Replace Linear Accelerator at Greater Los Angeles VAHCS Federal contract opportunity
Solicitation number
36C25821B0004
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 22

About this file

This solicitation is for a construction project to replace a linear accelerator at the Greater Los Angeles VA Healthcare System campus in West Los Angeles, California. The selected contractor will demolish the existing linear accelerator, prepare the site, and assist with installing the new Accuray Radixact linear accelerator. Additional requirements include renovating rooms for the linear accelerator, control room, mechanical room, and data server room. The contractor must provide structural supports, foundations, bracing, electrical power, controls, and communications systems. Ceiling, flooring, walls, HVAC, and other infrastructure must be replaced or added. The expected construction period is included in the solicitation. The Department of Veterans Affairs and Veterans Health Administration are the federal agencies procuring this project.

View the file

Other files for this federal contract opportunity

Other files attached to Replace Linear Accelerator at Greater Los Angeles VAHCS, newest first.
File Type Posted
Sign in Sheet for 11 Jan Site Walk Linear Accelerator.pdf PDF
36C25821B0004 0002.pdf PDF
36C25821B0004 0001 RFI Amendment.pdf PDF
36C25821B0004.pdf PDF
Wage Rate Determination CA20200022.docx DOCX document
Atch 1 Self Performance Calculations.pdf PDF
Contractor Background Investigation Request Form (May 2019).pdf PDF
6 8Infection Control Permit_revised.pdf PDF
of0306.pdf PDF
VA0710.pdf PDF
6 2 Drawings 2 of 2.pdf PDF
6 3Specification CSI 121.pdf PDF
6 1 Drawings1 of 2.pdf PDF
Self Certification of Continuous Service.pdf PDF
Fingerprint Record Prep Sheet BLANK.PDF PDF
Show all 15

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

ACCURAY INCORPORATED

1310 Chesapeake Terrace

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E N

T

S

P D

R

E V

C

4.6.2020B

LOS ANGELES, CA

RADIXACT ™ TREATMENT DELIVERY SYSTEM

WEST LA VA

GENERAL

G-001 COVER PAGE

G-002 GENERAL NOTES

ARCHITECTURAL

A-101 FLOOR PLAN

A-102 RIGGING PLAN

A-201 INTERIOR ELEVATIONS

STRUCTURAL

S-101 FLOOR PLAN

S-301 INTERIOR SECTIONS

S-501 PLAN AND SECTION DETAILS

MECHANICAL

M-101 FLOOR PLAN

M-501 PLAN AND SECTION DETAILS

ELECTRICAL

E-101 FLOOR PLAN

E-111 REFLECTED CEILING PLAN

E-301 INTERIOR SECTIONS

E-501 PLAN AND SECTION DETAILS

E-502 ELECTRICAL DETAILS

E-503 ELECTRICAL DETAILS

E-901 INTERIOR PERSPECTIVES

NETWORK

N-101 FLOOR PLAN

EQUIPMENT

Q-501 PLAN AND SECTION DETAILS

Q-901 INTERIOR PERSPECTIVE

ACCURAY INCORPORATED

FOR REFERENCE ONLY

G-001

COVER SHEET

CONTROL

ROOM

MP

1'-9"

4'-10"

3'-4

4'-9"

TREATMENT

ROOM

7'-11"

31'-3

10'-2

18'-9"

NOTE 3

NOTE 2

NOTE

1 TYP.

3'-0"

3'-0"

1'-7

1'-7

19'-9"

15'-2"

2'-0"

3'-7"

DATA

SERVER ROOM

MECHANICAL

ROOM

TREATMENT

PLANNING

ROOM

(PRECISION)

MECHANICAL ROOM

6a/6b

ACCURAY INCORPORATED

1310 Chesapeake Terrace

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E

S

P D

R

A-101

ARCHITECTURAL

FLOOR PLAN

GENERAL NOTES

SHEET NOTES

LEGEND

A. THESE DRAWINGS ARE FOR

REFERENCE ONLY. THEY

ARE NOT TO REPLACE

CONSTRUCTION

DOCUMENTS AND

SHIELDING STUDIES BY

LICENSED PROFESSIONALS.

B. FINAL RADIATION SHIELDING

SHALL BE COMPLETED BY

THE PHYSICIST OF RECORD.

C. UNLESS OTHERWISE NOTED,

ALL MATERIALS SUPPLIED

AND INSTALLED ARE THE

RESPONSIBILITY OF THE

CUSTOMER.

D. REFER TO SITE PLANNING

GUIDE (SPG) FOR

ADDITIONAL

SPECIFICATIONS.

E. ALL STRUCTURAL

CALCULATIONS TO BE

COMPLETED BY

STRUCTURAL ENGINEER OF

RECORD.

F. ALL FINAL DIMENSIONS TO

BE FIELD VERIFIED.

1. MINIMUM 2 WALL MOUNTED

CCTV (FACILITY PROVIDED).

EXACT LOCATION TO BE

DETERMINED BY CUSTOMER.

CCTV AT FOOT OF PATIENT BED

TO BE MOTORIZED.

2. FINISHED WALLS ARE FOR

REFERENCE ONLY. REFER TO

CONTRACT DOCUMENTS FOR

DIMENSIONS/DETAILS.

3. CLEARANCE AREA FOR

SERVICE.

ISOCENTER

MP MAIN POWER DISCONNECT

[ ] RECOMMENDED

OFCI OWNER FURNISHED/

CONTRACTOR INSTALLED

# EQUIPMENT NAME IN LBS

1 RADIXACT™ TREAMENT DELIVERY SYSTEM 77.5 x 110 x 99 13000

2 RADIXACT™ SYSTEM COUCH 108.6 x 25.6 x 44 1100

3 POWER DISTRIBUTION UNIT 31 x 22 x 56 980

4 DORADO LASER SYSTEM 7 x 31.25 x 7.75 37

5 APOLLO LASER SYSTEM 4 x 4.25 x 8.5 2.6

6a POWER CONDITIONER (OFCI) 31.6 x18.9 x 73.7 640

6b FREQUENCY CONVERTER (OFCI) 51.5 x26.75x71.25 1502

7 AIR COMPRESSOR N/A N/A

8 iDMS DATA MANAGEMENT SYSTEM 38 x 24 x 59 900

10 OPERATOR STATION CONTROL 8.5 x 4.5 x 3 2.6

12 CONTROL WORKSTATION N/A N/A

13 TREATMENT PLANNING SYSTEM N/A N/A

14 INTERCOM SPEAKER SYSTEM 7 x 6.3 x 9.4 5

15 PRINTER N/A N/A

16 SYNCHRONY 28 x 7.5 x 3 10

OVERALL FLOOR PLAN

SCALE: 1/8" = 1'-0"

A1

DIMENSION FORMAT: FEET & INCHES

THE FOLLOWING ROOMS ARE NOT SITE SPECIFIC

AND ARE FOR REFERENCE ONLY:

- TREATMENT PLANNING (DOSIMETRY)

- MECHANICAL ROOM

NOTE 1,2,3,4 TYP.

DATA

SERVER ROOM

5'-11"

MIN.

5'-11"

MIN.

82"

MINIMUM

HEIGHT

FOR DOOR

OPENING.

4'-0" MIN.

6'-10" MIN.

7'- 9

8'-4"

9'- 5

6'- 9

7'- 9

9'- 5

3' - 5

ACCURAY INCORPORATED

1310 Chesapeake Terrace

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E

S

P D

R

A-102

ARCHITECTURAL

RIGGING PLAN

1. RIGGING PLAN IS FOR

REFERENCE ONLY. ACTUAL

ROUTE TO BE DETERMINED BY

OWNER.

2. GANTRY IS 77.5"(L) x 110"(W) x

99(H) AND WEIGHTS 13000 LBS.

3. FACILITY TO PROVIDE CLEAR

ACCESS PATH MIN. 5'-11" WIDE x

82" HIGH TO OUTSIDE DELIVERY

LOCATION.

4. ALL FLOOR SURFACES ALONG

THE RIGGING PATH MUST BE

FULLY SUPPORTED AND

CAPABLE OF WITHSTANDING

13,000 LBS.

OVERALL RIGGING PLAN

SCALE: 3/16" = 1'-0"

A1

RIGGING THROUGH MAZE- WHEELS

NOT TO SCALE

D1

RIGGING THROUGH DOOR- WHEELS

NOT TO SCALE

D2

GANTRY ON WHEELS

NOT TO SCALE

D3

GANTRY ON WHEELS

NOT TO SCALE

D5

GENERAL NOTES

SHEET NOTES

LEGEND

A. THESE DRAWINGS ARE FOR

REFERENCE ONLY. THEY

ARE NOT TO REPLACE

CONSTRUCTION

DOCUMENTS AND

SHIELDING STUDIES BY

LICENSED PROFESSIONALS.

B. FINAL RADIATION SHIELDING

SHALL BE COMPLETED BY

THE PHYSICIST OF RECORD.

C. UNLESS OTHERWISE NOTED,

ALL MATERIALS SUPPLIED

AND INSTALLED ARE THE

RESPONSIBILITY OF THE

CUSTOMER.

D. REFER TO SITE PLANNING

GUIDE (SPG) FOR

ADDITIONAL

SPECIFICATIONS.

E. ALL STRUCTURAL

CALCULATIONS TO BE

COMPLETED BY

STRUCTURAL ENGINEER OF

RECORD.

F. ALL FINAL DIMENSIONS TO

BE FIELD VERIFIED.

ISOCENTER

DIMENSION FORMAT: FEET & INCHES

ISO CENTER

7'-0"

4'-0"

I S

O

C E

N T

E R

[4'-0"]

4'-9"

THERMOSTAT (NOTE 1)

CCTV (NOTE 1)

[4'-0"]

I S

O

C E

N T

E R

3'-8

" A.F.F

1'-0"

1'-6"

INTERCOM

NOTE 2

CCTV (NOTE 1,4)

APOLLO LASER POWER SWITCH

EMO (NOTE 1)

2'-3

6'-7

" A.F.F

C

L

BEAM

ON

3'-8

" A.F.F

4'-0"

HEIGHT MAY VARY

I S

O

C E

N T

E R

1'-4"

3'-7

RAD WARNING LIGHT

(NOTE 1,3)

EMO (NOTE 1)

THERMOSTAT (NOTE 1)

3'-8

" A.F.F

4'-0"

HEIGHT MAY VARY

1'-4"

3'-7

2'-3

6'-7

" A.F.F

C

L

BEAM

ON

I S

O

C E

N T

E R

EMO (NOTE 1)

RAD WARNING LIGHT

(NOTE 1,3)

CCTV (NOTE 1)

ACCURAY INCORPORATED

1310 Chesapeake Terrace

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E

S

P D

R

A-201

ARCHITECTURAL

ELEVATION

PLAN ELEVATION NORTH

SCALE: 1/4" = 1'-0"

A1

PLAN ELEVATION EAST

SCALE: 1/4" = 1'-0"

A4

PLAN ELEVATION SOUTH

SCALE: NOT TO SCALE

C1

PLAN ELEVATION WEST

SCALE: NOT TO SCALE

C4

NOT SITE SPECIFIC NOT SITE SPECIFIC

NOT SITE SPECIFICNOT SITE SPECIFIC

1. OWNER FURNISHED,

CONTRACTOR INSTALLED.

EXACT LOCATION TO BE

DETERMINED BY CUSTOMER.

2. AIR SHUT OFF VALVE AND

QUICK DISCONNECT, LOCATION

MAY VARY. SEE SITE PLANNING

GUIDE (SPG) FOR ALTERNATE

LOCATIONS. TO CONNECT BACK

TO AIR COMPRESSOR WITH

FLOW RATE OF 15 scfm AT 90 psig

3. REFER TO LOCAL

REGULATIONS. COORDINATE

ELECTRICAL AS NECESSARY.

4. CCTV AT THE BACK OF THE

GANTRY AND NEIS NEEDS TO

BE INSTALLED IN THE AXIS OF

ISOCENTER AND RIGHT ABOVE

THE LASER PLATE.

AFF ABOVE FINISHED FLOOR

4 DORADO (RED) LASER

5 APOLLO (GREEN) LASER

16 SYNCHRONY

[ ] RECOMMENDED

GENERAL NOTES

SHEET NOTES

LEGEND

A. THESE DRAWINGS ARE FOR

REFERENCE ONLY. THEY

ARE NOT TO REPLACE

CONSTRUCTION

DOCUMENTS AND

SHIELDING STUDIES BY

LICENSED PROFESSIONALS.

B. FINAL RADIATION SHIELDING

SHALL BE COMPLETED BY

THE PHYSICIST OF RECORD.

C. UNLESS OTHERWISE NOTED,

ALL MATERIALS SUPPLIED

AND INSTALLED ARE THE

RESPONSIBILITY OF THE

CUSTOMER.

D. REFER TO SITE PLANNING

GUIDE (SPG) FOR

ADDITIONAL

SPECIFICATIONS.

E. ALL STRUCTURAL

CALCULATIONS TO BE

COMPLETED BY

STRUCTURAL ENGINEER OF

RECORD.

F. ALL FINAL DIMENSIONS TO

BE FIELD VERIFIED.

A1

S301

C1

S301

PB1

PB2

D1

Q501

A1

Q501

C

L

M-501

A4

TYP.

G

C

D

NOTE 2

NOTE 1,3,7 TYP.

8'-6

7'-11"

NOTE 5

NOTE 6

NOTE 3

6'-9

5'-6

4"

2'-5"

8"

NOTE 4

PB5 PB5

TRENCH

ACCURAY INCORPORATED

1310 Chesapeake Terrace

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E

S

P D

R

S-101

STRUCTURAL

FLOOR PLAN

1. THIS FLOOR PLAN IS CUT AT

THE SUB-BASE LEVEL, WITHIN

THE CONCRETE.

2. MOISTURE SENSOR WITH

ALARM. A FLOOR DRAIN THAT

CAN HANDLE 14 L/min. IS A

POSSIBLE ALTERNATIVE.

3. SEISMIC BRACKET MAY BE

REQUIRED, REFER TO CODE.

4. WHERE TRENCHES ARE NOT

POSSIBLE, CONDUIT IS AN

ACCEPTABLE ALTERNATIVE.

5. SUPPLEMENTAL AIR DUCT (SEE

M-501)

6. COMPRESSED AIR LINE

(SEE M-501)

7. REFER TO A1 ON S-501

6" AREA AROUND

EPOXY SET ANCHOR,

CLEAR OF ALL REBAR

TO DEPTH OF 10".

PB1 2'-0" x 1'-6" x 1'-0" PDU

PULL BOX

PB2 6" x 6" x 6" PATIENT

TABLE PULL BOX

PB5 1'-0" x 8" x 8" GANTRY

BULL BOX

SIGNAL (3"Ø) - FROM

PATIENT TABLE TO

BACK OF GANTRY.

(20'-0" MAX).

POWER (3"Ø) - FROM

PATIENT TABLE TO

BACK OF GANTRY.

(55'-0" MAX).

SIGNAL (1"Ø) - INTERCOM

TO BACK OF GANTRY (15'

MAX)

4 (4"Ø) FROM PB1 TO PB5

TREATMENT ROOM SUB-BASE PLAN

SCALE: 1/4" = 1'-0"

A1

C

D

SUB-BASE BEFORE POURING CONCRETE

NOT TO SCALE

A5

SUB-BASE AFTER POURING CONCRETE

NOT TO SCALE

D3

SHEET NOTES

LEGEND

A. THESE DRAWINGS ARE FOR

REFERENCE ONLY. THEY

ARE NOT TO REPLACE

CONSTRUCTION

DOCUMENTS AND

SHIELDING STUDIES BY

LICENSED PROFESSIONALS.

B. FINAL RADIATION SHIELDING

SHALL BE COMPLETED BY

THE PHYSICIST OF RECORD.

C. UNLESS OTHERWISE NOTED,

ALL MATERIALS SUPPLIED

AND INSTALLED ARE THE

RESPONSIBILITY OF THE

CUSTOMER.

D. REFER TO SITE PLANNING

GUIDE (SPG) FOR

ADDITIONAL

SPECIFICATIONS.

E. ALL STRUCTURAL

CALCULATIONS TO BE

COMPLETED BY

STRUCTURAL ENGINEER OF

RECORD.

F. ALL FINAL DIMENSIONS TO

BE FIELD VERIFIED.

ISOCENTER

DIMENSION FORMAT: FEET & INCHES

G

GENERAL NOTES

B3

Q501

D3

Q501

A3

Q501

A4

S501

D1

Q501

10"

8'-10

11'-0"

3'-8

3'-7

2'-3

5'-0"

19'-9"

NOTE 2

NOTE 1

12'-8"

7'-8"

NOTE 4

A5

Q501

A5

Q501

C4

S501

CONCRETE DECK

HEIGHT IS NOT SITE

SPECIFIC AND MUST

BE FIELD VERIFIED

A1

Q501

TYP.

TYP.

8'-10

6'-7

3'-8

3'-0"

3'-0"

10"

NOTE 3

NOTE 1

NOTE 2

ACCURAY INCORPORATED

1310 Chesapeake Terrace

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E

S

P D

R

S-301

INTERIOR

SECTIONS

1. SLOPE PIT TOWARD MOISTURE

SENSOR. A FLOOR DRAIN THAT

CAN HANDLE 14 L/min. IS A

POSSIBLE ALTERNATIVE.

2. WORKING SYSTEM CLEARANCE

3. THE FLOOR MUST BE LEVEL

ENOUGH TO MEET A REQUIRED

FLATNESS TOLERANCE OF ±

1/4" OVER 10'-10"

4. SYNCHRONY MOUNTING

LOCATION. KEEP 2' CLEAR FOR

INSTALLATION AND SERVICING

1 GANTRY AND EQUIPMENT

ENCLOSURES

2 TREATMENT COUCH

4 DORADO (RED) LASER

5 APOLLO (GREEN)

LASER

16 SYNCHRONY

PLAN WEST-EAST TREATMENT ROOM SECTION

SCALE: 1/4" = 1'-0" NOT SITE SPECIFIC

A1

PLAN NORTH/SOUTH TREATMENT ROOM SECTION

SCALE: 1/4" = 1'-0" NOT SITES SPECIFIC

C1

GENERAL NOTES

SHEET NOTES

LEGEND

A. THESE DRAWINGS ARE FOR

REFERENCE ONLY. THEY

ARE NOT TO REPLACE

CONSTRUCTION

DOCUMENTS AND

SHIELDING STUDIES BY

LICENSED PROFESSIONALS.

B. FINAL RADIATION SHIELDING

SHALL BE COMPLETED BY

THE PHYSICIST OF RECORD.

C. UNLESS OTHERWISE NOTED,

ALL MATERIALS SUPPLIED

AND INSTALLED ARE THE

RESPONSIBILITY OF THE

CUSTOMER.

D. REFER TO SITE PLANNING

GUIDE (SPG) FOR

ADDITIONAL

SPECIFICATIONS.

E. ALL STRUCTURAL

CALCULATIONS TO BE

COMPLETED BY

STRUCTURAL ENGINEER OF

RECORD.

F. ALL FINAL DIMENSIONS TO

BE FIELD VERIFIED.

ISOCENTER

DIMENSION FORMAT: FEET & INCHES

C5

FLOOR FINISH TOLERANCES

NOT TO SCALE

D5

NOT TO SCALE

DEFINITION OF FF25/FL25: COMES

FROM A FLOOR FINISH

TOLERANCES SYSTEM ON

F-NUMBER CODES WITHIN

ASTM-E1155, AC1 #117, & CSA

#23.1. THESE CODES GIVE

TOLERANCE REQUIREMENTS

USING ABBREVIATION F -

NUMBERS.

DEFINITION OF FF25: "FF"

REPRESENTS FLOOR FLATNESS &

THE NUMBER REPRESENTS 0.25"

[0.6] MAX. HEIGHT WITHIN A 10'

[304.8] PLANE. DETAIL "B" ON THIS

SHEET SHOWS AN EXAMPLE OF

FF25.

DEFINITION OF FL25: "FL"

REPRESENTS FLOOR LEVELNESS

& THE NUMBER REPRESENTS 0.25"

[0.6] MAX. HEIGHT ON A CURVE

DISTANCE OF 24" [610] WIDE. THE

MAX. HEIGHT SHOULD BE TAKEN

FROM THE POINT LOCATED AT

MID-CURVE OF A 12" [30.5]

DISTANCE. DETAIL "A" ON THIS

SHEET SHOWS AN EXAMPLE OF

FL25.

HATCH AREA REPRESENTS FLOOR

GRADE FLATNESS AND LEVELNESS

WHICH MUST HAVE FLATNESS (FF)

OF 25 FOR TOTAL AREA &

LEVELNESS (FL) OF 25 PER ASTM-E

1155.

*FLOOR FINISH TOLERANCES

SYSTEM ON F-NUMBER CODES

WITHIN ASTM-E1155 (US

STANDARD TEST METHOD FOR

DETERMINING (FF) FLOOR

FLATNESS AND FL FLOOR

LEVELNESS NUMBERS)

10'-0"

1/8"

1/8"

EQUIVALENT TO A SINGLE ± 1/8"

ALLOWABLE FLATNESS TOLERANCE

WHEN MEASURED BETWEEN ANY

TWO PERIPHERAL PTS. 10'-0" WITHIN

SHADED AREA.

* REFERENCE FF25:

F-NUMBER

SPECIFICATION

SYSTEM THAT

CODE FLOORING

FLATNESS

TOLERANCE.

0.000 MIDPOINT

PLANE

EXISTING

FLOORING

EQUIVALENT TO A SINGLE ± 1/4"

ALLOWABLE LEVELNESS TOLERANCE

WHEN MEASURED BETWEEN TWO POINT

OF A 24" CURVE DISTANCE. THE

TOLERANCE CANNOT BE GREATER THEN

1/4" WITHIN THE MID DISTANCE (12")

* REFERENCE FL25:

F-NUMBER SPECIFICATION SYSTEM

THAT CODE FLOORING LEVELNESS

TOLERANCE.

STRAIGHT EDGE

± 1/32"

ACCURACY

24"

1/4" MAX.

12"

0.000 MIDPOINT PLANE

EXISTING

FLOORING

FLOOR FLATNESS

FLOOR LEVELNESS NOTES

1'-2"

MIN.

1'-1

3'-1

2'-3"

1'-10"

SUPPLEMENTAL AIR DUCT

COMPRESSED AIR PIPE

NOTE 1

ELECTRICAL CONDUITS

1'-0" MIN1'-0" MIN

5'-7

3'-0"

" 8

1'-7

2'-2

3'-0

3'-8

C

L

ISO CENTER

2'-2

6'-2

1'-2" MIN.3'-1

3'-8

1'-1

D C

2'-5"

6'-11

NOTE 3

NOTE 4

2'-8

PB2

1'-9

1'-0

2'-0"

1'-0

1'-0"

2'-8"

G

2'-4"

PB5PB5

2'-3"

1'-1 3/8"

10" MIN @

HIGH PT.

PB2

1'-10" MIN.

6'-11 1/2"

1'-0" MIN.

I S

O C

E N

T E

R

10" MIN DEPTH

NOTE 1

1'-1 3/8"

1'-0 5/8"

3'-1

3'-1

NOTE 2

C

L

I S

O

C E

N T

E R

10" MIN @

HIGH PT.

10" MIN DEPTH

1'-2"

MIN.

NOTE 1

NOTE E

ACCURAY INCORPORATED

1310 Chesapeake Terrace

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E

S

P D

R

S-501

STRUCTURAL

DETAILS

PLAN ELEVATION NORTH

SCALE: NOT TO SCALE

A1

PLAN ELEVATION EAST

SCALE: NOT TO SCALE

A4

PLAN ELEVATION SOUTH

SCALE: NOT TO SCALE

D1

SEISMIC BRACKETS (IF APPLICABLE)

SCALE: NOT TO SCALE

D4

1. SLOPE PIT TOWARDS

MOISTURE SENSOR.

2. RADIUS FOR ROTATING

EQUIPMENT CLEARANCE.

3. GANTRY SUB-BASE: THE

GANTRY SUB-BASE MUST

SUPPORT THE GANTRY'S

WEIGHT AT 13,000 lbs AND

LOCAL REQUIREMENTS.

4. PATIENT TABLE SUB-BASE: THE

PATIENT TABLE SUB-BASE

MUST SUPPORT THE PATIENT

TABLE WEIGHT AT 1100 lbs AND

MEET LOCAL REQUIREMENTS.

6" AREA AROUND ANCHOR,

CLEAR OF ALL REBAR TO

DEPTH OF 10". 4 ANCHORS

REQUIRED FOR COUCH

AND 6 ANCHORS FOR

GANTRY.

SIGNAL (3"Ø) - FROM

PATIENT TABLE TO

BACK OF GANTRY.(20'MAX).

POWER (3"Ø) - FROM

PATIENT TABLE TO BACK

OF GANTRY.(55' MAX).

SIGNAL (1"Ø) - INTERCOM

TO BACK OF GANTRY (15'

MAX).

PATIENT TABLE

PULL BOX (FLOOR)

6" x 6" x 6" WITH COVER

PLATE FLUSH WITH

FINISHED FLOOR

TREATMENT VAULT

PULL BOX (FLOOR)

1'-0" x 8" x 8"

C

D

PB2

SEISMIC BRACKETS (IF APPLICABLE)

SCALE: NOT TO SCALE

D5

PLAN ELEVATION EAST

SCALE: NOT TO SCALE

C4

PB5

GENERAL NOTES

SHEET NOTES

LEGEND

A. THESE DRAWINGS ARE FOR

REFERENCE ONLY. THEY

ARE NOT TO REPLACE

CONSTRUCTION

DOCUMENTS AND

SHIELDING STUDIES BY

LICENSED PROFESSIONALS.

B. FINAL RADIATION SHIELDING

SHALL BE COMPLETED BY

THE PHYSICIST OF RECORD.

C. UNLESS OTHERWISE NOTED,

ALL MATERIALS SUPPLIED

AND INSTALLED ARE THE

RESPONSIBILITY OF THE

CUSTOMER.

D. REFER TO SITE PLANNING

GUIDE (SPG) FOR

ADDITIONAL

SPECIFICATIONS.

E. ALL STRUCTURAL

CALCULATIONS TO BE

COMPLETED BY

STRUCTURAL ENGINEER OF

RECORD.

F. ALL FINAL DIMENSIONS TO

BE FIELD VERIFIED.

ISOCENTER

DIMENSION FORMAT: FEET & INCHES

L5X3X

"X18" STEEL ANGLE (LEFT &

RIGHT SIDES) PROVIDED BY

ACCURAY WHEN REQUIRED BY

LOCAL REGULATORY AGENCY.

DSU

(TOP)

(FACE OF ANGLE)

(RIGHT SIDE)

L8X4X

"X36" STEEL ANGLE (LEFT &

RIGHT SIDES) PROVIDED BY

ACCURAY WHEN REQUIRED BY

LOCAL REGULATORY AGENCY.

iDMS

(TOP)

(FACE OF ANGLE)

(RIGHT SIDE)

2'-2 3/4"

6'-2 1/2"

M

I

N

GANTRY SUB-BASE

PATIENT

TABLE

SUB-BASE

GANTRY PIT

G

DIMENSION FORMAT: L x W x H

2'-5"

NOTE 7

D4

M501

A

M

NOTE 4

NOTE 2

NOTE 1

4'-0"

NOTE 5

4'-0"

VAULT

SUPPLY

NOTE 9

6'-0" - 8'-0"

6'-0" - 8'-0"

(3) GANTRY SUPPLY (NOTE 6)

NOTE 3

SUPPLY

RETURN

ACCURAY INCORPORATED

1310 Chesapeake Terrace

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E

S

P D

R

M-101

MECHANICAL

FLOOR PLAN

1. FACILITY SUPPLEMENTAL AIR

DUCT: ROUTE UNDER SLAB.

2. COMPRESSED AIR PIPE:

COPPER PIPE UNDER SLAB TO

AIR COMPRESSOR. USE

THICK-BODY OR WIDE-BODY

COPPER PIPE. USE 3/4" INSIDE

DIAMETER FOR UP TO 300' OR

1" FOR UP TO 500'.

3. ROOM THERMOSTAT

MAINTAINED AT 68°F OR

COOLER AND 30-60% RELATIVE

HUMIDITY.

4. AIR SHUT OFF CABINET.

LOCATION MAY VARY. SEE SITE

PLANNING GUIDE (SPG) FOR

ALTERNATE LOCATIONS.

5. DEDICATED EQUIPMENT

THERMOSTAT BEHIND THE

MACHINE ISOCENTER ON WALL

A 5'-0" A.F.F. MUST SUPPLY

AND MAINTAIN 65°F COOL AIR

AND 30-60% RELATIVE

HUMIDITY.

6. HEAT OUTPUT IN THE

TREATMENT VAULT CAN

REACH 51228 BTU/H (14.3 KW)

AND THE PDU GENERATES UP

TO 2391 BUT/H (0.7 KW)

7. MOISTURE SENSOR WITH

ALARM. A FLOOR DRAIN THAT

CAN HANDLE 14 L/min. IS A

POSSIBLE ALTERNATIVE.

8. INLINE AIR REGULATOR TO BE

PLACED IN TREATMENT ROOM.

SEE SITE PLANNING GUIDE

(SPG) FOR ALTERNATE

LOCATIONS.

9. ROOM TEMPERATURE

MAINTAINED AT 68-75°F AND

30-60% RELATIVE HUMIDITY.

POWER CONDITIONER/

FREQUENCY CONVERTER

TREATMENT ROOM SUB-BASE FLOOR PLAN

SCALE: 1/8" = 1'-0"

A1

6a/6b

AIR

COMPRESSOR

DRYER

TANK

NOTE:

THE FOLLOWING EQUIPMENT MUST BE PROVIDED

BY THE FACILITY AND MEET SPECIFICATIONS (SEE

SITE PLANNING GUIDE): AIR TANK, DRYER, & AIR

COMPRESSOR

6b

TO AIR REGULATOR IN

TREATMENT VAULT.

(SEE NOTE 8)

AIR TANK, 60 US GALLON MIN.

REQUIRED

AIR

MECHANICAL ROOM PLAN

SCALE: NOT TO SCALE

A4

TREATMENT ROOM MECHANICAL PLAN

SCALE: 1/8" = 1'-0"

C1

DATA SERVER ROOM HEATING, VENTILATION & AIR CONDITIONING

SCALE: 1/4" = 1'-0" NOT SITE SPECIFIC

D4

SUPPLY

RETURN

GENERAL NOTES

SHEET NOTES

LEGEND

A. THESE DRAWINGS ARE FOR

REFERENCE ONLY. THEY

ARE NOT TO REPLACE

CONSTRUCTION

DOCUMENTS AND

SHIELDING STUDIES BY

LICENSED PROFESSIONALS.

B. FINAL RADIATION SHIELDING

SHALL BE COMPLETED BY

THE PHYSICIST OF RECORD.

C. UNLESS OTHERWISE NOTED,

ALL MATERIALS SUPPLIED

AND INSTALLED ARE THE

RESPONSIBILITY OF THE

CUSTOMER.

D. REFER TO SITE PLANNING

GUIDE (SPG) FOR

ADDITIONAL

SPECIFICATIONS.

E. ALL STRUCTURAL

CALCULATIONS TO BE

COMPLETED BY

STRUCTURAL ENGINEER OF

RECORD.

F. ALL FINAL DIMENSIONS TO

BE FIELD VERIFIED.

A

N

T

R

Y

P

T

FACILITY SUPPLIED

SUPPLEMENTAL AIR

DUCT TERMINATE 2"

ABOVE FINISHED

FLOOR.

FACILITY-SUPPLIED

SUPPLEMENTAL

AIR DUCT

G

A

N

T

S

U

B

B

A

S

E

ELECTRICAL

FLOOR

CONDUIT

I

S

O

C

E

N

T

E

R

ELECTRICAL

FLOOR BOX

C

L

C

L

C L

C L

C L

G

C

PB5

D

PB5

COMPRESSED

AIR LINE

ISO CENTER

GANTRY PIT

CENTERLINE

6'-2 1/2"

6"

6"

2'-3"

7"

1'-4"

DRAIN/ MOISTURE

SENSOR

2'-5"

2'-3"

2'-8

2'-9"

PB2

COMPRESSED AIR

LINE

1'-3

1'-1

6"

2'-8"

ELECTRICAL CONDUIT

SUPPLEMENTAL AIR

PB5 PB5

ACCURAY INCORPORATED

1310 Chesapeake Terrace

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E

S

P D

R

M-501

MECHANICAL

DETAILS

SIGNAL (3"Ø) - FROM

PATIENT TABLE TO BACK OF

GANTRY (20'-0" MAX).

POWER (3"Ø) - FROM

PATIENT TABLE TO BACK OF

GANTRY (55'-0" MAX).

FLOOR PULL BOX

(1'-0" x 8" x 8")

PATIENT TABLE PULL BOX

(6" x 6" x 6")

TREATMENT ROOM SUB-BASE FLOOR PLAN (TRENCH OPTION)

NOT TO SCALE

A1

MECHANICAL ROOM PLAN

NOT TO SCALE

A4

BACK OF GANTRY AXON (CONDUIT OPTION)

NOT TO SCALE

D1

AIR SHUT OFF VALVE ACCESS PANEL

NOT TO SCALE

D4

C

D

PB5

1. COMPRESSED AIR SUPPLY SHUT

OFF VALVE.

2. 3/4" THICK BODY OR WIDE WALL

COPPER PIPE BACK TO AIR

COMPRESSOR WITH FLOW RATE

OF 15 SCFM AT 90 PSIG.

3. QUICK DISCONNECT FOR

SERVICE AIR HOSE.

4. BARBED FITTING FOR 1/2" INSIDE

DIAMETER AIR HOSE.

CONTRACTOR SUPPLIED AND

INSTALLED JUST ABOVE

FINISHED FLOOR.

5. SHUT-OFF VALVE TO POINT

DOWN IN CLOSED POSITION.

6. MESH PROTECTION ON TOP OF

SUPPLEMENTAL AIR DUCT.

7. FACILITY TO PROVIDE 55°F

SUPPLEMENTAL AIR FLOW

GREATER THAN 264 CFM.

INSTALL AN UNDER-SLAB 12"x10"

AIR DUCT OR A 12" ROUND (OR

EQUIVALENT MINIMUM

CROSS-SECTIONAL AREA) PVC

DUCT TO THE UNDERSIDE OF

THE GANTRY COVERS.

TO

TREATMENT

SYSTEM

FROM AIR

COMPRESSOR

1'-0"

1'-6"

NOTE 1

NOTE 2

PB2

GENERAL NOTES

SHEET NOTES

LEGEND

A. THESE DRAWINGS ARE FOR

REFERENCE ONLY. THEY

ARE NOT TO REPLACE

CONSTRUCTION

DOCUMENTS AND

SHIELDING STUDIES BY

LICENSED PROFESSIONALS.

B. FINAL RADIATION SHIELDING

SHALL BE COMPLETED BY

THE PHYSICIST OF RECORD.

C. UNLESS OTHERWISE NOTED,

ALL MATERIALS SUPPLIED

AND INSTALLED ARE THE

RESPONSIBILITY OF THE

CUSTOMER.

D. REFER TO SITE PLANNING

GUIDE (SPG) FOR

ADDITIONAL

SPECIFICATIONS.

E. ALL STRUCTURAL

CALCULATIONS TO BE

COMPLETED BY

STRUCTURAL ENGINEER OF

RECORD.

F. ALL FINAL DIMENSIONS TO

BE FIELD VERIFIED.

ISOCENTER

DIMENSION FORMAT: FEET & INCHES

NOTE 3

NOTE 4

COMPRESSED

AIR LINE (NOTE

2)

C

L

SUPPLEMENTAL

AIR DUCT.

(NOTE 6,7)

ELECTRICAL

TRENCH/

CONDUITS

C

L

C

L

7"

MACHINE

CENTERLINE

2"

NOTE 5

ROOM

B

A

F

EO

MP

RL

JB4

TS

TO DEDICATED

HVAC SYSTEM

S

NOTE 1, TYP.

NOTE 2, TYP.

NOTE 1, TYP.

TREATMENT

ROOM

NOTE 4

PB1

TRENCHTRENCH

D

C

1'-4"

2'-8"

PB2

2'-5"

2'-3"

PB5

PB5

RL

ES

RL

ES

ES IC

ES

DS

3 4

B A iDMS

ACCURAY INCORPORATED

1310 Chesapeake Terrace

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E

S

P D

R

E-101

ELECTRICAL

FLOOR PLAN

PDU PULL BOX (FLOOR). 2'-0"

x 1'-6" x 1'-0" WITH COVER

PLATE. LAST 6" OPEN TO

REMAIN OPEN. COVER TO

WITHSTAND WEIGHT OF PDU.

PATIENT TABLE PULL BOX

WITH DIVIDER

(6" x 6" x 6")

OPERATOR STATION PULL

BOX (WALL). CONTRACTOR

TO DETERMINE SIZE.

CONDUIT FLOOR PULL BOX

(1'-0" x 8" x 8")

PB5

1. CCTV WIRING RUN TO CONTROL

ROOM TO BE DETERMINED BY

CUSTOMER OR CONTRACTOR.

2. PULL STRINGS MUST BE

INCLUDED IN ALL NON-WIRED

CONDUITS.

3. 90 DEGREE ELBOWS OR

FLEXIBLE CONDUIT ARE NOT TO

BE USED IN CONDUIT RUNS.

4. SEE DETAIL A1 ON SHEET E-901

PB2

S

FACILITY PROVIDED

EO EMERGENCY OFF BUTTON

ES EMERGENCY STOP BUTTON

DS DOOR INTERLOCK SWITCH

RS RESET SWITCH(DOOR-LESS)

RL RADIATION WARNING LIGHT

MP MAIN POWER DISCONNECT

IC INTERCOM

TS THERMOSTAT

PC POWER CONDITIONER

FQ FREQUENCY CONVERTER

SIGNAL CONDUIT

POWER CONDUIT

NON-WIRED CONDUITS (EMPTY

CONDUITS)

A SIGNAL (3"Ø) OPERATOR

STATION TO PDU. (80' MAX).

B POWER (3"Ø) - OPERATOR

STATION TO PDU. (80' MAX).

C SIGNAL (3"Ø) - FROM PATIENT

TABLE TO BACK OF GANTRY 20'

MAX).

D POWER (3"Ø) - FROM PATIENT

ABLE TO BACK OF GANTRY 55'

MAX).

E SIGNAL (1

"Ø) - DORADO

ASERS TO PDU (65' MAX).

F PHYSICS PORT (4"Ø)

G SIGNAL (1"Ø) - INTERCOM TO

BACK OF GANTRY (15' MAX).

PB4

PB1

OVERALL FLOOR PLAN

SCALE: 1/4" = 1'-0"

A1

GENERAL NOTES

SHEET NOTES

LEGEND

A. THESE DRAWINGS ARE FOR

REFERENCE ONLY. THEY ARE NOT

TO REPLACE CONSTRUCTION

DOCUMENTS AND SHIELDING

STUDIES BY LICENSED

PROFESSIONALS.

B. FINAL RADIATION SHIELDING SHALL

BE COMPLETED BY THE PHYSICIST

OF RECORD.

C. UNLESS OTHERWISE NOTED, ALL

MATERIALS SUPPLIED AND

INSTALLED ARE THE

RESPONSIBILITY OF THE

CUSTOMER.

D. REFER TO SITE PLANNING GUIDE

(SPG) FOR ADDITIONAL

SPECIFICATIONS.

E. ALL STRUCTURAL CALCULATIONS

TO BE COMPLETED BY STRUCTURAL

ENGINEER OF RECORD.

F. ALL FINAL DIMENSIONS TO BE FIELD

VERIFIED.

ISOCENTER

CONDUIT

LOCATION NOTE

POWER (1/2"Ø) -

EMERGENCY

OFF BUTTON AT

OPERATOR

STATION TO MAIN POWER

DISCONNECT PANEL.

POWER (4"Ø) - MAIN

POWER DISCONNECT TO

PDU. SEE PAGE E-501 FOR

ELECTRICAL WIRING

OPTIONS

POWER (1"Ø) -

EMERGENCY STOP

SWITCHES TO PDU.

POWER (1"Ø) - RADIATION

WARNING LIGHTS TO PDU.

POWER (1"Ø) - DOOR

INTERLOCK SWITCH TO

PDU.

POWER (1"Ø) - DORADO

LASERS TO WALL SWITCH.

7 LOCAL GROUNDING

SIGNAL (1"Ø) INTERCOM TO

CONTROL ROOM

POWER 4" PDU TO POWER

CONDITIONER TO MP

(4) 4"Ø CONDUIT FROM JB1

TO JB5

WIRED CONDUIT LEGEND

APOLLO LASER WALL

POWER SWITCH

1 POWER OUTLET (NEMA

6-15R OR EQUAL)

2 POWER OUTLET (MATCH

LOCAL)

3 POWER OUTLET (220V

12AMP)

ROOM

TREATMENT

ROOM

PB3

E

E

S

PB1

NOTE 3

NOTE 1, TYP.

NOTE 2

NOTE 2

EH

12'-8"

NOTE 4

H

ACCURAY INCORPORATED

1310 Chesapeake Terrace

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E

S

P D

R

E-111

ELECTRICAL

CEILING PLAN

1. PULL STRINGS MUST BE

INCLUDED IN ALL NON-WIRED

CONDUITS.

2. MINIMUM 2 CEILING MOUNTED

CCTV WITH PAN, TILT AND

ZOOM @ 45° FROM ISOCENTER

(FACILITY PROVIDED). EXACT

LOCATION TO BE DETERMINED

BY CUSTOMER.

3. APOLLO (GREEN AND DORADO

(RED) LASER ALUMINUM

MOUNTING PLATED TO BE

PROVIDED AND INSTALLED BY

CONTRACTOR. ACCURAY

PROVIDES AND INSTALLS ALL

LASERS.

4. SYNCHRONY MOUNTING

LOCATION. KEEP 2' CLEAR FOR

INSTALLATION AND SERVICING

WIRED CONDUITS

6 POWER (1"Ø) - DORADO

LASERS TO PDU.

8 SIGNAL (3/4"Ø) INTERCOM TO

CONTROL ROOM

10 POWER (1"Ø) APOLLO LASER

NON-WIRED CONDUITS

(EMPTY CONDUITS)

E SIGNAL (1

"Ø) - DORADO

LASERS TO PDU. (65' MAX).

H (2"Ø)- SYNCHRONY CONDUIT

PB1 PDU PULL BOX (FLOOR).

2'-0" x 1'-6" x 1'-0"

PB3 DORADO POWER PULL

BOX (CEILING). SIZE TO BE

DETERMINED BY

CONTRACTOR.

SIGNAL CONDUIT

POWER CONDUIT

TREATMENT ROOM CEILING PLAN

SCALE: 1/4" = 1'-0"

A1

GENERAL NOTES

SHEET NOTES

LEGEND

A. THESE DRAWINGS ARE FOR

REFERENCE ONLY. THEY

ARE NOT TO REPLACE

CONSTRUCTION

DOCUMENTS AND

SHIELDING STUDIES BY

LICENSED PROFESSIONALS.

B. FINAL RADIATION SHIELDING

SHALL BE COMPLETED BY

THE PHYSICIST OF RECORD.

C. UNLESS OTHERWISE NOTED,

ALL MATERIALS SUPPLIED

AND INSTALLED ARE THE

RESPONSIBILITY OF THE

CUSTOMER.

D. REFER TO SITE PLANNING

GUIDE (SPG) FOR

ADDITIONAL

SPECIFICATIONS.

E. ALL STRUCTURAL

CALCULATIONS TO BE

COMPLETED BY

STRUCTURAL ENGINEER OF

RECORD.

F. ALL FINAL DIMENSIONS TO

BE FIELD VERIFIED.

ISOCENTER

DIMENSION FORMAT: FEET & INCHES

$ WALL POWER SWITCH

1 POWER OUTLET (NEMA 6-15R)

2 POWER OUTLET

(MATCH LOCAL)

PB3

S

TO BE

DETERMINED

BY FACILITY

2'-3

5'-0"

6'-7

3'-8

E

E

E

CCTV

G

INTERCOM

JB2

PB1

NOTE 2

PB5

I S

O

C E

N T

E R

CCTV

H

12'-8"

PB3

E

E

NOTE 1

NOTE 2

3'-8

6'-7

E

E

I S

O

C E

N T

E R

ACCURAY INCORPORATED

1310 Chesapeake Terrace

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E

S

P D

R

E-301

STRUCTURAL

FLOOR PLAN

WIRED CONDUITS

6 POWER (1"Ø) - DORADO

LASERS TO PDU.

8 SIGNAL (1"Ø) INTERCOM TO

CONTROL ROOM

10 POWER (1"Ø) APOLLO LASER

NON-WIRED CONDUITS

(EMPTY CONDUITS)

E SIGNAL (1

"Ø) - DORADO

LASERS TO PDU (65' MAX).

G SIGNAL (1/2"Ø) - INTERCOM TO

BACK OF GANTRY (15' MAX).

H (2" Ø) SYNCHRONY CONDUIT

JUNCTION BOXES

PB1 PDU PULL BOX

(FLOOR) 2' x x 1'-6" x 12"

WITH COVER PLATE. LAST

6" TO REMAIN OPEN.

PB2 COUCH PULL BOX

(FLOOR) 6" x 6" x 6"

PB3 DORADO POWER

PULL BOX (CEILING).

SIZE TO BE DETERMINED

BY CONTRACTOR.

PB5 FLOOR PULL BOX

1'-0" x 8" x 8"

PLAN WEST-EAST TREATMENT ROOM SECTION

SCALE: 1/4" = 1'-0" NOT SITE SPECIFIC

A1

PLAN NORTH-SOUTH TREATMENT ROOM SECTION

SCALE: 1/4" = 1'-0" NOT SITE SPECIFIC

C1

1 LOCATION TO BE DETERMINED

BY CONTRACTOR

2 MOISTURE SENSOR WITH

ALARM. A FLOOR DRAIN THAT

CAN HANDLE 14 L/min IS AN

ACCEPTABLE ALTERNATIVE

$ WALL POWER SWITCH

1 POWER OUTLET (NEMA 6-15R)

2 POWER OUTLET (MATCH

LOCAL)

GENERAL NOTES

SHEET NOTES

LEGEND

A. THESE DRAWINGS ARE FOR

REFERENCE ONLY. THEY

ARE NOT TO REPLACE

CONSTRUCTION

DOCUMENTS AND

SHIELDING STUDIES BY

LICENSED PROFESSIONALS.

B. FINAL RADIATION SHIELDING

SHALL BE COMPLETED BY

THE PHYSICIST OF RECORD.

C. UNLESS OTHERWISE NOTED,

ALL MATERIALS SUPPLIED

AND INSTALLED ARE THE

RESPONSIBILITY OF THE

CUSTOMER.

D. REFER TO SITE PLANNING

GUIDE (SPG) FOR

ADDITIONAL

SPECIFICATIONS.

E. ALL STRUCTURAL

CALCULATIONS TO BE

COMPLETED BY

STRUCTURAL ENGINEER OF

RECORD.

F. ALL FINAL DIMENSIONS TO

BE FIELD VERIFIED.

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E

S

P D

R

E-501

ELECTRICAL

DETAILS

C1

ELECTRICAL DIAGRAM - OPTION 1

NOT TO SCALE

A1

ELECTRICAL DIAGRAM - OPTION 2

NOT TO SCALE

380 - 480 VAC DELTA 3 PH

TO WYE 3 PH PRIMARY

TRANSFORMER SECONDARIES

Magnetic Contactor

A A

BB

CC

GG

B

G

N

C

AFACILITY POWER

A

B

C

G

*SITE SPECIFIC OPTION SUPPLIED BY ACCURAY

SUPPLIED BY FACILITY

G

GROUND

CONNECTION

FROM BUILDING

GROUND

380 - 480 VAC 3 PH

CANNOT HAVE

BACK-UP BATTERY

IN CONFIGURATION

(LOCATE OUTSIDE ROOM NEAR

OPERATOR STATION)

EPO

O

E

A T

M E

S

T E

M

NOT SITE SPECIFIC

NOT SITE SPECIFIC

*LOCAL GROUND TO BUILDING STEEL

OR GROUNDING ROD. TO BE

DETERMINED BY ELECTRICIAN. (NOTE 4)

1. IN CONFIGURATION OPTION 1,

WHERE FACILITY POWER

GOES DIRECTLY TO MP, THE

FCU MUST BE RESET AFTER

POWER EMERGENCY STOP.

2. CABLING FROM THE SUPPLY

TO PDU IS THE FACILITY'S

RESPONSIBILITY.

3. 70 AMP MAIN DISCONNECT

PANEL. FACILITY TO PROVIDE

MAIN POWER SUPPLY BOX

FROM DEDICATED LINE OR

POWER PROTECTION DEVICE.

4. GROUND TO HAVE EARTH

CONDUCTED IMPEDANCE OF

25 OHMS OR LESS. MUST

COMPLY WITH LOCAL CODES.

PDU POWER DISTRIBUTION UNIT

FCU FREQUENCY CONVERTER

UNIT (50Hz SITES ONLY)

MP MAIN DISCONNECT PANEL

PC POWER CONDITIONER

A PHASE A

B PHASE B

C PHASE C

G GROUND

N NEUTRAL

WIRED CONDUITS

1 POWER (1/2"Ø) - PDU TO

PC/FCU TO MP

NOTE 2

MP

*FCU/PC

PDU

380 - 480 VAC DELTA 3 PH

TO WYE 3 PH PRIMARY

TRANSFORMER SECONDARIES

Magnetic Contactor

A A

BB

CC

GG

B

G

N

C

AFACILITY POWER

A

B

C

G

SUPPLIED BY ACCURAY

SUPPLIED BY FACILITY

G

GROUND

CONNECTION

FROM BUILDING

GROUND

380 - 480 VAC 3 PH

CANNOT HAVE

BACK-UP BATTERY

IN CONFIGURATION

(LOCATE OUTSIDE ROOM NEAR

OPERATOR STATION)

EPO

E

A T

M E

S

T E

M

*LOCAL GROUND TO BUILDING STEEL

OR GROUNDING ROD. TO BE

DETERMINED BY ELECTRICIAN. (NOTE 4)

NOTE 2

MP

*FCU/PC

PDU

GENERAL NOTES

SHEET NOTES

LEGEND

A. THESE DRAWINGS ARE FOR

REFERENCE ONLY. THEY

ARE NOT TO REPLACE

CONSTRUCTION

DOCUMENTS AND

SHIELDING STUDIES BY

LICENSED PROFESSIONALS.

B. FINAL RADIATION SHIELDING

SHALL BE COMPLETED BY

THE PHYSICIST OF RECORD.

C. UNLESS OTHERWISE NOTED,

ALL MATERIALS SUPPLIED

AND INSTALLED ARE THE

RESPONSIBILITY OF THE

CUSTOMER.

D. REFER TO SITE PLANNING

GUIDE (SPG) FOR

ADDITIONAL

SPECIFICATIONS.

E. ALL STRUCTURAL

CALCULATIONS TO BE

COMPLETED BY

STRUCTURAL ENGINEER OF

RECORD.

F. ALL FINAL DIMENSIONS TO

BE FIELD VERIFIED.

DIMENSION FORMAT: FEET & INCHES

NOTE 3

NOTE 3

NOTE 1

*SITE SPECIFIC OPTION

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E

S

P D

R

E-502

ELECTRICAL

DETAILS

A1

DATA SERVER UNIT

NOT TO SCALE

(Power Distribution Unit)

PDU

TB1

L1

L2

L3

PDU

Pull Box

(PB1, in floor)

A B C GND

shunt trip coil

60 Hz

480 VAC, +/- 5%

80 Amp facility supplied control power

N.O.

One EO normally open switch located within reach of Operator Station. "Push to lock, twist to release" switch recommended.

- The system Emergency Off (EO) located on the power control panel is responsible for shutting off all electrical power from the PDU by tripping CB1 (on the PDU) with the exception of the

Un-interruptible Power Supply (UPS)

-The UPS remains on and provides power to the Data Receiver Server

(DRS) for data logging.

Facility Responsibility Accuray Responsibility

UPS

(Cluster/Optimizer)

Data Server Unit

UPS

DSU

UPS INPUT CORDS (10 foot / 3 meters)

USA domestic = Nema L6-20P to internal wired

International = site specific plug to IEC 320-C20

FACILITY POWER REQUIREMENTS FOR DSU

2 circuits @ 200-240 VAC, 20A, 50/60 Hz

G

X

Y

X

Y

Facility Responsibility Accuray Responsibility

Note: The DSU communication cable connections

(LC Duplex Fiber Optic and CAT5) under the responsibility of the facility are referenced in the Site

Planning Guide.

Facility Responsibility

PDU Junction Box

(PB1 in floor)

Ceiling Junction

Box

(JB5)

B

Y-RY-L

A

X receptacles for dorado lasers typical 5 places

NEMA 6-15R

USA Domestic

SITE SPECIFIC

International

LASER POWER CORDS (type SJT 18AWG)

USA domestic = Nema 6-20P to IEC320-C13

International = site specific plug to IEC320-C13

5X

Dorado Laser

Load = 240VAC, 30W each

Facility Responsibility Accuray Responsibility

Apollo Laser (ceiling)

Load = 120/240VAC, 10W

Apollo Laser (rear wall)

Load = 120/240VAC, 10W

LASER POWER CORDS

Plug end converts 120/240 VAC to 5VDC

Site Specific plug adaptors provided by Accuray

USA domestic = NEMA 5-15P

International = site specific

FACILITY POWER

SPST switch in treatment room controls both lasers.

N

L

POWER RECEPTACLES

USA domestic = 120V NEMA 5-15R

International = site specific

UPS

To DRS

Notes:

- Ground conductor to be the same diameter as phases. 4 AWG MIN.

- Neutral conductor not used

- PF = 0.90 at max level

-50 kVA max

120-240 VAC FACILITY NEUTRAL

120-240 VAC FACILITY LINE

PDU TB6

N

L

SOLID

STATE

RELAY

Accuray Responsibility

CAT5 PATCH PANEL

DSU Room

LC F.O. PATCH PANEL

DSU Room

1x

4x

Note: F.O. AND CAT5 PATCH

CABLES PROVIDED BY ACCURAY.

(3 METERS)

FC/PC NOT SHOWN

5x

5x

If NEMA version of receptacle is not available in your locale, contact the

Accuray Project

Manager to coordinate the use of locally approved receptacles and plugs, USA domestic = 120 VAC 60 Hz, 15A

International = 120-240 VAC 50/60 Hz, 15/16A iDMS™

CAT5 Receptacles 3x

Planning Room

LC F.O. PATCH PANEL

Control Room

4x

Duplex Pair 3x (or) 6 Strand Bundle

Extended Network Optional

OR

C1

EMERGENCY OFF AND PDU POWER INPUT

NOT TO SCALE

C4

DORADO LASER (RED)

NOT TO SCALE

A4

APOLLO LASER (GREEN)

NOT TO SCALE

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E

S

P D

R

E-503

ELECTRICAL

DETAILS

A1

STATUS INDICATORS

NOT TO SCALE

N.C.

Facility Responsibility Accuray Responsibility

Circuit Specifications

- 48-240 VAC, 50/60 Hz

- 24-10 AWG

- Min ON current = 40mA

- Max OFF current = 5mA

- Dedicated circuit protection recommended.

- Provide panel/breaker designation # for the facility indicator circuit. De-energizing label to be affixed by

Accuray to PDU access panel.

N.C.

N.C.

Additional E-STOP switches as needed wired normally closed in

SERIES.

PDU Pull Box

(PB1 in floor)

-TB5 connections made by Accuray

-Label wires "E-STOP"

24VDC

"Push to lock, twist to release"' switch

Min. 20 AWG shielded, twisted-pair wire

E-STOP prevents and/or terminates radiation by shutting off electrical power to the gantry Slip Rings and all other rotating components.

- Stops gantry rotation.

- Stops automatic couch motion.

- Stops Radiation Delivery Subsystem (RDS)

- Leads to an Unrecoverable State

Facility Responsibility Accuray Responsibility

PDU

PDU Pull Box

(PB1 in floor)

-Leave 6 feet (2 meters) of extra wire length in junction box.

- TB5 connections performed by Accuray

-Label wires

"DOOR SW"

24V

TB5

Wire size = 20 AWG min.

24 VDC, 3A

Use only momentary type switch.

Contacts are open when door is open

Facility Responsibility

Accuray Responsibility

PDU Pull Box

(PB1 in floor)

Indicator Specifications

- 8A max each

- Incandescent lamps and/or auditory indicators.

- Fluorescent lamps (40 to 200 W)

- Inductive or electronic ballasts are ok to use.

- LED lamps may require additional facility supplied coil type relay circuits.

Power

On

Room

Ready

Standby

TB5

* Radiation

On

240 VAC FACILITY LINE

240 VAC FACILITY LINE

240 VAC FACILITY LINE

240 VAC FACILITY LINE

L

L

L

L

240 VAC FACILITY

NEUTRAL

PDU

TB6

* Min. 1 lamp circuit required (Radiation On).

Other lamp circuits optional per local requirements.

Door Switch

PDU

TB6 wires connected by Accuray

** Status indicators located inside and/or outside treatment vault as needed.

RJ45

receptacle for Accuray

Speaker Unit on rear wall

RJ45

receptacle for Accuray

Intercom, Desk Unit

1x

Facility supplied CAT5 ethernet cable between Control Room and

Rear Wall of Vault

Label receptacles

"NEIS"

B1

ACCURAY INTERCOM (NEIS)

NOT TO SCALE

C1

FACILITY EMERGENCY STOP (E-STOP)

NOT TO SCALE

C4

DOOR INTERLOCK -TREATMENT ROOM

NOT TO SCALE

C2

FACILITY STANDBY

NOT TO SCALE

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E

S

P D

R

E-901

ELECTRICAL

DETAILS

C1

PDU FLOOR PULL BOX

NOT TO SCALE

A1

PATIENT TABLE FLOOR PULL BOX

NOT TO SCALE NOT SITE SPECIFIC

1. POWER DISTRIBUTION UNIT

2. COVER PROVIDED BY ACCURAY.

3. INCOMING POWER ENCASED IN 2" FLEXIBLE

CONDUIT ON LEFT SIDE OF PDU.

CONTRACTOR PROVIDED.

4. LAST 6" OF FLOOR BOX TO REMAIN OPEN

FOR WIRE CONNECTION. CONTRACTOR

PROVIDED.

5. TERMINATE COVER PLATE FLUSH WITH

FINISHED CONCRETE FLOOR. CONTRACTOR

TO PROVIDE ALL COVER PLATES.

6. INCOMING 230VAC POWER TO DORADO

LASERS

F

R

O

N

T

E

D

F

B

O

X

N

T

L

N

E

PATIENT

TABLE

FLOOR BOX

(PB2)

D

NOTE 5

C3 iDMS

NOT TO SCALE

C5

POWER DISTRIBUTION UNIT (PDU)

NOT TO SCALE

WEIGHT: 900 lb

18,000 BTUs/HOUR

F

R

O

N

T

5'-3

AIR

VENT

WEIGHT: 980 lb

3370 BTUs/HOUR

SEISMIC

BRACKETS (IF

APPLICABLE)

GENERAL NOTES

SHEET NOTES

LEGEND

A. THESE DRAWINGS ARE FOR REFERENCE

ONLY. THEY ARE NOT TO REPLACE

CONSTRUCTION DOCUMENTS AND

SHIELDING STUDIES BY LICENSED

PROFESSIONALS.

B. FINAL RADIATION SHIELDING SHALL BE

COMPLETED BY THE PHYSICIST OF RECORD.

C. UNLESS OTHERWISE NOTED, ALL

MATERIALS SUPPLIED AND INSTALLED ARE

THE RESPONSIBILITY OF THE CUSTOMER.

D. REFER TO SITE PLANNING GUIDE (SPG) FOR

ADDITIONAL SPECIFICATIONS.

E. ALL STRUCTURAL CALCULATIONS TO BE

COMPLETED BY STRUCTURAL ENGINEER OF

RECORD.

F. ALL FINAL DIMENSIONS TO BE FIELD

VERIFIED.

ISOCENTER

JUNCTION BOXES

PB1 PDU PULL BOX (FLOOR). 2'-0" x 1'-2" x 1'-0"

WITH COVER PLATE. LAST 6" OPEN TO

REMAIN OPEN.

PB2 PATIENT TABLE PULL BOX (FLOOR). 6" x

6" x 6" WITH COVER PLATE FLUSH

TO FINISHED FLOOR.

4'-8"

1'-10"

2'-7"

FRONT

NOTE 1

NOTE 3

NOTE 6

4 3 3

E

4 3 2 B

A

NOTE 4

PB1

WIRED CONDUITS

POWER (4"Ø)- MAIN POWER DISCONNECT TO

PDU

POWER (1"Ø)- EMERGENCY STOP SWITCHES

TO PDU

POWER (1"Ø)- RADIATION WARNING LIGHTS

TO PDU

POWER (1"Ø)- DOOR INTERLOCK SWITCH TO

PDU

POWER (1"Ø)- DORADO LASERS TO PDU

LOCAL GROUNDING

SIGNAL (3"Ø)- OPERATOR STATION TO PDU.

80' MAX

POWER (3"Ø)- OPERATOR STATION TO PDU.

80' MAX.

SIGNAL (3"Ø)- FROM PATIENT TABLE TO

BACK OF GANTRY.

20' MAX

POWER (3"Ø)- FROM PATIENT TABLE TO

BACK OF GANTRY TO PDU. 55' MAX.

SIGNAL (1 1/2"Ø)- DORADO LASER TO PDU.

65' MAX.

(2" Ø) -SYNCHRONY CONDUIT

(4"Ø)- FROM PB1 TO PB5 (CONDUIT OPTION

ONLY)

A

B

C

D

E

NON-WIRED CONDUITS

A3

NOT USED

NOT TO SCALE NOT SITE SPECIFIC

H

H

WEIGHT: 1,502 lb

12,000 BTU/HOUR

ONLY FOR

FACILITIES

WITH 50Hz POWER

5'-11 1/4"

A5

FREQUENCY CONVERTER UNIT (FCI)

NOT TO SCALE NOT SITE SPECIFIC

TREATMENT ROOM

CONTROL ROOM

MECHANICAL

ROOM

TREATMENT

PLANNING ROOM

DATA SERVER

ROOM

ACCURAY INCORPORATED

1310 Chesapeake Terrace

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E

S

P D

R

NOTE 5

NOTE 3

NOTE 2

LC

= LC CONNECTOR

NOTE 1

LC LC LC

N-101

NETWORK

FLOOR PLAN

A1

OVERALL FLOOR PLAN

SCALE: 3/16" = 1'-0" NOT SITE SPECIFIC

LCLC LC

NOTE 4

PHONE

PHONE

D1

TREATMENT PLANNING ISOMETRIC

NOT TO SCALE

ALL ITEMS IN PLANNING STATION

ARE ON FACILITY POWER SUPPLY.

D5

CONTROL ROOM ISOMETRIC

NOT TO SCALE

D3

DATA SERVER ROOM ISOMETRIC

NOT TO SCALE

1'-0"

1. CAT6E CABLE (OR GREATER)

FROM WALL MOUNTED

SPEAKER TO CONTROL ROOM.

2. CAT6E CABLES (OR GREATER)

FROM DATA SERVER ROOM TO

TREATMENT PLANNING.

3. FIBER OPTIC NETWORK

CONNECTIONS FROM DATA

SERVER ROOM TO CONTROL

ROOM.

4. CAT6E CABLES (OR GREATER)

FROM DATA SERVER ROOM TO

FACILITY NETWORK.

5. DIGITAL INTERNET DATA JACK.

6. FIBER OPTICS REQUIRED OVER

300' ALL DIRECT ROUTE.

7. BACKUP COPPER CAT6E

CABLES (OR GREATER)

= DATA OUTLET

= PHONE

OUTLET/INTERCOM

GENERAL NOTES

SHEET NOTES

LEGEND

A. THESE DRAWINGS ARE FOR

REFERENCE ONLY. THEY

ARE NOT TO REPLACE

CONSTRUCTION

DOCUMENTS AND

SHIELDING STUDIES BY

LICENSED PROFESSIONALS.

B. FINAL RADIATION SHIELDING

SHALL BE COMPLETED BY

THE PHYSICIST OF RECORD.

C. UNLESS OTHERWISE NOTED,

ALL MATERIALS SUPPLIED

AND INSTALLED ARE THE

RESPONSIBILITY OF THE

CUSTOMER.

D. REFER TO SITE PLANNING

GUIDE (SPG) FOR

ADDITIONAL

SPECIFICATIONS.

E. ALL STRUCTURAL

CALCULATIONS TO BE

COMPLETED BY

STRUCTURAL ENGINEER OF

RECORD.

F. ALL FINAL DIMENSIONS TO

BE FIELD VERIFIED.

ISOCENTER

DIMENSION FORMAT: FEET & INCHES

PHONE

NOTE 5

NOTE 6,

7 TYP.

DIRECT WIRING TO

DOSIMETRY (NOTE 2)

8 RJ45 WALL

JACK (CAT6 MIN.)

TO HOSPITAL

NETWORK (NOTE 4)

8 RJ45 WALL

JACK (CAT6

MIN.) FOR

ACCESS TO

HOSPITAL

NETWORK

DIRECT WIRING TO

CONTROL STATION

(NOTE 3,6 & 7)

WALL RJ45 CONNECTOR

TO HOSPITAL NETWORK

WITH INTERNET ACCESS

WIRED BY ACCURAY

WIRED BY FACILITY

NOTE 1

NOTE 6

65°

CABINETSCABINETS

LASER BEAM

LASER

DORADO MOUNTING PLATE

6 x

1'-4"

3'-7

1"

1"

4 x

1'-6"

1'-0"

1"

1"

FINISHED CEILING

OPENING VARIES

(COORDINATE WITH

ACCURAY STAFF).

RECEPTACLE

CONNECTED TO

WALL SWITCH

1'-6"

10"

1'-4"

10"

FINISHED CEILING

RECEPTACLE

CONNECTED TO PDU

OPENING VARIES

(COORDINATE WITH

ACCURAY STAFF).

NOT TO SCALE

Q-501

EQUIPMENT

DETAILS

A1

APOLLO WALL LASER - SECION

NOT TO SCALE

A3

DORADO WALL LASER- SECTION

NOT TO SCALE

A5

DORADO MOUNTING PLATE

NOT TO SCALE

B1

APOLLO MOUNTING PLATE

NOT TO SCALE

D1

3/8" ALUMINUM OR 1/4"

STEEL MOUNTING

PLATE, TYP.

SPACING BETWEEN

PLATE AND RIGID

SURFACE (WALL)

1/2"

POWER CONDUIT

(NOTE 4), TYP.

NOTE 6

3/8" ALUMINUM OR 1/4"

STEEL MOUNTING

PLATE, TYP.

SPACING BETWEEN

PLATE AND RIGID

SURFACE (WALL)

POWER CONDUIT

(NOTE5), TYP.

1/2"

SIGNAL CONDUIT, TYP.

1/2"

NOTE 6

SHEET NOTES

LEGEND

A. THESE DRAWINGS ARE FOR

REFERENCE ONLY. THEY ARE NOT

TO REPLACE CONSTRUCTION

DOCUMENTS AND SHIELDING

STUDIES BY LICENSED

PROFESSIONALS.

B. FINAL RADIATION SHIELDING

SHALL BE COMPLETED BY THE

PHYSICIST OF RECORD.

C. UNLESS OTHERWISE NOTED, ALL

MATERIALS SUPPLIED AND

INSTALLED ARE THE

RESPONSIBILITY OF THE

CUSTOMER.

D. REFER TO SITE PLANNING GUIDE

(SPG) FOR ADDITIONAL

SPECIFICATIONS.

E. ALL STRUCTURAL CALCULATIONS

TO BE COMPLETED BY

STRUCTURAL ENGINEER OF

RECORD.

F. ALL FINAL DIMENSIONS TO BE

FIELD VERIFIED.

1. RECEPTACLE CONNECTED TO

WALL SWITCH.

2. RECEPTACLE CONNECTION TO

POWER DISTRIBUTION UNIT.

3. OPENING VARIES. TO BE

COORDINATED WITH SITE

PLANNING TEAM AT ACCURAY.

4. APOLLO LASER POWER

REQUIRES LOCAL STANDARD

OUTLETS DUE TO ADAPTOR

PROVIDED WITH LASER.

5. DORADO LASER POWER

REQUIRES NEMA 6-15R OUTLET.

6. PROVIDE ADDITIONAL SUPPORT

AS NECESSARY TO RIGID

SERVICE.

SIGNAL (1

"Ø) - DORADO

LASERS TO PDU (65' MAX).

E

GENERAL NOTES

APOLLO CEILING LASER - SECTION

NOT TO SCALE

D3

DORADO CEILING LASER - SECTION

NOT TO SCALE

B3

EE

ACCURAY INCORPORATED

1310 Chesapeake Terrace

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E

S

P D

R

D

R

A

D

O

N

O

DORADO

(NOTE 2)

DORADO

D O

R A

D O

N

O T

E

APOLLO

(NOTE 1)

APOLLO

(NOTE 1)

6'-7

3'-8

2'-11

TO THE

ROOM

POWER

SWITCH

JB1

RECEPTACLE (220V)

RECEPTACLES

RECEPTACLES

GANTRY PIT

LASERS PLATES HAVE TO

BE MOUNTED AT 250mm

ABOVE THE FINISHED

CEILING

JB3

5'-0"

CENTERLINE

DORADOS (WALLS) AND APOLLO (CEILING)

CENTERLINE

APOLLO &

DORADOS

(WALLS)

2'-3

JB2

E

E

E

E

C E

N T

E R

L I N

E

A P

O L L O

S

C

E I L I N

W

A L L

CENTERLINE

DORADOS

(WALLS)

CENTERLINE

DORADO (CEILING)

JB5 JB5

E

MATCH

LOCAL

MATCH

LOCAL

(220V)

(220V)

D O

R A

D O

N

O T

E

(220V)

(220V)

ACCURAY INCORPORATED

1310 Chesapeake Terrace

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E

S

P D

R

Q-901

EQUIPMENT

PERSPECTIVE

CONDUIT LOCATIONS

8"

MIN.

SURFACE-MOUNTED GUARD

PROTECTION

10"

MAX.

8"

MIN.

CUT SLOT AFTER THE

LASER HAS BEEN

INSTALLED.

1. APOLLO LASER:

CONNECTED IN SERIES WITH

STANDARD LOCAL DUPLEX

OUTLET ATTACHED TO FACILITY

POWER. NO SIGNAL CONDUIT

REQUIRED.

2. DORADO LASER:

RECEPTACLE NEMA 6-15R

OUTLET AND 1

" SIGNAL

CONDUIT REQUIRED BACK TO

PULL BOX 1 (PB1).

3. ALUMINUM LASER MOUNTING

PLATES ARE SUPPLIED AND

INSTALLED BY CONTRACTOR.

ACCURAY SUPPLIES AND

INSTALLS ALL LASERS.

SURFACE-MOUNTED

CABINET PROTECTION

A1

TREATMENT ROOM PERSPECTIVE

NOT TO SCALE NOT SITE SPECIFIC

D1

DORADO WALL LASER - OPTION 1

NOT TO SCALE

D3

DORADO WALL LASER - OPTION 2

NOT TO SCALE

D5

RECESSED LASER ALCOVE

NOT TO SCALE

B5

RECESSED LASER ALCOVE

NOT TO SCALE

A5

RECESSED LASER ALCOVE

NOT TO SCALE

-OWNER TO PROVIDE POWER

-MOUNT TO 12" x 18" x 3/8" ALUMINUM PLATE

-PROVIDE LOCAL POWER OUTLET

-2 PROVIDED/INSTALLED BY ACCURAY

-OWNER TO PROVIDE POWER & SIGNAL

-MOUNT TO 16" x 42

" x 3/8" ALUMINUM PLATE

-PROVIDE NEMA 6-20R POWER OUTLET

-5 PROVIDED/INSTALLED BY ACCURAY

(2) x 1 1/2"Ø

SIGNAL

CONDUIT (1

IN/ 1 OUT)

CABINETS,

OWNER

FURNISHED,

CONTRACTOR

INSTALLED

(2) x 1 1/2"Ø

SIGNAL

CONDUIT (1

IN/ 1 OUT)

CONDUIT LOCATIONS

RECESSED BOXSURFACE MOUNTED BOX

PULL BOXES

PB1 PDU PULL BOX (FLOOR)

2'-0" x 1'-6" x 1'-0" WITH

COVER PLATE. LAST 150

OPEN TO REMAIN OPEN.

PB2 PATIENT TABLE PULL

BOX (FLOOR) 6" x 6" x 6"

WITH COVER PLATE FLUSH

TO FINISHED FLOOR.

PB3 DORADO POWER PULL

BOX (CEILING). SIZE TO BE

DETERMINED BY

CONTRACTOR.

PB5 FLOOR PULL BOX

1'-0" x 8" x 8"

EMPTY DATA

CONDUIT 1 1/2"Ø

(min)

WIRED POWER

CONDUIT WITH

OUTLET AT

DORADO LASER

SIDE.

WIRED POWER

CONDUIT WITH

OUTLET AT

APOLLO LASER

SIDE, POWERED

FROM THE

FACILITY VIA THE

WALL SWITCH.

OUTLET

MEETING LOCAL

ELECTRICAL

REQUIREMENTS.

E

GENERAL NOTES

SHEET NOTES

LEGEND

A. THESE DRAWINGS ARE FOR

REFERENCE ONLY. THEY

ARE NOT TO REPLACE

CONSTRUCTION

DOCUMENTS AND

SHIELDING STUDIES BY

LICENSED PROFESSIONALS.

B. FINAL RADIATION SHIELDING

SHALL BE COMPLETED BY

THE PHYSICIST OF RECORD.

C. UNLESS OTHERWISE NOTED,

ALL MATERIALS SUPPLIED

AND INSTALLED ARE THE

RESPONSIBILITY OF THE

CUSTOMER.

D. REFER TO SITE PLANNING

GUIDE (SPG) FOR

ADDITIONAL

SPECIFICATIONS.

E. ALL STRUCTURAL

CALCULATIONS TO BE

COMPLETED BY

STRUCTURAL ENGINEER OF

RECORD.

F. ALL FINAL DIMENSIONS TO

BE FIELD VERIFIED.

ISOCENTER

DIMENSION FORMAT: FEET & INCHES

RECEPTACLE

RECEPTACLE

Sunnyvale, CA 94089

SHEET NUMBER

PROJECT INFORMATION

DRAWING INFORMATION

A

B

C

D

E

2 3 4 5 6

4.1.2020A

WEST LA VA

TOMOTHERAPY RADIXACT

LOS ANGELES, CA

REGIONAL PROJECT MANAGER

BETHANY NEIGEBAUER

SITE PLANNING COORDINATOR

JASON LENTZ

PAGE NUMBER

Q U

A L

I T

Y

S

Y S

T E

M

D

O C

U M

E

S

P D

R

G-002

GENERAL

NOTES

GENERAL

NOTES

Sheets and Views
Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-G-001
Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-A-101
Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-A-102
Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-A-201
Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-S-101
Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-S-301
Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-S-501
Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-M-101
Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-M-501
Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-E-101
Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-E-111
Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-E-301
Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-E-501
Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-E-502
Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-E-503
Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-E-901
Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-N-101
Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-Q-501
Los Angeles, CA - West LA VA - Radixact Full Set_462020_B-Q-901