J01k - Dental Equipment Electrical Requirements.pdf

PDF 943 KB Posted

Attached to
Peach Springs Sprinkler System, HVAC, and Communication Cabling
Solicitation number
75H701-23-R-00048
Issued by

About this file

This document provides the requirements for a construction services project to replace sprinkler, HVAC, and communication systems at the Peach Springs Health Center in Peach Springs, Arizona. The Indian Health Service seeks a single firm fixed price contract valued between $1,000,000 and $5,000,000 to remove an existing dry-pipe sprinkler system and install a new wet-pipe system, replace HVAC units beyond their useful lives, and upgrade outdated communication cabling to current standards. Interested offerors must register in the System for Award Management and attend a mandatory site visit detailed in solicitation number 75H701-23-R-00048. Proposals will be evaluated using lowest price technically acceptable criteria, and work must be completed within 365 calendar days of notice to proceed.

View the file

Other files

File Type Posted
B04 Pre-Proposal Inquiries (23R00048).pdf PDF
Amendment A00006 75H701-23-R-00048 2023.06.20.pdf PDF
Amendment A00005 75H701-23-R-00048 2023.06.15.pdf PDF
Amendment A00004 75H701-23-R-00048 2023.06.12.pdf PDF
Amendment A00003 75H701-23-R-00048 2023.06.11.pdf PDF
J02 - WD AZ20230026 Mod 2 2023.06.09.pdf PDF
Pre-Proposal Inquiries (23R00048).pdf PDF
Amendment A00002 75H701-23-R-00048 2023.06.06.pdf PDF
B04 Pre-Proposal Inquiries (23R00048).pdf PDF
J01g - ICRA Template.pdf PDF
Amendment A00001 75H701-23-R-00048 2023.06.01.pdf PDF
J01j - Hydrant Flow Tests.pdf PDF
J01a - Project Drawings.pdf PDF
J06 - Past Performance Questionnaire.docx DOCX document
B01 RFP 23-R-00048 PS Sprinkler Final.pdf PDF
J04 - 2023 Exemption Letter for a Qualifying Hospital.pdf PDF
J05 - Company Specialized Experience Form Construction.docx DOCX document
J02 - WD AZ20230026 Mod 1 2023.02.03.pdf PDF
J04a - AZ Form 5000HC-TPT Healthcare Exemption Certificate.pdf PDF
J01d - PSHC Fire Sprinkler Record Drawings.pdf PDF
J01g - ICRA Template.docx DOCX document
J01 - SOW - PS Sprinkler Project 2023.04.13.pdf PDF
J01f - Existing GFE to Move to Swing Space.pdf PDF
J01i - Electrical Panel Photos.pdf PDF
J01b - PSHC - Original As-built Drawings.pdf PDF
J01c - PSHC IPC- Record Drawings.pdf PDF
J03 - Self Performed Calculation Sheet (Construction).docx DOCX document
J01e - Dental Swing Space Needs.pdf PDF
J01h - Dental Clinic photos.pdf PDF
Show all 29

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

A-dec Regulatory Information, Specifications, and Warranty

Frequency (Hz)

Voltage Range

(VAC)

Maximum (Amps)

24 VDC Power Supply/LED light 50-60 100-240 1.25

24 VDC Power Supply (small)/cabinets 50-60 100-240 1.6

24 VDC Power Supply (large)/cabinets 50-60 100-240 2.5

24 VDC Power Supply (60W)/carts 50-60 100-240 1.6

25W Power Supply 50-60 100/110-120/220-240 0.3/0.3/0.15

80W Power Supply 50-60 100/110-120/220-240 0.9/0.8/0.4

300W Power Supply 50-60 100/110-120/220-240 3.1/2.8/1.4

Dental Furniture

Preference Collection® 60 120 20

Preference ICC® 50-60 100-120 15

ICV 50-60 110-120/220-240 0.5/0.5

A-dec Inspire Cabinet Models 591, 592, 593, 594, and 595

50-60 100-120 20

A-dec Inspire Cabinet Model 391 47-63 100-240 0.45

A-dec Inspire Distribution Box 50-60 100-240 10 Duplex output = 7 Amps max.

A-dec Inspire Power Box 50-60 100-240 10

Simulator 41L/41pro/42L/42pro 50-60 100/110-120/220-240 10/10/5 Duplex output = 7 Amps max.

Bitewing X-Ray Viewer 50-60 24 0.5

Monitor Mounts Performer 880X,381, 382, 482, 581, 584, 585, 586, and 587

50-60 100-240 10

NOTE

Electrical Rating

Frequency (Hz)

Voltage Range

(VAC)

(Amps)

A-dec 200 and Performer 8000

50-60 100/110-120/220-240 Input = 10/10/10 Duplex output = 10 Amps max.

Chair circuit board output = 2 Amps max.

Chair pump typical = 4/4/2

A-dec 311, A-dec 411, and A-dec 511 (chair rating includes optional chair-powered modules)

50-60 100/110-120/220-240 Input = 10/10/10 Duplex output = 10 Amps max.

511 Chair power supply = 4 Amps max.

Chair pump typical = 4/4/2

Systems with 300W Power Supply, including: A-dec 200, Performer 8100/8200/8500, 2671/2615, 2561/2562, 4631/4635, 3072, 7004, and A-dec 342.

50-60 100/110-120/220-240 Input = 3.1/2.8/1.4 Output with optional duplex on 2671/2615, 2561/2562

= 7 Amps max.

3420 Pac 1 Field and Institutional Units, N57D Bench Control, N74 M.O.M.

50-60 100-240 1.6

(Low Voltage

A-dec 200 Chair-Mount 50-60 12.1/17 5.5

(Mains Voltage)

A-dec 573L Post-Mount, A-dec 374L, 574L Cabinet- Mount, A-dec 375L, 575L Wall-Mount, A-dec 376L, 576L Ceiling-Mount, and A-dec 377L, 577L Track- Mount; A-dec 378L, 578L Universal Single

50-60 100-240 1.25

A-dec 578L Universal Dual 50-60 100-240 2.5

(Low Voltage)

A-dec 371L/372L/571L/572L Chair-Mount, A-dec 378L, 578L Stationary/In-Bench Simulator

50-60 16-24 (AC or DC) 1.5

J01k - Dental Equipment Electrical Requirements

AS10 AS12 AS21 AS22 AS30 AS40 AS50 AS70 AS100

Note 1: Install a buck or boost transformer if actual facility service is above or below the supply voltage fluctuation of ±10% of nominal voltage ratings listed, install a buck/boost transformer with the corresponding part number as shown below.

Nominal Supply Voltage (VAC, see note) 120 220 120 220 220 220 220 220 3 220

Maximum Current (Amps) 8 4 15 8 8 12 16 24 20

Minimum Panel Breaker Rating (Amps) 20 10 20 20 20 20 20 40 30

Minimum Wire Size (AWG) 12 16 12 12 12 12 12 8 10

Maximum Watts 864 792 1,620 1,584 1,584 2,376 3,168 4,752 3,960

BTU/hr @ 50% 1,473 1,350 2,201 2,701 2,701 4,051 5,401 8,201 6,752

BTU/hr @ 75% 2,210 2,026 4,143 4,051 4,051 6,077 8,102 12,153 N/A

Note 2: The 220 volt 3-phase, 4.6 KVa transformer is not available from Air Techniques.

Buck/Boost Transformer Air Techniques Part No. and Size

67006

2.5 KVa

67002

3.9 KVa

67006

2.5 KVa

67002

3.9 KVa

67002

3.9 KVa

67002

3.9 KVa

67002

3.9 KVa

67000-1

7.8 KVa

4.6 KVa

N/A

Site Requirements

All AirStar compressors comply with NFPA 99C level 3 requirements

Electrical Connections

AS50 and AS70 Electrical Connections

L1

L2

GND

WHITE

GREEN

BLACK

WHITE

BLACK

GREEN

HARD WIRED

CONNECTED

AIRSTAR MODELS

AS50 & AS70

AS100 3-Phase Electrical Connections

BLUE

L1

L2

GND

L3

GREEN

BLACK

RED

RED

BLACK

BLUE

GREEN

HARD WIRED

CONNECTED

AIRSTAR MODEL

AS100

24V Circuit Connections

AirStar Electrical Connection

Interconnect Cable

Yellow 2

Brown 4

Orange 3

See Note 3

Future Use

Remote Panel

Yellow 2 Brown 4 Orange 3

White

White

See Note 4

Connection without

24V Switch

4 BRN3 ORG

2 YEL

Notes:

1. Use 18 Gauge, 4 conductor, interconnect cable between the

AirStar unit and Remote Switch Panel.

When any 24V circuit exceeds 150 feet, use #16 AWG.

2. As shown, 3 conductors of the 24V circuit cable from each compressor connect via the user-supplied interconnect cable.

3. The fourth conductor of interconnect cable to be used for future equipment options/enhancements.

4. Leave with factory connection, without a 24V Switch, or connect the associated interconnect cable directly to remote switch.

Please note that one switch is used for each compressor.

Pressure Switch

From Electrical Box

3 ORN

YEL

BRN

BLU

BLK

BRN

YEL

BLK

BRN

BLU

ORN

To Remote Panel

YELORN

ORN

YEL

AirStar is a trademark of Air Techniques, Inc. 2009 Air Techniques, Inc. • PN 87104 Rev. P May 2021 www.air techniques.com

Corporate Headquarters

1295 Walt Whitman Road Melville, New York 11747- 3062

Phone: 800-247-8324 Fax: 888-247-8481

Product Specifications - Dimensions

AS10 AS12 AS21 AS22 AS30 AS40 AS50 AS70 AS100

Input Voltage Phase 1 1 1 1 1 1 1 1 3

System Power

HP

(kW)

0.75 (0.56)

0.75 (0.56)

1.5 (1.1)

1.5 (1.1)

1.5 (1.1)

2.25 (1.68)

3.0 (2.2)

4.5 (3.4)

4.8 (3.6)

Nominal Supply Voltage VAC 120 220 120 220 220 220 220 220 220

Frequency Hz 60 60 60 60 60 60 60 60 60

Maximum Current Amps 8 4 15 8 8 12 16 24 20

Maximum Simultaneous Air Users 2 2 3 3 4 6 8 12 15

System Output Flow Rate at

80 PSI CFM 2.5 2.5 5.0 5.0 5.0 7.5 10.0 15.0 20.0

Maximum Pump-up Time

0-115 PSI (M:SS) 2:55 2:55 3:10 3:10 3:10 1:40 2:50 2:40 2:30

Maximum Recovery Time

85-115 PSI (M:SS) 0:48 0:48 0:47 0:47 0:47 0:34 0:42 0:40 0:30

Tank Size US Gal.

(ft3)

(0.8)

(0.8)

(1.6)

(1.6)

(1.6)

(1.6)

(2.7)

(4.0)

(4.0)

Shipping Weight lbs.

(Approximate)

No Sound Cover With Sound Cover

N/A

N/A

Dimensions in.

(See note.)

No Sound Cover

H W D

28.50 25.00 19.75

28.50 25.00 19.75

30.50 29.00 20.00

30.50 29.00 20.00

30.50 29.00 20.00

30.50 32.50 20.00

33.50 35.50 20.50

35.00 47.75 21.75

34.50 46.00 24.50

With Sound Cover H W D

30.00 25.00 22.50

30.00 25.00 22.50

32.00 31.00 22.25

32.00 31.00 22.25

32.00 31.00 22.25

32.25 33.25 22.50

33.50 36.50 22.75

36.00 51.00 29.50

N/A N/A N/A

Description Model

Note: Height measured without leveling feet for all units with or without sound cover.

Speci cations & Site Requirements

Model QT 2 115V QT 2 230V

Maximum Number of Users 4 4

Phase 1 1

Horsepower (kW) 1.0 (0.7) 1.0 (0.7)

Voltage Rating (Minimum/Maximum)

104/127

208/253

Frequency (Hz) 60 60

Breaker Rating Size (Recommended Amps) 20 15

Full Load Amps 9.0 4.5

Fusetron Size (Recommended) FRN-25 FRN-8

THHN Wire Size (AWG) 12 12

BTU/HR (Heat Released to Room) 1,082 1,082

Sound Levels (dBA) 71 71

Operating Temperature (Min/Max) 32°F/104°F

Operating Relative Humidity Range 0-95%, No Condensing Moisture

Shipping/Transport Temperature (Min/Max) -20°F/165°F

Operating Atmospheric Pressure Range 63-105 kPa

DentalEZ 2500 Highway 31 South Bay Minette, AL 36507

Phone: 866-DTE-INFO Fax: 251-937-0461 www.dentalez.com

PN: 7565-049A

November, 2019

Bulldog® QT Dry VacuumPreinstallation Guide

Product Dimensions* Bulldog QT 15-Gallon Otter Tank Bulldog & 15-Gallon Otter Tank Stacked

Nominal Width 21" (533 mm) 25" (635 mm) 25" (635 mm)

Nominal Depth 17" (432 mm) 19" (483 mm) 19" (483 mm)

Nominal Height 28" (711 mm) 24" (610 mm) 44" (1,118 mm)

Nominal Height with S-Type Exhaust Filter

48" (1,219 mm) N/A 67" (1,702 mm)

Drain Valve Centerline to Floor N/A 8" (203 mm) 8" (203 mm)

*Add 6" (155 mm) to each dimension for working space.

Conforms to:

ANSI/AAMI STD ES60601

Certi ed to:

CAN/CSA STD C22.2 NO. 60601-1

ETL CLASSIFIED

4007136

7 Technical data

51774-IMG rev 13 Instrumentarium Dental

7 Technical data

7.1 Technical specifications

Manufacturer: Instrumentarium Dental, PaloDEx Group Oy, Nahkelantie 160 FI-04300 Tuusula, FINLAND

Manufacturer’s Quality System: In accordance with ISO13485 and ISO9001 standard

Environmental Management System:

In accordance with ISO 14001 standard

Conformity to standards: IEC 60601-1

IEC 60601-1-4

IEC 60601-1-3

IEC 60601-2-7

IEC 60601-2-28

IEC60601-2-65

CAN/CSA –C22.2 No. 601-1-M90 CE models marked according to the Medical Device Directive 93/

42/ EEC

Product Name FOCUS Type Intraoral X-ray unit

UNIT DATA

Protection against electric shock

Class I

Degree of protection Type B Protection against the ingress of liquids

IP20

Mode of operation Continuous operation with intermittent loading

Power supply Mains connection, plug or fixed Software version 3.0 or higher

X-RAY GENERATOR

Generator type Constant potential Nominal power 490 W High voltage DC Supply frequency 100-200 kHz Number of phases 1 Reference current time product 7 mAs (70 kV, 7 mA, 1 s) Lowest current time product 0.14 mAs (70 kV, 7 mA, 0.02 s) Coefficient of variation of DAP < 0,05

32 Instrumentarium Dental

DENTAL CARE UNIT MOUNT MODEL FOCUS

Horizontal arm length 330 mm (from axl to axl), 418 mm (total length) Generator module length 500 mm Horizontal arm shaft diameter 32 mm Main cable length 5230 mm

(measured from generator module to the scissors arm)

TUBE HEAD ASSEMBLY

Tube head assembly type FOCUS Tube type Toshiba D-0711SB or Kailong KL

21 SB or equivalent Max. tube voltage 60 or 70 kV Max. tube current 7 mA Max. electric output 490 W (70kV; 7mA) Reference axis Runs axially with the cone Target angle 16 degrees Focal spot 0.7 mm

(according to IEC 60336/2005) Nominal anode input power 940 W Max. symmetrical radiation field

60 mm at a 200 mm focal length

Total filtration 2,0 mm Al (70 kV) Inherent filtration 1,5 mm Al (70 kV) Fixed additional filtration 1,0 mm Al (70 kV) Max. anode heat content 7 kJ Maximum X-ray tube assembly heat content

140 kJ

Maximum continuous heat dissipation of the X-ray tube head assembly

19 W

ELECTRICAL CONNECTIONS

Nominal mains voltage 115 VAC +/- 10%

230 VAC +/- 10%

Input power frequency 60 Hz

50 Hz Nominal current 7.63A

3.55A Mains fuse, slow blow 6.25AT Apparent resistance of supply mains

0.68

Power consumption 816 VA / 230 VAC

877 VA / 115 VAC

US/Canada mains connector type

115 V / NEMA 6-15P or similar

Power supply cords type H05VV5-F / AWG 14 (UL 2587)

51774-IMG rev 13 Instrumentarium Dental

Nominal Shortest Irradiation Time 0.02 S Exposure Time Range 0.02 - 3.2 S Exposure Time Range In Aec Mode 0.02- 1.6 S

BEAM LIMITING DEVICE

Cone dimensions Round: 60 mm

Rectangular: 35 x 45 mm

PHYSICAL MEASURES AND AMBIENT TEMPERATURES:

Focal length (Standard/Long) 229 mm (9 inches) /

305 mm (12 inches) Installation Standard wall mount Optional base for free standing unit Height x Width x Depth (mm) Unit: 1059 mm x 279 mm x 946/

1096/1346 mm Tube head assembly: 112 mm x 260 mm x 201 mm

Weight Unit: approximately 30 kg (66 pounds) Tube head assembly:

approximately 4.5 kg (10 pounds)

Type and length of the cable of the remote exposure switch

Length approximately 10 m (32.5 feet), RJ-45 plug at both ends (8 wires)

Transportation and Storage -40°...+70°C (-40F...+158F), RH 10...100%

Operating environment +10°...+40°C (+50F...+104F), RH max. 70%, 700 – 1060 mbar