6 4Radixact Full Set.pdf
PDF 4 MB Posted
- Attached to
- Replace Linear Accelerator at Greater Los Angeles VAHCS Federal contract opportunity
- Solicitation number
- 36C25821B0004
About this file
This solicitation is for a construction project to replace a linear accelerator at the Greater Los Angeles VA Healthcare System campus in West Los Angeles, California. The selected contractor will demolish the existing linear accelerator, prepare the site, and assist with installing the new Accuray Radixact linear accelerator. Additional requirements include renovating rooms for the linear accelerator, control room, mechanical room, and data server room. The contractor must provide structural supports, foundations, bracing, electrical power, controls, and communications systems. Ceiling, flooring, walls, HVAC, and other infrastructure must be replaced or added. The expected construction period is included in the solicitation. The Department of Veterans Affairs and Veterans Health Administration are the federal agencies procuring this project.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Sign in Sheet for 11 Jan Site Walk Linear Accelerator.pdf | ||
| 36C25821B0004 0002.pdf | ||
| 36C25821B0004 0001 RFI Amendment.pdf | ||
| 36C25821B0004.pdf | ||
| Wage Rate Determination CA20200022.docx | DOCX document | |
| Atch 1 Self Performance Calculations.pdf | ||
| Contractor Background Investigation Request Form (May 2019).pdf | ||
| 6 8Infection Control Permit_revised.pdf | ||
| of0306.pdf | ||
| VA0710.pdf | ||
| 6 2 Drawings 2 of 2.pdf | ||
| 6 3Specification CSI 121.pdf | ||
| 6 1 Drawings1 of 2.pdf | ||
| Self Certification of Continuous Service.pdf | ||
| Fingerprint Record Prep Sheet BLANK.PDF |
Show all 15
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
ACCURAY INCORPORATED
1310 Chesapeake Terrace
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E N
T
S
P D
R
E V
C
4.6.2020B
LOS ANGELES, CA
RADIXACT ™ TREATMENT DELIVERY SYSTEM
WEST LA VA
GENERAL
G-001 COVER PAGE
G-002 GENERAL NOTES
ARCHITECTURAL
A-101 FLOOR PLAN
A-102 RIGGING PLAN
A-201 INTERIOR ELEVATIONS
STRUCTURAL
S-101 FLOOR PLAN
S-301 INTERIOR SECTIONS
S-501 PLAN AND SECTION DETAILS
MECHANICAL
M-101 FLOOR PLAN
M-501 PLAN AND SECTION DETAILS
ELECTRICAL
E-101 FLOOR PLAN
E-111 REFLECTED CEILING PLAN
E-301 INTERIOR SECTIONS
E-501 PLAN AND SECTION DETAILS
E-502 ELECTRICAL DETAILS
E-503 ELECTRICAL DETAILS
E-901 INTERIOR PERSPECTIVES
NETWORK
N-101 FLOOR PLAN
EQUIPMENT
Q-501 PLAN AND SECTION DETAILS
Q-901 INTERIOR PERSPECTIVE
ACCURAY INCORPORATED
FOR REFERENCE ONLY
G-001
COVER SHEET
CONTROL
ROOM
MP
1'-9"
4'-10"
3'-4
4'-9"
TREATMENT
ROOM
7'-11"
31'-3
10'-2
18'-9"
NOTE 3
NOTE 2
NOTE
1 TYP.
3'-0"
3'-0"
1'-7
1'-7
19'-9"
15'-2"
2'-0"
3'-7"
DATA
SERVER ROOM
MECHANICAL
ROOM
TREATMENT
PLANNING
ROOM
(PRECISION)
MECHANICAL ROOM
6a/6b
ACCURAY INCORPORATED
1310 Chesapeake Terrace
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E
S
P D
R
A-101
ARCHITECTURAL
FLOOR PLAN
GENERAL NOTES
SHEET NOTES
LEGEND
A. THESE DRAWINGS ARE FOR
REFERENCE ONLY. THEY
ARE NOT TO REPLACE
CONSTRUCTION
DOCUMENTS AND
SHIELDING STUDIES BY
LICENSED PROFESSIONALS.
B. FINAL RADIATION SHIELDING
SHALL BE COMPLETED BY
THE PHYSICIST OF RECORD.
C. UNLESS OTHERWISE NOTED,
ALL MATERIALS SUPPLIED
AND INSTALLED ARE THE
RESPONSIBILITY OF THE
CUSTOMER.
D. REFER TO SITE PLANNING
GUIDE (SPG) FOR
ADDITIONAL
SPECIFICATIONS.
E. ALL STRUCTURAL
CALCULATIONS TO BE
COMPLETED BY
STRUCTURAL ENGINEER OF
RECORD.
F. ALL FINAL DIMENSIONS TO
BE FIELD VERIFIED.
1. MINIMUM 2 WALL MOUNTED
CCTV (FACILITY PROVIDED).
EXACT LOCATION TO BE
DETERMINED BY CUSTOMER.
CCTV AT FOOT OF PATIENT BED
TO BE MOTORIZED.
2. FINISHED WALLS ARE FOR
REFERENCE ONLY. REFER TO
CONTRACT DOCUMENTS FOR
DIMENSIONS/DETAILS.
3. CLEARANCE AREA FOR
SERVICE.
ISOCENTER
MP MAIN POWER DISCONNECT
[ ] RECOMMENDED
OFCI OWNER FURNISHED/
CONTRACTOR INSTALLED
# EQUIPMENT NAME IN LBS
1 RADIXACT™ TREAMENT DELIVERY SYSTEM 77.5 x 110 x 99 13000
2 RADIXACT™ SYSTEM COUCH 108.6 x 25.6 x 44 1100
3 POWER DISTRIBUTION UNIT 31 x 22 x 56 980
4 DORADO LASER SYSTEM 7 x 31.25 x 7.75 37
5 APOLLO LASER SYSTEM 4 x 4.25 x 8.5 2.6
6a POWER CONDITIONER (OFCI) 31.6 x18.9 x 73.7 640
6b FREQUENCY CONVERTER (OFCI) 51.5 x26.75x71.25 1502
7 AIR COMPRESSOR N/A N/A
8 iDMS DATA MANAGEMENT SYSTEM 38 x 24 x 59 900
10 OPERATOR STATION CONTROL 8.5 x 4.5 x 3 2.6
12 CONTROL WORKSTATION N/A N/A
13 TREATMENT PLANNING SYSTEM N/A N/A
14 INTERCOM SPEAKER SYSTEM 7 x 6.3 x 9.4 5
15 PRINTER N/A N/A
16 SYNCHRONY 28 x 7.5 x 3 10
OVERALL FLOOR PLAN
SCALE: 1/8" = 1'-0"
A1
DIMENSION FORMAT: FEET & INCHES
THE FOLLOWING ROOMS ARE NOT SITE SPECIFIC
AND ARE FOR REFERENCE ONLY:
- TREATMENT PLANNING (DOSIMETRY)
- MECHANICAL ROOM
NOTE 1,2,3,4 TYP.
DATA
SERVER ROOM
5'-11"
MIN.
5'-11"
MIN.
82"
MINIMUM
HEIGHT
FOR DOOR
OPENING.
4'-0" MIN.
6'-10" MIN.
7'- 9
8'-4"
9'- 5
6'- 9
7'- 9
9'- 5
3' - 5
ACCURAY INCORPORATED
1310 Chesapeake Terrace
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E
S
P D
R
A-102
ARCHITECTURAL
RIGGING PLAN
1. RIGGING PLAN IS FOR
REFERENCE ONLY. ACTUAL
ROUTE TO BE DETERMINED BY
OWNER.
2. GANTRY IS 77.5"(L) x 110"(W) x
99(H) AND WEIGHTS 13000 LBS.
3. FACILITY TO PROVIDE CLEAR
ACCESS PATH MIN. 5'-11" WIDE x
82" HIGH TO OUTSIDE DELIVERY
LOCATION.
4. ALL FLOOR SURFACES ALONG
THE RIGGING PATH MUST BE
FULLY SUPPORTED AND
CAPABLE OF WITHSTANDING
13,000 LBS.
OVERALL RIGGING PLAN
SCALE: 3/16" = 1'-0"
A1
RIGGING THROUGH MAZE- WHEELS
NOT TO SCALE
D1
RIGGING THROUGH DOOR- WHEELS
NOT TO SCALE
D2
GANTRY ON WHEELS
NOT TO SCALE
D3
GANTRY ON WHEELS
NOT TO SCALE
D5
GENERAL NOTES
SHEET NOTES
LEGEND
A. THESE DRAWINGS ARE FOR
REFERENCE ONLY. THEY
ARE NOT TO REPLACE
CONSTRUCTION
DOCUMENTS AND
SHIELDING STUDIES BY
LICENSED PROFESSIONALS.
B. FINAL RADIATION SHIELDING
SHALL BE COMPLETED BY
THE PHYSICIST OF RECORD.
C. UNLESS OTHERWISE NOTED,
ALL MATERIALS SUPPLIED
AND INSTALLED ARE THE
RESPONSIBILITY OF THE
CUSTOMER.
D. REFER TO SITE PLANNING
GUIDE (SPG) FOR
ADDITIONAL
SPECIFICATIONS.
E. ALL STRUCTURAL
CALCULATIONS TO BE
COMPLETED BY
STRUCTURAL ENGINEER OF
RECORD.
F. ALL FINAL DIMENSIONS TO
BE FIELD VERIFIED.
ISOCENTER
DIMENSION FORMAT: FEET & INCHES
ISO CENTER
7'-0"
4'-0"
I S
O
C E
N T
E R
[4'-0"]
4'-9"
THERMOSTAT (NOTE 1)
CCTV (NOTE 1)
[4'-0"]
I S
O
C E
N T
E R
3'-8
" A.F.F
1'-0"
1'-6"
INTERCOM
NOTE 2
CCTV (NOTE 1,4)
APOLLO LASER POWER SWITCH
EMO (NOTE 1)
2'-3
6'-7
" A.F.F
C
L
BEAM
ON
3'-8
" A.F.F
4'-0"
HEIGHT MAY VARY
I S
O
C E
N T
E R
1'-4"
3'-7
RAD WARNING LIGHT
(NOTE 1,3)
EMO (NOTE 1)
THERMOSTAT (NOTE 1)
3'-8
" A.F.F
4'-0"
HEIGHT MAY VARY
1'-4"
3'-7
2'-3
6'-7
" A.F.F
C
L
BEAM
ON
I S
O
C E
N T
E R
EMO (NOTE 1)
RAD WARNING LIGHT
(NOTE 1,3)
CCTV (NOTE 1)
ACCURAY INCORPORATED
1310 Chesapeake Terrace
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E
S
P D
R
A-201
ARCHITECTURAL
ELEVATION
PLAN ELEVATION NORTH
SCALE: 1/4" = 1'-0"
A1
PLAN ELEVATION EAST
SCALE: 1/4" = 1'-0"
A4
PLAN ELEVATION SOUTH
SCALE: NOT TO SCALE
C1
PLAN ELEVATION WEST
SCALE: NOT TO SCALE
C4
NOT SITE SPECIFIC NOT SITE SPECIFIC
NOT SITE SPECIFICNOT SITE SPECIFIC
1. OWNER FURNISHED,
CONTRACTOR INSTALLED.
EXACT LOCATION TO BE
DETERMINED BY CUSTOMER.
2. AIR SHUT OFF VALVE AND
QUICK DISCONNECT, LOCATION
MAY VARY. SEE SITE PLANNING
GUIDE (SPG) FOR ALTERNATE
LOCATIONS. TO CONNECT BACK
TO AIR COMPRESSOR WITH
FLOW RATE OF 15 scfm AT 90 psig
3. REFER TO LOCAL
REGULATIONS. COORDINATE
ELECTRICAL AS NECESSARY.
4. CCTV AT THE BACK OF THE
GANTRY AND NEIS NEEDS TO
BE INSTALLED IN THE AXIS OF
ISOCENTER AND RIGHT ABOVE
THE LASER PLATE.
AFF ABOVE FINISHED FLOOR
4 DORADO (RED) LASER
5 APOLLO (GREEN) LASER
16 SYNCHRONY
[ ] RECOMMENDED
GENERAL NOTES
SHEET NOTES
LEGEND
A. THESE DRAWINGS ARE FOR
REFERENCE ONLY. THEY
ARE NOT TO REPLACE
CONSTRUCTION
DOCUMENTS AND
SHIELDING STUDIES BY
LICENSED PROFESSIONALS.
B. FINAL RADIATION SHIELDING
SHALL BE COMPLETED BY
THE PHYSICIST OF RECORD.
C. UNLESS OTHERWISE NOTED,
ALL MATERIALS SUPPLIED
AND INSTALLED ARE THE
RESPONSIBILITY OF THE
CUSTOMER.
D. REFER TO SITE PLANNING
GUIDE (SPG) FOR
ADDITIONAL
SPECIFICATIONS.
E. ALL STRUCTURAL
CALCULATIONS TO BE
COMPLETED BY
STRUCTURAL ENGINEER OF
RECORD.
F. ALL FINAL DIMENSIONS TO
BE FIELD VERIFIED.
A1
S301
C1
S301
PB1
PB2
D1
Q501
A1
Q501
C
L
M-501
A4
TYP.
G
C
D
NOTE 2
NOTE 1,3,7 TYP.
8'-6
7'-11"
NOTE 5
NOTE 6
NOTE 3
6'-9
5'-6
4"
2'-5"
8"
NOTE 4
PB5 PB5
TRENCH
ACCURAY INCORPORATED
1310 Chesapeake Terrace
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E
S
P D
R
S-101
STRUCTURAL
FLOOR PLAN
1. THIS FLOOR PLAN IS CUT AT
THE SUB-BASE LEVEL, WITHIN
THE CONCRETE.
2. MOISTURE SENSOR WITH
ALARM. A FLOOR DRAIN THAT
CAN HANDLE 14 L/min. IS A
POSSIBLE ALTERNATIVE.
3. SEISMIC BRACKET MAY BE
REQUIRED, REFER TO CODE.
4. WHERE TRENCHES ARE NOT
POSSIBLE, CONDUIT IS AN
ACCEPTABLE ALTERNATIVE.
5. SUPPLEMENTAL AIR DUCT (SEE
M-501)
6. COMPRESSED AIR LINE
(SEE M-501)
7. REFER TO A1 ON S-501
6" AREA AROUND
EPOXY SET ANCHOR,
CLEAR OF ALL REBAR
TO DEPTH OF 10".
PB1 2'-0" x 1'-6" x 1'-0" PDU
PULL BOX
PB2 6" x 6" x 6" PATIENT
TABLE PULL BOX
PB5 1'-0" x 8" x 8" GANTRY
BULL BOX
SIGNAL (3"Ø) - FROM
PATIENT TABLE TO
BACK OF GANTRY.
(20'-0" MAX).
POWER (3"Ø) - FROM
PATIENT TABLE TO
BACK OF GANTRY.
(55'-0" MAX).
SIGNAL (1"Ø) - INTERCOM
TO BACK OF GANTRY (15'
MAX)
4 (4"Ø) FROM PB1 TO PB5
TREATMENT ROOM SUB-BASE PLAN
SCALE: 1/4" = 1'-0"
A1
C
D
SUB-BASE BEFORE POURING CONCRETE
NOT TO SCALE
A5
SUB-BASE AFTER POURING CONCRETE
NOT TO SCALE
D3
SHEET NOTES
LEGEND
A. THESE DRAWINGS ARE FOR
REFERENCE ONLY. THEY
ARE NOT TO REPLACE
CONSTRUCTION
DOCUMENTS AND
SHIELDING STUDIES BY
LICENSED PROFESSIONALS.
B. FINAL RADIATION SHIELDING
SHALL BE COMPLETED BY
THE PHYSICIST OF RECORD.
C. UNLESS OTHERWISE NOTED,
ALL MATERIALS SUPPLIED
AND INSTALLED ARE THE
RESPONSIBILITY OF THE
CUSTOMER.
D. REFER TO SITE PLANNING
GUIDE (SPG) FOR
ADDITIONAL
SPECIFICATIONS.
E. ALL STRUCTURAL
CALCULATIONS TO BE
COMPLETED BY
STRUCTURAL ENGINEER OF
RECORD.
F. ALL FINAL DIMENSIONS TO
BE FIELD VERIFIED.
ISOCENTER
DIMENSION FORMAT: FEET & INCHES
G
GENERAL NOTES
B3
Q501
D3
Q501
A3
Q501
A4
S501
D1
Q501
10"
8'-10
11'-0"
3'-8
3'-7
2'-3
5'-0"
19'-9"
NOTE 2
NOTE 1
12'-8"
7'-8"
NOTE 4
A5
Q501
A5
Q501
C4
S501
CONCRETE DECK
HEIGHT IS NOT SITE
SPECIFIC AND MUST
BE FIELD VERIFIED
A1
Q501
TYP.
TYP.
8'-10
6'-7
3'-8
3'-0"
3'-0"
10"
NOTE 3
NOTE 1
NOTE 2
ACCURAY INCORPORATED
1310 Chesapeake Terrace
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E
S
P D
R
S-301
INTERIOR
SECTIONS
1. SLOPE PIT TOWARD MOISTURE
SENSOR. A FLOOR DRAIN THAT
CAN HANDLE 14 L/min. IS A
POSSIBLE ALTERNATIVE.
2. WORKING SYSTEM CLEARANCE
3. THE FLOOR MUST BE LEVEL
ENOUGH TO MEET A REQUIRED
FLATNESS TOLERANCE OF ±
1/4" OVER 10'-10"
4. SYNCHRONY MOUNTING
LOCATION. KEEP 2' CLEAR FOR
INSTALLATION AND SERVICING
1 GANTRY AND EQUIPMENT
ENCLOSURES
2 TREATMENT COUCH
4 DORADO (RED) LASER
5 APOLLO (GREEN)
LASER
16 SYNCHRONY
PLAN WEST-EAST TREATMENT ROOM SECTION
SCALE: 1/4" = 1'-0" NOT SITE SPECIFIC
A1
PLAN NORTH/SOUTH TREATMENT ROOM SECTION
SCALE: 1/4" = 1'-0" NOT SITES SPECIFIC
C1
GENERAL NOTES
SHEET NOTES
LEGEND
A. THESE DRAWINGS ARE FOR
REFERENCE ONLY. THEY
ARE NOT TO REPLACE
CONSTRUCTION
DOCUMENTS AND
SHIELDING STUDIES BY
LICENSED PROFESSIONALS.
B. FINAL RADIATION SHIELDING
SHALL BE COMPLETED BY
THE PHYSICIST OF RECORD.
C. UNLESS OTHERWISE NOTED,
ALL MATERIALS SUPPLIED
AND INSTALLED ARE THE
RESPONSIBILITY OF THE
CUSTOMER.
D. REFER TO SITE PLANNING
GUIDE (SPG) FOR
ADDITIONAL
SPECIFICATIONS.
E. ALL STRUCTURAL
CALCULATIONS TO BE
COMPLETED BY
STRUCTURAL ENGINEER OF
RECORD.
F. ALL FINAL DIMENSIONS TO
BE FIELD VERIFIED.
ISOCENTER
DIMENSION FORMAT: FEET & INCHES
C5
FLOOR FINISH TOLERANCES
NOT TO SCALE
D5
NOT TO SCALE
DEFINITION OF FF25/FL25: COMES
FROM A FLOOR FINISH
TOLERANCES SYSTEM ON
F-NUMBER CODES WITHIN
ASTM-E1155, AC1 #117, & CSA
#23.1. THESE CODES GIVE
TOLERANCE REQUIREMENTS
USING ABBREVIATION F -
NUMBERS.
DEFINITION OF FF25: "FF"
REPRESENTS FLOOR FLATNESS &
THE NUMBER REPRESENTS 0.25"
[0.6] MAX. HEIGHT WITHIN A 10'
[304.8] PLANE. DETAIL "B" ON THIS
SHEET SHOWS AN EXAMPLE OF
FF25.
DEFINITION OF FL25: "FL"
REPRESENTS FLOOR LEVELNESS
& THE NUMBER REPRESENTS 0.25"
[0.6] MAX. HEIGHT ON A CURVE
DISTANCE OF 24" [610] WIDE. THE
MAX. HEIGHT SHOULD BE TAKEN
FROM THE POINT LOCATED AT
MID-CURVE OF A 12" [30.5]
DISTANCE. DETAIL "A" ON THIS
SHEET SHOWS AN EXAMPLE OF
FL25.
HATCH AREA REPRESENTS FLOOR
GRADE FLATNESS AND LEVELNESS
WHICH MUST HAVE FLATNESS (FF)
OF 25 FOR TOTAL AREA &
LEVELNESS (FL) OF 25 PER ASTM-E
1155.
*FLOOR FINISH TOLERANCES
SYSTEM ON F-NUMBER CODES
WITHIN ASTM-E1155 (US
STANDARD TEST METHOD FOR
DETERMINING (FF) FLOOR
FLATNESS AND FL FLOOR
LEVELNESS NUMBERS)
10'-0"
1/8"
1/8"
EQUIVALENT TO A SINGLE ± 1/8"
ALLOWABLE FLATNESS TOLERANCE
WHEN MEASURED BETWEEN ANY
TWO PERIPHERAL PTS. 10'-0" WITHIN
SHADED AREA.
* REFERENCE FF25:
F-NUMBER
SPECIFICATION
SYSTEM THAT
CODE FLOORING
FLATNESS
TOLERANCE.
0.000 MIDPOINT
PLANE
EXISTING
FLOORING
EQUIVALENT TO A SINGLE ± 1/4"
ALLOWABLE LEVELNESS TOLERANCE
WHEN MEASURED BETWEEN TWO POINT
OF A 24" CURVE DISTANCE. THE
TOLERANCE CANNOT BE GREATER THEN
1/4" WITHIN THE MID DISTANCE (12")
* REFERENCE FL25:
F-NUMBER SPECIFICATION SYSTEM
THAT CODE FLOORING LEVELNESS
TOLERANCE.
STRAIGHT EDGE
± 1/32"
ACCURACY
24"
1/4" MAX.
12"
0.000 MIDPOINT PLANE
EXISTING
FLOORING
FLOOR FLATNESS
FLOOR LEVELNESS NOTES
1'-2"
MIN.
1'-1
3'-1
2'-3"
1'-10"
SUPPLEMENTAL AIR DUCT
COMPRESSED AIR PIPE
NOTE 1
ELECTRICAL CONDUITS
1'-0" MIN1'-0" MIN
5'-7
3'-0"
" 8
1'-7
2'-2
3'-0
3'-8
C
L
ISO CENTER
2'-2
6'-2
1'-2" MIN.3'-1
3'-8
1'-1
D C
2'-5"
6'-11
NOTE 3
NOTE 4
2'-8
PB2
1'-9
1'-0
2'-0"
1'-0
1'-0"
2'-8"
G
2'-4"
PB5PB5
2'-3"
1'-1 3/8"
10" MIN @
HIGH PT.
PB2
1'-10" MIN.
6'-11 1/2"
1'-0" MIN.
I S
O C
E N
T E
R
10" MIN DEPTH
NOTE 1
1'-1 3/8"
1'-0 5/8"
3'-1
3'-1
NOTE 2
C
L
I S
O
C E
N T
E R
10" MIN @
HIGH PT.
10" MIN DEPTH
1'-2"
MIN.
NOTE 1
NOTE E
ACCURAY INCORPORATED
1310 Chesapeake Terrace
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E
S
P D
R
S-501
STRUCTURAL
DETAILS
PLAN ELEVATION NORTH
SCALE: NOT TO SCALE
A1
PLAN ELEVATION EAST
SCALE: NOT TO SCALE
A4
PLAN ELEVATION SOUTH
SCALE: NOT TO SCALE
D1
SEISMIC BRACKETS (IF APPLICABLE)
SCALE: NOT TO SCALE
D4
1. SLOPE PIT TOWARDS
MOISTURE SENSOR.
2. RADIUS FOR ROTATING
EQUIPMENT CLEARANCE.
3. GANTRY SUB-BASE: THE
GANTRY SUB-BASE MUST
SUPPORT THE GANTRY'S
WEIGHT AT 13,000 lbs AND
LOCAL REQUIREMENTS.
4. PATIENT TABLE SUB-BASE: THE
PATIENT TABLE SUB-BASE
MUST SUPPORT THE PATIENT
TABLE WEIGHT AT 1100 lbs AND
MEET LOCAL REQUIREMENTS.
6" AREA AROUND ANCHOR,
CLEAR OF ALL REBAR TO
DEPTH OF 10". 4 ANCHORS
REQUIRED FOR COUCH
AND 6 ANCHORS FOR
GANTRY.
SIGNAL (3"Ø) - FROM
PATIENT TABLE TO
BACK OF GANTRY.(20'MAX).
POWER (3"Ø) - FROM
PATIENT TABLE TO BACK
OF GANTRY.(55' MAX).
SIGNAL (1"Ø) - INTERCOM
TO BACK OF GANTRY (15'
MAX).
PATIENT TABLE
PULL BOX (FLOOR)
6" x 6" x 6" WITH COVER
PLATE FLUSH WITH
FINISHED FLOOR
TREATMENT VAULT
PULL BOX (FLOOR)
1'-0" x 8" x 8"
C
D
PB2
SEISMIC BRACKETS (IF APPLICABLE)
SCALE: NOT TO SCALE
D5
PLAN ELEVATION EAST
SCALE: NOT TO SCALE
C4
PB5
GENERAL NOTES
SHEET NOTES
LEGEND
A. THESE DRAWINGS ARE FOR
REFERENCE ONLY. THEY
ARE NOT TO REPLACE
CONSTRUCTION
DOCUMENTS AND
SHIELDING STUDIES BY
LICENSED PROFESSIONALS.
B. FINAL RADIATION SHIELDING
SHALL BE COMPLETED BY
THE PHYSICIST OF RECORD.
C. UNLESS OTHERWISE NOTED,
ALL MATERIALS SUPPLIED
AND INSTALLED ARE THE
RESPONSIBILITY OF THE
CUSTOMER.
D. REFER TO SITE PLANNING
GUIDE (SPG) FOR
ADDITIONAL
SPECIFICATIONS.
E. ALL STRUCTURAL
CALCULATIONS TO BE
COMPLETED BY
STRUCTURAL ENGINEER OF
RECORD.
F. ALL FINAL DIMENSIONS TO
BE FIELD VERIFIED.
ISOCENTER
DIMENSION FORMAT: FEET & INCHES
L5X3X
"X18" STEEL ANGLE (LEFT &
RIGHT SIDES) PROVIDED BY
ACCURAY WHEN REQUIRED BY
LOCAL REGULATORY AGENCY.
DSU
(TOP)
(FACE OF ANGLE)
(RIGHT SIDE)
L8X4X
"X36" STEEL ANGLE (LEFT &
RIGHT SIDES) PROVIDED BY
ACCURAY WHEN REQUIRED BY
LOCAL REGULATORY AGENCY.
iDMS
(TOP)
(FACE OF ANGLE)
(RIGHT SIDE)
2'-2 3/4"
6'-2 1/2"
M
I
N
GANTRY SUB-BASE
PATIENT
TABLE
SUB-BASE
GANTRY PIT
G
DIMENSION FORMAT: L x W x H
2'-5"
NOTE 7
D4
M501
A
M
NOTE 4
NOTE 2
NOTE 1
4'-0"
NOTE 5
4'-0"
VAULT
SUPPLY
NOTE 9
6'-0" - 8'-0"
6'-0" - 8'-0"
(3) GANTRY SUPPLY (NOTE 6)
NOTE 3
SUPPLY
RETURN
ACCURAY INCORPORATED
1310 Chesapeake Terrace
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E
S
P D
R
M-101
MECHANICAL
FLOOR PLAN
1. FACILITY SUPPLEMENTAL AIR
DUCT: ROUTE UNDER SLAB.
2. COMPRESSED AIR PIPE:
COPPER PIPE UNDER SLAB TO
AIR COMPRESSOR. USE
THICK-BODY OR WIDE-BODY
COPPER PIPE. USE 3/4" INSIDE
DIAMETER FOR UP TO 300' OR
1" FOR UP TO 500'.
3. ROOM THERMOSTAT
MAINTAINED AT 68°F OR
COOLER AND 30-60% RELATIVE
HUMIDITY.
4. AIR SHUT OFF CABINET.
LOCATION MAY VARY. SEE SITE
PLANNING GUIDE (SPG) FOR
ALTERNATE LOCATIONS.
5. DEDICATED EQUIPMENT
THERMOSTAT BEHIND THE
MACHINE ISOCENTER ON WALL
A 5'-0" A.F.F. MUST SUPPLY
AND MAINTAIN 65°F COOL AIR
AND 30-60% RELATIVE
HUMIDITY.
6. HEAT OUTPUT IN THE
TREATMENT VAULT CAN
REACH 51228 BTU/H (14.3 KW)
AND THE PDU GENERATES UP
TO 2391 BUT/H (0.7 KW)
7. MOISTURE SENSOR WITH
ALARM. A FLOOR DRAIN THAT
CAN HANDLE 14 L/min. IS A
POSSIBLE ALTERNATIVE.
8. INLINE AIR REGULATOR TO BE
PLACED IN TREATMENT ROOM.
SEE SITE PLANNING GUIDE
(SPG) FOR ALTERNATE
LOCATIONS.
9. ROOM TEMPERATURE
MAINTAINED AT 68-75°F AND
30-60% RELATIVE HUMIDITY.
POWER CONDITIONER/
FREQUENCY CONVERTER
TREATMENT ROOM SUB-BASE FLOOR PLAN
SCALE: 1/8" = 1'-0"
A1
6a/6b
AIR
COMPRESSOR
DRYER
TANK
NOTE:
THE FOLLOWING EQUIPMENT MUST BE PROVIDED
BY THE FACILITY AND MEET SPECIFICATIONS (SEE
SITE PLANNING GUIDE): AIR TANK, DRYER, & AIR
COMPRESSOR
6b
TO AIR REGULATOR IN
TREATMENT VAULT.
(SEE NOTE 8)
AIR TANK, 60 US GALLON MIN.
REQUIRED
AIR
MECHANICAL ROOM PLAN
SCALE: NOT TO SCALE
A4
TREATMENT ROOM MECHANICAL PLAN
SCALE: 1/8" = 1'-0"
C1
DATA SERVER ROOM HEATING, VENTILATION & AIR CONDITIONING
SCALE: 1/4" = 1'-0" NOT SITE SPECIFIC
D4
SUPPLY
RETURN
GENERAL NOTES
SHEET NOTES
LEGEND
A. THESE DRAWINGS ARE FOR
REFERENCE ONLY. THEY
ARE NOT TO REPLACE
CONSTRUCTION
DOCUMENTS AND
SHIELDING STUDIES BY
LICENSED PROFESSIONALS.
B. FINAL RADIATION SHIELDING
SHALL BE COMPLETED BY
THE PHYSICIST OF RECORD.
C. UNLESS OTHERWISE NOTED,
ALL MATERIALS SUPPLIED
AND INSTALLED ARE THE
RESPONSIBILITY OF THE
CUSTOMER.
D. REFER TO SITE PLANNING
GUIDE (SPG) FOR
ADDITIONAL
SPECIFICATIONS.
E. ALL STRUCTURAL
CALCULATIONS TO BE
COMPLETED BY
STRUCTURAL ENGINEER OF
RECORD.
F. ALL FINAL DIMENSIONS TO
BE FIELD VERIFIED.
A
N
T
R
Y
P
T
FACILITY SUPPLIED
SUPPLEMENTAL AIR
DUCT TERMINATE 2"
ABOVE FINISHED
FLOOR.
FACILITY-SUPPLIED
SUPPLEMENTAL
AIR DUCT
G
A
N
T
S
U
B
B
A
S
E
ELECTRICAL
FLOOR
CONDUIT
I
S
O
C
E
N
T
E
R
ELECTRICAL
FLOOR BOX
C
L
C
L
C L
C L
C L
G
C
PB5
D
PB5
COMPRESSED
AIR LINE
ISO CENTER
GANTRY PIT
CENTERLINE
6'-2 1/2"
6"
6"
2'-3"
7"
1'-4"
DRAIN/ MOISTURE
SENSOR
2'-5"
2'-3"
2'-8
2'-9"
PB2
COMPRESSED AIR
LINE
1'-3
1'-1
6"
2'-8"
ELECTRICAL CONDUIT
SUPPLEMENTAL AIR
PB5 PB5
ACCURAY INCORPORATED
1310 Chesapeake Terrace
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E
S
P D
R
M-501
MECHANICAL
DETAILS
SIGNAL (3"Ø) - FROM
PATIENT TABLE TO BACK OF
GANTRY (20'-0" MAX).
POWER (3"Ø) - FROM
PATIENT TABLE TO BACK OF
GANTRY (55'-0" MAX).
FLOOR PULL BOX
(1'-0" x 8" x 8")
PATIENT TABLE PULL BOX
(6" x 6" x 6")
TREATMENT ROOM SUB-BASE FLOOR PLAN (TRENCH OPTION)
NOT TO SCALE
A1
MECHANICAL ROOM PLAN
NOT TO SCALE
A4
BACK OF GANTRY AXON (CONDUIT OPTION)
NOT TO SCALE
D1
AIR SHUT OFF VALVE ACCESS PANEL
NOT TO SCALE
D4
C
D
PB5
1. COMPRESSED AIR SUPPLY SHUT
OFF VALVE.
2. 3/4" THICK BODY OR WIDE WALL
COPPER PIPE BACK TO AIR
COMPRESSOR WITH FLOW RATE
OF 15 SCFM AT 90 PSIG.
3. QUICK DISCONNECT FOR
SERVICE AIR HOSE.
4. BARBED FITTING FOR 1/2" INSIDE
DIAMETER AIR HOSE.
CONTRACTOR SUPPLIED AND
INSTALLED JUST ABOVE
FINISHED FLOOR.
5. SHUT-OFF VALVE TO POINT
DOWN IN CLOSED POSITION.
6. MESH PROTECTION ON TOP OF
SUPPLEMENTAL AIR DUCT.
7. FACILITY TO PROVIDE 55°F
SUPPLEMENTAL AIR FLOW
GREATER THAN 264 CFM.
INSTALL AN UNDER-SLAB 12"x10"
AIR DUCT OR A 12" ROUND (OR
EQUIVALENT MINIMUM
CROSS-SECTIONAL AREA) PVC
DUCT TO THE UNDERSIDE OF
THE GANTRY COVERS.
TO
TREATMENT
SYSTEM
FROM AIR
COMPRESSOR
1'-0"
1'-6"
NOTE 1
NOTE 2
PB2
GENERAL NOTES
SHEET NOTES
LEGEND
A. THESE DRAWINGS ARE FOR
REFERENCE ONLY. THEY
ARE NOT TO REPLACE
CONSTRUCTION
DOCUMENTS AND
SHIELDING STUDIES BY
LICENSED PROFESSIONALS.
B. FINAL RADIATION SHIELDING
SHALL BE COMPLETED BY
THE PHYSICIST OF RECORD.
C. UNLESS OTHERWISE NOTED,
ALL MATERIALS SUPPLIED
AND INSTALLED ARE THE
RESPONSIBILITY OF THE
CUSTOMER.
D. REFER TO SITE PLANNING
GUIDE (SPG) FOR
ADDITIONAL
SPECIFICATIONS.
E. ALL STRUCTURAL
CALCULATIONS TO BE
COMPLETED BY
STRUCTURAL ENGINEER OF
RECORD.
F. ALL FINAL DIMENSIONS TO
BE FIELD VERIFIED.
ISOCENTER
DIMENSION FORMAT: FEET & INCHES
NOTE 3
NOTE 4
COMPRESSED
AIR LINE (NOTE
2)
C
L
SUPPLEMENTAL
AIR DUCT.
(NOTE 6,7)
ELECTRICAL
TRENCH/
CONDUITS
C
L
C
L
7"
MACHINE
CENTERLINE
2"
NOTE 5
ROOM
B
A
F
EO
MP
RL
JB4
TS
TO DEDICATED
HVAC SYSTEM
S
NOTE 1, TYP.
NOTE 2, TYP.
NOTE 1, TYP.
TREATMENT
ROOM
NOTE 4
PB1
TRENCHTRENCH
D
C
1'-4"
2'-8"
PB2
2'-5"
2'-3"
PB5
PB5
RL
ES
RL
ES
ES IC
ES
DS
3 4
B A iDMS
ACCURAY INCORPORATED
1310 Chesapeake Terrace
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E
S
P D
R
E-101
ELECTRICAL
FLOOR PLAN
PDU PULL BOX (FLOOR). 2'-0"
x 1'-6" x 1'-0" WITH COVER
PLATE. LAST 6" OPEN TO
REMAIN OPEN. COVER TO
WITHSTAND WEIGHT OF PDU.
PATIENT TABLE PULL BOX
WITH DIVIDER
(6" x 6" x 6")
OPERATOR STATION PULL
BOX (WALL). CONTRACTOR
TO DETERMINE SIZE.
CONDUIT FLOOR PULL BOX
(1'-0" x 8" x 8")
PB5
1. CCTV WIRING RUN TO CONTROL
ROOM TO BE DETERMINED BY
CUSTOMER OR CONTRACTOR.
2. PULL STRINGS MUST BE
INCLUDED IN ALL NON-WIRED
CONDUITS.
3. 90 DEGREE ELBOWS OR
FLEXIBLE CONDUIT ARE NOT TO
BE USED IN CONDUIT RUNS.
4. SEE DETAIL A1 ON SHEET E-901
PB2
S
FACILITY PROVIDED
EO EMERGENCY OFF BUTTON
ES EMERGENCY STOP BUTTON
DS DOOR INTERLOCK SWITCH
RS RESET SWITCH(DOOR-LESS)
RL RADIATION WARNING LIGHT
MP MAIN POWER DISCONNECT
IC INTERCOM
TS THERMOSTAT
PC POWER CONDITIONER
FQ FREQUENCY CONVERTER
SIGNAL CONDUIT
POWER CONDUIT
NON-WIRED CONDUITS (EMPTY
CONDUITS)
A SIGNAL (3"Ø) OPERATOR
STATION TO PDU. (80' MAX).
B POWER (3"Ø) - OPERATOR
STATION TO PDU. (80' MAX).
C SIGNAL (3"Ø) - FROM PATIENT
TABLE TO BACK OF GANTRY 20'
MAX).
D POWER (3"Ø) - FROM PATIENT
ABLE TO BACK OF GANTRY 55'
MAX).
E SIGNAL (1
"Ø) - DORADO
ASERS TO PDU (65' MAX).
F PHYSICS PORT (4"Ø)
G SIGNAL (1"Ø) - INTERCOM TO
BACK OF GANTRY (15' MAX).
PB4
PB1
OVERALL FLOOR PLAN
SCALE: 1/4" = 1'-0"
A1
GENERAL NOTES
SHEET NOTES
LEGEND
A. THESE DRAWINGS ARE FOR
REFERENCE ONLY. THEY ARE NOT
TO REPLACE CONSTRUCTION
DOCUMENTS AND SHIELDING
STUDIES BY LICENSED
PROFESSIONALS.
B. FINAL RADIATION SHIELDING SHALL
BE COMPLETED BY THE PHYSICIST
OF RECORD.
C. UNLESS OTHERWISE NOTED, ALL
MATERIALS SUPPLIED AND
INSTALLED ARE THE
RESPONSIBILITY OF THE
CUSTOMER.
D. REFER TO SITE PLANNING GUIDE
(SPG) FOR ADDITIONAL
SPECIFICATIONS.
E. ALL STRUCTURAL CALCULATIONS
TO BE COMPLETED BY STRUCTURAL
ENGINEER OF RECORD.
F. ALL FINAL DIMENSIONS TO BE FIELD
VERIFIED.
ISOCENTER
CONDUIT
LOCATION NOTE
POWER (1/2"Ø) -
EMERGENCY
OFF BUTTON AT
OPERATOR
STATION TO MAIN POWER
DISCONNECT PANEL.
POWER (4"Ø) - MAIN
POWER DISCONNECT TO
PDU. SEE PAGE E-501 FOR
ELECTRICAL WIRING
OPTIONS
POWER (1"Ø) -
EMERGENCY STOP
SWITCHES TO PDU.
POWER (1"Ø) - RADIATION
WARNING LIGHTS TO PDU.
POWER (1"Ø) - DOOR
INTERLOCK SWITCH TO
PDU.
POWER (1"Ø) - DORADO
LASERS TO WALL SWITCH.
7 LOCAL GROUNDING
SIGNAL (1"Ø) INTERCOM TO
CONTROL ROOM
POWER 4" PDU TO POWER
CONDITIONER TO MP
(4) 4"Ø CONDUIT FROM JB1
TO JB5
WIRED CONDUIT LEGEND
APOLLO LASER WALL
POWER SWITCH
1 POWER OUTLET (NEMA
6-15R OR EQUAL)
2 POWER OUTLET (MATCH
LOCAL)
3 POWER OUTLET (220V
12AMP)
ROOM
TREATMENT
ROOM
PB3
E
E
S
PB1
NOTE 3
NOTE 1, TYP.
NOTE 2
NOTE 2
EH
12'-8"
NOTE 4
H
ACCURAY INCORPORATED
1310 Chesapeake Terrace
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E
S
P D
R
E-111
ELECTRICAL
CEILING PLAN
1. PULL STRINGS MUST BE
INCLUDED IN ALL NON-WIRED
CONDUITS.
2. MINIMUM 2 CEILING MOUNTED
CCTV WITH PAN, TILT AND
ZOOM @ 45° FROM ISOCENTER
(FACILITY PROVIDED). EXACT
LOCATION TO BE DETERMINED
BY CUSTOMER.
3. APOLLO (GREEN AND DORADO
(RED) LASER ALUMINUM
MOUNTING PLATED TO BE
PROVIDED AND INSTALLED BY
CONTRACTOR. ACCURAY
PROVIDES AND INSTALLS ALL
LASERS.
4. SYNCHRONY MOUNTING
LOCATION. KEEP 2' CLEAR FOR
INSTALLATION AND SERVICING
WIRED CONDUITS
6 POWER (1"Ø) - DORADO
LASERS TO PDU.
8 SIGNAL (3/4"Ø) INTERCOM TO
CONTROL ROOM
10 POWER (1"Ø) APOLLO LASER
NON-WIRED CONDUITS
(EMPTY CONDUITS)
E SIGNAL (1
"Ø) - DORADO
LASERS TO PDU. (65' MAX).
H (2"Ø)- SYNCHRONY CONDUIT
PB1 PDU PULL BOX (FLOOR).
2'-0" x 1'-6" x 1'-0"
PB3 DORADO POWER PULL
BOX (CEILING). SIZE TO BE
DETERMINED BY
CONTRACTOR.
SIGNAL CONDUIT
POWER CONDUIT
TREATMENT ROOM CEILING PLAN
SCALE: 1/4" = 1'-0"
A1
GENERAL NOTES
SHEET NOTES
LEGEND
A. THESE DRAWINGS ARE FOR
REFERENCE ONLY. THEY
ARE NOT TO REPLACE
CONSTRUCTION
DOCUMENTS AND
SHIELDING STUDIES BY
LICENSED PROFESSIONALS.
B. FINAL RADIATION SHIELDING
SHALL BE COMPLETED BY
THE PHYSICIST OF RECORD.
C. UNLESS OTHERWISE NOTED,
ALL MATERIALS SUPPLIED
AND INSTALLED ARE THE
RESPONSIBILITY OF THE
CUSTOMER.
D. REFER TO SITE PLANNING
GUIDE (SPG) FOR
ADDITIONAL
SPECIFICATIONS.
E. ALL STRUCTURAL
CALCULATIONS TO BE
COMPLETED BY
STRUCTURAL ENGINEER OF
RECORD.
F. ALL FINAL DIMENSIONS TO
BE FIELD VERIFIED.
ISOCENTER
DIMENSION FORMAT: FEET & INCHES
$ WALL POWER SWITCH
1 POWER OUTLET (NEMA 6-15R)
2 POWER OUTLET
(MATCH LOCAL)
PB3
S
TO BE
DETERMINED
BY FACILITY
2'-3
5'-0"
6'-7
3'-8
E
E
E
CCTV
G
INTERCOM
JB2
PB1
NOTE 2
PB5
I S
O
C E
N T
E R
CCTV
H
12'-8"
PB3
E
E
NOTE 1
NOTE 2
3'-8
6'-7
E
E
I S
O
C E
N T
E R
ACCURAY INCORPORATED
1310 Chesapeake Terrace
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E
S
P D
R
E-301
STRUCTURAL
FLOOR PLAN
WIRED CONDUITS
6 POWER (1"Ø) - DORADO
LASERS TO PDU.
8 SIGNAL (1"Ø) INTERCOM TO
CONTROL ROOM
10 POWER (1"Ø) APOLLO LASER
NON-WIRED CONDUITS
(EMPTY CONDUITS)
E SIGNAL (1
"Ø) - DORADO
LASERS TO PDU (65' MAX).
G SIGNAL (1/2"Ø) - INTERCOM TO
BACK OF GANTRY (15' MAX).
H (2" Ø) SYNCHRONY CONDUIT
JUNCTION BOXES
PB1 PDU PULL BOX
(FLOOR) 2' x x 1'-6" x 12"
WITH COVER PLATE. LAST
6" TO REMAIN OPEN.
PB2 COUCH PULL BOX
(FLOOR) 6" x 6" x 6"
PB3 DORADO POWER
PULL BOX (CEILING).
SIZE TO BE DETERMINED
BY CONTRACTOR.
PB5 FLOOR PULL BOX
1'-0" x 8" x 8"
PLAN WEST-EAST TREATMENT ROOM SECTION
SCALE: 1/4" = 1'-0" NOT SITE SPECIFIC
A1
PLAN NORTH-SOUTH TREATMENT ROOM SECTION
SCALE: 1/4" = 1'-0" NOT SITE SPECIFIC
C1
1 LOCATION TO BE DETERMINED
BY CONTRACTOR
2 MOISTURE SENSOR WITH
ALARM. A FLOOR DRAIN THAT
CAN HANDLE 14 L/min IS AN
ACCEPTABLE ALTERNATIVE
$ WALL POWER SWITCH
1 POWER OUTLET (NEMA 6-15R)
2 POWER OUTLET (MATCH
LOCAL)
GENERAL NOTES
SHEET NOTES
LEGEND
A. THESE DRAWINGS ARE FOR
REFERENCE ONLY. THEY
ARE NOT TO REPLACE
CONSTRUCTION
DOCUMENTS AND
SHIELDING STUDIES BY
LICENSED PROFESSIONALS.
B. FINAL RADIATION SHIELDING
SHALL BE COMPLETED BY
THE PHYSICIST OF RECORD.
C. UNLESS OTHERWISE NOTED,
ALL MATERIALS SUPPLIED
AND INSTALLED ARE THE
RESPONSIBILITY OF THE
CUSTOMER.
D. REFER TO SITE PLANNING
GUIDE (SPG) FOR
ADDITIONAL
SPECIFICATIONS.
E. ALL STRUCTURAL
CALCULATIONS TO BE
COMPLETED BY
STRUCTURAL ENGINEER OF
RECORD.
F. ALL FINAL DIMENSIONS TO
BE FIELD VERIFIED.
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E
S
P D
R
E-501
ELECTRICAL
DETAILS
C1
ELECTRICAL DIAGRAM - OPTION 1
NOT TO SCALE
A1
ELECTRICAL DIAGRAM - OPTION 2
NOT TO SCALE
380 - 480 VAC DELTA 3 PH
TO WYE 3 PH PRIMARY
TRANSFORMER SECONDARIES
Magnetic Contactor
A A
BB
CC
GG
B
G
N
C
AFACILITY POWER
A
B
C
G
*SITE SPECIFIC OPTION SUPPLIED BY ACCURAY
SUPPLIED BY FACILITY
G
GROUND
CONNECTION
FROM BUILDING
GROUND
380 - 480 VAC 3 PH
CANNOT HAVE
BACK-UP BATTERY
IN CONFIGURATION
(LOCATE OUTSIDE ROOM NEAR
OPERATOR STATION)
EPO
O
E
A T
M E
S
T E
M
NOT SITE SPECIFIC
NOT SITE SPECIFIC
*LOCAL GROUND TO BUILDING STEEL
OR GROUNDING ROD. TO BE
DETERMINED BY ELECTRICIAN. (NOTE 4)
1. IN CONFIGURATION OPTION 1,
WHERE FACILITY POWER
GOES DIRECTLY TO MP, THE
FCU MUST BE RESET AFTER
POWER EMERGENCY STOP.
2. CABLING FROM THE SUPPLY
TO PDU IS THE FACILITY'S
RESPONSIBILITY.
3. 70 AMP MAIN DISCONNECT
PANEL. FACILITY TO PROVIDE
MAIN POWER SUPPLY BOX
FROM DEDICATED LINE OR
POWER PROTECTION DEVICE.
4. GROUND TO HAVE EARTH
CONDUCTED IMPEDANCE OF
25 OHMS OR LESS. MUST
COMPLY WITH LOCAL CODES.
PDU POWER DISTRIBUTION UNIT
FCU FREQUENCY CONVERTER
UNIT (50Hz SITES ONLY)
MP MAIN DISCONNECT PANEL
PC POWER CONDITIONER
A PHASE A
B PHASE B
C PHASE C
G GROUND
N NEUTRAL
WIRED CONDUITS
1 POWER (1/2"Ø) - PDU TO
PC/FCU TO MP
NOTE 2
MP
*FCU/PC
PDU
380 - 480 VAC DELTA 3 PH
TO WYE 3 PH PRIMARY
TRANSFORMER SECONDARIES
Magnetic Contactor
A A
BB
CC
GG
B
G
N
C
AFACILITY POWER
A
B
C
G
SUPPLIED BY ACCURAY
SUPPLIED BY FACILITY
G
GROUND
CONNECTION
FROM BUILDING
GROUND
380 - 480 VAC 3 PH
CANNOT HAVE
BACK-UP BATTERY
IN CONFIGURATION
(LOCATE OUTSIDE ROOM NEAR
OPERATOR STATION)
EPO
E
A T
M E
S
T E
M
*LOCAL GROUND TO BUILDING STEEL
OR GROUNDING ROD. TO BE
DETERMINED BY ELECTRICIAN. (NOTE 4)
NOTE 2
MP
*FCU/PC
PDU
GENERAL NOTES
SHEET NOTES
LEGEND
A. THESE DRAWINGS ARE FOR
REFERENCE ONLY. THEY
ARE NOT TO REPLACE
CONSTRUCTION
DOCUMENTS AND
SHIELDING STUDIES BY
LICENSED PROFESSIONALS.
B. FINAL RADIATION SHIELDING
SHALL BE COMPLETED BY
THE PHYSICIST OF RECORD.
C. UNLESS OTHERWISE NOTED,
ALL MATERIALS SUPPLIED
AND INSTALLED ARE THE
RESPONSIBILITY OF THE
CUSTOMER.
D. REFER TO SITE PLANNING
GUIDE (SPG) FOR
ADDITIONAL
SPECIFICATIONS.
E. ALL STRUCTURAL
CALCULATIONS TO BE
COMPLETED BY
STRUCTURAL ENGINEER OF
RECORD.
F. ALL FINAL DIMENSIONS TO
BE FIELD VERIFIED.
DIMENSION FORMAT: FEET & INCHES
NOTE 3
NOTE 3
NOTE 1
*SITE SPECIFIC OPTION
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E
S
P D
R
E-502
ELECTRICAL
DETAILS
A1
DATA SERVER UNIT
NOT TO SCALE
(Power Distribution Unit)
PDU
TB1
L1
L2
L3
PDU
Pull Box
(PB1, in floor)
A B C GND
shunt trip coil
60 Hz
480 VAC, +/- 5%
80 Amp facility supplied control power
N.O.
One EO normally open switch located within reach of Operator Station. "Push to lock, twist to release" switch recommended.
- The system Emergency Off (EO) located on the power control panel is responsible for shutting off all electrical power from the PDU by tripping CB1 (on the PDU) with the exception of the
Un-interruptible Power Supply (UPS)
-The UPS remains on and provides power to the Data Receiver Server
(DRS) for data logging.
Facility Responsibility Accuray Responsibility
UPS
(Cluster/Optimizer)
Data Server Unit
UPS
DSU
UPS INPUT CORDS (10 foot / 3 meters)
USA domestic = Nema L6-20P to internal wired
International = site specific plug to IEC 320-C20
FACILITY POWER REQUIREMENTS FOR DSU
2 circuits @ 200-240 VAC, 20A, 50/60 Hz
G
X
Y
X
Y
Facility Responsibility Accuray Responsibility
Note: The DSU communication cable connections
(LC Duplex Fiber Optic and CAT5) under the responsibility of the facility are referenced in the Site
Planning Guide.
Facility Responsibility
PDU Junction Box
(PB1 in floor)
Ceiling Junction
Box
(JB5)
B
Y-RY-L
A
X receptacles for dorado lasers typical 5 places
NEMA 6-15R
USA Domestic
SITE SPECIFIC
International
LASER POWER CORDS (type SJT 18AWG)
USA domestic = Nema 6-20P to IEC320-C13
International = site specific plug to IEC320-C13
5X
Dorado Laser
Load = 240VAC, 30W each
Facility Responsibility Accuray Responsibility
Apollo Laser (ceiling)
Load = 120/240VAC, 10W
Apollo Laser (rear wall)
Load = 120/240VAC, 10W
LASER POWER CORDS
Plug end converts 120/240 VAC to 5VDC
Site Specific plug adaptors provided by Accuray
USA domestic = NEMA 5-15P
International = site specific
FACILITY POWER
SPST switch in treatment room controls both lasers.
N
L
POWER RECEPTACLES
USA domestic = 120V NEMA 5-15R
International = site specific
UPS
To DRS
Notes:
- Ground conductor to be the same diameter as phases. 4 AWG MIN.
- Neutral conductor not used
- PF = 0.90 at max level
-50 kVA max
120-240 VAC FACILITY NEUTRAL
120-240 VAC FACILITY LINE
PDU TB6
N
L
SOLID
STATE
RELAY
Accuray Responsibility
CAT5 PATCH PANEL
DSU Room
LC F.O. PATCH PANEL
DSU Room
1x
4x
Note: F.O. AND CAT5 PATCH
CABLES PROVIDED BY ACCURAY.
(3 METERS)
FC/PC NOT SHOWN
5x
5x
If NEMA version of receptacle is not available in your locale, contact the
Accuray Project
Manager to coordinate the use of locally approved receptacles and plugs, USA domestic = 120 VAC 60 Hz, 15A
International = 120-240 VAC 50/60 Hz, 15/16A iDMS™
CAT5 Receptacles 3x
Planning Room
LC F.O. PATCH PANEL
Control Room
4x
Duplex Pair 3x (or) 6 Strand Bundle
Extended Network Optional
OR
C1
EMERGENCY OFF AND PDU POWER INPUT
NOT TO SCALE
C4
DORADO LASER (RED)
NOT TO SCALE
A4
APOLLO LASER (GREEN)
NOT TO SCALE
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E
S
P D
R
E-503
ELECTRICAL
DETAILS
A1
STATUS INDICATORS
NOT TO SCALE
N.C.
Facility Responsibility Accuray Responsibility
Circuit Specifications
- 48-240 VAC, 50/60 Hz
- 24-10 AWG
- Min ON current = 40mA
- Max OFF current = 5mA
- Dedicated circuit protection recommended.
- Provide panel/breaker designation # for the facility indicator circuit. De-energizing label to be affixed by
Accuray to PDU access panel.
N.C.
N.C.
Additional E-STOP switches as needed wired normally closed in
SERIES.
PDU Pull Box
(PB1 in floor)
-TB5 connections made by Accuray
-Label wires "E-STOP"
24VDC
"Push to lock, twist to release"' switch
Min. 20 AWG shielded, twisted-pair wire
E-STOP prevents and/or terminates radiation by shutting off electrical power to the gantry Slip Rings and all other rotating components.
- Stops gantry rotation.
- Stops automatic couch motion.
- Stops Radiation Delivery Subsystem (RDS)
- Leads to an Unrecoverable State
Facility Responsibility Accuray Responsibility
PDU
PDU Pull Box
(PB1 in floor)
-Leave 6 feet (2 meters) of extra wire length in junction box.
- TB5 connections performed by Accuray
-Label wires
"DOOR SW"
24V
TB5
Wire size = 20 AWG min.
24 VDC, 3A
Use only momentary type switch.
Contacts are open when door is open
Facility Responsibility
Accuray Responsibility
PDU Pull Box
(PB1 in floor)
Indicator Specifications
- 8A max each
- Incandescent lamps and/or auditory indicators.
- Fluorescent lamps (40 to 200 W)
- Inductive or electronic ballasts are ok to use.
- LED lamps may require additional facility supplied coil type relay circuits.
Power
On
Room
Ready
Standby
TB5
* Radiation
On
240 VAC FACILITY LINE
240 VAC FACILITY LINE
240 VAC FACILITY LINE
240 VAC FACILITY LINE
L
L
L
L
240 VAC FACILITY
NEUTRAL
PDU
TB6
* Min. 1 lamp circuit required (Radiation On).
Other lamp circuits optional per local requirements.
Door Switch
PDU
TB6 wires connected by Accuray
** Status indicators located inside and/or outside treatment vault as needed.
RJ45
receptacle for Accuray
Speaker Unit on rear wall
RJ45
receptacle for Accuray
Intercom, Desk Unit
1x
Facility supplied CAT5 ethernet cable between Control Room and
Rear Wall of Vault
Label receptacles
"NEIS"
B1
ACCURAY INTERCOM (NEIS)
NOT TO SCALE
C1
FACILITY EMERGENCY STOP (E-STOP)
NOT TO SCALE
C4
DOOR INTERLOCK -TREATMENT ROOM
NOT TO SCALE
C2
FACILITY STANDBY
NOT TO SCALE
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E
S
P D
R
E-901
ELECTRICAL
DETAILS
C1
PDU FLOOR PULL BOX
NOT TO SCALE
A1
PATIENT TABLE FLOOR PULL BOX
NOT TO SCALE NOT SITE SPECIFIC
1. POWER DISTRIBUTION UNIT
2. COVER PROVIDED BY ACCURAY.
3. INCOMING POWER ENCASED IN 2" FLEXIBLE
CONDUIT ON LEFT SIDE OF PDU.
CONTRACTOR PROVIDED.
4. LAST 6" OF FLOOR BOX TO REMAIN OPEN
FOR WIRE CONNECTION. CONTRACTOR
PROVIDED.
5. TERMINATE COVER PLATE FLUSH WITH
FINISHED CONCRETE FLOOR. CONTRACTOR
TO PROVIDE ALL COVER PLATES.
6. INCOMING 230VAC POWER TO DORADO
LASERS
F
R
O
N
T
E
D
F
B
O
X
N
T
L
N
E
PATIENT
TABLE
FLOOR BOX
(PB2)
D
NOTE 5
C3 iDMS
NOT TO SCALE
C5
POWER DISTRIBUTION UNIT (PDU)
NOT TO SCALE
WEIGHT: 900 lb
18,000 BTUs/HOUR
F
R
O
N
T
5'-3
AIR
VENT
WEIGHT: 980 lb
3370 BTUs/HOUR
SEISMIC
BRACKETS (IF
APPLICABLE)
GENERAL NOTES
SHEET NOTES
LEGEND
A. THESE DRAWINGS ARE FOR REFERENCE
ONLY. THEY ARE NOT TO REPLACE
CONSTRUCTION DOCUMENTS AND
SHIELDING STUDIES BY LICENSED
PROFESSIONALS.
B. FINAL RADIATION SHIELDING SHALL BE
COMPLETED BY THE PHYSICIST OF RECORD.
C. UNLESS OTHERWISE NOTED, ALL
MATERIALS SUPPLIED AND INSTALLED ARE
THE RESPONSIBILITY OF THE CUSTOMER.
D. REFER TO SITE PLANNING GUIDE (SPG) FOR
ADDITIONAL SPECIFICATIONS.
E. ALL STRUCTURAL CALCULATIONS TO BE
COMPLETED BY STRUCTURAL ENGINEER OF
RECORD.
F. ALL FINAL DIMENSIONS TO BE FIELD
VERIFIED.
ISOCENTER
JUNCTION BOXES
PB1 PDU PULL BOX (FLOOR). 2'-0" x 1'-2" x 1'-0"
WITH COVER PLATE. LAST 6" OPEN TO
REMAIN OPEN.
PB2 PATIENT TABLE PULL BOX (FLOOR). 6" x
6" x 6" WITH COVER PLATE FLUSH
TO FINISHED FLOOR.
4'-8"
1'-10"
2'-7"
FRONT
NOTE 1
NOTE 3
NOTE 6
4 3 3
E
4 3 2 B
A
NOTE 4
PB1
WIRED CONDUITS
POWER (4"Ø)- MAIN POWER DISCONNECT TO
PDU
POWER (1"Ø)- EMERGENCY STOP SWITCHES
TO PDU
POWER (1"Ø)- RADIATION WARNING LIGHTS
TO PDU
POWER (1"Ø)- DOOR INTERLOCK SWITCH TO
PDU
POWER (1"Ø)- DORADO LASERS TO PDU
LOCAL GROUNDING
SIGNAL (3"Ø)- OPERATOR STATION TO PDU.
80' MAX
POWER (3"Ø)- OPERATOR STATION TO PDU.
80' MAX.
SIGNAL (3"Ø)- FROM PATIENT TABLE TO
BACK OF GANTRY.
20' MAX
POWER (3"Ø)- FROM PATIENT TABLE TO
BACK OF GANTRY TO PDU. 55' MAX.
SIGNAL (1 1/2"Ø)- DORADO LASER TO PDU.
65' MAX.
(2" Ø) -SYNCHRONY CONDUIT
(4"Ø)- FROM PB1 TO PB5 (CONDUIT OPTION
ONLY)
A
B
C
D
E
NON-WIRED CONDUITS
A3
NOT USED
NOT TO SCALE NOT SITE SPECIFIC
H
H
WEIGHT: 1,502 lb
12,000 BTU/HOUR
ONLY FOR
FACILITIES
WITH 50Hz POWER
5'-11 1/4"
A5
FREQUENCY CONVERTER UNIT (FCI)
NOT TO SCALE NOT SITE SPECIFIC
TREATMENT ROOM
CONTROL ROOM
MECHANICAL
ROOM
TREATMENT
PLANNING ROOM
DATA SERVER
ROOM
ACCURAY INCORPORATED
1310 Chesapeake Terrace
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E
S
P D
R
NOTE 5
NOTE 3
NOTE 2
LC
= LC CONNECTOR
NOTE 1
LC LC LC
N-101
NETWORK
FLOOR PLAN
A1
OVERALL FLOOR PLAN
SCALE: 3/16" = 1'-0" NOT SITE SPECIFIC
LCLC LC
NOTE 4
PHONE
PHONE
D1
TREATMENT PLANNING ISOMETRIC
NOT TO SCALE
ALL ITEMS IN PLANNING STATION
ARE ON FACILITY POWER SUPPLY.
D5
CONTROL ROOM ISOMETRIC
NOT TO SCALE
D3
DATA SERVER ROOM ISOMETRIC
NOT TO SCALE
1'-0"
1. CAT6E CABLE (OR GREATER)
FROM WALL MOUNTED
SPEAKER TO CONTROL ROOM.
2. CAT6E CABLES (OR GREATER)
FROM DATA SERVER ROOM TO
TREATMENT PLANNING.
3. FIBER OPTIC NETWORK
CONNECTIONS FROM DATA
SERVER ROOM TO CONTROL
ROOM.
4. CAT6E CABLES (OR GREATER)
FROM DATA SERVER ROOM TO
FACILITY NETWORK.
5. DIGITAL INTERNET DATA JACK.
6. FIBER OPTICS REQUIRED OVER
300' ALL DIRECT ROUTE.
7. BACKUP COPPER CAT6E
CABLES (OR GREATER)
= DATA OUTLET
= PHONE
OUTLET/INTERCOM
GENERAL NOTES
SHEET NOTES
LEGEND
A. THESE DRAWINGS ARE FOR
REFERENCE ONLY. THEY
ARE NOT TO REPLACE
CONSTRUCTION
DOCUMENTS AND
SHIELDING STUDIES BY
LICENSED PROFESSIONALS.
B. FINAL RADIATION SHIELDING
SHALL BE COMPLETED BY
THE PHYSICIST OF RECORD.
C. UNLESS OTHERWISE NOTED,
ALL MATERIALS SUPPLIED
AND INSTALLED ARE THE
RESPONSIBILITY OF THE
CUSTOMER.
D. REFER TO SITE PLANNING
GUIDE (SPG) FOR
ADDITIONAL
SPECIFICATIONS.
E. ALL STRUCTURAL
CALCULATIONS TO BE
COMPLETED BY
STRUCTURAL ENGINEER OF
RECORD.
F. ALL FINAL DIMENSIONS TO
BE FIELD VERIFIED.
ISOCENTER
DIMENSION FORMAT: FEET & INCHES
PHONE
NOTE 5
NOTE 6,
7 TYP.
DIRECT WIRING TO
DOSIMETRY (NOTE 2)
8 RJ45 WALL
JACK (CAT6 MIN.)
TO HOSPITAL
NETWORK (NOTE 4)
8 RJ45 WALL
JACK (CAT6
MIN.) FOR
ACCESS TO
HOSPITAL
NETWORK
DIRECT WIRING TO
CONTROL STATION
(NOTE 3,6 & 7)
WALL RJ45 CONNECTOR
TO HOSPITAL NETWORK
WITH INTERNET ACCESS
WIRED BY ACCURAY
WIRED BY FACILITY
NOTE 1
NOTE 6
65°
CABINETSCABINETS
LASER BEAM
LASER
DORADO MOUNTING PLATE
6 x
1'-4"
3'-7
1"
1"
4 x
1'-6"
1'-0"
1"
1"
FINISHED CEILING
OPENING VARIES
(COORDINATE WITH
ACCURAY STAFF).
RECEPTACLE
CONNECTED TO
WALL SWITCH
1'-6"
10"
1'-4"
10"
FINISHED CEILING
RECEPTACLE
CONNECTED TO PDU
OPENING VARIES
(COORDINATE WITH
ACCURAY STAFF).
NOT TO SCALE
Q-501
EQUIPMENT
DETAILS
A1
APOLLO WALL LASER - SECION
NOT TO SCALE
A3
DORADO WALL LASER- SECTION
NOT TO SCALE
A5
DORADO MOUNTING PLATE
NOT TO SCALE
B1
APOLLO MOUNTING PLATE
NOT TO SCALE
D1
3/8" ALUMINUM OR 1/4"
STEEL MOUNTING
PLATE, TYP.
SPACING BETWEEN
PLATE AND RIGID
SURFACE (WALL)
1/2"
POWER CONDUIT
(NOTE 4), TYP.
NOTE 6
3/8" ALUMINUM OR 1/4"
STEEL MOUNTING
PLATE, TYP.
SPACING BETWEEN
PLATE AND RIGID
SURFACE (WALL)
POWER CONDUIT
(NOTE5), TYP.
1/2"
SIGNAL CONDUIT, TYP.
1/2"
NOTE 6
SHEET NOTES
LEGEND
A. THESE DRAWINGS ARE FOR
REFERENCE ONLY. THEY ARE NOT
TO REPLACE CONSTRUCTION
DOCUMENTS AND SHIELDING
STUDIES BY LICENSED
PROFESSIONALS.
B. FINAL RADIATION SHIELDING
SHALL BE COMPLETED BY THE
PHYSICIST OF RECORD.
C. UNLESS OTHERWISE NOTED, ALL
MATERIALS SUPPLIED AND
INSTALLED ARE THE
RESPONSIBILITY OF THE
CUSTOMER.
D. REFER TO SITE PLANNING GUIDE
(SPG) FOR ADDITIONAL
SPECIFICATIONS.
E. ALL STRUCTURAL CALCULATIONS
TO BE COMPLETED BY
STRUCTURAL ENGINEER OF
RECORD.
F. ALL FINAL DIMENSIONS TO BE
FIELD VERIFIED.
1. RECEPTACLE CONNECTED TO
WALL SWITCH.
2. RECEPTACLE CONNECTION TO
POWER DISTRIBUTION UNIT.
3. OPENING VARIES. TO BE
COORDINATED WITH SITE
PLANNING TEAM AT ACCURAY.
4. APOLLO LASER POWER
REQUIRES LOCAL STANDARD
OUTLETS DUE TO ADAPTOR
PROVIDED WITH LASER.
5. DORADO LASER POWER
REQUIRES NEMA 6-15R OUTLET.
6. PROVIDE ADDITIONAL SUPPORT
AS NECESSARY TO RIGID
SERVICE.
SIGNAL (1
"Ø) - DORADO
LASERS TO PDU (65' MAX).
E
GENERAL NOTES
APOLLO CEILING LASER - SECTION
NOT TO SCALE
D3
DORADO CEILING LASER - SECTION
NOT TO SCALE
B3
EE
ACCURAY INCORPORATED
1310 Chesapeake Terrace
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E
S
P D
R
D
R
A
D
O
N
O
DORADO
(NOTE 2)
DORADO
D O
R A
D O
N
O T
E
APOLLO
(NOTE 1)
APOLLO
(NOTE 1)
6'-7
3'-8
2'-11
TO THE
ROOM
POWER
SWITCH
JB1
RECEPTACLE (220V)
RECEPTACLES
RECEPTACLES
GANTRY PIT
LASERS PLATES HAVE TO
BE MOUNTED AT 250mm
ABOVE THE FINISHED
CEILING
JB3
5'-0"
CENTERLINE
DORADOS (WALLS) AND APOLLO (CEILING)
CENTERLINE
APOLLO &
DORADOS
(WALLS)
2'-3
JB2
E
E
E
E
C E
N T
E R
L I N
E
A P
O L L O
S
C
E I L I N
W
A L L
CENTERLINE
DORADOS
(WALLS)
CENTERLINE
DORADO (CEILING)
JB5 JB5
E
MATCH
LOCAL
MATCH
LOCAL
(220V)
(220V)
D O
R A
D O
N
O T
E
(220V)
(220V)
ACCURAY INCORPORATED
1310 Chesapeake Terrace
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E
S
P D
R
Q-901
EQUIPMENT
PERSPECTIVE
CONDUIT LOCATIONS
8"
MIN.
SURFACE-MOUNTED GUARD
PROTECTION
10"
MAX.
8"
MIN.
CUT SLOT AFTER THE
LASER HAS BEEN
INSTALLED.
1. APOLLO LASER:
CONNECTED IN SERIES WITH
STANDARD LOCAL DUPLEX
OUTLET ATTACHED TO FACILITY
POWER. NO SIGNAL CONDUIT
REQUIRED.
2. DORADO LASER:
RECEPTACLE NEMA 6-15R
OUTLET AND 1
" SIGNAL
CONDUIT REQUIRED BACK TO
PULL BOX 1 (PB1).
3. ALUMINUM LASER MOUNTING
PLATES ARE SUPPLIED AND
INSTALLED BY CONTRACTOR.
ACCURAY SUPPLIES AND
INSTALLS ALL LASERS.
SURFACE-MOUNTED
CABINET PROTECTION
A1
TREATMENT ROOM PERSPECTIVE
NOT TO SCALE NOT SITE SPECIFIC
D1
DORADO WALL LASER - OPTION 1
NOT TO SCALE
D3
DORADO WALL LASER - OPTION 2
NOT TO SCALE
D5
RECESSED LASER ALCOVE
NOT TO SCALE
B5
RECESSED LASER ALCOVE
NOT TO SCALE
A5
RECESSED LASER ALCOVE
NOT TO SCALE
-OWNER TO PROVIDE POWER
-MOUNT TO 12" x 18" x 3/8" ALUMINUM PLATE
-PROVIDE LOCAL POWER OUTLET
-2 PROVIDED/INSTALLED BY ACCURAY
-OWNER TO PROVIDE POWER & SIGNAL
-MOUNT TO 16" x 42
" x 3/8" ALUMINUM PLATE
-PROVIDE NEMA 6-20R POWER OUTLET
-5 PROVIDED/INSTALLED BY ACCURAY
(2) x 1 1/2"Ø
SIGNAL
CONDUIT (1
IN/ 1 OUT)
CABINETS,
OWNER
FURNISHED,
CONTRACTOR
INSTALLED
(2) x 1 1/2"Ø
SIGNAL
CONDUIT (1
IN/ 1 OUT)
CONDUIT LOCATIONS
RECESSED BOXSURFACE MOUNTED BOX
PULL BOXES
PB1 PDU PULL BOX (FLOOR)
2'-0" x 1'-6" x 1'-0" WITH
COVER PLATE. LAST 150
OPEN TO REMAIN OPEN.
PB2 PATIENT TABLE PULL
BOX (FLOOR) 6" x 6" x 6"
WITH COVER PLATE FLUSH
TO FINISHED FLOOR.
PB3 DORADO POWER PULL
BOX (CEILING). SIZE TO BE
DETERMINED BY
CONTRACTOR.
PB5 FLOOR PULL BOX
1'-0" x 8" x 8"
EMPTY DATA
CONDUIT 1 1/2"Ø
(min)
WIRED POWER
CONDUIT WITH
OUTLET AT
DORADO LASER
SIDE.
WIRED POWER
CONDUIT WITH
OUTLET AT
APOLLO LASER
SIDE, POWERED
FROM THE
FACILITY VIA THE
WALL SWITCH.
OUTLET
MEETING LOCAL
ELECTRICAL
REQUIREMENTS.
E
GENERAL NOTES
SHEET NOTES
LEGEND
A. THESE DRAWINGS ARE FOR
REFERENCE ONLY. THEY
ARE NOT TO REPLACE
CONSTRUCTION
DOCUMENTS AND
SHIELDING STUDIES BY
LICENSED PROFESSIONALS.
B. FINAL RADIATION SHIELDING
SHALL BE COMPLETED BY
THE PHYSICIST OF RECORD.
C. UNLESS OTHERWISE NOTED,
ALL MATERIALS SUPPLIED
AND INSTALLED ARE THE
RESPONSIBILITY OF THE
CUSTOMER.
D. REFER TO SITE PLANNING
GUIDE (SPG) FOR
ADDITIONAL
SPECIFICATIONS.
E. ALL STRUCTURAL
CALCULATIONS TO BE
COMPLETED BY
STRUCTURAL ENGINEER OF
RECORD.
F. ALL FINAL DIMENSIONS TO
BE FIELD VERIFIED.
ISOCENTER
DIMENSION FORMAT: FEET & INCHES
RECEPTACLE
RECEPTACLE
Sunnyvale, CA 94089
SHEET NUMBER
PROJECT INFORMATION
DRAWING INFORMATION
A
B
C
D
E
2 3 4 5 6
4.1.2020A
WEST LA VA
TOMOTHERAPY RADIXACT
LOS ANGELES, CA
REGIONAL PROJECT MANAGER
BETHANY NEIGEBAUER
SITE PLANNING COORDINATOR
JASON LENTZ
PAGE NUMBER
Q U
A L
I T
Y
S
Y S
T E
M
D
O C
U M
E
S
P D
R
G-002
GENERAL
NOTES
GENERAL
NOTES
| Sheets and Views |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-G-001 |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-A-101 |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-A-102 |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-A-201 |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-S-101 |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-S-301 |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-S-501 |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-M-101 |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-M-501 |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-E-101 |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-E-111 |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-E-301 |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-E-501 |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-E-502 |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-E-503 |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-E-901 |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-N-101 |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-Q-501 |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-Q-901 |
| Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-G-002 |