This document package includes architectural plans and drawings for a federal construction contract opportunity to relocate nuclear medicine services. Key details include renovating and reconfiguring existing spaces on multiple floors to accommodate new PET/CT and nuclear medicine scanner equipment, support spaces including control rooms, dosing rooms, waiting areas, and staff and patient facilities. The project also involves installing new mechanical, electrical and plumbing systems to support the medical equipment. The solicitation number is listed as 36C25021B0036 with a project title to renovate nuclear medicine services at the Richard L. Roudebush VA Medical Facility in Indianapolis, Indiana. The contract is through the Department of Veterans Affairs Veterans Health Administration.
View the file
Other files
Show 40 more
Latest 50 of 66 listed here. See the rest on GovTribe.
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
270 SF
A-B005
COMBINED PET
CONTROL ROOM
NMCR1
316 SF
A-B007
HOT LAB
NMRP1
114 SF
A-B008
DOSE ADMIN. ROOM
NMIR1
114 SF
A-B009
DOSE ADMIN. ROOM
NMIR1
145 SF
A-B010
DOSE ADMIN. ROOM
NMIR1
520 SF
A-B013
READING &
CONSULTATION
XVC01
106 SF
A-B012
DOSE ADMIN. ROOM
NMIR1
235 SF
A-B014
EQUIP CALIBRATION
NMDC1
92 SF
A-B01
SOILED UTILITY
USCL1
65 SF
A-B026
TOILET, STAFF
TNPG1
59 SF
A-B027
TOILET, STAFF
TNPG1
64 SF
A-B028
PET TOILET,
PATIENT
TPG01
452 SF
A-B029a
NM SCAN 1
NMGS1
338 SF
A-B029
NM CONTROL RM
NMCR1
69 SF
A-B016
NM TOILET, PATIENT
TPG01
60 SF
A-B017
NM TOILET, PATIENT
TPG01
130 SF
A-B018
SUPERVISOR
OFFICE
OFA09
760 SF
A-B020
OUTPATIENT
WAITING
WTG06
1190 SF
CORRIDOR
CB-9
717 SF
AUDITORIUM
CORRIDOR
CB-10
A
457 3.5
B
C
D
E
F
2.58910
CLR
7'-2"
17 SF
A-B025
ELEC ROOM
CLR
9'-8 1/2"
'-7
/8
6'
-2
/8
'-4
/8
C
LR
8' -0
7'
-1
3/ 8"
CLR
8'-0"
'-9
/2
'-1
3/ 4"
D.5
C
LR
5' -0
'-9
/8
9'
-0
2'-2" 8'-5 3/8" 28'-11 1/8"
CLR
8'-0"
8'
/8
112 SF
A-B004a
PET EQUIPMENT
ROOM
XMRC2
464 SF
A-B004
PET / CT 2
NMSS1
CLR
8' - 0"
150 SF
A-B002
DOSE ADMIN. ROOM
NMRI1
150 SF
A-B003
DOSE ADMIN. ROOM
NMRI1
126 SF
A-B006a
PET EQUIPMENT
ROOM
XMRC2
136 SF
A-B024
PATIENT INTERVIEW
ROOM
PAIA1
457 SF
A-B023
CARDIAC STRESS
TESTING
OPPE2
202 SF
A-B021
INPATIENT HOLDING
ORPP1
123 SF
A-B022
THYROID UPTAKE
NMUR1
32 SF
C-5A
ALCOVE
SB001
150 SF
A-B030
DIGI ROD, MOBILE
EQUIPMENT
XRM01
136 SF
A-B031
CLEAN SUPPLY
UCCL1
456 SF
A-B029b
NM SCAN 2
NMGS1
8 SF
CLOSET
C-6a
CLR
5' - 11"
/2
7'
/8
111213
'-0
/8
9'
-9
17'-1 3/8" 21'-1 5/8" 9'-8 1/2"
'-1
/4
8'
-1
7/ 8"
CLR
8'-0"
C
LR
3' -6
9'-2 5/8" 25'-8 3/4"
CLR
8'-0" 10 3/8"
S3A.Gi
S3A.Gi
S3A.Gi
S3A.Gi
S3A.Gi
S3A.Gi
S3A.Gi
S3A.Gi
S3A.Gi
S3A.Gi
S3A.Gi
S3A.Gi
S3A.Gi
S3A.Gi
S3A.Gi
S3A.Gi
S3A.Gi
S3A.Gi
S3A.1F
S3A.1F
S3A.1F
S3A.Gi
S6A.Gi
S6A.GiL
S6A.GiL
S6A.GiL
S6A.GiL
S6A.GiL S6A.GiL
S6A.GiL
S6A.GiL
S3A.Gi
S6A.GiL
S6A.GiL
S6A.GiL
S3A.Gi
S3A.Gi
S3A.Gi
S3A.Gi
215 SF
A-B019
MSA / SCHEDULING
RCP02
S6A.GiL
S6A.GiL
S3B.SP C
LR
8' -0
S3A.Gi
S3A.Gi
11'-9 1/4" 11'-9 1/4" 10'-2 7/8"
3' -1
1/
4" 3'
-0
/8
CLR
8'-0"
F0D.G
1'-6"
1 1
CLR
6'-9 1/2"
S3A.Gi
488 SF
A-B006
PET / CT 1
NMSS1
13'-9 7/8" 9'-8 3/4"
8' -9
/8
10'-3 1/8"
19'-4 5/8"
16'-0 3/8"
R .O
4' -4
/2
4'-8 3/8"
R.O.
4'-4 1/2"
2'-6"
R.O.
6'-0" 6'-10"
R.O.
4'-4 1/2"
401 SF
A-B015
STAFF LOUNGE
SL001
BUILDING EXIT
STAIR
#9B
PUBLIC
TOILET,
MEN'S
A-B156
PUBLIC
TOILET,
WOMEN'S
A-B157
ELEV.
P-15
ELEV.
P-16
ELEV.
P-14
ELEV.
P-13
ELEV.
S-11
ELEV.
S-12
FUTURE
S-13
LOBBY
CB-11
LOBBY
CB-10
UP
ELEC.
A-B153
MECHANICAL
ROOM
A-B150
UP
CORRIDOR
CB-7
4' -4
/8
'-6
/4
31 SF
1A
PATIENT
EDUCATION KIOSK
CLSC1
74 SF
3A
HOLDING BAY 1
ORPP1
74 SF
3B
HOLDING BAY 2
ORPP1
102 SF
RADIOACTIVE
WASTE STORAGE
A-B007a
S6A.GiL
F3D.G
S6A.GiL
16'-10 3/8"
71 SF
3C
HOLDING BAY 3
ORPP1
6' -8
/4
1' -7
/4
R .O
4' -0
6"
5'-11 5/8"
1' -0
/8
3' -6
3'
-0
R .O
6' -0
6'
-6
/8
R .O
6' -0
3'
-1
3' -6
3/
4"
11'-7 1/2" 11'-7 1/2"
3' -0
1'-1 1/8"
5'-4 7/8" 13'-7 1/2"
10 1/2"
C
LR
3' -6
'-7
/2
AI203
AI203
S3A.G
3' -0
S6A.GiL
S6A.GiL
S6A.GiL
S6A.GiL
S6A.GiL
7'-10 1/4" 13'-3 3/8"
4' -4
/8
S3A.G
S3A.Gi
S3A.Gi
S3A.Gi
S3A.Gi
F3D.G
S6A.GiL
S6A.GiL
S3A.G S3A.G
S6A.Gi
16'-11 3/8"
S3A.Gi
S3A.Gi
F6A.GiL
S6A.GiL
'-6
/8
9'
-0
21'-2"
2'-2 1/2"
3' -0
/4
9'
-0
21'-2"
2'-3 1/2"
2092 SF
AUDITORIUM
A-B066
1741 SF
OFFICE
A-B040
153 SF
QCIS
A-B039
234 SF
TEAM ROOM
A-B038
105 SF
PATIENT SAFETY
MANAGER
A-B043
270 SF
ASSISTANT LOG
A-B041152 SF
CHIEF LOG.
A-B042
84 SF
NOURISHMENT
A-B047
68 SF
INFECTION
CONTROL MANAGER
A-B046
118 SF
CHIEF QUALITY
MANAGEMENT
A-B045
76 SF
RISK MANAGER
A-B044
EQ EQ
EQ EQ
EQ EQ
1 1
MECHANICAL ROOM
A-B048
MIN
3'-0"
NEW PAD TO BE 6" LARGER THAN NEW AHU
VERIFY IN FIELD
N E
W P
A D
T O
B E
LA
R
G E
R T
H A
N N
E W
A H
U
V E
R
IF
Y
IN
F
IE
LD
324 SF
EAST CORRIDOR
C-3
897 SF
NORTH CORRIDOR
C-2
413 SF
WEST CORRIDOR
C-5
509 SF
SOUTH CORRIDOR
C-4
238 SF
STAFF ENTRANCE
CORRIDOR
C-1
N
KEY PLAN
PROJECT
LOCATION
DIMENSION FLOOR PLAN GENERAL NOTES
DIMENSION FLOOR PLAN NOTES
1. ALIGN FINISH FACE OF WALL WITH OUTER FINISH FACE OF EXISTING WALL.
PROVIDE SEAMLESS TRANSITION AND NEW FINISH. ALIGN FINISH FACE OF
WALL, BOTH SIDES WHERE APPLICABLE.
2. INFILL EXISTING WALL WITH WALL OF SAME HEIGHT, WIDTH, FIRE OR SMOKE
RATING, ETC. ALIGN FINISH FACE OF NEW WALL WITH EXISTING, BOTH
SIDES.PROVIDE SMOOTH TRANSITION. PREP FOR NEW FINISHES.
3. LINE OF NEW SLAB ON GRADE.
4. 1/16" LEAD WALL / DOOR SHIELDING, REFER TO WALL TYPE
5. 1/8" LEAD WALL / DOOR SHIELDING, REFER TO WALL TYPE
6. 1/4" LEAD WALL SHIELDING, REFER TO WALL TYPE
7. 1/2" LEAD WALL SHIELDING, REFER TO WALL TYPE
8. 1/32" LEAD WALL SHIELDING, REFER TO WALL TYPE
9. RAISED 6" CONCRETE PAD, REFER TO MECHANICAL DRAWINGS.
CONTRACTOR TO VERIFY ALL EXISTING AND NEW DIMENSIONS.
10. PET/CT EQUIPMENT INSTALLED BY VENDOR - REFER TO SD300 SERIES
SHEETS FOR VENDOR INFORMATION.
11. PET/CT EQUIPMENT INSTALLED BY VENDOR - REFER TO SD400 SERIES
SHEETS FOR VENDOR INFORMATION.
12. NUC MED SCANNER EQUIPMENT INSTALLED BY VENDOR - REFER TO SD100
SERIES SHEETS FOR VENDOR INFORMATION.
13. NUC MED SCANNER EQUIPMENT RELOCATED AND INSTALLED BY VENDOR -
REFER TO SD200 SERIES SHEETS FOR VENDOR INFORMATION.
A. ALL DIMENSIONS SHOWN ARE TO FACE OF WALL FRAMING/ MASONRY
UNLESS NOTED OTHERWISE. DIMENSIONS DESIGNATES AS "CLR" OR
"CLEAR" INDICATE A CLEAR DIMENSION FROM FACE OF FINISH TO FACE OF
FINISH.
B. VERIFY EXISTING FIELD CONDITIONS PRIOR TO BEGINNING WORK, AND
REPORT DISCREPANCIES IMMEDIATELY TO ARCHITECT FOR CLARIFICATION.
C. CONTRACTOR TO MAINTAIN/REPAIR RATING OF EXISTING PARTITIONS AS
AFFECTED BY DEMOLITION/NEW CONSTRUCTION, TYPICAL THROUGHOUT.
D. PROVIDE CONTROL JOINT ON CONTINUOUS GYPSUM BOARD WALLS
GREATER THAN 30' LONG.
E. PROVIDE CONTROL JOINTS AT CORNERS OF DOORS OR WINDOWS.
F. WHERE NO DIMENSION IS INDICATED, FINISH DOORS 6" FROM EDGE OF
STUD FRAMING AT CORNERS UNLESS OTHERWISE NOTED.
G. ALL NEW PARTITIONS ARE WALL TYPE S3A.G UNLESS NOTED OTHERWISE.
REFER TO WALL TYPE GENERAL NOTES AE002 FOR WALL BOARD MATERIAL.
H. NON-RATED WALL CONSTRUCTION THAT FORMS CHASE AREAS DOES NOT
REQUIRE GYPSUM BOARD INSTALLATION ON INTERIOR OF CHASE WALL.
REFER TO WALL TYPES F0D.G, F1D.G, AND F3D.G.
I. COLUMN DIMENSIONS ARE FOR REFERENCE AND SHALL BE VERIFIED IN THE
FIELD.
J. REFER TO ENLARGED FLOOR PLANS FOR ADDITIONAL DIMENSIONS FOR
TOILET AND BATH ROOMS.
K. DIMENSIONS INDICATED AS 'VIF' MAY INCREASE OR DECREASE TO MATCH
FIELD CONDITIONS. ALL OTHER DIMENSIONS WITHIN THE DIMENSION STRING
MUST BE MAINTAINED.
L. PROVIDE BLOCKING AND STRAPPING FOR WALL MOUNTED CASEWORK,
FURNITURE, AND ACCESSORIES. COORDINATE WITH FLOOR PLANS,
SCHEDULES, ELEVATIONS, AND ENGINEERING DRAWINGS FOR FULL SCOPE
OF WORK.
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
Office of Construction and Facilities Management
U.S. Department of Veterans Affairs
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
Phase
ARCHITECT/ENGINEER OF RECORD
A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
Consultant:
C
U s e rs \d c h ri s te n s e n \D o c u m e n ts \1
.1
.0
N u c M e d v d c h ri s te n s e n M
N F
X
.r v t
/2
/2
:5
:2
A
M
AE101
LOWER LEVEL DIMENSION PLAN
GUIDON DESIGN, INC.
1221 N. PENNSYLVANIA STREET
SUITE 200
INDIANAPOLIS, INDIANA 46202
317.800.6388
HEAPY
1400 W. DOROTHY LANE
DAYTON, OH 45409
937.224.0861
RELOCATE NUCLEAR MEDICINE
SERVICES FOR CENTRALIZED
SERVICES
INDIANAPOLIS, IN
583-20-102
01/28/2021 JLG DMC
BID DOCUMENTS
FULLY SPRINKLERED
SCALE: 1/8" = 1'-0"AE101
1 DIMENSION PLAN
SPACE PLANNING SCHEDULE
ROOM
ROOM
NUMBER
ROOM
CODE SEPS SF ACTUAL SF
PATIENT EDUCATION KIOSK 1A CLSC1 30 SF 31 SF
HOLDING BAY 1 3A ORPP1 80 SF 74 SF
HOLDING BAY 2 3B ORPP1 80 SF 74 SF
HOLDING BAY 3 3C ORPP1 71 SF
SOILED UTILITY A-B01 USCL1 92 SF
DOSE ADMIN. ROOM A-B002 NMRI1 150 SF
DOSE ADMIN. ROOM A-B003 NMRI1 150 SF
PET / CT 2 A-B004 NMSS1 464 SF
PET EQUIPMENT ROOM A-B004a XMRC2 112 SF
COMBINED PET CONTROL ROOM A-B005 NMCR1 270 SF
PET / CT 1 A-B006 NMSS1 488 SF
PET EQUIPMENT ROOM A-B006a XMRC2 126 SF
HOT LAB A-B007 NMRP1 316 SF
RADIOACTIVE WASTE STORAGE A-B007a NMDC1 102 SF
DOSE ADMIN. ROOM A-B008 NMIR1 114 SF
DOSE ADMIN. ROOM A-B009 NMIR1 114 SF
DOSE ADMIN. ROOM A-B010 NMIR1 145 SF
DOSE ADMIN. ROOM A-B012 NMIR1 106 SF
READING & CONSULTATION A-B013 XVC01 520 SF
EQUIP CALIBRATION A-B014 NMDC1 235 SF
STAFF LOUNGE A-B015 SL001 401 SF
NM TOILET, PATIENT A-B016 TPG01 SS-4 UP TO 5'-0"
A.F.F.
69 SF
NM TOILET, PATIENT A-B017 TPG01 SS-4 UP TO 5'-0"
A.F.F.
60 SF
SUPERVISOR OFFICE A-B018 OFA09 130 SF
MSA / SCHEDULING A-B019 RCP02 215 SF
OUTPATIENT WAITING A-B020 WTG06 760 SF
INPATIENT HOLDING A-B021 ORPP1 202 SF
THYROID UPTAKE A-B022 NMUR1 123 SF
CARDIAC STRESS TESTING A-B023 OPPE2 457 SF
PATIENT INTERVIEW ROOM A-B024 PAIA1 136 SF
ELEC ROOM A-B025 -- 17 SF
TOILET, STAFF A-B026 TNPG1 SS-4 UP TO 5'-0"
A.F.F.
65 SF
TOILET, STAFF A-B027 TNPG1 SS-4 UP TO 5'-0"
A.F.F.
59 SF
PET TOILET, PATIENT A-B028 TPG01 SS-4 UP TO 5'-0"
A.F.F.
64 SF
NM CONTROL RM A-B029 NMCR1 338 SF
NM SCAN 1 A-B029a NMGS1 452 SF
NM SCAN 2 A-B029b NMGS1 456 SF
DIGI ROD, MOBILE EQUIPMENT A-B030 XRM01 150 SF
CLEAN SUPPLY A-B031 UCCL1 136 SF
TEAM ROOM A-B038 234 SF
QCIS A-B039 153 SF
OFFICE A-B040 1741 SF
ASSISTANT LOG A-B041 270 SF
CHIEF LOG. A-B042 152 SF
PATIENT SAFETY MANAGER A-B043 105 SF
RISK MANAGER A-B044 76 SF
CHIEF QUALITY MANAGEMENT A-B045 118 SF
INFECTION CONTROL MANAGER A-B046 68 SF
NOURISHMENT A-B047 84 SF
MECHANICAL ROOM A-B048 6264 SF
AUDITORIUM A-B066 2092 SF
STAFF ENTRANCE CORRIDOR C-1 238 SF
NORTH CORRIDOR C-2 -- 897 SF
EAST CORRIDOR C-3 324 SF
SOUTH CORRIDOR C-4 509 SF
WEST CORRIDOR C-5 413 SF
ALCOVE C-5A SB001 32 SF
CORRIDOR C-6 156 SF
CLOSET C-6a -- 8 SF
CORRIDOR CB-9 -- 1190 SF
AUDITORIUM CORRIDOR CB-10 -- 717 SF
N
Revisions: Date:
NOTE:
REFER TO SHEETS SD100-SD409 FOR PET/CT AND NUC MED
SCANNER EQUIPMENT VENDOR INFORMATION.
UP
ARCHITECTURAL GENERAL NOTES
ARCHITECTURAL PLAN NOTES #
A. REFERENCE STRUCTURAL, MECHANICAL, PLUMBING, FIRE PROTECTION
AND ELECTRICAL FOR COORDINATION ITEMS. COORDINATE AS REQUIRED
INCLUDING NECESSARY FRAMING, BLOCKING, ETC.
B. VERIFY EXISTING FIELD CONDITIONS PRIOR TO BEGINNING WORK, AND
REPORT DISCREPANCIES IMMEDIATELY TO ARCHITECT FOR CLARIFICATION.
C. PATCH EXISTING WALLS WHERE NEW WALLS MEET.
D. ALL HANDRAILS SHALL BE FABRICATED AND INSTALLED IN ACCORDANCE
WITH ALL REQUIREMENTS OF ADA/ABA GUIDELINES AND ALL CODES AND
REGULATIONS HAVING JURISDICTION, THE MOST STRINGENT SHALL
PREVAIL.
E. PROVIDE MINIMUM CLEARANCES AT ALL DOORS PER ADA/ABA
REQUIREMENTS.
F. WHERE NO DIMENSION IS INDICATED, FINISH DOORS 6" FROM EDGE OF
STUD FRAMING AT CORNERS UNLESS OTHERWISE NOTED.
G. REFER TO SHEET QH102 FOR EQUIPMENT SCHEDULE AND INFORMATION.
H. REFER TO SHEET AI201 FOR ACCESSORIES TO BE PROVIDED IN THE
TOILET/BATH ROOMS AND AT THE HAND LAVATORIES.
I. LEVEL AND INFILL FLOOR SLAB TO RECEIVE NEW FLOORING. REFER TO
SPECIFICATIONS AND FLOORING MANUFACTURE'S TOLERANCES FOR
ADDITIONAL INFORMATION.
J. PROVIDE METAL PLATE ON UNDERSIDE OF FLOOR SLAB WHERE CONCRETE
INFILL IS REQUIRED AND CHIP AWAY CONCRETE AT FLOOR LEVEL TO
CREATE A SLOPED OPENING/LIP TO PATCH PENETRATIONS AND PREVENT
CONCRETE INFILL CORE FROM FALLING THROUGH OPENING TO FLOOR
BELOW.
1. OVERHEAD CURTAIN AND TRACK. REFER TO REFLECTED CEILING PLAN FOR
MORE INFORMATION.
2. EXISTING CHASE TO REMAIN. PROTECT DURING DEMOLITION AND
CONSTRUCTION PHASES. IF ACCESS IS REQUIRED DURING CONSTRUCTION,
MAINTAIN FIRE RATING AND FINISH AS REQUIRED.
3. EXISTING 3'-6" DIA. X 3'-0" DEEP SLAB RECESS FOR LIFT PUMP TO REMAIN.
4. SHADED AREA INDICATES AREA OF NEW REINFORCED SLAB ON GRADE.
5. LOCATION FOR CREDENTIAL READER. REFER TO DOOR SCHEDULE AND
ENGINEERING DRAWINGS FOR ADDITIONAL INFORMATION.
6. LOCATION FOR AUTOMATIC DOOR OPERATOR DOOR ACTUATOR. REFER TO
DOOR SCHEDULE AND ENGINEERING DRAWINGS FOR ADDITIONAL
INFORMATION.
7. EXISTING RAISED CONCRETE PAD w/ ADDITION. REFER TO MECHANCIAL
DRAWINGS FOR ADDITIONAL INFORMATION. CONTRACTOR TO VERIFY ALL
EXISTING AND NEW DIMENSIONS.
8. PET/CT EQUIPMENT INSTALLED BY VENDOR - REFER TO SD300 SERIES
SHEETS FOR VENDOR INFORMATION.
9. PET/CT EQUIPMENT INSTALLED BY VENDOR - REFER TO SD400 SERIES
SHEETS FOR VENDOR INFORMATION.
10. NUC MED SCANNER EQUIPMENT INSTALLED BY VENDOR - REFER TO SD100
SERIES SHEETS FOR VENDOR INFORMATION.
11. NUC MED SCANNER EQUIPMENT RELOCATED AND INSTALLED BY VENDOR -
REFER TO SD200 SERIES SHEETS FOR VENDOR INFORMATION.
A
457 3.5
B
C
D
E
F
2.58910
D.5
111213
COMBINED PET
CONTROL ROOM
A-B005
HOT LAB
A-B007
DOSE ADMIN. ROOM
A-B008
DOSE ADMIN. ROOM
A-B009
DOSE ADMIN. ROOM
A-B010
READING &
CONSULTATION
A-B013
DOSE ADMIN. ROOM
A-B012
EQUIP CALIBRATION
A-B014
SOILED UTILITY
A-B01
TOILET, STAFF
A-B026
TOILET, STAFF
A-B027
PET TOILET,
PATIENT
A-B028
NM SCAN 1
A-B029a
NM CONTROL RM
A-B029
NM TOILET, PATIENT
A-B016
NM TOILET, PATIENT
A-B017
SUPERVISOR
OFFICE
A-B018
OUTPATIENT
WAITING
A-B020
CORRIDOR
CB-9
AUDITORIUM
CORRIDOR
CB-10
ELEC ROOM
A-B025
CB-9
9A
A -B
0a
A -B
A-B025
A-B022
A -B
A-B029b
A-B026
A-B027
A -B
A-B006
A -B
A -B
A -B
A -B
9B
A-B016
A-B018
A-B017
A -B
0b
9C A-B029
W1
W2
W2
A-B023
A-B019
A-B029a
A-B012
A -B
A -B
A-B011
A -B
PET EQUIPMENT
ROOM
A-B004a
PET / CT 2
A-B004
6'
/8
8' - 0"
DOSE ADMIN. ROOM
A-B002
DOSE ADMIN. ROOM
A-B003
PET EQUIPMENT
ROOM
A-B006a
PATIENT INTERVIEW
ROOM
A-B024
CARDIAC STRESS
TESTING
A-B023
202 SF
INPATIENT HOLDING
A-B021
THYROID UPTAKE
A-B022
ALCOVE
DIGI ROD, MOBILE
EQUIPMENT
A-B030
CLEAN SUPPLY
A-B031
NM SCAN 2
A-B029b
A-B002
A-B003
A-B004
A -B
4b
CLOSET
C-6a
5' - 11"
9'
/2
/8
A -B
AE501
AE501
AE501
AOCR
C R
AOCR CR
CR
AO
AO
CR
AO
W2
2 5 5 6
6 5
CR
A-B029d
A-B029c
MSA / SCHEDULING
A-B019
AE501
A -B
4a
488 SF
PET / CT 1
A-B006
102 SF
RADIOACTIVE
WASTE STORAGE
A-B007a
401 SF
STAFF LOUNGE
A-B015
A-B007a
W2
HOLDING BAY 3
HOLDING BAY 1
HOLDING BAY 2
A-B006a
NORTH CORRIDOR
C-2
WEST CORRIDOR
C-5
EAST CORRIDOR
C-3
238 SF
STAFF ENTRANCE
CORRIDOR
C-1
A-B005a
AE502
AE502
AE502
AE502
AE502
AE502
AE502
AE502
A-B020
AE501
CORRIDOR
C-6
SOUTH CORRIDOR
C-4
AUDITORIUM
A-B066
OFFICE
A-B040
QCIS
A-B039
TEAM ROOM
A-B038
PATIENT SAFETY
MANAGER
A-B043
ASSISTANT LOG
A-B041
CHIEF LOG.
A-B042
NOURISHMENT
A-B047
INFECTION
CONTROL MANAGER
A-B046
CHIEF QUALITY
MANAGEMENT
A-B045
RISK MANAGER
A-B044
AE501
AE501
AE501
AE501
9' -0
21'-2"
21'-2"
9' -0
A -B
A-B029f
CR
AOCR
CR
AO
CR
AE501
AE501
NEW PAD TO BE 6" LARGER THAN NEW AHU
VERIFY IN FIELD
N E
W P
A D
T O
B E
LA
R
G E
R T
H A
N N
E W
A H
U
V E
R
IF
Y
IN
F
IE
LD
MECHANICAL ROOM
A-B048
MIN
3'-0"
N
KEY PLAN
PROJECT
LOCATION
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
Office of Construction and Facilities Management
U.S. Department of Veterans Affairs
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
Phase
ARCHITECT/ENGINEER OF RECORD
A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
Consultant:
C
U s e rs \d c h ri s te n s e n \D o c u m e n ts \1
.1
.0
N u c M e d v d c h ri s te n s e n M
N F
X
.r v t
/2
/2
:5
:2
A
M
AE102
LOWER LEVEL ARCHITECTURAL
PLANGUIDON DESIGN, INC.
1221 N. PENNSYLVANIA STREET
SUITE 200
INDIANAPOLIS, INDIANA 46202
317.800.6388
HEAPY
1400 W. DOROTHY LANE
DAYTON, OH 45409
937.224.0861
RELOCATE NUCLEAR MEDICINE
SERVICES FOR CENTRALIZED
SERVICES
INDIANAPOLIS, IN
583-20-102
01/28/2021 JLG PAE, II
BID DOCUMENTS
FULLY SPRINKLERED
SCALE: 1/8" = 1'-0"AE102
1 PLAN DETAIL
N
Revisions: Date:
1' - 8"
2'
10" DIA. STAINLESS
STEEL WELDED DUCT,
FIRE WRAPPED, SEE
MECHANICAL
20" x 26" INSIDE
CLEAR DIMENSION
MECHANICAL
CHASE, NON RATED,
SEE MECHANICAL
F3D.Gi
REFER TO OPENING THRU
SLAB DETAIL, 2/AE104 FOR
SLAB REINFORCING.
6'-0" MIN.
PROVIDE TEMPORARY BARRIERS AS
REQUIRED TO MEET ICRA RATING BY
STARC WALL SYSTEMS OR EQUAL 'LITE
WALL'. THIS AREA SHALL BE AVAILABLE
TO THE CONTRACTOR FOR 48 HRS MAX,
DAYS AND TIMES TO BE COORDINATED
WITH VA COR. SEE MECHANICAL FOR
SCOPE OF MECHANICAL WORK.PROVIDE TEMPORARY BARRIERS AS
REQUIRED TO MEET ICRA RATING
BY STARC WALL SYSTEMS OR
EQUAL 'LITE WALL'.
*REFER TO DRAWINGS AND
SPECIFICATIONS FOR OTHER INFECTION
CONTROL RELATED REQUIREMENTS IN
THIS AREA. COORDINATE CLOSELY WITH
VA COR AS THIS IS A 24 HR EMERGENCY
CARE SUITE.
6' -0
M
IN
LIBRARY
A-2009
MICROFORM
A-2014
CLERICAL
A-2015
OFFICE
A-2017
ROOM
A-2021
TRAINING
ROOM
A2022
SIMULATION
LAB
A2023
RUN EXHAUST DUCT ACROSS
ROOF, COORD. W/ MECHANICAL
DRAWINGS
RUN EXHAUST DUCT UP
WALL, COORD. W/
MECHANCIAL DRAWINGS
RUN EXHAUST DUCT UP THRU ROOF
SLAB, INSULATION, AND ROOFING,
SEE MECHANICAL DRAWINGS.
REFER TO OPENING THRU SLAB
DETAIL, 2/AE104 FOR SLAB
REINFORCING.
N
KEY PLAN
PROJECT
LOCATION
EXISTING ROOF DECK, CUT FOR NEW
PENETRATION AS REQUIRED, REFER TO
OPENING THRU SLAB DETAIL, 11/AE501 FOR
SLAB REINFORCING
EXISTING ROOF INSULATION, CUT FOR NEW
PENETRATION AS REQUIRED
NEW ROOFING MEMBRANE TO MATCH EXISTING
NEW .063" ALUMINUM COPING WITH KYNAR FINISH
NEW TREATED WOOD BLOCKING
NEW TREATED WOOD CANT
NEW EXHAUST DUCT
NEW DUCT PENETRATION CURB
CONCRETE
BEAM
C9X20 FRAME
DECK
OPENING
NOTE: COORDINATE OPENING SIZE WITH MECHANICAL
CONTRACTOR.
FOR OPENINGS
12" OR GREATER
1/2" PLATE. DRILL AND EPXOY W/ MIN 2 ANCHORS DESIGNED TO RESIT 5
KIPS. FOLLOW MANUFACTUERS RECOMENDATION FOR MINIUM EMBED,
TYP
CONCRETE
BEAM
STANDARD AISC SHEAR CONNECTION
DESIGNED FOR 5 KIPS ASD, TYP
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
Office of Construction and Facilities Management
U.S. Department of Veterans Affairs
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
Phase
ARCHITECT/ENGINEER OF RECORD
A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
Consultant:
C
U s e rs \d c h ri s te n s e n \D o c u m e n ts \1
.1
.0
N u c M e d v d c h ri s te n s e n M
N F
X
.r v t
/2
/2
:1
:2
A
M
AE104
PARTIAL FIRST & SECOND FLOOR
/ ROOF PLANSGUIDON DESIGN, INC.
1221 N. PENNSYLVANIA STREET
SUITE 200
INDIANAPOLIS, INDIANA 46202
317.800.6388
HEAPY
1400 W. DOROTHY LANE
DAYTON, OH 45409
937.224.0861
RELOCATE NUCLEAR MEDICINE
SERVICES FOR CENTRALIZED
SERVICES
INDIANAPOLIS, IN
583-20-102
01/28/2021 JF EB
BID DOCUMENTS
FULLY SPRINKLERED
Revisions: Date:
SCALE: 1/8" = 1'-0"AE104
4 PARTIAL FIRST FLOOR PLAN
SCALE: 1/8" = 1'-0"AE104
3 PARTIAL SECOND FLOOR ROOF PLAN
N N
SCALE: 3" = 1'-0"AE104
1 DUCT PENETRATION CURB
SCALE: 3/4" = 1'-0"AE104
2 TYPICAL OPENING THRU SLAB DETAIL
NOTE:
INSTALL DUCT ALONG SPANDREL GLAZING
WHERE APPLICABLE.
DUCT ACROSS
ROOF, SEE MECHANICAL
DUCT UP
WALL, SEE
MECHANCIAL
DRAWINGS
DUCT UP
WALL, SEE
MECHANICAL
DUCT UP WALL, SEE MECHANICAL
DRAWINGS
N
KEY PLAN
PROJECT
LOCATION
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
Office of Construction and Facilities Management
U.S. Department of Veterans Affairs
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
Phase
ARCHITECT/ENGINEER OF RECORD
A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
Consultant:
C
U s e rs \d c h ri s te n s e n \D o c u m e n ts \1
.1
.0
N u c M e d v d c h ri s te n s e n M
N F
X
.r v t
/2
/2
:5
:3
A
M
AE105
PARTIAL THIRD & FOURTH FLOOR
/ ROOF PLANSGUIDON DESIGN, INC.
1221 N. PENNSYLVANIA STREET
SUITE 200
INDIANAPOLIS, INDIANA 46202
317.800.6388
HEAPY
1400 W. DOROTHY LANE
DAYTON, OH 45409
937.224.0861
RELOCATE NUCLEAR MEDICINE
SERVICES FOR CENTRALIZED
SERVICES
INDIANAPOLIS, IN
583-20-102
01/28/2021 JF EB
BID DOCUMENTS
FULLY SPRINKLERED
Revisions: Date:
SCALE: 1/8" = 1'-0"AE105
1 PARTIAL THIRD FLOOR ROOF PLAN
SCALE: 1/8" = 1'-0"AE105
2 PARTIAL FOURTH FLOOR PLAN
NN
NOTE:
1. SEE DETAIL 4/AE104 FOR PIPE SUPPORT CURB DETAIL.
2. INSTALL DUCT ALONG SPANDREL GLAZING WHERE APPLICABLE.
KEY PLAN
PROJECT
LOCATION
DUCT UP WALL, SEE
MECHANICAL DRAWINGS
DUCT UP WALL, SEE
MECHANICAL DRAWINGS
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
Office of Construction and Facilities Management
U.S. Department of Veterans Affairs
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
Phase
ARCHITECT/ENGINEER OF RECORD
A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
Consultant:
C
U s e rs \d c h ri s te n s e n \D o c u m e n ts \1
.1
.0
N u c M e d v d c h ri s te n s e n M
N F
X
.r v t
/2
/2
:5
:3
A
M
AE106
PARTIAL FIFTH & SIXTH FLOOR /
ROOF PLANSGUIDON DESIGN, INC.
1221 N. PENNSYLVANIA STREET
SUITE 200
INDIANAPOLIS, INDIANA 46202
317.800.6388
HEAPY
1400 W. DOROTHY LANE
DAYTON, OH 45409
937.224.0861
RELOCATE NUCLEAR MEDICINE
SERVICES FOR CENTRALIZED
SERVICES
INDIANAPOLIS, IN
583-20-102
01/28/2021 JF EB
BID DOCUMENTS
FULLY SPRINKLERED
Revisions: Date:
SCALE: 1/8" = 1'-0"AE106
1 PARTIAL FIFTH FLOOR ROOF PLAN
SCALE: 1/8" = 1'-0"AE106
2 PARTIAL SIXTH FLOOR PLAN
NN
NOTE:
KEY PLAN
PROJECT
LOCATION
DUCT UP
WALL, SEE
MECHANICAL
DUCT UP WALL, SEE
MECHANICAL DRAWINGS
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
Office of Construction and Facilities Management
U.S. Department of Veterans Affairs
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
Phase
ARCHITECT/ENGINEER OF RECORD
A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
Consultant:
C
U s e rs \d c h ri s te n s e n \D o c u m e n ts \1
.1
.0
N u c M e d v d c h ri s te n s e n M
N F
X
.r v t
/2
/2
:5
:4
A
M
AE107
PARTIAL SEVENTH & EIGHTH
FLOOR PLANSGUIDON DESIGN, INC.
1221 N. PENNSYLVANIA STREET
SUITE 200
INDIANAPOLIS, INDIANA 46202
317.800.6388
HEAPY
1400 W. DOROTHY LANE
DAYTON, OH 45409
937.224.0861
RELOCATE NUCLEAR MEDICINE
SERVICES FOR CENTRALIZED
SERVICES
INDIANAPOLIS, IN
583-20-102
01/28/2021 JF EB
BID DOCUMENTS
FULLY SPRINKLERED
Revisions: Date:
SCALE: 1/8" = 1'-0"AE107
1 PARTIAL SEVENTH FLOOR PLAN
SCALE: 1/8" = 1'-0"AE107
2 PARTIAL EIGHTH FLOOR PLAN
N N
NOTE:
DUCT UP WALL,
SEE MECHANICAL
DUCT ALONG ROOF
SEE MECHANICAL
EXHAUST FAN AND STACK,
SEE 2/AE108. COORDINATE
WITH MECHANICAL DRAWINGS
NOTE:
INSTALL DUCT ALONG SPANDREL GLAZING WHERE
APPLICABLE.
N
KEY PLAN
PROJECT
LOCATION
ROOF
EXHAUST
FAN
GUY WIRE (4 REQUIRED)
WELDED STAINLESS
STEEL EXHAUST STACK
SECURE GUY WIRE
TO ROOF (TYP.)
SECURE
STACK
TO
ROOF
(TYP.)
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
Office of Construction and Facilities Management
U.S. Department of Veterans Affairs
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
Phase
ARCHITECT/ENGINEER OF RECORD
A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
Consultant:
C
U s e rs \d c h ri s te n s e n \D o c u m e n ts \1
.1
.0
N u c M e d v d c h ri s te n s e n M
N F
X
.r v t
/2
/2
:5
:4
A
M
AE108
PARTIAL NINTH FLOOR ROOF
PLANGUIDON DESIGN, INC.
1221 N. PENNSYLVANIA STREET
SUITE 200
INDIANAPOLIS, INDIANA 46202
317.800.6388
HEAPY
1400 W. DOROTHY LANE
DAYTON, OH 45409
937.224.0861
RELOCATE NUCLEAR MEDICINE
SERVICES FOR CENTRALIZED
SERVICES
INDIANAPOLIS, IN
583-20-102
01/28/2021 JF EB
BID DOCUMENTS
FULLY SPRINKLERED
Revisions: Date:
SCALE: 1/8" = 1'-0"AE108
1 PARTIAL NINTH FLOOR ROOF PLAN
N
NOTE:
1. SEE DETAIL 4/AE104 FOR PIPE SUPPORT CURB DETAIL, COORDINATE WITH
MECHANICAL DRAWINGS.
SCALE: 1/2" = 1'-0"AE108
2 EXHAUST FAN AND STACK
NOTE: COORDINATE FINAL DESIGN WITH MECHANICAL DRAWINGS.