This document contains the specifications for the calibration and preventative maintenance of Landtec GEM 5000 Plus Portable Gas Analyzers for a state and local contract opportunity. The scope of work includes annual factory calibration, instrument inspection, calibration of various gas levels, oxygen cell programming, firmware updates, and replacement of components as needed. All equipment must be calibrated according to the manufacturer's protocols using traceable equipment. A unique and signed Statement of Calibration must accompany each newly calibrated instrument. If any mechanical problems are discovered during the inspection, a quotation will be provided to the customer prior to servicing the instrument.

The related state and local contract opportunity is for calibration services for GEM 5000 and GEM 5000 Plus devices in the state of Texas. No additional pricing terms, set asides, incumbents, funding requirements, or other salient information is provided in the document.

View the file

Other files

Show all 16

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Specifications for Calibration of Landtec GEM 5000 Plus Portable Gas Analyzers

MINIMUM SPECIFICAITONS

Calibration of Landtec GEM 5000 Plus Portable Gas Analyzers

Perform annual factory calibration and preventative maintenance on Landtec GEM 5000 Plus Portable Gas Analyzers against LandTech/QED Environmental Systems equipment or equivalent. This calibration and assessment process shall include the following:

1. Instrument Inspection

2. Instrument Calibration:

• Methane: 0-100% (volume)

• Carbon Dioxide: 0-100% (volume)

• Oxygen: 0-25% (volume)

• Carbon Monoxide: 0-2000 parts per million

• Hydrogen Sulfide: 0-1000 parts per million

• Atmospheric pressure, differential pressure and calculates gas flow

3. Oxygen cell programming (as needed)

4. Firmware Updates (as needed)

5. Replacement of sensors, transducers, pump, keypad, lens, tubing, USB cable, battery, shipping box (as needed)

6. Return freight

7. All equipment must be calibrated according to manufacturer’s protocols using traceable equipment that meets manufacturer’s calibration specifications.

8. A unique and signed Statement of Calibration must accompany each newly calibrated Landtec GEM 5000 Plus instrument.

9. If during the inspection of the instrument a mechanical problem is discovered that would affect its ability to pass the calibration process, a written quotation will be provided to the customer prior to servicing the instrument.

Specifications for Calibration of Landtec GEM 5000 Plus Portable Gas Analyzers