Technical specification document for STERIS AMSCO 400 Series Medium Steam Sterilizers, detailing medical equipment for sterilization processes. The document covers multiple sterilizer configurations with chamber dimensions of 26 x 37.5 x 36", 26 x 37.5 x 48", and 26 x 37.5 x 60", offering single and double door options in horizontal-sliding and hinged door models. Key features include prevacuum sterilization capabilities, touch-sensitive control screens, multiple processing cycles (Prevac 4-minute and 3-minute, Gravity, and Liquid cycles), and comprehensive utility requirements including steam (50-80 psig), electrical (various voltage options), and specific water quality recommendations.
The specification provides detailed technical insights including standards compliance (UL, ASME), construction materials (Type 316L stainless steel), safety features, control system design, and utility conservation features. Operational weights range from 3800 lb (1720 kg) for 36" models to 4700 lb (2125 kg) for 60" models. Accessories include loading cars, transfer carriages, extra shelves, and seismic tie-down kits. The document also highlights ProConnect Technical Support Services for equipment monitoring and maintenance, with availability limited to U.S. and Canada.
View the file
Other files
Show all 22
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
AMSCO® 400 SERIES
MEDIUM STEAM STERILIZERS
Horizontal-Sliding Door Model Hinged-Door Model
(Typical – details may vary.)
ns Checked Below Apply To This Equipment
APPLICATION
This Prevacuum sterilizer is designed for fast, efficient sterilization of heat- and moisture-stable materials. Prevacuum configuration sterilizers are equipped with prevacuum, gravity, liquid, leak test and daily air removal (Bowie-Dick) test cycles.
DESCRIPTION
AMSCO 400 Series Medium Steam Sterilizers are equipped with the latest features in both state-of-the-art technology and ease of use.
Product Configurations
Interior Chamber Dimensions
• 26 x 37.5 x 36" (660 x 950 x 910 mm)
• 26 x 37.5 x 48" (660 x 950 x 1220 mm)
• 26 x 37.5 x 60" (660 x 950 x 1520 mm)
36", 48" and 60" configurations include the choice of a single or double door.
• Hinged door with fast operating, low-effort door lock mechanism. Door
The Selectio
SIZE
❑ 26 x 37.5 x 36" (660 x 950 x 910 mm) ❑ 26 x 37.5 x 48" (660 x 950 x 1220 mm) ❑ 26 x 37.5 x 60" (660 x 950 x 1520 mm)
VACUUM PUMP ELECTRIC SERVICE
❑ 208/240 Vac, 60 Hz, 3-phase, 6A per phase ❑ 380/415 Vac, 50/60 Hz, 3-phase, 3 A per phase ❑ 480 Vac, 60 Hz, 3-phase, 3A per phase
DOOR CONFIGURATIONS
Horizontal Sliding Door– Only A and C apply Hinged Door– A through F apply
HORIZONTAL SLIDING DOOR
❑ SINGLE DOOR - Horizontal Sliding ❑ (A) Left-Hand ❑ Cabinet Enclosed/Freestanding ❑ Recessed
❑ DOUBLE DOOR - Horizontal Sliding ❑ (C) Left-Hand/Right Hand 1
❑ Recessed through One Wall ❑ Recessed through Two Walls handle lock lever requires a single 30° handle movement to lock or unlock.
• Horizontal-sliding door with quiet, motor-driven cable and pulley mechanism.
Door travels horizontally right to left to open, and is controlled from the touch screen.
Item ________________________
Location(s)___________________
SD958 Rev B
HINGED DOOR
❑ SINGLE DOOR - Hinged ❑ (A) Left-Hand ❑ (B) Right-Hand ❑ Cabinet Enclosed/Freestanding ❑ Recessed
❑ DOUBLE DOOR - Hinged ❑ (C) Left-Hand/Right-Hand 1
❑ (D) Left-Hand/Left-Hand 1
❑ (E) Right-Hand/Left-Hand 1
❑ (F) Right-Hand/Right-Hand 1
❑ Recessed through One Wall ❑ Recessed through Two Walls
ACCESSORIES
36" UNITS
❑ Loading Car (AX067013081) ❑ Transfer Carriage (Hinged Door Only)
(AX0640230830000013)
❑ Transfer Carriage (Sliding Door Only)
(AX063023083)
❑ Extra Shelf Support (AX067823081) ❑ Extra Side Gate (AX067813081) ❑ Chamber Rack and Two Shelves (AX067111) ❑ Seismic Tie-Down Kit 2
48" UNITS
❑ Loading Car (AX068013081) ❑ Transfer Carriage (Hinged Door Only)
(AX0650230830000013)
❑ Transfer Carriage (Sliding Door Only)
(AX063023083)
❑ Shelf with Supports (AX068803081) ❑ Extra Side Gate (AX068813081) ❑ Extra Shelf Support (AX068823081) ❑ Seismic Tie-Down Kit 2
60" UNITS
❑ Loading Car (AX070013081) ❑ Transfer Carriage (Hinged Door Only)
(AX0660230830000013)
❑ Transfer Carriage (Sliding Door Only)
(AX063023083)
❑ Shelf with Supports (AX070803081) ❑ Extra Side Gate (AX070813081) ❑ Extra Shelf Support (AX070823081) ❑ Seismic Tie-Down Kit 2
REMOTE MONITORING
❑ ProConnect® Technical Support Services (Remote
Monitoring, Priority Technical Support, Customer Care Center Access, Equipment Performance Reports)
(GP09162)
Notes:
1. Hinged and Horizontal Sliding Double-Door Units – Operating End (OE) door opening direction is listed first, Non-operating End (NOE) direction second. Directions are given in relation to facing OE and NOE, respectively.
Refer to configuration figure.
2. Based on CA requirements.
STANDARDS
Each sterilizer meets the applicable requirements of the following listings and standards, and carries the appropriate symbols:
• Underwriters Laboratory (UL) Standard 61010-1 as certified by ETL Testing Laboratories, Inc.
• CAN/CSA-C22.2 No. 61010-1.
• ASME Code, Section VIII, Division 1 for unfired pressure vessels. The pressure vessel is so stamped; ASME Form U-1 is provided with shipment. Shell and door are constructed to withstand working pressure of 45 psig.
FEATURES
26 x 37.5" chamber cross-sections are sized for efficient, high-volume sterilization processing.
Fast-operating, low-effort manual door lock mechanism (hinged-door models) allows door to be locked or unlocked using a single 30° handle motion.
Power door drive systems consists of a cable and pulley arrangement driven by an electric motor. Power door is arranged horizontally and controlled from the touch screen.
Door slides horizontally on a rigid rail assembly housed within stainless-steel paneling.
The control system for the AMSCO 400 Series Medium Steam Sterilizer features enhanced functionality and user-friendly interface screen.
• Touch-sensitive screen with 30-line x 40-character display area.
• Ink-on-paper impact printer.
• Help screens for programming and troubleshooting alarm conditions.
• Automatic check of control program and cycle data maintains process integrity.
Software calibration is performed in service mode using external or internal temperature and pressure sources. Control system provides a printed record of all calibration data for verification.
Steam purge feature is provided to assist in air removal and to preheat the load.
Lighted DIN connectors are installed on all steam, water and exhaust valves, for reliability and ease of maintenance.
ProConnect® Technical Support Services – Maximize operational efficiencies with secure, internet-based, real-time equipment monitoring. Data from your equipment is used by STERIS to provide pro-active Customer alert notifications, technical support, and predictive maintenance. Online parts ordering, equipment performance dashboards, and online service scheduling at steris.com is also available. (ProConnect Technical Support Services is available in U.S. and Canada only.) Refer to Tech Data sheet SD983, PROCONNECT TECHNICAL SUPPORT SERVICES, for details.
UTILITIES CONSERVATION FEATURES
Vacuum pump effectively pulls chamber to specified vacuum levels and reduces water consumption.
Resistance Temperature Detectors (RTD) are installed for sterilizer temperature control. Dual element chamber drain line RTD senses and controls temperature variations within sterilizer chamber. Dedicated RTD provides jacket-space temperature control.
Electronic water saving control includes an RTD to control water quantity used in condensing chamber steam exhaust.
Automatic utilities start-up/shutdown permits utilities conservation. Shutdown may be programmed to activate at end of designated cycle or time of day. When used, utilities control automatically shuts off utility valves, conserving steam and water. Sterilizer utilities can be restarted automatically or manually. Different daily shutdown and restart times can be programmed.
Insulation, one-inch thick, asbestos-free spin-glass (rated at 500°F [260°C] continuous) encompasses the exterior of the sterilizer vessel and is sealed in an oil and water resistant outer jacket.
PROCESSING CYCLES
All processing cycles factory programmed into the sterilizer control have been validated to AAMI ST8.
AMSCO 400 Series Medium Steam Sterilizers, models 36 H, 48 H, 60 H, 36 SL, 48 SL and 60 SL are provided with the following default cycles:
Prevac Cycles (4-minute exposure): Cycle type is for sterilizing porous and non-porous loads. Wrapped 25 lb (11 kg) instrument tray(s) or fabric packs.
» Sterilize exposure temperature: 270°F (132°C)
» Sterilize exposure time: 4 minutes
» Dry time: 30 minutes (full load of instruments trays), 20 minutes (full load of fabric packs) or 5 minutes (Customer option, for a single fabric pack)
Prevac Cycle (3-minute exposure): This cycle is for sterilizing porous and non-porous loads. Wrapped 25 lb (11 kg) instrument trays.
» Sterilize exposure temperature: 275°F (135°C)
» Sterilize exposure time: 3 minutes
» Dry time: 30 minutes
Gravity Cycle (30-minute exposure): This cycle is for sterilizing items such as a single fabric pack.
» Sterilize exposure temperature: 250°F (121°C)
» Sterilize exposure time: 30 minutes
» Dry time: 15 minutes
Liquid Cycle (45-minute exposure): This cycle is used for sterilizing liquids in borosilicate containers with vented closures.
» Sterilize temperature: 250°F (121°C)
» Factory programmed sterilize time: 45 minutes
» Dry time: not applicable
TESTING CYCLES
• Vacuum Leak Test is used for testing vacuum integrity of sterilizer piping. Sterilizer chamber must be empty while running this test. Preprogrammed cycle parameters cannot be adjusted by user.
• A preprogrammed Bowie-Dick Test Cycle is used to test for adequate air removal from the sterilizer chamber.
Recommended load is a Dart® testing apparatus from STERIS, or a properly prepared Bowie-Dick test pack.
Preprogrammed cycle parameters cannot be adjusted by user.
» Sterilize temperature: 270°F (132°C)
» Sterilize time: 3.5 minutes
» Dry time: 1 minute
CONTROL SYSTEM
Design Features The Control System for AMSCO 400 Series Medium Steam Sterilizers monitors and controls all sterilizer operations and functions. The control system is factory-programmed with standard sterilizing cycles. Cycle values and operating features may be adjusted and verified prior to cycle operation.
Important: If cycle parameters (sterilize time, dry time, temperature) other than those listed are required, it is the responsibility of the healthcare facility to consult and follow the device manufacturer’s written instructions.
Once the cycle is started, cycle values cannot be changed until cycle completes.
Critical control system components are housed within a sealed compartment for protection from moisture and heat. A compartment cooling fan with filter is provided.
Operator interface control panel consists of a touch screen and impact printer located at sterilizer’s operating end (OE – loading or non-sterile side). If sterilizer is equipped with double doors, an additional touch screen is provided at sterilizer's non-operating (i.e., unload or sterile) end.
• Touch-Sensitive Screen features a multi-color, touch-screen graphics display. Sterilizer is operated by pressing touch-sensitive areas on display (touch pads). Appropriate touch pads, operator prompts and status messages necessary for sterilizer operation appear on display as needed. Display also indicates any alarm conditions either in or out of cycle.
• Ink-On-Paper Impact Printer, located near touch screen, provides an easy-to-read printed record of all pertinent cycle data on 2-1/4" wide paper. Printer take-up spool stores an entire roll of paper. Three paper tape rolls and two printer ribbons are furnished with each unit.
Non-Operating End (NOE) Control Panel (double-door sterilizers only) includes a touch screen similar to operating end, but without printer. Preprogrammed cycles can be started from NOE control panel. OE and NOE displays concurrently show identical information.
Cycle Configuration is accomplished at OE control Change Values menu. Cycle values and following operating parameters can be changed using Change Values:
• Time Display and Printout Units AM/PM or 24-hour notation.
• Access Code requires entry of a four-digit access code to change cycle values or change operating parameters.
• Audible Signals are adjustable. Touch pad and end-of-cycle signals can be adjusted to low, medium, high or off.
Alarm signal can be adjusted to low, medium or high; but cannot be turned off.
• Print Format allows selection of full or condensed cycle printout.
• Temperature Display and Printout Units are Fahrenheit (°F) or Celsius (°C). Temperature is set, displayed, controlled and printed to nearest 1°. Recalibration is not required when changing temperature units.
• Pressure/Vacuum Display and Printout Units in psig/InHg or bar. Recalibration is not required when changing pressure units.
SAFETY FEATURES
Control lockout switch prevents cycle from starting until limit switch signal is received. If signal is lost during cycle, alarm activates, cycle aborts and chamber safely vents controlled exhaust.
Chamber float switch activates alarm, aborts cycle and safely vents chamber if detecting excessive condensate.
Pressure relief valve releases pressure buildup in chamber if approaching or reaching rated pressure.
Power door safety feature causes door movement to stop if sliding door encounters an obstruction.
CONSTRUCTION
Shell Assembly Sterilizer vessel is formed from two fabricated Type 316L stainless-steel shells, welded one within the other. Type 316L stainless-steel end frame(s) is welded to the door end. On single door units, the back of the chamber is fitted with a welded 316L stainless-steel dished head.
Sterilizer vessel is ASME rated at 45 psig and insulated. Vessel includes one 1"-NPT chamber port for Customer use.
Steam-supply opening inside chamber is shielded by a stainless-steel baffle.
Chamber Door(s) Door is constructed of Type 316L stainless steel.
During operation, door is sealed by steam-activated door seal.
Door seal is constructed of long-life rubber compound. When cycle is complete, seal retracts (under vacuum) into a machined groove in end frame.
Long-life proximity switch detects if door is closed, or locking plates are engaged. An additional seal pressure switch prevents starting a cycle if door is not sealed.
Hinged-door sterilizer is equipped with a solenoid door lock.
Horizontal sliding-door units are provided with a mechanically-locking door assembly. Locking assemblies prevent door from opening as long as seal is intact, energized and chamber is pressurized at 2 psi or more.
The sterilizer door is fitted with a stainless-steel panel that insulates operator from chamber end frame, reducing chances of accidental contact with hot metal surfaces.
Chamber Drain System Drain system is designed to prevent sterilizer effluent from entering water-supply system and chamber. An automatic system of a stainless-steel plate-type condenser, converts chamber steam to condensate, disposing condensate to waste. Cooling water flow is regulated by waste line RTD, minimizing water usage. Water supply shut off valve is located in unit’s recessed area.
Vacuum System Vacuum pump reduces chamber pressure during prevacuum and post-drying phases. Following dry phase, chamber vacuum is relieved to atmospheric pressure by admitting air through a bacteria-retentive filter.
Steam Source Sterilizers are piped, valved and trapped to receive building-supplied steam at 50 to 80 psig, dynamic. Steam piping is brass and includes a shut-off valve, steam strainer and pressure regulator.
Piping All piping connections terminate within confines of sterilizer frame and are accessible from front, left side.
• Solenoid Valves with DIN connectors simplify sterilizer piping, and can be serviced individually.
• Manual Shut-off Valves are pressure rated at 125 psig (8.62 bar) for saturated steam. Valve handles are heat-insulated.
MOUNTING ARRANGEMENT
Sterilizers are arranged for either freestanding or recessed installation, as specified. Each sterilizer is height-adjustable.
Sterilizer subframe is equipped with a synthetic rubber gasket to provide a tight fit between cabinet panels on freestanding units, or between front cabinet panel and wall partition on recessed units.
On freestanding units, stainless-steel side panels and a louvered top panel enclose sterilizer body and piping.
ACCESSORY
Seismic Tie-Down Kit conforms to the California Code of Regulations.
Material Handling Accessories include stainless-steel chamber tracks and stainless-steel loading cars with painted-steel carriages. Stainless-steel chamber rack and shelf are available for 36" sterilizers. See TECH DATA SD978 for details.
PREVENTIVE MAINTENANCE
A global network of skilled service specialists can provide periodic inspections and adjustments to help ensure low-cost peak performance. STERIS representatives can provide information regarding annual maintenance agreements.
NOTES
1. The sterilizer is not supplied with a backflow preventer and, where required by local codes, installation of a backflow preventer in the water line is not provided by STERIS.
2. Pipe sizes shown indicate terminal outlets only. Building service lines, provided by third parties, must supply the specified pressures and flow rates.
3. Disconnect switches (with OFF position lockout only; by third parties) should be installed in electric supply lines near the equipment.
4. Access to the recessing area from the control end of the sterilizer is recommended.
5. Clearances shown are minimal for installing and servicing the equipment.
6. Depending on the loading equipment used, additional clearance is required:
• If shelves are used, clearance should be length of the sterilizer plus 24" (610 mm) at each door.
• If loading car and carriage are used, clearance should be twice the length of the sterilizer at each door.
7. Floor drain should be provided within confines of the sterilizer framework.
8. Units require a minimum 38" door opening for transport within facility prior to installation.
UTILITY REQUIREMENTS
Steam 1" NPT, 50 to 80 psig (3.5 to 5.6 bar) dynamic, 97% to 100% vapor quality.
Drain 2" ODT drain terminal. (Floor drain capacity must handle peak water consumption; refer to Utilities Consumption, page 6.)
Electrical – Controls 120 Volt, 50/60 Hz, 1-phase, 2.0 Amp
Electrical – Vacuum Pump
• 208/240 Vac, 50/60 Hz, 3-phase, 6 Amp per phase
• 380/415 Vac, 50/60 Hz, 3-phase, 3 A per phase
• 480 Vac, 50/60 Hz, 3-phase, 3 Amp per phase
Sterilizer Feed Water 1" NPT, 20 to 50 psig (1.4 to 3.5 bar) dynamic. Water is used for vacuum pump, heat exchanger and trap cooling. Refer to Water Quality Recommendations listed on page 6.
NOTE: Backflow prevention is by third parties; not supplied on unit.
Requirements for ProConnect Technical Support Services
Refer to Tech Data sheet SD983, PROCONNECT TECHNICAL SUPPORT SERVICES. (Available in U.S. and Canada only.)
CUSTOMER IS RESPONSIBLE FOR COMPLIANCE WITH
APPLICABLE LOCAL AND NATIONAL CODES AND
REGULATIONS.
The base language of this document is ENGLISH. Any translations must be made from the base language document.
Reference the Following Equipment Drawings for Installation Details
Equipment Drawing Number
Chamber Size Equipment Configuration
129394-138 26 x 37.5 x 36"
Single-Door, Hinged, Cabinet Enclosed26 x 37.5 x 48" 26 x 37.5 x 60"
129394-139 26 x 37.5 x 36"
Single-Door, Hinged, Recessed, One Wall26 x 37.5 x 48" 26 x 37.5 x 60"
129394-140 26 x 37.5 x 36, Double-Door, Hinged, Recessed, One Wall26 x 37.5 x 48" 26 x 37.5 x 60"
129394-141 26 x 37.5, x 36
Double-Door, Hinged, Recessed, Two Wall26 x 37.5 x 48" 26 x 37.5 x 60"
129394-142 26 x 37.5 x 36"
Single-Door, Horizontal-Sliding, Cabinet Enclosed26 x 37.5 x 48" 26 x 37.5 x 60"
129394-143 26 x 37.5 x 36"
Single-Door, Horizontal-Sliding, Recessed, One Wall26 x 37.5 x 48" 26 x 37.5 x 60"
129394-144 26 x 37.5 x 36"
Double-Door, Horizontal-Sliding, Recessed, One Wall26 x 37.5 x 48" 26 x 37.5 x 60"
129394-145 26 x 37.5 x 36"
Double-Door, Horizontal-Sliding, Recessed, Two Wall26 x 37.5 x 48" 26 x 37.5 x 60"
Operating Weight*: 26 x 37.5 x 36" 3800 lb (1720 kg) 26 x 37.5 x 48" 4200 lb (1900 kg) 26 x 37.5 x 60" 4700 lb (2125 kg)
* Operating weight includes a full load in the chamber.
UTILITIES CONSUMPTION
Consumption Values for AMSCO 400 Series Medium Steam Sterilizers
Water Consumption Steam Consumption
Peak Average Average Peak Average Average gal/min gal/cycle gal/hr lb/hr lb/cycle lb/hr
Equipment Drawing Number
Chamber Size Configuration in cycle in cycle out of cycle in cycle in cycle out of cycle
129394-138
26 x 37.5 x 36"
Single-Door, Hinged, Cabinet Enclosed 13
115 10 190 112 22
26 x 37.5 x 48" 120 12 255 148 28
26 x 37.5 x 60" 125 15 335 185 34
129394-139
26 x 37.5 x 36"
Single-Door, Hinged, Recessed, One Wall
115 10 190 112 22
26 x 37.5 x 48" 120 12 255 148 28
26 x 37.5 x 60" 125 15 335 185 34
129394-140
26 x 37.5 x 36"
Double-Door, Hinged, Recessed, One Wall
115 10 190 112 22
26 x 37.5 x 48" 120 12 255 148 28
26 x 37.5 x 60" 125 15 335 185 34
129394-141
26 x 37.5, x 36"
Double-Door, Hinged, Recessed, Two Wall
115 10 190 112 22
26 x 37.5 x 48" 120 12 255 148 28
26 x 37.5 x 60" 125 15 335 185 34
129394-142
26 x 37.5 x 36"
Single-Door, Horizontal-Sliding, Cabinet Enclosed
115 10 190 112 22
26 x 37.5 x 48" 120 12 255 148 28
26 x 37.5 x 60" 125 15 335 185 34
129394-143
26 x 37.5 x 36"
Single-Door, Horizontal-Sliding, Recessed, One Wall
115 10 190 112 22
26 x 37.5 x 48" 120 12 255 148 28
26 x 37.5 x 60" 125 15 335 185 34
129394-144
26 x 37.5 x 36"
Double-Door, Horizontal-Sliding, Recessed, One Wall
115 10 190 112 22
26 x 37.5 x 48" 120 12 255 148 28
26 x 37.5 x 60" 125 15 335 185 34
129394-145
26 x 37.5 x 36"
Double-Door, Horizontal-Sliding, Recessed, Two Wall
115 10 190 112 22
26 x 37.5 x 48" 120 12 255 148 28
26 x 37.5 x 60" 125 15 335 185 34
WATER QUALITY RECOMMENDATIONS
CONDITION NOMINAL CONDITIONS MAXIMUM CONDITIONS
Temperature 40°-60°F (4°-16°C) 70°F (21°C)
Total Hardness as CaCO3* 50-120 mg/L 171 mg/L
Total Dissolved Solids 100-200 mg/L 500 mg/L
Total Alkalinity as CaCO3 70-120 mg/L 180 mg/L pH 6.8-7.5 6.5-8.5
Total Silica 0.1 - 1.0 mg/L 2.5 mg/L
* 17.1 mg/L = 1.0 grain hardness
Healthcare
Capital Equipment
For further information, contact:
STERIS Corporation 5960 Heisley Road Mentor, OH 44060-1834 • USA 440-354-2600 • 800-548-4873 www.steris.com
This document is intended for the exclusive use of STERIS Customers, including architects or designers. Reproduction in whole or in part by any party other than a Customer is prohibited.
©2017, STERIS Corporation. All rights reserved. SD958 Rev B
| Application |
| Description |
| Standards |
| Features |
| Utilities Conservation Features |
| Processing Cycles |
| Testing Cycles |
| Control system |
| Safety Features |
| Construction |
| Mounting Arrangement |
| Accessory |
| Preventive Maintenance |
| NOTES |
| UTILITY REQUIREMENTS |
| 26 x 375 x 36 660 x 950 x 910 mm: Off |
| 26 x 375 x 48 660 x 950 x 1220 mm: Off |
| 26 x 375 x 60 660 x 950 x 1520 mm: Off |
| Shelf with Supports AX068803081: Off |
| Extra Side Gate AX068813081: Off |
| Extra Shelf Support AX068823081: Off |
| Seismic TieDown Kit 2: Off |
| SINGLE DOOR Hinged: Off |
| DOUBLE DOOR Hinged: Off |
| A LeftHand: Off |
| B RightHand: Off |
| Cabinet EnclosedFreestanding: Off |
| Recessed: Off |
| 208240 Vac 60 Hz 3phase 6A per phase: Off |
| 380415 Vac 5060 Hz 3phase 3 A per phase: Off |
| 480 Vac 60 Hz 3phase 3A per phase: Off |
| SINGLE DOOR Horizontal Sliding: Off |
| DOUBLE DOOR Horizontal Sliding: Off |
| A LeftHand_2: Off |
| Cabinet EnclosedFreestanding_2: Off |
| Recessed_2: Off |
| C LeftHandRight Hand 1: Off |
| Recessed through One Wall: Off |
| Recessed through Two Walls: Off |
| C LeftHandRightHand 1: Off |
| D LeftHandLeftHand 1: Off |
| E RightHandLeftHand 1: Off |
| F RightHandRightHand 1: Off |
| Recessed through One Wall_2: Off |
| Recessed through Two Walls_2: Off |
| Loading Car AX067013081: Off |
| Transfer Carriage Hinged Door Only: Off |
| Transfer Carriage Sliding Door Only: Off |
| Extra Shelf Support AX067823081: Off |
| Extra Side Gate AX067813081: Off |
| Chamber Rack and Two Shelves AX067111: Off |
| Seismic TieDown Kit 2_2: Off |
| Loading Car AX068013081: Off |
| Transfer Carriage Hinged Door Only_2: Off |
| Transfer Carriage Sliding Door Only_2: Off |
| Loading Car AX070013081: Off |
| Transfer Carriage Hinged Door Only_3: Off |
| Transfer Carriage Sliding Door Only_3: Off |
| Shelf with Supports AX070803081: Off |
| Extra Side Gate AX070813081: Off |
| Extra Shelf Support AX070823081: Off |
| Seismic TieDown Kit 2_3: Off |
| ProConnect Technical Support Services Remote: Off |
| Item: |
| undefined: |
| Locations: |
File details come from the government source that posted it. Updated .