ESBD_377870_1727471354712_582-25-03322 Attachment B Specifications.pdf
PDF 108 KB Posted
- Attached to
- Calibration Services
- Solicitation number
- 582-25-03322
- Issued by
- Texas
About this file
This document outlines the specifications for the annual factory calibration and preventative maintenance of Landtec GEM 5000 Portable Gas Analyzers. The services are requested by an unspecified state or local government entity. The calibration and assessment process includes instrument inspection, calibration of various gas measurements, oxygen cell programming, water trap kit replacement, firmware updates, and replacement of various components as needed. All equipment must be calibrated according to the manufacturer's protocols using traceable equipment that meets the manufacturer's calibration specifications. Additionally, a unique and signed Statement of Calibration must accompany each newly calibrated Landtec GEM 5000 instrument.
The document does not provide details on pricing terms, set asides, incumbents, funding requirements or sources, award dates, or the term of the contract. It focuses solely on the technical specifications and requirements for the calibration and maintenance services for the Landtec GEM 5000 Portable Gas Analyzers.
View the file
Other files
Show all 16
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Specifications for Calibration of Landtec GEM 5000 Portable Gas Analyzers
MINIMUM SPECIFICAITONS
Perform annual factory calibration and preventative maintenance on Landtec GEM 5000 Portable Gas Analyzers against Landtec/QED Environmental Systems equipment or equivalent.
This calibration and assessment process shall include the following:
1. Instrument Inspection
2. Instrument Calibration:
• Methane: 0-100% (volume)
• Carbon Dioxide: 0-100% (volume)
• Oxygen: 0-25% (volume)
• Carbon Monoxide: 0-2000 parts per million
• Hydrogen Sulfide: 0-1000 parts per million
• Atmospheric pressure, differential pressure and calculates gas flow
3. Oxygen cell programming (as needed)
4. Water trap kit (as needed)
5. Firmware Updates (as needed)
6. Replacement of sensors, transducers, pump, keypad, lens, tubing, USB cable, battery, shipping box (as needed)
7. Return freight
8. All equipment must be calibrated according to manufacturer’s protocols using traceable equipment that meets manufacturer’s calibration specifications.
9. A unique and signed Statement of Calibration must accompany each newly calibrated
Landtec GEM 5000 instrument.
10. If during the inspection of the instrument a mechanical problem is discovered that would affect its ability to pass the calibration process, a written quotation will be provided to the customer prior to servicing the instrument.
Specifications for Calibration of Landtec GEM 5000 Portable Gas Analyzers
File details come from the government source that posted it. Updated .