This is a user manual for the LDO (Luminescent Dissolved Oxygen) Sensor, Edition 6, December 2022, provided in eight languages (English, French, Spanish, Portuguese, Chinese, Japanese, Korean, and Russian). The manual covers comprehensive specifications and operational guidance for dissolved oxygen measurement equipment manufactured by Hach Company.
The sensor specifications include measurement ranges of 0-20 ppm (0-20 mg/L) or 0-200% saturation, with measurement accuracy below 5 ppm at ±0.1 ppm and above 5 ppm at ±0.2 ppm. The device is IP68 rated and features wetted materials including CPVC, polyurethane, 316 stainless steel, Viton, and Noryl components. Standard and seawater probe variants are available, with Class 1, Division 2 hazardous location certified options. The manual provides detailed instructions for installation in both hazardous and non-hazardous locations, sensor configuration including temperature units, measurement units, altitude/pressure settings, and salinity correction values. Calibration procedures cover air calibration (recommended method) and sample calibration by comparison with portable dissolved oxygen meters. Maintenance requirements include sensor cleaning every 90 days, equipment inspection every 90 days, and sensor cap replacement every two years or as needed. Replacement parts and accessories are listed with corresponding item numbers for both U.S. and EU markets, including various probe types, extension cables, mounting hardware kits, and calibration bags. Troubleshooting sections address diagnostic menus, error lists, warning lists, and event logs to facilitate system monitoring and maintenance.
View the file
Other files
Show all 17
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
DOC023.97.80170
LDO Sensor 12/2022, Edition 6
User Manual Manuel d'utilisation Manual del usuario Manual do Usuário
用户手册 取扱説明書 사용 설명서
คู่มือผู้ใช้
Table of Contents
English
Français
Español
Português
中文
日本語
한국어
ไทย
Table of Contents 1 Specifications on page 3 2 General information on page 4 3 Installation on page 7 4 Operation on page 9
5 Calibration for measurements on page 12 6 Maintenance on page 14 7 Troubleshooting on page 15 8 Replacement parts and accessories on page 17
Section 1 Specifications Specifications are subject to change without notice.
The product has only the approvals listed and the registrations, certificates and declarations officially provided with the product. The usage of this product in an application for which it is not permitted is not approved by the manufacturer.
Specification Details
Wetted materials Standard Probe, Standard Class 1-Div 2 Probe
• CPVC, sensor end and cable end
• Polyurethane, over-molding on cable end and cable jacket
• 316 stainless steel body and screws
• Viton, O-ring
• Noryl, nut on the cable end
Standard Seawater Probe, Seawater Class 1-Div 2 Probe
• CPVC, sensor end and cable end
• Polyurethane, over-molding on cable end and cable jacket
• PVC seawater body
• Seawater epoxy sealant
• Noryl, nut on the cable end
IP classification IP68
Wetted materials (sensor cap) Acrylic
Measurement range (dissolved oxygen) 0 to 20 ppm (0 to 20 mg/L)
0 to 200% saturation
Measurement accuracy (dissolved oxygen)
Below 5 ppm: ± 0.1 ppm
Above 5 ppm: ± 0.2 ppm
Repeatability (dissolved oxygen) 0.1 ppm (mg/L)
Response time (dissolved oxygen) T90<40 seconds
T95<60 seconds
Resolution, sensor (dissolved oxygen) 0.01 ppm (mg/L); 0.1% saturation.
Measurement range (temperature) 0 to 50 °C (32 to 122 °F)
Measurement accuracy (temperature) ± 0.2 °C (± 0.36 °F)
Interferences No interferences from the following: H2S, pH, K+, Na+, Mg2+, Ca2+, NH4 +, Al3+, Pb2+, Cd2+, Zn2+, Cr (total), Fe2+, Fe3+, Mn2+, Cu2+, Ni2+, Co2+, CN–, NO3
–, SO4 2–, S2–, PO4 3–, Cl–, Anion Active Tensides, Crude Oils, Cl2 < 4 ppm
Storage temperature –20 to 70 °C (–4 to 158 °F)
Maximum temperature 0 to 50 ºC (32 to 122 °F)
English 3
Specification Details
Hazardous location classification (9020000-C1D2 sensor only)
Class I Division 2, Groups A–D, T4 / Class I, Zone 2 Group 2C, T4 Note: This product does not fulfill the requirements of the 94/9/EC Directive (ATEX Directive).
Certifications (9020000-C1D2 sensor only)
ETL listed to ANSI/ISA, CSA and FM standards for use in hazardous location.
Note: This product does not fulfill the requirements of the 94/9/EC Directive (ATEX Directive).
Minimum flow rate Not required
Calibration/verification Air calibration: One point, 100% water-saturated air
Sample calibration: Comparison with standard instrument
Probe immersion depth and pressure limits
Pressure Limits at 34 m (112 ft.), 345 kPa (50 psi) maximum; accuracy may not be maintained at this depth
Sensor cable 10 m (30 ft) integral cable with quick disconnect plug (all sensor types) Up to 100 m possible with extension cables (non-Class I, Division 2 sensor types only) Up to 1000 m with junction box (non-Class I, Division 2 sensor types only)
Probe weight 1.0 kg (2 lb, 3 oz)
Probe dimensions Standard probe (diameter x length): 49.53 x 255.27 mm (1.95 x
10.05 in.)
Seawater probe (diameter x length): 60.45 x 255.27 mm (2.38 x
10.05 in.)
Power requirements 12 VDC, 0.25 A, 3 W
Warranty Probe: 3 years against manufacturing defects
Sensor cap: 2 years against manufacturing defects
Section 2 General information In no event will the manufacturer be liable for damages resulting from any improper use of product or failure to comply with the instructions in the manual. The manufacturer reserves the right to make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.
2.1 Safety information
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including, without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent permitted under applicable law. The user is soley responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this equipment in any manner other than that specified in this manual.
2.2 Use of hazard information
D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
4 English
W A R N I N G Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.
C A U T I O N Indicates a potentially hazardous situation that may result in minor or moderate injury.
N O T I C E Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis.
2.3 Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.
This is the safety alert symbol. Obey all safety messages that follow this symbol to avoid potential injury. If on the instrument, refer to the instruction manual for operation or safety information.
This symbol indicates the presence of a light source that may have the potential to cause minor eye injury. Obey all messages that follow this symbol to avoid potential eye injury.
This symbol indicates the presence of devices sensitive to Electro-static Discharge (ESD) and indicates that care must be taken to prevent damage with the equipment.
Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.
2.4 Compliance and certification
C A U T I O N
This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments.
Canadian Radio Interference-Causing Equipment Regulation, ICES-003, Class A:
Supporting test records reside with the manufacturer.
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne sur les équipements provoquant des interférences.
FCC Part 15, Class "A" Limits Supporting test records reside with the manufacturer. The device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions:
1. The equipment may not cause harmful interference.
2. The equipment must accept any interference received, including interference that may cause undesired operation.
Changes or modifications to this equipment not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. This equipment has been tested
English 5 and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at their expense. The following techniques can be used to reduce interference problems:
1. Disconnect the equipment from its power source to verify that it is or is not the source of the interference.
2. If the equipment is connected to the same outlet as the device experiencing interference, connect the equipment to a different outlet.
3. Move the equipment away from the device receiving the interference.
4. Reposition the receiving antenna for the device receiving the interference.
5. Try combinations of the above.
2.5 Product overview
D A N G E R
Chemical or biological hazards. If this instrument is used to monitor a treatment process and/or chemical feed system for which there are regulatory limits and monitoring requirements related to public health, public safety, food or beverage manufacture or processing, it is the responsibility of the user of this instrument to know and abide by any applicable regulation and to have sufficient and appropriate mechanisms in place for compliance with applicable regulations in the event of malfunction of the instrument.
This sensor is designed to work with a controller for data collection and operation. The sensor can be used with several controllers. Refer to the controller-specific user manual for more information.
The primary applications for this sensor are municipal and industrial wastewater applications. LDO sensor technology does not consume oxygen, and can measure DO concentration in low or no-flow applications. Refer to Figure 1.
Figure 1 LDO sensor
1 Sensor cap 3 1-inch NPT
2 Temperature sensor 4 Connector, quick-connect (standard)
6 English
2.6 LDO Sensor component list
Make sure that all components shown in Figure 2 have been received. If any items are missing or damaged, contact the manufacturer or a sales representative immediately. Refer to Figure 2.
Figure 2 Sensor component list
1 LDO sensor1 2 Calibration bags (2x)
Section 3 Installation
3.1 Validate the sensor type
D A N G E R
Explosion hazard. Connect only peripheral components that are clearly marked as certified for Class 1, Division 2 Hazardous Locations.
N O T I C E The hazardous location certified version of this product does not fulfill the requirements of the 94/9/EC Directive (ATEX Directive).
1. Go to the connector end of the cable.
2. Read the label on the connector end of the cable. For hazardous location certified sensors, the label will show "Rated: Class 1 Division 2".
3. Examine the connector.
• Hazardous location certified sensors have a safety lock connector. Refer to Figure 3 on page 8.
• Sensors that are not certified for hazardous locations have a quick-connect connector, without a safety lock.
3.2 Connect the sensor in a hazardous location
D A N G E R
Explosion hazard. This equipment is suitable for use in non-hazardous locations or Class 1, Division 2, Groups A, B, C, D Hazardous Locations with specified sensors and options when installed per the Hazardous Location Installation Control Drawing. Always refer to the Control Drawing and applicable electrical code regulations for proper installation instructions.
1 Included user manual is not shown.
English 7
D A N G E R
Explosion hazard. Do not connect or disconnect electrical components or circuits to the equipment unless power has been removed or the area is known to be non-hazardous.
N O T I C E Use only a hazardous location certified sensor and cable lock in hazardous locations. The hazardous location certified version of this product does not fulfill the requirements of the 94/9/EC Directive (ATEX Directive).
For more information, refer to Validate the sensor type on page 7 .
1. Remove the connector cap from the controller. Keep the connector cap to seal the connector opening when the sensor is removed.
2. Connect the sensor to the controller. Refer to the controller manual for more information.
3. Close the safety lock over the connector.
4. To remove the connector safety lock, use a small flat screwdriver. Refer to Figure 3.
Figure 3 Connector safety lock
3.3 Connect the sensor in a non-hazardous location
Refer to Figure 4 to connect an LDO sensor to an sc controller. Refer to the specific sc controller manual for hard-wiring instructions.
Figure 4 Connect the LDO sensor (non-hazardous location sensor shown)
After the sensor is attached, scan for the sensor. Refer to Install the sensor on page 8.
3.4 Install the sensor
There are two options to install the sensor:
• Connect the sensor while power to the controller is off. The controller will look for and install new sensors when it is turned on.
• Connect the sensor while power to the controller is on. Use the Scan Devices command to install the new sensor:
Option Description sc200 controller Go to MENU>TEST/MAINT>SCAN DEVICE
8 English
Option Description sc100 controller Go to MENU>TEST/MAINT>SCAN SENSORS sc1000 controller Go to MENU>SYSTEM SETUP>DEVICE MANAGEMENT>SCANNING FOR NEW
DEVICES
Refer to Connect the sensor in a non-hazardous location on page 8 for digital sensor connection.
3.5 Sensor installation options
The installation and accessory options available for the sensor are supplied with installation instructions in the hardware kit. Figure 5 shows several installation options. To order installation hardware, refer to Replacement parts and accessories on page 17.
Figure 5 Installation options
1 Rail mount 4 Chain mount
2 Float mount 5 Union mount (not seawater probe compatible)
3 Air blast system mount (not seawater probe compatible)
Section 4 Operation
4.1 User navigation
Refer to the controller documentation for keypad description and navigation information.
Push the RIGHT arrow key on the controller multiple times to show more information on the home screen and to show a graphical display.
English 9
4.2 Configure the sensor
Use the Configure menu to enter identification information for the sensor and to change options for data handling and storage.
For information about sensor installation, refer to Install the sensor on page 8.
Make sure that all of the Configuration menu values are correct for the application.
1. Go to MENU, SENSOR SETUP, [Select Sensor], LDO SETUP, CONFIGURE.
2. Select an option, ENTER.
Option Description
EDIT NAME Changes the name that corresponds to the sensor on the top of the measure screen. The name is limited to 10 characters in any combination of letters, numbers, spaces or punctuation.
SET UNITS TEMP–Sets the temperature units to °C (default) or °F.
MEASURE–Set the measurement units in mg/L, ppm or % .
ALT/PRESS–Set the altitude in m or ft or set the atmospheric pressure units in mmHg or torr.
ALT/PRESS Enter the value of the altitude or atmospheric pressure.
SALINITY Enter the salinity value. Salinity range: 0.00 to 250.00 parts per thousand. Refer to Enter a salinity correction value on page 11 for more information.
SIGNAL AVERAGE Set the time interval to average signal in seconds
CLEAN INTRVL Set the time interval for sensor cleaning in days
RESET CLN INTRVL Set the time interval to the factory default value
LOG SETUP Sets the time interval for data storage in the data log—1, 2, 5, 10, 15 (default), 30, 60 minutes.
SET DEFAULTS Restores the configurable default values for the sensor. Does not change the setting for slope or offset.
4.3 Enter the atmospheric pressure value
1. Go to MENU, SENSOR SETUP, [Select Sensor], LDO SETUP, CONFIGURE, SET UNITS, AIR
PRESS/ALT UNITS.
2. Select the units, ENTER.
3. Select AIR PRESS/ALT. Enter and confirm a value from Table 1.
Table 1 Elevation (ft) and barometric pressure (mm Hg)
Ft mm Hg Ft mm Hg Ft mm Hg Ft mm Hg
0 760 3000 683 6000 613 9000 548
500 746 3500 671 6500 601 9500 538
1000 733 4000 659 7000 590 10,000 527
1500 720 4500 647 7500 579 10,500 517
2000 708 5000 635 8000 568 11,000 506
2500 695 5500 624 8500 559 11,5000 —
10 English
4.4 Enter a salinity correction value
Dissolved oxygen measurements in saline samples can show an apparent DO value that is very different from the actual DO value. To correct for the influence of dissolved salts in a sample, enter a salinity correction factor.
Note: If the presence or amount of salinity in the process is unknown, consult with the treatment facility engineering staff.
1. Use a conductivity meter to measure the conductivity of the sample in mS/cm at a reference temperature of 20 °C (68 °F).
2. Use Table 2 to estimate the salinity correction factor in parts per thousand (‰) saturation.
Note: The chloride ion concentration, in g/kg is equal to the chlorinity of the sample. Salinity is calculated with the formula: Salinity = 1.80655 × chlorinity.
Salinity can be calculated with the relationship in section 2520 B of Standard Methods for the Examination of Water and Wastewater.2,
3. Go to MENU>SENSOR SETUP>[Select Sensor]>CONFIGURE>SALINITY.
4. Enter the salinity correction factor and confirm.
Table 2 Salinity saturation (‰) per conductivity value (mS/cm) mS/cm ‰ mS/cm ‰ mS/cm ‰ mS/cm ‰
5 3 16 10 27 18 38 27
6 4 17 11 28 19 39 28
7 4 18 12 29 20 40 29
8 5 19 13 30 21 42 30
9 6 20 13 31 22 44 32
10 6 21 14 32 22 46 33
11 7 22 15 33 23 48 35
12 8 23 15 34 24 50 37
13 8 24 17 35 25 52 38
14 9 25 17 36 25 54 40
15 10 26 18 37 26
4.5 Configure linear output on the controller
Linear outputs send probe data back to the facility PLC, SCADA or other data collection system.
1. Go to the controller output setup menu.
Option Description sc200 Go to MENU>SETTINGS>sc200 SETUP>OUTPUT SETUP>[Select Output]>SET FUNCTION.
2 Standard Methods for the Examiniation of Water and Wastewater, 20th Edition. Editors Lenore S. Clesceri, Arnold E. Greenberg and Andrew D. Eaton, p. 2-48-2-29 (1998). The relationship between Chlorinity and Oxygen Solubitity is provided in the same reference in 4500-O:I p. 4-131.
English 11 sc100 Go to MENU>SYSTEM SETUP>OUTPUT SETUP>[Select Output]>SET FUNCTION.
sc1000 Go to MENU>SYSTEM SETUP>OUTPUT SETUP>[Select Output]>SET FUNCTION.
2. Set the function for the controller.
Option Description sc200 LINEAR sc100 LINEAR CONTROL (Default value) sc1000 LINEAR CONTROL (Default value)
4.6 Modbus registers
A list of Modbus registers is available for network communication. Refer to the manufacturer's website for more information.
Section 5 Calibration for measurements The sensor is calibrated to specification at the factory. The manufacturer does not recommend calibration unless periodically required by regulatory agencies. If calibration is required, let the sensor come to equilibrium with the process before calibration. Do not calibrate the sensor at setup.
Table 3 shows options for calibration.
Table 3 Calibration options
Option Description
AIR CAL Recommended calibration method. This calibration modifies the slope.
SAMPLE CAL Calibration by comparison with a hand-held DO meter . This calibration modifies the offset.
RESET DFLT CAL Resets the calibration gain (slope) and offset to the factory default: default gain=1.0; default offset=0.0
5.1 Calibration with air
Air calibration is recommended for best accuracy and repeatability.
1. Remove the sensor from the process. Use a wet cloth to clean the sensor.
2. Put the cap end of the sensor in a calibration bag with 25-50 mL of water.
3. Attach the bag to the sensor body. Make sure that the sensor cap is not in contact with the water inside the calibration bag and that no water drops are on the sensor cap. Refer to Figure 6.
4. Select MENU, SENSOR SETUP, LDO SETUP, [Select Sensor], CALIBRATE, AIR CAL, ENTER.
5. Select the option for the output signal during calibration:
Option Description
Active The instrument sends the current measured output value during the calibration procedure.
Hold The sensor output value is held at the current measured value during the calibration procedure.
Transfer A preset value is sent during calibration. Refer to the controller user manual to change the preset value.
6. The controller will show "Move the sensor to air". Wait for the value to stabilize. The display will show "Complete" and the slope correction. Push ENTER.
7. When the sensor is calibrated, put the sensor back into the process. Push ENTER.
12 English
Figure 6 LDO air calibration
If the value does not stabilize, the display will show "Unable to Calibrate" followed by an error message. Table 4 shows the error message and resolution for calibration problems.
Table 4 Calibration error messages
Message Description Resolution
Cal fail, offset high The calculated gain value is too high. Repeat the calibration.
Cal fail, offset low The calculated gain value is too low. Repeat the calibration.
Cal fail, unstable The value did not stabilize in the maximum allowed calibration time. Repeat the calibration.
5.2 Sample CAL - calibration by comparison
This calibration method uses an alternate sensor attached to a hand-held meter.
1. Put the alternate sensor into the process. Put the second sensor as close as possible to the first sensor.
2. Wait for the DO value to stabilize .
3. On the controller for the first sensor, go to MENU>SENSOR SETUP>[Select
Sensor]>CALIBRATE>SAMPLE CAL.
4. Select the option for the output signal during calibration:
Option Description
Active The instrument sends the current measured output value during the calibration procedure.
Hold The sensor output value is held at the current measured value during the calibration procedure.
Transfer A preset value is sent during calibration. Refer to the controller user manual to change the preset value.
5. The controller will show:
• "Press ENTER when stabilized"
• The current dissolved oxygen measurement
• The current temperature measurement
6. When the measurement is stable, push ENTER. The display will show an entry screen.
Note: The measurement will usually stabilize in 2 to 3 minutes.
If the value does not stabilize, the display will show "Unable to Calibrate" followed by an error message. Table 5 shows the error message and resolution for calibration problems.
English 13
Table 5 Sample cal error messages
Message Description Resolution
Cal fail, offset high The calculated offset value is too high. Repeat the calibration.
Cal fail, offset low The calculated offset value is too low. Repeat the calibration.
Cal fail, unstable The value did not stabilize in the maximum allowed calibration time. Repeat the calibration.
5.3 Exit the calibration procedure
1. During calibration, push the BACK key. Three options are shown:
Option Description
ABORT Stop the calibration. A new calibration must start from the beginning.
BACK TO CAL Return to the current calibration.
LEAVE Exit the calibration temporarily. Access to other menus is allowed while the calibration continues in the background. A calibration for a second sensor (if present) can be started.
To return to the calibration, push the MENU key and select Sensor Setup, [Select Sensor].
2. Select one of the options. Confirm.
5.4 Reset calibration defaults
Calibration settings can be reset to the factory defaults. Gain and offset values are set to 1.0 and 0.0, respectively.
1. Go to MENU>SENSOR SETUP>[Select Sensor]>CALIBRATE>RESET CAL DEFLT.
2. The display will show a confirmation message. Confirm to reset the sensor to the factory default calibration curve.
Section 6 Maintenance D A N G E R
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.
D A N G E R Explosion hazard. Do not connect or disconnect electrical components or circuits to the equipment unless power has been switched off or the area is known to be non-hazardous.
D A N G E R Explosion hazard. Substitution of components may impair suitability for Class 1, Division 2. Do not replace any component unless power has been switched off and the area is known to be non-hazardous.
N O T I C E The hazardous location certified version of this product does not fulfill the requirements of the 94/9/EC Directive (ATEX Directive).
6.1 Maintenance schedule
The maintenance schedule shows minimum intervals for regular maintenance tasks. Perform maintenance tasks more frequently for applications that cause electrode fouling.
14 English
Note: Do not disassemble the probe for maintenance or cleaning.
Maintenance task Recommended minimum frequency
Clean the sensor 90 days
Inspect the sensor for damage 90 days
Calibrate the sensor As recommended by regulatory agencies
6.2 Clean the sensor
Clean the exterior of the sensor with a soft, wet cloth.
Note: If the sensor cap must be removed for cleaning, do not expose the interior of the cap to direct sunlight for extended periods of time.
6.3 Set or change the clean interval
Application conditions may need shorter or longer durations between manual sensor cleanings. The default clean interval is 0 days. To change the interval, refer to the steps in this procedure.
1. Go to MENU>SENSOR SETUP>[Select Sensor]>CONFIGURE>CLEAN INTRVL.
2. Change the shown value as needed. Confirm the change.
• To turn off the clean interval, set the value to '0'.
6.4 Change the sensor cap
W A R N I N G
Potential explosion hazard. The sensor setup cap is not rated for hazardous location use.
Replacement sensor caps and setup caps are shipped with installation instructions. Refer to the included instructions to change the cap. After the new sensor cap is installed, make sure that the lot number on the sensor cap is the same lot number that is read by Modbus. The sensor uses the calibration information from the sensor cap.
For best performance and accuracy, replace the sensor cap:
• Every two years, or more frequently if necessary
• When routine inspection shows significant erosion of the sensor cap
Section 7 Troubleshooting
7.1 Diagnostic and test menu
The diagnostic and test menu shows current and historical information about the LDO sensor.
To access the diagnostic and test menu, go to MENU>SENSOR SETUP>[Select Sensor]>DIAG/TEST.
Refer to Table 6.
Table 6 DIAG/TEST menu
Option Description
SENSOR INFO SOFTWARE VERS—Shows the installed software version
BOOT VERSION—Shows the installed boot version
DRIVER VERS—Shows the installed software driver version
LOT CODE Shows the sensor cap manufacturing lot
English 15
Table 6 DIAG/TEST menu (continued)
SERIAL NUMBER Sensor serial number
GAIN CORR Adjust the calibration gain value.
Range: 0.50 to 2.00
OFFSET CORR Adjust the calibration offset value (mg/L or ppm).
Range: –3.00 to +3.00
PHASE DIAG Shows the phase for total, red and blue wavelengths. Updates once per second.
AMPL DIAG Shows the amplitude for red and blue wavelengths. Updates once per second.
DAYS TO CLEAN Shows the number of days until the next scheduled manual cleaning.
SENSOR LIFE Shows the number of days until the next scheduled sensor cap replacement
7.2 Error list
If an error occurs, the reading on the measurement screen flashes. Output behavior is determined by controller settings. Refer to the controller manual for details.
To show the current sensor errors, go to MENU>DIAGNOSTICS>[Select Sensor]>ERROR LIST.
Refer to Table 7.
Table 7 Error list for the LDO sensor
Error Possible cause Resolution
RED AMPL LOW (Value is below 0.01)
OR
BLUE AMPL LOW (Value is below 0.01)
The sensor cap is not installed, or is not installed correctly.
Remove the sensor cap and install it again.
The light path is blocked in the sensor cap.
Inspect the inside of the sensor cap and lens.
The sensor is not operating correctly.
Make sure that the LED is flashing.
Contact the manufacturer.
7.3 Warning list
When the warning icon flashes (sc100 and sc200) or when the screen turns yellow (sc1000), a message is shown on the bottom of the measurement screen. On the sc1000, the screen turns yellow to show a warning. To show the current sensor warnings, go to MENU> DIAGNOSTICS>[Select Sensor]>WARNING LIST. Refer to Table 8.
Table 8 Sensor warning list
Warning Definition Resolution
EE SETUP ERR Storage is corrupt. The values have been set to the factory default.
Contact technical support.
EE RSRVD ERR
TEMP < 0 C The process temperature is below 0 °C (32 °F)
Increase the process temperature or stop use until the process temperature is in the sensor specification range.
TEMP > 50 C The process temperature is above 50 °C (120 °F)
Decrease the process temperature or stop use until the process temperature is in the sensor specification range.
RED AMPL LOW Value falls below 0.03 Refer to Table 7 on page 16 .
RED AMPL HIGH Value is greater than 0.35 Call technical support.
16 English
Table 8 Sensor warning list (continued)
Warning Definition Resolution
BLUE AMPL LOW Value is below 0.03 Refer to Table 7 on page 16.
BLUE AMPL HIGH Value is greater than 0.35 Call technical support.
CAP CODE FAULT The sensor cap code has become corrupt.
The code has been reset automatically to the default cap and lot codes.
Complete the sensor setup cap procedure.
If no setup cap is available for the sensor cap, call technical support.
7.4 Event list
The Event list keeps a log of changes to how data is recorded by the sensor. To show sensor events, go to MENU>DIAGNOSTICS>[Select Sensor]>EVENT LIST.
Refer to Table 9.
Table 9 Event list for the sensor
Event Description
ALT/PRESSURE UNIT CHANGE Atmospheric pressure or altitude units have changed.
ALT/PRESSURE CHANGE The value for altitude or atmospheric pressure has changed.
TEMP UNIT CHANGE The units for temperature have changed.
MEAS UNIT CHANGE A new unit of measurement has changed.
SALINITY CHANGE The value for salinity has changed.
SET DEFAULT Sensor settings have been reset to the default values.
SENSOR SETUP CHANGE The sensor setup has changed.
CLEAN INTERVAL TIMER CHANGE The time between sensor cleaning has changed.
SENSOR CAP LIFE TIMER CHANGE The time between sensor cap replacements has changed.
Section 8 Replacement parts and accessories Use only replacement parts approved by the manufacturer. Use of non-approved parts may cause personal injury, damage to the instrument or equipment malfunction.
Replacement items
Description Item no. (US / EU)
LDO Probe, with one sensor cap and 2 calibration bags 9020000 / LXV416.99.20001
LDO Probe for seawater, with one sensor cap and two calibration bags 9020000-SW / —
LDO Probe for seawater hazardous locations, with one sensor cap and two calibration bags
9020000-C1D2-SW / —
LDO Probe for hazardous locations, with one sensor cap and 2 calibration bags 9020000-C1D2 / —
Sensor cap, replacement (includes the sensor setup cap, which is not rated for use in Class 1, Division 2 hazardous locations)
9021100 / 9021150
Accessories
Description Item no. (US / EU)
Sensor cable lock for hazardous locations 6139900 / —
Cable, sensor extension, Class 1, Division 2 Hazardous Location, 1 m (3.3 ft) 6122402 / —
English 17
Description Item no. (US / EU)
Cable, sensor extension, Class 1, Division 2 Hazardous Location, 7m (23 ft) 5796002 / —
Cable, sensor extension, Class 1, Division 2 Hazardous Location, 15 m (49.21 ft)
5796102 / —
Cable, sensor extension, Class 1, Division 2 Hazardous Location, 31 m (101.71 ft)
5796202 / —
High output air blast cleaning system, 115 V (not rated for use in hazardous locations)
6860000 / 6860003.99.0001
High output air blast cleaning system, 230 V (not ATEX rated for use in hazardous locations)
6860100 / 6860103.99.0001
Calibration bag (1x) 5796600 / 5796600
Cable, sensor extension, non-hazardous location, 7.7 m (25 ft)3 US: 5796000, 7.7 m (25 ft)
EU: LZX849, 10 m (33 ft)
Hardware kit for pipe mount (PVC) 9253000 / LZY714.99.21810
Hardware kit for float mount (PVC) 9253100 / LZX914.99.42200
Hardware kit for air blast mount 9253500 / LZY812
Hardware kit for chain mount (stainless steel) — / LZX914.99.11200
Hardware kit for union mount 9257000 / 9257000
HQd meter with LDO rugged probe (not rated for use in hazardous locations) 8505200 / HQ40D.99.310.000
3 15 m (49 ft) and 30 m (98 ft) also available
18 English
Table des matières 1 Caractéristiques à la page 19 2 Généralités à la page 20 3 Installation à la page 23 4 Fonctionnement à la page 26
5 Etalonnage pour les mesures à la page 28 6 Maintenance à la page 31 7 Dépannage à la page 32 8 Pièces de rechange et accessoires à la page 35
Section 1 Caractéristiques Les caractéristiques techniques peuvent être modifiées sans préavis.
Le produit ne possède que les homologations mentionnées et les enregistrements, certificats et déclarations officiellement fournis avec le produit. L'utilisation de ce produit dans une application pour laquelle il n'est pas autorisé n'est pas approuvée par le fabricant.
Caractéristiques Détails
Matériaux immergés Sonde standard, sonde standard classe 1 division 2
• CPVC, extrémité du capteur et extrémité du câble
• Polyuréthanne, surmoulage de l'extrémité du câble et de la gaine de câble
• Corps et vis en acier inoxydable 316
• Viton, joint torique
• Noryl, écrou sur l'extrémité du câble
Sonde standard pour eau de mer, sonde pour eau de mer classe 1 division 2
• CPVC, extrémité du capteur et extrémité du câble
• Polyuréthanne, surmoulage de l'extrémité du câble et de la gaine de câble
• Corps en PVC pour eau de mer
• Matériau d'étanchéité époxy pour eau de mer
• Noryl, écrou sur l'extrémité du câble
Classification IP IP68
Matériaux immergés (Capsule de sonde)
Acrylique
Plage de mesure (oxygène dissous) 0 à 20 ppm (0 à 20 mg/l)
0 à 200 % de saturation
Précision des mesures (oxygène dissous)
En dessous de 5 ppm : ± 0,1 ppm
Au-dessus de 5 ppm : ± 0,2 ppm
Répétabilité (oxygène dissous) 0,1 ppm (mg/l)
Temps de réponse (oxygène dissous) T90<40 secondes
T95<60 secondes
Résolution, capteur (oxygène dissous) 0,01 ppm (mg/l) ; 0,1 % de saturation
Plage de mesure (température) 0 à 50 °C (32 à 122 °F)
Précision des mesures (température) ± 0,2 °C
Interférences Les éléments suivants ne causent aucune interférence : H2S, pH, K+, Na+, Mg2+, Ca2+, NH4 +, Al3+, Pb2+, Cd2+, Zn2+, Cr (total), Fe2+, Fe3+, Mn2+, Cu2+, Ni2+, Co2+, CN–, NO3
–, SO4 2–, S2–, PO4 3–, Cl–, surfactants anioniques actifs, pétrole, Cl2 < 4 ppm
Français 19
Caractéristiques Détails
Température de stockage -20 à 70 °C (–4 à 158 °F)
Température maximale 0 à 50 °C
Classification pour zone dangereuse (capteur 9020000-C1D2 uniquement)
Classe I Division 2, Groupes A–D, T4 / Classe I, Zone 2 Groupe 2C, T4 Remarque : ce produit n'est pas conforme à la norme 94/9/EC (norme ATEX).
Certifications (capteur 9020000- C1D2 uniquement)
Certifié conforme aux normes ANSI/ISA, CSA et FM par l'ETL pour une utilisation en environnement dangereux.
Remarque : ce produit n'est pas conforme à la norme 94/9/EC (norme ATEX).
Débit minimal Non requis
Etalonnage/vérification Etalonnage à l'air : un point, air saturé d'eau à 100 %
Etalonnage par échantillon : comparaison avec un instrument de référence (2)
Profondeur d'immersion et limites de pression de la sonde
Limites de pression à 34 m, 345 kPa maximum ; il est possible que la précision ne soit plus assurée à cette profondeur
Câble du capteur Câble intégral de 10 m avec connecteur rapide (tous les types de capteur) Extension possible jusqu'à 100 m à l'aide de câbles supplémentaires (uniquement les types de capteur non-classe I, Division 2) Extension possible jusqu'à 1 000 m à l'aide d'une boîte de raccordement (uniquement les types de capteur non-classe I, Division 2)
Poids de la sonde 1 kg
Dimensions de la sonde Sonde standard (diamètre x longueur) : 49,53 x 255,27 mm
Sonde pour eau de mer (diamètre x longueur) : 60,45 x 255,27 mm
Alimentation électrique requise 12 V CC, 0,25 A, 3 W
Garantie Sonde : 3 ans de couverture des défauts de fabrication
Capsule du capteur : 2 ans de couverture des défauts de fabrication
Section 2 Généralités En aucun cas le fabricant ne pourra être tenu responsable des dommages résultant d'une utilisation incorrecte du produit ou du non-respect des instructions du manuel. Le constructeur se réserve le droit d'apporter des modifications à ce manuel et aux produits décrits, à tout moment, sans avertissement ni obligation. Les éditions révisées se trouvent sur le site Internet du fabricant.
2.1 Consignes de sécurité
Le fabricant décline toute responsabilité quant aux dégâts liés à une application ou un usage inappropriés de ce produit, y compris, sans toutefois s'y limiter, des dommages directs ou indirects, ainsi que des dommages consécutifs, et rejette toute responsabilité quant à ces dommages dans la mesure où la loi applicable le permet. L'utilisateur est seul responsable de la vérification des risques d'application critiques et de la mise en place de mécanismes de protection des processus en cas de défaillance de l'équipement.
Veuillez lire l'ensemble du manuel avant le déballage, la configuration ou la mise en fonctionnement de cet appareil. Respectez toutes les déclarations de prudence et d'attention. Le non-respect de cette procédure peut conduire à des blessures graves de l'opérateur ou à des dégâts sur le matériel.
Assurez-vous que la protection fournie avec cet appareil n'est pas défaillante. N'utilisez ni n'installez cet appareil d'une façon différente de celle décrite dans ce manuel.
20 Français
2.2 Informations sur les risques d'utilisation
Indique une situation de danger potentiel ou imminent qui, si elle n'est pas évitée, entraîne des blessures graves, voire mortelles.
A V E R T I S S E M E N T Indique une situation de danger potentiel ou imminent qui, si elle n'est pas évitée, peut entraîner des blessures graves, voire mortelles.
A T T E N T I O N Indique une situation de danger potentiel qui peut entraîner des blessures mineures ou légères.
A V I S Indique une situation qui, si elle n'est pas évitée, peut occasionner l'endommagement du matériel. Informations nécessitant une attention particulière.
2.3 Etiquettes de mise en garde
Lisez toutes les informations et toutes les étiquettes apposées sur l’appareil. Des personnes peuvent se blesser et le matériel peut être endommagé si ces instructions ne sont pas respectées. Tout symbole sur l'appareil renvoie à une instruction de mise en garde dans le manuel.
Ceci est le symbole d'alerte de sécurité. Respectez tous les messages de sécurité qui suivent ce symbole afin d'éviter tout risque de blessure. S'ils sont apposés sur l'appareil, se référer au manuel d'utilisation pour connaître le fonctionnement ou les informations de sécurité.
Ce symbole signale la présence d'une source lumineuse entraînant un risque de détérioration légère de la vue. Veillez à vous conformer à tous les messages de sécurité qui suivent ce symbole afin d'éviter une potentielle détérioration de votre vue.
Ce symbole indique la présence d'appareils sensibles aux décharges électrostatiques et indique que des précautions doivent être prises afin d'éviter d'endommager l'équipement.
Le matériel électrique portant ce symbole ne doit pas être mis au rebut dans les réseaux domestiques ou publics européens. Retournez le matériel usé ou en fin de vie au fabricant pour une mise au rebut sans frais pour l'utilisateur.
2.4 Conformité et certification
A T T E N T I O N
Cet équipement n'est pas conçu pour être utilisé dans des environnements résidentiels et peut ne pas offrir une protection adéquate à la réception radio dans de tels environnements.
Règlement canadien sur les équipements causant des interférences radio, ICES-003, Classe A :
Les données d'essai correspondantes sont conservées chez le constructeur.
Cet appareil numérique de classe A respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.
Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne sur les équipements provoquant des interférences.
FCC part 15, limites de classe A :
Les données d'essai correspondantes sont conservées chez le constructeur. L'appareil est conforme à la partie 15 de la règlementation FCC. Le fonctionnement est soumis aux conditions suivantes :
Français 21
1. Cet équipement ne peut pas causer d'interférence nuisible.
2. Cet équipement doit accepter toutes les interférences reçues, y compris celles qui pourraient entraîner un fonctionnement inattendu.
Les modifications de cet équipement qui n’ont pas été expressément approuvées par le responsable de la conformité pourraient annuler l’autorité dont l’utilisateur dispose pour utiliser cet équipement.
Cet équipement a été testé et déclaré conforme aux limites définies pour les appareils numériques de classe A, conformément à la section 15 de la réglementation FCC. Ces limites ont pour but de fournir une protection raisonnable contre les interférences néfastes lorsque l’équipement fonctionne dans un environnement commercial. Cet équipement génère, utilise et peut irradier l'énergie des fréquences radio et, s'il n'est pas installé ou utilisé conformément au mode d'emploi, il peut entraîner des interférences dangereuses pour les communications radio. Le fonctionnement de cet équipement dans une zone résidentielle risque de causer des interférences nuisibles, dans ce cas l'utilisateur doit corriger les interférences à ses frais Les techniques ci-dessous peuvent permettre de réduire les problèmes d'interférences :
1. Débrancher l'équipement de la prise de courant pour vérifier s'il est ou non la source des perturbations
2. Si l'équipement est branché sur le même circuit de prises que l'appareil qui subit des interférences, branchez l'équipement sur un circuit différent.
3. Eloigner l'équipement du dispositif qui reçoit l'interférence.
4. Repositionner l’antenne de réception du périphérique qui reçoit les interférences.
5. Essayer plusieurs des techniques ci-dessus à la fois.
2.5 Présentation du produit
D A N G E R
Dangers chimiques ou biologiques. Si cet appareil est utilisé pour la surveillance d'un procédé de traitement et/ou d'un système de dosage de réactifs chimiques auxquels s'appliquent des limites réglementaires et des normes de surveillance motivées par des préoccupations de santé et de sécurité publiques ou de fabrication et de transformation d'aliments ou de boissons, il est de la responsabilité de l'utilisateur de cet appareil de connaître et d'appliquer les normes en vigueur et d'avoir à sa disposition suffisamment de mécanismes pour s'assurer du respect de ces normes dans l'éventualité d'un dysfonctionnement de l'appareil.
Ce capteur est conçu pour fonctionner avec un contrôleur assurant la collecte de données et le fonctionnement. Il est possible d'utiliser le capteur avec plusieurs transmetteurs. Pour plus d'informations, reportez-vous au manuel d'utilisation du transmetteur.
Ce capteur est essentiellement conçu pour le traitement municipal ou industriel des eaux usées. La technologie de capteur LDO ne consomme pas d'oxygène et permet de mesurer la concentration d'oxygène dissous pour un débit faible ou nul. Voir la Figure 1.
Figure 1 Capteur LDO
1 Capsule de sonde 3 Filetage 1 pouce NPT (2)
2 Capteur de température 4 Connecteur, connexion rapide (standard)
22 Français
2.6 Liste des composants du capteur LDO
Assurez-vous d'avoir bien reçu tous les composants illustrés dans la Figure 2. Si des éléments manquent ou sont endommagés, contactez immédiatement le fabricant ou un représentant commercial. Voir la Figure 2.
Figure 2 Liste des composants du capteur
1 Capteur LDO1 2 Sachets d'étalonnage (2)
Section 3 Installation
3.1 Validation du type de capteur
D A N G E R
Risque d'explosion. Ne branchez que des composants périphériques clairement indiqués comme certifiés pour usage en endroits dangereux de Classe 1, Division 2.
A V I S La version de ce produit certifiée pour les environnements dangereux n'est pas conforme à la norme 94/9/EC (norme ATEX).
1. Saisissez l'extrémité du câble comportant le connecteur.
2. Lisez l'étiquette placée à cette extrémité du câble. Sur les capteurs certifiés pour une utilisation en environnement dangereux, l'étiquette affiche « Certifié : Classe 1 Division 2 ».
3. Examinez le connecteur.
• Les capteurs certifiés pour une utilisation en environnement dangereux comportent un connecteur avec verrouillage de sécurité. Voir la Figure 3 à la page 24.
• Les capteurs non certifiés pour une utilisation en environnement dangereux comportent une prise de déconnexion rapide, sans verrouillage de sécurité.
3.2 Connexion du capteur dans un environnement dangereux
D A N G E R
Risque d’explosion. Cet équipement convient à une utilisation au sein d'environnements non dangereux ou au sein d'environnements dangereux Classe 1, Division 2, Groupes A, B, C et D avec capteurs et options spécifiées installées selon la fiche technique relative à l'installation au sein d'environnements dangereux. Reportez-vous toujours aux instructions de la fiche technique et des réglementations en vigueur en matière de normes électriques pour une installation conforme.
1 Le manuel fourni n'est pas représenté.
Français 23
Risque d'explosion. Ne branchez ni ne débranchez aucun composant électrique ou circuit sur l'équipement avant de vous être assuré que l'alimentation a été coupée et que l'emplacement est sécurisé.
A V I S Dans un environnement dangereux, utilisez uniquement un capteur certifié pour ce type d'environnement et un verrou de câble. La version de ce produit certifiée pour les environnements dangereux n'est pas conforme à la norme 94/9/EC (norme ATEX).
Pour plus de renseignements, référez-vous à la section Validation du type de capteur à la page 23 .
1. Retirez le bouchon (2) du connecteur du transmetteur. Conservez le bouchon (2) du connecteur afin de protéger l'ouverture du connecteur une fois le capteur retiré.
2. Connectez le capteur au transmetteur. Pour plus d'informations, reportez-vous au manuel du transmetteur.
3. Verrouillez le dispositif de sécurité du connecteur.
4. Pour retirer le verrouillage de sécurité du connecteur, utilisez un petit tournevis plat. Voir la
Figure 3.
Figure 3 Verrouillage de sécurité du connecteur
3.3 Connexion du capteur dans un environnement non dangereux
Voir la Figure 4 pour la connexion d'un capteur LDO à un transmetteur sc. Pour les instructions relatives au câblage, reportez-vous au manuel du transmetteur sc.
Figure 4 Connexion du capteur LDO (illustration correspondant à un capteur pour environnement non dangereux)
Une fois le capteur fixé, détectez-le à l'aide du logiciel. Voir la section Installation du capteur à la page 24.
3.4 Installation du capteur
Deux méthodes permettent d'installer le capteur :
24 Français
• Connectez le capteur lorsque le transmetteur est hors tension. Une fois sous tension, le transmetteur recherche et installe tout nouveau capteur.
• Connectez le capteur lorsque le transmetteur est sous tension. Installez le nouveau capteur à l'aide de la commande Scan Devices (Rechercher les périphériques) :
Option Description
Transmetteur sc200 Sélectionnez MENU>TEST/MAINT>RECHERCHE CAPTEUR
Transmetteur sc100 Sélectionnez MENU>TEST/MAINT>RECHERCHE CAPTEUR
Transmetteur sc1000 Sélectionnez MENU>SYSTEM SETUP (CONFIG SYSTEME)>CONTROLE
APPAREILS>RECHERCHE NOUVEAUX APPAREILS
Pour plus d'informations sur la connexion d'un capteur numérique, voir la section Connexion du capteur dans un environnement non dangereux à la page 24.
3.5 Options d'installation du capteur
Les options d'installation et les accessoires du capteur sont fournis avec les instructions nécessaires dans le kit de fixation. La Figure 5 présente plusieurs options d'installation. Pour commander du matériel d'installation, voir la section Pièces de rechange et accessoires à la page 35.
Figure 5 Options d'installation
1 Montage sur rail 4 Montage sur chaîne
2 Montage sur flotteur 5 Montage raccord union (non compatible avec sonde pour eau de mer)3 Montage sur système de soufflage d'air (non compatible avec sonde pour eau de mer)
Français 25
Section 4 Fonctionnement
4.1 Navigation utilisateur
Consultez la documentation du transmetteur pour une description du clavier et des informations de navigation.
Appuyez plusieurs fois sur la touche flèche vers la DROITE du transmetteur pour afficher plus d'informations sur l'écran d'accueil ainsi qu'un graphique.
4.2 Configuration du capteur
Utiliser le menu Configurer pour entrer les informations d'identification du capteur et modifier les options de gestion et de stockage de données.
Pour plus d'informations sur l'installation du capteur, voir Installation du capteur à la page 24.
Assurez-vous que toutes les valeurs du menu de configuration sont appropriées à l'utilisation de l'appareil.
1. Allez dans MENU >CONFIGURATION CAPTEUR > [Sélectionner le capteur] >CONFIGURATION LDO, CONFIGURER.
2. Sélectionnez une option, puis appuyez sur ENTREE.
Option Description
EDITER NOM Modifie le nom correspondant au capteur en haut de l'écran de mesure. Le nom est limité à 10 caractères avec n'importe quelle combinaison de lettres, chiffres, espaces ou ponctuation.
DEFINITION UNITES TEMP : règle les unités de température en °C (par défaut) ou °F.
MEASURE (MESURE) : règle les unités de mesure en mg/l, ppm ou %.
ALT./PRESS. : définit l'altitude en m ou ft, ou les unités de pression atmosphérique en mmHg ou torr.
ALT/PRESS Saisissez la valeur d'altitude ou de pression atmosphérique.
SALINITE Saisissez la valeur de salinité. Plage de salinité : 0 à 250 parties par milliers.
Pour en savoir plus, reportez-vous à la section Saisie d'une valeur de correction de la salinité à la page 27.
MOYENNE SIGNAL Définissez l'intervalle de temps pour la moyenne du signal en secondes.
INTERVALLE NETTOYAGE Définit l'intervalle de temps en jours pour le nettoyage du capteur
RAZ INT NETTOY Rétablit toutes les options de l'appareil de mesure à leur valeur par défaut
PROGRAMMATION
HISTORIQUE
Définit l'intervalle de temps pour l'enregistrement des données dans le journal des données ; 1, 2, 5, 10, 15 (par défaut), 30, 60 minutes.
SET VAL/DEFAUT Restaure les valeurs par défaut des options de configuration du capteur. Les réglages de pente et de décalage ne sont pas affectés.
4.3 Saisie de la valeur de pression atmosphérique
1. Allez dans MENU, CONFIGURATION CAPTEUR, [Sélectionner capteur], CONFIGURATION
LDO, CONFIGURER, DEFINIR LES UNITES, UNITES DE PRESS./D'ALT. DE L'AIR.
2. Sélectionnez les unités et appuyez sur ENTREE
3. SELECTIONNEZ PRESS./ALT. DE L'AIR. Saisissez et confirmez une valeur à partir du
Tableau 1.
26 Français
Tableau 1 Altitude (pi) et pression barométrique (mm Hg)
Pi mm Hg Pi mm Hg Pi mm Hg Pi mm Hg
0 760 3 000 683 6 000 613 9 000 548
500 746 3 500 671 6 500 601 9 500 538
1 000 733 4 000 659 7 000 590 10 000 527
1 500 720 4 500 647 7 500 579 10 500 517
2 000 708 5 000 635 8 000 568 11 000 506
2 500 695 5 500 624 8 500 559 11 500 —
4.4 Saisie d'une valeur de correction de la salinité
Il est possible que la valeur DO affichée par les mesures d'oxygène dissous des échantillons salins soit très différente de la valeur DO réelle.