AMSCO 400 STERILIZER PREVAC SINGLE DOOR RECESSED STEAM HEAT .pdf

PDF 975 KB Posted

Attached to
Z1DA--SPS Equipment Site Preparation (VA-25-00046418) | EDA: 6/5/25
Solicitation number
36C24525B0007
Issued by

About this file

This engineering technical file contains detailed installation specifications and technical drawings for a 26 x 37.5 AMSCO 400 Sterilizer with a prevac, single-hinge door, recessed steam heat configuration. The technical drawings provide comprehensive dimensional plans, plumbing connection details, and electrical connection specifications for domestic and international installations, with specific requirements including:

Plumbing Requirements: Cold water supply temperatures must be 70°F (21°C) maximum with vacuum efficiency reduced at water temperatures above 70°F. Steam requirements include 1" NPT with 100% vapor quality, dynamic pressure ranges of 20-50 PSIG, and specific flow rate specifications for different sterilizer models. Electrical requirements specify domestic and international power connections, including control power (120V, 50/60Hz), vacuum pump power, and step-down transformer power configurations with recommended circuit breaker specifications. The documentation includes precise measurements, installation guidelines, and technical notes for proper sterilizer setup, emphasizing requirements for safe and efficient equipment placement and connection.

View the file

Other files

Show all 22

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

26 x 37.5 AMSCO 400 STER.

PREVAC

SINGLE (HINGE) DOOR

RECESSED

STEAM HEAT

SIDE VIEW

PLAN VIEW

AutoCAD SHX Text 48 [1219]

AutoCAD SHX Text 'S1'

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

ALSO REFER TO GENERAL NOTES

AutoCAD SHX Text

APPLICABLE TO EQUIPMENT DRAWINGS

AutoCAD SHX Text

ALL DIMENSIONS ARE IN INCHES

AutoCAD SHX Text

DWG. NO. 62941-091

AutoCAD SHX Text

(MILLIMETERS)

AutoCAD SHX Text STERIS Corporation

AutoCAD SHX Text Mentor, OH

AutoCAD SHX Text

ITEM _____________________

AutoCAD SHX Text

EQUIPMENT DRAWING NO.

AutoCAD SHX Text

LOCATION(S) ______________

AutoCAD SHX Text

AutoCAD SHX Text

SHT. OF

AutoCAD SHX Text

AUTOCAD. DWG.

AutoCAD SHX Text

REV. ______

AutoCAD SHX Text

AC

AutoCAD SHX Text 129394-139

AutoCAD SHX Text

WALL OPENING

AutoCAD SHX Text

SERVICE CLEARANCE

AutoCAD SHX Text

SERVICE CLEARANCE

AutoCAD SHX Text

FINISHED FLOOR

AutoCAD SHX Text

OPERATING END

AutoCAD SHX Text

2. ALL DIMENSIONS IN INCHES AND (MM).

AutoCAD SHX Text

3. THIS SERVICE CLEARANCE MUST BE MAINTAINED TO ALLOW

AutoCAD SHX Text

ACCESS TO STERILIZER FOR SERVICEABILITY.

AutoCAD SHX Text

1. WALL THICKNESS: 4" TO 8" (102-203).

AutoCAD SHX Text

NOTES:

AutoCAD SHX Text

4. THE SHADED PLUMBING SERVICE CLEARANCE FLOOR SURFACE MUST BE SMOOTH WITH NO OBSTRUCTIONS (SUCH AS PIPING, WIRING, FLOOR SINKS, ETC.) THIS OBSTRUCTION FREE CLEARANCE MUST EXTEND TO HEIGHT OF AT LEAST 77" (1956) IN ORDER TO PROPERLY SERVICE

AutoCAD SHX Text

STERILIZER CHAMBER

AutoCAD SHX Text C

AutoCAD SHX Text L

AutoCAD SHX Text

5. RECOMMENDED WALL OPENING FOR NEW CONSTRUCTION IS 40".

AutoCAD SHX Text 26 X 37.5 X 36

AutoCAD SHX Text (660 X 953 X 914)

AutoCAD SHX Text

SIZE

AutoCAD SHX Text 81 (2057)

AutoCAD SHX Text 93 (2362)

AutoCAD SHX Text 'S'

AutoCAD SHX Text +1/4

AutoCAD SHX Text -0

AutoCAD SHX Text +6

AutoCAD SHX Text -0

AutoCAD SHX Text 26 X 37.5 X 48

AutoCAD SHX Text (660 X 953 X 1219)

AutoCAD SHX Text (660 X 953 X 1524)

AutoCAD SHX Text 26 X 37.5 X 60

AutoCAD SHX Text

AutoCAD SHX Text 69 (1753)

AutoCAD SHX Text 'S1'

AutoCAD SHX Text 61 1/4 (1556)

AutoCAD SHX Text 73 1/4 (1861)

AutoCAD SHX Text 49 1/4 (1251)

POWER

PREVAC

SINGLE (HINGE) DOOR

RECESSED

STEAM HEAT

SIDE VIEW

PLAN VIEW

GENERAL NOTES:

²

AutoCAD SHX Text

OFF

AutoCAD SHX Text

ON

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

ALSO REFER TO GENERAL NOTES

AutoCAD SHX Text

APPLICABLE TO EQUIPMENT DRAWINGS

AutoCAD SHX Text

ALL DIMENSIONS ARE IN INCHES

AutoCAD SHX Text

DWG. NO. 62941-091

AutoCAD SHX Text

(MILLIMETERS)

AutoCAD SHX Text STERIS Corporation

AutoCAD SHX Text Mentor, OH

AutoCAD SHX Text

ITEM _____________________

AutoCAD SHX Text

EQUIPMENT DRAWING NO.

AutoCAD SHX Text

LOCATION(S) ______________

AutoCAD SHX Text

AutoCAD SHX Text

SHT. OF

AutoCAD SHX Text

AUTOCAD. DWG.

AutoCAD SHX Text

REV. ______

AutoCAD SHX Text

AC

AutoCAD SHX Text 129394-139

AutoCAD SHX Text

FRONT CLEARANCE FOR DOOR SWING

AutoCAD SHX Text

CLEARANCE FOR R.H. DOOR SWING

AutoCAD SHX Text

SEE NOTE #2

AutoCAD SHX Text

FINISHED FLOOR

AutoCAD SHX Text

OPERATING END

AutoCAD SHX Text

DRAWING AND NOTE PAGES. SPECIFICATIONS AND DESCRIPTIONS ARE

AutoCAD SHX Text

THROUGH FAILURE TO OBSERVE THE SPECIFICATIONS ON EQUIPMENT

AutoCAD SHX Text

SUBJECT TO CHANGE WITHOUT NOTICE.

AutoCAD SHX Text

8. STERIS ASSUMES NO RESPONSIBILITY FOR CHANGES MADE NECESSARY

AutoCAD SHX Text

2. DISPLAY THIS SIDE FOR LH DOOR SWING AND OPPOSITE SIDE

AutoCAD SHX Text

FOR RH DOOR SWING.

AutoCAD SHX Text

26 X 37.5 X 36 - 3800 LBS (1720 KG)

AutoCAD SHX Text

26 X 37.5 X 48 - 4200 LBS (1901 KG)

AutoCAD SHX Text

26 X 37.5 X 60 - 4700 LBS (2127 KG)

AutoCAD SHX Text

7. LEVELLING FEET ARE PROVIDED FOR PROPER INSTALLATION.

AutoCAD SHX Text

9. IF LOADING CAR AND CARRIAGE ARE TO BE USED, FRONT

AutoCAD SHX Text

CLEARANCE SHOULD EQUAL TWICE THE LENGTH OF THE

AutoCAD SHX Text

STERILIZER.

AutoCAD SHX Text

1. UNIT WIDTH IS 37 1/2". DOOR OPENINGS IN FACILITY MUST

AutoCAD SHX Text

BE 38" WIDE OR LARGER TO ACCOMODATE STERILIZER.

AutoCAD SHX Text

6. MAX. OPERATING WEIGHT BASED ON CHAMBER FULLY LOADED WITH WATER FLASKS:

AutoCAD SHX Text

3. CURB WEIGHTS:

AutoCAD SHX Text 26 X 37.5 X 36 SD VESSEL = 2866 LBS (1310 KG) 26 X 37.5 X 48 SD VESSEL = 3053 LBS (1386 KG) 26 X 37.5 X 60 SD VESSEL = 3428 LBS (1556 KG)

AutoCAD SHX Text

CLEARANCE FOR L.H. DOOR SWING

AutoCAD SHX Text

STERILIZER CHAMBER

AutoCAD SHX Text C

AutoCAD SHX Text L

AutoCAD SHX Text

4. DIMENSIONS SHOWN FOR LH DOOR ONLY.FOR RH DOOR,DIMENSIONS WILL INTERCHANGE.

AutoCAD SHX Text

SIZE

AutoCAD SHX Text 26 X 37.5 X 48

AutoCAD SHX Text (660 X 953 X 1219)

AutoCAD SHX Text 26 X 37.5 X 36

AutoCAD SHX Text (660 X 953 X 914)

AutoCAD SHX Text (660 X 953 X 1524)

AutoCAD SHX Text

HEAT LOSS,BTU/HR (HINGED DOOR)

AutoCAD SHX Text 26 X 37.5 X 60

AutoCAD SHX Text inches

AutoCAD SHX Text (mm)

AutoCAD SHX Text

SINGLE DOOR

AutoCAD SHX Text

RECESSED

AutoCAD SHX Text

BACK OF

AutoCAD SHX Text

WALL

AutoCAD SHX Text

FRONT OF

AutoCAD SHX Text

WALL

AutoCAD SHX Text 2,675

AutoCAD SHX Text 2,675

AutoCAD SHX Text 2,675

AutoCAD SHX Text

VALUES ARE BASED ON PRE-VACUUM CONDITION

AutoCAD SHX Text 6,880

AutoCAD SHX Text 7,565

AutoCAD SHX Text 8,250

5. AIRFLOW GAP FROM CLEAN SIDE TO SERVICE AREA = 27.8 IN²..

PREVAC

SINGLE (HINGE) DOOR

RECESSED

STEAM HEAT

SIDE VIEW

PLAN VIEW

PLUMBING CONNECTIONS DOMESTIC

NOTE:

UNLOCK

LOCK

FRONT VIEW

AutoCAD SHX Text C

AutoCAD SHX Text

OFF

AutoCAD SHX Text

ON

AutoCAD SHX Text C

AutoCAD SHX Text B

AutoCAD SHX Text A

AutoCAD SHX Text A

AutoCAD SHX Text B

AutoCAD SHX Text 74 116

AutoCAD SHX Text 74 116

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

ALSO REFER TO GENERAL NOTES

AutoCAD SHX Text

APPLICABLE TO EQUIPMENT DRAWINGS

AutoCAD SHX Text

ALL DIMENSIONS ARE IN INCHES

AutoCAD SHX Text

DWG. NO. 62941-091

AutoCAD SHX Text

(MILLIMETERS)

AutoCAD SHX Text STERIS Corporation

AutoCAD SHX Text Mentor, OH

AutoCAD SHX Text

ITEM _____________________

AutoCAD SHX Text

EQUIPMENT DRAWING NO.

AutoCAD SHX Text

LOCATION(S) ______________

AutoCAD SHX Text

AutoCAD SHX Text

SHT. OF

AutoCAD SHX Text

AUTOCAD. DWG.

AutoCAD SHX Text

REV. ______

AutoCAD SHX Text

AC

AutoCAD SHX Text 129394-139

AutoCAD SHX Text

FINISHED FLOOR

AutoCAD SHX Text

OPERATING END

AutoCAD SHX Text

TO COLD WATER

AutoCAD SHX Text

A STERILIZER COLD WATER

AutoCAD SHX Text

B STERILIZER STEAM

AutoCAD SHX Text

C STERILIZER DRAIN

AutoCAD SHX Text

1. SEE SHEET 4 FOR PLUMBING INSTALLATION SPECS

AutoCAD SHX Text

AND PLUMBING REQUIREMENTS.

AutoCAD SHX Text

WARNING

AutoCAD SHX Text DOOR LOCKS ON LOSS OF POWER To open door in the event of a power failure. 1. Open lower access panel. 2. Pull down on the red handle located near lower left corner of door while moving door lock handle to the "OPEN" position.

AutoCAD SHX Text

STERILIZER CHAMBER

AutoCAD SHX Text C

L

SINGLE (HINGE) DOOR

RECESSED

STEAM HEAT

PLUMBING REQUIREMENTS

INSTALLATION SPECIFICATIONS:

PLACEMENT OF PIPING SHUTOFFS:

HARMS WAY

PLUMBING REQUIREMENTS-DOM. AND INT'L.

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

ALSO REFER TO GENERAL NOTES

AutoCAD SHX Text

APPLICABLE TO EQUIPMENT DRAWINGS

AutoCAD SHX Text

ALL DIMENSIONS ARE IN INCHES

AutoCAD SHX Text

DWG. NO. 62941-091

AutoCAD SHX Text

(MILLIMETERS)

AutoCAD SHX Text STERIS Corporation

AutoCAD SHX Text Mentor, OH

AutoCAD SHX Text

ITEM _____________________

AutoCAD SHX Text

EQUIPMENT DRAWING NO.

AutoCAD SHX Text

LOCATION(S) ______________

AutoCAD SHX Text

AutoCAD SHX Text

SHT. OF

AutoCAD SHX Text

AUTOCAD. DWG.

AutoCAD SHX Text

REV. ______

AutoCAD SHX Text

AC

AutoCAD SHX Text 129394-139

AutoCAD SHX Text

AS PART OF EQUIPMENT UNLESS UNDER WRITTEN AGREEMENT WITH STERIS.

AutoCAD SHX Text

GUIDES. REFERENCED COMPONENTS AND SERVICE CONNECTIONS ARE NOT FURNISHED

AutoCAD SHX Text

STATION FOR STEAM LINE AND MAIN WATER HEADERS. TO SUPPLY

AutoCAD SHX Text

INSTALLATION SPECIFICATION IS LISTED AS ENGINEERING AND INSTALLATION

AutoCAD SHX Text

EQUIPMENT TERMINATION SIZES ONLY. SIZE PIPING TO EQUIPMENT

AutoCAD SHX Text

DEPENDING ON LENGTH OF PIPE RUN FROM PRESSURE REGULATING

AutoCAD SHX Text

5. BLOW DOWN BUILDING STEAM & WATER SUPPLY LINES BEFORE FINAL

AutoCAD SHX Text

THE SPECIFIED SERVICE PRESSURE AND FLOW RATE AT EQUIPMENT

AutoCAD SHX Text

STERIS DWG. NO. 62941-091.

AutoCAD SHX Text

3. RECOMMEND PROVISION OF BLOW DOWN VALVE AT EACH STEAM AND

AutoCAD SHX Text

TERMINALS, INCLUDE EFFECT OF COINCIDENT DRAW OF MULTIPLE

AutoCAD SHX Text

4. FOR RECOMMENDED FEED WATER QUALITY FOR STERILIZERS SEE

AutoCAD SHX Text

CONNECTION TO EQUIPMENT.

AutoCAD SHX Text

WATER STRAINER TO ENABLE STRAINER CLEAN OUT.

AutoCAD SHX Text

UNIT INSTALLATIONS.

AutoCAD SHX Text THE INSTALLATION OF THE AMSCO 400 STERILIZER MUST MEET ALL FEDERAL, STATE AND LOCAL REGULATIONS.

AutoCAD SHX Text

BACKFLOW PREVENTER AND WHERE REQUIRED BY LOCAL CODES,

AutoCAD SHX Text 6.

AutoCAD SHX Text

THE STERILIZER IS NOT SUPPLIED WITH A VACUUM BREAKER OR

AutoCAD SHX Text

1. PIPE SIZES LISTED UNDER PLUMBING REQUIREMENTS INDICATE THEINDICATE THE

AutoCAD SHX Text

AND ELECTRICAL CONTROL PANEL.

AutoCAD SHX Text

STUB OUTS.THESE SHOULD BE LOCATED WITHIN ROOM, BELOW CEILING,

AutoCAD SHX Text

2. PROVIDE PIPING, SHUT-OFF VALVE, PIPE PLUGGED TEE, AND UNION

AutoCAD SHX Text

IN STEAM & WATER SUPPLY CONNECTION BETWEEN EQUIPMENT AND

AutoCAD SHX Text

EASILY ACCESSIBLE FOR EMERGENCY SHUT OFF SITUATIONS.PLUGGED

AutoCAD SHX Text

TEE CAN LATER BE USED FOR TEST PRESSURE GAUGE CONNECTION.

AutoCAD SHX Text

ARRANGE CONNECTION PIPING TO ALLOW ACCESS TO MACHINE COMPONENTS

AutoCAD SHX Text

BE LOCATED IN A SUITABLE LOCATION WITHIN LINE OF SIGHT AND

AutoCAD SHX Text

CLEAR OF ANY OBSTRUCTIONS THAT WOULD PUT THE SERVICE PERSON

AutoCAD SHX Text

7. PLACEMENT OF PIPING SHUTOFFS: WHEN INSTALLING; SHUTOFFS MUSTWHEN INSTALLING; SHUTOFFS MUST

AutoCAD SHX Text IN HARMS WAY IN ORDER TO SHUT IT OFF.IN HARMS WAY IN ORDER TO SHUT IT OFF. IN ORDER TO SHUT IT OFF.

AutoCAD SHX Text

A STERILIZER COLD WATER (SEE NOTE #4):

AutoCAD SHX Text

B STERILIZER STEAM:

AutoCAD SHX Text

C STERILIZER DRAIN TERMINAL: 2" ODT

AutoCAD SHX Text

PEAK- 26 X 37.5 X 60 - 335 LB/HR (152 KG/HR)

AutoCAD SHX Text

PEAK WATER CONSUMPTION).

AutoCAD SHX Text

(FLOOR DRAIN CAPACITY MUST HANDLE

AutoCAD SHX Text

AVERAGE- 26 X 37.5 X 60 - 185 LB/HR (84 KG/HR)

AutoCAD SHX Text

PEAK- 13 GPM (49.2 LPM)

AutoCAD SHX Text

IS REDUCED AT WATER TEMPERATURES ABOVE

AutoCAD SHX Text

70%%DF (21%%DC) MAX.. VACUUM EFFICIENCY

AutoCAD SHX Text

AVERAGE-26 X 37.5 X 36 - 115 GPH (7.6 LPM)

AutoCAD SHX Text

26 X 37.5 X 48 - 120 GPH (8.0 LPM)

AutoCAD SHX Text

26 X 37.5 X 60 - 125 GPH (8.3 LPM)

AutoCAD SHX Text

20-50 PSIG DYNAMIC (1.38-3.45 BAR)

AutoCAD SHX Text

PEAK- 26 X 37.5 X 48 - 255 LB/HR (116 KG/HR)

AutoCAD SHX Text

AVERAGE- 26 X 37.5 X 48 - 148 LB/HR (67 KG/HR)

AutoCAD SHX Text

PEAK- 26 X 37.5 X 36 - 190 LB/HR (86 KG/HR)

AutoCAD SHX Text

DYNAMIC PRESSURE (SEE SHEET 2 NOTE 5):

AutoCAD SHX Text

AVERAGE- 26 X 37.5 X 36 - 112 LB/HR (51 KG/HR)

AutoCAD SHX Text

CONSUMPTION:

AutoCAD SHX Text

1" NPT

AutoCAD SHX Text

70%%DF (21%%DC).

AutoCAD SHX Text

CONSUMPTION:

AutoCAD SHX Text

1" NPT

AutoCAD SHX Text

97 TO 100% VAPOR QUALITY.

AutoCAD SHX Text

NON-MDD UNITS - 50-80 PSIG (3.45-5.52 BAR) DYNAMIC

AutoCAD SHX Text

MDD UNITS- 50-55 PSIG (3.45-3.79 BAR) DYNAMIC

AutoCAD SHX Text

SUPPLY TEMPERATURE REQUIREMENT IS

AutoCAD SHX Text

INSTALLATION OF SUCH A DEVICE IN WATER LINE IS BY OTHERS.

FACILITY MUST PROVIDE REGULATED STEAM PRESSURE IN THE DYNAMIC RANGE SPECIFIED. FAILURE TO DO SO WILL RESULT IN IMPROPER EQUIPMENT OPERATION (SEE SHEET 4 NOTE B).

PREVAC

SINGLE (HINGE) DOOR

RECESSED

STEAM HEAT

SIDE VIEW

PLAN VIEW

ELECTRICAL CONNECTIONS-DOM. AND INT'L.

NOTE:

AutoCAD SHX Text

OFF

AutoCAD SHX Text

ON

AutoCAD SHX Text F

AutoCAD SHX Text F

AutoCAD SHX Text E

AutoCAD SHX Text D

AutoCAD SHX Text D

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

ALSO REFER TO GENERAL NOTES

AutoCAD SHX Text

APPLICABLE TO EQUIPMENT DRAWINGS

AutoCAD SHX Text

ALL DIMENSIONS ARE IN INCHES

AutoCAD SHX Text

DWG. NO. 62941-091

AutoCAD SHX Text

(MILLIMETERS)

AutoCAD SHX Text STERIS Corporation

AutoCAD SHX Text Mentor, OH

AutoCAD SHX Text

ITEM _____________________

AutoCAD SHX Text

EQUIPMENT DRAWING NO.

AutoCAD SHX Text

LOCATION(S) ______________

AutoCAD SHX Text

AutoCAD SHX Text

SHT. OF

AutoCAD SHX Text

AUTOCAD. DWG.

AutoCAD SHX Text

REV. ______

AutoCAD SHX Text

AC

AutoCAD SHX Text 129394-139

AutoCAD SHX Text

FINISHED FLOOR

AutoCAD SHX Text

OPERATING END

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

E STERILIZER VACUUM PUMP

AutoCAD SHX Text

F STERILIZER OPT. STEP DOWN TRANSFORMER POWER

AutoCAD SHX Text

SEE SHEET 6 FOR ELECTRICAL INSTALLATION SPECS

AutoCAD SHX Text

AND ELECTRICAL REQUIREMENTS.

AutoCAD SHX Text

D STERILIZER CONTROL POWER

AutoCAD SHX Text

AutoCAD SHX Text

= TO CONTROL BOX KNOCK OUT

AutoCAD SHX Text

= TO TRANSFORMER BOX KNOCK OUT

AutoCAD SHX Text

= TO POWER INPUT HOLE

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text F

AutoCAD SHX Text D

AutoCAD SHX Text E

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

SINGLE (HINGE) DOOR

RECESSED

STEAM HEAT

INSTALLATION SPECIFICATIONS:

PLACEMENT OF ELECT. DISCONNECTS:

HARMS WAY

ELECTRICAL REQUIREMENTS-DOM. AND INT'L.

AutoCAD SHX Text

AutoCAD SHX Text

AutoCAD SHX Text

ALSO REFER TO GENERAL NOTES

AutoCAD SHX Text

APPLICABLE TO EQUIPMENT DRAWINGS

AutoCAD SHX Text

ALL DIMENSIONS ARE IN INCHES

AutoCAD SHX Text

DWG. NO. 62941-091

AutoCAD SHX Text

(MILLIMETERS)

AutoCAD SHX Text STERIS Corporation

AutoCAD SHX Text Mentor, OH

AutoCAD SHX Text

ITEM _____________________

AutoCAD SHX Text

EQUIPMENT DRAWING NO.

AutoCAD SHX Text

LOCATION(S) ______________

AutoCAD SHX Text

AutoCAD SHX Text

SHT. OF

AutoCAD SHX Text

AUTOCAD. DWG.

AutoCAD SHX Text

REV. ______

AutoCAD SHX Text

AC

AutoCAD SHX Text 129394-139

AutoCAD SHX Text

2. PROVIDE GROUNDED METAL CONDUIT AND WIRING BETWEEN EQUIPMENT

AutoCAD SHX Text

1. PROVIDE GROUPED OR GANGED CIRCUIT PROTECTION AND DISCONNECT

AutoCAD SHX Text

FOR CONTROL POWER AND VACUUM PUMP POWER AS REQUIRED

AutoCAD SHX Text

BY NEC AND LOCAL CODES.INDIVIDUAL POWER SHUTOFFS RECOMMENDED

AutoCAD SHX Text

TERMINALS AND STUB OUTS OR DISCONNECTS. 14 AWG. MINIMUM SIZE

AutoCAD SHX Text

NEAR EACH MACHINE FOR SERVICING.

AutoCAD SHX Text D

AutoCAD SHX Text E

AutoCAD SHX Text

RECOMMENDED.

AutoCAD SHX Text

3. FOR GENERAL INSTALLATION INFORMATION SEE STERIS DRAWING

AutoCAD SHX Text

NO. 62941-091. (THIS DWG. SHOULD ALWAYS ACCOMPANY THE

AutoCAD SHX Text

EQUIPMENT DWGS.) IF DWG IS NOT ATTACHED , CONTACT STERIS

AutoCAD SHX Text

SERVICE ENGINEERING AT 1-800-333-8848 TO OBTAIN A COPY.

AutoCAD SHX Text

4. PLACEMENT OF ELECT. DISCONNECTS: WHEN INSTALLING;DISCONNECTS WHEN INSTALLING;DISCONNECTS MUST BE LOCATED IN A SUITABLE LOCATION WITHIN LINE OF SIGHT AND CLEAR OF ANY OBSTRUCTIONS THAT WOULD PUT THE SERVICE PERSON IN HARMS WAY IN ORDER TO SHUT IT OFF. ALSO, THE IN ORDER TO SHUT IT OFF. ALSO, THE LOCATION OF THE DISCONNECTS SHOULD ALLOW THE SERVICE PERSON TO SHUTOFF POWER FROM THE SIDE TO PREVENT POSSIBLE ARC FLASH.

AutoCAD SHX Text

E STERILIZER PUMP POWER:

AutoCAD SHX Text

DOMESTIC: 208/240V, 50/60HZ, 6A, (3) PHASE.MINIMUM

AutoCAD SHX Text

8A CIRCUIT BREAKER RECOMMENDED.

AutoCAD SHX Text

OR 480V, 50/60HZ, 3A, (3) PHASE.MINIMUM

AutoCAD SHX Text

4A CIRCUIT BREAKER RECOMMENDED.

AutoCAD SHX Text

INTERNATIONAL: 400V, 50HZ, 4A, (3) PHASE. MINIMUM

AutoCAD SHX Text

5A CIRCUIT BREAKER RECOMMENDED.

AutoCAD SHX Text

F STERILIZER OPTIONAL STEP-DOWN TRANSFORMER POWER:

AutoCAD SHX Text

INTERNATIONAL: 230V, 50/60HZ, 1.5A, SINGLE PHASE

AutoCAD SHX Text

ELECTRIC SERVICE.MINIMUM 2A CIRCUIT

AutoCAD SHX Text

BREAKER RECOMMENDED.

AutoCAD SHX Text

D STERILIZER CONTROL POWER:

AutoCAD SHX Text

DOMESTIC: 120V, 50/60HZ, 2A, SINGLE PHASE ELECTRIC

AutoCAD SHX Text

SERVICE. MINIMUM 3A CIRCUIT BREAKER

AutoCAD SHX Text

RECOMMENDED.

AutoCAD SHX Text D

E

File details come from the government source that posted it. Updated .