H4247213 - Bay Pines VA Hospital CT Update - ACM LBP Report Revised Final Signed.pdf

PDF 4 MB Posted

Attached to
Z2DA--FY26: NRM (PROJ: 516-CSI-134) BB Prepare Site for installation of a new PET/CT
Solicitation number
36C24826Q0459
Issued by

About this file

This is a Limited Pre-Renovation Asbestos and Lead Paint Survey Report for the C.W. (Bill) Young, Bay Pines VA Medical Center in Bay Pines, Florida, prepared by Terracon Consultants, Inc. for Above Group Inc. The survey was conducted on December 17-18, 2024, and June 12, 2025, to identify asbestos-containing materials (ACM) and lead-based paint in designated Building 100 areas scheduled for renovation to install new PET/CT equipment. The survey covered interior areas including rooms 2C118, 2C107, 2C107A, 2C112D, 2C129, SC120, 2C121, 2C124, 2C125, and various hallways, with floor finishes of vinyl tile, faux wood plank, carpet, and carpet tile; wall finishes of painted drywall, vinyl cove base, ceramic tile, and plastic wall panels; and suspended ceiling systems.

The asbestos survey identified 29 homogeneous areas with 79 bulk samples collected and analyzed. Results detected asbestos in multiple materials classified as Non-Friable Category I (NESHAP/OSHA classification), including 3-5% chrysotile in mixed yellow/black mastics under vinyl floor tiles and carpet (approximately 1,414 square feet total), 2-10% chrysotile in white paper-wrapped thermal system insulation (TSI) on mechanical equipment (80 linear feet total), 2% chrysotile in tan duct mastic (20 linear feet), and 20 wood interior doors assumed to contain asbestos. The lead paint survey of eight samples showed no detectable lead concentrations above laboratory reporting limits. All identified asbestos-containing materials require removal by a Florida-licensed asbestos abatement contractor as OSHA Class II work activities prior to renovation, with compliance required under EPA NESHAP regulations (40 CFR 61, Subpart M) and Florida Department of Environmental Protection requirements, including third-party air monitoring during abatement. The survey was limited to accessible areas, and additional suspect materials may exist in concealed locations.

View the file

Other files

File Type Posted
36C24826Q0459 0003.docx DOCX document
S02_36C24826Q0459 0002.docx DOCX document
Bay Pines PET-CT Room 2025.07.24 - Shielding Report.pdf PDF
S05_Site visit Sign in sheet 05262026 .jpg JPG image
Change Summary.docx DOCX document
H4247213 Bay Pines PET-CT Mold Sampling Final - signed.pdf PDF
516-CSI-134_Bay Pines PET-CT_100 percent IFC Drawings.pdf PDF
516-CSI-144_Submittal_Log.xlsx XLSX spreadsheet
516-CSI-144_Submittal_Log.pdf PDF
S05_RFIs Prepare Site PET.xlsx XLSX spreadsheet
516-CSI-144_Install New PETCT_100 percent IFC Specs.pdf PDF
516-CSI-134_Bay Pines PET CT_IFC Change Summary.pdf PDF
Amendment 0001.pdf PDF
S02_RFP Attachment WD FL20250222 01022026_Pinellas 10 pages.pdf PDF
S02_RFP Attachment 07a-516-CSI-134_BayPines_PET-CT_100percent_IFB_Drawings_REV31MAR2026 104 pages.pdf PDF
S02_RFP Attachment NARA 3 pages Bay Pines 2026.pdf PDF
S02_RFP Attachment 07b-516-CSI-144_New_PET-CT_100percent_IFB_Specs_REV_30MAR2026 945 pages.pdf PDF
S02_RFP Attachment Price Schedule 1 page.pdf PDF
S02_RFP Attachment SOW-516-CSI-134_2026-04-02 4 pages.pdf PDF
S02_RFP Attachment Past Performance Questionnaire 4 pages.pdf PDF
S02_RFP Attachment EMR 1 page.pdf PDF
S02_RFP Attachment LOS 2 pages.pdf PDF
S02_Solicitation 36C24826R0112 FINAL.pdf PDF
S02_RFP Attachment Specialized Experience Form 3 pages.pdf PDF
S02_RFP Attachment Divisional Price Breakdown.xlsx XLSX spreadsheet
S02_RFP Attachment H4247213 Bay Pines PET-CT Mold Sampling Final 23 pages.pdf PDF
S02_RFP Attachment H4247213-BayPinesVA_CT_Update-ACM_LBP_Report_Revised 41 pages.pdf PDF
S02_RFP Attachment VA_Submittal_Register 17 pages.pdf PDF
Show all 28

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Limited Pre-Renovation Asbestos and Lead Paint Survey Report

C.W. (Bill) Young, Bay Pines VA Medical Center

Project No. 516-CSI-134 –New 1.5T PET/CT 10000 Bay Pines Boulevard

Bay Pines, Florida 33744

August 27, 2025 Terracon Project No. H4247213, Revised

Prepared for:

Above Group Inc.

400 E. Pratt Street, 8th Floor

Baltimore, MD 21202

Prepared by:

Terracon Consultants, Inc.

Tampa, Florida

5463 West Waters Avenue, Suite 830

Tampa, Florida 33634

P (813) 221-0050

F (813) 221-0051

Terracon.com

Facilities | Environmental | Geotechnical | Materials

Above Group, Inc.

400 E. Pratt Street, 8th Floor

Baltimore, MD 21202

Attn: Ms. Jennifer Knight

P: (405) 206-0437

E: jknight@abovegroupinc.com

RE: Limited Pre-Renovation Asbestos and Lead Paint Survey Report

C.W. (Bill) Young, Bay Pines VA Medical Center

Project No. 516-CSI-134 – Site Prep for the Installation of New PET/CT

Designated Building 100 Areas

10000 Bay Pines Boulevard

Terracon Report No. H4247213, Revised

Dear Ms. Knight:

Terracon Consultants, Inc. is pleased to submit the attached report to Above Group, Inc. (Client). The purpose of this report is to present the results of a limited pre-renovation asbestos and lead paint survey performed within the designated areas of the C.W. (Bill) Young, Bay Pines VA Medical Center located in

Bay Pines, Florida. This survey was conducted in general accordance with PH4247213, dated July 1, 2024. Terracon understands this survey was requested in connection with the installation of PET/CT equipment within the designated areas of Building 100 of the subject structure.

Based on the laboratory results, asbestos was detected in building materials assessed during our survey. Laboratory analysis of the paint chip samples indicated lead was not detected in samples above the laboratory detection/reporting limit. Please refer to the attached report for further details.

Terracon appreciates the opportunity to provide this service to Above Group, Inc. If you have any questions regarding this report, or if you need assistance with project oversight and sampling during renovation, please contact the undersigned at 727-470-1104.

Sincerely, Terracon Consultants, Inc.

Florida Asbestos Business License Number ZA-337

Karl Kologiski Mark Fricke, CIH, CSP, LAC AHERA Building Inspector FL Licensed Asbestos Consultant AX132 mailto:jknight@abovegroupinc.com Kologiski, Karl V Rectangle

Kologiski, Karl V Rectangle

Kologiski, Karl V Stamp

Facilities | Environmental | Geotechnical | Materials 1

TABLE OF CONTENTS

Page

1.0 INTRODUCTION

1.1 Project Objective

1.2 Reliance

2.0 BUILDING DESCRIPTION

3.0 ASBESTOS FIELD ACTIVITIES

3.1 Visual Assessment

3.2 Physical Assessment

3.3 Sample Collection

3.4 Sample Analysis

4.0 LEAD-BASED PAINT FIELD ACTIVITIES

4.1 Visual Observations

4.2 Visual Assessment

4.3 Sample Collection

4.4 Sample Analysis

5.0 ASBESTOS REGULATORY OVERVIEW

6.0 LEAD-BASED PAINT REGULATORY OVERVIEW

7.0 ASBESTOS FINDINGS AND RECOMMENDATIONS

7.1 Findings

7.2 Recommendations

8.0 LEAD-BASED PAINT FINDINGS AND RECOMMENDATIONS

8.1 Findings

9.0 LIMITATIONS/GENERAL COMMENTS

APPENDIX A ASBESTOS ANALYTICAL LABORATORY RESULTS

APPENDIX B LEAD PAINT ANALYTICAL LABORATORY RESULTS

APPENDIX C REPRESENTATIVE PHOTOGRAPHS

APPENDIX D LICENSES AND CERTIFICATIONS

APPENDIX E ASBESTOS SAMPLE LOCATIONS

Asbestos and Lead Paint Survey Report

C.W. (Bill) Young, Bay Pines VA Medical Center

10000 Bay Pines Boulevard

Terracon Project No. H4247213 (Revised)

1.0 INTRODUCTION

Terracon Consultants, Inc. (Terracon) conducted a limited pre-renovation asbestos survey, and limited lead paint survey of the designated Building 100 areas of the C.W. (Bill) Young, Bay Pines VA Medical

Center in Bay Pines, Florida. Surveys and sampling were limited to suspect materials located within interior areas that are part of the PET/CT installation in the designated areas.

The survey was conducted from December 17-18, 2024 and June 12, 2025, by Terracon representative

Mr. Karl Kologiski, an Asbestos Hazard Emergency Response Act (AHERA)-accredited asbestos inspector in general accordance with Terracon Proposal PH4247213, dated July 1, 2024. Although reasonable effort was made to survey all accessible suspect materials within the designated interior and roof locations, additional suspect but un-sampled materials could be located in walls, in voids, or in other inaccessible or concealed areas. Applicable licenses and certifications are presented in Appendix D.

Homogeneous areas of suspected asbestos-containing materials (ACM) and suspect lead paint were visually identified, surveyed, and documented. Asbestos sampling was performed in general accordance with the United States (U.S.) Environmental Protection Agency (EPA) 40 Code of Federal Regulations

(CFR) Part 763 Subpart E 763, known as AHERA. Suspect lead paint samples were obtained from representative surfaces potentially containing lead coatings.

1.1 Project Objective

Terracon’s services were performed to accommodate planned renovation activities and to satisfy requirements of the EPA regulation 40 CFR 61, Subpart M, National Emission Standards for Hazardous

Air Pollutants (NESHAP). NESHAP regulations prohibit the release of asbestos fibers to the atmosphere during renovation or demolition activities. The asbestos NESHAP requires that potentially regulated ACM

(RACM) be identified, classified, and quantified prior to planned disturbances or demolition activities.

Lead paint testing was performed to comply with applicable sections of the U.S. Occupational Safety and

Health Administration (OSHA) Lead Standard (29 CFR 1926.62) for disturbance of paint/coatings with lead that will be encountered during renovation or demolition activities.

1.2 Reliance

This report is for the exclusive use of Above Group, Inc. (Above Group) for the project being discussed.

Reliance by any other party on this report is prohibited without written authorization of Terracon and

Above Group. Reliance on this report by Above Group and all authorized parties will be subject to the terms, conditions, and limitations stated in the proposal, this report, and the existing agreement between Above Group and Terracon. The limitations of liability defined in Terracon’s Agreement for

Services is the aggregate limit of Terracon’s liability to Above Group.

Asbestos & Lead Paint Survey Report

Project No. 516-CSI-134 – Site Prep for the Installation of New PET/CT

Bay Pines VA Hospital ■ Bay Pines, Florida

August 27, 2025 ■ Terracon Project No. H4247213 Revised

Facilities | Environmental | Geotechnical | Materials 2

2.0 BUILDING DESCRIPTION

The designated interior areas of Building 100 included in the survey consisted of the mechanical system and the following rooms: 2C118, 2C107 & 2C107A, 2C112D, 2C129, SC120, 2C121, 2C124, 2C125 ,Hallways C2-47, C2-49, C2-50 C2-52 and mechanical system mezzanine. Interior floor finishes of the Building 100 areas in the subject areas included vinyl floor tile, faux wood plank, carpet, and carpet tile. Wall finishes included painted drywall - joint compound, vinyl cove base, ceramic tile, and plastic wall panel. Ceilings consisted of a suspended ceiling grid and ceiling tiles.

3.0 ASBESTOS FIELD ACTIVITIES

The asbestos portion of the survey was conducted on December 17-18, 2024 and June 12, 2025, by

Terracon representative Mr. Karl Kologiski an AHERA-accredited asbestos inspector. A copy of the inspector certificates and applicable licenses are attached as Appendix F. The survey was conducted in general accordance with the sample collection protocols established in EPA 40 CFR Part 763 Subpart E

763.86, AHERA. A summary of survey activities is provided below.

3.1 Visual Assessment

Survey activities were initiated with limited visual observations of the accessible, designated areas to identify homogeneous areas of suspect ACM. A homogeneous area (HA) consists of building materials that appear similar throughout in terms of color and texture with consideration given to the date of application. The limited assessment was conducted in readily accessible areas of the buildings.

3.2 Physical Assessment

A physical assessment of each homogeneous area of suspect ACM was conducted to assess the friability and condition of the materials. A friable material is defined by the EPA as a material which can be crumbled, pulverized, or reduced to powder by hand pressure when dry. Friability was assessed by physically touching suspect materials.

3.3 Sample Collection

Based on results of the visual observations, bulk samples of suspect ACM were collected in general accordance with AHERA sampling protocols. Samples of suspect materials were collected in randomly selected locations in each homogeneous area. Samples were placed in sealable containers and labeled with unique sample numbers using an indelible marker.

The selection of sample locations and frequency of sampling were based on Terracon’s observations and the assumption that like materials in the same area are homogeneous in content. A total of seventy-nine (79) bulk samples were collected by Terracon from twenty-nine (29) homogeneous areas of suspect ACM.

Project No. 516-CSI-134 – Site Prep for the Installation of New PET/CT

Facilities | Environmental | Geotechnical | Materials 3

3.4 Sample Analysis

Bulk samples were submitted under secure chain of custody protocols to Scientific Analytical Institute

(SAI) of Greensboro North Carolina for analysis by polarized light microscopy with dispersion staining techniques per EPA methodology 600/R-93/116. The percentage of asbestos, where applicable was determined by microscopic visual estimation. SAI is accredited under the National Voluntary

Laboratory Accreditation Program (NVLAP) Accreditation No. 200664-0. Asbestos laboratory data is presented in Appendix A. Representative asbestos sample photographs are presented in Appendix C.

Asbestos sample locations are presented in Appendix E.

4.0 LEADPAINT FIELD ACTIVITES

4.1 Visual Observations

The lead paint survey was limited to readily observable and accessible surfaces that may be disturbed during the planned renovations. Terracon cannot guarantee that all lead paint coated surfaces were identified during our survey as some lead paint may be hidden from sight or in inaccessible locations, or the homogeneous construction areas identified may not be truly homogeneous. This limited lead paint survey was not performed in accordance with U.S. Department of Housing and Urban

Development (HUD) Guidelines for the Evaluation and Control of Lead-Based Paint (LBP)

Hazards in Housing or applicable to EPA 40 CFR 745 requirements. It should be understood that this limited lead paint survey was not considered to be comprehensive in nature, and the results are not intended to be used to determine all lead hazards, develop lead abatement plans, or prepare detailed cost estimates for lead abatement.

4.2 Visual Assessment

A visual assessment of each unique combination, defined as a color, component, and substrate, of suspect lead paint was conducted in addition to an assessment of the paint condition.

4.3 Sample Collection

A total of eight (8) paint chip samples of suspect lead paint were collected. Paint chip samples were placed in sealable containers and labeled with unique sample numbers using an indelible marker.

Laboratory results are presented in Appendix B. Representative lead paint sample photographs are presented in Appendix C.

4.4 Sample Analysis

Paint chip samples were submitted under chain of custody to Scientific Analytical Institute (SAI) for analysis by flame atomic absorption spectrometry per EPA SW-846/3050B/7000B. SAI is accredited under the Environmental Lead Laboratory Accreditation Program (ELLAP) (Accreditation No. 173190).

Project No. 516-CSI-134 – Site Prep for the Installation of New PET/CT

Facilities | Environmental | Geotechnical | Materials 4

5.0 ASBESTOS REGULATORY OVERVIEW

The asbestos NESHAP (40 CFR Part 61, Subpart M) regulates asbestos fiber emissions and asbestos waste disposal practices. The NESHAP regulation also requires the identification and classification of existing ACM according to friability prior to demolition or renovation activity.

Under NESHAP, asbestos-containing building materials are classified as:

1. Friable ACM, or

2. Category I Non-Friable ACM, or

3. Category II Non-Friable ACM.

Friable materials are those that, when dry, may be crumbled, pulverized, or reduced to powder by hand pressure. Category I Non-Friable ACM includes packings, gaskets, resilient floor coverings and asphalt roofing products containing more than 1% asbestos. Category II Non-Friable ACM are any materials other than Category I Non-Friable ACM that contain more than 1% asbestos.

Under NESHAP, Regulated Asbestos-Containing Materials (RACM are classified as:

1. Friable ACM.

2. Category I Non-Friable ACM that is in poor condition and/or has become friable.

3. Category I Non-Friable ACM that will or has been subjected to sanding, grinding, cutting, or abrading.

4. Category II Non-Friable ACM that has a high probability of becoming or has become crumbled, pulverized, or reduced to powder by the forces expected to act upon the material in the course of renovation or demolition operations.

In the State of Florida, asbestos activities are regulated by the Florida Department of Environmental

Protection (FDEP). Some counties have developed an enforcement division to carry out the responsibilities of the FDEP and have developed environmental and asbestos ordinances in which additional compliance is required. Since the facility is located in Pinellas County Florida, the NESHAP regulations are under the authority of the Pinellas County Air Quality Division (727-464-4425).

Per the NESHAP, RACM must be removed prior to renovation or demolition activities that will disturb the materials. Removal of any ACM or RACM must be conducted by a State of Florida-licensed asbestos abatement contractor, in compliance with Florida Statute (F.S) 469.001-014. In addition, third party air monitoring should be performed during and following the abatement.

The OSHA Asbestos standard for construction (29 CFR 1926.1101) regulates employee workplace exposure to asbestos. The OSHA standard has (2) Permissible Exposure Limits (PEL) that require employee exposure to airborne asbestos fibers be maintained below 0.1 asbestos fibers per cubic centimeter of air (0.1 f/cc) for an 8-hour Time-Weighted Average (8-Hr. TWA) and 1.0 f/cc for a 30-minute TWA (Excursion Limit). The OSHA standard classifies construction and maintenance activities which could disturb ACM and specifies work practices and precautions which employers must follow when engaging in each class of regulated work.

Project No. 516-CSI-134 – Site Prep for the Installation of New PET/CT

Facilities | Environmental | Geotechnical | Materials 5

6.0 LEAD-CONTAINING PAINT REGULATORY OVERVIEW

Lead is regulated by the EPA, HUD, Florida Department of Environmental Protection (FDEP) and OSHA.

The EPA and FDEP regulate lead use, removal, and disposal, and OSHA regulates lead exposure to workers.

The EPA defines Lead-Based Paint (LBP) as any paint, varnish, stain, or other applied coating that contains lead equal to or greater than 1.0 mg/cm2, 5,000 mg/kg, or 0.5% by dry weight as determined by laboratory analysis.

For the purpose of the OSHA lead standard, lead-containing materials includes any detectable concentrations of metallic lead, all inorganic lead compounds, and organic lead soaps. A synopsis of the OSHA regulations (29 CFR 1926.62) and the applicability are as follows:

The OSHA Lead Standard for Construction (29 CFR 1926.62) applies to all construction work where an employee may be occupationally exposed to lead. All work related to construction, alteration, or repair

(including painting and decorating) is included. The Lead-in-Construction standard applies to any detectable concentration of lead in paint, as even small concentrations of lead can result in unacceptable employee exposures depending upon on the method of disturbance-removal and other workplace conditions.

Under this standard, construction includes, but is not limited to, the following:

◼ Demolition or salvage of structures where lead or materials containing lead are present.

◼ Removal or encapsulation of materials containing lead.

◼ New construction, alteration, repair, or renovation of structures, substrates, or portions containing lead, or materials containing lead.

◼ Installation of products containing lead

◼ Lead contamination/emergency clean-up.

◼ Transportation, disposal, storage, or containment of lead or materials containing lead on the site or location at which construction activities are performed.

◼ Maintenance operations associated with construction activities described above.

7.0 FINDINGS AND RECOMMENDATIONS – ASBESTOS

7.1 Findings

HA#

HA

DESCRIPTION

SAMPLE

LOCATIONS

NESHAP / OSHA

CLASSIFICATION

CONDITION

MATERIAL

LOCATION

QUANTITY

(SF)

ASBESTOS TYPE

CX1

Non-asbestos 12” white/gray streak VFT/ and mixed yellow/black asbestos mastic C2-51 HA

2C-122D

C2-51 Non-Friable Cat I Good

2C-122D

C2-51

C2-50

C2-49

C2-47

3% Chrysotile -

Mixed Yellow/Black

Mastic

CX2

Non-asbestos 12” off white VFT and asbestos black mastic / 2C-10712

2C-107

2C107A Non-Friable Cat I Good

2C-107

2C107A

5% Chrysotile -

Mastic

CX4

Non asbestos 12” blue

VFT and mixed yellow/black asbestos mastic / 2C122D

2C-122D Non-Friable Cat I Good 2C-122D

(Patch)

2% Chrysotile -

Mixed Yellow/Black

Mastic

CX10

24” Black Carpet Squares w/ Green & Black Mastic 2C-120 Non-Friable Cat I Good 2C-120 140

2% Chrysotile -

Mixed Green/Black

Mastic

CX11

Non-asbestos blue sq pattern carpet and mixed yellow/black asbestos mastic / 2C121

2C-121 Non-Friable Cat I Good 2C-121 180

3% Chrysotile -

Mixed Yellow/Black

Mastic

Project No. 516-CSI-134 – Site Prep for the Installation of New PET/CT

Facilities | Environmental | Geotechnical | Materials 6

HA#

HA

DESCRIPTION

SAMPLE

LOCATIONS

NESHAP / OSHA

CLASSIFICATION

CONDITION

MATERIAL

LOCATION

QUANTITY

(SF)

ASBESTOS TYPE

CX24 4” White paper wrap TSI

Upper

Mechanical

Mezzanine

Non-Friable Cat I Good

Upper

Mechanical

Mezzanine

60 Liner

Feet

7-10% Chrysotile -

White Mastic

CX27 Tan Duct Mastic

Upper

Mechanical

Mezzanine

Non-Friable Cat I Good

Upper

Mechanical

Mezzanine

20 Liner

Feet 2% Chrysotile

CX30 Wood Interior Door NA Assumed ACM Good Throughout

Work Area 20 Doors NA

Laboratory results are presented in Appendix A. Representative asbestos sample photographs are presented in Appendix C. Asbestos sample locations are presented in Appendix E. Estimated quantities are based on a cursory field evaluation, and actual quantities may vary significantly, especially if asbestos containing materials are present in hidden and/or inaccessible areas not evaluated as part of this survey. It should be noted that the quantity provided is an estimate only and should not be used for bidding or notification purposes. It should be noted that suspect materials other than those identified may exist. Should suspect materials other than those which were identified during this survey be uncovered prior to or during the renovation process, those materials should be assumed asbestos-containing until sampling and analysis can confirm or refute their asbestos content.

7.2 Recommendations

Non-Friable Cat I Mixed Yellow/Black Mastics In its current state, the noted mastics can be considered a Non-Friable, Category I ACM. As such, this material should be removed by a Florida licensed asbestos abatement contractor as an OSHA Class II work activity. In addition, The State of Florida requires any concrete waste (CMU, Slab, etc.) to be free of any ACM material (paint, mastics, etc.) prior to disposal or recycling.

Non-Friable Cat I, Tan Duct Mastic and White TSI Mastic

In its current state, the noted mastics can be considered a Non-Friable, Category I ACM. As such, this material should be removed by a Florida licensed asbestos abatement contractor as an OSHA Class II work activity. In addition, The State of Florida requires any concrete waste (CMU, Slab, etc.) to be free of any ACM material (paint, mastics, etc.) prior to disposal or recycling.

Assumed ACM, Interior Door and Frame Assemblies

Due to the on-going operations in the designated areas of the PET/CT scan areas, the doors and frames could not be tested for asbestos at the time of the survey. The metal door frames were labeled as fire-rated door frames, however there were no markings on the wood doors. The door and frame assemblies are assumed to be asbestos containing and are to be treated as such, until the designated areas are vacated, at which time, the doors and frames can be drilled so that the interior contents can be sampled for asbestos content.

Terracon recommends that if any materials not sampled in this survey are uncovered during planned work that these suspect materials should be considered asbestos containing until sampled and proven otherwise. It is recommended that any materials identified as containing asbestos be removed prior to renovation or demolition activities by a Florida-Licensed Asbestos Abatement Contractor. We also recommend direct contact with the Pinellas County Air Quality Division (727-464-4425) for confirmation and clarifications of their asbestos program requirements.

Project No. 516-CSI-134 – Site Prep for the Installation of New PET/CT

Facilities | Environmental | Geotechnical | Materials 7

8.0 FINDINGS AND RECOMMENDATIONS – Lead Paint

8.1 Findings

Detectable concentrations of lead were not identified in the various coatings evaluated during the assessment. The lead paint summary table presented below summarizes the results of the 7 paint chip samples:

Sample

ID

Sample

Location Item Substrate Color

Reporting

Limit

(PPM)

Concentration

(PPM)

Paint Chip

Results %

Weight

CP-1 2C-122D Wall Drywall Blue 38 <38 0.0038

CP-2 2C-118 Wall Drywall Brown 53 <53 0.0053

CP-3 2C-124 Wall Drywall Teal 48 <48 0.0048

CP4 2C-48 Wall Drywall Green 46 <46 0.0046

CP5 2C-107 Wall Drywall Off White 62 <62 0.0062

CP6 2C-50 Wall Drywall Beige 35 <35 0.0035

CP7 2C-107 Door Frame Metal White 40 <86 0.0086

CP8 2C-120 Wall Drywall Purple 43 <43 0.0043

9.0 LIMITATIONS/GENERAL COMMENTS

This Limited Interior Pre-Renovation Asbestos and Lead Paint Survey was conducted in a manner consistent with the level of care and skill ordinarily exercised by members of the profession currently practicing under similar conditions in the same locale. The results, findings, conclusions, and recommendations expressed in this report are based on conditions observed during our survey of the designated areas. The information contained in this report is relevant to the date on which this survey was conducted and should not be relied upon to represent conditions at a later date.

This report has been prepared on behalf of and exclusively for use by Above Group for specific application to their project as discussed. This report is not a bidding document. Contractors or consultants reviewing this report must draw their own conclusions regarding further investigation or remediation deemed necessary. Terracon does not warrant the work of regulatory agencies, laboratories or other third parties supplying information which may have been used in the preparation of this report.

No warranty, express or implied, is made.

APPENDIX A

ASBESTOS ANALYTICAL LABORATORY RESULTS

Bulk Asbestos Analysis By Polarized Light Microscopy EPA Method: 600/R-93/116 and

40 CFR, Part 763, Subpart E, App.E

Lab Order ID:Attn: 10071152Customer: Terracon 5463 W Waters Ave Suite 830 Tampa, FL 33634-1253

Analysis:

Date Received:

Date Amended:

Karl Kologiski

PLM

12/19/2024

Project: Bay Pines VA CT - H4247213 Date Reported: 12/26/2024

01/23/2025

Sample ID

Lab Sample ID

Description

Lab Notes Asbestos

Fibrous Components

Non-Fibrous Components Treatment

Attributes

CX1-1 - A

10071152_0001

12" white/gray streak

VFT/2C-122D

None Detected 100% Other

Non-Fibrous

Dissolved

White

Homogeneous tile

CX1-1 - B

10071152_0053

12" white/gray streak

VFT/2C-122D

None Detected 100% Other

Non-Fibrous

Dissolved

Yellow

Homogeneous mastic

CX1-2 - A

10071152_0002

12" white/gray streak

VFT/2C-51 HA

None Detected 100% Other

Non-Fibrous

Dissolved

White

Homogeneous tile

CX1-2 - B

10071152_0054

12" white/gray streak

VFT/2C-51 HA

3% Chrysotile 97% Other

Non-Fibrous

Dissolved

Yellow, Black

Homogeneous mixed mastic

CX1-3 - A

10071152_0003

12" white/gray streak

VFT/2C-51 HA

None Detected 100% Other

Non-Fibrous

Dissolved

White

Homogeneous tile

CX1-3 - B

10071152_0055

12" white/gray streak

VFT/2C-51 HA

3% Chrysotile 97% Other

Non-Fibrous

Dissolved

Yellow, Black

Homogeneous mixed mastic

CX2-4 - A

10071152_0004

12" off white VFT/2C-107

None Detected 100% Other

Non-Fibrous

Dissolved

White

Homogeneous tile

CX2-4 - B

10071152_0056

12" off white VFT/2C-107

5% Chrysotile 95% Other

Non-Fibrous

Dissolved

Black

Homogeneous mastic

Disclaimer: Due to the nature of the EPA 600 method, asbestos may not be detected in samples containing low levels of asbestos. We strongly recommend that analysis of floor tiles, vermiculite, and/or heterogenous soil samples be conducted by TEM for confirmation of "None Detected" by PLM. This report relates only to the samples tested and may not be reproduced, except in full, without the written approval of SAI. This report may not be used by the client to claim product endorsement by NVLAP or any other agency of the U.S. government. Analytical uncertainty available upon request. Scientific Analytical Institute participates in the NVLAP Proficiency Testing program.

Unless otherwise noted blank sample correction was not performed. Estimated MDL is 0.1%.

Charmel Dozier (85)

Analyst Approved Signatory

Page 1 of 11P-F-002 r15 1/15/2028 Scientific Analytical Institute, Inc. 4604 Dundas Dr. Greensboro, NC 27407 (336) 292-3888

EPA Method: 600/R-93/116 and

40 CFR, Part 763, Subpart E, App.E

Lab Order ID:Attn: 10071152Customer: Terracon 5463 W Waters Ave Suite 830 Tampa, FL 33634-1253

Analysis:

Date Received:

Date Amended:

Karl Kologiski

PLM

12/19/2024

Project: Bay Pines VA CT - H4247213 Date Reported: 12/26/2024

01/23/2025

Sample ID

Lab Sample ID

Description

Lab Notes Asbestos

Fibrous Components

Non-Fibrous Components Treatment

Attributes

CX2-5 - A

10071152_0005

12" off white VFT/2C-107A

None Detected 100% Other

Non-Fibrous

Dissolved

White

Homogeneous tile

CX2-5 - B

10071152_0057

12" off white VFT/2C-107A

5% Chrysotile 95% Other

Non-Fibrous

Dissolved

Black

Homogeneous mastic

CX2-6 - A

10071152_0006

12" off white VFT/2C-107A

None Detected 100% Other

Non-Fibrous

Dissolved

White

Homogeneous tile

CX2-6 - B

10071152_0058

12" off white VFT/2C-107A

5% Chrysotile 95% Other

Non-Fibrous

Dissolved

Black

Homogeneous mastic

CX3-7 - A

10071152_0007

6" faux wood plank/2C-118

None Detected 100% Other

Non-Fibrous

Dissolved

Brown

Homogeneous wood plank

CX3-7 - B

10071152_0059

6" faux wood plank/2C-118

None Detected 100% Other

Non-Fibrous

Dissolved

Yellow

Homogeneous mastic

CX3-8 - A

10071152_0008

6" faux wood plank/2C-124

None Detected 100% Other

Non-Fibrous

Dissolved

Brown

Homogeneous wood plank

CX3-8 - B

10071152_0060

6" faux wood plank/2C-124

None Detected 100% Other

Non-Fibrous

Crushed, Dissolved

Yellow, Gray

Homogeneous mastic/leveling

Disclaimer: Due to the nature of the EPA 600 method, asbestos may not be detected in samples containing low levels of asbestos. We strongly recommend that analysis of floor tiles, vermiculite, and/or heterogenous soil samples be conducted by TEM for confirmation of "None Detected" by PLM. This report relates only to the samples tested and may not be reproduced, except in full, without the written approval of SAI. This report may not be used by the client to claim product endorsement by NVLAP or any other agency of the U.S. government. Analytical uncertainty available upon request. Scientific Analytical Institute participates in the NVLAP Proficiency Testing program.

Unless otherwise noted blank sample correction was not performed. Estimated MDL is 0.1%.

Charmel Dozier (85)

Page 2 of 11P-F-002 r15 1/15/2028 Scientific Analytical Institute, Inc. 4604 Dundas Dr. Greensboro, NC 27407 (336) 292-3888

EPA Method: 600/R-93/116 and

40 CFR, Part 763, Subpart E, App.E

Lab Order ID:Attn: 10071152Customer: Terracon 5463 W Waters Ave Suite 830 Tampa, FL 33634-1253

Analysis:

Date Received:

Date Amended:

Karl Kologiski

PLM

12/19/2024

Project: Bay Pines VA CT - H4247213 Date Reported: 12/26/2024

01/23/2025

Sample ID

Lab Sample ID

Description

Lab Notes Asbestos

Fibrous Components

Non-Fibrous Components Treatment

Attributes

CX3-9 - A

10071152_0009

6" faux wood plank/2C-125

None Detected 100% Other

Non-Fibrous

Dissolved

Brown

Homogeneous wood plank

CX3-9 - B

10071152_0061

6" faux wood plank/2C-125

None Detected 100% Other

Non-Fibrous

Crushed, Dissolved

Yellow, Gray

Homogeneous mastic/leveling

CX4-10 - A

10071152_0010

12" blue VFT/2C122D

None Detected 100% Other

Non-Fibrous

Dissolved

Blue

Homogeneous tile

CX4-10 - B

10071152_0062

12" blue VFT/2C122D

2% Chrysotile 98% Other

Non-Fibrous

Dissolved

Black, Yellow

Homogeneous mixed mastic

CX4-11 - A

10071152_0011

12" blue VFT/2C122D

None Detected 100% Other

Non-Fibrous

Dissolved

Blue

Homogeneous tile

CX4-11 - B

10071152_0063

12" blue VFT/2C122D

2% Chrysotile 98% Other

Non-Fibrous

Dissolved

Black, Yellow

Homogeneous mixed mastic

CX5-12 - A

10071152_0012

4" tan cove base/2C122D

None Detected 100% Other

Non-Fibrous

Dissolved

Blue

Homogeneous cove base

CX5-12 - B

10071152_0064

4" tan cove base/2C122D

None Detected 100% Other

Non-Fibrous

Dissolved

Yellow

Homogeneous mastic

Disclaimer: Due to the nature of the EPA 600 method, asbestos may not be detected in samples containing low levels of asbestos. We strongly recommend that analysis of floor tiles, vermiculite, and/or heterogenous soil samples be conducted by TEM for confirmation of "None Detected" by PLM. This report relates only to the samples tested and may not be reproduced, except in full, without the written approval of SAI. This report may not be used by the client to claim product endorsement by NVLAP or any other agency of the U.S. government. Analytical uncertainty available upon request. Scientific Analytical Institute participates in the NVLAP Proficiency Testing program.

Unless otherwise noted blank sample correction was not performed. Estimated MDL is 0.1%.

Charmel Dozier (85)

Page 3 of 11P-F-002 r15 1/15/2028 Scientific Analytical Institute, Inc. 4604 Dundas Dr. Greensboro, NC 27407 (336) 292-3888

EPA Method: 600/R-93/116 and

40 CFR, Part 763, Subpart E, App.E

Lab Order ID:Attn: 10071152Customer: Terracon 5463 W Waters Ave Suite 830 Tampa, FL 33634-1253

Analysis:

Date Received:

Date Amended:

Karl Kologiski

PLM

12/19/2024

Project: Bay Pines VA CT - H4247213 Date Reported: 12/26/2024

01/23/2025

Sample ID

Lab Sample ID

Description

Lab Notes Asbestos

Fibrous Components

Non-Fibrous Components Treatment

Attributes

CX5-13 - A

10071152_0013

4" tan cove base/2C122D

None Detected 100% Other

Non-Fibrous

Dissolved

Blue

Homogeneous cove base

CX5-13 - B

10071152_0065

4" tan cove base/2C122D

None Detected 100% Other

Non-Fibrous

Dissolved

Yellow

Homogeneous mastic

CX6-14

10071152_0014

DW/JC/2C118

None Detected 10% Cellulose 90% Other

Fibrous

Crushed

White

Homogeneous drywall: none detected, joint compound: none detected

CX6-15

10071152_0015

DW/JC/2C107

None Detected 10% Cellulose 90% Other

Fibrous

Crushed

White

Homogeneous drywall: none detected, joint compound: none detected

CX6-16

10071152_0016

DW/JC/2C122D

None Detected 10% Cellulose 90% Other

Fibrous

Crushed

White

Homogeneous drywall: none detected, joint compound: none detected

CX6-17

10071152_0017

DW/JC/2C121

None Detected 10% Cellulose 90% Other

Fibrous

Crushed

White

Homogeneous drywall: none detected, joint compound: none detected

CX6-18

10071152_0018

DW/JC/2C-51-hallway

None Detected 10% Cellulose 90% Other

Fibrous

Crushed

White

Homogeneous drywall: none detected, joint compound: none detected

CX6-19

10071152_0019

DW/JC/C2-48

None Detected 10% Cellulose 90% Other

Fibrous

Crushed

White

Homogeneous drywall: none detected, joint compound: none detected

Disclaimer: Due to the nature of the EPA 600 method, asbestos may not be detected in samples containing low levels of asbestos. We strongly recommend that analysis of floor tiles, vermiculite, and/or heterogenous soil samples be conducted by TEM for confirmation of "None Detected" by PLM. This report relates only to the samples tested and may not be reproduced, except in full, without the written approval of SAI. This report may not be used by the client to claim product endorsement by NVLAP or any other agency of the U.S. government. Analytical uncertainty available upon request. Scientific Analytical Institute participates in the NVLAP Proficiency Testing program.

Unless otherwise noted blank sample correction was not performed. Estimated MDL is 0.1%.

Charmel Dozier (85)

Page 4 of 11P-F-002 r15 1/15/2028 Scientific Analytical Institute, Inc. 4604 Dundas Dr. Greensboro, NC 27407 (336) 292-3888

EPA Method: 600/R-93/116 and

40 CFR, Part 763, Subpart E, App.E

Lab Order ID:Attn: 10071152Customer: Terracon 5463 W Waters Ave Suite 830 Tampa, FL 33634-1253

Analysis:

Date Received:

Date Amended:

Karl Kologiski

PLM

12/19/2024

Project: Bay Pines VA CT - H4247213 Date Reported: 12/26/2024

01/23/2025

Sample ID

Lab Sample ID

Description

Lab Notes Asbestos

Fibrous Components

Non-Fibrous Components Treatment

Attributes

CX7-20 - A

10071152_0020

4" tan ceramic wall tile/2C122D

None Detected 100% Other

Non-Fibrous

Crushed

White

Homogeneous ceramic tile

CX7-20 - B

10071152_0066

4" tan ceramic wall tile/2C122D

None Detected 100% Other

Non-Fibrous

Dissolved

White

Homogeneous adhesive

CX7-21 - A

10071152_0021

4" tan ceramic wall tile/2C122D

None Detected 100% Other

Non-Fibrous

Crushed

White

Homogeneous ceramic tile

CX7-21 - B

10071152_0067

4" tan ceramic wall tile/2C122D

None Detected 100% Other

Non-Fibrous

Crushed

White

Homogeneous grout

CX8-22 - A

10071152_0022

4" brown cove base/2C107A

None Detected 100% Other

Non-Fibrous

Dissolved

Brown

Homogeneous cove base

CX8-22 - B

10071152_0068

4" brown cove base/2C107A

None Detected 100% Other

Non-Fibrous

Dissolved

Yellow

Homogeneous mastic

CX8-23 - A

10071152_0023

4" brown cove base/2C-121

None Detected 100% Other

Non-Fibrous

Dissolved

Brown

Homogeneous cove base

CX8-23 - B

10071152_0069

4" brown cove base/2C-121

None Detected 100% Other

Non-Fibrous

Dissolved

Yellow

Homogeneous mastic

Disclaimer: Due to the nature of the EPA 600 method, asbestos may not be detected in samples containing low levels of asbestos. We strongly recommend that analysis of floor tiles, vermiculite, and/or heterogenous soil samples be conducted by TEM for confirmation of "None Detected" by PLM. This report relates only to the samples tested and may not be reproduced, except in full, without the written approval of SAI. This report may not be used by the client to claim product endorsement by NVLAP or any other agency of the U.S. government. Analytical uncertainty available upon request. Scientific Analytical Institute participates in the NVLAP Proficiency Testing program.

Unless otherwise noted blank sample correction was not performed. Estimated MDL is 0.1%.

Charmel Dozier (85)

Page 5 of 11P-F-002 r15 1/15/2028 Scientific Analytical Institute, Inc. 4604 Dundas Dr. Greensboro, NC 27407 (336) 292-3888

EPA Method: 600/R-93/116 and

40 CFR, Part 763, Subpart E, App.E

Lab Order ID:Attn: 10071152Customer: Terracon 5463 W Waters Ave Suite 830 Tampa, FL 33634-1253

Analysis:

Date Received:

Date Amended:

Karl Kologiski

PLM

12/19/2024

Project: Bay Pines VA CT - H4247213 Date Reported: 12/26/2024

01/23/2025

Sample ID

Lab Sample ID

Description

Lab Notes Asbestos

Fibrous Components

Non-Fibrous Components Treatment

Attributes

CX8-24 - A

10071152_0024

4" brown cove base/C2-51-hallway

None Detected 100% Other

Non-Fibrous

Dissolved

Brown

Homogeneous cove base

CX8-24 - B

10071152_0070

4" brown cove base/C2-51-hallway

None Detected 100% Other

Non-Fibrous

Dissolved

Yellow

Homogeneous mastic

CX9-25

10071152_0025

24" pinhole Ct/2C 121

None Detected 50% 25%

Mineral Wool Cellulose

25% Other

Fibrous

Teased

White

Homogeneous

CX9-26

10071152_0026

24" pinhole Ct/2C 118

None Detected 50% 25%

Mineral Wool Cellulose

25% Other

Fibrous

Teased

White

Homogeneous

CX10-27 - A

10071152_0027

24" black carpet sq/2C-120

None Detected 70% Synthetic Fibers 30% Other

Fibrous

Ashed

Black, Blue

Homogeneous carpet

CX10-27 - B

10071152_0071

24" black carpet sq/2C-120

None Detected 100% Other

Non-Fibrous

Dissolved

Green

Homogeneous mastic

CX10-28 - A

10071152_0028

24" black carpet sq/2C-120

None Detected 70% Synthetic Fibers 30% Other

Fibrous

Ashed

Black, Blue

Homogeneous carpet

CX10-28 - B

10071152_0072

24" black carpet sq/2C-120

2% Chrysotile 98% Other

Non-Fibrous

Dissolved

Green, Black

Homogeneous mastic - amended - additional QC requested

Disclaimer: Due to the nature of the EPA 600 method, asbestos may not be detected in samples containing low levels of asbestos. We strongly recommend that analysis of floor tiles, vermiculite, and/or heterogenous soil samples be conducted by TEM for confirmation of "None Detected" by PLM. This report relates only to the samples tested and may not be reproduced, except in full, without the written approval of SAI. This report may not be used by the client to claim product endorsement by NVLAP or any other agency of the U.S. government. Analytical uncertainty available upon request. Scientific Analytical Institute participates in the NVLAP Proficiency Testing program.

Unless otherwise noted blank sample correction was not performed. Estimated MDL is 0.1%.

Charmel Dozier (85)

Page 6 of 11P-F-002 r15 1/15/2028 Scientific Analytical Institute, Inc. 4604 Dundas Dr. Greensboro, NC 27407 (336) 292-3888

EPA Method: 600/R-93/116 and

40 CFR, Part 763, Subpart E, App.E

Lab Order ID:Attn: 10071152Customer: Terracon 5463 W Waters Ave Suite 830 Tampa, FL 33634-1253

Analysis:

Date Received:

Date Amended:

Karl Kologiski

PLM

12/19/2024

Project: Bay Pines VA CT - H4247213 Date Reported: 12/26/2024

01/23/2025

Sample ID

Lab Sample ID

Description

Lab Notes Asbestos

Fibrous Components

Non-Fibrous Components Treatment

Attributes

CX11-29 - A

10071152_0029

Blue sq pattern carpet/2C121

None Detected 70% Synthetic Fibers 30% Other

Fibrous

Ashed

Blue

Homogeneous carpet

CX11-29 - B

10071152_0073

Blue sq pattern carpet/2C121

None Detected 100% Other

Non-Fibrous

Dissolved

Yellow

Homogeneous mastic

CX11-29 - C

10071152_0074

Blue sq pattern carpet/2C121

None Detected 100% Other

Non-Fibrous

Dissolved

White

Homogeneous tile

CX11-29 - D

10071152_0075

Blue sq pattern carpet/2C121

3% Chrysotile 97% Other

Non-Fibrous

Dissolved

Yellow, Black

Homogeneous mixed mastic

CX11-30 - A

10071152_0030

Blue sq pattern carpet/2C121

None Detected 70% Synthetic Fibers 30% Other

Fibrous

Ashed

Blue

Homogeneous carpet

CX11-30 - B

10071152_0076

Blue sq pattern carpet/2C121

None Detected 100% Other

Non-Fibrous

Dissolved

Yellow

Homogeneous mastic

CX11-30 - C

10071152_0077

Blue sq pattern carpet/2C121

None Detected 100% Other

Non-Fibrous

Dissolved

White

Homogeneous tile

CX11-30 - D

10071152_0078

Blue sq pattern carpet/2C121

3% Chrysotile 97% Other

Non-Fibrous

Dissolved

Black, Yellow

Homogeneous mixed mastic

Disclaimer: Due to the nature of the EPA 600 method, asbestos may not be detected in samples containing low levels of asbestos. We strongly recommend that analysis of floor tiles, vermiculite, and/or heterogenous soil samples be conducted by TEM for confirmation of "None Detected" by PLM. This report relates only to the samples tested and may not be reproduced, except in full, without the written approval of SAI. This report may not be used by the client to claim product endorsement by NVLAP or any other agency of the U.S. government. Analytical uncertainty available upon request. Scientific Analytical Institute participates in the NVLAP Proficiency Testing program.

Unless otherwise noted blank sample correction was not performed. Estimated MDL is 0.1%.

Charmel Dozier (85)

Page 7 of 11P-F-002 r15 1/15/2028 Scientific Analytical Institute, Inc. 4604 Dundas Dr. Greensboro, NC 27407 (336) 292-3888

EPA Method: 600/R-93/116 and

40 CFR, Part 763, Subpart E, App.E

Lab Order ID:Attn: 10071152Customer: Terracon 5463 W Waters Ave Suite 830 Tampa, FL 33634-1253

Analysis:

Date Received:

Date Amended:

Karl Kologiski

PLM

12/19/2024

Project: Bay Pines VA CT - H4247213 Date Reported: 12/26/2024

01/23/2025

Sample ID

Lab Sample ID

Description

Lab Notes Asbestos

Fibrous Components

Non-Fibrous Components Treatment

Attributes

CX12-31

10071152_0031

24" random divet CT/2C- 122D

None Detected 55% 15%

Cellulose Mineral Wool

30% Other

Fibrous

Teased

White

Homogeneous

CX12-32

10071152_0032

24" random divet CT/2C-120

None Detected 55% 15%

Cellulose Mineral Wool

30% Other

Fibrous

Teased

White

Homogeneous

CX12-33

10071152_0033

24" random divet CT/2C-124

None Detected 55% 15%

Cellulose Mineral Wool

30% Other

Fibrous

Teased

White

Homogeneous

CX13-34

10071152_0034

2x4 random divit CT/C2-51 hallway

None Detected 55% 15%

Cellulose Mineral Wool

30% Other

Fibrous

Teased

White

Homogeneous

CX13-35

10071152_0035

2x4 random divit CT/C2-51 hallway

None Detected 55% 15%

Cellulose Mineral Wool

30% Other

Fibrous

Teased

White

Homogeneous

CX14-36 - A

10071152_0036

4" white cove base/C2-48

None Detected 100% Other

Non-Fibrous

Dissolved

White

Homogeneous cove base

CX14-36 - B

10071152_0085

4" white cove base/C2-48

None Detected 100% Other

Non-Fibrous

Dissolved

Yellow

Homogeneous mastic

CX14-37 - A

10071152_0037

4" white cove base/2C118

None Detected 100% Other

Non-Fibrous

Dissolved

White

Homogeneous cove base

Disclaimer: Due to the nature of the EPA 600 method, asbestos may not be detected in samples containing low levels of asbestos. We strongly recommend that analysis of floor tiles, vermiculite, and/or heterogenous soil samples be conducted by TEM for confirmation of "None Detected" by PLM. This report relates only to the samples tested and may not be reproduced, except in full, without the written approval of SAI. This report may not be used by the client to claim product endorsement by NVLAP or any other agency of the U.S. government. Analytical uncertainty available upon request. Scientific Analytical Institute participates in the NVLAP Proficiency Testing program.

Unless otherwise noted blank sample correction was not performed. Estimated MDL is 0.1%.

Charmel Dozier (85)

Page 8 of 11P-F-002 r15 1/15/2028 Scientific Analytical Institute, Inc. 4604 Dundas Dr. Greensboro, NC 27407 (336) 292-3888

EPA Method: 600/R-93/116 and

40 CFR, Part 763, Subpart E, App.E

Lab Order ID:Attn: 10071152Customer: Terracon 5463 W Waters Ave Suite 830 Tampa, FL 33634-1253

Analysis:

Date Received:

Date Amended:

Karl Kologiski

PLM

12/19/2024

Project: Bay Pines VA CT - H4247213 Date Reported: 12/26/2024

01/23/2025

Sample ID

Lab Sample ID

Description

Lab Notes Asbestos

Fibrous Components

Non-Fibrous Components Treatment

Attributes

CX14-37 - B

10071152_0079

4" white cove base/2C118

None Detected 100% Other

Non-Fibrous

Dissolved

Yellow

Homogeneous mastic

CX14-38 - A

10071152_0038

4" white cove base/2C129

None Detected 100% Other

Non-Fibrous

Dissolved

White

Homogeneous cove base

CX14-38 - B

10071152_0080

4" white cove base/2C129

None Detected 100% Other

Non-Fibrous

Dissolved

Yellow

Homogeneous mastic

CX15-39

10071152_0039

Caulking - wall panel/2C118

None Detected 100% Other

Non-Fibrous

Ashed

Tan

Homogeneous

CX15-40

10071152_0040

Caulking - wall panel/2C118

None Detected 100% Other

Non-Fibrous

Ashed

Tan

Homogeneous

CX16-41

10071152_0041

24" solid ceiling tile/2C118

None Detected 40% 30%

Cellulose Mineral Wool

30% Other

Fibrous

Teased

White

Homogeneous

CX16-42

10071152_0042

24" solid ceiling tile/2C118

None Detected 40% 30%

Cellulose Mineral Wool

30% Other

Fibrous

Teased

White

Homogeneous

CX17-43 - A

10071152_0043

Plastic wall panel/2C118

None Detected 100% Other

Non-Fibrous

Ashed

White

Homogeneous wall panel

Disclaimer: Due to the nature of the EPA 600 method, asbestos may not be detected in samples containing low levels of asbestos. We strongly recommend that analysis of floor tiles, vermiculite, and/or heterogenous soil samples be conducted by TEM for confirmation of "None Detected" by PLM. This report relates only to the samples tested and may not be reproduced, except in full, without the written approval of SAI. This report may not be used by the client to claim product endorsement by NVLAP or any other agency of the U.S. government. Analytical uncertainty available upon request. Scientific Analytical Institute participates in the NVLAP Proficiency Testing program.

Unless otherwise noted blank sample correction was not performed. Estimated MDL is 0.1%.

Charmel Dozier (85)

Page 9 of 11P-F-002 r15 1/15/2028 Scientific Analytical Institute, Inc. 4604 Dundas Dr. Greensboro, NC 27407 (336) 292-3888

EPA Method: 600/R-93/116 and

40 CFR, Part 763, Subpart E, App.E

Lab Order ID:Attn: 10071152Customer: Terracon 5463 W Waters Ave Suite 830 Tampa, FL 33634-1253

Analysis:

Date Received:

Date Amended:

Karl Kologiski

PLM

12/19/2024

Project: Bay Pines VA CT - H4247213 Date Reported: 12/26/2024

01/23/2025

Sample ID

Lab Sample ID

Description

Lab Notes Asbestos

Fibrous Components

Non-Fibrous Components Treatment

Attributes

CX17-43 - B

10071152_0081

Plastic wall panel/2C118

None Detected 100% Other

Non-Fibrous

Dissolved

Yellow

Homogeneous mastic

CX17-44 - A

10071152_0044

Plastic wall panel/2C118

None Detected 100% Other

Non-Fibrous

Ashed

White

Homogeneous wall panel

CX17-44 - B

10071152_0082

Plastic wall panel/2C118

None Detected 100% Other

Non-Fibrous

Dissolved

Yellow

Homogeneous mastic

CX18-45

10071152_0045

Fire stop Pu/2C120

None Detected 1% Fiber Glass 99% Other

Non-Fibrous

Dissolved

Red

Homogeneous

CX18-46

10071152_0046

Fire stop Pu/2C120

None Detected 1% Fiber Glass 99% Other

Non-Fibrous

Dissolved

Red

Homogeneous

CX19-47 - A

10071152_0047

4" faux wood plank/2C129

None Detected 100% Other

Non-Fibrous

Dissolved

Brown

Homogeneous wood plank

CX19-47 - B

10071152_0083

4" faux wood plank/2C129

None Detected 100% Other

Non-Fibrous

Crushed, Dissolved

Yellow, Gray

Homogeneous mastic/leveling

CX19-48 - A

10071152_0048

4" faux wood plank/C248

None Detected 100% Other

Non-Fibrous

Dissolved

Brown

Homogeneous wood plank

Disclaimer: Due to the nature of the EPA 600 method, asbestos may not be detected in samples containing low levels of asbestos. We strongly recommend that analysis of floor tiles, vermiculite, and/or heterogenous soil samples be conducted by TEM for confirmation of "None Detected" by PLM. This report relates only to the samples tested and may not be reproduced, except in full, without the written approval of SAI. This report may not be used by the client to claim product endorsement by NVLAP or any other agency of the U.S. government. Analytical uncertainty available upon request. Scientific Analytical Institute participates in the NVLAP Proficiency Testing program.

Unless otherwise noted blank sample correction was not performed. Estimated MDL is 0.1%.

Charmel Dozier (85)

Page 10 of 11P-F-002 r15 1/15/2028 Scientific Analytical Institute, Inc. 4604 Dundas Dr. Greensboro, NC 27407 (336) 292-3888

EPA Method: 600/R-93/116 and

40 CFR, Part 763, Subpart E, App.E

Lab Order ID:Attn: 10071152Customer: Terracon 5463 W Waters Ave Suite 830 Tampa, FL 33634-1253

Analysis:

Date Received:

Date Amended:

Karl Kologiski

PLM

12/19/2024

Project: Bay Pines VA CT - H4247213 Date Reported: 12/26/2024

01/23/2025

Sample ID

Lab Sample ID

Description

Lab Notes Asbestos

Fibrous Components

Non-Fibrous Components Treatment

Attributes

CX19-48 - B

10071152_0084

4" faux wood plank/C248

None Detected 100% Other

Non-Fibrous

Dissolved, Crushed

Gray, Yellow

Homogeneous mastic/leveling

CX20-49

10071152_0049

Gray sink undercoat/2C107

None Detected 5% Cellulose 95% Other

Non-Fibrous

Dissolved

Gray

Homogeneous

CX20-50

10071152_0050

Gray sink undercoat/2C107

None Detected 5% Cellulose 95% Other

Non-Fibrous

Dissolved

Gray

Homogeneous

CX21-51

10071152_0051

Caulking back splash/2C107

None Detected 100% Other

Non-Fibrous

Dissolved

White, Tan

Homogeneous

CX21-52

10071152_0052

Caulking back splash/2C107

None Detected 100% Other

Non-Fibrous

Dissolved

Tan, White

Homogeneous

Disclaimer: Due to the nature of the EPA 600 method, asbestos may not be detected in samples containing low levels of asbestos. We strongly recommend that analysis of floor tiles, vermiculite, and/or heterogenous soil samples be conducted by TEM for confirmation of "None Detected" by PLM. This report relates only to the samples tested and may not be reproduced, except in full, without the written approval of SAI. This report may not be used by the client to claim product endorsement by NVLAP or any other agency of the U.S. government. Analytical uncertainty available upon request. Scientific Analytical Institute participates in the NVLAP Proficiency Testing program.

Unless otherwise noted blank sample correction was not performed. Estimated MDL is 0.1%.

Charmel Dozier (85)

Page 11 of 11P-F-002 r15 1/15/2028 Scientific Analytical Institute, Inc. 4604 Dundas Dr. Greensboro, NC 27407 (336) 292-3888

EPA Method: 600/R-93/116 and

40 CFR, Part 763, Subpart E, App.E

Lab Order ID:Attn: 10084739Customer: Terracon 5463 W Waters Ave Suite 830 Tampa, FL 33634-1253

Analysis:

Date Received:

Date Amended:

Karl Kologiski

PLM

06/16/2025

Project: Bay Pines VA PET/CT

08/11/2025

Sample ID

Lab Sample ID

Description

Lab Notes Asbestos

Fibrous Components

Non-Fibrous Components Treatment

Attributes

CX 24 60 - A

10084739_0008

4" White paper wrap TSI elbow/T joint

10% Chrysotile 90% Other

Non-Fibrous

Ashed

White

Homogeneous mastic

CX 24 60 - B

10084739_0028

4" White paper wrap TSI elbow/T joint

None Detected 60% Cellulose 40% Other

Fibrous

Ashed

White

Homogeneous paper

CX 24 61 - A

10084739_0009

4" White paper wrap TSI elbow/T joint

10% Chrysotile 90% Other

Non-Fibrous

Ashed

White

Homogeneous mastic

CX 24 61 - B

10084739_0029

4" White paper wrap TSI elbow/T joint

None Detected 60% Cellulose 40% Other

Fibrous

Ashed

White

Homogeneous paper

CX 24 62 - A

10084739_0010

4" White paper wrap TSI elbow/T joint

10% Chrysotile 90% Other

Non-Fibrous

Ashed

White

Homogeneous mastic

CX 24 62 - B

10084739_0030

4" White paper wrap TSI elbow/T joint

None Detected 60% Cellulose 40% Other

Fibrous

Ashed

White

Homogeneous paper

CX 24 63 - A

10084739_0011

4" White paper wrap TSI elbow/T joint

7% Chrysotile 93% Other

Non-Fibrous

Ashed

White

Homogeneous mastic

CX 24 63 - B

10084739_0031

4" White paper wrap TSI elbow/T joint

None Detected 60% Cellulose 40% Other

Fibrous

Ashed

White

Homogeneous paper

Disclaimer: Due to the nature of the EPA 600 method, asbestos may not be detected in samples containing low levels of asbestos. We strongly recommend that analysis of floor tiles, vermiculite, and/or heterogenous soil samples be conducted by TEM for confirmation of "None Detected" by PLM. This report relates only to the samples tested and may not be reproduced, except in full, without the written approval of SAI. This report may not be used by the client to claim product endorsement by NVLAP or any other agency of the U.S. government. Analytical uncertainty available upon request.