About this file

This document provides product specifications for the STERRAD 100NX Sterilization System with ALLClear Technology. Key details include that the system uses hydrogen peroxide gas plasma sterilization in standard, flex, express, and duo cycles ranging from 24 to 60 minutes. It has a single or double door configuration, touchscreen control interface, and connectivity options. Installation, space, and environmental requirements are specified. Related federal contract opportunity 36C25621R0014 seeks proposals to upgrade, renovate, and relocate approximately 16,500 square feet of sterile processing services at the Michael E. DeBakey Veterans Affairs Medical Center, including adherence to latest standards and performance without medical center operations disruption.

View the file

Other files

File Type Posted
087100 Door Hardware Revised.pdf PDF
36C25621R0014 0005.docx DOCX document
580-17-12_Renovate SPS_Project Drawings_1-15-21.pdf PDF
36C25621R0014 0004.docx DOCX document
36C25621R0014 0003.docx DOCX document
QH102.pdf PDF
36C25621R0014 0002.pdf PDF
Area B-F-compressed.pdf PDF
S3185_Steris_Vision1327-CartWash.pdf PDF
S9610_Steris_PrepTable.pdf PDF
S0235_Steris_Amsco400.pdf PDF
Area B-G-compressed.pdf PDF
Slab Infill Detail.pdf PDF
S0940_Steris_Amsco7052HPWasher.pdf PDF
105113 MetalLockers.docx DOCX document
Clarifications 12-15-2020.pdf PDF
S9815B_Steris_Amsco600.pdf PDF
Equipment Drawing - Vision 1327 - STEAM PIT.pdf PDF
S5505B_STERRAD-100S.pdf PDF
20201211_ADDENDUM_1.pdf PDF
Site Visit Attendees.docx DOCX document
DE-1R.pdf PDF
S5505C_STERRAD-NX.pdf PDF
36C25621R0014 0001.pdf PDF
580-17-121_Renovate SPS_Combined Specifications.pdf PDF
DOL Wate Determinations TX20200253 dated 9-11-20.pdf PDF
36C25621R0014.pdf PDF
580-17-121_Renovate SPS_Specifications.pdf PDF
Site Visit Instructions _ Renovate SPS_1.docx DOCX document
580-17-121 _ Renovate SPS _ Statement of Work.docx DOCX document
SF24 BID BOND.pdf PDF
580-17-121 _ Renovate SPS _ Drawings.pdf PDF
Show all 32

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

THE STERRAD® 100NX SYSTEM

with ALLClear™ Technology

Productive

Reduces workflow interruptions

Connected

Enhances compliance, automatically*

Easy

Designed with simplicity in mind to reduce the potential for human error

*STERRAD® Systems with ALLClear™ Technology have features that may enhance compliance, including greater adherence to device IFUs and improved record handling.

JUST LOAD AND GO

Cycle temperature 47°C-56°C

Sterilant Hydrogen peroxide

Sterilant delivery Delivered in closed system in cassettes with automatic detection of expiration date:

Standard/Express/Flex: 10.8 mL per cycle (double kill) (2 ampules, each ampule is 5.4 mL) Duo: 3.1 mL per cycle

Used cassette disposal Automatic and touchless ejection into cassette disposal container

Peroxide residual breakdown

Gas plasma technology breaks down H2O2 to safe elements of water and oxygen

Configurations Single and double doors

System dimensions (maximum)

Height: 1800 mm Width: 775 mm Depth: single door: 1055 mm; double door: 1095 mm

Chamber total volume 152 L

Chamber dimensions Height: 410 mm Width: 510 mm Depth: 735 mm

Chamber shape Rectangular

Shelf info Two-tiered shelf: Width: 444 mm Depth: 643 mm

Mobility 4 casters (2 locking)

System weight 1 door: 382 kg 2 door: 408 kg

User interface Touchscreen technology: projected capacitive touch Resolution: 800 x 600 pixels

Supported USB devices

• Barcode reader • External drives: USB that allows data upload and download

• External printer connection (PCL3 compatible or equivalent)

Standards/Compliance ISO 14937

TECHNICAL SPECIFICATIONS

Cycles Standard: sterilizes most general surgical instruments Flex: sterilizes up to 2 single-channel flexible endoscopes Express: sterilizes da Vinci® 3-D endoscopes and other delicate instruments without lumens Duo: sterilizes cameras and up to 2 single-channel flexible endoscopes

Time Standard: 47 minutes Flex: 42 minutes Express: 24 minutes Duo: 60 minutes

Lumen claims Standard: diameter: ≥0.7 mm; length: ≤500 mm Flex: diameter: ≥1 mm; length: ≤850 mm Express: nonlumen sterilization only Duo: diameter: ≥1 mm; length: ≤875 mm

CYCLE SPECIFICATIONS*

Sterilization process Terminal-sterilization, double-kill cycle to provide a Sterility Assurance Level (SAL) of 10-6; 2 injections and identical plasma phases

Delivered sterilant concentration

58%-59.6%

Sterilization cycle monitoring

Critical system parameters monitored with on-board sensors, biological indicators, and chemical indicators; IMS (independent monitoring system) available

H2O2 concentration continuous monitoring

Monitoring using UV sensor within the chamber

STERILIZATION SPECIFICATIONS

FOR MORE INFORMATION

Visit www.emea.aspjj.com or contact your local ASP representative

*Always refer to the STERRAD® Sterility Guide (www.sterradsterilityguide.com) for the latest updates on instruments’ validations and cycles and systems’ compatibility.

2 x 1080 [42.5]

2 x 1000 [39.4]

[30.4]

110º 2 x 1080

[42.5]

2 x 1000 [39.4]

[30.4]

110º

[30.4]

[39.4]

[30.4]

LIMIT TO

90º

SPACE PLANNING DIMENSIONS

TOP VIEWS

ONE-DOOR CONFIGURATION

TOP VIEW

TWO-DOOR CONFIGURATION

Legend

Millimeters

[Inches]

[41.5]

[30.5] 25

[1.0]

[70.9]

[43.1]

STERILIZER

FOOTPRINT

[39.4]

1084 MAX.

[42.7]

[39.4]

[39.4]

[39.4]

865.0 +/- 6.4 [34.1 +/- 0.25]

1842 +/- 6.4 [72.5 +/- 0.25]

STERILIZER

[69.9]

STERILIZER

FOOTPRINT

[39.4]

1084 MAX.

[42.7]

[39.4]

[39.4]

[39.4]

865.0 +/- 6.4 [34.1 +/- 0.25]

1842 +/- 6.4 [72.5 +/- 0.25]

STERILIZER

[69.9]

LEFT SIDE

SYSTEM LAYOUT

TWO-DOOR CONFIGURATION

For recessed system, clearance should be ~1.8” on each side:

(865-775)/2 = 45 mm (~1.8”)

LEFT SIDE INPUT SIDE

SYSTEM LAYOUT

ONE-DOOR CONFIGURATION

Clearance on each side of the system should be ~39.4”

SYSTEM DIMENSIONS

WALL OPENING

IN-WALL INSTALLATION

INPUT OR OUTPUT SIDE

SERVICE CLEARANCE

CODE PRODUCT DESCRIPTION

10104-005 STERRAD® 100NX Sterilization System with ALLClear™ Technology, single-door unit, STANDARD & FLEX cycles, includes installation

10104-006 STERRAD® 100NX Sterilization System with ALLClear™ Technology, double-door unit, STANDARD & FLEX cycles, includes installation

10104-007 STERRAD® 100NX Sterilization System with ALLClear™ Technology, single-door unit, STANDARD, FLEX and DUO cycles, includes installation

10104-008 STERRAD® 100NX Sterilization System with ALLClear™ Technology, double-door unit, STANDARD, FLEX and DUO cycles, includes installation

10144 STERRAD® 100NX System Cassette (2 cassettes/case)

20227 Cassette Disposal Box (10 boxes/case)

10135 STERRAD® 100NX System EXPRESS Cycle Kit

10137 STERRAD® 100NX System DUO Cycle Kit

10305 Thermal Printer Paper (12 rolls/case)

113617-01 STERRAD® Systems Bar Code Scanner Kit

SYSTEM ORDERING INFORMATION

Heat generation Idle: 4574 BTU/hour Maximum: 6187 BTU/hour

Electrical power specifications

• US/Canada: 208 VAC, 60 Hz, NEMA L21-30 5-wire grounding twist-lock outlet attached to a dedicated 30-amp, 3-phase wye configuration circuit with separate neutral and ground conductors

• International: 380-415 VAC, 50/60 Hz, 5-wire grounding outlet attached to a dedicated 30-amp, 3-phase wye configuration circuit with separate neutral and ground conductors

• Japan: 200 VAC, 50/60 Hz: 4-wire grounding outlet attached to a dedicated 30-amp, 3-phase delta configuration with separate ground conductors

Installation space requirements

In operation, the STERRAD® 100NX System with ALLClear™ Technology should not be placed closer than 50.8 mm from the rear wall. The power receptacle should be positioned 30.5 to 61 cm above the floor. For recessed systems, a clearance of 25 mm from the top of the system and 45 mm on each side of the system is required.

Service space requirements

Service access requires a minimum clearance of 610 mm above the top and approximately 991 mm in on all sides of the system (can be less if the system can be moved for servicing).

System performance data and reports

Cycle history, full 1-second data files, and reports available via ASP ACCESS™ Technology

Network connectivity Communication protocol for Instrument Tracking Systems (ITS) available via ASP ACCESS™ Technology

Data recording • Electronic data storage up to 200 cycles • Internal printer for manual recordkeeping

• Full electronic cycle data and reports via ASP ACCESS™ Technology

Temperature 18°C-35°C

Humidity 10%-85% RH (noncondensing)

INSTALLATION AND ELECTRICAL REQUIREMENTS

NETWORKING AND DATA RECORDING

OPERATIONAL ENVIRONMENT

STERRAD® is a registered trademark of Advanced Sterilization Products, division of Cilag GmbH International, a Johnson & Johnson company.

STERRAD ALLClear™ and ASP ACCESS™ are trademarks of Advanced Sterilization Products, division of Cilag GmbH International, a Johnson & Johnson company.

Advanced Sterilization Products, Gubelstrasse 34, 6300 Zug www.emea.aspjj.com

© All rights reserved. 2017 AD-160028-01-CT_B