Med Air System - VA duplex Reciprocating.pdf

PDF 176 KB Posted

Attached to
Z1DA--Upgrade Medical Gas & Communication System Design Build
Solicitation number
36C24426R0119
Issued by

About this file

This document is a maintenance service report for a medical air system at VA Medical Center Erie, dated December 11, 2024. The report documents the inspection and servicing of a Lifeline Duplex Medical Air system (Model OHT-7TDX-D120-072299, Serial 25551) manufactured August 20, 1999, located in the basement mechanical area. The system comprises two Hitachi 5.5OU-8.5CH compressors with run hours of 8,464 and 8,814 respectively. During the inspection, technicians tested system operation, all alarms, and replaced line filters. Motor amperage readings were documented for four motors, and both CO and dew point monitors were calibrated and tested to acceptable standards with readings of 20 PPM and 7 respectively.

The service report identifies multiple deficiencies requiring corrective action. Key issues include a bottom brass regulator requiring replacement with a gauge, right dryer #2 malfunction (not purging or functioning correctly), and a line regulator rated at 50 psi that needs replacement. Parts used included two cogged belts. The technician recommends complete replacement of the medical air package and installation of a new dryer package with line filters, CO monitor, and dew point controller, citing the system's age and multiple issues. A detailed quote was requested for replacement parts including two pressure tower gauges, one switching valve, one purge valve, two check valves, one muffler, and three bags of desiccant for dryer changes. MedPipe performed the service work with contact technician RG/TP at 814-853-5430.

View the file

Other files

Show all 15

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Facility Job # 622374

Address SO # 622374

Trip # 1

Contact Year 2024

Email Date 12/11/24

Phone Tech RG/TP

R 12/11/24 R 12/11/24

A A

A A

A 01/24/24 A 01/24/24

A A

A A

A A

A A

90-105 psi

85-100 psi Audible Yes psi Visual Yes psi

RPM

Manuf/Model/SN: Manuf/Model/SN:

Gas Conc.= 20 PPM

Pass

R 12/11/24

R 12/11/24

A

NR right

NA = Not Applicable

Local Lag Alarm

GFG 4021 AR KAHN-AA05-052 (SF72)

A

Motor Information Section

Volts L1 to L2

Frame Size Voltage Full Load Amps

Lag 3 Pressure Switch

Efficiency Service Factor

Reading after Cal 20 PPM

Dew Point Monitor

D.P. Reading (as found)

Measured D.P. (Medpipe)

Sensor Replaced ?

No

Test Gas Reading

Cal Performed ?

20 PPM

Yes

Sensor Replaced ? No

CO Monitor

D.P. Reading (as left) 7

Horse Power

Volts L2 to L3

Volts L1 to L3

Motor #1 Motor # 2 Motor # 3 Motor #4

19.50 AMPS

20.10 AMPS

20.40 AMPS

20.10 AMPS

20.40 AMPS

20.30 AMPS

Lag 2 Pressure Switch

28.50 AMPS

A

A

A

Thermal Overload Settings

Transformer Condition

Panel Lamps

Panel Relays

Lead Pressure Switch

Relief Valves

High Temp Shutdown

Crankcase Filter

Rider and Piston Rings

Starter Contacts

814-853-5430

Discharge Check Valves

Unloaders and Solenoids

Inlet Filters

After Cooler Drain Valves

Belt Tension and Condition

141°F

Model #

Serial #

Run Hours

Hitachi 5.5OU-8.5CH

28.50 AMPS

Lag 1 Pressure Switch

LEGEND

A = Acceptable

U = Unacceptable

R = Repaired

NR = Needs Replaced

Job Information

Pump # #1 #2 #3 #4

Customer System #

Lifeline Duplex Medical Air system

OHT-7TDX-D120-072299

25551

08/20/99

Basement floor mechanical

562 ee8386 pmi sn

System Description

System Model #

System Serial # david.dombkowsi@va.org

Manufacture Date

System Location

System Information

VA Medical Center Erie

Customer Information

135 East 38th Street

Erie,PA 16504

David Dombkowski

140°F private label

Discharge Temperature

74°FRoom Temperature

Hitachi 5.5OU-8.5CH private label

8,814

Auto Blow DownPurge Select Controlled

After Filters

Discharge Check Valve

Mufflers

Unit On Line (as found)

Unit On Line (as left)

Dryers

Dryer #2 Pre Filters

Dryer #1

Selector Switch On

Lamps

Suction Pressure

Blow Down Valve mailto:david.dombkowsi@va.org mailto:david.dombkowsi@va.org mailto:david.dombkowsi@va.org

NR right Float DrainLine Regulator 50ps (NR) Bottom Solenoids

Quantity

Part Number Part Description

Cogged belts belts 1,2 see above

Replace Medical air package with new and install new dryer package with line filters co monitor dew point controller upgrade system due to age and issues if possible-will quote finding for repairs of this system

Recommendations

Comments

Parts Used

Deficiencies

None

Regulator bottom 1/2" needs replaced with gauge rebuild right dryer #2 not purging or functioning correctly at this tim

Tested system operation and all alarms, replaced line filters found bottom brass regulator creeping up and down allow top line regulator to remain on-line right side dessicant dryer not functioning correctly

Lifeline Model DRY05-002L Serial: C82856

Quote Qty 2 pressure tower gauges, Qty 1 switching valve, Qty 1 purge valve, Qty 2 check valves Qty 1 muffler qty 3 bags of dessicant or enough to perform qty 2 dryer desiccant changes

File details come from the government source that posted it. Updated .