6 2 Drawings 2 of 2.pdf
PDF 37 MB Posted
- Attached to
- Replace Linear Accelerator at Greater Los Angeles VAHCS Federal contract opportunity
- Solicitation number
- 36C25821B0004
About this file
This solicitation is for replacing a linear accelerator at the Greater Los Angeles Veterans Affairs Health Care System campus in West Los Angeles. The contractor will demolish the existing linear accelerator, prepare the site for installation of a new linear accelerator and Accuray Radixact system to be furnished by the equipment provider. Renovations and construction services are required, including converting a room to a control room, adding a mechanical room and data server room, replacing ceiling, flooring, and rewiring. New lead shielding, air compressor and dryer system, HVAC units, ductwork and ceiling-mounted patient lift must also be installed. Electrical power and controls for the lift, linear accelerator and associated equipment must be provided. Structural bracing, foundation and supports meeting VA, city and state seismic requirements are included. The contractor must coordinate with equipment providers and ensure all existing equipment is removed. The expected construction period is not specified. The soliciting agency is the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 22.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Sign in Sheet for 11 Jan Site Walk Linear Accelerator.pdf | ||
| 36C25821B0004 0002.pdf | ||
| 36C25821B0004 0001 RFI Amendment.pdf | ||
| 6 8Infection Control Permit_revised.pdf | ||
| of0306.pdf | ||
| VA0710.pdf | ||
| 36C25821B0004.pdf | ||
| Wage Rate Determination CA20200022.docx | DOCX document | |
| Atch 1 Self Performance Calculations.pdf | ||
| 6 4Radixact Full Set.pdf | ||
| Contractor Background Investigation Request Form (May 2019).pdf | ||
| 6 3Specification CSI 121.pdf | ||
| 6 1 Drawings1 of 2.pdf | ||
| Self Certification of Continuous Service.pdf | ||
| Fingerprint Record Prep Sheet BLANK.PDF |
Show all 15
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
B500 KEY PLAN
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
FULLY SPRINKLERED
Office of Construction and Facilities Management
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT/ENGINEER OF RECORD A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
FD-100-1
FIRE PROTECTION DEMO RCP
PLAN - GROUND FLOOR
TRIPLE C - THE A/E GROUP
4401 ATLANTIC AVENUE, SUITE 200
LONG BEACH, CA 90807
562.452.7175
MAT PERKINS
REPLACE LINEAR ACCELERATOR
11301 Wilshire Blvd, West Los Angeles, CA 90073
691-CSI-121
07/29/2020 Checker Author
100% CONSTRUCTION
DOCUMENTS
SCALE: 1/4" = 1'-0"
1 GROUND FLOOR DEMOLITION RCP PLAN
1 (E) PENDENT SPRINKLER HEAD TO BE DEMOLISHED.
(E) SIDE WALL SPRINKLER HEAD TO BE DEMOLISHED.
THE HATCHED AREA IS NOT IN SCOPE.
1. ALL EXISTING SPRINKLER HEADS TO BE DEMOLISHED PRIOR
TO DEMOLITION OF CEILING.
Revisions: Date:
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T A
T E O F
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T A
T E O F
No. M35363
M
I K H A I L F U K
S
ME C H A N I C A L
06/30/201906/30/2021
REVISION 1 11/02/20
REVISION 2 11/24/20
7/29/20
Expires 6.30.16Expires: 6/30/1821
7/2 9/2
7/2 9/2
7/2 9/2
7/2 9/2
7/2 9/2
7/2 9/2
7/2 9/2
7/2 9/2
7/2 9/2
7/2 9/2
7/2 9/2
7/2 9/2
7/2 9/2
SYMBOL DESCRIPTION
NOTE CALLOUT
DETAIL CALLOUT
- NUMBER ON TOP DENOTES DETAIL NUMBER
- NUMBER ON BOTTOM DENOTES SHEET DETAIL IS SHOWN
MECHANICAL EQUIPMENT CALLOUT, SEE MECHANICAL PLANS FOR
EXACT LOCATION AND REQUIREMENTS
SECTION CALLOUT
FEEDER CALLOUT
EXISTING FEEDER CALLCOUT
NEW LINEWORK
EXISTING LINEWORK
DEMOLISHED LINEWORK
CONDUIT CONCEALED IN WALL OR ABOVE CEILING
CONDUIT EXPOSED
CONDUIT CONCEALED UNDERGROUND OR BELOW FLOOR
CONDUIT EMERGENCY
CONDUIT TURNED UP
CONDUIT CAPPED
BRANCH CIRCUIT HOMERUN TO PANELBOARD AND CIRCUITS AS
INDICATED
3/4" CONDUIT, TICK MARKS INDICATE QUANTITY OF #12 AWG WIRES
(UNLESS NOTED OTHERWISE, NO MARKS INDICATES 2#12 & 1#12 GND
WIRES)
- SMALL MARK DENOTES HOT WIRE
- LARGE MARK DENOTES NEUTRAL WIRE
- DIAGONAL DENOTES GROUND WIRE
GENERATOR
SWITCH
CIRCUIT BREAKER
2-WAY SWITCH, TRANSFER SWITCH
FUSE
TRANSFORMER
GROUND CONNECTION
MOTOR - SINGLE PHASE FRACTIONAL OR INTEGRAL HORSEPOWER
METER
ELECTRONIC CIRCUIT MONITOR
480V DRAWOUT BREAKER
VARIABLE FREQUENCY DRIVE
FUSED DISCONNECT SWITCH
NON-FUSED DISCONNECT SWITCH
COMBINATION STARTER/DISCONNECT SWITCH
SWITCH MOTOR RATED
SPLICE
TERMINATION
EXISTING TERMINATION
MEDIUM VOLTAGE - AIR CIRCUIT BREAKER
DRAWOUT BREAKER
MEDIUM VOLTAGE FUSED
DISCONNECT SWITCH
MEDIUM VOLTAGE MODULAR SPLICE
MEDIUM VOLTAGE EXISTING MODULAR SPLICE
2X4 LIGHT FIXTURE - UPPER CASE LETTER INDICATES LIGHT FIXTURE
CALLOUT. LOWER CASE LETTER INDICATES LIGHTING CONTROL ZONE.
2X4 EMERGENCY LIGHT FIXTURE FED FROM GENERATOR/ INVERTER/
BATTERY BACKUP
2X2 LIGHT FIXTURE - UPPER CASE LETTER INDICATES LIGHT FIXTURE
CALLOUT. LOWER CASE LETTER INDICATES LIGHTING CONTROL ZONE.
2X2 EMERGENCY LIGHT FIXTURE FED FROM GENERATOR/ INVERTER/
BATTERY BACKUP
LINEAR LIGHT FIXTURE, DIMENSIONS PER PLANS - UPPER CASE LETTER
INDICATES LIGHT FIXTURE CALLOUT. LOWER CASE LETTER INDICATES
LIGHTING CONTROL ZONE.
EMERGENCY LINEAR LIGHT FIXTURE, DIMENSIONS PER PLANS - LIGHT
FIXTURE FED FROM GENERATOR/ INVERTER/ BATTERY BACKUP
LINEAR PENDANT LIGHT FIXTURE, DIMENSIONS PER PLANS - UPPER
CASE LETTER INDICATES LIGHT FIXTURE CALLOUT. LOWER CASE
LETTER INDICATES LIGHTING CONTROL ZONE.
TRACK LIGHTING - UPPER CASE LETTER INDICATES LIGHT FIXTURE
CALLOUT. LOWER CASE LETTER INDICATES LIGHTING CONTROL ZONE.
UNDERCABINET / COVE FIXTURE - UPPER CASE LETTER INDICATES
LIGHT FIXTURE CALLOUT. LOWER CASE LETTER INDICATES LIGHTING
CONTROL ZONE.
SYMBOL DESCRIPTION
DOWNLIGHT FIXTURE - UPPER CASE LETTER INDICATES LIGHT FIXTURE
CALLOUT. LOWER CASE LETTER INDICATES LIGHTING CONTROL ZONE.
EMERGENCY DOWNLIGHT FIXTURE FED FROM GENERATOR/ INVERTER/
BATTERY BACKUP
PENDANT LUMINAIRE - UPPER CASE LETTER INDICATES LIGHT FIXTURE
CALLOUT. LOWER CASE LETTER INDICATES LIGHTING CONTROL ZONE.
WALLWASH LIGHT FIXTURE - UPPER CASE LETTER INDICATES LIGHT
FIXTURE CALLOUT. LOWER CASE LETTER INDICATES LIGHTING
CONTROL ZONE.
WALL MOUNTED LIGHT FIXTURE - UPPER CASE LETTER INDICATES
LIGHT FIXTURE CALLOUT. LOWER CASE LETTER INDICATES LIGHTING
CONTROL ZONE.
EMERGENCY WALL MOUNTED LIGHT FIXTURE FED FROM GENERATOR/
INVERTER/ BATTERY BACKUP
IN GRADE LUMINAIRE
PATHWAY LUMINAIRE
LANDSCAPE FIXTURE
EXIT LIGHT FIXTURE WITH DIRECTIONAL ARROWS AS INDICATED.
SHADED SIDE DENOTES NUMBER OF FACES
JUNCTION BOX
PHOTOCELL FOR EXTERIOR APPLICATIONS
DAYLIGHT SENSOR - CEILING MOUNTED
RELAY
EMERGENCY RELAY UL 924 COMPLIANT
MOTION SENSOR - CEILING MOUNTED
MOTION SENSOR - CORNER OR WALL MOUNTED
MOTION SENSOR WITH AISLE/CORRIDOR LENS - CEILING MOUNTED
COMBINATION MOTION AND DAYLIGHT SENSOR
LIGHTING CONTROL NETWORK DEVICE
DIGITAL TIMER SWITCH
MOTION SENSOR SWITCH
LOW VOLTAGE SWITCH
DIMMER MASTER SWITCH
DIGITAL DIMMING SWITCH
GRAPHICAL TOUCH SCREEN - LIGHTING CONTROL STATION
THERMOSTAT WITH A 3/4" CONDUIT TO ACCESSIBLE CEILING SPACE
MODULAR FURNITURE - BASE POWER WHIP FEED CONNECTION
MODULAR FURNITURE - FLOOR BOX FEED CONNECTION
MODULAR FURNITURE - POWER POLE FEED CONNECTION
LIGHTING CONTROL PANEL - SURFACE MOUNTED
PANELBOARD - RECESSED MOUNTED
PANELBOARD - SURFACE MOUNTED
DISTRIBUTION PANEL/ BOARD
SINGLE POLE SWITCH, DEVICE SHALL BE MOUNTED +48" MAX AND +36"
MIN FROM THE CENTER OF DEVICE:
SWITCH 3-WAY (48" AFF MAXIMUM)
TIMER SWITCH (48" AFF MAXIMUM)
DUAL SWITCH (48" AFF MAXIMUM)
PUSHBUTTON SWITCH
RECESSED ON WALL G=GFCI, WP=WEATHER PROOF
SURFACE G=GFCI, WP=WEATHER PROOF
FLOOR OR CEILING C=CEILING
20A, 125V DUPLEX RECEPTACLE
MOUNTED +15" AFF, UNLESS OTHERWISE NOTED
20A, 125V QUAD RECEPTACLE
MOUNTED +15" AFF, UNLESS OTHERWISE NOTED
20A, 125V DUPLEX RECEPTACLE
RECEPTACLE ON DEDICATED CIRCUIT
20A, 125V CONTROLLED DUPLEX RECEPTACLE
20A, 125V QUAD RECEPTACLE
(HALF) CONTROLLED RECEPTACLE
SPECIAL RECEPTACLE
REFER TO DRAWINGS FOR NEMA CONFIGURATION
JUNCTION BOX
RECESSED POKE-THROUGH
RECESSED POKE-THROUGH - POWER/TEL/DATA
RECESSED FLOOR BOX - POWER/TEL/DATA
20A, 125V DUPLEX RECEPTACLE FIRE RATED TYPE
20A, 125V QUAD RECEPTACLE FIRE RATED TYPE
W W MULTI-CHANNEL RACEWAY
ABBREVIATIONSLEGEND
SHEET INDEX
A-1
G
M
ECM
VFD
X y
X y
X y
X y
X y
X y
X y
X y
X y
X y
X y
J
PC
DS
R
ER
M
M
M
MD
N
TM
MS
LV
DM
D
G
T
J
J
JJJ
M ab
T
J
SCOPE OF WORK
1. ELECTRICAL DESIGN TO SUPPORT INSTALLATION OF NEW LINEAR ACCELERATOR AND SUPPORT AREAS.
2. DEMOLITION OF EXISTING ELECTRICAL CONNECTIONS AND SUPPORT AREAS.
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
FULLY SPRINKLERED
Office of Construction and Facilities Management
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT/ENGINEER OF RECORD A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
E-001
GENERAL NOTES, LEGEND,
ABBREVIATIINS AND SHEET INDEX
TRIPLE C - THE A/E GROUP
4401 ATLANTIC AVENUE, SUITE 200
LONG BEACH, CA 90807
562.452.7175
MAT PERKINS
REPLACE LINEAR ACCELERATOR
11301 Wilshire Blvd, West Los Angeles, CA 90073
691-CSI-121
07/29/2020 Checker Author
100% CONSTRUCTION
DOCUMENTS
ABBREVIATION DESCRIPTION
LF LINEAR FEET
LFMC LIQUIDTIGHT FLEXIBLE METAL CONDUIT
LGST LARGEST
LIS LOAD INTERRUPTER SWITCH
LOC. LOCATION
LOTO LOCK-OUT & TAG-OUT
LSI LONG TERM, SHORT TERM,
INSTANTANEOUS
LTG LIGHTING
LV LOW VOLTAGE
M METER
MAX MAXIMUM
MCA MINIMUM CIRCUIT AMPS
MCC MOTOR CONTROL CENTER
MCP MOTOR CIRCUIT PROTECTOR
MFGR, MFR MANUFACTURER
MH MANHOLE
MI. MECHANICAL INTERLOCK
MIN MINIMUM
MOCP MAXIMUM OVERCURRENT PROTECTION
MRCT MULTI-RATIO CURRENT TRANSFORMER
MTD MOUNTED
MTG MOUNTING
MTR MOTOR
MTTB MAIN TELEPHONE TERMINAL BOARD
MV MEDIUM VOLTAGE
N NORTH
NAC NOTIFICATION APPLIANCE CIRCUIT
NC NORMALLY CLOSED
NEC NATIONAL ELECTRICAL CODE
NF NON-FUSED
NIC NOT IN CONTRACT
NL NIGHT LIGHT- 24HRS ON
NO. NUMBER
OC ON CENTER
OCPD OVERCURRENT PROTECTIVE DEVICE
OD OUTSIDE DIAMETER
OE OVERHEAD ELECTRICAL
OFC OIL FUSED CUTOUT
OH OVER HEAD
OL OIL LEVER SWITCH
P POLE
PAC PROGRAMMABLE AUTOMATION
CONTROLLER
PB PULL BOX
PC PHOTOCELL
PCB POLYCHLORINATED BIPHENYL
PDS PRESSURE DIFFERENTIAL SWITCH
PF POWER FACTOR
PH OR Ø PHASE
PILC PAPER INSULATED, LEAD COVER
PIV POST INDICATING VALVE
PL PLATE
PLC PROGRAMMABLE LOGIC CONTROLLER
PNL PANEL
POC POINT OF CONNECTION
PREF. PREFERRED
PRI. PRIMARY
PVC POLY-VINYL CHLORIDE
PWR POWER
REC/RECEPT RECEPTACLE
REQ'D REQUIRED
RGS RIGID GALVANIZED STEEL
RM ROOM
RMC RIGID METAL CONDUIT
RPBP REDUCED PRESSURE BACK FLOW
PREVENTER
RTAC REAL TIME
SCCR SHORT CIRCUIT CURRENT RATING
SCE SOUTHERN CALIFORNIA EDISON
SF SQUARE FEET
SHT SHEET
SIG. SIGNAL
SP SPARE
SPECS SPECIFICATIONS
ST STREET
STD STANDARD
STP SHIELDED TWISTED PAIR
SW SWITCH
SWBD SWITCHBOARD
SWGR SWITCHGEAR
SWST SWITCHING STATION
T.O.D. TOP OF DUCTBANK
T.O.M. TOP OF MANHOLE
TB TERMINAL BLOCK
TEL./TELE TELEPHONE
TMH TELEPHONE MANHOLE
TPS TWISTED SHIELDED PAIR
TRANSF, XFMR TRANSFORMER
TS TAMPER SWITCH
TYP TYPICAL
UG UNDERGROUND
UON UNLESS OTHERWISE NOTED
V VOLTS
VA VOLT-AMPERES
VB VIBRATION SWITCH
VFD VARIABLE FREQUENCY DRIVE
W WATTS
W/ WITH
W/O WITHOUT
WP WEATHERPROOF
Z IMPEDANCE
ABBREVIATION DESCRIPTION
& AND
1/C SINGLE CONDUCTOR
@ AT
A OR AMP AMPERES
A.C. ASPHALT CONCRETE
ABV ABOVE
AF AMPERE FUSE RATING
AFC AVAILABLE FAULT CURRENT
AFF ABOVE FINISHED FLOOR
AFG ABOVE FINISH GRADE
AIC AMPERE INTERRUPTING CAPACITY
AL ALUMINUM
APPROX. APPROXIMATE
ARCH. ARCHITECT; ARCHITECTURAL
AS AMPERE SWITCH RATING
ASCC AVAILABLE SHORT CIRCUIT CURRENT
ATC AIR TERMINAL CHAMBER
ATO AUTOMATIC THROW-OVER (SWITCH)
ATS AUTOMATIC TRANSFER SWITCH
AUTO AUTOMATIC
AUX AUXILIARY
AWG AMERICAN WIRE GAUGE
B.S. BARE STRANDED
BAT BATTERY
BEL BELOW
BKBD BACKBOARD
BKR BREAKER
BLDG BUILDING
C CONDUIT
C.O. CONDUIT ONLY WITH PULL WIRE
CB CIRCUIT BREAKER
CC CONSTANT CURRENT
CKT CIRCUIT
CL CENTER LINE
CLG CEILING
CMU CONCRETE MASONRY UNIT
COL COLUMN
CP COMMUNICATION PROCESSOR
CPT CONTROL POWER TRANSFORMER
CR CONTROL RELAY
CSFD COMBINATION SMOKE FIRE DAMPER
CT CURRENT TRANSFORMER
CU COPPER
CW COLD WATER
DIAG DIAGRAM
DIS DISCONNECT
DIST. DISTANCE
DL DAMP LOCATION LISTING
DM DIGITAL METER
DMM DIGITAL METER MODULE
DP DISTRIBUTION PANEL
DWG DRAWING
DWP DEPARTMENT OF WATER & POWER
EA EACH
ECM ELECTRIC CIRCUIT MONITOR
ELEC. ELECTRICAL
EM EMERGENCY
EMH ELECTRICAL MANHOLE
EMT ELECTRICAL METALLIC TUBING
EPO EMERGENCY POWER OFF
EPR ETHYLENE PROPYLENE RUBBER
EQUIP EQUIPMENT
ERR EXISTING TO BE RELOCATED AND
RECONNECTED
EXIST/(E) EXISTING
EXP EXPLOSION PROOF
FA FIRE ALARM
FACP FIRE ALARM CONTROL PANEL
FATC FIRE ALARM TERMINAL CABINET
FFE FINISHED FLOOR ELEVATION
FIN. FINISH
FIP. FIELD INTERFACE PANEL
FIXT FIXTURE
FLA FULL LOAD AMPS
FLR FLOOR
FLUOR FLUORESCENT
FMC FLEXIBLE METAL CONDUIT
FO FIBER OBTIC
FT FEET
FTG FOOTING
GEN GENERATOR
GFI GROUND FAULT INTERRUPTER
GFR GROUND FAULT RELAY
GG GREEN GROUND
GND GROUND
HOA HAND-OFF-AUTOMATIC
HP HORSEPOWER
HT HEIGHT
HTR HEATER
HV HIGH VOLTAGE
HZ HERTZ
ICON INTEGRATED COMMUNICATIONS OPTICAL
NETWORK
IE INVERT ELEVATION
IED INTELLEGENT ELECTRONIC DEVICE
IMC INTERMEDIATE METAL CONDUIT
INCAND INCADESCENT
ISC SHORT CIRCUIT CURRENT
J, JB, J-BOX JUNCTION BOX
KCMIL THOUSAND CIRCULAR MILS
KV KILOVOLT
KVA KILOVOLT-AMPERES
KW KILOWATT
SHEET DESCRIPTION
IN THE EVENT ABBREVIATIONS NOT MENTIONED HEREIN ARE USED, REFERENCE WILL BE MADE TO ANSI Y1.1, MILITARY
STANDARD ABBREVIATIONS AND OTHER STANDARD INDUSTRY CONVENTIONS.
Revisions: Date:
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T
A T E O F
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T
A T E O F
No. E17012
PA
T R
I CK WA LDENFELS
09/30/2021
E L E C T R I C A L
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
FULLY SPRINKLERED
Office of Construction and Facilities Management
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT/ENGINEER OF RECORD A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
E-002
NOTES
TRIPLE C - THE A/E GROUP
4401 ATLANTIC AVENUE, SUITE 200
LONG BEACH, CA 90807
562.452.7175
MAT PERKINS
REPLACE LINEAR ACCELERATOR
11301 Wilshire Blvd, West Los Angeles, CA 90073
691-CSI-121
07/29/2020 Checker Author
100% CONSTRUCTION
DOCUMENTS
GENERAL NOTES
ALL WORK SHALL COMPLY WITH THE 2019 NATIONAL ELECTRICAL CODE, AND ALL OTHER APPLICABLE
FEDERAL AND STATE. WHERE THE CONSTRUCTION DOCUMENTS INDICATE MORE RESTRICTIVE
REQUIREMENTS, THE CONSTRUCTION DOCUMENTS SHALL GOVERN BUT THE CONSTRUCTION DOCUMENTS
SHALL NOT BE INTERPRETED AS AUTHORITY TO VIOLATE ANY CODE OR REGULATION.
ALL MATERIALS AND EQUIPMENT SHALL BE NEW AND SHALL BEAR THE UNDERWRITERS' LABEL (UL)OR OTHER NATIONALLY RECOGNIZED TESTING LABORATORY THAT IS ACCEPTABLE TO THE CITY OF L.A. AND
SHALL BE INSTALLED IN THE MANNER FOR WHICH THEY ARE DESIGNED AND APPROVED.
THE CONTRACTOR SHALL NOT BORE, NOTCH OR IN ANY WAY CUT INTO ANY STRUCTURAL MEMBER
WITHOUT WRITTEN APPROVAL FROM THE STRUCTURAL ENGINEER.
MECHANICAL, ELECTRICAL AND PLUMBING EQUIPMENT ANCHORAGE NOTES:
ALL MECHANICAL, PLUMBING, AND ELECTRICAL COMPONENTS SHALL BE ANCHORED AND INSTALLED PER THE DETAILS ON THE APPROVED CONSTRUCTION DOCUMENTS. WHERE NO DETAIL IS INDICATED, THE FOLLOWING COMPONENTS SHALL BE ANCHORED OR BRACED TO MEET THE FORCES AND DISPLACEMENT
REQUIREMENTS.
A. ALL PERMANENT EQUIPMENT AND COMPONENTS.
B. TEMPORARY OR MOVABLE EQUIPMENT THAT IS PERMANENTLY ATTACHED (E.G. HARD WIRED) TO THE
BUILDING UTILITY SERVICES SUCH AS ELECTRICITY, GAS OR WATER.
C. MOVABLE EQUIPMENT WHICH IS STATIONED IN ONE PLACE FOR MORE THAN 8 HOURS AND HEAVIER
THAN 400 POUNDS ARE REQUIRED TO BE ANCHORED WITH TEMPORARY ATTACHMENTS.
THE ATTACHMENT OF THE FOLLOWING MECHANICAL AND ELECTRICAL COMPONENT SHALL BE POSITIVELY ATTACHED TO THE STRUCTURE, BUT NEED NOT BE DETAILED ON THE PLANS. THESE COMPONENTS SHALL HAVE FLEXIBLE CONNECTIONS PROVIDED BETWEEN THE COMPONENT AND ASSOCIATED DUCTWORK, PIPING, AND CONDUIT.
A. COMPONENTS WEIGHING LESS THAN 400 POUNDS AND HAVE A CENTER OF MASS LOCATED 4 FEET OR LESS ABOVE THE ADJACENT FLOOR OR ROOF LEVEL THAT DIRECTLY SUPPORTS THE COMPONENT.
B. COMPONENTS WEIGHING LESS THAN 20 POUNDS, OR IN THE CASE OF DISTRIBUTED SYSTEMS, LESS THAN 5 POUNDS PER FOOT, WHICH ARE SUSPENDED FROM A ROOF OR FLOOR OR HUNG FROM A WALL.
FOR THOSE ELEMENTS THAT DO NOT REQUIRE DETAILS ON THE APPROVED DRAWINGS, THE INSTALLATION SHALL BE SUBJECT TO THE APPROVAL OF THE STRUCTURAL ENGINEER OF RECORD AND THE STRUCTURAL ENGINEER. THE PROJECT INSPECTOR WILL VERIFY THAT ALL COMPONENTS AND EQUIPMENT HAVE BEEN
ANCHORED IN ACCORDANCE WITH ABOVE REQUIREMENTS.
PIPING, DUCTWORK, AND ELECTRICAL DISTRIBUTION SYSTEM BRACING NOTES:
PIPING, DUCTWORK, AND ELECTRICAL DISTRIBUTION SYSTEMS SHALL BE BRACED TO COMPLY WITH THE
FORCES AND DISPLACEMENTS PRESCRIBED IN LATEST SECTIONS OF CBC AND ASCE.
THE BRACING AND ATTACHMENTS TO THE STRUCTURE SHALL BE DETAILED ON THE APPROVED DRAWINGS.
COPIES OF THE MANUAL SHALL BE AVAILABLE ON THE JOBSITE PRIOR TO THE START OF HANGING AND
BRACING OF THE PIPE, DUCTWORK, AND ELECTRICAL DISTRIBUTION SYSTEMS.
THE STRUCTURAL ENGINEER OF RECORD SHALL VERIFY THE ADEQUACY OF THE STRUCTURE TO SUPPORT
THE HANGER AND BRACE LOADS.
OBTAIN BUILDING AND LICENSING PERMITS AND PAY ASSOCIATED COSTS AND EXPENSES REQUIRED BY
FEDERAL, STATE, AND LOCAL AGENCIES.
VISIT THE SITE INCLUDING AREAS INDICATED ON THE DRAWINGS. OBTAIN AND INSPECT CONTRACT DOCUMENTS IMMEDIATELY UPON RECEIPT INCLUDING CONTRACT DOCUMENTS OF OTHER DISCIPLINES.
NOTIFY OWNER OF DISCREPANCIES WITHIN 24 HOURS.
SUBMIT A CHANGE ORDER TO THE OWNER DETAILING AND SPECIFYING REQUIRED WORK AND OBTAIN
APPROVAL BEFORE PROCEEDING WITH WORK WHERE CONDITIONS DEVELOP NOT COVERED BY THE
CONTRACT DOCUMENTS OR WHERE FINISHED WORK WILL NOT COMPLY WITH TITLE 24, CALIFORNIA
CODE OF REGULATIONS (CCR).
COORDINATE AND SCHEDULE WORK WITH OTHER TRADES AND OWNER.
MAINTAIN RECORD DRAWINGS DAILY TO REFLECTING CHANGES MADE TO THE CONTRACT DOCUMENTS DURING CONSTRUCTION. IDENTIFY AND DIMENSION UNDERGROUND CONDUIT LOCATIONS AND DEPTHS.
PROVIDE CONNECTION AND TERMINATION TO OWNER FURNISHED EQUIPMENT COMPLIANT WITH THE
LISTING REQUIREMENTS AND THE MANUFACTURER'S INSTALLATION REQUIREMENTS AND
RECOMMENDATIONS.
NOTIFY OWNER 48 HOURS PRIOR TO PROJECT COMPLETION TO ALLOW FOR OWNER'S INSPECTION OF
WORK. PROVIDE COMPETENT PERSONNEL TO ASSIST AND DEMONSTRATE EQUIPMENT OR SYSTEM
OPERATION. FINAL ACCEPTANCE OF WORK BY THE OWNER WILL BE AFTER APPROVAL AND ACCEPTANCE
BY AHJ.
FURNISH A ONE-YEAR WRITTEN GUARANTEE FOR MATERIALS AND WORKMANSHIP FROM THE DATE OF
SUBSTANTIAL COMPLETION.
PROVIDE AND MAINTAIN REQUIRED WORKING SPACE, DEDICATED EQUIPMENT SPACE, AND EGRESS
REQUIREMENTS PER NEC 110.26.
INSTALL CONDUITS AND EQUIPMENT PLUMB AND LEVEL. ROUTE CONDUITS PARALLEL TO WALLS AND OTHER PIPING. PROVIDE CONDUIT OFF-SETS AND SADDLES TO AVOID OBSTACLES, MAINTAIN CONDUIT
PATHWAYS, AND PROVIDE CLEARANCES.
PROVIDE FIELD BENDS FOR EMT AND RIGID CONDUIT 1 INCH AND SMALLER. DO NOT KINK OR REDUCE
CROSS SECTIONAL AREA OF CONDUIT. USE BENDING EQUIPMENT APPROVED BY THE CONDUIT
MANUFACTURER. REPAIR SCRATCHES AND MARS WITH CONDUIT MANUFACTURER APPROVED PRODUCT.
AVOID CUTTING OR DAMAGING REINFORCING BARS OR TENDONS WHEN CORING OR WHEN INSTALLING DRILLED-IN ANCHORS OR POWDER DRIVEN PINS IN REINFORCED CONCRETE. LOCATE BARS AND PRE- STRESSED (PRE- OR POST-TENSIONED) TENDONS BY USING A NON-DESTRUCTIVE METHOD PRIOR TO CORING OR INSTALLATION OF ANCHORS OR PINS. MAINTAIN MINIMUM CLEARANCE OF ONE INCH BETWEEN
REINFORCEMENT AND CORED HOLE OR DRILLED-IN ANCHOR OR PIN. COORDINATE ANY WORK IN
REINFORCED CONCRETE WITH SEOR PRIOR TO ROUGH-IN.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
CONTRACTOR SHALL FIELD DEVELOP THE CONDUIT INSTALLATION ALONG ROUTE SHOWN ON PLANS WITH PULLBOXES AND/OR LB FITTINGS INSTALLED EVERY 270 DEGREES (MAXIMUM) OF CONDUIT BENDS TO AVOID OBSTRUCTIONS. PROVIDE FIRE-RATED PENETRATIONS FOR ALL FIRE RATED WALLS AND FLOORS TO
MATCH OR EXCEED THE RATING OF WALL OR FLOOR.
PROVIDE FIRE-RATED PUTTY PADS ON ALL NEW ELECTRICAL OUTLET BOXES LOCATED IN RATED WALLS.
FUSES TO BE PROVIDED WITH REJECTION TYPE FUSE HOLDERS.
ALL CONDUITS PENETRATING CONCRETE FLOORS OR WALLS SHALL BE COORDINATED WITH SEOR PRIOR TO
COMENCEMENT OF WORK.
COMPLY WITH ARTICLE 517.13 (A) AND (B) - WIRING METHOD FOR GROUNDING OF RECEPTACLES AND FIXED ELECTRIC EQUIPMENT IN PATIENT CARE AREAS: ALL BRANCH CIRCUITS SERVING PATIENT AREA AREAS SHALL BE PROVIDED WITH A GROUND PATH FOR FAULT CURRENT BY INSTALLATION IN A METAL RACEWAY SYSTEM, OR A CABLE HAVING A METALLIC ARMOR OR SHEATH ASSEMBLY. THE METAL RACEWAY SYSTEM, OR METALLIC CABLE ARMOR, OR SHEATH ASSEMBLY SHALL ITSELF QUALIFY AS AN EQUIPMENT GROUNDING
RETURN PATH IN ACCORDANCE WITH 250.118.
COMPLY WITH ARTICLE 517.13(A) AND (B) - PATIENT CARE RECEPTACLE REQUIREMENTS: THE GROUNDING TERMINALS OF ALL RECEPTACLES AND ALL NON-CURRENT CARRYING CONDUCTIVE SURFACES IN PATIENT CARE AREAS SHALL BE GROUNDED BY AN INSULATED COPPER CONDUCTOR. PER SECTION 517.13(B).
ALL RECEPTACLES IN PATIENT CARE AREAS SHALL BE HOSPITAL GRADE.
33.
34.
35.
36.
37.
38.
39.
OBTAIN WRITTEN PERMISSION FROM THE STRUCTURAL ENGINEER PRIOR TO USE OF POWER DRIVEN PINS.
ORDER EQUIPMENT IN SPLITS SIZED TO ALLOW ENTRY INTO BUILDING AND SPACE.
COORDINATE AND SCHEDULE INTERRUPTIONS OF ELECTRICAL POWER WITH OWNER IN WRITING MINIMUM TWO WEEKS IN ADVANCE AND AT TIMES CONVENIENT TO THE OWNER. INCLUDE COST OF PREMIUM TIME
WORK IN BID.
CUT AND PATCH EXISTING CEILINGS AND WALLS WHERE SPECIFIED FOR INSTALLATION OF CONDUIT, OUTLET BOXES, SUPPORTS, AND EQUIPMENT. REPAIR CEILINGS AND WALLS TO MATCH EXISTING
INSTALLATION.
REFER TO ARCHITECTURAL REFLECTED CEILING PLANS FOR LOCATIONS OF CEILING MOUNTED EQUIPMENT.
REFER TO ARCHITECTURAL FLOOR AND ELEVATION PLANS FOR DEVICE AND OUTLET LOCATIONS. LOCATE COUNTERTOP OUTLETS 6 INCHES ABOVE BACKSPLASH. DO NOT SCALE DEVICE AND OUTLET LOCATIONS
FROM ELECTRICAL PLANS.
VERIFY ACTUAL CONNECTION LOCATIONS OF PLUMBING AND MECHANICAL EQUIPMENT WITH EQUIPMENT
SHOP DRAWINGS. COORDINATE EXACT DISCONNECT LOCATION WITH PLUMBING AND MECHANICAL
CONTRACTORS. LOCATE ROOF PENETRATIONS AND DISCONNECT SWITCHES TO PROVIDE CODE REQUIRED CLEARANCES, ACCESS, AND LINE OF SIGHT TO EQUIPMENT. INSTALL DISCONNECT SWITCHES 30 INCHES
MINIMUM ABOVE ROOF. PROVIDE ANGLE GALVANIZED STEEL SUPPORT STRUCTURE.
CONTRACTOR TO USE FIRE BARRIER MOLDABLE PUTTY ON ELECTRICAL BOXES LOCATED IN RATED WALLS.
PROVIDE TESTED FIRE STOPPING ASSEMBLY APPROVED BY THE AHJ FOR CONDUIT PENETRATIONS OF FIRE
RATED WALLS, FLOORS, AND CEILINGS.
PROVIDE GROUNDING CONDUCTOR WITH FEEDERS AND BRANCH CIRCUITS SIZED PER CEC 250.122.
PROVIDE 3 FEET MAXIMUM LIQUIDTIGHT FLEXIBLE METAL CONDUIT (LFMC) CONNECTION TO
TRANSFORMERS AND VIBRATING EQUIPMENT. USE ARMORED CABLE OR BARE CONDUCTOR, WHERE NOT
SUBJECT TO PHYSICAL DAMAGE (4 AWG OR LARGER), FOR GROUNDING ELECTRODE CONDUCTOR
TRANSFORMER CONNECTIONS. REFERENCE NEC 250.64.
COORDINATE WORK WITH OTHER DISCIPLINES TO ALLOW FOR PHASED INSTALLATION. SCHEDULE AND
PERFORM CONSTRUCTION TO MINIMIZE AND AVOID IMPACTS TO ADJOINING SPACES.
PROVIDE COPPER CONDUCTORS. USE TYPE XHHW-2 CONDUCTOR INSULATION UNDERGROUND EXTERIOR TO BUILDING AND ON ROOF TOPS. USE TYPE THHN/THWN-2 CONDUCTOR INSULATION ABOVE GRADE EXTERIOR TO BUILDING AND UNDERGROUND INTERIOR TO BUILDING. USE TYPE THHN/THWN CONDUCTOR
INSULATION ABOVE GRADE INTERIOR TO BUILDING.
USE CIRCUITS WITH DEDICATED NEUTRALS FOR BRANCH CIRCUITS. MULTIWIRE BRANCH CIRCUITS ARE NOT
ALLOWED.
PROVIDE INSTALLATION IN PATIENT CARE AREAS USING CONDUIT APPROVED AS A GROUND RETURN
BACKBOXES IN COMMON WALLS OF PATIENT CARE AREAS.
PROVIDE RIGID METAL CONDUIT FOR EXPOSED CONDUITS BELOW 8 FEET ABOVE FINISHED FLOOR.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29.
31.
32.
WORK INCLUDED: ALL LABOR, MATERIAL, APPLIANCES, TOOLS, EQUIPMENT, FACILITIES, TRANSPORTATION AND SERVICES NECESSARY AND INCIDENTAL TO PERFORMING ALL OPERATIONS IN CONNECTION WITH THE
DEMOLITION, DISPOSAL, TRANSPORTATION AND RECONNECTION OF EXISTING REMAINING
EQUIPMENT/DEVICES. COMPLETE AS INDICATED IN THE CONTRACT DOCUMENTS AS SPECIFIED HEREIN.
THE FOLLOWING IS INTENDED TO COVER THE COMPLETE DEMOLITION, DISPOSAL, TRANSPORTATION AND RECONNECTION OF EXISTING ELECTRICAL EQUIPMENT. THE OMISSION OF EXPRESSED REFERENCE TO ANY ITEM OF LABOR OR MATERIAL FOR THE PROPER EXECUTION OF THE WORK IN ACCORDANCE WITH PRESENT PRACTICE OF THE TRADE SHALL NOT RELIEVE THE CONTRACTOR FROM PROVIDING SUCH ADDITIONAL
LABOR AND MATERIALS.
REFER TO THE CONTRACT DOCUMENTS FOR ADDITIONAL DEMOLITION REQUIREMENTS WHICH AFFECT THE PROPER EXECUTION OF THIS WORK. DIAGRAMS AND SYMBOLS SHOWING ELECTRICAL CONNECTIONS ARE
DIAGRAMMATIC ONLY. WIRING DIAGRAMS DO NOT NECESSARILY SHOW THE EXACT PHYSICAL
ARRANGEMENT OF THE EQUIPMENT.
IT IS THE CONTRACTOR'S RESPONSIBILITY TO VISIT THE SITE AND BECOME THOROUGHLY FAMILIAR WITH ALL FEATURES OF THE BUILDING AND SITE WHICH MAY AFFECT THE PROPER PERFORMANCE OF THIS WORK.
PORTIONS OF THESE PLANS HAVE BEEN DERIVED FROM INFORMATION TAKEN FROM ORIGINAL ELECTRICAL PLANS. THE INTENT OF THE DRAWINGS AND SPECIFICATIONS IS TO PROVIDE FOR THE REMOVAL AND DISPOSAL OF ALL ELECTRICAL EQUIPMENT, LIGHTING FIXTURES, DEVICES, OUTLET BOXES, CONDUIT, WIRING, ETC. AS DEFINED HEREIN AND AS INDICATED ELSEWHERE IN THE CONTRACT DOCUMENTS.
IT IS UNDERSTOOD AND AGREED THAT THIS CONTRACT DOES NOT CONTEMPLATE THE HANDLING OF ASBESTOS, PCB OR ANY HAZARDOUS WASTE MATERIAL. IF ASBESTOS, PCB OR ANY HAZARDOUS WASTE MATERIAL IS ENCOUNTERED, NOTIFY THE ELECTRICAL ENGINEER IMMEDIATELY. DO NOT DISTURB, HANDLE
OR ATTEMPT TO REMOVE.
THE SCOPE OF DEMOLITION WORK SHALL INCLUDE, BUT IS NOT LIMITED TO THE FOLLOWING:
A. SUBMIT PROPOSED OUTAGE SCHEDULE. PROVIDE A SEQUENCE OF DEMOLITION TO INSURE
THE UNINTERRUPTED USE OF OCCUPIED PORTIONS OF THE FACILITY WHICH ARE TO
REMAIN OPERATIONAL DURING THE CONTRACT PERIOD.
B. SYSTEM OUTAGES SHALL BE PERMITTED ONLY AT TIMES APPROVED BY OWNER, IN WRITING.
WORK WHICH COULD RESULT IN A ACCIDENTAL OUTAGE (BEYOND BRANCH CIRCUITS)
SHALL BE PERFORMED WITH THE OWNER'S MAINTENANCE PERSONNEL ADVISED OF SUCH
WORK.
1) COORDINATE WITH THE OWNER AND DESIGNATE A PORTION OF THE
FACILITY FOR SAFELY STORING ALL REMOVED AND SALVAGED MATERIALS.
C. ALL REMOVED MATERIALS AND EQUIPMENT WHICH IN THE OPINION OF THE
OWNER/ARCHITECT ARE SALVAGEABLE, SHALL REMAIN THE PROPERTY OF THE OWNER.
DELIVER SUCH SALVAGED MATERIALS AND EQUIPMENT ON PREMISES AS DIRECTED, NEATLY
PILE OR STORE THEM AND PROTECT FROM DAMAGE. DO NOT REUSE MATERIALS AND
EQUIPMENT UNLESS SPECIFICALLY INDICATED ON PLANS OR SPECIFIED. REMOVE FROM
PREMISES AND DISPOSE OF ALL MATERIALS CONSIDERED BY THE OWNER/ARCHITECT TO BE
SCRAP.
D. PERFORM CUTTING AND PATCHING OF THE CONSTRUCTION WORK WHICH MAY BE
REQUIRED FOR THE PROPER DEMOLITION OF THE ELECTRICAL WORK. PATCHING SHALL BE
OF THE SAME MATERIAL, THICKNESS, WORKMANSHIP AND FINISH AS EXISTING AND
ACCURATELY MATCH SURROUNDING WORK TO THE SATISFACTION OF THE OWNER.
CUTTING OF STRUCTURAL MEMBERS SHALL NOT BE DONE WITHOUT FIRST OBTAINING
APPROVAL FROM THE OWNER AND/OR STRUCTURAL ENGINEER OF RECORD.
1) PATCHING OF OPENINGS IN RATED PARTITIONS, BARRIERS, FLOORS,
CEILINGS,ETC. SHALL BE EXECUTED USING UL AND NEPA FIRE STOP
MATERIAL EQUAL TO THE FIRE RATINGS OF THE PENETRATED SURFACE.
2) WHERE MATERIALS AND EQUIPMENT HAVE BEEN REMOVED AND NOT
REPLACED THE EXPOSED SURFACE SHALL BE PAINTED TO MATCH
SURROUNDING SURFACE COLOR.
E. AS INDICATED, DISCONNECT AND REMOVE EXISTING ELECTRICAL EQUIPMENT, DISCONNECT
SWITCHES, RECEPTACLES, MOUNTING HARDWARE AND ASSOCIATED RACEWAYS,
CONDUCTORS AND OUTLET BOXES. WHERE OUTLET BOXES ARE FLUSH AND CONDUITS ARE
CONCEALED IN EXISTING REMAINING WALLS, REMOVE WIRING AND ABANDON CONDUIT IN
PLACE.
F. EXCEPT AS NOTED OTHERWISE DISCONNECT AND REMOVE EXISTING FIRE ALARM SYSTEM
EQUIPMENT, DEVICES, MOUNTING HARDWARE AND ASSOCIATED RACEWAYS, CONDUCTORS
AND OUTLET BOXES. WHERE OUTLET BOXES ARE FLUSH AND CONDUITS ARE CONCEALED
IN EXISTING REMAINING WALLS, REMOVE WIRING AND ABANDON CONDUITS IN PLACE.
G. WHERE INDICATED BY THE DRAWINGS REMOVE EXISTING ELECTRICAL DISTRIBUTION
EQUIPMENT INCLUDING SWITCHBOARDS, DISTRIBUTION BOARDS, PANELBOARDS, MOTOR
CONTROL CENTERS, TRANSFORMERS, LOAD CENTERS, ETC. MODIFICATIONS TO THE MAIN
SWITCHBOARD SHALL NOT TAKE PLACE UNTIL THE SERVING UTILITY POWER COMPANY IS
NOTIFIED. COMPLY WITH ALL POWER OUTAGE RESTRICTIONS INDICATED ELSEWHERE IN THE
CONTRACT DOCUMENTS.
H. EXCEPT AS NOTED OTHERWISE REMOVE EXISTING TIME SWITCHES CONTACTORS,
RELAYS AND ASSOCIATED CONDUITS AND CONDUCTORS CONTROLLING HVAC SYSTEMS.
I. WHERE EXISTING WALL AND CEILINGS ARE TO REMAIN, PROVIDE BLANK COVER PLATES FOR
OUTLETS WHERE EQUIPMENT OR DEVICES ARE REMOVED UNDER THIS CONTRACT. PRIME
BLANK PLATES AND PAINT TO MATCH SURROUNDING AREA.
J. REFER TO THE DRAWINGS FOR ALL EQUIPMENT TO BE RELOCATED AND REUSED.
K. THE CONTRACTOR SHALL FIELD VERIFY EXISTING EQUIPMENT, DEVICES, AND/OR CIRCUITS
THAT ARE REMAINING. CIRCUITS SHALL BE RECONNECTED TO NEW OR EXISTING POWER
DISTRIBUTION EQUIPMENT AND FUNCTION IN THE MANNER THEY WERE ORIGINALLY
DESIGNED. PROVIDE ADDITIONAL EQUIPMENT, DEVICES, OUTLET BOXES, CONDUIT, WIRING,
ETC. AS REQUIRED TO RESTORE CONTINUITY TO THESE CIRCUITS.
L. WHERE FIXTURES, EQUIPMENT, DEVICES, ETC. ARE SPECIFIED BY THE CONTRACT
DOCUMENTS FOR REMOVAL, THE CONTRACTOR SHALL REMOVE ALL CIRCUIT
CONDUCTORS/CABLING BACK TO THE NEAREST REMAINING JUNCTION BOX AND/OR POINT
OF TERMINATION.
M. PROVIDE ADDITIONAL CONDUIT, CONDUCTORS, CABLING, OUTLET BOXES, ETC. AS
REQUIRED FOR THE RELOCATION OF ELECTRICAL APPARATUS.
N. RELOCATION AND/OR REMOVAL OF EXISTING EQUIPMENT, DEVICES, OUTLET BOXES,
CONDUIT, WIRING, ETC. MAY AFFECT THE OPERATION OF EXISTING, REMAINING ELECTRICAL
EQUIPMENT /DEVICES. THE CONTRACTOR SHALL PROVIDE ADDITIONAL MATERIALS AS
REQUIRED TO MAINTAIN AND/OR RESTORE CONTINUITY OF SERVICE TO EXISTING,
REMAINING ELECTRICAL/DEVICES.
O. PROVIDE ALL NECESSARY DEMOLITION TO REMOVE EXISTING UNUSED RACEWAYS,
CONDUCTORS, CABLING, OUTLET BOXES, RECEPTACLES, SWITCHES, EQUIPMENT, ETC.
P. WHERE IT IS NOT FEASIBLE TO REMOVE UNUSED FLUSH MOUNTED OUTLET BOXES AND
CONCEALED RACEWAYS IN EXISTING, REMAINING WALLS AND/OR CEILINGS REMOVE ALL
WIRING AND PROVIDE BLANK COVER PLATES OVER OUTLET BOXES.
Q. ALL CONDUITS RISING FROM BELOW GRADE TO AREAS WHERE PARTITIONS, WALLS, AND/OR
OTHER CONSTRUCTION ENTITIES ARE INDICATED AS BEING REMOVED SHALL BE CUT TO
BELOW FINISH FLOOR, CAPPED AND ABANDONED. PROVIDE PATCHING AS REQUIRED. SEAL
IN A MANNER ACCEPTABLE TO THE OWNER.
R. COVER UNUSED AND/OR ABANDONED OUTLETS WITH BLANK COVER PLATES.
S. SEAL ALL ABANDONED FLOOR PENETRATIONS IN A MANNER ACCEPTABLE TO THE OWNER.
T. DISCONNECT ABANDONED CIRCUITS AT EXISTING PANELBOARDS AND REMOVE WIRE TO
LAST REMAINING DEVICE. LABEL ALL ABANDONED CIRCUIT BREAKERS "SPARE".
U. PROVIDE WEATHERPROOF CAPS ON ABANDONED CONDUITS PENETRATING THE ROOF.
REPAIR ROOFING DAMAGED BY REMOVAL OF EXISTING ELECTRICAL EQUIPMENT.
V. PROVIDE NEW TYPEWRITTEN DIRECTORIES FOR RE-USED PANELBOARDS THAT ARE ALTERED
AS PART OF THE DEMOLITION.
W. EXISTING CIRCUITS THAT ARE REMOVED AND NOT RE-USED SHALL BE IDENTIFIED ON THE
PANEL DIRECTORIES AS "SPARE".
X. THE CONTRACTOR SHALL PROVIDE CIRCUIT NUMBERS AND LOADS FOR ALL EXISTING
REMAINING EQUIPMENT. PRIOR TO RECONNECTION, REASSIGN CIRCUITS AS REQUIRED TO
BALANCE LOADS EVENLY ACROSS PHASES WHEN RECONNECTED. PROVIDE COMPLETE "AS-
BUILTS" DRAWINGS.
1.
2.
3.
4.
5.
6.
7.
DEMOLITION NOTES
LOW VOLTAGE NOTES
1.
2.
3.
ELECTRICAL CONTRACTOR TO PROVIDE 120V TO POWER SUPPLIES FOR NURSE CALL, FIRE ALARM, SECURITY AND AV SYSTEMS AS REQUIRED BY TECHNOLOGY DRAWINGS.
ELECTRICAL CONTRACTOR TO PROVIDE CONDUIT FOR VOICE/DATA, NURSE CALL, FIRE ALARM, SECURITY
AND AV SYSTEMS AS REQUIRED BY TECHNOLOGY DRAWINGS.
NURSE CALL AND FIRE ALARM DRAWINGS SUBMITTED UNDER DEFERRED APPROVAL.
PRECONSTRUCTION NOTES
1.
2.
CONTRACTOR SHALL PROVIDE A 30 DAY LOAD MEASUREMENT OF EQUIPMENT INDICATED AND PROVIDE THE RESULTS TO THE ENGINEER OF RECORD (EEOR) TO VERIFY THE EQUIPMENT HAS CAPACITY TO SUPPORT ADDITIONAL LOAD. LOAD MEASUREMENTS SHALL BE SUBMITTED PRIOR TO OR ALONG WITH EQUIPMENT SUBMITTALS. EQUIPMENT SUBMITTALS WITHOUT LOAD MEASUREMENTS SHALL BE RETURNED
AND NOT REVIEWED.
CONTRACTOR SHALL PROVIDE A SUBMITTAL INDICATING CIRCUIT TRACING OF ALL EXISITNG CIRCUITS THAT ARE BEING REUSED. ALONG WITH OR PRIOR TO EQUIPMENT SUBMITTALS. EQUIPMENT SUBMITTALS
SUBMITTED WITHOUT CIRCUIT TRACING SHALL BE RETURNED AND NOT REVIEWED.
Revisions: Date:
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T
A T E O F
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T
A T E O F
No. E17012
PA
T R
I CK WA LDENFELS
09/30/2021
E L E C T R I C A L
LIGHT FIXTURE SCHEDULE
TYPE FIXTURE DESCRIPTION MANUFACTURER & MODEL TOTAL
V-A
LAMP
TYPE LUMENS COLOR
TEMP CRI VOLTAGE MTG CEILING TYPE REMARKS
L1 6" LED SEALED, RECESSED DOWNLIGHT
MRI RATED FIXUTRE.
KENALL
MRIDL6-PAFW-23L-35K9-RIMRI6-
24V-DIM1
23 LED 1504 3500 90 24 R - NOTES : 1, 2
L2 6" LED SEALED, RECESSED DOWNLIGHT
WITH FLUSH LENS TRIM.
LITHONIA LIGHTING
LDN6-35/30-LO6-WR-MVOLT 35 LED 3139 3500 90 MVOLT R - NOTES : 1, 2
L3 2 X 4 LED FIXTURE. LITHONIA LIGHTING
2BLT4-30L-ADSM-MVOLT-LP935 24 LED 3000 3500 90 MVOLT R - NOTES : 1, 2
NOTES: MOUNTING ABBREVIATIONS:
P = PENDANT
R = RECESSED
S = SURFACE
U = UNDERCABINET
1. COLOR/FINISH AS SELECTED BY ARCHITECT.
GENERAL
NOTES:
1. LIGHTING FIXTURE TYPE SHALL BE TYPE AS INDICATED ON FIXTURE SCHEDULE
AND SHALL BE U.L APPROVED.
2.
IT IS THE CONTRACTORS RESPONSIBILITY TO DETERMINE ACTUAL CEILING
CONSTRUCTION TYPE AS DEFINED ON THE ARCHITECTURAL DRAWINGS AND
FURNISH ALL LIGHTING FIXTURES WITH THE CORRECT MOUNTING DEVICES AND
PROPER OPERATING VOLTAGE.
3. LAMPS SHALL BE TYPE AS INDICATED ON DRAWINGS.
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
FULLY SPRINKLERED
Office of Construction and Facilities Management
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT/ENGINEER OF RECORD A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
E-003
SCHEDULES
TRIPLE C - THE A/E GROUP
4401 ATLANTIC AVENUE, SUITE 200
LONG BEACH, CA 90807
562.452.7175
MAT PERKINS
REPLACE LINEAR ACCELERATOR
11301 Wilshire Blvd, West Los Angeles, CA 90073
691-CSI-121
07/29/2020 Checker Author
100% CONSTRUCTION
DOCUMENTS
TOTAL PANEL = 20,028 VA @ 208V, 3Ø = 56 AMPS
LCL = 0 VOLT-AMPS
TOTAL ØC = 3,620 VOLT-AMPS
TOTAL ØB = 7,448 VOLT-AMPS 1.
TOTAL ØA = 8,960 VOLT-AMPS * "L" DENOTES LONG CONTINUOUS LOAD
NOTES:
SPACE 41 - - * 42 SPACE
SPACE 39 - * - 40 SPACE
SPACE 37 * - - 38 SPACE
SPACE 35 - - * 36 SPACE
SPACE 33 - * - 34 SPACE
SPACE 31 * - - 32 SPACE
SPACE 29 - - * 30 SPACE
SPARE 27 20/1 - * - 28 SPACE
SPARE 25 20/1 * - - 26 SPACE
SPARE 23 20/1 - - * 24 SPACE
RM.0687 REC. 3 540 21 20/1 - * - 20/1 22 SPARE
RM.0689 REC. 4 720 19 20/1 * - - 20/1 20 SPARE
RM.0689 S-4 2 17 20/1 - - * 20/1 18 SPARE
RM.0689 S-3 2 1,000 15 20/1 - * - 20/1 16 540 3 TREATMENT REC.
RM.0689 S-2 2 1,000 13 20/1 * - - 14 1,000 3 DATA ROOM-4
RM.0689 PRINT-1 1 500 11 20/1 - - * - 12 -
RM.0689 S-1 2 1,000 9 20/1 - * - 20/2 10 1,248 1 DATA ROOM-3
- 1,872 7 - * - - - 8 1,248 -
OU-2 1,872 5 25/2 - - * 20/2 6 1,248 1 DATA ROOM-2
- 1,872 3 - - * - - 4 1,248 -
OU-1 1,872 1 25/2 * - - 20/2 2 1,248 1 DATA ROOM-1
LOADS NOTE LTG REC MISC A B C CKT POLE A B C POLE CKT A B C LTG REC MISC NOTE LOADS
SEE * OU... VOLT... BKR/ BKR/ VOLT... OU... * SEE
MOUNTING : FLU... MAIN BREAKER : 100A/3P
FLOOR : FIRST FLOOR BUS AMPS : 100A MINIMUM BUS BRACING : 10KIAC
LOCATION : RM. 0689 VOLTAGE/PHASE : 120/208V, 3Ø, 4W FED FROM : XMFR "TL"
PANEL: (N)LIN2
TOTAL PANEL = 98,532 VA @ 480V, 3Ø = 119 AMPS
LCL = 0 VOLT-AMPS
TOTAL ØC = 32,844 VOLT-AMPS
TOTAL ØB = 32,844 VOLT-AMPS
TOTAL ØA = 32,844 VOLT-AMPS * "L" DENOTES LONG CONTINUOUS LOAD
NOTES:
SPACE 41 - - * - 42 10,000 -
SPACE 39 - * - - 40 10,000 -
SPACE 37 * - - 50/3 38 10,000 TRANSFORMER "TL"
SPACE 35 - - * 36 SPACE
SPACE 33 - * - 34 SPACE
SPACE 31 * - - 32 SPACE
SPACE 29 - - * 30 SPACE
SPACE 27 - * - 28 SPACE
SPACE 25 * - - 26 SPACE
SPACE 23 - - * 24 SPACE
SPACE 21 - * - 22 SPACE
SPACE 19 * - - 20 SPACE
SPACE 17 - - * 20/1 18 SPARE
SPACE 15 - * - 20/1 16 SPARE
SPACE 13 * - - 20/1 14 SPARE
SPARE 11 20/1 - - * - 12 3,878 -
SPARE 9 20/1 - * - - 10 3,878 -
SPARE 7 20/1 * - - 20/3 8 3,878 AC-1
- 16,666 5 - - - * - 6 2,300 -
- 16,666 3 - - * - - 4 2,300 -
PDU 16,666 1 70/3 * - - 20/3 2 2,300 FCU-1
LOADS NOTE LTG REC MISC A B C CKT POLE A B C POLE CKT A B C LTG REC MISC NOTE LOADS
SEE * OU... VOLT... BKR/ BKR/ VOLT-A... OU... * SEE
MOUNTING : FLUSH MAIN BREAKER : 200/3P
FLOOR : BASEMENT BUS AMPS : 225A MINIMUM BUS BRACING : 22KIAC
LOCATION : BASEMENT VOLTAGE/PHASE : 277/480V, 3Ø, 4W FED FROM : US #3
PANEL: (N)LIN1
1 AIC RATING OF THE PANEL IS PRELIMINARY. CONTRACTOR SHALL UPDATED THE AIC RATING ONCE THE SHORT CIRCUIT, OVERCURRENT COORDINATION AND ARC FLASH STUDY IS COMPLETED. REFER TO SINGLE LINE DIAGRAM ON SHEET E-601 FOR MORE DETAILS.
1 1
1 AIC RATING OF THE PANEL IS PRELIMINARY. CONTRACTOR SHALL UPDATED THE AIC RATING ONCE THE SHORT CIRCUIT, OVERCURRENT COORDINATION AND ARC FLASH STUDY IS COMPLETED. REFER TO SINGLE LINE DIAGRAM ON SHEET E-601 FOR MORE DETAILS.
Revisions: Date:
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T
A T E O F
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T
A T E O F
No. E17012
PA
T R
I CK WA LDENFELS
09/30/2021
E L E C T R I C A L
'(E) LA''(E) BXR2-3'
(E)(E)(E)(E)
(E)
(E)
(ER)
(ER)
(ER)
(E)
(E)
(E)
(E)
(E)
1 (TYPICAL)
2(TYPICAL)
(ER)
(ER)
(ER)
(ER)
(ER)
B500 KEY PLAN
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
FULLY SPRINKLERED
Office of Construction and Facilities Management
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT/ENGINEER OF RECORD A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
ED100-0
ELECTRICAL DEMO PLAN - GROUND
FLOOR
TRIPLE C - THE A/E GROUP
4401 ATLANTIC AVENUE, SUITE 200
LONG BEACH, CA 90807
562.452.7175
MAT PERKINS
REPLACE LINEAR ACCELERATOR
11301 Wilshire Blvd, West Los Angeles, CA 90073
691-CSI-121
07/29/2020 Checker Author
100% CONSTRUCTION
DOCUMENTS
1 EXISTING DEVICE TO REMAIN.
REFER TO DEMOLITION NOTED SHEET E-002 FOR ADDITIONAL
REQUIREMENTS.
ALL REMAINING ABOVE CEILING CONDUITS SERVING AREAS
OUTSIDE OF THE PROJECT AREA THAT OBSTRUCT THE
INSTALLATION OF THE NEW CEILING AND HVAC ELEMENTS ARE TO
BE INTERCEPTED, RE-ROUTED AND RECONNECTED. CONTRACTOR
TO CIRCUIT TRACE AND DOCUMENT THE FOLLOWING TO
COORDINATE WITH OWNER TO MINIMIZE DISRUPTIONS TO THE
FACILITY.
ALL DEVICES OUTSIDE THE PROJECT AREA TO REMAIN IN
SERVICE.
PROVIDE SUBMITTAL INDICATING CIRCUIT TRACING ALONG WITH
OR PRIOR TO EQUIPMENT SUBMITTALS. EQUIPMENT SUBMITTALS
SUBMITTED WITHOUT CIRCUIT TRACING SHALL BE RETURNED AND
NOT REVIEWED.
1.
2.
3.
4.
RELOCATE EXISTING PANEL TO NEW LOCATION SHOWN
ON PLAN. EXTEND EXISTING CONDUIT/FEEDERS TO NEW
LOCATION.
2 DEMOLISH EXISTING DEVICE BACK TO LAST EXISTING
JUNCTION BOX OUTSIDE OF THE SCOPE OF WORK.
COORDINATE EXTENT OF DEMOLITION WITH ARCHITECTURAL
DRAWINGS.
DEMOLISH EXISTING FEEDERS SERVING THE EXISTING
LINER ACCELERATOR BACK TO THE SOURCE.
DEMOLISH EXISTING POWER DISTRIBUTION UNIT (PDU),
AND FEEDERS BACK TO THE SOURCE.
Revisions: Date:
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T
A T E O F
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T
A T E O F
No. E17012
PA
T R
I CK WA LDENFELS
09/30/2021
E L E C T R I C A L
1 (TYPICAL)
1 (TYPICAL)
1 (TYPICAL)
2(TYPICAL)
2(TYPICAL)
GLB-20
GLAE-3
GLB-22
B500 KEY PLAN
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
FULLY SPRINKLERED
Office of Construction and Facilities Management
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT/ENGINEER OF RECORD A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
ED100-1
ELECTRICAL DEMO RCP PLAN -
GROUND FLOOR
TRIPLE C - THE A/E GROUP
4401 ATLANTIC AVENUE, SUITE 200
LONG BEACH, CA 90807
562.452.7175
MAT PERKINS
REPLACE LINEAR ACCELERATOR
11301 Wilshire Blvd, West Los Angeles, CA 90073
691-CSI-121
07/29/2020 Checker Author
100% CONSTRUCTION
DOCUMENTS
EXISTING LIGHTING FIXTURE TO BE DISCONNECTED AND
REMOVED. EXISTING LIGHTING CIRCUIT TO REMAIN TO BE
REUSED.
EXISTING LIGHTING CONTROLS TO BE DISCONNECTED
AND REMOVED.
REFER TO DEMOLITION NOTED SHEET E-002 FOR ADDITIONAL
REQUIREMENTS.
ALL REMAINING ABOVE CEILING CONDUITS SERVING AREAS
OUTSIDE OF THE PROJECT AREA THAT OBSTRUCT THE
INSTALLATION OF THE NEW CEILING AND HVAC ELEMENTS ARE TO
BE INTERCEPTED, RE-ROUTED AND RECONNECTED. CONTRACTOR
TO CIRCUIT TRACE AND DOCUMENT THE FIXTURE CIRCUITS TO
COORDINATE WITH OWNER TO MINIMIZE DISRUPTIONS TO THE
FACILITY.
PROVIDE SUBMITTAL INDICATING CIRCUIT TRACING ALONG WITH
OR PRIOR TO EQUIPMENT SUBMITTALS. EQUIPMENT SUBMITTALS
SUBMITTED WITHOUT CIRCUIT TRACING SHALL BE RETURNED AND
NOT REVIEWED.
1.
2.
3.
3 INTERCEPT EXISTING CONDUIT AND EXISTING CIRCUIT.
REUSE AND EXTEND EXISTING CIRCUIT TO NEW LOCATION.
REFER TO RENOVATION PLANS FOR NEW LOCATION.
CONTRACTOR TO CIRCUIT TRACE EXISTING CIRCUIT AND
FIELD VERIFY EXISTING CONDUCTOR SIZE AND PROVIDE
SHOP DRAWINGS AND SUBMIT TO EEOR FOR REVIEW
PRIOR TO START OF WORK.
PROVIDE SUBMITTAL INDICATING CIRCUIT TRACING ALONG
WITH OR PRIOR TO EQUIPMENT SUBMITTALS. EQUIPMENT
SUBMITTALS SUBMITTED WITHOUT CIRCUIT TRACING
SHALL BE RETURNED AND NOT REVIEWED.
Revisions: Date:
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T
A T E O F
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T
A T E O F
No. E17012
PA
T R
I CK WA LDENFELS
09/30/2021
E L E C T R I C A L
(E)(E)(E)(E)
(E)
(E)
'BXR2-3'
'LA'
(E)
(E)
(E)
(E)
(E)
PDU PULL BOX
PULL BOX
PULL BOX
EMERGENCY OFF BUTTON
MAIN POWER DISCONNECT "MB"
LIN2-1,3
LIN2-5,7
LIN1-2,4,6
LIN2-10,12
LIN2-6,8
LIN2-2,4
OU-1
OU-2
LIN2-14
FCU-1
5 6
'LIN2' 7
LIN2-9
9 9
LIN2-11
LIN2-13
13 13
LIN2-15
LIN2-17
TRENCH
ES
RL
ES
DS
RL
181716
ES
RL
ES
+48"
LIN2-16
191919
19 LIN2-19
+48"
LIN2-21
POWER CONDITIONER 21
LIN1-8,10,1222 14
AC-1
B500 KEY PLAN
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
FULLY SPRINKLERED
Office of Construction and Facilities Management
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT/ENGINEER OF RECORD A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
E-100-0
ELECTRICAL RENOVATION PLAN -
GROUND FLOOR
TRIPLE C - THE A/E GROUP
4401 ATLANTIC AVENUE, SUITE 200
LONG BEACH, CA 90807
562.452.7175
MAT PERKINS
REPLACE LINEAR ACCELERATOR
11301 Wilshire Blvd, West Los Angeles, CA 90073
691-CSI-121
07/29/2020 Checker Author
100% CONSTRUCTION
DOCUMENTS
1 RELOCATE EXISTING PANEL TO NEW LOCATION SHOWN ON
PLAN. EXTEND EXISTING CONDUIT/FEEDERS AND BRANCH
CIRCUITS TO NEW LOCATION. COORDINATED SHUTDOWNS
WITH THE OWNER/USER.
ALL CONDUITS PENETRATING CONCRETE FLOORS OR WALLS
SHALL BE COORDINATED WITH SEOR PRIOR TO COMMENCEMENT
OF WORK.
ELECTRICAL CONTRACTOR SHALL VERIFY THE EXACT LOCATIONS
OF ALL CONNECTION POINTS WITH EQUIPMENT INSTALLER PRIOR
TO ROUGH-IN.
ELECTRICAL CONTRACTOR SHALL REFER TO MECHANICAL AND
ARCHITECTURAL DRAWINGS FOR EXACT LOCATIONS AND
CHARACTERISTICS OF ALL MECHANICAL EQUIPMENT.
CONTRACTOR TO PROVIDE DETAILED METHOD OF PROCEDURE
(MOP) WITH CONTINGENCY PLAN FOR THE REVIEW BY OWNER,
ARCHITECT AND EEOR PRIOR TO POWER SHUT DOWN.
VERIFY ALL VERIFY ALL REQUIREMENTS ASSOCIATED WITH THE
LINEAR ACCELERATOR EQUIPMENT WITH THE LINEAR
ACCELERATOR VENDOR PRIOR TO ROUGH-IN.
2 PER VENDOR DRAWING SHEET E101 PDU PULL BOX
(FLOOR). PROVIDE 2'-0" X 1'-6" X 1'-0" WITH COVER PLATE.
LAST 6" OPEN TO REMAIN OPEN. COVER RATED FOR WEIGHT
OF PDU.
3 PER VENDOR DRAWING SHEET E101 PROVIDE CONDUIT
FLOOR PULL BOX 1'-0" X 8" X 8".
4 PER VENDOR DRAWING SHEET E101 PROVIDE 6" X 6" X 6"
PATIENT TABLE PULL BOX WITH DIVIDER.
5 REFER TO SINGLE LINE DIAGRAM FOR REQUIREMENTS.
6 PROVIDE STRUCTURAL SUPPORT PER SEISMIC ANCHORAGE
SPECIFICATION REQUIREMENTS.
7 PER VENDOR DRAWING SHEET E101 PROVIDE NEMA L6-20
RECEPTACLE.
8 PROVIDE 208V, 30AS, 25AF,2P DISCONNECT SWITCH IN THE
INTERSTITIAL SPACE. PROVIDE UNISTRUT MOUNTING FOR
DISCONNECT SWITCH AS REQUIRED. COORDINATE EXACT
LOCATION IN FIELD.
1.
2.
3.
4.
5.
9 PROVIDE 480V, 30AS, 20AF,3P DISCONNECT SWITCH IN THE
INTERSTITIAL SPACE. PROVIDE UNISTRUT MOUNTING FOR
DISCONNECT SWITCH AS REQUIRED. COORDINATE EXACT
LOCATION IN FIELD.
10 PER VENDOR DRAWING SHEET E101 APOLLO LASER WALL
POWER SWITCH PROVIDED BY VENDOR.
11 PER VENDOR DRAWING SHEET E101 PROVIDE EMERGENCY
OFF BUTTON.
12 PER VENDOR DRAWING SHEET E101 PROVIDE EMERGENCY
STOP BUTTON.
13 PER VENDOR DRAWING SHEET E101 PROVIDE DOOR
INTERLOCK SWITCH.
14 PER VENDOR DRAWING SHEET E101 PROVIDE RADIATION
WARNING LIGHT.
15 PER VENDOR DRAWING SHEET E101 PROVIDE 4#4 + 1#8 GND.
IN 4" CONDUIT.
16 PER VENDOR DRAWING SHEET E101 PROVIDE 2#12 IN 1"C
FROM EMERGENCY STOP SWITCHES TO PDU.
17 PER VENDOR DRAWING SHEET E101 PROVIDE 2#12 IN 1"C
FROM RADIATION WARNING LIGHTS TO PDU.
18 PER VENDOR DRAWING SHEET E101 PROVIDE 2#12 IN 1"
PROVIDE 1"C FROM DOOR INTERLOCK SWITCH TO PDU.
19 PER VENDOR DRAWING SHEET E101 PROVIDE (4) 4" CONDUIT.
20 PER VENDOR DRAWING SHEET E101 PROVIDE 4" PHYSICS
PORT CONDUIT.
Revisions: Date:
21 PER VENDOR DRAWING PROVIDE NEW POWER CONDITIONER.
COORDINATE EXACT LOCATION IN FIELD.
22 PROVIDE 480V, 30AS, 20AF,3P DISCONNECT SWITCH IN THE
PROVIDE UNISTRUT MOUNTING FOR DISCONNECT SWITCH AS
REQUIRED. COORDINATE EXACT LOCATION IN FIELD.
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T
A T E O F
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T
A T E O F
No. E17012
PA
T R
I CK WA LDENFELS
09/30/2021
E L E C T R I C A L
L1
L1
L1
L1
L1
L1
L1
L1 L1
GLAE-3
L1 L1
L1
L1L1
L1
L3
L2
L2
L3
L2 L2
L2 L2
GLB-20
GLB-22
B500 KEY PLAN
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
FULLY SPRINKLERED
Office of Construction and Facilities Management
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT/ENGINEER OF RECORD A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
E-100-1
ELECTRICAL RENOVATION RCP
PLAN - GROUND FLOOR
TRIPLE C - THE A/E GROUP
4401 ATLANTIC AVENUE, SUITE 200
LONG BEACH, CA 90807
562.452.7175
MAT PERKINS
REPLACE LINEAR ACCELERATOR
11301 Wilshire Blvd, West Los Angeles, CA 90073
691-CSI-121
07/29/2020 Checker Author
100% CONSTRUCTION
DOCUMENTS
1 INSTALL AND CONNECT NEW FIXTURE TO EXISTING
LIGHTING CIRCUIT AND NEW LIGHTING CONTROLS. MATCH
EXISTING CONDUCTOR TYPE AND SIZE WHEN MODIFYING
AND EXTENDING EXISTING CIRCUITS.
ELECTRICAL CONTRACTOR SHALL VERIFY THE EXACT LOCATIONS
OF ALL CONNECTION POINTS WITH EQUIPMENT INSTALLER PRIOR
TO ROUGH-IN.
ALL DEVICES OUTSIDE THE PROJECT AREA TO REMAIN IN SERVICE.
1.
2.
Revisions: Date:
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T
A T E O F
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T
A T E O F
No. E17012
PA
T R
I CK WA LDENFELS
09/30/2021
E L E C T R I C A L
GROUND FLOOR
GROUND FLOOR
INTERSTITIAL
(E) DIST. SECTION BXR-2
2500A, 480/277V, 3PH, 4W
PANEL
"BXR2-3"
(E)
(E) 150A 3P
BASEMENT FLOOR
(E) "T1"
15KVA
480-120/208,3PH,4W
x.x%IMP
PANEL
"LA"
(E)
PANEL
"LIN2"
(N)
PANEL
"LIN1"
(N)
(N) "TL"
30KVA
480-120/208,3PH,4W
TO (E) T3
(E)
1500KVA
4.16KV-
480/277V,3PH,4W
6.7%IMP
(E) 2500A 3P
(E) UNIT SUBSTATION #3 (E) DIST. SECTION 2500A, 480/277V, 3PH, 4W
(E)
600AS
600AT
3P
(E)
800AS
800AT
3P
(E)
600AS
600AT
3P
(E)
600AS
600AT
3P
(N)
200AS
200AT
3P
(N) 80A 3P
ENCLOSED CIRCUIT
BREAKER "MB"4
(N)30AS
20AF
3P
2"C.-4#3/0 & 1#6 GND
2"C.4#1+1#8 GND
FCU
8HP
8.3FLA
3/4"C.-3#12 & 1#10 GND.
1"C.-4#6 & 1#10 GND
(N)100AS
100AF
3P
OU
OU
(N)30AS
25AF
2P
(N)30AS
25AF
2P
3/4"C.-3#12 & 1#12 GND.
(N) POWER
CONDITIONER
(N)30AS
20AF
3P
AC
10HP
14FLA
3/4"C.-3#12 & 1#10 GND.
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
FULLY SPRINKLERED
Office of Construction and Facilities Management
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT/ENGINEER OF RECORD A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
E-601
SINGLE LINE DIAGRAM
TRIPLE C - THE A/E GROUP
4401 ATLANTIC AVENUE, SUITE 200
LONG BEACH, CA 90807
562.452.7175
MAT PERKINS
REPLACE LINEAR ACCELERATOR
11301 Wilshire Blvd, West Los Angeles, CA 90073
691-CSI-121
07/29/2020 Checker Author
100% CONSTRUCTION
DOCUMENTS
ALL NEW FEEDER CONDUCTORS SHALL BE CONTINUOUS WITH NO
SPLICES.
ALL FEEDERS SHALL BE COPPER WITH 90° C. (THHN/THWN)
INSULATION.
NEW CIRCUIT BREAKERS SHALL BE TYPE TO MATCH EXISTING,
UON. PROVIDE ALL REQUIRED MOUNTING HARDWARE, LUGS AND
COVERPLATES FOR A COMPLETE INSTALLATION.
PROVIDE SCHNEIDER SURFACE WALL MOUNTED ENCLOSED
80A CIRCUIT BREAKER WITH 277V SHUNT TRIP AND BUILT-IN
EPO SWITCH ON NEMA 1 ENCLOSURE.
REFER TO SHEETS E-602 FOR ADDITIONAL REQUIREMENTS.4
CONTRACTOR SHALL PROVIDE A 30 DAY LOAD MEASUREMENT
OF EQUIPMENT INDICATED AND PROVIDE THE RESULTS TO
THE ENGINEER OF RECORD (EEOR) TO VERIFY THE
EQUIPMENT HAS CAPACITY TO SUPPORT ADDITIONAL LOAD.
LOAD MEASUREMENTS SHALL BE SUBMITTED PRIOR TO OR
ALONG WITH EQUIPMENT SUBMITTALS. EQUIPMENT
SUBMITTALS WITHOUT LOAD MEASUREMENTS SHALL BE
RETURNED AND NOT REVIEWED.
PROVIDE NEW CIRCUIT BREAKER INDICATED IN AVAILABLE
SPACE. TYPE AND RATING TO MATCH EXISTING. PROVIDE ALL
REQUIRED MOUNTING HARDWARE, LUGS AND COVERPLATES
FOR A COMPLETE INSTALLATION.
3/4"C.-1#6 GND. WIRE TO EXISTING GROUNDING SYSTEM.5
1"C.-4#4 & 1#6 GND.6
PER VENDOR DRAWING SHEET E101 PROVIDE 4#4 + 1#6
GND. IN 4" CONDUIT.
CONTRACTOR TO PROVIDE SHORT CIRCUIT, OVERCURRENT
PROTECTIVE DEVICE COORDINATION AND ARC FLASH ANALYSIS
FOR ALL NEW PANELS, SWITCHBOARDS AND SWITCHGEAR. STUDY
SHALL INCLUDE EXISTING SYSTEM TO THE EXTENT THAT THE FAULT
CURRENT, COORDINATION AND ARC FLASH PREPARED IN
COMPLIANCE WITH 2018 NFPA 70E FOR THE NEW PANELS CAN BE
CALCULATED. CONTRACTOR IS RESPONSIBLE FOR ALL DATA
COLLECTION REQUIRED TO COMPLETE THE ANALYSIS. REPORT OF
THE ANALYSIS SHALL BE PREPARED SUMMARIZING THE INPUT
DATA, SINGLE-LINE USED IN CALCULATION, AND RESULTS
INCLUDING ARC FLASH LABELS. REPORT SHALL BE STAMPED AND
SIGNED BY PROFESSIONAL ENGINEER UNDER WHOSE DIRECTION
THE STUDY WAS PERFORMED. CONTRACTOR TO INSTALL ARC
FLASH LABELS ON PANELS IN COMPLIANCE WITH 2018 NFPA 70E.
ELECTRICAL EQUIPMENT SUBMITTALS SHALL REFLECT RESULTS OF
SHORT CIRCUIT ANALYSIS. CONTRACTOR SHALL SET ALL
ADJUSTABLE CIRCUIT BREAKERS PER OVERCURRENT PROTECTIVE
DEVICE STUDY. MINIMAL INFORMATION INCLUDED ON ARC FLASH
LABELS INCLUDE:
1. NOMINAL SYSTEM VOLTAGE
2. ARC FLASH BOUNDARY
3. AT LEAST ONE OF THE FOLLOWING:
a.EITHER THE AVAILABLE INCIDENT ENERGY AND
THE CORRESPONDING WORKING DISTANCE, OR
THE ARC FLASH PPE CATEGORY FOR THE
EQUIPMENT, BUT NOT BOTH.
b.MINIMUM ARC RATING OF CLOTHING.
c.SITE-SPECIFIC LEVEL OF PPE.
ANALYSIS SHALL BE PREPARED PRIOR TO OR CONCURRENT WITH
ELECTRICAL SWITCHGEAR, SWITCHBOARDS AND PANEL
SUBMITTALS. EQUIPMENT SUBMITTALS WITHOUT ANALYSIS SHALL
BE REJECTED AND NOT REVIEWED.
EEOR CAN NOT BE RESPONSIBLE FOR ANALYSIS UNLESS
SEPARATELY AND SPECIFICALLY CONTRACTED TO DO SO.
1.
2.
3.
Revisions: Date:
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T
A T E O F
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T
A T E O F
No. E17012
PA
T R
I CK WA LDENFELS
09/30/2021
E L E C T R I C A L
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
FULLY SPRINKLERED
Office of Construction and Facilities Management
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT/ENGINEER OF RECORD A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
E-602
LINAC EQUIPMENT DRAWINGS
TRIPLE C - THE A/E GROUP
4401 ATLANTIC AVENUE, SUITE 200
LONG BEACH, CA 90807
562.452.7175
MAT PERKINS
REPLACE LINEAR ACCELERATOR
11301 Wilshire Blvd, West Los Angeles, CA 90073
691-CSI-121
07/29/2020 Checker Author
100% CONSTRUCTION
DOCUMENTS
LINEAR ACCELERATOR EQUIPMENT INFORMATION BASED ON THE
LATEST PROVIDED BY ACCURAY. LATER REVISIONS PROVIDED BY
ACCURAY MAY REQUIRE CHANGES TO THE CONTRACT
DOCUMENTS. REFER TO ACCURAY DRAWINGS FOR ADDITIONAL
REQUIREMENTS.
ELECTRICAL CONTRACTOR TO PROVIDE COORDINATION DRAWINGS
INDICATING ALL REQUIREMENTS FOR ACCURAY LINEAR
ACCELERATOR EQUIPMENT AS INDICATED ON LATEST DRAWINGS
PROVIDED BY THE ACCURAY DESIGN TEAM, OWNER, AND VENDOR
TO REVIEW WITH NO EXCEPTION TAKEN PRIOR TO CONSTRUCTION.
ELECTRICAL CONTRACTOR TO COORDINATE SPECIFIC RACEWAY,
CONDUIT, DEVICE, JUNCTION BOX, ETC. , REQUIREMENTS WITH
SUPPLIER/VENDOR ( ACCURAY) FOR LINEAR ACCELERATOR
EQUIPMENT. RATING OF DEVICES, CONDUCTORS, AND CONDUIT TO
BE COORDINATED WITH VENDORS SITE SPECIFIC DOCUMENTS.
1.
2.
3.
Revisions: Date:
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T
A T E O F
EFORP
DE
R E
T
SI
G E
R R E
E
NI
G
NE
LANOISS
C A L I F ORN
I AS
T
A T E O F
No. E17012
PA
T R
I CK WA LDENFELS
09/30/2021
E L E C T R I C A L
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
FULLY SPRINKLERED
Office of Construction and Facilities Management
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT/ENGINEER OF RECORD A/E:
STAMPCONSULTANT
321 4 5 6 7 8 9 10
E-603
LINAC EQUIPMENT DRAWINGS
TRIPLE C - THE A/E GROUP
4401 ATLANTIC AVENUE, SUITE 200
LONG BEACH, CA 90807
562.452.7175
MAT PERKINS
REPLACE LINEAR ACCELERATOR
11301 Wilshire Blvd, West Los Angeles, CA 90073
691-CSI-121
07/29/2020 Checker Author
100% CONSTRUCTION
DOCUMENTS
LINEAR ACCELERATOR EQUIPMENT INFORMATION BASED ON THE
LATEST PROVIDED BY ACCURAY. LATER REVISIONS PROVIDED BY
ACCURAY MAY REQUIRE CHANGES TO THE CONTRACT
DOCUMENTS. REFER TO ACCURAY DRAWINGS FOR ADDITIONAL
REQUIREMENTS.
ELECTRICAL CONTRACTOR TO PROVIDE COORDINATION DRAWINGS
INDICATING ALL REQUIREMENTS FOR ACCURAY LINEAR
ACCELERATOR EQUIPMENT AS INDICATED ON LATEST DRAWINGS
PROVIDED BY THE ACCURAY DESIGN TEAM, OWNER, AND VENDOR
TO REVIEW WITH NO EXCEPTION TAKEN PRIOR TO CONSTRUCTION.
ELECTRICAL CONTRACTOR TO COORDINATE SPECIFIC RACEWAY,
CONDUIT,…