This pre-solicitation announcement describes an upcoming invitation for bid for construction project number 506-CSI-113 at the Ann Arbor VA Medical Center. The project involves reconfiguring operating room number six for the installation of new high-cost medical equipment, including site preparation work for floors, walls, ceilings, and utilities. Interested service-disabled veteran-owned small businesses will be able to download the solicitation and submit bids through the Contract Opportunities website between March 9th and the due date specified in the solicitation. The contract will be a firm-fixed-price award between $1 million and $2 million, with a 180-day performance period. Bidders must register in SAM, VetBiz, and submit current VETS-4212 reports (if applicable) prior to bid submission.
View the file
Other files
Show all 25
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Pre-Installation Manual
SkyVision Linx 300 Integration Control System
TEC-R-0049 REV4
06/01/2020
CAST Manufactured by:
US - SKYTRON
5085 Corporate Exchange Blvd. S.E.
Grand Rapids, MI 49512
(616) 656-2900 www.skytron.us
Distributed by:
US - SKYTRON
5085 Corporate Exchange Blvd. S.E.
Grand Rapids, MI 49512
(616) 656-2900 www.skytron.us
The base language for this document is ENGLISH. Any translations must be from the base language document.
Printed copies are not controlled documents.
Although current at the time of publication, SKYTRON'S policy of continuous development makes this manual subject to change without notice. If current manuals are required, contact your local SKYTRON representative or contact SKYTRON directly at the distribution addresses listed above.
SkyVision Linx 300 Pre-Install• REV4
TABLE OF CONTENTS
TITLE PAGE
SECTION 1. INTRODUCTION AND SAFETY
1-1. Safety Precautions
SECTION 2. REQUIREMENTS
2-1. Environmental 2-2. Power 2-3. Grounding 2-4. Redundant Connection and Emergency Operation 2-5. Rack Mounting 2-6. Display Power Management 2-7. Optional Equipment 2-8. Remote Rack Locations 2-9. Connections 2-10. Network
SECTION 3. INSTALLATION RESPONSIBILITIES
3-1. SKYTRON Responsibilities 3-2. Customer Responsibilities
SECTION 4. CAST NETWORK RESOURCE PLAN
4-1. Central Storage System (CSS) 4-2. Small Scale CSS - Recommended Servers 4-3. Large Scale CSS - Recommended Servers 4-4. Server Virtualization 4-5. Storage Resource Allocation 4-6. Network Resource Allocation 4-7. CAST Network Connection 4-8. Network Address Range 4-9. Operating in Unicast 4-10. Mobile Server Module – MSM 4-11. MSM in a Virtual DMZ 4-12. Full MedicVOD Connection Scheme
SECTION 5. CAST STORAGE POLICY PLAN
5-1. Real-Time Compression 5-2. Offline Compression 5-3. Storage space 5-4. Archival Policies to Define 5-5. Storage Scenarios 5-6. Backup Policy
SECTION 6. REVISION HISTORY
SkyVision Linx 300 Pre-Install • REV4
SECTION 1. INTRODUCTION AND SAFETY
The SKYTRON SkyVision Linx 300 Integration Control System installation can vary according to the specific equipment needs of the OR facility. What is common in all installations is that video and data input sources are routed to various outputs such as displays, video recorders, or networks providing communication and video conferences.
This SkyVision Pre-Installation Manual provides an overview of typical installation considerations and
| specific | installation | requirements | associated | with | a | SkyVision | system. | |||||||||
| Only | SKYTRON | certified | technicians | are | authorized | to | install | a | system. | The | SkyVision | installation | team | |||
| may | consist | of | two | (2) | or | more | technicians | who | coordinate | with | the | hospital | staff | and | local | subcontractors |
| to | complete | an | installation. | Technicians | follow | specific | installation | details | mandated | by | the | customer | and | |||
| specified | in | the | SkyVision | OPI | (Order | / | Production | / | Installation) | document. | SKYTRON | supports | SkyVision |
installations with thorough documentation, classroom training, and technical support.
1-1. Safety Precautions The following is a summary CAUTIONS denoted in this manual. These precautions are found throughout the manual where they are applicable. Carefully read the manual before proceeding to operate or service the equipment.
CAUTION
CAUTION without the safety alert symbol, is used to address practices not related to personal injury but with a possibility of damage to equipment.
With Multicast on Parallel, it is very important to properly configure the network segment (VLAN) using multicast to prevent video packets to inadvertently “breach” the “isolated” multicast setup and be transferred over the unicast.
If this occurs, that packet will reach the server and may adversely affect the quality, size and integrity of the video that is being recorded at the same time.
SECTION 2. REQUIREMENTS
2-1. Environmental The SkyVision system must be stored and operated in a clean, dust free environment. Ambient room temperatures should be maintained between 60˚ and 75˚ Fahrenheit with non-condensing humidity between 20% to 50%.
The SkyVision Linx 300 system components will generate up to 2900 BTU's/hr. when fully equipped and under maximum load during peak operating times. Mechanical and HVAC designs of the SKYTRON Nurse Documentation Center (NDC), SORIC system, and Third Party provided versions must address heating, ventilation and exhaust requirements in a manner consistent with the operational requirements of the SkyVision system.
When installing the system within custom cabinetry or a NDC, a minimum of 50 cfm through a 12" square opening is required for constant return HVAC. This will allow the systems to operate within the equipment temperature specifications and ensure product performance, reliability, and longevity.
2-2. Power The SKYTRON Linx 300 System requires two 120VAC/20A, GFCI protected outlets. One circuit is dedicated to the component rack and the second circuit is dedicated to the monitor power supplies.
All receptacle outlet strips, adapters, receptacle covers, and faceplates shall be marked with the following words or equivalent:
"WARNING - TECHNICAL POWER
DO NOT connect to lighting equipment For electronic equipment use only
60/120VAC
GFCI protected" The Uninterruptible Power Supply (UPS) that is provided for power conditioning on the Linx 300 system is not designed for use with Isolated Power Supply systems.
2-3. Grounding SKYTRON requires a minimum 10 AWG copper ground wire to be provided to the 12" x 12" junction box for connection to the equipment rack.
Refer to Article 647.6(B) (Grounding Conductors Required for Sensitive Electronic Equipment) of the NEC.
Permanently wired utilization equipment and receptacles shall be grounded by means of an equipment grounding conductor run with circuit conductors to an equipment grounding bus prominently marked "Technical Equipment Ground" in the originally branch circuit panel board.
2-4. Redundant Connection and Emergency Operation The SKYTRON Linx system MUST BE configured with a redundant back up video. This video delivers a backup video signal to a dedicated display when the system is unable to deliver the primary care video to the display interface. The default primary care video source is dedicated to the endoscopic camera. The customer collaborates with SKYTRON during the installation to identify the source video to connect as the primary and back up video sources.
If the primary video source loses the signal to the display interface(s), and the manual rerouting or use of the preset fails to restore the video image, switch the system to back up video mode. Use the following steps to switch the system to the back up video mode:
1. Manually switch the video mode on the monitor or display 1 to the analog input source.
2. Select the designated back up video input source on the monitor or display 1.
3. If a secondary back up display monitor was designated, manually switch the video source mode on monitor or display 2.
These steps deliver the primary care video source to display on monitor 1 and secondary video on monitor 2.
Contact SKYTRON for assistance to resolve this issue.
2-5. Rack Mounting
| The | SKYTRON | SkyVision | System | is | offered | in | a | single | 18U | wall | mounted | rack, | that | does | NOT | stand | on |
| the | floor, | or | as | part | of | a | NDC | in | two | (2) | 12U | racks. |
NOTICE
| For | Seismic | mounting | specifications, | refer | to | the | Middle | Atlantic | Products | |
| letter | for | "Seismic | Certification | of | the | DWR | Series | Wall | Mounted | Electronics |
| Rack | Enclosures" | available | on | SkyNet. |
A. Rack - 18U Wall Mount
Overall Width Overall Height Overall Depth Weight
Capacity
Minimum Structural
Requirements
Mounting Hole Spacing
23-3/8” [594mm]
38-1/2” [978mm]
26-5/16” [668mm] 250 lbs 500 lbs 16”
[406mm]
Minimum Clearance - 6” from non-hinged side, 20” from hinged side, 4” on top and bottom.
NOTICE
| It | is | the | responsibility | of | the | contractor | to | provide | sufficient | backing/studs | in | ||
| the | wall | to | support | the | rack | (the | rack | does | NOT | stand | on | the | floor). |
An alternative to mounting directly to the studs is to mount plywood to the entire wall and use toggle bolts to mount the rack to the plywood.
Whatever method is used to mount the rack, the contractor is responsible for making sure it will hold the minimum structural required load of 500 pounds.
6" MIN.
CLEARANCE
W A
LL
W A
LL
WALL
26 5/16”
6-1/2"
20-1/2"
20" MIN.
CLEARANCE
90°
90°
(4) MOUNTING
HOLE LOCATIONS
AB x B
C D
D
E
E
F
F
G
G
J
H
H
B
K
L
I
Letter Dimension A 38-1/2" B 12-1/2" C 22-3/4" D 1-3/8" E 16" F 5-3/8" G 2-11/16" H 13" I 4" J 33-1/4" K 5-7/8" L 4-3/8"
B
CONDUIT
BOXES
RACK
Figure 1. Rack Mounting and Junction Box Location (Left-Hand Swing Cabinet)
W A
LL
6" MIN.
CLEARANCE
20" MIN.
CLEARANCE
6-1/2"
W A
LL
WALL
20-1/2"
90°
90°
26-5/16"
(4) MOUNTING
HOLE LOCATIONS
AB x B
C
D E
G
G
H
H
B
I
J
K
L
D E
F
Letter Dimension A 38-1/2" B 12-1/2" C 22-3/4" D 1-3/8" E 16" F 5-3/8" G 2-11/16" H 13" I 4" J 33-1/4" K 5-7/8" L 4-3/8"
RACK
CONDUIT
BOXES
B
F
Figure 2. Rack Mounting and Junction Box Location (Right-hand Swing Cabinet)
12" x 12" x 4"
JUNCTION BOX
(PROVIDED BY
CUSTOMER) TO BE
ROUGHED IN BY THE
CONTRACTOR.
MOUNT THE POWER
OUTLETS IN BOX AS
SHOWN.
10 AWG GREEN
STRANDED WIRE
Figure 3. Left-Hand Swing Installation with Junction Box Termination
B. Nurse Documentation Center (NDC) Additional NDC installation information can be found for the SKYTRON Model SORIC 48 and SORIC 72 in the SKYTRON SORIC Installation Manual.
NOTICE
NDC systems require a minimum 50 cfm HVAC ventilation system.
MOUNT POWER
OUTLETS AS
SHOWN
REAR CABINET TO
CABINET CONNECTIONS
REQUIRE 15 FOOT LONG
CABLES.
CENTER OF 12” X 12” X 4”
JUNCTION BOX IS
LOCATED 12” IN FROM
THE SIDE AND 15” UP
FROM THE FLOOR
IF THE JUNCTION BOX MUST
BE LOCATEDD ON THIS SIDE,
THE CONTENTS OF EACH
RACK MUST BE SWITCHED.
THE CONDUITS MUST EXIT AT
THE DIGITAL MATRIX SWITCH.
50 CFM HVAC VENTILATION
AIR SYSTEM (DUCTS)
Figure 4. NDC Junction Box Location
2-6. Display Power Management
• The SKYTRON Panasonic and Foreseeson's (FSN) displays can be powered in one of two ways:
• Mount the low voltage power supply onto the display bracket powered by a low leakage 120VAC power line (P/N H5-010-71) run through the boom arm. Mounting the power supply in the back of the display provides open service access to the power supply. Mounting the power supply at this location may not be desirable by the customer for aesthetic reasons, therefore, customer approval is required for this mounting configuration.
• Place the low voltage power supply in a hospital provided NEMA 4 enclosure located in the ceiling or at the equipment rack and run the low voltage power line through the boom arm to the display.
• The Optik View 24" displays come with a power supply and includes 75 feet of cable. The power supply should be located remotely, such as a hospital provided NEMA 4 enclosure (supplied by others) located in the ceiling.
• The remote location of the FSN 24" and 26" display power supply is limited to 75 feet (cable length)
(PN MON-CBL-001).
• The remote location of the Panasonic 26" display power supply is limited to 90 feet (cable length)
(PN DJ-NC0275)
• For integration installations where the rack is within a 75 feet (cable length) of the display, the power supply can be located in the rack. If not, a NEMA 4 encased power supply in the ceiling is recommended (supplied by others).
• The minimum size NEMA 4 enclosure for each power supply is 14”L x 12”W x 7”H (L-Com Global Connectivity – Model: NB1207-10FS-UL or equivalent).
• Appropriate power supplies and cables required for displays not provided by SKYTRON must be
| provided | by | the | equipment | vendor | to | meet | the | specific | needs | of | the | display | location. | ||
| • Not | all | displays | use | low | voltage | power | supplies. | Larger | displays | (typically | 32" | and | above) | normally |
use direct 120VAC input from an outlet as they are intended to be wall mounted. Remote cable kits are not intended for use with these displays as they do not use external low voltage power.
2-7. Optional Equipment A. Image/Video Capture Devices The SkyVision system offers a Capture, Archive, and Streaming system ( SVIS-CAST) for digital capture and streaming, however, some hospitals will maintain existing capture devices and cameras.
When the capture device is located close to the camera, connections can be made by a short copper cable (provided by the Minimally Invasive Surgery (MIS) vendor). Typically, the capture device is located at the nurse’s station far from the camera. The appropriate cable between the camera and the capture device must be provided by the equipment vendor. The SkyVision system will process the capture device output signal, rather than the camera output signal (to maximize functionality), and route the output signal, as limited by the technology of the capture device unless specified otherwise on the OPI document.
B. Optional Remote Trigger Image/Video Capture Capture devices specific to endoscopic cameras allow the surgeon to capture images or video using the trigger on the camera. This method is limited to capturing only the camera source as the surgeon would need to have the camera in hand when capturing. If capturing endoscope images with the endoscope trigger is required, this must be specified by the sales representative as a "special" on the OPI documentation.
The hospital must supply SKYTRON with the make and model number of the camera and capture device so SKYTRON can verify compatibility, proper operation, and provide the required items to quote.
C. Optional Keyboard/Mouse Switch The KM Switch (SVIS-KM) allows a single keyboard and mouse to control up to four (4) PCs or one CAST and three (3) PCs.
When a CAST is purchased in addition to the on-board PC, the KM Switch must be quoted. If the hospital uses their PCs instead of purchasing the SkyVision PCs, they may use their keyboard and mouse instead of the SkyVision keyboard and mouse. The hospital computer can be placed anywhere in the room with their keyboard, mouse, and monitor. If they would like to use the SkyVision keyboard and mouse, a KM Switch is required and the hospital PC must be located within 10 feet of the rack.
D. Additional Touch Panels Optional second and/or third touch panels (SVIS-SECTP) may be selected with a SkyVision system to enable the system to be operated from more than one nurse documentation station location. In this configuration, the second and/or third touch panels are routed through the primary touch panel through VNC connections.
2-8. Remote Rack Locations The stock cable length for a SKYTRON SkyVision System installation is 100 feet. 200 and 300 feet cable lengths may be selected from the OPI document if longer cable lengths are required.
For installations requiring greater distances than 300 feet, consult with the SkyVision Product Manager so the options can be specified and project details/requirements resolved.
2-9. Connections The SkyVision system connects multiple types of video sources (VGA, DVI, SDI, HD-SDI or S-Video/ Composite) to the control rack through conduit runways over single strand multi-mode optical fibers via source conversion faceplates (Figure 5) located near the source. The conversion faceplates are installed in the boom arm equipment carriers or in double-gang electrical wall boxes (Figure 6) depending on the location of the source. The conversion faceplates convert a copper wire signal (digital or analog) to a fiber-optic digital signal. Low voltage power lines are the only other type of connection utilized to and from the SkyVision system. Copper wiring is only employed between components at the rear of the SkyVision control rack. See Figure 7 and for a typical system build drawing set (plan and elevation views) that shows a layout of the affected room(s), equipment, and cable routing.
Figure 5. Conversion Faceplates
4” x 4” x 3-1/2” DEEP DOUBLE-GANG
ELECTRICAL WALL BOX
RACO MODEL 256 OR EQUIVALENT
GARVIN 52181-S RAC0 256
SINGLE-GANG MUD
RING
Figure 6. Electrical Boxes
4'- 6"
2'- 6"
7'
6'- 6"
CA
UT
IO
N
TH
IS
BO
OM
HA
S
RO
T A
TI
ON
AL
CO
NF
LIC
TS
B C
9A 6A
5A 2A TO
NE
TW
OR
K
DESCRIPTION:ROOMLAYOUT
a
SH
EE
T
IN
IT
IA
L:
DA
TE
Sk yV is io n No te s:
Sk yV isi on ra ck re qu ire sa
"s q.
x4 "d ee pw all bo xp er
Sk yV isi on sp ec
.s up pli ed by co ntr ac tor
Ne tw or kc on ne cti on si ns tal led to ra ck by oth er s.
St ru ctu ra ls up po rt for
Sk yV isi on ra ck de sig ne db ye ng ine er of re co rd st os up po rt
0l bs
.o flo ad
Mo un tin gh ar dw ar es up pli ed by oth er st om ee ts up po rt re qu ire me nts
FI
NA
L
PL
AC
EM
EN
T O F
RE
Q
UI
RE
D
CO
ND
UI
T
IS
TH
E
RE
SP
O
NS
IB
IL
IT
Y O F
AN
D
DE
TE
RM
IN
ED
BY
TH
E
AR
CH
IT
EC
T.
ACMEHOSPITALPROJECT#:XX-123-XX
FABRICATION
PLOTDATE:
ROOMTYPE:GENOR
REV.#: 1
EQ
UI
PM
EN
T
SC
HE
DU
LE
ID
DE
SC
RI
PT
IO
N
A
EQ
UI
PM
EN
T
BO
OM
W
IT
H
LIG
HT
S
B
UT
ILI
TY
/M
ON
IT
OR
BO
OM
C
AN
ES
TH
ES
IA
BO
OM
IN
TH
E
EV
EN
T
TH
AT
FI
XT
UR
E
LO
CA
TI
O
NS
NE
ED
TO
CH
AN
G E
PL
EA
SE
CH
EC
K W
IT
H
SK
YT
RO
N O N
TH
E
RE
VI
SE
D
LA
YO
UT
LO
CA
TI
O N
*S
IT
E
SP
EC
IF
IC
LA
YO
UT
*I F
SK
YV
IS
IO
N
IS
AP
PL
IC
AB
LE
RE
FE
R
TO
TH
E
LA
ST
PO
RT
IO
N
OF
BO
OM
PA
CK
AG
E*
Figure 7. Typical System Build Drawing (Plan View)
FL
OO
R
IN
IT
IA
L:
DA
TE
FI
XT
UR
E
W
EI
GH
T:
MO
ME
NT
LO
AD
7L
BS
0F
T.
LB
S.
"A
BO
VE
FI
NI
SH
ED
FL
OO
R
"B
EL
OW
FI
NI
SH
ED
CE
ILI
NG
AU
DI
O
PA
CK
AG
E
FI
NI
SH
ED
CE
ILI
NG
MI
CR
OP
HO
NE
FI
NI
SH
ED
CE
ILI
NG
TY
PI
CA
L1
6"
AB
OV
E
FI
NI
SH
ED
FL
OO
R
"B
EL
OW
FI
NI
SH
ED
CE
ILI
NG
G
EN
ER
IC
M O
UN
T
SH
O W N
CE
ILI
NG
MO
UN
TE
D
EQ
UI
PM
EN
T (2
"C
ON
DU
IT
S)
TE
RM
IN
AT
ED
W
IT
H
IN
SU
LA
TE
D
BU
SH
IN
GS
AN
D
PU
LL
ST
RI
NG
S
W
IT
HI
N
"O
F
TH
E
CE
NT
ER
OF
TH
E
CE
ILI
NG
MO
UN
T.
TH
ES
E
CO
ND
UI
TS
MU
ST
BE
AC
CE
SS
IB
LE
FR
OM
TH
E
EQ
UI
PM
EN
T
BO
OM
AC
CE
SS
PA
NE
L.
RE
FE
R
TO
IN
DI
VI
DU
AL
EQ
UI
PM
EN
T
BO
OM
FO
R
MO
RE
IN
FO
RM
AT
IO
N.
NO
TE
:A
LL
JU
NC
TI
ON
BO
XE
S
AN
D
CO
ND
UI
T
TO
BE
SU
PP
LIE
D
BY
CO
NT
RA
CT
OR
TY
PI
CA
L5
5"
AB
OV
E
FI
NI
SH
ED
FL
OO
R
OR
PE
R
HO
SP
IT
AL
ACMEHOSPITAL
EL
EV
AT
IO
N S
SH
O
W N
BE
LO
W
AR
E
GE
N
ER
IC
AN
D
PR
O
VI
DE
TY
PI
CA
L
EL
EV
AT
IO
N
S
FO
R
EA
CH
IT
EM
HY
BR
ID
II
SK
YV
IS
IO
N
CO
ND
UI
T
LE
G
EN
D (N O
TE
-I
TE
M S
LI
ST
ED
M
AY
NO
T
AL
L
BE
PA
RT
O F
TH
IS
PR
O
JE
CT
RE
FE
R
TO
RO
O M
LA
YO
UT
S
FO
R
SI
TE
SP
EC
IF
IC
IN
FO
RM
AT
IO
N.
ID
IT
EM
CO
ND
UI
T
DI
AM
ET
ER
DE
SC
RI
PT
IO
N
TE
RM
IN
AT
IO
N
LO
CA
TI
O N O F
JU
NC
TI
O N
BO
X
TO
SK
YV
IS
IO
N
EN
CL
OS
UR
E N
/A
SK
YV
IS
IO
N
RO
UG
H-
IN
"S
Q.
X
4"
DE
EP
JU
NC
TI
ON
BO
X
SE
E
RO
UG
H-
IN
DE
TA
IL
N /A
2"
NE
TW
OR
K
CO
NN
EC
TI
ON
S
RO
UT
E
T O
NE
TW
OR
K
PE
R
HO
SP
IT
AL
HO
SP
IT
AL
NE
TW
OR
K
TO
UC
H
PA
NE
L
1"
TO
UC
H
PA
NE
L, KE
YB
OA
RD
,M
OU
SE
4"
SQ
.X 3.5
"D
EE
PB
OX
W /1
GA
NG
P-
RI
NG
P E
R
HO
SP
IT
AL
RA
CK
W
AL
LS
OU
RC
E 1"
SO
UR
CE
DE
SC
RI
PT
IO
N 4"
SQ
.X
3.5 "D
EE
P
BO
X
W /1
GA
NG
P-
RI
NG
PE
R
HO
SP
IT
AL
RA
CK
W
AL
LC
AM
ER
A 1.2
5" W
AL
LC
AM
ER
A 4"
SQ
.X
3.5 "D
EE
P
BO
X
W /1
GA
NG
P-
RI
NG
"B
EL
OW
FI
NI
SH
ED
CE
ILI
NG
RA
CK
SH
AR
ED
W
AL
LC
AM
ER
A
CO
NT
RO
L 4"
SQ
.X
2"
DE
EP
BO
X W
/1
GA
NG
P-
RI
NG
"B
EL
OW
FI
NI
SH
ED
CE
ILI
NG
CO
NN
EC
TE
D
TO
3.5
"B
OX
7*
IP
CA
ME
RA
1"
IP
CA
ME
RA
4"
SQ
.X
2"
DE
EP
BO
X W
/1
GA
NG
P-
RI
NG
"B
EL
OW
FI
NI
SH
ED
CE
ILI
NG
NE
TW
OR
K
CE
ILI
NG
MO
UN
TE
D
EQ
UI
PM
EN
T
2"
DE
SC
RI
PT
IO
N
OF
MO
UN
T
PE
R
EQ
UI
PM
EN
T
MA
NU
FA
CT
UR
ER
MO
UN
T
ID
RA
CK
W
AL
LD
IS
PL
AY
1"
W
AL
LD
IS
PL
AY
4"
SQ
.X 3.5
"D
EE
P
BO
X W
/1
GA
NG
P-
RI
NG
"A
BO
VE
FI
NI
SH
ED
FL
OO
R
RA
CK
AU
DI
O
PA
CK
AG
E 3/8
"F
LE
X
CE
ILI
NG
SP
EA
KE
RS
TE
RM
IN
AT
ES
TO
SP
EA
KE
RS
PE
R
IN
ST
RU
CT
IO
NS
FI
NI
SH
ED
CE
ILI
NG
PE
R
HO
S P
IT
AL
RA
CK
MI
CR
OP
HO
NE
3/8
"F
LE
X
MI
CR
OP
HO
NE
ST
UB
OU
T
FI
NI
SH
ED
CE
ILI
NG
PE
R
HO
SP
IT
AL
RA
CK
PE
R
HO
SP
IT
AL
W
AL
LC
ON
TR
OL
0.7
5" W
AL
LC
ON
TR
OL
-L
IG
HT
IN
TE
NS
IT
Y
SI
NG
LE
/D
UA
L-
"H x8
"W x5
.87 5"
D
PE
R
HO
SP
IT
AL
LIG
HT
MO
UN
T
TR
IP
LE
-1
3.5 "H x1 0"
W X
6.3
"D
LIG
HT
MO
UN
T
ID
IT
EM
CO
ND
UI
T
DI
AM
ET
ER
DE
SC
RI
PT
IO
N
TE
RM
IN
AT
IO
N
LO
CA
TI
O N O F
JU
NC
TI
O N
BO
X A
DE
VI
CE
PO
W
ER
N /A
0V
,2 0A
DU
PL
EX
PE
R
EN
GI
NE
ER
PE
R
HO
SP
IT
AL
EL
EC
TR
IC
AL
B
SK
YV
IS
IO
N
PO
W
ER
N
/A (2
)1
V, A
DE
DI
CA
TE
D
CI
RC
UI
TS
PE
R
EN
GI
NE
ER
SK
YV
IS
IO
N
JU
NC
TI
ON
BO
X
EL
EC
TR
IC
AL
C
GR
OU
ND
LE
AD
N /A
PE
R
EN
GI
NE
ER
FA
CI
LIT
Y
GR
OU
ND
RA
CK
D
P.
O.
E.
N
/A
PO
W
ER
OV
ER
ET
HE
RN
ET
SH
AR
ED
PE
R
IT
EM
"B
EL
OW
FI
NI
SH
ED
CE
ILI
NG
HO
SP
IT
AL
NE
TW
OR
K
-I
F P.
O.
E.
IS
NO
T
AV
AI
LA
BL
E, US
E
EL
EC
TR
IC
AL
RE
QU
IR
EM
EN
TS
ID
A
HV
AC
Re qu ire dt om ain tai n
St or ag ec ab ine tte mp er atu re no tto ex ce ed de gr ee sF ah re nh eit .A irfl ow to be ad jus ted ac co rd ing ly
4"
AB
OV
E
FI
NI
SH
ED
FL
OO
R
3/1 6"
AB
OV
E
FI
NI
SH
ED
FL
OO
R
SK
YT
RO
N
LIG
HT
CA
ME
RA
CO
NT
RO
L
1"
LIG
HT
CA
M
VI
DE
O
TO
RA
CK
4"
SQ
.X
3.5 "D
EE
P
BO
X
W /1
GA
NG
P-
RI
NG
PE
R
HO
SP
IT
AL
RA
CK
1"
LIG
HT
CA
M
VI
DE
O
FR
OM
CA
ME
RA
4"
SQ
.X 2"
DE
EP
BO
X
W /1
GA
NG
P -
RI
NG
PE
R
HO
SP
IT
AL
LIG
HT
CA
M
MO
UN
T
EX
TE
RN
AL
CO
MP
UT
ER
1"
PC
VI
DE
O
4"
SQ
.X 3.5
"D
EE
P
BO
X W
/1
GA
NG
P-
RI
NG
PE
R
HO
SP
IT
AL
RA
CK
HP
NE
TW
OR
K
PR
IN
TE
R
1"
HP
NE
TW
OR
K
PR
IN
TE
R
4"
SQ
.X 2"
DE
EP
BO
X
W /1
GA
NG
P-
RI
NG
PE
R
HO
SP
IT
AL
HO
SP
IT
AL
NE
TW
OR
K
TO
NE
TW
OR
K
TO
SU
RG
IC
AL
LIG
HT
MO
UN
TS
TO
LIG
HT
CA
ME
RA
MO
UN
T
1"
HO
SP
IT
AL
NE
TW
OR
K
CO
NN
EC
TI
ON
HO
SP
IT
AL
SU
PP
LIE
D
RJ
NE
TW
OR
K
JA
CK
PE
R
HO
SP
IT
A L
HO
SP
I T
AL
NE
TW
OR
K
TO
Figure 8. Typical System Build Drawing (Elevation View)
2-10. Network A. Network Drops and Required Ports Up to eight (8) Static IP network drops are required for the following components (if installed), to enable full system function, configuration, and diagnostics.
Required or w/Option
Device Lan Drop Location
TCP Ports Required
UDP Ports Required
Notes
Required Primary Touch Panel
Touch Panel 22
SVIS-AUD Audio DSP (Nexia)
Rack 12001 12000
SVIS-CAST CAST Rack 80 1234 (If Multicast is not available use 5000-5500)
224.10.10.0 through 224.255.255.255
SVIS-OBC PC1 Rack unless hospital provided and not at rack
Standard
SVIS-OBC PC2 Rack unless hospital provided and not at rack
Standard
SVIS-OBC PC3 Rack unless hospital provided and not at rack
Standard
B. Analog Phone Line Drop One analog phone line is required for the audio conference option
Required or w/Option Device Drop Location SVIS-AUD Audio DSP (Nexia) Rack
C. VPN and Remote Diagnostics The following ports must be open for remote VPN and diagnostics:
• 5900 - Remote VNC
• 3000 - Remote Diagnostics
SECTION 3. INSTALLATION RESPONSIBILITIES
3-1. SKYTRON Responsibilities
1. SKYTRON requires a minimum of 60 days delivery time from receipt of signed drawings and OPI document for manufacturing of the proposed SkyVision system.
2. SKYTRON will provide customer with the appropriate signal cable and wiring for the system described in this proposal.
3. SKYTRON will provide customer with a suggested conduit layout drawing for the system described in the proposal.
NOTICE
An inspection and evaluation of the installation site must be performed and the SkyVision Pre-Installation Evaluation form must be completed prior to the arrival of installation technicians (see attached sample form).
4. SKYTRON will provide the appropriate number of specialized technical staff to complete the cable terminations, equipment rack assembly, and testing of the completed SkyVision system.
5. SKYTRON will provide the redundant video cable as part of the quote process and identified video format from the customer and video scope vendor.
6. Installation per unit includes two (2) full days of clear access to each room for cable pulling and terminations, plus at least two (2) days for fitting of connectors, final system configuring, system power up, and complete system testing.
7. Installation should take place on consecutive days and must be coordinated by a customer representative.
8. Installation will include mounting the SkyVision rack if necessary.
3-2. Customer Responsibilities
1. Customer will provide SKYTRON with the appropriate room drawings to include plan, electrical, and mechanical drawings.
2. Customer will be responsible for supply and installation of all conduit and electrical junction boxes.
3. Customer will provide proper HVAC ventilation of NDC compatible with system requirement.
4. Customer will comply with grounding requirements.
5. Customer will provide a dedicated and isolated electrical supply to the SkyVision system.
6. Customer will be responsible for supplying all specified junction boxes.
7. Customer will be responsible for supply and installation of drywall and backing for installation of wall or ceiling equipment respectively.
8. Customer will be responsible for supply and installation of appropriate nurse documentation station(s).
9. Customer will provide SKYTRON with all customer provided equipment source signal specifications including signal output type, power requirements, and cable connector types.
10. Customer will provide SKYTRON with required IP addressing information as specified on the OPI document prior to production.
11. Customer will provide SKYTRON with the dates for cable installation, boom and light installation, and agree on specific dates for SkyVision system installation. Notices for cable installation and system installation must be given no less than 21 days prior to start of work.
12. At time of installation, all Customer provided systems & equipment must be in place and working in order for SKYTRON to complete system installation and testing. If customer provided systems are not in place and working during the SKYTRON installation, delays in the SkyVision installation may occur and result in additional daily charges of $1900.00 per day, per technician, to be added to the final invoice. If return trips are necessary due to lack of customer suppled equipment, additional charges may occur.
13. Customer will assign a designated representative to coordinate installation schedules and approve the SkyVision installation within ten (10) days of completion of work. Customer representatives from BioMed, IT, and the OR Director will sign off on completion, which will signal approval of final SKYTRON invoice.
SECTION 4. CAST NETWORK RESOURCE PLAN
This section describes the storage capabilities and resources of the CAST system, the recommended type of storage sever, and the network resource being used for both real-time broadcasting and off line viewing modes.
4-1. Central Storage System (CSS) The CSS is a server or a cluster of servers performing the following tasks:
• General storage and online video storage
• On-demand broadcasting
• Authentication, encryption, and conditional access
• Mobile interface
4-2. Small Scale CSS - Recommended Servers Recommended small scale servers for pilot and trial options
DELL servers HP servers
• SFF Dell 7010 OptiPlex series
• 2x i7, CPY 3.6 GHz, quad core
• 8GB RAM
• HD: 1TB
• HP 8300 I5 4G 500G SSF
• MEMORY: HYN/TRA/CRU DDR3 1333 4GB
• 8GB RAM
• HD: 1TB
4-3. Large Scale CSS - Recommended Servers DELL PowerEdge R720 Rack Server Virtual server (Recommended)
• 4 CPU thread E5xxx
• 8GB RAM
• Windows 2012 server
• 4 CPU cores, 8 threads (statistics)
• 8GB RAM
• Windows 2008 RD or 2012
NOTICE
External storage is provided by the hospital as an external SNAP or NAS.
Both server types are fully redundant.
4-4. Server Virtualization CAST server software components are capable of being virtualized on any standard virtual machine environment.
4-5. Storage Resource Allocation The video assets are saved as a compressed MP2 transport H.264 files with a bit rate of 1.2 Mbps VBR for SD and 4 to 12 to full HD. Following is a table to calculate an average storage resource per broadcast type.
Broadcast type Avg. size per hour Total hours per 1TB SD, PAL, 720x576 250MB 4,000 Hours Full HD, 1080p 3.6GB 300 Hours
It is recommended that the system be initialized with 1TB of initial storage, thus providing 4,000 hours of storage for SD broadcasting and 300 hours for HD broadcasting.
4-6. Network Resource Allocation All CAST videos are multicast on the network with a broadband of 1.2 Mbps VBR for SD or 4 to 12 Mbps for full HD. Network calculation is simply a linear calculation of the number of concurrently used streams.
In a typical hospital environment, a network allocation for 10 concurrent capture devices capturing an SD stream uses an average sum of 12 Mbps.
NOTICE
CAST broadcasts all streams on a multicast basis. Optionally, network segmentation, to gap over VLANs, would benefit large hospital environments.
| When | multicast | is | not | enabled | in | the | network, | the | CAST | system | switches |
| automatically | (on | a | configuration | basis) | into | a | unicast | broadcast | environment. |
4-7. CAST Network Connection
| The | CAST | could | be | connected | on | network | cables | especially | when | installed | on | a | fixed | boom. | In | cases | ||||
| where | the | CAST | is | installed | on | an | independent | tower | there | is | an | option | to | use | the | built | in | WiFi | device. | |
| The | WiFi | inside | the | CAST | is | of | the | type | a/b/g/n | and | supports | both | 2.4 | and | 5 | Ghz | frequencies | to | be | used |
within a capacity of up to 300 Mbps.
NOTICE
In cases where there are several CAST machines connected to the network
| and | on | a | WiFi | connection, | creating | a | WDS | connection | and | enabling | the |
| WMM | bit | is | recommended. |
4-8. Network Address Range The CAST system uses the following address range and ports (all are configured and can be changed according to local policies);
Protocol Source -> Destination Address range Port Web Services, HTTP Any -> CSS Dynamic 80 Unicast, UDP (when multicast is not available)
Orpheus CAST -> CSS CSS -> Any
Dynamic 5000-5500
Multicast, UDP Orpheus CAST -> Any CSS -> Any Any -> Any (Talkback)
224.10.10.0- 224.255.255.255
4-9. Operating in Unicast In some cases, network segments using CAST devices may not be connected to the server using multicast, for example due to hospital policy or network layout. This section specifies limitations and solutions for devices not connected to the server using multicast.
There are two possible ways to operate in unicast mode:
| • Configure | the | CAST | to | operate | in | "Video | Upload" | mode | |
| • Maintain | the | configuration | of | CAST | to | broadcast, | but | over | unicast |
A. Video Upload Mode When CAST is configured to operate in Video Upload mode, the video is still being recorded on the server while the procedure is being captured. However, CAST uses HTTP to transfer video packets of 2MB to the server. On the server, the file is being created based on these packets.
Depending on network bandwidth allocation and latencies between CAST and the server, these packets may be “backlogged”. If this occurs, CAST stores these packets locally on the device and completes the transfer automatically. Therefore, it may take between 1-5 minutes after the procedure recording has been stopped on the device before the entire video is available on the server. The main difference between this mode and the multicast mode is, in the latter the video is available within a few seconds.
B. Mixed Mode - Multicast on Parallel The hospital network setup may be mixed. This is especially useful if the data center is in a remote location.
For example, if the CAST is deployed in a facility the facility network can be set up as multicast. However between the facility and the data center, the network can be set up as unicast.
In this situation, it is possible to enjoy multicast live streaming between the CAST device and the Show application from anywhere within a facility. However, the video recording to the server will be done through the unicast network (as described in the Video Upload Mode section).
CAST DEVICEMULTICAST
SWITCH
SERVER
SHOW APPLICATION
UNICAST
(HTTP)
MULTICAST
Figure 9. Mixed Mode
CAUTION
With Multicast on Parallel, it is very important to properly configure the network segment (VLAN) using multicast to prevent video packets from inadvertently “breaching” the “isolated” multicast setup and being transferred over the unicast. If this occurs, that packet will reach the server and may adversely affect the quality, size and integrity of the video that is being recorded at the same time.
C. Is Live Stream Possible using Unicast?
Yes, but it’s not recommended due to poor performance issues.
The next diagram shows how the CAST, the Server, and the Show client applications will operate to provide live streaming over unicast. For the example below, a video at 10Mbps bitrate is assumed to be broadcast from the CAST.
1. A single CAST will transmit to the server the video stream. This requires allocating the necessary bandwidth (e.g. 10Mbps) for the stream upload.
2. A single Server will transmit the video stream using the Peer-to-Peer method (like YouTube, Netflix, etc.) to each Show client application. This requires allocating the necessary bandwidth for each Show client (e.g. the Peer) from the server.
As depicted in the diagram below, for 3 Show client applications, the stream will consume 30Mbps (10Mbps x 3) of bandwidth from the Server.
CAST DEVICE SERVER
SHOW APPLICATION
UNICAST
SHOW APPLICATION
UNICAST
UNICAST
UNICAST
SHOW APPLICATION
Figure 10. Unicast Livestream To support the peer-to-peer broadcast from the server, the server will require dedicated Network Interface Cards (NICs). The server is commonly deployed as a virtual machine on a VM farm. In order to support a peer-to-peer broadcast in such a deployment, NICs will need to be exclusively allocated to the server, considerably reducing the efficiency of the virtual server farm.
The load on the server will increase considerably, both on disk I/O and CPU to support more than two CAST and/or more than two Show client applications. To support more concurrent streams, more servers will be needed.
4-10. Mobile Server Module – MSM The mobile Server Module is either a SW module in the CSS or a separate module on a separate unit – either a separate machine or on a separate VM. It is important to separate the CSS and the MSM because of a DMZ access to the server – usually mobiles access the system by utilizing the cellular network.
A DMZ could be created by locating the MSM on a separate DMZ or creating a virtual DMZ on a firewall enabling the MSM to connect on a separate NIC in the CSS.
The options are detailed below:
A. MSM on the CSS This option is usually created to serve requests in an internal network (intranet) for the mobile interface.
CSS
MSM
HOSPITAL
NETWORK
Figure 11. MSM on the CSS
B. MSM in DMZ
CSS
HOSPITAL
NETWORK
MSM
FW
CELLULAR
NETWORK
Figure 12. MSM in DMZ
4-11. MSM in a Virtual DMZ
CSS
MSM
HOSPITAL
NETWORK
FW
CELLULAR
NETWORK
Figure 13. MSM in a Virtual DMZ
4-12. Full MedicVOD Connection Scheme Below is a connection diagram for a video on demand application.
Figure 14. Full MedicVOD connection scheme
SECTION 5. CAST STORAGE POLICY PLAN
The CAST system compresses the video signal on the source side and delivers the compressed signal to be stored in real-time. The real-time compression, therefore, points directly to the amount of data to be stored.
5-1. Real-Time Compression The real-time compression is a direct parameter of the source. In some cases it is up to the hospital to define that a specific video source should be compressed only after it has been scaled down:
1. A scaled down source prior of compression could never be reconstructed.
2. No point of scaling down on real-time basis since there is always the option of offline, on a later stage, scale down process.
3. Images taken are not affected by a real-time scaled down compression, images are always kept in the native resolution.
A. Video source type The CAST system can capture the following video source types:
1. Composite PAL/NTSC
2. Analog VESA VGA/XGA/WXGA
3. Digital HD DVI/HDMI/HD-SDI
B. Video Resolution The CAST system can capture the following video resolution types:
1. 720x480/576 P25/30
2. Any VESA 50/60 HZ
3. 720i/P, 1080i/P 30/60
5-2. Offline Compression Offline compression is done after the file had been closed (the video part of the procedure is controlled and stopped using the CAST interface). The compression is done while performing a full transcoding process on the file and it is usually involved in changing or keeping the resolution.
Typical codecs and encapsulations are as the following:
| 1. Windows | 7 | and | above: | MP4 | encapsulating | h.264 | codec |
| 2. Windows | Vista | and | down: | WMV | encapsulating | WV8 |
3. Macintosh OSX: MOV encapsulating h.264
4. IOS and Android devices: MP2 encapsulating h.264
5-3. Storage space All compressions mentioned above are on base of VBR (Variable Bit Rate) averaging on a target bitrate within the following table:
Resolution Type Screen Pixel Size Width x Height
Average bitrate Average Size in bytes for Hour of video
| SD | (Standard | Definition) 720x576 1.2Mbps 500MB |
| HD | (High | Definition) 1280x720 4Mbsp 1.8GB |
Full HD 1920x1080 6Mbsp 2.7GB Full HD 3D 1920x1080 8Mbsp 3.6GB
A. Storage Space Policies The storage space policies are mainly involved in transcoding the original file into other formats and the period to keep them. Following are samples of transcoding to other formats:
1. MOV files for Macintosh users
2. MP4 low resolution to be used by HTML interface
3. MP2 low resolution to be used by smart phone devices
4. MP4 low resolution file to be used a long term storage
B. Long Term Storage
| Files | can | be | archived. | A | hospital | can | define | how | long | an | original | file | should | be | kept | before | it | is | transcoded |
| and | scaled | to | a | low | resolution | file | for | archiving. | |||||||||||
| Usually | the | long | term | policy | is | to | reduce | the | resolution | to | 25% | of | the | original | resolution, | creating | smaller | ||
| files | for | archiving. |
5-4. Archival Policies to Define Each department should define archival policies for each procedure. The policies should be defined per hospital server with each department having its own policies. For example, the storage policy for the
| surgery | department | could | be | different | then | the | policy | storage | for | the | ENT | department. |
| Following | are | storage | policies | that | must | be | defined: |
1. Final resolution: should the original be kept or scaled down?
2. Final transcoding: encapsulation and video codec
3. Expiration: how long should the final video be kept?
5-5. Storage Scenarios The following are some hospital storage policy scenarios. In all scenarios an operating room is active on a net base of 4 hours a day 5 days a week (totaling of 1040 hours a year) and is equipped with one
| scope | camera | (Net | base | of | a | procedure | is | usually | measured | by | 70% | of | the | actual | procedure | time, | e.g. | if |
| a | procedure | is | of | 1 | hour, | the | video | taken | will, | effectively, | last | 42 | minutes). |
No .of operating rooms / scope cameras
No. of SD video signals
No. of HD video signals (Full
HD 3D)
Policies 1 Year Storage effected
After long term
5 3 2 - Long term in 1 year
- 30 days for MOV
- 30 days for WMV
- 60 days for low res MP4
SD - 1.5 TB
HD – 7.5 TB
Total – 9 TB
2.25 TB
8 6 2 - No Long term
- No formats
SD – 3.2 TB
HD – 7.5 TB
Total – 10.7 TB
10.7 TB
The above is just an example of storage and the connection to its policies. All combinations of scaling/ transcoding and expiration (period to keep the file) are acceptable.
5-6. Backup Policy Orpheus backup policy recommend to apply the following rules:
Data Base
• Full Database Backup – Daily; Keeping last 60 days;
• Transactional Backup Storage
• Incremental Backup of source files
FDA 0900-36 REV0
SkyVision Pre-Installation Evaluation
Facility
PO# Date
The Pre-Installation Evaluation must be completed prior to the arrival of the SkyVision integration system installation team. This form can be used to describe a multi-room installation if the details that are room speci�c can be noted on the �nal page. In the event this evaluation is not completed properly and additional visits are required, the customer will be billed at $1900.00 per day, per technician.
Pre-Install Inspection The inspection must be performed by one Skytron SkyVision Technician, the local Skytron Account Representative, the on-site customer contact and pertinent contractor representatives.
Pre-install site evaluation The evaluation consists of a review of the project documentation during a “walk through” of the entire project site and completion of the checklist.
** RETURN VISITS DUE TO INACCURATE/ INCOMPLETE INSTALLATION VERIFICATION ARE SUBJECT TO BILLING **
Environmental Considerations:
Do installed Conduit Runs match drawings in both sizing and locations? Yes or no.
Are any conduit runs missing, changed or different from the submittal drawings? Yes or no.
Are the Nurse Document Centers installed and placed in their final location? Yes or no.
Is there a 12” x 12” box for conduit termination installed at the NDC or Rack location? Yes or no.
Are the ceilings installed Yes or no.
What is the ceiling material?
Will there be sufficient ceiling / mount access to pull cables through conduit? Yes or no.
Is all drywall work finished? Yes or no.
Are the finished floors installed? Yes or no.
Has all work involving dust and paint has been completed? Yes or no.
Is HVAC system completed? Yes or no.
Power Considerations:
Will permanent power be live prior to Skytron installers arriving on site? Yes or no.
Are two 120 VAC, 20A, properly circuit protected, dedicated circuits provided to power each
SkyVision system? Yes or no.
Are properly sized NEMA4 enclosures and proper power circuit provided for all monitor power supplies that are to be placed above the ceiling? Yes, no or NA
Sample
For R eference
Only
Grounding Considerations:
Has proper, 10AWG braided, ground wire been provided to the 12”x12” box? Yes or no.
Network Considerations:
Are STATIC IP addresses assigned and confirmed? (static IP, subnet mask, default gateway) Yes or no.
What is the hospitals network bandwidth?
If video conferencing is installed, what is the hospitals external IP and are the correct ports open?
IP ports open? Yes or no.
Are all required Network drops in place? Yes or no.
Equipment Mounting Considerations:
Is proper backing material installed for wall mounted equipment? (monitors, equipment rack, wall camera)
Yes or no. Type Are the Booms installed? Yes or no.
Non-Skytron Equipment Considerations:
Is there any special or custom equipment the installer(s) may not be familiar with? Yes or no.
Are all customer provided systems and equipment in place and in working order? Yes or no.
Are non Skytron supplied displays on site? Yes or no.
Have proper power supplies been provided for non Skytron displays? Yes or no.
Are proper mounts or adapters provided for monitors? Yes or no.
** RETURN VISITS DUE TO INACCURATE/ INCOMPLETE INSTALLATION VERIFICATION ARE SUBJECT TO BILLING **
Printed Name: Date:
Skytron Representative :
Signature: ____________________________________________________
Hospital Representative:
Signature: ____________________________________________________
Contractor Representative:
Signature: ____________________________________________________
Other Representative (A):
Signature: ____________________________________________________
Other Representative (B):
Signature: ____________________________________________________
Sample
For R
Are there any changes or additions not addressed by this list?
Notes:
Sample
For R
SECTION 6. REVISION HISTORY
Date Revision Revision History
03/04/2014 0 Initial Release 10/07/2014 1 • Page 4, revised Section 2.2 (Power) in its entirety.
• Page 4, added additional specification on 10 AWG copper ground wire.
• Page 10, changed "DCSS" to "SVIS-DCSS" as a clarification.
• Page 11 and 16, changed "VIS" to "SVIS" in several places as a correction.
| • Page | 18, | changed | "dedicated | and | isolated | electrical | ||
| supply" | to | "GFCI | protected | electrical | supply" | in | Section | 3.2 |
(Customer Responsibilities).
08/05/2015 2 Corrected the TEC number from TEC-R-0047 to TEC-R-0049 06/27/2016 3 • Added chapters 4 and 5 relating to the CAST system
• Reference to DCSS changed to CAST throughout
• Figure 8 - Updated typical system build drawing without iPod
06/01/2020 4 • Pg 4 - added statement on redundant cabling
• Pg 17 - Under Skytron Responsibilities added redundant cabling requirement
NOTES
5085 Corporate Exchange Blvd. S.E.
Grand Rapids, MI 49512 • 616.656.2900 • FAX 616.656.2906
NOTES