This is a User Manual for the SC4500 Controller (Document DOC343.52.90690, Edition 13, April 2025) manufactured by Hach Company. The SC4500 is a 2-channel controller designed for digital analytical devices such as sensors and analyzers, with optional expansion modules for various applications.

The controller features a 3.5-inch TFT color touchscreen display and supports multiple installation methods including wall, pole, and panel mounting. Electrical specifications include AC power requirements of 100–240 VAC ±10%, 50/60 Hz at 1 A, or DC power of 18–28 VDC at 2.5 A. The device is rated UL50E type 4X with IP 66 protection and operates in temperatures from –20 to 60°C. Key capabilities include two high-voltage relays (SPDT) with maximum switching ratings of 5 A resistive/1 A pilot duty at 100–240 VAC for AC controllers, optional low-voltage relay modules with four relays, configurable analog outputs (0–20 mA or 4–20 mA), and digital communication options including Profibus DP, Modbus RS232/RS485, Modbus TCP, EtherNet/IP, and PROFINET. The manual provides comprehensive installation guidance covering mechanical mounting, electrical connections, network connectivity (LAN, WiFi, or cellular), sensor configuration, firmware updates, and system operation. Optional expansion modules support analog inputs/outputs, pH/ORP/conductivity measurements, amperometry, and advanced control functions through various RTC (Real Time Control) modules for water treatment applications.

View the file

Other files

File Type Posted
Atch-01 PWS Attachment 8.8.4. Digital Analyzer US – Field Service Report.pdf PDF
Atch-01 WWTP OM Services PWS dated 16 Jan 2026.pdf PDF
Atch-01 PWS Attachment 8.7. Short Stay Wastewater Treatment Facility O&M Manual Book 2.pdf PDF
Atch-01 PWS Attachment 8.1. Short NPDES Stay Permit Jan 2026.pdf PDF
Atch-01 PWS Attachment 8.3. Short Stay Wastewater Treatment Plant Site Layout Drawings and Photos.pdf PDF
Atch-01 PWS Attachment 8.8.3. Digital Analyzer Install Inspection.pdf PDF
Atch-01 PWS Attachment 8.9. Short Stay Wastewater Treatment Plant PMI Spreadsheet.pdf PDF
Atch-01 PWS Attachment 8.7. Short Stay Wastewater Treatment Facility O&M Manual Book 1.pdf PDF
Atch-01 PWS Attachment 8.2. Short Stay Wastewater Treatment Plant Layout As-builts.pdf PDF
Atch-01 PWS Attachment 8.7. Short Stay Wastewater Treatment Facility O&M Manual Book 3.pdf PDF
Atch-01 PWS Attachment 8.8.2. Digital Analyzer Certificate of Instrument Performance.pdf PDF
Atch-01 PWS Attachment 8.8.7. pH ORP Sensor Manual.pdf PDF
Atch-01 PWS Attachment 8.8.6. Output Module Manual.pdf PDF
Atch-01 PWS Attachment 8.8.5. LDO Sensor Manual.pdf PDF
Atch-01 PWS Attachment 8.5. Short Stay Water Distribution Map.pdf PDF
Atch-01 PWS Attachment 8.6. Short Stay RV Dump Station Operation and Maintenance Manual.pdf PDF
Atch-01 PWS Attachment 8.4. Short Stay Fire Line and Hydrants Map.pdf PDF
Show all 17

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

DOC343.52.90690

SC4500

04/2025, Edition 13

User Manual

DOC343.52.90690

Table of Contents Section 1 Specifications Section 2 General information

2.1 Safety information

2.1.1 Use of hazard information

2.1.2 Precautionary labels

2.1.3 EMC compliance

2.2 Product overview

2.2.1 Optional expansion modules

2.3 Product components

Section 3 Installation

3.1 Installation guidelines

3.2 Mechanical installation

3.2.1 Attach the instrument to a wall

3.2.2 Attach the instrument to a pole

3.2.3 Install the instrument in a panel

3.3 Electrical installation

3.3.1 Electrical connectors and fittings

3.3.2 Electrostatic discharge (ESD) considerations

3.3.3 Power connections

3.3.3.1 Open the controller cover

3.3.3.2 Remove the high-voltage barrier

3.3.3.3 Wiring for power

3.3.3.4 Connect conduit or a power cord

3.3.4 Connect the high-voltage relays

3.3.5 Install an expansion module

3.4 Close the cover

3.5 Connect measurement devices

Section 4 User interface and navigation

4.1 Measurement window

4.2 Graph

Section 5 Startup

5.1 Menu structure

5.2 Enter initial settings

Section 6 Operation

6.1 Configure the controller settings

6.2 Configure menu access

6.3 Connect the instrument to a network

6.3.1 LAN connection

6.3.2 WiFi network connection

6.3.3 Cellular network connection

6.3.3.1 Configuration with Hach Cellular (Telenor) SIM

6.3.3.2 Configuration with customer supplied SIM

6.4 Configure the sensors or analog inputs

6.5 Connect a USB flash drive (optional)

6.5.1 Data download

6.5.2 Install controller firmware updates

6.5.3 Install sensor firmware updates

6.5.4 Restore the controller firmware

6.5.5 Manage a software module

6.5.5.1 Install a software module

6.5.5.2 Delete a software module

6.5.5.3 Install a firmware update for an RTC module

6.6 Configure the outputs

6.6.1 Configure the high voltage relays

6.6.2 Configure the low voltage relays

6.6.3 Configure the analog outputs

6.6.4 Configure the Modbus RS232/RS485

6.6.4.1 Diagnostics/Test menu—Modbus RS232/RS485

6.6.5 Configure the telegram

6.6.6 Configure the Modbus TCP

6.6.7 Configure the Profibus DP

6.6.8 Configure the EtherNet/IP

6.6.9 Configure the PROFINET

6.6.10 Hold or release the analog outputs

6.7 Hach Controller Network Bus configuration

6.8 Prognosys diagnostic system

6.9 Claros

6.10 Enter license key

6.11 Show the instrument information

Section 7 Maintenance

7.1 Clean the instrument

7.2 Unlock the controller cover

7.3 Fuse replacement

7.4 Battery replacement

Section 8 Troubleshooting Section 9 Replacement parts and accessories Appendix A Telegram tag list Appendix B Open-source code and licenses

Table of Contents

Section 1 Specifications Specifications are subject to change without notice.

Specification Details

Dimensions (W x H x D) ½ DIN-144 x 144 x 192 mm (5.7 x 5.7 x 7.6 in.)

Enclosure UL50E type 4X, IEC/EN 60529–IP 66, NEMA 250 type 4X Metal enclosure with a corrosion-resistant finish

Weight 1.7 kg (3.7 lb) (Controller weight without optional expansion modules)

Pollution degree Environment: 4; instrument: 2

Overvoltage category II

Protection class I, connected to protective earth

Environmental conditions Indoor and outdoor use

Power requirements AC controller: 100–240 VAC ±10%, 50/60 Hz; 1 A (50 VA with 8W sensor load, 100VA with 28W sensor load) DC controller: 18–28 VDC; 2.5 A (12W with 9W sensor load, 36W with 20 W sensor load)

Operating temperature –20 to 60 °C (–4 to 140 °F) (8 W (AC)/9 W (DC) sensor load) –20 to 45 °C (–4 to 113 °F) (28 W (AC)/20 W (DC) sensor load) Linear derating between 45 and 60 °C (–1.33 W/°C)

Storage temperature –20 to 70 °C (–4 to 158 °F)

Relative humidity 0 to 95%, non-condensing

Altitude 3000 m (9842 ft) maximum

Display 3.5-inch TFT color display with capacitive touchpad

Measurement Two device, digital SC connectors

Relays (high voltage) Two relays (SPDT);

Wire gauge: 0.75 to 1.5 mm2 (18 to 16 AWG) AC controller Maximum switching voltage: 100–240 VAC Maximum switching current: 5 A Resistive/1 A Pilot Duty Maximum switching power: 1200 VA Resistive/360 VA Pilot Duty DC controller Maximum switching voltage: 30 VAC or 42 VDC Maximum switching current: 4 A Resistive/1 A Pilot Duty Maximum switching power: 125 W Resistive/28 W Pilot Duty

Low voltage relays (optional)1,2

Four relays (SPDT);

Wire gauge: 0.08 to 1.5 mm2 (28 to 16 AWG) Maximum switching voltage: 30 VDC Maximum switching current: 1 A Resistive Maximum switching power: 30 VA Resistive

1 Dependent on controller configuration.

2 The low voltage relays cannot be used in hazardous locations.

English 3

Specification Details

Analog inputs (optional)1 One 0-20 mA (or 4-20 mA) analog input on each analog input module One analog sensor input on each sensor module Maximum of two analog inputs

Analog outputs (optional)1 Five 0–20 mA (or 4-20 mA) analog outputs on each analog output module3

Digital communication (optional)1

Profibus DPV1 module, Modbus RS232/RS485 module, Modbus TCP, PROFINET module, EtherNet/IP™4 module

Software module (optional) Contact sales or technical support for information.

Note: Only one software module can be installed on a controller at the same time.

Network connection1 LAN version (optional): Two Ethernet connectors (10/100 Mbps), M12 female D-coding connector; Cellular version and WiFi version (optional)5

USB port Used for data download and software upload. The controller records approximately 20,000 data points for each connected sensor.

Compliance information CE. ETL certified to UL and CSA safety standards (with all sensor types), FCC, ISED, KC, RCM, EAC, UKCA, SABS, CMIM, Morocco

Warranty 1 year (EU: 2 years)

Section 2 General information In no event will the manufacturer be liable for damages resulting from any improper use of product or failure to comply with the instructions in the manual. The manufacturer reserves the right to make changes in this manual and the products it describes at any time, without notice or obligation.

Revised editions are found on the manufacturer’s website.

2.1 Safety information

The manufacturer is not responsible for any damages due to misapplication or misuse of this product including, without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent permitted under applicable law. The user is solely responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction.

Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.

If the equipment is used in a manner that is not specified by the manufacturer, the protection provided by the equipment may be impaired. Do not use or install this equipment in any manner other than that specified in this manual.

2.1.1 Use of hazard information

D A N G E R Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.

W A R N I N G Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.

3 Refer to the module documentation for additional information.

Note: Install only one module in one of the available slots.

4 EtherNet/IP is a trademark of OVDA Inc.

5 An external USB box WiFi is necessary for network connection on WiFi versions. An external

USB box cellular is necessary for network connection on cellular versions.

4 English

C A U T I O N Indicates a potentially hazardous situation that may result in minor or moderate injury.

NOT I CE

Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis.

2.1.2 Precautionary labels

Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.

This is the safety alert symbol. Obey all safety messages that follow this symbol to avoid potential injury. If on the instrument, refer to the instruction manual for operation or safety information.

This symbol indicates that a risk of electrical shock and/or electrocution exists.

This symbol indicates the presence of devices sensitive to Electro-static Discharge (ESD) and indicates that care must be taken to prevent damage with the equipment.

Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.

2.1.3 EMC compliance

C A U T I O N This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments.

CE (EU)

The equipment meets the essential requirements of EMC Directive 2014/30/EU.

UKCA (UK)

The equipment meets the requirements of the Electromagnetic Compatibility Regulations 2016 (S.I.

2016/1091).

Canadian Radio Interference-Causing Equipment Regulation, ICES-003, Class A:

Supporting test records reside with the manufacturer.

This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.

Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne sur les équipements provoquant des interférences.

FCC Part 15, Class "A" Limits Supporting test records reside with the manufacturer. The device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions:

1. The equipment may not cause harmful interference.

2. The equipment must accept any interference received, including interference that may cause undesired operation.

English 5

Changes or modifications to this equipment not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules.

These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at their expense. The following techniques can be used to reduce interference problems:

1. Disconnect the equipment from its power source to verify that it is or is not the source of the interference.

2. If the equipment is connected to the same outlet as the device experiencing interference, connect the equipment to a different outlet.

3. Move the equipment away from the device receiving the interference.

4. Reposition the receiving antenna for the device receiving the interference.

5. Try combinations of the above.

2.2 Product overview

D A N G E R

Chemical or biological hazards. If this instrument is used to monitor a treatment process and/or chemical feed system for which there are regulatory limits and monitoring requirements related to public health, public safety, food or beverage manufacture or processing, it is the responsibility of the user of this instrument to know and abide by any applicable regulation and to have sufficient and appropriate mechanisms in place for compliance with applicable regulations in the event of malfunction of the instrument.

NOT I CE

Network and access point security is the responsibility of the customer that uses the wireless instrument. The manufacturer will not be liable for any damages, inclusive however not limited to indirect, special, consequential or incidental damages, that have been caused by a gap in, or breach of network security.

NOT I CE

Perchlorate Material—Special handling may apply. Refer to www.dtsc.ca.gov/perchlorate. This perchlorate warning applies only to primary batteries (provided singly or installed on this equipment) when sold or distributed in California, USA.

NOT I CE

The controller is supplied with a protection foil installed on the display. Make sure to remove the protection foil before the controller is used.

The SC4500 Controller is a 2-channel controller for digital analytical devices (e.g., sensors and analyzers) and analog sensors that are connected to a digital gateway or expansion module. Refer to Figure 1.

The controller shows sensor measurements and other data on the display, can transmit analog and digital signals, and can interact with and control other devices through outputs and relays. Outputs, relays, sensors and expansion modules are configured and calibrated through the user interface on the front of the controller or remotely for network connected controllers. The controller connects to Claros with a cellular network6, WiFi network6 or through LAN connection. The Prognosys diagnostic system6 shows the status of maintenance tasks and gives the status of the instrument condition.

The instrument display is a touchscreen. The instrument enclosure has a protective vent in the bottom. Do not cover or remove the protective vent. Replace the protective vent if damage is seen.

Note: Some controller configurations have two Ethernet ports for LAN connectivity or for Modbus TCP/IP based fieldbus communication with PLC/SCADA. Refer to LAN connection on page 29.

The controller is available with optional expansion modules. Refer to Optional expansion modules on page 7.

6 English

HTTP://WWW.DTSC.CA.GOV/PERCHLORATE

Figure 1 Product overview

1 Label for module installation and wiring information 7 USB cover

2 USB connection for external USB box (WiFi or cellular connection)

8 Electrical connections and fittings

3 Expansion module (Slot 0)6 9 Protective vent

4 Additional expansion module slots (Slots 1, 2, 3 and 4)

10 Cover for module installation

5 Touchpad display 11 High-voltage barrier

6 USB connection for data download and firmware update

2.2.1 Optional expansion modules

The controller is available with optional expansion modules:

• Profibus DP module (Slot 0)—Used to connect the controller to a Profibus DP (decentralized peripherals) network. Profibus DP is used to operate sensors and actuators with a centralized controller in production (factory) automation applications.

• Modbus RS232/RS485 module (Slot 0)—Used to connect the controller to a Modbus RS232/RS485 network.

• 4–20 mA output module (Slot 1)—Used for analog signaling or to control other external devices.

Note: Only one 4–20mA output module can be installed.

• EtherNet/IP module (Slot 2)—The controller can connect to a PLC through Industrial Ethernet Protocol including a EtherNet/IP solution. Line, Star and Ring topologies are available.

• PROFINET module (Slot 2)—The controller can connect to a PLC through the Industrial Ethernet Protocol, which includes a PROFINET solution. Line, Star and Ring topologies are available.

• Low voltage relay module (Slot 2)—The controller can control external devices (e.g., an alarm or dispenser system).

• 4–20 mA input module (Slot 3 or 4)—Lets the controller accept one analog input (0–20 mA or 4–20 mA) from an external device.

• pH/ORP and conductivity modules (Slot 3 or 4)—Lets the controller accept measurements from one pH, ORP or conductivity analog sensor.

• Amperometry module (Slot 3)—Lets the controller accept measurements from the dissolved oxygen sensor of the 2582sc analyzer or the electrodes for the oxygen scavenger application used in the 2586sc analyzer.

Note: An expansion module can be installed in only one of Slots 0 ,1 and 2.

6 Dependent on controller configuration. The expansion modules are factory-installed based on controller configuration.

English 7

Note: Only two input devices7 can be connected to the controller at the same time. If more than two input devices are connected, the controller will only see two of the devices. Refer to Sensor is not recognized in Troubleshooting on page 51.

2.3 Product components

Make sure that all components have been received. Refer to Figure 2. If any items are missing or damaged, contact the manufacturer or a sales representative immediately.

Figure 2 Product components

1 SC4500 controller 7 Mounting foot (mounting bracket inserts) (2x)

2 Mounting hardware 8 Flat washer, ¼-inch ID (4x)

3 Sealing gasket for panel mount, Neoprene 9 Lock washer, ¼-inch ID (4x)

4 Vibration isolation gasket for pipe mount 10 Keps hexnut, M5 x 0.8 (4x)

5 Vibration isolation washer for pipe mount (4x) 11 Pan head screws, M5 x 0.8 x 15 mm (4x)

6 Bracket for wall and pipe mounting 8 12 Pan head screws, M5 x 0.8 x 100 mm (4x)9

Section 3 Installation D A N G E R

Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.

7 Input devices are analog inputs from external devices, analog sensors and digital sensors.

8 A bracket for panel mounting is available as an optional accessory. Refer to Replacement parts and accessories on page 54.

9 Used for variable diameter pipe mount installations.

8 English

3.1 Installation guidelines

D A N G E R

Electrical shock hazard. Externally connected equipment must have an applicable country safety standard assessment.

W A R N I N G Explosion hazard. This manual is only for installation of the unit in a non-hazardous location. For installation of the unit in hazardous locations, use only the instructions and approved control drawing provided in the hazardous location installation manual.

NOT I CE

Do not install the controller in an environment with a caustic atmosphere without a protective enclosure. A caustic atmosphere will cause damage to electronic circuitry and components.

NOT I CE

Do not install the controller outdoors in an environment that receives direct sunlight or UV radiation or damage to the controller can occur. Install the optional UV protection screen with sunroof to prevent damage from UV exposure when installed outdoors in direct sunlight.

Note: (Network and Claros version only) Make sure that your IT department has approval for the installation and commissioning of the device. Administrator rights are not necessary. The email address "No-reply@hach.com" sends the setup email and "donotreply@hach.com" sends the system notifications that are necessary for the installation. Add the two email addresses to the safe senders list to make sure to receive mails from these senders.

Hach does not send a request to confirm that the sender is not a robot.

• Install the controller in a location where the power disconnect device for the controller is easily operated.

• Attach the controller upright and level on a flat, vertical surface.

• As an alternative, attach the instrument to a panel, vertical pole or horizontal pole.

• Make sure that the device is in a location where there is sufficient clearance around it to make connections and to do maintenance tasks.

• Make sure that there is a minimum of 16 cm (6.30 in.) of clearance for the controller door to open.

• Install the instrument in a location with minimum vibration.

• The optional holder for mobile phones is recommended for all installations.

• The optional sunroof or the optional UV protection screen with sunroof is recommended for all outdoor installations.

• Give protection to computers or other connected equipment that may not have equivalent environmental ratings based on the enclosure rating of the equipment.

• Obey specified ambient ratings on the internal side of panels for panel mount installations.

• Make sure that the maximum power rating is correct for the ambient temperature.

3.2 Mechanical installation

3.2.1 Attach the instrument to a wall

Attach the controller upright and level on a flat, vertical surface. Make sure that the wall mounting is able to hold 4 times the weight of the equipment. Refer to the illustrated steps in Figure 3 and Product components on page 8 for the necessary mounting hardware.

English 9

Figure 3 Wall mounting

3.2.2 Attach the instrument to a pole

Attach the controller upright to a pole or pipe (horizontal or vertical). Make sure that the pipe diameter is 19 to 65 mm (0.75 to 2.5 in.) Refer to the illustrated steps in Figure 4 and Product components on page 8 for the necessary mounting hardware.

10 English

Figure 4 Pole mounting

3.2.3 Install the instrument in a panel

A rectangular hole is necessary for panel installation. Use the supplied sealing gasket for panel mount as a template to cut the hole in the panel. Make sure to use the template in the up position to install the controller vertical. Refer to Figure 5.

Note: If using the bracket (optional) for panel mounting, push the controller through the hole in the panel and then slide the bracket over the controller on the back side of the panel. Use the four 15 mm pan head screws (supplied) to attach the bracket to the controller and secure the controller to the panel.

English 11

Figure 5 Panel mounting dimensions

3.3 Electrical installation

3.3.1 Electrical connectors and fittings

Figure 6 shows the electrical connectors and fittings on the instrument. To keep the environmental rating of the enclosure, make sure that there is a plug in the strain relief fittings that are not used and a connector cap on the unused connectors.

Based on the controller configuration, the controller has:

• Ethernet connectors (LAN) to give internet access to the controller through a customer network.

• Ethernet connectors for Industrial Ethernet Protocols: EtherNet/IP or PROFINET.

• Digital SC connectors for sc digital sensors, sc digital gateways and analyzers.

A color code identifies the connectors. The LAN connectors are green with a red dot. The EtherNet/IP or PROFINET connectors are yellow with a red dot. The sc digital sensor connectors are black with a green dot. Refer to Table 1 for the applicable options for each connector and fitting.

Note: The controller is supplied without strain relief fittings installed. The user must supply the necessary strain reliefs. Refer to Replacement parts and accessories on page 54.

12 English

Figure 6 Electrical connectors and fittings

1 Ethernet connector (optional) for LAN port 1 or EtherNet/IP or PROFINET connector

5 Strain relief fitting for USB box and expansion modules: Analog inputs/outputs, Profibus DP

2 Ethernet connector (optional) for LAN port 2 or EtherNet/IP or PROFINET connector

6 Power cord (or conduit hub)11

3 Digital SC connector: Channel 1. Optional: Analog sensor connection to sensor module or analog input connection to 4-20 mA input module10

7 Strain relief fitting for high voltage relay

4 Digital SC connector: Channel 2. Optional: Analog sensor connection to sensor module or analog input connection to 4-20 mA input module

Table 1 Options for each connector and fitting

Device 112 2 Option13 3 4 5 6 7 sc digital sensor, sc digital gateway or analyzer X X

Analog sensor X X

Sensor analog module X X

4-20 mA output X

Low voltage relay module X

Profibus DP module X

Modbus RS232/RS485 module X

USB Box X

LAN + LAN Split / Chaining

LAN + Modbus TCP Split / Chaining

EtherNet/IP IEP only

LAN + EtherNet/IP Mix IEP

10 To connect an analog sensor or 4-20 mA input to the controller, install the applicable expansion module, if not already installed. Refer to the documentation supplied with the expansion module for additional information.

11 The power cord is factory-installed based on the controller configuration.

12 A color code identifies the connectors. The LAN connectors are green. The EtherNet/IP or

PROFINET connectors are yellow.

13 Refer to LAN connection on page 29 for Ethernet port configuration options.

English 13

Table 1 Options for each connector and fitting (continued)

Device 112 2 Option13 3 4 5 6 7

PROFINET IEP only

LAN + PROFINET Mix IEP

High voltage relay X

Power supply X

3.3.2 Electrostatic discharge (ESD) considerations

NOT I CE

Potential Instrument Damage. Delicate internal electronic components can be damaged by static electricity, resulting in degraded performance or eventual failure.

Refer to the steps in this procedure to prevent ESD damage to the instrument:

• Touch an earth-grounded metal surface such as the chassis of an instrument, a metal conduit or pipe to discharge static electricity from the body.

• Avoid excessive movement. Transport static-sensitive components in anti-static containers or packages.

• Wear a wrist strap connected by a wire to earth ground.

• Work in a static-safe area with anti-static floor pads and work bench pads.

3.3.3 Power connections

D A N G E R

Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.

D A N G E R

Electrocution hazard. Always remove power to the instrument before making electrical connections.

If the controller does not have an installed power cord, connect power with conduit or a power cord.

Refer to the sections that follow to connect power with conduit or a power cord.

3.3.3.1 Open the controller cover

Open the controller cover to get access to the wiring connections. Refer to Figure 7.

12 A color code identifies the connectors. The LAN connectors are green. The EtherNet/IP or PROFINET connectors are yellow.

13 Refer to LAN connection on page 29 for Ethernet port configuration options.

14 English

Figure 7 Open the controller

3.3.3.2 Remove the high-voltage barrier

High-voltage wiring for the controller is located behind a high-voltage barrier in the controller enclosure. Do not remove the barrier while power is supplied to the controller. Make sure that the barrier is installed before power is supplied to the controller.

Remove the high-voltage barrier to get access to the high-voltage wiring. Refer to Figure 8.

Figure 8 High-voltage barrier

3.3.3.3 Wiring for power

D A N G E R

Electrocution hazard. Protective Earth Ground (PE) connection is required.

English 15

D A N G E R

Electrical shock and fire hazards. Make sure to identify the local disconnect clearly for the conduit installation.

W A R N I N G Potential Electrocution Hazard. If this equipment is used outdoors or in potentially wet locations, a Ground Fault Interrupt device must be used for connecting the equipment to its mains power source.

W A R N I N G Electrocution hazard. The local disconnection means must disconnect all the electrical current-carrying conductors. Mains connection must keep supply polarity. The separable plug is the disconnect means for cord connected equipment.

W A R N I N G

Electrical shock and fire hazards. Make sure that the user-supplied power cord and non‐locking plug meet the applicable country code requirements.

W A R N I N G Explosion hazard. This manual is only for installation of the unit in a non-hazardous location. For installation of the unit in hazardous locations, use only the instructions and approved control drawing provided in the hazardous location installation manual.

NOT I CE

Install the device in a location and position that gives easy access to the disconnect device and its operation.

The controller can be purchased as either a 100-240 VAC powered model or a 18-28 VDC powered model. Follow the appropriate wiring instructions for the purchased model.

Supply power to the instrument with conduit or a power cable. Make sure that a circuit breaker with sufficient current capacity is installed in the power line. The circuit breaker size is based on the wire gauge used for installation.

For installation with conduit:

• Install a local disconnect for the instrument within 3 m (10 ft) of the instrument. Put a label on the disconnect that identifies it as the main disconnect device for the instrument.

• Rated for at least 90 °C (194 °F) and applicable to the installation environment

• For permanent connections use only solid wires. Use cable dimensions between 0.75 to 1.5 mm2

(18 to 16 AWG). Flexible wires must have a crimped ferrule or pin type terminal on the end.

• Connect equipment in accordance with local, state or national electrical codes.

• Connect the conduit through a conduit hub that holds the conduit securely and seals the enclosure when tightened.

• If metal conduit is used, make sure that the conduit hub is tightened so that the conduit hub connects the metal conduit to safety ground.

• The DC power source that supplies power to the DC controller must maintain voltage regulation within the specified 18-28 VDC voltage limits. The DC power source must also provide adequate protection against surges and line transients.

For installation with a power cable, make sure that the power cable is:

• Less than 3 m (10 ft) in length

• Rated sufficient for the supply voltage and current.

• Rated for at least 90 °C (194 °F) and applicable to the installation environment

16 English

• Not less than 0.75 mm2 (18 AWG) with applicable insulation colors for local code requirements.

Flexible wires must have a crimped ferrule or pin type terminal on the end.

• A power cable with a three-prong plug (with ground connection) that is applicable to the supply connection

• Connected through a cable gland (strain relief) that holds the power cable securely and seals the enclosure when tightened

• Does not have a locking type device on the plug

3.3.3.4 Connect conduit or a power cord

D A N G E R

Electrocution hazard. Always remove power to the instrument before making electrical connections.

NOT I CE

The manufacturer recommends the use of manufacturer-supplied electrical components, such as power cord, connectors and strain relief fittings.

NOT I CE

Make sure that the cable sheath goes through the inner side of the enclosure to keep the environmental rating of the enclosure.

The controller can be wired for line power by hard-wiring in conduit or wiring to a power cord.

Regardless of the wire used, the connections are made at the same terminals.

The power cable plug is used to connect and disconnect power to the controller. For installation in conduit, the installed local disconnect is used to connect and disconnect power to the controller.

Refer to Figure 9 and Table 2 or Table 3 to connect conduit or a power cord. Connect the protective earth ground first.

Insert each wire into the appropriate terminal until the insulation is seated against the connector with no bare wire exposed. Tug gently after insertion to make sure that there is a secure connection. If necessary, remove the connector from the PCBA for easier wiring of the terminals.

Note: Make sure that all of the cables stay below the cable limit line printed on the PCBA to prevent interferences with the high-voltage barrier. Refer to Figure 9.

English 17

Figure 9 Connect conduit or a power cord

1 AC and DC power terminal 3 Protective earth ground

2 Cables limit: do not put cables above the line. 4 Conduit hub (or strain relief fitting for power cord)

Table 2 Wiring information—AC power

Terminal Description Color—North America Color—EU

L Hot (Line 1) Black Brown

N Neutral (N) White Blue

Protective earth ground Green Green with yellow stripe

Table 3 Wiring information—DC power

Terminal Description Color—North America Color—EU

L +24 VDC Red Red

N 24 VDC return Black Black

Protective earth ground Green Green with yellow stripe

3.3.4 Connect the high-voltage relays

D A N G E R

Electrocution hazard. Always remove power to the instrument before making electrical connections.

18 English

W A R N I N G Potential Electrocution Hazard. Power and relay terminals are designed for only single wire termination. Do not use more than one wire in each terminal.

W A R N I N G Potential fire hazard. Do not daisy-chain the common relay connections or jumper wire from the mains power connection inside the instrument.

W A R N I N G Explosion hazard. This manual is only for installation of the unit in a non-hazardous location. For installation of the unit in hazardous locations, use only the instructions and approved control drawing provided in the hazardous location installation manual.

C A U T I O N

Fire hazard. Relay loads must be resistive. Always limit current to the relays with an external fuse or breaker. Obey the relay ratings in the Specifications section.

NOT I CE

Make sure that the cable sheath goes through the inner side of the enclosure to keep the environmental rating of the enclosure.

The instrument has two non-powered relays, each with a single-pole change-over contact. For AC controllers, the wiring compartment is not made for voltage connections more than 264 VAC.

The relay terminals are located behind a high-voltage barrier in the controller enclosure. Do not remove the barrier while power is supplied to the relay terminals. Do not supply power to the relay terminals when the barrier is not installed.

Connect each relay to a control device or an alarm device as necessary. Refer to Figure 10 and Table 4 to connect the relays. Refer to Configure the high voltage relays on page 36 to configure the trigger condition for each relay.

Refer to Specifications on page 3 for the relay specifications. The relays are isolated from each other and the low-voltage input/output circuitry.

The largest gauge wire the power and relay plugs are rated for 1.5 mm2 (16 AWG). The relay terminals accept 0.75 to 1.5 mm2 (18 to 16 AWG) wire (as determined by load application). Use wire with an insulation rating of 300 VAC or higher. Insert each wire into the appropriate terminal until the insulation is seated against the connector with no bare wire exposed. Tug gently after insertion to make sure that there is a secure connection. If necessary, remove the connector from the PCBA for easier wiring of the terminals. Flexible wires must have a crimped ferrule or a pin type terminal on the end.

Note: Make sure that all of the cables stay below the cable limit line printed on the PCBA to prevent interferences with the high-voltage barrier.

The current to the relay contacts must be 5 A (resistive only load), 1250 VA 125 W (resistive only load) or less. Make sure to have a second switch available to remove power from the relays locally in case of an emergency or for maintenance.

For AC controllers, use the relays at high voltage. For DC controllers, use the relays at low voltage.

Refer to Specifications on page 3 for the relay specifications. Do not configure a combination of both high and low voltage.

Relay terminal connections to the mains circuit in permanent connection applications must have insulation rated for a minimum of 300 V, 90 °C (194 °F). Terminals connected to the mains circuit with a power cord connection must be double insulated and rated 300 V, 90 °C (194 °F) at both the inner and outer insulation levels.

Note: Put the wires for relays through the strain relief fitting for the high voltage relay. Refer to Figure 6 on page 13.

English 19

Figure 10 Connect the relays

Table 4 Wiring information—relays

Terminal Description Terminal Description

1 Relay 2, NC 4 Relay 1, NC

2 Relay 2, common 5 Relay 1, common

3 Relay 2, NO 6 Relay 1, NO

NC = normally closed; NO = normally open

3.3.5 Install an expansion module

W A R N I N G Explosion hazard. This manual is only for installation of the unit in a non-hazardous location. For installation of the unit in hazardous locations, use only the instructions and approved control drawing provided in the hazardous location installation manual.

Expansion modules for analog outputs, analog inputs, analog sensors and Profibus communication are available for the controller. Refer to the documentation supplied with the expansion module for additional information.

3.4 Close the cover

D A N G E R

Electrocution hazard. High voltage wiring for the controller is connected behind the high voltage barrier in the controller enclosure. The barrier must remain in place except when installing modules, or when a qualified installation technician is wiring for power, relays or analog and network cards.

20 English

NOT I CE

Close the controller cover and make sure that the cover screws are tight to keep the environmental rating of the enclosure.

After the power connections are made, install the high-voltage barrier. Make sure that the high-voltage barrier is correctly put on the enclosure guides and fixed to the main PCBA. A click sound is heard when the high-voltage barrier is correctly installed. Make sure that the lower part of the high-voltage barrier (soft rubber lip) is correctly installed and has no deformation. Refer to Figure 11.

Close the controller cover. Tighten the cover screws with 2 Nm (17.70 lbf-in) torque. Refer to Figure 7 on page 15.

Figure 11 Install the high voltage barrier

3.5 Connect measurement devices

Connect digital devices (e.g., sensors and analyzers) to the device connectors on the instrument.

Refer to Figure 12. Keep the device connector caps for future use.

Make sure that the device cables do not cause a trip hazard and do not have sharp bends.

English 21

Figure 12 Connect a device

Section 4 User interface and navigation

NOT I CE

Do not use writing tips of pens or pencils or other sharp objects to make selections on the screen or damage to the screen will occur.

Figure 13 shows an overview of the main screen. Refer to Table 5 for the descriptions of the icons in the display.

The instrument display is a touchscreen. Only use a clean, dry finger tip to navigate the functions of the touchscreen. To prevent unwanted touchs, the screen is automatically lock after a period of inactivity. Touch the screen and swipe up to set the screen back to operation again.

Note: Go to the General configuration menu to disable the Screen lock setting or to set the Waiting time for the inactivity period.

22 English

Figure 13 Main screen

1 Measurement window: shows the device name and a measurement, push on the tile to show the device detail screen

4 Prognosys icon (optional)

2 Diagnostic bar: shows the system messages and alarm conditions, push the bar to see the system errors and warnings. Shows pending tasks and information about the system.

5 Carousel icon: swipe left or right on the screen to show the other screen views.

3 Status bar

Table 5 Icon descriptions

Icon Description Icon Description

Push to show the Main menu. 3G/4G signal strength. Shows when a USB box with a cellular modem is connected to the controller.

Claros connection USB connection. Shows when a USB flash drive is connected to the controller. Blinks when there is data transmission.

WiFi connection. Shows when a USB box with a WiFi adapter is connected to the controller.

Remote user. Shows when a remote user is connected to the controller.

Screen lock. Shows when the screen is locked.14 Swipe up to unlock the screen.

Push to enter a submenu or go back to the previous menu.

When in a sub menu, push to go to the main screen.

Push a measurement window to show the device detail screen. Refer to Figure 14.

14 The Screen lock option is enabled by default.

English 23

Figure 14 Device screen

1 Device name

4.1 Measurement window

The main screen shows the measurement windows. Refer to Figure 15. The carousel icon shows at the bottom of the main screen when there are other screen views available.

Each measurement window screen view has one, two or four measurement windows, based on the connected sensors and the parameters of each sensor. If an RTC module is installed on the controller and the license key is activated, the last measurement window screen of the carousel is for the RTC module.

The controller automatically fills the measurement windows to show all of the information for the connected devices. The measurement windows show the data that follows:

• Name, value, parameter and unit of the measurement device and RTC module output, if installed Note: Measurement devices are the connected digital sensors, analog sensors connected with analog modules or digital gateways, and external devices that supply 4-20 mA inputs.

• Sensor errors and warnings. If the background color of the measurement window is amber, there is a warning for the sensor. If the background color of the measurement window is red, there is an error for the sensor. A small right-arrow icon shows when the error or warning is in a different screen view. Push the small right-arrow icon to go to the device notifications screen.

• RTC module warnings. If the background color of the measurement window screen for the RTC module is amber (last measurement window screen of the carousel), there is a warning for the RTC module.

• Prognosys (if available). The Prognosys icon shows when there is no issue or maintenance planned, or if maintenance is due or scheduled. Go to the device detail screen to see more Prognosys information.

• The task in operation (maintenance or calibration task)

24 English

Figure 15 Measurement window screen and device detail screen

1 Device error 6 Measured value

2 Device parameter 7 Device detail screen

3 Units 8 Task

4 Device warning 9 Error or warning in a different screen view

5 Device name

4.2 Graph

From the main screen, swipe right until the graph screens show for the connected sensors. Refer to Figure 16. Each graph screen shows the four parameters (maximum) measured by a sensor.

If an RTC module is installed on the controller and the license key is activated, the last screen of the carousel is the graph screen for the RTC module. The graph screen shows the same parameters that show on the measurement window screen for the RTC module.

Figure 16 Graph screen

1 Legend: select the parameter shown 7 Error in a different screen view

2 Sensor name (or RTC module) 8 Zoom in

3 Main menu icon 9 Zoom out

4 Status bar 10 Move back in time

5 Number of graph screens (three maximum) 11 Move forward in time

6 Warning in a different screen view 12 Refresh icon

English 25

Section 5 Startup Connect the power cord to an electrical outlet with protective earth ground or set the circuit breaker for the controller to on.

26 English

5.1 Menu structure

The menu options available depend on the controller accessories that are installed.

English 27

5.2 Enter initial settings

At initial startup, follow the prompts on the display to set up the language, the date, the time and network information. Refer to the expanded user manual on the manufacturer's website for instructions.

Section 6 Operation

6.1 Configure the controller settings

Set the controller language, time, date, facility, location and display options.

1. Push the main menu icon, then select Controller > General.

2. Select and configure each option.

Option Description

Language Sets the language that shows on the controller display and in the log files.

Time zone Sets the time zone. Select the Region and City for the time zone.

Note: The Time zone option is not available when the controller is connected to Claros.

Time format Sets the time format: 12 h (default) or 24 h.

Time Sets the time.

Note: The Time option is not available when the controller is connected to Claros.

Date Sets the date.

Note: The Date option is not available when the controller is connected to Claros.

Facility Sets the name of the facility (32 characters maximum). Default: not selected

Location Sets the name of the location (32 characters maximum). Default: controller serial number

Device menu Shows the controller name and serial number . Change the Name setting as necessary.

Display Sets the display options:

• Brightness—Sets the brightness of the display. Options: 20%, 40%, 60% (default), 80% or 100%.

• Screen lock—When set to on (default), the screen automatically locks after the inactivity period. When the screen is locked, the touchscreen is disabled and there are no active areas in the display. Touch the screen and swipe up to set the screen back to on.

Note: The manufacturer strongly recommends not to disable the Screen lock setting. The Screen lock setting prevents unwanted touches on the screen (especially in outdoor installations).

• Waiting time—Sets the inactivity period, after which the controller locks the screen.

Options: 1, 3, 5, 10 or 15 minutes

6.2 Configure menu access

Set the password protection to on to prevent unwanted changes to specific controller settings and some of the device menus (e.g., sensor and RTC menus).

When password protection is set to on, a prompt to enter the password shows when a menu is protected. After the password is entered, access to the protected menus is available for 30 minutes.

1. To set password protection to on, do the steps that follow:

a. Push the main menu icon, then select Controller > Password.

b. Select Password protection, then select ON.

The default password is "SC4500". There is only one password for each controller.

2. To change the password, do the steps that follow:

a. Push the main menu icon, then select Controller > Password.

28 English

b. Select Change password, then enter the old password.

c. Enter the new password. The password must be 4 to 8 characters. The password is case sensitive and can include alphanumeric and special characters.

3. If the password is forgotten, set the password back to the default as follows:

a. Push the main menu icon, then select Controller > Service.

b. Select Reset password. The password is set to "SC4500".

4. To set password protection to off, do the steps that follow:

a. Push the main menu icon, then select Controller > Password.

b. Select Password protection, then enter the password.

c. Select OFF.

The password is saved.

6.3 Connect the instrument to a network

Based on instrument configuration, the instrument can connect to a network with internet connectivity for configuration and operation. Based on the version, the controller connects to the internet with a cellular network, WiFi network or a LAN connection.

6.3.1 LAN connection

The controller is supplied with two Ethernet connectors based on the controller configuration.

Regardless of the supplied controller, the user must configure the Ethernet ports to the applicable option.

Note: An Ethernet port upgrade may be necessary based on the controller configuration. Refer to the documentation supplied with the upgrade kit.

1. Use a user-supplied Ethernet cable (M12 to RJ45)15 to connect the controller to a device with internet connection. Refer to Electrical connectors and fittings on page 12.

2. Push the main menu icon, then select Controller > Connectivity > LAN > Ethernet ports.

3. Enter the settings.

Option Description

None The controller does not have Ethernet ports or the Ethernet ports are not configured. The controller is supplied with the option None by default.

Note: When the controller configuration is set back to the factory settings, all of the LAN connection settings are lost and the Ethernet ports are set back to None.

Chaining The two Ethernet ports are configured for LAN connectivity. When selected, the LAN port configuration setting shows on the LAN menu. The two Ethernet ports have the same port configuration. Select this option to share the internet connection with other controllers (daisy chain topology) or to connect the controller to Claros and HCNB through the Ethernet cable.

Split One Ethernet port is configured for LAN connectivity and the other Ethernet port is configured for a separate LAN or Modbus TCP. When selected, the Lan port 1 configuration and Lan port 2 configuration settings show on the LAN menu. The two Ethernet ports have different port configurations. Select the Split option to have different controller connections to Claros and fieldbus (Modbus TCP or HCNB).

Mix IEP One Ethernet port is configured for LAN connectivity and the other Ethernet port is configured for IEP16. When selected, the LAN port configuration setting shows on the LAN menu. Select this option to use port 1 for LAN connectivity and port 2 for IEP (compatible with EtherNet/IP or PROFINET star topology).

IEP only The two Ethernet ports are configured for IEP. The two ports are used for EtherNet/IP or PROFINET with chain, ring or star topology.

15 Refer to Replacement parts and accessories on page 54.

16 Based on the controller configuration. Make sure that a PROFINET or EtherNet/IP module is installed in the controller.

English 29

4. Push SAVE.

5. When applicable, select LAN port configuration. Enter the settings as follows:

a. If the customer network supports Dynamic Host Configuration Protocol (DHCP), set the option to on (default) to get an IP address automatically. The display shows the MAC address of the selected port.

b. If the customer network does not support DHCP, set the option to off and complete the settings with IP addresses supplied by the network administrator. Enter the correct settings based on the network topology: IP address, Netmask, Default gateway and DNS address.

c. When Chaining is selected, the LAN server option shows on the LAN port configuration menu.

Set the LAN server to on to share the internet connection with other controllers through LAN connectivity.

6. Select APPLY.

6.3.2 WiFi network connection

An external USB box WiFi with a WiFi adapter is necessary to connect the controller to a WiFi network. Refer to the documentation supplied with the USB box WiFi for more information about the WiFi installation. Follow all WiFi module manufacturer recommendations and guidance.

If the controller connects to the internet with a WiFi network, do the steps that follow:

1. Push the main menu icon, then select Controller > Connectivity > Wi-Fi.

2. Set the Wi-Fi connectivity button to on. A list of available access points shows.

3. Select Network to set the applicable WiFi access point. If necessary, enter the correct credentials.

The WiFi connectivity currently accepts the authentication methods that follows: User name and Wi-Fi password.

a. WEP: Not recommended. The WEP authentication method has low security.

b. WPA2 Personal / PSK: Standard authentication. Enter the Wi-Fi password to access the network.

c. WPA2 Enterprise: Authentication for enterprise networks. Enter the enterprise credentials, User name andWi-Fi password, to access the network.

4. Select APPLY.

6.3.3 Cellular network connection

Note: Only for Claros connection.

The external USB box cellular is necessary to connect the controller to a mobile network. Refer to the documentation supplied with the USB box cellular for installation instructions.

6.3.3.1 Configuration with Hach Cellular (Telenor) SIM

If the controller connects to the internet with the factory installed SIM card and the default Hach data plan (Telenor), do the steps that follow:

1. Push the main menu icon, then select Controller > Connectivity >…