Summary of MSCL Calibration Disciplines and Ranges
Attachment J-5, Page 1
Parameter Range Standard Acceleration - Linear 1 to 10 G's ± 0.50% Rate Table
Acceleration - Dynamic 1 to 10 G's, 30 Hz-900 Hz ± 1% to ± 3% Standard Accelerometers 1 to 10 G's, 900 Hz-10 kHz ± 1% to ± 3%
Attribute Gages Radius ± 25 seconds Optical Comparator, Overlay Angle ± 1 minute
Capacitance 1000 pF, 100 pF, 10 pF ± 5 ppm Standard Capacitors
0.00001 µF to 1µF, 1 kHz ± 0.001% to ± 0.3% Capacitance Bridge
0.00001 µF to 10 mF, 12 Hz to 100 kHz ± 0.02% to 5% LCR Meter
Crimp Tools Dimension ± 100 microinches Pin Gages Crimp Height Digital Point Micrometer
Current-DC 100 µA to 10 A ± 0.015% to 0.05% Current Calibrator 10 A to 100 A ± 0.04% Potentiometer and Precision Shunts 100 A to 500 A ± 0.05% Potentiometer and Precision Shunts 100 nA to 1A ± 0.01% Precision DMM
Current-AC 100 µA to 10 A, 40 Hz to 10 kHz ± 0.04% to 0.40% Current Calibrator 10 to 100 A, 60 Hz to 400 Hz ± 0.05% Precision Shunts 100 µA to 1A, 10 Hz to 50 kHz ± 0.065% to 0.45% Precision DMM
Dial Indicators to 4 inches ± 50 microinches Gage Blocks Supermicrometer
Durometers Type A ± 0.3 duro Durocalibrator Type D ± 0.4 duro
Fiber Optic Attenuation 0-70 dB ± 0.5 dB Fiber Optic Power Meter 850 & 1300 nm Up to 100 um Core
Fiber Optic Power 850 nm, 1300 nm, 1310 nm & 1550 nm ± 1.30% Fiber Optic Power Meter 100 uW Level Up to 100 um Core
Fiber Optic Wavelength 600 to 1750 nm ± 1nm Optical Analyzer
Flow-Gas 0 to 1000 SLPM ± 0.2% reading Mol-Blox
Flow-Gas [Oxygen Clean] 0 to 1000 SLPM ± 0.20% Laminar Flow Elements, Mol-Blox
Accuracy/Uncertainty
Attachment J-5, Page 2
Parameter Range StandardAccuracy/Uncertainty Flow-Liquid 0.4 to 400 GPM (1.5 to 1500 L/min) ± 0.05% Water Flow Calibrator
.001 to 1 GPM (3.8 x10-3 to 3.8 L/min) ± 0.05%
Flow-Linear 20 to 10,000 ft/min ± 3% Wind Tunnel
Frequency 1, 5 & 10 MHz ± 7 x 10-12 Cesium Beam Standard 1 GHz to 26 GHz ± 3 x 10-11 High Frequency Counter
Force 0 to 500 lbs (0 to 2224 Newtons) ± 0.01% Class-S Mass Standard & Moorehouse Dead Weight Tester Compression 0 to 100,000 lbs ± 0.25% Hydraulic Force Cal Sub-System Tension 0 to 50,000 lbs ± 0.25% Hydraulic Force Cal Sub-System
Torque - Wrench 50 to 1000 lbf-ft (68 to 1355 N-m) ± 0.0065 lbf-ft/lbf-ft + 0.1 lbf-ft Torque Standard 1 to 50 lbf-ft (1.36 to 68 N-m) ± 0.0012 lbf-ft/lbf-ft + 0.032 lbf-ft
1.2 to 100 lbf-in (136 mN-m to 11300mN-m) ± 0.0023 lbf-in/lbf-in + 0.086 lbf-in
Torque - Transducer 50 to 1000 lbf-ft (68 to 1355 N-m) ± 0.00032 lbf-ft/lbf-ft + 0.065 lbf-ft Torque Arms, Mass 1 to 50 lbf-ft (1.36 to 68 N-m) ± 0.00017 lbf-ft/lbf-ft + 0.008 lbf-ft
1.2 to 100 lbf-in (136 mN-m to 11300mN-m) ± 0.0002 lbf-in/lbf-in + 0.056 lbf-in
Humidity-Absolute -70 Deg C to -30 Deg C ± 0.3°C Low Humidity Generator -30 Deg C to 10 Deg C ± 0.13°C 10 Deg C to 25 Deg C ± 0.15°C 25 Deg C to 90 Deg C ± 0.3°C
Humidity-Relative 0 to 90% RH ± 0.6% (15% to 90% RH) Humidity Standard
Inductance 1 x 10-4 to 10 H ± 0.02% to ± 0.1% RCL Digibridge
1 x 10-4 to 1200 H @ 1 KHz ± 0.10% to ± 0.3% Standard Inductors to 4 inches ± 25 microinches Gage Blocks, Supermicrometer to 24 inches ± 60 microinches to 72 inches ± 100 microinches to 4 inches ± 15 microinches Gage Blocks to 24 inches ± 45 mciroinches to 72 inches ± 100 microinches to 12 inches ± 25 microinches Labmaster Universal, Gage Blocks to 24 inches ± 60 microinches Supermicrometer, Gage Blocks
Liquids:
Buffer Solution pH 2-10 ± 0.01 pH pH Standards Specific Gravity .001 to 2.11 ± 0.50% Troemner Balance
Inside Dimensional Measuring Devices
Outside Dimensional Measuring Devices
Length/End Standards/Pins & Plugs
Attachment J-5, Page 3
Parameter Range StandardAccuracy/Uncertainty
Mass to 1 lb ± 9 micropounds Standard Weights 5 lb ± 51 micropounds Mass Comparators 10 lb ± 80 micropounds 20 lb ± 130 micropounds 50 lb ± 510 micropounds
100 mg - 1 mg ± 0.004 mg 1 g - 100 mg ± 0.007 mg 1 kg -1 g ± 0.13 mg 2 kg ± 2.6 mg 5 kg ± 5.8 mg 10 kg ± 16 mg 25 kg ± 34 mg
Magnetics 100 to 10 K Gauss (0.01 to 1 T) ± 1% Standard Magnets
Optical Comparators Stage Travel ± 50 microinches Glass Scales Angle Measurement ± 30 seconds Angle Blocks
Surface Plate Flatness Class AA ± 10 microinches Dynamic Measuring System
Phase 0 to 360 Deg (0 to 6.28 Rad) Phase Bridges 1 KHz ± 0.005 Deg. (8.7 x10-5 Rad) < 10 kHz ± 0.030 Deg. (5.2 x10-4 Rad) < 100 kHz ± 0.100 Deg. (1.7 x10-3 Rad) < 1 MHz ± 0.500 Deg. (8.7 x10-3 Rad) < 2 MHz ± 1.000 Deg. (1.7 x10-2 Rad) 0 to 360 Deg, 10 Hz to 50 kHz ± 0.05 Deg. Phase Meter
Power 1 to 2000 Watts, DC to 400 Hz ± 0.10% Precision Voltage and Current Sources
Protractors, Levels to 90 degrees ± 30 seconds Sine Bar, Gage Blocks
Pressure-Static 1e-07 to 4e-3 Torr ± 2.2% reading Stabil Ion Gauges 0 to 104" Hg (0 to 210 kPa) ± 0.004" Hg (±6.75 Pa) Pressure Standard 0 to 10,000 PSI ± 0.02% Pressure Calibration System, Piston Gauge 0 to 1,000 PSI ± 30 ppm RDG Piston Gauge
Pressure [Oxygen Clean] 0 to 1000 PSI ± 0.02% Pressure Calibration System 1000 to 10,000 PSI ± 0.02% Standard Gage
Ratio-DC 1:1 to 1:1000 ± 10 ppm Divider, Resistive
Attachment J-5, Page 4
Parameter Range StandardAccuracy/Uncertainty Ratio-AC 1:1 to 1:1000 ± 10 ppm Ratio Transformers
Resistance 1.0 Ohm ± 1 ppm Thomas 1 Ohm Resistors 10 K Ohms ± 3 ppm Resistors
1 x 10-5 to 10 MOhms ± 0.005% Resistance Bridge
10 M to 1 x 1011 Ohms ± 0.0025% to ± 0.5% Electrometer Resistance Meter
RF-Attenuation 1 to 100 db at 30 MHz ± 0.01 db/10 db Waveguide Beyond Cutoff Attenuator 1 to 20 db, DC to 18 GHz ± 0.2 db Coaxial Attenuators 0 to 100 db, 2.5 MHz to 18 GHz ± 0.3 db maximum IF Substitution
RF-Power .5 to 10 mW, 10 MHz to 18 GHz ± 2% Thermistor Mount 10 to 30 W, 10 MHz to 18 GHz ± 5% Power Sensor/Power Meter 1 to 50W, 10 MHz to 500 MHz ± 5% Calorimetric Test Set and Power Meters
Ring Gages to 3.5 inches ± 20 microinches Labmaster Universal, Gage Blocks, Standard Rings
Rise Time 17.5 pS ± 2.5 pS Rise Time Standard
Sound PressureLevel Sensitivity (Ref. 250 Hz) Referenced to 250 Hz ± 0.08 db to 0.20 db SPL 1", 1/2" and 1/4" Microphones Calibration System Acoustic Response 250 Hz to 25 kHz ± 0.10 db to 0.50 db 1", 1/2" and 1/4" Microphones Calibration System
Steel Rules to 72 inches ± 100 microinches Optical Comparator Dynamic Measuring System
Temperature Ref. -195.881 Deg C (Comparison) ± 0.04° C LN2 Comparator Source Ref. -38.8344 Deg C ± 0.014°C Mercury Triple Point Cell
Ref. 0.01 Deg C ± 0.013° C Water Triple Point Cell Ref. 29.7646 Deg C ± 0.014° C Gallium Melting Point Cell Ref. 231.928 Deg C ± 0.018° C Tin Freeze Point Cell Ref. 419.5276 Deg C ± 0.029° C Zinc Freeze Point Cell
Measure -200 Deg to 660 Deg C ± 0.01° C Platinum Resistance Thermometer 800 Deg C to 2300 Deg C ± 1.5° C to 3.5° C Thermocouple/Thermocouple Simulation
Infrared (IR Thermometry) -15°C to 500°C ± 0.53° C to 2.4° C Black Body Calibrator
Thread Plugs 2-80 TPI, to 3 inches ± 100 microinches Measuring Thread Wires, Supermicrometer, Gage Blocks
Thread Rings 2-80 TPI, to 3 inches ± 200 microinches Setting Plugs
Voltage-DC 10 V ± 0.5 ppm 10 V Standards 100 mV to 1 kV ± 2 ppm Reference divider / Null Detector
1x10-6 to 1.5 kV ± 0.001% Voltage Divider
0.5 kV to 30 kV ± 0.01% Voltage Divider
Attachment J-5, Page 5
Parameter Range StandardAccuracy/Uncertainty
Voltage-AC 2 mV to 70 V, 20 Hz to 1 MHz ± 25 ppm to 1500 ppm AC Measurement Standard 70 V to 1 kV, 20 Hz to 100 kHz ± 50 ppm to 850 ppm AC Measurement Standard 500 mV to 3V, to 1 kHz to 100 MHz ± 0.01% to 1.0% Thermal Voltage converter
VSWR 10MHz to 18 GHz, 40 dB Return Loss ± 2% Network Analyzer
Vacuum 4x10-7 to 1 x10-3 Torr (2.7 x 10-7Pa to 0.13 Pa) ± 5% Ion Gages
1 x 10-5 to 1 Pa. ± 2% Spinning Rotor Gage
1 x 10-3 to 1000 Torr (0.13 Pa to 1.3 x105 Pa) ± 0.08% Vacuum System
Summary