This document provides details for a federal solicitation seeking construction services for infrastructure upgrades at the VA Sepulveda Ambulatory Care Center in North Hills, California. Work includes general construction, alterations, mechanical and electrical work, and utility systems necessary to complete the Electronic Health Records Modernization project. The solicitation value is between $10-20 million. Responses are due on June 23, 2021 via SAM.gov. The project will be set aside for Service-Disabled Veteran-Owned Small Businesses. The NAICS code is 236220 and size standard is $39.5 million. A site visit will be scheduled. The point of contact is James Barber at the VA Network Contracting Office 22 in Long Beach.

View the file

Other files

Show all 21

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

S-001

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

A

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 BS IMM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

1.

1.

A/E:

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

A

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 BS IMM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

1.

1.

A/E:

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

S-002

ABBREVIATIONSYMBOL DESCRIPTIONABBREVIATION DESCRIPTION ABBREVIATION DESCRIPTION

MECHANICAL ABBREVIATIONS AND LEGEND

ABBREVIATIONSYMBOL DESCRIPTION

POINT OF CONNECTION

POINT OF DISCONNECTPOD

POC

T'STAT

RR

CD

CFM

ER

SYMBOL, SEE EQUIPMENT SCHEDULE

CUBIC FEET PER MINUTE

CEILING DIFFUSER

RETURN REGISTER

EXHAUST REGISTER

HUMIDITY LEVEL SENSOR

ROUND DUCT

ROUND ELBOW ( LONG RADIUS )

12 INCH DIAMETER DUCT12"Ø

STRAIGHT TEE

TRANSITION - ROUND TO

RECTANGULAR DUCTTRANS.

EXHAUST DUCT DOWNEA

EXHAUST DUCT UPEA

SUPPLY AIR DUCT UPSA

SUPPLY AIR DUCT DOWNSA

T=3'' T

TRANS.

FLEX. CONN. FLEXIBLE DUCT CONNECTION

BD BALANCING DAMPER WITH

LOCKING DEVICE

RETURN AIR

DUCT UP

RETURN AIR

DUCT DOWN

CEILING SUPPLY DIFFUSER

SEE SCHEDULE

FLEXIBLE DUCTWORK

CEILING RETURN GRILLE

SEE SCHEDULE

SIDEWALL REGISTER

SEE SCHEDULE

MANUAL VOLUME DAMPER

RA

RA

MVD

MOTOR OPERATED DAMPERMOD

CEILING EXHAUST GRILLE

SEE SCHEDULE

THERMOMETERTI

PG

CONTROL VALVE (2-WAY)CV (2W)

CAPPED LINE

VALVE IN RISE

VALVE IN DROP

PIPE UP (RISE)

PIPE DN (DROP)

CD CONDENSATE DRAIN

SHUT-OFF VALVE

UNIONU

STR

FLEXIBLE CONNECTION (PIPE)FC

SUCTION DIFFUSERS

BV

CONTROL VALVE (3-WAY)CV (3W)

BACKFLOW PREVENTERBFP

PRESSURE RELIEF VALVEPR

PRV PRESSURE REDUCING VALVE

THERMOSTAT OR TEMPERATURE SENSOR

(NUMBER INDICATES EQUIPMENT OR

ZONE SERVED)

REMOVE/DEMO EXISTING EQUIPMENT

REMOVE/DEMO EXISTING PIPING

REMOVE/DEMO EXISTING DUCTWORK

REMOVE/DEMO EXISTING EQUIPMENT

TRANSITION RECTANGULAR DUCT

BALANCING VALVE

STRAINER W/ BLOW

DOWN VALVE

PRESSURE GAUGE WITH

SHUTOFF VALVE

TEST WELL (PETE'S PLUG)TW

SOV

CV CHECK VALVE

CONT CONTINUATION

x1

SPLITTER THROAT SIZE

D CFM

RECTANGULAR DUCT DIMENSIONS

A

ACM ASBESTOS CONTAINING MATERIALS

A/E ARCHITECT / ENGINEER

ABV ABOVE

ACD AUTOMATIC CONTROL DAMPER,

MODULATING

ACH AIR CHANGE RATE PER HOUR

AD ACCESS DOOR

AF AFTER FILTER

AFF ABOVE FINISHED FLOOR

AFG ABOVE FINISHED GRADE

AHU AIR-HANDLING UNIT

AMP AMPERE

APD AIR PRESSURE DROP

B

BHP BRAKE HORSEPOWER

BLDG BUILDING

BOD BOTTOM OF DUCT

BTU BRITISH THERMAL UNIT

BTUH BRITISH THERMAL UNIT PER HOUR

C

CC COOLING COIL

CD CEILING DIFFUSER OR CONDENSATE DRAIN

CFM CUBIC FEET PER MINUTE

CFT CUBIC FEET

CG CEILING GRILLE

CO CLEAN OUT/CARBON MONOXIDE

CO2 CARBON DIOXIDE

COP COEFFICIENT OF PERFORMANCE

CP CONDENSATE PUMP

CR CEILING REGISTER

CU CONDENSING UNIT

CV CONSTANT VOLUME

D

DB DRY-BULB TEMPERATURE

DB DECIBELS

DDC DIRECT DIGITAL CONTROLS

DEG DEGREE

DEMO DEMOLITION OR DEMOLISH

DF DIFFUSER

DN DOWN

DX DIRECT EXPANSION

DWGS. DRAWINGS

E

(E) EXISTING

EER ENERGY EFFICIENCY RATIO

EHRM ELECTRONIC HEALTH RECORDS

MODERNIZATION

ENT ENTERING

ESP EXTERNAL STATIC PRESSURE

EX EXISTING

EOL END OF LINE

F

F DEGREES FAHRENHEIT

FT FEET

FV FACE VELOCITY

G

GA GAUGE

GAL GALLONS

H

H HUMIDIFER

HP HORSEPOWER

HZ HERTZ

I

I/O INPUT/OUTPUT

IN INCHES

J

JCI JOHNSON CONTROL INC.

K

KW KILOWATT

KWH KILOWATT HOUR

L

LAT LEAVING AIR TEMPERATURE

LH LATENT HEAT

LVG LEAVING

M

M METER, SI UNIT

MAX MAXIMUM

MBH 1,000 BTUH

MCA MINIMUM BRANCH CIRCUIT AMPACITY

MFGR MANUFACTURER

MIN MINIMUM

MISC MISCELLANEOUS

N

NA NOT APPLICABLE

NAE NETWORK AUTOMATION ENGINE

NC NORMALLY CLOSED

NO NORMALLY OPEN

NTS NOT TO SCALE

O

OR OPERATING ROOM

OSA OUTSIDE AIR

OSHA OCCUPATIONAL SAFETY AND HEALTH ASSOCIATION

OSP OSHPD SEISMIC CERTIFICATION PRE-APPROVAL

P

PID PROPORTIONAL INTEGRAL DERIVATIVE

R

RA RETURN AIR

RL REFRIGERANT LIQUID

RS REFRIGERANT SUCTION

RENO RENOVATION

REQ. REQUIREMENT

S

SA SUPPLY AIR

SEN SENSIBLE HEAT

SQ FT SQUARE FOOT (FEET)

SS SPLIT SYSTEM

STRUCT. STRUCTURAL

T

T&PCV TEMPERATURE AND PRESSURE CONTROL VALVE

TAB TESTING, ADJUSTING AND BALANCING

TD TEMPERATURE DIFFERENCE

TDS TOTAL DISSOLVED SOLIDS

TG TRANSFER GRILLE

THRU THROUGH

TR TELECOMMUNICATION ROOM

TSP TOTAL STATIC PRESSURE

TU TERMINAL UNIT

TYP TYPICAL

U

UNO UNLESS OTHERWISE NOTED

UTR UP THROUGH ROOF

V

V VALVE

VA VETERAN AFFAIRS

VAV VARIABLE AIR VOLUME

VFD VARIABLE FREQUENCY DRIVE

W

W WATTS

WB WET-BULB (TEMPERATURE)

WC WATER COOLED

WG WATER GAUGE

W/ WITH

SQUARE ELBOW

RECTANGULAR ELBOW (LONG RADIUS)

HVAC EQUIPMENT SEE SCHEDULE

ABBREVIATION DESCRIPTION ABBREVIATION DESCRIPTION

CD

- CFM

C CFM

B CFM

A CFM

T

H

- PHOTO REFERENCE

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

M-001

MECHANICAL LEGEND &

ABBREVIATIONS

1. THESE DRAWINGS ARE A GENERAL GRAPHIC PRESENTATION OF THE WORK.

DUCTWORK, PIPING, AND EQUIPMENT, AS SHOWN, ARE DIAGRAMMATIC. DO

NOT SCALE DRAWINGS. PROVIDE ADDITIONAL OFFSETS AND SECTIONS OF

DUCTWORK AND PIPING AS REQUIRED TO MEET THE PROJECT REQUIREMENTS.

2. COORDINATE WITH OTHER TRADES. PROVIDE A COMPLETE SET OF

FABRICATION DRAWINGS REFLECTING ACTUAL DIMENSIONS, ACCESS

REQUIREMENTS, AND DETAILS BASED UPON THE ACTUAL EQUIPMENT

PROCURED. MAINTAIN AN UP TO DATE SET OF AS-BUILT DRAWINGS AT THE

JOB SITE.

3. COMPLY WITH APPLICABLE MECHANICAL CODE, PLUMBING CODE, FIRE

PROTECTION CODE, AND ALL OTHER GOVERNING CODES. THERE SHALL BE NO

EXCEPTION. REPORT DEFICIENCIES WITHIN THIRTY (30) DAYS UPON

AUTHORIZATION TO PROCEED.

4. REVIEW ALL DRAWINGS AND SPECIFICATIONS INCLUDING ARCHITECTURAL,

STRUCTURAL, CIVIL, MECHANICAL, PLUMBING, FIRE PROTECTION,

TELECOMMUNICATIONS AND ELECTRICAL. ANY QUESTIONS SHALL BE

BROUGHT UP, IN WRITING, TO THE ATTENTION OF THE ENGINEER BEFORE THE

START OF CONSTRUCTION.

5. PROVIDE ACCESS AND CLEARANCE FOR MAINTENANCE OF MECHANICAL

EQUIPMENT AND COMPONENTS AS RECOMMENDED BY EQUIPMENT

MANUFACTURER AND APPLICABLE CODES, BUT NO LESS THAN 3 FEET ON ALL

SIDES.

6. ALL SEALS, BEARINGS, PACKINGS, AND ACCESSORIES FOR ALL EQUIPMENT

AND PIPING SPECIALTIES SHALL BE SUITABLE FOR THE CONTINUOUS

OPERATIONAL TEMPERATURES, PRESSURES, AND CHARACTERISTICS OF THE

SYSTEMS THEY SERVE.

7. HANDLE, STORE, AND INSTALL EQUIPMENT PER MANUFACTURER'S

INSTRUCTIONS.

8. SUPPORT PIPING IN ACCORDANCE WITH ANSI/MSS SP-58 PIPE HANGERS AND

SUPPORTS - MATERIALS, DESIGN, MANUFACTURE, SELECTION, APPLICATION,

AND INSTALLATION.

9. SUPPORT DUCTWORK IN ACCORDANCE WITH ANSI/SMACNA 006 HVAC DUCT

CONSTRUCTION STANDARDS - METAL AND FLEXIBLE.

10. BRACE PIPING, EQUIPMENT, DUCTWORK, AND CONDUIT IN ACCORDANCE WITH

ISAT DESIGN, INSTALLATION & INSPECTION MANUAL FOR NON-STRUCTURAL

SEISMIC BRACING UNLESS THE AUTHORITY HAVING JURISDICTION HAS

ADDITIONAL REQUIREMENTS.

11. COMMISSION AND START-UP THE MECHANICAL SYSTEMS TO ASSURE A

COMPLETE AND OPERATIONAL HVAC SYSTEM IN ACCORDANCE WITH ASHRAE,

NEBB, OR AABC.

12. ALL PENETRATIONS THROUGH FIRE RATED CONSTRUCTION SHALL BE FIRE

SEALED WITH UL LISTED THROUGH-PENETRATION SEALING SYSTEMS SUITABLE

FOR WALL/FLOOR ASSEMBLIES AND PENETRATING MATERIAL.

13. MAINTAIN FIRE PROOFING OF STRUCTURAL BEAMS WHERE APPLICABLE.

14. LOCATE ALL ROOF EQUIPMENT IN ACCORDANCE WITH OSHA.

MECHANICAL GENERAL NOTES

1. FIELD VERIFY ALL EXISTING CONDITIONS INCLUDING STRUCTURAL,

MECHANICAL, PLUMBING, ELECTRICAL, TELECOMMUNICATIONS, EQUIPMENT,

AND ALL OTHER EXISTING SYSTEMS. MAKE NECESSARY PROVISIONS TO

MAINTAIN THE INTEGRITY OF SYSTEMS PRIOR TO THE COMMENCEMENT OF

DEMOLITION.

2. COORDINATE PHASING OF ALL DEMOLITION AND NEW WORK WITH OTHER

TRADES. REVIEW RENOVATION DRAWINGS TO VERIFY AND DETERMINE

EXTENT OF, AND SCHEDULING FOR, ALL DEMOLITION PRIOR TO PERFORMING

ANY WORK.

3. FIELD VERIFY ALL SIZES, LOCATIONS, AND ROUTING OF EXISTING ITEMS TO

REMAIN, AND OF NEW WORK INDICATED ON THE PLANS. NOTIFY THE DESIGN

AUTHORITY OF ANY DEVIATIONS WHICH MAY AFFECT RENOVATION WORK OR

SYSTEM OPERATION PRIOR TO PROCEEDING WITH THE WORK.

4. ESP DATA PROVIDED ON EQUIPMENT SCHEDULES DOES NOT INCLUDE ANY

ALLOWANCE FOR SYSTEM EFFECTS, DRIVE LOSSES, OR OTHER APPLIED

MATERIALS AND/OR COMPONENTS PROVIDED WITH OR FIELD INSTALLED

WITHIN MANUFACTURED EQUIPMENT. EXCEPT AS NOTED OTHERWISE, FINAL

FAN AND MOTOR SELECTIONS SHALL INCLUDE THE ABOVE AS WELL AS

ALLOWANCES FOR ACCUMULATIONS ON FILTERS AND COILS. ESP INCLUDES

ONLY THOSE ITEMS 'EXTERNAL' TO EQUIPMENT. MANUFACTURER'S SELECTION

OF FANS AND MOTORS SHALL BE BASED UPON CALCULATED TSP AND BHP MAY

NOT EXCEED 85% OF NOMINAL HP.

MECHANICAL PROJECT NOTES

SPLIT SYSTEM SCHEDULE (HEAT PUMP)

INDOOR

SYMBOL AREA SERVED

INDOOR FAN SECTION

OUTDOOR

SYMBOL

OUTDOOR SECTION

MAX.

CFM V Ø Hz

WEIGHT

(LBS) MCA V Ø Hz

WEIGHT

(LBS)

COOLING CAPACITY

MIN

SEER

MAX. PIPING

LENGTH (FT)NOMINAL

TONNAGE

SENS.

CAP.

(MBH)

AMB

(°F)

DETAIL REMARKSREFRIGERANT

TYPE

BLDG 1 ROOM B-120A 568 1 60 31 13.75 1 60 104SS

1-A154 R410A 97.7 14 98

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

SS

1-B120A

SS

10-BB110

SS

10-A29B

SS

10-C00A

SS

20-1A15

SS

20-2A115

SS

22-103

SS

22-202

SS

23-106

SS

25-A109

SS

45-105

SS

99-G57A

SS

99-112

SS

200-1525

SS

200-1533

SS

200-1001B

SS

200-1927

SS

200-2466

SS

200-2165

SS

200-2664

SS

200-3326

SS

200-3406

SS

200-3143

SS

200-3573

M-002

MECHANICAL SCHEDULES

CU

10-BB110

CU

10-A29B

CU

10-C00A

CU

20-1A15

CU

20-2A115

CU

22-103

CU

22-202

CU

23-106

CU

25-A109

CU

45-105

CU

99-G57A

CU

99-112

CU

200-1525

CU

200-1533

CU

200-1001B

CU

200-1927

CU

200-2466

CU

200-2165

CU

200-2664

CU

200-3326

CU

200-3406

CU

200-3143

CU

200-3573

SS

200-2366

CU

200-2366

BLDG 1 ROOM A-154 1 60 1 60 R410A 97.7

1 60 1 60 R410A 97.7

1 60 1 60 R410A 97.7

1 60 1 60 R410A 97.7

1 60 R410A 97.7

1 60 1 60 R410A 97.7

1 60 31 13.75 1 60 R410A 97.7

60 1 60 R410A 97.7

1 60 1 60 R410A 97.7

1 60 31 13.75 1 60 R410A 97.7

1 60 31 13.75 1 60 R410A 97.7

1 60 31 1 60 R410A 97.7

1 60 31 1 60 R410A 97.7

1 60 31 1 60 R410A 97.7

1 60 31 1 60 R410A 97.7

1 60 31 1 60 R410A 97.7

1 60 31 1 60 R410A 97.7

1 60 31 1 60 R410A 97.7

1 60 31 1 60 R410A 97.7

1 60 31 1 60 R410A 97.7

1 60 31 1 60 R410A 97.7

1 60 31 1 60 R410A 97.7

1 60 1 60 R410A 97.7 98

1 60 1 60 R410A 97.7

1 60 1 60 R410A 97.7

1.25 9.2

5.0

8.5

8.5

8.5

5.0

6.4

5.6

4.6

3.1

8.8

4.6

9.1

3.4

9.1

7.3

8.6

7.0

7.3

6.0

11.3

7.2

5.8

8.7

4.5

7.1

EXHAUST FAN SCHEDULE

SYMBOL DESCRIPTION AREA SERVED CFM ESP DRIVE RPM

MOTOR OPER

WEIGHT

(LBS)

MTG

DETAIL REMARKS

BHP HP V Ø Hz VFD

INLINE EXHAUST FAN BUILDING 70 TR 330 0.3 DIRECT 1184 0.04 1/6 120 1 60 NO 50EF

70-1

DESCRIPTION

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DUCTLESS SPLIT SYSTEM

DESCRIPTION

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

CONDENSING UNIT

BLDG 10 ROOM B-B110

BLDG 10 ROOM A29B

BLDG 10 ROOM C00A

BLDG 20 ROOM 1-A15

BLDG 20 ROOM 2-A115

BLDG 22 ROOM 103

BLDG 22 ROOM 202

BLDG 23 ROOM 106

BLDG 25 ROOM A 109

BLDG 45 ROOM 105

BLDG 99 ROOM G57A

BLDG 99 ROOM 112

SS

200-1356

CU

200-1356DUCTLESS SPLIT SYSTEM 1 60 1 60 R410A 97.7 989.8CONDENSING UNIT

PROVIDE FACTORY WIRED WALL-MOUNTED

THERMOSTAT.

BLDG 200 ROOM 1356

BLDG 200 ROOM 1525

BLDG 200 ROOM 1533

BLDG 200 ROOM 1001B

BLDG 200 ROOM 1927

BLDG 200 ROOM 2366

BLDG 200 ROOM 2466

BLDG 200 ROOM 2165

BLDG 200 ROOM 2664

BLDG 200 ROOM 3326

BLDG 200 ROOM 3406

BLDG 200 ROOM 3143

BLDG 200 ROOM 3573

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

DO NOT EXCEED MANUFACTURER'S LISTED MAXIMUM PIPING LENGTH REQUIREMENTS. COORDINATE SELECTION WITH ACTUAL LENGTHS

FROM SHOP DRAWINGS.

PROVIDE W/ FACTORY PROVIDED FIELD INSTALLED CONDENSATE PUMP.1 2 3 4 INDOOR UNIT POWERED FROM THE OUTDOOR UNIT. REFER TO ELECTRICAL

55 PROVIDE PER APPLICABLE OSP.

C8/M-502 1 2

PROVIDE WITH BACK DRAFT DAMPER.1 2 PROVIDE WITH A SPEED CONTROLLER TO BE USED FOR BALANCING ONLY. CONTROLLER SHALL NOT RESET UPON LOSS OF POWER.

13.75

13.75

13.75

13.75

13.75

13.75

13.75

13.75

13.75

13.75

13.75

13.75

13.75

13.75

SERVICE

AIR DISTRIBUTION SCHEDULE

SYMBOL MOUNTING

SURFACEDESCRIPTION NECK

SIZE PHOTO REMARKS

EXHAUST

AIR EXHAUST GRILLE

CFM

RANGE

T-BAR/

HARD

CEILING

ALUMINUM

CONSTRUCTION

1908"Ø

34010"Ø

56012"Ø

89014"Ø

1,190

18"Ø 1,610PROVIDE PER APPLICABLE OSP.3

CU

1-A154

CU

1-B120A

568 1 60 31 104 9813.75 1.25

1.25

1.25

1.25

1.25

1.25

1.25

1.25

1.25

1.25

1.25

1.25

1.25

1.25

1.25

1.25

1.25

1.25

6VERIFY AND COORDINATE CONNECTION SIDE PRIOR TO PROCUREMENT.

5 6

5 6

5 6

5 6

5 6

5 6

5 6

5 6

5 6

5 6

5 6

5 6

5 6

5 6

5 6

5 6

5 6

5 6

5 6

5 6

5 6

5 6

5 6

5 6

5 6

5 6

7 PROVIDE WITH INTEGRAL OVERFLOW ALARM TO SHUFT OFF THE EVAPORATOR IN CASE OF BLOCKAGE OR

FAILURE.

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

B5 / M-501

16"Ø

1. PROVIDE MECHANICAL SHOP DRAWINGS IN ACCORDANCE WITH GI003.

MECHANICAL SHOP DRAWINGS

568 31 13.75 104 981.25

568 31 13.75 104 981.25

568 31 13.75 104 981.25

568 31 13.75 104 981.25

568 31 13.75 104 981.25

568 31 13.75 104 981.25

98568 31 13.75 104 1.25

98568 31 13.75 104 1.25

CORRIDOR

PASSAGE

B124

STAIR

PASSAGE

B124

FISCAL

PAYROLL OFFICE

B169

EMPLOYMENT

OFFICE

FISCAL

B168

FISCAL

PAYROLL OFFICE

B167B170AB170

LOBBY

B100AB100B

IFCAP TRAINING

BID OPENING

B101B102

KEN REINBOLD

JAMES KREVITZ

B103

FISCAL OOFICE

A165

FINANCIAL

MGR. OFFICE

A165A

CHIEF

OF STAFF

SECRETARY

TO CHIEF

SECRETARY

A-162

FEMALE

TOILET

B155

MALE

TOILET

A157

JAN.

CLO. TOI.

CONTRACT

SERVICE

C106

JANITOR

C105

PROCUREMENT

SECTION

C112

DIRECTOR'S

CONF. ROOM

B125

PERSONNEL

OFFICE

B146

B150

MAIL

CENTER

B119

POST OFFICE

CHIEF PURCHASE

CONTRACT

C110

PETER EGGENA

DEPUTY ACCS.

CHIEF OF SUPPLY

SERVICE

C108

C113

MEN'S

TOILET

C114

WOMEN'S

TOILET

C116

BREAK

ROOM

C118

C117

JIM YEANCE

OFFICE

B126

PROCESSING

AND RECORDS

B-127

RETIREMENT

B131

B132

B133

ENVIRONMENTAL

MGMT.

B135

CHIEF OF

ACOS.

LABOR REL.

SPECIALIST

B145

CHIEF OF

PERSONNEL SER.

B144A

B144

PERSONNEL

B142

WOMEN'S

LOUNGE

B140

SPIRIT 1776

CONF. ROOM

WOMEN'S

TOILET

MEN'S

TOILET

CONFERENCE

ROOM

A158

ASS. DIRECTOR'S

OFFICE PASSAGE

CONFERENCE

A160

AA TO

DIRECTOR

DIRECTOR'S

OFFICE

C108

VESTIBULE

PERSONAL PROPERTY

MANAGEM'T SECTION

SUPPLY

STAFFING

B151 B152

STAFFING

B153B

PERSONNEL

SERVICE

B153

B143

PERSONNEL

B141

CLOS.

STOR.

JAN.

CLOS.

C115

CORRIDORCORRIDOR CORRIDORCO

RR

ID

OR

CO

RR

ID

OR

CO

RR

ID

OR

CO

RR

ID

OR

LOADING

DOCK

LOADING

DOCK

FIRE EXIT

B147

B148

CAREER

MGMT

B149

BUDGET

ASSISTANT

C111A

FIRE EXIT

C107

TELECOM RM

B120A

TELECOM RM

A154

3/4" CD

3/4" CD

C1 M-401

E1 M-401

C4 M-401

E4 M-401

CD

CD

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

1.M-111

MECHANICAL FIRST FLOOR PLAN -

BUILDING 1

MECHANICAL FIRST FLOOR PLAN - BUILDING 1F1 SCALE: 3/32" = 1'-0" 0' 4' 16' 24'2' 8'

BUILDING 1

NORTH

1. DISCHARGE CONDENSATE DRAIN TO UTILITY SINK. SEE

DETAIL F5/M-501.

2. RS/RL UP TO ROOF. REFER TO 1.M-1R1 FOR

CONTINUATION.

3. ROUTE CONDENSATE DRAIN EXPOSED BELOW CEILING.

PATCH AND REPAIR CEILING AS REQUIRED.

4. ROUTE RS/RL IN ACCORDANCE WITH DETAIL D5/M-501.

NORTH

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

1C

1A

BUILDING 1

CU

1-A154

CU

1-B120A

1. TERMINATE CONDENSATE DRAIN ABOVE MOPSINK. SEE DETAIL C5/M-502.2. PROVIDE RECOMMENDED MANUFACTUREREQUIPMENT CLEARANCE. REFER TOSTRUCTURAL PLANS FOR ANCHORAGEDETAILS.

1. TERMINATE CONDENSATE DRAIN ABOVE MOPSINK. SEE DETAIL C5/M-502.2. PROVIDE RECOMMENDED MANUFACTUREREQUIPMENT CLEARANCE. REFER TOSTRUCTURAL PLANS FOR ANCHORAGEDETAILS.

1(TYP)

3 (TYP)

2 (TYP)

4(TYP)

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

1.M-1R1

MECHANICAL ROOF PLAN -

BUILDING 1

MECHANICAL ROOF PLAN - BUILDING 1F1 SCALE: 3/32" = 1'-0" 0' 4' 16' 24'2' 8'

BUILDING 1

NORTH

1. PROVIDE CODE REQUIRED AND MANUFACTURER'S

RECOMMENDED EQUIPMENT CLEARANCE.

2. RS/RL DOWN TO FLOOR BELOW. REFER TO 1.M-111 FOR

CONTINUATION.

3. PROVIDE CONCRETE HOUSEKEEPING PAD.

4. ROUTE RS/RL IN ACCORDANCE WITH DETAIL D5/M-501.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

1B

1D

BUILDING 1

3/4" CD

CU

10-BB110

C7 M-401

E7 M-401

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

10.M-101

MECHANICAL BASEMENT PLAN -

BUILDING 10

MECHANICAL BASEMENT PLAN - BUILDING 10F1 SCALE: 1/16" = 1'-0"

NORTH

0' 8' 16' 32'

BUILDING 10

NORTH

BUILDING 10

1. DISCHARGE CONDENSATE DRAIN TO MOP SINK. SEE

DETAIL C5/M-502.

2. PROVIDE CODE REQUIRED AND MANUFACTURER'S

RECOMMENDED EQUIPMENT CLEARANCE.

3. PROVIDE CONCRETE HOUSEKEEPING PAD ON GRADE.

4. ROUTE RS/RL IN ACCORDANCE WITH DETAIL D5/M-501.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

1E

3/4" CD 3/4" CD

C4 M-402

E4 M-402

C7 M-402

E7 M-402

CD33

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

10.M-121

MECHANICAL SECOND FLOOR PLAN

- BUILDING 10

MECHANICAL SECOND FLOOR PLAN - BUILDING 10F1 SCALE: 1/16" = 1'-0"

NORTH

0' 8' 16' 32'

BUILDING 10

NORTH

BUILDING 10

1. DISCHARGE CONDENSATE DRAIN TO MOP SINK. SEE

DETAIL C5/M-502.

2. RS/RL UP TO ROOF. REFER TO 10.M-1R1 FOR

CONTINUATION.

3. ROUTE RS/RL IN ACCORDANCE WITH DETAIL D5/M-501.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

1H

1I

10-C00A

CU

10-A29B

(TYP)

(TYP)3

(TYP)

4(TYP)

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

10.M-1R1

MECHANICAL ROOF PLAN -

BUILDING 10

MECHANICAL ROOF PLAN - BUILDING 10F1 SCALE: 1/16" = 1'-0"

NORTH

0' 8' 16' 32'

BUILDING 10

NORTH

BUILDING 10

1. PROVIDE CODE REQUIRED AND MANUFACTURER'S

RECOMMENDED EQUIPMENT CLEARANCE.

2. RS/RL DOWN TO FLOOR BELOW. REFER TO 10.M-121 FOR

CONTINUATION.

3. PROVIDE CONCRETE HOUSEKEEPING PAD.

4. ROUTE RS/RL IN ACCORDANCE WITH DETAIL D5/M-501.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

1J 1H 1I

C1 M-403

E1 M-403

CD

CU

20-1A15

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

20.M-111

MECHANICAL FIRST FLOOR PLAN -

BUILDING 20

MECHANICAL FIRST FLOOR PLAN - BUILDING 20F1 SCALE: 3/32" = 1'-0" 0' 4' 16' 24'2' 8'

BUILDING 20

NORTH

BUILDING 20

1. DISCHARGE CONDENSATE DRAIN TO UTILITY SINK. SEE

DETAIL F5/M-501.

2. PROVIDE CODE REQUIRED AND MANUFACTURER'S

RECOMMENDED EQUIPMENT CLEARANCE.

3. PROVIDE CONCRETE HOUSEKEEPING PAD.

4. ROUTE RS/RL IN ACCORDANCE WITH DETAIL D5/M-501.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

1K

M-403

E4 M-403

CD

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

20.M-121

MECHANICAL SECOND FLOOR PLAN

- BUILDING 20

MECHANICAL SECOND FLOOR PLAN - BUILDING 20F1 SCALE: 3/32" = 1'-0"

BUILDING 20

NORTH

BUILDING 20

1. DISCHARGE CONDENSATE DRAIN TO MOP SINK. SEE

DETAIL C5/M-502.

2. RS/RL UP TO ROOF. REFER TO 20.M-1R1 FOR

CONTINUATION.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

1L

0' 4' 16' 24'2' 8'

20-2A115

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

20.M-1R1

MECHANICAL ROOF PLAN -

BUILDING 20

MECHANICAL ROOF PLAN - BUILDING 20F1 SCALE: 3/32" = 1'-0"

BUILDING 20

NORTH

BUILDING 20

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

1. PROVIDE CODE REQUIRED AND MANUFACTURER'S

RECOMMENDED EQUIPMENT CLEARANCE.

2. RS/RL DOWN TO FLOOR BELOW. REFER TO 20.M-121

FOR CONTINUATION.

3. PROVIDE CONCRETE HOUSEKEEPING PAD.

4. ROUTE RS/RL IN ACCORDANCE WITH DETAIL D5/M-501.

1M

0' 4' 16' 24'2' 8'

C7 M-403

E7 M-403

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

22.M-111

MECHANICAL FIRST FLOOR PLAN -

BUILDING 22

MECHANICAL FIRST FLOOR PLAN - BUILDING 22F1 SCALE: 1/8" = 1'-0"

NORTH

0' 4' 8' 16'

BUILDING 22

NORTH

BUILDING 22

1. CONDENSATE DRAIN TO UTILITY SINK, SEE DETAIL

F5/M-501.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

1N

C1 M-404

E1 M-404

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

22.M-121

MECHANICAL SECOND FLOOR PLAN

- BUILDING 22

MECHANICAL SECOND FLOOR PLAN - BUILDING 22F1 SCALE: 1/8" = 1'-0"

NORTH

0' 4' 8' 16'

BUILDING 22

NORTH

BUILDING 22

1. CONDENSATE DRAIN TO UTILITY SINK, SEE DETAIL

F5/M-501.

2. RS/RL UP TO ROOF ABOVE. REFER TO 22.M-1R1 FOR

CONTINUATION.

3. ROUTE RS/RL IN ACCORDANCE WITH DETAIL D5/M-501.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

1O

1 (TYP)

(TYP)

(TYP)

CU

22-103

CU

22-202

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

22.M-1R1

MECHANICAL ROOF PLAN -

BUILDING 22

MECHANICAL ROOF PLAN - BUILDING 22F1 SCALE: 1/8" = 1'-0"

NORTH

0' 4' 8' 16'

BUILDING 22

NORTH

BUILDING 22

1. PROVIDE CODE REQUIRED AND MANUFACTURER'S

RECOMMENDED EQUIPMENT CLEARANCE.

2. RS/RL DOWN TO FLOOR BELOW. REFER TO 22.M-121 FOR

CONTINUATION.

3. PROVIDE CONCRETE HOUSEKEEPING PAD.

4. ROUTE RS/RL IN ACCORDANCE WITH DETAIL D5/M-501.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

1P

M-404

E4 M-404

CD

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

23.M-111

MECHANICAL FIRST FLOOR PLAN -

BUILDING 23

MECHANICAL FIRST FLOOR PLAN - BUILDING 23F1 SCALE: 1/8" = 1'-0"

NORTH

0' 4' 8' 16'

BUILDING 23

NORTH

BUILDING 23

1. DISCHARGE CONDENSATE DRAIN TO UTILITY SINK. SEE

DETAIL F5/M-501.

2. RS/RL UP TO FLOOR ABOVE. REFER TO 23.M-121 FOR

CONTINUATION.

3. ROUTE RS/RL IN ACCORDANCE WITH DETAIL D5/M-501.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

1Q

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

23.M-121

MECHANICAL SECOND FLOOR -

BUILDING 23 BUILDING 23

MECHANICAL SECOND FLOOR - BUILDING 23F1 SCALE: 1/8" = 1' - 0"

NORTH

0' 4' 8' 16'

BUILDING 23

NORTH

BUILDING 23

1. RS/RL UP TO ROOF. REFER TO 23.M-1R1 FOR

CONTINUATION.

2. RS/RL DOWN TO CEILING OF FLOOR BELOW. REFER TO

23.M-111 FOR CONTINUATION.

3. ROUTE RS/RL IN ACCORDANCE WITH DETAIL D5/M-501.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

23-106

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

23.M-1R1

MECHANICAL ROOF PLAN -

BUILDING 23 BUILDING 23

MECHANICAL ROOF PLAN - BUILDING 23F1 SCALE: 1/8" = 1' - 0"

1. PROVIDE CODE REQUIRED AND MANUFACTURER'S

RECOMMENDED EQUIPMENT CLEARANCE.

2. RS/RL DOWN TO FLOOR BELOW. REFER TO 23.M-121

FOR CONTINUATION.

3. PROVIDE CONCRETE HOUSEKEEPING PAD.

NORTH

BUILDING 23

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

NORTH

0' 4' 8' 16'

1R

CD

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

25.M-111

MECHANICAL FIRST FLOOR PLAN -

BUILDING 25

MECHANICAL FIRST FLOOR PLAN - BUILDING 25F1 SCALE: 1/8" = 1'-0"

NORTH

0' 4' 8' 16'

BUILDING 25

NORTH

BUILDING 25

1. DISCHARGE CONDENSATE DRAIN TO MOP SINK. SEE

DETAIL C5/M-502.

2. RS/RL UP TO ROOF. REFER TO 25.M-1R1 FOR

CONTINUATION.

C1 M-405

E1 M-405

1S

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

25-A109

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

25.M-1R1

MECHANICAL ROOF PLAN -

BUILDING 25

MECHANICAL ROOF PLAN - BUILDING 25F1 SCALE: 3/32" = 1'-0"

NORTH

0' 4' 16' 24'2' 8'

BUILDING 25

NORTH

BUILDING 25

1. PROVIDE CODE REQUIRED AND MANUFACTURER'S

RECOMMENDED EQUIPMENT CLEARANCE.

2. RS/RL DOWN TO FLOOR BELOW. REFER TO 25.M-111 FOR

CONTINUATION.

3. PROVIDE CONCRETE HOUSEKEEPING PAD.

4. ROUTE RS/RL IN ACCORDANCE WITH DETAIL D5/M-501.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

1T

M-405

E4 M-405

3/4" CD

CD

CU

45-105

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

45.M-111

MECHANICAL FIRST FLOOR PLAN -

BUILDING 45

MECHANICAL FIRST FLOOR PLAN - BUILDING 45F1 SCALE: 1/8" = 1'-0" 0' 4' 8' 16'

BUILDING 45

NORTH

BUILDING 45

KEY NOTES-

1. CONDENSATE DRAIN TO LAVATORY TAILPIECE. SEE

DETAIL F8/M-501.

2. PROVIDE CODE REQUIRED AND MANUFACTURER'S

RECOMMENDED EQUIPMENT CLEARANCE.

3. PROVIDE CONCRETE HOUSEKEEPING PAD ON GRADE.

4. ROUTE RS/RL IN ACCORDANCE WITH DETAIL D5/M-501.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

NORTH

2B

2A

M-405

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

63.M-111

MECHANICAL FIRST FLOOR PLAN -

BUILDING 63

MECHANICAL FIRST FLOOR PLAN - BUILDING 63F1 SCALE: 1/4" = 1'-0"

NORTH0' 4' 8' 16'

BUILDING 63

NORTH

BUILDING 63

2C

1. REFER TO M-80X FOR PHOTO REFERENCES ( ).-

DINING

ROOM

KITCHEN

STORAGE

STAFF

TOILET

RECREATION AREA

RECEPTION

CONFERENCE ROOM

OFFICE

O.T. OFFICE

MEDICAL

TREATMENT

PATIENT

TOILET

CLOS.

TELECOM

RM

TDU#201-70

CLINICAL

DIRECTOR MALE

TOILET

C1 M-406

E1 M-406

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

70.M-111

MECHANICAL FIRST FLOOR PLAN -

BUILDING 70

MECHANICAL FIRST FLOOR PLAN - BUILDING 70F1 SCALE: 3/8" = 1'-0"

BUILDING 70

NORTH NORTH

BUILDING 70

0' 1' 4' 6'2'

HZ

HZHZ

HZHZ

HZ

HZ

HZ

HZ

HZ

HZ

3/4" CD

C4 M-406

E4 M-406

CD

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

99.M-101

MECHANICAL GROUND FLOOR

PLAN - BUILDING 99

MECHANICAL GROUND FLOOR PLAN - BUILDING 99F1 SCALE: 1/16" = 1'-0" 0' 8' 16' 32'

BUILDING 99

NORTH

BUILDING 99

1. DISCHARGE CONDENSATE DRAIN TO MOP SINK.

SEE DETAIL C5/M-502.

2. RS/RL UP TO LOWER ROOF. REFER TO 99.M-1R1

FOR CONTINUATION.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

NORTH

2D

M-406

E7 M-406

CD

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

99.M-111

MECHANICAL FIRST FLOOR PLAN -

BUILDING 99

MECHANICAL FIRST FLOOR PLAN - BUILDING 99F1 SCALE: 3/16" = 1'-0"

NORTH

0' 2' 6' 12'1' 4'

BUILDING 99

NORTH

BUILDING 99

1. DISCHARGE CONDENSATE DRAIN TO MOP SINK. SEE

DETAIL C5/M-502.

2. RS/RL UP TO ROOF. REFER TO 99.M-1R1 FOR

CONTINUATION.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

2F

99-G57A

CU

99-112

(TYP)

(TYP)

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

99.M-1R1

MECHANICAL ROOF PLAN -

BUILDING 99 BUILDING 99

MECHANICAL ROOF PLAN - BUILDING 99F1 SCALE: 1/16" = 1' - 0" 0' 8' 16' 32'

NORTH

BUILDING 99

1. PROVIDE CODE REQUIRED AND MANUFACTURER'S

RECOMMENDED EQUIPMENT CLEARANCE.

2. RS/RL DOWN TO FLOOR BELOW. REFER TO 99.M-101 FOR

CONTINUATION.

3. PROVIDE CONCRETE HOUSEKEEPING PAD.

4. LOWER ROOF.

5. UPPER ROOF.

6. RS/RL DOWN TO FLOOR BELOW. REFER TO 99.M-111 FOR

CONTINUATION.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

2E

2G

3/4" CD

3/4" CD

3/4" CDC1

M-407

E1 M-407

C4 M-407

E4 M-407

C7 M-407

E7 M-407

C1 M-408

F1 M-408

C4 M-408

F4 M-408

CD

CD

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

200.M-111

MECHANICAL FIRST FLOOR PLAN -

BUILDING 200

MECHANICAL FIRST FLOOR PLAN - BUILDING 200F1 SCALE: 1/16" = 1'-0"

BUILDING 200

NORTH

0' 8' 16' 32'

NORTH

BUILDING 200

1. DISCHARGE CONDENSATE DRAIN TO MOP SINK. SEE

DETAIL C5/M-502.

2. DISCHARGE 3/4" CD WITH MINIMUM 1" AIR GAP

INDIRECTLY TO EXISTING FLOOR SINK.

3. RS/RL UP TO FLOOR ABOVE. REFER TO 200.M-121 FOR

CONTINUATION.

4. DISCHARGE CONDENSATE DRAIN TO LAVATORY

TAILPIECE. SEE DETAIL F8/M-501.

5. ROUTE RS/RL IN ACCORDANCE WITH DETAIL D5/M-501.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

2H

2I

2J

2K

3/4" CD

C4 M-409

F4 M-409

C7 M-409

F7 M-409

C1 M-409

F1 M-409

C7 M-408

F7 M-408

3/4" CD

CD

CU

200-1356

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

200.M-121

MECHANICAL SECOND FLOOR PLAN

- BUILDING 200

MECHANICAL SECOND FLOOR PLAN - BUILDING 200F1 SCALE: 1/16" = 1'-0"

BUILDING 200

NORTH

0' 8' 16' 32'

NORTH

BUILDING 200

1. DISCHARGE CONDENSATE DRAIN TO MOP SINK. SEE

DETAIL C5/M-502.

2. DISCHARGE CONDENSATE DRAIN TO LAVATORY

TAILPIECE, SEE DETAIL F8/M-501.

3. RS/RL DOWN TO FLOOR BELOW. REFER TO 200.M-111 FOR

CONTINUATION.

4. PROVIDE CONCRETE HOUSEKEEPING PAD.

5. PROVIDE CODE REQUIRED AND MANUFACTURER'S

RECOMMENDED EQUIPMENT CLEARANCE.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

2L

2M

3/4" CD 3/4" CD

3/4" CD

C1 M-410

F1 M-410

C4 M-410

F4 M-410

C1 M-411

F1 M-411

C7 M-410

F7 M-410

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

200.M-131

MECHANICAL THIRD FLOOR PLAN -

BUILDING 200

MECHANICAL THIRD FLOOR PLAN - BUILDING 200F1 SCALE: 1/16" = 1'-0"

BUILDING 200

NORTH

0' 8' 16' 32'

NORTH

BUILDING 200

1. DISCHARGE CONDENSATE DRAIN TO MOP SINK. SEE

DETAIL C5/M-502.

2. NOT USED.

3. DISCHARGE CONDENSATE DRAIN TO LAVATORY

TAILPIECE, SEE DETAIL F8/M-501.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

2N

2P

2O

3 (TYP)

(TYP)

(TYP)

(TYP)

2 (TYP)

2(TYP)

CU

200-1001B

CU

200-2165

CU

200-3143

CU

200-1927

CU

200-2664

CU

200-3573

CU

200-1525

CU

200-3326

CU

200-2366

CU

200-1533

CU

200-3406

CU

200-2466

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

Key Plan

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

KEY NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

200.M-1R1

MECHANICAL ROOF PLAN -

BUILDING 200 BUILDING 200

MECHANICAL ROOF PLAN - BUILDING 200F1 SCALE: 1/16" = 1'-0"

NORTH

0' 8' 16' 32'

NORTH

BUILDING 200

1. PROVIDE CODE REQUIRED AND MANUFACTURER'S

RECOMMENDED EQUIPMENT CLEARANCE.

2. RS/RL DOWN TO FLOOR BELOW. REFER TO 200.M-131

FOR CONTINUATION.

3. PROVIDE CONCRETE HOUSEKEEPING PAD.

1. RS/RL SIZED PER MANUFACTURER'S RECOMMENDATIONS.

2. REFER TO M-80X FOR PHOTO REFERENCES ( ).

3. ROUTE CONDENSATE DRAIN PIPING CONCEALED ABOVE

CEILING, EXCEPT IN JANITORS CLOSET / HOUSEKEEPING

ROOMS WITH HARD LID CEILINGS. PATCH AND REPAIR

CEILING WHERE REQUIRED TO PENETRATE CEILING.

2R2Q

2S 2T

FEMALE

TOILET

B155

CO

RR

ID

OR

TELECOM RM

A154

TELECOM RM

A154

10"Ø (E)TERMINAL UNIT

1T H

10"Ø (E)TERMINAL UNIT

STORAGE

B120A

TELECOM RM

B120A

Project Number

Building Number

Drawing Number

Project Title

Location

Issue Date Checked Drawn

Drawing Title

Approved:

VA FORM 08 - 6231

A

Fi le

P at h U.S. Department of Veterans Affairs

B

E

F

A

B

C

D

E

F

21 4 5 6 7 8 9 10

Phase

STAMPCONSULTANT

321 4 5 6 7 8 9 10

Project Director

Revisions: Date:

691A4-20-201FINAL CONSTRUCTION

DOCUMENTS

May 17, 2021 MS TS / JM

FULLY SPRINKLERED

A

GENERAL NOTES

ARCHITECT/ENGINEER OF RECORD

C

D

1611 PLUMMER ST, NORTH HILLS, CA, 91343

20801.13

Electronic Health Records Modernization (EHRM) Infrastructure Upgrades - Sepulveda, CA

M-401

MECHANICAL ENLARGED PLANS

SCALE: 1/4" = 1' -0"

SCALE: 1/4" = 1' -0"

0' 2' 4' 8'

0' 2' 4' 8'

MECHANICAL ENLARGED FIRST FLOOR

DEMOLITION PLAN (BLDG 1) - ROOM A154

MECHANICAL ENLARGED FIRST FLOOR

RENOVATION PLAN (BLDG 1) - ROOM A154E4

1.M-111

C4

1.M-111

MECHANICAL ENLARGED BASEMENT

DEMOLITION PLAN (BLDG 10) - ROOM B-B110

SCALE: 1/4" = 1' -0" 0' 2' 4' 8'

C7

10.M-101

MECHANICAL ENLARGED BASEMENT

RENOVATION PLAN (BLDG 10) - ROOM B-B110

SCALE: 1/4" = 1' -0" 0' 2' 4' 8'

E7

10.M-101

SCALE: 1/4" = 1' -0"

SCALE: 1/4" = 1' -0"

0' 2' 4' 8'

0' 2' 4' 8'

MECHANICAL ENLARGED FIRST FLOOR

DEMOLITION PLAN (BLDG 1) - ROOM B-120A

MECHANICAL ENLARGED FIRST FLOOR

RENOVATION PLAN (BLDG 1) - ROOM B-120AE1

1.M-111

C1

1.M-111

1T H

DEMO KEY NOTES-

SS

10-BB110

SS

1-A154

SS

1-B120A

1HT

TAB NOTES

1. FIELD SURVEY: PERFORM A FIELD SURVEY TO DOCUMENT

EXISTING AIR DISTRIBUTION CONFIGURATION OF EACH

TR. IDENTIFY BRANCH DUCT SERVING EACH TR AS WELL

AS SOURCE AND DIRECTION OF AIRFLOW TO EACH TR

AND IMPACTED AREAS.

2. TAB PLAN: PREPARE A TAB PLAN FOR EACH AREA OF

WORK. THE TAB PLAN SHALL INCLUDE A TAB OF

DIFFUSERS, REGISTERS, AND GRILLES OF IMPACTED

AREAS AND TRAVERSE LOCATIONS OF UPSTREAM

AND/OR DOWNSTREAM OF MAIN DUCT IF REQUIRED.

SUBMIT TAB PLAN TO VA FOR APPROVAL.

3. PRE-TAB: PERFORM A PRE-TAB PRIOR TO

COMMENCEMENT OF WORK IN THE SPACE. THE PRE-TAB

SHALL DOCUMENT THE EXISTING CONDITIONS OF

IMPACTED AREAS AND SHALL SERVE AS THE BASIS TO

RE-BALANCE THE HVAC SYSTEMS. MEASURE, RECORD,

AND SUBMIT PRE-TAB TO VA FOR DOCUMENTATION.

4. POST-TAB: PERFORM A POST-TAB UPON COMPLETION OF

WORK. RE-BALANCE IMPACTED AREAS TO PRE-TAB AIR

VOLUMES. IN ADDITION, PERFORM A POST-TAB ANYTIME

MODIFICATIONS ARE MADE TO THE HVAC SYSTEMS.

PROVIDE BALANCING DAMPERS AS REQUIRED TO

BALANCE THE SYSTEM. SUBMIT POST-TAB TO VA FOR

APPROVAL.

1. NO MECHANICAL DEMO.

2. DEMOLISH DUCTWORK, DIFFUSER/ GRILLE, SUPPORTS

AND ALL ASSOCIATED APPURTENANCES.

3. CAP & SEAL DUCT AIR TIGHT.

4. RELOCATE SENSORS TO OTHER SIDE OF WALL.

5. INFILL, REPAIR AND PATCH WALL OPENING TO MATCH

ADJACENT FINISH SURFACES.

6. PIPES SHOWN FOR REFERENCE.

T H

- RENO KEY NOTES

1. PROVIDE DDC SENSORS. ROUTE CONTROL

WIRING TO DDC I/O MODULE OR CONTROLLER.

2. LOCATE DDC I/O MODULE AND NEMA-1 ENCLOSURE IN

TELECOM ROOM. COORDINATE EXACT LOCATION WITH VA

AND ROOM SPACE AVAILABILITY.

3. REFER TO ELECTRICAL FOR RACK TEMPERATURE

SENSORS.

4. PROVIDE DRAIN PAN UNDER (E) PIPES, SEE DETAIL

C1/M-502.

5. PROVIDE MANUFACTURER'S AND CODE REQUIRED

CLEARANCE.

6. RS, RL, AND PUMPED CD. REFER TO OVERALL FLOOR PLAN

FOR CONTINUATION.

7. LOCATION FOR RELOCATED (E) SENSORS.

(TYP)

(E)

1. IN ADDITION…