36E77619R0011-0002005.pdf

PDF 312 KB Posted

Attached to
Construct New Radiation Oncology Suite Milwaukee VAMC (Minor)
Solicitation number
36E77619R0011
Issued by

About this file

36E77619R0011 0002 ES602 - ELECTRICAL SCHEDULES.pdf

View the file

Other files

File Type Posted
36E77619R0011-0007001.pdf PDF
36E77619R0011-0007000.docx DOCX document
36E77619R0011-0006003.pdf PDF
36E77619R0011-0006000.docx DOCX document
36E77619R0011-0006001.docx DOCX document
36E77619R0011-0006002.pdf PDF
36E77619R0011-0005005.pdf PDF
36E77619R0011-0005006.pdf PDF
36E77619R0011-0005003.pdf PDF
36E77619R0011-0005002.pdf PDF
36E77619R0011-0005004.pdf PDF
36E77619R0011-0005000.docx DOCX document
36E77619R0011-0005001.docx DOCX document
36E77619R0011-0004000.docx DOCX document
36E77619R0011-0004002.pdf PDF
36E77619R0011-0004003.pdf PDF
36E77619R0011-0004001.pdf PDF
36E77619R0011-0003000.docx DOCX document
36E77619R0011-0002002.pdf PDF
36E77619R0011-0002007.pdf PDF
36E77619R0011-0002010.docx DOCX document
36E77619R0011-0002000.docx DOCX document
36E77619R0011-0002001.pdf PDF
36E77619R0011-0002009.pdf PDF
36E77619R0011-0002008.pdf PDF
36E77619R0011-0002003.docx DOCX document
36E77619R0011-0002006.pdf PDF
36E77619R0011-0002004.pdf PDF
36E77619R0011-0001000.docx DOCX document
36E77619R0011-0001001.pdf PDF
36E77619R0011-000.docx DOCX document
36E77619R0011-005.pdf PDF
36E77619R0011-007.pdf PDF
36E77619R0011-012.pdf PDF
36E77619R0011-003.pdf PDF
36E77619R0011-008.pdf PDF
36E77619R0011-002.pdf PDF
36E77619R0011-001.pdf PDF
36E77619R0011-009.docx DOCX document
36E77619R0011-006.pdf PDF
36E77619R0011-010.docx DOCX document
36E77619R0011-011.pdf PDF
36E77619R0011-004.pdf PDF
36E77619R0011_1.docx DOCX document
Show all 44

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

LIGHT FIXTURE SCHEDULE

FIXTURE LIGHT SOURCE BALLASTS/DRIVERS INPUT

VOLT MOUNT

SURFACE FIXTURE FIXTURE SPECIFICATION AND ACCEPTABLE MANUFACTURERS

OPTIONS

SEE

ID DESCRIPTION TYPE NO. DESCRIPTION WATTS K CRI (MIN.) TYPE CONTROL NO. WATTS TYPE DEPTH MANUFACTURER CATALOG SERIES NOTE

= SPARE

= SPARE

= SPARE = SPARE

= SPARE

= SPARE = SPARE

= SPARE

= SPARE

= SPARE

= SPARE

= SPARE = SPARE = SPARE

= SPARE = SPARE = SPARE = SPARE

OA

4" DIA APERTURE DOWNLIGHT 1000

R DW <7"

GOTHAM EVO4

WIDE BEAM DISTRIBUTION LED - DELIVERED 13 4000 80 DR SW 1 13 277 8,41 1C

LUMENS

OD

16IN.W X 7IN. H WALL MOUNTED LIGHT 3000 GARDCO 101L 16L 530

TYPE 4 DISTRIBUTION LED - DELIVERED 28 4000 80 DR SW 1 28 MVOLT S W - 10,28 1D

LUMENS

OE

16IN.W X 7IN. H WALL MOUNTED LIGHT 3000 GARDCO 101L 16L 530

TYPE 2 DISTRIBUTION LED - DELIVERED 28 4000 80 DR SW 1 28 MVOLT S W - 10,28 1D

LUMENS

UC

12" NOMINAL UNDERCABINET 300 TECH LIGHTING 700UCFDW

SOLID FRONT; INTEGRAL ROCKER SWITCH LED - DELIVERED 14 4000 80 DR SW 1 14 120 S CAB - 25,28,54 1F,5

LUMENS

WA

2FT VANITY LIGHT 1200 OXYGEN 3-524

HORIZONTAL MOUNT LED - DELIVERED 26 4000 80 DR SW 1 26 277 S W - SCOTT ARCHITECTURAL S3720 8,28

MATTE WHITE ACRYLIC DIFFUSER LUMENS BROWNLEE 5190-26

XA

SINGLE FACE EXIT SIGN LITHONIA LQM

THERMOPLASTIC LED - IN UNIT 3 - - - - - 3 277 S V - 28,61,62 1G

WHITE

XB

DOUBLE FACE EXIT SIGN LITHONIA LQM

THERMOPLASTIC LED - IN UNIT 3 - - - - - 3 277 S V - 28,61,62 1G

WHITE

XG

EXIT SIGN - SINGLE FACE "IN USE" LITHONIA LQM

THERMOPLASTIC HOUSING LED - IN UNIT 3 - - - - - 3 277 S V - 28,61,62 1G

UNIVERSAL MOUNTING

NOTE:

SEE SPECIFICATION SECTIONS 26 50 00 FOR ADDITIONAL INFORMATION REGARDING LIGHTING FIXTURE AND INSTALLATION REQUIREMENTS. PROVIDE OPTIONS AND ACCESSORIES REFERENCED BY THE COLUMN TITLED "OPTIONS".

MANUFACTURERS AS ACCEPTABLE SHALL MEET ALL REQUIREMENTS AND FEATURES INDICATED. ACCEPTABLE MANUFACTURERS MUST MEET THE PHOTOMETRIC PERFORMANCE OF THE LISTED UNIT.

ABBREVIATIONS:

AR = AS REQUIRED DR = DRIVER GRD = GRADE PD = PENDANT SUSP = SUSPENDED

BL = BI-LEVEL SWITCHING DW = DRYWALL H = HID PL = PLASTER SW = SWITCHING

CAB = UNDERCABINET ELV = ELECTRONIC LOW VOLTAGE DIMMING HAL = HALOGEN PS = PULSE START T = CERAMIC/GLASS TILE

CL = CLEAR EM = EMERGENCY HBF = HIGH BALLAST FACTOR R = RECESS TW = TUNABLE WHITE LED

DALI = ADDRESSABLE, DIMMING <1-100% ES = EXPOSED STRUCTURE I = INCANDESCENT RGB = RED, GREEN, BLUE LEDS V = VARIES

DM0 = LED DIMMING 0-10V, 0 -100% ET = ELECTRONIC BALLAST LBF = LOW BALLAST FACTOR RM = REMOTE W = WALL MOUNTED

DM1 = LED DIMMING 0-10V, 1 -100% F = FLUORESCENT LG = LAY-IN GRID S = SURFACE WD = WOOD

DM5 = LED DIMMING 0-10V, 5 -100% FD1 = FLUORESCENT DIMMING 1-100% MLV = MAGNETIC LOW VOLTAGE DIMMING SPEC = SPECIAL XFMR = TRANSFORMER

DM10 = LED DIMMING 0-10V, 10 -100% FD5 = FLUORESCENT DIMMING 5-100% NBF = NORMAL BALLAST FACTOR ST = STANDARD

DMW = DIM TO WARM LED FD10 = FLUORESCENT DIMMING 10-100% OT = OTHER STN = STONE

DMX = ADDRESSABLE, DIMMING <1-100%

OPTIONS:

1 = POST PAINTED FINISH 21 = FURNISH WITH AIR HANDLING CAPABILITY 41 = MATTE DIFFUSE LOW IRIDESCENT REFL. 61 = FURNISH WITH UNIVERSAL ARROWS AND RED STENCIL FACE

2 = REGRESSED ALUMINUM DOOR FRAME 22 = FURNISH WITH WIRE GUARD 42 = FURNISH WITH SYMMETRICAL REFLECTOR 62 = FURNISH WITH FIELD ADJUSTABLE CHEVRONS

3 = FLAT STEEL DOOR FRAME 23 = FURNISH WITH ENDCAPS 43 = FURNISH WITH ASYMMETRICAL REFLECTOR 63 = FURNISH WITH MIRRORED BACKGROUND

4 = FLAT ALUMINUM DOOR FRAME - MITERED CORNERS 24 = FURNISH IN LENGTH AS SHOWN 44 = FURNISH WITH TOP AND BOTTOM LENS 64 = FURNISH WITH INTEGRAL BATTERY

5 = STAINLESS STEEL TRIM AND DOOR FRAME 25 = FURNISH WITH SOLID FRONT 45 = FURNISH WITH LOUVER 65 = FURNISH WITH INTEGRAL HIGH CAPACITY BATTERY (FOR REMOTE HEADS)

6 = FURNISH WITH SPRING LOADED LATCHES 26 = FURNISH WITH DUST COVER 46 = FURNISH WITH PERFORATED DIFFUSER 66 = FURNISH WITH SELF DIAGNOSTICS

7 = FURNISH WITH HINGED DOOR 27 = CUSTOM FINISH - SELECTION TBD 47 = FURNISH WITH EGGCRATE LOUVER 67 = FURNISH WITH AUXILIARY EMERGENCY BATTERY BALLAST/DRIVER

8 = DAMP LOCATION CONSTRUCTION 28 = STANDARD FINISH - SELECTION TBD 48 = FURNISH WITH SOFTENING LENS 68

9 = WET LOCATION CONSTRUCTION 29 = FURNISH WITH HOUSE SIDE SHIELD 49 = FURNISH WITH PRISMATIC LENS 69

10 = IP65 RATING 30 50 = FURNISH WITH LINEAR SPREAD LENS 70

11 = IP66 RATING 31 51 = FURNISH WITH SPREAD LENS 71 = FURNISH WITH ANTI-MICROBIAL FINISH

12 = IP67 RATING 32 52 72 = CERTIFIED FOR MIL-STD461F

13 = NEMA 4X RATED 33 = FURNISH WITH CHAIN MOUNTING 53 73 = FURNISH WITH RF NOISE SUPPRESSORS (ADVANCE RIF-1) FOR EACH DRIVER

14 = IC RATING 34 = FURNISH WITH RIGID STEM MOUNTING 54 = FURNISH WITH ROCKER SWITCH 74 = FURNISH WITH ACRYLIC LENS WITH INTEGRAL RFI SHIELDING

15 = ADA COMPLIANT 35 = FURNISH WITH SWIVEL CANOPY 55 = FURNISH WITH INTEGRAL OCC. SENSOR 75 = FURNISH WITH LOW VOLTAGE CONTROL (LVC)

16 = DOUBLE GASKETED-LENS TO TRIM TO CEILING 36 = FURNISH WITH ADJUSTABLE AIRCRAFT CABLE MOUNTING 56 = FURNISH WITH INTEGRAL DAYLIGHT SENSOR 76

17 = TRIPLE GASKETED-LENS TO DOOR TO HOUSING TO CEILING 37 = FURNISH WITH ADJUSTABLE MOUNTING HARDWARE 57 = FURNISH WITH INTEGRAL PHOTOCELL 77

18 = FURNISH WITH TAMPER RESISTANT HARDWARE 38 = FURNISH WITH MANUFACTURER'S CUSTOM JUNCTION BOX 58 = FURNISH WITH REMOTE DRIVER 78

19 39 = FURNISH WITH RECESSED SLOPE ADAPTER 59 79

20 40 60 80

GENERAL NOTES:

1. VERIFY CEILING T-BAR WIDTH (I.E. 9/16", 15/16", 1.5") WITH THE ARCHITECT/INTERIORS SPECIFICATION PRIOR TO ORDERING LAY-IN FIXTURES.

2. IN ALL INSULATED OR FIRE RATED CEILINGS, THE EC SHALL BE RESPONSIBLE FOR THE PROPER TENTING (OR FIRE BOXING) OF THE RECESSED FIXTURES. FOR PREFAB PRODUCTS REFER TO TENMAT OR SPECIALTY PRODUCTS AND INSULATION.

3. VERIFY DIMMING BALLAST / DRIVER COMPATIBILITY WITH DIMMING CONTROL.

4. REFER TO ARCHITECTURAL ELEVATIONS FOR EXACT LOCATION AND MOUNTING HEIGHT.

5. TO MEET SPECIFIED DELIVERED LUMEN PACKAGE PROVIDE PROGRAMMED DRIVER AS REQUIRED.

6. FIXTURE PERFORMANCE TOLERANCES: CCT +/- 200 DEGREES; DELIVERED LUMENS +/- 5%; ENERGY CONSUMPTION +/- 5%; BEAM SPREAD DISTRIBUTION +/- 5 DEGREES AND AS OUTLINED IN 26 50 00.

7. IF SUBMITTING A FIXTURE FOR APPROVAL AS AN EQUAL, PROVIDE CUT SHEET AND IES PHOTOMETRIC FILE PRIOR TO BIDDING. A WORKING SAMPLE OF THE FIXTURE MAY BE REQUIRED FOR APPROVAL BY THE A/E.

LIGHT FIXTURE SCHEDULE NOTES:

1. MANUFACTURERS, PRODUCTS, AND DESCRIPTIONS LISTED IN THE LIGHTING FIXTURE SCHEDULE ARE THE BASIS OF DESIGN SPECIFICATION FOR EACH FIXTURE TYPE. EQUAL PRODUCTS WILL BE ENTERTAINED UNLESS OTHERWISE NOTED. ALL PROPOSED EQUAL PRODUCTS ARE REQUIRED TO MEET OR EXCEED ALL OF THE SPECIFICATION REQUIREMENTS LISTED IN THE LIGHTING FIXTURE SCHEDULE INCLUDING BUT NOT LIMITED TO: PRODUCT GRADE, LIGHT SOURCE TECHNOLOGY, CONTROL TECHNOLOGY, OPTICAL PERFORMANCE (DELIVERED LUMENS OR CBCP), DISTRIBUTION, FIXTURE ACCESSORIES, CONSTRUCTION, DIMENSIONS, MOUNTING, FINISH, ENERGY PERFORMANCE, AND LISTINGS/RATINGS.

THE SPECIFIER HAS EVALUATED ALL OF THE MANUFACTURERS LISTED BELOW AND CONSIDERS THESE MANUFACTURERS TO BE OF A QUALITY THAT MEET GENERAL SPECIFICATION REQUIREMENTS FOR THE CATEGORY OF PRODUCT UNDER WHICH THEY ARE LISTED. ALTHOUGH MANUFACTURERS ARE LISTED BELOW IT DOES NOT ENSURE THAT THEY HAVE PRODUCTS THAT MEET ALL OF THE SPECIFICATION REQUIREMENTS FOR EACH FIXTURE TYPE. IT MEANS THAT THESE MANUFACTURERS HAVE ALREADY BEEN EVALUATED AND HAVE PRODUCTS THAT SHALL BE REVIEWED FOR APPLICABILITY AS EQUALS.

MANUFACTURERS THAT ARE NOT LISTED BELOW WILL REQUIRE A WORKING SAMPLE FOR SPECIFIER TO REVIEW IN ORDER TO BE EVALUATED FOR APPLICABILITY.

THE BURDEN OF MEETING THE BASIS OF DESIGN SPECIFICATION IS ON THE DISTRIBUTOR QUOTING THE PROJECT. DETERMINATION OF EQUIVALENCY IS THE RESPONSIBILITY OF THE SPECIFIER. EVEN IF A PROPOSED EQUAL PRODUCT IS FROM ONE OF THE MANUFACTURERS LISTED BELOW, IF IT DOES NOT MEET THE BASIS OF DESIGN SPECIFICATION REQUIREMENTS THEN IT WILL BE REJECTED. THERE SHALL BE NO ADDITIONAL COST ASSOCIATED WITH REJECTION OF A FIXTURE NOT DEEMED EQUIVALENT BY THE SPECIFIER.

A. SPECIFICATION GRADE TROFFERS – FINELITE, MARK ARCHITECTURAL LIGHTING, PRUDENTIAL

B. COMMERCIAL TROFFERS AND WHITEGOODS – COLUMBIA, CREE, DAYBRITE, DAY-O-LITE, HE WILLIAMS, LITHONIA, METALUX, MERCURY , PMC

C. SPECIFICATION GRADE DOWNLIGHTS – FOCAL POINT, GOTHAM EVO, INTENSE MX, JUNO-INDY, LEDRA ALPHABET, LIGHTOLIER, PORTFOLIO, PRESCOLITE LITEFRAME

D. SPECIFICATION GRADE EXTERIOR:

LEVEL 1: GARDCO, HYDREL, INVUE, KIM LIGHTING

LEVEL 2: HUBBELL, LITHONIA, MCGRAW EDISON

E. HEALTHCARE DOWNLIGHTS: EVERBRITE MEDLUX (MRI ONLY), FAILSAFE, HUBBELL HEALTHCARE, KENALL, KIRLIN, KURTZON, NEW STAR

F. ARCHITECTURAL UNDERCABINET: AFX NOBLE PRO, FEELUX TUNELIGHT2, TECH LIGHTING AMBIANCE GLYDE, TECH LIGHTING UNILUME

G. EXIT AND EMERGENCY – CHLORIDE, DUAL-LITE, EMERGI-LITE, EVENLITE, HE WILLIAMS, ISOLITE, LITHONIA, MULE, SURE-LITES

2. PROVIDE A BROKEN-OUT COST IN THE BID FORM FOR ALL SINGLE NAME SPECIFIED FIXTURE TYPES.

3. PROVIDE REMOTE TRANSFORMER IN RATING NECESSARY TO ACCOMMODATE LOAD. REMOTE TRANSFORMER SHALL BE LOCATED IN THE NEAREST ACCESSIBLE LOCATION OUT OF VIEW. FIELD VERIFY WITH ARCHITECT.

4. ARCHITECT/OWNER SHALL SELECT IMAGE PRIOR TO ORDER RELEASE.

5. REFER TO DETAIL 6/ES502 FOR MOUNTING INFORMATION.

6. PROVIDE WITH CONTINUOUS ILLUMINATION ENTIRE LENGTH OF FIXTURE. LENSING SHALL BE FASTENED SO THAT AT NO POINT ARE LED NODES VISIBLE.

GENERAL EQUIPMENT SCHEDULE

FEED FROM OCPD BRANCH WIRING MOTOR CONTROLLER DISCONNECT SWITCH CONNECTION REQUIREMENTS SEE

NO. SERVING LOC. VOLT/ HP FLA KW PANEL CKT BRNH TYPE SIZE POLE NO. SIZE GRD COND GRD FURN INST WIRE LOC. TYPE OPTIONS FURN INST WIRE LOC. OPTIONS NEMA TYPE OPTIONS MTG HEIGHT NOTE

PHASE (Y/N) SIZE BY BY BY BY BY BY

AO AUTO DOOR OPERATOR SFP 120/1 - - - SFP SFP SFP CB 20 1 2 12 Y 3/4 12 - - - - - - EC EC EC NU 4 - 3,10 - -

AS AIR SEAL MOTOR B0015 120/1 - - - SFP SFP CR CB 20 1 2 12 Y 3/4 12 - - - - - - EC EC EC NU 4 - 3,10 - -

CM COFFEE MACHINE B0200 120/1 - - 1 SFP SFP N CB 20 1 2 12 Y 3/4 12 - - - - - - - - - - - - 2,14 ABV COUNTER -

MG MED GAS ALARM B0201 120/1 - - - EPL-LS-11-2 4 LS CB 20 1 2 12 Y 3/4 12 - - - - - - - - - - - - 3 10 -

OD OVERHEAD DOOR B0015 120/1 - - - SFP SFP CR CB 20 1 2 12 Y 3/4 12 - - - - - - EC EC EC NU 4 - 3 10 -

PL PATIENT LIFT B0015 120/1 - - - SFP SFP CR CB 20 1 2 12 Y 3/4 12 - - - - - - - - - - - - 2 - 1

PS PAGING SYSTEM B0201 120/1 - - - SFP SFP CR CB 20 1 2 12 Y 3/4 12 - - - - - - - - - - - - 3 10

RF REFRIGERATOR B0221 120/1 - - - SFP SFP CR CB 20 1 2 12 Y 3/4 12 - - - - - - - - - - - - 2,14 18"AFF -

SC SECURITY CABINET SFP 120/1 - - - EUL-CR-B001 23 CR CB 20 1 2 12 Y 3/4 12 - - - - - - - - - - - - 3,10 - -

SV SOLENOID VALVE SFP 120/1 - - - SFP SFP N CB 20 1 2 12 Y 3/4 12 MFR EC EC NU - - EC EC EC NU 4 - 3 - -

WC WATER COOLER B0200 120/1 - - 0.6 SFP SFP N CB 20 1 2 12 Y 3/4 12 - - - - - - - - - - - - 2,14 10 -

ABBREVIATIONS:

N = NORMAL BRANCH CB = AMPERE RATED CIRCUIT BREAKER MFR = MANUFACTURER 2SP = 2 SPEED MAGNETIC CONTROLLER SFS = SEE FOOD SERVICE PLANS SA = SEE ARCHITECTURAL DETAILS

LS = LIFE SAFETY BRANCH F = AMPERE RATED FUSES PC = PLUMBING CONTRACTOR VFD = VARIABLE FREQUENCY DRIVE SD = SEE ONELINE DIAGRAM CR = CRITICAL CARE BRANCH OCPD = OVERCURRENT PROTECTIVE DEVICE MAN = MANUAL CONTROLLER IU = IN UNIT BOL = BUILT-IN OVERLOAD EQ = EQUIPMENT SYSTEM EC = ELECTRICAL CONTRACTOR MAG = MAGNETIC CONTROLLER NU = NEAR UNIT WP = WEATHER PROOF

EM = EMERGENCY SYSTEM FSEC = FOOD SERVICE EQUIP. CONTRACTOR CS = COMBINATION CONTROLLER OU = ON UNIT FD = FUSIBLE DISCONNECT

ES = ESSENTIAL SYSTEM MC = MECHANICAL CONTRACTOR RVS = REDUCE VOLTAGE CONTROLLER SFP = SEE FLOOR PLANS LMRS = LOCKABLE MOTOR RATED SWITCH

OPTIONS:

1 = SINGLE RECEPTACLE 4 = NON FUSIBLE DISCONNECT 7 = LOCKABLE 10 = VERIFY WITH MANUFACTURER PRIOR TO INSTALLATION 13 = GFI RECEPTACLE

2 = DUPLEX RECEPTACLE 5 = FUSIBLE DISCONNECT 8 = CONCEALED 11 = SURFACE MOUNTED 14 = FED FROM UPSTREAM GFI BREAKER

3 = DIRECT CONNECTION 6 = WEATHERPROOF 9 = TOGGLE SWITCH 12 = FLUSH MOUNTED

GENERAL EQUIPMENT SCHEDULE NOTES:

1. COORDINATE FINAL MOUNTING HEIGHT WITH MANUFACTURER'S INSTALLATION DRAWINGS.

LOW VOLTAGE SWITCH STATION SCHEDULE - BASEMENT LEVEL

RELAY DESIGNATION LOCATION BUTTON RELAYS CONTROLLED DESCRIPTIONS SEE

CABINET NOTE

LCP/B00-1

0.01 MRI VEST. a R-1 CORRIDOR

0.02 MRI VEST. a R-1 CORRIDOR

0.03 CORR A17 a R-2 CORRIDOR

0.04 CORR A17 a R-2 CORRIDOR

0.05 CORR A15

a R-3,4 CORRIDOR - DAY 1 b R-3,4 CORRIDOR - NIGHT 1

0.06 CORR A15

a R-3,4 CORRIDOR - DAY 1 b R-3,4 CORRIDOR - NIGHT 1

0.07 CORR A14

a R-5 CORRIDOR - DAY 1 b R-5 CORRIDOR - NIGHT 1 c R-11 ACCENT RAISE/LOWER 2

0.08 CORR A14

a R-5 CORRIDOR - DAY 1 b R-5 CORRIDOR - NIGHT 1

0.09 CORR A12

a R-6 CORRIDOR - DAY 1 b R-6 CORRIDOR - NIGHT 1

0.10 CORR A12

a R-6 CORRIDOR - DAY 1 b R-6 CORRIDOR - NIGHT 1

0.11 CORR A13 a R-7 CORRIDOR - DAY

0.12 CORR A13 a R-7 CORRIDOR - DAY

0.13 CORR A13 a R-7 CORRIDOR - DAY

0.14 CORR A15 a R-12 ACCENT RAISE/LOWER 2

LOW VOLTAGE SWITCH STATION GENERAL NOTES:

1. REFER TO LIGHTING CONTROL PANEL SCHEDULE FOR MORE INFORMATION.

2. CONFIRM ALL SCENES/GROUPS WITH ARCHITECT AND OWNER DURING ON SITE PROGRAMMING.

3. DESCRIPTIONS ARE SUGGESTED BUTTON LABELS, OWNER TO CONFIRM SWITCH LABELS PRIOR TO ORDER RELEASE.

LOW VOLTAGE SWITCH STATION SCHEDULE NOTES:

1. SWITCH BUTTON SHALL BE PROGRAMMED AS A SCENE, WHICH MAY INCLUDE MULTIPLE RELAY GROUPINGS AND DIFFERENT DIMMED LEVELS.

2. PROVIDE SWITCH CONFIGURATION THAT SUPPORTS RAISE/LOWER FUNCTION OF RELAY(S) CONTROLLED BY THIS BUTTON.

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Office of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ARCHITECT/ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Consultant:

GUIDON DESIGN INC.

905 N. CAPITOL AVE. SUITE 100 | INDIANAPOLIS, IN. 46204

317.800.6388 | WWW.GUIDONDESIGN.COM

SUSTAINABLE ARCHITECTURE + ENGINEERING

PROJECT LEAD

Architect, Structural Engineer, Civil Engineer

FULLY-SPRINKLERED

PHASING IS DEFINED IN PROJECT SPECIFICATIONS AND PHASING PLANS. CONTRACTOR

TO COORDINATE THE FULL SCOPE OF WORK TO BE PHASED SO AS TO NOT IMPACT THE

OPERATION OF SURROUNDING SPACES OR PREVIOUS PHASES. SPECIFIC COORDINATION

SHOULD BE GIVEN FOR DEMOLITION, UTILITY ROUTING AND SHUTDOWNS.

C

R e v it L o c a lF ile s \7

.0

E c k ro llG T

Z L J .r v t

/2

/2

:5

:4

P

M

ES602

ELECTRICAL SCHEDULES CONSTRUCT NEW RADIATION

ONCOLOGY SUITE

MILWAUKEE VAMC

695-404

04 OCTOBER 2018 Checker MTP

FINAL CONSTRUCTION

DOCUMENTS BLDG 111

17400 West Capitol Drive, Brookfield, WI 53045 Phone: 414.778.1700 / Fax: 414.778.2360 / r-d@ringdu.com

R&D Project No.:

THIS BAR IS 1" LONG.

IF IT MEASURES ANYTHING OTHER THAN 1"

ADJUST SCALE ACCORDINGLY.

217164.00

LIGHT FIXTURE SCHEDULE CONTINUED FROM ES602

Revisions: Date:

ADDENDUM 1 02/22/2019

Sheets
ES601 - ELECTRICAL SCHEDULES
ES602 - ELECTRICAL SCHEDULES