Attachment 3 Med Air System -Dental duplex Reciprocating.xlsx
XLSX spreadsheet 62 KB Posted
- Attached to
- J045--Medical Gas Preventative Maintenance and Inspection Services for Erie VA Medical Center
- Solicitation number
- 36C24426Q0958
About this file
This is a Medical Gas System Inspection Report documenting preventative maintenance performed on a RAMVAC dental air duplex compressor system at VA Medical Center Erie. The inspection was conducted on May 28, 2026, by technician R. Garmon on System Model OSP25 (Serial #4773030), which has accumulated 1,065 run hours since manufacture on July 1, 2019. The report details comprehensive system evaluation including two Gast OF1202 pump units, motor performance metrics (208V, approximately 9.3-9.7 amps), and component condition assessments using a standardized legend (A=Acceptable, U=Unacceptable, R=Repaired, NR=Needs Replaced, NA=Not Applicable). All major components including discharge check valves, unloaders, solenoids, inlet filters, after cooler drain valves, crankcase filters, piston rings, and relief valves were marked acceptable, with lead pressure switch readings of 80-100 psi and room temperature at 77°F.
The report notes that the customer facility is currently servicing dental units with patient backlog constraints and has installed an additional compressor system, requesting information and parts compatibility details for the new dental air system. The deficiencies section indicates that parts may be required for service depending on the specifications of the newly installed equipment. No CO Monitor or Dew Point Monitor was present on the current system, and auto blow-down line functionality was not assessed. This inspection documentation directly supports the VA's solicitation for Medical Gas Preventative Maintenance and Inspection Services (Solicitation 36C24426Q0958, a 100% SDVOSB set-aside with a September 25, 2026 quote deadline).
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Attachment 6 Med Vac System - Penthouse Oil Lubricated duplex vacuum.xlsx | XLSX spreadsheet | |
| Attachment 2 Med Air System -Dental duplex Reciprocating Dental EZ.xlsx | XLSX spreadsheet | |
| Attachment 5 Med Air System -VA duplex Reciprocating.xlsx | XLSX spreadsheet | |
| Attachment 4 Med Air System -SPS Duplex Instrument Scroll air system.xlsx | XLSX spreadsheet | |
| Attachment 7 July 2026 final report.pdf | ||
| 36C24426Q0958 0002.pdf | ||
| Attachment 1 RFI 36C24426Q0958 QA.pdf | ||
| 36C24426Q0958 0001.docx | DOCX document | |
| Atch 1 SCA WD No. 2015-4223 Rev.35 Dated 08-22-2026.pdf | ||
| 36C24426Q0958_1.docx | DOCX document | |
| 36C24426Q0958.pdf |
Show all 11
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Sheet1
| Customer Information | System Information | Job Information | ||||||||
| Facility | VA Medical Center Erie | System Description | RAMVAC dental air duplex | Job # | 627910 | |||||
| Address | 135 East 38th Street | System Model # | OSP25 | SO # | 627910 | |||||
| Erie,PA 16504 | System Serial # | 4773030 | Trip # | 1 | ||||||
| Contact | Davvid Dombkowski | Manufacture Date | 07/01/19 | Year | 2026 | |||||
| david.dombkowski@va.org | System Location | Dental Mechanical room | Date | 05/28/26 | ||||||
| Phone | 814-860-2924 | Customer System # | Tech | Rgarmon | ||||||
| Pump # | #1 | #2 | #3 | #4 | ||||||
| Model # | OF1202 | OF1202 | ||||||||
| Serial # | Gast Item#004822 | Gast Item#004822 | ||||||||
| Run Hours | 1065 | |||||||||
| Room Temperature | 77°F | |||||||||
| Discharge Temperature | ||||||||||
| Discharge Check Valves | A | A | ||||||||
| Unloaders and Solenoids | A | A | ||||||||
| Inlet Filters | A | A | ||||||||
| After Cooler Drain Valves | A | A | ||||||||
| High Temp Shutdown | ||||||||||
| Crankcase Filter | ||||||||||
| Rider and Piston Rings | A | A | ||||||||
| Starter Contacts | A | A | ||||||||
| Thermal Overload Settings | ||||||||||
| Transformer Condition | ||||||||||
| Panel Lamps | A | LEGEND | ||||||||
| Relief Valves | A | A = Acceptable | ||||||||
| Panel Relays | A | U = Unacceptable | ||||||||
| Lead Pressure Switch | 80-100 | psi | Local Lag Alarm | R = Repaired | ||||||
| Lag 1 Pressure Switch | psi | Audible | NR = Needs Replaced | |||||||
| Lag 2 Pressure Switch | psi | Visual | NA = Not Applicable | |||||||
| Lag 3 Pressure Switch | psi | |||||||||
| Motor Information Section | Motor #1 | Motor # 2 | Motor # 3 | Motor #4 | ||||||
| Volts L1 to L2 | 208 | 9.50 AMPS | 9.30 AMPS | |||||||
| Volts L2 to L3 | 208 | 9.40 AMPS | 9.50 AMPS | |||||||
| Volts L1 to L3 | 208 | 9.40 AMPS | 9.70 AMPS | |||||||
| Efficiency | Service Factor | Horse Power | RPM | 1,650 | ||||||
| Frame Size | Voltage | Full Load Amps | ||||||||
| CO Monitor | Dew Point Monitor | |||||||||
| Manuf/Model/SN: | None | Manuf/Model/SN: | None | |||||||
| Test Gas Reading | Gas Conc.= | 20 PPM | D.P. Reading (as found) | |||||||
| Cal Performed ? | Fail | Measured D.P. (Medpipe) | ||||||||
| Sensor Replaced ? | Sensor Replaced ? | |||||||||
| Reading after Cal | D.P. Reading (as left) | |||||||||
| Dryers | ||||||||||
| Unit On Line (as found) | Pre Filters | Lamps | ||||||||
| Unit On Line (as left) | After Filters | Suction Pressure | ||||||||
| Selector Switch | Discharge Check Valve | Blow Down Valve | ||||||||
| Purge Select | Mufflers | Auto Blow Down | ||||||||
| Line Regulator | Solenoids | Float Drain |
Sheet2
| Parts Used | ||
| Quantity | Part Number | Part Description |
Deficiencies
| Comments |
| customer servicing units for dental due to patient backlog |
| Customer installed additional compressor system- |
| Customer would like information for the new dental air system. Will require parts for service if different from |
| current compressor. |
Recommendations image1.png image2.png