This technical equipment drawing provides detailed specifications for the STERIS AMSCO® 7052HP Single-Chamber Washer/Disinfector, a steam-heated medical device designed for instrument and utensil processing. The comprehensive 10-page document details precise utility requirements, including plumbing (hot/cold/pure water, compressed air, steam, drain connections), electrical specifications (multiple voltage configurations), and physical characteristics. Key technical specifications include a chamber volume of 10.3 ft³, operating dimensions of 42 x 80.75 x 32 inches, an operating weight of 1,254 lbs when fully loaded, and specific requirements for installation such as floor load capacity, environmental operating conditions, and ventilation exhaust options.
The document provides extensive technical guidance for installation, specifying detailed connection requirements for various utilities, recommended circuit protection, network connectivity, and environmental operating parameters. Notably, the washer requires specific utility conditions including steam inlet pressure of 30-80 psig, compressed air at minimum 80 psig, and electrical configurations ranging from 200-208V to 460-480V with corresponding amperage requirements. The drawing includes comprehensive notes on installation considerations, such as recommended water quality parameters, air purity standards, and environmental conditions for optimal performance, making it a critical reference for facilities planning to integrate this medical washer/disinfector into their infrastructure.
View the file
Other files
Show all 22
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
42 [1067]
UNIT WIDTH
80-3/4 [2051]
UNIT
HEIGHT
31 [787]
LOADING
HEIGHT
26-1/4 [667]
LOADING
CLEARANCE
HEIGHT
8-1/2 [216]
[381]
91 [2311]
RAISED DOOR
HEIGHT
34 [864]
MAX LENGTH
32 [813]
UNIT LENGTH
90 [2286]
SERVICE
AREA
HEIGHT
45 [1143]
WALL OPENING
*(NOTE F)
1-1/2 [38]
MINIMUM
94 [2388] *(NOTE G)
RECOMMENDED
98 [2489] *(NOTE H)
MAX 1 [25]
ADJUSTMENT WITH
WASHER LEGS
(UNIT HEIGHT 81-3/4 [2077] MAX)
MINIMUM
90 [2286] *(NOTE G)
RECOMMENDED
96 [2438] *(NOTE H)
81-7/8 [2080]
MAX
HEIGHT
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292221
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SERVICE AREA
REQUIRED FOR
DRYING MOTOR
REMOVAL
TYPICAL INSTALLATION
NOTES:
(A) UNLESS OTHERWISE SPECIFIED, ALL DIMENSIONS ARE IN INCHES [mm].
(B) SERVICE ACCESS SIDE OF UNIT IS RECOGNIZED BY THE TOUCH SCREEN AT THE RIGHT OF
WASH CHAMBER. ELECTRICAL AND CONTROL COMPONENTS, AS WELL AS MOST OF
SERVICABLE PARTS ARE ACCESSIBLE FROM THAT SIDE OF UNIT. NON-SERVICE SIDE IS
EQUIPPED WITH THE TOUCH SCREEN ON THE LEFT OF THE WASH CHAMBER.
(C) CUSTOMER MUST BE SURE THAT MACHINE STANDS ON A NON-COMBUSTIBLE FLOOR.
(D) WEIGHTED EQUIVALENT SURFACE SOUND PRESSURE LEVEL: 64.0 dB(A) CALCULATED AS DESCRIBED IN ISO 3746.
(E) VERTICAL BARRIER WALL FLANGES ARE PROVIDED WITH UNIT TO FILL GAP BETWEEN
UNIT AND WALL ON BOTH SIDES. DIMENSIONS ARE 3-1/2 in [89 mm] WIDTH X 80 in [2032 mm] LONG.
(F) HORIZONTAL BARRIER WALL FLANGES ARE PROVIDED WITH UNIT TO FILL GAP BETWEEN
UNIT AND WALL ON BOTH SIDES. DIMENSIONS ARE 49 in [1245 mm] WIDTH X 16 in [406 mm] HEIGHT.
(G) MINIMUM REQUIRED DIMENSION FOR DOOR REMOVAL WITH LIMITED ADJUSTMENT AVAILABLE: +0 in [0 mm] THE STANDARD BARRIER WALL FLANGES CAN'T BE USED AT THAT DIMENSION. -3/4 in [19.1 mm]
(H) RECOMMENDED DIMENSION FOR DOOR REMOVAL WITH FULL LEVELING ADJUSTMENT AVAILABLE: +1 in [25.4 mm] THE STANDARD BARRIER WALL FLANGES CAN BE USED AT THAT DIMENSION. *(NOTE E AND F) -3/4 in [19.1 mm]
FRONT VIEW
SERVICE ACCESS
SIDE OF UNIT
*(NOTE B)
TOP VIEW
PASS THROUGH
DOOR
SIDE VIEWLEVEL FLOOR
*(NOTE C)
W A
LL
O
P E
N
IN
G H
E
IG
H T
C E
IL
IN
G H
E
IG
H T
FO
R
D O
O R
R E
M O
V A
L
[102]
Page 1 of 10 10292221.D
Released
2X 33-7/16 [849]
2X 27-11/16 [704]
5-11/16 [144]
7-1/8 [181]
4X 15 [381]
1-7/8 [47]
17-3/8 [441]
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292221
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
*NOTES:
(A) CONCRETE FLOOR MUST REMAIN FREE OF DEFECTS IN THESE AREA, AS PER ESR-1917 & ESR-3187.
FOR INSTALLATION OF SEISMIC ANCHORS (NOT PROVIDED BY STERIS).
(B) THE FLOOR SINK AND DRAIN SHALL NOT OCCUPY THESE AREAS. 5 in x 5 in [127 mm x 127 mm] FLOOR SINK WITH Ø4 in [102 mm] DRAIN IS RECOMMENDED WHEN USING SEISMIC MOUNTS.
SEISMIC MOUNTS
SEISMIC ANCHORAGE
FOR DETAIL SEE FD354
QTY 4 ANCHORS
FINISHED WALL
PREFERENCE
LOCATION OF
FLOOR SINK
*(NOTE A AND B)
Page 2 of 10 10292221.D
80-3/4 [2051]
7-1/8 [181]
[381]
31-1/8 [791]
[432]
19-1/32 [483]
16-17/32 [420]
15-1/2 [394]
3-7/8 [99]
[356]
11-1/2 [292]
32-11/32 [822]
HW HOT WATER
CW COLD WATER
PW PURE WATER
A COMPRESSED AIR
D FLOOR DRAIN LOCATION
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292221
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
UTILITY (PLUMBING)
CONNECTION POSITION
FRONT VIEW
A
D
PW
HW
CW
SIDE VIEW
HW
PW
CW
A
D
CR
STOP VIEW
HW
CW
PW
A
CR
S
D
S STEAM
CR CONDENSATE RETURN
S
CR
Page 3 of 10 10292221.D
79-1/4 [2013]
35-13/16 [910]
18-3/8 [467]
[534]
38-1/8 [967]
39-5/8 [1007]
E ELECTRICAL
N NETWORK
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292221
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
UTILITY (ELECTRICITY)
CONNECTION POSITION
FRONT VIEW SIDE VIEW
E
N
E
N
TOP VIEW
E
N
Page 4 of 10 10292221.D
6-5/32 [156]3
[76]
81-29/32 [2080]
EX VENTILATION EXHAUST
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292221
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
UTILITY (VENTILATION)
CONNECTION POSITION (OPTIONAL)
FRONT VIEW SIDE VIEW
EX
SUGGESTED DUCTWORK
(BY OTHERS)
EX
VAPOR CONDENSATE
COLLECTING MANAGED
BY THE UNIT
TOP VIEW
EX
VENTILATION CONNECTION (OPTIONAL)
(SEE PAGE 9 *(NOTE 1)
DRY CONTACT (SEE PAGE 10 *(NOTE 3))
EX
BY OTHERS
Page 5 of 10 10292221.D
60 [1524]
REF
8-1/2 [216]
132 [3350]
MAX.
17 [432]
INSTRUMENT
(With PUC HP)
UTENSIL
(With PUC HP)
CHEMICAL
(CH)
ENZYME CONSUMPTION PER CYCLE
(SEE PAGE 7; *(NOTE 4)) 0.74 oz [22 ml] 0.74 oz [22 ml]
DETERGENT CONSUMPTION PER CYCLE
(SEE PAGE 7; *(NOTE 4)) 0.74 oz [22 ml] 0.74 oz [22 ml]
LUBRICANT CONSUMPTION PER CYCLE
(SEE PAGE 7; *(NOTE 4)) 0.2 oz [6 ml] N/A
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292221
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
*NOTES:
(A) DOSING STATION AND CHEMICAL CONTAINERS MUST BE ON THE SAME FLOOR OF THE BUILDING AS THE WASHER/DISINFECTOR.
CHEMICAL CONTAINER MUST BE LOCATED AT A LOWER HEIGHT THAN DOSING STATION. MAXIMUM CONTAINER INCLINATION
IS 15 DEG. FROM FLOOR. AMBIENT TEMPERATURE SHALL NOT EXCEED 32 °C [90 °F]
(B) HP PICKUP TUBE SYSTEM CAN PROVIDE UP TO TWO WASHERS WITH CHEMISTRY FROM THE SAME CONTAINER.
DOSING STATION
CEILING
ROUTE CHEMICAL LINES
AND SIGNAL CABLE
THROUGH PIPE TO REACH
BOTTOM OF UNIT
CHEMICAL CONTAINERS
*(NOTE A)
12 ft [3.6 m] OF INJECTION LINE
(FROM CONTAINERS TO
DOSING STATION)
*(NOTE B)
DOSING STATION
WEIGHT : 15 lb [6.8 kg]
2X FD349 SHOWN
UP TO 5 TUBES
100 ft [30 m] OF INJECTION LINE
AND SIGNAL CABLE
DETERGENT LINES SHALL NOT BE IN CONTACT WITH HOT SURFACES.
STERIS RECOMMENDS CUSTOMER TO ROUTE CHEMICAL LINES AND
SIGNAL CABLE THROUGH A 3 in [76 mm] ID CONDUIT.
EXCESS LENGTH MUST BE CUT BEFORE CONNECTING TO
WASHER/STATION TO AVOID FORMING BUNDLES THAT COULD
AFFECT THE CHEMICAL INJECTION SYSTEM'S PERFORMANCE
(DO NOT CUT SIGNAL CABLE).
Page 6 of 10 10292221.D
UTILITY REQUIREMENTS CHART (PLUMBING)
UTILITY PARAMETER REQUIREMENT NOTES
HOT WATER
(HW)
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
1. STERIS REQUIRES TO INSTALL WATER HAMMER
ARRESTOR, UNION & PRESSURE GAUGE.
2. CUSTOMER MUST PROVIDE UTILITY
CONNECTIONS WITH SHUTOFFS DISCONNECTS
WITHIN 2 FEET OF THE PERIMETER OF THE
EQUIPMENT AND BELOW THE CEILING DECK.
*STERIS OFFERS FLEXIBLE HOSES
ACCESSORY (FD088 OR FD037)
WITH INSTRUCTIONS.
3. SAMPLING VALVES INSTALLATION AS CLOSE AS
POSSIBLE TO WASHER WATER CONNECTIONS IS
RECOMMENDED.
PRESSURE RANGE
15 - 50 psig DYNAMIC
[103-345 kPa] MAX. 90 psig STATIC [620 kPA]
OPERATING FLOW RATE 6.8 US GPM [25.7 L/min] TO
14.1 US GPM [53.4 L/min]
TOTAL CONSUMPTION PER
INSTRUMENT CYCLE
*(NOTE 4)
9.9 US GAL [37.5 L]
WITH ACU-WASH RESERVOIR:
6.6 US GAL [25.0 L]
TEMPERATURE RANGE 110 °F [43 °C] TO 150 °F [65.5 °C]
WATER QUALITY *(NOTE 5) HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)
COLD WATER
(CW)
*(NOTES 6)
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
4. CONSUMPTION BASED ON INSTRUMENT CYCLE WITH
PROLYSTICA ULTRA CONCENTRATE HIGH
PERFORMANCE CHEMICALS (PUC HP) WITH
THE FOLLOWING UTILITY CONDITIONS:
THREE-LEVEL MANIFOLD RACK
(CW) COLD WATER PRESSURE: 30 psig DYNAMIC [207 kPa]
(CW) COLD WATER TEMPERATURE: 70 °F [21 °C]
(HW) HOT WATER TEMPERATURE: 150 °F [65 °C]
(PW) PURE WATER TEMPERATURE 70 °F [21 °C]
(S) STEAM SUPPLY: 80 psig DYNAMIC [552 kPa]
UTENSIL CYCLE CONSUMPTION:
TWO-LEVEL MANIFOLD RACK
CW: 10.4 US GAL [39.4 L]
HW: 9.6 US GAL [36.3 L]
PW: 6.4 US GAL [24.2 L]
S: 14.1 lb [6.4 kg]
5. STERIS RECOMMENDS THE FOLLOWING WATER
QUALITY PARAMETERS:
TOTAL DISSOLVED SOLIDS: 100-200 ppm (MAX. 500 ppm) TOTAL ALKALINITY (CaCO3) : 70-120 ppm (MAX. 180 ppm) pH : 6.8-7 (MAX. 7.5) TOTAL SILICA : 0.1-0.5 ppm (MAX. 1 ppm)
PRESSURE RANGE
30 - 50 psig DYNAMIC
[206-345 kPa] MAX. 90 psig STATIC [620 kPa]
OPERATING FLOW RATE 10.7 US GPM [40.5 L/min] TO
14.1 US GPM [53.4 L/min]
TOTAL CONSUMPTION PER
INSTRUMENT CYCLE
*(NOTE 4)
FOR SUMP FILLING: 10.5 US GAL [39.8 L]
(ADD THE FOLLOWING CONSUMPTION
FOR OPTIONS/CONFIGURATIONS)
CONDENSATE RETURN COOL DOWN:
6.2 US GAL [23.5 L]
NON-VENTED SYSTEM:
12.9 US GAL [48.8 L]
DRAIN COOLDOWN:
14.5 US GAL [54.9 L]
DRAIN ACU-COOLDOWN:
10.9 US GAL [41.3 L]
TEMPERATURE RANGE 70 °F [21 °C] MAXIMUM
WATER QUALITY *(NOTE 5) HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)
PURE WATER
(PW)
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
6. BACK FLOW PREVENTER AGAINST BACK
SIPHONAGE MAY BE REQUIRED (LOCAL
PLUMBING CODE)
7. PW IS OPTIONAL. IF UNAVAILABLE, UNIT WILL
PROCESS THERMAL RINSE USING HOT WATER
SUPPLY.
PRESSURE RANGE 5 - 30 psig DYNAMIC [30 - 206 kPa] MAX. 90 psig STATIC [620 kPa]
OPERATING FLOW RATE 5.5 US GPM [21 L/min] TO
12.6 US GPM [48 L/min]
TOTAL CONSUMPTION PER
INSTRUMENT CYCLE
* (NOTE 4 AND NOTE 7 )
6.6 US GAL [25.0 L]
TEMPERATURE RANGE 70 °F [21.1 °C] TO 150 °F [65.5 °C]
RESISTIVITY
(CONDUCTIVITY)
0.1-0.5 M -cm (2-10 µS/cm) pH 6.8 - 7.5
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292221
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
Page 7 of 10 10292221.D
UTILITY REQUIREMENTS CHART (PLUMBING)
COMPRESSED
AIR
(A)
CONNECTION SIZE / TYPE 1/8 in / FEMALE NPT [1/8 in / MALE BSPT]
1. AIR PURITY PER ISO-8573-1: 2010 [7:4:4]
MAX. PARTICLE DENSITY: 10 mg/m³
MAX. DEW POINT FOR WATER CONTENT:
37 °F [3 °C]
MAX. OIL CONCENTRATION FOR THE OIL
CONTENT: 5 mg/m³
2. STERIS RECOMMENDS A REFRIGERATED AIR
DRYER WHERE ENVIRONMENTAL DEW POINT
CONDITIONS ARE HIGHER THAN
RECOMMENDED.
3. CUSTOMER MUST PROVIDE UTILITY
CONNECTIONS WITH SHUTOFFS DISCONNECTS
WITHIN 2 FEET OF THE PERIMETER OF THE
EQUIPMENT AND BELOW THE CEILING DECK.
PRESSURE RANGE MIN 80 psig DYNAMIC [552 kPa] MAX. 125 psig STATIC [860 kPa]
OPERATING FLOW RATE 1.75 SCFM [50 L/min]
DRAIN
(D)
FLOOR DRAIN SIZE MIN 4 in [102 mm] O.D. WITH 8 in x 8 in [204 mm x 204 mm] FLOOR SINK
4. WASHER DRAIN OUTLET SIZE 1-1/2 in / MALE NPT
5. MAX OFFSET OF 3/4 in [19 mm] FROM
OUTLET (PAGE 3, D).
6. FOR DRAIN REQUIREMENTS WHEN USING
SEISMIC MOUNTS, SEE PAGE 2.
OPERATING FLOW RATE GRAVITY DRAIN : 50 US GPM
[189.3 L/min]
CHEMICAL
DOSING
SYSTEM
CEILING ROUTING
CONDUIT 3 in CONDUIT
11. STERIS RECOMMENDS TO ROUTE CHEMICAL
LINES AND SIGNAL CABLE THROUGH A
3 in [76 mm] ID CONDUIT (BY OTHERS).
100 ft [30 m] OF INJECTION LINE AND
SIGNAL CABLE ARE PROVIDED WITH THE UNIT.
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292221
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
STEAM INLET
(S)
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
7. STERIS REQUIRES AT LEAST 97% SATURATED
DRY PLANT STEAM, FREE OF DIRT, RESIDUES
AND CONTAMINANTS.
8. STERIS RECOMMENDS TO INSTALL A STEAM
TRAP AS CLOSE AS POSSIBLE TO UNIT
CONNECTION.
9. PROPER OPERATION OF THE WASHER
REQUIRES A MINIMUM PRESSURE DIFFERENTIAL
OF 21.5 psig [148 kPa] BETWEEN STEAM
INLET AND CONDENSATE RETURN. IN CASES OF
LOW STEAM INLET PRESSURE, CONDENSATE
CAN BE ROUTED TO DRAIN TO INCREASE
DIFFERENTIAL.
*STERIS OFFERS CONDENSATE RETURN
TO DRAIN ACCESSORY (FD353)
10. CUSTOMER MUST PROVIDE UTILITY
CONNECTIONS WITH SHUTOFFS DISCONNECTS
WITHIN 2 FEET OF THE PERIMETER OF THE
EQUIPMENT AND BELOW THE CEILING DECK.
PRESSURE RANGE
30 - 80 psig DYNAMIC
[206-552 kPa] MAX. 90 psig STATIC [620 kPa]
OPERATING
MASS FLOW RATE
169 lb/h [76.5 kg/h] AT 30 psig [207 kPa]
271 lb/h [123 kg/h AT 80 psig [550 kPa]
TOTAL CONSUMPTION
PER INSTRUMENT CYCLE
(SEE PAGE 7, *(NOTE 4))
14.1 lb [6.4 kg]
CONDENSATE
RETURN
(CR)
PRESSURE RANGE MAX. 8.5 psig [68.9 kPa] BACK PRESSURE (17 ft [5.2 m] ELEVATION)
OPERATING FLOW RATE PEAK : 0.54 US GPM [2.05 L/min]
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
Page 8 of 10 10292221.D
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292221
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
UTILITY REQUIREMENTS CHART (ELECTRIC)
UTILITY
PARAMETER AND REQUIREMENT NOTES
CONNECTION
SIZE/TYPE
NOMINAL
AMPERAGE
RECOMMENDED
CIRCUIT
ELECTRICITY
(EL)
200-208V 60 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
36 A 50 A
1. ALL CONNECTIONS SHOULD BE IN
ACCORDANCE WITH LOCAL CODES.
2. NEUTRAL WIRE NOT REQUIRED.
3. ALWAYS FOLLOW LOCAL ELECTRICAL CODES
AND SAFETY-RELATED WORK PRATICES TO
DETERMINE THE VALUE OF PROTECTION.
4. WIRING ON THE EQUIPMENT TERMINATES AT
A JUNCTION BOX. WIRING BETWEEN JUNCTION
BOX AND BUILDING SERVICE LINES ARE NOT
SUPPLIED BY STERIS.
5. DISCONNECT SWITCHES WITH OFF POSITION
LOCKOUT (SUPPLIED BY OTHERS) MUST BE
INSTALLED IN ELECTRIC SUPPLY LINES NEAR THE
EQUIPMENT.
6. STERIS RECOMMENDS A LIGHT IN SERVICE
AREA.
200-208V 60 Hz, 3-Ph
WITH 40A
MIN. CIRCUIT
PROTECTION
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
28.5 A 40 A
200-208V 50 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
35 A 50 A
460-480V 60 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
16.5 A 30 A
380-400V 60 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
19 A 30 A
380-400-415V 50 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
18.5 A 30 A
NETWORK
(N)
OPTIONAL
100/1000 BASE T SUPPORTING
TCP/IP PROTOCOL
RJ45F CONNECTION
7. NETWORK CONNECTION (IF APPLICABLE)
- ACTIVE NETWORK DROP WITH CAT 5/6 CABLE
WITHIN 10 ft [3 m] OF UNIT
- STATIC OR DHCP RESERVED IP ADDRESS
(FOR EACH UNIT)
- NETWORK CABLE (SUPPLIED BY OTHER)
Page 9 of 10 10292221.D
UTILITY REQUIREMENTS CHART
VENTILATION
EXHAUST
(EX)
OPTIONAL
CONNECTION SIZE 3 in [76 mm]
OUTSIDE DIAMETER
1. NOT REQUIRED WITH NON VENTED UNIT.
2. STERIS REQUIRES CORROSION-RESISTANT AND
WATER-TIGHT DUCT:
- MIN OPERATING TEMPERATURE 210 °F [99 °C]
- SLOPED TOWARDS THE WASHER AND NO DEAD LEGS
- INSTALL DRAIN PART AT LOWEST POINT NEAR UNIT
3. EXHAUST FAN NORMALLY OPEN-AUXILIARY CONTACTS
IS AVAILABLE TO SUPPLY A SIGNAL TO THE FACILITY HVAC
CONTROL SYSTEM WHENEVER THE WASHER VENTILATION
EXHAUST FAN IS IN OPERATION.
PRESSURE RANGE 1/2 in [50 mm]
VACUUM AT CONNECTION
OPERATING FLOW RATE 60 SCFM [1.7 m³/min]
TEMPERATURE RANGE MAX. 240 °F [116 °C]
PHYSICAL SPECIFICATIONS
SHIPPING DIMENSION W x H x L 50 in x 91 in x 42 in [1270 mm x 2311 mm x 1067 mm]
1. LEVELING FEET ARE PROVIDED FOR PROPER INSTALLATION.
2. PERMISSIBLE ENVIRONMENTAL CONDITIONS.
THIS EQUIPMENT IS DESIGNED TO GIVE OPTIMAL RESULTS
UNDER THE FOLLOWING CONDITIONS:
- INDOOR USE ONLY;
- ALTITUDE OF OPERATION UP TO 6562 ft [2000 m];
- TEMPERATURE OF 41 TO 104 °F [5 to 40 °C];
- MAXIMUM RELATIVE HUMIDITY IS 80% FOR TEMPERATURE UP
TO 88 °F [31 °C], DECREASING LINEARLY TO 50% RELATIVE
HUMIDITY AT 104 °F [40 °C];
- POLLUTION DEGREE 2: EQUIPMENT MUST BE INSTALLED IN AN
ENVIRONMENT WHERE NORMALLY ONLY NON-CONDUCTIVE
POLLUTION OCCURS BUT WHERE OCCASIONAL, TEMPORARY
CONDUCTIVITY CAUSED BY CONDENSATION CAN BE EXPECTED
(AS DETERMINED ACCORDING TO INTERNATIONAL STANDARD
EN/IEC 61010-1).
3. STERIS ASSUMES NO RESPONSIBILITY FOR CHANGES MADE TO,
OR FAILURE TO OBSERVE THE SPECIFICATIONS ON EQUIPMENT
DRAWING AND NOTE PAGES. SPECIFICATIONS AND
DESCRIPTIONS ARE SUBJECT TO CHANGE WITHOUT NOTICE.
SHIPPING WEIGHT 1268 lb [575 kg]
OPERATING DIMENSION W x H x L 42 in x 80.75 in x 32 in [1067 mm x 2051 mm x 813 mm]
OPERATING WEIGHT
(WITH WATER AND FULLY LOADED) 1254 lb [569 kg]
AVERAGE FLOOR LOAD
OVER WASHER FOOTPRINT 0.93 psi [6.43 kPa]
MAX. FLOOR LOAD
AT WASHER FOOT PADS
(3 in [76.2 mm] DIAMETER PADS) 67 psi [462 kPa]
NOISE LEVEL:
A-WEIGHTED EQUIVALENT
SURFACE SOUND PRESSURE LEVEL
60.6 dB(A)
CALCULATED AS DESCRIBED
IN ISO 3746.
OPENING AREA FROM
CLEAN SIDE TO SOILED SIDE
151.8 in² [979 cm²]
*(NOTE 4)
4. OPENING AREA CAN BE REDUCED TO 26 in² [168 cm²]
WITH INSTALLATION OF AIR MANAGEMENT PANELS
ACCESSORY (FD352)
CHAMBER DIMENSION W x H x L
VOLUME
26.5 in x 26.25 in x 25.5 in [673 mm x 667 mm x 648 mm]
10.3 ft³ [0.29 m³]
TYPICAL HEAT LOSS AT 75 °F [24 °C]
CONTINUOUS OPERATION
NON-VENTED UNIT
WITHOUT ACU-RINSE RES.
VENTED UNIT (OPTIONAL)
WITHOUT ACU-RINSE RES.
CLEAN SIDE 2049 Btu/h [600 W]
2049 Btu/h [600 W]
SOILED SIDE 6557 Btu/h [1920 W]
4747 Btu/h [1390 W]
TOTAL 8606 ± 854 Btu/h [2520 ± 250 W]
6796 ± 854 Btu/h [1990 ± 250 W]
VALUES INCLUDE HEAT LOSS DURING DOOR OPENING FOR LOADING AND UNLOADING OPERATIONS. "TOTAL" WILL VARY DEPENDING ON DISINFECTION TIME AND TEMPERATURE, DRYING TIME AND TEMPERATURE AND USE RATE OF UNIT.
AN ADDITIONAL HEAT LOSS FOR LOAD COOL DOWN IN CLEAN SIDE CAN BE ADDED TO THE TOTAL HEAT LOSS OF THE WASHER:
MAXIMUM 1713 Btu [1807 kJ]
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT OFFSET FROM WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292221
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
Page 10 of 10 10292221.D
| Feuille 1 |
| Vues |
| new_view_35 |
| new_view_37 |
| new_view_57 |
| new_view_66 |
| Zones |
| A1 |
| Feuille 2 |
| Vues |
| new_view_55 |
| Zones |
| A1 |
| Feuille 3 |
| Vues |
| new_view_38 |
| new_view_39 |
| new_view_40 |
| Zones |
| A1 |
| Feuille 4 |
| Vues |
| new_view_44 |
| new_view_45 |
| new_view_46 |
| Zones |
| A1 |
| Feuille 5 |
| Vues |
| new_view_41 |
| new_view_42 |
| new_view_43 |
| new_view_65 |
| Zones |
| A1 |
| Feuille 6 |
| Vues |
| new_view_47 |
| Zones |
| A1 |
| Feuille 7 |
| Zones |
| A1 |
| Feuille 8 |
| Zones |
| A1 |
| Feuille 9 |
| Vues |
| new_view_64 |
| Zones |
| A1 |
| Feuille 10 |
| Zones |
| A1 |
File details come from the government source that posted it. Updated .