2023 10 17 SD Report Fluoro 1C204 and 1C205 VA.pdf
PDF 461 KB Posted
- Solicitation number
- 36C26125R0032
About this file
This document is a Radiation Shielding Design Evaluation prepared by West Physics for the VA Medical Center Sacramento, covering two fluoroscopic rooms (1C204 and 1C205) in the facility's urology area. The evaluation provides detailed technical specifications for radiation shielding requirements, including minimum shielding thicknesses for various room barriers, based on calculations performed in accordance with NCRP Report No. 147. Both rooms utilize a GE OEC Elite imaging unit, with identical specifications: maximum 125 kVp, 35 patients per week, 64 mA-min per patient, and a weekly workload of 2240 mA-min/week. The report specifies shielding design goals of 0.10 mGy/week for controlled areas and 0.02 mGy/week for uncontrolled areas, with recommendations for barrier materials including lead, drywall, and concrete, tailored to the specific distances and functions of adjacent spaces.
View the file
Other files
Show all 25
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
The Medical Physics Experts
W E S TW E S T P H Y S I C S
3 0 3 3 9S u i t e 2 5 0 A t l a n t a , G A3 8 2 5 P a c e s W a l k S E 866-275-WEST (9378) www.westphysics.com
October 17, 2023
Daniel Chaparro VA Medical Center Sacramento 10535 Hospital Way Mather, California 95655
RE: Fluoroscopic Suite Shielding Design Evaluation – VA Medical Center Sacramento 10535 Hospital Way, Mather, California 95655
Dear Daniel Chaparro, Please find enclosed the radiation shielding design evaluation for your Fluoroscopic suite. The first page of the shielding plan summarizes the total minimum required shielding for each applicable room barrier.
Additional information supporting the shielding recommendations constitutes the remaining sheets. You should retain a copy of this shielding design evaluation at your facility in the event that a regulatory agency wishes to verify the room’s shielding. We expect that a regulatory audit of the installation will take place, and an inspector will likely request this document at that time.
Thank you for the opportunity to take care of your radiation protection needs. Should you have any questions or comments about this shielding design, we would be happy to speak with you at your convenience. Please feel free to contact us at (866) 275-9378 or at shielding@westphysics.com.
Thank you, David Fair, B.S.
Medical Physicist
Isaiah Robertson, B.S.
Medical Physicist
Enclosure: Radiation Shielding Design Evaluation
Radiation Shielding Design Evaluation
Performed by
Site: VA Medical Center Sacramento
10535 Hospital Way Mather, California 95655
Room: 1C204
Unit: GE OEC Elite
Shielding Design Results:
Barrier Function of Space Behind Barrier
Controlled or Uncontrolled
Occupancy Factor (T)
Distance
(m) Total Minimum Shielding Required
A-B Corridor Uncontrolled 0.200 16' 3-3/8" 1/64" lead or equivalent
B-C Med Storage Uncontrolled 0.050 15' 11-7/8" 7/8" drywall
(Standard 2 × 5/8" drywall panel construction sufficient)
C-D Outdoor Area Uncontrolled 0.025 5' 5-1/8" 1/64" lead or equivalent
D-E Procedure 1C230 Uncontrolled 0.500 3' 5-5/8" 3/64" lead or equivalent
E-F Soiled / Clean Utility Uncontrolled 0.050 9' 11-1/4" 1/64" lead or equivalent
F-A Toilet Uncontrolled 0.200 13' 4-7/8" 1/64" lead or equivalent
Floor Slab N/A N/A N/A N/A
Ceiling Roof Uncontrolled 0.025 13' 0" 1/64" lead, 3/8" standard density concrete, or equivalent*
*Required concrete ceiling may already be in place. Please double check concrete ceiling thickness.
Facility Diagram (1C204):
Note: Cardinal directions are approximate only. Shielding should be installed carefully and with respect to the function of the surrounding areas to ensure proper barrier thicknesses.
Radiation Shielding Design Evaluation
Performed by
Site: VA Medical Center Sacramento
10535 Hospital Way Mather, California 95655
Room: 1C205
Unit: GE OEC Elite
Shielding Design Results:
Barrier Function of Space Behind Barrier
Controlled or Uncontrolled
Occupancy Factor (T)
Distance
(m) Total Minimum Shielding Required
A-B Toilet Uncontrolled 0.200 3' 3-3/8" 1/32" lead or equivalent
B-C HAC Uncontrolled 0.025 3' 7" 1/64" lead or equivalent
C-D Police Office Uncontrolled 1.000 7' 7-5/8" 1/32" lead or equivalent
D-E Police Holding Uncontrolled 1.000 7' 7-3/4" 1/32" lead or equivalent
E-F Outside Uncontrolled 0.025 14' 2-3/8" 1/2" drywall
(Standard 2 × 5/8" drywall panel construction sufficient)
F-G Med Storage Uncontrolled 0.050 18' 10-1/8" 5/8" drywall
(Standard 2 × 5/8" drywall panel construction sufficient)
G-H Corridor Uncontrolled 0.200 18' 9-3/4" 1/64" lead or equivalent
H-A Toilet Uncontrolled 0.200 3' 3-3/4" 1/32" lead or equivalent
Floor Slab N/A N/A N/A N/A
Ceiling Roof Uncontrolled 0.025 13' 0" 1/64" lead, 3/8" standard density concrete, or equivalent*
*Required concrete ceiling may already be in place. Please double check concrete ceiling thickness.
Facility Diagram (1C205):
Note: Cardinal directions are approximate only. Shielding should be installed carefully and with respect to the function of the surrounding areas to ensure proper barrier thicknesses.
Other Recommendations and Requirements
• If the X-Ray room ceiling and the ceiling of adjacent rooms does not meet the minimum requirement specified in the “Shielding Design Results” table, then all vertical barriers shall extend to overlap the ceiling shielding. Otherwise, all vertical barrier shielding shall extend to a height of at least 7 feet.
• Doors, windows, frames, conduits, and wall openings must be of the specified shielding equivalencies or greater and should have no voids.
• Joints between lead sheets should be constructed so that their surfaces are in contact and with an overlap of at least 1 cm.
• Any penetrations in a given barrier should be designed to afford the same shielding equivalency as that specified for the barrier. For example, pipes, electrical outlets, or other barrier penetrations must be wrapped, covered, or backed up with appropriate shielding-equivalent material.
• A qualified contractor with experience in shielded construction should be utilized and a follow-up inspection performed to verify the adequacy of the installation after completion.
• This shielding design is for the equipment layout and occupancy of adjacent areas as indicated on the attached drawing. If any of these parameters are changed, the shielding design will require re-evaluation by a qualified physicist.
• The shielding requirements specified in this plan review were based on the stated conservative workload(s). If the patient volume increases significantly, the shielding will require re-evaluation by a qualified physicist.
Regulatory Recommendations and Requirements
• The registrant must ensure that each radiation machine is labeled in a conspicuous manner, which cautions individuals that radiation is produced when it is energized.
• Viewing system(s) shall be provided to permit continuous observation of the patient during irradiation and shall be located so that the operator can observe the patient from the control panel.
• The operator shall be allotted no less than 7.5 ft2 of unobstructed floor space in the booth, exclusive of any encumbrance by the X-Ray control panel such as overhangs, cables, etc.
• The X-Ray control shall be permanently mounted in a protected area so that the operator must remain in that protected area during the entire exposure. This control shall be at least 40 inches from any point subject to direct scatter, leakage or primary beam radiation. If a mobile barrier is used to create the protected area, the barrier must be permanently mounted to the wall and/or floor.
Additional Shielding Design Calculation Information VA Medical Center Sacramento
All calculations were performed in accordance with industry-standard guidance as detailed in NCRP Report No. 147, “Structural Shielding Design and Evaluation for Medical X-Ray Imaging Facilities.” Wherever possible, vendor and model-specific information for the x-ray units and/or radioactive sources in question were used. Where appropriate, calculations account for attenuation by primary beam pre-shielding using the equivalent thickness values listed in Table 4.6 of NCRP Report No. 147. Patient throughput estimates as well as workload kVp distributions were used in the following calculations, based on average values contained in NCRP Report No. 147. The recommended shielding amounts are specified in a highly conservative fashion, which will allow for a broad range of patient workloads, technique factors, etc.
without creating a radiation exposure issue. However, if any major changes occur (i.e., replacement of the x-ray unit, facility/x-ray room structural layout and/or changes in the usage of the x-ray room surrounding areas), the facility administration is advised to seek the services of a qualified medical physicist to determine whether shielding additions would be needed in those cases.
All shielding design goals, occupancy factors and workloads are as specified in NCRP Report No. 147 unless stated otherwise.
Equipment Specifications and Workload (W):
Room 1C204:
Equipment Type kVp (maximum)
Patients per Week mA-min per patient Workload (W)
GE OEC Elite 125 35 64 2240 mA-min/week
Room 1C205:
Equipment Type kVp (maximum)
Patients per Week mA-min per patient Workload (W)
GE OEC Elite 125 35 64 2240 mA-min/week
Shielding Design Goals (P):
Area Type Shielding Design Goal (P) (mGy/week)
Controlled 0.10 Uncontrolled 0.02
Methodology:
For this shielding design evaluation, the fitting parameter method for shielding calculation was employed, based on manufacturer data and information provided by the user. The room will be primarily used for the purpose of urodynamics and cystograms, and all barriers have been treated as secondary barriers. This shielding evaluation assumed a conservative number of exposures per week, based on facility estimates of its typical weekly workload. Equation 4.10 of NCRP Report No. 147 was used to compute required barrier thicknesses, and fitting parameters for this equation for lead, concrete, glass, and gypsum wallboard thicknesses were found in Table C.1. Patient exam technique was based upon the typical workload distributions defined in Table 4.3 of NCRP Report No. 147 along with facility estimates.
All barrier thickness calculations were performed in Excel. These are available from West Physics and may be furnished upon request of the detailed calculations.
Summary:
The above referenced calculations and recommendations are made by West Physics based on data provided by the facility planners and unit manufacturer. As such, the recommendations made herein are subject to that information’s accuracy. This shielding design is hereby certified and executed:
Medical Physicist
10/171/2023 Date