This equipment drawing document details the technical specifications for a STERIS AMSCO® 7052HP Single-Chamber Washer/Disinfector, a steam-heated medical/surgical instrument processing device designed for flush-to-wall installation. The comprehensive 10-page technical drawing provides extensive utility requirements, including precise plumbing, electrical, compressed air, steam, and ventilation connection specifications for healthcare facility installations.

Key technical details include: operating dimensions of 42 in x 80.75 in x 32 in, shipping weight of 1,268 lbs, operating weight of 1,254 lbs when fully loaded, and specific utility connection requirements such as 200-208V 3-phase electrical connection, steam inlet pressure range of 30-80 psig, and compressed air requirements of minimum 80 psig. The device features a 10.3 ft³ chamber volume, supports multiple voltage configurations, and has optional network connectivity. The document provides detailed installation guidelines, environmental operating conditions, and heat loss calculations, indicating it is intended for precise medical equipment procurement and installation planning.

View the file

Other files

Show all 22

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

42 [1067]

UNIT WIDTH

80-3/4 [2051]

UNIT

HEIGHT

31 [787]

LOADING

HEIGHT

26-1/4 [667]

LOADING

CLEARANCE

HEIGHT

8-1/2 [216]

[381]

91 [2311]

RAISED DOOR

HEIGHT

34 [864]

MAX LENGTH

32 [813]

UNIT LENGTH

90 [2286]

SERVICE

AREA

HEIGHT

45 [1143]

WALL OPENING

*(NOTE F)

1-1/2 [38]

MINIMUM

94 [2388] *(NOTE G)

RECOMMENDED

98 [2489] *(NOTE H)

MAX 1 [25]

ADJUSTMENT WITH

WASHER LEGS

(UNIT HEIGHT 81-3/4 [2077] MAX)

MINIMUM

94 [2388] *(NOTE G)

RECOMMENDED

96 [2438] *(NOTE H)

81-7/8 [2080]

MAX

HEIGHT

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

SERVICE AREA

REQUIRED FOR

DRYING MOTOR

REMOVAL

TYPICAL INSTALLATION

NOTES:

(A) UNLESS OTHERWISE SPECIFIED, ALL DIMENSIONS ARE IN INCHES [mm].

(B) SERVICE ACCESS SIDE OF UNIT IS RECOGNIZED BY THE TOUCH SCREEN AT THE RIGHT OF

WASH CHAMBER. ELECTRICAL AND CONTROL COMPONENTS, AS WELL AS MOST OF

SERVICABLE PARTS ARE ACCESSIBLE FROM THAT SIDE OF UNIT. NON-SERVICE SIDE IS

EQUIPPED WITH THE TOUCH SCREEN ON THE LEFT OF THE WASH CHAMBER.

(C) CUSTOMER MUST BE SURE THAT MACHINE STANDS ON A NON-COMBUSTIBLE FLOOR.

(D) WEIGHTED EQUIVALENT SURFACE SOUND PRESSURE LEVEL: 64.0 dB(A) CALCULATED AS DESCRIBED IN ISO 3746.

(E) VERTICAL BARRIER WALL FLANGES ARE PROVIDED WITH UNIT TO FILL GAP BETWEEN

UNIT AND WALL ON BOTH SIDES. DIMENSIONS ARE 3-1/2 in [89 mm] WIDTH X 80 in [2032 mm] LONG.

(F) HORIZONTAL BARRIER WALL FLANGES ARE PROVIDED WITH UNIT TO FILL GAP BETWEEN

UNIT AND WALL ON BOTH SIDES. DIMENSIONS ARE 49 in [1245 mm] WIDTH X 16 in [406 mm] HEIGHT.

(G) MINIMUM REQUIRED DIMENSION FOR DOOR REMOVAL WITH LIMITED ADJUSTMENT AVAILABLE: +0 in [0 mm] THE STANDARD BARRIER WALL FLANGES CAN'T BE USED AT THAT DIMENSION. -3/4 in [19.1 mm]

(H) RECOMMENDED DIMENSION FOR DOOR REMOVAL WITH FULL LEVELING ADJUSTMENT AVAILABLE: +1 in [25.4 mm] THE STANDARD BARRIER WALL FLANGES CAN BE USED AT THAT DIMENSION. *(NOTE E AND F) -3/4 in [19.1 mm]

FRONT VIEW

SERVICE ACCESS

SIDE OF UNIT

*(NOTE B)

TOP VIEW

PASS THROUGH

DOOR

SIDE VIEWLEVEL FLOOR

*(NOTE C)

W A

LL

O

P E

N

IN

G H

E

IG

H T

C E

IL

IN

G H

E

IG

H T

FO

R

D O

O R

R E

M O

V A

L

Page 1 of 10 10292220.D

Released

2X 33-7/16 [849]

2X 27-11/16 [704]

5-11/16 [144]

7-1/8 [181]

4X 15 [381]

2-1/8 [54]

17-3/8 [441]

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

*NOTES:

(A) CONCRETE FLOOR MUST REMAIN FREE OF DEFECTS IN THESE AREA, AS PER ESR-1917 & ESR-3187.

FOR INSTALLATION OF SEISMIC ANCHORS (NOT PROVIDED BY STERIS).

(B) THE FLOOR SINK AND DRAIN SHALL NOT OCCUPY THESE AREAS. 5 in x 5 in [127 mm x 127 mm] FLOOR SINK WITH Ø4 in [102 mm] DRAIN IS RECOMMENDED WHEN USING SEISMIC MOUNTS.

SEISMIC MOUNTS

SEISMIC ANCHORAGE

FOR DETAIL SEE FD354

QTY 4 ANCHORS

FINISHED WALL

PREFERENCE

LOCATION OF

FLOOR SINK

*(NOTE A AND B)

Page 2 of 10 10292220.D

80-3/4 [2051]

7-1/8 [181]

[381]

31-1/8 [791]

[533]

23-1/32 [585]

20-17/32 [521]

19-1/2 [495]

3-7/8 [99]

[457]

15-1/2 [394]

32-11/32 [822]

HW HOT WATER

CW COLD WATER

PW PURE WATER

A COMPRESSED AIR

D FLOOR DRAIN LOCATION

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

UTILITY (PLUMBING)

CONNECTION POSITION

FRONT VIEW

A

D

PW

HW

CW

SIDE VIEW

HW

PW

CW

A

D

CR

STOP VIEW

HW

CW

PW

A

CR

S

D

S STEAM

CR CONDENSATE RETURN

S

CR

Page 3 of 10 10292220.D

79-1/4 [2013]

35-13/16 [910]

22-3/8 [568]

[636]

38-1/8 [967]

39-5/8 [1007]

E ELECTRICAL

N NETWORK

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

UTILITY (ELECTRICITY)

CONNECTION POSITION

FRONT VIEW SIDE VIEW

E

N

E

N

TOP VIEW

E

N

Page 4 of 10 10292220.D

6-5/32 [156]

[178]

81-29/32 [2080]

EX VENTILATION EXHAUST

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

UTILITY (VENTILATION)

CONNECTION POSITION (OPTIONAL)

FRONT VIEW SIDE VIEW

EX

SUGGESTED DUCTWORK

(BY OTHERS)

EX

VAPOR CONDENSATE

COLLECTING MANAGED

BY THE UNIT

TOP VIEW

EX

VENTILATION CONNECTION (OPTIONAL)

(SEE PAGE 9 *(NOTE 1)

DRY CONTACT (SEE PAGE 10 *(NOTE 3))

EX

BY OTHERS

Page 5 of 10 10292220.D

60 [1524]

REF

8-1/2 [216]

132 [3350]

MAX.

17 [432]

INSTRUMENT

(With PUC HP)

UTENSIL

(With PUC HP)

CHEMICAL

(CH)

ENZYME CONSUMPTION PER CYCLE

(SEE PAGE 7; *(NOTE 4)) 0.74 oz [22 ml] 0.74 oz [22 ml]

DETERGENT CONSUMPTION PER CYCLE

(SEE PAGE 7; *(NOTE 4)) 0.74 oz [22 ml] 0.74 oz [22 ml]

LUBRICANT CONSUMPTION PER CYCLE

(SEE PAGE 7; *(NOTE 4)) 0.2 oz [6 ml] N/A

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

*NOTES:

(A) DOSING STATION AND CHEMICAL CONTAINERS MUST BE ON THE SAME FLOOR OF THE BUILDING AS THE WASHER/DISINFECTOR.

CHEMICAL CONTAINER MUST BE LOCATED AT A LOWER HEIGHT THAN DOSING STATION. MAXIMUM CONTAINER INCLINATION

IS 15 DEG. FROM FLOOR. AMBIENT TEMPERATURE SHALL NOT EXCEED 32 °C [90 °F]

(B) HP PICKUP TUBE SYSTEM CAN PROVIDE UP TO TWO WASHERS WITH CHEMISTRY FROM THE SAME CONTAINER.

DOSING STATION

CEILING

ROUTE CHEMICAL LINES

AND SIGNAL CABLE

THROUGH PIPE TO REACH

BOTTOM OF UNIT

CHEMICAL CONTAINERS

*(NOTE A)

12 ft [3.6 m] OF INJECTION LINE

(FROM CONTAINERS TO

DOSING STATION)

*(NOTE B)

DOSING STATION

WEIGHT : 15 lb [6.8 kg]

2X FD349 SHOWN

UP TO 5 TUBES

100 ft [30 m] OF INJECTION LINE

AND SIGNAL CABLE

DETERGENT LINES SHALL NOT BE IN CONTACT WITH HOT SURFACES.

STERIS RECOMMENDS CUSTOMER TO ROUTE CHEMICAL LINES AND

SIGNAL CABLE THROUGH A 3 in [76 mm] ID CONDUIT.

EXCESS LENGTH MUST BE CUT BEFORE CONNECTING TO

WASHER/STATION TO AVOID FORMING BUNDLES THAT COULD

AFFECT THE CHEMICAL INJECTION SYSTEM'S PERFORMANCE

(DO NOT CUT SIGNAL CABLE).

Page 6 of 10 10292220.D

UTILITY REQUIREMENTS CHART (PLUMBING)

UTILITY PARAMETER REQUIREMENT NOTES

HOT WATER

(HW)

CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]

1. STERIS REQUIRES TO INSTALL WATER HAMMER

ARRESTOR, UNION & PRESSURE GAUGE.

2. CUSTOMER MUST PROVIDE UTILITY

CONNECTIONS WITH SHUTOFFS DISCONNECTS

WITHIN 2 FEET OF THE PERIMETER OF THE

EQUIPMENT AND BELOW THE CEILING DECK.

*STERIS OFFERS FLEXIBLE HOSES

ACCESSORY (FD088 OR FD037)

WITH INSTRUCTIONS.

3. SAMPLING VALVES INSTALLATION AS CLOSE AS

POSSIBLE TO WASHER WATER CONNECTIONS IS

RECOMMENDED.

PRESSURE RANGE

15 - 50 psig DYNAMIC

[103-345 kPa] MAX. 90 psig STATIC [620 kPA]

OPERATING FLOW RATE 6.8 US GPM [25.7 L/min] TO

14.1 US GPM [53.4 L/min]

TOTAL CONSUMPTION PER

INSTRUMENT CYCLE

*(NOTE 4)

9.9 US GAL [37.5 L]

WITH ACU-WASH RESERVOIR:

6.6 US GAL [25.0 L]

TEMPERATURE RANGE 110 °F [43 °C] TO 150 °F [65.5 °C]

WATER QUALITY *(NOTE 5) HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)

COLD WATER

(CW)

*(NOTES 6)

CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]

4. CONSUMPTION BASED ON INSTRUMENT CYCLE WITH

PROLYSTICA ULTRA CONCENTRATE HIGH

PERFORMANCE CHEMICALS (PUC HP) WITH

THE FOLLOWING UTILITY CONDITIONS:

THREE-LEVEL MANIFOLD RACK

(CW) COLD WATER PRESSURE: 30 psig DYNAMIC [207 kPa]

(CW) COLD WATER TEMPERATURE: 70 °F [21 °C]

(HW) HOT WATER TEMPERATURE: 150 °F [65 °C]

(PW) PURE WATER TEMPERATURE 70 °F [21 °C]

(S) STEAM SUPPLY: 80 psig DYNAMIC [552 kPa]

UTENSIL CYCLE CONSUMPTION:

TWO-LEVEL MANIFOLD RACK

CW: 10.4 US GAL [39.4 L]

HW: 9.6 US GAL [36.3 L]

PW: 6.4 US GAL [24.2 L]

S: 14.1 lb [6.4 kg]

5. STERIS RECOMMENDS THE FOLLOWING WATER

QUALITY PARAMETERS:

TOTAL DISSOLVED SOLIDS: 100-200 ppm (MAX. 500 ppm) TOTAL ALKALINITY (CaCO3) : 70-120 ppm (MAX. 180 ppm) pH : 6.8-7 (MAX. 7.5) TOTAL SILICA : 0.1-0.5 ppm (MAX. 1 ppm)

PRESSURE RANGE

30 - 50 psig DYNAMIC

[206-345 kPa] MAX. 90 psig STATIC [620 kPa]

OPERATING FLOW RATE 10.7 US GPM [40.5 L/min] TO

14.1 US GPM [53.4 L/min]

TOTAL CONSUMPTION PER

INSTRUMENT CYCLE

*(NOTE 4)

FOR SUMP FILLING: 10.5 US GAL [39.8 L]

(ADD THE FOLLOWING CONSUMPTION

FOR OPTIONS/CONFIGURATIONS)

CONDENSATE RETURN COOL DOWN:

6.2 US GAL [23.5 L]

NON-VENTED SYSTEM:

12.9 US GAL [48.8 L]

DRAIN COOLDOWN:

14.5 US GAL [54.9 L]

DRAIN ACU-COOLDOWN:

10.9 US GAL [41.3 L]

TEMPERATURE RANGE 70 °F [21 °C] MAXIMUM

WATER QUALITY *(NOTE 5) HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)

PURE WATER

(PW)

CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]

6. BACK FLOW PREVENTER AGAINST BACK

SIPHONAGE MAY BE REQUIRED (LOCAL

PLUMBING CODE)

7. PW IS OPTIONAL. IF UNAVAILABLE, UNIT WILL

PROCESS THERMAL RINSE USING HOT WATER

SUPPLY.

PRESSURE RANGE 5 - 30 psig DYNAMIC [30 - 206 kPa] MAX. 90 psig STATIC [620 kPa]

OPERATING FLOW RATE 5.5 US GPM [21 L/min] TO

12.6 US GPM [48 L/min]

TOTAL CONSUMPTION PER

INSTRUMENT CYCLE

* (NOTE 4 AND NOTE 7 )

6.6 US GAL [25.0 L]

TEMPERATURE RANGE 70 °F [21.1 °C] TO 150 °F [65.5 °C]

RESISTIVITY

(CONDUCTIVITY)

0.1-0.5 M -cm (2-10 µS/cm) pH 6.8 - 7.5

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

Page 7 of 10 10292220.D

UTILITY REQUIREMENTS CHART (PLUMBING)

COMPRESSED

AIR

(A)

CONNECTION SIZE / TYPE 1/8 in / FEMALE NPT [1/8 in / MALE BSPT]

1. AIR PURITY PER ISO-8573-1: 2010 [7:4:4]

MAX. PARTICLE DENSITY: 10 mg/m³

MAX. DEW POINT FOR WATER CONTENT:

37 °F [3 °C]

MAX. OIL CONCENTRATION FOR THE OIL

CONTENT: 5 mg/m³

2. STERIS RECOMMENDS A REFRIGERATED AIR

DRYER WHERE ENVIRONMENTAL DEW POINT

CONDITIONS ARE HIGHER THAN

RECOMMENDED.

3. CUSTOMER MUST PROVIDE UTILITY

CONNECTIONS WITH SHUTOFFS DISCONNECTS

WITHIN 2 FEET OF THE PERIMETER OF THE

EQUIPMENT AND BELOW THE CEILING DECK.

PRESSURE RANGE MIN 80 psig DYNAMIC [552 kPa] MAX. 125 psig STATIC [860 kPa]

OPERATING FLOW RATE 1.75 SCFM [50 L/min]

DRAIN

(D)

FLOOR DRAIN SIZE MIN 4 in [102 mm] O.D. WITH 8 in x 8 in [204 mm x 204 mm] FLOOR SINK

4. WASHER DRAIN OUTLET SIZE 1-1/2 in / MALE NPT

5. MAX OFFSET OF 3/4 in [19 mm] FROM

OUTLET (PAGE 3, D).

6. FOR DRAIN REQUIREMENTS WHEN USING

SEISMIC MOUNTS, SEE PAGE 2.

OPERATING FLOW RATE GRAVITY DRAIN : 50 US GPM

[189.3 L/min]

CHEMICAL

DOSING

SYSTEM

CEILING ROUTING

CONDUIT 3 in CONDUIT

11. STERIS RECOMMENDS TO ROUTE CHEMICAL

LINES AND SIGNAL CABLE THROUGH A

3 in [76 mm] ID CONDUIT (BY OTHERS).

100 ft [30 m] OF INJECTION LINE AND

SIGNAL CABLE ARE PROVIDED WITH THE UNIT.

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

STEAM INLET

(S)

CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]

7. STERIS REQUIRES AT LEAST 97% SATURATED

DRY PLANT STEAM, FREE OF DIRT, RESIDUES

AND CONTAMINANTS.

8. STERIS RECOMMENDS TO INSTALL A STEAM

TRAP AS CLOSE AS POSSIBLE TO UNIT

CONNECTION.

9. PROPER OPERATION OF THE WASHER

REQUIRES A MINIMUM PRESSURE DIFFERENTIAL

OF 21.5 psig [148 kPa] BETWEEN STEAM

INLET AND CONDENSATE RETURN. IN CASES OF

LOW STEAM INLET PRESSURE, CONDENSATE

CAN BE ROUTED TO DRAIN TO INCREASE

DIFFERENTIAL.

*STERIS OFFERS CONDENSATE RETURN

TO DRAIN ACCESSORY (FD353)

10. CUSTOMER MUST PROVIDE UTILITY

CONNECTIONS WITH SHUTOFFS DISCONNECTS

WITHIN 2 FEET OF THE PERIMETER OF THE

EQUIPMENT AND BELOW THE CEILING DECK.

PRESSURE RANGE

30 - 80 psig DYNAMIC

[206-552 kPa] MAX. 90 psig STATIC [620 kPa]

OPERATING

MASS FLOW RATE

169 lb/h [76.5 kg/h] AT 30 psig [207 kPa]

271 lb/h [123 kg/h AT 80 psig [550 kPa]

TOTAL CONSUMPTION

PER INSTRUMENT CYCLE

(SEE PAGE 7, *(NOTE 4))

14.1 lb [6.4 kg]

CONDENSATE

RETURN

(CR)

PRESSURE RANGE MAX. 8.5 psig [68.9 kPa] BACK PRESSURE (17 ft [5.2 m] ELEVATION)

OPERATING FLOW RATE PEAK : 0.54 US GPM [2.05 L/min]

CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]

Page 8 of 10 10292220.D

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

UTILITY REQUIREMENTS CHART (ELECTRIC)

UTILITY

PARAMETER AND REQUIREMENT NOTES

CONNECTION

SIZE/TYPE

NOMINAL

AMPERAGE

RECOMMENDED

CIRCUIT

ELECTRICITY

(EL)

200-208V 60 Hz, 3-Ph

1 in [27 mm] 1-1/4 in [35 mm]

CONDUIT

36 A 50 A

1. ALL CONNECTIONS SHOULD BE IN

ACCORDANCE WITH LOCAL CODES.

2. NEUTRAL WIRE NOT REQUIRED.

3. ALWAYS FOLLOW LOCAL ELECTRICAL CODES

AND SAFETY-RELATED WORK PRATICES TO

DETERMINE THE VALUE OF PROTECTION.

4. WIRING ON THE EQUIPMENT TERMINATES AT

A JUNCTION BOX. WIRING BETWEEN JUNCTION

BOX AND BUILDING SERVICE LINES ARE NOT

SUPPLIED BY STERIS.

5. DISCONNECT SWITCHES WITH OFF POSITION

LOCKOUT (SUPPLIED BY OTHERS) MUST BE

INSTALLED IN ELECTRIC SUPPLY LINES NEAR THE

EQUIPMENT.

6. STERIS RECOMMENDS A LIGHT IN SERVICE

AREA.

200-208V 60 Hz, 3-Ph

WITH 40A

MIN. CIRCUIT

PROTECTION

1 in [27 mm] 1-1/4 in [35 mm]

CONDUIT

28.5 A 40 A

200-208V 50 Hz, 3-Ph

1 in [27 mm] 1-1/4 in [35 mm]

CONDUIT

35 A 50 A

460-480V 60 Hz, 3-Ph

1 in [27 mm] 1-1/4 in [35 mm]

CONDUIT

16.5 A 30 A

380-400V 60 Hz, 3-Ph

1 in [27 mm] 1-1/4 in [35 mm]

CONDUIT

19 A 30 A

380-400-415V 50 Hz, 3-Ph

1 in [27 mm] 1-1/4 in [35 mm]

CONDUIT

18.5 A 30 A

NETWORK

(N)

OPTIONAL

100/1000 BASE T SUPPORTING

TCP/IP PROTOCOL

RJ45F CONNECTION

7. NETWORK CONNECTION (IF APPLICABLE)

- ACTIVE NETWORK DROP WITH CAT 5/6 CABLE

WITHIN 10 ft [3 m] OF UNIT

- STATIC OR DHCP RESERVED IP ADDRESS

(FOR EACH UNIT)

- NETWORK CABLE (SUPPLIED BY OTHER)

Page 9 of 10 10292220.D

UTILITY REQUIREMENTS CHART

VENTILATION

EXHAUST

(EX)

OPTIONAL

CONNECTION SIZE 3 in [76 mm]

OUTSIDE DIAMETER

1. NOT REQUIRED WITH NON VENTED UNIT.

2. STERIS REQUIRES CORROSION-RESISTANT AND

WATER-TIGHT DUCT:

- MIN OPERATING TEMPERATURE 210 °F [99 °C]

- SLOPED TOWARDS THE WASHER AND NO DEAD LEGS

- INSTALL DRAIN PART AT LOWEST POINT NEAR UNIT

3. EXHAUST FAN NORMALLY OPEN-AUXILIARY CONTACTS

IS AVAILABLE TO SUPPLY A SIGNAL TO THE FACILITY HVAC

CONTROL SYSTEM WHENEVER THE WASHER VENTILATION

EXHAUST FAN IS IN OPERATION.

PRESSURE RANGE 1/2 in [50 mm]

VACUUM AT CONNECTION

OPERATING FLOW RATE 60 SCFM [1.7 m³/min]

TEMPERATURE RANGE MAX. 240 °F [116 °C]

PHYSICAL SPECIFICATIONS

SHIPPING DIMENSION W x H x L 50 in x 91 in x 42 in [1270 mm x 2311 mm x 1067 mm]

1. LEVELING FEET ARE PROVIDED FOR PROPER INSTALLATION.

2. PERMISSIBLE ENVIRONMENTAL CONDITIONS.

THIS EQUIPMENT IS DESIGNED TO GIVE OPTIMAL RESULTS

UNDER THE FOLLOWING CONDITIONS:

- INDOOR USE ONLY;

- ALTITUDE OF OPERATION UP TO 6562 ft [2000 m];

- TEMPERATURE OF 41 TO 104 °F [5 to 40 °C];

- MAXIMUM RELATIVE HUMIDITY IS 80% FOR TEMPERATURE UP

TO 88 °F [31 °C], DECREASING LINEARLY TO 50% RELATIVE

HUMIDITY AT 104 °F [40 °C];

- POLLUTION DEGREE 2: EQUIPMENT MUST BE INSTALLED IN AN

ENVIRONMENT WHERE NORMALLY ONLY NON-CONDUCTIVE

POLLUTION OCCURS BUT WHERE OCCASIONAL, TEMPORARY

CONDUCTIVITY CAUSED BY CONDENSATION CAN BE EXPECTED

(AS DETERMINED ACCORDING TO INTERNATIONAL STANDARD

EN/IEC 61010-1).

3. STERIS ASSUMES NO RESPONSIBILITY FOR CHANGES MADE TO,

OR FAILURE TO OBSERVE THE SPECIFICATIONS ON EQUIPMENT

DRAWING AND NOTE PAGES. SPECIFICATIONS AND

DESCRIPTIONS ARE SUBJECT TO CHANGE WITHOUT NOTICE.

SHIPPING WEIGHT 1268 lb [575 kg]

OPERATING DIMENSION W x H x L 42 in x 80.75 in x 32 in [1067 mm x 2051 mm x 813 mm]

OPERATING WEIGHT

(WITH WATER AND FULLY LOADED) 1254 lb [569 kg]

AVERAGE FLOOR LOAD

OVER WASHER FOOTPRINT 0.93 psi [6.43 kPa]

MAX. FLOOR LOAD

AT WASHER FOOT PADS

(3 in [76.2 mm] DIAMETER PADS) 67 psi [462 kPa]

NOISE LEVEL:

A-WEIGHTED EQUIVALENT

SURFACE SOUND PRESSURE LEVEL

60.6 dB(A)

CALCULATED AS DESCRIBED

IN ISO 3746.

OPENING AREA FROM

CLEAN SIDE TO SOILED SIDE

46.5 in² [300 cm²]

*(NOTE 4)

4. OPENING AREA CAN BE REDUCED TO 13 in² [84 cm²]

WITH INSTALLATION OF AIR MANAGEMENT PANELS

ACCESSORY (FD352)

CHAMBER DIMENSION W x H x L

VOLUME

26.5 in x 26.25 in x 25.5 in [673 mm x 667 mm x 648 mm]

10.3 ft³ [0.29 m³]

TYPICAL HEAT LOSS AT 75 °F [24 °C]

CONTINUOUS OPERATION

NON-VENTED UNIT

WITHOUT ACU-RINSE RES.

VENTED UNIT (OPTIONAL)

WITHOUT ACU-RINSE RES.

CLEAN SIDE 2049 Btu/h [600 W]

2049 Btu/h [600 W]

SOILED SIDE 6557 Btu/h [1920 W]

4747 Btu/h [1390 W]

TOTAL 8606 ± 854 Btu/h [2520 ± 250 W]

6796 ± 854 Btu/h [1990 ± 250 W]

VALUES INCLUDE HEAT LOSS DURING DOOR OPENING FOR LOADING AND UNLOADING OPERATIONS. "TOTAL" WILL VARY DEPENDING ON DISINFECTION TIME AND TEMPERATURE, DRYING TIME AND TEMPERATURE AND USE RATE OF UNIT.

AN ADDITIONAL HEAT LOSS FOR LOAD COOL DOWN IN CLEAN SIDE CAN BE ADDED TO THE TOTAL HEAT LOSS OF THE WASHER:

MAXIMUM 1713 Btu [1807 kJ]

T I T L E

EQUIPMENT DRAWING OF

AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR

STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Jean Tanguay 10292220

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

Page 10 of 10 10292220.D

Feuille 1
Vues
new_view_35
new_view_37
new_view_57
new_view_66
Zones
A1
Feuille 2
Vues
new_view_55
Zones
A1
Feuille 3
Vues
new_view_38
new_view_39
new_view_40
Zones
A1
Feuille 4
Vues
new_view_44
new_view_45
new_view_46
Zones
A1
Feuille 5
Vues
new_view_41
new_view_42
new_view_43
new_view_65
Zones
A1
Feuille 6
Vues
new_view_47
Zones
A1
Feuille 7
Zones
A1
Feuille 8
Zones
A1
Feuille 9
Vues
new_view_64
Zones
A1
Feuille 10
Zones
A1

File details come from the government source that posted it. Updated .