TO 00-20-14
28 FEBRUARY 2022
TECHNICAL MANUAL
AIR FORCE METROLOGY AND
CALIBRATION PROGRAM
This publication replaces TO 00-20-14 dated 28 February 2021.
Distribution Statement A - Approved for public release; distribution is unlimited. AFLCMC PA Case Number AFLCMC-2022-0028
Published under Authority of the Secretary of the Air Force
TO 00-20-14 28 February 2022
A USAF
INSERT LATEST CHANGED PAGES. DESTROY SUPERSEDED PAGES.
NOTE: The portion of the text and illustrations affected by the changes is indicated by a vertical line in the outer margins of the page.
Date of issue for original and changed pages are:
Original ..................... 0 ..................... 28 February 2022
TOTAL NUMBER OF PAGES IN THIS PUBLICATION IS 98, CONSISTING OF THE FOLLOWING:
Page * Change Page * Change Page * Change No. No. No. No. No. No.
Title A i - iii iv Blank 1-1 - 1-8 2-1 - 2-4 3-1 - 3-19 3-20 Blank 4-1 - 4-10 5-1 - 5-15 5-16 Blank 6-1 6-2 Blank 7-1 - 7-3 7-4 Blank 8-1 - 8-8 9-1 - 9-6 10-1 - 10-3 10-4 Blank 11-1 11-2 Blank 12-1 - 12-3 12-4 Blank 13-1 – 13-3 13-4 Blank
*Zero in this column indicates an original page
LIST OF EFFECTIVE PAGES
28 February 2022 TO 00-20-14 i
TABLE OF CONTENTS
Section/Paragraph Page
1 INTRODUCTION ........................................................................................................................................... 1-1
1.1 Purpose .............................................................................................................................................. 1-1
1.2 General .............................................................................................................................................. 1-1
1.3 Scope ................................................................................................................................................. 1-3
1.4 Definitions ......................................................................................................................................... 1-3
1.5 Program Abbreviations ..................................................................................................................... 1-7
2 AFMETCAL PROGRAM ORGANIZATION AND INTERFACE ............................................................... 2-1
2.1 General .............................................................................................................................................. 2-1
2.2 AFMETCAL Program ...................................................................................................................... 2-1
2.3 Program Policy and Direction ........................................................................................................... 2-2
2.4 Periodic Reports Utilized .................................................................................................................. 2-2
2.5 Program Requirements ...................................................................................................................... 2-3
2.6 Program Material .............................................................................................................................. 2-3
2.7 Program Integration .......................................................................................................................... 2-3
3 OPERATION .................................................................................................................................................. 3-1
3.1 Calibration of Standards and Other TMDE ....................................................................................... 3-1
3.2 Exceptions from Periodic Calibration ............................................................................................... 3-5
3.3 Limited or Special Calibration .......................................................................................................... 3-7
3.4 Calibration Intervals .......................................................................................................................... 3-8
3.5 Precise Time Support ........................................................................................................................ 3-10
3.6 TMDE User Responsibilities ............................................................................................................ 3-11
3.7 Input Priority System ........................................................................................................................ 3-14
3.8 Care of TMDE .................................................................................................................................. 3-14
3.9 PMEL Mobile/On-Site Measurement Capabilities ........................................................................... 3-16
3.10 Responsibilities of PMEL Managers ................................................................................................. 3-18
4 CALIBRATION ASSISTANCE AND SUPPORT AGREEMENTS ............................................................. 4-1
4.1 Requests for Calibration Assistance .................................................................................................. 4-1
4.2 Technical/Engineering Assistance. ................................................................................................... 4-2
4.3 Calibration, Repair & Return, and Loan ........................................................................................... 4-2
4.4 PMEL Support Agreements .............................................................................................................. 4-2
4.5 Commercial and N49/59/64 Calibration of TMDE ........................................................................... 4-4
4.6 Calibration Support for Site Activation Task Force (SATAF), Operational Test and
Evaluation (OT&E), and Initial Operational Test and Evaluation (IOT&E)..................................... 4-8
4.7 Security Assistance (SA) Programs and Foreign Military Sales (FMS) ........................................... 4-8
4.8 TMDE Warranties ............................................................................................................................. 4-8
4.9 Leased/Borrowed Equipment ............................................................................................................ 4-9
4.10 Depot Public-Private Partnerships (PPP) .......................................................................................... 4-9
5 PREPARATION AND USE OF CALIBRATION FORMS AND LABELS ................................................. 5-1
5.1 TMDE Certification Stamp (K Stamp) ............................................................................................. 5-1
5.2 General Instructions .......................................................................................................................... 5-1
5.3 TMDE Bar Code Labels .................................................................................................................... 5-2
5.4 Certification Labels ........................................................................................................................... 5-3
5.5 AFTO Form 256, No Calibration Required Label ............................................................................ 5-8
5.6 NIST Report of Measurement ........................................................................................................... 5-8
5.7 Calibration Certificates or Calibration Reports ................................................................................. 5-8
5.8 AFTO FORM 249, TMDE Calibration Data .................................................................................... 5-10
5.9 AFTO FORM 250, TMDE Calibration Correction Chart ................................................................. 5-10 ii
TABLE OF CONTENTS (Cont.)
Section/Paragraph Page
5.10 Measurement Standard Decals .......................................................................................................... 5-10
5.11 AFTO FORM 255, Certification Void if Broken .............................................................................. 5-10
5.12 AFTO FORM 45, Request for Calibration Responsibility Determination ........................................ 5-11
5.13 AFTO FORM 57, PMEL Technical Order Information .................................................................... 5-14
6 RESERVED .................................................................................................................................................... 6-1
7 PMEL ASSESSMENT PROGRAM ............................................................................................................... 7-1
7.1 General .............................................................................................................................................. 7-1
7.2 Assessment Intervals ......................................................................................................................... 7-1
7.3 Assessment Notification .................................................................................................................... 7-1
7.4 Assessment Process ........................................................................................................................... 7-1
7.5 Certification Status ............................................................................................................................ 7-2
7.6 Assessment Reports .......................................................................................................................... 7-3
8 PMEL FACILITY OPERATIONAL REQUIREMENTS ............................................................................... 8-1
8.1 General Requirements ....................................................................................................................... 8-1
8.2 PMEL Environment .......................................................................................................................... 8-2
8.3 TMDE Environmental Restrictions ................................................................................................... 8-4
8.4 Dust Levels. ...................................................................................................................................... 8-4
8.5 Noise ................................................................................................................................................. 8-4
8.6 Vibration ........................................................................................................................................... 8-4
8.7 Electromagnetic Interference (EMI)/Radio Frequency Interference (RFI) ....................................... 8-4
8.8 Lighting ............................................................................................................................................. 8-5
8.9 Doors & Windows............................................................................................................................. 8-6
8.10 Air Locks ........................................................................................................................................... 8-6
8.11 Controlled Areas ............................................................................................................................... 8-6
8.12 Flooring ............................................................................................................................................. 8-6
8.13 Furniture and Fixtures ....................................................................................................................... 8-6
8.14 FMS Type IV and Type II( ) PMEL Facility Requirements ............................................................. 8-6
8.15 Environmental Limits for RASCAL ................................................................................................. 8-8
9 QUALITY PROGRAM (QP) .......................................................................................................................... 9-1
9.1 Quality Program (QP) ....................................................................................................................... 9-1
9.2 QP Reviews ....................................................................................................................................... 9-1
9.3 Nonconformity (NC) ......................................................................................................................... 9-2
9.4 Root Cause (RC) Analysis ................................................................................................................ 9-3
9.5 Process Review Plan (PRP)............................................................................................................... 9-3
9.6 NC and RC Codes ............................................................................................................................. 9-4
10 MANAGEMENT SYSTEM (MS) .................................................................................................................. 10-1
10.1 Management System (MS) ................................................................................................................ 10-1
10.2 Quality Manual (QM) ....................................................................................................................... 10-1
10.3 Management Internal Control Toolset (MICT) ................................................................................. 10-1
10.4 Internal Audit .................................................................................................................................... 10-1
10.5 Management Review (MR) ............................................................................................................... 10-2
10.6 PMEL Activity Summary.................................................................................................................. 10-3
10.7 Trainer Proficiency Evaluation (TPE) ............................................................................................... 10-3
11 PROFICIENCY TESTING/MEASUREMENT ASSURANCE PROGRAM (PT/MAP) AND ACCEPTANCE
TESTING .......................................................................................................................................... 11-1
11.1 Proficiency Testing/Measurement Assurance Program (PT/MAP) .................................................. 11-1
11.2 Acceptance Testing ........................................................................................................................... 11-1 iii/(iv Blank)
TABLE OF CONTENTS (Cont.)
Section/Paragraph Page
12 AIR FORCE METROLOGY AND CALIBRATION LABORATORIES .................................................... 12-1
12.1 Major Command (MAJCOM) and Agency Codes .......................................................................... 12-1
12.2 Air Force Primary Standards Laboratory (AFPSL) ......................................................................... 12-1
12.3 Metrology & Calibration Flight (MCF) Sites .................................................................................. 12-1
12.4 Precision Measurement Equipment Laboratories (PMELs) ............................................................ 12-2
13 METROLOGY & CALIBRATION FLIGHT (MCF) ................................................................................... 13-1
13.1 Purpose ............................................................................................................................................ 13-1
13.2 Operation ......................................................................................................................................... 13-1
13.3 Exceptions ....................................................................................................................................... 13-1
13.4 MCF Workload Definition .............................................................................................................. 13-2
13.5 Additional Requirements................................................................................................................. 13-3
LIST OF FIGURES
Number Title Page 1-1 PMEL Support Matrix .............................................................................................................................. 1-4 1-2 Measurement Traceability of the Air Force Metrology and Calibration Program ................................... 1-5 2-1 AFMETCAL Program Resources and Planning Interfaces ...................................................................... 2-2 4-1 Type IIA PMEL Areas of Responsibility ................................................................................................. 4-1 5-1 TMDE Certification Stamp (K Stamp) ..................................................................................................... 5-1 5-2 AFTO FORM 65, TMDE Bar Code Label, Aluminum Stock .................................................................. 5-2 5-3 AFTO FORM 66, TMDE Bar Code Label, Polyester Stock .................................................................... 5-3 5-4 AFTO FORM 108, TMDE Certification Label (White) .......................................................................... 5-6 5-5 AFTO FORM 394, TMDE Certification Label (White) .......................................................................... 5-6 5-6 AFTO FORM 99, Limited/Special TMDE Certification Label (Yellow) ................................................ 5-7 5-7 AFTO FORM 398, Limited TMDE Certification Label (Yellow) ........................................................... 5-7 5-8 AFTO FORM 256, No Calibration Required Label ................................................................................. 5-8 5-9 AFTO FORM 255, Certification Void If Broken Label ........................................................................... 5-10 5-10 AFTO FORM 45, Request for Calibration Responsibility Determination .............................................. 5-13 5-11 AFTO FORM 57, PMEL Technical Order Information .......................................................................... 5-15
LIST OF TABLES
Number Title Page 3-1 Calibration Software Classes .................................................................................................................... 3-3 9-1 Summary of QP Codes ............................................................................................................................. 9-4 12-1 MAJCOM and Agency Codes .................................................................................................................. 12-1 12-2 AFPSL Site Information ........................................................................................................................... 12-1 12-3 PMEL Site Information ............................................................................................................................ 12-2
1-1
SECTION 1
INTRODUCTION
1.1 PURPOSE.
This Technical Order (TO) contains the methods and procedures for the management of the Air Force Metrology and Calibration (AFMETCAL) Program. This TO implements AFMAN 21-113, Air Force Metrology and Calibration (AFMETCAL) Management.
1.2 GENERAL.
The AFMETCAL Program is an Air Force program that provides measurement standards and equipment, professional and technical metrologists, Performing Work Centers (PWCs), a system of worldwide Precision Measurement Equipment Laboratory (PMEL) facilities, measurement equipment users, calibration data and integrated planning. This program ensures the reliability and accuracy of systems, subsystems and equipment. The program provides for the calibration and repair of Test, Measurement and Diagnostic Equipment (TMDE). It also ensures measurement traceability of the TMDE through the Air Force Primary Standards Laboratory (AFPSL) to National Institute of Standards and Technology (NIST) or other AFLCMC/WNM (AFMETCAL) approved sources. To accomplish this traceability, the AFMETCAL program requires that Air Force users and Precision Measurement Equipment Laboratories (PMELs) obtain calibration service from Air Force PMELs or the AFPSL. AFMETCAL must approve calibration service from other sources. The responsibilities of organizations in the AFMETCAL Program are identified in AFMAN 21-113.
1.2.1 AFLCMC/WNM (AFMETCAL). AFLCMC/WNM, referred to as AFMETCAL throughout this document, serves as the Air Force technical authority on metrology issues and is the Air Force single point of contact for calibration services and traceability of measurements to NIST. AFMETCAL makes TMDE calibration and repair responsibility determinations.
AFMETCAL publishes responsibility determinations in TO 33K-1-100-1 and TO 33K-1-100-2 and Calibration and Measurement Summary (CMS) TOs. TO 33K-1-100-2 and the workload summary tables in the CMS TOs are electronic data tables displayable with the Air Force Calibration Authority Viewer (AFCAV) application included with these TOs.
The PMEL Automated Management System (PAMS), Facility and Equipment Maintenance System (FEMS) and LabMet systems maintain authorized facsimiles of TO 33K-1-100-2 and certain CMS workload summary tables. All references to TO 33K-1-100-2 and CMS TOs apply equally to the corresponding authorized electronic facsimiles maintained in PAMS, FEMS and LabMet. The AFMETCAL Point of Contact (POC) list is available on the AFMETCAL SharePoint AFMETCAL Leadership & POCs page. This POC list should be used to determine the appropriate AFMETCAL personnel/office to be contacted for specific issues addressed by this TO.
1.2.1.1 Key documents referenced above:
a. TO 33K-1-100-1 (Calibration Procedure for Maintenance Data Collection Codes And Calibration Measurement Summaries)
b. TO 33K-1-100-2 (TMDE Calibration Notes, Calibration Interval, Technical Order, And Work Unit Code Reference Guide)
NOTE
These TOs are distributed via CD (33K-1-100-2-CD-1), which contains a desktop version of AFCAV. They are also available to all AF users via a web-based version (WebAFCAV), which is available on the Metrology Web Site (MetWeb) at url https://metweb.afmetcal.af.mil. In order to access WebAFCAV, submit a request for a MetWeb account by submitting the customized DD Form 2875 downloaded from this site. On this form, identify a supporting TODO account that has established a requirement for 33K-1-100-2-CD-1.
1.2.1.2 Calibration Repair Network. The AFMETCAL Program has been recognized as a Calibration Repair Network under the Repair Network Integration (RNI) initiative. Under RNI, the AFMETCAL Director is also known as the Product Repair Manager, AFMETCAL’s Plans and Analysis Section is the Repair Network Manager, the PMEL MAJCOM Functional Managers (MFM) serve as the MAJCOM leads, and PMEL Managers are referred to as Node Managers.
1-2
1.2.2 Air Force Primary Standards Laboratory (AFPSL). The Air Force Primary Standards Laboratory (AFPSL) maintains the Air Force Measurement Standards. The standards are traceable to NIST or other sources as approved by AFMETCAL. Air Force Measurement Standards are used to ensure the accuracy and traceability of Base Measurement Standards. Base Measurement Standards are provided to PMELs. Unless specified elsewhere throughout this TO, the following clarifies what requirements apply to the AFPSL:
1.2.2.1 Section 3 OPERATION. All paragraphs with the exception of para 3.10.8. The AFPSL will provide customer support IAW contract requirements.
1.2.2.2 Section 4 CALIBRATION ASSISTANCE AND SUPPORT AGREEMENTS. All paragraphs (as applicable).
1.2.2.3 Section 5 PREPARATION AND USE OF CALIBRATION FORMS AND LABELS. All paragraphs (as applicable).
1.2.2.4 Section 7 PMEL ASSESSMENT PROGRAM. All paragraphs. To determine the AFPSL certification status, the AFPSL will be assessed by AFMETCAL IAW Section 7 of this TO and AFI90-201, AF Inspection System. However, the PT/MAP assessment will be limited to interservice or interdepartmental proficiency testing processes/results (if applicable).
1.2.2.5 Section 8 PMEL FACILITY OPERATIONAL REQUIREMENTS. Not applicable. AFPSL facility requirements will be documented in an AFPSL Operating Instruction maintained by AFMETCAL. The AFPSL will submit ECS documentation IAW contract requirements.
1.2.2.6 Section 9 QUALITY PROGRAM (QP). All paragraphs. The AFPSL will submit QP documents to AFMETCAL IAW contract requirements (this includes any QP upload requirements associated with the AFMAN 21-113 TMDE Self- Assessment Communicator (SAC) in the Manager’s Internal Control Toolset (MICT)).
1.2.2.7 Section 10 MANAGEMENT SYSTEM (MS). All paragraphs. The AFPSL will submit MS documents to AFMETCAL IAW contract requirements (this includes any MS upload requirements associated with the AFMAN 21-113 SAC in MICT).
1.2.2.8 Section 11 PROFICIENCY TESTING/MEASUREMENT ASSURANCE PROGRAM (PT/MAP) AND ACCEPTANCE TESTING. AFMETCAL manages the PT/MAP program with the assistance of the AFPSL. The AFPSL may also perform interservice or interdepartmental proficiency testing as directed by AFMETCAL. The AFPSL will participate in Acceptance Testing IAW contract requirements.
1.2.3 Precision Measurement Equipment Laboratories (PMELs).
Headquarters AF/A4LM approves the establishment of PMELs at selected installations. The PMEL is the base-level AFMETCAL Program focal point. It is the activity authorized to possess and use base measurement standards. Operational details of the program are set forth in Section 3 of this TO and related command directives.
1.2.4 Metrology & Calibration Flight (MCF).
Metrology & Calibration Flights (MCF), formerly known as DIMFs, are currently established at the three Air Logistics Complexes (ALC) located at Ogden, UT (OO-ALC), Oklahoma City, OK (OC-ALC), and Warner Robins, GA (WR-ALC).
The MCF has some regional workload responsibilities, but they are not officially part of the Calibration Repair Network.
They have a unique function and mission separate from the co-located Type IIA PMEL. Operational details of the MCFs are set forth in Section 13 of this TO. Unless specifically stated in this TO or excluded in Section 13, the use of the term “PMEL” throughout this TO and AFMAN 21-113 is synonymous with MCF.
1.2.5 Other DoD Services Calibration Laboratories. The use of other DOD services calibration laboratories is authorized when approved by the applicable MFM and AFMETCAL. The Army, Navy and Marine Corps operate calibration laboratories using similar measurement techniques and management concepts. These laboratories provide measurement traceability to NIST. Units requesting support from another service shall follow the guidance in Section 4, which includes the interservice support agreement guidance.
1.3 SCOPE.
This TO applies to all activities that possess, use, calibrate, certify or maintain TMDE.
1-3
1.4 DEFINITIONS.
The following definitions apply for the purpose of this TO.
1.4.1 Absolute Measurement Standard. Designated measurement standards based on the absolute value of natural physical constants whose values can be accurately repeated under controlled conditions. Also known as an Intrinsic Standard. Examples are Monochromatic (Single Wavelength) Light Source, Josephson Junction Voltage Standards and Hall Effect Magnetic Field Standards.
1.4.2 Air Force Primary Standards Laboratory (AFPSL). The highest-level laboratory in the AFMETCAL Program. It maintains Air Force measurement standards certified by the NIST, the US Naval Observatory (USNO), or other AFMETCAL approved sources.
1.4.3 Calibration. Calibration is a comparison between equipment items, one of which is a measurement standard of known accuracy, to detect, correlate, adjust and report any variation in the accuracy of the other item(s).
1.4.4 Calibration, Limited. Calibration of TMDE to less than the accuracy or functional capabilities specified in the authorized calibration procedure or data.
1.4.5 Calibration, Special. Calibration of TMDE to all specifications called for in the authorized calibration procedure plus additional requirements.
1.4.6 Calibration and Measurement Summary (CMS). A TO which identifies calibration support necessary to ensure the operational readiness of a specific weapon system, weapon subsystem or mission. The summary describes the calibration concept and is calibration authority for the applicable weapon system, weapon subsystem or mission. The CMS is printed as a weapon system or 33K series technical order. AFMETCAL maintains Air Force CMSs (see paragraph 2.5.3).
1.4.7 Calibration and Measurement Requirement Summary (CMRS). A CMRS is a three category, in-line summary of measurement parameters. Normally, a Program Office or Department of Defense (DOD) system contractor prepares and maintains the CMRS (DI-QCIC-80278B). A CMRS identifies all measurement requirements within a specific system or item of equipment. The CMRS further displays the proposed solutions for maintaining the system measurement requirements within stated limits. It is also used to identify the need for new calibration standards. This data is provided to AFMETCAL, 813 Irving-Wick Dr W, Horton Building, Heath OH, 43056-1199. AFMETCAL uses it for ensuring calibration supportability and planning for PMEL support for a particular weapon system/subsystem (see paragraph 2.5.2).
1.4.8 Calibration Certificate or Calibration Report. Also referred to as, "Report of Measurement". It is a document containing data relevant to the calibration of a specific test instrument or standard. The AFMETCAL Program does not differentiate between a calibration certificate and a calibration report provided the document contains the required information to interpret the calibration results for the specific calibration.
1.4.9 Calibration Chart. A chart prepared for a specific item. It shows the difference between the nominal value and the actual value of a measurement or group of measurements.
1.4.10 Centralized Repair Facility (CRF). Centralized repair facilities provide regional intermediate-level maintenance repair support for designated USAF units. This regional repair capability replaces the decentralized maintenance concept in which base-level units performed all intermediate-level maintenance for deployed aircraft.
1.4.11 Certification. The documented designation that standards and TMDE have been calibrated and meet established technical requirements.
1.4.12 Commercial Calibration. Calibration of TMDE obtained from a commercial source where the TMDE already has a published calibration determination but the calibrating activity (PMEL or USER) asks for an exception. This should be considered only after organic (Air Force) support options have been exhausted. AFMETCAL approval is required prior to obtaining the calibration from a commercial source. Commercial calibrations will be funded by the user.
1-4
1.4.13 Contingency Calibration Operation. A deployed calibration capability that may include but is not limited to Rapid Assistance Support for Calibration (RASCAL), Transportable Field Calibration Unit (TFCU), mobile calibration teams, combined operations with a foreign military, etc. See Figure 1-1 for the PMEL support matrix.
DEPLOYED TMDE
THAT IS NOT
RAINBOWED
DEPLOYED TMDE
THAT IS
RAINBOWED
TMDE PERMANENTLY
ASSIGNED WITHIN AOR
STEADY-STATE/
ROTATIONAL
OPERATIONS
Customer Uses Reachback Support From Their Home
Station
Customer Uses Steady State CRF 2, 3,4
OR
Customer Uses
Reachback Support From Their Home Station
Customer Uses Contingency State CRF 1, 3, 4
OR
Customer Uses Nearest
USAF Activity with PMEL Capability 3,4
CONTINGENCY/ CRISIS
OPERATIONS
Customer Uses Contingency State CRF 1, 3, 4
OR
Customer Uses Nearest USAF Activity With PMEL Capability 3, 4
NOTES: 1. Contingency CRFs provide regional intermediate repair to support small-scale rapid response deployments to multiple global locations.
2. Steady state CRFs provide regional intermediate repair to support AEF/AEW rotations.
3. CRF and PMEL as applicable shall request manning assistance from the LGM/A4 of the MAJCOM/AFFOR (Ref: AFI 21-101, Aircraft and Equipment Maintenance Management).
4. CRF and PMEL as applicable shall use the customer’s AF Form 616 Fund Cite Authorization to fund repair parts (Ref: AFI 21-101).
Figure 1-1. PMEL Support Matrix.
1.4.14 Contract Calibration. Calibration obtained from a contracted source, funded by the user or an Item/Program Manager. Neither the PMEL nor AFMETCAL centrally manages or funds these calibrations. These items are identified with a calibration responsibility of “CONTRACT”. All CONTRACT TMDE shall be coordinated with the PMEL for reporting purposes.
1.4.15 Correction Chart. A chart prepared for a specific item. The chart shows corrections to apply to indicated values to obtain true value. In instances where an item is charted and the item’s accuracy or uncertainty is applied, the charted values can be used through the item’s calibration interval. However, in instances where a better accuracy or uncertainty of another standard is “transferred” and being used, the charted values may only be used to complete the immediate calibration requiring the transfer/charting. The AFPSL is authorized to assign an interval only after a documented engineering analysis is completed that supports the transfer of accuracy or uncertainty to the assigned interval.
1.4.16 Equipment Types:
a. Mission Equipment. Part of operational ground or airborne systems.
b. Non-TMDE. SE not meeting the TMDE definition. This equipment does not require calibration and is the responsibility of the user.
c. Precision Measurement Equipment (PME). For purposes of this TO, PME is the same as TMDE. This is the former title for TMDE.
d. Support Equipment (SE). A broad category of equipment and tools used to verify and maintain mission equipment. SE can include equipment used to verify the operation of other support equipment. SE is categorized as either TMDE or Non-TMDE. The AFMETCAL Program is primarily concerned with TMDE support equipment.
e. Test, Measurement and Diagnostic Equipment (TMDE). Devices used to maintain, evaluate, measure, calibrate, test, inspect, diagnose or otherwise examine materials, supplies, equipment and systems to identify or isolate actual or potential malfunction, or decide if they meet operational specifications established in technical documents. ISO 17025 refers to this equipment as “measuring and test equipment.”
1-5
1.4.17 ISO/IEC Standard 17025. The international standard for commercial metrology, “General Requirements for the Competence of Calibration and Testing Laboratories.” It includes the relevant requirements of ISO 9000 series standards.
The AFMETCAL program does not approve this standard as the sole requirement for operating a USAF PMEL.
1.4.18 Manager. The individual responsible for day-to-day management and operation of the PMEL and its associated work centers, such as production control, quality program and material control. Synonymous with PMEL superintendent, PMEL branch chief, TMDE branch chief, TMDE flight chief, TMDE section chief and Laboratory Manager. The PMEL manager is referred to as the PMEL node manager under RNI.
NATIONAL MEASUREMENT STANDARD
AIR FORCE MEASUREMENT STANDARDS AIR FORCE PRIMARY
AFPSL WORKING STANDARDS STANDARDS LABORATORY
TRANSFER STANDARDS (AFPSL)
BASE MEASUREMENT STANDARDS PRECISION MEASUREMENT
WORKING STANDARDS EQUIPMENT LABORATORY
(PMEL)
TMDE USER
PERFORMING WORK CENTER
SUPPORT EQUIPMENT
MISSION EQUIPMENT
Figure 1-2. Measurement Traceability of the Air Force Metrology and Calibration Program.
1.4.19 Measurement or Reference Standards. Equipment by which the accuracy and traceability of a precision measurement is derived. The typical chain of measurement standards is shown in Figure 1-2. The following are types of standards:
a. Air Force Measurement Standards. Equipment certified by NIST or other AFMETCAL approved sources.
These standards are used by the AFPSL as a basic measurement reference for the Air Force.
b. Air Force Primary Standards Laboratory (AFPSL) Working Standards. TMDE certified by the AFPSL and used to calibrate other TMDE.
c. Base Measurement Standards. Equipment certified by the AFPSL or AFMETCAL approved sources. These standards are used by an Air Force PMEL as a measurement reference. Base Measurement Standards are designated in TO 33K-1-100-2.
d. Exchange Standard. A calibration standard calibrated by the AFPSL through a calibrate-and-replace methodology to minimize PMEL down time. An exchange standard is calibrated by the AFPSL and sent to a PMEL to replace a like base measurement standard before its calibration has expired. The exchange standard whose calibration interval is expiring is returned to the AFPSL for calibration, and then forwarded to another PMEL. TO 33K-2-11, Detailed Instruction and Information Pertaining to Air Force Primary Standards, Laboratory Calibration Exchange Standards, and Artifact Proficiency Testing Standards, PMELs Only, contains detailed instructions and information for Exchange Standards.
1-6
e. National Measurement Standards. Equipment or physical constants identified and normally used by NIST.
These standards serve as a basic measurement reference for use throughout the United States and at U.S.
installations/activities overseas.
f. Working Standards. TMDE certified by PMEL and used to calibrate other TMDE.
g. Transfer Standard. A measurement standard calibrated by the AFPSL and circulated to multiple PMELs to calibrate base measurement standards or working standards. A transfer standard is calibrated by the AFPSL, forwarded to a PMEL, used by the PMEL to calibrate a base or working standard, and returned to the AFPSL or forwarded to the next PMEL as directed. TO 33K-2-11 contains detailed instructions and information for Transfer Standards.
1.4.20 Metrology. The science or system of weights and measures used to determine conformance to technical requirements. This includes the development of standards and systems for absolute and relative measurements.
1.4.21 Precision Measurement Equipment Laboratory (PMEL). A laboratory authorized to own and use base measurement standards to maintain working standards. The working standards are used along with PMEL-owned TMDE to maintain (troubleshoot, align, repair, and calibrate) TMDE designated as PMEL responsibility. Authorizations for PMEL equipment and facility requirements are tailored to meet specific requirements for supported missions. PMELs are the base-level link for measurement transfer and maintenance self-sufficiency for all systems in the Air Force. PMEL types are:
a. Type IIA. A base-level PMEL with a 68 °F room providing support to an Air Logistics Complex (ALC) and/or a large geographical area. Type IIA PMELs supporting ALCs are operated by AFMC. PACAF theater support command operates the Type IIA PMELs at Kadena AB, JA and Elmendorf AFB, AK. USAFE theater support command operates the Type IIA PMEL at RAF Feltwell, GB.
b. Type IIB. A base-level PMEL established to support aircraft, missiles, ground systems and/or other equipment on base or in the local area.
c. Type IIC. A base-level PMEL with a 68°F room supporting research, development, test or evaluation programs, as well as other operational and support functions.
d. Type III. A PMEL supporting a specific mission. It normally receives calibration support from a Type II( ) PMEL. With the exception of Rapid Assistance Support for Calibration (RASCAL) sets, a Type III PMEL is not authorized at an installation where a Type II( ) PMEL exists.
e. Type IV. A PMEL supporting a specific weapon system. It uses a transportable measurement system in fixed and deployed locations. A Type IV PMEL receives calibration support from a Type II( ) PMEL. It may be located at an installation where a Type II( ) PMEL exists. Type IV PMELs are no longer authorized in the USAF metrology program. They still exist, however, in FMS country metrology programs.
1.4.22 PMEL Mobile and On-Site Measurement Capabilities.
a. Rapid Assistance Support for Calibration (RASCAL). A portable PMEL designed for rapid deployment to support emergency calibration needs. It consists of five environmentally controlled shelters containing measurement standards and other TMDE. The shelters are joined together when in use. They are taken apart and airlifted to the needed location.
b. Jet Engine Test Stand Calibrator (JETSC). TMDE and accessories needed to perform on-site calibration of jet engine test stands, housed in a trailer or set of ruggedized cases.
c. Portable Automatic Test Equipment Calibrator (PATEC). A PATEC is used to calibrate Automatic Test Equipment (ATE) on-site. A PATEC consists of a set of portable standards. A Core PATEC meets the calibration needs of a variety of test stations.
d. Transportable Field Calibration Unit (TFCU). The TFCU consists of working standards in hardened cases selected to function within given performance tolerances as long as the equipment is operated within the published environmental restrictions of the TFCU. The TFCU is used to calibrate TMDE located at remote sites or bases.
1-7
1.4.23 PMEL Quality Assurance (PQA) Evaluator. Shall be a highly qualified 2P071 or equivalent when possible. PQA evaluators are selected by PMEL management and appointed in writing. PQA evaluators are the only personnel designated to perform Quality Reviews and Process Reviews applicable to the Quality Program.
1.4.24 Proficiency Testing/Measurement Assurance Program (PT/MAP). The PT/MAP is a process for assessing laboratory capability by means of inter-laboratory data comparisons and exchange standard intercomparisons. Results from PT/MAP are an indication of a laboratory’s competence and the integrity of measurement standards and can be an integral part of the assessment process.
1.4.25 Rainbowed Equipment. Non-prepositioned equipment that is sourced from multiple units, left in place, and used by multiple units over time to enable the best support of vulnerable AEFs while minimizing home station impact on supporting units.
1.4.26 Reachback. The process of obtaining products, services and applications of forces, equipment or materiel from Air Force organizations not forward deployed. This capability allows commanders to obtain or coordinate support from units not physically located with the forward force. See AFI 21-101 for more information.
1.4.27 Risk. Any potential that negatively affects SART, capability, mission or the RNI network (based on time, cost, quality, performance, etc). When required, risk shall be described as Low, Moderate or High.
1.4.28 Risk Management. The process of identifying, assessing, and prioritizing risks to develop mitigation strategies in an effort to accept, control, avoid, minimize negative impacts. Risks identified as Moderate or High must have a corrective action documented and performed.
1.4.29 Shall. Shall is a directive statement. Reference TO 00-5-1, AF Technical Order System.
1.4.30 Test Accuracy Ratio/Test Uncertainty Ratio (TAR/TUR). The ratio of the uncertainty (accuracy) of the TI to the uncertainty (accuracy) of the standard (for example 4:1). TAR and TUR are equivalent in the Air Force Metrology and Calibration Program, as the term Accuracy has been used for the (more correct) term Uncertainty in Air Force Calibration procedures.
1.4.31 Traceability. The property of a measurement result whereby the result can be related to a reference through a documented unbroken chain of calibrations, each contributing to the measurement uncertainty. Mandatory USAF traceability relates individual measurement results to national standards at NIST or AFMETCAL approved sources through an unbroken chain of calibrations all having stated uncertainties as shown in Figure 1-2.
1.4.32 User. User refers to the using activity or owning organization responsible for calibration and maintenance of items designated “USER” responsibility in a CMS or TO 33K-1-100-2. In most cases, the ‘user’ and the ‘owner’ are the same.
In general, the ‘user’ is the activity that physically uses the item. The ‘owner’ is the organization that owns the item via a supply account. The 'user' performs USER calibrations or coordinates with the PMEL for assistance when resources are not available. The PMEL will assist to identify required equipment or approved sources of support. As a last resort, the PMEL may perform the calibration. The owner is responsible for all funding associated with these efforts.
1.5 PROGRAM ABBREVIATIONS.
a. ABS Absolute Standard
b. AEF Air Expeditionary Force
c. ALC Air Logistics Complex
d. ASC Allowance Source Code
e. CAP Corrective Action Plan
f. CBU Calibrate Before Use
g. CEE Contingency Equipment Exception
h. CPIN Computer Program Identification Number
i. CRF Centralized Repair Facility
1-8
j. ECS Environmental Control System
k. ESD Electrostatic Discharge
l. FMS Foreign Military Sales
m. GEOLOC Geographic Location
n. MAD Measurement Area Discipline
o. MCF Metrology & Calibration Flight
p. OEM Original Equipment Manufacturer
q. OWC Owning Work Center
r. PAMS PMEL Automated Management System
s. PAS PMEL Activity Summary
t. PNC Process Nonconformity
u. PPP Public Private Partnership
v. PQA PMEL Quality Assurance
w. PR Process Review
x. PRP Process Review Plan
y. PT/MAP Proficiency Testing/ Measurement Assurance Program
z. PWC Performing Work Center
aa. QNC Quality Nonconformity
bb. QP Quality Program
cc. QR Quality Review
dd. RADIAC Radiation, Detection, Indication, and Computation
ee. RC Root Cause
ff. RNI Repair Network Integration
gg. SA Security Assistance
hh. SART Safety, Accuracy, Reliability and Traceability
ii. SATAF Site Activation Task Force
jj. SBSS Standard Base Supply System
kk. SICL See Individual Component Listing
ll. SICW See Item Calibrated With
mm. TCM Technical Content Manager
nn. TI Test Instrument
oo. TPE Trainer Proficiency Evaluation
pp. TRC Technical Repair Center
qq. WRM War Reserve Material
rr. WUC Work Unit Code
2-1
SECTION 2
AFMETCAL PROGRAM ORGANIZATION AND INTERFACE
2.1 GENERAL.
The AFMETCAL Program includes AFMETCAL, PMELs and their resources, and the TMDE in the Air Force. The directives that provide the policy and guidance are also part of the program. The purpose of the AFMETCAL Program is to ensure that all Air Force TMDE used to make measurements is accurate, reliable, provides standardized measurements, and is traceable through the AFPSL to NIST or other AFMETCAL approved sources. This TMDE is used to make sure that each system and subsystem is accurate and reliable and ensures that they properly interface with other systems.
2.2 AFMETCAL PROGRAM.
2.2.1 Figure 2-1 Description. Figure 2-1 depicts each element of the AFMETCAL Program and the interrelationship of major Air Force organizations. The center circle identifies the major elements of the AFMETCAL Program: AFMETCAL Director (AFLCMC/WNM), the PMELs and TMDE users. The focus of the AFMETCAL Program is to ensure accuracy and traceability of TMDE. This TMDE is used to ensure the accuracy and reliability of all Air Force systems. The resources needed to operate the AFMETCAL Program are identified in the rectangles. In cooperation with the other elements, AFMETCAL provides program guidance, PMEL standards, traceability of standards, calibration procedures, PMEL evaluations, metrology engineering support, and determines calibration requirements of systems. The PMELs are the performing work centers containing the personnel that calibrate, certify and repair the TMDE owned and used by Air Force organizations. The AFMETCAL program requires users and PMELs to obtain calibration service from Air Force PMELs or the AFPSL as appropriate. AFMETCAL must approve obtaining calibration service from other sources. Users of TMDE utilize their PMEL calibrated items to verify their support equipment (test benches, simulators, etc.) as well as major systems or subsystems. Users of TMDE are assigned owning work center (OWC) codes.
2.2.2 Resources used by AFMETCAL Program. There are four primary groups of resources used by the AFMETCAL Program to do its mission (personnel, equipment/material, documentation and facilities). A basic need is sufficient personnel, having the necessary education, training, technical knowledge and experience for their assigned functions. These personnel require directives/procedures, material, and facilities. Calibration procedures provide the calibration methods.
The PMEL requires parts and material to perform calibration and repair of TMDE. They also need their own measurement standards. The PMEL facility has some special requirements based on the system(s) it supports. The requirements of the program are based upon the measurement needs, and the accuracy of the measurements required of all systems supported by the Air Force. The outer ring of Figure 2-1 identifies the organizations that interface with the AFMETCAL Program.
They provide, or coordinate, in the provisioning of the resources for the program. The outer ring of Figure 2-1 also shows the planning required to ensure that the proper AFMETCAL Program support is available. That support is needed for all systems from the time a system is under development through its entire life cycle.
2.2.3 Funding. Material and personnel needs are satisfied according to the amount of funding available.
2.2.3.1 Funds for daily operating expenses of the PMELs are provided by the owning command. They generally come from operation and maintenance (O&M) appropriations. Air Force regional workload support is generally funded by O&M appropriations.
2.2.3.2 Centrally procured investment equipment funding for equipment costing less than $250K per item is provided to the AFMC procuring agency by MAJCOMs.
2.2.3.3 Centrally procured expense (parts) item funding comes from the USAF Stock Fund.
2.2.3.4 Operating command or base procured investment and expense items are generally funded from O&M and equipment funds.
2.2.3.5 Facility funds are part of 3300 (Military Construction – Air Force) appropriations. This includes major and some minor construction projects as well as facility project planning efforts.
2-2
2.2.3.6 AFMC is…