This document outlines the specifications for the calibration of Landtec GEM 5000 Plus Portable Gas Analyzers for a state or local government contract. The calibration and preventive maintenance services requested include instrument inspection, calibration of various gas measurements, oxygen cell programming, firmware updates, and replacement of related components as needed. The calibration must be performed according to the manufacturer's protocols using traceable equipment that meets the calibration specifications, and a unique and signed Statement of Calibration must accompany each calibrated instrument.
The document specifies that if a mechanical problem is discovered during the inspection that would affect the instrument's ability to pass the calibration process, a written quotation will be provided to the customer prior to servicing the instrument. No other pricing terms, set asides, incumbent information, funding requirements, or additional details are provided in the document.
View the file
Other files
Show all 16
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Specifications for Calibration of Landtec GEM 5000 Plus Portable Gas Analyzers
MINIMUM SPECIFICAITONS
Calibration of Landtec GEM 5000 Plus Portable Gas Analyzers
Perform annual factory calibration and preventative maintenance on Landtec GEM 5000 Plus Portable Gas Analyzers against LandTech/QED Environmental Systems equipment or equivalent. This calibration and assessment process shall include the following:
1. Instrument Inspection
2. Instrument Calibration:
• Methane: 0-100% (volume)
• Carbon Dioxide: 0-100% (volume)
• Oxygen: 0-25% (volume)
• Carbon Monoxide: 0-2000 parts per million
• Hydrogen Sulfide: 0-1000 parts per million
• Atmospheric pressure, differential pressure and calculates gas flow
3. Oxygen cell programming (as needed)
4. Firmware Updates (as needed)
5. Replacement of sensors, transducers, pump, keypad, lens, tubing, USB cable, battery, shipping box (as needed)
6. Return freight
7. All equipment must be calibrated according to manufacturer’s protocols using traceable equipment that meets manufacturer’s calibration specifications.
8. A unique and signed Statement of Calibration must accompany each newly calibrated Landtec GEM 5000 Plus instrument.
9. If during the inspection of the instrument a mechanical problem is discovered that would affect its ability to pass the calibration process, a written quotation will be provided to the customer prior to servicing the instrument.
Specifications for Calibration of Landtec GEM 5000 Plus Portable Gas Analyzers