SPLICES IN THE SPLICE BOX TO

REDUCE CABLES TO THE MAXIMUM SIZE THAT THE BREAKER LUGS CAN ACCOMMODATE.

2. ALL ELECTRICAL SYSTEMS SHALL BE INSTALLED FOR FINAL INSPECTIONS.

3. ALL CONDUCTORS SHALL BE 600V. RATED WITH THHN/THWN INSULATION AT 75° CELSIUS.

4. ALL CIRCUIT BREAKER TERMINALS IN SWITCHBOARDS AND PANEL BOARDS SHALL BE U.L LISTED

AND APPROVED FOR USE COPPER 75° CELSIUS CONDUCTORS.

5. ALL FINAL TERMINATION'S OF CONDUCTORS TO ELECTRICAL EQUIPMENT AND DEVICES SHALL BE

TORQUE WRENCH TIGHTENED TO THE MANUFACTURER'S RECOMMENDED SPECIFICATION, NO

EXCEPTION.

6. MINIMUM AIC SHALL BE 10,000 FOR 208/120-VOLT EQUIPMENT UNLESS NOTED.

AFFECTED BY SCOPE OF WORK.

REPLACE EXISTING 400AT/800AF/3P TO 600AT/800AF/3P. CIRCUIT BREAKER SHALL BE SAME

MANUFACTURER AS EXISTING AND SHALL HAVE A MINIMUM AIC RATING OF EQUIPMENT'S

AIC RATING.

PROVIDE AN INSULATED CONTINUOUS COPPER CONDUCTOR SIZED PER NEC 250.122 AS

REQUIRED PER NEC 517.14 AND BOND ALL PANELBOARDS SERVING PATIENT CARE AREAS.

PROVIDE 3#3/0, 1#6 GND., 2"C.

NOT FOR CONSTRUCTION

GENERAL NOTES

KEYNOTES

NO SCALEPARTIAL SINGLE LINE DIAGRAM 1

1 1

(TYP)

(TYP)

(TYP)

ISSUED FOR 100% C.D. 2020-10-02

CRITICAL NORMAL NORMAL

CRITICAL

NORMAL

NORMAL

NORMAL

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

CONSTRUCTION DOCUMENTSOffice of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

664-CSI-1154W NUCLEAR MEDICINE PET- CT 2

SAN DIEGO, CA

IDS GROUP INC.

1 PETERS CANYON RD. STE. 130

IRVINE, CA 92606

949.387.8500

FULLY SPRINKLERED

C U s e rs \A le ja n d ro .A g u ila r\ D o c u m e n ts

\1 X

.0

-V

A S

D -N u c -M e d

E le c R

T V ra m i.

e lh a s s a n id s g i.

c o m .r v t

/2 /2

:5

:4

P

M

EP-003

ELECTRICAL PANEL SCHEDULES

10/02/20 RO GA

CONTRACTOR SHALL TRACE, LOCATE AND VERIFY LOADS. REVISE DIRECTORY CARD WITH

DESCRIPTION OF LOAD AND INCLUDE ROOM NUMBER PER NEC 408.4.

PROVIDE NEW CIRCUIT BREAKER SIZE AND TYPE INDICATED. CIRCUIT BREAKER SHALL MATCH

EXISTING IN MANUFACTURER, TYPE, STYLE AND AIC RATING, AND UPDATE PANEL CARD

INFORMATION.

DISCONNECT AND REMOVE EXISTING CONDUIT AND CONDUCTORS COMPLETELY. THERE

SHALL BE NO ABANDONED CONDUITS. LEAVE BREAKER AS SPARE, AND UPDATE PANEL CARD

INFORMATION.

REMOVE EXISTING CIRCUIT BREAKER 30A/2P AN RETURN TO THE OWNER. PROVIDE NEW

CIRCUIT BREAKER SIZE AND TYPE INDICATED. CIRCUIT BREAKER SHALL MATCH EXISTING IN

MANUFACTURER, TYPE, STYLE AND AIC RATING, AND UPDATE PANEL CARD INFORMATION.

REMOVE EXISTING CIRCUIT BREAKERS 20A/1P AN RETURN TO THE OWNER. PROVIDE NEW

CIRCUIT BREAKER SIZE AND TYPE INDICATED. CIRCUIT BREAKER SHALL MATCH EXISTING IN

MANUFACTURER, TYPE, STYLE AND AIC RATING, AND UPDATE PANEL CARD INFORMATION.

1. ALL PANELS ARE EXISTING TO REMAIN. MODIFY AS INDICATED.

NOT FOR CONSTRUCTION

GENERAL NOTES

KEYNOTES

PANEL 4CEXA

PANEL 4XCA

PANEL 4A

PANEL 4XB

PANEL 4XE

PANEL 5CE

PANEL 4XC

PANEL PP4D

ISSUED FOR 100% C.D. 2020-10-02

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

CONSTRUCTION DOCUMENTSOffice of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

664-CSI-1154W NUCLEAR MEDICINE PET- CT 2

SAN DIEGO, CA

IDS GROUP INC.

1 PETERS CANYON RD. STE. 130

IRVINE, CA 92606

949.387.8500

FULLY SPRINKLERED

C U s e rs \A le ja n d ro .A g u ila r\ D o c u m e n ts

\1 X

.0

-V

A S

D -N u c -M e d

E le c R

T V ra m i.

e lh a s s a n id s g i.

c o m .r v t

/2 /2

:5

:4

P

M

EP-004

VOLTAGE DROP CALCULATIONS &

LOAD SUMMARY

10/02/20 RO GA

NOT FOR CONSTRUCTION

NOT SCALEVOLTAGE DROP CALCULATION 2

NOT SCALELOAD SUMMARY 1NO SCALEFAULT CURRENT CALCULATIONS 3

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

CONSTRUCTION DOCUMENTSOffice of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

664-CSI-1154W NUCLEAR MEDICINE PET- CT 2

SAN DIEGO, CA

IDS GROUP INC.

1 PETERS CANYON RD. STE. 130

IRVINE, CA 92606

949.387.8500

FULLY SPRINKLERED

C U s e rs \A le ja n d ro .A g u ila r\ D o c u m e n ts

\1 X

.0

-V

A S

D -N u c -M e d

E le c R

T V ra m i.

e lh a s s a n id s g i.

c o m .r v t

/2 /2

:5

:4

P

M

EP-005

TITLE 24 COMPLIANCE DOCUMENTS

10/02/20 RO GA

UPDNUPDN

CAB DESK

1 2 2.1 3 3.1 3.2 4 4.2 5 5.1 6 7 7.1 8 9 9.1 9.5 9.4 10 11 11.1 12 13 13.3 14 14.1 15 15.1 15.3 16 16.116.3 17 18

B

B.1

C

D

D.2

D.4

E

E.2

E.3

F

G

G.1

H

J

J.1

J.5

J.4

K

L

L.1

M

N

N.2

N.4

N.5

N.6

N.7

O

O.1

P

Q

Q.1

R

S

EP-101

EP-101

FOURTH FLOOR

ELECTRICAL

ROOM

EP-103

EP-105

MH-102IA

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

CONSTRUCTION DOCUMENTSOffice of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

664-CSI-1154W NUCLEAR MEDICINE PET- CT 2

SAN DIEGO, CA

IDS GROUP INC.

1 PETERS CANYON RD. STE. 130

IRVINE, CA 92606

949.387.8500

FULLY SPRINKLERED

C U s e rs \A le ja n d ro .A g u ila r\ D o c u m e n ts

\1 X

.0

-V

A S

D -N u c -M e d

E le c R

T V ra m i.

e lh a s s a n id s g i.

c o m .r v t

/2 /2

:5

:4

P

M

EP-100

ELECTRICAL 4TH FLOOR - OVERALL

10/02/20 RO GA

1" = 20'-0"LEVEL 4 - OVERALL REFERENCE PLAN 1

N

1. ALL EXISTING ELECTRICAL INFORMATION SHOWN ON THIS PLAN IS BASED ON VISUAL SITE

SURVEYS AND AS-BUILT DRAWINGS. THE CONTRACTOR SHALL VERIFY AND CONFIRM ALL

WORK PRIOR TO ROUGH-IN.

2. PRIOR TO ANY WORK, THE CONTRACTOR SHALL VERIFY AND CONFIRM ALL EXISTING

CIRCUITS AND LOADS PRIOR TO DISCONNECTING POWER,OUTAGES, OR THE LIKE OF. THE

CONTRACTOR SHALL VERIFY AND CONFIRM LOCATIONS AND REQUIREMENTS OF EXISTING TO

REMAIN DEVICES AND EQUIPMENT IN BOTH AREAS OF WORK AND AREAS OUTSIDE OF AREA

OF WORK. PROVIDE ADDITIONAL CONDUIT AND CONDUCTORS, PERFORM SPLICES, AND

EXTEND CIRCUITS AS NECESSARY TO MAINTAIN THE CONTINUITY OF CIRCUIT TO EXISTING TO

REMAIN DEVICES AND EQUIPMENT.

NOT FOR CONSTRUCTION

GENERAL NOTES

KEYPLAN - LEVEL 4 INTERSTITIAL

A4012

CORRIDOR

A4021.4

CORRIDOR

A4018.1

CORRIDOR

A4018.4

CORRIDOR

A4018.2

A4018

STAIR

NO 9

CORRIDOR

A4018.3

A4012

CORRI DOR

A4021.4

CORRI DOR

A4018.1

CORRI DOR

A4018.4

CORRI DOR

A4018.2

A4018

STAIR

NO 9

CORRI DOR

A4018.3

9.1

9.1

E.2 E.2

E.3 E.3

F F

(E)4XE

120/208V

225A BUS

3PH, 4W

7 7.1 8 9

K

L

L.1

(E) PANEL PP4C

120/208V

400A BUS 3PH, 4W

(E) PANEL PP4A

277/480V

400A BUS 3PH, 4W

(E) PANEL PP4AA

277/480V

225A BUS 3PH, 4W

(E) PANEL D4XA

120/208V

400A BUS 3PH, 4W

(E) PANEL PP4D

277/480V

400A BUS3PH, 4W

(E)TELEPHONE

BACKBOARD

(E)TELEPHONE

BACKBOARD

(E) PANEL D4A

277/480V

225A BUS 3PH, 4W

(E) T16 480-208/120V

112.5 KVA

3PH, 4W

(E) T4 480-208/120V

75 KVA

3PH, 4W

4.2 5

J.5

(E) PANEL 4CEXA

120/208V

225A BUS 3PH, 4W

(E)4XB

120/208V

225A BUS

3PH, 4W

(E)4XCA

120/208V

225A BUS

3PH, 4W

(E)4XC

120/208V

225A BUS

3PH, 4W

(E) PANEL 4A

277/480V

225A BUS 3PH, 4W

(E) PANEL A

120/208V

225A BUS 3PH, 4W

(E) PANEL UPS

120/208V

100A BUS 3PH, 4W

CONTRACTOR SHALL COMPLETELY

REMOVE EXISTING ABANDONED CABLE

TRAY, ASSOCIATED SUPPORTS,

SEISMIC BRACING, ANCHORS, AND

APPURTENANCES. PATCH AND REPAIR

SURFACES.

(E) LOW VOLTAGE PANELS

MH-102IA

EP-101

EP-101

EP-101

EP-101

5 5.1

G.1

H

J

CONTRACTOR SHALL

COMPLETELY REMOVE EXISTING

ABANDONED CABLE TRAY,

ASSOCIATED SUPPORTS, SEISMIC

BRACING, ANCHORS, AND

APPURTENANCES. PATCH AND

REPAIR SURFACES.

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

CONSTRUCTION DOCUMENTSOffice of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

664-CSI-1154W NUCLEAR MEDICINE PET- CT 2

SAN DIEGO, CA

IDS GROUP INC.

1 PETERS CANYON RD. STE. 130

IRVINE, CA 92606

949.387.8500

FULLY SPRINKLERED

C U s e rs \A le ja n d ro .A g u ila r\ D o c u m e n ts

\1 X

.0

-V

A S

D -N u c -M e d

E le c R

T V ra m i.

e lh a s s a n id s g i.

c o m .r v t

/2 /2

:5

:4

P

M

EP-101

ELECTRICAL 4TH INTERSTITIAL -

ENLARGED PLANS

10/02/20 RO GA

AFFECTED BY SCOPE OF WORK. REFER TO PANEL SCHEDULE ON EP-003

REFER TO PHOTO

1. ALL EXISTING ELECTRICAL INFORMATION SHOWN ON THIS PLAN IS BASED ON VISUAL SITE

SURVEYS AND AS-BUILT DRAWINGS.

1/4" = 1'-0"LEVEL 4 - INTERSTITIAL - PANEL LOCATION 1 3 1/4" = 1'-0"LEVEL 4 - INTERSTITIAL ELECT. RM. A4018 1

1/4" = 1'-0"LEVEL 4 - INTERSTITIAL - PANEL LOCATION 2 2

N N

N

NOT FOR CONSTRUCTION

GENERAL NOTES

KEYNOTES

KEYPLAN - LEVEL 4 INTERSTITIAL

1 1

ISSUED FOR 100% C.D. 2020-10-02

1/4" = 1'-0"LEVEL 4 - INTERSTITIAL - PLATFORM AREA 4

J

J

SCAN

EXAM RM

NUCMED ALCOVE

EQUIPMENT RM

CORRIDOR

TOILET

4070B

LAB

4070D

EXAM RM

4070A

LAB

4070C

RECEPTION

CORRIDOR

HAC

GAMMA WELL

COUNTER

4500A

LAB

SCAN

EXAM RM

CLEAN UTILITY

TOILET

OFFICE

BREAK

TOILET

CONFERENCE

OFFICE (S)

OFFICE (S)

OFFICE (S)

OFFICE (S)

SCOPE OF WORK

(E)POWER OUTLETS

COMPUTER RM

(E) RESTROOM

(E)

(E) (E)

(E)

(E)

(D)

(D)

(D)

(D)

(D)

(D)

(D)

(D)

(D)

(D)(D)(D)(D)(D)

(D)

(D)

(D) (D)(D)(D)

SCAN

(E)

EP-102

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

CONSTRUCTION DOCUMENTSOffice of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

664-CSI-1154W NUCLEAR MEDICINE PET- CT 2

SAN DIEGO, CA

IDS GROUP INC.

1 PETERS CANYON RD. STE. 130

IRVINE, CA 92606

949.387.8500

FULLY SPRINKLERED

C U s e rs \A le ja n d ro .A g u ila r\ D o c u m e n ts

\1 X

.0

-V

A S

D -N u c -M e d

E le c R

T V ra m i.

e lh a s s a n id s g i.

c o m .r v t

/2 /2

:5

:5

P

M

EP-102

ELECTRICAL 4TH FLOOR - AREA OF

WORK DEMOLITION

10/02/20 RO GA

1/4" = 1'-0"LEVEL 4 - AREA OF WORK DEMOLITION 1

DISCONNECT AND REMOVE EXISTING COMMUNICATION DEVICE, PLATE AND DEVICE BOX.

REMOVE CABLING AND ALL FITTINGS, BACK TO THE SERVICE POINT.

DISCONNECT AND REMOVE EXISTING DISCONNECT SWITCH. REMOVE EXISTING WIRE AND

CONDUIT, BACK TO THE SERVICE POINT. REMOVE ALL FITTING, ATTACHING HARDWARE,

BRACKETS, SUPPORTS AND APPLICABLE APPURTENANCES.

REMOVE EXISTING SWITCH DEVICE, PLATE AND DEVICE BOX. REMOVE EXISTING WIRING AND

ALL EXPOSED CONDUIT.

DISCONNECT AND REMOVE ALL LIGHTING FIXTURES. JUNCTION BOXES, RECEPTACLES.

COMPLETELY REMOVE ALL EXPOSED ATTACHING HARDWARE, BRACKETS, SUPPORT,

PROTRUDING BOLTS AND SLEEVES. MAINTAIN CONTINUITY OF CIRCUIT TO BE USED ON NEW

LIGHT CIRCUITS. REFER TO NEW CONSTRUCTION PLAN FOR MORE INFORMATION.

DISCONNECT AND REMOVE EXISTING EQUIPMENT, SALVAGE AND RETURN TO THE OWNER.

DISCONNECT AND RELOCATE EXISTING UPS PANEL. COORDINATE LOCATION WITH OWNER

PRIOR TO ROUGH-IN.

DISCONNECT AND RELOCATE EXISTING OUTLETS. ACCOMMODATE ELECTRICAL DEVICES,

FOLLOWING THE NEW ROOM DISTRIBUTION. MAINTAIN CONTINUITY OF CIRCUIT TO BE USED ON

NEW POWER CIRCUITS. REFER TO NEW CONSTRUCTION PLAN FOR MORE INFORMATION.

DISCONNECT AND RELOCATE EXISTING LIGHTING FIXTURES. ACCOMMODATE LIGHTING

FIXTURES, FOLLOWING THE NEW ROOM DISTRIBUTION. MAINTAIN CONTINUITY OF CIRCUIT TO

BE USED ON THE NEW LIGHTING FIXTURE LOCATION. REFER TO NEW CONSTRUCTION PLAN FOR

MORE INFORMATION.

EXISTING DUPLEX RECEPTACLE AND FACEPLATE TO BE REMOVED. SAFE-OFF EXISTING BRANCH

CIRCUIT AND CONDUCTORS. MAINTAIN EXISTING BACKBOX AND CONDUIT. SEE EP-104 FOR NEW

WORK.

DISCONNECT CEILING OUTLETS AND PROTECT IN PLACE. MAINTAIN CONTINUITY OF CIRCUIT

FOR BEEN RECONNECTED AFTER DEMOLITION IS COMPLETED. REFER TO NEW CONSTRUCTION

PLAN FOR MORE INFORMATION. COORDINATE WITH ELECTRICAL CONTRACTOR FOR

RECONNECTION.

N

1 2 3 4

1. ALL EXISTING ELECTRICAL INFORMATION SHOWN ON THIS PLAN IS BASED ON VISUAL SITE

SURVEYS AND AS-BUILT DRAWINGS.

NOT FOR CONSTRUCTION

GENERAL NOTES

KEY NOTES

KEYPLAN - LEVEL 4

N

9 9

D

D D

J

J

OS

OS

OS

OS

OSOS

J

JJJJ

EP-103

2 2.1 3 3.1 3.2 4 4.2 5 5.1

J

J.1

J.5

J.4

SCOPE OF WORK.

(TYP)

SCOPE OF WORK

a a

F3

(E) RESTROOM

4054TOILET

4070B

GAMMA WELL

COUNTER

4500ALAB

NUCMED ALCOVE

CORRIDOR

RECEPTION

OFFICE (S)

OFFICE (S)

OFFICE

TOILET

OFFICE (S)

OFFICE (S)

CORRIDOR

BREAK

CLEAN UTILITY

EXAM RM

CONFERENCE

TOILET

SCAN

CORRIDOR

SCAN

EXAM RM

COMPUTER RM

EQUIPMENT RM

F3E

PDC

X1

X1

EQUIPMENT RM

SCAN

CONTROL

SCAN

X-RAY

F1

F1

F1E

F1 F2

F2E

F2E F2 F2

F2

F1 F1E

F1 F1

F2

F2 F2

F2E

F2

F1E

F1

F2E

F2E

F1

F1E

5CE - 1

a a a b b b ab

1 a 1 a b b bb b b

F2E

5CE - 1

4XE - 14

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

CONSTRUCTION DOCUMENTSOffice of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

664-CSI-1154W NUCLEAR MEDICINE PET- CT 2

SAN DIEGO, CA

IDS GROUP INC.

1 PETERS CANYON RD. STE. 130

IRVINE, CA 92606

949.387.8500

FULLY SPRINKLERED

C U s e rs \A le ja n d ro .A g u ila r\ D o c u m e n ts

\1 X

.0

-V

A S

D -N u c -M e d

E le c R

T V ra m i.

e lh a s s a n id s g i.

c o m .r v t

/2 /2

:5

:5

P

M

EP-103

4TH FLOOR-AREA OF WORK

REMODEL- LIGHTING

10/02/20 RO GA

RE-CONNECT NEW LIGHTS TO EXISTING CIRCUIT J-BOX (277V) IN CEILING SPACE. VERIFY CIRCUIT

NUMBER IN FIELD.

PROVIDE (2)-#10 & (1)-#10 GND IN EXISTING 3/4"C.

PROVIDE ENGRAVED SWITCH WITH SEPARATE ON/OFF/DIM 0-10V CONTROLS FOR GENERAL

LIGHTING AND FOR DOWNLIGHTING . LOCATE AT SCAN ENTRANCE AND CONTROL ROOM FOR

CONTROL FROM BOTH LOCAITONS.

PROVIDE WARNING LIGHT WITH WORDING "X-RAY IN USE, DO NOT ENTER". PROVIDE INTERFACE

WITH CT CONTROLLER VIA INTERFACE RELAY. WARNING LIGHT TO BE PROVIDED BY OTHERS.

INTERCEPT EXISTING BRANCH CIRCUIT AND EXTEND AS INDICATED.

PROVIDE (2)-#12 & (1)-#12 GND IN 3/4"C. FOR LOW VOLTAGE TRANSFORMER. COORDINATE

LOCATION WITH OWNER PRIOR TO ROUGH-IN.

PROVIDE NURSE CALL LIGHT. INSTALL 2" ABOVE DOOR JAMB AND CENTERED ON DOOR.

PROVIDE 120V TO WALL MOUNTED LIGHT PANELS MANUFACTURER SKY FACTORY SERIES #ALS-

BL-TWIN-2240-4IV. COORDINATE WITH ARCHITECTURAL PLANS FOR EXACT LOCATIONS AND

INSTALLATION REQUIREMENTS.

PROVIDE 120V TO LUMINOUS VIRTUAL WINDOW. COORDINATE CONNECTION REQUIREMENTS

WITH VENDOR PRIOR TO ROUGH-IN.

N

1. ALL EXISTING ELECTRICAL INFORMATION SHOWN ON THIS PLAN IS BASED ON VISUAL SITE SURVEYS

AND AS-BUILT DRAWINGS.

2. MULTIPLE HOME RUNS MAYBE SHOWN ON THESE PLANS FOR CLARITY. E.C. IS TO COMBINE HOME RUN CONDUIT AND WIRES SO AS TO OBSERVE ALL REQUIREMENTS FOR N.E.C.110.14. IT IS NOT THE

INTENTION OF THIS DESIGN TO ALLOW MULTIPLE TAPPED CIRCUIT BREAKERS.

3. CONTRACTOR TO VERIFY ALL CONDITIONS, LOCATIONS, AND CONNECTIONS TO EQUIPMENT WITH

OWNER AND/OR ARCHITECT PRIOR TO BID AND INCLUDE ALL COST IN BID.

4. PROVIDE SHIELDING BEHIND ALL BOXES AND OTHER PENETRATIONS IN SHIELDED SCANNING

ROOM SURFACES.

(TYP)

NOT FOR CONSTRUCTION

GENERAL NOTES

KEY NOTES

KEYPLAN - LEVEL 4

1/4" = 1'-0"LEVEL 4 - AREA OF WORK - REMODEL LIGHTING 1

ISSUED FOR 100% C.D. 2020-10-02

T T

T

T

T T

T

G F

C I

G F

C I

G F

C I

GFCI

J

J

J

J

J

JJ

J

J

J

2.1 3 3.1 3.2 4 4.2 5 5.1

H

J

J.1

J.5

J.4

SCAN

EXAM RM

NUCMED ALCOVE

EQUIPMENT RM

CORRIDOR

TOILET

4070B

LAB

4070D

EXAM RM

4070A

LAB

4070C

RECEPTION

CORRIDOR

HAC

GAMMA WELL

COUNTER

4500A

LAB

SCAN

EXAM RM

CLEAN UTILITY

TOILET

OFFICE

BREAK

TOILET

CONFERENCE

OFFICE (S)

OFFICE (S)

OFFICE (S)

OFFICE (S)

COMPUTER RM

SCOPE OF WORK.

19 19 19

21 21

SCOPE OF WORK

(E) RESTROOM

LAB

4070C

AC

VOLTAGE LOAD

AC

4XE - 19

4XE - 21

4XCA - 22

EQUIPMENT RM

4XCA - 24

4XCA - 2424

CORRIDOR

PP4D - 18,20,22

(N)HEAT EXCHANGER CABINET

PP4D - 7,9,11

PP4D - 2,4,6

EPO

4XE - 1,3

4XE - 1,3

4XE - 5

BLANKET

WARMER

4XCA - 24

+48"

+48"

+48"

PDC

cPDC

4XE - 15

4XE - 17

4CEXA - 8

+48"

4CEXA - 11

4CEXA - 13

EPO

+48"

DS

(E)

(E) (E)

(E)

(E)

EPO

CONTROL

COMMON POWER

DISTRIBUTION UNIT

POWER DISTRIBUTION

CABINET (PDC) AND UPS

LEAD

PIG

IMAGE

RECONSTRUCTION

SYSTEM

PATIENT TABLE

SCAN

KEYBOARD

& CONTROL

BOX

MEDRAD

WORKSTATION

& BASE UNIT

GATING LAPTOP

COMPUTER & DECK PORT

HEAT EXCHANGER

WATER/AIR SPLIT (ID#11)

4XE - 2,4,6

4XE - 17

4XE - 7,15,8

(E)

4CEXA -

12,14,16,18

4CEXA - 1

+48"

EPO 5

4CEXA - 3

EP-104

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

CONSTRUCTION DOCUMENTSOffice of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

664-CSI-1154W NUCLEAR MEDICINE PET- CT 2

SAN DIEGO, CA

IDS GROUP INC.

1 PETERS CANYON RD. STE. 130

IRVINE, CA 92606

949.387.8500

FULLY SPRINKLERED

C U s e rs \A le ja n d ro .A g u ila r\ D o c u m e n ts

\1 X

.0

-V

A S

D -N u c -M e d

E le c R

T V ra m i.

e lh a s s a n id s g i.

c o m .r v t

/2 /2

:5

:5

P

M

EP-104

4TH FLOOR-AREA OF WORK

REMODEL-POWER

10/02/20 RO GA

1/4" = 1'-0"LEVEL 4 - AREA OF WORK - REMODEL POWER 1

N

NOT FOR CONSTRUCTION

PROVIDE POWER CONNECTION ABOVE CEILING TO MODULAR SERVICE COLUMN.

EXISTING OUTLET TO REMAIN.

PROVIDE EMERGENCY POWER OFF PUSH BUTTON STATION A22EL-M-12A-11-B, WITH ELECTRIC

GUARD, METAL FRAME AND A TEXT ABOVE WHICH READS "EMERGENCY SHUT OFF". CONNECT

TO SHUNT TRIP AT MAIN DISCONNECT.

COORDINATE WITH VENDOR FOR EXACT LOCATION OF CONDUIT ENTRY WINDOW IN SLAB.

PROVIDE 1" CONDUIT AND FLOOR BOX ROUTED IN INTERSTITIAL SPACE BELOW. ROUTE FROM

PDC UNIT AT EQUIPMENT ROOM 4520 TO LOCATION OF PATIENT TABLE. CONDUIT SHALL BE

ROUTED WITHIN 12" BELOW UNDERSIDE OF SLAB BELOW. CONDUITS BELOW 12" SHALL NOT BE

ALLOWED WITHOUT THE CONTRACTOR PROVIDING ADDITIONAL SEISMIC BRACING OF CONDUIT

FOR SEISMIC ZONE 4.

PROVIDE (4)-#1/0 & (1)-#6 GND IN 2"C.

PROVIDE DOOR SWITCH PER DETAIL X/E-XX FOR SAFETY CONTROL OF CT SYSTEM. POWER TO

THE CT EQUIPMENT SHALL BE INTERRUPTED IF DOOR IS OPENED VIA SHUNT TRIP DEVICE.

PROVIDE (3)-#10 & (1)-#10 GND IN 3/4"C.

PROVIDE 200AS/200AF/3P STAINLESS STEEL FUSED DISCONNECT AND LIQUIDTIGHT FLEXIBLE

METAL CONDUIT OR LFMC FOR POWER CONNECTION TO UNIT. PROVIDE FUSE SIZE PER

MANUFACTURER'S RECOMMENDATION.

PROVIDE 200AS/150AF/3P STAINLESS STEEL FUSED DISCONNECT AND LIQUIDTIGHT FLEXIBLE

METAL CONDUIT OR LFMC FOR POWER CONNECTION TO UNIT. PROVIDE FUSE SIZE PER

MANUFACTURER'S RECOMMENDATION. COORDINATE EXACT LOCATION WITH SIEMENS AND

ARCHITECT PRIOR ROUGH-IN.

PROVIDE 30AS/30AF/3P STAINLESS STEEL FUSED DISCONNECT AND LIQUIDTIGHT FLEXIBLE

METAL CONDUIT OR LFMC FOR POWER CONNECTION TO UNIT. PROVIDE FUSE SIZE PER

MANUFACTURER'S RECOMMENDATION.

PROVIDE RECESSED CEILING MOUNT DUPLEX CONVENIENCE OUTLET 125V, 20A TYPE 5-20R FOR

OVERHEAD LIFT SYSTEM. COORDINATE EXACT LOCATION PRIOR TO ROUGH-IN.

PROVIDE DOUBLE POLE THERMAL RATED MOTOR STARTER SWITCH.

PROVIDE POWER TO AED. COORDINATE EXACT LOCATION WITH OWNER PRIOR ROUGH-IN.

RE-CONNECT CEILING OUTLET TO NEW OVERHEAD LIFT SYSTEM AS INDICATED. COORDINATE

EXACT LOCATION PRIOR TO ROUGH-IN.

RE-CONNECT WALL OUTLET AS INDICATED.

PROVIDE SURFACE MOUNTED 10"X3-1/2" METALLIC DUCT AT THE FLOOR LINE FOR ENTIRE

LENGTH OF DESK. DUCT SHALL HAVE THREE SEPARATE DIVIDED SECTIONS. COORDINATE WITH

SIEMENS VENDOR.

PROVIDE (4) 120V DEDICATED CIRCUITS TO THE MEDRAD BOOM PER DETAIL 2/EP-601.

PROVIDE (4) CAT5E ETHERNET CONNECTIONS PER DETAIL 2/EP-601.

1. RECEPTACLES ON COMMON WALLS SHALL BE SEPARATE BOXES AND OFFSET 12" MINIMUM. FIRE RATED WALLS SHALL HAVE AN OFFSET OF 24" MINIMUM. COORDINATE w/ ARCH'L FIRE RATED WALL

DETAILS

2. PENETRATIONS THROUGH FIRE RATED WALLS SHALL BE PROTECTED FROM THE SPREAD OF FIRE W/ AN APPROVED FIRESTOP SYSTEM EQUAL OR GREATER THAN THE FIRE RATING OF THE WALL.

3. PROVIDE 1-1/2" C.O. FROM EACH TELE/DATA OUTLET AND STUB 6" INTO ACCESIBLE CEILING SPACE, U.N.O. SEE DETAIL 3/EP-601 FOR FURTHER REQUIREMENTS.

4. REFER TO ARCHITECTURAL PLAN FOR ELECTRICAL DEVICE HEIGHTS AND EXACT LOCATION.

5. PROVIDE SHIELDING FOR ALL BOXES, CONDUITS AND OTHER PENETRATIONS IN SHIELDED

SCANNING ROOM SURFACES. REFER TO SHIELDING REQUIREMENTS AND DETAILS ON

ARCHITECTURAL SHEET AE-608.

6. PROVIDE 3/4" C.O. FOR EACH THERMOSTAT BACK TO EACH CORRESPONDING HVAC UNIT AND

BETWEEN MECHANICAL UNITS AS NEEDED FOR CONTROLS. REFER TO MECHANICAL DRAWINGS

FOR EXACT QUANTITY AND LOCATION.

GENERAL NOTES

KEY NOTES

KEYPLAN - LEVEL 4

15 15

ISSUED FOR 100% C.D. 2020-10-02

Revisions: Date:

J

J

J

J

J

J

J

EP-105

2.1 3 3.1 3.2 4 4.2 5 5.1

G

G.1

H

J

J.1

J.5

CAV

4-5

CAV

4-2

CAV

4-1

CAV

4-4

GX

AC

AV

(E)

0.40

AV

(E)

0.30

DV

(E)

2.23

AV

(E)

0.30

BV

(E)

0.55

AV

(E)

0.30

AV

(E)

0.30

AV

(E)

0.30

AC

(E)

PP4D - 1,3,5

4XCA - 15

4XCA - 31,33

31,33 31,33

31,33

31,33

31,33

4XCA - 40,42

GX

(E)

AC

DV

(E)

1.30

BV

(E)

0.63

AV

(E)

0.27

BV

(E) 0.55

VAV

E

4-5

VAV

E

VAV

E

VAV

E

VAV

E

VAV

E

AV

(E)

0.29

CAV

(E)

4-3

BV

(E)

0.55

BV

(E)

0.65

F3

F3

F3E

+6'-6" A.F.F.

+6'-6" A.F.F.

+6'-6" A.F.F.

4CEXA - 10

4CEXA - 17

TO RH SENSOR

RH SENSOR

AND AIR FLOW

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

CONSTRUCTION DOCUMENTSOffice of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

664-CSI-1154W NUCLEAR MEDICINE PET- CT 2

SAN DIEGO, CA

IDS GROUP INC.

1 PETERS CANYON RD. STE. 130

IRVINE, CA 92606

949.387.8500

FULLY SPRINKLERED

C U s e rs \A le ja n d ro .A g u ila r\ D o c u m e n ts

\1 X

.0

-V

A S

D -N u c -M e d

E le c R

T V ra m i.

e lh a s s a n id s g i.

c o m .r v t

/2 /2

:5

:5

P

M

EP-105

ELECTRICAL ENLARGED PLAN -

INTERSTITIAL

10/02/20 RO GA

PROVIDE (2)-#12 & (1)-#12 GND IN 3/4"C.

PROVIDE 60AS/40AF/3P STAINLESS STEEL FUSED DISCONNECT AND LIQUIDTIGHT FLEXIBLE

METAL CONDUIT OR LFMC FOR POWER CONNECTION TO UNIT. PROVIDE FUSE SIZE PER

MANUFACTURER'S RECOMMENDATION. COORDINATE EXACT LOCATION WITH ARCHITECT PRIOR

ROUGH-IN.

PROVIDE (3)-#8 & (1)-#10 GND IN 1"C.

PROVIDE GROUND-FAULT PROTECTED DUPLEX CONVENIENCE OUTLET 125V, 20A TYPE 5-20R IN

WEATHERPROOF BOX WITH WEATHERPROOF COVER. COORDINATE EXACT LOCATION PRIOR TO

ROUGH-IN.

PROVIDE 1"C.O. FOR LOW VOLTAGE HVAC CONTROLS.

PROVIDE (2) 1"C.O. FOR LOW VOLTAGE HVAC CONTROLS AND F.P. LIFE SAFETY INTERLOCK.

PROVIDE 1"C.O. AND CAT5E ETHERNET CONNECTION TO IDF ROOM.

GENERAL NOTES

KEY NOTES

1. ALL EXISTING ELECTRICAL INFORMATION SHOWN ON THIS PLAN IS BASED ON VISUAL SITE

SURVEYS AND AS-BUILT DRAWINGS.

2. THE CONTRACTOR SHALL COORDINATE WITH THE ARCHITECT ALL CONDUIT AND PIPING

ROUTING THROUGH THE INTERSTITIAL SPACE, CONDUIT CHASE, ABOVE CEILING AREAS NOT

ACCESSIBLE AT THE INTERSTITIAL SPACE, AND PROVIDE A CONDUIT ROUTING PLAN FOR

REVIEW AND APPROVAL BY THE AOR AND COR PRIOR TO ROUGH-IN.

NOT FOR CONSTRUCTION

N

KEYPLAN - LEVEL 4 INTERSTITIAL

1/4" = 1'-0"LEVEL 4 INTERSTITIAL - AREA A 1

ISSUED FOR 100% C.D. 2020-10-02

T

W P

WP

J

4.2 5

J.4

K

L

TRANSFER

CENTER

RECEPTION

CORRIDOR

HP

4CEXA - 7,9

4XCA - 24 WP

SCAN

5 5.1

G

G.1

H

J

OFFICE (S)

6083B

HP

WP

4XE - 8,10

4XE - 26

EP-106

1LEVEL 4

EP-106

LEVEL 6

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

CONSTRUCTION DOCUMENTSOffice of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

664-CSI-1154W NUCLEAR MEDICINE PET- CT 2

SAN DIEGO, CA

IDS GROUP INC.

1 PETERS CANYON RD. STE. 130

IRVINE, CA 92606

949.387.8500

FULLY SPRINKLERED

C U s e rs \A le ja n d ro .A g u ila r\ D o c u m e n ts

\1 X

.0

-V

A S

D -N u c -M e d

E le c R

T V ra m i.

e lh a s s a n id s g i.

c o m .r v t

/2 /2

:5

:0

P

M

EP-106

ELECTRICAL ENLARGED PLAN -

EXTERIOR

10/02/20 RO GA

N

NOT FOR CONSTRUCTION

1. PENETRATIONS THROUGH FIRE RATED WALLS SHALL BE PROTECTED FROM THE SPREAD OF FIRE W/ AN APPROVED FIRESTOP SYSTEM EQUAL OR GREATER THAN THE FIRE RATING OF THE WALL.

2. REFER TO ARCHITECTURAL PLAN FOR ELECTRICAL DEVICE HEIGHTS AND EXACT LOCATION.

3. COORDINATE WITH MECHANICAL CONTRACTOR AND ROUTE BRANCH CIRCUIT CONDUITS WITH

PRIMARY COOLING CIRCUIT PIPES. VERIFY ROUTE IN PIPE CHASE PRIOR TO WORK.

GENERAL NOTES

KEYPLAN - LEVEL 4

1/4" = 1'-0"LEVEL 4 - AREA OF WORK REMODEL 1

PROVIDE (2)-#12 & (1)-#12 GND IN 3/4"C.

PROVIDE WEATHER PROOF DOUBLE POLE THERMAL RATED MOTOR STARTER SWITCH.

PROVIDE GROUND-FAULT PROTECTED DUPLEX CONVENIENCE OUTLET 125V, 20A TYPE

5-20R IN WEATHERPROOF BOX WITH WEATHER PROOF COVER.COORDINATE EXACT

LOCATION PRIOR TO ROUGH-IN.

COORDINATE WITH VA FACILITY FOR ACCESS INTO THE PIPE CHASE. ONLY QUALIFIED AND

TRAINED PERSONNEL SHALL HAVE ACCESS INTO THE PIPE CHASE.

COORDINATE ALL PIPING AND BRANCH CIRCUIT CONDUITS ORIGINATING FROM LEVEL D6 WITH

WORK IN THIS AREA AND PROVIDE SINGLE ROUTE AND RACK.

PROVIDE (2) 3" CONDUITS TO HEAT EXCHANGER COORDINATE WITH SIEMENS VENDOR.

PLAN NOTES

1/4" = 1'-0"LEVEL 6 - AREA OF WORK REMODEL 2

N

ISSUED FOR 100% C.D. 2020-10-02

PROVIDE (2)-#12 & (1)-#12 GND IN 3/4"C.

PROVIDE WEATHER PROOF DOUBLE POLE THERMAL RATED MOTOR STARTER SWITCH.

PROVIDE GROUND-FAULT PROTECTED DUPLEX CONVENIENCE OUTLET 125V, 20A TYPE

5-20R IN WEATHERPROOF BOX WITH WEATHER PROOF COVER.COORDINATE EXACT

LOCATION PRIOR TO ROUGH-IN.

COORDINATE WITH VA FACILITY FOR ACCESS INTO THE PIPE CHASE. ONLY QUALIFIED AND

TRAINED PERSONNEL SHALL HAVE ACCESS INTO THE PIPE CHASE.

COORDINATE ALL PIPING AND BRANCH CIRCUIT CONDUITS ORIGINATING FROM LEVEL D6 WITH

WORK IN THIS AREA AND PROVIDE SINGLE ROUTE AND RACK.

PROVIDE (2) 3" CONDUITS TO HEAT EXCHANGER COORDINATE WITH SIEMENS VENDOR.

PROVIDE (2) 3" C.O. TO THE LOCATION OF THE HEAT EXCHANGER CABONET PER THE

REQUIREMENTS OF SEIMENS EQUIPMENT SUPPLIER.

PLAN NOTES

Revisions: Date:

nPODM

DX

nP D nPD

ER

nP D nPD

ER

1nPODM

DX

nP D nPD

ER

nPD

ER

nPODM

2P DX

nPD

ER

nPODM

DX

nPODM

2P DX

nPODM

DX

nP D nPD

ER

nP D

J

3.2 4 4.2 5 5.1

J

J.1

J.5

J.4

EXAM RM

COMPUTER RM

EQUIPMENT RM

SCAN

CONTROL

EP-102

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

CONSTRUCTION DOCUMENTSOffice of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

664-CSI-1154W NUCLEAR MEDICINE PET- CT 2

SAN DIEGO, CA

IDS GROUP INC.

1 PETERS CANYON RD. STE. 130

IRVINE, CA 92606

949.387.8500

FULLY SPRINKLERED

C U s e rs \A le ja n d ro .A g u ila r\ D o c u m e n ts

\1 X

.0

-V

A S

D -N u c -M e d

E le c R

T V ra m i.

e lh a s s a n id s g i.

c o m .r v t

/2 /2

:5

:0

P

M

EP-300

ELECTRICAL LIGHTING CONTROL

10/02/20 RO GA

CONTROLS LINETYPE LEGEND

SECONDARY DAYLIT ZONE BOUNDARY

SKYLIGHT DAYLIT ZONE BOUNDARY

nLIGHT AIR GROUP PAIRING

SYA CABLE A4

LOW VOLTAGE WIRING - SEE DRAWINGS FOR WIRING DETAILS

DMX512 - WIRING MUST FOLLOW THE DMX512-A/RDM STANDARD - ANSI E1.1

FRESCO NETWORK - CAT5, CAT5E OR CAT6; TERMINATED AS RJ45 TIA/EIA-568B

nLIGHT NETWORK - CAT5, CAT5E OR CAT6; TERMINATED AS RJ45 TIA/EIA-568B

PRIMARY DAYLIT ZONE BOUNDARY

DAYLIGHT LINETYPE LEGEND

CONTROL NOTES

THE LIGHTING CONTROL DESIGN IS SUBJECT TO CHANGE BASED ON ADDITIONAL

PROJECT SPECIFIC INFORMATION.

• THIS IS A STAND ALONE NLIGHT DESIGN- TIME CLOCK AND ADR EQUIPMENT NOT INCLUDED

• PLUG LOAD CONTROLLERS NOT INCLUDED IN THIS DESIGN

• ASSUMING 0-10V DIMMING ON ALL FIXTURES

0-10V CONTINUOUS DIMMING PROVIDED IN LIEU OF A/B SWITCHING

SYMBOLS LIST

nPODM

2P DX

nPODM 2P DX [COLOR] 2 Channel On/Off Toggle With Dimming nPODM

DX

nPODM DX [COLOR] 1 Channel On/Off Toggle With Dimming nPP16 D Power/Relay Pack: 16A 120/277 VAC w/ 0-10VDC Dimming nP D nPP16 D ER Relay Pack: 16A 120/277 VAC UL-924 Emergency Relay Pack w/ 0-10VDC Dimming nPD

ER

NOT FOR CONSTRUCTION

N a ae a a ae a ae a/ae/ce a ae ce ce ce cece cece ae a a ae ae ae a/ae/ce b/be ae b be b be a a

1/4" = 1'-0"LEVEL 4 - AREA OF WORK - REMODEL LIGHTING AND CONTROL 1

a

KEYPLAN - LEVEL 4

ISSUED FOR 100% C.D. 2020-10-02

a a ae aea a aa

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

CONSTRUCTION DOCUMENTSOffice of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

664-CSI-1154W NUCLEAR MEDICINE PET- CT 2

SAN DIEGO, CA

IDS GROUP INC.

1 PETERS CANYON RD. STE. 130

IRVINE, CA 92606

949.387.8500

FULLY SPRINKLERED

C U s e rs \A le ja n d ro .A g u ila r\ D o c u m e n ts

\1 X

.0

-V

A S

D -N u c -M e d

E le c R

T V ra m i.

e lh a s s a n id s g i.

c o m .r v t

/2 /2

:5

:0

P

M

EP-301

ELECTRICAL LIGHTING CONTROL

SINGLE LINE DIAGRAM

10/02/20 RO GA

1. nPODM DX SWITCH PROVIDES MANUAL OVERRIDE OF THE LIGHTS: ON/OFF/RAISE/LOWER.

2. nPP16 D PROVIDES RELAY AND 0-10V DIMMING CONTROL OF NORMAL POWERED GENERAL LIGHTING

3. nPP16 D ER PROVIDES RELAY AND 0-10V DIMMING CONTROL OF EMERGENCY CIRCUIT POWERED LIGHTING WITH INTEGRAL UL924 SHUNT. 1. nPODM DX SWITCH PROVIDES MANUAL OVERRIDE OF THE LIGHTS: ON/OFF/RAISE/LOWER.

2. nPODM 2P DX SWITCH PROVIDES 2 ZONES OF MANUAL OVERRIDE OF THE LIGHTS: ON/OFF/RAISE/LOWER.

3. nPP16 D PROVIDES RELAY AND 0-10V DIMMING CONTROL OF NORMAL POWERED GENERAL LIGHTING

4. nPP16 D ER PROVIDES RELAY AND 0-10V DIMMING CONTROL OF EMERGENCY CIRCUIT POWERED LIGHTING WITH INTEGRAL UL924 SHUNT.

BILL OF MATERIAL

nPP16 D

PRODUCT # DESCRIPTION

16 AMP RELAY PACK WITH 0-10V DIMMING CONTROL;

CHASE NIPPLE MOUNTING

nPODM DX WH SINGLE CHANNEL ON/OFF TOGGLE WITH DIMMING

QUANTITY OF nLIGHT DEVICES MAY VARY BASED ON LIGHTING FLOOR PLAN nPP16 D ER

16 AMP RELAY PACK WITH 0-10V DIMMING CONTROL;

EMERGENCY CIRCUIT CONTROL; INTEGRAL UL924

SHUNT; CHASE NIPPLE MOUNTING

BILL OF MATERIAL

nPP16 D

PRODUCT # DESCRIPTION

16 AMP RELAY PACK WITH 0-10V DIMMING CONTROL;

CHASE NIPPLE MOUNTING

nPODM 2P DX WH 2 CHANNEL ON/OFF TOGGLE WITH DIMMING

QUANTITY OF nLIGHT DEVICES MAY VARY BASED ON LIGHTING FLOOR PLAN nPODM DX WH

SINGLE CHANNEL ON/OFF TOGGLE WITH

DIMMING

nPP16 D ER

16 AMP RELAY PACK WITH 0-10V DIMMING CONTROL;

EMERGENCY CIRCUIT CONTROL; INTEGRAL UL924

SHUNT; CHASE NIPPLE MOUNTING

BLUE

W /R

Y E

L /R

E D

RJ45 PORTS

nPP16 D ER

HN

W H

T

B L K

EHEN

B L U

E

EN

VIOLET

GRAY

YEL- 120V/ RED- 277V

NORMAL (sense only)

G R

N

OPTION FOR

NETWORKED

SYSTEM

NOTE 1

BLK - 120VAC

ORN - 277VAC

RJ45

PORTS

+40mA nPP16 D

VIOLET

GRAY

GRN

BLUE

HN

W H

T

B L K

/O R

N1

B L U

E

N277/120VAC 0-10V DIMMING

GENERAL LIGHTING

R J

R J nPODM DX WH

277/120VAC 0-10V DIMMING

EMERGENCY

CIRCUIT LIGHTING

NOTE 1

BLK - 120VAC

ORN - 277VAC

nPODM 2P DX WH

R J

R J

R J

R J nPODM DX WH

RJ45

PORTS

+40mA nPP16 D

VIOLET

GRAY

GRN

BLUE

HN

W H

T

B L K

/O R

N1

B…