583-20-102 Relocate Nuclear Medicine Drawing SetPart-1_01_Part8.pdf

PDF 4 MB Posted

Attached to
Z1DA--Renovate Nuclear Med 583-20-102 (VA-21-00037478)
Solicitation number
36C25021B0036
Issued by

About this file

This solicitation is for renovation services to relocate Nuclear Medicine services at the Richard L. Roudebush VA Medical Facility in Indianapolis, Indiana. The solicitation was issued by the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 10. Renovation work includes relocating Nuclear Medicine services to a centralized location and includes architectural, mechanical, electrical, and structural upgrades. The estimated award value is $2-5M. Responses are due by February 28, 2021 and award is expected in April 2021. The set-aside type was not specified.

View the file

Other files

File Type Posted
Amendment A00006.docx DOCX document
Amendment A00006 RFI Responses.docx DOCX document
Amendment 36C25021B0036 A00005.pdf PDF
A00004- Corrected Statement of Work.pdf PDF
Amendment 36C25021B0036 A00004.pdf PDF
A00004 - RFI Responses.docx DOCX document
A0004 - 08 71 00 - Door Hardware.pdf PDF
A00004 - 10 25 13 - Patient Bed Service Walls.pdf PDF
A00004 - Door Schedule.pdf PDF
Amendment 36C25021B0036 A00003.pdf PDF
Amendment A00002.pdf PDF
Site Visit Sign-in Sheet.pdf PDF
A0001 - RFI 1 ShieldingReportComplete.pdf PDF
A0001 - RFI 2- GC101.pdf PDF
A0001 - RFI Responses.docx DOCX document
Amendment A0001.docx DOCX document
A0001 - RFI 4- Door Schedule.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-2_Part1.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-1_01_Part12.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-2_Part18.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-1_01_Part10.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-1_01_Part5.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-2_Part23.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-2_Part22.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-1_01_Part7.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-1_01_Part14.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-1_01_Part11.pdf PDF
Specifications.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-2_Part21.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-1_01_Part13.pdf PDF
Wage Determination.txt TXT text file
583-20-102 Relocate Nuclear Medicine Drawing SetPart-2_Part15.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-2_Part16.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-2_Part4.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-1_01_Part9.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-2_Part20.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-2_Part5.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-2_Part10.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-2_Part7.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-2_Part12.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-2_Part6.pdf PDF
36C25021B0036 Solicitation Document.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-2_Part11.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-2_Part14.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-1_01_Part16.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-2_Part24.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-2_Part9.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-1_01_Part1.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-1_01_Part2.pdf PDF
583-20-102 Relocate Nuclear Medicine Drawing SetPart-1_01_Part3.pdf PDF
Show 40 more

Latest 50 of 66 listed here. See the rest on GovTribe.

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

N

KEY PLAN

PROJECT

LOCATION

A. THE ARCHITECTURAL REFLECTED CEILING PLAN SHALL GOVERN THE

LAYOUT OF ALL CEILING ELEMENTS, HEIGHTS AND/OR PENETRATIONS.

COORDINATION OF THE ARCHITECTURAL AND ENGINEERING ELEMANTS ARE

THE RESPONSIBILITY OF THE CONTRACTOR. THE CONTRACTOR IS TO

VERIFY FIELD CONDITIONS AND REPORTS ANY DISCREPANCIES TO

ARCHITECT AND COR.

B. SHOWN SIZES OF MECHANICAL, ELECTRICAL AND PLUMBING EQUIPMENT

ARE FOR REFERENCE ONLY. REFER TO ENGINEERING PLANS FOR

ADDITIONAL INFORMATION. COORDINATE SCOPE OF WORK WITH

ENGINEERING DRAWINGS, UNLESS OTHERWISE NOTED.

C. ACOUSTIC PANEL CEILING SHALL BE CENTERED IN THE ROOM AND

INSTALLED TO ELIMINATE NARROW CUT STRIPS OF PANELS ALONG THE

EDGES OF WALLS AND BULKHEADS, UNLESS OTHERWISE NOTED.

COORDINATE WITH THE LIGHTING LAYOUT. THE LIGHTING SHALL BE

CENTERED IN ROOMS.

D. CEILING MOUNTED ITEMS (SPRINKLER HEADS, MOUNTING DEVICES, ETC.)

SHALL BE CENTERED IN THE CEILING PANELS UNLESS NOTED OTHERWISE.

E. ABOVE CEILING WORK REQUIRED BEYOND THE SCOPE OF THE WORK SHALL

BE COORDINATED WITH PATIENT SAFETY, INFECTION CONTROL, AND COR.

F. CEILING TILE REMOVED IN ORDER TO COMPLETE ABOVE CEILING WORK

SHALL BE STORED IN A SAFE DRY LOCATION. DAMAGED OR BROKEN CEILING

TILES SHALL BE REPLACED. DRYWALL CEILINGS SHALL BE PATCHED,

PREPPED AND NEW FINISH APPLIED TO MATCH EXISTING.

G. REFER TO FINISHES SHEETS AND SPECIFICATIONS FOR INFORMATION ON

GYPSUM BOARD PAINT FINISH.

1. SALVAGED PATIENT LIFTS - PROVIDE AND INSTALL NECESSARY BRACKETS.

SEE DETAIL 10/AE501.

2. SYMBOL NOTES CENTER POINT OF EQUIPMENT BELOW. CEILING GRID AND

ELEMENTS TO ALIGN TO EQUIPMENT CENTER POINT.

REFLECTED CEILING GENERAL NOTES

REFLECTED CEILING PLAN NOTES

A

TYPE GRAPHIC DESCRIPTION

D

B

2x2 ACOUSTICAL CEILING TILE (AP-1)

GYPSUM BOARD CEILING. CEILING PAINT.

2x2 ACOUSTICAL CEILING TILE (AP-2)

REFLECTED CEILING PLAN TYPE LEGEND

E NO CEILING / EXISTING CEILING TO REMAIN

2x4 ACOUSTICAL CEILING TILE (AP-3)C

A

457 3.5

B

C

D

E

F

2.58910

D.5

111213

AE501

235 SF

EQUIP CALIBRATION

A-B014

520 SF

READING &

CONSULTATION

A-B013

106 SF

DOSE ADMIN. ROOM

A-B012

92 SF

SOILED UTILITY

A-B01

145 SF

DOSE ADMIN. ROOM

A-B010 114 SF

DOSE ADMIN. ROOM

A-B009

114 SF

DOSE ADMIN. ROOM

A-B008

316 SF

HOT LAB

A-B007

270 SF

COMBINED PET

CONTROL ROOM

A-B005

488 SF

PET / CT 1

A-B006

NORTH CORRIDOR

C-2

452 SF

NM SCAN 1

A-B029a

456 SF

NM SCAN 2

A-B029b

338 SF

NM CONTROL RM

A-B029

150 SF

DIGI ROD, MOBILE

EQUIPMENT

A-B030

136 SF

CLEAN SUPPLY

A-B031

123 SF

THYROID UPTAKE

A-B022

32 SF

ALCOVE

C-5A

202 SF

INPATIENT HOLDING

A-B021

457 SF

CARDIAC STRESS

TESTING

A-B023

760 SF

OUTPATIENT

WAITING

A-B020

717 SF

AUDITORIUM

CORRIDOR

CB-10

215 SF

MSA / SCHEDULING

A-B019

150 SF

DOSE ADMIN. ROOM

A-B002

150 SF

DOSE ADMIN. ROOM

A-B003

112 SF

PET EQUIPMENT

ROOM

A-B004a

464 SF

PET / CT 2

A-B004

65 SF

TOILET, STAFF

A-B026

59 SF

TOILET, STAFF

A-B027

64 SF

PET TOILET,

PATIENT

A-B028

130 SF

SUPERVISOR

OFFICE

A-B018

60 SF

NM TOILET, PATIENT

A-B017

69 SF

NM TOILET, PATIENT

A-B016

126 SF

PET EQUIPMENT

ROOM

A-B006a

136 SF

PATIENT INTERVIEW

ROOM

A-B024

A 9' - 0" A 9' - 0"

A 9' - 0"

A 9' - 0"

A 9' - 0"

A 9' - 0"

A 9' - 0"

A 9' - 0"

A 9' - 0"

A 9' - 0"

A 9' - 0"

D 9' - 0"

A 9' - 0"

401 SF

STAFF LOUNGE

A-B015

A 9' - 0"

A 9' - 0"

A 9' - 0"

A 9' - 0"

A 9' - 0"

A 9' - 0"

A 9' - 0"

A 9' - 0"

A 9' - 0"

A 9' - 0"

A 9' - 0"

413 SF

WEST CORRIDOR

C-5

509 SF

SOUTH CORRIDOR

C-4

156 SF

CORRIDOR

C-6

324 SF

EAST CORRIDOR

C-3

238 SF

STAFF ENTRANCE

CORRIDOR

C-1

A 9' - 0"

102 SF

RADIOACTIVE

WASTE STORAGE

A-B007a

2092 SF

AUDITORIUM

A-B066

1741 SF

OFFICE

A-B040

153 SF

QCIS

A-B039

234 SF

TEAM ROOM

A-B038105 SF

PATIENT SAFETY

MANAGER

A-B043

270 SF

ASSISTANT LOG

A-B041

152 SF

CHIEF LOG.

A-B042

84 SF

NOURISHMENT

A-B047

68 SF

INFECTION

CONTROL MANAGER

A-B046

118 SF

CHIEF QUALITY

MANAGEMENT

A-B045

76 SF

RISK MANAGER

A-B044

D 8' - 6"

2' -2

D 8' - 6"

A 9' - 0"

A 9' - 0"

A 9' - 0"

A 9' - 0"

A 9' - 0"

AE501

AE501

AE501

AE501

AE501

AE501

D 8' - 6" D 8' - 6"

D 8' - 6"

A 9' - 0"

A 9' - 0" A 9' - 0"

A 9' - 0"A 9' - 0"

A 9' - 0"

A 9' - 0"

A 9' - 0"

A 9' - 0"

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Office of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ARCHITECT/ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Consultant:

C

U s e rs \d c h ri s te n s e n \D o c u m e n ts \1

.1

.0

N u c M e d v d c h ri s te n s e n M

N F

X

.r v t

/2

/2

:5

:4

A

M

AE111

LOWER LEVEL REFLECTED

CEILING PLANGUIDON DESIGN, INC.

1221 N. PENNSYLVANIA STREET

SUITE 200

INDIANAPOLIS, INDIANA 46202

317.800.6388

HEAPY

1400 W. DOROTHY LANE

DAYTON, OH 45409

937.224.0861

RELOCATE NUCLEAR MEDICINE

SERVICES FOR CENTRALIZED

SERVICES

INDIANAPOLIS, IN

583-20-102

01/28/2021 JLG PAE, II

BID DOCUMENTS

FULLY SPRINKLERED

SCALE: 1/8" = 1'-0"AE111

4 REFLECTED CEILING PLAN

Revisions: Date:

6" M

IN

V A

R

IE

S

2 1/2" METAL STUD BRACING ON BOTH

SIDES AT 48" O.C. - MIN. 3 LOCATIONS.

ANCHOR TO STRUCTURE ABOVE

5/8" TYPE "X" GYPSUM BOARD

4 7/8"

3-5/8" METAL STUD FRAMING AT 16" O.C.

ANCHOR TO STRUCTURE ABOVE

BOTTOM OF BULKHEAD

VARIES

BOTTOM OF CEILING

VARIES

2 1/2" METAL STUD BRACING ON BOTH

SIDES AT 48" O.C. - MIN. 3 LOCATIONS.

ANCHOR TO STRUCTURE ABOVE

5/8" TYPE "X" GYPSUM BOARD

3-5/8" METAL STUD FRAMING AT 16" O.C.

ANCHOR TO STRUCTURE ABOVE

BOTTOM OF CEILING

VARIES

BOTTOM OF CEILING

VARIES

M

IN6"

V A

R

IE

S

2 1/2" METAL STUD BRACING AT 32" O.C.

ANCHOR TO STRUCTURE ABOVE

5/8" TYPE "X" GYPSUM BOARD

3-5/8" METAL STUD FRAMING AT 16" O.C.

ANCHOR TO STRUCTURE ABOVE

V A

R

IE

S 6"

M

IN

REFER TO REFLECTED CEILING PLAN

VARIES

BOTTOM OF BULKHEAD

VARIES

BOTTOM OF CEILING

VARIES

5/8" TYPE "X" GYPSUM BOARD

Lower Level 100' - 0"

CEILING, RE: REFLECTED CEILING PLAN

5/8" TYPE X GYP BD,BOTH SIDES

BEAD OF CONT. SEALANT, BOTH SIDES

6" STEEL STUDS @ 16" O.C.

6" SOUND ATTENUATION BATT

INSULATION

LEAD SHIELDING PER PHYSICIST REPORT

TO 7'-0" AFF MIN.

EXISTING SLAB ON GRADE TO REMAIN.

LINE OF STRUCTURAL DECK, ABOVE

STEEL STUD BRACING @ 4'-0" O.C.

DEEP LEG SLIP TRACK AT DECK

HOLD GYPSUM BACK FROM DECK FOR

DEFLECTION, BOTH SIDES

RADIATION SIDE

BEAD OF CONT. SEALANT, BOTH SIDES

FILL FLUTES OF DECK WITH FIRE SAFING

AT FIRE RATED WALLS

9' -0

Lower Level 100' - 0"

CEILING, RE: REFLECTED CEILING PLAN

5/8" TYPE X GYP BD,BOTH SIDES

BEAD OF CONT. SEALANT, BOTH SIDES

6" STEEL STUDS @ 16" O.C.

6" SOUND ATTENUATION BATT

INSULATION

LEAD SHIELDING PER PHYSICIST REPORT

TO 7'-0" AFF MIN.

EXISTING SLAB ON GRADE TO REMAIN.

LINE OF STRUCTURAL DECK, ABOVE

STEEL STUD BRACING @ 4'-0" O.C.

DEEP LEG SLIP TRACK AT DECK

HOLD GYPSUM BACK FROM DECK FOR

DEFLECTION, BOTH SIDES

RADIATION SIDE

BEAD OF CONT. SEALANT

9' -0

6"

Lower Level 100' - 0"

CEILING, RE: REFLECTED CEILING PLAN

5/8" TYPE X GYP BD

BEAD OF CONT. SEALANT

6" STEEL STUDS @ 16" O.C.

6" SOUND ATTENUATION BATT

INSULATION

EXISTING WALL TO REMAIN, VIF.

LEAD SHIELDING PER PHYSICIST REPORT

TO 7'-0" AFF MIN.

EXISTING SLAB ON GRADE TO REMAIN.

LINE OF STRUCTURAL DECK, ABOVE

STEEL STUD BRACING @ 4'-0" O.C.

DEEP LEG SLIP TRACK AT DECK

HOLD GYPSUM BACK FROM DECK FOR

DEFLECTION

RADIATION SIDE

BEAD OF CONT. SEALANT

9' -0

Lower Level 100' - 0"

CEILING, RE: REFLECTED

CEILING PLAN

WALL PER PLAN, RE: WALL

TYPES

WALL PER PLAN, RE: WALL

TYPES. ANCHOR TO

STRUCTURAL DECK,

ABOVE

1/2" TEMPERED GLASS

SLIDING BARN DOOR

TRANSACTION WINDOW

EXISTING SLAB ON GRADE TO

REMAIN.

LINE OF STRUCTURAL DECK, ABOVE

STEEL STUD BRACING @ 4'-0" O.C.

CASEWORK, RE:INTERIORS

DRAWINGS

COUNTER, RE: INTERIORS

DRAWINGS

LINE OF WALL, BEYOND

2' -4

/2

1/ 2"

4' -6 max

6"

Lower Level 100' - 0"

CEILING, RE: REFLECTED

CEILING PLAN

WALL PER PLAN, RE: WALL

TYPES

WALL PER PLAN, RE: WALL

TYPES. ANCHOR TO

STRUCTURAL DECK,

ABOVE

1/2" TEMPERED GLASS

SLIDING BARN DOOR

TRANSACTION WINDOW

EXISTING SLAB ON GRADE TO

REMAIN.

LINE OF STRUCTURAL DECK, ABOVE

STEEL STUD BRACING @ 4'-0" O.C.

CASEWORK, RE:INTERIORS

DRAWINGS

COUNTER, RE: INTERIORS

DRAWINGS

1' -1

7/

8" 3'

-6

3' -6

MAX

6"

1/

2"

9' -0

#4 REBAR @ 16" O.C. - EPOXY INTO

CONCRETE A MIN. DEPTH OF 4 1/2"

TYPICAL AROUND PERIMETER

EXISTING CONCRET SLAB

MINIMUM 6" CONCRETE SLAB ON

1 1/2" RIGID INSULATION ON

10 MIL. VAPOR BARRIER ON

6" GRAVEL BASE

REFER TO PLAN FOR DIMENSIONS

C O

V E

R 3"

4 1/2" 4 1/2"

C O

V E

R 3"4 1/2"4 1/2"

EXISTING CONCRET SLAB

REFER TO PLANS FOR DIMENSIONS

CONCRETE PAD ADDITION

COVER

3"

#4 REBAR @ 16" O.C. - EPOXY INTO

CONCRETE A MIN. DEPTH OF 4 1/2",

TYPICAL AROUND PERIMETER

CHAMFERED CORNERS-

TYPICAL AROUND PERIMETER

EXISTING RAISED CONCRETE

PAD

#3 REBAR @ 12" O.C. - EPOXY INTO

CONCRETE A MIN. DEPTH OF 3",

TYPICAL AROUND PERIMETER

VERIFY IN FIELD VERIFY IN FIELD

ALIGN

C O

V E

R 3"

4 1/2" 4 1/2"

1/

2"

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Office of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ARCHITECT/ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Consultant:

C

U s e rs \d c h ri s te n s e n \D o c u m e n ts \1

.1

.0

N u c M e d v d c h ri s te n s e n M

N F

X

.r v t

/2

/2

:5

:5

A

M

AE501

BUILDING DETAILS

GUIDON DESIGN, INC.

1221 N. PENNSYLVANIA STREET

SUITE 200

INDIANAPOLIS, INDIANA 46202

317.800.6388

HEAPY

1400 W. DOROTHY LANE

DAYTON, OH 45409

937.224.0861

RELOCATE NUCLEAR MEDICINE

SERVICES FOR CENTRALIZED

SERVICES

INDIANAPOLIS, IN

583-20-102

01/28/2021 JLG PAE, II

BID DOCUMENTS

FULLY SPRINKLERED

SCALE: 1 1/2" = 1'-0"AE501

7 TYPICAL BULKHEAD DETAIL - SINGLE STUD

SCALE: 1 1/2" = 1'-0"AE501

8 CEILING TRANSITION DETAIL

SCALE: 1 1/2" = 1'-0"AE501

6 TYPICAL BULKHEAD DETAIL

SCALE: 3/4" = 1'-0"AE501

1 1-HOUR RATED WALL W/ LEAD SHIELDING

SCALE: 3/4" = 1'-0"AE501

5 LEAD SHIELD SMOKE BARRIER

SCALE: 3/4" = 1'-0"AE501

4 LEAD SHIELDED FURRING

SCALE: 3/4" = 1'-0"AE501

3 MSA/SCHEDULING SEATED COUNTER SECTION

SCALE: 3/4" = 1'-0"AE501

2 MSA/SCHEDULING STANDING COUNTER SECTION

SCALE: 1 1/2" = 1'-0"AE501

9 INFILL SLAB DETAIL

SCALE: 3" = 1'-0"AE501

10 PATIENT LIFT MOUNTING DETAIL

Revisions: Date:

SCALE: 1 1/2" = 1'-0"AE501

12 RAISED CONCRETE PAD DETAIL

D

S3A.Gi

EXISTING

CONSTRUCTION

F0D.G

1'-4 3/4" 2'-2 3/8"

1' -1

/8

1'

-0

/8

F3D.G

F3D.G

F3D.G

11 1/4"

S3A.Gi

E

S3A.GiF6D.G

F3D.G 9" 9 5/8"

ALIGN

2' -1

1/

8" 1'

-4

/8

ALIGN

EXISTING

CONSTRUCTION

S3A.Gi

ALIGN

FIRE EXTINGUISHER

CABINET

EQ EQ

EXISTING

CONSTRUCTION

S6A.GiL

F3D.G

1' -1

/8

135.00°

F6D.Gi

C

5/

8" 1' -1

/8

F3D.G

S6A.GiL

135.00°

EXISTING

CONSTRUCTION

F3D.G

F3D.G

A

3.5

1'-0 5/8" 3'-2 1/8"

/8

1' -1

/8

F3D.G

EXISTING

CONSTRUCTION

ALIGN WITH

EXISTING

CONSTRUCTION

ALIGN WITH

EXISTING

CONSTRUCTION

B

2'-5 1/8" 1'-1 1/4"

ALIGN WITH

EXISTING

CONSTRUCTION S6A.GiL

ALIGN WITH

EXISTING

CONSTRUCTION

/4

1' -1

/8

F3D.G

EXISTING

CONSTRUCTION F3D.G

C

10 1/2"

F3D.G

1/

8"

1/ 2"

ALIGN WITH

EXISTING

CONSTRUCTION

F3D.GiF3D.G

F3D.G

F3D.G

1' -2

/4

10" 1'-2 1/8"

ALIGN

D.5

F3D.G

F3D.G

10 3/4" 1'-0 7/8"

1' -2

/2

1'

-4

/2

2' -2

S3A.Gi

F3D.G

S3A.Gi

ALIGN WITH

EXISTING

CONSTRUCTION

ALIGN

WITH EXISTING

CONSTRUCTION

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Office of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ARCHITECT/ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Consultant:

C

U s e rs \d c h ri s te n s e n \D o c u m e n ts \1

.1

.0

N u c M e d v d c h ri s te n s e n M

N F

X

.r v t

/2

/2

:5

:5

A

M

AE502

COLUMN DETAILS

GUIDON DESIGN, INC.

1221 N. PENNSYLVANIA STREET

SUITE 200

INDIANAPOLIS, INDIANA 46202

317.800.6388

HEAPY

1400 W. DOROTHY LANE

DAYTON, OH 45409

937.224.0861

RELOCATE NUCLEAR MEDICINE

SERVICES FOR CENTRALIZED

SERVICES

INDIANAPOLIS, IN

583-20-102

01/28/2021 JF EB

BID DOCUMENTS

FULLY SPRINKLERED

SCALE: 3/4" = 1'-0"AE502

1 D/6 COLUMN DETAIL

SCALE: 3/4" = 1'-0"AE502

2 E/5 COLUMN DETAIL

SCALE: 3/4" = 1'-0"AE502

3 2 COLUMN DETAIL

SCALE: 3/4" = 1'-0"AE502

4 C/2 COLUMN DETAIL

SCALE: 3/4" = 1'-0"AE502

5 A/3.5 COLUMN DETAIL

SCALE: 3/4" = 1'-0"AE502

6 B/4 COLUMN DETAIL

SCALE: 3/4" = 1'-0"AE502

7 C/6 COLUMN DETAIL

SCALE: 3/4" = 1'-0"AE502

8 D.5/4 COLUMN DETAIL

F1 F2

AS SCHEDULED 2" TYP

A

S S

C H

E D

U

LE

D

AS SCHEDULED 2" TYP2"

A S

S C

H E

D U

LE

D

D1 D2

A S

S C

H E

D U

LE

D

AS SCHEDULED AS SCHEDULED

A S

S C

H E

D U

LE

D

4"6"

A F

F

3'

2'

D5

A S

S C

H E

D U

LE

D

4"6"

A F

F

3'

2'

D3

AS SCHEDULED

A S

S C

H E

D U

LE

D

4"6"

A F

F

3'

2'

AS SCHEDULED

DOOR AND FRAME TYPE LEGEND

F3

AS SCHEDULED 2" TYP

RESCUE FRAME

A S

S C

H E

D U

LE

D

AE601

TYP

AE601

TYP

AE601

TYP

AE601

TYP

AE601

TYP

AE601

TYP

LEAD LINED

FRAME

D4

AS SCHEDULED

A S

S C

H E

D U

LE

D

8"

8"

6" 6" 6" 6"

F4

AS SCHEDULED

A S

S C

H E

D U

LE

D

SLIDING DOOR

RE: 4/AS601

D6

AS SCHEDULED

A S

S C

H E

D U

LE

D

F5

AS SCHEDULED

A S

S C

H E

D U

LE

D

LEAD-LINED AUTOMATIC

SLIDING DOOR

RE: 4/AS601

AE601

TYP

AE601

TYP

AE601

TYP

AE601

TYP

D7

AS SCHEDULED

A S

S C

H E

D U

LE

D

4"6"

A F

F

3'

2'

LEAD LINED

WHERE INDICATED

PER PHYSICIST

REPORT

LEAD LINED

WHERE INDICATED

PER PHYSICIST

REPORT

LEAD LINED

WHERE INDICATED

PER PHYSICIST

REPORT

D8

AS SCHEDULED

A S

S C

H E

D U

LE

D

2'-0"

LEAD LINED

WHERE INDICATED

PER PHYSICIST

REPORT

D9

A S

S C

H E

D U

LE

D

AS SCHEDULED

HOLLOW METAL FRAME,

RE: DOOR SCHEDULE

WALL TYPE,

RE: PLANS

DOUBLE STUD FRAMING

SEALANT BOTH SIDES

HOLLOW METAL FRAME,

RE: DOOR SCHEDULE

WALL TYPE,

RE: PLANS

FRAMING STUD HEADER

SEALANT BOTH SIDES

2"

DOUBLE STUD FRAMING

RESCUE DOOR HOLLOW METAL

DOOR FRAME

SEALANT - BOTH SIDES

WALL TYPE,

RE: PLANS

RESCUE STOP/ STRIKE

TOILET SIDE

CORRIDOR SIDE

2"

FRAMING STUD HEADER

RESCUE DOOR HOLLOW METAL

DOOR FRAME

SEALANT - BOTH SIDES

WALL TYPE,

RE: PLANS

RESCUE STOP/ STRIKE

CORRIDOR

SIDE

TOILET

SIDE

DOOR SCHEDULE COMMENTS

GLAZING LEGEND

1. PROVIDE CARD READER AND ASSOCIATED ACCESSORIES. PROVIDE

ELECTRIC STRIKE AND OR ELECTRIFIED EXIT DEVICES AS REQUIRED. REFER

TO ENGINEERING DRAWINGS AND ACCESS CONTROL FOR ADDITIONAL

INFORMATION.

2. PROVIDE AUTOMATIC DOOR OPENER AND ASSOCIATED ACCESSORIES.

REFER TO ENGINEERING DRAWINGS FOR ADDITIONAL INFORMATION.

3. PROVIDE INDICATOR LIGHT OUTSIDE OF TOILET ROOM. REFER TO

ENGINEERING DRAWINGS.

4. PROVIDE ANTI-BARRIER HARDWARE AND FRAME TO COORDINATE WITH

INDICATOR LIGHT IN EXAM ROOMS. REFER TO ENGINEERING DRAWINGS.

5. PROVIDE ALARMED, DELAYED EGRESS ACCESS CONTROL. REFER TO

ENGINEERING DRAWINGS.

6. PROVIDE HOLD OPENS.

7. TEMPORARY INFECTION CONTROL DOOR. SEE SHEET GC101 AND

SPECIFICATIONS FOR MORE INFORMATION. DOOR, HARDWARE AND CARD

READER ARE THE RESPONSIBILITY OF THE CONTRACTOR.

8. REFER TO DETAILS INDICATED FOR HEAD, JAMB, AND SILL CONDITIONS.

VERIFY ALL DIMENSIONS IN FIELD PRIOR TO FABRICATION.

9. PROVIDE LEAD LINED DOOR, FRAME & HARDWARE PER PHYSICIST REPORT.

DOOR AND FRAME GENERAL NOTES

A. CONTRACTOR IS RESPONSIBLE TO COORDINATE ALL HARDWARE

REQUIREMENTS BETWEEN HARDWARE SUPPLIERS, ACCESS CONTROL

VENDOR, AND ANY OTHER TRADES THAT ARE INVOLVED. REQUIRED POWER

SUPPLIES AND AUXILIARY COMPONENTS THAT IS REQUIRED TO MAKE A

HARDWARE SET WORK BETWEEN ACCESS CONTROL AND HARDWARE SHALL

BE PROVIDED.

B. ALL HOLLOW METAL DOORS AND FRAMES SHALL BE FACTORY PRIMED AND

RECEIVE PAINT FINISH ON SITE. REFER TO FINISH SCHEDULE FOR HOLLOW

METAL PAINT COLOR.

C. PROVIDE SEALANT AROUND PERIMETER OF DOOR/ WINDOW FRAME

ADJACENT TO WALL AND FLOOR CONSTRUCTION, BOTH SIDES.

D. LOCATE HINGE SIDE OF DOOR 6" OFF OF ADJACENT WALL UNLESS

OTHERWISE INDICATED ON ARCHITECTURAL PLAN AE102.

E. MAINTAIN ADA/ABA CLEARANCE AROUND DOORS AND REFER ANY

DISCREPANCIES TO COR AND ARCHITECT FOR CLARIFICATION.

F. RATED FRAMES, DOORS AND GLAZING SHALL BE UL LABELED IN A VISIBLE

LOCATION.

G. REFER TO DOOR/ WINDOW AND FRAME LEGEND FOR REFERENCE TO

TYPICAL HEAD, JAMB AND SILL DETAILS.

H. ALL HOLLOW METAL JAMBS TO BE PREPARED FOR ACCESS CONTROL AND

SECURITY HARDWARE. REFER TO ENGINEERING DRAWINGS FOR

ADDITIONAL INFORMATION.

I. REFER TO SPECIFICATIONS AND ENGINEERING DRAWINGS FOR ADDITIONAL

HARDWARE INFORMATION.

J. COORDINATE POWER AND DATA REQUIREMENTS AT DESIGNATED DOOR

OPENINGS WITH ENGINEERING DRAWINGS.

CLEAR FULLY TEMPERED FLOAT GLASS

CLEAR LEADED GLASS PER PHYSICIST REPORT

CLEAR LAMINATED GLASS

GL1

GL2

GL3

LEAD LINED HOLLOW

METAL FRAME,

RE: DOOR SCHEDULE AND

PHYSICIST REPORT

WALL TYPE, RE: PLANS

STEEL STUD BOX BEAM HEADER

SEALANT BOTH SIDES

5"

LEAD LINING IN WALL,

RE: PHYSICIST REPORT

LEAD LINED DOOR,

RE: DOOR SCHEDULE &

PHYSICIST REPORT

ROOM SIDE

LEAD LINED HOLLOW METAL FRAME,

RE: DOOR SCHEDULE AND

PHYSICIST REPORT

WALL TYPE, RE: PLANS

DOUBLE STUD JAMB FRAMING

SEALANT BOTH SIDES

R E

: W A

LL

T

Y P

E

V A

R

IE

S

LEAD LINED DOOR,

RE: DOOR SCHEDULE & PHYSICIST

REPORT

LEAD LINING IN WALL,

RE: PHYSICIST REPORT

ROOM SIDE

BOTTOM TRACK

SILL GUIDE

ANCHOR

6" BOX HEADER

EPDM GASKET BOTH

SIDES

OUTSIDE

CLOSER AND

CARRIAGE TRACK

HANGERS

FLOOR

WOOD DOOR,

RE: DOOR SCHEDULE

ACOUSTIC BOTTOM SEAL

ROOM SIDE

SLOPED TOP

EPDM GASKET BOTH SIDES

MIN. 16 GA STUD AT JAMBS &

16" O.C. FOR ENTIRE LENGTH

OF VALANCE

EDGE GASKET

SILL GUIDE

RE: DOOR SCHEDULE

DOOR LEAF

ROOM SIDE

OUTSIDE

FRAME OPENING

VALANCE AND END CAP ABOVE

STILE SUPPORT BRACKET

STILE POCKET

BACK TO BACK 1" LADDER PULLS

94" TOTAL VALANCE WIDTH - ON PATIENT SIDE DOORS

82" TOTAL VALANCE WIDTH - ON STAFF SIDE DOORS

AE601

TYP

AE601

TYP

MIN. 16 GA. BOX HEADER

FIRE RETARDENT TREATED BLOCKING

MIN. 16 GA STUDS AT JAMBS & AT 16"

O.C. FOR ENTIRE LENGTH OF VALANCE

EXISTING FLOOR SLAB ABOVE

EXISTING FLOOR

MIN. 16 GA STUDS AT DOOR OPENING WIDTH

R E

: D O

O R

S C

H E

D U

LE

D O

O R

O P

E N

IN

G

RE: DOOR SCHEDULE

DOOR OPENING

A

B

A

B

BRUSHED STAINLESS

HEADER COVER

INTERLOCK RAIL

DOOR OPERATOR

LEAD LINED DOOR

PER PHYSICIST

REPORT

FINISH FLOOR

WALL AS INDICATED RE:

PLANS AND WALL

TYPES. LEAD LINING

PER PHYSICIST REPORT

3/ 4"

1/

4"

FLOOR GUIDE

3/ 8"

D O

O R

O P

E N

IN

G

8"

LEAD LINED

INTERLOCK RAIL PER

PHYSICIST REPORT

SLOPED TOP

6"

W A

LL

O

P E

N

IN

G

6" 18GA BOX HEADER

LEAD LINING ON

INTERLOCK CHANNEL

AND JAMB PER

PHYSICIST REPORT

WALL / CLEAR SLIDE OPENING

TOP OF

FLOOR

GUIDE

WALL AS INDICATED RE:

PLANS AND WALL

TYPES. LEAD LINING

PER PHYSICIST REPORT

WALL AS INDICATED RE:

PLANS AND WALL

TYPES. LEAD LINING

PER PHYSICIST REPORT

RADIATION SIDE

JA

M

B

4"

JA

M

B

4"

OVERLAP

2 1/2"

OVERLAP

2 1/2"

JAMB

1 3/4"

LEAD LINED DOOR

PER PHYSICIST

REPORT

1/

2"SAFETY BEAM

JAMB

1 3/4"

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Office of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ARCHITECT/ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Consultant:

C

U s e rs \d c h ri s te n s e n \D o c u m e n ts \1

.1

.0

N u c M e d v d c h ri s te n s e n M

N F

X

.r v t

/2

/2

:5

:0

A

M

AE601

DOOR SCHEDULE

GUIDON DESIGN, INC.

1221 N. PENNSYLVANIA STREET

SUITE 200

INDIANAPOLIS, INDIANA 46202

317.800.6388

HEAPY

1400 W. DOROTHY LANE

DAYTON, OH 45409

937.224.0861

RELOCATE NUCLEAR MEDICINE

SERVICES FOR CENTRALIZED

SERVICES

INDIANAPOLIS, IN

583-20-102

01/28/2021 JLG PAE, II

BID DOCUMENTS

FULLY SPRINKLERED

DOOR SCHEDULE

DOOR

ROOM DIMENSIONS DOOR

ELEVATION

DOOR FRAME DETAIL

GLAZING

FIRE

RATING

HARDWARE

SET COMMENTSNUMBER NAME WIDTH HEIGHT THICKNESS MATERIAL FINISH TYPE MATERIAL FINISH HEAD JAMB SILL

9A CB-10 AUDITORIUM CORRIDOR 6' - 0" 7' - 0" 1 3/4" D4 WD STAIN F1 HM PT 9/AE601 8/AE601 -- GL1 -- HW-SH-8 1, 2, 6, 8

9B CB-10 AUDITORIUM CORRIDOR 3' - 8" 7' - 0" 1 3/4" D1 WD STAIN F1 HM PT 9/AE601 8/AE601 -- -- 20 Minute HW-SH-3B 1, 8

9C CB-10 AUDITORIUM CORRIDOR 6' - 0" 7' - 0" 1 3/4" D3 WD STAIN F1 HM PT 9/AE601 8/AE601 -- -- -- HW-SH-8 1, 2, 8

39 C-6a CLOSET 3' - 0" 7' - 0" 1 3/4" D1 WD STAIN F1 HM PT 9/AE601 8/AE601 -- -- -- HW-1A 8

A-B002 C-1 STAFF ENTRANCE CORRIDOR 3' - 8" 7' - 0" 1 3/4" D2 WD STAIN F2 HM PT 7/AE601 6/AE601 -- -- HW-3C 8, 9

A-B003 C-1 STAFF ENTRANCE CORRIDOR 3' - 8" 7' - 0" 1 3/4" D2 WD STAIN F2 HM PT 7/AE601 6/AE601 - -- HW-3C 8, 9

A-B004 A-B004 PET / CT 2 4' - 4 1/2" 7' - 0" 1 3/4" D6 WD STAIN F5 ALUM CLEAR 2/AE601 1/AE601 2/AE601 -- -- HW-16 2, 8, 9

A-B004a A-B004 PET / CT 2 5' - 0" 7' - 0" 1 3/4" D8 WD STAIN F2 HM PT 7/AE601 6/AE601 HW-4G

A-B004b A-B004 PET / CT 2 3' - 0" 7' - 0" 1 3/4" D1 WD STAIN F2 HM PT 7/AE601 6/AE601 -- -- HW-3B 8, 9

A-B005 A-B005 COMBINED PET CONTROL ROOM 3' - 0" 7' - 0" 1 3/4" D1 WD STAIN F2 HM PT 7/AE601 6/AE601 -- -- HW-3A 8, 9

A-B005a A-B005 COMBINED PET CONTROL ROOM 4' - 0" 7' - 0" 1 3/4" D6 WD STAIN F5 HM PT 2/AE601 1/AE601 2/AE601 - -- HW-16A 8

A-B006 A-B006 PET / CT 1 4' - 0 1/2" 7' - 0" 1 3/4" ETR (1) ETR ETR ETR ETR PT - - - - HW-16A

A-B006a A-B006a PET EQUIPMENT ROOM 5' - 0" 7' - 0" 1 3/4" D8 WD STAIN F2 HM PT 7/AE601 6/AE601 HW-4G

A-B007 A-B007 HOT LAB 3' - 0" 7' - 0" 1 3/4" D3 WD STAIN F2 HM PT 7/AE601 6/AE601 GL1 -- HW-SH-3A 1, 8, 9

A-B007a A-B007 HOT LAB 3' - 0" 7' - 0" 1 3/4" D1 WD STAIN F2 HM PT 7/AE601 6/AE601 HW-5A

A-B008 C-3 EAST CORRIDOR 4' - 0" 7' - 0" 1 3/4" ETR (1) ETR ETR ETR ETR PT - - - - HW-17 EXISTING TO

REMAIN

A-B009 C-3 EAST CORRIDOR 4' - 0" 7' - 0" 1 3/4" ETR (1) ETR ETR ETR ETR PT - - - - HW-17 EXISTING TO

REMAIN

A-B010 C-4 SOUTH CORRIDOR 4' - 0" 7' - 0" 1 3/4" ETR (1) ETR ETR ETR ETR PT - - - - HW-17 EXISTING TO

REMAIN

A-B011 A-B01 SOILED UTILITY 3' - 0" 7' - 0" 1 3/4" D1 WD STAIN F1 HM PT 9/AE601 8/AE601 -- -- 20 Minute HW-5G 8

A-B012 A-B012 DOSE ADMIN. ROOM 3' - 8" 7' - 0" 1 3/4" D9 WD STAIN F3 HM PT 7/AE601 6/AE601 -- - 20 Minute HW-3D 8, 9

A-B013 A-B013 READING & CONSULTATION 3' - 0" 7' - 0" 1 3/4" D2 WD STAIN F1 HM PT 9/AE601 8/AE601 -- - -- HW-3A 8

A-B014 A-B014 EQUIP CALIBRATION 3' - 0" 7' - 0" 1 3/4" D2 WD STAIN F1 HM PT 9/AE601 8/AE601 -- - -- HW-3A 8

A-B015 A-B015 STAFF LOUNGE 3' - 6" 7' - 0" 1 3/4" D9 WD STAIN F1 HM PT 9/AE601 8/AE601 -- - -- HW-3 8

A-B016 A-B016 NM TOILET, PATIENT 3' - 6" 7' - 0" 1 3/4" D1 WD STAIN F5 HM PT 11/AE601 10/AE601 -- -- -- HW-2B 3, 4, 8

A-B017 A-B017 NM TOILET, PATIENT 3' - 6" 7' - 0" 1 3/4" D1 WD STAIN F4 HM PT 11/AE601 10/AE601 -- -- -- HW-2B 3, 4, 8

A-B018 A-B018 SUPERVISOR OFFICE 3' - 0" 7' - 0" 1 3/4" D1 WD STAIN F1 HM PT 9/AE601 8/AE601 -- -- -- HW-SH-3 8

A-B019 A-B019 MSA / SCHEDULING 4' - 0" 7' - 0" 1 3/4" D7 WD STAIN F4 ALUM CLEAR 5/AE601 3/AE601 5/AE601 GL1 -- HW-15 8

A-B020 A-B020 OUTPATIENT WAITING 3' - 0" 7' - 0" 1 3/4" D4 WD STAIN F2 HM PT 9/AE601 8/AE601 20 Minute HW-3 1, 6, 8

A-B020a A-B020 OUTPATIENT WAITING 6' - 0" 7' - 0" 1 3/4" D4 WD STAIN F2 HM PT 9/AE601 8/AE601 -- GL1 20 Minute HW-SH-7 1, 6, 8

A-B020b C-4 SOUTH CORRIDOR 6' - 0" 7' - 0" 1 3/4" D5 WD STAIN F1 HM PT 9/AE601 8/AE601 -- -- -- HW-SH-9 1, 6, 8

A-B021 A-B021 INPATIENT HOLDING 4' - 0" 7' - 0" 1 3/4" D7 WD STAIN F2 HM PT 9/AE601 10/AE601 GL1 20 Minute HW-3 1, 6, 8

A-B022 A-B022 THYROID UPTAKE 3' - 0" 7' - 0" 1 3/4" D2 WD STAIN F2 HM PT 7/AE601 6/AE601 -- - -- HW-3A 8, 9

A-B023 C-2 NORTH CORRIDOR 4' - 0" 7' - 0" 1 3/4" D2 WD STAIN F4 HM PT 5/AE601 3/AE601 5/AE601 GL1 -- HW-15 8

A-B024 A-B024 PATIENT INTERVIEW ROOM 3' - 6" 7' - 0" 1 3/4" D1 WD STAIN F1 HM PT 9/AE601 8/AE601 -- -- -- HW-3 8

A-B026 C-2 NORTH CORRIDOR 3' - 0" 7' - 0" 1 3/4" ETR (2) ETR ETR ETR ETR PT - - - - -- HW-17 EXISTING TO

REMAIN

A-B027 C-2 NORTH CORRIDOR 3' - 0" 7' - 0" 1 3/4" ETR (2) ETR ETR ETR ETR PT - - - - -- HW-17 EXISTING TO

REMAIN

A-B028 A-B028 PET TOILET, PATIENT 3' - 0" 7' - 0" 1 3/4" ETR (2) ETR ETR ETR ETR PT - - - - -- HW-17 EXISTING TO

REMAIN

A-B029 A-B029 NM CONTROL RM 3' - 0" 7' - 0" 1 3/4" D2 WD STAIN F2 HM PT 7/AE601 6/AS601 GL2 -- HW-SH-3 8

A-B029a A-B029a NM SCAN 1 4' - 4 1/2" 7' - 0" 1 3/4" D6 WD STAIN F5 ALUM CLEAR 2/AE601 1/AE601 2/AE601 -- -- HW-16 2, 8, 9

A-B029b A-B029b NM SCAN 2 4' - 4 1/2" 7' - 0" 1 3/4" D6 WD STAIN F5 ALUM CLEAR 2/AE601 1/AE601 2/AE601 -- -- HW-16 2, 8, 9

A-B031 A-B031 CLEAN SUPPLY 3' - 6" 7' - 0" 1 3/4" D1 WD STAIN F1 HM PT 9/AE601 8/AE601 -- -- 45 Minute HW-5H 8

CB-9 CB-9 CORRIDOR 5' - 0" 7' - 0" 1 3/4" D1 WD STAIN F1 HM PT 9/AE601 8/AE601 -- -- 20 Minute HW-SH-3B 1

SCALE: 3" = 1'-0"AE601

8 DOOR JAMB DETAIL, HM

SCALE: 3" = 1'-0"AE601

9 DOOR HEAD DETAIL, HM

SCALE: 3" = 1'-0"AE601

10 DOOR HAMB DETAIL, ANTI-BARRICADE

SCALE: 3" = 1'-0"AE601

11 DOOR HEAD DETAIL, ANTI-BARRICADE

SCALE: 3" = 1'-0"AE601

7 DOOR HEAD DETAIL, LEAD LINED HM

SCALE: 3" = 1'-0"AE601

6 DOOR JAMB DETAIL, LEAD LINED HM

SCALE: 3" = 1'-0"AE601

5 TYPICAL SLIDING DOOR HEAD & SILL DETAIL

SCALE: 3" = 1'-0"AE601

3 TYPICAL SLIDING DOOR PLAN & JAMB DETAIL

SCALE: 1/2" = 1'-0"AE601

4 TYPICAL DOOR FRAMING @ SLIDING DOOR

SCALE: 3" = 1'-0"AE601

2 AUTOMATIC SLIDING DOOR HEAD & SILL DETAIL

SCALE: 3" = 1'-0"AE601

1 AUTOMATIC SLIDING DOOR PLAN & JAMB DETAIL

RADIATION SIDE

WINDOW COMMENTS

1. PAINT HM WINDOW FRAME PT-2.

2. LEAD LINING AND LEADED GLASS PER PHYSICIST REPORT.

W1

AS SCHEDULED

A S

S C

H E

D U

LE

D

3'

FIXED W2

AS SCHEDULED

A S

S C

H E

D U

LE

D

3'

FIXED

PAINTED

LEAD LINED

HM FRAME

LEADED

GLASS

PAINTED

HM

FRAME

1/4"

TEMPERED

GLASS

HOLLOW METAL FRAME,

RE: WINDOW SCHEDULE

WALL TYPE,

RE: PLANS

STEEL STUD FRAMING HEADER

SEALANT, BOTH SIDES

GLAZING PER SCHEDULE

GLAZING TAPE, BOTH SIDES

REMOVEABLE STOP

HOLLOW METAL FRAME

SEE WINDOW SCHEDULE

WALL TYPE, SEE PLANS

DOUBLE STUD FRAMING

SEALANT BOTH SIDES

GLAZING PER SCHEDULE

GLAZING TAPE, BOTH SIDES

REMOVEABLE STOP

LEAD LINED HOLLOW

METAL FRAME,

RE: WINDOW SCHEDULE

WALL TYPE,

RE: PLANS

STEEL STUD FRAMING HEADER

SEALANT, BOTH SIDES

LEADED GLASS PER

PHYSICIST REPORT

GLAZING TAPE, BOTH SIDES

REMOVEABLE STOP

LEAD LINING PER

PHYSICIST REPORT

HOLLOW METAL FRAME

SEE WINDOW SCHEDULE

WALL TYPE, SEE PLANS

DOUBLE STUD FRAMING

SEALANT, BOTH SIDES

LEADED GLASS PER PHYSICIST REPORT

GLAZING TAPE, BOTH SIDES

REMOVEABLE STOP

LEAD LINING PER PHYSICIST REPORT

LEAD LINED HOLLOW

METAL FRAME,

RE: WINDOW SCHEDULE

WALL TYPE,

RE: PLANS

STUD WALL FRAMING

SEALANT BOTH SIDES

LEADED GLASS PER

PHYSICIST REPORT

GLAZING TAPE, EACH SIDE

REMOVEABLE STOP

LEAD LINING PER

PHYSICIST REPORT

HOLLOW METAL FRAME,

RE: WINDOW SCHEDULE

SOUND ATTENUATION BATT

INSULATION WHERE INDICATED

STUD WALL FRAMING

SEALANT BOTH SIDES

GLAZING PER

SCHEDULE

GLAZING TAPE, EACH SIDE

REMOVEABLE STOP

WALL TYPE, RE: PLANS

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Office of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ARCHITECT/ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Consultant:

C

U s e rs \d c h ri s te n s e n \D o c u m e n ts \1

.1

.0

N u c M e d v d c h ri s te n s e n M

N F

X

.r v t

/2

/2

:5

:0

A

M

AE602

WINDOW SCHEDULE

GUIDON DESIGN, INC.

1221 N. PENNSYLVANIA STREET

SUITE 200

INDIANAPOLIS, INDIANA 46202

317.800.6388

HEAPY

1400 W. DOROTHY LANE

DAYTON, OH 45409

937.224.0861

RELOCATE NUCLEAR MEDICINE

SERVICES FOR CENTRALIZED

SERVICES

INDIANAPOLIS, IN

583-20-102

01/28/2021 JLG PAE, II

BID DOCUMENTS

FULLY SPRINKLERED

WINDOW SCHEDULE

TYPE

MARK TYPE MATERIAL FINISH WIDTH HEIGHT

DETAIL

COMMENTSHEAD JAMB SILL

W1 FIXED HM PT 6' - 0" 4' - 2" 1/AE602 2/AE602 3/AE602 1

W2 FIXED HM PT 6' - 0" 3' - 10" 6/AE602 5/AE602 4/AE602 2

SCALE: 3" = 1'-0"AE602

1 WINDOW HEAD DETAIL, HM

SCALE: 3" = 1'-0"AE602

2 WINDOW JAMB DETAIL, HM

SCALE: 3" = 1'-0"AE602

6 WINDOW HEAD DETAIL, LEAD LINED HM

SCALE: 3" = 1'-0"AE602

5 WINDOW JAMB DETAIL, LEAD LINED HM

SCALE: 3" = 1'-0"AE602

4 WINDOW SILL DETAIL, LEAD LINED HM

SCALE: 3" = 1'-0"AE602

3 WINDOW SILL DETAIL, HM

FINISHES GENERAL NOTES

A. COORDINATE CONSTRUCTION WITH OWNER FURNISHED AND CONTRACTOR

PROVIDED EQUIPMENT. REFER TO EQUIPMENT SCHEDULE AND EQUIPMENT

PLAN FOR ADDITIONAL INFORMATION. FURNTIURE AND EQUIPMENT SHOWN

FOR REFERENCE ONLY.

B. COORIDNATE WITH ENGINEERING DRAWINGS. REFER ANY DISCREPANCIES

TO COR AND ARCHITECT.

C. REFER TO PLUMBING ENGINEERING DRAWINGS FOR ADDITIONAL

COORDINATION INFORMATION ON SINKS.

D. PROVIDE BLOCKING AS REQUIRED. REFER TO SPECIFICATIONS FOR

ADDITIONAL INFORMATION.

E. REFER TO TOILET ACCESSORY SCHEDULE FOR INFORMATION REGARDING

ACCESSORIES. PROVIDE SOAP AND PAPER TOWEL DISPENSERS AT EACH

SINK WITHIN PROJECT BOUNDARY.

F. REFER TO SPECIFICATIONS FOR CASEWORK HARDWARE.

G. REFER TO FINISH SCHEDULE ON SHEET AI601 FOR ADDITIONAL

INFORMATION.

H. REFER TO INTERIOR ELEVATION AND FINISHES SHEET AI200 SERIES FOR

ADDITIONAL NOTES AND LEGENDS.

I. PROVIDE WORK SURFACE SUPPORT BRACKETS AS NECESSARY WITH

SPACING RECOMMENDED BY MANUFACTURER. COORDNATE SUPPORT

LOCATIONS WITH UNDERCOUNTER EQUIPMENT AND FURNITURE.

J. PROVIDE FILLER PANELS BETWEEN WALLS AND CABINETS, NOT TO EXCEED

3". VERIFY DIMENSIONS IN THE FIELD.

K. PROVIDE PLASTIC LAMINATE ON ALL EXPOSED SURFACES, EDGES AND ENDS

OF WALL MOUNTED AND BASE CABINETS. INTERIORS SHALL BE WHITE

MELAMINE. PROVIDE FURNITURE FINISHED ENDS ON ALL EXPOSED SIDES OF

CASEWORK.

L. WALLS AND INFILL WALL CONSTRUCTION AND FINISHES SHALL BE PROVIDED

FROM WALL CORNER TO WALL CORNER.

M. REFER TO SHEET AI101 AND AI200 SERIES SHEETS FOR INFORMATION

REGARDING ACCENT PAINT LOCATIONS.

N. REFER TO SHEET AI102 FOR FLOORING PATTERNS.

O. PROVIDE P-1 ON ALL WALLS UNLESS OTHERWISE NOTED.

P. PROVIDE P-2 ON HOLLOW METAL DOOR FRAMES AND METAL DOORS, IN SEMI-

GLOSS FINISH, UNLESS OTHER WISE NOTED.

Q. PATCH AND/OR REPAIR, FINISH AND PAINT ALL WALLS IN AREA OF WORK

BEING DISTRUBED. BRING AREA BACK TO NEW LIKE CONDITIONS.

R. REFER TO INTERIOR ELEVATIONS SHEETS AI200 SERIES FOR CERAMIC WALL

TILE PATTERN IN STAFF BREAK ROOM. IN RESTROOMS PAINT WALLS ABOVE

CORIAN SHEETS AS IDENTIFIED IN FINISH SCHEDULE AND ON ELEVATIONS.

S. PROVIDE TS-1 FINISH END TRIM ON RIGID VINYL WALCOVERING. INSTALL RVW

WITH BUTT END INSTALLATION.

T. INSTALL CORNER GUARDS MOUNTED ABOVE INTERGERAL BASE UP TO VA

STANDARD HEIGHT.

U. INSTALL WALL BASE OVER RIGID VINYL WALLCOVERING. SEE DETAIL SHEET

AI501.

#FINISHES PLAN NOTES

1. PROVIDE CORNER GUARD CG-1. MOUNT ABOVE WALL BASE UP TO VA

STANDARD HEIGHT.

2. PROVIDE END WALL GUARD CG-2. MOUNT ABOVE WALL BASE UP TO VA

STANDARD HEIGHT.

3. APPLY SS-4 FULL HEIGHT ON THESE WALLS.

4. CURTAIN TRACK ABOVE.

5. PAINT ALL SIDES OF COLUMN P-3.

6. PAINT ALL SIDES OF COLUMN P-3 AND RWC-2. SEE DETAIL 3/AI501

7. PAINT ALL WALLS IN CORRIDOR P-1 UNLESS OTHERWISE NOTED.

8. ALL WALLS IN THIS ROOM WILL HAVE RWC-1 UNLESS OTHERWISE NOTED.

SEE LEGEND FOR DETAIL.

RIGID VINYL WALL PROTECTION LEGEND

PROVIDE HANDRAIL WITH CRASH RAIL AND IMPACT RESISTANT

WALLCOVERING RWC THROUGHOUT CORRIDOR AS INDICATD. SEE

DETAIL 1/AI501. WHERE HANDRAIL AND BUMPER RAIL ARE NOT

PRESENT IMPACT RESISTANT WALLCOVERING WILL CONTINUE @

42" A.F.F. WITH TS-1 AS TOP TRIM. SEE DETAIL 3/AI501.

PROVIDE DOUBLE CRASH RAIL WITH IMPACT RESISTANT

WALLCOVERING AS INDICATD. CRASH RAIL TO BE STAINLESS

STEEL AND WILL HAVE RWC-1 UNLESS OTHERWISE NOTED. SEE

DETAIL 2/AI501. WHERE CRASH RAILS ARE NOT PRESENT IMPACT

RESISTANT WALLCOVERING WILL CONTINUE @ 42" A.F.F. WITH

TS-1 AS TOP TRIM. SEE DETAIL 3/AI501.

PROVIDE RIGID VINYL WALL PROTECTION RWC FROM FLOOR TO

42"A.F.F. WITH TS-1 AS TOP TRIM. SEE DETAILED 3/AI501.

A

457 3.5

B

C

D

E

F

2.58910

DOSE ADMIN. ROOM

A-B003

PET EQUIPMENT

ROOM

A-B004a

PET / CT 2

A-B004

PET EQUIPMENT

ROOM

A-B006a

PET / CT 1

A-B006

HOT LAB

A-B007

DOSE ADMIN. ROOM

A-B008

DOSE ADMIN. ROOM

A-B009

DOSE ADMIN. ROOM

A-B010

READING &

CONSULTATION

A-B013

DOSE ADMIN. ROOM

A-B012

EQUIP CALIBRATION

A-B014

SOILED UTILITY

A-B01

NM SCAN 2

A-B029b

NM SCAN 1

A-B029a

NM CONTROL RM

A-B029

NM TOILET, PATIENT

A-B016

NM TOILET, PATIENT

A-B017

SUPERVISOR

OFFICE

A-B018

CLEAN SUPPLY

A-B031

THYROID UPTAKE

A-B022

CARDIAC STRESS

TESTING

A-B023

PATIENT INTERVIEW

ROOM

A-B024

INPATIENT HOLDING

A-B021

OUTPATIENT

WAITING

A-B020

DIGI ROD, MOBILE

EQUIPMENT

A-B030

CORRIDOR

CB-9

AUDITORIUM

CORRIDOR

CB-10

NORTH CORRIDOR

C-2

AI203

AI203

AI203

AI203

AI203

D.5

AI201

AI201

AI201

AI201

AI202

AI202

AI201

CORRIDOR

C-6

SOUTH CORRIDOR

C-4

WEST CORRIDOR

C-5

EAST CORRIDOR

C-3

ALCOVE

C-5A

STAFF ENTRANCE

CORRIDOR

C-1

2 2

1 1

STAFF LOUNGE

A-B015

P-4

P-4

P-4

AI201

AI201

AI201

AI201

AI201

AI201

AI202

COMBINED PET

CONTROL ROOM

A-B005

TOILET, STAFF

A-B026

TOILET, STAFF

A-B027

PET TOILET,

PATIENT

A-B028

P-4

P-4 P-4 P-4

P-4

P-4

2092 SF

AUDITORIUM

A-B066

1741 SF

OFFICE

A-B040

153 SF

QCIS

A-B039

234 SF

TEAM ROOM

A-B038

105 SF

PATIENT SAFETY

MANAGER

A-B043

270 SF

ASSISTANT LOG

A-B041

152 SF

CHIEF LOG.

A-B042

84 SF

NOURISHMENT

A-B047

68 SF

INFECTION

CONTROL MANAGER

A-B046

118 SF

CHIEF QUALITY

MANAGEMENT

A-B045

76 SF

RISK MANAGER

A-B044

P-4

P-4

DOSE ADMIN. ROOM

A-B002

P-4

P-4

P-4

P-4

RWC-3

RWC-3

8 8

7 8

7 8

7 8

7 8 7 8

7 8

7 8

7 8

RWC-3

P-4

RWC-3

RWC-3

RWC-3

P-4

RWC-3

RWC-3RWC-3RWC-3

RWC-3

MSA / SCHEDULING

A-B019

N

KEY PLAN

PROJECT

LOCATION

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Office of Construction and Facilities Management

U.S. Department of Veterans Affairs

B

C

D

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

ARCHITECT/ENGINEER OF RECORD

A/E:

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Consultant:

C

U s e rs \d c h ri s te n s e n \D o c u m e n ts \1

.1

.0

N u c M e d v d c h ri s te n s e n M

N F

X

.r v t

/2

/2

:5

:1

A

M

AI101

FINISH TREATMENT PLAN

GUIDON DESIGN, INC.

1221 N. PENNSYLVANIA STREET

SUITE 200

INDIANAPOLIS, INDIANA 46202

317.800.6388

HEAPY

1400 W. DOROTHY LANE

DAYTON, OH 45409

937.224.0861

RELOCATE NUCLEAR MEDICINE

SERVICES FOR CENTRALIZED

SERVICES

INDIANAPOLIS, IN

583-20-102

01/28/2021 JLG LF

BID DOCUMENTS

FULLY SPRINKLERED

SCALE: 1/8" = 1'-0"AI101

2 FINISH TREATMENT PLAN

File details come from the government source that posted it. Updated .