Attachment 6 Med Vac System - Penthouse Oil Lubricated duplex vacuum.xlsx
XLSX spreadsheet 60 KB Posted
- Attached to
- J045--Medical Gas Preventative Maintenance and Inspection Services for Erie VA Medical Center
- Solicitation number
- 36C24426Q0958
About this file
This is a medical vacuum system inspection and maintenance record for a Rietschle Duplex Medical Vacuum system installed at the VA Medical Center in Erie, Pennsylvania. The inspection was conducted on May 29, 2026, by technician Rgarmon on a CHBD-341-243 model system (Serial #3000496097) located in the penthouse on the 8th floor. The duplex system operates two CLFH/EH341 pumps with run hours of 80,809 and 80,922 respectively, maintaining discharge temperatures of 139°F and 141°F with room temperature at 78°F. Most system components including exhaust drip legs, filters, high-temperature shutdown coupling, motor greasing, starter contacts, transformer, panel relays, and lamps were rated as acceptable. The motors operate at 230 volts, 40 full-load amps, 15 horsepower, 1,165 RPM, with 90.20% efficiency and 1.15 service factor.
The inspection identified that both pumps require oil changes and cooling system flushing due to high operating hours, with five units of OEM synthetic oil (Part #R590) from customer stock to be used for this maintenance. The recommendations include maintaining the current preventive maintenance frequency and replacing exhaust and intake filters on both pumps, with the customer confirming availability of replacement parts. No deficiencies were noted in the final assessment. This inspection record directly supports the related federal contract opportunity (36C24426Q0958) for medical gas preventative maintenance and inspection services at the Erie VA Medical Center, a 100% Service-Disabled Veteran-Owned Small Business Set-Aside with quotes due by September 25, 2026.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Attachment 3 Med Air System -Dental duplex Reciprocating.xlsx | XLSX spreadsheet | |
| Attachment 2 Med Air System -Dental duplex Reciprocating Dental EZ.xlsx | XLSX spreadsheet | |
| Attachment 5 Med Air System -VA duplex Reciprocating.xlsx | XLSX spreadsheet | |
| Attachment 4 Med Air System -SPS Duplex Instrument Scroll air system.xlsx | XLSX spreadsheet | |
| Attachment 7 July 2026 final report.pdf | ||
| 36C24426Q0958 0002.pdf | ||
| Attachment 1 RFI 36C24426Q0958 QA.pdf | ||
| 36C24426Q0958 0001.docx | DOCX document | |
| Atch 1 SCA WD No. 2015-4223 Rev.35 Dated 08-22-2026.pdf | ||
| 36C24426Q0958_1.docx | DOCX document | |
| 36C24426Q0958.pdf |
Show all 11
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Sheet1
| Customer Information | System Information | Job Information | ||||||
| Facility | VA Medical Center Erie | System Description | Rietschle Duplex Medical Vacuum | Job # | 627910 | |||
| Address | 135 East 38th Street | System Model # | CHBD-341-243 SYSTEM | SO # | 627910 | |||
| Erie,PA 16504 | System Serial # | 3000496097 | Trip # | 1 | ||||
| Contact | David Dombkowski | Manufacture Date | None | Year | 2026 | |||
| david.dombkowski@va.org | System Location | Penthouse 8th floor (7th fl up stairs) | Date | 05/29/26 | ||||
| Phone | 814-860-2924 | Customer System # | None | Tech | Rgarmon | |||
| Pump # | #1 | #2 | #3 | #4 | ||||
| Model # | CLFH/EH341(12)100923-1200 | CLFH/EH341(12)100923-1200 | ||||||
| Serial # | 1742146 (96) | 1735925 (96) | ||||||
| Run Hours | 80,809 | 80,922 | ||||||
| Room Temperature | 78°F | |||||||
| Discharge Temperature | 139°F | 141°F | ||||||
| Exhaust Drip Leg | A | A | ||||||
| Exhaust Filters | A | A | ||||||
| Ballast Filters | A | A | ||||||
| Inlet Filters | A | A | ||||||
| Oil and Oil Filters | R | 05/29/26 | R | 05/29/26 | ||||
| High Temp Shutdown | ||||||||
| Coupling | A | A | ||||||
| Motor Greased | A | A | ||||||
| Starter Contacts | A | A | ||||||
| Thermal Overload Settings | 48.00 AMPS | 48.00 AMPS | ||||||
| Transformer Condition | A | |||||||
| Panel Lamps | A | LEGEND | ||||||
| Panel Relays | A | A = Acceptable | ||||||
| Lead Vac Switch | 18-23 | inHG | Local Lag Alarm | U = Unacceptable | ||||
| Lag 1 Vac Switch | 16-21 | inHG | Audible | Yes | R = Repaired | |||
| Lag 2 Vac Switch | inHG | Visual | Yes | NR = Needs Replaced | ||||
| Lag 3 Vac Switch | inHG | NA = Not Applicable |
| Motor Information Section | Motor #1 | Motor # 2 | Motor # 3 | Motor #4 | ||||
| Volts L1 to L2 | 206 | 33.00 AMPS | 32.00 AMPS | |||||
| Volts L2 to L3 | 203 | 33.00 AMPS | 32.00 AMPS | |||||
| Volts L1 to L3 | 203 | 33.00 AMPS | 33.00 AMPS | |||||
| Efficiency | 90.20% | Service Factor | 1.15 | Horse Power | 15 | RPM | 1,165 | |
| Frame Size | 284TC | Voltage | 230 | Full Load Amps | 40 |
Sheet2
| Parts Used | ||
| Quantity | Part Number | Part Description |
| 5 | R590 | OEM SYNTHETIC OIL CUSTOMER STOCK |
| Deficiencies |
| None |
| Comments |
| Change oil both pumps, flush oil out of coolers and exhaust box sumps 1,2. |
Pumps have high hours
| Recommendations |
| Maintain PM Frequency |
Exhaust filters and intake filters should be replaced both pumps. Customer has parts-Will Quote image1.png image2.png