Med Air System - Dental duplex Reciprocating.pdf
PDF 162 KB Posted
- Solicitation number
- 36C24426R0119
About this file
This is a preventive maintenance service report for a dental air compressor system at the VA Medical Center in Erie, Pennsylvania. The document records the inspection and testing of a RAMVAC dental air duplex system (Model OSP25, Serial #4773030) performed on December 11, 2024, by technician RG/TP. The system, manufactured on July 1, 2019, is located in the dental mechanical room and operates two Gast OF1202 compressor units (Item #004822 each).
The inspection findings indicate that all major components evaluated—including discharge check valves, unloaders and solenoids, inlet filters, after cooler drain valves, rider and piston rings, and starter contacts—were rated as "Acceptable" across both pump units. Motor testing recorded voltage readings of 208V across all three phase combinations and amperage readings ranging from 9.30 to 9.70 AMPS across Motor #1 and Motor #2, with the system operating at 1,650 RPM. The CO monitor showed no manufacturer/model information available, and the dew point monitor similarly had no data recorded. The lead pressure switch was set to 80-100 psi with local lag alarm capability (audible and visual indicators). The report notes that the customer has installed an additional compressor system and is currently servicing dental units due to patient backlog. No parts were used during this service visit, and no specific deficiencies or recommendations were documented in the respective sections.
View the file
Other files
Show all 15
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Facility Job # 622374
Address SO # 622374
Trip # 1
Contact Year 2024
Email Date 12/11/24
Phone Tech RG/TP
A A
A A
A A
A A
A A
A A
80-100 psi psi Audible psi Visual psi
RPM 1,650
Manuf/Model/SN: Manuf/Model/SN:
Gas Conc.= 20 PPM
Fail
Unit On Line (as found)
Unit On Line (as left)
Dryers
Pre Filters
Selector Switch
Lamps
Suction Pressure
Blow Down Valve
Auto Blow Down
Float Drain
Purge Select
Line Regulator
After Filters
Discharge Check Valve
Mufflers
Solenoids
Davvid Dombkowski
Gast Item#004822
Discharge Temperature
77°FRoom Temperature
OF1202
Gast Item#004822
Job Information
Pump # #1 #2 #3 #4
Customer System #
RAMVAC dental air duplex
OSP25
4773030
07/01/19
Dental Mechanical room
System Description
System Model #
System Serial # david.dombkowski@va.org
Manufacture Date
System Location
System Information
VA Medical Center Erie
Customer Information
135 East 38th Street
Erie,PA 16504
Lag 1 Pressure Switch
LEGEND
A = Acceptable
U = Unacceptable
R = Repaired
NR = Needs Replaced
High Temp Shutdown
Crankcase Filter
Rider and Piston Rings
Starter Contacts
814-860-2924
Discharge Check Valves
Unloaders and Solenoids
Inlet Filters
After Cooler Drain Valves
Model #
Serial #
Run Hours
OF1202
Lag 2 Pressure Switch
A
A
Thermal Overload Settings
Transformer Condition
Panel Lamps
Panel Relays
Lead Pressure Switch
Relief Valves
Horse Power
Volts L2 to L3
Volts L1 to L3
Motor #1 Motor # 2 Motor # 3 Motor #4
9.50 AMPS
9.40 AMPS
9.40 AMPS
9.30 AMPS
9.50 AMPS
9.70 AMPS
Reading after Cal
Dew Point Monitor
D.P. Reading (as found)
Measured D.P. (Medpipe)
Sensor Replaced ?
Test Gas Reading
Cal Performed ?
Sensor Replaced ?
CO Monitor
D.P. Reading (as left)
NA = Not Applicable
Local Lag Alarm
None None
A
Motor Information Section
Volts L1 to L2
Frame Size Voltage Full Load Amps
Lag 3 Pressure Switch
Efficiency Service Factor
Quantity
Parts Used
Deficiencies
Customer installed additional compressor system customer servicing units for dental due to patient backlog
Recommendations
Comments
Part Number Part Description